Datex-Ohmeda Avance - User Reference Manual PDF
Datex-Ohmeda Avance - User Reference Manual PDF
Datex-Ohmeda Avance - User Reference Manual PDF
User Responsibility
This Product will perform in conformity with the description thereof
contained in this Users Reference manual and accompanying labels
and/or inserts, when assembled, operated, maintained, and repaired
in accordance with the instructions provided. This Product must be
checked periodically. A defective Product should not be used. Parts
that are broken, missing, plainly worn, distorted, or contaminated
should be replaced immediately. Should repair or replacement
become necessary, Datex-Ohmeda recommends that a telephonic or
written request for service advice be made to the nearest
Datex-Ohmeda Customer Service Center. This Product or any of its
parts should not be repaired other than in accordance with written
instructions provided by Datex-Ohmeda and by Datex-Ohmeda
trained personnel. The Product must not be altered without the prior
written approval of Datex-Ohmeda. The user of this Product shall
have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or
alteration by anyone other than Datex-Ohmeda.
CAUTION
AAAX11111
The X represents an alpha
character indicating the year
the product was manufactured;
H = 2004, J = 2005, etc. I and
O are not used.
AAAXX111111AA
The XX represents a number
indicating the year the product
was manufactured; 04 = 2004,
05 = 2005, etc.
Table of Contents
1 Introduction
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Symbols used in the manual or on the equipment . . . . . . . . 1-4
Typeface conventions used . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
M1145956
Avance
3 Operation
Turning on the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Start case (start gas flow) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Using default settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Using customized settings . . . . . . . . . . . . . . . . . . . . . . . 3-4
Minimum Alveolar Concentration (MAC) . . . . . . . . . . . . 3-5
End case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Ventilator setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Using quick keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Using Vent Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Changing ventilator modes and settings . . . . . . . . . . . . 3-8
Gas setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Using quick keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Using Gas Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Changing gas and settings . . . . . . . . . . . . . . . . . . . . . . 3-9
Changing circuit type . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Spirometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Setting loop type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Spirometry setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Setting patient and sensor type . . . . . . . . . . . . . . . . . . 3-14
Selecting a data source . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Setting spirometry split screen . . . . . . . . . . . . . . . . . . . 3-15
Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Cardiac bypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Fresh gas usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Screen configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Select page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Screen setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
ii
M1145956
Table of Contents
Alarm setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Setting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Volume apnea . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
MV/TV alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
CO2 alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Auto MV limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Alternate O2 control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
EZchange canister (optional) . . . . . . . . . . . . . . . . . . . . . . . 3-27
Condenser (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
Passive AGSS (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Active AGSS (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
Connecting active AGSS with a flow indicator . . . . . . . 3-31
Connecting active adjustable AGSS . . . . . . . . . . . . . . 3-31
4 Preoperative Checkout
Every day before your first patient . . . . . . . . . . . . . . . . . . . . 4-2
Before every patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
5 Preoperative Tests
Inspect the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Vaporizer installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Flow and pressure calibration . . . . . . . . . . . . . . . . . . . . . . . 5-4
Circuit compliance compensation . . . . . . . . . . . . . . . . . . . . 5-4
Checkout menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Leak < 250 ml . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
No . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Yes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Machine check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Machine check - system . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Machine check - circuit . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Machine check -circuit O2 . . . . . . . . . . . . . . . . . . . . . . . 5-7
Machine check - monitor . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Individual checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
M1145956
iii
Avance
Circuit O2 cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Low P leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Low P leak (machines with ACGO) . . . . . . . . . . . . . . . . 5-9
Positive low pressure leak test (ACGO systems only) . . . . 5-10
Vaporizer back pressure test . . . . . . . . . . . . . . . . . . . . . . . 5-11
6 Airway Modules
Airway modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Connection to a patient . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Parameters setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Data source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
CO2 setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
O2 setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Agent setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Spirometry setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Automatic agent identification . . . . . . . . . . . . . . . . . . . . . . . 6-6
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
iv
M1145956
Table of Contents
8 Setup and Connections
Setup warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Canister setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
When to change the absorbent . . . . . . . . . . . . . . . . . . . 8-5
Removing a canister . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Removing an EZchange canister . . . . . . . . . . . . . . . . . . 8-6
Reusable Multi Absorber canister filling . . . . . . . . . . . . . 8-7
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Mains inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Outlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Serial port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Pneumatic connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Pipeline inlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Scavenging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Sample gas return port . . . . . . . . . . . . . . . . . . . . . . . . 8-12
Vacuum suction regulator (optional) . . . . . . . . . . . . . . 8-12
Venturi suction regulator (optional) . . . . . . . . . . . . . . . 8-13
Auxiliary O2 flowmeter (optional) . . . . . . . . . . . . . . . . . 8-13
How to install gas cylinders . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Pin indexed cylinder yokes . . . . . . . . . . . . . . . . . . . . . 8-14
DIN cylinder connections . . . . . . . . . . . . . . . . . . . . . . . 8-14
High-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . 8-15
How to attach equipment to the top of the machine . . . . . . 8-16
9 User Maintenance
Repair policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Maintenance summary and schedule . . . . . . . . . . . . . . . . . 9-2
Datex-Ohmeda approved service . . . . . . . . . . . . . . . . . 9-3
Circuit O2 cell replacement . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Calibration menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Flow and pressure calibration . . . . . . . . . . . . . . . . . . . . . . . 9-5
Circuit O2 cell calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
21% O2 calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Airway gas calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
M1145956
Avance
Backlight test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
How to help prevent water buildup . . . . . . . . . . . . . . . . . . . . 9-7
10 Parts
Flow sensor module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Breathing circuit module . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Complete Advanced Breathing System . . . . . . . . . . . . . . . 10-5
Absorber canister . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Exhalation valve assembly . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
EZchange canister system . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
Condenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-10
Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . 10-11
vi
M1145956
Table of Contents
Environmental requirements . . . . . . . . . . . . . . . . . . . . . . 11-12
Airway module specifications . . . . . . . . . . . . . . . . . . . . . . 11-13
Gas specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-13
Typical performance . . . . . . . . . . . . . . . . . . . . . . . . . 11-14
Suction regulators (optional) . . . . . . . . . . . . . . . . . . . . . . 11-15
Ventilator theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-16
O2 monitoring theory of operation . . . . . . . . . . . . . . . 11-17
Ventilation modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-17
Ventilator operating specifications . . . . . . . . . . . . . . . . . . 11-25
Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-25
Fresh gas compensation . . . . . . . . . . . . . . . . . . . . . . 11-25
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-25
Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-25
Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-25
Ventilator accuracy data . . . . . . . . . . . . . . . . . . . . . . . . . 11-26
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . 11-27
Guidance and manufacturers declaration electromagnetic emissions . . . . . . . . . . . . . . . . . . . . . 11-27
Guidance and manufacturers declaration electromagnetic immunity . . . . . . . . . . . . . . . . . . . . . 11-28
Power immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-28
Radiated immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-29
Recommended separation distances . . . . . . . . . . . . . 11-30
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-31
IEC 60601-1 Classification . . . . . . . . . . . . . . . . . . . . . . . . 11-32
Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-32
System components . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-33
Integral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-33
Not integral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-33
M1145956
vii
Avance
12 Super User Mode
Install/Service menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-2
Using super user mode . . . . . . . . . . . . . . . . . . . . . . . . 12-3
Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-4
Cumulative gas usage . . . . . . . . . . . . . . . . . . . . . . . . . 12-4
Volume apnea setup . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5
Setting time and date . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6
Trends setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-7
Setting colors and units . . . . . . . . . . . . . . . . . . . . . . . . 12-9
Parameter settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-10
Page Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-10
Configuring case defaults . . . . . . . . . . . . . . . . . . . . . 12-12
Index
Warranty
viii
M1145956
1 Introduction
WARNING
In this section
M1145956
1-1
Avance
Intended use
The Avance anesthesia system combines our highly advanced
anesthesia delivery, the very best in patient monitoring, and care
information management. The contemporary, compact design
addresses many ergonomic considerations including an efficient
cable management solution and storage system and an extra large
work surface area. Optional, integrated features include auxiliary O2
and suction control. The Avance system offers improved usability
through integration of ventilation, gas delivery and gas monitoring (E
and M-Series respiratory gas module capable) on a full-color 12-inch
display. This system also features electronic gas mixing of up to three
gases.
The Avance system uses SmartVent ventilation technology offering
Volume Control Ventilation with tidal volume compensation and
electronic PEEP. The proven SmartVent technology also features
optional Pressure Control Ventilation, Pressure Support Ventilation
with an Apnea Backup (PSVPro) that is used for spontaneously
breathing patients, Synchronized Intermittent Mandatory Ventilation
(SIMV) modes, Pressure Control Ventilation-Volume Guarantee
(PCV-VG), and VCV cardiac bypass. These advanced features allow
for the ventilation of a broad patient range: from neonates and
compromised patients to routine cases.
The Avance system uses the Advanced Breathing System (ABS).
This integrated breathing system is easy to remove and disassemble,
is made of highly durable polymers, and is fully autoclavable. Its fully
integrated design enhances the systems elegance while minimizing
tube connections, minimizing circuit volume, and increasing the work
surface area.
The anesthesia system is designed for expansion and upgrades, so it
is easy to add new technologies and ventilation capabilities without
investing in a new system.
This anesthesia system is not suitable for use in an MRI environment.
This system must only be operated by authorized medical personnel
well trained in the use of this product. It must be operated according
to the instructions in this Users Reference manual.
1-2
WARNING
Note
M1145956
AB.91.108
1 Introduction
M1145956
1-3
Avance
On (power)
Standby
1-4
Off (power)
O2 +
O2 Flush button
Type BF equipment
Type B equipment
ACGO active
Dangerous voltage
Earth ground
Direct current
Alternating current
Caution/Warning
Pinch hazard
Electrical input
Electrical output
Electrical input/output
M1145956
1 Introduction
Pneumatic inlet
SN
134C
Serial number
REF
Stock number
Equipotential
Variability
Variability in steps
Vacuum inlet
Maximum
Vacuum
Exhaust
Mechanical ventilation
Inspiratory flow
Expiratory flow
Lock
Unlock
Isolation transformer
Autoclavable
Not autoclavable
This way up
M1145956
Pneumatic outlet
O2%
O2 cell connection
1-5
Avance
Caution: federal law prohibits
dispensing without prescription.
Alarm silence
Submenu
No battery/battery failure
Close drain
Pipeline
Cylinder
Date of manufacture
Manufacturer
1-6
M1145956
1 Introduction
Abbreviations
Abbreviation
Definition
A
AA
Anesthetic agent
ABS
ACGO
AGSS
Alt O2
Alternate O2
APL
Adjustable pressure-limiting
APN
Apnea
C
CGO
CO2
Carbon dioxide
Compl
Compliance
E
ET
End-tidal concentration
EtCO2
EtO2
End-tidal oxygen
Exp
Expiratory
M1145956
FI
FiCO2
FI-ET
FiO2
F-V
Flow-volume loop
1-7
Avance
Abbreviation
Definition
I
I:E
Inspiratory-expiratory ratio
Insp
Inspiratory
Insp Pause
M
MAC
MV
Minute volume
MVexp
MVinsp
N
N2O
Nitrous oxide
O
O2
Oxygen
P
Pair
Paux
Auxiliary pressure
Paw
Airway pressure
PCV
PCV-VG
PEEP
PEEPe
Pexp
Expiratory pressure
P-F
Pressure-flow loop
Pinsp
Inspiratory pressure
Plimit
Pmax
Maximum pressure
Pmean
Mean pressure
PO2
Ppeak
Peak pressure
Pplat
Plateau pressure
Psupp
Support pressure
PSV
PSVPro
P-V
Pressure-volume loop
R
Rate
1-8
Respiratory rate
Raw
Airway resistance
RR
Respiratory rate
M1145956
1 Introduction
Abbreviation
Definition
S
SIMV/PSV
SIMV-PC
T
TV
Tidal volume
TVexp
TVinsp
M1145956
VCO2
VCV
Vol
Volume
1-9
Avance
1-10
M1145956
WARNING
In this section
M1145956
2-1
Avance
System overview
12
11
1
2
3
4
5
10
6
7
1.
