Astm E202
Astm E202
Astm E202
1. Scope
1.1 These test methods cover the chemical and physical
analysis of the commonly available grades of ethylene glycol,
diethylene glycol, triethylene glycol, propylene glycol, and
dipropylene glycol. The key sections appear in the following
order:
Sections
Purity of Reagents
Specific Gravity
Distillation Range
Acidity
Water
Iron
Color
Gas Chromatographic Analysis
4
5-7
8-10
11-13
14-16
17-25
26-28
29-39
4. Purity of Reagents
4.1 Purity of ReagentsReagent grade chemicals shall be
used in all tests. Unless otherwise indicated, it is intended that
all reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society,
where such specifications are available.6 Other grades may be
used, provided it is first ascertained that the reagent is of
sufficiently high purity to permit its use without lessening the
accuracy of the determination.
4.2 Unless otherwise indicated, references to water shall be
understood to mean reagent water conforming to Specification
D 1193, Type II or III.
2. Referenced Documents
2.1 ASTM Standards:
D 891 Test Methods for Specific Gravity, Apparent, of
Liquid Industrial Chemicals2
D 1078 Test Method for Distillation Range of Volatile
Organic Liquids3
D 1193 Specification for Reagent Water4
D 1209 Test Method for Color of Clear Liquids (PlatinumCobalt Scale)3
D 1613 Test Method for Acidity in Volatile Solvents and
Chemical Intermediates Used in Paint, Varnish, Lacquer,
and Related Products3
E 60 Practice for Photometric and Spectrophotometric
SPECIFIC GRAVITY
5. Procedure
5.1 Determine the specific gravity of the sample at 20/20C
1
These test methods are under the jurisdiction of ASTM Committee E15 on
Industrial and Specialty Chemicals and are the direct responsibility of Subcommittee E15.01 on General Standards.
Current edition approved October 10, 2000. Published November 2000. Originally published as E 202 62. Last previous edition E 202 94a.
2
Annual Book of ASTM Standards, Vol 15.05.
3
Annual Book of ASTM Standards, Vol 06.04.
4
Annual Book of ASTM Standards, Vol 11.01.
Copyright ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
E 202
TABLE 1 Thermometers for Distillation Range
Glycol
Ethylene glycol
Diethylene glycol
Triethylene glycol
Propylene glycol
Dipropylene glycol
Thermometer
Number
104C
106C
107C
104C
106C
this case, up to 30 min can be allowed to collect the first drop, and a
distillation rate of 2 to 3 mL/min is satisfactory. Alternatively, the flask
chamber may be covered with a suitable shield so that only the upper neck
and thermometer are exposed to room air to achieve the specified rates.
Thermometer
Range
173
223
248
173
223
to
to
to
to
to
227C
277C
302C
227C
277C
Chemical
Ethylene glycol
Diethylene glycol
Triethylene glycol
Propylene glycol
Dipropylene glycol
0.045
0.050
0.055
0.043
0.051
9. Report
9.1 Report the corrected temperatures to the nearest 0.1C at
each volume required by the specification for the glycol being
analyzed.
6. Report
6.1 Report the specific gravity at 20/20C (in air) to the
nearest 0.0001 unit.
7.2 BiasThe bias of this test method has not been determined due to the unavailability of suitable reference materials.
DISTILLATION RANGE
8. Procedure
8.1 Determine the distillation range of the sample in accordance with Test Method D 1078. Use the conditions as specified in Test Method D 1078, and the ASTM Solvents Distillation Thermometer shown in Table 1 . (See Note 2 for certain
allowable exceptions in applying this test method to triethylene
glycol.)
ACIDITY
11. Procedure
11.1 Determine the acidity of the sample in accordance with
Test Method D 1613.
12. Report
12.1 Report the acidity, expressed as weight percent of
acetic acid, to the nearest 0.0001 %.
E 202
TABLE 2 Distillation Range Precision Values
0 to 5C/qa
5 to 15C
0.154 (54A)
0.4
0.154 (54)
0.4
0.218 (52)
0.6
0.148 (56)
0.4
0.173
0.5
(27)
0.173 (27)
0.5
0.119 (23)
0.4
0.0901(28)
0.2
0.414
1.2
(8)
0.414
1.2
0.489
1.4
0.190
0.5
0.118
0.3
(54)
0.118 (54)
0.3
...
