Panametrics Mms3 0707 en

Download as pdf or txt
Download as pdf or txt
You are on page 1of 80

GE Industrial

Sensing

Moisture Monitor Series 3


Hygrometer
Abridged Manual

GE Industrial
Sensing

Moisture Monitor Series 3


Hygrometer

Abridged Manual
914-110A4
August 2004
Moisture Monitor Series 3 Hygrometer is a GE Panametrics product. GE Panametrics has joined other GE
high-technology sensing businesses under a new nameGE Industrial, Sensing.

August 2004

Warranty

Each instrument manufactured by GE Infrastructure Sensing, Inc. is


warranted to be free from defects in material and workmanship.
Liability under this warranty is limited to restoring the instrument to
normal operation or replacing the instrument, at the sole discretion of
GE Infrastructure Sensing, Inc. Fuses and batteries are specifically
excluded from any liability. This warranty is effective from the date of
delivery to the original purchaser. If GE Infrastructure Sensing, Inc.
determines that the equipment was defective, the warranty period is:

one year for general electronic failures of the instrument

one year for mechanical failures of the sensor

If GE Infrastructure Sensing, Inc. determines that the equipment was


damaged by misuse, improper installation, the use of unauthorized
replacement parts, or operating conditions outside the guidelines
specified by GE Infrastructure Sensing, Inc., the repairs are not
covered under this warranty.

The warranties set forth herein are exclusive and are in lieu of
all other warranties whether statutory, express or implied
(including warranties of merchantability and fitness for a
particular purpose, and warranties arising from course of
dealing or usage or trade).

Return Policy

If a GE Infrastructure Sensing, Inc. instrument malfunctions within the


warranty period, the following procedure must be completed:
1. Notify GE Infrastructure Sensing, Inc., giving full details of the
problem, and provide the model number and serial number of the
instrument. If the nature of the problem indicates the need for
factory service, GE Infrastructure Sensing, Inc. will issue a RETURN
AUTHORIZATION number (RA), and shipping instructions for the
return of the instrument to a service center will be provided.
2. If GE Infrastructure Sensing, Inc. instructs you to send your
instrument to a service center, it must be shipped prepaid to the
authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE Infrastructure Sensing, Inc. will evaluate the
instrument to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:

If the damage is covered under the terms of the warranty, the


instrument will be repaired at no cost to the owner and returned.

If GE Infrastructure Sensing, Inc. determines that the damage is not


covered under the terms of the warranty, or if the warranty has
expired, an estimate for the cost of the repairs at standard rates
will be provided. Upon receipt of the owners approval to proceed,
the instrument will be repaired and returned.

iii

August 2004

Table of Contents
Chapter 1: General Information
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Unpacking the Series 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Checking the Delta F Oxygen Cell for Leakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Choosing a Site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Grounding the Series 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Moisture/Temperature Probe Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Moisture Condensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Static or Dynamic Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Long-Term Storage & Operational Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Freedom from Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Corrosive Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Sample System Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Moisture Sample Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Oxygen Sample Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9

August 2004

Table of Contents (cont.)


Chapter 2: Installation
Mounting the Hygrometer System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mounting the Electronics Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mounting the Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mounting the Oxygen Cell Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Installing the Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Pressure Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Delta F Oxygen Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Making Basic Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Making Channel Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Connecting the Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Connecting Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Connecting the Delta F Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Establishing a Gas Flow Through the Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Connecting Optional Recorder Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Accessing the Channel Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Setting the Switch Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Replacing the Channel Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Connecting the Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Connecting Optional Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18

vi

August 2004

Table of Contents (cont.)


Chapter 3: Setup & Operation
Startup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Using the Keypad and Passcode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Displaying Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Displaying Measurement Mode and Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Adjusting the Screen Contrast. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Entering System Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Entering a Saturation Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Setting Up the Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Setting Up the Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10

vii

August 2004

Table of Contents (cont.)


Chapter 4: Calibration & Maintenance
Aluminum Oxide Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Probe Cable Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Probe Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Delta F Oxygen Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Electrolyte Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Calibrating the Delta F Oxygen Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Entering Background Gas Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Setting Up a New Probe or Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Reconfiguring a Channel for a New Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Entering Calibration Data for New Probes/Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Setting Up a New Channel Card. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Entering Moisture Reference Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Entering Oxygen Reference Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Entering Pressure Reference Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18

viii

Chapter 1

General Information
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Unpacking the Series 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Checking the Delta F Oxygen Cell for Leakage . . . . . . . . . . . . . . . . . . . . . 1-2
Choosing a Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Grounding the Series 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Moisture/Temperature Probe Considerations . . . . . . . . . . . . . . . . . . . . . . 1-5
Sample System Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8

August 2004

Introduction

The GE Infrastructure Sensing, Inc. Moisture Monitor Series 3 is a


one- or two-channel analyzer designed to measure dissolved moisture
concentration in gases and non-aqueous liquids, and dissolved
oxygen concentration in gases.
A microprocessor-based instrument, the Series 3 combines hardware
and software to make various measurements. The user connects the
needed inputs (moisture probes, pressure transmitters, oxygen cells,
etc.) to the back panel of the electronics unit using the appropriate
cables. Typically, the user installs moisture probes and oxygen cells
into the process using a sample system that is specifically designed
for the application. The sample system delivers a sample of the
process gas or liquid to the probes. The probes then send signals to
the Series 3 electronics unit, which interprets the signals and converts
them into measurements.
Users typically install the Series 3 as part of a complex process
system, which includes components such as filters, pumps, and
pressure regulators. In such an environment, probes and other parts of
the system may be subjected to environmental hazards, such as high
temperature, pressure extremes, corrosive elements, and mechanical
vibrations.
!WARNING!
To ensure the safe operation of this unit, you must install
and operate the Series 3 as described in this manual. In
addition, be sure to follow all applicable safety codes and
regulations for installing electrical equipment in your area.

Unpacking the Series 3

Upon receipt, unpack the Series 3 and make sure all the parts and
documentation listed on the packing slip are included. The packing
slip may not list the Calibration Data Sheet(s), which are usually
packed in the plastic storage case with the moisture, oxygen, and
pressure probes. You may also find the Calibration Data Sheet(s) in
an envelope taped to the Series 3. There should be one Calibration
Data Sheet for each probe.
Be sure to inspect each component, including the sample system, for
evidence of mishandling. If anything has been damaged, report this to
the carrier and to GE Infrastructure Sensing, Inc. immediately. You
should leave the plastic caps on the probes and the pressure
transmitters when they are not installed in the process stream. If
anything is missing, contact GE Infrastructure Sensing, Inc.
immediately.

General Information

1-1

August 2004

Checking the Delta F


Oxygen Cell for Leakage

Before connecting the Delta F Oxygen Cell(s), you must check it for
damage and/or leakage. Depending on the application, the oxygen
cell may have a top drain or both a top and bottom drain for the
electrolyte reservoir. It is important to identify your cell for the
following procedure. Use Figure 1-1 below to identify your cell.
1. Remove the top of the electrolyte reservoir.
IMPORTANT:

If your cell also has a bottom drain, make sure that


the electrolyte discharge valve, mounted on the rear
of the oxygen cell, is closed (in the vertical position).
See Figure 1-1 below.

2. Add approximately three ounces (100 ml) of distilled water to the


reservoir and replace the top.
3. Using the min/max window (see Figure 1-2 on the next page) on
the oxygen cell, check the water level. The water should cover
about 60% of the window.
Top Drain

Bottom Drain
Electrolyte Discharge Valve
(in vertical, closed position)
Figure 1-1: Delta F Oxygen Cell - Drain Locations
4. Let the oxygen cell stand for about 6 hours; then check for any
leakage.
5. If there is no leakage, drain the cell completely.
If the cell leaks, see the warranty information at the beginning of this
manual.

1-2

General Information

August 2004

Checking the Delta F


Oxygen Cell for Leakage
(cont.)

Min/Max Window

Water Level
Figure 1-2: Min/Max Window and Water Level

Choosing a Site

The Series 3 is available in rack, bench or panel mounts that are


suitable for most indoor installations, as well as weatherproof and
explosion-proof configurations. See the drawings at the end of this
chapter for an example of each enclosure.
You should have discussed environmental and installation factors
with an applications engineer or field sales person by the time you
receive the Series 3. The equipment should be suited to the
application and installation site.
Before installing the unit, read the guidelines below to verify that you
have selected the best installation site.
IMPORTANT:

General Information

For compliance with the EUs Low Voltage Directive


(IEC 1010), this unit requires an external power
disconnect device. The disconnect device for this unit
is its power cord.

1-3

August 2004

Choosing a Site (cont.)

!WARNING!
Division 2 applications may require special installation.
Consult the National Electric Code
and/or the Canadian Electrical Code for proper installation
requirements. The analyzer must be configured in a
suitable equipment enclosure and installed according to
the sections of the National Electric Code, Article 500, and
Canadian Electrical Code, Section 18, that pertain to the
hazardous environment classification in which the
electronics will be used.

Choose an installation site for the probes and sample systems that
is as close to the process line as possible. Avoid long runs of
connecting tubing. If long distances are unavoidable, a fast
sampling by-pass loop is recommended. Do not install any other
components, such as filters, ahead of the probes or sample system
unless instructed by GE Infrastructure Sensing, Inc. to do so.

Observe all normal safety precautions. Use probes within their


maximum pressure and temperature ratings.

Although the Series 3 may not need to be accessed during normal


operation, install the electronics unit at a convenient location for
programming, testing and servicing. A control room or instrument
shed are typical locations.

Locate the electronics unit away from high temperatures, strong


electrical transients, mechanical vibrations, corrosive atmospheres,
and any other conditions that could damage or interfere with the
Series 3 operation.

Protect the probe cables from excessive physical strain (bending,


pulling, twisting, etc.). In addition, do not subject the cables to
temperatures above 65C (149F) or below 50C (58F).

Observe the proper cable restrictions for the probes as follows:

The M Series and TF Series probes require specially shielded


cable. You can locate the M and TF probes up to 600 m (2,000
ft) from the Series 3. If you are measuring pressure with a TF
probe, the cable length should not exceed 152 m (500 ft).

The Delta-F Oxygen Cell uses a four-wire shielded cable 22


AWG). Cells with a range from 0 to 1/10/100 ppmv or 0 to 0.5/
5/50 ppmv can be located up to 15 m (50 ft) away. All other
cells can be located up to 91 m (300 ft) away.

Consult GE Infrastructure Sensing, Inc. for remote location of the


oxygen cell and cable restrictions for other sensors.