2.
3.
4.
5.
6.
Light switch
Dovetail
Vaporizer
Alternate O2 control
System switch
Mains indicator
AB.91.111
2-2
M1145956
Description
Alternate O2 control Alternate O2 control activates automatically in the case of certain failures or errors. It
can also be activated manually by pushing the Alternate O2 control button if the display
fails. Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow. See Alternate O2 control in the Operation section.
System switch
Set the switch to the On (|) position to permit gas flow and to turn on the system.
6
7
Mains indicator
Integrated suction
(optional)
Brake
O2 flush button
Push the O2 flush button to supply high flows of O2 to the breathing system.
11
Auxiliary O2 flow
control (optional)
Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow.
M1145956
2-3
Avance
1
2
3
4
5
6
11
9
8
1.
2.
3.
4.
5.
6.
AB.91.029
10
2-4
M1145956
15
14
13
12
11
10
1
2
3
4
6
1.
2.
3.
4.
5.
6.
7.
8.
AB.91.110
M1145956
2-5
Avance
Description
3, 4
Adjustable pressure-limiting
(APL) valve
10
N
MI
30
70
20
11
2-6
Bag/Vent switch
M1145956
Optional ABS
components
7
5
6
1.
2.
3.
4.
5.
6.
7.
AB.82.043
AB.91.109
Description
Set switch to ACGO position for fresh gas to flow through the ACGO port. The
ACGO may be used to provide fresh gas to an auxiliary manual breathing circuit.
Push to drop the canister to EZchange position. This seals the breathing circuit,
permitting continued ventilation and rebreathing of exhaled gases. Be sure to
hold the canister by the handle before releasing the canister.
Push to drain water out of the condenser.
M1145956
2-7
Avance
Non-circle circuit
Fresh gas flow is diverted around the inspiratory check valve and out
through the inspiratory port when non-circle ventilation is selected.
This fresh gas source may be used with circuits without CO2
absorbent capability (for example, Mapleson variants). Mechanical
ventilation is not available when using the non-circle circuit. Tidal
volume monitoring is not available.
O2 monitoring of fresh gas is available automatically when using the
non-circle circuit if the system has the airway module option or the O2
cell monitoring option.
Fresh gas oxygen concentration is displayed on the screen. Set the
alarm limits appropriately. Note that fresh gas oxygen concentration
may not reflect FiO2 when using these types of circuits. Use an
external O2 monitor when using a rebreathing circuit with the noncircle circuit.
Systems with both an airway module and an O2 cell will display the
circuit O2 value obtained from the airway module.
Do not use an external ventilator when using the non-circle circuit. Do
not use the non-circle circuit to drive external ventilators or for jet
ventilation.
WARNING
2-8
M1145956
ACGO (optional)
Fresh gas flow is directed through the Auxiliary Common Gas Outlet
(ACGO) on the front of the machine when the ACGO switch is in the
ACGO position. Mechanical ventilation is not available when
operating an auxiliary manual breathing circuit with fresh gas from the
ACGO. The Bag/Vent switch, APL valve, and bag arm are not part of
the external circuit. Volume and pressure monitoring are not
available.
O2 monitoring of fresh gas is available automatically when the ACGO
is selected if the system has the airway module option or the O2 cell
monitoring option. A sample of the fresh gas is diverted to the O2 cell
in the breathing system. The sample flow to the O2 cell is dependent
on the pressure in the external circuit. The sample flow reduces the
fresh gas flow rate to the auxiliary breathing circuit equal to the
amount diverted to the O2 cell.
Fresh gas oxygen concentration is displayed on the screen. Set the
alarm limits appropriately. Note that fresh gas oxygen concentration
may not reflect FiO2 during spontaneous breathing or in rebreathing
circuits. Use an external O2 monitor if using a rebreathing circuit on
ACGO.
Systems with both an airway module and an O2 cell will display the
circuit O2 value obtained from the airway module.
Do not use an external ventilator on the ACGO. Do not use the ACGO
to drive external ventilators or for jet ventilation.
WARNING
Scavenging the
ACGO sample flow
M1145956
2-9
Avance
Scavenging from an
auxiliary manual
breathing circuit
Scavenging a gas
monitor sample flow
2-10
Sample gas from a gas monitor can be scavenged using the sample
gas return port or the AGSS.
To scavenge from a gas monitor using the sample gas return
port, connect the tubing from the monitor to the sample gas return
port.
To scavenge from a gas monitor using the AGSS, connect tubing
from the monitor to the male luer inlet on the bottom of the AGSS
underneath the breathing system.
M1145956
Vaporizer controls
Refer to the vaporizer Users Reference manual for more detailed
information on the vaporizer.
1.
2.
3.
4.
5.
6.
AA.43.051
AB.80.009
Tec 6 Plus
Tec 7
Lock lever
Concentration control and release
Indicators (Tec 6 Plus)
Silence alarm touch key (Tec 6 Plus)
M1145956
2-11
Avance
Description
Lock lever
Push the release and turn the concentration control to set the agent
concentration. The Tec 6 Plus concentration control does not turn as long
as the warm-up indicator is on.
n
be
2-12
All indictors come on briefly at the start. The warm-up indicator goes off
after approximately 10 minutes and the operational indicator comes on.
Other indicators come on to advise the user of required action.
Push to silence alarms. Hold for 4 seconds to sound the speaker and light
all indicators (alarm test).
M1145956
Display controls
8
1.
2.
3.
Menu keys
ComWheel
4.
5.
6.
7.
8.
Timer keys
MV/TV Alarms key
Alarm LEDs
AB.91.107
Push to silence any active, silenceable high and medium priority alarms or to
suspend/acknowledge any non-active medium or high priority alarms. Alarm is
silenced for 120 seconds or alarm is suspended for 90 seconds.
Push to show corresponding menu.
Push to select a menu item or confirm a setting. Turn clockwise or counterclockwise
to scroll menu items or change settings.
Push to remove all menus from the screen.
Push to change corresponding gas setting or ventilator setting. Turn the ComWheel
to make a change. Push the ComWheel to activate the change.
Push to start or stop the timer. Push to reset the timer back to zero.
Push to turn off the MV and TV alarms. Push again to turn the MV and TV alarms on.
Turn on solid or flash to indicate alarm priority.
M1145956
2-13
Avance
4 5
3
Plug in power
cable. On
battery.
Paw
Timer:
7
cmH2O
Measured
l/min
Agent
Ppeak
Pmean
PEEP
RR /min
Et %
Fi %
8
MAC 40
CO2
Air
EtCO2 %s
O2
EtO2 %
FiCO2 %s
PCV-VG: Ventilator On
Total Flow
l/min
TV
ml
FiO2 %
Exp Flow
RR
/min
I:E
PEEP
cmH2O
MV l/min
TVexp ml
12
11
10
AB.91.125
Off
2-14
M1145956
cmH2O
Main Menu
Ppeak
Pmean
PEEP
RR /min
Et %
Fi %
Trends
System Status
Cardiac Bypass Off
Fresh Gas Usage
Screen Setup
Parameters Setup
Calibration
Normal Screen
MAC 40
EtCO2 %s
Show trends.
EtO2 %
PCV-VG: Ventilator On
Total Flow
l/min
TV
ml
FiCO2 %s
FiO2 %
Exp Flow
RR
/min
I:E
PEEP
cmH2O
MV l/min
TVexp ml
AB.91.124
Off
1. Menu
2. Waveform fields
M1145956
2-15
Avance
Waveform fields
Digit field
The digit field can be set to show specific information such as gas
supply, flow, or agent. If the digit field is set to show agent and no
airway module is inserted, the area is blank.
Paw, O2, and either TVexp or CO2 must show on the display during a
case. If any of these parameters is not selected to show on the
display, the digit field information is replaced with the missing
parameter.
See Screen configuration in the Operation section for more
information.
2-16
M1145956
Using menus
Push a menu key to display the corresponding menu. Use the
ComWheel to navigate the menu.
1
Main Menu
3
Trends
On
System Status
Cardiac Bypass
On
Off
Screen Setup
Parameters Setup
Calibration
Normal Screen
1.
2.
3.
4.
5.
6.
5
AB.91.126
Menu title
Current selection
Adjustment window
Submenu
Instructions or help information
Menu items
M1145956
2-17
Avance
2-18
M1145956
3 Operation
WARNING
In this section
M1145956
3-1
Avance
AB.91.044
1. System switch
2. Mains indicator
CAUTION
Note
3-2
M1145956
3 Operation
Start Case
Default Settings
Patient Weight
Set Vent By Weight
Patient Age
Other Gas
CO2 Alarms
Volume Apnea
Start Case Now
The Default Settings has five selections. The first four selections are
case types. The fifth selection is Last Case. The case type in the first
slot shows as the default when the Start Case menu is opened.
Case types are configured in the Install/Service - Case Defaults
menu. The Install/Service menu is only accessible by a Super User
or a service person. See the Super User Mode section for more
information on the Start Case menu defaults.
The values of Patient Weight, Patient Age, Other Gas, and Volume
Apnea are configured to the default case type by the Super User.
CO2 Alarms setting is set to On when the Start Case menu is
opened.
Note
Volume Apnea is not shown on the Start Case menu when the
Volume Apnea Selection is set to Disable in the Install/Service
menu.
The Set Vent By Weight menu item is accessible when the
ventilation mode is set to VCV or PCV-VG. Use this setting for breath
rate and tidal volume calculations based on the set patient weight.
The Default Settings selection shows Weight when Set Vent By
Weight is selected.
M1145956
3-3
Avance
Using default
settings
Using customized
settings
3-4
M1145956
3 Operation
7. To change the gas settings:
Push the Gas Setup key.
Select Adjust Settings.
Change the settings.
Select Exit to return to the Start Case menu.
8. To change the balance gas:
Push the Gas Setup key.
Select the balance gas to use with O2.
Select Exit or adjust the settings and select Exit to return to
the Start Case menu.
9. To change the circuit type:
Push the Gas Setup key.
Select the circuit type.
Select Confirm or adjust the settings and select Confirm to
return to the Start Case menu.
10. From the Start Case menu, select Start Case Now. Gas flow
starts.
Minimum Alveolar
Concentration (MAC)
M1145956
3-5
Avance
End case
Use the End Case menu to stop gas flow and end the patient alarms.
End Case
End Case Now
Normal Screen
3-6
M1145956
3 Operation
Ventilator setup
Access the Vent Setup menu by pushing the Vent Setup key. The
system has up to six modes of mechanical ventilation:
Volume Control Ventilation (VCV).
Pressure Control Ventilation (PCV) (optional).
Synchronized Intermittent Mandatory Ventilation/Pressure
Support (SIMV/PSV) (optional).
Pressure Support Ventilation (PSVPro) (optional).
Synchronized Intermittent Mandatory Ventilation-Pressure
Control (SIMV-PC) (optional).
Pressure Control Ventilation-Volume Guarantee (PCV-VG)
(optional).
Vent Setup
Adjust Settings
Mode:
VCV
PCV
Spirometry
Normal Screen
WARNING
Important
M1145956
3-7
Avance
Changing ventilator
modes and settings
3-8
M1145956
3 Operation
Gas setup
Access the Gas Setup menu by pushing the Gas Setup key. Use
the Gas Setup menu to adjust the O2% and total flow, to change the
balance gas, and to change the circuit type.
Gas Setup
Adjust Settings
Gas:
N2O
Air
Circuit:
Circle
Non-Circle
Normal Screen
The O2% and total flow can easily be changed using the gas quick
keys.
1. Push a gas quick key to select the corresponding gas setting.
2. Turn the ComWheel to make a change.
3. Push the ComWheel to activate (confirm) the change.
M1145956
3-9
Avance
3-10
M1145956
3 Operation
Spirometry
There are three types of spirometry loops: Pressure-Volume (P-V),
Flow-Volume (F-V) and Pressure-Flow (P-F). View, save, and erase
spirometry loops and set the loop scaling in the Spirometry menu.
Access the Spirometry menu by pushing the Vent Setup key and
selecting Spirometry.
Use the Spirometry menu to:
View a specific loop type; select Loop Type and set the loop
type.
Adjust the loop scaling; select Scaling and set the scale type.
Access the Spirometry Setup menu; select Spiro Setup.
