Baby Warmer
Operation and Maintenance Manual
Approved
200964 16:13 EET DST
66000371000DWG
100
Reproduced from the electronic master in MATRIX
Barbara A Hairfield
MANUALDOC, User Reference Manual, OM Panda, Englishdefault
TableResponsibility
User
of Contents
This Product will perform in conformity with the description thereof contained
in this operating manual and accompanying labels and/or inserts, when
assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product
should not be used. Parts that are broken, missing, plainly worn, distorted or
contaminated should be replaced immediately. Should such repair or replacement become necessary, Ohmeda Medical recommends that a telephone or
written request for service advice be made to the nearest Ohmeda Medical
Regional Service Center. This Product or any of its parts should not be
repaired other than in accordance with written instructions provided by
Ohmeda Medical and by Ohmeda Medical trained personnel. The Product
must not be altered without the prior written approval of Ohmeda Medicals
Quality Assurance Department. The user of this Product shall have the sole
responsibility for any malfunction which results from improper use, faulty
maintenance, improper repair, damage, or alteration by anyone other than
Ohmeda Medical.
66000371000DWG
CAUTION: U.S. Federal law restricts this device to sale by or on the order
of a licensed medical practitioner.
6600-0371-000
100
10/27/03
Table of Contents
Definitions ....................................................................................................................... ii
General Precautions ..................................................................................................... iv
Warnings ........................................................................................................ iv
Cautions ......................................................................................................... iv
1/Description
1.1
1.2
1.3
1.4
1.5
1.6
1.7
2.2
2.3
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3/Operation
66000371000DWG
6600-0371-000
100
10/27/03
Table of Contents
A.
B.
3.8
4.2
4.3
5.1
5.2
5.3
5.4
5.5
5.6
5/Maintenance
6/Illustrated Parts
Appendix
Temperature Conversion Chart .................................................................. A-1
Servo Mode Algorithm ................................................................................ A-2
Infant Warmer System Specifications ......................................................... A-2
Electrical ............................................................................................. A-2
Controller ............................................................................................ A-3
Alarms ................................................................................................ A-4
Environmental Specifications ............................................................. A-5
Electromagnetic Compatibility (EMC) ................................................ A-6
Mechanical (Without Accessories) ..................................................... A-6
Bassinet ............................................................................................. A-6
Accessories ........................................................................................ A-7
Radiant Energy Distribution ............................................................... A-8
Installing wall mounted units ....................................................................... A-9
Pre-Installation Preparation .............................................................. A-10
Warmer Installation .......................................................................... A-10
Warranty
ii
66000371000DWG
6600-0371-000
100
10/27/03
ii
Definitions
Table of Contents
Note: A Note provides additional information to clarify a point in the text.
Important: An Important statement is similar to a note but used for greater
emphasis.
w
w
Symbols
w
m
y
y
M
?
Type B equipment.
Alarm Silence
Alternating Current
iii
66000371000DWG
6600-0371-000
100
10/27/03
iii
Notes
Table of Contents
iv
66000371000DWG
6600-0371-000
100
10/27/03
iv
General
Table of Precautions
Contents
wWarnings
Before using the Ohmeda Medical PandaTM Baby Warmer System, read this
entire manual. Attempting to use this device without a thorough understanding
of its operation may result in patient or user injury. This device should only be
operated by personnel trained in its operation and under the direction of qualified
medical personnel familiar with the risks and benefits of this type of device.
Additional precautions specific to certain procedures are found in the
text of this manual.
Do not use the Warmer in the presence of flammable anesthetics; a possible
explosion hazard exists under these conditions.
Do not perform the Checkout Procedures (Mechanical and Control Unit) while a
patient occupies the Warmer.
Complete the Checkout Procedures section of this manual before putting the
unit into operation. If the warmer fails any portion of the checkout procedures it
must be removed from use and repaired.
Do not use the Warmer system if the system failure alarm is activated. Remove
the unit from service and refer to qualified personnel for repair.
wCautions
Only competent individuals trained in the repair of this equipment should
attempt to service it as detailed in the service manual. The Service Manual
provides detailed information solely for use by individuals having proper
knowledge, tools and test equipment, and for service representatives trained
by Ohmeda Medical.
66000371000DWG
6600-0371-000
100
10/27/03
Notes
Table of Contents
vi
66000371000DWG
6600-0371-000
100
10/27/03
vi
1/Description
In this section
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.1 General
The Ohmeda Medical PandaTM Baby Warmer provides a controlled source of radiant heat
for infants and pediatric patients. The control system uses a microprocessor and provides
both manual and servo modes of operation. The patient temperature, control temperature,
and elapsed time displays are digital for ease of viewing. The control panel includes a one
hour elapsed timer with time displayed in minutes and seconds. The timer features
optional audible tones for use during Apgar scoring. A complete audio and visual alarm
system is included on the control panel.
WARNING:
w Bed-to-heater spacing less than the recommended 64 cm will result in incorrect operation and may adversly affect the patients condition.
For your convenience, on the free standing unit the proper bed-to-heater spacing is indicated by a label located on the right aluminum upright (as viewed from the front).
The heater assembly consists of a radiant heater, parabolic reflector, observation light, and
a visual alarm light. The parabolic reflector focuses radiant energy on the bed surface,
minimizing energy loss due to scattering and providing an even field of radiant heat over the
bed surface. The observation light provides intense light for procedures. The alarm light is
located on the end of the heater assembly for ease of viewing. The entire heater assembly
rotates to the side for X-ray procedures and for observation lamp replacement.
1-1 66000371000DWG
6600-0371-000
100
10/27/03
1-1
1/Description
When the unit is first turned on, it automatically operates in the preheat mode, and will
continue to do so until the mode, increase/decrease or Apgar buttons are pressed. Then
a tone will prompt the operator to select a heater power percentage in the manual mode.
Either manual or servo mode of operation may be selected. In the manual mode of
operation, select the level of radiant heat output as indicated by the % power display on
the control panel. The control circuit then maintains the selected level of radiant heat. In
the servo mode of operation, select the patients control temperature. A skin temperature
probe is used to monitor the patient skin temperature. The control system modulates the
radiant heat to maintain the patient at the selected control temperature. The patients
skin temperature is continuously displayed.
Alarms activate to alert the operator of a low or high patient temperature, a skin temperature probe failure, a power failure, equipment failure or a check patient prompt.
1.6 Bassinet
Two bassinets models are available for use with the free standing warmers. One with
removabe drop side panels and mattress, and one with a removable tub. Transparent
locking side panels fold down for easy patient access and are removable for cleaning. A
tiltable lock block allows Trendelenburg and reverse Trendelenburg positioning at the 8
tilt position. Three drawers open from the side for storing equipment or supplies. The top
drawer is provided with a key lock. Front casters lock to hold the bassinet in place.
1.7 Accessories
The rail mounting system is the basis for most of the accessories for the Warmer. This
patented design consists of a dovetail shaped aluminum extrusion and a positive locking
mounting block. Mounting blocks are attached to various accessories for mounting to the
rail system. The mounting blocks are produced in two standard lengths and are machined
to accommodate each specific accessory.
Accessories which mount on a single upright are interchangeable between all model
Warmers.
Accessories can only be mounted to the inside dovetail rail on each upright of the free
standing warmer.
1-2 66000371000DWG
6600-0371-000
100
10/27/03
1-2
2.2
2.3
2.1 Setup
Refer to the setup instructions shipped with the warmer for initial unpacking and setup of
the unit after shipment.
After removal from the shipping containers, inspect the Ohmeda Panda Baby Warmer and all
accessory items for any signs of damage which may have occurred during shipment. File a
damage claim with the shipping carrier if damage has occurred. Also confirm the presence of
all accessory items or factory installed options as listed on the packing slip.
w Before using the Ohmeda Panda Baby Warmer, read this entire manual.
Attempting to use this device without a thorough understanding of its
operation may result in patient or user injury.
A. Overall Appearance
1. Disconnect the power cord from the ac power source for the mechanical checks
portion of this procedure.
2. Check the overall appearance of the Panda Baby Warmer/ Bassinet System. There
should be no obvious damage.
