Industrial Radiography Image Forming Techniques

Download as pdf or txt
Download as pdf or txt
You are on page 1of 99

GE

Inspection Technologies

Industrial Radiography
Image forming techniques

GE imagination at work
2007 General Electric Company. All Rights Reserved.
We reserve the right to technical modifications without prior notice.
GEIT-30158 (01/07)

GE imagination at work

Industrial Radiography
Image forming techniques

Digital radiography

CR-image of a weld
1

Introduction to the overview of


Industrial Radiography
Image forming techniques
The first overview Industrial Radiography was published by Agfa in the sixties for
educational and promotional purposes. Since then, a few improved editions were released,
each one with the latest image forming radiographic techniques added.
This updated version is compiled by Mr. J.A. de Raad, NDT-expert and consultant, who has a
considerable number of publications on the subject of non-destructive testing to his name.
Mr. A. Kuiper, an experienced specialist and tutor on industrial radiography, assisted him.
Both were involved in NDT during the many decades they were employed by RTD bv
(Non-Destructive Testing) with its headquarters in Rotterdam, the Netherlands.
Apart from the developments in conventional radiography with X-ray films, this overview
also describes the now mature methods of digital radiography using radiation sensitive
plate- and panel detectors including digitization of traditional film.
Moreover the separate small booklet entitled Radiographers Weld Interpretation
Reference, with a number of radiographs and their interpretation, now also has been
inserted as an additional chapter in this new issue.
We trust that this revised edition of Industrial Radiography will once again fulfill a need.

GE Inspection Technologies 2007

Contents
Introduction to the overview of Industrial Radiography
Image forming techniques
Preface

11

1.

Introduction to industrial radiography

13

2.

Basic properties of ionising radiation

17

Natural radioactive sources


Artificial radioactive sources
4.5 Advantages and disadvantages of artificial radioactive sources
4.6 Properties of radioactive sources
Activity (source strength)
Specific activity
Specific gamma-ray emission (k-factor)
Half-life of a radioactive source
5.

34
35

37

NDT equipment

23
24
24

5.1 X-ray equipment


Types of X-ray tubes
Bipolar X-ray tubes
Unipolar X-ray tubes
Special types of X-ray tubes
5.2 High voltage generators
5.3 Megavolt equipment
The Btatron
The linear accelerator (linac)
5.4 Radioactive sources
Average energy level (nominal value)
5.5 Source holders (capsules)
5.6 Transport- and exposure containers
5.7 Checking for container leakage

44
44
47

27

6.

49

3.1 Units
3.2 Definitions
Radioactivity
Ionisation dose rate
Ionisation dose
Absorbed energy dose
Equivalent dose (man dose)

27
28

6.1 Radiation images


6.2 Radiation filters
6.3 Intensifying screens
Lead screens
Steel and copper screens
Fluorescent screens
Salt screens
Fluorometallic screens

49
51
51

4.

31

7.

The X-ray film and its properties

57

7.1
7.2

Structure of the X-ray film


Radiographic image
Latent image
Developing the latent image
Characteristics of the X-ray film
Density (optical)
Contrast

57
57

2.1 Wavelengths of electromagnetic radiation


2.2 X-rays
2.3 Gamma-rays (-rays)
2.4 Main properties of X-rays and -rays
2.5 Radiation energy-hardness
2.6 Absorption and scattering
Photoelectric effect
Compton effect
Pair production
Total absorption/attenuation
2.7 Penetrating power
2.8 Filtering (hardening)
2.9 Half-value thickness
3.

Units and definitions

Radiation sources

4.1 X-ray tube


4.2 The anode
Cooling the anode
The focal spot
Effective focal spot size
4.3 Tube voltage and tube current
4.4 Radioactive sources (isotopes)

17
18
19
20
20
21

31
31

33
34
4

7.3

Radiation images, filters and intensifying screens

37

39
40

43

58

7.4

59

Replenishing
Stopbath
Fixing
Final wash
Drying in the drying cabinet
Roller dryers
Recommendations for the darkroom
Silver recovery
Automatic film processing
NDT-U (universal) film processor
NDT-E (economy) film processor
Checking the developing process and film archiving properties
PMC-strips to check the developing process
Thiosulphate-test to check the film archival properties
Storage of exposed films

7.5
7.6

Characteristic curve (density curve)


Gradient of the density curve
Average gradient
Effect of developing conditions on the density curve
Film speed (sensitivity)
Graininess

63
63

8.

Film types and storage of films

65

8.1
8.2
8.3
8.4

The Agfa assortment of film types


Film type selection
Film sizes
Handling and storage of unexposed films

65
68
68
68

9.

Exposure chart

71

10.7

71

11. Defect discernibility and image quality

95

11.1 Unsharpness
Geometric unsharpness
Inherent (film) unsharpness
Total unsharpness
11.2 Selection source-to-film distance
11.3 Other considerations with regard to the source-to-film distance
Inverse square law
Selection of radiation energy (kV)
Selection of gamma source
11.4 Radiation hardness and film contrast
11.5 Summary of factors that influence the image quality

95

102
103

12. Image distortion and useful film length

105

12.1 Image distortion


12.2 Useful film length

105
106

13. Image quality

109

13.1 Factors influencing image quality


13.2 Image quality indicators (IQI)
Wire type IQI according to EN 462-1
13.3 List of common IQIs
American IQIs
French IQIs
Duplex IQIs
13.4 Position of the IQI
13.5 IQI sensitivity values

109
110

9.1 Exposure chart parameters


Type of X-ray equipment
The radioactive source
Source-to-film distance
Intensifying screens
Type of film
Density
Developing process
9.2 Densitometer
9.3 Producing an exposure chart for X-rays
9.4 The exposure chart
9.5 Use of the exposure chart
9.6 Relative exposure factors
9.7 Absolute exposure times
9.8 Use of the characteristic (density) curve with an exposure chart

73
73
76
76
78
78
79

10. Processing and storage of X-ray films

83

10.1 The darkroom


Entrance and colour
Darkroom lighting
Darkroom layout
Tanks
10.2 Chemicals and film-development
Making-up processing solutions
Developer
Fixer
Developing time and bath temperatures
Film agitation

83

84

10.3
10.4
10.5

10.6

88
88
88

91

93

99
100

112

115
115

14. Film exposure and handling errors

117

15. Film interpretation and reference radiographs

121

15.1 Film interpretation


15.2 The film-interpreter
15.3 Reference radiographs
Weld inspection
Casting radiography
Examination of assembled objects

121
122
122

16. Digital Radiography

143

16.1
16.2
16.3
16.4
16.5
16.6
16.7
16.8

Digitisation of radiographs
Computed radiography (CR)
Direct radiography (DR)
Image quality of digital radiography - MTF and DQE
Comparison of film, CR and DR methods
Selection of the CR and DR methods
Applications of the CR and DR methods
Work station and software

144
145
148
150
152
153
154
156

17.

Special radiographic techniques

161

17.1 Image magnification techniques


17.2 Fluoroscopy, image intensifiers
Stationary real-time installations
Portable real-time equipment
17.3 Computer Tomography (CT)
17.4 Neutron radiography (neutrography)
17.5 Compton back scatter technique

165
166
167

18. Special radiographic applications

169

18.1 Measuring the effective focal spot


18.2 Radiographs of objects of varying wall thickness
18.3 Radiography of welds in small diameter pipes
Elliptical technique
Perpendicular technique
18.4 Determining the depth position of a defect
18.5 Determining the depth position and diameter
of reinforcement steel in concrete
18.6 On-stream inspection
Projection technique
Tangential technique
Selection of source, screens and filters

169
170
171

161
162

Exposure time
18.7 Flash radiography

177

19. Radiation hazards, measuring- and recording instruments

179

19.1 The effects of radiation on the human body


19.2 Responsibilities
The client
The radiographer
19.3 The effects of exposure to radiation
19.4 Protection against radiation
19.5 Permissible cumulative dose of radiation
19.6 Radiation measurement and recording instruments
Radiation measuring instruments
Dose rate meters
Scintillation counter
Personal protection equipment
Pendosismeter (PDM)
Thermoluminescent dose meter (TLD-badge)
Film dose meter (film-badge)
19.7 Dose registration
19.8 Radiation shielding
Distance
Absorbing barrier and distance

179
179

20

Standards, literature / references ,


acknowledgements and appendices
European norms (EN-standards)
Literature and references
Acknowledgements
Appendices: tables and graphs.

173
174
174

180
180
181
182

184
184

187
187
188
188
189

Preface
To verify the quality of a product, samples are taken for examination or a non-destructive
test (NDT) is carried out. In particular with fabricated (welded) assemblies, where a high
degree of constructional skill is needed, it is necessary that non-destructive testing is
carried out.
Most NDT systems are designed to reveal defects, after which a decision is made as to
whether the defect is significant from the point of view of operational safety and/or
reliability. Acceptance criteria for weld defects in new constructions have been specified
in standards.
However, NDT is also used for purposes such as the checking of assembled parts, the
development of manufacturing processes, the detection of corrosion or other forms of
deterioration during maintenance inspections of process installations and in research.
There are many methods of NDT, but only a few of these allow the full examination of a
component. Most only reveal surface-breaking defects.
One of the longest established and widely used NDT methods for volumetric examination
is radiography: the use of X-rays or gamma-rays to produce a radiographic image of an
object showing differences in thickness, defects (internal and surface), changes in
structure, assembly details etc. Presently, a wide range of industrial radiographic equipment, image forming techniques and examination methods are available. Skill and experience are needed to select the most appropriate method for a particular application.
The ultimate choice will be based on various factors such as the location of the object to
be examined, the size and manoeuvrability of the NDT equipment, the image quality
required, the time available for inspection and last but not least financial considerations.
This book gives an overview to conventional industrial radiography, as well as digital
(computer-aided) techniques and indicates the factors which need to be considered for
selection of the most suitable system and procedures to be followed.
At the end of the book a chapter is added describing aspects of radiation safety.

10

11

Introduction to industrial radiography


Image forming techniques

source
In industrial radiography, the usual procedure for producing a radiograph is to have a
source of penetrating (ionising) radiation (X-rays or gamma-rays) on one side of the
object to be examined and a detector of the radiation (the film) on the other side as show
in figure
1-1. The energy level of the radiation must be well chosen so that sufficient radiation is
transmitted through the object onto the detector.
The detector is usually a sheet of photographic film, held in a light-tight envelope or cassette having a very thin front surface that allows the X-rays to pass through easily.
Chemicals are needed to develop the image on film, which is why this process is called the
classic or wet process.

homogeneous
radiation

Nowadays, different kinds of radiation-sensitive films and detectors not requiring the use
of chemicals to produce images, the so-called dry process, are used increasingly. These
techniques make use of computers, hence the expressions; digital or computer aided
radiography (CR) or direct digital radiography (DR), see chapter 16.

object

Summarising, the image of radiation intensities transmitted through the component can
be recorded on:

cavity

screens
X-ray film
projection of defect on film

Fig. 1-1. Basic set-up for film radiography

12

A DR related technique that has been available for many decades is the one in which images are formed directly with the aid of (once computerless) radiation detectors in combination with monitor screens (visual display units: VDUs), see chapter 17. This is in fact is
an early version of DR.
These through transmission scanning techniques (known as fluoroscopy) the storage of
images and image enhancement are continually improved by the gradual implementation
of computer technology. Nowadays, there is no longer a clear division between conventional fluoroscopy with the aid of computers and the entirely computer-aided DR. In time
DR will, to some extent, replace conventional fluoroscopy.

The conventional X-ray film with chemical development, the wet process, or one of
the following dry processes:
A film with memory phosphors and a work station for digital radiography, called
computer-assisted radiography or CR.
Flat bed detectors and a computer work station for direct radiography, called DR.
A phosphorescent or fluorescent screen (or similar radiation sensitive medium)
and a closed-circuit television (CCTV) camera as in conventional fluoroscopy,
an early version of direct radiography.

13

By means of radiation detectors, e.g.: crystals, photodiodes or semiconductors in a


linear array by which in a series of measurements an image is built up of a moving
object. This method is applied in systems for luggage checks on airports.
The source of radiation should be physically small (a few millimetres in diameter), and as
X-rays travel in straight lines from the source through the specimen to the film, a sharp
image is formed of the specimen and discontinuities. This geometric image formation is
identical to the shadow image with a visible light source. The sharpness of the image
depends, in the same way, on the radiation source diameter and its distance away from
the surface on which the image is formed.
The classic film in its light-tight cassette (plastic or paper) is usually placed close behind
the specimen and the X-rays are switched on for an appropriate time (the exposure time)
after which the film is taken away and processed photographically, i.e. developed, fixed,
washed and dried. In direct radiography (DR), a coherent image is formed directly by
means of an computerised developing station. The two methods have a negative image in
common. Areas where less material (less absorption) allows more X-rays to be transmitted to the film or detector will cause increased density. Although there is a difference how
the images are formed, the interpretation of the images can be done in exactly the same
way. As a result, the DR- technique is readily accepted.

At the limits of image detection it can be shown that contrast and unsharpness are interrelated and detectability depends on both factors.
As an image on a photographic film is made up of grains of silver, it has a grainy appearance, dependent on the size and distribution of these silver particles. This granular appearance of the image, called film graininess, can also mask fine details in the image.
Similarly, in all other image forming systems these three factors are fundamental parameters. In electronic image formation, e.g. digital radiography or scanning systems with
CCTV and screens, the factors contrast, sharpness and noise are a measure for the image
quality; pixel size and noise being the (electronic) equivalent of graininess (pixel size).
The three factors: contrast, sharpness and graininess or noise are the fundamental parameters that determine the radiographic image quality. Much of the technique in making
a satisfactory radiograph is related to them and they have an effect on the detectability of
defects in a specimen.
The ability of a radiograph to show detail in the image is called radiographic sensitivity.
If very small defects can be shown, the radiographic image is said to have a high (good)
sensitivity. Usually this sensitivity is measured with artificial defects such as wires or
drilled holes. These image quality indicators (IQIs) are described in chapter 13.

The classic film can be viewed after photochemical treatment (wet process) on a film
viewing screen. Defects or irregularities in the object cause variations in film density
(brightness or transparency). The parts of the films which have received more radiation
during exposure the regions under cavities, for example appear darker, that is, the film
density is higher. Digital radiography gives the same shades of black and white images,
but viewing and interpretation is done on a computer screen (VDU).
The quality of the image on the film can be assessed by three factors, namely :
1. Contrast
2. Sharpness
3. Graininess
As an example, consider a specimen having a series of grooves of different depths machined in the surface. The density difference between the image of a groove and the background density on the radiograph is called the image contrast. A certain minimum image
contrast is required for the groove to become discernible.
With increased contrast:
a. the image of a groove becomes more easily visible
b. the image of shallower grooves will gradually also become discernible
Assuming the grooves have sharp-machined edges, the images of the grooves could
still be either sharp or blurred; this is the second factor: image blurring, called image
unsharpness.
14

15

Basic properties
of ionising radiation
In 1895 the physicist Wilhelm Conrad Rntgen discovered a new kind of radiation, which
he called X-rays. The rays were generated when high energy electrons were suddenly
stopped by striking a metal target inside a vacuum tube the X-ray tube.
It was subsequently shown that X-rays are an electromagnetic radiation, just like light,
heat and radiowaves.

2.1 Wavelengths of electromagnetic radiation


The wavelength lambda () of electromagnetic radiation is expressed in m, cm, mm,
micrometer (m), nanometer (nm) and ngstrom (1 = 0.1 nm).
Electromagnetic radiation

Wavelength

10 km

104

1 km

103

100 m

102

10 m

101

1m

10 cm

10-1

1 cm

10-2

1 mm

10-3

100 m

10-4

10 m

10-5

1 m

10-6

Heat-rays, Infra-red rays,


micro waves

100 nm

10-7

Visible light and Ultraviolet (UV)

100 eV

10 nm

10-8

1 keV

1 nm

10-9

10 keV

0.1 nm

10-10

100 keV

0.01 nm

10-11

1 MeV

1 pm

10-12

10 MeV

0.1 pm

10-13

100 MeV

0.01 pm

10-14

X-ray energy

Table 1-2. Overview of wavelength, energy and type of electromagnetic radiation

16

17

Type

X-rays and Gamma-rays


(Radiography)

2.3 Gamma-rays (-rays)

The radiation which is emitted by an X-ray


tube is heterogeneous, that is, it contains
X-rays of a number of wavelengths, in the
form of a continuous spectrum with some
superimposed spectrum lines.
See fig. 1-2.

Fig. 1-2. X-ray spectrum intensity/wavelength


distribution
The small peaks are the characteristic radiation of the
target material

Radioactivity is the characteristic of certain elements to emit alpha (), beta () or


gamma () rays or a combination thereof. Alpha and beta rays consist of electrically charged particles, whereas gamma rays are of an electromagnetic nature.

intensity

2.2 X-rays

Gamma rays arise from the disintegration of atomic nuclei within some radioactive substances, also known as isotopes. The energy of gamma-radiation cannot be controlled; it
depends upon the nature of the radioactive substance. Nor is it possible to control its
intensity, since it is impossible to alter the rate of disintegration of a radioactive substance.

The shortest wavelength of the spectrum is


given by the Duane-Hunt formula:

1,234
kV
wavelength

In which :
= wavelength in nanometers (10 -9 m)
kV = voltage in kilovolts

The average shape of the X-ray spectrum is generally the same however not truely identical
for different X-ray sets; it depends chiefly on the energy range of the electrons striking the
X-ray tube target and, therefore, on the voltage waveform of the high-voltage generator.
A constant potential (CP) X-ray set will not have the same spectrum as a self-rectified set
operating at the same nominal kV and current. The spectrum shape also depends on the
inherent filtration in the X-ray tube window (glass, aluminium, steel or beryllium).

Figure 2-2 shows the energy spectrum lines for Selenium75, Cobalt60 and Iridium192.
In practical NDT applications, sources (radio active isotopes) are allocated an average
nominal energy value for calculation purposes, see section 5.4. Spectrum components
with the highest energy levels (keV values) influence radiographic quality the most.
relative intensity

min=

Unlike X-rays, generated to a continuous spectrum, Gamma-rays are emitted in an isolated line spectrum, i.e. with one or more discrete energies of different intensities.

The energy imparted to an electron having a charge e, accelerated by an electrical potential V is (eV) so the energy of the electrons can be quoted in eV, keV, MeV. These same
units are used to denote the energy of an X-ray spectrum line.
The energy of a single wavelength is :

=h.v

.v=c

In which:
E = the energy in electronVolt (eV)
h = Plancks constant
v = frequency
c = the velocity of electromagnetic radiation, such as light (300,000 km/s)

Fig. 2-2. Energy spectrum (lines) for Se75, Ir192 and Co60

The heterogeneous X-rays emitted by an X-ray tube do not however have a single
wavelength, but a spectrum, so it would be misleading to describe the X-rays as (say)
120 keV X-rays. By convention therefore, the e- in keV- is omitted and the X-rays
described as 120 kV, which is the peak value of the spectrum.
18

19

energy (keV)

2.4 Main properties of X-rays and -rays

2.6 Absorption and scattering

X-rays and -rays have the following properties in common:

The reduction in radiation intensity on penetrating a material is determined by the


following reactions :
1. Photoelectric effect
2. Compton effect
3. Pair production

1. invisibility; they cannot be perceived by the senses


2. they travel in straight lines and at the speed of light
3. they cannot be deflected by means of a lens or prism, although their path can be bent
(diffracted) by a crystalline grid
4. they can pass through matter and are partly absorbed in transmission.
5. they are ionising, that is, they liberate electrons in matter
6. they can impair or destroy living cells

2.5 Radiation energy-hardness


Radiation hardness (beam quality) depends on wavelength. Radiation is called hard
when its wavelength is small and soft when its wavelength is long. In industry the quality
of the X-ray tube ranges from very soft to ultra hard. The beam quality is related to a tube
voltage (kV) range, or keV for isotopes.
The first two columns of table 2-2 below indicate the relationship hardness/tube voltage
range applied in NDT. The third column gives the related qualification of the radiation
effect, i.e. half-value thickness (HVT), described in detail in section 2.9.
Radiation quality

Tube voltage

Which of these reactions will predominate depends on the energy of the incident
radiation and the material irradiated.
Photoelectric effect
When X-rays of relatively low
energy pass through a material
and a photon collides with an
atom of this material, the total
energy of this photon can be
used to eject an electron from the
inner shells of the atom, as figure
3-2 illustrates. This phenomenon
is called the photoelectric effect
and occurs in the object, in the
film and in any filters used.

incident
X-rays
ejected
electron

Global half-value

Hardness

thickness for steel (mm)

Very soft

Less than 20 kV

Soft

20 60 kV

Fairly soft

60 150 kV

0.5-2

Hard

150 300 kV

2-7

Very hard

300 3000 kV

7-20

Ultra hard

meer dan 3000 kV

> 20

Table 2-2. Comparative values of radiation quality (hardness) against tube voltage.

20

Fig. 3-2. Photoelectric effect

Compton effect
With higher X-ray energies
ejected
(100 keV to 10 MeV), the interacelectron
tion of photons with free or
X-ray
weakly bonded electrons of the 100keV - 10 MeV
outer atom layers causes part of
the energy to be transferred to
these electrons which are then
ejected, as illustrated i figure 4-2.
scattered
radiation
At the same time the photons
will be deflected from the initial
angle of incidence and emerge
from the collision as radiation of
reduced energy, scattered in all
directions including backward, Fig. 4-2. Compton effect
known as back-scatter, see
section 17.5. In this energy band, the absorption of radiation is mainly due to the Compton
effect and less so to the photoelectric effect.
21

2.7 Penetrating power


ejected
electron

The penetrating power of X-radiation increases with the energy (hardness).


The relationship of energy and penetrating power is complex as a result of the various
mechanisms that cause radiation absorption. When monochromatic ( homogeneous single wave length) radiation with an intensity Io passes through matter, the relative
intensity reduction I/Io is proportional to the thickness t. The total linear absorption
coefficient () consisting of the three components described in section 2.6 is defined by
the following formula:

I
= .t
Io

ejected
positron
Fig. 5-2. Pair production

Total absorption/attenuation
The total linear absorption or
attenuation of X-rays is a combination of the three absorption
processes described above, in
which the primary X-ray energy
changes to a lower form of energy. Secondary X-ray energy arises of a different wavelength and
a different direction of travel.
Some of this secondary (scattered) radiation does not contribute to radiographic image forming
and may cause loss of image
quality through blurring or fog.

Expressed differently:

In which:
Io = intensity at material entry
I = intensity at material exit
= total absorption coefficient

I = Io.e-t

t = thickness
e = logarithm: 2.718

Figure 7-2 shows the resulting radiation


intensity (logarithmic) as a function of
increased material thickness, for soft and
hard homogeneous radiation.
When radiation is heterogeneous the
graphs are not straight, see figure 7-2, but
slightly curved as in figure 8-2.

Fig. 6-2 Absorption coefficient for steel plotted against radiation energy
PE = Photoelectric effect
C = Compton effect
PP = Pair production

The contribution of the various


causes of X-ray absorption to the total linear absorption coefficient () for steel plotted
against radiation energy, are shown in figure 6-2.

Fig. 7-2. Intensity of homogeneous


radiation as function of increasing
thickness

intensity

X-ray
> 1 MeV

The slope of the curves becomes gradually


shallower (because of selective absorption
of the softer radiation) until it reaches the
so-called point-of-homogeneity. After passing this point the coefficient of absorption
remains virtually unchanged, as if the radiation had become homogeneous.
The position of the point of homogeneity
varies with the nature of the material irradiated. The graph shows that with increasing
material thickness, softer radiation is filtered out, more than hard radiation.
This effect is called hardening.

hard radiation,
high tube voltage
soft radiation,
low tube voltage

penetrated material thickness

Fig. 8-2. Intensity of heterogeneous


radiation as function of increasing
thickness hard radiation
hard radiation
intesity

Pair production
The formation of ion pairs, see
figure 5-2, only occurs at very
high energy levels
(above 1 MeV). High-energy
photons can cause an interaction
with the nucleus of the atom
involved in the collision. The
energy of the photon is here used
to eject an electron (e-) and a
positron (e+).

soft radiation
points of homogeneity

penetrated material thickness

22

23

Table 2-2 shows the average HVT-values for steel, table 3-2 shows the values for lead.

2.8 Filtering (hardening)


All materials, for example a metal layer between the radiation source and the film, cause
absorption and filtering. The position of the metal layer plays an important role in the
effect it has. A metal layer in front of the object will harden the radiation because it filters out the soft radiation. The degree of hardening depends on the type and the thickness
of the material. This phenomenon is used to reduce excessive contrast (variation in density) when examining objects of which the thickness varies greatly.
A metal layer between the object and the film filters the soft scattered radiation that
occurs in the object, thereby increasing the contrast and consequently the image quality.
This method of filtering is for example applied in the use of Cobalt60 in combination with
exposure time reducing intensifying screens, which are sensitive to scattered radiation.
Lead, copper and steel are suitable filtering materials.

2.9 Half-value thickness

Element/Isotope

Symbol

Ceasium137
Cobalt60
Iridium192
Selenium75
Ytterbium169
Thulium170

Cs137
Co60
Ir192
Se75
Yb169
Tm170

Average energy level Half-value thickness


in MeV
in mm lead
0.66
8.4
1.25
13
0.45
2.8
0.32
2
0.2
1
0.072
0.6

Table 3-2. Half-value thickness for lead

For a heterogeneous beam the HVT is not constant; the second HVT is slightly larger
than the first. In general, in industry where relatively hard radiation is used, a fixed
average HVT is applied.

A convenient practical notion (number) of the linear absorption coefficient is the introduction of the half-value thickness (HVT). It quantifies the penetrating power of radiation for a particular type of material and is defined as the thickness of a particular material necessary to reduce the intensity of a monochromatic beam of radiation by half, as
shown in figure 9-2. This HVT-value depends on the hardness of radiation.

HVT

HVT

HVT

intensity

HVT

thickness

Fig. 9-2. Illustration of half-value thickness

24

25

Units
and definitions
3.1 Units
Until 1978 the International Commission of Radiation Units and Measurements
(ICRU) used the conventional radiation units of roentgen (R), rad (rd), and curie (Ci).
Since 1978 the ICRU has recommended the use of the international system units (SI)
with special new units for radiation quantities; the Becquerel, Gray and Sievert.
Table 1-3 shows the relationships of these new units to the older units.

Designation of quantity
Activity (A)
Ionization dose rate
Ionization dose
Absorbed energy
dose (D)
Equivalent dose (H)
H=D x RBE**

SI units
Name
Unit
Designation
Becquerel
1/s*
(Bq)
Coulomb (C) C/kg
Coulomb (C) C/kg.s
Ampre (A) or A/kg
Gray
J/kg
(Gy)
Sievert
J/kg
(Sv)

Table 1-3. Overview of new and old units


* disintegrations per second
** RBE = Relative Biological Effect

Formerly used
Name
Unit
Designation
Curie
Ci

Conversion
Old to SI
1 Ci = 37 GBq

Rntgen

R
R/s

1 R=2.58 x 10-4 C/kg

Rad

Rad

1 Rad = 0.01 Gy

Rem

Rem

1 Rem = 0.01 Sv

C = Coulomb = A.s
A = Ampre

J = Joule

In radiography and radiation safety, units are preceded by prefixes.