1-4

General Information

August 2004

Grounding the Series 3

The Series 3 case is grounded to the electrical system safety ground


through the third wire in the power connector and cord (refer to
Figure 1-3 below). This ground should not be removed.
Fuse
STD/TF

STD/TF

PROBE

PROBE

3
4
5

3
4
5

8
9

4
5

8
9

4
5

OXYGEN

CHANNEL 1
ALM A
NO C

ALM B

NC RTN

A RECB

NO C

NC

AUX
RTN 1

2 +24V

1/2 AMP

250V
SLO-BLO

3AG

OXYGEN

CHANNEL 2
ALM A
NO C

Line
Power

ALM B

NC RTN

A RECB

L ine
G nd
N eut

NO C

NC

AUX
RTN 1

2 +24V

GND

Figure 1-3: Series 3 Back Panel - Ground Location

Moisture/Temperature
Probe Considerations

The M Series and TF Series probes consist of an aluminum oxide


sensor mounted on a connector head. Standard probe mounts include
a protective stainless-steel shield.
The probe sensor materials and housing maximize durability and
insure a minimum of water-adsorbing surfaces in the vicinity of the
aluminum oxide surface. A sintered stainless-steel shield is used to
protect the sensor from high flow rates and particulate matter. The
shield should not be removed except upon advice from GE
Infrastructure Sensing, Inc..
The sensor has been designed to withstand normal shock and
vibration. You should make sure that the active sensor surface is
never touched or allowed to come into direct contact with foreign
objects, since this may adversely affect performance.
Observing these few simple precautions will result in a long and
useful probe life. GE Infrastructure Sensing, Inc. recommends that
probe calibration be checked routinely, at 6-month intervals, or as
recommended by our applications engineers for your particular
application.
The probe will measure the water vapor pressure in its immediate
vicinity; therefore, readings will be influenced by its proximity to the
system walls, materials of construction, and other environmental
factors. The sensor can be operated under vacuum or pressure,
flowing or static conditions.
Observe the environmental precautions noted on the next page.

General Information

1-5

August 2004

Temperature Range

The standard probe is operable from 110C to +70C (166F to


158F).

Moisture Condensation

Be sure the temperature is at least 10C (18F) higher than the dew/
frost point temperature. If this condition is not maintained, moisture
condensation could occur on the sensor or in the sample system,
which will cause reading errors. If this happens, dry out the probe
following the procedures outlined in Chapter 3.

Static or Dynamic Use

The sensor performs equally well in still air or where considerable


flow occurs. Its small size makes it ideal for measuring moisture
conditions within completely sealed containers or dry boxes. It will
also perform well under gas flow conditions as high as 10,000 cm/sec
and liquid flow conditions to 10 cm/sec. Refer to Table 1-1 below and
Table 1-2 on the next page for maximum flow rates in gases and
liquids.
Table 1-1: Maximum Gas Flow Rates
Based on the physical characteristics of air at a temperature of
77F and a pressure of 1 atm, the following flow rates will
produce the maximum allowable gas stream linear velocity of
10,000 cm/sec in the corresponding pipe sizes.
Inside Pipe Diameter (in.)
0.25
0.50
0.75
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
11.0
12.0

1-6

Gas Flow Rate (cfm)


7
27
60
107
429
966
1,718
2,684
3,865
5,261
6,871
8,697
10,737
12,991
15,461

General Information

August 2004

Static or Dynamic Use


(cont.)

Table 1-2: Maximum Liquid Flow Rates


Based on the physical characteristics of benzene at a
temperature of 77F, the following flow rates will produce the
maximum allowable fluid linear velocity of 10 cm/sec in the
corresponding pipe sizes.
Pipe I.D. (in.)
0.25
0.50
0.75
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
11.0
12.0

Flow Rate (gal/hr)


3
12
27
48
193
434
771
1,205
1,735
2,361
3,084
3,903
4,819
5,831
6,939

Flow Rate (l/hr)


11
46
103
182
730
1,642
2,919
4,561
6,567
8,939
11,675
14,776
18,243
22,074
26,269

Pressure

The moisture probe always senses the correct water vapor pressure
regardless of total ambient pressure. The moisture sensor measures
water vapor under vacuum or high pressure conditions from as little
as 5 m Hg to as high as 5,000 psi total pressure.

Long-Term Storage &


Operational Stability

Sensors are not affected by continuous abrupt humidity changes or


damaged by exposure to saturation conditions even when stored.
However, you should store probes in their original shipping container
in a clean, dry location. If the probe is saturated during storage, refer
to Moisture Condensation on later in this chapter before installing the
probe. For best performance, do not store probes longer than one to
two years from their calibration date.

Freedom from
Interference

The sensor is completely unaffected by the presence of a wide variety


of gases or organic liquids. Large concentrations of hydrocarbon
gases, Freon, ozone, carbon dioxide, carbon monoxide, and
hydrogen have no effect on sensor water vapor indications. The
sensor will operate properly in a multitude of gaseous or nonconductive liquid environments.

Corrosive Materials

Avoid all materials that are corrosive or otherwise damaging to


aluminum or aluminum oxide. These include strongly acidic or basic
materials and primary amines.

General Information

1-7

August 2004

Sample System
Guidelines

A sample system is required for oxygen measurement and, although


not mandatory, is highly recommended for moisture measurement.
The purpose of a sample system is to condition or control a sample
stream to within the specifications of a probe. The application
requirements determine the design of the sample system. GE
Infrastructure Sensing, Inc. applications engineers will make
recommendations based on the following general guidelines.

Moisture Sample Systems

Typically, sample systems should contain as few components as


possible and all or most of those components should be located
downstream of the measurement probe. Figure 1-4 below shows an
example of a basic sample system consisting of an explosion-proof
housing with a sample cell, a filter, a flow meter, a vent valve and
two-shut off valves, one at the inlet and one at the outlet.
The sample system components should not be made of any material
that will affect measurements. A sample system may include a filter
to remove particulates from the sample stream or a pressure regulator
to reduce or control the pressure of the stream. However, most
common filters and pressure regulators are not suitable for sample
systems because they have wetted parts that may absorb or release
components (such as moisture) into the sample stream. They may also
allow ambient contamination to enter the sample stream. In general,
you should use stainless steel material for all wetted parts.

Sample
Cell

Sample
Outlet

Sample
Inlet

Vent

Figure 1-4: Typical Moisture Sample System

1-8

General Information

August 2004

Oxygen Sample Systems

Oxygen sample systems are required and can be be ordered from GE


Infrastructure Sensing, Inc. for bench or wall mounting, or you can
build your own sample system by using the following guidelines.
IMPORTANT:

The GE Infrastructure Sensing, Inc. warranty is void,


if the sample system does not have a relief valve.

The basic sample system requirements are (see Figure 1-5 below):
1. The oxygen cell requires a sample gas flow of 2.0 to 2.5 SCFH.
2. The sample gas pressure in the cell must be between 0.0 and 1.0
psig. The pressure must not exceed 1.0 psig.
3. A 10 psig pressure relief valve installed upstream of the oxygen
cell is required to prevent over-pressure.
4. A flow meter is required to measure the flow.
5. A pressure gauge is required to measure the pressure.
6. A flow regulating or needle valve is required to regulate flow and
should be located upstream of the cell.
7. A pressure regulator is required for sample gas supplies of 50 psig
or greater.
If a sample pump is required to draw a sample to the oxygen cell, the
pump should be installed downstream of the oxygen cell. This will
also require you to install a vacuum relief valve set at 1.0 psig
between the oxygen cell and the pump.

Sample
Inlet

Sample
Outlet

Figure 1-5: Typical Oxygen Cell Sample System

General Information

1-9

Chapter 2

Installation
Mounting the Hygrometer System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Installing the Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Making Basic Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Establishing a Gas Flow Through the Oxygen Cell . . . . . . . . . . . . . . . . . 2-14
Connecting Optional Recorder Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Connecting Optional Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18

August 2004

Mounting the
Hygrometer System

Mounting the hygrometer system consists of mounting the electronics


unit, the probes, and the sample system(s). Refer to Figures 2-14 to 216 on pages 2-19 to 2-21 for the dimensions of your unit.

Mounting the Electronics


Unit

Use the outline and dimension drawings at the end of this chapter to
mount the Series 3. These drawings provide clearance and other
mounting dimensions needed to prepare the site for mounting.
IMPORTANT:

For compliance with the EUs Low Voltage Directive


(IEC 1010), this unit requires an external power
disconnect device such as a switch or circuit breaker.
The disconnect device must be marked as such,
clearly visible, directly accessible, and located
within 1.8 m (6 ft) of the Series 3. The power cord is
the main disconnect device.

Be sure to follow the guidelines outlined in Choosing a Site in


Chapter 1 before mounting the enclosure.
Note: You may want to make probe, Delta F Oxygen Cell, recorder,
and alarm connections before mounting the enclosure, if the
installation location does not provide enough room for these
connections to be made easily after installation.

Mounting the Sample


System

The sample system is normally fastened to a metal plate that has four
mounting holes. GE Infrastructure Sensing, Inc. also provides the
sample system in an enclosure if requested. In either case, mount the
sample system plate or enclosure with four bolts one in each
corner. If you ordered sample system outline and dimensions
drawings, they will be included in your shipment.
Connect the sample system inlet/outlet to the process/return using the
appropriate fittings or an appropriate NPT adapter.
Caution!
Do not start a flow through the system until all probes and
transmitters are properly installed.

Mounting the Oxygen Cell


Assembly

Installation

If your oxygen cell is not mounted into a sample system, refer to


Figure 2-18 on page 2-23 to mount the cell.

2-1

August 2004

Installing the Probes

After you mount the sample system you must insert moisture probes
into the sample cells. In addition, you must check, prepare, and
connect oxygen cells (if used) to the gas line.

Moisture Probes

GE Infrastructure Sensing M Series and TF Series moisture probes


have 3/4-16 straight threads with an o-ring to secure the probes into
the sample system or directly into the process line.
Caution!
If mounting the moisture probes directly into the process
line, you must consult the factory for proper installation
instructions and precautions.
Moisture probes are usually installed into a sample system to protect
the probes from coming into contact with damaging elements in the
process. The sample system includes a sample cell (see Figure 2-1
below) for mounting the moisture probe. For easy identification, the
sample cell is labeled on the sample system plate. To install a
moisture probe in the sample cell, insert it so that the probe is
perpendicular to the sample inlet. Screw the probe into the cell,
making sure not to cross thread it, and tighten it securely.

Probe

Inlet

Sample Cell

Outlet
Figure 2-1: Moisture Probe Installed in Sample Cell
Note: Standard moisture probes have a sintered stainless-steel
shield that protects the aluminum oxide sensor. Leave the
shield in place for maximum protection.
IMPORTANT:

2-2

You must eliminate all leaks for safety reasons and to


be sure that measurements are not affected by
ambient contamination. For gas applications, you
should check for leaks using a soap bubble solution.

Installation

August 2004

Pressure Sensor

If a pressure measurement is required, and for some reason the TF


probe pressure option is not used, you can connect a separate pressure
sensor to an auxiliary input. The Series 3 uses any type of 0/4 to 20
mA or a 0 to 2 V pressure transducer or transmitter. GE Infrastructure
Sensing, Inc. offers two types of pressure transmitters: the P40 and
P40X. The P40 has a 1/4-inch threaded NPTM fitting and the P40X
has a 1/2-inch threaded NPTF fitting for mounting directly into the
process line or into a sample system.
Caution!
If mounting the pressure transmitters directly into the
process line, consult the factory for proper installation
instructions and precautions.
Always mount the pressure transmitter directly downstream of the
moisture probe to measure the pressure as close as possible to the
point of the moisture measurement.

Delta F Oxygen Cell

Although the Series 3 can use other oxygen devices as auxiliary


inputs, it is designed to accept oxygen inputs directly from the Delta
F Oxygen Cell. There are three steps for installing the Delta F Oxygen
Cell: preparing the oxygen cell for operation, calibrating the oxygen
cell, and then connecting the cell to the gas line.
The Delta F Oxygen Cell is available in various enclosure types;
however, the cell itself looks like either of the cells shown in Figure
2-2 below. The oxygen cell may have both a bottom and top drain or
only a top drain.