3. Check that all casters are in firm contact with the floor and that the warmer is stable
and moves freely.
For the free standing warmer, check that all six casters on the warmer move freely.
Note: On the free steanding warmer, it is possible that the two center casters may not
be in contact with the floor at all times.
2-1 66000371000DWG
6600-0371-000
100
10/27/03
2-1
Examine the power cord for damage. Replace the power cord if damage is evident.
7. Examine the unit for objects placed on top of the heater assembly.
8.
WARNING:
Check heater-to-bed spacing falls within ranges specified in "1.3 Heater Assembly".
w Do not place any accessories or other objects directly over the bed surface.
This may block radiant heat energy and lead to cooling of the infant.
w Do not place items on top of the heater assembly. Items placed on top of
the heater assembly can fall and injure the patient, prevent adequate
ventilation of the heater assembly, and may pose a fire hazard.
C. Mechanical Checks
1. Check the operation of the bed sides. The bed sides should operate smoothly.
WARNING:
w Regularly inspect the bed side panel latching mechanism to ensure proper
operation and to prevent the infant falling off the bed.
2. Check the operation of the tilt mechanism. Verify that the bed platform operates
smoothly and locks in normal, Trendelenburg and reverse Trendelenburg positions.
3. If the drawer package is installed, check that all the drawers open and close freely. On
units with the rotating drawer package, make sure the drawers rotate from front to 90
left and right and firmly stop in the detented positions.
WARNING:
w Overloading the drawers can affect the stability of the unit. Limit the load
to 4.6 kg per drawer.
D. Accessory Checks
Perform these checks if they are applicable.
1. Check that all accessories are mounted securely and that the load limits are not
exceeded.
2.
Check that all gas accessories are installed and operating properly (refer to Section 3.9).
3. Where applicable, perform the checkout procedures detailed in the Operation and
Maintenance Manuals for the accessories.
2-2 66000371000DWG
6600-0371-000
100
10/27/03
2-2
CI.14.096
Figure 2-1
Wall mount pin insertion
1.
Verify that the warmer is rigidly secured to the wall and the heater assembly is level.
Note: To access the control unit and display module for service procedures, either
hinge pin may be removed, allowing the warmer to pivot away from the wall.
2. Check that both hinge pins are in place and fully inserted with the pin heads at the
top of the hinge.
WARNING:
w In the service position the strength of the hinge bracket is reduced. Never place a
patient in the bed when the heater is in the service position. Never leave the unit
unattended in the service position or with either pin removed.
Light
On/Off
Switch
0000
Control
Temperature
Heater
Power %
0000
Elapsed
Time Display
CI.44.005
Patient
Temperature
0000
Bed
Controls*
Alarm
Lights
Alarm
Silence
Mode
Switch
Mode
Indicators
Increase
Decrease
Apgar
Indicator
Apgar
Tones
On/Off
Elapsed Time
Start/Hold
and Reset
Figure 2-2
Control Panel
2-3 66000371000DWG
6600-0371-000
100
10/27/03
2-3
2-4 66000371000DWG
6600-0371-000
100
10/27/03
2-4
CAUTION:
w Do not continue to run the motor at the upper and lower limit positions; equipment damage may result.
1. Press the Raise Bed touch switch and verify that the bed raises to a maximum of
118 cm off the floor.
2-5 66000371000DWG
6600-0371-000
6600-0371-000
100
10/27/03
03/07/02
2-5
2-6 66000371000DWG
6600-0371-000
100
10/27/03
03/07/02
2-6
3/Operation
In this section
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
..................................................... 3-7
A. Displays
Patient Temperature
Control Temperature*
% Power
Elapsed Time*
*During the preheat mode "PrE HEAT" flashes in the control temperature and elapsed time displays.
Patient Temperature:
The patient temperature display indicates the temperature sensed at the skin temperature probe tip. The skin temperature probe must be properly connected to the warmer and
the infant for an accurate patient temperature measurement. The range of temperature
measurement is from 30.0 to 42.0C, with a resolution of 0.1C. Temperatures above this
range result in an HH.H patient temperature display and below this range result in an
LL.L patient temperature display .
Control Temperature:
The control temperature display indicates the control temperature you select in the servo
mode. The servo control temperature range is from 35.0 to 37.5C. The control temperature display is not active in the manual mode.
% Power:
The % power display indicates the percentage of maximum power that is being supplied
to the radiant heater in 5% increments.
Power settings between 0 to 25% allow operation without the 12 minute check patient
alarm in the manual mode.
3-1 66000371000DWG
6600-0371-000
100
10/27/03
3-1
3/Operation
Elapsed Time:
The elapsed time display indicates elapsed time in minutes and seconds up to a maximum of 60 minutes.
B. Indicator Lights
Servo Mode Indicator
Manual Mode Indicator
Apgar Tones Indicator
Servo Mode Indicator:
The servo mode indicator lights when the warmer is in the servo mode of operation.
For proper operation of the warmer in the Servo Mode see section 3.2.
Manual Mode Indicator:
The manual mode indicator lights when the warmer is in the manual mode of operation.
For proper operation of the warmer in the Manual Mode see section 3.2.
WARNING:
w Use the servo mode unless the manual mode is specifically prescribed.
While both modes require patient monitoring, the manual mode requires
constant attention. In the manual mode, you must take the responsibility
for making heater adjustments or environmental changes in response to
the following conditions:
patient size and weight;
bed to heater head height differences;
changes in the environment (drafts, direct sunlight, phototherapy lamp
usage, etc.);
changes in the patient condition
In the servo mode, the warmer automatically adjusts heater output to
maintain the desired skin temperature, reducing (but not eliminating) the
need to monitor the patient and make environmental adjustments to
ensure adequate patient temperature.
C. Switches
Mode Switch
Increase (M) and Decrease (?) Switches
Alarm Silence
Start/Hold
Reset
Apgar Tones On/Off
Light On/OffMode Switch
Raise Bed (M) (Elevating models only)
Lower Bed (?) (Elevating models only)
3-2 66000371000DWG
6600-0371-000
100
10/27/03
3-2
3/Operation
This switch is used to select either the Manual or Servo mode of operation. An audio tone
sounds momentarily when the mode switch is depressed.
?) Switches:
Increase (M
M) and Decrease (?
These switches are used to set the radiant power level (% power) in the manual mode
and to set the control temperature in the servo mode.
Alarm Silence:
This switch is used to silence the audible alarm. It silences all alarms except the system
failure alarm. Refer to Table 1 - Warmer Alarms for details on how long the audible alarms
are silenced.
Holding this switch depressed for five seconds or longer activates the audible alarm and
lights all the indicator lamps for testing purposes.
Start/Hold:
This switch activates the elapsed timer. Refer to the elapsed timer operation, Section 3.3,
for a detailed explanation of its function.
Reset:
This switch is used to reset the elapsed timer to 00:00.
Apgar Tones On/Off:
This switch activates the Apgar tones. When the Apgar tones are activated, a short alarm
tone is sounded at one minute and every five minute interval after the elapsed timer is
started.
Light On/Off:
This switch activates the observation light located in the heater assembly.
WARNING:
WARNING:
w Check for proper clearance above the Warmer and below the bed surface
before raising or lowering the bed.
w Inspect all patient connected tubes or wires before and after moving or
tilting the bed. Tilting or moving the warmer bed up or down can pull on
tubing or leads connected to the patient. This may disconnect tubes or
leads, restrict gas or liquid flow, or move sensors out of position.
CAUTION:
w Do not continue to run the motor at the upper and lower limit positions; equipment damage may result.
This switch raises the bed to a maximum height of 118 cm from the floor.
Lower Bed (?
?) (Elevating Models Only)
This switch lowers the bed to a minimum height of 98 cm from the floor.
3-3 66000371000DWG
6600-0371-000
100
10/27/03
3-3
3/Operation
On/Off Power Switch and Circuit Breaker:
(See Figure 3-1.)