Table 2-3 shows the ones mostly used.
Prefix

Meaning

Value

Notation

pico

0.000000000001

10-12

nano

0.000000001

10-9

micro

0.000001

10-6

milli

0.001

10-3

kilo

1000

103

Mega

1000000

106

Giga

1000000000

109

Table 2-3. Prefixes

26

27

3.2 Definitions
Radioactivity
The activity of a radioactive source of radiation (isotope) is equal to the number of disintegrations per second. The SI-unit is the Becquerel (Bq) and is equal to 1 disintegration per
second. The Becquerel is too small a unit to be used in industrial radiography. Source
strengths are, therefore, quoted in Giga Becquerel (GBq).
1 Curie = 37 GBq, see table 2-3.

Ionisation dose rate


The output of an X-ray set or isotope per unit of time is generally quoted at one metre
distance from the source, and designated in C/kg.s, see table 2-3.

Ionisation dose
The ionising effect of radiation in one kilogram of dry air is used to define the ionisation
dose. The dose of radiation delivered is equal to the ionisation dose rate multiplied by the
amount of time during which radiation takes place.
The designation used is C.kg.
The output of an X-ray set, however, is quoted in Sievert per hour, measured at 1 metre
distance.

Absorbed energy dose


The radiation energy that is absorbed is expressed in Joules per kilogram (J/kg).
The SI-unit is called Gray (Gy) whereby 1 J/kg = 1 Gy.

Equivalent dose (man dose)


The Sievert (Sv) is the SI-unit for the biological effect of ionising radiation upon man. It
corresponds with the product of the absorbed energy dose gray (Gy) with a factor that has
been experimentally determined and that indicates the relative biological effect (RBE) of
the ionising radiation. For X- and -radiation this factor is equal to one, so that the Sievert
is equal to the Gray.

28

29

Radiation sources
4.1 X-Ray tube
The X-ray tube, see figure 1-4, consists ofconsists of a glass (or ceramic) envelope containing a positive electrode (the anode) and a negative electrode (the cathode) evacuated to
an ultra high vacuum [10 - 9 hPa (hectoPascal)].
The cathode comprises a filament that generates electrons. Under the effect of the electrical tension set up between the anode and the cathode (the tube voltage) the electrons from
the cathode are attracted to the anode, which accelerates their speed.
This stream of electrons is concentrated into a beam by a cylinder or focusing cup.
When the accelerated electrons collide with a target on the anode, part of their energy is
converted to X-radiation, know as X-rays.

cathode
focusing
cylinder
or cup

anode

4.2 The anode


filament

target

The target is generally made of tungsten. Not only because it has a high atomic number, but
also because of its high melting point (approx. 3400C).
It is essential to use a material with a high melting point because of the substantial amount
of heat dissipated as the electron-bombardment is concentrated (focused) on a very
small surface. Only a part (approx. 0.1 % at 30 keV; 1 % at 200 keV; 40 % at 30 to 40 MeV)
of the kinetic energy of the electrons is converted into X-radiation; the remainder is transformed into heat.

electron beam

Cooling the anode


The heat which accompanies the production of X-radiation is quite considerable, so that
the anode has to be cooled. This can be done in a variety of ways :
1. by natural radiation
2. by convection
3. by forced circulation of liquid or gas
4. by conduction

X-ray beam
Fig 1-4. Glass envelope X-ray tube

The focal spot


The area of the target which is struck by the electrons, see figure 2-4, is called the focal spot
or the focus. It is essential that this area is sufficiently large to avoid local overheating,
which might damage the anode.
From the radiographic point of view, however, the focus has to be as small as possible in
order to achieve maximum sharpness in the radiographic image. This focal loading is
expressed in Joule/mm2. A tungsten target can take a maximum loading of
200 Joule/mm2. A higher loading might damage the anode.

30

31

Effective focal spot size


The projections of the focal spot on a surface perpendicular to the axis of the beam of Xrays is termed the effective focal spot size or focus size, see figure 2-4. The effective
focus size is one of the parameters in radiography, see section 11-1.
The effective focus size has to be as small as possible in order to achieve maximum sharpness in the radiographic image. The dimensions of the focus are governed by:
1. The size of the focal spot, and
2. The value of angle , see figure 2-4.

4.3 Tube voltage and tube current


The voltage across the X-ray tube determines the energy spectrum and so the hardness of
the radiation, see figure 3-4. The intensity is proportional to the tube current, see figure
4-4. This graph shows that, contrary to a change in tube voltage, a change in tube current
does not shift the spectrum (in other words: the hardness does not change).

relative intensity

Conventional X-ray tubes have effective focal spot sizes in the range 4 x 4 mm to 1 x 1 mm.
There are fine-focus tubes with focal spots from 0.5 x 0.5 mm ad micro-focus tubes down
to 50 m diameter or even less.

relative intensity

It should be noted that when in radiography we speak of the size of the focus without specifying this more exactly, it is normally the effective focal spot size which is meant.

The effective focal spot size is determined in accordance with the procedure described in
EN 12543. For a practical alternative method see section 18.1.
KeV
1.
2.
3.
4.
5.

Dimension of the electron beam


Focal spot
Effective focal spot size
Anode target
True focus size

KeV

Fig 3-4. Energy spectra at varying tube voltages and constant tube current (here 10mA)

Fig. 4-4. Energy spectra at varying values for tube current


and constant high voltage (here 200 kV)

The energy spectrum is also influenced by the characteristics of the high voltage applied to
the tube. When the spectrum of one X-ray tube on constant voltage is compared with that
of another with a current of pulsating voltage, of the same kV value, both spectra will be
slightly different. With a current of pulsating voltage there are, during each cycle, moments
of relatively low voltage, during which there will be a greater proportion of soft X- rays,
with their side-effects. This means that a set working on a constant voltage will provide a
higher intensity of hard radiation than one on a pulsating voltage; although both working
at the same nominal kV value.
However, even identical X-ray tubes may also show differences in generated energy.
The energy generated by one 200 kV X-ray tube will not be true identical to the energy
generated by another X-ray tube with the same applied voltage, not even if they are the
same type of tube.
This behaviour impedes calibration in kV of X-ray sets. Another reason why it is hard to calibrate an X-ray tube within a small tolerance band is, that the absolute level and wave characteristics of the supplied high voltage are difficult to measure.
It follows that it is difficult to standardise and calibrate X-ray equipment as far as spectra
and kV-settings is concerned, which precludes the exchange of exposure charts,
see section 9.1. Each X-ray set therefore requires its own specific exposure chart.

Fig. 2-4. Effective focal spot size

32

33

4.4 Radioactive sources (isotopes)

4.6 Properties of radioactive sources

Natural radioactive sources

Activity (source strength)

The elements from this group which have been used for the purposes of industrial radiography are radium and mesothorium. These give a very hard radiation, making them particularly suitable for examining very thick objects.

The activity of a radioactive substance is given by the number of atoms of the substance
which disintegrate per second.
This is measured in Becquerels (Bq), 1 Becquerel corresponds to 1 disintegration per
second(1 Bq = 1/s).

A disadvantage of natural sources, next to their high cost, is that it is not possible to make
them in dimensions small enough for good quality images and still give sufficient activity.

Specific activity

Artificial radioactive sources

The specific activity of a radioactive source is the activity of this substance per weight unit,
expressed in Bq/g.

Artificial radioactive sources for NDT are obtained by irradiation in a nuclear reactor. Since
1947, it has been possible to produce radioactive isotopes this way in relatively large quantities and in a reasonably pure state and particularly of sufficiently high concentration; the
latter being extremely important in NDT because the size of the source has to be as small
as possible. Among the many factors deciding a source suitability for non-destructive
testing are the wavelength and intensity of its radiation, its half-life and its specific radiation. In fact, only a few of the many artificial radio-isotopes available have been found to
be suitable for industrial radiography.

Specific gamma-ray emission factor (k-factor)


The k-factor is the generally used unit for radiation output of a source and is defined
as the activity measured at a fixed distance. It indicates the specific gamma-emission
(gamma constant) measured at 1 metre distance.
The higher the k-factor, the smaller the source can be for a particular source strength.
A source of small dimensions will improve the sharpness of a radiograph.
Table 1-4 shows the various k-factors and half-life values.

4.5 Advantages and disadvantages


of artificial radioactive sources
Advantages
1. require no electric power supply; easy to use in the field
2. can be obtained in a range of source diameters, so that if necessary a very short
source-to-film distance with a small diameter source can be used, for example,
for pipes of small diameter
3. a wide variety of radiation hardnesses
4. higher radiation hardness (more penetration power) than those of conventional X-ray
equipment can be selected

Isotope

Half-life

Ytterbium169
Iridium192
Selenium75
Cobalt60
Caesium137

31 days
74 days
120 days
5.3 years
30 years

Specific gamma constant


or k-factor
0.05
0.13
0.054
0.35
0.09

Table 1-4 Various k-factors and half-life values

Disadvantages

Half-life of a radioactive source

1.
2.
3.
4.
5.

Of an Iridium192 source with an activity of 40 GBq for example 10 GBq will remain after
two half-lives (148 days), 5 GBq after three half-lives (222 days) etc.

cannot be switched off


the energy level (radiation hardness) cannot be adjusted
the intensity cannot be adjusted
limited service life due to source deterioration (half-life)
less contrast than X-ray equipment

34

35

1a-5 Bipolar tube

NDT equipment
5.1 X-ray equipment
X-ray sets are generally divided in three voltage categories, namely:
1. Up to 320 kV, mainly for use on intermittent, ambulatory work. Tubes are generally of
the unipolar alternating current type. Higher voltages are hardly possible with this
type of equipment because of insulation problems.
2. Upto 450 kV, mainly for use on continuous, stationary or semi-ambulatory work,
because of their dimensions, limited manageability and weight.
Tubes are of the bipolar direct current type.
3. Upto 10 MeV, so called Megavolt equipment.
Virtually exclusively applied to stationary work.

Directional
X-ray beam
1b-5 Unipolar tube

Directional
X-ray beam

The first two categories are suitable for radiography on most common objects. Objects of
extreme thickness, however, require an energy even higher than 450 kV. In this case
Megavolt equipment is used, if alternative sources such as Cobalt60 prove unsuitable.
It will normally involve stationary installations of large dimensions and high weight. Lately,
portable versions have become available meant for ambulatory use.

Types of X-ray tubes


Depending on the shape of the anode, X-ray tubes produce :
a. a beam of radiation in one direction (directional tube)
b. an annular beam (panoramic tube)
X-ray tubes are either unipolar or bipolar.
1c-5 Hollow anode tube giving annular (panoramic) beam

Fig. 1-5. X-ray tubes


A = position of target

Panoramic
X-ray beam

36

Bipolar tubes
Figure 1a-5 shows a bipolar tube. The bipolar tube has the advantage that the potential difference with respect to earth on both the anode and the cathode is equal to one-half of the
tube voltage, which is a great help from the point-of-view of insulation. The exit window
is necessarily situated in the middle of the tube. Bipolar tubes usually operate on direct current and are generally air, oil or water cooled. They are designed to operate at voltages of
100 to 450 kV and a tube current of up to 20 mA.
Unipolar tubes
In these (shorter) tubes, as shown in figure 1b-5, the anode is held at earth potential and
the cathode only has a potential difference to earth. This makes anode cooling a simpler
operation. It also means that for low/medium kilo-voltage sets, up to approx. 300 kV as often
used in ambulant applications, a single simpler high voltage supply source will suffice.
The radiation window is placed asymmetric which can be advantageous in practice.
37

Special types of X-ray tubes


Unipolar X-ray tubes with a long hollow anode, as shown in fig. 1c-5, are generally known
as rod anode tube and can be inserted into pipes or vessels. These tubes produce an
annular (panoramic) beam over 360, so allowing a complete circumferential weld to be
radiographed in one exposure.
Figure 2-5 shows the conical anode
of a (360) panoramic tube, which
allows a circumferential weld to be
radiographed centrally, hence uniformly, from within. With this
anode the axis of the electron
beam must strike the top of the
cone in such a way that the centre
of the generated X-ray beam is
perpendicular to the longitudinal
axis of the tube.

Conventional (trans)portable X-ray equipment for use up to approximately 300 kV are provided with step-up HT transformers, rectifiers and smoothing capacitors. The X-ray tube
and the circuitry of this equipment are usually placed in an insulated tank. In most cases
these tank type sets use oil for insulation and cooling and weigh approximately 60 kg.
Gas is used when weight is important; the set than weighs approximately 30 kg.

X-ray

electron beam
cathode
anode
filament

Note: Anodes shaped so that the centre of the generated X-ray beam is
not perpendicular (oblique) to the
centre line of the tube (which was
acceptable in the past), are no longer
allowed when work is to be performed to official standards.

5.2 High voltage generators

X-ray

Figure 3-5 shows an integrated


(all-in-one) tank set for 300 kV. At
voltages over 300 kV housing everything in one tank becomes very
difficult because the high voltage
insulation would be inadequate.

Window

Equipment up to 450 kV operating


on direct current is connected to a
separate high tension (HT) supply
unit by means of HT leads, connectors and plug, see figure 4-5. As a
result this equipment is bigger and Fig. 3-5. All-in-one 300 kV tank set with an a-symmetric window
heavier than all-in-one tank sets and mostly meant for stationary or semi-ambulant use.
Window

Fig. 2-5. Anode configuration for an annular panoramic tube

There are also panoramic tubes in which the electron beam is focused over an extended
length by means of a magnetic lens or an electrostatic lens (Wehnelt-cylinder) to produce
a very small focal spot size. These sets are called micro-focus rod anode tubes with which
a very small focal spot size, of less than 10 micrometer, can be achieved. Since the anode
can be damaged relatively easy through overheating the anode is usually interchangeable.
This requires a separate vacuum unit in order to restore the vacuum after replacement. The
advantage of this construction is that with different types of anodes, different radiation
patterns can be obtained for special applications. The maximum energy level is usually
below 150 kV.
However, there are 150 kV micro-focus tubes with a fixed anode for enlarging or scanning
purposes, see section 17.1. With these tubes the tube current has to be kept low, because of
heat dissipation limitations of the non-interchangeable anode.

Fig. 4-5. Direct current X-ray tube for 450 kV with a symmetric window (Eresco)

A 300 kV tank set and a 450 kV direct current X-ray tube (the latter with separate high
voltage power supply) are of roughly the same dimensions.

Some X-ray tubes used in the radiography of plastics and aluminium are equipped with a
beryllium window to allow the softer radiation generated at the lower tube voltages of
5 to 45 kV, to pass.

Most tank sets are connected to a mains power supply with a frequency of 50 or 60 Hz. At
this frequency the supply voltage can be transformed upward.
This is followed by rectifying, which occurs in various forms. With some sets the X-ray tube
itself functions as rectifier, so called single-phase rectifying. If there is no smoothing
applied, considerable changes in voltage per cycle of alternating current will occur.
This periodic and greatly varying high voltage affects the intensity and spectrum of the
radiation generated, see section 4.3.

38

39

The intensity of radiation is increased by double-phased rectifying and varying degrees of


smoothing. At very low voltage ripple these sets are considered constant potential (CP)
equipment.

6
7

In the latest types of tank sets the mains frequency is first converted to a high frequency
alternating current and only then transformed upward, which makes it easier still to
smooth the ripple. At very high frequencies, up to 50 kHz, smoothing is hardly necessary
anymore and such x-ray sets can be called CP systems. Additional features may be built in,
for example an automatic warm-up facility, see note below. This type of circuitry with
advanced electronics leads to a higher degree of reliability and significant space and weight
reduction compared with earlier power supply systems. As a result of the various improvements that have gradually been implemented, present day AC-X-ray sets perform as well as
true CP sets.
Note: Because of the high vacuum prevailing inside the X-ray tube, it carefully has to be warmed-up
after a period of rest. During rest the vacuum deteriorates. This warm-up procedure has to be done
in accordance with the suppliers instructions, to prevent high voltage short-circuiting which might
damage the tube or render it useless.

5.3 Megavolt equipment


The equipment described in sections 5.1 and 5.2 is used to generate X-radiation up to
approximately 450 kV. However, sometimes higher energy levels are needed.
Several types of equipment have been built to operate in the 1 MeV to 10 MeV range. In
industrial radiography almost exclusively Btatrons or linear accelerators (linacs) are used.
Operating high-energy X-ray installations requires (costly) safety precautions.

The Btatron
The Btatron is an electron accelerator, which can produce X-radiation in the 2-30 MeV
energy range. The electrons are emitted into a round-sectioned donut shaped glass
vacuum tube, as shown in figure 5-5. After several millions of revolutions the electrons
reach maximum energy and are deflected towards the target. On the target, part of the
electron energy is converted into a tangentially directed beam of X-radiation.
To obtain a reasonably high radiation intensity, most Btatrons have been designed to operate in the 10-30 MeV energy range, as these voltages achieve maximum conversion rate of
electron energy into radiation. Even so the output of Btatrons is usually small compared
to linacs. Transportable low energy Btatrons (2-6 MeV) have been built, but these generally have a low radiation output, which limits their application.
One advantage of Btatrons is that they can be built with very small (micromillimeter)
focal spots.
A disadvantage is that with these very high energy levels the X-ray beam is usually narrow,
and the coverage of larger film sizes is only possible by using increased source-to-film
distances. The extended exposure times required can be a practical problem.
40

4
1

1.
2.
3.
4.&5
6.
7.

Ring-shaped accelerator tube


Anode
Cathode filament (emitting electrons)
magnetic fields
Coils
Auxiliary winding
X-rays

Fig. 5-5. Btatron

The linear accelerator (linac)


The energy levels mostly used for linacs (linear accelerators) are 4 MeV and 8 MeV. Linear
accelerators can be constructed for one or two energy levels.
In the travelling-wave linac, the acceleration of electrons from a heated filament to very
high energies results from the electrons riding a high-frequency (3-10 MHz) electromagnetic wave travelling in a straight line down an acceleration tube (the hollow guide).
The electrons are bunched into pulses at a frequency of a few hundred pulses per second.
The target, which the electrons strike to generate X-radiation, is at the opposite end of the
main wave guide of the filament assembly. This is a transmission type target from which
the radiation beam passes in a straight line.
The X-ray output from a linear accelerator is many times higher than from a Btatron of the
same energy. An 8 MeV linac with a 2 mm diameter focal spot can deliver a radiation dose
rate of 30 Sv/minute at 1 metre distance from the focus. Small light-weight portable linacs
of 3 MeV capacity can have outputs of 1.5 Sv/minute at 1 metre distance.

41

5.4 Radioactive sources


Table 1-5 shows various radioactive sources for industrial NDT. The most commonly
used ones are Cobalt, Iridium and increasingly Selenium. Seleneum is very attractive
while it permits lighter containers than Iridium. Due to its average energy level it often is
a good alternative for an X-ray tube, also attractive while no electricity is needed.

magnetron

vacuum pump

focus coils

Element

Mass
Number

Co
Cs
Ir
Se
Yb
Tm

60
137
192
75
169
170

X-rays

Symbol

target

electron gun
wave guide

Fig. 6-5. Linear electron accelerator (linac)

The main properties of a linear accelerator are:


1. very high output of radiation
2. very small focal spot dimensions (<2 mm)
3. considerable weight (approx. 1200 kg for an 8 MeV stationary installation)

Cobalt60
Caesium137
Iridium192
Selenium75
Ytterbium169
Thulium170

Average energy
level
in MeV
1.25
0.66
0.45
0.32
0.2
0.072

Table 1-5. Radioactive sources used in industrial radiography, in sequence of nominal (average) energy level

Average energy level (nominal value)


The spectrum of a source has one or more energy lines, as shown in figure 2-2. For sources
with multiple energy lines an average energy level is assumed, the so-called nominal value.
Source

Figure 7-5 shows an 8 MeV linac in a radiation bunker examining a pump housing.

Specific gamma
constant
k-factor
0.35
0.09
0.13
0.054
0.05
0.001

Number of
spectrum lines
Cobalt60
2
Caesium137
1
Iridium192
>10
Selenium75
>4
Ytterbium169
>6
Thulium170
2

Main energy
levels in MeV
1.17 en 1.34 MeV
0.66 MeV
0.3; 0.31; 0.32; 0.47 en 0.6 MeV
120, 140 and 400 keV
0.06 and 0.2 MeV
52 en 84 keV

Nominal value
in MeV
1.25 MeV.
0.66 MeV
0.45 MeV.
320 keV.
200 keV.
72 keV.

Table 2-5. Radiation spectra and nominal values

On the basis of these spectra data it is clear that Co60, CS137


and Ir192 sources produce high-energy radiation and are therefore well suited to irradiate thick materials.
source

Yb169, on the other hand, is a source that produces relatively


soft radiation and is of a very small size (0.5 mm), which makes
it particularly suitable for radiographic examination of circumferential welds in pipes of a small diameter and thin wall thickness, with the source centrally positioned so that the weld can
be exposed uniformly in one exposure, as shown in figure 8-5.

Fig.7-5. Linac and pump house in a radiation bunker

42

43

film
Fig. 8-5. Ytterbium169 source in
central position for exposure of
circumferential weld

that it has a certain minimal radioactivity, which is reason that in some countries the use of
depleted uranium is not allowed.

5.5 Source holders (capsules)


All gamma-ray sources for radiography are supplied in hermetically sealed, corrosion
resistant source holders (capsules), made out of monel, vanadium or titanium.
The Atomic Energy Authority in the country of origin encapsulates the radioactive material. The supplier will supply the source with a certificate which indicates the type of source,
its serial number, the activity at a certain date, and a disintegration graph.
The radiation material proper, also called the source or pellet, ranges in size from
1 to 4 mm. The size is dictated by the specific radiation activity of the source material.
The outside dimensions of the cylindrical capsule are approximately 5.5 x 15 mm, as shown
in figure 9-5.

Regardless of the shielding material used, all containers have a considerable weight in
common.
There are several solutions to the problem of safely storing a source on the one hand, and
of putting it in a simple but absolutely safe manner in its radiation position on the other
hand. Two regularly used constructions for this purpose are: source S is situated in a rotating cylinder, as shown in figure 11-5, or in an S-channel container as shown in figure 12-5.

15 mm
S

5.5 O
/

open
Fig. 9-5. Cross-section of a capsule for a radioactive source

Fig. 10-5. Sealed capsule

5.6 Transport- and exposure containers


Transportation and handling of sealed sources are subject to strict international safety
regulations, as a source is continuously emitting radiation in all directions, in contrast to an
X-ray tube which can be switched off. During transportation and use the source must be
surrounded by a volume of radiation absorbing material, which in turn is encapsulated in
a container. The level of radioactivity at the outside surface of the container shall not
exceed the legally established maximum limit.
Like the transport container, the exposure container needs to be robust and must function
safely at all times. The exposure container, also called camera, must be fail-safe and waterand dirt proof. It must also not be effected by impact. Moreover, if the radiation-absorbing
material, for example lead, melts (in a fire) the radiation absorbing qualities must not be
lost. This requires a casing made of a material with a high melting point, for example steel.
Besides lead, increasingly a new sintered material with very high tungsten content (97%)
is used as shielding material. This material is easily worked and finished and not prone to
melting.
Also greatly depleted uranium (with the highest radiation absorption) is used for shielding,
resulting in very compact exposure containers. A disadvantage of this material, however, is
44

closed

Fig 11-5. Exposure container with source S in a rotating inner cylinder

Fig. 12-5. S-channel container with


source S in storage position

The S-channel container is usually provided with a means to move the source out from a
distance (after all, distance is the safest protection from radiation). This may be done by
means of a flexible cable in a hose (Teleflex design) as shown in figures 13-5 and 14-5.
With this construction it is possible to extend the flexible hose in such a way that the source
can safely be moved several metres out of the container to the most favourable exposure
position.

casing/container

handling / operation side


shielding

exposure position
of the source
storage position
of the source
Fig. 13-5. Exposure container
with S-channel and flexible
operating hose ad cable

flexible connection

45

Figure 14-5 shows an S-channel container with a flexible


(metal) hose and cable in rolled up (transport) position.

handle
tungsten container

Figure 15-5 shows a more


recent (2006) S-channel
Selenium75 container with
operation hoses and pigtail.
Selenium75 radio-isotope is
becoming popular since new
production
(enrichment)
methods resulted in a much
better k-factor. Thus for a certain activity (source strength) Fig. 14-5. S-channel container with the flexible cable and deployment mechanism.
a much smaller source size (focus) is achieved. This results in a better/sharper image quality than could be achieved with the old Selenium75 production method.
Due to its average energy level of 320 kV, Selenium75 increasingly replaces X-ray
equipment for a thickness range from 5 mm to 30 mm of steel. This eliminates the need for
electric power, very attractive in the field for reasons of electrical safety and more
convenient at remote- or work locations with difficult access (high, deep, offshore,
refineries, etc). Last but not least, a Selenium container is of much lower weigth than
needed for an Iridium192 container with the same source strength.
To enable radiography on work sites with (many) people in the vicinity, for example on
offshore installations or in assembly halls, containers with rotating cylinders
and collimators were developed so that only the beam of radiation required for the
radiograph is emitted.
The remainder of radiation is absorbed by the collimator material which allows people to
work safely at a distance of a few metres while radiography is in progress. Such containers
with collimators are known by the name of CARE (Confined Area Radiation Equipment)
or LORA (Low Radiation) equipment.

rotating cylinder
source
Collimator

measuring tape

base

weld

pipe and weld


film
lead shielding

lead

boundaries of the beam of radiation


Fig. 16a-5. Gamma container with collimator
on a circumferential weld in a pipe

Fig. 16b-5. Cross-section of CARE/LORA container on the pipe

Without collimating the minimum safety distance is considerably more than 10 metres (in
all directions!).
Such containers with collimators are particularly suitable for frequent and identical repetitive
NDT work, for example radiographic testing of welds in pipes of < 300 mm diameter.
Figure 16a-5 shows such a special container with collimator set up for a double wall
radiograph. The cross-section drawing of figure 16b-5 shows the boundaries of the beam of
radiation. For bigger focus-to-film distances, longer collimators are used to restrict the
beam of radiation.
This type of container is suitable for Iridium sources up to 1000 GBq and weighs only
approx. 20 kg.

5.7 Checking for container leakage


A sealed radioactive source (capsule) might start to leak and become an open source as a
result of corrosion, mechanical damage, chemical reactions, fire, explosion etc.
Regular mandatory wipe-tests by specialists serve to detect leakage at an early stage.

Fig. 15-5. S-channel container for Selenium75 with pigtail (at right) and operating hoses (at left)

46

47

6
film
defect

negative
(shadow) image
on the film

primary radiation

Radiation images, filters and


intensifying screens
To influence the effects of radiation on an image, filters and intensifying screens are used to :
filter / harden the radiation to influence contrast and/or
to intensify the effect of radiation to improve contrast

6.1 Radiation images


object

The intensity of a beam of X-rays or gamma-rays undergoes local attenuation as it passes


through an object, due to absorption and scattering of the radiation. On a uniform object
attenuation of the primary beam will also be uniform and the film evenly exposed. If the
object contains defects or is of variable thickness, the surface of the film will be unevenly
exposed resulting in a shadow image of the object and the defects in it. When the film is
processed the variations in radiation intensity show up as varying film densities; higher
radiation intensity producing higher film density resulting in a negative X-ray image as
shown in figure 1-6.

radiation intensity
after passing
through the object

Fig. 1-6. The negative X-ray image

source (S)

Fig. 2-6. Image forming and non-image


forming radiation
Only the radiation from source (S) that
reaches the film in straight lines via beam
section DE, produces an image of cavity
N at P. The remainder, not reaching P
directly, is scattered radiation.