Store a loop to memory; select Save Loop. Up to six loops can
be saved.
View a saved loop; select Reference Loop and the time at which
the loop was saved.
Erase a saved loop; select Erase Loop and the time at which the
loop was saved.
Spirometry
Loop Type
Scaling
Spiro Setup
Save Loop
Reference Loop
Erase Loop
Previous Menu
M1145956
P-V
None
None
3-11
Avance
The spirometry loops show in the spirometry window and can be set
to show alongside the waveforms as the split screen.
AB.98.039
1.
2.
3.
4.
Volume axis
Pressure axis
Real-time loop
Reference loop (appears on display in white)
3-12
The loop type can also be set in the Spirometry Setup menu. Push
the Main Menu key, and select Parameters Setup - Spirometry
Setup - Loop Type.
M1145956
3 Operation
Scaling
The scales shown in the spirometry loop graph are set in the Scaling
menu. The available settings for the volume, Paw, and flow graph
axes are dependent on the set patient type of adult or pediatric.
The AUTO selection automatically adjusts the volume, Paw, and flow
axes of the loop graph based on the minimum and maximum breath
reading shown in the waveform.
The Linked selection links the adjustment of the volume, Paw, and
flow axes of the loop graph together. Change one of the scales and
the remaining two scales automatically change based on the one set
scale.
The Indep selection allows the axes of the loop graph to be changed
separately for the volume, Paw, and flow axes.
Scaling
Scaling AUTO
Vol Scale
Paw Scale
Flow Scale
Previous Menu
M1145956
3-13
Avance
Spirometry setup
Push the Vent Setup key and select Spirometry - Spiro Setup to
access the Spirometry Setup menu. Use the Spirometry Setup
menu to:
Set the patient and sensor type.
Select a data source.
Set the loop type.
Change the volume shown on the spirometry split screen to
MVexp or TVexp.
Set split-screen view.
Access the Paw and MVexp alarm limits.
Note
Spirometry Setup
Patient and Sensor Type
Data Source
Loop Type
TV or MV
Split Screen
Paw Alarm
MVexp Alarm
Previous Menu
Adult
Vent
P-V
TV
None
Patient and sensor type refer to the style of airway adapter used with
the airway module. If spirometry data is obtained from the airway
module, ensure that the sensor type matches the airway adapter
used.
If the sensor type is not set correctly, the information displayed may
not be accurate.
1. Push the Vent Setup key.
2. Select Spirometry.
3. Select Spiro Setup - Patient and Sensor Type.
4. Select Adult or Pedi depending on the sensor used.
Adult refers to the D-lite sensor.
Pedi refers to the Pedi-lite sensor.
3-14
M1145956
3 Operation
Selecting a data
source
Setting spirometry
split screen
M1145956
3-15
Avance
Main Menu
Menus and settings available through the Main Menu include
Trends, System Status, Cardiac Bypass, Fresh Gas Usage,
Screen Setup, Parameters Setup, and Calibration.
System Status shows the status of gas supplies, electrical supplies,
and software settings.
See Screen configuration for information on the Screen Setup
menu item.
See Parameters setup in the Airway Modules section for more
information on the Parameters Setup menu item.
See the User Maintenance section for more information on
calibrations.
Main Menu
Trends
System Status
Cardiac Bypass Off
Fresh Gas Usage
Screen Setup
Parameters Setup
Calibration
Normal Screen
3-16
M1145956
3 Operation
Trends
M1145956
3-17
Avance
Cardiac bypass
Manual ventilation
cardiac bypass
WARNING
WARNING
3-18
M1145956
3 Operation
Use the Fresh Gas Usage selection to view the volume of O2, Air,
N2O, and agents used for the three most recent cases. Data only
shows for gases available on the system. Agent data shows the three
most recently used agents.
1. Push the Main Menu key.
2. Select Fresh Gas Usage.
3. Select the patient case to view.
4. Push the Normal Screen key or select Previous Menu.
M1145956
3-19
Avance
Screen configuration
Each case type selected on the Start Case menu has a
corresponding screen configuration for the normal screen. Four
additional normal screen views are available through the Select Page
menu. The case type screen configurations and the normal screen
views are set by the Super User in the Install/Service menu.
Areas of the normal screen can be customized to show specific
information. Use the Screen Setup menu to customize the normal
screen view.
Select page
Use the Select Page menu to quickly change or return the normal
screen view to a case type view.
Select Page
Page:
Default Screen
PAW TUBES
LOOPS
BIG WAVE
LOCAL
Normal Screen
3-20
M1145956
3 Operation
Screen setup
Waveform, digit field, fresh gas control style, split screen, sweep
speed, and brightness are adjusted in the Screen Setup menu.
Changes made to all settings (except Brightness) during a case are
not saved when the case is ended. The set Brightness remains until
it is changed again or until the system power is turned off.
Screen Setup
Waveform Field 1
Waveform Field 2
Waveform Field 3
Digit Field
Fresh Gas Controls
Split Screen
Sweep Speed
Brightness
Install/Service
Previous Menu
Waveform fields
Paw
Flow
CO2
AA
O2%
None
Fast
4
The waveforms can be set to show agent, CO2, flow, Paw, or can be
set to off. Waveforms cannot be set to the same value (except for off).
If waveform is set to the same value as another waveform, the
previously set waveform changes to off and is removed from the
normal screen
1. Push the Main Menu key.
2. Select Screen Setup.
3. Select the waveform field and make the change.
4. Push the Normal Screen key or select Previous Menu.
Digit field
The digit field can be set to show gas supply, flow, or agent.
1. Push the Main Menu key.
2. Select Screen Setup - Digit Field.
3. Select the Supply, Flow, or AA.
If AA is selected, an airway module must be inserted or the
digit field will be blank.
4. Push the Normal Screen key or select Previous Menu.
M1145956
3-21
Avance
Setting gas controls and
screen configuration
There are two ways to configure gas controls: O2% with total flow or
individual gas flow. The Super User can set the gas control for the
facility. Or the Super User can set the gas control to allow the user to
select the setting.
Selecting O2% shows O2% as the first quick key and total flow l/min
as the second quick key. This gas control configuration adjusts the
balance automatically when either the O2% or the total flow is
changed.
Selecting Flow shows balance gas l/min as the one quick key and O2
l/min as the other quick key. This gas control/screen configuration
allows individual control of the gasses.
1. Push the Main Menu key.
2. Select Screen Setup - Fresh Gas Controls.
3. Select the desired view.
4. Push the Normal Screen key or select Previous Menu.
Split screen
Sweep speed
The waveform draw rate can be set to fast (6.25 mm/s) or slow
(0.625 mm/s). When the sweep speed is changed, the waveforms are
redrawn at the new rate.
1. Push the Main Menu key.
2. Select Screen Setup - Sweep Speed.
3. Set the speed.
4. Push the Normal Screen key or select Previous Menu.
3-22
M1145956
3 Operation
Alarm setup
Alarm limits, alarm volume, and other alarm settings are adjusted in
the Alarm Setup menu. Alarm history is also accessed through this
menu. Selecting Default Limits loads the default settings as set by
the Super User or the factory defaults if no Super User settings have
been entered.
Setting Leak Audio to Off silences audio alarms for small leaks. Leak
Audio is automatically set to On and cannot be changed when either
the Low MV alarm limits are off or the MV/TV Alarms is set to Off.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
The apnea time delay range is 10 to 30 seconds. Set the Apnea
Time in 1 second increments.
Alarm Setup
Adjust Settings
Volume Apnea
MV/TV Alarms
Leak Audio
CO2 Alarms
Auto MV Limit
Default Limits
Alarm Volume
Apnea Time
Alarm History
Previous Menu
Off
On
On
On
Off
3
30
M1145956
3-23
Avance
Volume apnea
Volume Apnea is not shown on the Alarm Setup menu when the
Volume Apnea Selection is set to Disable in the Install/Service
menu.
Set Volume Apnea to Off to disable the volume apnea alarm during
manual ventilation. Volume Apnea Off shows in the general
message field. The volume apnea alarm remains disabled until the
Bag/Vent switch is set to ventilator or Volume Apnea is set to On.
1. Push the Alarm Setup key.
2. Select Volume Apnea and set to On or Off.
3. Push the Normal Screen key or select Previous Menu.
Note
MV/TV alarms
When Volume Apnea is set to Off during manual ventilation and then
mechanical ventilation is started, the volume apnea alarms are active.
Volume Apnea - Off requires confirmation through a pop-up window
when manual ventilation is resumed.
The MV and TV alarms can be turned off. When the volume alarms
are disabled, MV/TV Alarms Off appears in the general message
field. The volume alarm limits waveform numerics show as dashes
during a case.
If the MV and TV alarms are set to Off during manual ventilation, the
alarms remain off until the Bag/Vent switch is set to ventilator or the
MV/TV Alarms is set to On.
The MV and TV alarms retain the last value set during mechanical
ventilation. If MV/TV Alarms is set to On during mechanical
ventilation, it remains on when manual ventilation starts. If MV/TV
Alarms is set to Off during mechanical ventilation it remains off when
manual ventilation starts.
1. Push the Alarm Setup key.
2. Select MV/TV Alarms and set to On or Off.
3. Push the Normal Screen key or select Previous Menu.
CO2 alarms
3-24
M1145956
3 Operation
Auto MV limit
M1145956
3-25
Avance
Alternate O2 control
WARNING
WARNING
1
2
AB.91.077
1. ON/OFF button
2. Flow tube
3. Flow control
3-26
M1145956
3 Operation
AB.82.042
To return to absorber mode, reinsert the canister into the holder and
push the canister back up and snap it into absorber position. When
the canister is in the absorber position, the exhaled gas flows through
the absorber, removing CO2.
Note
Check the absorber canister to ensure it has side rails. If the canister
does not have side rails, it will not work in the EZchange canister
holder.
AB.74p.043
1. Side rails
M1145956
3-27
Avance
Condenser (optional)
The condenser removes water in the system that is produced from
the reaction of CO2 gas with the absorbent. When fresh gas flow
settings of less than the patient minute volume are used during
ventilation, the amount of re-breathed gas increases. More CO2 flows
through the breathing system increasing the absorbent consumption.
The moisture buildup in the breathing system is also increased.
The condenser is connected between the outlet of the absorber
canister and the inlet of the circuit module. Moisture in the gas is
condensed into water droplets, which run into the condensers
reservoir.
Visually check the condenser reservoir daily. Drain the reservoir daily.
1. Place a container under the reservoir.
2. Push the drain button to empty any water in the condenser.
AB.75p080
1. Drain button
2. Reservoir
3. Condenser
3-28
M1145956
3 Operation
M1145956
3-29
Avance
3-30
M1145956
3 Operation
Connecting active
AGSS with a flow
indicator
AB.91p093
2. With the AGSS operating, verify that the flow indicator ball on the
flow indicator rises to the green zone, indicating adequate flow.
Note
The ball in the upper red zone indicates excessively high extraction
flow. The ball in the lower red zone indicates extraction flow rate is
too low or a blocked filter.
3. Complete the tests in the Preoperative Tests section of this
manual.
Connecting active
adjustable AGSS
The active adjustable AGSS option flow rate is limited to 30 l/min with
this option.
To use the optional active AGSS installed on the system which uses
the three-liter bag as a visual indicator, connect it as follows:
1. Connect a disposal hose to the DISS connector on the needle
valve on the bottom of the AGSS (underneath). The hose should
be flexible and reinforced to help prevent kinking and crushing.
2. Attach the other end of the hose to the hospital disposal system.
3. Attach the three-liter bag to the 30 mm auxiliary 1 port on the
bottom of the AGSS.
M1145956
3-31
Avance
AB.75p.110
4. Use the needle valve to adjust the flow rate to match the amount
of gas being scavenged. Use the visual indicator bag when
adjusting the flow rate. The bag should remain partially inflated
when the flow rate is adequate.
3-32
M1145956
4 Preoperative Checkout
WARNING
In this section
M1145956
4-1
Avance
4-2
Check that the equipment is not damaged and that components are
correctly attached.
Check that pipeline gas supplies are connected and cylinders are
installed.
Set the appropriate controls and alarm limits for the case.
M1145956
4 Preoperative Checkout
M1145956
Note
This check does not need to be done before the first case of the day if
the Every day before your first patient checklist was done.