The On/Off power switch is located on the left side of the controller assembly (as viewed
from the front) near the power cord socket. It is used for switching the power to the
warmer(On and Off). The switch is also a circuit breaker and limits the maximum current
drawn by the unit. If this circuit breaker trips when the warmer is operating, the power switch
is deactivated to the Off position. To reset the circuit breaker, return the power switch to the
On position. If the circuit breaker trips again, service is required. On/Off Switch
CI.05.009
(I-On, O-Off)
Figure 3-1
On/Off Power Switch with built-in circuit breaker
D. Alarms
All alarms except system failure and power failure are preceded by a 30 second intermittent operator prompt tone.
6600-0371-000
100
10/27/03
3-4
3/Operation
WARNING:
w In the servo mode, verify that the patient temperature probe is securely
attached to the patient at least once every half hour. A dislodged probe
may not trigger an alarm. If the probe becomes dislodged, the warmer
can over or under heat the infant.
WARNING:
w Do not use the Warmer system if the system failure alarm is activated.
Remove the unit from service and refer to qualified personnel for repair.
3-5 66000371000DWG
6600-0371-000
100
10/27/03
3-5
3/Operation
Table 1 Warmer Alarms
Alarm Condition
Probe Failure
Alarm Sound
Alternating Two-Tone
Heater
Off
Patient Temperature
is greater than 42C
Alternating Two-Tone
1 minute
Off
Patient Temperature
more than 2C from
Control temperature
Alternating Single-Tone
5 minutes
Patient Temperature
between 1 and 2C of
Control temperature
15 minutes
Patient Temperature
is less than 30C
Alternating Two-Tone
1 minute
Off
System Failure**
Alternating Two-Tone
Cannot be
silenced
Off
Heat Off
5 minutes
Off
Check Patient
12 minutes
15 minutes
Off
Cannot be silenced
Off
Power Failure
Error Codes:
In the event of a system failure, a nonsilenceable, high priority alarm sounds and the System
failure LED illuminates. Before shutting off the Warmer and removing it from service, make
note of the error code that appears in the elapsed time display. This code will aid qualified
service personnel in diagnosing the problem. Table 2 lists the possible error codes, for a more
detailed explanation of error codes see the Service manual.
E002
E003
E004
Checksum Failure
E005
E007
E010
E012
E013
E014
E015
E016
E017
E019
Software Upset
E025
3-6 66000371000DWG
6600-0371-000
100
10/27/03
3-6
3/Operation
3.2 Preheat, Manual and Servo Mode Operation
WARNING:
w Before using the Panda Baby Warmer System, read this entire manual.
Attempting to use this device without a thorough understanding of its
operation may result in patient or user injury. This device should only be
operated by personnel trained in its operation and under the direction of
qualified medical personnel familiar with the benefits and risks of this
type of device.
w Do not leave the patient unattended when using the Infant Warmer.
Check the patients temperature periodically using an independent temperature monitoring device to ensure the comfort and the safety of the
patient. If the warmer is used for an extended time, it is recommended
that the servo mode of operation be used. When an alarm is silenced,
close monitoring of the patients condition is required.
w When using a radiant warmer, change the patients diapers frequently. Radiant
energy causes more rapid urine evaporation, and may lead to inaccurate urine
diagnosis test analysis and inaccurate weight measurements.
w Do not touch the protective grill under the radiant heater or the top of the
heater assembly. These surfaces may be hot and a burn could result.
w Always set the caster brakes before placing a patient in the warmer.
1. Connect the power cord to a properly grounded AC power source.
2. Place the power switch in the On position. During the first seconds of operation, the
warmer performs a self check of the control system. The software is verified, calibration is checked and operation of the solid state relay controlling the heater is verified.
All displays and indicators are lit and the audible alarm is sounded. If the self check
detects a failure, the alarm stays on and service is required.
Note: The Panda Baby Warmer begins operation in the Preheat Mode, described in
Step 3. To operate the warmer in the Manual Mode, refer to step 4. To operate the
warmer in the Servo Mode, refer to Step 6.
3-7 66000371000DWG
6600-0371-000
100
10/27/03
3-7
3/Operation
WARNING:
w Do not operate the warmer in the preheat mode while a patient occupies the
bed. The preheat mode is used to quickly warm and maintain heat to an
empty bed. The preheat mode is not designed for clinical use with a patient
since the Check Patient Alarm is disabled.
3. Preheat Mode Operation
The Panda Baby Warmer automatically starts in the Preheat Mode. During the preheat
mode Check Patient Alarms are not activated.
In the Preheat Mode, the control temperature and elapsed time displays continiously
flash the message "PrE HEAT" and the heater operates at 100% power for the first 12
minutes. After 12 minutes, heater power drops to 50%, and will continue at 50% until
either the mode, increase/decrease or Apgar touch switches are pressed. Pressing the
mode, increase/decrease or Apgar touch switches exits the Preheat Mode and returns
the warmer to normal modes of operation.
If the Apgar touch switches are pressed, the manual mode indicator will light and a
prompt tone will sound until a power percentage is selected using the increase/decrease
touch switches. The Check Patient Alarm is again activated.
If the mode or the increase/decrease touch switches are pressed, the unit will automatically change to the Manual Mode with a power percentage setting of 50%. The Check
Patient Alarm is again activated.
The only way to re-enter the preheat mode is to shut off and restart the warmer.
4. Manual Mode Operation.
WARNING:
w Use the servo mode unless the manual mode is specifically prescribed. While
both modes require patient monitoring, the manual mode requires constant
attention. In the manual mode, you must take the responsibility for making
heater adjustments or environmental changes in response to the following
conditions:
patient size and weight;
bed to heater head height differences;
changes in the environment (drafts, direct sunlight, phototherapy lamp usage,
etc.);
changes in the patient condition
In the servo mode, the warmer automatically adjusts heater output to maintain the desired skin temperature, reducing (but not eliminating) the need to
monitor the patient and make environmental adjustments to ensure adequate
patient temperature.
In the manual mode of operation, an operator prompt tone and the flashing of the % power
display prompt you to select a level of radiant heat.
Use the increase (M) or decrease (?) touch switch to adjust the % power in 5% increments. The % power display indicates the power level selected. You must select a %
power level each time the warmer is switched to the manual mode.
When you select % power levels between 30% and 100%, a check patient alarm sounds
after 12 minutes to remind you to monitor the patient temperature. After checking the
patient, this alarm can be reset by pressing the alarm silence switch. This alarm recurs
every 12 minutes during operation in the manual mode. If the alarm silence switch is not
pressed within 3 minutes of the check patient alarm, an alternating two tone alarm
sounds and the heater is turned off.
In the manual mode, the warmer may be operated using 0% to 25% power level settings
without the 12 minute check patient alarm being activated.
3-8 66000371000DWG
6600-0371-000
100
10/27/03
3-8
3/Operation
The skin temperature probe may be used to monitor the patients temperature in the
manual mode but it does not control the radiant heat energy level. See step 5 for skin
temperature probe attachment.
Note: A patient placed in any warmer will normally develop temperature gradients
with hotter and cooler areas. This is due to radiant heat being applied above the
infant, the cooling effect from the mattress below the infant, the unequal skin cooling
effect from evaporative water loss, unequal internal heat generation within the patient,
and the environmental variables of room temperature, room air movement, incidental
sunlight, etc.
5. Skin Temperature Probe Attachment.
WARNING:
w Use only the Reusable Ohmeda skin temperature probe (Stock No. 02080697-700) and heat reflective patches (Stock No. 0203-1980-300, 50/pkg) or
the disposable probe (Stock No. 6600-0208-700, 10/pkg; Stock No. 66000196-700. 50/pkg) to monitor the patients skin temperature. Use of other
manufacturers probes may affect the accuracy of warmer operation and
the electrical safety of the patient.
The Ohmeda Patient probe lead is made from low mass wire that helps prevent probe
detachment while reducing pulling on the neonates skin.
Place the metal side of the skin temperature probe on the skin over the liver area of
the infants abdomen. Remove the paper protecting the hypoallergenic adhesive on
the Heat Reflective Patch. Secure the skin temperature probe to the patients skin
with the adhesive side of the patch (Figure 3-2). Do not remove the heat reflecting foil
which must be facing up.
Note: The probe jack is attached at a specific torque value. Loosening or tightening
the jack may break the electrical connector. See Figure 3-3.
