Colorimetric Test Instructions
Colorimetric Test Instructions
Colorimetric Methods
44720
Dilution Factors
2 : 1 Dilution Factor
Mix 5 ml of distilled water with 5 ml of sample water together. Run test according
to instructions. Multiply result x 2.
3 : 1 Dilution Factor
Mix 10 ml of distilled water with 5 ml of sample water together. Run test
according to instructions. Multiply result x 3.
5 : 1 Dilution Factor
Mix 40 ml of distilled water with 10 ml of sample water together. Run test
according to instructions . Multiply result x 5.
1 0 : 1 Dilution Factor
Mix 45 ml of distilled water with 5 ml of sample water together. Run test
according to instructions. Multiply result x 10.
25 : 1 Dilution Factor
Mix 48 ml of distilled water with 2 ml of samp le water together. Run test
according to instructions. Multiply result x 25.
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow 5 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to allow 5 minutes for reaction in step 3e
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of distilled water to the
vial
c. Recap and invert several times
d. Uncap vial and carefully add exactly 5 drops of Alkalinity Reagent # 1
e. Recap and invert several times to mix
f. Uncap and add 2 ml of the water to be tested to the vial
g. Recap and invert several times to mix
h. Thoroughly wipe the vial clean
i. Allow 30 seconds for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to use distilled water in step 3b
Carefully add exactly 5 drops of reagent # 1 in step 3d
Be sure to use exactly 2 ml of water to be tested in step 3f
If water contains more than 50 mg/l, then a dilution will be necessary prior to testing
Questions? Please call 330-437-0067 for technical assistance.
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert 20 times
d. Thoroughly wipe the vial clean
e. Allow 20 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Must allow 20 minutes for reaction in step 3e
Fluoride may interfere with test
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On I Enter key to power up the instrument. (Please refer to General Operating
Instructions for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Shake dropper bottle marked Ammonia Reagent # 2 and carefully add exactly 9 drops
d. Thoroughly wipe the vial clean
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to choose Ammonia Nes when selecting test on instrument
Very important to add exactly 9 drops of regent #2
If EPA approved method is required, then a distillation procedure is necessary
Interferences: Calcium, Iron, Magnesium and Sulfide may precipitate with the reagent and
should be removed with Zinc Sulfate and Alkali
For very low concentrations of ammonia, a distillation process may be used
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and shake 40 times to mix
d. Uncap vial and carefully add exactly 10 drops of Reagent A
e. Recap and shake 40 times to mix
f. Uncap vial and carefully add 2 drops of Reagent B
g. Recap and shake 40 times to mix
h. Thoroughly wipe the vial clean
i. Allow 25 minutes for reaction
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to add correct amount of drops in steps 4d and 4f
Be sure to shake 40 times to mix
Be sure to allow 25 minutes for reaction
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
Preliminary Procedure (Caution!! This test contains hazardous materials. Use of fume
hood is recommended. Refer to MSDS for proper handling.)
2. Transfer 40 ml of the water to be tested into the conical flask while stirring using a Hotplate I
Stirrer (It is recommended that a magnetic stir bar be used also)
3. Carefully add exactly 5 mL of Reagent A (Use caution: contains Concentrated
Hydrochloric Acid. Extremely corrosive!)
4. Carefully add exactly 2 mL of Reagent B
5. Carefully add exactly 8 drops of Reagent C
6. Allow solution to stir for 10-15 minutes
7. While the solution is stirring in step 3e, insert glass wool (provided) loosely into scrubber
assembly as shown in figure 1
8. Impregnate the wool with 8 drops of Reagent D (Be sure Reagent D does not drip out of
scrubber tube)
9. Carefully add exactly 5 mL of the indicator solution to the absorber tube (see figure 1)
44720
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times to mix
d. Thoroughly wipe the vial clean
e. Allow 5 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to allow 5 minutes for reaction in step 3e
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with distilled water
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial (Caution!! This test contains Concentrated Sulfuric Acid and
Hydrochloric Acid. Refer to MSDS for proper handling.) Uncap one of the reagent test vials
a. Pipette 5 ml of the Concentrated Sulfuric Acid into the vial
b. Recap vial and set aside
c. Rinse the pipette several times with distilled water before next step
d. Pipette I ml of the water to be tested into a conical flask
e. Carefully add exactly 1 drop of Concentrated Hydrochloric Acid to the flask. (Make sure
the drop hits the water or inconsistent readings may occur..) Carefully add exactly 5 ml
of Concentrated Sulfuric Acid to the flask
f. Swirl solution in flask for at least 1 minute. (Make sure the solution is thoroughly mixed.)
