This document provides an operations manual for the Serospin centrifuge. It describes the centrifuge specifications including maximum speed and capacity. It outlines safety devices and cautions regarding imbalance and overspeed. It details installation requirements regarding location, balance, and electrical needs. It describes the control panel and buttons. It provides step-by-step instructions for standard and programmed operation. It also includes maintenance instructions regarding cleaning and decontamination.

Uploaded by

Verdana
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
50% found this document useful (2 votes)
697 views19 pages

Serospin Operating Manual 20100407

This document provides an operations manual for the Serospin centrifuge. It describes the centrifuge specifications including maximum speed and capacity. It outlines safety devices and cautions regarding imbalance and overspeed. It details installation requirements regarding location, balance, and electrical needs. It describes the control panel and buttons. It provides step-by-step instructions for standard and programmed operation. It also includes maintenance instructions regarding cleaning and decontamination.

Uploaded by

Verdana
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

2

Serospin

Operations Manual

Contents

1. Introduction
-

Specifications
Caution
Safety Devices

Location
Balance
Electrical Requirements

2. Note for Installation

3. Parts Description
- Control Panel

4. Operation Procedure
-

Standard Operation
Programmed Operation

5. Maintenance

Page 2

Hanil Science Industrial Co., Ltd.

Serospin

Operations Manual

Serospin
1. INTRODUCTION
- SPECIFICATIONS Serospin
Max. Speed

3400 rpm

Max. RCF

1,020 x g

Max. Capacity

75ml x 12 slide

Timer

99 minutes 59seconds

Acceleration/Deceleration

3 steps (Positive Action Break System)

Programs
Display

10 memory
Digital type
Speed, Time, Program, Breaking Steps

Power requirement

Single Phase, 220V/50, 60Hz

Dimension(W x D x H)

270W x 355D x 280H mm

Weight

11 kg

Page 3

Hanil Science Industrial Co., Ltd.

Serospin

Operations Manual

- CAUTION Be sure that Rotor speed is not more than maximum

OVERSPEED

rotation speed. Dont be over the maximum speed of


rotor.
When the rotor is given to relative centrifugal force
over the allowed seal intensity, the destruction of rotor
occurs because the shape of rotor is designed to be able
to stand an external force in accordance with the
allowed seal intensity of the rotor.

IMBALANCE
OF CYTO-UNIT

Put samples measured exactly into tube each and load


tubes symmetrically each other in the rotor. If the
volume of opposite sample is different, the serious
turbulence occurs during rotation and a motor, rotor
and shaft are damaged.

When you put the tubes on rotor, those must be loaded


symmetrically

Page 4

Hanil Science Industrial Co., Ltd.

Serospin

Operations Manual

If the tubes are not symmetrical in the


.

number of tubes load the another tube


having as same as others weight.

Page 5

Hanil Science Industrial Co., Ltd.

Serospin

Operations Manual

- NOTE FOR SAFETY

POWER ON/OFF

An automatic circuit breaker, for emergency situations, such as


power surges, which could damages the unit, protects instruments
circulation when it is overpowered.

DOOR
When a door opens, the door limit switch by sensor makes the
rotor be still.

IMBALANCE
& VIBRATION

Page 6

If the rotor is operated with imbalance over standard during


rotation, motor is also moved. In this case, the danger is detected
by measuring the vibration of motor. With a alarm imbalance
LED lit up and the rotation stops by preset deceleration time.
Safety device as above keeps the instrument from an accident
during operators absence in lab.

Hanil Science Industrial Co., Ltd.

Serospin

Operations Manual

2. NOTE FOR INSTALLATION


Location
BEST LOCATION
The rotating instrument should be set on the flat and solid
surface. In the case of setting the instrument at the incline
surface, it is possible the shaft become bent by the heavy
weight rotor because of a long time of rotation in inclining of
the shaft and ground.

