Laser light opens new frontiers in endoscopy

New-generation Endoscope System

Ref. No. XB-000E

(SK1402F1079F9711) Printed in Japan 2014 FUJIFILM Corporation

A new generation
of the endoscope system

Fujifilms pursuit of a new light source that illuminates lesions


more clearly has resulted in an endoscope system equipped
with the laser light source.

Lasereo is a trademark of FUJIFILM Corporation.

Two laser sources


LASEREO is equipped with laser light sources that emit two
different laser beams at different wavelengths: white light
mode laser and BLI mode laser. The white light mode laser
excites phosphors to create white light illumination
(Oscillation wavelength: 450 nm 10 nm). The BLI mode
laser enables the structures of vessels and mucosal surface
to be captured through high contrast signals (Oscillation
wavelength: 410 nm 10 nm). By controlling the emission
intensity ratio of the two lasers and combining it with
image processing, four types of observation modes are
available.

White light

White light mode laser


Phosphor

Narrowband
light

BLI mode laser

Conceptual diagram of laser illumination


(The illustrations differ from laser light sources in actual use)

BLI (Blue LASER Imaging)


Image obtained from narrowband light

Image obtained from white light


White light

Narrowband light

Mucous
membrane
surface layer

Mucosa

Signal
processing
BLI image

BLI produces high-contrast


images of superficial mucosal
structures and microvessels by
combining two image
components induced by
narrowband light and white
light.

Blood vessels

Four observation modes


The white light and BLI mode laser are emitted at a controlled intensity ratio to obtain the optimal light according
to the observation purposes. In combination with image processing, four types of observation modes are available.

[White light/FICE* mode]

White light mode

[BLI-bright mode]

[BLI mode]

White light mode laser

BLI mode laser

BLI mode laser

BLI mode laser

White light mode laser

White light mode laser

Provides images having the same color tone as with the conventional system (xenon light source).

FICE* mode

Constructs special images from rays having specific wavelenghs which are useful for better enhancemente of tissue aspects and vessels.

BLI-bright mode

Emphasizes blood vessels and mucosal surface structures in middle to near distant views, which are brighter than the BLI mode.

BLI mode

Emphasizes blood vessels and mucosal surface structures in near distant views or close up.

* Flexible spectral Imaging Color EnhancementImage-Enhanced Endoscopy

Double laser imaging stepping into the new world

Lower power consumption and longer life than


with conventional lighting
This laser light source consumes only 10 W while a conventional xenon lamp needs 300 W to generate the same
amount of light. In case a xenon lamp-installed endoscope has to be replaced regularly, there is no need with the
laser source.

Four observation modes


to suit various purposes

Up to four observation modes can be allocated to the scope switch.


White light

White light

White light

BLI

FICE

FICE

BLI

BLI-bright
BLI

FICE* mode

White light mode


The white light mode provides bright, sharp stereoscopic images. The color tones are the same as with the conventional xenon light source.

Stomach

Colon

FICE emphasizes minute differences in color tones between normal mucosa and lesions by the spectral image
processing to clearly show mucosal structures and such lesions. The images are bright from near to distant, making
it easy to identify a lesion and observe the boundary with the normal tissue.

OFF (Stomach)

ON

* Flexible spectral Imaging Color EnhancementImage-Enhanced Endoscopy

BLI-bright mode
The BLI-bright mode, in which the white light and narrowband light are used at an optimal balance, provides
brighter images even from a middle to distant view compared to those obtained under typical narrowband light.
This supports a wide-ranging observation including middle to near distant, close-up and magnified images.

Stomach

Colon

BLI mode
By increasing the ratio of the narrowband light, BLI made aims at depicting sharp images of superficial microvessels
and mucosal surface structures. BLI images support close-up and magnified observations.

Esophagus

Esophagus

Gastroscopes for use with the laser light source


For the upper Optical magnification

For the upper G.I. Tract-Standard Type

EG-L590ZW

EG-L590WR
This endoscope is designed to support identification of lesions by using
FICE or BLI images. The water and air supply systems are improved for
better draining to ensure a clear view.

Capturing clear images never seen before

This endoscope offering close-up BLI images is suitable for observation


of superficial microvessels and mucosal surface structures.
The maximum optical magnification is approximately 135 times (on
19-inch LCD monitors).

Viewing direction

0 (Forward)

Viewing direction

0 (Forward)

Field of view

140

Field of view

WD: 140/TL: 55

Observation range

4-100 mm

Observation range

WD: 6-100 mm/TL: 2-3 mm

Distal end diameter

9.6 mm

Distal end diameter

10.8 mm

Flexible portion diameter 9.8 mm

Flexible portion diameter 9.3 mm


Bending capability

An endoscope specially designed to make the best use of laser lights

UP: 210 / DOWN: 90

Bending capability

RIGHT: 100 / LEFT: 100

RIGHT: 100/ LEFT: 100

Light guide

210UP
90DOWN

Working length

1,100 mm

Working length

1,100 mm

Total length

1,400 mm

Total length

1,400 mm

Minimum diameter of
forceps channel

2.8 mm

Minimum diameter of
Forcept channel

2.8 mm

Light guide

Image area &


Forcept entry position

Forceps channel

210UP
90DOWN

Objective lens

Air/water nozzle

UP: 210 / DOWN: 90

Objective lens
Forceps channel

100LEFT
100RIGHT

Image area &


Forcept entry position

100LEFT
100RIGHT

Air/water nozzle

Colonoscopes for use with the laser light source


For the Lower G.I. Tract-Standard Type

For the Lower G.I. Tract-Optical magnification

EC-L590WM

EC-L590ZW

This endoscope is designed to support identification of lesions by


using FICE and BLI images. It has the large distal end of 12.0 mm in
diameter, 140-degree field of view, 3.8-mm large forceps channel and
water jetting function.

