BMD-M1

Single Channel Por table Coagulometer

Single Channel Optical Detection System


Reduce Volumes up to One Quarter
Onboard Computer with an Interactive Program
Automatic INR Calculation
User-Selectable Program Options
Adaptable to Most Standard and Non-Standard Clotting Assays
PC Connectivity
Optional External Printer
CSA, UL, TV-GS, CE-Mark Approved

BioMedica Diagnostics Inc.


Affordable Diagnostic Solutions

BMD-M1

Single Channel Por table Coagulometer


The BMD-M1 is ideal for small and medium size laboratories, blood banks, intensive care
units and emergency services. Its optical detection system uses only one-quarter of the
volume of sample and reagents generally required by other analyzers.
The coefficients of variation obtained are lower than 2 to 5%, depending on the test and
reagents used. A new double micro-cuvette allows for effective and rapid processing of a
larger number of samples. Automatic amplification has been especially designed for new
sensitive thromboplastin reagents and equalizes high differences of turbidity from reagents
and samples. All specific test parameters are preset in its memory and its interactive program
can be adjusted according to the preferences of the operator.
The measurement can be started either manually when the reagent is added or with a connected autopipette and electronic start.
The results are displayed in the measured dimension and can be converted into other desired units with the press of a button.
A connected printer will automatically print out the results in all dimensions with optic channel, patient number and test name.
Computer connectivity allows measuring curve data to be transmitted to use in spreadsheet or report programs. Software updates
can be easily applied to the instrument from a personal computer.

Specications
Volume Usage

Measure all clotting based Coagulation tests with a quarter of the usual volumes.
e.g. PT: 25 l sample + 50 l reagent

Tests

PT, APTT, TT, FIB, Fa-I (Factor Intrinsic), Fa-E (Factor Extrinsic)

Selectable Dimensions

s, INR for PT, U (units) g/L for Fibrinogen, % for Factors

Standard Package

1 Pc Coatron M1
1 Pc Power Supply
25 Pcs Single cuvette
5 Pcs Reagent tubes
2 Pcs Reagent adaptors (pre-mounted)
1 Pc Operation Manual
1 Pc Warranty card

Incubation Block

1 Detection position, 6 Incubation wells


Temperature: 37.0 C 0.2 C

Optic Block

1 Channel Optics

Device

SMD Board, Microcontroller, 16-bit Analog-Digital Converter, Memory: 128 Kb

Ambient Temperature

From 18 C to 23 C

Display

2 lines x 16 Characters LCD

Keyboard

Foil keyboard, 3 x 3 matrix with Test, Functions and Numerical keys

Printer
(Optional)

External, serial thermal printer with print-out of the results in all selected dimensions and print-out of
system parameters

Dimensions

13 x 24.5 x 6 cm (5 x 9.5 x 2.5 in) (w/d/h)

Weight

0.51 kg (1.1 lbs) (w/o Power supply)

Power Supply

Input: 100-240 V AC, 50/60 Hz


Power: 12 W

Certications

Approved by the Canadian Standards Association (CSA), the US Underwriters Laboratories (UL), and
the European Union (CE-Mark and TV-GS)