When the primary beam is partly absorbed in the object, some radiation,
as shown in figure 2-6, will be scattered and reach the film as secondary radiation by an
indirect path.
The quality of the radiograph is reduced by this scattered radiation, and it is important to
keep its effects to a minimum.
At any point P on the film, therefore, the total radiation reaching that point is made up of
some transmitted primary radiation forming the image of cavity (N), the image formingor direct radiation intensity Ip, and some secondary non-image forming , scattered radiation, intensity Is. Hence, the total radiation intensity at P is (Ip + Is).
The ratio Is/Ip is called the scattered radiation factor and can be as high as 10 for great
wall thicknesses, which means that the scattered radiation is ten times higher than the
image-forming radiation. The ratio (Ip+Is)/Ip = 1+Is/Ip is called the build-up factor and
is of considerable importance for the detectability of defects. It usually has a value between
2 and 20, depending on radiation energy and object thickness.
C

It must also be appreciated that any object in the neighbourhood of the object being examined (table, walls, ground and so on) which is struck by the gamma- or X-rays will partially reflect these rays in the form of back scatter which is liable to fog the film.

N
D

object

A
B

F
film

48

Backscatter coming from the object under examination is less hard than the primary radiation that has caused it and can be intercepted by a metal filter between object and film.
Radiation scattered by objects nearby the film can be intercepted by means of a protective
sheet of lead at the rear face of the film cassette.
49

Scattered radiation also occurs in radiographic examination of cylindrical objects,


as shown in figure 3-6.

6.2 Radiation filters


When a metal plate, usually lead or copper, is placed between the tube window and the
object, radiation hardening occurs leading to a lower image contrast.
This may be counter-balanced by a metal filter placed immediately behind the object (i.e.
between object and film). This filter will cause the (softer) scattered radiation passing
through the object to be absorbed by the filter to a greater extent than the primary (harder) radiation. This also improves the image quality.

Fig. 3-6.
Scattered radiation in radiography of cylindrical objects.
Scattered radiation from object 1
causes a spurious band at B,
object 2 at A etc, unless lead strips
are used as shown in the lower
part of this figure

If the edges of an object being radiographed are not close to the film (as in the case of a
cylindrical body in figure 3-6) considerable scatter of the primary radiation can occur,
leading to fogging. This scatter can be prevented by positioning sheets of lead foil between
the object and the film as illustrated in this figure.

X-rays

Reducing the contrast by filtration is also desirable when a radiographic image of an object
of widely varying thicknesses has to be obtained on a single film see section 18.2.
Typical filter thicknesses are :
0.1 0.25 mm lead for 300 kV X-rays
0.25 1.0 mm lead for 400 kV X-rays

6.3 Intensifying screens


1

A B

film

C D

The radiographic image is formed by only approximately 1 % of the amount of radiation


energy exposed at the film. The rest passes through the film and is consequently not used.
To utilise more of the available radiation energy, the film is sandwiched between two intensifying screens. Different types of material are being used for this purpose.

Lead screens
lead strips

film

The effects of scattered radiation can be further reduced by :


limiting the size of the radiation beam to a minimum with a diaphragm in front
of the tube window
using a cone to localise the beam, a so called collimator
the use of masks: lead strips around the edges of the object.
50

Under the impact of X-rays and gamma-rays, lead screens emit electrons to which the film
is sensitive. In industrial radiography this effect is made use of: the film is placed between
two layers of lead to achieve the intensifying effect and intensity improvement of approximately factor 4 can be realised. This method of intensification is used within the energy
range of 80 keV to 420 keV, and applies equally to X-ray or gamma-radiation, such as produced by Iridium192.
Intensifying screens are made up of two homogeneous sheets of lead foil (stuck on to a thin
base such as a sheet of paper or cardboard) between which the film is placed: the so called
front and back screens.
The thickness of the front screen (source side) must match the hardness of the radiation
being used, so that it will pass the primary radiation while stopping as much as possible of
the secondary radiation (which has a longer wavelength and is consequently less penetrating).
51

The lead foil of the front screen is usually 0.02 to 0.15 mm thick. The front screen acts not
only as an intensifier of the primary radiation, but also as an absorbing filter of the softer
scatter, which enters in part at an oblique angle, see figure 2-6. The thickness of the back
screen is not critical and is usually approx. 0.25 mm.
The surface of lead screens is polished to allow as close a contact as possible with the surface of the film. Flaws such as scratches or cracks on the surface of the metal will be visible
in the radiograph and must, therefore, be avoided. There are also X-ray film cassettes on
the market with built-in lead-screens and vacuum packed to ensure perfect contact between emulsion and lead foil surface.
Figure 4a-6 and figure 4b-6 clearly show the positive effect of the use of lead screens.
Summarizing, the effects of the use of lead screens are :
improvement in contrast and image detail as a result of reduced scatter
decrease in exposure time

Steel and copper screens


For high-energy radiation, lead is not the best material for intensifying screens. With
Cobalt60 gamma-rays, copper or steel have been shown to produce better quality radiographs than lead screens. With megavoltage X-rays in the energy range 5-8 MeV (linac)
thick copper screens produce better radiographs than lead screens of any thickness.

Fluorescent screens
The term fluorescence (often mistaken for phosphorescence) is used to indicate the characteristic of a substance to instantly instantly emit light under the influence of electromagnetic radiation. The moment radiation stops, so does the lighting effect. This phenomenon is made good use of in film based radiography. Certain substances emit so much
light when subjected to ionising radiation, that they have considerably more effect on the
light sensitive film than the direct ionising radiation itself..
The term phosphorescence is used to describe the same luminescent phenomenon, but
once the electromagnetic radiation ceases, light fades slowly (so called after-glow).
NDT additionally uses the memory effect of some phosphorous compounds to store
a latent radiographic image in order to develop it later into a visible image with the aid
of laser stimulation, see section 16.2. The image quality is mediocre because relatively
coarse phosphorous crystals are used. The possibility of producing memory phosphors
with smaller crystals is studied.

Salt screens
Fluorescent screens consist of a thin, flexible base coated with a fluorescent layer made up
from micro-crystals of a suitable metallic salt (rare earth; usually calcium tungstate) which
fluoresce when subjected to radiation. The radiation makes the screen light up. The light
intensity is in direct proportion to the radiation intensity. With these screens a very high
intensification factor of 50 can be achieved, which means a significant reduction in exposure time. The image quality, however, is poor because of increased image unsharpness.
Fluorescent screens are only used in industrial radiography when a drastic reduction of
exposure time, in combination with the detection of large defects, is required.

Fluorometallic screens
Fig. 4a-6. Radiograph of a casting without lead
intensifying screens

Fig. 4b-6. Radiograph of a casting with lead intensifying screens

Apart from fluorescent and lead intensifying screens, there are fluorometallic screens
which to a certain extent combine the advantages of both. These screens are provided with
a lead foil between the film base and the fluorescent layer. This type of screen is intended
to be used in combination with so-called RCF-film (Rapid Cycle Film) of the type Structurix
F6 or F8, see section 8.1.
The degree of intensification achieved largely depends on the spectral sensitivity of the
X-ray film for the light emitted by the screens.

52

53

To achieve satisfactory radiographs with fluorometallic screens, they should be used in


combination with the appropriate F-film type.
When used correctly and under favourable conditions, exposure time can be reduced by a
factor 5 to 10, compared with D7 film in combination with lead screens. This is not a constant factor because the energy level applied (radiation hardness) and ambient temperature also affects the extent of fluorescence. For example, at 200 kV a factor 10 can be achieved, but with Iridium192 (nominal value 450 kV) it will only be a factor 5 compared to D7
film. Table 1-6 shows the relative exposure factors for the RCF-technique.
Film system
200 kV
0.1
1.0

F6 + RCF screens
D7 + lead screens

Relative exposure time


Ir192 (450 kV)
0.2
1.0

Table 1-6. Relative exposure factors for RCF technique

F6+
RCF screen

Figure 6-6 gives an overview of graphs from which the relative exposure times can be deduced when using different films and screens at 200 kV, (for film-density 2). The graph shows
that an F8-film with RCF screen (point C) is approximately 8 times faster than a D8-film
with lead (point B) and approximately 15 times faster than a D7-film with lead (point A).
Since on-stream examination as well as examination of concrete, and also flash radiography (see section 18.7) allow concessions to image quality, a special fluorometallic screen
(NDT1200) has been developed with extremely high light emission. In combination with
an F8-film it may result in a reduction in exposure time at a factor 100 at 200 kV, against a
D7-film with lead (point D as opposed to point A in figure 6-6), or even a factor 140 to 165,
depending on source selection, see table 2-6. The intensification factor of the NDT1200
screens increases significantly at lower temperatures.
Table 2-6 shows the effect of radiation hardness on relative exposure times for the various
film/screen combinations compared with D7 film with lead screen.
Noticeably, for the NDT1200 screen and F-8 film the factor increases with the increase in
energy, but for the F6 film the factor decreases at energy levels exceeding 300 keV.es at
energy levels exceeding 300 keV.
Energy level
100 kV

density

density

A total processing cycle of a few minutes is possible with the use of an automatic film processor which makes it a very attractive system to deploy offshore (on lay barges) where
weld examination has to be done at a very fast rate and few concessions are made towards
image quality. Fig. 5-6 shows that a time saving at 10 (3.7-2.8) or 10 0.9 works out at approximately a factor 8. The actual time saving is often closer to factor 10.

On balance, the relative time saving is much smaller; usually no more than a factor 2 for an
F6-film (at Ir192 and Co60) instead of 10 in the D7 lead screen technique.
See the bold figures (2.5 and 1.7) in table 2-6.

D8+lead

D7+lead
F8+RCF
F8 +
NDT 1200

D7+lead

300 kV

Ir192
450 keV
Co60
1.25 MeV
log. rel.bel.

log. rel.bel.

Fig. 5-6. Relative exposure time RCF and lead


intensifying screen, for 300kV

Fig. 6-6. Speed comparison F8 film +NDT1200


and RCF versus D7 and D8 +lead, for 200kV

These RCF screens are also used for on-stream examination (see section 18.6), whereby
long exposure times and mostly hard (gamma) radiation are applied because of the penetrating power required. However, the relatively long exposure time (causing reciprocity)
and hard radiation (Cobalt60) together considerably reduce the light emission effect, as
tables 1-6 and 2-6 show.
54

Screen type
NDT1200
RCF
none
NDT1200+Pb
RCF
Lead
NDT1200+Pb
RCF
Lead
NDT1200
RCF
Lead

Relative exposure times


Film F8
Factor
Film F6
0.01
100
0.05
0.03
33
0.17

Factor
20
6

Film D7

1
0.008
0.02

125
50

0.04
0.13

25
8

0.007
0.035

140
30

0.06
0.4

17
2.5

0.006
0.04

165
25

0.1
0.6

10
1.7

Table 2-6. Relative exposure times for NDT1200, RCF and lead screens.

It is clear from the above tables and graphs that there are many ways to reduce the exposure time or radiation dose needed. The required image quality is decisive (a higher exposure rate automatically means reduced image quality), and next the economic factors, for
example the cost of the screens against time saved need to be weighed.

55

a.layer of
hardened gelatine
b. emulsion layer
c. substratum
(bonding layer)

d. cellulose
triacetate or
polyester base

c. substratum
b. emulsion layer
a.layer of
hardened gelatine

The X-ray film


and its properties

7.1 Structure of the X-ray film


An X-ray film, total thickness approx. 0.5 mm, is made up of seven layers,
see figure 1-7:
a transparent cellulose triacetate or polyester base (d).
On both sides of this base are applied:
a layer of hardened gelatine (a) to protect the emulsion
emulsion layer (b) which is suspended in gelatine, sensitive to radiation
a very thin layer called the substratum (c)
which bonds the emulsion layer to the base
The normal X-ray film, therefore, has two coatings of emulsion doubling the speed
compared to a film with a single emulsion layer. Photographic emulsion is a substance
sensitive to ionising radiation and light, and consists of microscopic particles of silver
halide crystals suspended in gelatine.
Note: In the past radiography on paper was not unusual. In this instant cycle process
results became available within 60 seconds. The quality of the images, however, was extremely poor and the life of the film limited to a few months.
The availability of better and faster instant cycle techniques such as digital radiography
(see chapter 16), has made radiography on paper obsolete.

7.2 Radiographic image


Fig. 1-7. Schematic cross-section of an X-ray film, total thickness approx. 0.5 mm.

Latent image
When light or X-radiation strikes a sensitive emulsion, the portions receiving a sufficient quantity of radiation undergo a change; extremely small particles of silver halide
crystals are converted into metallic silver.
These traces of silver are so minute that the sensitive layer remains to all appearances
unchanged. The number of silver particles produced is higher in the portions struck by
a greater quantity of radiation and less high where struck by a lesser quantity.
In this manner a complete, though as yet invisible, image is formed in the light-sensitive layer when exposure takes place, and this image is called the latent image.
Before and after exposure, but prior to development of the film, the latent image has a
shiny pale green appearance.

56

57

Developing the latent image

7.4 Characteristic curve (density curve)

Development is the process by which a latent image is converted into a visible image. This
result is obtained by selective reduction into black metallic silver of the silver halide crystals
in the emulsion. These crystals carry traces of metallic silver and in doing so form the latent
image. Several chemical substances can reduce the exposed silver halides to metallic silver:
these are called developing agents.

The characteristic or density curve indicates the relationship between increasing exposures
and resulting density. By exposure (E) is meant the radiation dose on the film emulsion. It
is the product of radiation intensity (Io) and exposure time (t), therefore: E = Io.t

The use of X-ray film and the definition of its characteristics call for an adequate knowledge of sensitometry. This is the science which studies the photographic properties of a film,
and the methods enabling these to be measured.
The density (or blackness) of the photographic layer, after development under closely defined conditions, depends on exposure. By exposure is meant a combination of radiation
dose striking the emulsion, that is to say intensity (symbol I) and the exposure time (symbol t). In sensitometry, the relationship between exposure and density (I.t) is shown in the
so-called characteristic curve or density curve.

Density (optical)
When a photographic film is placed on an illuminated screen for viewing, it will be observed that the image is made up of areas of differing brightness, dependent on the local optical densities (amount of silver particles) of the developed emulsion.
Density (D) is defined as the logarithm to base 10 of the ratio of the incident light Io and
the transmitted light through the film It, therefore: D = log (Io/ It) . Density is measured
by a densitometer, see section 9.2.
Industrial radiography on conventional film covers a density range from 0 to 4, a difference corresponding with a factor 10,000.

The curve is obtained by applying increasing exposures to a series of successive


areas of a strip of film, whereby each following exposure is a certain factor (for
example 2) greater than the previous one.
After development, the densities (D) are
a
measured by means of a densitometer and
then plotted against the logarithmic
log E (exposure) relative units
values of the corresponding exposures
(log E). The points obtained are then joi- Fig. 2-7. The characteristic curve for an
industrial X-ray film
ned together by a continuous line. It is not
necessary to know the absolute exposure values; relative values can be used, so at a fixed
X-ray intensity only exposure time needs to be changed.
density D

7.3 Characteristics of the X-ray film

The ratio between different exposures


and related densities is not usually plotted
on a linear scale but on a logarithmic
scale; i.e. density D versus log E.

Density (D) of a photographic emulsion does not increase linearly with exposure (E) over
the entire density range, but has a shape as in figure 2-7. The lower part of the curve (a-b)
is called the toe, the middle part (b-c) is called the straight line (linear) portion, and the
upper part (c-d) is called the shoulder.

Contrast
The contrast of an image is defined as the relative brightness between an image and the
adjacent background.
The contrast between two densities D1 and D2 on an X-ray film is the density difference
between them and is usually termed the radiographic contrast.

The shoulder of a characteristic curve relating to industrial X-ray film corresponds to densities higher than 6. Since such densities are too high for normal film viewing, the curve from
density D = 3.5 upwards is shown as a broken line.
It should be noted that the straight-line portion (b-c) is not truly straight, but slightly
continues the trend of the toe of the curve.

Film contrast, or emulsion contrast, are rather vague terms used to describe the overall
contrast inherent in a particular type of film. When an emulsion type shows most of the
image contrasts present, the film is said to be of high contrast or hard.
For the measurement of film contrast, the term film gradient is used, for which the symbol is GD. Suffix D indicates the density at which G is measured.
58

59

Gradient of the density curve


The density curve shows one of the most important characteristics of a film. The slope of
the characteristic curve at any given point is equal to the slope of the tangent line at this
point. This slope (a/b in figure 3-7), is called the film gradient GD, film contrast or the
film gamma.
D

therefore necessary to use the highest possible density radiograph, while remaining within
the acceptable density range of the viewing screen so as not to impede film interpretation..
Most codes of good practice ask for densities
between 2.0 and 3.0 in the relevant area of
the image.
Table 1-7 shows the loss in contrast on
typical film as density values obtained fall
below 3.0 .
Density D
D = 3.0

3.0
2.5
2.0
1.5
1.0

Film contrast as a %
of the value at

100
85
71
54
35

density

Fig. 5-7. Gradient/density curves


for three types of film:
A low sensitivity: very fine grain film (D2)
B average sensitivity: fine grain film (D4)
C high sensitivity: medium grain film (D8)

Table 1-7. Contrast loss with reduced film density

Fig. 4-7. Average gradient (a/b)


of an X-ray film

Fig. 3-7. Gradient of an X-ray film

Average gradient
The straight line connecting two points on a characteristic curve, as figure 4-7 shows, is
equal to the average gradient of the segment of the curve linking these two points. This
gradient (GD) is the average of all gradients in the segment between density values 3.50
and 1.50, and is a standard characteristic of a particular type of radiographic film.
In all films (for example D2 through to D8) the gradient (a/b) increases with increasing
density within the for conventional viewing screens useful density range of D<5.

The specimen in figure 6-7 containing a


small step is radiographed with an exposure
time resulting in a density difference of 0.5
(B minus A). If now, using the same type of
film and the same tube voltage, a longer
exposure time is given, the density difference
is 0.9 (D minus C).
The second radiograph, therefore, shows
more contrast.

The various types of films are not identical. This becomes clear if plotting the values of
gradient Gd against the density resulting in the gradient/density curves, as shown in
figure 5-7. At higher film sensitivity the gradient is lower and, hence, the density curve
less steep.

Fig. 6-7. Illustration of enhanced contrast


at increasing density

A steeper gradient means an increase in density difference at equal radiation dose and so a
greater contrast, resulting in better defect discernibility. If one requires high contrast, it is
60

density

log.rel.exp.

log.rel.exp.

61

Effect of developing conditions on the density curve

7.5 Film speed (sensitivity)

The characteristic curve of an X-ray film is not only determined by the emulsion characteristics but also by the way the film is developed. Parameters which can influence the characteristic curve are: developing time and its temperature, developer concentration and
agitation.
The effect of, for example, the developing time on speed (relative exposure factor),
contrast and fog, has been made visible in figure 7-7. Initially, up to approx. 4 minutes,
speed and contrast are low but increase rapidly with developing time.
From 8 minutes on, a further increase in developing time increases the background fog,
and eventually a decrease in contrast will occur.

relative exposure factor

developing time (min)

The generally accepted method of measuring the film speed of radiographic films is to measure the exposure required to achieve a density of 2.0 above base and fog, using a specific
processing technique. The various relative exposure values are shown in table 1-8.

7.6 Graininess
When a developed X-ray film is viewed in detail on an illuminated screen, minute density
variations are visible in a grainy sort of structure. This visual impression is called graininess
and a measurement of this phenomenon establishes a degree of granularity.

contrast

speed

contrast effect
(average gradient for
densities between
1.5 and 3.5)

development fog
Fig. 7-7.
Film characteristics at various
developing times

In radiography the relationship between exposure (in C/kg) and resulting density is commonly referred to as film speed. Other than in normal photography where film speed is
indicated by a DIN or ASA number, films for industrial radiography do not carry an internationally recognised speed number.

development fog
industrial X-ray film developer G128 at 20C

Although it is possible to compensate, to a certain extent, for minor variations from the correct radiation exposure by adapting the developing time, normally a fixed time is maintained. In manual developing the standard time is 5 minutes. Developer type, film agitation
in the tank and temperature also influence density. That is why the overall developing process should preferably be standardised or automated. In most cases, deviating from the
optimum developing conditions leads to reduced image quality.

62

63

image quality - better

Film types
and storage of films
Industrial X-ray films are produced by a limited number of manufacturers in an assortment
for use with or without intensifying screens and filters. The selection of a particular film
type not only depends on economics but in particular on the required, often prescribed,
image quality.

D2

8.1 The Agfa assortment of film types

D3

Agfas assortment of industrial standard radiographic film comprises the following types in
sequence of increased speed and granularity : D2, D3, D4, D5, D7 and D8, complemented
with the very fast films F6 and F8.

D4

Note: Industrial X-ray films made by Agfa carry the generic name STRUCTURIX. The various types
of STRUCTURIX film are identified by the number which follows the generic name, e.g. STRUCTURIX D4. For ease of reading the indication STRUCTURIX is omitted in the text from here on.

D5

D7
D8

Fig. 1-8. Image quality versus film speed


The film speed is presented on a linear scale

faster
film speed

In addition to these graphs, figure 13-16 gives a graphical representation of relative image
quality as a function of relative dose and exposure time (film speed) for D-films and
computer-assisted CR and DR techniques.
Agfa has developed special intensifying screens specifically for use in combination with F6
and F8 films, see section 6.3. These so-called rapid cycle film screens are usually referred
to as RCF-screens. F8 has the highest film speed. Depending on quality requirements, F6
is mostly used for weld inspection on lay-barges; since it shortens examination time by a
factor 10, see section 6.3.

D2
image quality - better

The ultra-fine-grain D2-film is used in the radiography of very small components, when
optical magnification is applied to allow very fine details to be observed.
D8 is used for the examination of big castings and steel reinforced concrete. D10 film is also
produced for exposure monitoring purposes, see section 19.6. Figures 1-8 and 2-8 show the
relationship between film speed and image quality and film contrast respectively.

D3
D4
D5
D7

D8
faster

Fig. 2-8. Contrast versus film speed

64

film speed
65

Figure 3-8 shows graphs of relative exposure time versus density for the entire Agfa D-film
range. For density 2, the difference between a D8 and a D2 film is a factor 14 (10(3.25-2.1) ),
at 200 kV.

Part of the Agfa film range with relative exposure factors and code classification has been
listed in table 1-8 for various radiation intensities :

density

Film type
Fig. 3-8. Density graphs of the
Agfa film range D2 through to D8
with lead intensifying screens at 200 kV
and automatic film development.

D2
D3
D4
D5
D7
D8
F6+RCF (5)
F8+RCF (5)

Filmtype

Relative exposure factors


100 kV
(1)
9.0
4.1
3.0
1.7
1.0
0.6
0.174
0.03

200 kV 300 kV
(2)
(3)
7.0
4.3
2.7
1.5
1.0
1.0
0.6
0.132
0.022

Ir192
(4)
8.0
5.0
3.0
1.5
1.0
0.6
0.389
0.035

Co60
9.0
5.0
3.0
1.5
1.0
0,6
0.562
0.040

Film system
Class
EN 584 -1
ASTM
E 1815
C1
Special
C2
1
C3
1
C4
1
C5
2
C6
3

Table 1-8.
Listing of various Agfa films and their relative exposure factors and film system classification

Note I: It is common practice to compare relative exposure factors with those of D7 film, which
are shown bold as reference value 1.0 in the table.
Note II: The numbers (1) to (5) used in the table indicate the use of the following screen types:
1 without lead screens
2 with lead screen 0.027 mm thickness
3 with lead screen 0.027 mm thickness
4 with lead screen, front 0.10 mm, back 0.15 mm thickness
5 with fluorometallic screen (RCF)
Note III: Developing process for table 1-8: automatic, 8 minute-cycle, 100 seconds immersion
time in developer G135 at 28C.

log.rel.exp.

Note IV: The relative exposure factor depends not only on radiation intensity, but also on exposure time and is, therefore, not a constant value.

Note: Developing process for figure 3-8 above:


Automatic, 8 min cycle, 100 seconds immersion time in developer G135 at 28C.

66

67

8.2 Film type selection


Most procedures and codes of good practice for the performance of industrial radiography
base the choice of type of film for a specific application on the EN or ASTM classification
systems. For weld inspection, when one is attempting to detect small cracks, a film of class
C2 or C3 would be specified. For the examination of castings or general radiography a film
of class C4 or C5 would normally be used. For small component inspection, where the
image might be viewed under magnification to reveal small details, a film of class C2 or
possibly even a single emulsion film of class C1 would be desirable.
In megavoltage radiography, because most equipment have a very high radiation output,
class C3 films can be used for objects of great wall thickness. This has the advantage that a
high film gradient can be achieved.

8.3 Film sizes


Film sizes in industrial radiography are to a large extent standardised according to
ISO 5565. Non-standard sizes are possible. Standard film sizes and metal screens are
supplied separately, but can also be supplied vacuum-packed so that the risk of film faults
is considerably reduced. For weld inspection there is so-called Rollpac film strip on the
market which is available on a roll together with the lead screen. For very large projects the
strip film can even be pre-cut to suit a particular weld length or pipe/vessel circumference.

8.4 Handling and storage of unexposed films


The conditions under which unexposed films are handled and stored play a very important
role in the final quality of the exposed film. Recommendations for handling and storage
are contained in, for example, ASTM E1254. Pre-exposure as a result of background radiation must be avoided as it causes unacceptable fogging of the film.
If films are to be kept for a longer period, the following storage conditions must
be adhered to:
background radiation levels below 90 nGray
temperatures below 24C
relative humidity levels below 60 %
away from X-ray film chemicals
preferably stacked on edge
In the long run, minor fogging will occur to films stored. Background fog to a maximum
density of 0.3 is considered acceptable.
68

X-ray photograph of a Van Gogh painting, presumably a self-portrait, on canvas.


X-rays are made to prove a paintings authenticity, check the condition of the canvas material,
or determine possible paint-overs.
Cadmium and/or lead in the paint absorb radiation, thus forming an image.
Applied X-ray process data: 30 kV, 10 mA.min, Agfa D4 without lead screens and a source-to-film
distance of 100 cm.

69

Exposure chart
9.1 Exposure chart parameters
Codes for the inspection of welds and castings specify the maximum allowed radiation
intensity, based on the type of material and the thickness of the object. Exposure charts are
necessary to etablish the correct exposure value. A universal exposure graph or slide-rule
can be used for radioactive sources, as these have a fixed natural radiation spectrum.
The radiation spectrum of X-ray tubes varies with each tube, even if they are of the same
type. This problem is easily solved by using a universal exposure chart for the specific type
of tube, and then individualise it for each tube, the so-called curve fitting. The adaptation
is normally limited to a zero-point correction, based on a few measured values obtained by
trial. Sometimes the gradient of the exposure graph needs to be adjusted as well.
An exposure chart is produced by making a series of radiographs of a step wedge as
illustrated in figure 1-9.
The radiation intensity level of most X-ray equipment is expressed by the amperage of the
current through the X-ray tube, measured in milliampres (mA).
The exposure (radiation dose) is specified as the product of radiation intensity and
exposure duration in mA.min. (intensity x time).
The exposure chart shows the relationship between the thickness of the object (in mm) and
the exposure value (for X-ray tubes in kV and mA.min; for sources in GBq/h).
The exposure chart is applied for:
1. a given density, for example: 2 or 2.5
2. a given film-screen combination, for example D7 with lead screens
3. a given type of material, for example steel
The chart depends amongst others on:
4. type of X-ray equipment or radioactive source
5. source-to-film distance, usually 800 mm
6. development conditions, for example: automatic, 8 minutes at 28C.