g. Allow the flask to cool to room temperature
h. Shake contents in the vial in step 4a 4c until thoroughly dissolved
i. Pour contents of the vial into the flask
j. Swirl solution in flask for at least 1 minute. (Make sure the solution is thoroughly mixed.)
k. Allow solution in flask to set for 30 to 45 minutes for full color development
l. Uncap clean empty vial and carefully fill with solution from flask
m. Recap and thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
44720
Page 2
Use Caution when performing procedure. Consult MSDS for proper handling.
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Be sure to follow instructions and add exact amounts called for in procedure
When finished, rinse all glassware with 1 + 1 Nitric Acid and follow with several rinses of
distilled water
Questions? Please call 800-543-9283 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow one minute for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or
smudges from handling may affect results
Rinse 5 ml syringe with distilled water when finished
Be sure to allow 1 minute for reaction in step 4e
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions) Scroll down to desired test and press enter. This will
allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial (Caution! ! This test contains Chloroform. Use of a fum e hood is
recommended. Refer to MSDS for proper handling.)
a. Select one clean empty vial (supplied with each test kit) and fill with Chloroform and
recap.
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Transfer 10 ml of the water to be tested into a 125 ml Separatory funnel
b. Carefully add exactly 0.4 ml of Reagent A to the funnel.
c. Insert the stopper and carefully shake for 10 seconds to mix
d. Carefully add exactly 2 ml of Reagent B to the funnel
e. Insert the stopper and carefully shake for 10 seconds to mix
f. Carefully add exactly 0.4 ml of Reagent C to the funnel
g. Insert the stopper and carefully shake for 10 seconds to mix
h. Carefully add exactly 6 ml of Stock Dithizone solution to the funnel
i. Insert the stopper and carefully shake for 1 minute to mix. Open stopper about every 15
seconds to relieve pressure
j. Drain bottom layer into a 25 ml glassware and keep for additional steps
k. Carefully add exactly 4 ml of Chloroform to the funnel
l. Insert the stopper and carefully shake for 10 seconds to mix. Open stopper to relieve
pressure
m. Drain bottom layer into the 25 ml glassware used in step 4j
n. Drain remaining solution from funnel and discard according to EPA Regulations
o. Rinse funnel thoroughly with distilled water
p. Pour solution from 25 ml glassware in step 4m back into funnel
q. Carefully add exactly 10 ml of Reagent D to the funnel
r. Insert the stopper and carefully shake for 10 seconds to mix. Open stopper to relieve
pressure
s. Drain bottom layer from funnel and discard according to EPA Regulations
t. Carefully add exactly 2 ml of Chloroform to the funnel
u. Insert the stopper and carefully shake for 10 seconds to mix. Open stopper to relieve
pressure
v. Drain bottom layer from funnel and discard according to EPA Regulations
w. Carefully add exactly 0. 1 ml of Reagent C to the funnel
x. Carefully add exactly 6 ml of Working Dithizone solution (1 0 % of Stock Solution) to the
funnel. Working Dithizone Solution must be prepared daily or before each test.
(See Helpful Hints for preparation)
y. Carefully add exactly 2 ml of Reagent E to the funnel
44720
z. Insert the stopper and carefully shake for 10 seconds to mix. Open stopper to relieve
pressure
aa. Drain the bottom layer into a clean empty vial with this solution and cap. This is the vial
for the test
bb. Allow I minute for reaction
cc. Thoroughly wipe the vial
dd. Discard remaining solution according to EPA Regulations
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening ofthe Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard solution from test preparation and vials according to EPA
regulations
IMPORTANT & HELPFUL STEPS
(Caution!! This test contains Chloroform. Use of a fume hood is recommended. Refer
to MSDS for proper handling.)
