0% found this document useful (0 votes)
284 views2 pages

Direct-Pure Edi Lab Pure Water System

The document describes a water purification system that uses electrodeionization (EDI) to produce Type II water directly from tap water. The system incorporates RO and EDI modules to consistently deliver high quality purified water at a relatively low cost. It provides fully automated control and monitoring to ensure the water meets Type II specifications.

Uploaded by

Elida Rondić
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
284 views2 pages

Direct-Pure Edi Lab Pure Water System

The document describes a water purification system that uses electrodeionization (EDI) to produce Type II water directly from tap water. The system incorporates RO and EDI modules to consistently deliver high quality purified water at a relatively low cost. It provides fully automated control and monitoring to ensure the water meets Type II specifications.

Uploaded by

Elida Rondić
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Direct-Pure

EDI

Produce Type II water from tap water directly

Resistivity of the product water is above 5 Mcm at 25 , which meets or exceeds Type II water quality as defined by
ASTM, CAP, CLSI and ISO 3696 / BS 3997 and also complies with the Purified Water requirements from the European and
U.S. Pharmacopoeia. The system is CE certified.

Features
The system incorporates the best in class Siemens EDI module and stringently selected RO membrane, to deliver high quality
Type II water consistently at a relatively low cost.
The system is equipped with a safeguard cartridge to provide complete protection of RO membrane and EDI module.
Automatic water quality control programs ensure EDI product water meeting or exceeding Type II water requirements.
The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane life monitoring as
well as many other programs for convenient operation.
The pre-rinsing program starts automatically once a new RO membrane is installed so there is no need for users to set up
additional cleaning procedures.
User can set water dispensing time for easy water collection.
EDI product water can be dispensed from the water system directly.
The system can be fit with a 30 or 60-liter PE tank with a liquid level sensor and a tank vent filter
with CO2 scavenger to preserve EDI water from the air contamination.
RephiLe can provide full document support to meet user's GMP, GLP, FDA
and other certification requirements.

Main Applications
Type II pure water can be used in many areas. Some of the typical applications are:
Preparation of chemical and bio-reagents
Preparation of culture media
Preparation of solutions for chemical analysis such as HPLC and ICP
For clinical analyzers
Medical device and equipment rinsing
For serum and blood fractionation
For ophthalmics
Feed water for ultrapure water systems

A combination of advanced technologies

Accessories
Water Softener

To prevent the RO membrane

and EDI module from scaling

Consistant and Reliable


Optimized Operation Costs

Pre-filtration Kit

To protect RO membrane and


EDI module to maximize the
lifetime of the system

PE Water Reservoir

Designed to minimize the risk


of bacteria growth and air-borne
contamination during the storage

www.rephile.com

Tank Sanitization Module

To inhibit growth of
microbes inside the tank

Specifications

Direct-Pure EDI

Feed Water Requirements


Tap water
Operating temperature
Feedwater pressure
Product Water Flow Rate
Type II water ( 25 )
Product Water Quality
Resistivity ( 25 )
TOC*
Particles ( > 0.2 m )
Microorganisms
Pyrogens (endotoxins)
Dimensions
Length x Depth x Height

Municipal water, TDS <1000 ppm ( 2000 S/cm )


5 - 45
15 - 90 psi ( 1 - 6 bar )
5 and 10 L/hr
> 5 Mcm ( typically 10 -15 Mcm )
< 30 ppb
< 1/ml ( with a 0.2 m final filter or terminal ultrafiltration cartridge)
< 1 cfu/ml ( with a 0.2 m final filter or terminal ultrafiltration cartridge)
< 0.001 Eu/ml ( with a terminal ultrafiltration cartridge)
303 mm 511 mm 482 mm

* Product water quality may vary due to local feed water conditions.

Ordering Information
Description

Catalogue Number

Direct-Pure Water System, EDI 5


Direct-Pure Water System, EDI 5 UV
Direct-Pure Water System, EDI 10
Direct-Pure Water System, EDI 10 UV

RD0E00500
RD0E005UV
RD0E01000
RD0E010UV

Consumables

Catalogue Number

RephiDuo P Pack Cartridge


0.2 m Final Capsule Filter, 1/Pk
254 nm UV Lamp
Reverse Osmosis Membrane with housing
30-liter PE Tank
60-liter PE Tank
Tank Sanitization Module
Tank Vent Filter
Leak Protector with Auto Water Shutoff Valve
Prefiltration Kit (10", 3-stage)
Polypropylene Depth Filter, 10 Inches, 1 m, 12/Pk
Polypropylene Depth Filter, 10 Inches, 10 m, 12/Pk
Activated Carbon Cartridge, 10 Inches, 12/Pk
Softening Cartridge, 10 Inches, 12/Pk
200lph Water Softener System (Auto-regeneration, a complete set)
200lph Activated Carbon Filter (Auto-backwashing)

RR300CP01
RAFFC0200
RAUV357A1
RAR0100HG
RATANK030
RATANK060
RAPRC0102
RATANKVN1
RAPRC0111
RW10311KT
LABPA0112
LABPA1012
LABAC1012
LABAT1012
LAB0200AT
LAB0200AC

RephiLe provides complete water purification solutions


RephiLe not only supplies the lab with Type II water systems, but also offers total water purification solutions for pretreatment,
storage and production of ultrapure water.

Water Softener

Direct-Pure EDI

PE Tank

PURIST UV

All rights reserved 2014 RephiLe Bioscience, Ltd.


RephiLe, Direct-Pure, PURIST are registered trademarks of RephiLe Bioscience, Ltd.. TM and (R) may be omitted in this brochure.

RephiLe Bioscience, Ltd.


Toll Free: +1-855-RephiLe (+1-855-737-4453) E-mail: [email protected]

www.rephile.com

Literature: RFPR0951402

You might also like