Bottom/Top Drain

Top Drain

Figure 2-2: Delta F Oxygen Cell Drains

Installation

2-3

August 2004

Preparing the Oxygen Cell

Prepare the oxygen cell for operation by filling it with the electrolyte
that has been supplied in a plastic bottle (see Figure 2-3 below).
!WARNING!
The electrolyte contains potassium hydroxide, which is
harmful if it comes in contact with eyes or skin. Consult
your company safety personnel for proper procedures for
handling the electrolyte.
1. Unscrew the top on the reservoir. If you are using an explosionproof or bottom-top drain cell, make sure the electrolyte discharge
valve is in the vertical (closed) position.

Electrolyte Level

Bottom Drain
Electrolyte Discharge Valve
(in vertical, closed position)
Figure 2-3: Bottom Drain Delta F Oxygen Cell
2. Slowly add the entire contents of the bottle, approximately three
ounces (90 ml), making sure not to spill any on the outside of the
cell. Be especially careful that the electrolyte does not come in
contact with any of the cells electrical connections.
3. Using the min/max window on the oxygen cell, check the
electrolyte level. The electrolyte should cover about 60% of the
window. The cell is now ready to be connected to the gas line.
4. Replace the top cover of the oxygen cell.
Note: DO NOT add additional electrolyte to the reservoir. If the
level falls below the minimum level,see page 4-4 for
instructions on electrolyte maintenance.
5. Calibrate the oxygen cell as described on page 4-4. After you
calibrate the cell, connect it to the gas line as described in the
following section.
Note: Oxygen cells are calibrated using nitrogen. If you plan to use
your cell with a gas other than nitrogen, you must enter a
current multiplier as described on page 4-7.

2-4

Installation

August 2004

Connecting the Oxygen


Sample System to the Gas
Line

To connect the oxygen sample system to the process line, attach 1/8
tubing to the sample gas inlet fitting using a Swagelok or equivalent
mating connector. Avoid using plastic or rubber for any tubing or
fittings that are included in the inlet gas lines.
Caution!
Do not connect the oxygen cell outlet to flow restricting
devices, pressure lines, or vacuum lines. Pressure
differentials across the cell sensor in excess of 1 psig
could damage the cell.
If the sample gas is not a safety hazard, vent it to atmosphere at the
sample system outlet. If venting the gas to atmosphere causes a safety
hazard, vent the gas to a safe location. Make sure the venting system
does not create a back pressure for the oxygen cell.
Note: The vented sample will not be corrosive if you install and
operate the oxygen cell properly.

Making Basic Electrical


Connections

!WARNING!
To ensure the safe operation of this unit, you must install
and operate the Series 3 as described in this manual. In
addition, be sure to follow all applicable safety codes and
regulations for installing electrical equipment in your area.
Turn the Series 3 OFF before making any connections.

STD/TF

STD/TF
PROBE

3
4
5

OXYGEN

8
9

4
5

CONNECTIONS

HAZARDOUS AREA

PROBE

1/2 AMP

250V

SLO-BLO

3
4
5

3AG

OXYGEN

8
9

4
5

CHANNEL 1
ALM A
NO C

NC RTN

A RECB

ALM B
NO C

NC

AUX
RTN 1

2 +24V

CHANNEL 2
ALM A
NO C

ALM B

NC RTN

A RECB

L ine
G nd
N eut

NO C

NC

AUX
RTN 1

2 +24V

Figure 2-4: Series 3 Back Panel

Installation

2-5

August 2004

Making Channel
Connections

Make all connections to the back of the meter (see Figure 2-4 on the
previous page and Figure 2-17 on page 2-22). The larger panel is
separated into two sections, one for each channel. Make connections
by placing the press lock lever into the desired terminal (one presslock lever is supplied with each terminal block). Press and hold the
lever against the terminal block and insert the stripped and tinned
portion of the wire into the terminal. Then, release the lever to secure
the connection.
IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.

Proper connections and cabling are extremely important for accurate


measurements. Be sure to use the correct cable type for each probe,
and make sure that the cables are not damaged during installation. If
you are not using a factory-supplied cable, or you are using a
modified cable, read the following section carefully.

Connecting the Power

!WARNING!
Division 2 applications may require special installation.
Consult the National Electric Code and/or the Canadian
Electrical Code for proper installation requirements. The
analyzer must be mounted in a suitable equipment
enclosure and installed according to the sections of the
National Electric Code, Article 500, and Canadian Electrical
Code, Section 18, that pertain to the hazardous environment
classification in which the electronics will be used.
Note: The power line is the main disconnect device. However, GE
Infrastructure Sensing, Inc. does not provide power supply
cords with CSA Div. 2 hygrometers.
IMPORTANT:

2-6

For compliance with the EUs Low Voltage Directive


(IEC 1010), this unit requires an external power
disconnect device such as a switch or circuit breaker.
The disconnect device must be marked as such,
clearly visible, directly accessible, and located
within 1.8 m (6 ft) of the Series 3.

Installation

August 2004

Connecting Moisture
Probes

GE Infrastructure Sensing, Inc. manufactures a variety of moisture


probes for the Series 3. The most commonly used are the M Series
and TF Series. If you ordered one or more of these probe types, the
probe setup data has been entered on a preassigned channel. Channel
designations and probe serial numbers are listed on the Calibration
Data Sheet that is shipped with each probe. The serial number is also
engraved on the probe hex nut, as shown in Figure 2-5 below.
Probe Serial Number

Figure 2-5: Moisture Probe Serial Number

M Series Probes

M Series probes are primarily used for moisture measurement, but


they can be also equipped with an optional thermistor to measure
temperature. The optional temperature thermistor is included in the
moisture probe and requires an additional connection.
The M Series Probe is connected with a four-wire shielded cable
having a bayonet-type connector and a length of up to 600 m (2,000
ft). Consult the factory for longer cables. Before making the electrical
connections, connect the cable to the probe by inserting the bayonettype connector onto the probe and twisting the shell clockwise until it
snaps into a locked position (approximately 1/8 of a turn). If you are
not using a factory-supplied cable, refer to Figure 2-6 below to make
proper pin connections to a bayonet-type connector.
STD Probe
Terminal Block
H1
H2
T1
T2

M Series Probe

Red

H1

Green

H2

Black

T1

White

T2

To
Moisture
Sensor
To
Temp.
Sensor

Bayonet
Connector
Pins

6
7
8
RTN

Shield

Figure 2-6: M Series Probe Cable Assembly

Installation

2-7

August 2004

M Series Probes (cont.)

NOTICE FOR BASEEFA CERTIFICATION


The M Series probe may not be capable of withstanding the
500 V insulation test required by clause 5.7 of EN50 020 when
installed in the process media. This must be taken into
account in any installation in which it is used. (See Cert. #
Ex95C2002X in its entirety.) Copies of official BASEEFA
documentation (certificates of compliance, licenses, etc.) are
to be made in their entirety.
IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.

Use Table 2-1 below to connect the remaining end of the cable to the
terminal block labeled STD/TF PROBE on the back of the electronics
unit (see Figure 2-4 on page 2-5 for the terminal block location). All
the connections listed in the table must be made, even if you do not
have the temperature option.
IMPORTANT:

For intrinsically safe installations, refer to Figure


2-19 on page 2-24.
Table 2-1: M Series Probe Connections
Connect:

To STD/TF PROBE TB:

red wire (moisture H1)

pin 1

green wire (moisture H2)

pin 2

white wire (temperature T1)

pin 3

black wire (temperature T2)

pin 4

shield

pin 9

After you connect the cables, perform a calibration test, as described


on page 4-1, to compensate for any small electrical offsets in the
cable.

2-8

Installation

August 2004

TF Series Probe

Using the special factory-supplied cable, connect the TF Series Probe


to the designated terminal blocks on the back panel of the Series 3.
For moisture and temperature measurements, the TF Series Probe
may be located to 600 m (2,000 ft) from the meter. For pressure
measurements, the cable length should not exceed 152 m (500 ft).
Connect the cable to the probe by inserting the connector onto the
probe and securing it. If you are not using a factory-supplied cable,
refer to Figure 2-7 below to make proper pin connections to a
bayonet-type connector.
STD Probe
Terminal Block
H1
H2
T1
T2
P1+
P1
P2+
P2
RTN

TF Probe

Red

H1

Green

H2

White

T1

Black

Violet

Orange

Yellow

Blue

Shield

10 T2

To
Moisture
Sensor
To
Temp.
Sensor
Excitation
To
Pressure
Transducer
Output

Figure 2-7: Cable Assembly for TF Series Probe


IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.

NOTICE FOR BASEEFA CERTIFICATION


The TF Series probe may not be capable of withstanding the
500 V insulation test required
by clause 5.7 of EN50 020 when installed in the
process media. This must be taken into account
in any installation in which it is used.
(See Cert. # Ex95C2002X in its entirety.)
Copies of official BASEEFA documentation
(certificates of compliance, licenses, etc.)
are to be made in their entirety.
IMPORTANT:

Installation

For intrinsically safe installations, refer to Figure


2-19 on page 2-24.

2-9

August 2004

TF Series Probe (cont.)

Use Table 2-2 below to connect the remaining end of the cable to the
terminal blocks labeled STD/TF PROBE on the back of the electronics
unit (see Figure 2-4 on page 2-5 for the terminal block location).
Table 2-2: TF Series Probe Connections
Connect:

To STD/TF PROBE TB:

red wire (moisture H1)

pin 1

green wire (moisture H2)

pin 2

white wire (temperature T1)

pin 3

black wire (temperature T2)

pin 4

violet wire (IN +)

pin 5

orange wire (IN -)

pin 6

yellow wire (OUT +)

pin 7

blue wire (OUT -)

pin 8

shield

pin 9

After you connect the probe, perform a calibration test. as described


on page 4-1, to compensate for any small electrical offsets in the
cable.

Connecting the Delta F


Oxygen Cell

The Delta F Oxygen Cell is available in a general purpose model with


standard or VCR fittings. The oxygen cell can also be mounted in a
weatherproof enclosure (R4) for outdoor applications or an
explosion-proof enclosure (R7) for hazardous areas.
Caution!
Do not power up your Series 3 without establishing a flow
through the Delta F Oxygen Cell (see page 2-14).
Each oxygen cell has a set of sensing and secondary electrodes that
must be connected to the terminal block labeled OXYGEN on the back
of the meter (see Figure 2-4 on page 2-5). Instructions for connecting
each type of cell are given in the following sections (for intrinsically
safe installations, refer to the following section and to Figure 2-19 on
page 2-24, for special installation requirements.
Caution!
The Delta F Oxygen Cell is BASEEFA approved for use in
intrinsically safe areas when connected to a BASEEFA
approved Series 3 Hygrometer.The certification numbers are
as follows: Series 3 I.S. Cert. Ex95D242432/2Delta F Oxygen
Cell Ex96D2191XInstall the apparatus so the terminals are
protected to at least the requirements of IP20. Copies of
official BASEEFA documentation (certificates of compliance,
licenses, etc.) are to be made in their entirety.

2-10

Installation

August 2004

Standard Delta-F Oxygen


Cells and Cells with VCR
Fittings

Figure 2-8 below shows the standard oxygen cell and identifies the
sensing and secondary electrodes.

Secondary
Electrodes

Sensing
Electrodes

Figure 2-8: Standard Delta F Oxygen Cell


IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.

