D6414. Mercury
D6414. Mercury
D6414. Mercury
1. Scope
1.1 These test methods cover procedures to determine the
total mercury content in a sample of coal or coal combustion
residue.
1.2 The values stated in SI units are regarded as the
standard.
1.3 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D 121 Terminology of Coal and Coke2
D 1193 Specification for Reagent Water3
D 2013 Method of Preparing Coal Samples for Analyses2
D 3173 Test Method for Moisture in the Analysis Sample of
Coal and Coke2
D 5142 Test Methods for the Proximate Analysis of the
Analysis Sample of Coal and Coke by Instrumental
Procedures2
D 3180 Practice for Calculating Coal and Coke Analyses
from As-Determined to Different Bases2
IEEE/ASTM SI 10 Standard for Use of the International
System of Units (SI): The Modern Metric System4
6. Apparatus
6.1 Apparatus for Test Method A:
6.1.1 Analytical Balance, with a sensitivity of 0.1 mg.
6.1.2 Atomic Absorption Spectrophotometer, with a flameless cold-vapor mercury analysis system.
6.1.3 Digestion Vessels, 100- to 250-mL bottles with an
O-ring seal and screw cap. Bottle must be compatible for use
with aqua regia. Polycarbonate and HDPE are acceptable.
Bottles and cap assemblies shall be washed in 1-to-1 HCl then
dried before each use.
3. Terminology
3.1 For definitions of terms used in this standard, refer to
Terminology D 121.
NOTE 1Other bottle and cap assemblies may be used provided they
are compatible for use with aqua regia at a temperature of 80C.
D 6414
6.2.4 Heat Source, an aluminum block with 18-mm holes to
accommodate the disposable test tubes. The block shall be
capable of slowly reaching and maintaining a final temperature
of 150C. The block can be heated by placing it on a hot plate
or it can contain its own internal heating elements.
7. Sample
7.1 Prepare the analysis sample in accordance with Method
D 2013 by pulverizing the material to pass a 250-m (No. 60)
sieve.
7.2 Analyze separate test portions for moisture content in
accordance with Test Method D 3173 or Test Methods D 5142.
9. Procedure
9.1 Preparation of Test Solution A (Extraction Step):
9.1.1 Weigh a test portion of approximately 1 g of the
sample into a digestion bottle. Record the weight (Ws) to the
nearest 0.0001 g.
9.1.2 Quantitatively add 2 mL of concentrated nitric acid
and 6 mL of concentrated hydrochloric acid to the digestion
bottle and secure the cap.
9.1.3 Transfer the digestion bottle and contents to a water
bath that has been heated to 80C and heat for 1 h. Secure the
digestion bottle in such a way as to keep the contents below the
surface of the water.
9.1.4 After 1 h, remove the digestion bottle and allow to
cool to room temperature.
(1)
where:
Wcrm
5 dry weight of the CRM coal used for preparation of the quality control sample, g;
CRM
5 certified value of mercury in the quality control sample;
Vstandard 5 volume of mercury standard solution added to
the digestion bottle, mL; and
0.1
5 the concentration of the mercury standard
solution, g/mL.
6
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia and
National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, MD.
NOTE 3A test portion of 0.9-g dry weight of a CRM coal with 0.11
g/g (110 ng/g) of mercury is weighed out as a quality control sample. To
D 6414
10.2 Purity of AcidsUse trace metal purity grade acids or
equivalent. Redistilled acids are acceptable.
10.3 Purity of WaterUse water equivalent to ASTM Type
II reagent water of Specification D 1193.
10.4 Mercury Standard Stock Solution [1000 ppm (1000
g/mL)]Dissolve 1.0800 g of mercury (II) oxide (HgO) in a
minimum volume of HCl (1+1) and dilute to 1 L with water.
Alternatively, use a commercially available stock solution
specifically prepared for atomic absorption spectrometry.
10.5 Mercury Standard Solution (100 ng/mL)Prepare the
mercury standard solution fresh daily. Dilute 5 mL of the
mercury standard stock solution to 500 mL with deionized
water. Further dilute 10 mL of this intermediate solution to 1 L
with deionized water.
