Aplication Fat Extraction Food

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Application Note

No. 173/2014
Extraction Unit E-816 ECE:
Fat Determination in Food using Twisselmann Continuous Extraction

www.buchi.com

Quality in your hands

1. Introduction
An effective procedure for fat determination in food according to 64 LFGB L 17.00-4 and ISO
1444:1996 is presented [1, 2]. The sample is hydrolyzed with the Hydrolysis Unit E-416. The
Twisselmann extraction is performed with the Extraction Unit E-816 ECE (Economic Continuous
Extraction).This allows the sample to be constantly kept in hot solvent vapor whilst efficiently
rinsed with freshly distilled solvent. The total fat content is determined gravimetrically after the
extract has been dried to a constant weight.

2. Equipment

Hydrolysis Unit E-416


Extraction Unit E-816 ECE
Analytical balance (accuracy 0.1 mg)
Mixer, Retsch Grindomix GM 200
Microwave oven
Drying oven / Vacuum drying oven

3. Chemicals and Materials


Chemicals:

Quartz sand, particle size 0.3-0.9 mm, BUCHI (037689)


Celite 545, Macherey-Nagel (815560)
10 L of 4 M Hydrochloric acid (HCl) are prepared by dilution of 4 L HCl 32 % (Hnseler
20-2000-5) to 10 L with deionised water
Petroleum ether, boiling range 40-60 C, analytical grade, ACS, EGT (ET0093005M)
Diethyl ether, puriss. meets analytical specification of Ph. Eur. BP 99.5 % (GC),
Sigma-Aldrich (24004)
Hexane, puriss. p.a. ACS reagent, reag. Ph. Eur, 99 %(GC), Sigma-Aldrich (32293)
Chloroform, Chromasolv Plus, for HPLC, 99.9 %, contains amylenes as stabilizer,
Sigma-Aldrich (650498)

For a safe handling please pay attention to all corresponding MSDS!


Samples:

Madeira Cake, certified reference material LGC7107, specified fat content:


13.4 0.7 g/100 g
Processed meat, certified reference material ERM-BB501b, specified fat content:
11.57 0.44 g/100 g

The samples were purchased from LGC Standards GmbH (Wesel, Germany).

4. Procedure
The determination of fat includes the following steps:

Homogenization of the sample by grinding


Acid hydrolysis of the sample, using the E-416
Filtration of the hydrolysed solution to separate the fat
Drying of the filtered sample
Extraction of the sample, using the E-816 ECE
Drying of the extract
Weighing of the extract
Calculation of the fat content

Application Note 173/2014

November 2014

2/6

4.1. Acid hydrolysis


4.1.1. Preparation of the glass sample tubes
1. Add approx. 20 g of quartz sand to the glass sample tube and compact the sand by
gently tapping the glass sample tube onto the table
2. Add approx. 2 g Celite 545 and spread it evenly using a spoon
The sand and the Celite layer should not be mixed together. Otherwise the Celite
phase may breakthrough the frit and affect the results either by a increasing the
recovery or by blocking the frit.
4.1.2. Hydrolyzing the sample matrix
3. Place 2 g Celite 545 in the digestion vessel
4. Add up to 10 g homogeneous sample1 to the digestion vessel and note the accurate
weight of the sample
5. Add 50 mL hydrochloric acid (4 M) and form a suspension by gently swirling the tube
6. Add another 50 mL hydrochloric acid (4 M) making sure to rinse any remaining sample
off the glass wall
7. Preheat the Hydrolysis Unit for 10 min
8. Insert the samples into the unit and lower the vessels
9. Connect the aspiration tubes, reduce the heat to level 3 and start the water-jet pump
after boiling begins
Violent foaming can be prevented by adding 4 M hydrochloric acid dropwise.
The degree of foaming depends on the sample and on the preheating time of the
unit. Do not extend preheating excessively.
10. Hydrolyze the sample for 30 min after constant boiling is observed in each position
11. Add 100 mL of warm (40-50 C) deionised water to each digestion vessel at the end of
the hydrolysis time
12. Switch off the heating and lift the digestion vessels to the top position in order to filter
the hydrolyzate
13. Wash each of the vessels by gradually adding a total of at least 400 mL warm deionised
water, until a neutral pH is reached
14. Check the pH with a pH paper on the bottom of the frit
For maximum efficiency, aspire aspirate all samples/rinsing water at the same time.
15. Stir the Celite layers (without touching the sand layer) with a spatula to loosen the pulp
16. Carefully wipe off the spatula with a piece of tissue and add it on the top of the sample
17. Dry the glass sample tubes in a vacuum oven ( 4 h at 100 C/200 mbar), in a drying
oven ( 8 h at 100 C) or in a microwave oven
Using a microwave oven accelerates the drying process. However, its control is more delicate.
This is due to the fact that the sample can easily overheat (> 105 C) if an inappropriate heating
power is chosen. The following suggestion is valid for the drying of six hydrolyzed samples at
the same time. First step: 15 min 640 W, second step: 9 min 480 W, power of microwave oven
800 W (the optimal parameters may depend on the model of microwave).
Faster drying at higher temperatures is not recommended because fat may
decompose at temperatures above 105 C. Oxidized fat can result in an excessive
recovery.
18. Allow the glass sample tubes to cool down to room temperature in a desiccator
19. Add another layer of quartz sand (20 g). This prevents the Celite from being resuspended in the condensed solvent