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
When finished, rinse all glassware with I + 1 Nitric Acid and follow with several rinses of
distilled water
Fill Blank vial with Chloroform and cap
Working Dithizone solution is 10 % of Stock Dithizone solution. Mix .6 ml of Stock Solution with
5.4 ml of Chloroform.
Working Dithizone solution must be prepared daily or before each test.
Care must be taken when draining layers
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of distilled water to the
vial
c. Recap and invert several times
d. Uncap vial and carefully add exactly 5 drops of Calcium Reagent #2
e. Recap and invert 5 times to mix
f. Uncap the vial and use the 1 ml syringe (supplied with box of reagents) to add 0.2 ml of
the water to be tested to the vial
g. Recap and invert 5 times to mix
h. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in wa ter and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to use distilled water in step 3b
Carefully add exactly 5 drops of reagent # 2 in step 3d
Carefully add exactly 0.2 ml of the water to be tested In step 3f
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert until all chemicals have dissolved. Do not shake vial
d. Uncap vial and carefully add exactly 1 drop of Carbon Dioxide reagent A
e. Recap and invert 5 times to mix
f. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affec t
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow 5 minutes for reaction
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and shake for 5 seconds
d. Uncap vial and carefully add exactly 10 drops of Chloride Reagent #2
e. Recap and invert several times to mix
f. Thoroughly wipe the vial clean
g. Allow 10 minutes for reaction
h. Un-dissolved reagent should settle on bottom and not be disturbed when handling
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, gently insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Be sure to gently place the test vial into Colorimeter in step 5c so that the un-dissolved reagent
is not disturbed
44720
1.
Press On/Enter key to power up the instrument. (Please refer to General Operating
Instructions for the Colorimeter for further instructions.)
2.
Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3.
4.
5.
6.
After completion of test, discard the reagent vial according to EPA regulations
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow one minute for reaction
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & ITELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Shelf life is no longer than 6 months
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the vial marked Blank (supplied with each test kit and to be used for all 50
tests)
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap
d. Thoroughly wipe the vial clean
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with distilled w ater
Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials (all vials are empty) and fill with the water to be
tested
b. Recap and thoroughly wipe the vial clean
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
pH of the water to be tested should be between 3 and 6
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial.
c. Uncap Test Vial (test vial #1) and add 1 level spoon of Reagent #1, cap, and invert 20
times.
d. Fill an empty vial (test vial #2) with 5mL of sample to test and add 1 level scoop of
Reagent #2 to this vial. Invert 60 times and allow 2 minutes for reaction.
e. Combine the contents of Test Vial #1 and Test Vial #2 into a 10 ml graduated cylinder.
Invert 20 times and allow to react for 5 minutes. Re-pour a portion of test solution back
into Test Vial #1 (3/4 full).
f. Thoroughly wipe the vial clean
4. Preparing the blank vial
a. Add 5 mL of sample water to be tested into test vial #2. This will become the blank.
b. Thoroughly wipe the vial clean
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe and capped vial with distilled water when finished
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times to mix
d. Thoroughly wipe the vial clean
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
44720
Page 2
Use caution when performing procedure. Consult MSDS for proper care.