Make oxygen cell connections for to the OXYGEN terminal block on


the back of the meter using Table 2-3 below. For each set of
electrodes use a four-wire shielded (22 AWG) cable. Oxygen cells
with a range from 0 to 1/10/100 ppmv or 0 to 0.5/5/50 ppmv can be
located up to 50 ft (15 m) away from the analyzer. All other oxygen
cells may be located up to 300 ft (100 m) away.
Table 2-3: Standard Delta F Oxygen Cell Connections*
Connect:

To Delta F Oxygen Cell:

To OXYGEN TB:

red wire

+ sensing electrode

pin 1

green wire

sensing electrode

pin 2

white wire

+ secondary electrode

pin 3

black wire

secondary electrode

pin 4

shield

cable shield

pin 5

*includes cells with VCR Fittings

Installation

2-11

August 2004

Weatherproof Oxygen Cell

Figure 2-9 below shows the weatherproof oxygen cell and the
terminal block for wiring the sensing and secondary electrodes.

Terminal
Block
1
2
3
4
5
6

MAX

MIN

Figure 2-9: Weatherproof Delta F Oxygen Cell


IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.

Using Table 2-4 below, connect each set of electrodes using a fourwire shielded (22 AWG) cable. Connect one end of the cable to the
terminals in the oxygen cell enclosure and the other end of the cable
to the OXYGEN terminal block on the back of the meter (see Figure 2-4
on page 2-5). Oxygen cells with a range from 0 to 1/10/100 ppmv or 0
to 0.5/5/50 ppmv can be located up to 50 ft (15 m) from the analyzer;
all other oxygen cells can be located up to 300 ft (100 m) away.
Table 2-4: Weatherproof Delta F Oxygen Cell Connections
Connect:
To Delta F Enclosure: To OXYGEN TB:

2-12

red wire (+sensing)

pin 1

pin 1

green wire (sensing)

pin 3

pin 3

white wire (+secondary)

pin 4

pin 4

black wire (secondary)

pin 5

pin 5

shield

pin 2

pin 2

Installation

August 2004

Explosion-Proof Oxygen
Cell

Figure 2-10 below shows an explosion-proof oxygen cell and the


terminal block for wiring the sensing and secondary electrodes.

Figure 2-10: Explosion-proof Oxygen Cell


IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.

Using Table 2-5 below, connect each set of electrodes using a fourwire shielded (22 AWG) cable. Connect one end of the cable to the
terminals in the oxygen cell enclosure and the other end of the cable
to the OXYGEN terminal block on the back of the meter (see Figure 2-4
on page 2-5). Oxygen cells with a range from 0 to 1/10/100 ppmv or 0
to 0.5/5/50 ppmv can be located up to 50 ft (15 m) from the analyzer;
all other oxygen cells can be located up to 300 ft (100 m) away.
Table 2-5: Explosion-Proof Oxygen Cell Connections
Connect:
To Oxygen Cell TB: To OXYGEN TB:

Installation

red wire (+sensing)

pin 1

pin 1

green wire (sensing)

pin 2

pin 2

white wire (+secondary)

pin 3

pin 3

black wire (secondary)

pin 4

pin 4

shield

N/C

pin 5

2-13

August 2004

Establishing a Gas Flow


Through the Oxygen Cell

Caution!
Establish a gas sample flow before you power up the
Series 3 or damage may occur to the oxygen cell.
If you are using an oxygen cell, you must establish a gas flow through
the cell before powering up (if you are not using an oxygen cell,
proceed directly to the Startup Procedure on page 3-1). The oxygen
cell requires a flow rate of 2-2.5 SCFH through the cell, and the
oxygen cell inlet pressure should be 0.2-1.0 psig. Refer to Figure 2-11
below when establishing a gas sample flow.

Optimum Inlet
Pressure Limits
0.2 1.0 psig

Sample
Gas In

Optimum Sample
Flow Range
2.0 2.5 SCFH

Flow Control
Valve

Tee

Electrolyte
Reservoir
In

Out

Oxygen Cell

Sample
Gas Out

Relief
Valve
Assembly
Vent

Flowmeter

Sample Gas Outlet and Relief Valve


should vent to atmosphere.

Figure 2-11: Gas Flow Schematic


Caution!
Do not operate the Delta F oxygen cell for extended
periods of time at oxygen concentrations that are over
range. Trace and low percent range sensors may be
damaged if exposed to high levels of oxygen, such as air,
for long periods (>1 hour) while the Series 3 is on. If
exposure is unavoidable, either disconnect the oxygen cell
from the Series 3 or equip the sample system with a valve
that allows the cell to be switched to purge gas.

2-14

Installation

August 2004

Establishing a Gas Flow


Through the Oxygen Cell
(cont.)

To establish the gas flow, complete the following steps:


1. Close the flow control valve and adjust the upstream pressure as
required. GE Infrastructure Sensing, Inc. recommends about 2-10
psig upstream of the flow control valve, depending on the type of
valve installed in the sample system.
2. To safeguard against over-pressurizing the oxygen cell, install a
relief valve rated at 10 psig into the gas flow system. If the
pressure exceeds 10 psig, the relief valve will open; therefore,
there should be no restrictions downstream of the oxygen cell. Use
1/4-inch tubing or larger on the oxygen cell outlet and relief valve
outlet. Both outlets should vent to atmosphere if possible.
Caution!
Do not connect the relief valve and oxygen cell outlet to a
common outlet line smaller than 1/4 inch. This pressure
restriction will damage the oxygen cell. In addition, a relief
valve should be installed in the oxygen sample system. If
either of these conditions is not met, the Delta F oxygen
cell warranty is void.
3. Slowly open the flow control valve until the recommended flow of
2 to 2.5 SCFH is established on the flow meter.
4. When the proper flow is achieved, make sure the relief valve is
closed by placing an object (e.g., your finger, if the gas is noncorrosive) over the relief valve vent. Cover and uncover the relief
valve vent and verify that the flow meter shows no change in the
flow rate. Keep the relief valve closed during operation to
minimize leakage in the sample system.

Installation

2-15

August 2004

Connecting Optional
Recorder Outputs

The Series 3 has two optically isolated recorder outputs. These


outputs provide either a current or voltage signal, as selected by
switch blocks on the channel card. Although the Series 3 is
configured to your specifications at the factory, you should check the
switch block positions before making connections. Refer to Figure
2-12 below and check or reset these switch blocks as follows:

Channel
Cards
Retainer Bar

Screw

Screw

Top View
Figure 2-12: Channel Cards Location

Accessing the Channel


Cards

1. Remove the screws on the Series 3 front panel and slide the
electronics unit out of its enclosure.
2. Remove the retainer bar by removing the two screws on the
outside of the chassis.
3. Remove the desired channel card by sliding it straight up.

Setting the Switch Blocks

1. Locate switch blocks S2 and S3 (see Figure 2-13 on the next


page). Switch block S2 controls the output signal for Recorder A
and switch block S3 controls the output signal for Recorder B.
2. Set the switch blocks to the appropriate positions: I for current or
V for voltage.

2-16

Installation

August 2004

S3
S2
Figure 2-13: Channel Card - S2 and S3 Locations

Replacing the Channel


Card

1. Once the switches are set, replace the channel card.


Note: If you intend to connect pressure inputs or other input devices
to the Series 3, do not replace the retainer bar and cover,
because you will need to set switches on the channel card for
those inputs as well.
2. Replace the retainer bar. Make sure the slots on the retainer bar are
seated correctly against the printed circuit boards. Secure the bar
with the two screws that were previously removed.
3. Slide the electronics units into its enclosure and reinstall the
screws. Tighten the screws until they are snug, but do not over
tighten them.

Connecting the Recorders

Connect the recorders to the REC terminal block on the back panel
(see Figure 2-4 on page 2-5). Use Table 2-6 below to make the
recorder connections
IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.
Table 2-6: Recorder Connections

Installation

Connect:

To REC Terminal Block:

Recorder A out (+)

pin A+

Recorder A return ()

pin A

Recorder B out (+)

pin B+

Recorder B return ()

pin B

2-17

August 2004

Connecting Optional
Alarms

You can order the Series 3 with optional high and low alarm relays in
either general-purpose or hermetically-sealed styles. Each alarm relay
is a single-pole, double throw relay with the following contacts:

normally-closed (NC)

armature/common (C)

normally-open (NO)

Make connections for the high and low alarm relays on the desired
channel terminal blocks labeled ALM A and ALM B on the back
panel of the electronics unit (see Figure 2-4 on page 2-5). Use Table
2-7 below to make the alarm connections.
IMPORTANT:

To maintain good contact at each terminal block and


to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.
Table 2-7: Alarm Connections

Connect Low Alarm:

To ALM A Terminal Block:

NC contact

pin NC

C contact

pin C

NO contact

pin NO

Connect High Alarm:

To ALM B Terminal Block:

NC contact

pin NC

C contact

pin C

NO contact

pin NO

Note: The alarm terminal block has an additional Return connection


that you can use to ground the alarms if desired.

2-18

Installation

Figure 2-14: Rack-Mount Outline & Installation Drawing (712-240)

August 2004

Installation
2-19

Figure 2-15: Panel-Mount Outline & Installation Drawing (712-241)

August 2004

Installation
2-20

Figure 2-16: Bench-Mount Outline & Installation Drawing (712-242)

August 2004

Installation
2-21

Figure 2-17: Interconnection Diagram (702-196)

August 2004

Installation
2-22

Figure 2-18: Delta F Oxygen Cell (Ref. 752-064

August 2004

Installation
2-23

4, 5 & 8

4, 5 & 8

Installation

Figure 2-19: Intrinsically-Safe Installation Requirements (752-138)

August 2004

MMS 3

2-24

Chapter 3

Setup & Operation


Startup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Displaying Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Adjusting the Screen Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Entering System Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Entering a Saturation Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Setting Up the Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Setting Up the Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10

August 2004

Startup Procedure

Sensor configuration and system calibration data for your Series 3 are
entered at the factory. After you unpack the unit, use the sections that
follow to verify that your system is working properly and to set up the
screen to display the desired measurements.

Powering Up

The Series 3 has a universal power supply that automatically adjusts


to line voltages from 95-260 VAC. After making the electrical
connections as described in Chapter 2, Installation, press the [ON]
switch to power up the unit.
IMPORTANT: For compliance with the EUs Low Voltage Directive
(IEC 1010), this unit requires an external power
disconnect device such as a switch or circuit breaker.
The disconnect device must be marked as such, clearly
visible, directly accessible, and located within 1.8 m
(6 ft) of the Series 3. The power line is the main
disconnect device.
Caution!
Do not operate the Delta F Oxygen Cell for extended
periods of time at oxygen concentrations that are over
range. Trace and low percent range sensors may be
damaged if exposed to high levels of oxygen, such as
air, for long periods (>1 hour) while the Series 3 is on. If
exposure is unavoidable, either disconnect the oxygen
cell from the Series 3 or equip the sample system with a
valve that allows the cell to be switched to purge gas.

Using the Keypad and


Passcode

Use the following sections to become familiar with the key functions
and the passcode number used to access the user program.

Key Functions

The function keys to the right of the keypad are used as follows:

[MODE] - modifies or selects the measurement mode to display.

[CHAN] - toggles between channels (only works with units that have

two channels installed).

[PROG] - accesses the User Program.

[RUN] - exits the Modify Display Mode or the User Program

(except during numeric entry) and returns to displaying


measurements.
The keypad beneath the Series 3 screen consists of 16 keys, including
the [.] and [] symbols, two arrow keys, and two response keys, ([YES]
and [NO]). The numeric keys are for numeric entry only; however, the
arrow and response keys have more than one function.