70

71

Type of X-ray equipment

9.2 Densitometer

Among the factors to be taken into account are: the voltage (in kV), whether alternating or
direct current, the limits of voltage adjustment and the current through the tube (in mA).
It follows that the exposure chart is unique for a particular X-ray set.

The radioactive source


Radiation intensity and half-life-time of the source have to be taken into account.

Source-to-film distance
The exposure chart for an X-ray set is produced for a specified source-to-film distance. If
another distance is used, corrections will be necessary, using the inverse square law.

Intensifying screens
When drawing up the exposure chart, intensifying screens used must be recorded and the
same type of screens used again when making radiographs.

Type of film
The type of film must be indicated on the exposure chart, since the various types of
industrial X-ray films are substantially different in sensitivity (speed).

A densitometer is used to accurately measure the photographic (optical) density at any spot
on a radiographic film. For most types of densitometers the size of the measured area is
approx. 1 mm2. The measuring range runs from density 0 to 4.
Since it is a logarithmic scale, this equals a factor 10,000 (104) in density.
It is very important to regularly recalibrate these instruments, particularly around values
2 and 2.3, since those are the minimum densities (depending on class: A or B) which a film
must have in accordance with standard EN-444, to allow it to be interpreted.
Densitometers are supplied with reference material (density strips) to re-calibrate them.
Regular recalibration, at least once a year according to code, is essential.
The most commonly used density strips deteriorate quickly as a result of scratching and disintegration of the sealed transparent wrapping in which they are usually kept.
Their service-life, depending on use, is usually not much longer than six months. Agfa has
developed the Denstep density step wedge film and has succeeded in considerably
extending the service-life of these strips by supplying them in special wear proof wrapping.
These density strips are certified and have a guaranteed minimum service life of four years.
The 15 steps of the Denstep comprise a density range from 0.3 to 4.

9.3 Producing an exposure chart for X-rays

Density
An exposure chart must be as accurate as possible. Densities indicated are to be measured
by a densitometer, see section 9.2. The radiographs that form the basis for the chart must
have been made under controlled and reproducible conditions, whereby quality monitoring tools such as PMC strips as described in section 10.6 are used.

Developing process
Developer formula, processing temperature and developing time all influence the final
result. The exposure chart produced will be related to a particular well-defined developing
process.

The step wedge


The production of an exposure chart calls for either a large
step wedge or a series of
plates of different thicknesses
made from the same material
to which the chart relates.
The increase in thickness
between each consecutive
step is constant, but varies for
different materials from
0.5mm to several millimetres.

Fig. 1-9. Step wedge

For examinations using a tube voltage of less than 175 kV the thickness of the wedge might
increase by 0.5 or 1.0 mm at each step, while for radiographs using a higher tube voltage
the increase could be in the order of 2-3 mm. In addition several flat plates made from the
same material and of a specified thickness (e.g. 10 mm) should be available.
If the thickness range of a step wedge runs from, say, 0.5 to 10 mm, the step wedge and flat
plate together would give a thickness range of 10.5 20 mm
72

73

Preliminary charts
Before producing an exposure chart it is useful to first draw up preliminary charts, the
so-called density-thickness chart for the voltage range of the specific X-ray set and a
kV- thickness chart.
The two preliminary charts are produced on the basis of the following data:
1.
2.
3.
4.
5.
6.
7.
8.

X-ray set: tube voltage 60-200 kV, tube current 5-10 mA


Filter: none
Source-to-film distance: 80 cm
Material: steel
Intensifying screens: none
Type of film: D7
Density: 2.0
Development: automatic, 8 minutes at 28C in G135 developer

The density-thickness (preliminary) charts as described, provide the data needed to


prepare the final exposure chart. In order to eliminate any inaccuracies, an intermediate
chart (based on the preliminary charts) is prepared for density 2, using the data already
recorded in the first charts.
This is how the thickness-tube voltage chart of figure 3-9 is arrived at. Points relating to
the same series of exposures are then joined in a smooth line producing the intermediate
curves for 8 mA.min and 200 mA.min. In this way deviations in the results of any of the
radiographs can be compensated for.

Exposures
Using a tube current of say 8 mA and an exposure time of 1 minute (i.e. 8 mA.min) radiographs of the step wedge are made at voltages of, for example 75, 90, 105, 120, 135, 150,
165, 180 and 195 kV. Only a narrow strip of the film is used for each exposure. The same
process is repeated at, say 10 mA with an exposure of 20 minutes (i.e. 200 mA.min).
Measuring the density
After development of the radiographs, the density of all steps is measured by a densitometer, see section 9.2.

density=2
Fig. 3-9. Thickness versus tube-voltage preliminary chart

Drawing up the preliminary charts


The densities measured are plotted graphically against the material thickness for which
they were obtained. A smoothly curved line then joins the points relating to one particular
voltage. The result is two preliminary charts (figure 2-9), made at 8 mA.min and
200 mA.min.

Fig. 2-9. Density-thickness preliminary charts

74

75

9.4 The exposure chart


The exposure chart should be drawn on uni-directional logarithmic paper. The material
thickness (in mm) is plotted on the horizontal linear axis and the exposure value (in
mA.min) on the vertical logarithmic axis. For a given kilovoltage (for example 150 kV), we
can, using the previously described intermediate kV-thickness chart, determine that for an
exposure dose of 8 mA.min a density of 2 can be obtained at a thickness of 4.5 mm and for
an exposure dose of 200 mA.min, at a thickness of 15.2 mm.
These thicknesses, and the corresponding exposures, are then plotted on the graph paper
to give points A and B, see figure 4-9. Drawing a straight line linking points A and B, the
150 kV line is obtained. In a similar way the lines for other kV-values can be drawn in the
diagram, eventually resulting in the complete exposure chart of figure 4-9.

be of influence. Therefore, an exposure chart for each individual X-ray set should be drawn
up. This is an excellent way to become familiar with the equipment, while time and money
put into this work will be amply repaid at a later stage.
Exposure charts for gamma-ray examination are drawn up in a similar way as described
above. Figure 5-9 shows one for a Cobalt60 source. A specially designed slide-rule can also
be used, since there is no need to consider individual radiation spectra as for X-ray tubes.
Figure 6-9 shows a similar exposure chart for an Iridium192 source.

Cobalt60
Film type C3
(Agfa D4)
Steel screens
Density 2.5
mm steel

Fig. 5-9. Exposure chart for examination of steel with a Cobalt60 source

density=2
Fig. 4-9. Exposure chart for a 200 kV X-ray set

9.5 Use of the exposure chart


While it may be possible to gradually build up a store of information which can be
consulted in day-to-day work, it is better to make use of good exposure charts.
This system has many advantages to offer, particularly when it comes to choosing the most
suitable working method. Apart from saving time, it gives a guarantee of efficiency and
moreover does away with, or reduces to an acceptable extent, the need for trial exposures
on jobs which are a little outside the normal routine.

Iridium192
Film type C5
(Agfa D7)
Lead screens 0.027 mm
Density 2.5
mm steel

Different X-ray tubes can in practice give quite different results, even though they may be
of the same type. Even a different cable length between the control panel and the tube may

Fig. 6-9. Exposure chart for examination of steel with an Iridium192 source
FFD = focus-to-film distance

76

77

9.6 Relative exposure factors


Relative exposure factors can be used to convert an exposure chart for one type of film
to another film, although still for the same radiation energy.
These factors are not constant for different radiation energies and should, therefore, be
used with some caution. Some examples of relative exposure factors for Agfa Structurix
films are shown in table 1-9.
These are the factors by which to increase or decrease the exposure-time when using the
types of film other than those for which the exposure charts have been prepared. In view
of the widely-varying quality of the radiation emitted by different types of X-ray equipment and the appreciably different characteristics of the various types of X-ray films made
for industrial use, caution should be exercised in applying these relative exposure factors
generally.
Type of film

D2
D3
D4
D5
D7
D8
RCF+F6 (5)
RCF+F8 (5)

100 kV
(1)
9.0
4.1
3.0
1.7
1.0
0.6
0.174
0.03

Relative exposure factors


200 kV
300 kV
Ir192
(2)
(3)
(4)
7.0
8.0
4.3
5.0
2.7
3.0
1.5
1.5
1.0
1.0
1.0
0.6
0.6
0.132
0.389
0.022
0.035

Co60
9.0
5.0
3.0
1.5
1.0
0.6
0.562
0/040

X-ray tube
Gamma source
Energy
100 kV
250 kV
300 kV
450 keV
1,25 MeV
mA
3
10
10
Exp. C/Kg.s
1.8
4.7
FFD. mm
500
700
700
1000
2000
Film type D4 D7 D4 D7 D4 D7 D4 D7
D7
Mat.thickness
Exposure time in seconds
15 mm
50 10
25 mm
100 20
70
15 80
50 mm
1080 210 660 210 6300 1980
1680
100 mm
6300
150 mm
32400
200 mm
400 mm

In the following examples the tube voltage and focus-to-film distance (FFD) are assumed
to be constant, and automatic development is for 8 minutes in G135 developer at 28C.

Table 2-9, derived from reference [2], lists the widely varying absolute exposure times
when different radiation sources are used for radiography on steel of varying thickness.
The relative exposure factors from table 1-9 for both types of film can be recognised in this
table.

Question: What image density


will be obtained behind the section measuring 12 mm under
these given conditions?

With radioactive sources, which give a constant quality (hardness/energy) of radiation, the
relative exposure factors listed can be used quite safely.

78

10
30
4200

9.8 Use of the characteristic (density) curve


with an exposure chart

9.7 Absolute exposure times

Darkroom technique too, plays an important role and a uniform manual or automatic
development process is, therefore, essential.

5000
2000
D7

Table 2-9. Absolute exposure times for steel of varying thicknesses, derived from [ref. 2]

Example 1:
Effect of the thickness of the
object on the density of the radiographic image
It is required to radiograph, on
D7 film, a steel object comprising
two sections of different thickness of 12 mm and 15 mm.
The exposure chart figure 7-9
shows that at 160 kV and an
FFD. of 70 cm, using 10mA.min,
a density of 2 behind the section
measuring 15 mm in thickness
will be obtained.

Table 1-9. Relative exposure factors. For (1) to (5) refer table 1-8.

Linac
8 MeV

Fig. 7-9 Exposure chart for D7


Material = steel; density = 2; ffd = 70 cm;
screens = 0.02 mm lead;
Automatic processing, with developer G135 at 28C,
8 min. cycle.

79

Method and answer

Method and answer

The exposure chart (fig.7-9) shows that under the conditions mentioned above density
D = 2 is obtained on D7-film through the 15 mm thick section, using an exposure of
10 mA.min, point A on the chart.

The characteristic curve (fig. 9-9) shows that at the measured densities of 1.5 and 0.5
respectively, the corresponding logarithm of relative exposures are 2.15 and 1.65.

The characteristic curve (fig. 8-9)


of the D7-film shows that density 2
corresponds to log relative exposure 2.2 (point C in fig. 8-9).
At 12 mm the log. rel. exposure is
2.2 + 0.3 = 2.5.
The corresponding density is then
3.5 (point D in fig. 8-9)

density

Since density 3.0 should not be exceeded, the area which is most important for interpretation, which showed density 1.5 on the first exposure, must now display 3.0
Characteristic curve, figure 9-9, shows that density 3.0 corresponds with log.rel.exp.
2.45 and the difference between the two values amounts to 2.45 - 2.15 = 0.3.
This means that the exposure time must be doubled (10 0.3 = 2), resulting in a radiation
dose of 30 mA.min. This answers the first question.
If the exposure time is doubled, the log.rel.exposure of the lowest density value
originally measured will increase by 0.3, i.e. 1.65 + 0.3 = 1.95. The corresponding
density will be 1.0 (fig. 9-9).

film
base

log.rel.exp.

The average gradient between the upper and lower densities on the original radiograph
was (1.5 - 0.5) / (2.15 - 1.65) = 2.0.
The average gradient on the new radiograph is (3.0 - 1.0) / (2.45 - 1.95) = 4.0, so the
average contrast has doubled.

Fig. 8-9. Characteristic (density) curve of the D7-film

density

Under the same conditions the


12 mm section would require an
exposure of 5 mA.min (point B in
the chart), which means an exposure ratio of 10/5.
The exposure through the 12 mm
section is two times greater than
through the 15 mm section.
The logarithm of this ratio equals:
0.3 (log 2 = 0.3).

Example 2:
Effect of exposure on contrast
Assume that when an exposure
of 15 mA.min is used for a
radiograph on D7-film, both
average density and contrast
prove to be too low after processing. The highest and
lowest density in the most
relevant section of the image
film
are only 1.5 and 0.5.
base
The intention was to make a
log.rel.exp.
radiograph with a maximum
Fig.
9-9.
Characteristic
density
curve
of
the
D7-film
density of 3.0.
Questions: What exposure
time would be required for the same radiation intensity and what contrast
increase would be achieved?

80

81

and
10 Processing
storage of X-ray films
revolving door

labyrinths

Film developing is the process by which a latent image, see section 7.2, is converted into
a visible image. The crystals in the emulsion - carriers of the silver traces forming the
latent image - are transformed into metallic silver by selective reduction as a result of
which the visible image is created. The development procedure must be carried out
carefully to achieve this and guarantee successful archiving over a longer period.
Manual developing is a laborious process that must be carried out meticulously in order
to get the high quality results.
For increased efficiency and uniform quality, X-ray films are more commonly processed
processed automatically. The manual process is, however, still frequently applied.
It will therefore be useful to describe manual processing in this chapter and so become
familiar with the developing process.

10.1 The darkroom


Fig. 1-10. Entrances to a darkroom

Entrance and colour


For practical reasons the darkroom needs to be as close as possible to the place where
the exposures are made, although naturally out of reach of radiation.
The darkroom needs to be completely lightproof, so the entrance must be a light-trap
usually in the form of two doors, (one after the other), a revolving door or a labyrinth,
see figure 1-10.
In practice the labyrinth is found to be the best arrangement, although it does take up
a comparatively large space. The walls of the passage are painted matt black, and a
white stripe about 10 cm wide running along its walls at eye-level is enough as a guide.
Inside the darkroom itself, the walls should preferably be painted in a light colour; light
walls reflect the little light there is and are easier to keep clean.

Darkroom lighting
X-ray films are best-processed in normal orange-red (R1) or green (D7) darkroom
lights. The distance between film and darkroom lighting needs to be considered, depending on the sensitivity of the film and the duration of the development process.
The light safety of the darkroom lighting can be tested by covering half of a pre-exposed film (density 2) lengthways, leave it for 5 minutes and then process it as usual.
The difference in density may not exceed 0.1.

82

83

Another method is to place an unexposed film on the workbench and cover part of it up
with a sheet of cardboard, which is then gradually removed so as to produce a series of
different exposures. By developing the film in the usual way, it will then be possible to
see how safe the light is, and how long a film can be exposed to it before it exceeds the
maximum acceptable difference in density of 0.1.

Darkroom layout
The darkroom should preferably be divided into a dry side and a wet side. The dry side
will be used for loading and emptying cassettes, fitting films into developing frames and
so on - in short, for all the work that does not allow dampness.

Developer
Development fog, graininess and contrast are dependent on the type of developer,
which is preferably made up to suit the film used.
If a concentrated manual developer is used, for example G128 made by Agfa, and the
developer tank has a capacity of, say, 25 litres, then all to do is pour 5 litres of the concentrated developer into the tank and add 20 litres of water (ratio 1 part of concentrate to 4 parts of water). G128 developer is also used as a replenisher, in which case 3
parts of water are added to 1 part of concentrate.

Fixer
On the wet side, the films will be processed in the various tanks of chemical solution.
For efficient working, and to ensure uniform quality, there should be automatic control
of the temperature of the solutions.

Developing times and bath temperatures

Tanks
In processing tanks used in the manual process, films are held vertically in their frames.
These tanks can be made of stainless steel or plastic. The dimensions of the tank must
be suited to the size and number of films to be processed. There must be a space of at
least 1.5 cm between films. The top edge of the films must be approx. 2 cm below the
surface of the solution.
The wet side of the darkroom will have five tanks, arranged in the following sequence:
1.
2.
3.
4.
5.

Fixer too is supplied as a concentrated liquid (G328). The same instructions as for preparing developer apply here.

developer tank
stopbath or rinse tank
fixer tank
final wash tank
tank for wetting solution

10.2 Chemicals and film-development


Making-up processing solutions
Nowadays, chemicals are supplied as a liquid concentrate, suitable for the particular
type of film used.
The processing solutions can be prepared either directly in the tanks or in plastic buckets. In the latter case each type of solution must be prepared in a separate bucket,
which is never used for other chemicals.

84

The film is clipped on or slipped into a frame, depending on the type of frame, and hung
in the developer tank. As soon as the film is submerged in the developer, the darkroom
timer is set for the required number of minutes. The optimal developing time is the time
at which the most favourable contrast to fog ratio is achieved. Minor deviations from
the correct exposure time may be compensated by adjusting the developing time.
The recommended developing time for Agfa films in G128- manual developer is
5 minutes at 20C. In the automatic process using G135 developer, the developing time
is 100 seconds at 28C. Deviating from the recommended developing times and
temperatures will almost always lead to reduced image quality (e.g. increased coarsegraininess).
Raising the tank temperature will speed up the development process as table 1-10
shows, but the developer will oxidise more rapidly. Should it not be possible to achieve
a bath temperature of 20C, the following developing times can be used at the
temperatures as indicated in table 1-10. This applies to all D-type films.
Temp. C
Time/mins

18
6

20
5

22
4

24
3.5

26
3

28
2.5

30
2

Table 1-10. Developing time versus developer temperature.

The temperature of the developer shall never be less than 10C, but is preferably
higher than 18C to obtain optimal image contrast. It is best to always maintain the
same developing conditions, so that the exposure technique can be matched to these
and uniform results obtained.

85

Film agitation

Final wash

To prevent air bubbles from forming on the surface of the emulsion (which will cause
spots on the finished radiographs), and to make sure that the developer penetrates all
areas of the emulsion evenly, the films should be kept moving during their first 30
seconds in the developer. After that, it is recommended to move the film from time to
time to prevent film faults such as lines or streaks.

The final wash is intended to remove to remove the residual fixer and the soluble silver
compounds left behind in the emulsion, which if not flushed out, would reduce film
shelf life. Washing should preferably be done with running water, ensuring that all
parts of the film are reached by fresh water. The duration of the final wash depends on
the temperature of the water. See table 2-10. Temperatures over 25C must be avoided.

Replenishing
Up to 400 ml of liquid per square meter of film processed may be carried over to the next
tank. When developing frames used it is, therefore, preferable to hold the film
2-3 seconds over the developer tank to drip.
After each square meter of film developed, 600 ml of replenisher must be added to the
bath regardless of the quantity of developer lost from the tank. Up to about 4 litres of
replenisher can be added in this way, for every litre of the original developer in the tank.
The solution must be discarded and replaced with fresh developer when the total
quantity of replenisher added is three times the original total contents, but in any case
after eight weeks, irrespective of the number of X-ray films processed.

Stopbath
Before transferring the developed film to the fixer tank, it is placed in a stopbath
(consisting of 30 ml glacial acetic acid to 1 litre of water) for 30 seconds to prevent the
fixer solution from being neutralized too rapidly by the developer, and stripes or
dichroitic fog from forming on the film.
If a film is not passed through a stopbath, it must be rinsed in running water for a few
minutes immediately after leaving the developer.

Fixing
Fixing renders the image formed during development permanent, by removing undeveloped silver halide salts from the emulsion. When the film is taken from the stopbath
it still has a milky appearance; this changes in the fixer and the light areas of the film
become transparent.
As a rule the film is left in the fixer twice as long as it takes to clear or become transparent. Fixing time (in a fresh solution approx. 3 minutes) is twice the clearing time
(1.5 min.). As soon as it takes double that time to clear a film in G328 fixer solution
at 20C, it must be replaced.
For every liter of solution in the fixing tank, a square meter of film can be treated.
Films have to be kept moving during the first 30 seconds in the fixing bath.
86

Temperature range (C)


5-12
13-25
26-30
> 30

Washing time (minutes)


30
20
15
10

Table 2-10. Relationship between water temperature and washing time

Drying in the drying cabinet


When the film is taken out of the water, the water on the film, as a result of its surface
tension, runs together to form droplets of varying size. The film will, therefore, dry unevenly, causing drying marks. For this reason it is advisable to immerse the films in a
solution of 5-10 ml wetting agent to each litre of water. Wetting agent reduces the surface tension of the water so that, after the film has drained, the surface will be evenly
wetted and will dry evenly with no risk of marks. Films should be hung to drain for
about 2 minutes before they are placed in the drying cabinet.
Drying should preferably be done in a drying cabinet, or alternatively in a dry and dust
free room. No drops of water must be allowed to fall on films that are already drying, as
this will cause marks. Wet films should, therefore, always be hung below already
drying films.
Drying time will depend on temperature, air circulation and relative humidity of the air
in the cabinet. Films will dry more quickly when they have first been put into a wetting
agent.
Before a film is taken out of the drying cabinet, it must be checked that the corners and
edges of the film are thoroughly dry. Air temperatures above 40C should be avoided as
this may cause ugly drying marks. There must be free circulation of air between the
films in a cabinet; if they cannot dry evenly on both sides, they may curl or distort.

Roller dryers
Industrial dryers can be used to dry films quickly and uniformly after washing.
This mechanised drying process only takes minutes. Dryers and chemicals should
preferably be matched. There are compact roller dryers on the market which are
capable of developing approx. 15 cm of film per minute and take up far less space than
a drying cabinet.
87

10.3 Recommendations for the darkroom

10.5 Automatic film processing

Cleaning of tanks

NDT-U (universal) film processor

Whenever the processing solution is renewed the tank must be cleaned, preferably with
hot water and soap. If this proves inadequate, polyester tanks can be cleaned using a
bleach solution (100-200 ml/litre of water), hydrochloric acid (10 ml/litre of water) or
acetic acid (50 ml/litre of water). Stainless steel tanks may be cleaned with a solution
of nitric acid (10 ml/litre of water) or acetic acid (50 ml/litre of water). Hydrochloric
acid must never be used for stainless steel tanks.
There are industrial cleaning agents on the market (for example Devclean and the, environment-friendly, Fixclean), specially developed for cleaning of darkrooms.

Over the last few years there has been a vast increase in the use of automatic processors for
handling industrial X-ray films. Not only is it a faster and more efficient process, the
uniform process also leads to improved image quality. The total processing time may be
between 1.5 and 12 minutes (nominally 8 minutes), significantly shorter than in manual
processing. Of these 8 minutes, the film will be in the developer solution for only 100
seconds, the so-called immersion time. These shorter processing times have been made
possible by the use of special chemicals (G135 and G335), and by a higher temperature of
the solutions: 28C instead of 20C.

Stained fingers

The shortest processing time of 1.5 minute is essential for the development of the special
films used on board lay-barges, where the results must be available quickly.

Brown stains on the fingers can be avoided by rinsing the hands in water whenever they
come into contact with developer. If fingers do become stained, they should be immersed
in a solution of:
a. 1 litre of water
b. 2 gr of Potassium-permanganate
c. 10 ml of concentrated sulphuric acid
Next the hands should be rinsed in an acid fixer solution, and finally washed with soap and
water.

Chalky water
If hard, chalky water is used for mixing the solutions, troublesome processing faults
may occur. Calcium salts may, in the presence of carbonates and sulphites, result in a
whitish deposit on the films which is insoluble in water. To prevent this, the diluant can
be softened by using a special filter, or by boiling it first and letting it cool down before
making up the chemical solutions.
To remove chalk deposit from films, they may be soaked in a solution of 7 ml glacial
acetic acid to a litre of water.

10.4 Silver recovery


The silver halides in the emulsion which were not reduced during development, are
dissolved in the fixer. Silver can be recovered from the fixer in order to keep the silver
content of the fixer solution as low as possible so that the fixer lasts two to four times
longer, and sell the silver.
Silver recovery can, for example, be done by electrolysis. In addition to electrolysis
equipment, there are other silver recovery systems commercially available.
It is worthwhile considering subcontracting this work to a specialised firm in view of
secondary aspects such as organisation, logistics, storage and environmental
requirements.

88

The chemicals used are more active at higher temperatures. The higher temperature of the
solutions makes the emulsion layers swell, resulting in a faster diffusion of the liquid
through the layers and, consequently, more rapid action of the chemicals.
Swollen emulsion coatings do, however, have the disadvantage of being softer and hence
more vulnerable to damage; a compromise between the advantages and drawbacks is reached by adding a carefully determined proportion of hardening ingredients to the fixer.
Chemicals for use in automatic processors also have additives to inhibit oxidation of the
solutions and formation of fog in the emulsions.
Automatic film processing not only makes the results available sooner, it also standardises
(improved reproducibility/uniformity) the development process and, consequently, the
exposure technique. This increases the quality and reliability of radiography as a method
of non-destructive testing.
GE Inspection Technologies supplies integrated Agfa-systems in which X-ray films, chemicals and processing equipment are all adapted to each other. Through the uniform characteristics of its films, carefully formulated chemicals, continuous agitation, automatic replenishment and accurate temperature control of the solutions in the processors, Agfa systems
ensure top-quality results.
The Agfa NDT-U processor is equipped with an infrared film drier while its functions are
controlled by a microprocessor. Its throughput depends on the required cycle time (adjustable between 1.5 and 12 minutes) and film size. All normal film sizes, including roll film,
can be processed. When set for an 8-minute cycle (100 seconds immersion time) for example, approximately 100 films of size 10 x 48 cm can be processed per hour.

89

NDT-E (economy) film processor


In order to limit any detrimental effects on the environment, Agfa has developed the Eco
(Ecology and economy) designated processors. Here, too, equipment and chemicals are
carefully matched, thus complying with strict ecological requirements such as a maximum
of 50 mg silver per square metre of processed film, for the disposal of rinse water.
This figure for silver content is at least fifteen times lower than for conventional developing
systems. This is achieved through the considerably improved (cascade) fixing process,
which additionally results in a bigger quantity of recovered silver.
Furthermore, measures have been taken to save on energy, chemical and water usage, thereby making the eco range of film processors as environmentally friendly as possible.
Figure1-10 shows the schematic lay-out of this high-tech processor.
The S eco-version has a 50 % higher production capacity than the U-version.
A very useful option is its suitability for use in daylight, in combination with a matching
film feeding system.

10.6 Checking the development process


and film archiving properties
Besides exposure technique, many aspects influence the quality of the final radiograph.
An important factor is the development system. Monitoring and quantifying the proper
functioning of a development system is an essential part of quality control, as a properly
exposed radiograph can be spoilt if the processes that follow are performed incorrectly.
For the monitoring of the development process and archival properties of X-ray-films,
Agfa has produced two methods: the so-called PMC-strips, and the Thio-Test.
Both methods are based on the international standard ISO 11699 part 2, and the
European standard EN 584 part 2, which describe a standard development process and
the means to control its execution.

PMC-strips to check the developing process


To facilitate ongoing quality control, and ensure compliance with existing standards on
systems classification, certified PMC-strips are used to monitor the development process.
PMC is short for Processing Monitoring Control.
The purpose is to:
demonstrate conformance with the standard film system as described in the
standards ISO 11699 or EN 584
demonstrate the consistency of the development system
monitor and promote uniformity of the various development systems in
different locations
initiate timely corrective action if deviations occur
PMC-strips are film strips that have been pre-exposed in a regular step-pattern by the
supplier, under special conditions and within narrow tolerances, but have not as yet been
developed. They are supplied with a certificate of compliance with EN 584-2 and
ISO 11699-2.