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
When finished, rinse all glassware with 1 + 1 Nitric Acid and follow with several rinses of
distilled water
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 8 ml (4 ml twice) of the water
to be tested to the vial
c. Carefully add exactly 2 drops of reagent A and 2 drops of reagent B
d. Recap and invert for 1 minute to mix
e. Allow 2 minutes for reaction
f. Uncap vial and carefully add exactly 20 drops of Reagent C
g. Recap and invert for I minute to mix
h. Allow 5 minutes for reaction (If reagent is not dissolved, shake I minute and allow
another 5 minutes for reaction)
i. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (P1ease refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select one of the reagent vials and uncap
b. Use the 5 ml syringe (supplied with each test kit) and add 2 ml of the 3 ppm Standard
c. Recap and invert several times
d. Place a mark on the cap of this vial (This blank will be used for the entire box)
e. Thoroughly wipe the vial clean
f. Rinse the 5 ml syringe with distilled water before next step
3. Preparing the reagent test vial (If Chlorine is present, follow Step 4)
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 2 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Follow Step 5
4. Preparing the reagent test vial (only if Chlorine is present)
a. In a separate clean container, fill with 25 ml of the water to be tested
b. Carefully add exactly 4 drops from dropper bottle labeled Sodium Arsenite to the 25 ml
container and mix
c. Uncap one of the reagent test vials
d. Use the 5 ml syringe (supplied with Colorimeter) and add 2 ml of the 25 ml solution
mixed with Sodium Arsenite to the vial
e. Recap and invert several times
f. Thoroughly wipe the vial clean
g. Discard the remaining 23 ml Sodium Arsenite solution
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
6657 Frank Avenue NW
Suite 200
North Canton, Ohio
www.hydrodynesystems.com
44720
Page 2
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
The blank should only be used with the box of reagents it came from. Discard blank after each
box of reagents and prepare a new blank with each new box
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select one of the reagent vials and uncap
b. Use the 5 ml syringe (supplied with each test kit) and add 5 ml of the 1 ppm Standard
c. Recap and invert several times
d. Place a mark on the cap of this vial (This blank will be used for the entire box)
e. Thoroughly wipe the vial clean
f. Rinse the 5 ml syringe with distilled water before next step
3. Preparing the reagent test vial (If Chlorine is present, go to Step 4)
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Follow Step 5
4. Preparing the reagent test vial (only if Chlorine is present)
a. In a separate clean container, fill with 25 ml of the water to be tested
b. Carefully add exactly 4 drops from dropper bottle labeled Sodium Arsenite to the 25 ml
container and mix
c. Uncap one of the reagent test vials
d. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the 25 ml solution
mixed with Sodium Arsenite to the vial
e. Recap and invert several times
f. Thoroughly wipe the vial clean
g. Discard the remaining 20 ml Sodium Arsenite solution
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
44720
Page 2
44720
1.
Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2.
Scroll down to Fluoride Low test and press enter. This will allow the instrument to warm up
while preparing the reagent test vial, thus saving time.
3.
4.
Preparing the reagent test vial (If Chlorine is present, follow Step 5)
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 2 ml of the water to be tested to
the vial
3. Recap and invert several times
4. Thoroughly wipe the vial clean
5. Follow Step 6
44720
Page 2
44720
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Use caution when performing procedure. Consult MSDS for proper care.
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow ten minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of vial is thoroughly cleaned by dipping vial in water and wiping clean with
a dry paper towel. Foreign residue of any kind or smudges from handling may affect results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert 1 0 times to mix
d. Thoroughly wipe the vial clean
e. Allow 10 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may, affect
results
Be sure to analyze the test vial immediately to avoid auto-oxidation. To prevent, place 5 ml of
HCl (1 + 9) in a 50 ml volumetric flask and fill to 50 ml with water to be tested
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times to mix
d. Thoroughly wipe the vial clean
e. Allow one minute for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
7. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
8. Scroll down to desired test and press enter. This will allow the instrum ent to warm up while
preparing the reagent test vial, thus saving time.
9. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
10. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow one minute for reaction
11. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
12. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or
smudges from handling may affect results
Rinse 5 ml syringe with distilled water when finished
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On I Enter key to power up the instrument. (Please refer to General Operating
Instructions for the Colorimeter for further instructions.) Scroll down to desired test and
press enter. This will allow the instrument to warm up while preparing the reagent test vial,
thus saving time.
2. Preparing the blank vial
a. Select one clean empty vial (supplied with each test kit) and fill with chloroform.