Setup & Operation

3-1

August 2004

Key Functions (cont.)

The arrow keys perform three functions:

[W] or [X] - scroll through the screen selections by moving the


brackets forward or backward to the desired location.

[W] - moves backward and erases during numeric entry.

[X] - moves the cursor to the desired location during numeric entry
and accepts a change at the present location.

The response keys perform three functions:

[YES] or [NO] - respond to questions and/or exit a screen

[YES] - confirms an entry or retrieves a previous number

[NO] - moves backward through the menu

IMPORTANT: After pressing a key, wait for the unit to perform the
desired function before pressing another key.

Passcode

To enter the User Program you must enter a passcode. The passcode
prevents unauthorized users from changing data (the default
passcode for this unit is 2719).

Displaying
Measurements

The Series 3 begins displaying measurements in a two-line format


immediately after power-up. Due to display size limitations, the
Series 3 uses the abbreviations shown in Table 3-1 below to represent
the displayed measurement mode and units.
Table 3-1: DSP Modes
Display Abbreviation

Measurement Mode

Oxygen

Hygrometry

Temperature

Pressure

AUX1

Auxiliary 1

AUX2

Auxiliary 2

USER

User Function

VREF

Volt Reference

GND

Signal Ground

Proceed to the next section for instructions on selecting the


measurement mode for each display line.

3-2

Setup & Operation

August 2004

Displaying Measurement
Mode and Units

Press the [MODE] key to enter the Modify DSP mode.


Display Page: 1
Press YES to Edit

The Series 3 can display six


pages of measurements, each
showing two different
parameters. Use the numeric
keys to enter the desired page
and press [YES].

Modify DSP Mode


LINE1
[LINE2]

P1

Use the arrow keys to move the


brackets to the desired line and
press [YES].

Line 2 DSP Mode


[CH 1]
CH 2

P1

Use the arrow keys to move the


brackets to the desired channel
and press [YES].

Select DSP Mode

P1

Use the arrow keys to move the


brackets to the desired DSP
mode and press [YES].

_[H]

Aux1`

Note: See Table 3-1 on page 3-2 for the available DSP modes.
Select HYGRO Unit
[DP/C]

DP/F`

P1

Use the arrow keys to move the


brackets to the measurement
units and press [YES].

Note: Those measurement units listed in Table 3-2 below require


multiple inputs to obtain proper readings. If the Series 3
displays No Data, you may have to reconfigure the channel.
Table 3-2: Measurement Units Requiring Multiple Inputs
To measure:
You need the following inputs:
RH

Temperature and moisture

PPMv

Moisture and pressure

PPMw

Moisture, temperature and


saturation constant data

MCF/IG

Moisture and pressure

MCF/NG

Moisture and pressure

PPMv/NG

Moisture and pressure

For a list of all the available measurement units, see Table 3-3 on the
next page. Repeat the above procedure to set up the other line, and
when you are done press [RUN].

Setup & Operation

3-3

August 2004

Table 3-3: Measurement Modes and Units


Selected
Meas.
Mode
Oxygen

Hygrometry

Temperature

Pressure

Auxiliary 1

Description of Units

Displayed Displaye
Meas. Mode d Units

%= Percent Oxygen default

ppmv = Parts Per Million by volume

PPMv

A = Microamps (Diagnostic Mode)

DVM = Digital Voltmeter (Diagnostic Mode)

VDC

DP

F = Dew/Frost Point F

DP

K = Dew/Frost Point K (Kelvin)

DP

R.H. = Relative Humidity

H/ppMv = Parts per Million of Water by Volume

HPMv

H/ppMw = Parts per Million of Water by Weight (for liquids only)

ppMw

MCF/IG = Pounds of Water per Million Std. Cubic Feet in Ideal Gas

Ilbs

MCF/NG = Pounds of Water per Million Std. Cubic Feet in Natural Gas

Nlbs

ppMv/NG = Parts Per Million by Volume in Natural Gas

NPMv

Pw/mmHg = Vapor Pressure of Water in mmHg

mmHg

Pw/Pa = Vapor Pressure of Water in Pascals

Pa

MH = MH* (Diagnostic Mode)

MH

DVM = Digital Voltmeter (Diagnostic Mode)

VDC

C = Degrees Celsius default

F = Degrees Fahrenheit

K = Kelvin

DVM = Digital Voltmeter (Diagnostic Mode)

VDC

PSIg = Pounds per Square Inch Gauge default

PSIg

Bars = Bars

Bars

mbs = Millibars

mbs

mm/Hg = Millimeters of Mercury

mmHg

C = Dew/Frost Point default

Pas = Pascal

Pg

kPas = KiloPascal

KPg

pMV = Pressure in millivolts

mV

DVM = Digital Voltmeter (Diagnostic Mode)

VDC
VDC

VDC

X1

mA

X1

mA

Func

X1

none

VDC

X2

mV

mA

X2

mA

Func

X2

none

Volt Reference

VREF = Volts default (Diagnostic Mode)

VR

VDC

Signal Ground

GND = Volts default (Diagnostic Mode)

Auxiliary 2

*The MH value is the moisture sensors response value and is the value which is recorded during calibration.

3-4

Setup & Operation

August 2004

Menu Map

Figure 3-1 below is a menu map of the User Program options that are
discussed in this manual.
View system information: program code, serial and PCI numbers, uptime and
type of boot device.

SysInfo
RECORDERS

Set up recorder range, mode, unit, zero and span.

Set alarm mode, unit, trip type, setpoint and deadband.

ALARMS

Enter communications port settings.

PORT

View or set the clock and calendar.

CLK
BACKLIGHT

Set the backlight timer.

PROG
Enter passcode.

CONTRAST Increase or decrease the screen contrast.


BATTERY

TEST

Check or recharge the battery.

RECORDER Scale or trim recorder outputs .

ALARM

Trip or reset alarm relays.

LOGGER

Set up, view or stop a log, or store logged data.

USER

Enter the saturation constants,


or turn on/off Computer Enhanced Response.

AUTOCAL

Enter the AutoCal interval in hours and minutes.


ERROR

Set up error processing.

CONSTANT Set up system constants.


CURVES
SYSTEM

Enter input probe calibration data.

Measurement
Mode
REF

Enter reference values for a channel card.

CONFIG

Configure channel for a different sensor.

Figure 3-1: Series 3 User Program Menu Map

Setup & Operation

3-5

August 2004

Adjusting the Screen


Contrast

To adapt to various ambient lighting conditions, the Series 3 provides


a screen contrast adjustment. To make this adjustment, press the
[PROG] key to enter the user program:
Enter Passcode: XXXX

Enter the passcode.

Note: Refer to Figure 3-1 on page 3-5 to navigate to the Contrast


option.
Programming Menu

_ TEST

Use the arrow keys to move the


brackets to CONTRAST and press
[YES].

Use the arrow keys to move the


brackets to one of the options.

[CONTRAST]`

Contrast Control
[INCR] DECR

Press and hold [YES] until the required contrast is achieved.

Entering System
Constants

A system constant is a fixed value for moisture (Kh), temperature


(Kt), or pressure (Kp) entered into the Series 3 in lieu of a live
input. For example, if the process runs at a steady pressure, that
pressure is entered into the Series 3 memory rather than using a
pressure transmitter. In addition, you can multiply the ppmv value by
a constant for special purposes (the default multiplier is 1.000).
Note: You must use a current multiplier for oxygen if you are using a
different background gas than the one for which the cell was
calibrated (see page 4-7).
IMPORTANT: In order for the constant to work properly, you must
make sure you also configured the channel for a
constant (Kh, Kt or Kp) as described on page 4-9.
To enter a system constant, press the [PROG] key to enter the User
Program and proceed as follows:
Enter Passcode: XXXX

Enter the passcode.

Note: After you enter the user program, see Figure 3-1 on page 3-5
to navigate to the System menu.

3-6

Setup & Operation

August 2004

Entering System
Constants (cont.)

Be sure the number displayed in the upper right-hand corner of the


screen is the channel you want to program. If not, press the [CHAN]
key to select the desired channel.
Programming Menu

Measurement Mode
[H]

Use the arrow keys to move the


brackets to SYSTEM and press
[YES].

Use the arrow keys to move to


the desired measurement mode
(see Table 3-4 below) and press
[YES].

AUTOCAL`

[SYSTEM]

AUX1`

Table 3-4: Available Measurement Modes


Display Abbreviation

Measurement Mode

O*

Oxygen

Hygrometry

Temperature

Pressure

AUX1*

Auxiliary 1

AUX2*

Auxiliary 2

CNST-PPMV

PPMv Multiplication Factor

*System constants are not available for these measurement modes.

System Menu
CURVES

Use the arrow keys to move the


brackets to CONSTANT and press
[YES].

Enter the constant value and


press [YES].

[CONSTANT]`

Const. Dew Point C


+20.0

Note: The resulting screen display varies, depending on the


measurement mode selected.
When the Series 3 displays measurements that are using a constant, it
replaces the first letter of the mode on the display with K. For
example, the Series 3 will display DP as KP (dew point) or H as KH
(hygrometry).

Setup & Operation

3-7

August 2004

Entering a Saturation
Constant

The Series 3 requires a saturation constant to calculate parts per


million by weight (ppmw) in non-aqueous Henrys law type liquids.
To enter a saturation constant, the user must enter 1-6 data points of a
Cs (saturation constant) vs. temperature curve for each channel. To
accomplish this, press the [PROG] key and proceed as follows:
Enter Passcode: XXXX

Enter the passcode.

Note: After you enter the user program, see Figure 3-1 on page 3-5
to navigate to the User menu.
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the [CHAN]
key to select the desired channel.
Programming Menu

_[USER]

Use the arrow keys to move the


brackets to USER and press
[YES].

Use the arrow keys to move the


brackets to SAT-CONST and press
[YES].

LOGGER`

User Menu
[SAT-CONST] `

#
1

T(C)
+56

Cs
0

To enter data at the data point


displayed, use the right arrow
key to move the bracket to T/C.
Otherwise, enter the desired
data point number and press
[YES] to confirm the entry. Then
press the right arrow key.

#
1

T(C)
+56

Cs
33

Enter the desired temperature


and press [YES] to confirm the
entry. Press the right arrow key.

#
1

T(C)
+56

Cs
33

Enter the corresponding


saturation constant. Press [YES]
to confirm the entry.

Repeat the last three steps to enter each data point (1-6 points). Enter
the next data point number over the present data point number and
press the right arrow key. When you do this, T/C and Cs display
values of the current saturation constant curve. This indicates you are
ready to begin entering information for the selected data point.

3-8

Setup & Operation

August 2004

Setting Up the Recorders

The Series 3 has two recorder outputs (A and B) for each channel. To
set up these outputs, press the [PROG] key to enter the user program:
IMPORTANT: The switch blocks must be set to the appropriate
positions for current or voltage output (see page 2-16).
Enter Passcode: XXXX

Enter the passcode.

Note: After you enter the user program, see Figure 3-1 on page 3-5
to navigate to the Recorder menu.
Programming Menu

_[RECORDERS]`

Use the arrow keys to move the


brackets to RECORDER and press
[YES].

If necessary, press the [CHAN] key to display the desired channel in


the upper right-hand corner of the screen.
Select Recorders
[A] B

Use the arrow keys to move the


brackets to the desired recorder
and press [YES].