Set the appropriate controls and alarm limits for the case.
4-3
Avance
4-4
M1145956
5 Preoperative Tests
In this section
M1145956
5-1
Avance
5-2
M1145956
5 Preoperative Tests
Vaporizer installation
WARNING
M1145956
5-3
Avance
WARNING
5-4
M1145956
5 Preoperative Tests
Checkout menu
The Checkout menu shows on the display after turning on the
system. To access the Checkout menu between cases, push the
Checkout key. Step-by-step instructions show in the right window
next to the Checkout menu during the checks. Use the Checkout
menu to:
Perform a Machine Check.
Perform any of the individual checks.
Set the Leak < 250 ml setting.
View the Check Log.
Start a case.
Checkout
Machine Check
Leak < 250 ml No
Individual Checks
System
Circuit
Circuit O2 Cell
Low P Leak
Check Log
Start Case
M1145956
No
When No is selected, the leak test will pass for leaks below 250 ml at
3 kPa (30 cmH2O) pressure with no user interaction required. For
leaks between 250 ml and 750 ml, the user can fix the leak and rerun
the test or accept the leak and continue. For leaks above 750 ml, the
test will fail and the user must fix the leak and rerun the test.
Yes
Set to Yes to measure small leaks above 100 ml during the checkout
procedures. Selecting Yes will display the measured leak at 3 kPa
(30 cmH2O) pressure and result in the test taking somewhat longer.
5-5
Avance
Machine check
The Machine Check or the individual checks must be performed at
least once within every 24-hour period.
Perform the Machine Check at the start of each day. The machine
check runs automatically and beeps to indicate when it is finished or if
interaction is required.
The Machine Check does a Machine Check-System check,
Machine Check-Circuit check, and a Machine Check-Circuit O2
cell check (if circuit O2 cell is present). When one of the checks is
completed, the next check begins.
1. Turn the System switch to On.
2. Select Machine Check and follow the instructions.
3. If a check fails, follow the instructions to perform a recheck or
accept the results.
4. When the Machine Check is completed, start a case.
Note
5-6
M1145956
5 Preoperative Tests
M1145956
When External Gas Monitor is set to Yes by the Super User, the
Machine Check-Monitor check occurs. This check is not a test. This
check is a reminder to connect a respiratory gas monitor.
5-7
Avance
Individual checks
The Machine Check or the individual checks must be performed at
least once within every 24-hour period.
Individual checks allow the user to perform any combination of single
checks. These checks are helpful if there is a specific problem/alarm
and the user wishes to test only that portion of the system.
The checks do not automatically move on to the next check. After
completing a check, do another check or start a case. If a check fails,
follow the instructions to perform a recheck or accept the results.
System
The System check checks the Bag/Vent switch, proper gas supply
pressures, ventilator operation and leak, battery and electrical power,
circuit compliance, and flow control operation. This is a two-step
check.
1. Set the Bag/Vent switch to Vent.
2. Open the patient Y.
3. (ACGO option only.) Set the ACGO switch to Circle.
4. Select Start. The display shows the checks being run.
The system beeps when this portion of the check is done.
The results are shown on the display.
5. Make sure the bellows is fully collapsed.
6. Occlude the patient Y.
7. Select Continue. The display shows the checks being run.
8. When the check passes, select Back.
9. Select another check or select Start Case to go to the Start Case
menu.
Circuit
The Circuit check checks the Bag/Vent switch, proper gas supply
pressures, airway pressure measurement transducer, APL valve, and
manual circuit leak.
1. Occlude the patient Y.
2. Set Bag/Vent switch to Bag.
3. Set the APL valve halfway between 30 and 70.
4. (ACGO option only.) Set the ACGO switch to Circle.
5. Select Start. The display shows the checks being run.
6. When the check passes, select Back.
7. Select another check or select Start Case to go to the Start Case
menu.
5-8
M1145956
5 Preoperative Tests
Circuit O2 cell
Low P leak
Low P leak
(machines with
ACGO)
The negative Low P Leak check measures machine leaks before the
breathing system, between the common gas outlet and the high
pressure pneumatics and includes the gas mixer and vaporizer. It
measures low pressure pneumatic leaks with a pass/fail limit of 50 ml.
1. Make sure that the vaporizers to be used during the case are
mounted on the machine and turned off.
2. Make sure the ACGO switch is set to ACGO.
3. Attach the squeeze bulb to the ACGO outlet.
4. Squeeze (collapse) the bulb.
5. If the bulb inflates in less than 30 seconds, select Fail.
6. If the bulb remains collapsed, repeat the check once for each
vaporizer with one vaporizer turned on at a time.
7. When the checks pass, turn the vaporizers off and remove the
squeeze bulb from the ACGO outlet.
M1145956
5-9
Avance
CAUTION
CAUTION
If the needle valve is not fully open, this test can damage
the pressure gauge on the test device.
3. Turn the ACGO switch to the ACGO position.
4. Set the O2 flow to 500 ml.
5. Make sure that the total flow through the flowmeter on the test
device is 0.5 l/min.
6. Make sure that the pressure gauge on the test device reads zero.
7. Close the needle valve on the test device until the test gauge
reads 20 kPa (3 psi) (BSI) or 3 kPa (0.4 psi) (ISO).
8. If the flow through the test device is less than 0.45 l/min (ISO)
or 0.4 l/min (BSI), there is a low pressure leak in the anesthesia
machine. See the Alarms and Troubleshooting section for more
information.
9. Repeat this low-pressure leak test for each vaporizer.
Set the applicable vaporizer to 1%.
Turn the vaporizer off after the test.
WARNING
5-10
M1145956
5 Preoperative Tests
M1145956
5-11
Avance
5-12
M1145956
6 Airway Modules
In this section
M1145956
6-1
Avance
Airway modules
The optional compact airway modules measure and monitor gases
delivered to the patient and exhaled through the breathing circuit. The
modules consist of an infrared sensor for measuring CO2, N2O, and
anesthetic agents; a paramagnetic O2 sensor; and a gas sampling
system with the D-fend water separation system.
Systems with both an airway module and an O2 cell will display the
patient inspired O2 value obtained from the airway module.
Respiratory rate is the frequency of peak (end tidal) CO2
measurements per minute. A breath is defined as a change in the
CO2 signal that exceeds 1% (8 mmHg). All concentrations are
measured and displayed breath by breath.
WARNING
6-2
CAUTION
M1145956
6 Airway Modules
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. The following modules can be used on
this system: E-CAiO, E-CAiOV, E-CAiOVX, M-CAiO, M-CAiOV, and
M-CAiOVX. (E series modules must be software version 4.5 and
above. M series modules must be software version 3.2 and above.)
Letters in the name of the airway modules stand for:
E = plug-in gas module
M = plug-in gas module
C = CO2 and N2O
A = anesthetic agents
i = agent identification
O = patient O2
V = patient spirometry
X = Gas exchange
2
3
4
6
MD.60.008
1.
2.
3.
4.
5.
6.
M1145956
6-3
Avance
Connection to a
patient
WARNING
AB.91.010
1. Airway module
2. Gas sampling line
3. Airway adapter with sampling line connector
6-4
M1145956
6 Airway Modules
Parameters setup
Use the Parameters Setup menu to change the monitoring settings
of the data source, CO2, O2, agent, and spirometry. Push the Main
Menu key. Select Parameters Setup.
Data source
CO2 setup
O2 setup
Change the size of the CO2 waveform by changing the scale height.
The scale units (%, kPa, and mmHg) are set by the Super User.
Select CO2 Alarm or Resp Rate Alarm to access and change the
alarm limits for the corresponding alarm.
Select O2 Alarm to access and change the O2 alarm limits.
Agent setup
Change the size of the agent waveform by changing the scale height.
Select Agent Alarm to access and change the agent alarm limits.
Spirometry setup
Change the size of the Paw and Flow waveforms by changing the
corresponding scale heights. Select Paw Alarm or MVexp Alarm to
access and change the alarm limits for the corresponding alarm.
M1145956
6-5
Avance
Calibration
Calibrate airway modules once every six months or whenever there
are indications of errors in the gas readings. Use a Datex-Ohmeda
calibration gas and regulator to calibrate the modules. See the
Parts section for the stock numbers of the calibration gas and
regulator.
WARNING
6-6
M1145956
6 Airway Modules
9. If adjustments are needed:
Do not close the regulator until all the adjustments have been
made.
Select the gas to be adjusted and press the ComWheel.
Use the ComWheel to change the value until it matches the
calibration gas cylinder value. Push the ComWheel to confirm
the change.
Repeat for each gas requiring adjustment.
M1145956
6-7
Avance
6-8
M1145956
CAUTION
WARNING
In this section
M1145956
7-1
Avance
Alarms
Alarms are divided into technical alarms and parameter alarms.
These alarms may be high priority, medium priority, or informational.
When an alarm occurs during a case, an alarm tone sounds and the
alarm message is displayed in the alarm message field.
Technical alarms result from a technical problem and occur whether
or not a patient is connected to the system. Parameter alarms are
calculated limits and limits that are set by the user on the Alarm
Setup menu. Parameter alarms occur only during a case.
Alarm priorities
Alarm priority is indicated by the color of the alarm message and the
alarm LED located next to the alarm silence button.
High-priority alarm messages appear in white text on a red
background. During a high-priority alarm, the red LED flashes.
Medium-priority alarm messages appear in yellow text on a gray
background. During a medium-priority alarm, the yellow LED flashes.
Informational alarms appear in white text on a gray background.
During an informational alarm, the yellow LED is on solid.
When a high-priority alarm is active at the same time as a mediumpriority alarm or an informational alarm, the red and yellow LEDs
flash. When a high-priority alarm is active at the same time as an
informational alarm, the red LED flashes and the yellow LED is on
solid. When a medium-priority alarm is active at the same time as an
informational alarm, the yellow LED flashes. Pushing the Silence
Alarms key changes the LED from flashing to on solid until the end
of the alarm silence.
Silencing alarms
Silencing an alarm stops the audible tone for 120 seconds. The alarm
message shows in the alarm message field. Pushing the Silence
Alarms key when no medium or high priority alarms are active
suspends audible alarm tones for 90 seconds.
Alarms in the apnea alarm family have special silence behavior to
reduce apnea nuisance alarms. Apnea family alarms include Apnea,
TVexp low, EtCO2 low, MVexp low, and RR low.
When silencing an apnea family alarm, the audio tone for the active
alarm is silenced for 120 seconds. The audible tone for any additional
apnea family alarm that occurs during the silence period is silenced
for the remain time shown on the alarm silence countdown. Only the
audible alarm tone is silenced. The alarm messages still show in the
alarm message fields. APN shows above the alarm silence
countdown when the audible tone silence is in affect for the apnea
family alarms.
Display changes
during alarms
7-2
M1145956
De-escalating alarms
Battery indicator
Internal failure
M1145956
7-3
Avance
List of alarms
If the corrective action does not resolve the alarm message, contact a
Datex-Ohmeda trained service representative
Circuit pressures and volumetric flows are measured by the ventilator
and airway gas module. If the Data Source is set to Patient, the
displayed waveforms and numeric information are measured by the
airway gas module. Although not displayed, the ventilator
measurements continue and if a measured value violates an alarm
setting, the appropriate alarm occurs. The value highlighted in the
parameter numeric box may not appear to have violated the alarm
setting. Changing the Data Source to Vent will display the circuit
pressures and volumetric flows measured by the ventilator.
Message
Priority
Cause
Action
AA, CO2
monitoring not
connected
Medium
Air supply
pressure low
Medium
Apnea
Medium
Apnea >120 s
High
Backup Mode
active
Informational
Breathing
system loose
Informational
7-4
M1145956
Message
Priority
Cause
Calibrate
(remove) flow
sensors
Informational
Calibrate O2
sensor
Informational
Cannot monitor
Air pipeline
Medium
Cannot monitor
gas supplies
Informational
Hardware failure.
Cannot monitor
O2 pipeline
Medium
Transducer failure.
Check circuit
connections
Medium
Check D-Fend
Medium
Check flow
sensors
Medium
Check sample
gas out
Medium
Circuit leak
Medium
Circuit leak
silenced
Informational
Circuitry >75C
shutdown
possible
Medium
Cooling fan
needs service.