...
...
...
0.147
0.4
(27)
0.147 (27)
0.4
...
...
...
...
WATER
0.317
0.9
(8)
0.317
0.9
...
...
...
...
Ethylene Glycol
Distillation Range
Diethylene Propylene
and
and
Triethylene Dipropylene
Glycols
Glycols
(8)
(8)
(6)
(6)
0.0783 (54)
0.2
0.0783(54)
0.2
0.129 (52)
0.4
0.0892(56)
0.2
0.0981 (27)
0.3
0.0981(27)
0.3
0.0961(26)
0.3
0.0505(28)
0.1
0.279
0.8
0.279
0.8
0.201
0.6
0.133
0.4
(8)
(8)
(6)
0.0837 (54)
0.2
0.0837(54)
0.2
...
...
...
...
0.126
0.4
(27)
0.126 (27)
0.4
...
...
...
...
0.336
0.9
(8)
0.336
0.9
...
...
...
...
(8)
0.384 (36)
1.1
0.272 (46)
0.8
0.193 (56)
0.5
0.122
0.3
(8)
0.640 (18)
1.8
0.103 (23)
0.3
0.250 (28)
0.7
0.466
1.3
(7)
2.77
7.8
1.12
3.1
0.896
2.5
(5)
14. Procedure
14.1 Determine the water content of the sample using any
suitable Karl Fischer reagent titration method. Test Method
E 203 is recommended.
15. Report
15.1 Report the water content to the nearest 0.01 %.
16. Precision and Bias
16.1 The following criteria should be used for judging the
acceptability of results (see Note 5):
16.1.1 Repeatability (Single Analyst)The standard deviation for a single determination has been estimated to be
0.000370 % absolute at 92 dF. The 95 % limit for the difference
between two such runs is 0.01 % absolute.
16.1.2 Laboratory Precision (Within-Laboratory, BetweenDays)The standard deviation of results (each the average of
duplicates), obtained by the same analyst on different days, has
been estimated to be 0.00819 % absolute at 58 df. The 95 %
limit for the difference between two such averages is 0.02 %
absolute.
16.1.3 Reproducibility (Multilaboratory)The standard deviation of results (each the average of duplicates), obtained by
analysts in different laboratories, has been estimated to be
0.0187 % absolute at 9 df. The 95 % limit for the difference
between two such averages is 0.05 % absolute.
(6)
0.0779 (14)
0.2
(8)
(6)
13.1.1 Repeatability (Single Analyst)The standard deviation for a single determination has been estimated to be
0.0000918 % absolute at 92 dF. The 95 % limit for the
difference between two such runs is 0.0003 % absolute.
13.1.2 Laboratory Precision (Within-Laboratory, BetweenDays)The standard deviation of results (each the average of
E 202
solution. This curve must be determined for each instrument
and should be checked periodically.
IRON
17. Summary of Test Method
17.1 The sample is diluted with water and the iron determined photometrically at approximately 510 nm by the orthophenanthroline method without a preliminary ashing. The
amounts and the concentrations of hydroxylamine hydrochloride and ortho-phenanthroline solutions added has been increased over that given in conventional write-ups in order to
overcome the depression of the ferrous-ortho-phenanthroline
complex color by the large amount of glycol present.
22. Procedure
22.1 For samples containing 0 to 0.5 ppm Fe, weigh about
80 g of sample to the nearest 0.1 g into a 100-mL glassstoppered graduate. For samples of larger iron content, use
proportionally smaller samples. Dilute with water to 80 mL.
Add 80 mL of water to another graduate as a blank.
22.2 Add 5 mL of NH2OHHCl solution to each graduate
and mix. Make up to the 100-mL mark with orthophenanthroline solution and again mix.
22.3 Allow to stand for 5 min, fill a 5-cm absorption cell,
and measure the absorbance at approximately 510 nm, corrected for the blank (Note 8). From the calibration curve, read
the milligrams of iron present.