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial (Caution!! This test contains Chloroform and
Potassium Cyanide solution. Use of a fume hood is recommended. Refer to MSDS
for proper handling.)
a. Transfer 10 ml of Chloroform and 10 ml of water to be tested into the Separatory
funnel
b. Carefully add exactly 0.2 ml of the dithizone solution to the funnel
c. Insert the stopper and carefully shake for 10 seconds to mix
d. Place funnel back in upright position and open stopper to relieve pressure
e. Carefully add exactly 2 drops of Reagent # 1 to the funnel
f. Insert the stopper and carefully shake for 30 seconds
g. Place funnel back in upright position and open stopper to relieve pressure
h. Reinsert the stopper and allow 60 seconds for reaction
i. The presence of Lead will cause the bottom half of the solution to be red or
pink.
1. If the solution in the bottom of the funnel is green remove the
stopper and add 1 or 2 drops of reagent Lead reagent# 2 to the
separatory funnel, insert stopper and shake for 30 seconds.
i.
ii. Allow layers to separate for 30 seconds and drain from the bottom layer of
the funnel into the empty vial and fill 3/4 full.
Thoroughly wipe the vial clean
44720
Page 2
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling m ay affect
results
When finished, rinse all glassware with 1 + 1 Nitric Acid and follow with several rinses of
distilled water
Interferences: Most metals which form dithizonates are eliminated by double extraction
controlled pH 8.5 9.0. Organic material must be removed by acid digestion.
Lead - Reagent Instructions Page 2
Use caution when performing procedure. Consult MSDS for proper care.
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 6 ml (3 ml twice) of the water
to be tested to the vial
c. Recap and invert 5 times
d. Uncap vial and carefully add exactly 20 drops of Magnesium Reagent A
e. Recap and invert 5 times to mix
f. Uncap vial and carefully add 5 drops of Magnesium Reagent B
g. Recap and invert 20 times
h. Thoroughly wipe the vial clean
i. Allow one minute for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert 60 times
d. Uncap vial and carefully add 1 level spoon of Manganese High Reagent # 1
e. Recap and invert 20 times to mix
f. Thoroughly wipe the vial clean
g. Allow one minute for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials (All vials are empty)
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Carefully add exactly 5 drops of reagent # I
d. Immediately add exactly 5 drops of reagent # 2
e. Recap and invert several times to mix
f. Allow (2) two minutes for reaction
g. Uncap vial and carefully add exactly 5 drops of reagent #3
h. Immediately add exactly 5 drops of reagent #4
i. Recap and invert several times to mix
j. Allow (10) ten minutes for reaction
k. Thoroughly wipe the vial clean
l. Allow one minute for reaction
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Questions? Please call 330-437-0067 for technical assistance
6657 Frank Avenue NW
Suite 200
North Canton, Ohio
www.hydrodynesystems.com
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)Scroll down to desired test and press enter. This will
allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select one clean empty vial (supplied with each test kit) and fill with chloroform.
b. Recap and thoroughly wipe the vial clean
3. Preparing the reagent test vial (Caution!! This test contains Chloroform. Use of a fume
hood is recommended. Refer to MSDS for proper handling.)
a. Transfer 10 ml of the water to be tested into the Separatory funnel
b. Carefully add exactly 6 drops of Reagent A to the funnel
c. Insert the stopper and carefully shake for 10 seconds to mix
d. Place funnel back in upright position and open stopper to relieve pressure
e. Carefully add exactly 10 ml of Chloroform to the funnel
f. Carefully add exactly 0.2 ml of Dithizone Reagent to the funnel
g. Insert the stopper and carefully shake for 20 seconds
h. Place funnel back in upright position and open stopper to relieve pressure
i. Carefully add exactly 4 drops of Reagent B to the funnel
j. Reinsert the stopper and shake for 20 seconds
k. Place the funnel back in upright position and open stopper to relieve pressure
l. Allow approximately 30 seconds for layers to separate in the funnel
i. The bottom layer should be a shade of orange.
m. Drain solution from bottom layer of the funnel into a clean empty vial
n. Recap and thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard solution in the funnel and vials according to EPA regulations
IMPORTANT & HELPFUL STEPS
Use caution when performing procedure. Consult MSDS for proper handling.