10.6 Nitric AcidConcentrated (HNO3, sp. gr. 1.42).
10.7 Hydrochloric AcidConcentrated (HCl, sp. gr. 1.19).
10.8 Sulfuric AcidH2SO4, sp. gr. 1.83).
10.9 Vanadium Pentoxide, V2O5Remove traces of mercury by roasting the V2O5 in a fume hood at a temperature
below 690C, the melting point of V2O5, in a porcelain dish
using a muffle furnace or a Fisher burner.
11. Procedure
11.1 Preparation of Test Solution B:
11.1.1 Weigh a test portion of 0.15 g of the test sample into
a 16- by 150-mm disposable test tube. Record the weight (Ws)
to the nearest 0.0001 g.
11.1.2 Add approximately 0.1 g of V2O5, 3.5 mL of concentrated HNO3, and 1.5 mL of concentrated H2SO4. Swirl to
mix the contents. Place the test tube in the aluminum heating
block and cover with a watch glass.
11.1.3 Adjust the temperature of the heating block so that it
gradually reaches a temperature of 150C in 4 h. Maintain
heating at this temperature for 16 h.
11.1.4 After heating, carefully remove the test tube from the
aluminum heating block, allow to cool, and dilute the solution
to a final volume of 15 mL (V) with water. Securely cap the test
tube and mix the contents. Centrifuge the test tube and its
D 6414
contents for 5 min at 1000 rpm.
11.2 Reagent BlankPrepare a reagent blank by repeating
the procedure in 11.1 but without the test portion of the sample.
11.3 Quality Control Sample:
11.3.1 Prepare a CRM coal for analysis using the procedure
described in 11.1. Record the certified value as CRM.
11.3.2 Alternatively, weigh a test portion of 0.15 g of a
CRM. After the addition of the nitric acid, add mercury
standard solution (10.5) to the test tube. The volume of the
mercury standard solution to be added should yield an amount
of mercury approximately equivalent to that in the CRM coal
(Note 7). Calculate the expected value CRME of mercury
using Eq 1.
NOTE 7A test portion of 0.15 g of a CRM coal with 0.11 g/g (110
ng/g) of mercury is weighed out as a quality control sample. To yield an
amount approximately equivalent to that present in the CRM coal, 0.15
mL of the mercury standard solution is added to the test tube after the
addition of the nitric acid. CRME calculates to 0.21 g/g.
12. Calculation
12.1 These calculations apply to both Test Method A and
Test Method B.
12.1.1 Calculate the dry weight of each test portion of
sample prepared for analysis as follows
(2)
where:
Wds 5 dry sample weight, g;
Ws 5 weight of test portion of sample; and
Ma 5 moisture in the analysis sample calculated according
to Test Method D 3173 or Test Methods D 5142.
12.2 Quality Control:
12.2.1 Calculation of Mercury Concentration of the Quality
Control Sample:
12.2.1.1 Absorbance matchingCalculate the concentration of mercury in ppm (g/g) in the quality control sample as
Mercury ppm ~g/g! dry basis 5 DF 3 ~V/Vds! 3 C
3 @~A Ab!/~Ac Ab!#
(3)
where:
A
5 absorbance of the quality control (Aqs) or analysis
sample (A);
Ac 5 absorbance of the calibration standard nearest A;
Ab 5 absorbance of the blank;
C
5 concentration of the calibration standard, ng/mL;
DF 5 dilution factor, if any;
V
5 volume of the analysis solution; and
Wds 5 dry sample weight, g.
12.2.1.2 Calibration curve:
12.2.1.2.1 Prepare a calibration curve by plotting absorbance of calibration standard minus blank response versus the
concentration of the calibration solution in ng/mL. For the
number of standards specified in this test method, it is
recommended that a linear fit of the form y 5 mx + b be used.
Other fits can be used but can require the preparation of
additional standards to ensure adequate degrees of freedom are
maintained (Note 9).
D 6414
the limits of MLR and MLR+, prepare another portion of the
standard upon which the original mercury recovery was based.
Determine the absorbance of the new calibration standard.