AIR CIRCULATION
For the circulation of air, the space of at least 30 cm must be
needed from the centrifuge during operation. Avoid to set it
on the dusty place.

TEMPERATURE, HUMIDITY
The centrifuge which is controlled by a high-tech
microprocessor is affected by the external conditions such as
temperature or humidity. If a room temperature is extremely
high by the direct ray of light or heater, or very low, the
accuracy and reliance of instrument are reduced by errors of
electronic parts. On top of that the high humidity makes the
corrosion of rotor or parts. A proper temperature and
humidity should be maintained accordingly.

Page 7

Hanil Science Industrial Co., Ltd.

Serospin

THE PLACE NOT


TO OCCUR
CORROSIVE GAS

Operations Manual

Place the centrifuge at the place which the corrosive gas cant
occur. If there is sulfur dioxide or chlorine gas in atmosphere, it
causes a corrosion of rotor and shaft, and cause great damage of
metals.

BALANCE
BALANCING
The shaft has to be perpendicular to the ground.

ELECTRICAL
REQUIREMENTS
P O W E R
The standard power requirements are a single phase and
220V but those are changeable depending on Users request.
Therefore, plug in after checking the name plate located on
the rear panel. Be sure not to be three phases.

Page 8

Hanil Science Industrial Co., Ltd.

Serospin

Operations Manual

3. PARTS DESCRIPTION
- CONTROL PANELS

SPEED SETTING
REPRESENTED SPEED IN RPM
AND g-FORCE IN RCF.

PROGRAM RECALL
/ SAVE

TIME
DISPLAYED AND SET UP TO 99 MINS

INPUT TIME, SPEED AND ETC,


SAVE AND RECALL FOR THE

AND IT IS
CONTINUOUS OPERATION WHEN YOU

REPEATING USE

SELECT0

Enter Key
DOOR

UP-Down Key

LAMP

TO INPUT THE DATA.

FLASHES

WHEN

TO SAVE THE DATA.

DOOR OPENS.

BRAKE
3 BRAKE STEPS.
(HIGH, MIDDLE, LOW).

Page 9

Hanil Science Industrial Co., Ltd.

10

Serospin

Operations Manual

4. OPERATION PROCEDURE
- STANDARD OPERATION 1.Turn the power on

ON/OFF
SWITCH

MANUAL SWITCH
TO OPEN DOOR

2.Press down RPM key and input the desired value and then press down ENTER key. Follow this step to input
all the desired values including TIME (T)

Page 10

Hanil Science Industrial Co., Ltd.

11

Serospin

Operations Manual

3.Especially, when you input TIME,


Press down T key one time input Minute showing the light off as shown on the below picture

LIGHT OFF POSITION

4. Press down T key two times input Second showing the light on as shown on the below picture

LIGHT ON POSITION

Page 11

Hanil Science Industrial Co., Ltd.

12

Serospin

Operations Manual

5.Press down ENTER key after inputting the desired value(TIME)

6.Press down START key to operate

Page 12

Hanil Science Industrial Co., Ltd.

13

Serospin

Operations Manual

7.Press down P key to save the desired values as a program showing the light on

LIGHT ON POSITION

8.Press down or key to designate the desired program No. from 0 to 9

Page 13

Hanil Science Industrial Co., Ltd.

14

Serospin

Operations Manual

9.Press down ENTER key after designating the desired program No.

10.Press down P key to recall the saved program showing the light off

LIGHT OFF POSITION

Page 14

Hanil Science Industrial Co., Ltd.

15

Serospin

Operations Manual

11. Press down or key to set the saved program from 0 to 9

12. Press down ENTER key after selecting the desired program No.

Page 15

Hanil Science Industrial Co., Ltd.

16

Serospin

Operations Manual

- PROGRAM FUNCTION -

Speed(RPM)

When you set the desired speed, press the RPM key and set it by
the up-down direction key. After that, press the Enter key.
It is not possible to input higher speed than max.speed.
This is to prevent the accident by overspeed.