If the patient is prone, place the skin temperature probe on the back, where it will not
be against the mattress. If the probe is between the patient and the mattress, it will
produce false readings.
WARNING:
w Intimate contact between the skin temperature probe tip and the patients
skin must be maintained for accurate skin temperature measurement.
Underheating or overheating may result from poor contact between the
skin temperature probe and the patient. Verify that the skin temperature
probe is securely attached to the patient at least once every half hour.
Connect the skin temperature probe to the Warmer by plugging the probe connector into
the left side of the display module as viewed from the front. (See Figure 3-3.)
3-9 66000371000DWG
6600-0371-000
100
10/27/03
3-9
3/Operation
CAUTION:
w Always remove the probe from the patient by grasping and removing the heat
reflective patch first, then remove the probe from the patient or the patch. Always
remove the probe from the Warmer by grasping the plug at the panel. Placing
excessive strain on the skin temperature probe lead wire can damage the probe.
White Paper
Protector
Patient Probe
(Place the metal side in contact with the
infants skin)
Figure 3-2
Skin Temperature Probe attachment
CI.05.011
Figure 3-3
Temperature Probe connection to the Display Module
6. Servo Mode Operation.
Note: The skin temperature probe must be properly attached before starting servo
mode operation. Refer to step 5.
You must select a servo control temperature setting when the Warmer is used in the
servo mode for the first time. You are prompted to make this setting with an operator
prompt tone and the flashing of the control temperature display.
The servo control temperature is adjusted by pressing the increase (M) or the
decrease (?) touch switches. The control temperature can be adjusted from 35 to
37.5C. In the servo mode, the temperature sensed by the skin temperature probe is
used by the control system to modulate the radiant heat and maintain the patients
skin temperature at the selected control temperature. The Appendix details the Servo
Control Mode Algorithm.
3-1066000371000DWG
6600-0371-000
100
10/27/03
3-10
3/Operation
Note: The Warmer cannot differentiate between an increase in core temperature with
cold skin (fever), and low core and skin temperature (hypo-thermia). Patient temperature should be verified with an axillary thermometer.
WARNING:
w In the servo mode, verify that the patient temperature probe is securely
attached to the patient at least every half hour. A dislodged probe does
not always trigger an alarm. If the probe becomes dislodged, the warmer
can over or under heat the infant.
Note: A patient placed in any warmer will normally develop temperature gradients
with hotter and cooler areas. This is due to radiant heat being applied above the
infant, the unequal skin cooling effect from evaporative water loss, unequal heat
generation within the patient, and the environmental variables of room temperature,
room air movement, incidental sunlight, etc.
WARNING:
WARNING:
w Inspect all patient connected tubes or wires before and after moving or
tilting the bed. Tilting or moving the warmer bed up or down can pull on
tubing or leads connected to the patient. This may disconnect tubes or
leads, restrict gas or liquid flow, or move sensors out of position.
Note: Tilting the bed mattress can affect the operation and performance of the radiant
warmer.
3-1166000371000DWG
6600-0371-000
100
10/27/03
3-11
3/Operation
3.5 Side Panel Operation
(See Figure 3-4.)
WARNING:
w Do not leave the patient unattended when the side panels are lowered.
w Do not move the warmer by pushing or pulling on the bedside panels.
This action may lead to the deterioration and breakage of the components which form a safety barrier around the infant.
w Ensure that the bedside panels are locked in position when a patient
occupies the bed. Blankets or other foreign objects may prevent the
latches from fully engaging.
To lower a side panel, pull it up and then pull the top edge away from the bed.
To raise a side panel, swing it to the upright position; then allow it to engage in the
latched position.
To remove a side panel, lower the side panel, press the end pins in and lift the side panel
out.
To replace a side panel, hold the end pins in, place the side panel in position and release
the end pins.
CI.05.010
Figure 3-4
Side Panel operation
The Tubing Organizer Rear Side Panel aids in routing tubing to patients receiving therapy
in warmer beds. The various size holes in the panel can accommodate large diameter
tubes, such as aerosol tubes or the fiber optic cable of a BiliBlanket System; or
smaller diameter tubes, such as ventilator or I.V. tubes. See Figure 3-5.
WARNING:
3-1266000371000DWG
6600-0371-000
100
10/27/03
3-12
3/Operation
CI.05.038
Tubes up to
9 mm O.D.
Tubes up to
13 mm O.D.
Tubes up to
21 mm O.D.
Figure 3-5
Tubing organizer panel
WARNING:
w Inspect all patient connected tubes or wires before and after moving or
tilting the bed. Tilting or moving the warmer bed up or down can pull on
tubing or leads connected to the patient. This may disconnect tubes or
leads, restrict gas or liquid flow, or move sensors out of position.
w Do not lower the Tubing Organizer Rear Side Panel with tubing attached.
Lowering can pull on the tubing, causing the tubes to dislodge from the patient.
Using the side panel as a tubing organizer restricts the panels ability to be lowered.
Install the tubing organizer side panel on the back of the bed platform (the side nearest
the aluminum uprights) to maintain easy access to the infant.
The tubing organizer side panel should be inspected, visually and by touch, for sharp
edges, cracks or crazing, before each use.
w Oxygen concentrations higher than 40% can increase the risk of retrolental
fibroplasia (retinopathy of prematurity). It is probable that even concentrations
of 40% or less oxygen (formerly considered safe) could be dangerous to some
infants. Therefore, arterial blood gas measurements are extremely important
for regulation of the concentration of inspired oxygen when an oxygen-enriched environment is considered necessary. (See current edition of Standards
and Recommendations for Hospital Care of Newborn Infants prepared by the
Committee of Fetus and Newborn of the American Academy of Pediatrics.)
Oxygen can be administered from a regulated pipeline source or from auxiliary oxygen
tanks mounted on the Warmer.
3-1366000371000DWG
6600-0371-000
100
10/27/03
3-13
3/Operation
3.7 Gas Cylinder Installation and Operation
(See Figure 3-6)
Note: Use United States E size pin indexed gas cylinders only.
U.S. E size cylinders, with 11 cm x 66 cm outside dimensions, may be used with
warmer units with elevating pedestals.
WARNING:
w Using medical gas cylinders that are not recommended for mounting on
the Infant Warmer System may result in a fire hazard or an explosion.
Cylinder Wrench
Open cylinder valves
slowly to avoid
damaging the
regulators.
Figure 3-6
Gas cylinder installation
Use a fresh
gasket each
time a cylinder
is replaced.
Use only one
cylinder gasket
per yoke.
6600-0152-400
CI.06.013
Cylinder Valve
Close Yoke
Gate. Tighten
tee handle by
hand only.
Cylinder Removal:
1. Close the valve on the cylinder to be removed.
2. While supporting the cylinder, back out the Tee handle until the tip of the screw is
flush with the inside surface of the gate.
3. Swing the gate open in a counter clockwise direction.
4. Remove the old cylinder.
5. Remove the old gasket from the yoke strainer nipple and discard.
Cylinder Installation:
1. Install a new gasket (Ohmeda Stock No. 6600-0152-400) on the strainer nipple.
CAUTION:
w Use only one cylinder gasket per yoke. Use of more than one gasket could
cause leakage of the cylinder gas.
6600-0371-000
100
10/27/03
3-14
3/Operation
a. Remove the cylinder dust cap, if present.
b. Briefly open then immediately close the cylinder valve to blow any foreign matter
out of the valve outlet. Do not use excessive force on the shut off valve.
3. Install the cylinder valve over the strainer nipple, making sure that the safety index
pins are engaged.
4. Swing the gate closed in a clockwise direction and tighten the Tee Handle sufficiently
(by hand only) to hold the cylinder firmly in place. Do not use wrenches or any other
tool on the Tee Handle screws.
5. If only one oxygen cylinder is used, a yoke plug (Ohmeda Stock No. 6600-0399-500)
and a gasket are required to seal the unused yoke cylinder port.
CAUTION:
w Yoke check valves are not intended to provide a leak free seal; always use a
yoke plug and a fresh cylinder gasket to seal an unused cylinder port.