NOTE 8Although the ferrous-ortho-phenanthroline complex is reported to have a maximum absorbance at approximately 510 nm, the
measurements on both the standard iron and sample solutions should be
made at or near the wavelength at which maximum absorbance is obtained
with each individual instrument, as indicated by scanning in this region of
the spectrum. This step will ensure that the measurements will be made
with a maximum of sensitivity and will reduce wavelength calibration
errors.
19. Apparatus
19.1 Spectrophotometer, capable of measuring light absorption at approximately 510 nm.
NOTE 6A discussion of photometers and photometric practice is
given in Practice E 60.
23. Calculation
23.1 Calculate the iron present, in parts per million, as
follows:
Iron, ppm 5
SD
W
S 3 1000
(1)
where:
W = iron found, mg, and
S = sample used, g.
24. Report
24.1 Report the iron content of the sample to the nearest
0.01 ppm.
25. Precision and Bias
25.1 The following criteria should be used for judging the
acceptability of results (see Note 9):
25.1.1 Repeatability (Single Analyst)The standard deviation for a single determination has been estimated to be
0.00744 ppm at 92 dF. The 95 % limit for the difference
between two such runs is 0.02 % ppm.
25.1.2 Laboratory Precision (Within-Laboratory, BetweenDays)The standard deviation of results (each the average of
duplicates), obtained by the same analyst on different days, has
been estimated to be 0.00971 at 45 dF. The 95 % limit for the
difference between two such averages is 0.03 % ppm.
25.1.3 Reproducibility (Multilaboratory)The standard deviation of results (each the average of duplicates), obtained by
analysts in different laboratories, has been estimated to be
0.0255 ppm absolute at 8 df. The 95 % limit for the difference
between two such averages is 0.07 ppm.
E 202
more than 1 %, each. Up to 1 % tetraethylene glycol in
triethylene glycol may be analyzed by this test method. The
isomers of dipropylene and tripropylene glycol are not completely resolved under the conditions used. Gas chromatographic test methods for determining less than 0.1 % diethylene glycol in ethylene glycol are in accordance with Test
Methods E 611.
26. Procedure
26.1 Determine the color of the sample in accordance with
Test Method D 1209.
27. Report
27.1 Estimate and report the color to the nearest one
platinum-cobalt unit.
28. Precision and Bias
28.1 The following criteria should be used for judging the
acceptability of results (see Note 10):
28.1.1 Repeatability (Single Analyst)The standard deviation for a single determination has been estimated to be 0.0 unit
at 40 dF. The 95 % limit for the difference between two such
runs is two units.
28.1.2 Laboratory Precision (Within-Laboratory, BetweenDays)The standard deviation of results (each the average of
duplicates), obtained by the same analyst on different days, has
been estimated to be 0.64 unit at 46 dF. The 95 % limit for the
difference between two such averages is two units.
28.1.3 Reproducibility (Multilaboratory)The standard deviation of results (each the average of duplicates), obtained by
analysts in different laboratories, has been estimated to be 2.47
units at 9 df. The 95 % limit for the difference between two
such averages is seven units.
These parameters are summarized in Table 3 as typical values. See Note 13.
Hamilton Microsyringe No. 705 has been found satisfactory and is available
from supply houses.
9
E 202
calibration factors may be obtained using standards prepared
fromhearts cuts from the distillation of each of the glycols, or
from commercial grades of each glycol as described in the
following test methods. For highest accuracy, glycols obtained
from hearts cuts should be used. The calibration factors
should be checked periodically or whenever there is evidence
of a change in the column or instrument.
35.2 Test Method A:
35.2.1 Purify the commercial grade of each glycol needed
by careful fractional distillation in glass at reduced pressure,
discarding the first 30 % and retaining the next 30 % as the
hearts cuts. These fractions should be analyzed in accordance with 36.2 or 36.3 to be sure they are free from other
homologues of the glycol.
35.2.2 Prepare a standard mixture of these glycols whose
composition approximates that of the glycols to be analyzed.
The composition of this standard should be known to the
nearest 0.01 %. Correct the composition for any water present
as determined in accordance with 14.1, using the equation in
37.2.4.