Select RCD Range

Use the arrow keys to move the


brackets to the desired recorder
range and press [YES].

Use the arrow keys to move the


brackets to the desired mode
(see Table 3-5 below) and press
[YES].

4-20mA `

[0-20mA]

Select Mode
O

[H]

Aux1`

Table 3-5: Available Measurement Modes


Display Abbreviation

Measurement Mode

Oxygen

Hygrometry

Temperature

Pressure

Aux 1

Auxiliary 1

Aux 2

Auxiliary 2

User*

User Function

Vref

Volt Reference

Gnd

Signal Ground

*User functions are available only through PanaView software.

Setup & Operation

3-9

August 2004

Setting Up the Recorders


(cont.)

Select Units
[DP/C]

1
DP/F`

Use arrow keys to move the


brackets to the desired units and
press [YES].

Note: The measurement choices and the available units vary


depending on the measurement mode selected.
Set RCD-A OUTPUT
[ZERO]
SPAN

Use the arrow keys to move the


brackets to ZERO and press
[YES].

RCD-A Zero Setup


20.0

Enter the zero value and press


[YES]. Then press [NO] to return
to the previous screen.

Repeat the above two steps to enter the span value.

Setting Up the Alarms

The Series 3 can be equipped with an optional high and low alarm
relay for each channel. To set up your alarms, press the [PROG] key
and proceed as follows:
Enter Passcode: XXXX

Enter the passcode.

Note: After you enter the user program, see Figure 3-1 on page 3-5
to navigate to the Alarms menu.
Programming Menu

_ [ALARMS] `

Use the arrow keys to move the


brackets to ALARMS and press
[YES].

If necessary, press the [CHAN] key to display the desired channel in


the upper right-hand corner of the screen.
Select Alarm
[A] B

Use the arrow keys to move the


brackets to the desired alarm
(A or B) and press [YES].

The Series 3 briefly flashes a message Relay Status -- Reset and then
displays the following prompt.
Select Mode
O

3-10

[H]

1
T

Aux1`

Use the arrow keys to move the


brackets to the desired mode
(see Table 3-6 on the next page)
and press [YES].

Setup & Operation

August 2004

Setting Up the Alarms


(cont.)

Table 3-6: Available Measurement Modes


Display Abbreviation

Measurement Mode

Oxygen

Hygrometry

Temperature

Pressure

Aux 1

Auxiliary 1

Aux 2

Auxiliary 2

User*

User Function

Vref

Volt Reference

Gnd

Signal Ground

*User functions are available only through PanaView software.

Select Unit
[DP/C]

DP/ F `

Use arrow keys to move the


brackets to the desired units and
press [YES].

Note: The measurement choices and the available units vary


depending on the measurement mode selected.
Select Trip Type
[Above] Below

Use the arrow keys to move the


brackets to the desired trip type
(Above or Below) and press
[YES].

A Setpoint
+0.0

DeadBD
+0.0

Use the numeric keys to enter


the Setpoint (the alarm trip
point) and press [YES].

A Setpoint
+0.0

DeadBD
+0.0

To enter a Deadband value,


press the right arrow key. Then
use the numeric keys to enter
the value and press [YES].

Note: The Deadband is a range of (normally small) values, beyond


the setpoint, within which the Series 3 will not trigger the
alarm. For a high alarm, the alarm will trip if the
measurement exceeds the setpoint, but it will not reset until
the measurement is less than the setpoint minus the deadband.
For a low alarm, the alarm will trip if the measurement is
below the setpoint, but it will not reset until the measurement
is greater than the setpoint plus the deadband.

Setup & Operation

3-11

Chapter 4

Calibration & Maintenance


Aluminum Oxide Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Delta F Oxygen Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Setting Up a New Probe or Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Setting Up a New Channel Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17

August 2004

Aluminum Oxide
Moisture Probes

The most common probe used with the Series 3 is the aluminum
oxide moisture probe. This section describes the procedures for
calibrating and maintaining the probe and its cable.

Probe Cable Calibration

Whenever a moisture probe cable is modified or replaced, the Series 3


must be calibrated to compensate for any offset introduced by the
cable. This procedure should also be performed after the initial
installation of either a factory-supplied or a use-supplied cable.
Before starting the cable calibration, do the following:

Power up the Series 3.

Set up the screen to display the MH values for each channel that
you will be calibrating.

Note the high, low and zero reference values that are recorded on
the sticker located on the Series 3 chassis.

Perform the calibration by completing the following steps:


1. Disconnect the cable from the moisture probe, but leave the cable
connected to the Series 3. Verify that the displayed MH value
equals the zero reference value within 0.0003 MH.

If the reading is within specification, no adjustment is required.

If the reading is greater than the zero reference value on sticker


0.0003, subtract this difference from the low reference value.

If the reading is less than the zero reference value on the sticker
0.0003, add this difference to the low reference value.

2. Record the final, corrected low reference value.


3. If a change was necessary, program the Series 3 with the new,
corrected low reference value.
4. Verify that the cable is not connected to the probe and verify that
the zero reference reading is now within 0.0003 MH.
5. Fill out a new high and low reference sticker with the final low
reference value. Make sure you record the information below:
HIGH REF = ORIGINAL VALUE
LOW REF = NEW CORRECTED VALUE
ZERO REF = ORIGINAL RECORDED VALUE
6. Reconnect the probe to the cable.
This completes the procedure. Be sure to repeat the cable calibration
whenever any changes are made to the cable.

Calibration & Maintenance

4-1

August 2004

Probe Cleaning

Other than periodic calibration checks, little or no routine moisture


probe maintenance is required. However, any electrically conductive
contaminant trapped on the aluminum oxide sensor will cause
inaccurate moisture measurements. If such a situation develops,
return of the moisture probe to the factory for analysis and
recalibration is recommended. In an emergency, cleaning of the
moisture probe in accordance with the following procedure may be
attempted by a qualified technician or chemist.
IMPORTANT: Moisture probes must be handled carefully and cannot
be cleaned in any fluid that will attack their
components. The probes materials of construction are
Al, Al2O3, nichrome, gold, stainless steel, glass and
Viton A. Also, the sensors aluminum sheet is very
fragile and can be easily bent or distorted. Do not
permit anything to touch it!
The following items are needed to properly complete the moisture
probe cleaning procedure:

approximately 300 ml of reagent grade hexane or toluene

approximately 300 ml of distilled (not deionized) water

two glass containers to hold above liquids (metal containers should


not be used).

To clean the moisture probe, complete the following steps:


1. Record the dew point of the ambient air.
2. Making sure not to touch the sensor, carefully remove the
protective shield from the sensor.
3. Soak the sensor in the distilled water for ten (10) minutes. Be sure
to avoid contact with the bottom and the walls of the container!
4. Remove the sensor from the distilled water and soak it in the clean
container of hexane or toluene for ten (10) minutes. Again, avoid
all contact with the bottom and the walls of the container!
5. Remove the sensor from the hexane or toluene, and place it face
up in a low temperature oven set at 50C 2C (122F 4F) for
24 hours.
6. Repeat steps 3-5 for the protective shield. During this process,
swirl the shield in the solvents to ensure the removal of any
contaminants that may have become embedded in the porous walls
of the shield.

4-2

Calibration & Maintenance

August 2004

Probe Cleaning (cont.)

After the cleaning steps have been completed, continue as follows:


1. Carefully replace the probes protective shield, making sure not to
touch the sensor.
2. Connect the probe cable to the probe, and record the dew point of
the ambient air, as in step 1 on the previous page. Compare the
two recorded dew point readings to determine if the reading after
cleaning is a more accurate dew point value.
3. If the sensor is in proper calibration (2C accuracy), reinstall the
probe in the sample cell and proceed with normal operation of the
hygrometer.
4. If the sensor is not in proper calibration, repeat the cleaning steps,
using time intervals 5 times those used in the previous cleaning
cycle. Repeat this procedure until the sensor is in proper
calibration.
A trained laboratory technician should determine if all electrically
conductive compounds have been removed from the aluminum oxide
sensor and that the probe is properly calibrated. Probes which are not
in proper calibration must be recalibrated. It is recommended that all
moisture probes be recalibrated by GE Infrastructure Sensing, Inc.
approximately once a year, regardless of the probes condition.

Delta F Oxygen Cells

The Delta F oxygen cell is a very reliable sensor, but it can only
provide accurate results if it is properly maintained. This section
discusses the necessary procedures to maintain the cell. Refer to
Figure 4-1 below and proceed to the next page for instructions

Level Indicator
Ma x

Mi n

Figure 4-1: Delta F Oxygen Cell Electrolyte Level

Calibration & Maintenance

4-3

August 2004

Electrolyte Maintenance

Over time, particularly when monitoring dry gases, there will be a


gradual loss of water from the electrolyte in the Delta F oxygen cell.
Thus, the electrolyte level should be checked at regular intervals and
adjusted if necessary.
Note: Some applications require that the electrolyte be completely
changed periodically (consult the factory for details).
!WARNING!
The electrolyte contains a strong caustic ingredient that is
harmful if it comes in contact with skin or eyes. Follow
proper procedures for handling the caustic (potassium
hydroxide) solution. Consult your company safety
personnel for recommendations.
Using the min/max window on the oxygen cell (see Figure 4-1 on the
previous page), check to be sure the electrolyte covers about 60% of
the window. If it is necessary to raise the fluid level in the reservoir,
add DISTILLED WATER slowly in small amounts. Check the level as
you add the distilled water, making sure you do not overfill the
reservoir.

Calibrating the Delta F


Oxygen Cell

The Delta F oxygen cell should be calibrated upon initial installation


and about once a month thereafter for the first three months.
Additional calibrations should be performed as needed and whenever
the electrolyte is changed. The process includes:

checking the calibration

Your system must have a calibration gas inlet valve.

You must know the PPMv value of the calibration gas.

entering the new span value

GE Infrastructure Sensing, Inc. recommends that the calibration


gas value be 80-100% of the span of the sensors overall range
(e.g., 80-100 PPMv oxygen for a 0-100 PPMv oxygen sensor).

GE Infrastructure Sensing, Inc. recommends that the calibration


gas be in a background of nitrogen.

Note: If your measurement range is significantly lower than your


span gas value, you may elect to directly input the PPMv
oxygen content of the span gas and the measured A value as
an alternative to the following procedure.

4-4

Calibration & Maintenance

August 2004

Checking the Oxygen Cell


Calibration

To check the oxygen cell calibration, complete the following steps:


1. Determine which channel is connected to the Delta F oxygen cell
2. Set up the display to read the oxygen content in PPMv and A.
3. Start the flow of calibration gas through the oxygen cell.
4. Read the PPMv value. If it is correct, your oxygen cell is properly
calibrated. However, if the reading is incorrect, you must calculate
the new span reading (x) as follows:
( OX 1 OX c ) ( IO c IO 0 )
x = IO c + -------------------------------------------------------------( OX OX )
c

where,

OXc = correct PPMv for the calibration gas


OX0 = zero value in PPMv*
OX1 = span value in PPMv*
Oc = actual reading for calibration gas in A
O0 = zero value in A*
x = new span reading in A

*See the Calibration Data Sheet for the oxygen cell to obtain the
applicable zero and span values.
Example: If the calibration data for your cell is as follows:
OXc = 75 PPMv
OX0 = 0.050 PPMv
OX1 = 100 PPMv
IOc = 290 A
O0 = 0.4238 A
Then,

( 100 75 ) ( 290 0.4238 )


x = 290 + -------------------------------------------------------------( 75 0.05 )

The new span value (x) is 100 PPMv 387 A. Enter the new value
as described in the next section.