Fig. 1-10. Schematic layout of the Agfa NDT S eco processor


1
2
3
4
5a/b
6

Film feeder
Film surface scanner
Developer tank
Rinse tank
Fixer tanks
Final wash tank

7
8
9
10
11
12

In the development system to be checked, a PMC-strip is processed routinely in a way


identical to a normal radiograph. Finally, the various densities are measured with
a densitometer.

Infrared dryer
Film exit
Film receiving tray
Fixer pump
Developer pump
Overheating protection

90

91

The unexposed area on the PMC-strip shown in figure 2-10, apart from providing a
reference for fog and base density, also allows for the Thio-Test to be carried out.
The components used for this test are:

Unexposed area for the Thio-Test


fog + base density

Fig. 2-10. PMC-strip with an unexposed area


for the Thio-Test

Reference step
for film sensitivity

Reference step
for film contrast

A PMC-strip as shown in figure 2-10 has to be used whenever the chemicals in an automatic or manual processing system are replenished or changed. It is also advisable to use
a PMC-strip regularly, but at least once a month, for a routine check of the development
system.
A calibrated densitometer measures the following steps:
D0: fog and base density ( 0.3)
D3: density of step 3
D7: density of step 7
The reference values according to the corresponding certificate are Sr and Cr.
The following calculations are then made:
Sensitivity index Sx = D3 D0
Contrast index
Cx = (D7 D3).Sr /Sx
The system is acceptable if the following criteria are met:
a. D0 =  0.3
b. Sx has a value  10 % of Sr
c. Cx has a value between Cr +15 % and Cr 10 %

1
2

The Thio-Test colour step-wedge


A dropper-bottle of Thio-Test reagent
The reagent consists of a 1 % silver nitrate solution in demineralised water

The working method, which only takes a few minutes, is as follows:


1 apply the test liquid (reagent) to an unexposed part of undeveloped, dry film
2 allow to soak for 2 minutes,  15 seconds
3 remove excess liquid with absorbent paper
4 leave to dry for 1 minute before treating the reverse side
5 repeat the above procedure on the other side of the film, in exactly the same spot.
Evaluate as follows, within 30 minutes:
6 the test zone of the film is put against a white background
7 the Thio-Test colour strip is put on the film, as close to the spot as possible
8 the colour step of the wedge that resembles the colour of the test zone closest, is
regarded definitive for life expectancy.
Colour wedge
(from dark to light)
Darkest

Thiosulphate
(*)g/m2
Min. 0.35

Dark
Light
Lightest

Max. 0.20
Max. 0.10
max. 0.04

Archival quality
L.E. (Life expectancy)
Film needs
repeat treatment
L.E. 10 years
L.E. 100 years
L.E. 500 years

Table 3-10. Colour steps for the Thio-Test . (*)These values apply to films with double-sided emulsion layers.

If one or more of these criteria are not met, the development process must be adjusted.

Thiosulphate-test to check film archiving properties


The archival properties of a radiographic image must also be determined in accordance
with the standards ISO and EN by analysing the quantity of residual thiosulphate in the
films emulsion layers. This quantity depends on the thoroughness with which the fixing
and rinsing processes have been carried out.
For storage over a period of 100 years, 100 g/m2 is allowed; for a period of 10 years double this figure is allowed, see table 3-10. These values are difficult to measure however.
The so-called Structurix Thio-Test, developed by Agfa, is a very useful and quick method
to quantify film-keeping properties in practice.
92

A regular Thio-Test provides early detection of deficiencies in the development process,


for example exhausted fixer solutions, irregular water supply or insufficient rinsing, and
so prevents poorly processed films being archived.

10.7 Storage of exposed films


The way in which radiographs are handled and stored plays a very important role in their
keeping properties. Films that must be kept for longer periods of time require the same
ambient conditions as new unexposed films, i.e.:
ambient temperatures below 24C
relative humidity of less than 60 %
preferably stacked on edge
93

discernibility
11 Defect
and image quality
anode

Three factors govern the discernibility of defects in a radiograph:

source
1. Geometrical effects:
Size of the source
Source-to-object distance
Defect-to-film distance

2. Film properties (governing image quality):

(ffd)

Graininess
Contrast
Fog
Inherent unsharpness

3. Quality of radiation applied.

11.1 Unsharpness
Geometric unsharpness
X-ray tubes and radioactive sources always produce radiographs with a certain amount
of blurring the geometric unsharpness, Ug in fig. 1-11, because of the finite dimensions of the focal spot or source size.
The magnitude of this unsharpness, Ug, is given in the following equation:

Ug = s.a
F-a
In which:

film

s is the effective focus (or source) size


F is the focus-to-film (or source-to-film) distance
a is the defect-to-film distance

density across film


The maximum value of Ug related to a defect situated at a maximum distance from the
film (and for which a = t) can be calculated from the formula:

Ug (max) = s.t
F-t

Fig. 1-11. Geometric unsharpness.


The source diameter, S, is shown very
large for clarity.
In which:

94

t = the thickness of the object

95

Consequently, Ug can be reduced to any required value by increasing the source-to-film


distance. However, in view of the inverse square law this distance cannot be increased
without limitation, as extremely long exposure-times would result. The formula also
indicates that geometric unsharpness assumes more and more importance as the distance between defect and film increases.

In this situation the unsharp images of each of the two edges of the defect may overlap,
as shown in example C. The result is that image C not only becomes unsharp, but also
suffers a reduction in contrast compared to image A, made with a point source and
image B made with a relatively small source.

Inherent unsharpness
Not only the silver halide crystals directly
exposed to X-radiation are formed into
grains of silver, but also (albeit to a lesser
degree) the surrounding volume of emulsion. This cross-sectional area represents
the inherent unsharpness or film unsharpness Uf .

Figure 2-11 shows the effect of geometric unsharpness on the image of a defect smaller
than the focus size.

So, even in the absence of geometric unsharpness, if the radiation energy is high
enough, film unsharpness can occur: the so
called inherent unsharpness. If a steel test
plate with a sharp thickness transition is
radiographed with high energy X-rays,
there will be a gradual transition of film
density across the image of the step from
A to B.

test plate

film

film density

A special case arises, however, when one uses a micro focus X-ray tube with a focal spot
size in the range 10-50 m. With such a small focus size, the image can be deliberately
magnified (see section 17.1) by using a short source-to-specimen distance, and a large
specimen-to-film distance, and still retain an acceptably small value of Ug. The advantage of this technique, called the projective magnification method, is that the graininess always present in a photographic image is less of a disturbing factor in the discernibility of very small defects.

Without inherent unsharpness, the film


would show an absolutely sharp transition
between the two densities, as shown in figure 3a-11. In practice, the density change
across the image is as shown in figures 3b,
3c and 3d-11.

film

film

film

The width of this transitional area (Uf ),


expressed in mm, is a measure of film
unsharpness.
For clarity, the density curves are magnified along the X-axis.
(a) density distribution across image of sharp edge,
assuming Uf = 0
(b) (c) and (d) density distribution due to film unsharpness
(b) theoretical; (c) with grain; (d) smoothed.

A. Point focus size - s -: no geometric unsharpness - defect image sharp


B. Small focus size s -: geometric unsharpness Ug defect image blurred
C. Increased focus size s -: still larger Ug - defect image blurred and loss of contrast - Co is less than in A and B
Co = contrast

Fig. 3-11. Inherent (film) unsharpness for X and


Gamma-radiation.

Fig. 2-11. Geometric unsharpness: effect on the image of a small defect.

96

97

Table 1-11 and figure 4-11 show experimentally determined values of inherent unsharpness for film exposed at various radiation energy levels.
These values are based on the use of filters and thin lead intensifying screens; thicker
screens produce slightly higher values. If no lead screens are used, Uf is 1.5 to 2 times
smaller. Uf is influenced mainly by radiation intensity and the type of intensifying
screens used; the type of film is hardly of any consequence.
The distance between film and intensifying screen is of great importance for the value of
Uf . Good contact between film and intensifying screen is imperative and can be achieved by vacuum-packing of film and screens together.
Radiation energy Uf in mm
50 kV
0.05
100 kV
0.10
200 kV
0.15
400 kV
0.20
2 MeV
0.32
8 MeV
0.60
31 MeV
1.00
Se75 (320 keV)
0.18
Ir192 (450 keV)
0.25
Co60 (1.25 MeV)
0.35

Preceding paragraphs of this chapter described the effects of geometric unsharpness and
the possibility to influence this by adjusting the source-to-film distance.
This section will expand on this.
To obtain a radiograph which is as sharp as possible, so as to show maximum detail, the
total unsharpness should be kept to a minimum. The radiation energy level selected for
making the radiograph, see chapter 9, can serve as a lead. It is, after all, determined by
the thickness of the material to be radiographed, but is at the same time also responsible for film unsharpness Uf , which can be extracted from table 1-11 and or figure 4-11.
It is no use to try and keep geometric unsharpness Ug far below the value of Uf , as in that
case Uf determines the total unsharpness anyhow.
If the aim is to make geometric unsharpness Ug equal to the value of Uf, the source-tofilm distance (F) required can be calculated from the following formula :

t(Ut+1,4s)
F=
Ut

= source-to-film-distance

Ut

= total unsharpness

= thickness of the object

= effective source size

Total unsharpness
Total film unsharpness Ut is determined by the combination of Ug and Uf . The two values
cannot be just added up to arrive at a figure for Ut. In practice, the following formula produces the best approximation for film unsharpness Ut:

U t=

U g+Uf
2

Broadly, if one value of unsharpness (Ug or Uf ) is more than twice the value of the other,
the total unsharpness is equal to the largest single value; if both values of unsharpness
are equal, total unsharpness is about 2 = 1.4 times the single value.
If necessary, Ug can be reduced by increasing the focus-to-film distance. This can only be
done to a limited extent because, due to the inverse square law, exposure times would
become extremely long. As a compromise an optimum focus-to-film distance F is chosen
whereby Ug = Uf .

All measurements in mm.

Instead of calculating F,
various code-based procedures and
guidelines provide graphs from
which minimum distance (Fmin)
can be determined. Figure 5-11
shows a nomogram on the basis of
EN 1435, from which the minimum
focus distance for two quality levels
(category A and B) can be
extracted.

Fig.5-11. Nomogram for minimum source-to-film distance fmin


according to EN 1435-criteria.
Catagory A - less critical applications (general techniques)
Catagory B - techniques with high requirements
of detail discernability
The above graph appears enlarged in the appendix on page 190.

98

99

distance (b)

From the above information it can be deduced that Uf increases at higher radiation energies.

In which:

distance fmin for catagory A

Fig. 4-11. Graphical representation of table 1-11.


Values of Uf for X - and Gamma radiation at increasing radiation energies

f min = minimum distance


source to source-side object (mm)
s = source size (mm)
b = distance source-side object to film (mm)

distance fmin for catagory B

Table 1-11. Empirical values of film unsharpness Uf at various radiation energies


using lead intensifying screen

11.2 Selection of source-to-film distance

11.3 Other considerations with regard


to the source-to-film distance
Inverse square law
As explained in the previous section, the effect
of Ug can be reduced by increasing the focus-tofilm distance F.
One of the properties of electromagnetic radiation is that its intensity is inversely proportional
to the square of the distance, better known as
the inverse square law. Both X - and Gamma
radiation follows that law.
The intensity of radiation per unit area of film is
inversely proportional to the square of the source-to-film distance (s-f-d).
As figure 6-11 shows, at a distance 2F from the
source, a beam of rays will cover an area (b)
Fig. 6-11. Inverse square law for distances
four times greater than area (a) at distance F.
Consequently, the intensity per unit of surface area for (b) will be only 1/4 of the value
for area (a). This means that, all other parameters being equal, at distance 2F exposure
time must be multiplied by four to obtain the same film density.
This principle obviously has its (economical and practical) limitations, beyond which a
further increase in s-f-d is just not feasible.

Selection of radiation energy (kV)


Once the appropriate source-to-film distance is chosen, the correct kilo voltage can be
determined from an exposure chart (see chapter 9).
The importance of choosing the exact kilo voltage varies considerably with the kilo voltage range being considered. For X-rays below 150 kV the choice is reasonably critical
and gets more critical still at lower kilo voltages.
The kilo voltage to be applied is specified in (EN) standards, see chapter 20.
Table 2-11 gives useful empirical rule-of-thumb values for radiographs of aluminium,
steel or plastic objects.
Material
Steel
Aluminium
Plastics

kV-value
100 kV + 8 kV/mm
50 kV + 2 kV/mm
20 kV + 0.2 kV/mm

Examples :
15 mm steel:
12 mm aluminium:
10 mm plastics:

100 + 15 x 8 = 220 kV
50 + 12 x 2 = 74 kV
20 + 10 x 0.2 = 22 kV

In the range 200-400 kV, only a significant change in voltage, say 30-40 kV, will cause a
noticeable difference in defect discernibility.

Selection of gamma source


As it is not possible to vary the radiation energy emitted by a gamma-ray source, it is
necessary to indicate a range of thickness which may be satisfactorily examined with
each type of radio-isotope.
The upper limit is decided by the source strengths commercially available and the maximum tolerable exposure time: the lower limit is determined by the decrease in contrast
and the related reduced image quality.
The lower limit, therefore, depends on the required degree of defect discernibility.
When this is insufficient in comparison to what is achievable by the use of X-ray equipment, another type of isotope providing a reduced energy radiation could be selected.
Table 3-11 shows the thickness range usually recommended for various gamma sources.
The table applies to steel. If, for reasons of convenience, gamma rays are used on thin
specimens which could also be X-rayed, it should be understood that the resulting radiographs will be of poorer quality compared to X-radiographs.

Source type
Co60
Ir192
Se70
Yb169
Tm170

Standard sensitivity
technique in mm
30 - 200
10 - 80
5 - 40
1 - 15
1- 10

Table 3-11.Thickness ranges in mm for examining steel with the usual types of gamma sources.

Note: Standard sensitivity permits a slightly poorer image quality than high sensitivity.
Thus a larger thickness range can be inspected coping with the quality requirements.

Table 2-11. Rule-of-thumb values for the selection of kilo voltage

100

High sensitivity
technique in mm
60 - 150
20 - 70
10 - 30
3- 10
4-8

101

11.4 Radiation hardness and film contrast

11.5 Summary of factors that influence image quality

When radiation hardness increases, the half-value thickness (HVT) also increases.
Tables 2-2 and 3-2 for steel and lead respectively show this in figures.

The factors that influence image quality are:

This is why in an object with different thicknesses, image contrast diminishes when radiation hardness increases. Figure 7-11 clearly illustrates this.
The left side of a step-wedge is radiographed with 150 kV, the right side with 80 kV.
The right side shows the greater contrast between two steps, whereas on the left the
contrast range is the biggest.

150 kV

80 kV

1. Contrast
2. Unsharpness
3. Graininess

1 Contrast depends on:

Radiation energy (hardness)


Variation in thickness
Backscatter
Front- and back screen
Film-screen combination
Film-screen contact
Defect location, depth as well as orientation

2 Unsharpness depends on:

Size of focus
Thickness of the object
Source-to-film distance
Radiation energy (hardness)
Film-screen combination
Film-screen contact

3 Graininess depends on:

Type of film
Type of screen
Developing procedure
Radiation energy (hardness)

Fig. 7-11. X-rays of a step-wedge with 150 kV (left) and 80 kV (right).

102

103

distortion
12 Image
and useful film length
12.1 Image distortion
source

source

On a radiograph, a three-dimensional object is presented in a two-dimensional plane


(the film). The appearance of both the object and its defects depends on the orientation
of radiation relative to the object.
As shown in figure 1-12, the image of a gas cavity in a casting may be circular or elongated depending on beam orientation.
In general, the beam of radiation should be at right angles to the film and a specimen
should whenever possible be laid flat on the film cassette. Special angle shots are, however, sometimes useful.
Figure 2-12 (A) shows a situation whereby detection of lack-of-side wall fusion in a
V-weld is not performed optimally. Angled radiation (B) is more likely to show up this
type of weld defect.

source

circular image
Fig. 1-12.

source

elongated image

Distortion of the image of a gas cavity due to beam orientation

lack of side
wall fusion
blurred image of the defect

lack of side
wall fusion

sharp image

Fig. 2-12. Lack of sidewall fusion in a V-shaped weld joint.


Shot A is unlikely to detect the defect; shot B will.

104

105

The number of radiographs necessary for 100 % examination of a circumferential weld


can, through calculation, also be obtained from the codes. When large numbers of similar welds are involved, this is an important figure, because too many radiographs would
be uneconomical and too few would lead to insufficient quality of the examination.
The minimum number of radiographs required for various pipe diameters and wall
thicknesses at varying source positions can be derived from the graph in figure 4-12.
The graph is applicable to single wall and double wall technique, whereby the maximum
increase in thickness to be penetrated is 20 %, in accordance with EN 1435 A.

12.2 Useful film length


When radiographing curved objects, for example a
circumferential weld in a
pipe, as figure 3-12 shows,
the resulting image will be
distorted. Variations in density will also occur. As a
result of the curvature of
the pipe with a wall thickness t, the material thickness to be penetrated increases to T, so film density is
lower at the ends of the film
than in the middle.

source

crack
T

T>t

Moreover, if defects are projected nearer the ends of a


film, distortion of the defect
image will become greater.
The film length suitable for
defect interpretation is therefore limited. This so-called useful film length is,
depending on the nature of
the work, defined in codes
e.g. in EN 1435.

film

distorted
image of
crack

Example 1:
An X-ray tube with an outside diameter of 300 mm is used to examine a circumferential
weld in a pipe of a diameter De of 200 mm and a wall thickness t of 10 mm.
The distance between the focal spot and the outside of the
X-ray tube is 300/2 = 150 mm.
F = half the X-ray tube diameter
+ De = 150 + 200 = 350 mm.
t/De = 10/200 = 0.05 and
De/F = 200/350 = 0.57
The intersection of the two
co-ordinates (0.05 and 0.57) is
in the range where n = 5, so the
number of radiographs must be
at least 5.

Fig. 3-12. Image distortion caused by the curved shape of the object

It is not always practicable to apply the single-wall technique as shown in figure 3-12.
In order to still achieve 100 % examination, the double-wall / single-image technique
(DW-SI) is applied. (In NDT jargon the abbreviations DW-SI and DW-DI are frequently
used for Double WallSingle Image and Double Wall-Double Image respectively.)
In that case several radiographs are made, spaced equally around the circumference of
the item under examination. The number of radiographs to be made depends on the standard or code to be complied with.

Operating range
not being used.

Example 2:
When using a source placed
against the pipe wall,
t/De = 10/200 = 0.05 and
De/F = 200/(200+10) =
200/210 = 0.95.
The intersection of the two coordinates now lies in the area Fig. 4-12. Graph for the minimum number of exposures in accordance
where n = 4. So, by using
with EN 1435 A at maximum thickness increase of 20 %.
a radioactive source which is
This graph appears enlarged in the appendix on page 191.
located closer to the pipe surface,
one less exposure would still ensure compliance with EN 1435A. Initially, the code
would however have to allow the use of an isotope instead of an X-ray tube.

In codes, useful film length is determined by the percentage of extra wall thickness
which may be penetrated in relation to the nominal wall thickness (t) of the pipe.
Percentages of 10, 20 and 30 are commonly applied. For general use, 20 % is a practical
value whereby the lightest section of the film shall have a density of at least 2.
106

107

13 Image quality
13.1 Factors influencing image quality
With regard to image quality, the term frequently used is sensitivity.
Sensitivity determines the extent to which a radiograph is able to clearly show (anomaly) details of a certain size. Sensitivity in this sense must not be confused with the
sensitivity or speed of the film. (see section 7.5).
Discernibility of defects on a radiograph depends in general on:
the quality of the radiation
the properties of the film
the film viewing conditions
Image quality is governed by contrast, sharpness and film graininess.
Image contrast is affected by :

Digitised and enhanced image of a radiograph of the (American) Liberty Bell.


The image was used to monitor possible lengthening of the crack.

differences in thickness of the specimen


the radio-opacity (radiation transparency) of the specimen and its defects
the shape and (depth)location of the defects
the quality (hardness) of the radiation
the amount and effects of scattered radiation
the effect of filters used

Film contrast depends on:


the type of film
the density level
Sharpness of an image is governed by:

108

the (effective) size of the focal spot or radiation source


the source-to-object distance
the object-to-film distance
the contact between film and intensifying screens
the type of intensifying screens used
the radiation energy used
109

The last factor, graininess, depends on :

the thickness of the emulsion layer


the concentration of silver crystals in the emulsion (silver/gelatine ratio)
the size of the silver crystals
the radiation energy used
the developing process employed

The radiation energy level is the only factor that can be influenced by the radiographer;
the other factors are determined by the film making process.

13.2 Image quality indicators (IQIs)


In the past it was thought possible to assess the smallest defect detectable, by fixing a
simple type of indicator on the test object during exposure.
This would supposedly guarantee that defects of a certain minimum size, expressed as a
percentage of the material thickness, could be detected. In practice, however, this proved not to be achievable.
In particular where small cracks and other two-dimensional defects are concerned, it
can never be guaranteed that they are not in fact present when no indication of them can
be found in the X-ray image.
However, it is reasonable to expect that at least the quality of the radiographs, and of
course the rest of the entire process the film undergoes, meets certain requirements.
The probability is high that defects will be more easily detected when the image quality
is high. The exposure technique and required image quality, described in the code,
depend on the purpose for which the object involved will be used.
In order to be able to assess and quantify the image quality of a radiograph, it needs to
be converted into a numerical value, and to do this image quality indicators (IQI) are
used, known in the USA as penetrameters.
Image quality indicators typically consist of a series of wires of increasing diameters, or
a series of small plates of different thicknesses, with holes drilled in them of increasing
diameters.
Although codes describe their techniques differently, they agree on the following points:
An image quality indicator shall be placed at the source-side of the object being
examined,
If it is not possible to place the indicator on the source-side, it may be located on the
film-side. This exceptional situation must be indicated by a lead letter F on
or directly adjacent to the indicator,
The material of the indicator must be identical to the material being examined.
110

The image quality of a radiograph is, for example, defined as the number of the thinnest
wire still visible, and is generally said to have image quality number -X-.
The image quality can also be expressed as a percentage of the object thickness examined. If, for instance, the diameter of the thinnest wire visible to the naked eye is 0.2 mm
and material thickness at the point of exposure is 10 mm, wire discernibility or wire
recognizability is quoted as 2 %.
As emphasised above, the use of an IQI does not guarantee detection of defects of comparable size.
It would be incorrect to say that because a wire of 2 % of the object thickness can be seen
on the radiograph, a crack of similar size can also be detected.
The orientation, relative to the X-ray beam, of a defect plays an important role in its discernibility (see section 12.1.)
There are various types of IQI, but the four most commonly used are:
1. the wire type (used in most European countries)
2. the step-hole type (still occasionally used in France, but the wire type is generally
accepted as well.)
3. small plates with drilled holes, called penetrameters, which are used for
ASME-work, although the ASME-code nowadays includes the wire-type IQI.
4. the duplex IQI.
In some countries (e.g. Japan and France) additional means (such as step-wedges) are
used, to verify contrast and check the kV-value used.
At the location of the (step)-wedge, there must be a minimum specified difference in
density compared to the density at a location on the film where penetrated material
thickness equals nominal wall thickness.

Wire-type IQI according to EN 462-1


EN 462-1 standardises
four wire-type IQIs. Each
one is made up of seven
equidistant parallel wires
of various diameters, as
shown in figure 1-13.
Batch number

Batch number

Fig.1-13. Wire-type IQIs with different wire diameters

111

Table 1-13 shows the wire combinations for the four IQIs according to EN 462-01.
The diameters of the wires are given in table 2-13.
IQI
1 EN
6 EN
10 EN
13 EN

Wire numbers
1 to 7 inclusive
6 to 12 inclusive
10 to 16 inclusive
13 to 19 inclusive

Wire diameter from/to (mm)


3.2 to 0.80 inclusive
1 to 0.25 inclusive
0.40 to 0.10 inclusive
0.2 to 0.05 inclusive

Table 1-13. Wire IQIs according to EN 462-01.

EN-type IQIs are manufactured with wires of steel, aluminium, titanium or copper,
depending on the type of material to be examined. On each IQI the wire material is indicated. Fe for steel, Al for aluminium, Ti for titanium and Cu for copper.
Diameter
(mm)
Wire nr.
Diameter
(mm)
Wire no.

3.20
1

2.50
2

2.00 1.60
3
4

1.25
5

0.32
11

0.25
12

0.20
13

0.125 0.10
15
16

0.16
14

1.00
6

0.80
7

0.63
8

0.50
9

0.08 0.063 0.05


17
18
19

13.3 List of common IQIs

EN 462-01
BS 3971
ASTM 747
ASTM 1025
AFNOR NF A 04-304

The plaques have markings


showing their thickness in
thousandths of an inch. Each
plaque has three holes of diameters 1T, 2T and 4T. T being
the thickness of the plaque.
Thin plaques with T < 0.01
form an exception to this rule.
Hole diameters for these
plaques are always 0.01, 0.02
Fig. 3-13. ASTM plaque type IQI
and 0.04, so do not comply
with the 1T, 2T, 4T rule.
These types of plaque are identifiable through notches cut in the edge, by which they can
also be identified on the radiograph.
Originally it was standard practice to use a plate of 2 % of the specimen thickness, but at
present 1 % and 4 % plates are used too.
If T is 2 % of the specimen thickness and the 2T hole can be seen on the radiograph, the
attained sensitivity level is said to be (2-2T), etc. Equivalent sensitivity values in percentages are shown in table 3-13.
At least three sides of a penetrameter must be visible on the radiograph. The thickness of
the penetrameter in relation to the specimen thickness defines the contrast sensitivity.
The size of the smallest hole visible defines the detail sensitivity.

Figure 2-13 shows the five most common


IQIs. Their origin and description can be
found in the following standards:

0.40
10

American IQIs

Europe
Great Britain
USA
USA
France
Level
1 - IT
1- 2T
2 - IT

Equivalent (%)
0.7
1.0
1.4

Level
2 - 2T
2 - 4T
4 - 2T

Fig. 2-13. Examples of image quality indicators


Table 3-13. ASTM Equivalent image quality indicator

112

113

Equivalent (%)
2.0
2.8
4.0

French IQIs

13.4 Position of the IQI

The AFNOR-type IQIs originate in


France. They consist of metal step
wedges of the same material as the
object to be examined. The thickness of the steps increases in arithmetical progression. Each step has
one or more holes with a diameter
equal to the thickness of that step.
There are various models of step
wedges. The most common types Fig. 4-13. French AFNOR IQIs
are rectangular with square steps
measuring 15x15 mm and hexagonal with triangular steps measuring 14mm.
See figure 4-13.
Steps thinner than 0.8 mm, have two holes of the same diameter. For a step to be regarded
as visible, all the holes in that particular step must be clearly seen on the film.

To be of any value in checking the factors defining sharpness and quality, the IQI must
be placed on the source side of the specimen. If the source side is not accessible, the IQI
is placed on the film side. In the latter position visibility is no longer an indication of
geometric unsharpness, but still a check on the developing process and radiation
energy used.

13.5 IQI sensitivity values


It is important to realise that any IQI acceptance-value must be based on a particular
type of IQI and the thickness of the object being examined. When IQI sensitivity is
expressed in a percentage of object thickness, a lower recorded value indicates a higher
radiographic sensitivity, hence better image quality.