35.2.3 Obtain at least two chromatograms of the standard
mixture in accordance with 36.2 or 36.3 and calculate the
average area percent for each of the glycols present in
accordance with 37.2.1 (Note 12). Do not include any areas
associated with air and water in calculating the area percentages. Using the weight percent in the standard mixture and the
average area percent, calculate the factor for each glycol in
accordance with 37.1.1.
E 202
TABLE 4 Retention Time Data
36. Procedure
36.1 In analyzing the sample, either of two modes may be
used in operating the gas chromatograph: linear programmed
temperature or isothermal. Except for the column temperatures,
the procedure is the same for either mode, but other parameters
will change depending on whether ethylene or propylene
glycols are being analyzed. The procedure using linear programmed temperature will be described first.
36.2 Linear Programmed Temperature Operation:
36.2.1 Mount the column in the chromatograph, and adjust
the operating conditions in accordance with the parameters
given in Table 3 (Note 13). Allow sufficient time for the
instrument to reach equilibrium as indicated by a stable base
line on the chart at the maximum sensitivity setting to be used.
Compound
Programmed
Temperature
Operation
Isothermal
Operation at
190C
0.5
4.0
8.1
9.3
10.3
12.4
0.3
1.3
3.3
3.9
4.8
7.9
Tripropylene glycolB
36.2.4 Draw base lines under each glycol peak and measure
the area of each peak with a planimeter, unless the recorder is
equipped with an integrator.
36.3 Isothermal Operation:
36.3.1 Follow the procedure in accordance with 36.2, except use a constant column temperature of 170 C for the
analysis of mixtures of ethylene, diethylene, and triethylene
glycols. A temperature of 200C is used for mixtures containing tetraethylene glycol (Note 13 and Note 17). Approximate
retention times for the glycols are given in Table 4 and Table
5. Typical chromatograms for a sample of ethylene and of
diethylene glycol at temperatures of 170 and 200C respectively, are shown in Fig. 3 and Fig. 4. A typical chromatogram
of dipropylene glycol is shown in Fig. 5.
0.3
0.8
2.3
5.8
16.3
Carrier gas
Detector current
Injection port temperature
200C
0.3
1.3
4.2
14.0
...
NOTE 14Direct on-column injection of the sample has been reported to result in better-shaped peaks and to eliminate, or greatly reduce,
the buildup of carbon in the injection port. If this method of sample
injection is used, the temperature of the injection port should be the same
as that of the column.
Column temperature
(a) Programmed
temperature
operation
(b) Isothermal operation
170C
0.4
2.2
7.0
11.9
19.3
Isothermal Operation
Strip-chart recorder
Chart speed
Column
Programmed
Temperature
Operation
A
Retention times vary with component concentration and from instrument to
instrument. Thus, the times listed, measured from the point of sample injection to
the peak maximum, are approximate.
NOTE 13The instrument parameters given in Table 3 may be considered to be typical values. For any specific instrument some adjustment of
column temperature, programming rate, helium flow rate, etc. will
probably be required to achieve retention times similar to those in Table
4 and Table 5. The parameters should be adjusted so that the peaks
obtained are reasonably symmetrical, sharp, and exhibit satisfactory
resolution.
Instrument
37. Calculation
37.1 Calibration Factors:
37.1.1 When hearts cuts of glycols are used to prepare the
standard mixture, obtain the calibration factor for each glycol
as follows:
E 202
FIG. 1 Chromatogram of Diethylene Glycol Linear Programmed Temperature Operation (Recorder Attenuation in Parentheses)
FIG. 2 Chromatogram of Dipropylene Glycol Linear Programmed Temperature Operation (Recorder Attenuation in Parentheses)
Wi
Fi 5 A
Bi
Fi 5 A 2 A
si
bi
(2)
%i
where:
= calibration factor for component i,
Fi
= weight percent of component i in standard mixture,
Wi
and
A%i = average area percent of component i in standard
mixture.
37.1.2 When commercial grades of glycols are used to
prepare the standard mixture, obtain the calibration factor for
each glycol present in minor concentration as follows:
(3)
where:
Fi = calibration factor for component i,
Bi = weight percent of minor component i added to
principal glycol in preparing the standard mixture,
Asi = average area percent of minor component i in standard mixture, and
Abi = average area percent of minor component i in principal component.