System OK.
Medium
Cooling fans
failed. May
overheat.
Medium
M1145956
Action
7-5
Avance
Message
Priority
Cause
Action
Display panel
controls failure
Medium
EtCO2 high
High
EtCO2 is greater than high alarm Check the patient and EtCO2 settings. Check
limit.
if absorbent needs to be changed.
EtCO2 low
Informational
Medium after 1
minute
EtDES high
Medium1
EtDES low
Informational
EtENF high
Medium1
EtENF low
Informational
EtHAL high
Medium1
EtHAL is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
EtHAL low
Informational
EtISO high
Medium1
EtISO is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
EtISO low
Informational
EtO2 high
Medium
EtO2 low
Medium
EtO2 is less than low alarm limit. Set the alarm limits appropriately. Increase
the O2 concentration.
EtSEV high
Medium
EtSEV low
Informational
FiCO2 high.
Absorbent OK?
High
FiDES high
Medium1
FiDES is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
7-6
M1145956
Message
Priority
Cause
Action
FiDES low
Informational
FiENF high
Medium1
FiENF is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiENF low
Informational
FiHAL high
Medium1
FiHAL is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiHAL low
Informational
FiISO high
Medium1
FiISO is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiISO low
Informational
FiO2 high
Medium
FiO2 low
High
FiO2 is less than low alarm limit. Check O2 setting. Check for leaks or
blockages in the patient circuit.
FiSEV high
Medium
FiSEV is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiSEV low
Informational
Informational
Gas monitoring
not available
Medium
Airway module hardware failure. Replace airway module. Then, turn power off
and back on between cases to clear the
alarm and receive module data.
Inspiration
stopped
Medium
Internal failure.
System may
shut down.
High
Internal failure.
System may
shut down.
Medium
M1145956
7-7
Avance
Message
Priority
Cause
Action
Memory
(EEPROM)
failure
Informational
Software error.
Module fail. No
CO2, AA, O2
data.
Medium
Module not
compatible
Informational
Move Bag/Vent
Switch to Bag
Medium
MVexp high
Medium
MVexp low
Medium
N2O supply
pressure low
Medium
No battery
backup
Medium
No exp flow
sensor
Medium
No fresh gas
flow!
High
No fresh gas
flow?
High
Possible fresh gas flow occlusion Switch to circle circuit or bag the patient.
or loss of gas pressure.
Check pipeline supply connection.
No insp flow
sensor
Medium
O2 flush stuck
on?
Informational
Switch is detected on
continuously for more than 30
seconds.
O2 monitoring
not connected
Medium
O2 supply
pressure low
High
O2 pipeline pressure is less than Ensure the O2 pipeline and cylinder are
252 kPa (36 psi) and the O2
properly connected.
cylinder pressure dropped below
2633 kPa (381 psi) for one
second.
7-8
M1145956
Message
Priority
Cause
Action
PEEP high.
Blockage?
High
Plug in power
cable. On
battery.
Medium
Ppeak high
High
Ppeak low.
Leak?
Medium
Replace exp
flow sensor
Informational
Replace insp
flow sensor
Informational
Replace O2
sensor
Informational
Reverse exp
flow. Check
valves OK?
Medium
Flow toward the patient seen in Check the flow sensor condition. Replace the
the expiratory flow sensor during expiratory check valve between cases. Zero
inspiration for six breaths in a
flow transducers between cases.
row.
Reverse insp
flow. Check
valves OK?
Medium
Flow away from the patient seen Check the flow sensor condition. Replace the
in the inspiratory sensor during
inspiratory check valve between cases. Zero
expiration for six breaths in a
flow transducers between cases.
row.
RR high
Medium
RR low
Medium
Sample line
blocked
Medium
Service
calibration
advised
Informational
Medium
System leak?
Informational
M1145956
7-9
Avance
Message
Priority
Cause
Action
System
shutdown in <5
min
High
Informational
Turn switch on
to continue use
High
TVexp high
Medium
TVexp low
Medium
Unable to drive
bellows
Informational
Bellows is collapsed.
Using battery.
Power
Controller fail.
Medium
Ventilate
manually!
High
Ventilate
manually!
Medium
Ventilator has
no drive gas
High
Drive gas supply is not sufficient Check drive gas supply. Use a manual bag to
to mechanically ventilate.
ventilate the patient until the drive gas supply
is restored.
Informational
Medium
7-10
M1145956
Message
Priority
Cause
Action
Informational
TVexp is greater than TVinsp for Calibrate the flow sensors. Replace the flow
six breaths.
sensors if the message does not clear.
The alarm priority escalates to High if the alarm is not resolved after 2 minutes.
2The
sustained pressure threshold is calculated from the pressure limit setting. The sustained limit is calculated as
follows:
Mechanical
Ventilation with
PEEP Off:
For Pmax less than 30 cmH2O, the sustained pressure limit is 6 cmH2O.
For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of Pmax.
For Pmax greater than to 60 cmH2O, the sustained pressure limit is 12 cmH2O.
Mechanical
Ventilation with
PEEP On:
For Pmax less than 30 cmH2O, the sustained pressure limit is 6 cmH2O plus set PEEP
minus 2 cmH2O.
For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of Pmax plus set
PEEP minus 2 cmH2O.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 12 cmH2O plus set PEEP
minus 2 cmH2O.
Mechanical
Ventilation Off:
For Pmax between 12 and 60 cmH2O, the sustained pressure limit is 50% of Pmax.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 30 cmH2O.
M1145956
7-11
Avance
Alarm ranges
The alarms names are listed in order they appear on the Alarm
Limits page accessible from the Alarm Setup menu. The alarms
setting ranges show in the order they appear in the setting selections
in the Alarm Settings menu. See the Super User Mode section for
more information on the alarm default settings.
7-12
Alarm
Range
Increment
12 -100 cmH2O
1.2 - 9.8 kPa
12 - 98 mbar, hPa
9 - 73 mmHg
1 cmH2O
0.1 kPa
1 mbar, 1hPa
1 mmHg
MV High
0.5 l/min
MV Low
0.1 l/min
TV High
20 - 1600, Off ml
20 ml
TV Low
Off, 5 - 20 ml
20 - 1500 ml
5 ml
20 ml
RR High
1 bpm
RR Low
Off, 1 - 99 bpm
1 bpm
EtCO2 High
0.1%
0.1 kPa
1 mmHg
EtCO2 Low
0.1%
0.1 kPa
1 mmHg
FiCO2 High
0.1%
0.1 kPa
1 mmHg
FiO2 High
19 - 100%, Off
1%
FiO2 Low
18 - 99%
1%
EtO2 High
19 - 100%, Off
1%
EtO2 Low
Off, 1 - 99%
1%
FiISO High
0.1 - 7.0%
0.1%
FiISO Low
0.1%
FiSEV High
0.1 - 10.0%
0.1%
FiSEV Low
0.1%
FiDES High
0.1 - 20.0%
0.1%
FiDES Low
0.1%
FiENF High
0.1 - 7.0%
0.1%
FiENF Low
0.1%
FiHAL High
0.1 - 7.0%
0.1%
FiHAL Low
0.1%
EtISO High
0.1%
EtISO Low
0.1%
M1145956
M1145956
Alarm
Range
Increment
EtSEV High
0.1%
EtSEV Low
0.1%
EtDES High
0.1%
EtDES Low
0.1%
EtENF High
0.1%
EtENF Low
0.1%
EtHAL High
0.1%
EtHAL Low
0.1%
7-13
Avance
Alarm tests
Test the system to verify that alarms are functioning.
Note
If an airway module is installed, the FiO2 readings are taken from the
module instead of from the O2 cell. A sample line must be connected
from the airway module to the breathing circuit in order to test the O2
alarms.
1. Connect a test lung to the patient connection.
2. Start a case.
3. Set the Bag/Vent switch to Vent.
4. Set the O2 concentration to 30%, and allow the O2 reading to
stabilize.
For machines configured to individual gas control, set the O2
flow to approximately 500 ml/min and Air flow to
approximately 5 l/min.
5. Test the O2 alarms:
Set the FiO2 Low alarm limit to 50%. Make sure an FiO2 low
alarm occurs.
Set the FiO2 Low alarm limit back to 21% and make sure
that the FiO2 low alarm cancels.
Set the FiO2 High alarm limit to 50%.
Push the O2 flush button.
Make sure the FiO2 high alarm occurs.
Set the FiO2 High alarm limit back to 100%. Make sure that
the FiO2 high alarm cancels.
6. Test the MVexp low alarm:
Go to the Alarm Setup menu.
Set the MV Low alarm limit to greater than the measured
minute volume.
Make sure that a MVexp low alarm occurs.
Set the MV Low alarm limit to off.
7. Test the Ppeak high alarm:
Set the Pmax to less than the peak airway pressure.
Make sure that the Ppeak high alarm occurs.
Set the Pmax to the desired level.
8. Test the PEEP high. Blockage? alarm:
Close the APL valve.
Set the Bag/Vent switch to Bag. Mechanical ventilation stops.
Block the patient connection and push the O2 flush button.
Make sure that the PEEP high. Blockage? alarm occurs
after approximately 15 seconds.
7-14
M1145956
M1145956
7-15
Avance
Problem
Solution
Use a different scavenging extraction
system. Verify flow is within specification.
Replace the filter. Refer to Remove the
AGSS receiver filter in the Advanced
Breathing System Cleaning and Sterilization
Users Reference manual.
Contact a Datex-Ohmeda trained service
representative to repair the system.
7-16
M1145956
Electrical problems
WARNING
Symptom
Problem
Solution
The system circuit breaker (switch) Turn the circuit breaker on.
is off.
M1145956
One electrical outlet does not have The outlet circuit breaker is off.
power.
7-17
Avance
Pneumatic problems
Symptom
Problem
Solution
7-18
M1145956
In this section
M1145956
8-1
Avance
Setup warnings
See Standards and System components in the
Specifications and Theory of Operation section for
information on specific monitoring requirements.
WARNING
8-2
O2 monitoring.
CO2 monitoring.
Always make sure that the pipeline supply hoses and the
breathing circuit components are not toxic and will not:
M1145956
M1145956
To help prevent false alarms from devices with highintensity electrical fields:
8-3
Avance
Canister setup
The absorber canister is available in two versions: Disposable Multi
Absorber and Reusable Multi Absorber. Both are removed and
installed on the breathing system in the same way.
Each canister holds 800 grams of loose absorbent. The manufacturer
recommends MedisorbTM absorbent.
Both absorber versions should only be used with mixtures of air,
oxygen, nitrous oxide, halothane, enflurane, isoflurane, desflurane
and sevoflurane.
4
3
AB.74p042
AB.74p043
1.
2.
3.
4.
5.
6.
7.
8-4
M1145956
M1145956
8-5
Avance
1. Hold the canister by the handle and push on the release latch to
unlock the canister.
AB.74p058
Removing a canister
AB.75p088
Removing an
EZchange canister
AB.75p089
8-6
M1145956
AB.74p044
Reusable Multi
Absorber canister
filling
AB.74p046
4. Remove and properly discard the foam filters, the absorbent, and
any water in the reservoir.
AB.74p047
WARNING
M1145956
8-7
Avance
6. Place a new filter in the bottom of the canister, pour absorbent
into the canister and place a new filter over the absorbent before
closing and locking the cover. Wipe off any absorbent dust.
7. Align the cover slots with the canister locking tabs and press the
cover down into place. Turn the cover locking ring clockwise to
lock the cover in place. Ensure cover is properly sealed to
prevent leaks and spillage. Alignment of the arrows helps to
indicate correct assembly.
WARNING
8-8
M1145956
Electrical connections
Arrow shows the mains power inlet and cord.
AB.74p048
Mains inlet
Labels show outlet voltage ratings and circuit breaker amp ratings.
These are isolated outlets. Regularly test the leakage current.
AB.91p048
Outlets
M1145956
8-9
Avance
Serial port
8-10
AB.91p085
AB.91p050
M1145956
Pneumatic connections
CAUTION
AB.91p040
Pipeline inlets
Scavenging
AB.91p045
M1145956
8-11
Avance
Connect the Datex-Ohmeda sample gas exhaust tube to the gas
return port. Exhaust gas will be directed to the scavenging system.