The French standard AFNOR NF A04.304 includes an addendum, which defines the
index of visibility.
For each radiograph a record is made of:
1. the number of visible holes (a)
2. the number of holes (b) of a diameter equal to or greater than 5 % of the material
thickness being radiographed.
The index of visibility N is given by the formula: N = a-b.
The value of N may be positive, zero or negative.
Image quality improves as the value of N increases.

Duplex IQIs
The duplex IQI consists
of a number of pairs
(duplex) of wires or
thin strips made of platinum or tungsten, of
increasingly smaller size
and diminishing distanFig. 5-13. Duplex wire IQI
ces for each pair.
Figure 5-13 shows such an IQI made up of pairs of wires.
The duplex IQI has been in existence for decades but is no longer current in conventional film radiography because of their high cost and limited possibilities of application.
It is, however, increasingly used in digital radiography, because it is perfectly suited to
determine contrast and (un)sharpness.
114

115

exposure
14 Film
and handling errors
Before a particular difference in density in a radiograph is attributed to a defect in the
object examined, it must be sure that it is not the result of incorrect handling- or processing of the film. It is, therefore, essential to be able to recognise such faults when examining the film in order to prevent their recurrence. It is often possible to identify faults
due to wrong processing by looking obliquely at the surface of the film while facing
towards the light, and comparing the two emulsion surfaces. The X-ray image usually is
identical on both sides of the film, while a fault in processing will frequently affect only
one surface, and can be seen as a change in reflection on the surface.
The most common faults, and their possible causes, are listed below:

Insufficient contrast
a: with normal density:
1. radiation too hard
2. over-exposure compensated by reduced developing time
3. unsuitable or wrongly mixed developer
4. prolonged development in too cold a developing bath
b: with insufficient density:
1. insufficient development
2. exhausted developer
3. unsuitable or wrongly mixed developer

Excessive contrast (i.e. lack of intermediate tones)


1. radiation too soft
2. under-exposure, compensated by prolonged developing
3. unsuitable or wrongly mixed developer

General lack of density


1. radiation too soft
2. under-exposure, compensated by prolonged developing
3. unsuitable or wrongly mixed developer

General excessive density


1. over-exposure
2. prolonged development or developing temperature too high
3. unsuitable or wrongly mixed developer

Insufficient sharpness
1.
2.
3.
4.
5.
6.
116

source-to-focus distance too short


source or object moved during exposure
film-to-object distance too great
dimensions of source or focus too big
poor contact between film and screens
wrong type of foil used
117

Grey fog (local or overall)


1.
2.
3.
4.
5.
6.
7.
8.
9.

unsuitable dark room safelighting


excessive exposure to safelight (i.e. too long or too close)
film accidentally exposed to X-ray or Gamma-ray or to white light
heavy scatter
film out-of-date or stored under unsuitable conditions (ground fog)
extreme under-exposure compensated by excessive developing
exhausted or wrongly mixed developer
film cassette with film exposed to heat (e.g. sunlight, heat from radiators etc.)
cassette not properly closed (edge fog)

Yellow fog
1. prolonged development in badly oxidised developer
2. exhausted fixing bath
3. insufficient rinsing between developing and fixing
Note: It may take months before yellow fog becomes apparent.

Dichroic fog
(i.e. greenish-yellow by reflected light, pink by transmitted light)
1. developer contaminated with fixer
2. film insufficiently rinsed after development and subsequently fixed
in exhausted fixer
3. film stuck to another film when placed in fixer (in which case the development
continues in the fixing bath)
4. prolonged development in exhausted developer
5. film partly fixed in an exhausted fixing bath, exposed to white light
and then fixed again

Mottled fog
A greyish, mottled fog generally means the film is out-of-date or that it has been stored
under unfavourable conditions, e.g. in damp surroundings.

Whitish deposit
1. water used to make up developer or fixer too hard
2. wash water too hard
3. film insufficiently rinsed after development

Clear patches
1. minute round spots with sharp edges: the film was not kept moving
in the first 30 seconds of development
2. drops of fixer or water fell onto the film before development
3. marks from mechanical damage to the emulsion before exposure
4. marks due to rapid and uneven drying of the film (this occurs when there are
still droplets of water on the film when placed in the drying cabinet)
5. clear patches can occur from the film sticking to another film or to the tank wall
during development
6. grease on the film slowing down or preventing the penetration of the developer
118

7. screen(s) in poor condition


8. foreign bodies (for example metal particles ) between film and screen
during exposure
9. small, clear, hollow spots (usually with dark edges) may occur when the emulsion
has been subjected to local attack of bacteria. This is generally the result of slow
drying in a warm damp climate, particularly if there are impurities in the wash water.

Clear lines or streaks


1.
2.
3.
4.

the film envelope has been scored with a pointed object before exposure.
film insufficiently moved during development
uneven drying (film has been carelessly wiped dry after washing)
drops of fixer or stopbath have fallen on the emulsion before development

Clear shapes
1. clear crescent shapes may appear when, before exposure, the film has been
bent between two fingers
2. fingerprints may occur when the film has been touched with dirty fingers,
contaminated for example with grease, fixer, stopbath or acid

Dark patches
1.
2.
3.
4.

drops of developer have fallen onto the film before development


drops of water have fallen onto the film before development
electrical discharge marks, especially at low relative humidity of the air
marks from mechanical damage to the emulsion after exposure

Dark lines or streaks


1. the emulsion has been scratched after exposure
2. the film envelope containing the film has been scored or written on
with a pointed object after exposure
3. insufficient agitation of the film during development
4. uneven drying
5. water or developer has trickled down the surface of the emulsion prior
to development

Dark shapes
1. dark crescent shapes (see clear shapes above); these are darker than
the surrounding area if the bending occurred after exposure
2. fingerprints: the film has been touched with dirty fingers
3. electrical discharge (see dark patches).

119

interpretation
15 Film
and reference radiographs
15.1 Film interpretation
The common term for film interpretation is film viewing. Film viewing in fact means the
evaluation of the image quality of a radiograph for compliance with the code requirements and the interpretation of details of any possible defect visible on the film. For this
purpose, the film is placed in front of an illuminated screen of appropriate
brightness/luminance. The edges of the film and areas of low density need to be masked
to avoid glare.
The following conditions are important for good film interpretation:
brightness of the illuminated screen (luminance)
density of the radiograph
diffusion and evenness of the illuminated screen
ambient light in the viewing room
film viewers eye-sight
Poor viewing conditions may cause important defect information on a radiograph to go
unseen.
EN 25880 provides detailed recommendations for good film viewing conditions.
The luminance of the light passing through a radiograph shall not be less than 30 cd/m2
and, whenever possible, not less than 100 cd/m2 (cd = candela). These minimum values require a viewing box luminance of 3000 cd/m2 for a film density of 2.0. The practical difficulties of providing the required luminance for a film density of 4.0 are considerable. The main problem with constructing a film-viewing box for these higher densities
is the dissipation of heat from the lamps. However, by limiting the film area requiring
such high power lighting, it becomes possible to view radiographs of a film density of 4.
The light of the viewing box must be diffuse and preferably white. Radiographs should
be viewed in a darkened room, although total darkness is not necessary. Care must be
taken that as little light as possible is reflected off the film surface towards the film
viewer. If the film viewer enters a viewing room from full daylight, some time must be
allowed for the eyes to adapt to the dark.
A yearly eye-test according to EN473 for general visual acuity is required while especially sight at close range needs to be checked. The film viewer must be able to read a Jaeger
number 1 letter at 300 mm distance with one eye, with or without corrective aids.
The trained eye is capable of discerning an abrupt density change/step of 1 %.
While interpreting, a magnifying glass of power 3 to 4 can be advantageous.
120

121

15.2 The film-interpreter


Apart from the requirements regarding viewing conditions and viewing equipment
the film-interpreter (film viewer) shall have thorough knowledge of the manufacturing
process of the object being examined and of any defects it may contain. The type of
defects that may occur in castings, obviously, differs from those in welded constructions.
Different welding processes have their own characteristic defects which the film interpreter must know to be able to interpret the radiograph.
To become a qualified NDT operator, various training courses, course materials and leaflets specifying the requirements they need to comply with, exist. The European NDTindustry conforms to the qualification standards of the American ASNT organisation.
So far, a training programme for film-interpreter has not been established in similar
fashion. Textbooks for example are not uniform. Sometimes, the IIW-weld defect reference collection is used, beside which the instructor usually has his own collection of
typical examples, supplemented with process-specific radiographs.
ASTM has a reference set of defects in castings available.
There are incidental initiatives to introduce classification of film-interpreters by level, in
a system comparable to the qualification of NDT-personnel. Some countries have already implemented such a system.

15.3 Reference radiographs


The two main areas for the application of radiography are weld examination and
examination of castings. Radiography is also used to check complex assemblies for proper construction, and for many other technical applications. The following selection of
radiographs illustrates the wide variety of possibilities for detection possibilities of
defects or errors.

Weld inspection:
The following examples are from the booklet published by GE Inspection Technologies,
called Radiographers Weld Interpretation Reference
Note: All of these examples illustrating a variety of defects in welds are also issued on poster
format (60 x 90 cm) by GE Inspectio technologies.

Offset or mismatch (Hi-Lo).


An abrupt change in film density across the width
of the weld image

122

Offset or mismatch with Lack of Penetration (LOP).


An abrupt density change across the width of the weld image
with a straight longitudinal darker density line at the centre of
the width of the weld image along the edge of the density change.

123

External concavity or insufficient fill.


The weld density is darker than the density of the pieces welded
and extending across the full width of the weld.

Excessive penetration.
A lighter density in the centre of the width of the weld image,
either extended along the weld or in isolated circular drops.

124

External undercut.
An irregular darker density along the edge of the weld image.
The density will always be darker than the density of the pieces
being welded.

Internal (root) undercut.


An irregular darker density near the centre of the width of the
weld image and along the edge of the root pass image.

125

Internal concavity (suck back).


An elongated irregular darker density with fuzzy edges, in the
centre of the width of the weld image.

Burn through.
Localized darker density with fuzzy edges in the centre of the
width of the weld image. It may be wider than the width of the
root pass image

126

Incomplete - or Lack of Penetration (LoP)


A darker density band, with very straight parallel edges, in the
center of the width of the weld image.

Interpass slag inclusions.


Irregularly-shaped darker density spot, usually slightly
elongated and randomly spaced.

127

Elongated slag lines (wagon tracks).


Elongated parallel or single darker density lines, irregular in
width and slightly winding lengthwise.

Lack of side wall fusion (LOF).


Elongated parallel, or single, darker density lines sometimes
with darker density spots dispersed along the LOF-lines which
are very straight in the lengthwise direction and not winding
like elongated slag lines

128

Interpass cold lap


Small spots of darker densities, some with slightly elongated
tails in the welding direction.

Scattered porosity.
Rounded spots of darker densities random in size and location.

129

Cluster porosity.
Rounded or slightly elongated darker density spots in clusters
with the clusters randomly spaced.

Root pass aligned porosity.


Rounded and elongated darker density spots that may be connected, in a straight line in the centre of the width of the weld
image.

130

Transverse crack
Feathery, twisting lines of darker density running across the
width of the weld image.

Longitudinal crack
Feathery, twisting line of darker density running lengthwise
along the weld at any location in the width of the weld image.

131

Casting radiography
For the interpretation of X-ray films of castings, thorough knowledge of the specific manufacturing process is required. The type of defects in castings varies for the different types of
materials and casting processes. Figures 15-1 and 15-2 show X-rays of complex castings.
These radiographs were made to check the overall shape and possible presence of casting
defects.
As it solidifies during the casting process, metal contracts and unless precautions are taken
shrinkage cavities can occur inside the casting.
These can take various forms, such as piping/worm-holes, (figure 15-3), sponginess or
filamentary cavities, depending on the rate at which the metal has solidified. When the
contracting spreads slowly through the metal, filamentary shrinkage (figure 15-4) or even
inter-crystalline shrinkage (figure 15-5) may occur, while if the solidification front shifts
rapidly, shrinkage cavities tend to occur (figure 15-6).
Gas cavities in the form of porosity or larger gas holes can occur either due to a damp mould
or release of gas from the molten metal, and can be particularly troublesome in cast light
alloys (figure 15-7). Cracks can also occur in castings.
If they are formed while the metal is still semi-solid they are usually called hot tears (figure 15-8); if they occur when the metal has solidified, they are called stress cracks or cold
tears (figure 15-9).
A collection of radiographs of defects in iron/steel castings is provided in ASTM E446, and
for aluminium in ASTM E155.

Longitudinal root crack.


Feathery, twisting lines of darker density along the edge of the
image of the root pass The twisting feature helps to distinguish the root crack from incomplete root penetration.

Tungsten inclusions.
Irregularly shaped lower density spots randomly located in the
weld image.
Fig. 15-1. Radiograph of an aluminium casting

132

133

Fig. 15-3. Shrinkage (worm-hole cavities) in a (high heat conductive) copper casting

Fig. 15-2. Radiograph of an aluminium precision casting. Exposure on D2 film at 75 kV/5 mA/3.5 min/film-focus distance 100 cm

134

135

Fig. 15-4. Filamentary shrinkage in an aluminium alloy casting

Fig. 15-6. Shrinkage cavities in a bronze casting

Fig. 15-7. Gas-holes and porosity in an aluminium alloy casting

Fig. 15-5. Micro shrinkage (layer porosity) in a magnesium alloy casting

136

137

Fig. 15-8. Hot cracks (hot tears)

Fig. 15-9. Stress cracks (cold tears)

Fig. 15-10. Radiograph of an aluminium casting with coarse porosity


Exposure on D7 film at 60 kV/5 mA/15 sec, film-focus distance 100 cm

138

139

Examination of assembled objects


In addition to radiography for detection of defects in welds and castings, it can also be applied
to check for proper assembly of finished objects as figures 15-12 and 15-13 illustrate

Fig. 15-12. Radiograph of transistors


Exposure on D2 film with 27 m lead screens at 100 kV/5 mA/2 min film-focus distance 70 cm.

Fig.15-11. Radiograph of 25mm thick aluminium-copper alloy casting with gas porosity
Exposure on D7 film at 140 kV/5 mA, film-focus distance 100 cm

Fig. 15-13. Radiograph of electric detonators, taken to check details of assembly

140

141

16 Digital Radiography
As in other NDT methods, the introduction of microprocessors and computers has
brought about significant changes to radiographic examination.
Chapter 17 describes a number of systems such as computed-tomography and radioscopy
which are made possible by newly developed technology for the rapid digital processing
of vast quantities of data.
But as this chapter (16) shows, computer technology has also entered the field of conventional image forming radiography, as applied in industry. The driving force was the
medical world where digital radiography already earned its credits and has become
standard technology. Along with a few other companies, GE Inspection Technologies has
developed digital systems with a wide range of computer-aided NDT applications. Partly
digital radiography does replace conventional film and to some extend also permit new
applications.
The following three main methods can be identified:
1. digitisation of conventional flexible X-ray films for the purpose of archiving and/or image
enhancement (manipulation)
2. digital radiography by means of phosphor coated semi-flexible imaging plates
and computer processing, so-called Computed Radiography or CR
3. digital radiography e.g. with rigid flat-panel detectors and instant computer
processing, referred to as Digital Radiography or DR, sometimes referred to as
Direct Radiography.
Each method has differing strengths, advantages and limitations that should be evaluated in terms of specific application, inspection requirements and economics: capital,
human investment and production (number of exposures in a certain time).
The major merits of digital radiography compared to conventional film are:

Digitally enhanced radiograph of a mermaid/man.


The radiograph negated a myth by demonstrating that it concerns
a man-made creature.
The British Museum

142

Shorter exposure time and thus potentially safer


Faster processing
No chemicals, thus no environmental pollution
No consumables, thus low operational costs
Plates and panels can be used repeatedly
A very wide dynamic exposure range/latitude thus fewer retakes

On the other hand image resolution of even the most optimised digital method is (still)
less than can be achieved with finest grain film. A few other limitations are also
explained in this chapter.
143

16.1 Digitisation of radiographs

16.2 Computed Radiography (CR)

Storing and archiving of chemically developed X-ray films not only demands special storage
conditions, see section 10.7, but also takes up quite a bit of space. Digitisation of these films
provides an excellent alternative that also prevents degrading. Special equipment has been
developed for this purpose.
Current digitisation equipment actually consists of a fast computer-controlled scanner that
scans the film spotwise in a linear pattern, identical to the formation of a TV-image, measuring densities while digitising and storing the results.
The laser beam spot can be as small as 50 m (micron, equivalent to one thousands of a millimetre) diameter, but the equipment can be adjusted for a coarser scan, for example
500 micron, and therefore shorter scanning times.

Digital radiography using storage phosphor plates is known as Computed Radiography


or CR for short. This filmless technique is an alternative for the use of medium to coarse-grain X-ray films. In addition to having an extremely wide dynamic range compared to
conventional film, CR- technique is much more sensitive to radiation, thus requiring a
lower dose, see figures 6-16 and 13-16. This results in shorter exposure times and a reduced safety area.
CR is a two step process. The image is not formed directly, but through an intermediate
phase as is the case with conventional X-ray films. Instead of storing the latent image in
silverhalide crystals and developing it chemically, the latent image with CR is stored (the
intermediate phase) in a radiation sensitive phosphor layer.
The image information is, elsewhere and later, converted into light in the CR-scanner by
laser stimulation and only then transformed into a digital image.
The phosphor layer consisting of fine grains has been applied to a flexible, transparent
carrier and been provided with a protective coating.
An additional laminate layer mainly determines the mechanical properties such as flexibility which is, however, not as flexible as that of an X-ray film.

The values measured are compared


with a calibrated density scale and
processed digitally. Density variations between 0.05 up to 4.7 can be
measured and digitised with e.g.
12 bit density steps ( 4096 grey
levels), equivalent to almost 0.001.
Films with a maximum width of
350 mm can be digitised in a single
run. Even for the smallest pixel size
Fig. 1-16. Desk- top film digitiser
of 50 m, approximately 4 mm of
film can be scanned per second, so for the largest standard film size (350 x 430 mm) this process would take approximately 2 minutes.
Scanners exist without length limitation of film, and adapters are available for digitisation
of roll films.
Apart from the greatly reduced storage space and (almost) deterioration-free archiving,
digitising also makes it possible to (re)analyse the films images on a computer screen
(see figure 18-16), with the possibility of electronic image manipulation, see section 16.8.
Thus defect indication details not discernible on the original film using a viewing screen can
be made visible.
Because scanners vary widely in resolution, dynamic range, and ability to scan dense films,
evaluation is required to ensure that adequate scanning fidelity is achieved. Depending on
the selected resolution many Megabytes are needed to store a single film, see paragraph
16.8. Archiving usually is done on a mass storage facility e.g.: CD-ROM, DVD etc.
For use in laboratory environments only, high resolution film digitisation systems exist
applying a scanning spot size of 10 m. This enables detailed analysis of particular film
areas.

144

Figure 2-16 shows the layered structure of this type of plate, which is generally called an
imaging plate or sometimes wrongly called imaging screen.
Note: Screens in the world of NDT, made of lead or another metal, are used to intensify
the effect of incident radiation or to reduce the effect of (scattered) radiation.
Protective coat
Storage phosphor layer

Conductive layer

Transparent P.E.T.carrier/base

Laminate

Fig. 2-16. Structure of the CR imaging plate

As a result of incident X-ray or gamma-ray radiation on the storage phosphor, part of its
electrons are excited and trapped in a semi-stable, higher-energy state. This creates the
latent image. These trapped electrons can be released again by laser beam energy, causing
visible light to be emitted, which can then be captured by a PMT (Photo-Multiplier Tube).
The wavelength of the laser beam and that of the visible light generated are of course
different to separate the two from each other.
145

Scanning (development) of the latent image is


performed by a laser-scanning device, containing the PMT and its electronics, which digitises the analogue light signal that is generated.
This process takes place in the phosphor scanner, or so called CR-scanner. There are various types of scanners. In the most professional
scanners, all that needs to be done is to insert
a cassette in the input tray and the machine
automatically completes the processing cycle.
When this process is completed including
erasing the latent image the cassette is released from the CR-scanner and ready for re-use.
Figure 3-16 shows a typical tower type - mansize - automated scanner.

LFront of cassette
Lead intensifying foil
Magnetic sheet
Lead screen
Lead screen
Steel sheet
Back of cassette

Fig. 5-16. Structure of a CR cassette with storage phosphor

Figure 5-16 shows a cross-section of the CR-imaging plate in a cassette. The steel and
magnetic plates ensure that the various layers are pressed evenly and closely together.
Fig. 3-16. Automated CR-scanner

In smaller and portable desk-top scanner models


intended for use at remote locations e.g on offshore platforms, the CR imaging plate is manually
removed from the cassette and inserted into the
scanner, which slightly increases the risk of the
plates being damaged.
To this end the cassette can be opened, as shown
in figure 4-16.
CR-plate

Fig . 4-16. Opened CR-cassette

CR-plates can be exposed to subdued light for a


few minutes without any consequences for image
quality. The scanned image is ultimately made
visible on a high-resolution monitor (computer
screen) of the workstation, see figure 18-16.

The phosphor-crystals on a CR-plate react almost linearly to incident radiation while with
a conventional film the silver-halide crystals react exponentially, see figure 6-16. As a
result the dynamic range of a CR-plate is much wider than for conventional film, which
makes exposure times less critical, reducing reshoots (retakes), and allows various material thicknesses to be examined at the same time. Furthermore, dose sensitivity (speed)
is five to ten times higher as well, compare point A and B at a density of 2 (see also figure
13-16) allowing shorter exposure times or weaker sources, reducing the controlled area,
or even for some thin wall exposures apply other sources, e.g. Iridium192 to replace
Cobalt60, which can be an advantage from a radiation safety point of view.
Unfortunately, the image quality reduces. Iridium192, with a lower energy than
Cobalt60, requires a longer exposure time and this in turn reduces the image quality due
to the larger quantity of scattered radiation.
Note: CR-plates are more sensitive to this scatter (more noise) than conventional film.

Digital

The somewhat flexible CR-cassette can be re-used many times ( > 1000 times), provided
it is handled with care. Cassettes are available with or without lead screens.

B
Dynamic
range film

Intensity digital methods

The plate is scanned in a linear pattern identical to the formation of a TV-image.


Depending on the line distance selected, typically 50 or 100 micron, scanning speed is
5 to 10 mm per second. This is similar to the speed for digitisation of a radiograph.
In the scanner, the latent image is not only read but also subsequently erased (reset), and
therefore the CR-imaging plate is immediately available for the next exposure.

Density film

Film

For on-stream applications Iridium


can replace Cobalt for pipes with a
diameter up to 6 (150 mm), with still
an acceptable image quality, or even
8 (200 mm) in case of thin wall pipe.
The general rule is: the shorter the
exposure time the less the scatter thus
the better the image quality.

Dynamic range DR

Those developed especially for the NDT-market have built-in intensifying lead screens at
the source side, and a second lead screen at the back to absorb radiation caused by
backscatter. These multi-layer cassettes are not flexible anymore but can be re-used more
often than the flexible cassettes (several 1000 times).

Due to ongoing improvement efforts


the relative image quality of the
Fig. 6-16. Density/intensity versus dose for film and digital methods phosphor plate is, in the meantime,
equal to the quality obtainable with a medium-grain conventional X-ray film,
see figure 13-16. In fine-grain films, graininess is only a few microns, while in current
(2006) phosphor plates this is still considerably more (25 micron).

146

147

Relative dose

After exposure the intensity of the


stored information, in the then semi
stable phosphor layer, decreases over
time. Scanning within 1 hour of exposure provides the best results, typically half of the information is lost after
24 hours. Thus to avoid this fading,
scanning of the CR-plate should not be
delayed any longer than necessary.
To optimise use of the CR-imaging
plates in practice, a small handheld
terminal as shown in figure 7-16 has
Fig. 7-16. Terminal for CR-imaging plates
been developed to superimpose specific project- and exposure information to the images. To this end the cassette contains a
micro-chip which can receive (wireless) information from the terminal. At site and prior
to the exposure the relevant information is sent from this terminal to the micro-chip on
the cassette. The specific data are ultimately added to the image in the CR-scanner.
Once the data from the micro-chip is erased the cassette is ready for re-use.

16.3 Direct Radiography (DR)


Digital radiography is also known as direct radiography, for short DR. With DR-technology,
there is an immediate conversion of radiation intensity into digital image information.
Exposure and image formation happen simultaneously, allowing near real-time image
capture, with the image/radiograph available for review only seconds after the exposure.
This almost instant image formation is the reason that DR is considered the only genuine
method of digital radiography.
Some of the devices even provide a true real-time (radioscopic) mode with display rates up
to 30 images per second.

For DR, flat panel detectors in a variety of sizes are used, up to approximately 400 x 400 mm
(maximum in 2006) as shown in figure 8-16, that convert incident radiation intensity into
proportional and digitised electronic signals.
These digital signals can, by means of a computer and screen (workstation), without
intermediate steps, be presented as a coherent radiographic image. A cable typically links
the detector to this workstation from which the panel is controlled as well.
There are different types and suppliers of DR-flat panel systems. A variety of flat panel
systems exist with a wide range of pixel sizes and resolutions. The more and smaller the
pixels the higher the potential resolution of the system.
As sensor materials amorphous silicon and amorphous selenium are in use.
As sensors CCDs (Charge Coupled Devices) and CMOS (Complimentary Metal Oxide
Semiconductor) are applied.
The most common high resolution flat panels use amorphous silicon technology.
This material converts incident radiation into light. The conversion is proportional to the
radiation dose. This light in turn is converted from light into a proportional electric signal
by a scintillator made of e.g. structured Cesium Iodide (CsI) photodiodes and integrated
thin film transistors (TFTs).
Each picture element (pixel) contributes to the radiographic image formed on the screen of
the workstation. Each element is square in effective area, with pixel pitch typically ranging
from 50 to 400 micron. The smaller the pixels the better the resolution. Development is in
progress to make sensor elements/pixels smaller.
Depending on overall active area and detector pixel pitch, a panel consists of up to several
millions of such elements/pixels.
Figure 9-16 shows the different active layers of a flat panel detector which are deposited on
a glass substrate with a graphite cover on top.

Radiation
(photons)
Amorphous silicon array (detector)
light
CsI Scintillator - TFT photodiode array
electrons
Read-out electronics + digitisation
Fig. 8-16. DR flat panel component
with 400 x 400 mm active area.

Assembled flat panel detector


Physical data:
Dimensions: ~500 x 600 x 100 mm
Weight: ~ 10 kg

148

Digital data to workstation

Fig. 9 -16. Schematic of a two-stage flat-panel digital detector.

149

In practice DR has proven to be an excellent tool for the NDT-industry, however some
limitations apply as well:
Flat panel detectors can be used continuously for years in mass production processes,
to some extent however their lifetime is limited by the accumulated radiation dose.
The ultimate lifetime is determined by a combination of total dose, the dose rate and
radiation energy. The plates are less tolerant for high than for low energy radiation,
hence extreme high energies should be avoided. Thus the ultimate lifetime is dependent
on its application.
With the millions of pixels it is normal that over time that a few pixels become less
responsive, similar to pixels of flat panels as used with (notebook)computers. Usually
the un-acceptable number and pattern of dead pixels is specified by the manufacturer.
Fortunately, in cases a small area of the panel is out of order, an experienced
interpreter of DR-images is able to differentiate (by pattern recognition and known
position on the panel) real component defects from less responsive pixels.
Flat plate detectors are also subject to some memory effect, in jargon called ghosting.
This is due to hysteresis of the scintillation layer after exposure. The image slowly fades
away, particularly in case of high energy levels exceeding a few hundred kV.
This hysteresis causes a certain dead time of the system, from seconds to minutes
depending on the radiation energy, during which the plate cannot be re-used.