RCF (x g)

To calculate the Relative Centrifugal Force, it is made to calculate


automatically RCF to Speed by using the Maximum Radius.
RPM and RCF are calculated mutually so if RPM is input, RCF is
calculated automatically and if reversely is same like above.
It is like either during operation.

- SETTING-

Press the RCF key and input the data by using up-down direction key.
After that, press the Enter key.

TIME

The display range of time is 99mins and shows the 2 digits.


The change of time counts down from setting time after running.
Timer is proceeding by a unit of 1 minute and under 1 minute is used
with a unit of second.

-SETTING-

Press the TIME key.


When Alarm sounds, it is possible to set.
Input the desired time by using the up-down direction key.
If you input the wrong data, Press T key and re-input the correct data.

Page 16

Hanil Science Industrial Co., Ltd.

17

Serospin

Operations Manual
Press the Enter key and if the Start key is pressed, it starts to count
down, and decelerate after over setting time.
Under one minute, it is showed as a second unit.
z Free run : You can use the Free run function for long time
Operation. For that, input the 0 in display of time
and start to run.

DECELERATION

- SETTING-

3 Steps : High, Middle, Low

Press the Brake key.


Whenever the brake key is pressed, the lamp of H, M, L light on and the
level of brake is changed.

PROGRAM SAVE

10 Programs all can be saved and the range is 0 to 9.


The data of RPM/Time, etc are saved in Program function.

- SETTING-

The desired data should be input by up-down key.


If press the program, the lamp of P key lights on with sound and
press Enter.

CALLING
The program saved between 0 to 9 can be called.

-SETTING-

Press the Program key to call the program saved before.


When the lamp lights on P key with alarm, press the P again and turn
off the lamp and alarm. After that, input the number to be called by up-

Page 17

Hanil Science Industrial Co., Ltd.

18

Serospin

Operations Manual
down direction key and press the Enter key. At that time, the data of
the pertinent program is showed.

IMBALANCE

When the rotor loses the balance of normal standard and vibrates
seriously,
Imbalance LED turns on with the alarm and stops according to Decel
Time as a pre-input.

DOOR
To open the door, turn the dial on the right side of the instrument.
When the door closes, Door lights off and when the door opens, the
Door lights on.
Close the door and press the Start key.
Caution ! : Dont open the door during the spinning.

Page 18

Hanil Science Industrial Co., Ltd.

19

Serospin

Operations Manual

5. MAINTENACE
This chapter explains how to keep your unit in good operating order.
It includes instructions for cleaning, decontaminating and storing. This chapter also covers the cover
interlock by pass.

Care and cleaning


Keep your centrifuge cleans, to ensure good operation, and to extend its life.
Clean the sample chamber, rotor and lid at the end of each works day and immediately after any spill. To
clean the chamber, use a damp sponge, warm water, and a mild liquid detergent, suitable for washing
dishes by hand. Dont use caustic detergents or detergents that contain chlorine ions. These attack metals.
Remove stubborn stains with a plastic scrub pad. Dont use steel wool, wire brushes, abrasives, or
sandpaper. They create corrosion sites. Never pour water directly into the rotor chamber. Scrub the rotors
tube cavities with a stiff test tube brush that has end bristles and a non-metallic tip. Dry each part, after
cleaning, with a clean, absorbent towel.
If glass breakage occurs, remove all broken glass embed in the plastic or rubber accessories.
Glass particles can come in contact with new glass tubes, creating pressure points that may results in
breakage recurring. Glass particles, in the chamber, grind to a fine gray dust, during centrifugation.
This dust can coat the inside of the centrifuge.

Storage
Store parts on a soft surface, to avoid damage.
Rotors and other parts should be clean and dry. Store them opened to the air, not in a plastic bag, so
that any residual moisture evaporates. Face the parts upward to avoid moisture retention in the cavities.