This endoscope offering close-up BLI images is suitable for observation


of superficial microvessels and mucosal surface structures.
The maximum optical magnification is approximately 135 times (on
19-inch LCD monitors).

Viewing direction

0 (Forward)

Viewing direction

0 (Forward)

Field of view

140

Field of view

WD: 140/TL: 55

Observation range

4-100 mm

Observation range

WD: 6-100 mm/TL:2-3 mm

Distal end diameter

12.0 mm

Distal end diameter

12.8 mm

Flexible portion diameter 12.8 mm

Flexible portion diameter 12.0 mm


Bending capability

UP: 180 / DOWN: 180

Bending capability

RIGHT: 160 / LEFT: 160

Objective lens
Light guide

180UP

RIGHT: 160 / LEFT: 160

Working length

1,330 mm

Working length

1,330 mm

Total length

1,630 mm

Total length

1,630 mm

Minimum diameter of
forceps channel

3.8 mm

Minimum diameter of
forceps channel

3.8 mm

Remark

Water jet function

Remark

Water jet function

Water jet nozzle

180DOWN

Image area &


Forcept entry position

180DOWN
Objective lens
Air/water nozzle

160LEFT
Water jet nozzle

Entire endoscope

180UP
Light guide

Air/water nozzle

Forceps channel

UP: 180 / DOWN: 180

160RIGHT

Forceps channel

Control portion

Image area &


Forcept entry position

160LEFT
160RIGHT

Laser light source connecting section

EXAMPLE OF SYSTEM CONFIGURATION

EXAMPLE OF SYSTEM CONFIGURATION

LASEREO System

EPX-4450HDLASEREO System

L590-series scopes

EG - L590WR

L590-series scope

EG - L590ZW

EC-L590WM

400-series scope

500,600-series scope

EC- L590ZW

Light Source XL-4450


19-inch
LCD monitor

19-inch
LCD monitor

24-inch
LCD monitor

24-inch
LCD monitor

Processor VP-4450HD

Processor VP-4450HD

Light Source LL-4450


Light Source LL-4450
System cart
System cart

Output devices

Output devices
Printer

HD recorder

Printer

HD recorder

Specifications
Processor VP-4450HD
Digital output

Analog output
Color adjustment
Detail
Contrast
Hyper-sharpness
Color emphasis
FICE
Iris
Image storage
Power rating
Dimensions(W H D)
Weight
DICOM

Light source LL-4450

HD-SDI: HDTV 1080 i(2 ch)


DVI(Digital Visual Interface): 1280 1024 p
Ethernet: 100/10 Base
RGB:1280 1024 p
SDTV(120 V/NTSC, 230 V/PAL): RGB, Y/C, composite
Brightness, Red, Green, Blue, R-Hue, Chroma; 9 steps
Hi, Lo; 9 steps
3 steps
High, Middle, Low, Off
Color Enhancement 10 presets
Flexible spectral Imaging
Color Enhancement 10 presets
Average/Peak/Auto
CF card
AC 120 V 10 , 60 Hz, 0.8 A
AC 230 V 10 , 50 Hz, 0.5 A
390 105 460 mm
9.5 kg
MWL, Store

Laser light 1

Light source XL-4450

Light output wave length: 450 10nm

Lamp rated value

Maximum laser output: 3 W(CW)

Main lamp300 W xenon lamp LMP-002


Emergency lamp75 W halogen lamp

Light output wave length: 410 10nm

Light control

Automatic light control

Maximum laser output: 1.5 W(CW)

Lamp cooling method

Forced air cooling

Light control

Automatic light control

Air supply pump

High, Middle, Low, Off

Light cooling method

Forced air cooling

Light save

On, Off

Air supply pump

High, Middle, Low, Off

Transmitted illumination

On, Off

Light control

Automatic light control by the control signal from the prosessor

Power rating

AC 120 V 10 , 60 Hz, 3.3 A

Tranmitted illumination

On, Off

Power rating

AC 230 V 10 , 50 Hz, 1.2 A

Dimensions(W H D)

390 155 450 mm

Dimensions(W H D)

385 125 505 mm(Maximum)

Weight

15 kg

Weight

12 kg

Laser light 2

AC 230 V 10 , 50 Hz, 1.7 A

Product nameLight source


GMDN35158
Generic nameEndoscopic light sources

Product nameLight source


GMDN35158
Generic nameEndoscopic light source

Product nameProcessor
GMDN18034
Generic nameProcessing unit

10

11

Laser light opens new frontiers in endoscopy

New-generation Endoscope System

Ref. No. XB-000E

(SK1402F1079F9711) Printed in Japan 2014 FUJIFILM Corporation

You might also like