Calibration & Maintenance

4-5

August 2004

Entering the New Span


Value

To enter the new span value calculated on the previous page, press the
[PROG] key and proceed as follows:
Enter Passcode: XXXX

Enter the passcode.

Be sure the number displayed in the upper right-hand corner of the


screen is the channel you want to program. If not, press the [CHAN]
key to select the desired channel.
Programming Menu

Use the arrow keys to move the


brackets to SYSTEM and press
[YES].

Use the arrow keys to move the


brackets to O and press [YES].

Use the arrow keys to move the


brackets to CURVES and press
[YES].

O2 Curve Menu
S/N [CURVE] BkGd

Use the arrow keys to move the


brackets to CURVE and press
[YES].

O2 Curve Menu
S/N [CURVE] BkGd

Use the arrow keys to move the


brackets to SPAN and press
[YES].

#1

O(ua) O(%)
0.721 0.0000

Enter the new span percentage


value and press [YES]. Then
press the left arrow key.

#1

O(ua) O(%)
0.721 0.0000

Enter the new span microamp


value and press [YES].

AUTOCAL`

[SYSTEM]

Measurement Mode
[O]

P Aux1`

System Menu

_ [CURVES]

CONSTANT`

To exit the User Program, press the [RUN] key.

4-6

Calibration & Maintenance

August 2004

Entering Background Gas


Calibration Data

The factory calibration for Delta F oxygen cells uses nitrogen as the
reference background gas, and the Series 3 will yield inaccurate
oxygen measurements if a background gas other than nitrogen is
used. However, you may use a background gas other than nitrogen if
you first recalibrate the Series 3 for the desired background gas.
To recalibrate your Series 3 for a different background gas, look up
the current multiplier for your background gas in Table 4-1 below and
proceed to the next page to enter this multiplier into your meter.

Table 4-1: Background Gas Current Multipliers


Current Multiplier*
Background Gas

1000 PPM

5000-10,000 PPM

2.5% to 10%

25%

Argon (Ar)

0.97

0.96

0.95

0.98

Hydrogen (H2)

1.64

1.96

2.38

1.35

Helium (He)

1.72

2.13

2.70

1.39

Methane (CH4)

1.08

1.09

1.11

1.05

Ethane (C2H6)

0.87

0.84

0.81

0.91

Propylene (C3H6)

0.91

0.88

0.87

0.93

Propane (C3H8)

0.79

0.76

0.72

0.58

Butene (C4H8)

0.69

0.65

0.60

0.77

Butane (C4H10)

0.68

0.63

0.58

0.76

Butadiene (C6H6)

0.71

0.66

0.62

0.79

Acetylene (C2H2)

0.95

0.94

0.93

0.97

Hexane (C6H14)

0.57

0.52

0.89

0.67

Cyclohexane (C6H12)

0.64

0.58

0.54

0.72

Vinyl Chloride (CH2CHCl)

0.74

0.69

0.65

0.81

Vinylidene Chloride (C2H2F2)

0.77

0.73

0.69

0.83

Neon (Ne)

1.18

1.23

1.28

1.11

Xenon (Xe)

0.70

0.65

0.61

0.78

Krypton (Kr)

0.83

0.79

0.76

0.88

Sulfur Hexaflouride (SF6)

0.54

0.49

0.44

0.64

Freon 318 (C4F8)

0.39

0.34

0.30

0.49

Tetrafluoromethane (CF4)

0.62

0.57

0.52

0.71

Carbon Monoxide (CO)

0.99

0.99

0.98

0.99

* To use the current multipliers in this table, your Calibration Data Sheet must contain calibration
data for nitrogen. If your Calibration Data Sheet contains calibration data for a background gas other
than nitrogen, contact the factory for the nitrogen calibration data.

Calibration & Maintenance

4-7

August 2004

Entering Background Gas


Calibration Data (cont.)

The oxygen cell factory calibration values, referenced to nitrogen, are


stored in the Series 3 User Program and a default current multiplier
of 1.00 is applied to these values. To enter the new current multiplier,
as determined on the previous page, press the [PROG] key and proceed
as follows:
Enter Passcode: XXXX

Enter the passcode.

Be sure the number displayed in the upper right-hand corner of the


screen is the channel you want to program. If not, press the [CHAN]
key to select the desired channel.
Programming Menu

Use the arrow keys to move the


brackets to SYSTEM and press
[YES].

Use the arrow keys to move the


brackets to O and press [YES].

Use the arrow keys to move the


brackets to CURVES and press
[YES].

O2 Curve Menu
S/N CURVE [BkGd]

Use the arrow keys to move the


brackets to BkGd and press [YES].

O2 uA Multiplier
1.00

Use the numeric keys to enter


the Current Multiplier. Press [YES]
to confirm your entry.

AUTOCAL`

[SYSTEM]

Measurement Mode
[O]

P Aux1`

System Menu

_ [CURVES]

CONSTANT`

To exit the User Program, press the [RUN] key.

4-8

Calibration & Maintenance

August 2004

Setting Up a New Probe


or Sensor

Whenever you connect a new probe or sensor to the Series 3 or move


an existing probe or sensor to a different channel, the meter must be
reconfigured. This section provides the necessary instructions.

Reconfiguring a Channel
for a New Sensor

If you decide to connect a different type of sensor to one of the


channels of your unit, you must reconfigure that channel. To do so,
press the [PROG] key and complete the following steps:
Note: If you are using a probe capable of multiple inputs or using
more than one input (including a constant), you must
configure the channel to measure all the possible inputs (see
the Multiple Input Instructions on page 4-11).
Enter Passcode: XXXX

Enter the passcode.

Be sure the number displayed in the upper right-hand corner of the


screen is the channel you want to program. If not, press the [CHAN]
key to select the desired channel.
Programming Menu

Measurement Mode
H

Use the arrow keys to move the


brackets to SYSTEM and press
[YES].

Select the desired measurement


mode and press [YES]. (See
Table 4-2 below for the
available measurement modes).

AUTOCAL`

[SYSTEM]

[O]

P Aux1`

Table 4-2: Available Measurement Modes


Display Abbreviation
Measurement Mode

Calibration & Maintenance

Oxygen

Hygrometry

Temperature

Pressure

AUX1

Auxiliary 1

AUX2

Auxiliary 2

CNST-PPMV

PPMv Multiplication Factor

4-9

August 2004

Reconfiguring a Channel
for a New Sensor (cont.)

System Menu
[CONFIG]

1
REF `

Use the arrow keys to move the


brackets to CONFIG and press
[YES].

Note: The word HYGRO in the nest display is just an example. Your
actual sensor type will be shown instead.
Sel. HYGRO sensor
---

TF

[Mxx]`

Use the arrow keys to move the


brackets to desired
configuration and press [YES]
(see Table 4-3 below).

Table 4-3: Probe Types and Configuration Options


Probe Type

Configurations

Oxygen

N/A
% (percent)
PPMv (parts per million by volume)

Hygrometry

N/A
TF (Three-Function probe)
Mxx (M Series probe)
Kh (constant dew point)

Temperature

N/A
TF (Three-Function probe)
Mxx (M-Series probe)
Kt (constant temperature)

Pressure

N/A
TF (Three-Function probe)
X1 (Auxiliary 1)
X2 (Auxiliary 2)
Kp (constant pressure)

Auxiliary 1

N/A
I (current)
V (voltage)

Auxiliary 2

N/A
I (current)
V (voltage)

IMPORTANT: Be sure to enter the calibration data for the new probe/
sensor after configuring the channel (see page 4-12).

4-10

Calibration & Maintenance

August 2004

Multiple Input Instructions

Table 4-4 below shows a list of measurement modes that require


multiple inputs and what you need to display them. For example, if
you are using a TF Probe, you must configure its channel for
moisture, temperature and pressure. To do this, you must select TF
under the CONFIG option for the H, T and P measurement modes.
Note: When activating pressure sensors, you must select the input
(Aux 1 or Aux 2) used for the pressure sensor. After selecting
the input, the display will prompt you to enter the probe signal
in either current (I) or volts (V).

Table 4-4: Measurement Modes, Required Inputs and Configuration Information


To measure: Use these inputs:
Provided by:
Configure the channel for:
RH

PPMv

PPMw

MCF/IG

MCF/NG

PPMv/NG

temperature
and moisture

moisture
and pressure

moisture,
temperature
and
saturation constant

moisture
and pressure

moisture
and pressure

moisture
and pressure

Calibration & Maintenance

TF Probe

TF (H, T, & P)

M Series Probe w/ temp. opt.

Mxx (H) and Kt (T)

M Series Probe &


constant temperature

Mxx (H & T)

TF Probe w/ pressure option

TF (H, T & P)

M Series Probe &


aux pressure input

Mxx (H) and X1 (P)

M Series Probe &


pressure constant

Mxx (H) and Kp (P)

TF Probe w/ temp. option


and sat. constant

TF (H & T) and sat. const.

M Series Probe w/ temp.


option & sat. constant

Mxx (H & T) and sat. const.

TF Probe, constant temp. &


sat. constant

TF (H), Kt (T) and sat. const.

M Series Probe, constant


temp. & sat. constant

M (H), Kt (T) and sat. const.

TF Probe w/pressure option

TF (H, T & P)

M Series Probe &


aux. pressure input

Mxx (H) and X1 (P)

M Series Probe &


pressure constant

Mxx (H) and Kp (P)

lTF Probe w/pressure option

TF (H, T & P)

M Series Probe &


aux. pressure input

Mxx (H) and X1 (P)

M Series Probe &


pressure constant

Mxx (H) and Kp (P)

TF Probe w/pressure option

TF (H, T & P)

M Series Probe &


aux. pressure input

Mxx (H) and X1 (P)

M Series Probe &


pressure constant

Mxx (H) and Kp (P)

4-11

August 2004

Entering Calibration Data


for New Probes/Sensors

For a new probe or sensor to function properly, you must configure


the channel AND enter the calibration data. This data, which is a list
of 1 to 16 data points, is found on the Calibration Data Sheet that is
supplied with each probe or sensor. To enter the calibration data, press
the [PROG] key and complete the following steps:
Note: If you choose to enter a constant value for moisture,
temperature, pressure, or PPMv multiplier, it is not necessary
to enter the calibration data for the sensor.
Enter Passcode: XXXX

Enter the passcode.

Be sure the number displayed in the upper right-hand corner of the


screen is the channel you want to program. If not, press the [CHAN]
key to select the desired channel.
Programming Menu

Measurement Mode
H

Use the arrow keys to move the


brackets to the desired
measurement mode and press
[YES]. (See Table 4-2 on page
4-9 for the available modes).

Use the arrow keys to move the


brackets to CURVES and press
[YES].

P Aux1`

System Menu
[CURVES]

Use the arrow keys to move the


brackets to SYSTEM and press
[YES].

AUTOCAL`

[SYSTEM]

[O]

CONSTANT`

The remaining prompts depend on the type of sensor you are using.
Refer to one of the following sections to complete the data entry:

Entering Moisture Curve Data on page 4-13

Entering Oxygen Curve Data on page 4-14

Entering Pressure Curve Data on page 4-15

Entering Auxiliary Inputs Curve Data on page 4-16

Note: You do not have to enter data for a temperature curve.