AB.91p093
Vacuum suction
regulator (optional)
4
2
AB.74p032
1.
2.
3.
4.
8-12
M1145956
Venturi suction
regulator (optional)
4
3
AB.74p189
1.
2.
3.
4.
Venturi muffler
Overflow safety trap
Splash guard
Collection bottle connection
Auxiliary O2
flowmeter (optional)
1
AB.91p039
1. Auxiliary O2 outlet
2. Auxiliary O2 flow control
M1145956
8-13
Avance
WARNING
DIN cylinder
connections
8-14
M1145956
High-pressure leak
test
M1145956
8-15
Avance
WARNING
8-16
M1145956
9 User Maintenance
WARNING
In this section
M1145956
9-1
Avance
Repair policy
Do not use malfunctioning equipment. Make all necessary repairs or
have the equipment serviced by a Datex-Ohmeda trained service
representative. After repair, test the equipment to ensure that it is
functioning properly, in accordance with the manufacturers published
specifications.
To ensure full reliability, have all repairs and service done by a
Datex-Ohmeda trained service representative. If this cannot be done,
replacement and maintenance of those parts listed in this manual
may be undertaken by a competent, trained individual having
experience in the repair of devices of this nature.
CAUTION
Minimum Frequency
Maintenance
Daily
Two weeks
Monthly
9-2
M1145956
9 User Maintenance
Minimum Frequency
Maintenance
As necessary
Datex-Ohmeda
approved service
M1145956
9-3
Avance
AB.74p052
1. Pull the latch to unlock the flow sensor module from the breathing
system.
AB.74p065
AB.74p083
5. Replace the flow sensor module on the system and push the
latch closed to secure the module.
6. Calibrate the new O2 cell according to the Circuit O2 cell
calibration procedure.
9-4
M1145956
9 User Maintenance
Calibration menu
Access the Calibration menu by pushing the Main Menu key and
selection Calibration from the Main Menu. Select the calibration
procedure and follow the instructions shown.
See Calibration in the Airway Modules section for information on
calibrating the airway modules.
See Backlight test for information on testing the backlights.
Calibration
Flow and Pressure
Circuit O2 Cell
Airway Gas
Backlight Test
Previous Menu
M1145956
9-5
Avance
100% O2 calibration
9-6
M1145956
9 User Maintenance
Backlight test
1. Push the Main Menu key.
2. Select Calibration.
3. Select Backlight Test.
4. Select Start Test.
5. The display will show the test running on light 1 and then on light
2. If the display goes completely blank or flickers during the test,
one of the lights has failed. Contact a Datex-Ohmeda trained
service representative to replace the backlights.
M1145956
9-7
Avance
9-8
M1145956
10 Parts
Note
In this section
M1145956
10-1
Avance
AB.82.019
3
2
Item
1
2
3
10-2
Description
Stock number
1407-7001-000
1407-3000-000
1407-3004-000
1503-3858-000
1503-3244-000
M1145956
10 Parts
AB.82.021
5
Item
1
2
3
4
5
-
M1145956
Description
Stock number
1407-7002-000
1407-3101-000
1406-8219-000
1406-3466-000
6050-0004-110
1009-5570-000
1503-3857-000
10-3
Avance
Bellows
2
8
3
7
10-4
AB.82.018
4
5
Item
Description
Stock number
1
2
3
4
5
6
7
8
-
Bellows housing
Bellows
Rim
Pressure relief valve assembly
Latch, rim
Manifold, bellows base
Bellows base with latch
Seal, base
Diaphragm, APL
Poppet, APL valve
Cage, APL
1500-3117-000
1500-3378-000
1500-3351-000
1500-3377-000
1500-3352-000
1407-3702-000
1407-7006-000
1500-3359-000
1406-3331-000
1406-3332-000
1406-3333-000
M1145956
10 Parts
AB.82.001
Item
Description
Stock number
Breathing circuit module (does not include APL valve,O2 cell, o-ring,
or cable)
M1145956
10-5
Avance
Absorber canister
2
3
1
AB.82.017
Item
Description
Stock number
1407-7004-000
2
3
4
5
-
10-6
1009-8240-000
1407-3201-000
1407-3204-000
1407-3200-000
8003138
8003963
M1145956
10 Parts
M1145956
Description
Stock number
1407-7005-000
10-7
Avance
AGSS
Description
Stock number
Common
Cap 3.18 barb silicone
Connector, inlet 30 mm male to 9 mm male
Connector, inlet 30 mm male to 30 mm male
O-ring for connector, 21.95 ID
O-ring for receiver, 22 ID
O-ring for thumbscrews, 4.47 ID
Reservoir scavenger
Seal, down tube scavenger
Seal, receiver scavenger
Thumbscrew M6 X 28.5
Thumbscrew, M6 X 43
Valve, unidirectional (complete assembly)
1406-3524-000
M1003134
M1003947
1406-3558-000
1407-3104-000
1407-3923-000
1407-3903-000
1407-3904-000
1407-3901-000
1406-3305-000
1406-3304-000
1406-8219-000
Passive AGSS
Adapter, outlet 30 mm female to 19 mm male (pack
of 5)
Exhaust hose
Plug assembly 30 mm ISO
Screw, shoulder 4 diameter X 4 L M3 X 0.5 sst
1500-3376-000
8004461
1407-3909-000
1407-3915-000
8004460
1407-3909-000
1406-3521-000
1407-3902-000
10-8
1406-3521-000
1407-3902-000
M1145956
10 Parts
AB.82.044
AB.82.057
Item
Description
Stock number
1407-7021-000
2
3
4
-
M1145956
1407-3126-000
1407-3130-000
1407-7024-000
1407-7027-000
10-9
Avance
Condenser
AB.82.045
Item
Description
Stock number
10-10
M1145956
10 Parts
M1145956
Description
Stock number
755583
755571
8004463
755534
M1006864
0210-5022-300
1202-3651-000
0219-3415-800
1202-7146-000
0219-3372-600
0309-1319-800
1001-8975-000
1001-8976-000
1009-3119-000
1009-3356-000
1202-3641-000
0219-7210-300
2900-0001-000
1006-4198-000
1006-4199-000
1006-4200-000
1102-3016-000
0206-3040-542
0236-0039-870
10-11
Avance
10-12
M1145956
Note
Note
In this section
M1145956
11-1
Avance
AB.91.103
11-2
M1145956
M1145956
36. Bag
37. APL valve 0-70 cmH2O
38. Optional factory connection
39. Default factory connection
40. Gas monitor
41. Negative pressure relief valve
42. Absorber
43. Drain
44. Expiratory flow sensor
45. Inspriatory flow sensor
46. Patient Y
47. Airway transducer
48. Inspiratory flow transducer
49. Expiratory flow transducer
50. Manifold pressure transducer
51. Passive gas scavenging interface
52. 30-mm male, to disposal system
53. 0.3 cmH2O entrainment
54. Reservoir
55. Adjustable gas scavenging interface
56. DISS EVAC connector
57. Active gas scavenging interface
58. Room air
59. Filter
60. High or low flow restrictor
61. Flow indicator
62. To disposal system
63. ACGO variant
64. Circle or ACGO mode switch
65. 22 mm ACGO port
66. O2 cell
67. Non-circle variant
68. Circle or non-circle mode switch
69. 150 cmH2O fresh gas pressure limiting valve
70. EZchange canister
71. Condenser
11-3
Avance
Gas supplies
WARNING
O2 flow
Mixed gas
Either Air or N2O can be selected as the balance gas. The balance
gas flow is controlled by the mixer. When Alternate O2 control is in
use, only O2 is flowing and balance gas is disabled.
The mixed gas goes from the mixer outlet through the vaporizer to the
fresh gas outlet and into the breathing system. A pressure-relief valve
sets the maximum outlet pressure.
EZchange canister
11-4
M1145956
Condenser
Pneumatic specifications
CAUTION
Gas supplies
Pipeline gases
Cylinder gases
Cylinder connections
M1145956
A relief valve on the vaporizer manifold limits the fresh gas pressure
at the ACGO port to 55 kPa (8 psi) at 45 l/min.
A relief valve limits the fresh gas pressure at the inspiratory port to
27 kPa (4 psi) at 55 l/min when non-circle circuit is selected.
11-5
Avance
AB.91.114
11-6
M1145956
1. Power cord
2. AC inlet and breaker
3. Inrush board
4. Transformer
5. Fuse block
6. Outlet box with breakers
7. Line filter
8. Power controller board
9. Power supply fan with cable
10. Power supply with batteries
11. Display connector board
12. Anesthesia control board
13. Airway module power supply board
14. Airway module
15. Display unit CPU board
16. Display unit system interface board
17. Display unit user interface board
18. Display unit fan with cable
19. LCD backlight 1
20. Compact flash I/O port
21. USB I/O port
22. USB I/O port
23. Network I/O port (Ethernet)
24. Patient monitoring on/standby
25. Serial I/O port
26. LCD backlight 2
27. Electronic gas mixer board
28. Pan connector board
29. Ventilator interface board
30. I/O port
31. Pan fan with cable
32. ABS filter board
33. Air cylinder pressure transducer
34. N2O cylinder pressure transducer
35. O2 cylinder pressure transducer
M1145956
11-7
Avance
Electrical power
Supply voltage
Inlet circuit breakers
Power cord
Length
Voltage rating
Current capacity
Type
WARNING
11-8
5 meters
100 to 240 Vac
10 A for 220-240 Vac
15 A for 100-120 Vac
Three conductor power supply cord (medical grade
where required).
M1145956
Battery information
The system is not a portable unit; a sealed lead acid battery supplies
backup power in the event of a power failure.
Capacity to operate for 90 minutes under typical operating
conditions; 30 minutes under extreme conditions.
The system functions to specifications through the transition to
battery power.
Only trained service representatives are to replace the battery.
Batteries must be disposed of in accordance with applicable
regulatory requirements in effect at the time and place of disposal.
Contact a trained service representative to disconnect the battery if
the equipment is not likely to be used for an extended time.
Flow specifications
Alternate O2
Flow range
Indicator
Indicator accuracy
Fresh gas
Flow range
M1145956
11-9
Avance
Absorbent
Connections
System leakage
System compliance
Pressure required to
open Inspiratory or
expiratory valves
Pressure generated by
a wet unidirectional
valve
APL valve
Approximately 0 to 70 cmH2O
kPa
cmH2O
5
0.06
0.6
30
0.22
2.2
60
0.52
5.3
Ezchange canister system and condenser, absorber mode
5
0.06
0.6
30
0.24
2.4
60
0.57
5.8
EZchange canister system and condenser, canister removed
5
0.06
0.6
30
0.24
2.4
60
0.49
5.0
*Values include patient circuit tubing and Y-piece 4.9 kPa (0.5 cmH2O)
at 1 l/s. Patient circuit tubing and breathing system configurations may
affect resistance.
11-10
M1145956
Flow (l/s)
3
10
30
60
70
0.05
0.17
0.51
1.0
1.17
0.78
1.14
1.43
2.61
3.21
Gas scavenging
All scavenging
Positive pressure relief
10 cmH2O
Passive scavenging
Negative pressure relief
Outlet connector
0.3 cmH2O
30 mm male taper ISO
Active scavenging
Disposal system
type
Outlet
connector*
Flow range
Pressure
Adjustable
High flow, low
vacuum
Low flow, high
vacuum
DISS EVAC
BS6834
Up to 30 l/min
50 to 80 l/min
DISS EVAC
36 +/- 4 l/min
not applicable
1.6 kPa
(12 mmHg)
305 mmHg
(12 inHg)
minimum
vacuum
not applicable
M1145956
11-11
Avance
Physical specifications
All specifications are approximate values and can change without
notice.