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
When finished, rinse all glassware with 1 + 1 Nitric Acid and follow with several rinses of
distilled water
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times until powder is dissolved
d. Uncap vial and carefully add exactly one drop of Molybdenum reagent # 1
e. Recap and invert 10 times to mix
f. Thoroughly wipe the vial clean
g. Allow one minute for reaction
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
g. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Powder must be dissolved in step 4c
Care must be taken to add only one drop in step 4d
Allow one minute for reaction in step 4g
If water contains more than 35 mg/i of Molybdenum, than a dilution must be performed before
Step 4
Questions? Please call 330-437-0067 for technical assistance
44720
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Take extra care when adding stated amount of drops from dropper bottles
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving tim e.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of distilled water to the
vial
c. Carefully add exactly 0.2 ml of the water to be tested to the vial
d. Recap and invert several times
e. Uncap vial and carefully add approximately 12 grains from the Cadmium capsule
(amount of grains should approximately be the diameter of an eraser floating in the vial)
f. Recap and shake for 5 7 minutes to mix
g. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
When grains are added in step 3d, color should be pink. If too many grains are added, color will
turn grayish purple
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of distilled water to the
vial
c. Carefully add exactly 0.2 ml of the water to be tested to the vial
d. Recap and invert several times
e. Uncap vial and carefully add approximately 12 grains from the Cadmium capsule
(amount of grains should approximately be the diameter of an eraser floating in the vial)
f. Recap and shake for 5 7 minutes to mix
g. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to add 5 ml of distilled water in step 3b
When grains are added in step 3d, color should be pink. If too many grains are added, color will
turn grayish purple
Questions? Please call 330-437-0067 for technical assistance
6657 Frank Avenue NW
Suite 200
North Canton, Ohio
www.hydrodynesystems.com
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Uncap vial and carefully add a few crystals of Reagent #1 (amount of grains should
approximately be the diameter of an eraser floating in the vial)
e. Recap and shake for 5 7 minutes to mix
f. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
When grains are added in step 3d, color should be pink. If too many grains are added, color will
turn grayish purple
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)Scroll down to desired test and press enter. This will
allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow 1 0 minutes for reaction
5. Performing the test
a. Insert the cleaned blank vial into the curvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into curvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert 1 0 times
d. Uncap vial and carefully add exactly 1 drop of Orthophosphate Low reagent # 1
e. Recap and invert 10 times to mix
f. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow 5 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to allow 5 minutes for reaction in step 3e
Questions? Please call 330-437-0067 for technical assistance
44720
Instructions for: pH
Method: Phenol Red (Standard Methods)
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial (If Chlorine is present, follow Step 5)
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested to
the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
5. Preparing the reagent test vial (Chlorine is present)
a. In a separate clean container, fill with 25 ml of the water to be tested
b. Carefully add exactly 4 drops from dropper bottle labeled Sodium Arsenite to the 25 ml
container and mix
c. Uncap one of the reagent test vials
d. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the 25 ml solution mixed
with Sodium Arsenite to the vial
e. Recap and invert several times
f. Thoroughly wipe the vial clean
g. Discard the remaining 20 ml Sodium Arsenite solution
6. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted,. insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
f. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
44720
Continued
Instructions for: pH
Method: Phenol Red Range: 6.4 8.3
44720
Instructions for: pH
Method: Methyl Blue (Standard Method)
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean .
3. Preparing the reagent test vial (If Chlorine is present, follow Step 4)
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Go to Step 5
4. Preparing the reagent test vial (only if Chlorine is present)
a. In a separate clean container, fill with 25 ml of the water to be tested
b. Carefully add exactly 4 drops from dropper bottle labeled Sodium Arsenite to the 25 ml
container and mix
c. Uncap one of the reagent test vials
d. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the 25 ml solution
mixed with Sodium Arsenite to the vial
e. Recap and invert several times
f. Thoroughly wipe the vial clean S
g. Discard the remaining 20 ml Sodium Arsenite solution
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
44720
Instructions for: pH
Page 2
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times being sure reagent is completely dissolved