CAUTION:
2. Disconnect all pipeline gas supply and outlet lines from the yoke assembly.
3. Use new gaskets and full cylinders at each cylinder port (optionally, one full cylinder
and a yoke plug may be used on a 2 cylinder oxygen manifold)
4. Open a cylinder valve at each manifold to charge the manifold(s) to a minimum of
5137 kPa and then close the valve(s).
5. Note the gauge pressure reading in the manifold(s).
6. After one minute, check the gauge pressure reading(s) again. The pressure gauge
needle should not drop visually. If there is a visual pressure drop, discontinue use of
the gas system and have the yoke assembly repaired by a qualified service person.
Using Two Cylinders as an Oxygen Supply:
When two cylinders are installed to provide an oxygen supply, a check valve permits
replacement of one depleted cylinder while the other continues to furnish oxygen. Leave
the valve on the new cylinder closed until the old cylinder requires replacement, then
open the valve on the new cylinder and immediately replace the old cylinder as described
above. This assures that a full back-up cylinder is always available for use.
WARNING:
w Leaving both oxygen cylinder valves open at the same time will allow simultaneous depletion of both cylinders with no reserve oxygen supply available.
3-1566000371000DWG
6600-0371-000
100
10/27/03
3-15
3/Operation
Use of Pipeline Air and Oxygen Supplies:
1. Connect pipeline supply(s) to the appropriate pipeline inlet(s).
2. Connect desired equipment to the appropriate outlet(s).
3. Ensure correct operation of the secondary pipeline equipment.
When pipeline air or oxygen supplies are used, install cylinders as described above to
provide an emergency gas supply in the event of a pipeline failure. Cylinder valves should
remain closed until and unless the emergency gas supply is needed.
WARNING:
w Do not leave gas cylinder valves open if the pipeline supply is in use.
Pressures from both oxygen supplies may become equal, and if simultaneously used, cylinder supplies may be depleted, leaving no reserve supply
in case of pipeline failure.
WARNING:
w Limit the load of the accessories to 23 kg per side on the Warmer to ensure
stability. Always try to evenly distribute the weight of accessories on both
sides of the unit for a more balanced load. Accessories should not be
mounted more than 142 cm above the floor. For the free standing warmer,
limit the load of accessories to 9 kg maximum per side mounted no more
than 112 cm above the floor.
CI.02.004
CI.02.005
Rail system components mount to the uprights and provide ready access to commonly
used equipment such as suction regulators, flowmeters, collection bottles, etc.
Figure 3-7
Mounting rail system components
3-1666000371000DWG
6600-0371-000
100
10/27/03
3-16
3/Operation
A. Mounting and Releasing Procedure
(See Figure 3-7)
1. Loosen the mounting screw on the mounting block.
2. Place the mounting block in position on the rail.
3. Tighten the mounting screw.
4. Release the rail system component by loosening the mounting screw.
CI.14.002
Figure 3-8
Mounting the Adapter Plate
3-1766000371000DWG
6600-0371-000
100
10/27/03
3-17
3/Operation
3.9 Rotating Drawer Option
The rotating drawer option allows access to the storage drawers from the front or from
either side of the warmer. The drawer cabinet rotates by pushing on the side of the
cabinet, stoping at 90 detent positions.
WARNING:
w Use caution when rotating the cabinet to avoid damage to the drawers or
CI.14.003
possible injury. Always ensure the drawers are fully closed before rotating
the cabinet.
Figure 3-9
Rotating Drawer Package
3-1866000371000DWG
6600-0371-000
100
10/27/03
3-18
4.2
4.3
4.1 Cleaning
WARNING:
w Disconnect power to the Warmer and allow the heat rod to cool before
cleaning to avoid the possibility of a burn.
Clean the Warmer at least once a week or after each patient. Cleaning procedures for
the Warmer and accessories are explained in the following sections.
CAUTION:
w Use the cleaning solution sparingly on a cloth when cleaning the warmer. Do
not saturate the unit - excessive solution causes damage to internal components.
CI.10.013
Foam Mattress
Side Panel
Front/Rear
Panel
Push end pins
in for
installation
and removal.
Figure 4-1
Disassembly for cleaning
4-1 66000371000DWG
6600-0371-000
100
10/27/03
4-1
CAUTION:
Generic Formulation
Hydrogen peroxide
6%
Quaternary ammonium
0.2 %
Iodophors
50 mg/liter
Glutaraldehyde
2%
CAUTION:
w Do not autoclave or gas sterilize the skin temperature probe. Do not immerse
the probe in liquid cleaner. Avoid placing excessive strain on the probe lead.
Always remove the probe by grasping the plug at the panel. Do not pull on the
probe lead. These precautions will help avoid damage to the probe.
Note: Some cleansers may discolor the skin temperature probe over time.
Note: Disposable skin temperature probes cannot be cleaned.
4-2 66000371000DWG
6600-0371-000
100
10/27/03
4-2
5/Maintenance
In this section
5.1
5.2
5.3
5.4
5.5
5.6
CAUTION:
w Only competent individuals trained in the repair of this equipment should attempt to
service it as detailed in the Service Manual (Stock Number 6600-0195-000).
Contact the nearest Ohmeda Medical Regional Service Office for assistance or before
returning equipment for servicing. Obtain a Return Authorization Number before retuning
equipment.
If you send equipment to the Datex-Ohmeda Service and Distribution Center, package it
securely for protection, in the original shipping container if possible, and ship it prepaid.
Enclose the following five items:
1. A letter describing in detail any difficulties experienced with the equipment. Reference the Return Authorization number
2. Warranty information a copy of the invoice or other applicable documentation
must be included.
3. Purchase order number to cover repair of equipment not under warranty.
4. Ship to and bill to information.
5. Person (name and telephone number) to contact for functional questions.
In all cases, other than where Ohmeda Medicals warranty is applicable, repairs will be
made at Ohmeda Medicals current list price for replacement part(s) plus a reasonable
labor charge.
5-1 66000371000DWG
6600-0371-000
100
10/27/03
5-1
5/Maintenance
5.2 Circuit Breaker Reset
(See Figure 5-1.)
The warmer is equipped with a combination power switch and manual-reset circuit
breaker located on the left side of the controller assembly (as viewed from the front),
near the power cord socket. The circuit breaker limits the maximum current drawn.
If this circuit breaker trips when the warmer is operating, the power switch is deactivated to the Off position. To reset the circuit breaker, return the power switch to the On
position. If the circuit breaker trips again, service is required.
CI.05.009
Figure 5-1
Circuit breaker reset
w Disconnect the Warmer power cord and allow the unit to cool before
replacing the alarm or observation lights.
5-2 66000371000DWG
6600-0371-000
100
10/27/03
5-2
5/Maintenance
2. Use a Phillips head screwdriver to remove the lens mounting screw located in the
center of the alarm light.
3. Remove the bulb by pushing in and turning it counterclockwise.
4. Install the new bulb by pushing in and turning it clockwise.
Note: Replace both bulbs for maximum lamp life.
5. Place the lens cover in position and secure it with the mounting screw.
6. Plug the power cord in and check for proper operation.
CI.05.077
Lamp
Lens
Mounting Screw
Figure 5-2
Alarm Lamp replacement
5-3 66000371000DWG
6600-0371-000
100
10/27/03
5-3
CI.10.011
5/Maintenance
Figure 5-3
Heater rotation to the side position
CI.10.012
3. Refer to Figure 5-4. Use a Phillips head screwdriver to remove the back panel of
the heater assembly.
Figure 5-4
Observation Lamp cover removal
4. Refer to Figure 5-5. While holding the lamp with one hand, use your other hand to
pull the lever next to the lamp toward you. Remove the lamp.
5. Place the new lamp in position and push it into the lamp socket.
6. Replace the cover and the mounting screws.
7. Rotate the heater assembly back to its normal operating position.
8. Plug the power cord in and switch the observation lamp on. Check for proper
operation.
5-4 66000371000DWG
6600-0371-000
100
10/27/03
5-4
5/Maintenance
CI.05.013
Lever
Lamp
Figure 5-5
Observation Lamp replacement
w Never oil or grease oxygen equipment. Oils and grease oxidize readily,
and in the presence of oxygen, will burn violently. Vac Kote is the oxygen
service lubricant recommended (Stock No. 6700-0092-200) if the use of a
lubricant is specified.