If the mercury recovery calculated from the absorbance of
the new standard is within the MLR and MLR+ limits, then
proceed with calculation of the concentration of mercury in the
analysis samples.
12.2.2.4.2 If a calibration curve was used to calculate the
concentration of the quality control sample, examine the curve
for points that deviate significantly from the curve. Recalculate
the regression eliminating the suspect point(s) (Note 9). If the
value for the CRM falls within the MLR and MLR+ limits
based on the recalculated regression, then prepare another
portion of the standard(s) eliminated from the regression.
Determine the absorbance of the new standard(s). Recalculate
the regression incorporating the new standards. If the calculated mercury recovery of for CRME is within the MLR and
MLR+ limits, then proceed with calculation of the concentration of mercury in the analysis samples.
12.2.2.4.3 Verify the concentration of the mercury standard
solution from which the calibration standards have been
prepared. Prepare a new mercury standard solution from the
mercury standard stock solution. From the new mercury
standard solution, prepare a 2-ng/mL mercury solution and a
4-ng/mL mercury solution. Determine the absorbance of the
two solutions and of the quality control sample. Recalculate
the mercury recovery. If the recalculated values both fall within
the MLR and MLR+ limits, then the original mercury standard
solution is suspect. Prepare fresh calibration solutions from the
new mercury standard solution and reanalyze a separate aliquot
of the test solutions.
12.2.2.5 If it is not possible to reconcile the mercury
recovery within the limits of MLR and MLR+, then discard all
solutions and repeat the analysis of all samples.
12.3 Calculations for Test SamplesProvided the conditions in 12.2 are satisfied, calculate the mercury content of the
test samples using the same equation (Eq 3, or Eq 4, Eq 5) as
for the quality control sample.
(4)
where:
A 5 absorbance of the quality control (Aqs) or analysis
sample (A) and
Ab 5 absorbance of the blank.
12.2.1.2.3 Read the concentration of the quality control
sample from the calibration curve as Cs.
12.2.1.2.4 Calculate the concentration of mercury in ppm
(g/g) in the quality control sample as follows
Mercury ppm ~g/g! dry basis 5 DF 3 ~V/Wds! 3 Cs
(5)
where:
Cs 5 concentration of the quality control sample read
from the calibration curve, ng/mL;
DF 5 is the dilution factor, if any;
V
5 volume of the analysis solution; and
Wds 5 dry sample weight, g.
12.2.2 Calculation of Mercury Recovery in the Quality
Control Sample:
12.2.2.1 Calculate the CRM confidence factor as follows
CRMF 5 100 * ~mercury limits/mercury value!
(6)
where:
mercury limits 5 the confidence limits from the certificate
issued by the certifying agency and
mercury value 5 the mercury value issued by the certifying
agency.
If CRMF is greater than 30 %, then the mercury recovery
limits (MLR) are:
MLR+ 5 100 + CRMF
MLR 5 100 CRMF.
If CRMF is less than or equal to 30 %, then the MLR are:
MLR+ 5 130
MLR 5 70.
12.2.2.2 Calculate the mercury recovery from the quality
control sample as follows
Mercury recovery 5 ~mercury ppm ~g/g!/CRME! 3 100
13. Report
13.1 Report the results of the mercury analysis on a dry
sample basis.
13.2 Procedures for converting the value obtained on the
analysis sample to other basis are described in Practice D 3180.
13.3 Report the recovery of mercury in the quality control
sample.
13.4 Report the digestion method used.
(7)
where:
mercury ppm (g/g) 5 the mercury value of sample calculated according to Eq 3 or Eq 5 in
g/g and
CRME
5 the expected value of mercury in the
quality control sample in g/g.
12.2.2.3 If the mercury recovery is between MLR and
MLR+, then proceed with calculation of the concentration of
mercury in the analysis samples.
12.2.2.4 If the mercury recovery is not between MLR and
MLR+, then check one or all of the following:
12.2.2.4.1 If absorbance matching was used to calculate the
concentration of the quality control sample, recalculate the
mercury recovery using two other calibration standards. If the
mercury recovery calculated using both standards falls within
D 6414
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.