d. Thoroughly wipe the vial clean
4. Performing the test .
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preliminary Procedure
a. Place 50 ml of the water to be tested into an appropriately size glassware
b. Carefully add exactly 1 ml of 1 + 1 Normal Sulfuric Acid
c. Add the contents of 1 capsule labeled Ammonium Persulfate
d. Using a Hotplate I Stirrer slowly bring to a boil until a volume of 10 ml is left
e. Remove solution from heat when 10 ml is reached
f. While stirring, use a pH meter to adjust the pH of the 10 ml solutio n to 7.00 by adding I
drop of 1 Normal Sodium Hydroxide. Continue to add 1 drop at a time until pH of 7.00
g. After solution cools down, bring back to 50 ml by diluting with distilled water
h. Filter solution if it is not clear
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
4. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the solution from step
3 above to the vial
c. Recap and invert several times
d. Thoroughly wipe the vial clean
e. Allow 10 minutes for reaction
5. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
6. Press enter to continue and follow instructions on screen
7. After completion of test, discard the reagent vial according to EPA regulations
44720
Page 2
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert 20 times to mix
d. Uncap vial and carefully add exactly 3 drops of Reagent A
e. Recap and invert 20 times to mix
f. Thoroughly wipe the vial clean
g. Allow one minute for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening (Invert
reagent test vial 1 time immediately before inserting into cuvette)
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Carefully add 3 drops of Reagent A in step 3d
Invert reagent test vial 1 time immediately before inserting into cuvette in step 4c
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Carefully add exactly 2 drops of Reagent # 1 (Shake before using)
d. Recap and invert 5 times
e. Uncap vial and carefully add exactly 4 drops of Reagent # 2 (Shake before using)
f. Recap and invert 20 times to mix
g. Allow 5 minutes for reaction
h. Uncap vial and carefully add exactly 2 drops of Reagent #3 (Shake before using)
i. Recap and invert 20 times to mix
j. Thoroughly wipe the vial clean
k. Allow 2 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Carefully add correct number of drops in steps 3c, 3e and 3h
Be sure to allow proper amount of time for reaction in steps 3g and 3k
Questions? Please call 330-437-0067 for technical assistance
6657 Frank Avenue NW
Suite 200
North Canton, Ohio
www.hydrodynesystems.com
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Uncap vial and carefully add 1 spoonful (5 mg spoon) from bottle labeled Reagent # 1
d. Recap and invert 5 times
e. Allow 60 seconds for reaction
f. Uncap vial and carefully add 1 spoonful (5 mg spoon) from bottle labeled Reagent # 2
g. Recap and invert 60 times to mix
h. Thoroughly wipe the vial clean
i. Allow 10 minutes for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Allow proper amount of times for reaction in steps 3e and 3i
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of Acetone (supplied) to
the vial
c. Recap and invert 20 times to mix
d. Rinse 5 ml syringe with distilled water before next step
e. Uncap vial and using 5 ml syringe, add 4 ml of the water to be tested to the vial
f. Recap and invert 20 times to mix
g. Uncap vial and carefully add exactly 3 drops of Silver reagent #2
h. Recap and invert 20 times to mix
i. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Be sure to add 5 ml in step 3b and 4 ml in step 3e
Carefully add exactly 3 drops of Silver reagent #2
Questions? Please call 330-437-0067 for technical assistance
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.)
2. Scroll down to desired test and press enter. This will allow the instrument to warm up while
preparing the reagent test vial, thus saving time.
3. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water
to be tested
b. Thoroughly wipe the vial clean
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping
clean with a dry paper towel. Foreign residue of any kind or
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Carefully add exactly I drop of Sulfide reagent # 2
d. Recap and invert 5 times
e. Allow 2 minutes for reaction
f. Uncap vial and carefully add exactly 3 drops of Sulfide reagent # 3
g. Recap and invert 5 times to mix
h. Thoroughly wipe the vial clean
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 5 ml syringe with distilled water when finished
Care must be taken to add correct number of drops in steps 3c and 3f
Questions? Please call 330-437-0067 for technical assistance
44720
44720
44720
Page 2
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of distilled water to the
vial
c. Recap and invert several times
d. Uncap vial and using the 1 ml Syringe (supplied with Colorimeter) carefully add exactly
0. 1 ml of the water to be tested to the vial
e. Recap and invert 5 times to mix
f. Thoroughly wipe the vial clean .