Periodically (at least once a year) lubricate the Tee handle screw threads with a small
amount of oxygen service lubricant. This prolongs their life and makes sealing of the
yoke gaskets easier.
(See Figure 3-9)
Periodically (at least once a year) replace the yoke check valve strainer nipples before
they become clogged with lint or dust.
CAUTION:
5-5 66000371000DWG
6600-0371-000
100
10/27/03
5-5
5/Maintenance
5.6 Maintenance Schedule
The unit should be maintained in accordance with the procedures detailed in the
Service Manual. Service maintenance must be performed by a technically competent
individual as described in the Repair Policy.
A. Operator Maintenance
This schedule lists the minimum frequencies. Always follow hospital and local regulations for required frequencies.
Weekly or after each
patient
Annually
B. Service Maintenance
This schedule lists the minimum frequencies. Always follow hospital and local regulations for required frequencies.
Per hospital protocol
Annually
Annually
5-6 66000371000DWG
6600-0371-000
100
10/27/03
5-6
6/Illustrated Parts
Figure 6-1
Illustrated parts
Stock Number
Description
0305-5060-300
0217-5221-300
6600-0228-801
6600-0004-853
6600-0004-855
1.
2.
3.
4.
5.
CI.06.002
Mattress 46.2 x 64 cm
Mattress Support Panel
Corner Block Kit, gray
Left and Right Side Panel, gray
Front and Rear Side Panel, gray
Tubing organizer panel, gray-model 2001 & series 3000 (Figure 3-6)
HFOV Panel, gray
Adapter Plate (Figure 6-2)
Air Flowmeter with DISS Fittings (50 psi) (Figure 6-3)
Air/Oxygen Yoke and Regulator (all model - U.S.)
Airway Manometer Assembly (Figure 6-5)
Alarm Lamp
Bird Dovetail Adapter (includes adapter plate show in Figure 6-2)
Collection Bottle, 1/4 Gallon
Collection Bottle Bracket
Collection Bottle Cap with Float Assembly
DISS Continuous 3 Mode Suction Regulator with Overflow Trap
Drawer Kit, Stationary, gray, 3 drawer
Drawer Kit, Stationary, gray, 2 drawer
Drawer Kit, Rotating, gray, 3 drawer
Drawer Kit, Rotating, gray, 2 drawer
Heat Reflective Probe Patch (Box of 50)
Instrument Shelf, gray 30 x 30 cm (Figure 6-12)
IV Pole, 30 cm (Figure 6-4)
Monitor Shelf, gray, 30 x 66 cm (Figure 6-13)
Observation Lamp
Oxygen Flowmeter with DISS Fittings (50 psi)
Oxygen Yoke and Regulator Assembly (Figure 6-6 - all models - U.S.)
Spot Phototherapy Lamp Rail Mounting Kit
Power Cord, 120 Volt
Routing Clip, gray (Package of 6 )
Service Manual, all models
6-1 66000371000DWG
6600-0371-000
100
10/27/03
6-1
6/Illustrated Parts
0208-0697-700
6600-0208-700
6600-0196-700
6600-0001-850
6600-0004-801
6700-0136-900
6600-0018-800
0217-5374-800
0217-5369-800
0217-5367-800
0217-5357-800
For more information about these and other accessories see the
Infant Care Equipment Accessories Catalog (Form No F058)
6-2 66000371000DWG
6600-0371-000
100
10/27/03
6-2
6/Illustrated Parts
Figure 6-2
Adapter Plate
(0217-5363-800)
Figure 6-3
Air Flowmeter (50 psi)
with DISS Fittings (02175372-800)
Figure 6-4
IV Pole, 30 cm
(0217-5378-800)
Figure 6-5
Airway Manometer
Assembly
(0217-5377-800)
Figure 6-6
Oxygen Yoke and
Regulator Assembly
(see list on 6-1 and 6-2)
Figure 6-7
Routing Clip
(Package of 6)
(0217-5290-870)
6-3 66000371000DWG
6600-0371-000
100
10/27/03
6-3
6/Illustrated Parts
Figure 6-8
Utility Post,
3.5 Inch stub
(0217-5374-800)
Figure 6-9
Vacuum Manifold
with DISS Fittings
(0217-5369-800)
Figure 6-10
Vacuum Slide Bracket
(0217-5367-800)
Figure 6-11
Drawer Module
Stationary
(See list on 6-1 and 6-2)
Figure 6-12
Instrument Shelf
(See list on 6-1 and 6-2)
Figure 6-13
Monitor Shelf
(See list on 6-1 and 6-2)
6-4 66000371000DWG
6600-0371-000
100
10/27/03
6-4
Appendix
In this section
Temperature Conversion Chart ..................................................................................... A-1
Servo Mode Algorithm .................................................................................................. A-2
Infant Warmer System Specifications ..........................................................................
Electrical .............................................................................................................
Controller ............................................................................................................
Alarms ................................................................................................................
Environmental Specification ................................................................................
Electromagnetic Compatibility (EMC) .................................................................
Mechanical (Without Accessories) .....................................................................
Bassinet .............................................................................................................
Accessories ........................................................................................................
Radiant Energy Distribution .................................................................................
A-2
A-2
A-3
A-4
A-5
A-6
A-6
A-6
A-7
A-8
20.0
21.5
22.0
22.5
23.0
23.5
24.0
24.5
25.0
25.5
26.0
26.5
27.0
27.5
28.0
28.5
29.0
29.5
30.0
30.5
31.0
31.5
32.0
32.5
33.0
33.5
34.0
68.0
70.7
71.6
72.5
73.4
74.3
75.2
76.1
77.0
77.9
78.8
79.7
80.6
81.5
82.4
83.3
84.2
85.1
86.0
86.9
87.8
88.7
89.6
90.5
91.4
92.3
93.2
34.5
35.0
35.1
35.2
35.3
35.4
35.5
35.6
35.7
35.8
35.9
36.0
36.1
36.2
36.3
36.4
36.5
36.6
36.7
36.8
36.9
37.0
37.1
37.2
37.3
37.4
37.5
94.1
95.0
95.2
95.4
95.4
95.7
95.9
96.1
96.3
96.4
96.6
96.8
97.0
97.2
97.3
97.5
97.7
97.9
98.1
98.2
98.4
98.6
98.8
99.0
99.1
99.3
99.5
37.6
37.7
37.8
37.9
38.0
38.1
38.2
38.3
38.4
38.5
38.6
38.7
38.8
38.9
39.0
39.5
40.0
40.5
41.0
41.5
42.0
42.5
43.0
43.5
44.0
44.5
45.0
99.7
99.9
100.0
100.2
100.4
100.6
100.8
100.9
101.1
101.3
101.5
101.7
101.8
102.0
102.2
103.1
104.0
104.9
105.8
106.7
107.6
108.5
109.4
110.3
111.2
112.1
113.0
A-166000371000DWG
6600-0371-000
100
10/27/03
A-1
Appendix
Servo Mode Algorithm
The servo mode uses the difference between the servo mode control temperature and the
patient skin temperature to determine the percent of heater power required:
Control Temp - Patient Temp
0.45C
0.44
0.29
0.24
0.19
0.14
0.09
0.04
0.05
0.10
0.15
0.20
0.25C
to
to
to
to
to
to
to
to
to
to
to
95
90
85
75
65
45
25
20
15
10
5
0
WARNING:
A-266000371000DWG
6600-0371-000
100
10/27/03
A-2
Appendix
CircuitBreaker:
Rated Current: 7.5 A.
Trip Point: 9.45 A Minimum.
Type: Manual Resetting.
Model: Airpax Snapak.
Isolation Voltage:
2500 Vrms 60 Hz from the patient probe to the ac phase and neutral lines for one minute.
WARNING:
w The patient probe is not isolated from earth ground. Any additional equipment used
with the Ohmeda Infant Warmer System must comply with UL 544, CSA 22.2, IEC
601, and VDE 750.