Resolution is expressed as the number of lines that can be distinguished in one mm,
see figure 10-16. The scientific measure to quantify the mutual effect of distance and
contrast on resolution is the Modulation Transfer Function, for short MTF.

16.4 Image quality of digital radiography - MTF and DQE

Fig. 11-16. Effect of contrast versus noise

Image quality is the total result of resolution, contrast and noise.


Conventional X-ray films exhibit an extreme high intrinsic resolution due to the fine granularity of the radiation sensitive crystals (a few microns).
The resolution of the resulting image is far better than the human eye can resolve. Hence the
use of IQIs for film is an adequate measure of resolution and image quality which meets the
qualification needs of industry. As such there is no need for any additional resolving criteria.
However, for digital radiography with a much lower intrinsic resolution (typical 100 micron
or less) a different situation exists. To select or purchase the proper digital system, information which quantifies the resolving power of a digital system is needed.
Although generic methods to measure optical resolution do exist, they are not yet specified
for digital radiographic systems.
To fulfil the need for a qualification of resolution, prior to anticipated release of future standards, suppliers of digital radiography systems already use methods and definitions which
are common in other sciences.

Exposure conditions such as thickness of the object, focal spot and source to film distance
influence geometric unsharpness (blur) as well as contrast and noise of the system and
image, see figure 11-16. Therefore the definition of Detective Quantum Efficiency, for
short DQE, has been introduced. The DQE value, is a mathematical expression indicating
the overall aspects of image quality and is a combination of MTF, sensitivity and noise:

Resolution is defined as the smallest separation (distance) between two objects that the
human eye can distinguish. Because the
human eye is not easily quantifiable, an
objective method to indicate resolution is
needed. Resolution is dependent on contrast
(grey levels) and separation (distance).

1 mm
1 Lp/mm

Increasing contrast

WORST
Decreasing noise

Each component in an imaging chain


has an individual MTF. The MTF of a
whole system, is the product of the
MTFs of the individual components.
By measuring MTF, a relative value ranging from 0 to 1 (or 0 to 100 %) , the resolving capability of a system is given.
This characteristic value is typical for a
system. This means that flaws will be more
easily visible at a higher MTF.
A system should have an MTF that matches the MTF associated with the most
demanding application. However the
quality of a digital image is not only
determined by the systems MTF-value
but also by the exposure conditions.

DQE Image quality / Dose


Noise in turn depends on speed, thus the time needed for an exposure to create an image
which might include signal averaging to achieve the required image quality. Reduction of
noise by averaging the signals from a number of exposures increases the image quality considerably as illustrated by the images of figures 11-16 and 12-16. Unfortunately signal averaging in turn lowers the DQE proportionally due to increased exposure time.

1 mm
4 Lp/mm

Fig. 10-16. Resolution: line pairs per mm

150

BEST

Fig. 12-16. Effect of image- averaging on noise

151

In general, the DQE-value consolidates the plate- or panel performance of many individual
parameters (resolution, efficiency/exposure time, noise etc.) in a single number. DQE value
ranges from 0.0 to 1.0, numbers in practice differ from 0.05 to 0.9.

The area for true real time (real instant images) shows that exposures can be made with
extreme low dose but at cost of image quality.

In summary: MTF quantifies the performance of the plate or panel, DQE quantifies the
overall performance, thus MTF of plate or panel including the exposure itself.
In other words: DQE quantifies the ability of a detector to accurately provide the information present in the X- or gamma rays including the imaging process. Thus the DQE value
indicates the final image quality and the inspection time required for a given application.
Remark: Despite the efforts to introduce MTF and DQE in the NDT industry - a value which
for the actual users of CR and DR is rather abstract and hard to comprehend - in practice the
IQI is still (in 2006) the only indicator of image quality. Almost always the duplex IQI is
applied for CR and DR. The duplex IQI consists of sets of two metal wires close together.

With storage phosphor based imaging plates for computed radiography, the conversion of
radiation into an image is a two-step process. The DR-technique, however, immediately
(instantly during exposure or within seconds following exposure) produces an image on the
screen of the workstation (see figure 18-16). That makes DR extremely useful in automated,
robotic, production processes.
Although DR, with the correct exposure parameters, offers a higher relative image quality
than CR, flat panel detectors are less suitable for field use and for applications with difficult
access requirements due to their physical size (thickness) and inflexibility/rigidness.
Moreover flat panel detectors require a considerable higher capital investment compared to
the CR-method. Although the electronics needed for both methods, e.g. workstation, cost
approximately the same, a flat panel detector however (~ 150,000 ) is roughly a 200 times
more expensive than a phosphor plate (~ 750 ). Hence selection of a DR-solution requires
careful considerations with regard to return of investment. (pay back period).

16.5 Comparison of film, CR and DR methods


The choice of which technique to use depends first of all on the requirements with regard to
image quality. In both the CR and DR methods the same IQIs as used in conventional radiography are applied to check the radiographic process and image quality.
The major parameters to compare the three methods are speed (dose needed for creating
the image) and image quality (noise, resolution, contrast). Figure 13-16 illustrates graphically the relative image quality of D-type films, RCF-film, CR- and DR techniques.

16.6 Selection of the CR and DR methods

Another aspect which influences the selection between CR or DR is the availability (or lack)
of industrial standards. Standards exist for CR, for DR they do not yet exist which hampers
wider use of the strong potentials of the DR-method. Initiatives are deployed (2006) to work
on compilation of such standards for DR.

This overview shows that the best


image quality (best IQI visibility) of
F
CR-plates is similar to what can be
achieved with medium to finer grain
True real-time
Radioscopie
film (compare point A with B) but is
DR-Panels
~ 5x faster. At point C the quality is
CR-Plates
less than what can be achieved with
RCF-Films
C
course grain film but the speed is
D7 (coarse grain)
>10x faster compared to point B.
D-Films
B
RCF-films (5 to 10x faster than D7A
E
film) are positioned in the same area
D
D2 (fine grain)
as CR-plates.
Relative dose
The graph for DR-panels is based on
Fig. 13-16. Relative image quality and speed of the various
the results obtained with common
radiographic methods
flat-panel detectors with different
This graph appears enlarged in the appendix on page 192.
number of pixels (25 to 400 micron).
The best quality that can be achieved with DR-panels comes close to fine grain film D3
(compare point D with point E). The graph also shows that the speed is much higher to
achieve the same image quality of D-type films.
Depending on the required image quality, a time saving of at least a factor 20 (D against E)
and roughly 200 (F against E) can be achieved, however with poorer quality.

In summary: Numerous aspects with a great diversity such as: image quality, process speed,
productivity, portability, robustness/fragility, (in)flexibility of plate or panel, available field
space, logistics, environmental issues, capital investment, human investment, (non)existence of industrial standards etc. play a role in the ultimate choice between conventional film
or CR or DR including type of DR plate (size and number of pixels) to be used.

152

153

Better

Relative image quality DQE

Faster

16.7 Applications of CR and DR methods


For certain applications, e.g. when the requirements for image quality are less stringent and
normal or coarse-grain film could be used, the CR-technique is an excellent alternative to
film. Examples include on-stream radiography (using isotopes) to detect internal erosion or
corrosion of non-isolated piping, see figure 14-16 and detection of internal and external
corrosion under thermal insulation (CUI ) see figure 15-16.
For wall thickness determination of (insulated) pipes the so-called projection (shadow technique) or tangential technique is applied, see paragraph 18.6.

The reduced exposure times - in practice factor 2 to10 - or sources with lower energy that
can be used, are deciding efficiency- and safety factors (smaller safety area). For the examples given, the CR-plates usually do not need to be bent and rigid cassettes can be used, which
prolongs the life of the plates.
DR-systems are more complex and vulnerable than equipment for CR or film radiography
thus less suitable for harsh field conditions. Moreover the DR-method (in year 2006 )
although better than CR still has a limited resolution compared to conventional film and
cannot (yet) replace high resolution field-radiography as usually is required by written standards for weld inspection. Work is in progress to improve the merits and resolution of the
DR method.
The choice of DR-flat panel detectors in suitable applications
depends on the image quality
required, described in product
specifications or in-house procedures, and the number of parts to
X-ray beam
be inspected to make it cost effective (return on investment).
Flat panel

Fig. 14-16. CR- image of bare pipe with areas marked


for WT measurements

Fig. 15-16. CR image of insulated pipe with WT-values

The CR-technique is also very suitable for detection and quantification of scaling or clogging, concrete inspection and non-critical castings. Although conventional film is still superior compared to the
CR-technique, in several cases CR provides
sufficient image quality for weld inspection.
Figure 16-16 shows an
image of a weld with a
clear indication of a
serious defect.

Fig. 16-16. CR-image of a weld with a longitudinal defect

154

High-performance DR-detectors
are most suitable on stationary
X-ray head
locations, for example as part of a
Component
production line where vast numbers of precision components are
Fig. 17-16. DR flat panel in use on turbine blades
checked at high speed with the
lowest possible radiation dose, or in situations where mechanical automation can be applied
to achieve significant throughput improvements, see figure 17-16.
CR-plates (by handling), like DR-detectors (by radiation) have a finite useful life.
The working life of flat-panel devices can range up to millions of images, dependent on
application-specific details, see paragraph 16.3.
Thus cost-per-image should be considered in any return-on-investment financial analysis.
CR plates in flexible cassettes can be used over a thousand times, if used in a rigid cassette
they can be used three times longer.
Not only are CR- (~ 5x to10x) and DR-techniques (~ 20x with film-quality, to 200x with low
quality) much faster than standard X-ray film exposures, another attractive feature is the far
greater dynamic range/latitude (> 1000x).
These methods are, therefore, not over-sensitive to variations in radiation dose and very
tolerant of less than exact exposure times, see figure 6-16. This can reduce so called reshoots
(retakes) and can decrease the need for multiple exposures in some parts with different
thicknesses, thus further improving inspection throughput.
155

In all applications, an analysis is appropriate to determine the best method (film/CR/DR)


and the appropriate technique (including film-grade and resolution of the scanner, or CRplate and reader, or DR-detector type). Key considerations should include image quality
requirements, do inspection standards or procedures allow application of DR, and the financial benefits from throughput and process improvement.
For the CR method a standard EN 14784 has been issued in year 2005, part 1 describing
classification of systems and part 2 describing principles and applications. ASME V CC 2476
also addresses the use of the CR-technique. These standards support industrial applications
and increase the use of the CR-method. That is the main reason that initiatives are deployed to develop standards for the DR-method as well.

16.8 Work station and software


A computer and extreme high-resolution display screen are recommended for digitised films
as well as for displaying and processing the images obtained with the CR- and DR-techniques.
The number of pixels of the display screen should at least match with the digitisation spot- or
pixel size of the applied CR- plates or DR- panels to achieve maximum resolution.
Radiographic images contain more information than the human eye can discern. For this
purpose workstations, as shown in figure 18-16, are used as an image-processing centre.
This workstation operates with powerful dedicated proprietary software (e.g. Rhythm of
GE Inspection Technologies) to manage, process and manipulate images.
Images can be manipulated and enhanced in many ways: brightness, contrast, sharpness,
noise suppression, rotation, filtering, inversion, colouring, magnification, zoom-pan-scroll,
etc. This way, hidden details can be made visible, see figure 12-16 and 19-16.
Integrity-procedures should be applied to prohibit possible forgery of digital images.

Fig. 19-16. Contrast enhancement without loss of data

Images can be manipulated and enhanced in many ways: brightness, contrast, sharpness,
noise suppression, rotation, filtering, inversion, colouring, magnification, zoom-pan-scroll,
etc. This way, hidden details can be made visible, see figure 12-16 and 19-16.
Integrity-procedures should be applied to prohibit possible forgery of digital images.
In addition algorithms have been developed for e.g. the comparison of parts of an image
with conformance criteria, carrying out dimensional checks (sizing), remaining wall thickness measurements (see figure 20-16), determination of metal loss due to corrosion, defect
area measurement, providing image statistics etc.
Apart from the original image and its exposure parameters, on a true copy comments and
display characteristics (e.g., zoom, contrast, filters) can be superimposed and archived as
well. This enables inspection professionals to streamline the process and improve the
quality of distributed inspection information.

OD

ID

WT

Fig. 20-16. Wall thickness profile from on-stream image of figure 14-16 and report with statistics of all WT measurements

Fig. 18-16. Workstation

156

157

Figure 21-16 shows a screen shot made from the workstation. The screen shows the results
of an on-stream exposure on a CR-plate taken of a valve with connecting pipe. The screen
shot includes one of the selectable frames of the report module.
The image itself shows marks (white lines) superimposed by the operator at the workstation
to establish the remaining wall thickness at those places to be calculated by the software.

Archiving can be done on almost all existing professional mass storage solutions, e.g.
CD-ROM up to 700 MB, single layer DVD (~5 GB), double layer DVD (~10 GB) or in the
near future, HD-DVD (~15 GB), blue ray disk (~25 GB) and more distant future holographic disk (~300 GB).
Such mass memory capacity of several GBs (GigaBytes) is needed to be able to store a
number of high resolution digital images.
A single image of a ~ 400 x 400 mm panel with a pixel size of 50 micron requires 120 Mb
(position and up to16 bit of density data). A pixel size of 100 micron needs only 30 MB.
The workstation can also transfer images electronically over great distances (through
internet, intranet or wireless), which can be viewed, interpreted or stored by remote users
on identical satellite workstations. This way information is send to the experts rather than
sending the experts to the information.
Because the images are digital, multiple copies of the images are always identical.
These capabilities are driving the latest trends of enhanced database capabilities and
common workstation standards for digital radiography software.

WT 1
3.5  0,1 mm

WT 2
1.3  0,2 mm

Figure 23-16 shows a block diagram of the various components that make up a complete
system for digital radiography.

Wireless
transmission

DR- panel

Fig. 21-16. On-stream image and report of a valve

Wall Thickness Result


Nr.
1
2

Material
Steel

DA

WDSoll

WDIst

3,5
3,5

3,5
1,3

Figure 22-16 shows a detail of the


selected pipe wall area with the
reported results.
This example of a valve with a
great variety of wall thicknesses
also shows one of the strengths of
a digital exposure.
If wanted the same image can be
used to study the thick wall parts
of the valve due to the large dynamic range contained in the image.

Main
workstation

Printer

CR - tower

Film
digitiser

Storage/
Archive

Satellite
workstation(s)
Fig. 23-16. Block diagram for digital radiography with workstation and supporting equipment

Fig. 22-16. Detail of pipe wall of figure 21-16 with report

158

159

CR-plate
in cassette

17 Special
radiographic techniques
The previous chapter (16) dealt with techniques that would be impossible without the aid
of computers. These techniques share a common feature, whereby the processing, interpretation and storage of data is done by a central computer and monitor, also called the work
station. In this chapter (17) computers also play an ever increasing important role in some
of the techniques discussed. Computertomography (CT) and the Compton back scatter
technique for example would not exist without them.

17.1 Image magnification techniques


With the introduction of micro-focus X-ray tubes, which have focal spots of 10-20 m, new
special techniques have been developed.
By positioning an object close to this micro-focus X-ray tube as illustrated in figure 1-17, and
by placing the film or detector at a greater distance, a magnified image is obtained. Any
defects will consequently be also magnified but still have sufficient sharpness.

geometric
unsharpness

focal spot

film,
detector,
image intensifier

Fig. 1-17. Image magnification

density
brightness

Any unsharpness, as illustrated in figure 1-17, is determined by the relationship between


F1 and F2, and the size of Uf .
The effective unsharpness is calculated as follows: Ug = Uf (F2 - F1) / F1
For example:
A 150 kV X-ray tube with a focal spot size of 20 m with focus-to-object distance (F1)
of 50 mm and focus-to-film distance (F2) of 550 mm will have a geometric unsharpness of:
0.02 (550 - 50)
= 0.20 mm
50
F
550
= 11
The magnification factor is: 2 =
50
F1
160

161

Compared to standard exposures, the image magnification technique onto film using microfocus X-ray tubes, has the following advantages:
Smaller defects are discernible,
Less back scatter because a smaller part of the object is being irradiated,
Higher resolution, as the image but not the film grain is magnified.

Stationary real-time installations


Display monitor systems, as illustrated in figure 2-17, are almost exclusively used in stationary set-ups for production line testing of varying types of objects, in particular in metalcasting plants, pipe mills and component assembly industries.

Disadvantages are:

Sometimes real-time systems are utilised in the food industry to check for instance for the
presence of glass fragments or other foreign objects.

Costly if separate high-vacuum equipment is required (see section 5.1),


Time-consuming, as for each exposure only a small part of the object is being irradiated,
hence more exposures are needed.

Being part of a production line and due to the necessary radiation safety provisions (such as
cabins) these systems can be very expensive.
The display monitors are located at a safe distance.

The magnification technique is mainly used in combination with a radiation-sensitive device such as fluorescent screen, image intensifier or flat panel detector, and a CCTV-system
placed at a safe distance.

micro
processor

The CCTV-system can be replaced by a computer workstation for image processing and/or
enhancement prior to interpretation, as figure 2-17 shows.

digital
signal

digital
signal

17.2 Fluoroscopy, image intensifiers


object

Fluoroscopy, also known as radioscopy, is a technique whereby real-time detection of


defects is achieved by the use of specialised fluorescent screen technology.
At present, there are many alternatives to photographic film for making an X-ray image
visible. In addition to the CR and DR techniques described before (chapter 16), a wide range
of real-time image forming techniques using display monitors are available.
These systems are considered also to belong in the DR category.

image processing
system
archiving

flat panel detector


or
image intensifier

display monitor

X-ray
tube
Fig. 2-17. Radiography with image magnification

The choice of a radiographic system to be used depends on a number of factors:


It can generally be said that the image quality of conventional X-ray film is superior to either
digital direct radiography (DR) or computer-aided film radiography (CR) techniques.
Thus these new techniques cannot be considered acceptable alternatives at all times.
However, when the installation is adjusted to optimal refinement for a single application, for
example weld inspection in a pipe mill, a film-equivalent image quality can be obtained,
which would only just comply with the requirements. This would e.g. demand the use of a
micro-focus tube, see section 5.1.

162

Hardness of the radiation required and proper detector.


Resolution or detail discernibility required. The type of defects to be detected in
mass-production is normally known.
Magnification factor required when it concerns small defects.
Image dynamics (density range) with regard to object thickness range.
Image contrast required facilitating ease of defect detection. Sometimes this can be
automated when it concerns common defects.
Time restraints, number of objects to be examined per unit of time.
Budget.
Space available.
Installation and specimen dimensions.
163

A number of these factors also influence the choice of detector system.


1. pipe
2. sheet metal (Al) cladding
3. insulation material
4. collimated radiation source
5. miniature flat panel detector

Some of the options are:


Phosphorescent screen (afterglow) with TV camera and display monitor (CCTV)
at a remote (safe) location
Fluorescent screen (instant image) with CCTV-system at a safe location
X-ray image intensifier with conversion screen, in combination with a CCTV-system
CCD-camera as a substitute for the relatively slow conversion screen
Photo array detector, minimal size per diode approx. 0.25 mm to inspect slowly moving
objects (airport luggage checks)
Flat panel detector, FPD, consisting of millions of light-sensitive pixels. The light pulses
can be converted in different ways into electric pulses, which in turn can be processed
digitally.
Although the image intensifier is still most commonly used, the flat panel detector is
becoming more and more attractive. Flat panel detectors provide various pixel sizes with
extensive image dynamics (a very wide density range, far greater than is possible with film).
Since the signals received by the computer are digital, the screen image can be optimised for
interpretation (contrast, brightness, sharpness, magnification, filtering, noise suppression)
and subsequently stored. These advanced systems also offer the possibility of comparing the
image obtained with a reference image and of automatic defect interpretation.
Selection of the most suitable (expensive) system is made even more difficult because of the
rapid development in sensor- and electronic technology.
Fluoroscopy, imageintensifiers and-magnifiers are more elaborately described in the
booklets: Die Rntgenprfung and its translation The X-ray inspection[3].

Portable real-time equipment


A portable version of real-time equipment is used to detect external corrosion under thermal insulation. It is generally very difficult to detect corrosion on piping with insulation still
in place, whereas removing and re-installing the insulation is a costly and time consuming
operation. Sometimes the likely presence of corrosion is indicated by moisture/water detected in the insulation, see section 17.4.
External corrosion in a low-alloy steel pipe becomes apparent by local swelling
of the pipe surface as a result of volume increase of the corrosion layer. Figure 3-17 illustrates a system by which the swelling and even severe pitting can be detected. On one side is a
strongly collimated source or X-ray tube that must be aligned in such a way as to direct a narrow beam of radiation along the tangent of the pipe towards a flat panel detector behind.
164

6. tangent ( pipe horizon)


7. portable monitor
8. adjustable bracket
9. sliders/rollers
10. corrosion spot

Fig. 3-17. Schematic arrangement of a portable real-time radiographic corrosion detector

This way an image is obtained of the pipe horizon with possible presence of corrosion
(swelling or pitting). The image is presented real-time on a portable monitor. The batterypowered equipment uses soft radiation of low intensity, so that it can manually be moved
along the pipe. The system can also be used to locate welds under insulation.

17.3 Computer Tomography (CT)


For medical diagnostic purposes, techniques have been developed to obtain a radiographic
image with a resolution of approx. 0.5 mm. Powerful computers are used to transform all
absorption variations that occur when irradiating with a moving source, into a
comprehensive image. The technique is now also used in industry, e.g. for checking miniature electronic circuits as built into mobile telephones. Systems with a resolution of only a
few micrometers have already been successfully applied.
In the arrangement shown in figure 2-17, now the object to be checked rotates (continuously
or stepwise by a robot) to obtain a CT-image. The object is scanned section by section with
a very narrow beam of radiation (X- or gamma-ray). The receiver in this case is a flat panel
detector.
Each individual detector element measures, during a short exposure period, the total
absorption across a certain sectional plane of the object. A large amount of data has to be
stored and processed, in particular when an image of high resolution is required. A threedimensional representation (3D-CT) of the radiographic image requires vast computing
capacity.
In CT, absorption values are determined with a very high degree of accuracy, which means
that the contrast of an image can be varied over a extraordinary wide range.
Absorption/density variations of 0.02 % can be displayed in a range of density 6 and over.
This offers great possibilities for image processing.
165

With present day computers, depending on resolution required, the reconstruction-time of


a 3D-image runs from 2 to 60 minutes. Increasingly, 3D-CT is used on high-quality castings,
even with the extra facility of automatic object and defect identification. The subject of CT
is more elaborately described in the booklets (German) Die Rntgenprfung and (English)
The X-ray Inspection, see literature reference[3].
To analyse defects, 3D-CT systems exist which are able to accurately size random orinted cracks or other (planar) defects with a width of only  25 micron. Such a system can
be used to evaluate (flaws detected with other methods) in for example austenitic welds
(e.g. nuclear components) a task almost impossible to perform with ultrasonics.

17.5 Compton back scatter technique


The Compton back scatter technique, see section 2.6, benefited from the introduction of
computer technology into NDT equipment, just as most other methods discussed in this
chapter.
It is now an accepted NDT-technique for plastics and light metals [2].
The scanner comprises an X-ray tube and a detector consisting of a number of elements as
illustrated in figure 4-17. A collimator reduces the beam of rays to 0.5 mm in diameter, so
that it cannot irradiate the detectors directly.

17.4 Neutron radiography (neutrography)


Neutrons, which are atomic particles without an electric charge will penetrate most materials, are attenuated in passage, and so can be used to produce radiographs.
There are various kinds of neutron energies, but only the thermal and cold neutrons are suitable for NDT applications. Contrary to ionising radiation in the keV and MeV range, neutron absorption is higher in light than in heavy materials.
Neutrons will be influenced strongly by hydrogenous materials, plastics (all types), explosives, oil, water etc., even when these materials are inside metal containments made of lead,
steel or aluminium.
There are many potential applications for neutrography, but its use is limited to a large
extent by the lack of suitable, portable neutron sources. A neutron window in an atomic
reactor is by far the best source, but such facilities are not commonly available. The only
neutron-emitting radioactive source is californium252, which is extremely costly and has a
half-life of only 2.65 years.
An X-ray film also reacts to neutron energy, but not until it is combined with gadolinium or
cadmium intensifying screens usable results are obtained. The Agfa D3SC (SC = single coated) film is frequently used for this purpose. The secondary radiation generated in the intensifying screens brings about the image formation.
Another filmless application of neutron radiography in NDT is moisture-detection in insulation. This portable equipment that is on the market uses a very weak neutron source.
With the aid of this neutron backscatter method, the presence of water, actually that of
hydrogen atoms, is established. The presence of moisture is generally an indication of external corrosion in a pipe, or the likelihood that corrosion will occur in the near future.
The portable real-time equipment as described in section 17.3, or flash radiography described in section 18.7, can in some cases confirm the presence or absence of corrosion without
removing the insulation.
166

When a photon and an electron collide in the material, the primary X-radiation is scattered
as somewhat softer radiation in all directions, and thus partly also back from the material to
the scanner. This secondary radiation is then caught by the detector through a specially formed diaphragm, see figure 4-17.
The detector is made up of 20 or more detector elements marked A, B, C etc. each of which
measures the quantity of back scattered radiation from a certain depth (A, B, C) in the
object, as figure 4-17 shows. Each sensor element is, say, focussed on a certain depth.
The cylindrical scanner measures only 7 x 7 cm and scans the
object in a grid.
By linking the scanning system
object
with a data processor, a comprehensive Compton image of
the object develops and any
diaphragm
possible defects in it.
This method has the advantage
collimator
that an object needs to be accessible from one side only.
detector
detector
It is for instance frequently
applied to honeycomb conX-ray beam
structions and composite mateFig. 4-17. The Compton back scatter technique
rials and has a penetration
depth of approx. 50 mm.
The method is (still) fairly slow; scanning a 50 cm2 surface takes approximately 5 minutes.
An added advantage, however, is that the depth position of defects becomes known immediately as a result of the quasi-focussing of each individual detector element.

167

18 Special
radiographic applications
There are many special applications of radiography in NDT.
This chapter describes a limited number of different examples to illustrate this diversity.
Apart from the use of radiation in image forming radiography, it is also used in, for instance, measuring instruments such as metal alloy analysing instruments (Positive Material
Identification, PMI). This type of non-image forming instruments and applications are
outside the sphere of this book.

18.1 Measuring the effective focal spot


The effective focal spot is an important feature of an X-ray tube and is specified by the
manufacturer. In general it can be said; the smaller the better.
As focal spot size is also a critical exposure parameter (see section 11.1), the accuracy of
the manufacturers information is of vital importance.
Until recently, the film or pinhole method was commonly used. Since 1999, EN 12543-1
prescribes a standardised method which, however, does not have the general support of
suppliers, as it requires expensive instrumentation and is time-consuming.
The EN-method, suitable for effective foci <0.2 mm, involves scanning the X-ray tube radiation beam with a scintillation counter through a double collimator with an extremely
small opening of 10 m. The resulting intensity values are then represented in a threedimensional (isometric) diagram from which the effective focal spot can be deduced.
The film or pinhole method is still used by X-ray tube operators, to verify the equipmentmanufacturers data. Following the camera obscura principle, the X-ray tube projects its
focus through a very small hole in a lead plate onto a film.
The lead plate is positioned exactly halfway between focus and film. To prevent scattered
radiation, sometimes the hole is made in a tungsten plug which forms part of the lead
plate. After development, the effective focal spot size can be measured on the film, with the
aid of a magnifying glass.
The latter method, still allowed and accepted by EN, results in marginally smaller effective
focal spot sizes. Establishing the effective focal spot size of a panoramic X-ray tube is considerably more complicated.
It is therefore recommended to just make a radiograph of the object with the right IQIs and
check the results for compliance with the quality requirements specified.