g. Allow 30 seconds for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the cleaned reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results
Rinse 1 ml syringe with distilled water when finished
Must use distilled water in Step 3b
Be sure to add exactly 0. 1 ml of water to be tested in Step 3d
Questions? Please call 330-437-0067 for technical assistance
44720
2. Mark two clean empty vials using a fine point permanent marker on the vial glass that is
exposed (when inserted into cuvette opening) slightly above the keypad. This will allow the user
to place the vial in the proper alignment each time in the cuvette opening of the Colorimeter.
3. Fill one marked empty vial with the proper NTU Standard and thorough ly wipe clean.
4. Choose the correct turbidity test on the Colorimeter and press enter. Please refer to General
Operating instructions for the Colorimeter for further information.)
5. Insert the filled standard vial into the Colorimeter taking special care t o align the mark on the
vial with the mark on the keypad next to the cuvette opening, place cover over vial and press
enter.
Important Note:
If performing the 0 1 range, then you must reuse the same vial the
standard is in for the water to be tested. Follow Step 6 to 8.
If performing 0 100 or 0 1000 range, then the standard vial may be kept for reuse with
other Turbidity tests. Follow Step 9 and 10.
6. Pour standard solution back into its original container for later use. Rinse the now empty vial
three or more times with distilled water.
7. Fill the cleaned vial with water to be tested, thoroughly wipe clean, insert into Colorimeter
taking special care align mark on the vial with mark on the keypad, place cover over vial and
press enter.
8. Read results on screen and record. Press enter to continue.
9. Fill second marked empty vial with the water to be tested, thoroughly wipe clean, insert into
Colorimeter taking special care to align the mark on the vial with the mark on the keypad, place
cover over the vial and press enter.
10. Read results displayed on screen and record. Press enter to continue.
IMPORTANT STEPS
Be sure the outside of the vial is thoroughly cleaned by dipping vial in water and wiping clean
with a dry paper towel. Foreign residue of any kind or smudges from handling may affect
results.
0 1 range must use the same vial for standard and water to be tested a
When finished, rinse vials with distilled water thoroughly for reu se
Questions? Please call 330-437-0067 for technical assistance
6657 Frank Avenue NW
Suite 200
North Canton, Ohio
www.hydrodynesystems.com
44720
1. Press On/Enter key to power up the instrument. (Please refer to General Operating Instructions
for the Colorimeter for further instructions.) Scroll down to desired test and press enter. This
will allow the instrument to warm up while preparing the reagent test vial, thus saving time.
2. Preparing the blank vial
a. Select the uncapped empty vial (supplied with each test kit) and fill with the water to be
tested
b. Thoroughly wipe the vial clean
3. Preparing the reagent test vial
a. Uncap one of the reagent test vials
b. Use the 5 ml syringe (supplied with Colorimeter) and add 5 ml of the water to be tested
to the vial
c. Recap and invert several times
d. Uncap vial and carefully add exactly 5 drops of Zinc Reagent A
e. Recap and invert 5 times to mix
f. Thoroughly wipe the vial clean
g. Allow one minute for reaction
4. Performing the test
a. Insert the cleaned blank vial into the cuvette opening of the Colorimeter
b. Place the black cover over the vial and press enter
c. When prompted, insert the reagent test vial into cuvette opening
d. Place the black cover over the vial and press enter
e. Read the results displayed on the screen and record
5. Press enter to continue and follow instructions on screen
6. After completion of test, discard the reagent vial according to EPA regulations
IMPORTANT & HELPFUL STEPS
Select desired test to allow the instrument to warm up
Be sure the outside of the vial is thoroughly cleaned by dipping in water and wiping clean with a
dry paper towel. Foreign residue of any kind or Smudges from handling may affect results
Rinse 5 ml syringe with distilled water when finished
Carefully add exactly 5 drops of Reagent A
Questions? Please call 330-437-0067 for technical assistance
44720