LeakageCurrent:
With ground wire open or connected and measured at an exposed metal surface, less than 100
microamperes on 120 V units (200 microamperes on 230 V units).
GroundResistance:
Less than 0.2 ohms as measured from the power cord plug
IEC 601-1 Specifications
Type of protection against electric shock:
Degree of protection against electric shock:
Mode of Operation:
Protection against hazards of explosion:
Protection against ingress of liquids:
Degree of electrical connection between
equipment and patient:
connection to the patient.
Degree of mobility:
Class1
Type B
Continuous
Not Protected
Not Protected
Equipment with an applied
part designed for non-conductive
Mobile
Controller
Electronics:
Microprocessor-based control system.
Self-test functions are performed at power up and during normal operation.
Power Control Method:
Proportional heat control with zero-voltage switching to minimize radiated and conducted EMI.
Observation Light:
Nominal illuminance output: 100 foot candles at center of mattress.
Estimated lamp life: 3,000 hours.
Temperature Sensing System:
Range: 30 - 42C
Accuracy: 0.3C
Resolution: 0.1C
Probe interchangeability: 0.1C
Probe Model Number: LA003 or LA005
A-366000371000DWG
6600-0371-000
100
10/27/03
A-3
Appendix
Elapsed Timer:
60 minute elapsed timer with hold mode and Apgar tones.
Manual Mode Heat Selector Range:
0 to 540 watts in 20 increments of 5% each.
Servo Mode Control Range:
35.0 to 37.5C in increments of 0.1C.
Alarms
Multiple audio tones:
Operator prompt tone.*
Alternating single tone.*
Alternating two tone.*
*For a more detailed description of the audible alarms see section 3.1D
6600-0371-000
100
10/27/03
A-4
Appendix
Note: Excessive EMI in the hospital environment can trigger the system failure alarm.
Note the error code, if any, and switch off the unit. Wait ten seconds and switch the
power back on. If the system fail alarm recurs, remove the warmer from use.
Check Patient Alarm:
Preheat Mode: Check Patient Alarm is deactivated.
Manual Mode: Single tone alarm activates if the heater has been energized at greater than
25% heat for 12 continuous minutes.
Servo mode: Alarm activates when the heater has been at full power for 12 continuous
minutes.
Alternating two tone alarm activates after 3 minutes if the Check Patient Alarm is not
silenced.
Power Failure Alarm:
Single tone alarm activates if the line power is interrupted. The power failure LED on the
display panel lights.
A rechargeable maintenance-free nickel cadmium battery powers the audio alarm and the
microprocessor for up to 10 minutes with a fully charged battery. If power is restored within
10 minutes, the mode of operation and the set point are recalled.
Heat Off Alarm:
The Light-Emitting-Diode (LED) indicator activates when the heater is in the side position.
The audio alarm activates after the heater has been in the side position for 5 minutes.
Environmental Specifications
Operating Temperature Range: 10 to 40C.
Transport and Storage Temperature Range: -25 to 60C.
Transport, Operating and Storage Pressure Range: 500 to 1060 hPa
Operating and Storage Relative Humidity Range: 0 to 95%.
Height - cm
Depth - cm
Width - cm
Integral bed
Free standing
Elevating bed
185
185
183-203
104
103
114
66
62
77
Minimum Wt**
kg
96
102
91
Includes bassinet.
**Drawers not included; for stationary drawers add 25 kg, for rotating drawers add 32 kg.
A-566000371000DWG
6600-0371-000
100
10/27/03
A-5
Appendix
Mattress dimensions:
Model
Integral bed
Drop side bassinet
Tub bassinet
Width - cm
46
46
33
Length - cm
64
67
69
Casters
Model
Integral bed
13 cm dia., 2 locking, 2 non locking
Free standing 5 cm dia., 2 locking, 4 non locking
Bassinet
13 cm dia., 2 locking, 2 non locking
Drawers
Model
Integral bed
Bassinet
3 drawers- 38 x 39 x 10 cm or
1 drawer 38 x 39 x 10 cm,
1 drawer 38 x 39 x 23 cm
3 drawers- 38 x 57 x 12 cm
Bassinet
Material:
Select oak and oak veneer. Optional custom stain colors are available.
Storage:
Three Drawer storage opening from either side or the front of the unit, depending on the
model ordered.
Accessories
Accessories include the following (see also Section 6/Illustrated Parts ):
Air/Oxygen Yoke and Regulator:
Pin indexed oxygen yokes accommodate two U.S. E size oxygen cylinders.
Additional yoke accommodates one U.S. E-size cylinder of compressed air.
Air and oxygen pipeline fittings.
Regulators: 358 14 kPa.
Cylinder pressure gauges: 0 to 20700 kPa.
A-666000371000DWG
6600-0371-000
100
10/27/03
A-6
Appendix
Oxygen Yoke and Regulator:
Pin indexed yokes accommodate two U.S. E size oxygen cylinders.
Oxygen pipeline fittings.
Regulator: 358 14 kPa.
Cylinder pressure gauge: 0 to 20700 kPa.
Rail Mounted Accessories:
WARNING:
w Overloading the shelves can affect the stability of the unit. Limit the load to
9 kg per instrument shelf, mounted to a single upright, and 23 kg per monitor shelf, mounted between the uprights.
WARNING:
w Overloading the drawers can affect the stability of the unit. Limit the load to
4.6 kg per drawer, 9 kg per deep drawer.
A-766000371000DWG
6600-0371-000
100
10/27/03
A-7
Appendix
Utility Post, 22 Inch (0217-5376-800)
2.5 cm diameter x 56 cm long post
Provides a rigid vertical mounting space for stack mounting of equipment. The post
should be mounted as low as possible to ensure the stability of the Warmer.
Vacuum Manifold with DISS Adapters
Two DISS vacuum adapters are mounted on a standard manifold block. The locations
of these adapters can be changed to any of the other tapped holes in the block to meet
special requirements.
Vacuum Bottle Slide Bracket
The standard vacuum bottle slide can be mounted on either upright.
Ventilator Mounting Post
The vertical mounting post is a non-locking 1 inch diameter shaft which pivots in an 22
cm radius from the upright. One end of the shaft extends 36 cm while the other
extends 16 cm. Ventilators, humidifiers, blenders or other user hardware can be
pivoted into optional position.
mW/cm2
Zone
Mattress
Average
Average
34
34
CI.14.001
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
xxxx
xxxx
xxxx Infant Zone - 15" (38 cm) wide
xxxx
A-866000371000DWG
6600-0371-000
100
10/27/03
A-8
Appendix
Installing wall mounted units
WARNING:
The model 3050, 3100, and 3150 wall mounted warmers were designed for locations where
permanent, infant radiant warmers attached directly to the room wall are desired. To adequately
install these models, pre-planning at the architectural design phase is required.
Drywall
Stud
Track
41 cm O.C.
Ceiling Track
CI.05.040
30 cm
Sheet Metal
Screws
Floor Track
Figure A-1
Wall cross section
A-966000371000DWG
6600-0371-000
100
10/27/03
A-9
Appendix
The following instructions describe the manufacturers recommended method of installation in
new construction. Use only the Ohmeda hardware provided to mount the model 3050, 3100,
and 3150. The installation should be approved by the appropriate State and Local authorities.
Deviation from these recommendations or attempts to retrofit existing construction should only
be undertaken by professionals experienced in structural design, who in turn are wholly responsible for the structural integrity of the mounting method they devise.
Pre-Installation Preparation
A pattern of two (2) Extra Duty Drywall Studs (STE-20 gauge) and one (1) 10 x 4 cm
Drywall Track (16 gauge) at 41 cm centers form the basic rough in for installing each
IWS 3050 unit. See Figure A-1. Verify that the studs and track are securely fastened
together at 30 cm intervals and to both the floor and ceiling tracks with #10 sheet
metal screws.
Warmer Installation
After the wall is finished and the room is completed you are ready to install the warmer.
The distance from the warmer heater head to the bassinet or bed mattress surface is
critical for proper operation of the warmer. This distance must be 69 - 5, +23 cm.
Before installing the model 3050 or 3100, you must determine the bed surface height
that will be used (for your convenience, a bed height label is provided to attach to the
wall to aid in visually aligning the bassinet/bed surface).