CAUTION
System
Casters
Drawers
Ventilator display
Height
136 cm
Width
76 cm
Depth
76 cm
Weight
135 kg
Top of machine weight 34 kg
limit
13 cm with brakes on the front casters
23 cm Height x 33 cm Width x 27 cm Depth
146 x 184 mm (31 cm diagonal)
Environmental requirements
Operation
Temperature 10 to 40C
Oxygen cell operates to
specifications at 10 to 40C
Humidity
Altitude
11-12
Storage
Compensation range
-25 to 60C
Oxygen cell storage is -15 to
50C, 10 to 95% RH, 500 to 800
mmHg
10 to 95%, non-condensing
Not applicable
Not applicable
M1145956
Gas specifications
Airway humidity
Sampling delay
Total system response time
Warm-up time
Diversion flow
0 to 100% condensing
2.5 seconds typical with a 3 m sampling line
2.5 seconds typical with a 3 m sampling line,
including a sampling delay and rise time
2 minutes for operation with CO2, O2, and
N2O
5 minutes for operation of anesthetic agents
30 minutes for full specifications
200 ml/min
M1145956
During warm-up 2 to 10
minutes (anesthetic agents 5
to 10 minutes), under normal
conditions.
11-13
Avance
Typical performance
CO2
11-14
M1145956
Pharyngeal Suction
Air or O2 from system gas supply
28 l/min with pipeline drive gas at 280 kPa
52 l/min with pipeline drive gas at 600 kPa
Maximum Vacuum*
600 mmHg with pipeline drive gas at 280 kPa
550 mmHg with pipeline drive gas at 600 kPa
Maximum Flow*
29 l/min with pipeline drive gas at 280 kPa
32 l/min with pipeline drive gas at 600 kPa
Vacuum Gauge Accuracy +/- 5% of full scale
*Values are approximate.
Pharyngeal Suction
External vacuum
540 mmHg with external vacuum appplied of
540 mmHg and 40 l/min free flow
Maximum Flow*
39 l/min with external vacuum applied of 540
mmHg and 40 l/min free flow
Vacuum Gauge Accuracy +/- 5% of full scale
*Values are approximate.
M1145956
11-15
Avance
Ventilator theory
The ventilator pneumatics are at the rear of the breathing system. A
precision valve controls gas flow to the patient. During inspiration, this
gas flow closes the exhalation valve and pushes the bellows down.
During expiration, a small flow pressurizes the exhalation diaphragm
to supply PEEP pressure.
Volume and flow measurements come from flow sensors in the flow
sensor module. Two tubes from each sensor connect to a transducer
that measures the pressure change across the sensor, which
changes with the flow. A third transducer measures airway pressures
at the inspiratory flow sensor.
The ventilator uses the data from the flow sensors for volume-related
numerics and alarms. The ventilator also uses the flow sensors to
adjust its output for changes in fresh gas flow, small leaks, and gas
compression upstream of the breathing circuit. There is adjustment
for compression in the patient circuit.
In volume ventilation modes, certain alarm conditions prevent the
automatic adjustment of ventilator delivery based on measured flow
values. In these cases, Vent mode: TV accuracy decreased. Adj.
manually. shows above the ventilator setting area on the screen.
When this message shows, the ventilator may not be able to deliver
within the accuracy range specified. When this occurs, manually
adjust the tidal volume until the volume delivered reaches the desired
level. If compensation stops for a number of breaths, the condition
causing the hold shows as an alarm. Automatic volume
compensation resumes when alarm conditions are resolved.
For better precision a small quantity of gas bleeds through a resistor
to help keep the pressure on the exhalation valve constant. At high
airway pressures, this can cause a slight hiss during inspiration.
WARNING
11-16
M1145956
O2 monitoring theory
of operation
Ventilation modes
M1145956
11-17
Avance
Volume control mode
AB.98.036
1.
2.
3.
4.
5.
6.
7.
Paw waveform
Tinsp
Insp Pause
Texp
PEEP
Flow waveform
TV
11-18
M1145956
AB.98.037
1.
2.
3.
4.
5.
6.
Paw waveform
Tinsp
Texp
Pinsp
PEEP
Flow waveform
M1145956
11-19
Avance
SIMV/PSV mode
4
5
6
1.
2.
3.
4.
5.
6.
AB.91.040
11-20
M1145956
M1145956
11-21
Avance
2
3
AB.91.041
1. Paw
2. PEEP
3. Time
11-22
M1145956
AB.91.038
1.
2.
3.
4.
5.
6.
7.
Paw waveform
Tinsp
Spontaneous breathing period
Trigger window
Pressure supported breath
Pinsp
Flow waveform
M1145956
11-23
Avance
PCV-VG mode
AB.98.034
1.
2.
3.
4.
5.
6.
7.
Paw waveform
Tinsp
Texp
Variable pressure to deliver desired TV
PEEP
Flow waveform
TV
11-24
M1145956
Fresh gas
compensation
Anesthesia system
Medical Air or O2
350 kPa (50 psi)
240 to 700 kPa (35 to 102 psi)
120 l/min at 240 kPa (35 psi), 0.75
seconds
80 l/min at 240 kPa (35 psi)
1 to 120 l/min at 240 kPa (35 psi)
Pressure
High pressure alarm set range
Sustained pressure alarm range
Display range
Volume
Tidal volume display range
Setting range
Minute volume
Breath rate
Oxygen
Display range
Display resolution
Sensor type
Measurement range
Measurement accuracy
Cell response time
M1145956
0 to 110% O2
1% increments
Galvanic fuel cell
0 to 100% O2
Better than 3% of full scale
35 seconds
Note: Response time of cell and
adapters is measured using the test
method described in ISO 7767 (1997).
18% to 99%
19% to 100% or Off
Note: Low O2 limit may not be set
above high O2 limit. High O2 limit may
not be set below the low O2 limit.
Four months of shelf life (23C room
air) and one year of normal operation.
11-25
Avance
Delivery accuracy
greater than 210 ml tidal volume accuracy better than 7%
less than 210 ml but greater than 60 ml
tidal volume - accuracy better than 15 ml
less than 60 ml tidal volume - accuracy
better than 10 ml
Volume monitoring accuracy
greater than 210 ml tidal volume accuracy better than 9%
less than 210 ml but greater than 60 ml
tidal volume - accuracy better than 18 ml
less than 60 ml tidal volume - accuracy
better than 10 ml
Inspiratory pressure delivery accuracy greater of +/- 10% or +/- 3 cmH2O
PEEP delivery accuracy
+/- 1.5 cmH2O
Pressure monitoring accuracy
greater of +/- 5% or +/- 2 cmH2O
Note: Gas composition errors may be in addition to the above normalized
accuracy. When adding errors, positive errors can have the effect of nulling out
negative errors.
Note: Use of anesthetic agent could affect the errors by approximately -0.95%/%
volume agent in normal mode.
AB.74.027
11-26
M1145956
Guidance and
manufacturers
declaration electromagnetic
emissions
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
The system uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The system is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic
Class A
emissions
IEC 61000-3-2
Voltage
Complies
fluctuations/
flicker emissions
IEC 61000-3-3
M1145956
11-27
Avance
Guidance and
manufacturers
declaration electromagnetic
immunity
Power immunity
Immunity test
Compliance level
6 kV contact
8 kV air
Electrical fast
transient/burst IEC
61000-4-4
11-28
M1145956
Radiated immunity
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
10 Vrms
150 kHz to 80 MHz in
ISM bands
10 V/m
80 MHz to 2.5 GHz
3 Vrms (V1)
10 Vrms (V2)
D=12P
M1145956
10 V/m (E1)
11-29
Avance
Recommended
separation distances
0.01
0.35
1.2
0.12
0.23
0.1
1.1
3.8
0.38
0.73
1
3.5
12
1.2
2.3
10
11
38
3.8
7.3
100
35
120
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz to 800 MHz the separation distance for the higher frequency range applies.
Note 2: The ISM (Industrial, Scientific and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the
ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the
likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into
patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
11-30
M1145956
Electrical safety
WARNING
WARNING
M1145956
11-31
Avance
Standards
Devices used with this anesthesia system shall comply with the
following standards where applicable:
Breathing system and breathing system components ISO 8835-2.
Anesthetic vapor delivery devices ISO 8835-4.
Anesthetic agent monitors ISO 11196.
Oxygen monitors ISO 7767.
Carbon dioxide monitors ISO 9918.
Exhaled volume monitors IEC 60601-2-13.
11-32
M1145956
System components
Integral
Not integral
WARNING
O2 monitoring.
CO2 monitoring.
M1145956
11-33
Avance
11-34
M1145956
In this section
M1145956
12-1
Avance
Install/Service menu
WARNING
Important
Install/Service - Page 1
Cumulative Gas Usage
Show Alarm Limits
Fresh Gas Controls
VCV Cardiac Bypass
External Gas Monitor
Volume Apnea Setup
Time and Date
Factory Defaults
Install/Service - Page 2
Exit
Yes
O2%
No
No
Yes, No
O2%, Flow, User
Yes, No
Yes, No
Install/Service - Page 2
Trends Setup
Colors and Units
Parameter Settings
Page Setup
Case Defaults
Installation
Service
Install/Service - Page 1
Exit
12-2
M1145956
M1145956
12-3
Avance
Menus
Not all menu items are available depending on the system
configuration. Inactive menu items are grayed out and are not
selectable.
Menu selections shown are the factory default values. The options
are listed to the right of the menu shown.
Cumulative gas
usage
View the total system fresh gas usage since the last reset or reset the
gas usage to 0 through the Cumulative Usage menu.
Cumulative Usage
Reset Usage
Desflurane
Enflurane
Halothane
Isoflurane
Sevoflurane
O2 (*1000 l)
Air (*1000 l)
N2O (*1000 l)
Previous Menu
12-4
M1145956
Use the Volume Apnea Setup menu to enable or disable the users
ability to turn the volume apnea alarms off during manual ventilation.
Set the Volume Apnea Selection to Enable to allow the user to turn
the volume apnea alarm on or off from the Start Case menu and the
Alarm Setup menu. Set the Volume Apnea Selection for each case
type default through this menu.
Set the Volume Apnea Selection to Disable to set the volume
apnea alarms to be always on during manual ventilation. When set to
Disable, the Volume Apnea menu selection is not available from the
Start Case menu and the Alarm Setup menu.
Menu selection names shown in all capital letters are configured from
the Case Defaults menu
Enable
Enable, Disable
Off
On, Off
Off
On, Off
Off
On, Off
Off
On, Off
M1145956
12-5
Avance
Use the Time and Date menu to set the time and date. The Hour
field is dependent on the Clock Format setting. The Clock Format
can be set to a 12 or 24 hour format.
12-6
M1145956
Trends setup
Use the Trends Setup menu to set the default trend that shows in the
right window on the user Trends menu. Access the Graphical
Trends menu from the Trends Setup menu.
Trends Setup
Default Trend Num
Graphical Trends
Previous Menu
Graphical Trends
Page 1
Page 2
Page 3
Page 4
Page 5
Previous Menu
M1145956
12-7
Avance
Page 1
Field 1 Pres
Field 2 TVexp
Field 3 CO2
Previous Menu
The available selections for Field 1, Field 2, and Field 3 are the same
for each field on each page. Selections are Off, rr+co2, Pres, MVexp,
CO2, O2, Bal, AA1, AA2, N2O, MAC, TVexp, Pmean, Spont, Compl,
VO2. The default settings for Field 1, Field 2, and Field 3 of each
page show in the table.
Page 1
Page 2
Page 3
Page 4
Page 5
Pres
TVexp
CO2
O2
N2O
AA1
AA2
N2O
MAC
Bal
MAC
MVexp
rr+co2
Compl
Off
12-8
M1145956
Set the units of the parameters through the Colors and Units menu.
Access the Colors menu through Colors and Units menu.
kg
%
kPa
cmH2O
kg, lb
%, kPa, mmHg
psi, kPa, bar
kPa, hPa, cmH2O, mmHg, mbar
Setting colors
Set the colors of the parameter waveform information, digit fields, and
trends through the Colors menu.
Colors
Paw
Flow
Resp
CO2
Previous Menu
Yellow
Green
White
White
M1145956
12-9
Avance
Parameter settings
Parameter Settings
TV Based on ATPD
CO2 Numbers Dry
Previous Menu
Page Setup
Set the preset normal screen layouts through the Page Setup menu.
Menu selection names shown in all capital letters are configured from
the corresponding submenu.