The sample weight has to be chosen according to the approximate fat content of the sample.
80-100 %: 0.7-1 g
20-50 % 1.5-3.5 g
<10 %: 7- 10 g
50-80 %: 1-1.5 g
10-20 % 3.5-7 g
Application Note 173/2014

November 2014

3/6

4.2. Fat extraction


4.2.1. Preparation of the beakers
Always use dry and clean beakers for the Twisselmann extraction. Add a boiling aid (e.g. boiling
stones) to each beaker and dry them for at least 30 min at 102 C. Let them cool down to
ambient temperature in a desiccator for at least 1 h. Record the exact weight prior to extraction.
4.2.2. Twisselmann extraction
Put the sample tubes into the extraction chamber using the pliers. See Figure 1.

Figure 1: Twisselmann extraction chamber before start

Fill the solvent directly into the beakers and place them on their corresponding heating plate.
Close the safety shield and lower the rack. Activate the occupied positions, open the cooling
water or switch on the connected chiller and start the extraction according to the parameters
listed in Table 1.
Table 1: Parameters for the extraction with the Extraction Unit E-816 ECE

Method parameters Extraction Unit E-816 ECE


Solvent

Petroleum ether / Diethyl ether / Hexane / Chloroform2

Extraction step

50 min (Heater 100 %)3

Drying step

10 min (Heater 100 %)4

Solvent volume

70 mL

4.2.3. Drying of the extract


Dry the beakers containing the extract in a drying oven at 102 C until a constant weight is
reached. Let the beakers cool down to ambient temperature for at least 1 h in a desiccator and
record the weight.
Make sure that the cooling down time of the beakers in the dessicator is the same
before and after extraction. Differences in beakers temperature falsify the results.

Please select the solvent used in the menu.


Choose the heater between 100 120 % so the boiling is sufficient.
4 Choose the same parameter for the heater as in the extraction step.
3

Application Note 173/2014

November 2014

4/6

4.3. Calculation
The results are calculated as percentage of the fat according to equation (1).

% Fat
% Fat
mTotal
mBeaker
mSample

(m Total - m Beaker )
100%
m Sample

(1)

: Percentage of fat in the sample


: Beaker + extract [g]
: Empty beaker weight with boiling aid [g]
: Sample weight [g]

5. Results
Determined fat contents for the certified reference materials are in line with the specified and
labelled values, independent of the solvent used. The relative standard deviations (rsd) are low,
i.e. < 2 % for all samples.
Depending on the type of solvent used, minor differences in the fat content are observed. This
can be explained as an effect of the solvent polarity which affects the mass transfer during the
extraction. With chloroform and diethyl ether being more polar than petroleum ether and
hexane slightly more fat is extracted. The complete findings are summarized in Tables 2 and
3.
Table 2: Maderia Cake LGC7107 (Specification 13.4 0.7 g/100 g)

Petroleum ether

Diethyl ether

Hexane

Chloroform

Sample 1

12.80

13.21

12.93

13.19

Sample 2

12.63

13.35

13.02

13.24

Sample 3

13.00

13.32

12.97

13.09

Sample 4

12.99

13.03

12.90

13.26

Sample 5

12.75

13.03

12.92

13.09

Sample 6

12.77

13.47

13.07

13.13

Mean value [g/100g]

12.82

13.24

12.97

13.16

rsd [%]

1.13

1.36

0.51

0.56

Table 3: Processed Meat ERM-BB501b (Specification 11.57 0.44 g/100g)

Petroleum ether

Diethyl ether

Hexane

Chloroform

Sample 1

11.11

11.67

11.58

11.88

Sample 2

11.21

11.56

11.29

11.49

Sample 3

11.23

11.55

11.15

11.44

Sample 4

11.12

11.61

11.44

11.79

Sample 5

11.06

11.66

10.735

11.57

Sample 6

11.59

11.63

11.33

11.88

Mean value [g/100g]

11.22

11.62

11.36

11.68

rsd [%]

1.70

0.42

1.42

1.70

This value is an outlier according to the statistical test of Dean-Dixon for outliers ( = 5 %).

Application Note 173/2014

November 2014

5/6

6. Conclusion
The determination of fat in different food products using the Hydrolysis Unit E-416 and the
Extraction Unit E-816 ECE provides reliable and reproducible results. These results correspond
well to the labelled values, with low relative standard deviations (rsd).
With the Extraction Unit E-816 ECE the time to results is significantly reduced by 75 %
compared to the use of classical glassware, and it offers an unattended automated process.

7. References
[1] 64 LFGB L 17.00-4:1982-05 Bestimmung des Gesamtfettgehaltes in Brot einschliesslich
Kleingebck aus Brotteigen
[2] ISO1444:1993 Meat and meat products Determination of fat content

Operation Manual of Hydrolysis Unit E-416


Operation Manual of Extraction Unit E-816 ECE

Application Note 173/2014

November 2014

6/6

You might also like