102.4 cm
41 cm O.C.
xxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxx
Mounting
Anchors
48 cm
CI.05.041
Hinge
Bracket
19.8 cm
68.6 cm
88.4 cm
Heater
Assembly
CL
Note: For proper ventilation and
access allow at least 10 cm clearance
around the heater assembly.
24 cm
All dimensions
are nominal
Figure A-2
Warmer mounitng overhead view
1. Locate the stud edges by pre-measurement or by using a stud finder. Find the center
of the drywall track 86 mm over from the stud edge. See Figure A-2.
2. Remove the hinge bracket from the warmer by removing the (2) hinge pins (one
on either side) that secure the bracket to the warmer heater head.
A-10
66000371000DWG
6600-0371-000
100
10/27/03
A-10
Appendix
3. Utilizing a level, position the hinge bracket on the wall. The bottom edge of the heater
mounting bracket on the model 3050 and 3100 must be 69 - 5, + 23 cm, inches from
the bed surface of the mattress that will be used. See Figure A-3. The model 3150
integral bed warmer already comes with 695 cm bed-to-heater spacing.
4. Use the hinge bracket as a template to mark the mounting holes on the wall.
5. Use a 19 mm diameter metal cutting hole saw to drill (4) holes through the dry wall
and through metal drywall tracks behind the wall.
WARNING:
w Enclosing the heater assembly inside a cabinet may prevent proper ventilation and may create a fire hazard. If the heater is enclosed in a cabinet, it
must be equipped with a power cut-off device that prevents operation
while the cabinet is closed.
Hinge
Bracket
16.5 cm
Heater Assembly
11.5 cm
69 - 5, + 23 cm
Bed Surface
CI.05.046
Figure A-3
Heater to bed distance
All dimensions
are nominal
A-11
66000371000DWG
6600-0371-000
100
10/27/03
A-11
Appendix
Straps
Metal
Channel
Ring
Cap
CI.05.042,043,044
Drywall
Track
Figure A-4
Toggler anchor installation
7.
Mount the hinge bracket on the wall with the flat washers, split ring washers and screws
provided. See Figure A-5. Tighten the screws with approximately 80 lb/in of torque.
Note: The screws provided (3/8 - 16 UNC, 2 a long) are for installations in 1/2
inch or 5/8 inch thick drywall. For applications involving double thicknesses of
drywall, longer screws will be required. To properly mount the warmer, at least 2 a
screw threads must be engaged.
Verify that the bracket is level and the distance to the bed surface is 69 - 5, + 23
cm.
WARNING:
w Keep hands clear of the hinge area when installing a wall mounted heater
assembly. A pinch hazard exists.
8. Mount the warmer by inserting the back of the unit into the hinge bracket mounted
on the wall, then securing it by tapping in the (2) hinge pins. Verify that the heater
assembly is rigidly secured to the wall, is level, and is parallel to the floor.
9. For the model 3050, attach the proper bed to heater spacing label provided with
the unit to the wall so that the bottom line of the label is 74 cm from the lower edge
of the hinge bracket.
WARNING:
A-12
66000371000DWG
6600-0371-000
100
10/27/03
A-12
Appendix
Drywall
19 mm Dia. Hole
44 cm
Flat Washer
Slit Ring Washer
Mounting Screw
CI.05.045
Stud
Track
Toggler
Hinge Pin
Hinge Bracket
Power Cord
Heater Assembly
For the model 3100 or 3150
perform steps 10 through 14. For the model 3050
skip to step 15.
Figure A-5
Mounting the warmer
10. Position the lower mounting bracket on the wall. Use the lower hinge bracket as a
template to mark the mounting holes on the wall for the lower bracket.
11. Remove the warmer from the wall by tapping out the (2) hinge pins from the upper
bracket.
12. Remove the lower hinge bracket from the warmer by removing the (2) hinge pins
(one on either side) that secure the bracket to the warmer.
13. Repeat steps 5, 6 and 7 to properly install the lower mounting bracket.
14. Mount the warmer by inserting the back of the unit into the upper hinge bracket
mounted on the wall, then securing by tapping in the (2) hinge pins. Secure the
warmer to the lower mounting bracket by tapping in the (2) hinge pins. Verify that
the heater assembly and side rails are rigidly secured to the wall and that the
heater is level and is parallel to the floor.
15. Refer to the Operation & Maintenance Manual for checkout procedures.
WARNING:
A-13
66000371000DWG
6600-0371-000
100
10/27/03
A-13
Notes
Appendix
A-14
66000371000DWG
6600-0371-000
100
10/27/03
A-14
Appendix
Warranty
This Product is sold by Ohmeda Medical under the warranties set forth in the following
paragraphs. Such warranties are extended only with respect to the purchase of this
Product directly from Ohmeda Medical or Ohmeda Medicals Authorized Dealers as
new merchandise and are extended to the Buyer thereof, other than for the purpose of
resale.
For a period of twelve (12) months from the date of original delivery to Buyer or to
Buyers order, but in no event for a period of more than two years from the date of
original delivery by Ohmeda Medical to an Ohmeda Medical Authorized Dealer, this
Product, other than its expendable parts, is warranted to be free from functional defects
in materials and workmanship and to conform to the description of the Product contained in this operation manual and accompanying labels and/or inserts, provided that
the same is properly operated under the conditions of normal use, that regular periodic
maintenance and service is performed and that replacements and repairs are made in
accordance with the instructions provided. This same warranty is made for a period of
thirty (30) days with respect to expendable parts. The foregoing warranties shall not
apply if the Product has been repaired other than by Ohmeda Medical or in accordance
with written instructions provided by Ohmeda Medical, or altered by anyone other than
Ohmeda Medical, or if the Product has been subject to abuse, misuse, negligence, or
accident.
Ohmeda Medicals sole and exclusive obligation and Buyers sole and exclusive remedy
under the above warranties is limited to repairing or replacing, free of charge, at
Ohmeda Medicals option, a Product, which is telephonically reported to the nearest
Ohmeda Medical Field Service Support Center and which, if so advised by Ohmeda
Medical, is thereafter returned with a statement of the observed deficiency, not later
than seven (7) days after the expiration date of the applicable warranty, to the Ohmeda
Medical Service and Distribution Center during normal business hours, transportation
charges prepaid, and which, upon Ohmeda Medicals examination, is found not to
conform with above warranties. Ohmeda Medical shall not be otherwise liable for any
damages including but not limited to incidental damages, consequential damages, or
special damages.
There are no express or implied warranties which extend beyond the warranties hereinabove set forth. Ohmeda Medical makes no warranty of merchantability or fitness for a
particular purpose with respect to the product or parts thereof.
A-15
66000371000DWG
6600-0371-000
100
10/27/03
A-15
World Headquarters
EC Representative
GE Healthcare
9900 West Innovation Drive
Wauwatosa, WI 53226-4856
USA
Tel 1 800 345 2700
Datex-Ohmeda Ltd.
Ohmeda House
71 Great North Road
+DWHOG+HUWIRUGVKLUH
AL9 5EN
Tel +44 1707 263570
Fax +44 1707 260065
Asia Representative
GE Healthcare
P.O. Box 900
FIN-00031 GE
Finland
Tel +358 10 39411
Fax +358 9 146 3310
GE Healthcare
Shanghai GE (China) Hi-tech Park
No1 Huatuo Road, Zhangjiang Hi-tech Park Pudong, Shanghai, P.R.China 201203
GE
1, 201203
Tel + (8621) 38777888
Fax + (8621) 38777402
Australia
1300 722 229
China
800 810 8188
India
1 800 425 7255
Brazil Only
GE Healthcare Clinical Sytems Equipamentos Mdicos Ltda
Av. Paulista, 37 - 13 andar
CEP: 01311-902 - Cerqueira Csar
So Paulo, SP - Brasil
Tel +55 11 3053 2500
Fax +55 11 3053 2573
Ohmeda Medical,
D'LYLVLRQRI'DWH[2KPHGD,QF
a General Electric Company
8880 Gorman Road
Laurel MD 20723
USA
66000371000DWG