Page Setup
Edit/View Page:
Available Pages
PAW TUBES
LOOPS
BIG WAVE
LOCAL
Previous Menu
12-10
ATPD, BTPS
Dry, Wet
PAW TUBES
Back
Name
Waveform Field 1
Waveform Field 2
Waveform Field 3
Digit Field
Split Screen
PAW TUBES
Paw
Flow
CO2
AA
Paw
M1145956
PAW TUBES
LOOPS
BIG WAVE
LOCAL
PAW TUBES
Paw
Flow
CO2
AA
Paw
LOOPS
Paw
Flow
CO2
AA
Spiro
BIG WAVE
Paw
Off
CO2
Flow
Gas
LOCAL
Off
Off
CO2
Supply
None
M1145956
12-11
Avance
Configuring case
defaults
Set the user selectable default case types that show in the users
Start Case menu through the Case Defaults menu. The Set to Last
Case selection is only available on submenus if a case has occurred.
Case Defaults
Edit/View Default:
Available Defaults:
ADULT
PEDIATRIC
LOCAL
CUSTOM 1
Previous Menu
ADULT
Back
Set to Last Case
Name ADULT
Patient and Sensor Type Adult
Patient Weight 70 kg
Patient Age 40 y
Other Gas Air
Circuit Circle
Data Source Vent
Vent Mode VCV
-More-
12-12
M1145956
M1145956
ADULT
PEDIATRIC
LOCAL
CUSTOM 1
ADULT
Adult
70 kg
40 y
Air
Circle
Vent
VCV
500
12
1:2
Off
Off
40
*
*
*
*
*
*
100
6.00
Fast
Paw
Paw
Flow
CO2
AA
Off
3
On
On
10.0
2.0
1000
Off
Off
Off
8.0
3.0
Off
Off
21
Off
Off
5
PEDIATRIC
Pedi
18 kg
5y
Air
Circle
Vent
PCV
*
*
*
*
*
*
5
12
1:2
Off
40
Auto
100
6.00
Fast
Gas
Paw
Flow
CO2
AA
Off
3
On
On
10.0
2.0
1000
Off
Off
Off
8.0
3.0
Off
Off
21
Off
Off
5
LOCAL
Adult
70 kg
40 y
Air
Circle
Vent
VCV
500
12
1:2
Off
Off
40
*
*
*
*
*
*
100
0.20
Fast
Gas
Paw
Flow
CO2
AA
Off
1
Off
Off
Off
Off
Off
Off
Off
Off
8.0
3.0
Off
Off
21
Off
Off
5
CUSTOM 1
Adult
70 kg
40 y
Air
Circle
Vent
VCV
500
12
1:2
Off
Off
40
*
*
*
*
*
*
100
6.00
Fast
Gas
Paw
Flow
CO2
AA
Off
3
On
On
10.0
2.0
1000
Off
Off
Off
8.0
3.0
Off
Off
21
Off
Off
5
12-13
Avance
Default settings for default case types
FiISO Low
EtISO High
EtISO Low
FiSEV High
FiSEV Low
EtSEV High
EtSEV Low
FiDES High
FiDES Low
EtDES High
EtDES Low
FiENF High
FiENF Low
EtENF High
EtENF Low
FiHAL High
FiHAL Low
EtHAL High
EtHAL Low
12-14
ADULT
PEDIATRIC
LOCAL
CUSTOM 1
Off
Off
Off
8
Off
Off
Off
15
Off
Off
Off
5
Off
Off
Off
5
Off
Off
Off
Off
Off
Off
8
Off
Off
Off
15
Off
Off
Off
5
Off
Off
Off
5
Off
Off
Off
Off
Off
Off
8
Off
Off
Off
15
Off
Off
Off
5
Off
Off
Off
5
Off
Off
Off
Off
Off
Off
8
Off
Off
Off
15
Off
Off
Off
5
Off
Off
Off
5
Off
Off
Off
M1145956
Index
A
Abbreviations 1-7
Absorber canister
changing absorbent 8-5
filling 8-7
parts 10-6
removing 8-6
setup 8-4
ACGO 2-9
port relief 11-5
positive low-pressure leak test 5-10
scavenging a gas monitor sample flow 2-10
scavenging from an auxiliary manual
breathing circuit 2-10
scavenging the ACGO sample flow 2-9
Active AGSS 3-30
Advanced breathing system 2-5
optional components 2-7
parts 10-5
AGSS
active 3-30
connecting active adjustable 3-31
connecting active with a flow indicator 3-31
parts 10-8
passive 3-29
Air
adjust 3-9
Airway modules 6-2
agent identification 6-6
calibration 6-6
connection to a patient 6-4
parameter setup 6-5
specifications 11-13
Alarm 7-2
battery indicator 7-3
de-escalating 7-3
display changes 7-2
internal failure 7-3
list of 7-4
priorities 7-2
ranges 7-12
silencing 7-2
tests 7-14
Alarm priorities 7-2
M1145956
B
Backlight test 9-7
Battery
indicator 7-3
information 11-9
Battery indicator 7-3
Bellows assembly
parts 10-4
Breathing circuit module
parts 10-3
Breathing system
problems 7-16
specifications 11-10
C
Calibration
100% O2 9-6
21% O2 9-6
airway module 6-6
backlight test 9-7
flow sensor 5-4
O2 cell 9-6
Calibration menu 9-5
Canister setup 8-4
Cardiac bypass 3-18
Changing circuit type 3-10
Checkout menu 5-5
I-1
Avance
Circuit
circle 3-10
non-circle 2-8, 3-10
non-circle relief 11-5
Circuit compliance compensation 5-4
CO2 alarms 3-24
Condenser
operation 3-28
parts 10-10
theory 11-5
Configuring case defaults 12-12
Connections
electrical 8-9
pneumatic 8-11
Controls
alternate O2 3-26
on the display 2-13
vaporizer 2-11
Cylinder
installation 8-14
D
De-escalating alarms 7-3
Digit field 2-16
setup 3-21
DIN cylinder connections 8-14
Display changes during alarms 7-2
Display controls 2-13
E
Electrical
power specifications 11-8
problems 7-17
safety 11-31
Electrical block diagram 11-6
Electromagnetic compatibility 11-27
End case 3-6
Environmental requirements 11-12
Exhalation valve assembly
parts 10-7
EZchange canister 3-27
parts 10-9
removal 8-6
theory 11-4
I-2
F
Flow and pressure calibration 9-5
Flow sensors
parts 10-2
prevent water buildup 9-7
Flow specifications 11-9
Fresh gas usage 3-19
G
Gas scavenging
specifications 11-11
Gas setup 3-9
Gas supplies 11-4
specifications 11-5
H
High-pressure leak test 8-15
How to attach equipment to the top of the
machine 8-16
How to install gas cylinders 8-14
I
IEC 60601-1 Classification 11-32
Individual checks 5-8
circuit 5-8
circuit O2 cell 5-9
Low P leak 5-9
Low P leak (machines with ACGO) 5-9
system 5-8
Install/Service menu 12-2
Intended use 1-2
Internal failure 7-3
L
Leak 5-5
List of alarms 7-4
Low Pressure Leak test 5-9
M1145956
M
MAC 3-5
Machine check 5-6
circuit 5-7
circuit O2 5-7
monitor 5-7
system 5-6
Main menu 3-16
Mains inlet 8-9
Maintenance schedule 9-2
Minimum Alveolar Concentration 3-5
MV/TV alarms 3-24
Q
Quick keys
changing gas settings using 3-9
changing ventilator settings using 3-7
O
O2 cell
calibration 9-6
replacement 9-4
theory 11-17
O2 flow
adjust 3-9
start 3-3
Outlets 8-9
Overview
system 2-2
P
Parameters setup 6-5
agent setup 6-5
CO2 setup 6-5
data source 6-5
O2 setup 6-5
spirometry setup 6-5
Passive AGSS 3-29
Pin indexed cylinder yokes 8-14
Pipeline inlets 8-11
Pneumatic
connections 8-11
problems 7-18
specifications 11-5
system pneumatic circuits 11-2
Positive low pressure leak test (ACGO
systems only) 5-10
Preoperative Checkout 4-1
M1145956
specifications 11-11
Screen configuration 3-20
Screen setup 3-21
Select page 3-20
Serial port 8-10
Service schedule 9-3
Setting alarm limits 3-23
Setting gas controls 3-22
Setup
changing circuit type 3-10
changing gas and settings 3-9
gas 3-9
vent 3-7
Setup warnings 8-2
Silencing alarms 7-2
Specifications
breathing system 11-10
flow 11-9
gas scavenging 11-11
physical 11-12
pneumatic 11-5
ventilator operating 11-25
Spirometry
menu functions 3-11
scaling 3-13
setting loop type 3-12
I-3
Avance
Spirometry setup 3-14
selecting a data source 3-15
setting patient and sensor type 3-14
setting spirometry split screen 3-15
Split screen 3-22
Standards 11-32
Start case 3-6
Start gas flow 3-3
Suction regulator
specifications 11-15
vacuum 8-12
Venturi 8-13
Super User menus
case defaults 12-12
colors and units 12-9
cumulative gas usage 12-4
graphical trends 12-7
page setup 12-10
parameter settings 12-10
time and date 12-6
trends setup 12-7
volume apnea setup 12-5
Super User mode 12-1
menus 12-4
using 12-3
Sweep speed 3-22
Symbols 1-4
System
components 11-33
overview 2-2
pneumatic circuits 11-2
turn on 3-2
System inspection 5-2
V
Vacuum suction regulator 8-12
Vaporizer
back pressure test 5-11
controls 2-11
installation 5-3
Vaporizer back pressure test 5-11
Ventilation modes 11-17
PCV-VG 11-24
pressure control 11-19
PSVPro 11-21
SIMV/PSV 11-20
SIMV-PC 11-23
volume control 11-18
Ventilator
accuracy data 11-26
changing modes and settings 3-8
modes 11-17
operating specifications 11-25
setup 3-7
theory 11-16
Venturi suction regulator 8-13
Volume apnea 3-24
W
Water buildup
prevention 9-7
Waveform 2-16
setup 3-21
T
Test tools and system parts 10-11
Trends 3-17
Troubleshooting
breathing system 7-16
electrical 7-17
pneumatic 7-18
Turning system on 3-2
U
Using menus 2-16
I-4
M1145956
Warranty
This Product is sold by Datex-Ohmeda under the warranties set forth
in the following paragraphs. Such warranties are extended only with
respect to the purchase of this Product directly from Datex-Ohmeda
or Datex-Ohmedas Authorized Dealers as new merchandise and are
extended to the Buyer thereof, other than for the purpose of resale.
For a period of twelve (12) months from the date of original delivery to
Buyer or to Buyers order, but in no event for a period of more than
two years from the date of original delivery by Datex-Ohmeda to a
Datex-Ohmeda Authorized Dealer, this Product, other than its
expendable parts, is warranted against functional defects in materials
and workmanship and to conform to the description of the Product
contained in this Users Reference manual and accompanying labels
and/or inserts, provided that the same is properly operated under the
conditions of normal use, that regular periodic maintenance and
service is performed and that replacements and repairs are made in
accordance with the instructions provided. This same warranty is
made for a period of thirty (30) days with respect to expendable parts.
The foregoing warranties shall not apply if the Product has been
repaired other than by Datex-Ohmeda or in accordance with written
instructions provided by Datex-Ohmeda, or altered by anyone other
than Datex-Ohmeda, or if the Product has been subject to abuse,
misuse, negligence, or accident.
Datex-Ohmedas sole and exclusive obligation and Buyers sole and
exclusive remedy under the above warranties is limited to repairing or
replacing, free of charge, at Datex-Ohmedas option, a Product,
which is telephonically reported to the nearest Datex-Ohmeda
Customer Service Center and which, if so advised by Datex-Ohmeda,
is thereafter returned with a statement of the observed deficiency, not
later than seven (7) days after the expiration date of the applicable
warranty, to the Datex-Ohmeda Customer Service and Distribution
Center during normal business hours, transportation charges prepaid,
and which, upon Datex-Ohmedas examination, is found not to
conform with above warranties. Datex-Ohmeda shall not be
otherwise liable for any damages including but not limited to
incidental damages, consequential damages, or special damages.
There are no express or implied warranties which extend beyond the
warranties hereinabove set forth. Datex-Ohmeda makes no warranty
of merchantability or fitness for a particular purpose with respect to
the product or parts thereof.
Avance
Users Reference Manual
English
M1145956
05 08 001 18 05 02
Printed in USA