Neutron radiograph of an iris flower made on Agfa D3 s.c. (single coated)


Organic substances are well suited to examination by this type of radiation.

168

169

18.2 Radiographs of objects of varying wall thickness

18.3 Radiography of welds in small diameter pipes

For radiographs of an object with limited differences in wall thickness, it is common to base
exposure time on the average thickness to obtain the required film density of at least 2. It
is possible that parts of the film are either under- or over-exposed if there are great differences in wall thickness. This can be explained by the shape of the toe (lower part) of the
characteristic curve of the X-ray film used. The film gradient (contrast) is lower and, consequently, so is the defect discernibility. In accordance with EN 1435, therefore, there is a
limit to the thickness range covered by one single exposure.

For pipe welds, the single wall-single image technique (SW-SI), or if this is not feasible, the
double wall-single image technique (DW-SI) is to be applied. For small diameter pipes this
alternative is not really practical, as a disproportionate number of double wall-single image
exposures needs to be made due to the limited effective film length (see section 12.2). In
such a case the double wall-double image technique should be used (DW-DI). Normally, the
DW-DI technique is only applied on diameters <75 mm and wall thickness of <8 mm. Both,
the weld on the source side and film side of the pipe are simultaneously interpreted.

There are a number of practical ways to prevent over-exposure of thinner and under-exposure of thicker sections. These can be divided in two groups: compensation by single film
or by two film techniques.

Two more DW-DI techniques are suitable for small diameter pipes:

For exposures on one film, the following techniques can be applied:


Reduce contrast by utilizing a filter on the X-ray tube to make the radiation harder.
Reduce contrast by increasing the radiation energy using higher tube voltage or using
Iridium192 or Cobalt60 sources.
Compensate the difference in wall thickness as the left sketch of figure 1-18 shows,
with material B of similar composition as object A.
Instead of insertion of B in the previous method, use a special putty (filling material)
mainly consisting of metal powder.
When two films are used, the following techniques can be applied:
Simultaneous use of two films of different sensitivity but similar screens, for a single
exposure. For example an Agfa D7 and D4 type film could be used. This is the least
complicated and most practical method (see figure 1-18 at right).
Simultaneous exposure on Agfa D7 and D4 films with different screens (see figure 1-18).
Make two exposures on film of the same sensitivity and screen type:
one with the exposure time based on the thinner and one on the thicker section.
Make two exposures on film of the same sensitivity but different screen types.

Single film technique


Material A - test object
Material B material of comparable absorption

Double film technique


Section C. evaluate on D4
Section D. evaluate on D7

Fig. 1-18. Compensating for differences in wall thickness

the elliptical technique and


the perpendicular technique
Elliptical technique
The elliptical technique, as illustrated in figure 2-18, is the preferred technique, but should
only be applied if the following conditions are met:
external diameter (De) is <100 mm (in practice 75 mm)
wall thickness (t) is <8 mm
weld width < De /4
The number of exposures is determined by the relation between wall thickness (t) and
diameter (De). If t / De is < 0.12, two images - rotated 90 in relation to each other - are
sufficient for 100 % coverage. If t / De is equal to or bigger than 0.12, three exposures rotated 60 or 120 in relation to each other (i.e. equally divided over the circumference) is considered to be a 100 % examination.

Film

Film

Fig. 2-18. Elliptical double walldouble image technique

170

171

When using the elliptical exposure technique, the images of the weld on the source side
and on the film side are shown separately, next to each other. The distance between two
weld images has to be approximately one weld width. This requires a certain amount of
source offset (V), relative to the perpendicular through the weld. The offset can be calculated with the following formula :
O= 1,2 . w . f/De

18.4 Determination the depth position of a defect


The depth position (d) of a defect can be determined by the parallax-method, as shown in
figure 4-18. The radiograph is exposed from two opposite angles. The required quantity of
radiation is equally divided over positions A and B. Only one film is used.
The shift in defect image on the film (G in mm) is a measure for the depth position; the
shift of the source (A to B in mm) and the source-to-film distance (H in mm) are important
data. The depth position is calculated with the formula: d = (GxH) / (AB+G).

In which:
w

= width of the weld

= distance from source to the source side of the object, measured perpendicularly

De
O

= external pipe diameter

Another, much more complex method of depth determination is stereo-radiography, by


which two separate films are exposed which are viewed simultaneously via mirrors.
However, this method is rarely used.

= Offset distance

Perpendicular technique

source shift

Alternatively, the perpendicular technique can be used if the elliptical technique is not
practical, (see fig. 3-18). This is the case when, for instance, pipes of different wall thickness are joined or a pipe is joined to a 45 / 90 bend.
Three exposures equally divided over the circumference are sufficient for 100% coverage.

Film

Film

Fig. 3-18. Perpendicular double wall-double image technique

= defect shift
Fig. 4-18. Determining the depth position of a defect

172

173

The actual pipe wall thickness (t) is equal to the


image on film (tf ) multiplied by the correction
factor (see fig. 5-18).

Determination the depth position and diameter


of reinforcement steel in concrete

Similar to the method for determination of the depth position of a defect in metals is the
determination of the depth position (cover) of reinforcement steel in concrete.
Subsequently, the true diameter of the reinforcing bar (D) can be calculated. Correction factor = d / (H-d).
The dimension of the radiographic image (Df) on the film is multiplied by this correction
factor. The true diameter of the reinforcing steel is therefore D = Df x d / (H-d).

18.6 On-stream inspection


On-stream inspection can be carried out on
pipes, valves, vessels, and distillation
columns while in operation, in order to establish the degree of deterioration of the
system either the projection or the tangential
technique can be used. Since the introduction of digital radiography, the CR-method
using storage phosphorplates, is increasingly
becoming an alternative for traditional film
in case of on-stream exposures, see chapter
16. The main advantage being that it reduces
the exposure time by a factor of 5 to 10, or if
lower energies (Iridium192 instead of
Cobalt60) can be applied it results in a reduced safety area, which is very attractive in
cramped spaces and personnel nearby e.g. on
offshore platforms.

(source)

Dpipe
Dinsulation
filter
film

Projection technique
or

The projection technique is most commonly


used. With this technique the two walls are Fig. 5-18. Projection technique for on-stream radiography
projected on film simultaneously, as shown in
figure 5-18. The image projected is larger than the actual object dimensions. It is important
to know the degree of magnification so as to be able to determine the true wall thickness.
If both walls of the pipe are projected on the film, it is straight forward to establish the correction factor, which is the true diameter (D) divided by the radiographic diameter Df .
This method should be used as much as possible.
With the projection technique, the source is placed at a certain distance from the pipe.
At a film-to-focus distance of 3 x Dinsulation and a source size of 3 mm, image quality requirement A of EN 1435 is met.
174

FFD

Dinsulation

Tangential technique

= source-to-insulation
distance

correction factor =

Most common is on-stream radiography of


insulated pipes, for which half the insulated
diameter determines the sharpness. In on-stream radiography it is important to know the
direction of the product flow, so that a existence of localised wall thickness reduction can be
better deduced. Films of 30 x 40 cm are generally used for pipe diameters up to 250 mm.
Larger diameters require more films.

In the pipe diameter range of 250 to 400 mm


the tangential technique, as shown in figure 618 is sometimes applied. Only one wall is projected. The perpendicular projection produces
a sharper image. This allows a shorter focus-tofilm distance, and consequently a shorter exposure time. Generally, a focus-to-film distance
of 2.5 x Dinsulation is chosen.
The correction factor would then be:
(2.5 x Dinsulation -0.5 x Dinsulation) / 2.5 x Dinsulation = 0.8.

filter
film
Correction factor =

(2.5 x Dinsulation -0.5 x Dinsulation) / 2.5 x Dinsulation = 0.8


Fig. 6-18. Tangential technique for on-stream radiography

wall thickness

18.5

Selection of source, screens and filters


The graph in figure 7-18 indicates which radioactive source is the most suitable, depending
on pipe diameter and wall thickness. The quality of the radiograph can be optimised by
applying filters and screens, see table 1-18.
Zone

Source type

Source size

I
II
III
IV
V

Iridium192
Iridium192
Cobalt60
Cobalt60
Cobalt60

2 mm
2 mm
3 mm
3 mm
4 mm

diameter in mm
Fig. 7-18. Areas of application for selection of source,
screen and filter in on-stream radiography
This graph appears enlarged in theappendix on page 193.

Screens
front and back
0.027 mm Pb
0.027 mm Pb
0.5 mm Cu of RVS
0.5 mm Cu of RVS
0.5 mm Cu of RVS

Table 1-18. Selection of source, screen and filter for the various areas in figure 7-18.

175

Filter
1 mm Pb
2 mm Pb
1 mm Pb
2 mm Pb
4 mm Pb

Exposure time
Obviously different exposure times are required for gas filled or liquid filled pipelines.
Below are a few examples.

Notes:
In the most commonly used insulation materials absorption is negligible.
The long exposure times cause over-irradiation at the edge of the pipe. As a result
the pipe wall shows up thinner.

For gas filled pipelines:


Depending on diameter and wall thickness
Focus-to-film distance
Irradiated thickness
Film type
Film density

: Iridium192 or Cobalt60, see figure 7-18


: minimum 3 x Dinsulation
: 2 x nominal wall thickness
: minimum C5 (EN584-1)
: minimum 2.5 in the centre of the pipe projection

For liquid filled pipelines:


Depending on the diameter, wall thickness
Focus-to-film distance
Irradiated thickness
Film type
Film density

: Iridium192 or Cobalt60
: minimum 3 x Dinsulation
: 2 x nominal wall thickness plus steel
equivalent of the pipe content
: minimum C5 (EN584-1)
: minimum 2.5 in the centre of the pipe projection

The steel equivalent of the pipe content is determined as follows:


(specific density in kg/m3 of content) / (specific density in kg/m3 of steel) x internal diameter
= .... mm of steel
Density of steel = 7.800 kg/m3
Density of content (oil and aqueous liquids) = 800 to 1.000 kg/m3

Fig. 8-18. Preparations for on-stream radiography

Fig. 9-18. On-stream radiography of pipe with corrosion

176

Figure 8-18 shows preparations for on-stream radiography being made. The end piece for
the gamma-source is positioned above the pipe, while the flat film cassette is placed below.
Figure 9-18 shows an on-stream radiograph of a pipe with severe pitting corrosion.
Since the introduction of digital radiography the CR-method, using storage fosforplates, is
rapidly becoming an alternative for traditional film. The main advantage being that it reduces the exposure time by a factor of up to 10, or if weaker sources can be applied a reduced
safety area which is very attractive in cramped spaces e.g. offshore platforms, see chapter 16.

18.7 Flash radiography


Flash radiography or pulse radiography can be carried out when information is required
about the condition of the outer surface of an insulated pipe, without having to remove the
insulation.
Figure 10-18 shows flash radiography in progress. Since only the aluminium cladding and
insulation need to be penetrated, relatively soft radiation is applied, while exposure time is
limited to only a fraction of a second. In that time sufficient pulsed radiation is generated.
to create an image on the superfast F8 + NDT 1200 (film+screen combination) see section
6.3. It is safe to make radiographs manually without the need for additional safety arrangements.

Fig. 10-18. Flash radiography of an insulated pipe section

177

hazards, measuring19 Radiation


and recording instruments
19.1 The effects of radiation on the human body
The human body is constantly exposed to natural radiation (e.g. from space, the soil and
buildings), also known as background radiation. All ionising radiation, whether electromagnetic (gamma-) or corpuscular (particles in the form of alpha- or beta-), and neutrons, are harmful to the human body. The unit absorbed dose (D) defines the effect of
radiation on various substances. D is the absorbed dose in J/kg or Gray (Gy).
The biological damage done by the various types of ionising radiation, , , or fast neutrons, differs and depends on the quality factor (Q). The unit to which the damage quality
factor is applied is the equivalent dose H.
The equivalent dose is the product of absorbed dose (D) and quality factor (Q), so the equivalent dose is calculated as H = D x Q [Sv], (Sv = Sievert).
The Q factors for various types of radiation are indicated in table 1-19.
Type of radiation
X and gamma radiation ()
Beta radiation ()
Alpha radiation ()
Fast neutrons

Quality factor (Q)


1
1
20
10

Table 1-19. Q-factors for various types of radiation

19.2 Responsibilities
The client
It is the clients responsibility to consider possible alternatives before utilising ionising radiation. Considering its purpose, the decision to use ionising radiation can only be justified
when the radiation hazard remains at an acceptable level.

The radiographer
It is primarily the radiographers responsibility to protect himself and others from exposure
to radiation.
On stream image of insulated pipe

CR-radiography

178

179

19.3 The effects of exposure to radiation


The understanding of the effect that exposure to radiation has on human beings has grown
over the past 50 years and has led to a substantial reduction of the maximum permissible dose.
There are two categories of biological effects that an overdose of radiation can cause: somatic and genetic effects. Somatic effects are the physical effects.
A reduction in the number of white blood cells is an example of a somatic effect. Much more
is known about the somatic than about the genetic effects of radiation.
Blood cells are very sensitive and the first signs of radiation are found in the blood, which is
why people working in radiology are subjected to periodic blood tests.
The most serious effects of radiation occur when a large dose is received in a short period of
time. Table 2-19 shows doses received over 24 hours and their effects:
Dose received
by the body
0.0 - 0.25 Sv
0.25 - 0.5 Sv
0.5 - 1.0 Sv
2.0 - 2.5 Sv
5.0 Sv

Effects
No noticeable effects
Limited temporary changes in the blood
Nausea, fatigue
First lethal cases
50 % lethal (MLD = medical lethal dose)

Table 2-19. Effects of radiation doses

The consequences of excess radiation are not necessarily noticeable immediately after the
irradiation. Frequently, they only show up after some time. The time lapse between irradiation and the moment the effects become apparent is called the latent period.
Genetic effects can only be assessed over generations.

19.4 Protection against radiation


The International Commission on Radiation Protection (ICRP), a division of the
International Atomic Energy Agency (IAEA), is engaged in providing rules and regulations
for the protection against radiation, as the name suggests. The ICRP has established the
values for radiological and non-radiological workers, as indicated in section 19.5.

As there is considered to be no totally to be no totally to be no totally safe lower limit below


which no damage would be sustained, the ALARA concept is being promoted. ALARA
(short for As Low As Reasonably Achievable), aims to achieve the lowest possible radiation
dose whereby economic and social factors are considered.
The protection from unwanted external irradiation is based on three principles:
Speed: by working fast, the exposure duration is reduced.
Distance: the greater the distance, the lower the rate of exposure
(remember the inverse square law).
Shielding and collimating: materials with high radiation absorbing properties,
such as lead, reduce the exposure rate to a level that can be calculated in advance.
Table 3-19 in section 19.8 shows the half-value thickness of lead for various gamma sources.

19.5 Permissible cumulative dose of radiation


Although the subject of permissible cumulative dose of radiation is complex, the values
given below apply to external irradiation of the whole body.
The values have been established by the ICRP.
Radiology workers, category A:
20 mSv/year
Radiology workers, category B:
5 mSv/year
Public, not being radiology workers: 1 mSv/year
The whole body level of 20 mSv per year is normally interpolated as 0.4 mSv per week and
10 Sv/h at a 40 hr working week.
These levels are acceptable but it is not to be automatically assumed that people working
with radiation actually should receive these doses.
When radiography is carried out in factories and on construction yards etc. special consideration is required for non-radiological workers and demarcation of the area in which radiation is used, and a maximum dose rate of 10 Sv/h applies, is essential.
This is also the maximum value to be measured at the outside surface of a charged isotope
container.

Practically all countries have brought their national legislation (laws) on ionising radiation
in line with the ICRP codes. The conditions for registration, transport, storage, protection
and the expertise of preparation and use of radiation sources have been laid down in regulations. The purpose of practical protection against radiation is to prevent any individual
receiving a harmful dose.
180

181

19.6 Radiation measurement and recording instruments

Personal protection equipment

From what has been said before, it follows that establishing the presence of ionising radiation and measuring its level is of paramount importance. Since ionising radiation cannot be
detected by the senses, detectors and measuring equipment are used. There are various
instruments with which the radiographer can measure or register radiation.

Pendosismeter (PDM)

The most common measuring instruments are:


1. Dose rate meters
2. Scintillation counters

The PDM consists of a quartz fibre electrometer and a simple optic lens system housed in a
fountain pen type holder, see figure 2-19.
A small charging unit is used to electrically charge the fibre, which can then be viewed
through the lens.
The fibre is set on the zero mark of the calibrated scale as initial setting for the work period.

The most common instruments for personal protection are:


3. Pendosismeter (PDM)
4. Thermoluminescent badge (TLD)
5. Film badge
Fig. 2-19. Pendosis meter

Radiation measuring instruments

Any radiation will cause the charge to leak away through its ionising effect and the fibre will
move across the scale. The amount of radiation received can be read off the calibrated scale.

Dose rate meters


A portable Geiger-Mller counter of 7 x 15 x 4 cm, see
figure 1-19, is the most commonly used instrument for
measuring dose rate, but the more accurate and more
expensive ionisation chamber is used as radiation
monitor as well. Both instruments measure the electric
current that is produced by ionisation.
The radiation level can be read instantly off a microampere meter with a Sv/h or mSv/h calibrated scale.
Some radiation monitors give an audible signal when
a pre-set dose is exceeded.
The instruments are used by personnel working with
radioactive material or X-ray equipment, to determine
the safe distance and the dose rate of for instance
10 Sv/h at the safety barrier. A GM-counter has a
measuring range from 1 Sv/h to 2 mSv/h.

This type of instrument is excellent for personal protection as it is small, inexpensive and
reasonably robust. It can be easily read and records the total amount of radiation received
for the work period with an accuracy of 10 %.

Thermoluminescent dose meter (TLD badge)

Fig.1-19. Geiger-Mller counter

Scintillation counter
This is an accurate and multifunctional instrument to measure and analyse radiation.
The incidence of ionising radiation on a Sodium-iodine crystal is converted into weak light
flashes, which are amplified into electric pulses by an integrated photo-multiplier.
By measuring amplitude and number of these electric pulses, energy and intensity (dose
rate) of the radiation can be determined.
These instruments are predominantly used for scientific purposes.
182

The TLD meter consists of an aluminium plate with circular apertures. Two of these contain
luminescent crystals. Figure 3-19 shows an open TLD-meter and the plate with crystals next
to it. The right side of the
illustration shows the same
meter, now closed. When the
meter is read only one crystal is
used to determine the monthly
dose. The other one is spare
and, if necessary, can be read to
determine the cumulative dose.
The TLD meter is sensitive to Xand gamma radiation of 30 keV
and higher. The dose measuring
range is large and runs from
0.04 mSv to 100 mSv with an
accuracy of 5 %.
The instrument measures
60 x 40 x 10 mm and is conve- Fig. 3-19. Open TLD meter
Closed TLD meter
nient to wear
183

Film dose meter (film badge)

Distance

The film badge consists of two pieces of X-ray film contained, with filters, in a special holder. At the end of a
specified period, the films are developed and the density measured.
The radiation dose received by the wearer can then be
determined by consulting the density/exposure curves,
and the type of radiation received can be established by
checking the densities behind the filters. Film dose Fig. 4-19. Filmdosismeter (film badge)
meters as illustrated in figure 4-19 are a very cheap and reasonably accurate method of
monitoring personnel in selected areas. They measure 25 x 25 x 5 mm, are robust and
convenient to wear.

Since radiation is subjected to the inverse square law, its intensity is reduced with the increase in distance to the square.

19.7 Dose registration

Absorbing barrier and distance


Whenever radiation penetrates a material, the absorption process reduces its intensity.
By placing a high-density material such as lead around the source of radiation, the quantity
of transmitted radiation will decrease. To determine the material thickness required for a
certain reduction in radiation, a factor known as the half-value thickness (HVT) is used.
Table 3-19 shows the HVT-values for lead for various types of gamma sources
Symbol

Due to legally required monitoring and registration of radiation doses received by radiological workers over a specified period of time, dose meters must be worn. Generally, these are
TLD or film badges. The TLD-meter is preferred over the film badge as it is read out electronically and can be linked to a data base. Processing film badges is more complicated. The
films must first be developed before they can be viewed to quantify and register the radiation dose.
Radiation dose monitoring is carried out by a government-authorised organisation which is
responsible for mailing, processing and viewing of the badges. This organisation generates
reports, which contain the individual irradiation doses over a specified period of time, as
well as the accumulated dose.

Cesium137
Cobalt60
Iridium192
Selenium75
Ytterbium169
Thulium170

Average energy in
MeV
0.66
1.25
0.45
0.32
0.2
0.072

Half-value thickness
in mm lead
8.4
13
2.8
2.0
1.0
0.6

Table 3-19. Half-value thicknesses for lead using different types of gamma sources

Example

19.8 Radiation shielding


Protection from radiation (best by distance) can consist of ribbons or ropes and warning
flags to demarcate the area where radiographs are made, or concrete bunkers with doors
which automatically switch off the X-ray equipment as soon as they are opened.
Both methods have the same objective: i.e. to prevent unauthorised people entering the
area of radiation.
An area of radiation can be defined as an area in which the radiation level exceeds the permitted value of 10 Sv/h.
There are three ways to achieve a reduction in intensity:

To reduce 2.56 mSv/h, measured at 1 meter distance, to 10 Sv/h the required distance
according the inverse square law is 2560/10 = 16 metres. To achieve the same by placing
a shield, the number of HVTs is calculated as follows:
Required intensity reduction is 2560 / 10 = 256 x
Number of HVTs is, log 256 / log 2 = 8
The example above demonstrates that an intensity of 2.56 mSv/h can be reduced to
10 Sv/h by increasing the distance to 16 metres, or place shielding material of 8 HVTs as
close as possible to the source. If either of these methods cannot be used on its own, a
combination of the two could be considered.

1. by erecting a demarcation barrier at an appropriate distance,


2. by erecting an absorbing barrier,
3. by a combination of methods 1 and 2.
184

185

literature/references,
20 Standards,
acknowledgements and appendices
European norms (EN-standards)
Ever since the introduction of industrial radiography, there has been a growing need for
standardisation of examination techniques and procedures. At first, these standards had
mainly a national character, e.g. ASTM and ASME, DIN, AFNOR, BS, JIS etc, but as a result
of industrial globalisation the need for international standards grew. The national standards
were, and still are, frequently used internationally, in particular the ASTM and ASME standards.
International standards are largely based on existing national standards. Organisations that
engage in international standardisation are ISO and CEN. These standards are developed
by working groups of experts, who present the newly adapted (harmonised) standards to
the ISO, CEN etc.
A number of European norms (EN) relevant to radiography are listed in table 1-20.
Norm number
EN 444
EN 462-1 through 5
EN 473
EN 584-1
Equivalents:
ASTM E-1815
ISO 11699-1
JIS-K7627
EN 584-2
ISO 11699-2
EN 1435
EN 12543-1 through 5
EN 12544-1 through 3
EN 13068
EN 25580
EN 14784 1 and 2

Subject
General principles for radiographic examination of
metallic materials by X- and gamma rays
Image quality of radiographs IQIs
Qualification and certification of NDT personnel
Classification of film systems

Verification of film systems


Radiographic examination of welded joints
Characteristics of focal spots in industrial
X-ray systems for use in NDT
Measurement and evaluation of the X-ray tube voltage
Fluoroscopic/radioscopic testing
Industrial radiographic illuminators
minimum requirements
Industrial CR with storage phosphor imaging plates
Classification of systems and general principles of application

Table 1-20. European norms for industrial radiography

186

187

Literature and references

Appendices: tables and graphs

1. Industrial Radiology: Theory and Practice (English)


R. Halmshaw. Applied Science Publishers Ltd. London and New Jersey, 1982.
2. Niet-destructief onderzoek. ISBN 90-407-1147-X (Dutch)
W.J.P. Vink. Delftse Universitaire Pers.
3. Die Rntgenprfung, Band 7 ISBN 3-934225-07-8 (German)
The X-ray Inspection Volume 7 ISBN 3-934255-22-1 (English translation)
Both compiled by Dr.Ing. M. Purschke. Castell-Verlag GmbH
4. Handbook of radiographic apparatus and techniques. (English) Publication for the IIW
by the Welding Institute, Abington, Cambridge, England.
5. Radiographic film systems: brochure issued by GE Inspection Technologies.
6.

Home page : www.geinspectiontechnologies.com

Acknowledgements
Figures 9-5 and 4-17, as well as table 2-9 were copied with the publishers consent from
reference book [2] Niet-destructief Onderzoek by W.J.P. Vink, Delftse Universitaire Pers.
Furthermore, Rntgen Technische Dienst bv Rotterdam consented to the use of a number of
their illustrations and graphs.

Designation of quantity
Activity (A)
Ionization dose rate
Ionization dose
Absorbed energy
dose (D)
Equivalent dose (H)
H=D x RBE**

SI units
Name
Unit
Designation
Becquerel
1/s*
(Bq)
Coulomb (C) C/kg
Coulomb (C) C/kg.s
Ampre (A) or A/kg
Gray
J/kg
(Gy)
Sievert
J/kg
(Sv)

Formerly used
Name
Unit
Designation
Curie
Ci
R
R/s

1 R=2.58 x 10-4 C/kg

Rad

Rad

1 Rad = 0.01 Gy

Rem

Rem

1 Rem = 0.01 Sv

kV
100 kV + 8 kV/mm
50 kV + 2 kV/mm
20 kV + 0.2 kV/mm

Table 2-11. Rule-of-thumb values for the selection of X-ray tube voltage.
See chapter 11.

188

189

1 Ci = 37 GBq

Rntgen

Table 1-3. Overview of new and old units.


* disintegrations per second
C = Coulomb = A.s
** RBE = Relative Biological Effect
J = Joule
A = Ampre
See chapter 3

Material
Steel
Aluminium
Plastics

Conversion
Old to SI

Fig.5-11. Nomogram for minimum source-to-film distance fmin


according to EN 1435-criteria.
Catagory A - less critical applications (general techniques)
Catagory B - techniques with high requirements
of detail discernability

distance (b)

distance fmin for catagory A

distance fmin for catagory B

f min = minimum distance


source to source-side object (mm)
s = source size (mm)
b = distance source-side object to film (mm)

Operating range
not being used.

Fig. 4-12. Graph for the minimum number of exposures in accordance with EN 1435 A
at maximum thickness increase of 20 %.
See chapter 12

Fig. 5-11. Nomogram for minimum source-to-film distance Fmin according to EN 1435 criteria.
See chapter 11

190

191

wall thickness

Relative image quality DQE

Faster

True real-time
Radioscopie

DR-Panels
CR-Plates
RCF-Films

C
Better

D7 (coarse grain)
A
D

D-Films

B
E

D2 (fine grain)
diameter in mm

Relative dose
Fig. 13-16. Relative image quality and speed of the various radiographic systems.
See chapter 16

192

Fig. 7-18. Areas of application for selection of source, screen and filter in on-stream radiography.
See chapter 18.

193

You might also like