37.2 Sample Composition:
8
E 202
FIG. 3 Chromatogram of Ethylene Glycol Isothermal Operation at 170C (Recorder Attenuation in Parentheses)
FIG. 4 Chromatogram of Diethylene Glycol Isothermal Operation at 200C (Recorder Attenuation in Parentheses)
A%i 5
AiTi 3 100
~A 1T1! 1 ~A2T2! 1 ~A 3T3! 1 ~A4T4!
(4)
E 202
FIG. 5 Chromatogram of Dipropylene Glycol Isothermal Operation at 190C (Recorder Attenuation in Parentheses)
Gi
where:
A% i
Ai
Ti
A1, A2, A3, A4
=
=
=
=
Ci
D
38. Report
38.1 Report the weight percent of each component to the
nearest 0.01 %.
39. Precision and Bias
39.1 The following criteria should be used for judging the
acceptability of results (see Note 18):
39.1.1 Repeatability (Single Analyst)The standard deviation for a single determination has been estimated to be the
value in Table 6 at the indicated degrees of freedom. The 95 %
limit for the difference between two such runs is the value in
the table.
39.1.2 Laboratory Precision (Within-Laboratory, BetweenDays)The standard deviation of results (each the average of
duplicates), obtained by the same analyst on different days, has
been estimated to be the value in Table 6 at the indicated
degrees of freedom. The 95 % limit for the difference between
two such averages is the value in the table.
39.1.3 Reproducibility (Multilaboratory)The standard deviation of results (each the average of duplicates), obtained by
analysts in different laboratories, has been estimated to be the
value in Table 6 at the indicated degrees of freedom. The 95 %
limit for the difference between two such averages is the value
in the table.
(5)
where:
Ac i = corrected area for glycol i,
A%i = area percent for glycol i, and
= factor for glycol i.
Fi
37.2.3 Calculate the weight percent of each glycol on an
anhydrous basis as follows:
Aci
C i 5 A 1 A 1 A 1 A 3 100
c1
c2
c3
c4
where:
Ci
Aci
Ac1, Ac2, Ac3, Ac4
(6)
Ci~100 2 D!
100
NOTE 18The precision estimates for the ethylene glycols are based on
a 19641965 interlaboratory study in which three samples of the glycols
were analyzed. The concentrations of the minor glycol components ranged
from 0.48 to 1.00 weight %. Seven laboratories cooperated in the study in
(7)
where:
10
E 202
TABLE 6 Gas Chromatographic Precision Values
Ethylene Glycols
Standard Deviation,
absolute %
Propylene Glycols
Major
ComponentA
Minor
ComponentB
Major
ComponentA
Minor
ComponentB
0.0477
(42C)
0.13
0.0272
(12)
0.076
0.0224
(64)
0.063
0.0172
(92)
0.048
0.0700
(21)
0.20
0.0363
(56)
0.10
0.0366
(32)
0.10
0.0260
(46)
0.073
0.1024 (6)
0.0527 (6)
0.0615 (7)
0.0499 (7)
0.29
0.15
0.17
0.14
0.0516
(42)
0.14
0.0277
(112)
0.078
0.0265
(70)
0.074
0.0197
(94)
0.055
0.0498
(21)
0.14
0.0282
(56)
0.079
0.0515
(35)
0.14
0.0494
(47)
0.14
0.1033 (6)
0.0528 (6)
0.0633 (8)
0.0536 (7)
0.29
0.15
0.18
0.15
Programmed
Temperature, Weight %:
Repeatability
Standard
deviation
95 % limit
Laboratory
Precision
(WithinLaboratory
BetweenDays)
Standard
deviation
95 % limit
Reproducibility
Standard
deviation
95 % limit
Isothermal
Temperature,
Weight %
Repeatability
Standard
deviation
95 % limit
Laboratory
Precision
(WithinLaboratory
BetweenDays)
Standard
deviation
95 % limit
Reproducibility
Standard
deviation
95 % limit
A
12
Details of the interlaboratory study are available from ASTM Headquarters.
Request RR: E15-0028.
11
E 202
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at
610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website (www.astm.org).
12