4-12

Calibration & Maintenance

August 2004

Entering Moisture Curve


Data

1MH Curve Menu


[S/N]
#PTS CURVE

Probe Serial #
H: 123456

Use the arrow keys to move the


brackets to S/N and press [YES].

Enter the serial number and


press [YES]. Then press [NO] to
return to the Curve menu.

1MH Curve Menu


S/N
[#PTS] CURVE

Use the arrow keys to move the


brackets to #PTS and press [YES].

1Enter Total #PTS


14

Enter the total number of data


points (1 to 16) and press [YES].
Then press [NO] to return to the
Curve menu.

1MH Curve Menu


S/N
#PTS
[CURVE]

Use the arrow keys to move the


brackets to CURVE and press
[YES].

Note: The data points on the Calibration Data Sheet begin


numbering at 0 (zero). Thus, the Series 3 displays the selected
number of data points you entered minus 1.
# DP/C
12 +0.0

MH
0.0000

To begin entering data at the


point displayed, press the right
arrow key to move the bracket
to DP/C. Otherwise, enter the
desired data point number, press
[YES], and then press the right
arrow key.

#
12

DP/C
+20.0

MH
0.0000

Enter the dew point. Then press


[YES] and press the right arrow
key.

#
12

DP/C
+20.0

MH
1.046

Enter the corresponding MH


value. Then press [YES] and
press the right arrow key.

Repeat the last three steps until all of the data points are entered.

Calibration & Maintenance

4-13

August 2004

Entering Oxygen Curve


Data

4-14

O2 Curve Menu
[S/N]
CURVE

Use the arrow keys to move the


brackets to S/N and press [YES].

Probe Serial #
O2: 123456

Enter the serial number and


press [YES]. Then press [NO] to
return to the Curve menu.

O2 Curve Menu
S/N
[CURVE]

Use the arrow keys to move the


brackets to CURVE and press
[YES].

BkGd

BkGd

Sel. O2 Curve Pts#


[ZERO] SPAN

Use the arrow keys to move the


brackets to ZERO and press
[YES].

#1

O(uA)
0.721

O(ppm)
0.0000

Enter the zero microamp value


(IO0). Press [YES] and then press
the right arrow key.

#1

O(uA)
0.721

O(ppm)
0.5000

Enter the zero ppmv or % value


(OX0) and press [YES]. Then
press [NO].

Sel. O2 Curve Pts#


ZERO
[SPAN]

Use the arrow keys to move the


brackets to SPAN and press
[YES].

#1

O(uA)
0.721

O(ppm)
0.0000

Enter the span microamp value


(IO1). Press [YES] and then press
the right arrow key.

#1

O(uA)
0.721

O(ppm)
0.5000

Enter the zero ppmv or % value


(OX1) and press [YES]. Then
press [NO].

Calibration & Maintenance

August 2004

Entering Pressure Curve


Data

O2 Curve Menu
[S/N]
CURVE

Use the arrow keys to move the


brackets to S/N and press [YES].

Probe Serial #
O2: 123456

Enter the serial number and


press [YES]. Then press [NO] to
return to the Curve menu.

Pressure Curve Menu


S/N
[CURVE]

Calibration & Maintenance

Use the arrow keys to move the


brackets to CURVE and press
[YES].

#
1

mA
Psig
+0.000 0.000

Making sure the cursor is


positioned below the # sign,
enter a 1 to indicate the zero
point. Press [YES] and then press
the right arrow key.

#
1

mA
Psig
+1.000 0.000

Enter the mA value. Press [YES]


and press the right arrow key.

#
1

mA
Psig
+1.000 2.000

Enter the psig value. Then,


press [YES].

#
2

mA
Psig
+0.000 0.000

Making sure the cursor is


positioned below the # sign,
enter a 2 to indicate the span
point. Press [YES] and then press
the right arrow key.

#
2

mA
Psig
+4.000 0.000

Enter the mA value. Press [YES]


and press the right arrow key.

#
2

mA
Psig
+4.000 8.000

Enter the psig value. Then,


press [YES].

4-15

August 2004

Entering Auxiliary Input


Curve Data

Aux Curve Menu


[AUX1]
AUX2

Use the arrow keys to move the


brackets to AUX1 (or AUX2),
depending on where the sensor
is connected, and press [YES].

Aux Curve Menu


[#PTS]
CURVE

Use the arrow keys to move the


brackets to #PTS and press [YES].

1Enter Total #PTS


2

Enter the number of data points


and press [YES]. Then press [NO]
to return to the Curve menu.

Note: For linear devices only two data points are required, but up to
eight data points may be entered for other devices. Each data
point requires a current or voltage value with a corresponding
scale value in the desired output units.
Aux Curve Menu
#PTS
[CURVE]

Use the arrow keys to move the


brackets to CURVE and press
[YES].

#
2

mA/VDC
+0.000

Units
0.000

To begin entering data at the


data point displayed, press the
right arrow key to move the
bracket to X1 (mA or V).
Otherwise, enter the desired
data point number. Press [YES]
and press the right arrow key.

#
2

mA/VDC
+1.000

Units
0.000

Enter the mA/VDC value. Press


[YES] and press the right arrow
key.

#
2

mA/VDC
+1.000

Units
20.000

Enter the Units value and press


[YES].

Repeat the last three steps until all of the remaining data has been
entered.

4-16

Calibration & Maintenance

August 2004

Setting Up a New
Channel Card

If you replace a channel card, you will have to re-enter the high and
low reference values for moisture, oxygen, and pressure. The
reference values are unit-specific factory calibration values that are
located on a label placed on the left side of the Series 3 chassis. Enter
the new reference values from the label or tag attached to the new
channel card by pressing the [PROG] key and proceeding as follows:
Enter Passcode: XXXX

Enter the passcode.

IMPORTANT: Be sure the number displayed in the upper right-hand


corner of the screen is the channel you want to
program. If not, press the [CHAN] key to select the
desired channel.
Programming Menu

Measurement Mode
[H]

Use the arrow keys to move the


brackets to the desired
measurement mode and press
[YES]. (See Table 4-2 on page
4-9 for the available modes).

Use the arrow keys to move the


brackets to REF and press [YES].

P Aux1`

System Menu
CONFIG

Use the arrow keys to move the


brackets to SYSTEM and press
[YES].

AUTOCAL`

[SYSTEM]

[REF]`

The remaining prompts depend on the measurement mode you


selected. Refer to one of the following sections on the next page to
properly program your unit:
Note: You do not have to enter reference data for temperature,
auxiliary 1, auxiliary 2, or constant ppmv.

Entering Moisture Reference Data

Entering Oxygen Reference Data

Entering Pressure Reference Data

Note: The reference values shown on the following page are only
examples. You must enter the actual reference values supplied
with the new channel card.

Calibration & Maintenance

4-17

August 2004

Entering Moisture
Reference Data

MH

Hi Ref
0.1660

Lo Ref
0.0000

Enter the low reference value.


Press [YES] and then press the
left arrow key.

MH

Hi Ref
0.1660

Lo Ref
2.9335

Enter the high reference value


and press [YES].

Oxygen Ref Menu


[LOW]
HIGH

Use the arrow keys to move the


brackets to LOW and press [YES].

Lo O2 Zero
Span
+0.0499 +0.0000

Enter the low oxygen zero


value. Press [YES] and then press
the right arrow key.

Lo O2 Zero
Span
+0.0499 +1.9923

Enter the low oxygen span


value. Press [YES] and then press
the [NO] key.

Oxygen Ref Menu


LOW
[HIGH]

Use the arrow keys to move the


brackets to HIGH and press [YES]

Hi O2 Zero
Span
+0.0499 +0.0000

Enter the high oxygen zero


value. Press [YES] and then press
the right arrow key.

Hi O2

Enter the high oxygen span


value. Press [YES] and then press
the [NO] key.

Entering Oxygen
Reference Data

Entering Pressure
Reference Data

4-18

Zero
Span
+0.0499 +1.9923

Hi Ref
0.05

Lo Ref
0.00

Enter the low pressure value.


Press [YES] and then press the
left arrow key.

Hi Ref
0.05

Lo Ref
99.89

Enter the high pressure value


and press [YES].

Calibration & Maintenance

GE Industrial
Sensing

DECLARATION
OF
CONFORMITY
Panametrics Limited
Shannon Industrial Estate
Shannon, County Clare
Ireland

We,

declare under our sole responsibility that the

Moisture Image Series 1 Analyzer


Moisture Image Series 2 Analyzer
Moisture Monitor Series 3 Analyzer
to which this declaration relates, are in conformity with the following standards:
EN 50014:1997+A1+A2:1999
EN 50020:1994
II (1) G [EEx ia] IIC
BAS01ATEX7097
Baseefa (2001) Ltd/EECS, Buxton SK17 9JN, UK
EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation
EN 61010-1:1993+A2:1995, Overvoltage Category II, Pollution Degree 2
following the provisions of the 89/336/EEC EMC Directive, the 73/23/EEC Low Voltage Directive and the 94/9/EC ATEX
Directive.
The units listed above and any sensors and ancillary sample handling systems supplied with them do not bear CE
marking for the Pressure Equipment Directive, as they are supplied in accordance with Article 3, Section 3 (sound
engineering practices and codes of good workmanship) of the Pressure Equipment Directive 97/23/EC for DN<25.

Shannon - July 1, 2003


Mr. James Gibson
GENERAL MANAGER

TV

TV ESSEN
ISO 9001
U.S.

CERT-DOC-H2

August 2004

GE Industrial
Sensing

ATEX COMPLIANCE

GE Infrastructure Sensing, Inc.


1100 Technology Park Drive
Billerica, MA 01821-4111
U.S.A.

We,

as the manufacturer, declare under our sole responsibility that the product

Moisture Monitor Series 3 Analyzer


to which this document relates, in accordance with the provisions of ATEX Directive 94/9/EC Annex II, meets the
following specifications:
II 1 G EEx ia IIC (-20C to +50C)
1180

BAS01ATEX7097

Furthermore, the following additional requirements and specifications apply to the product:
Having been designed in accordance with EN 50014 and EN 50020, the product meets the fault tolerance
requirements of electrical apparatus for category ia.
The product is an electrical apparatus and must be installed in the hazardous area in accordance with the
requirements of the EC Type Examination Certificate. The installation must be carried out in accordance with all
appropriate international, national and local standard codes and practices and site regulations for flameproof
apparatus and in accordance with the instructions contained in the manual. Access to the circuitry must not be
made during operation.
Only trained, competent personnel may install, operate and maintain the equipment.
The product has been designed so that the protection afforded will not be reduced due to the effects of corrosion
of materials, electrical conductivity, impact strength, aging resistance or the effects of temperature variations.
The product cannot be repaired by the user; it must be replaced by an equivalent certified product. Repairs
should only be carried out by the manufacturer or by an approved repairer.
The product must not be subjected to mechanical or thermal stresses in excess of those permitted in the
certification documentation and the instruction manual.
The product contains no exposed parts which produce surface temperature infrared, electromagnetic ionizing, or
non-electrical dangers.

CERT-ATEX-D (Rev. August 2004)

USA
1100 Technology Park Drive
Billerica, MA 01821-4111
Web: www.gesensing.com
Ireland
Shannon Industrial Estate
Shannon, County Clare
Ireland

You might also like