Department of Transportation: California Test 670 February 2011
Department of Transportation: California Test 670 February 2011
Department of Transportation: California Test 670 February 2011
DEPARTMENT OF TRANSPORTATION
DIVISION OF ENGINEERING SERVICES
Transportation Laboratory
5900 Folsom Boulevard
Sacramento, California 95819 - 4612
SCOPE
This method presents the testing procedures for determining mechanical properties of spliced
reinforcing steel.
B.
REFERENCES
APPARATUS
1. A tensile test machine able to apply a tensile force greater than the ultimate tensile
strength of the sample. This machine must be accurate and calibrated in
accordance with ASTM A 370.
2. A slip-measuring device consisting of two calibrated dial indicators that measure
displacement across the splice to the nearest 0.001 inch. A typical test set-up is
shown in Figure 1a. The dial indicators may have an analog dial or be digital.
Alternatively, extensometers accurate to within 0.001 inch can be used.
3. A calibrated caliper accurate to 0.001 inch.
D.
DEFINITIONS
E.
Slip Test
The slip test is required for all mechanical splices except mechanical lap splices or splices that
are used on hoops. There are two acceptable options for measuring slip.
a. Slip Test (Option I)
Option I requires the use of two dial indicators to measure displacement across the
splice. Testing is performed as follows:
(1) Mount the sample in the tensile test machine.
(2) Attach the slip-measuring device so that the dial indicators are 180 apart as shown
in Figure 1a. During this step, the vertical bars shall be securely attached with the
hex socket bolts shown in Figure 1b.
(3) Once the measuring device is securely attached, release any restraining bars to
allow free movement of the device. For the testing device shown in 1a, this is done
by removing the hex socket bolts from the bottom end of the vertical bars as shown
in Figures 2a and 2b.
(4) Pre-stress the sample to 3,000 psi and then zero out the dial indicators.
(5) Apply an axial stress of 30,000 psi. Maintain this stress until a steady reading is
obtained on both dial indicators.
(6) Reduce the stress to 3,000 psi and record the two dial indicator readings. Take an
average of the two readings. This is the total slip. Record the total slip on the Test
Form (Figure 3).
(7) Remove the slip-measuring device.
-2-
-3-
Leave a gap of at least two bar diameters between the weld splice and the
8-inch gage marks (see figure below), and proceed to step 5.
2d
2d
2. For hoops, first determine the outside diameter of the hoop from
submittal documents. Record this information on the Test Report Form
(Figure 3). Next calculate the bar center to bar center hoop diameter (D)
as follows:
D = Dod bar nominal diameter
where:
Dod = hoop outside diameter
3. Place punch marks along the sides of the samples (not along the concave
or convex sides of the bar) to create a total of two 8-inch gage lengths,
one on each side of the weld. Label these gage lengths as A and B. Leave
a gap of at least two bar diameters between the weld splice and the
8-inch gage marks. See figure below:
2d
2d
8
4. Straighten the ends of the hoop samples to fit the sample into the testing
grips. This straightening should be outside of the gage marks, in
accordance with the Concrete Reinforcing Steel Institutes Manual of
Standard Practice MSP-1-90.
5. Mark each one of the 8-inch nominal gage lengths sequentially as A and B.
6. Measure and record the initial (Lo) of the two 8-inch gage lengths A and B
to the nearest 0.001 inch.
7. Tensile test the sample following the procedure in Section E.3.
8. Measure and record, to the nearest 0.001 inch, the final gage length (Lf)
not encompassing the location of the rupture.
9. For straight samples go to step 10 to calculate the percent strain. For
hoop samples, calculate the corrected final gage length as follows:
Corrected final gage length (Ln) = measured gage length
correction factor
Lf
Lf
D ArcSin
D
-4-
correction factor
where:
Lf = measured final gage length in inches
D = diameter of the hoop (rebar center to rebar center in inches)
ArcSin should be expressed in radians.
10. Calculate the percent strain as follows:
For straight samples:
% strain = [(Lf Lo)/Lo]
100
100
11. Record the calculated strain on the Test Report Form (Figure 3).
b. Mechanical Splices:
(1) Place punch marks as shown in the figure below:
Grip
2d
8
2d
d = bar diameter
Lc
2d
8
2d
Grip
(2) Mark each one of the 8-inch nominal gage lengths sequentially as A and
B.
(3) Individually, measure and record the initial (Lo) 8 inch gage lengths A
and B, to the nearest 0.001 inch.
(4) Tensile test each sample following the procedure in Section E.3.
(5) Measure and record the final gage length (Lf) not encompassing the
location of rupture, to the nearest 0.001 inch.
(6) Calculate the percent strain as follows:
% strain = [(Lf Lo)/Lo]
100
(7) Record the measured strain on the Test Report Form (Figure 3).
-5-
F.
1. Cyclical Testing
a. Cyclically load the sample from 5% to 90% of the specified yield strength (y) of the
sample for 100 cycles. Use a haversine waveform at 0.5 cps for #10, #11, #14, and #18
bars, and a haversine waveform at 0.7 cps for smaller bars. Record whether or not the
sample fractures.
b. If sample does not fracture during the cyclical test, increase the axial tensile load until
the sample fractures.
c. If applicable, record the ultimate load, ultimate tensile strength, location of failure, and
any necking on the Test Report Form (Figure 3).
2. Fatigue Testing
a. Fatigue load the sample from + 25,000 psi to 25,000 psi for 10,000 cycles. Use a sine
waveform at 0.083 cps for #10, #11, #14, and #18 bars, and a sine waveform at
0.35 cps for smaller bars. Record whether or not the sample fractures.
b. If the sample does not fracture during the fatigue test, increase axial tensile load until
the sample fractures.
c. If applicable, record the ultimate load, ultimate tensile strength, location of failure, and
any necking on the Test Report Form (Figure 3).
G.
REPORT
The test report form shown in Figure 3 is the recommended form to report test results.
Modification of this form is allowed as long as all the information listed in this figure is
reported in the modified form.
H.
HAZARDS
The test samples are heavy and may contain sharp edges or burrs. Sample rupture may
involve brittle fractures and ejection of sample fragments. Use appropriate safety measures.
I.
It is the responsibility of the user of this test method to establish appropriate safety and health
practices and determine the applicability of regulatory limitations prior to use. Prior to
handling, testing or disposing of any materials, testers must be knowledgeable about safe
laboratory practices, hazards and exposure, chemical procurement and storage, and personal
protective apparel and equipment.
Caltrans Laboratory Safety Manual is available at:
http://www.dot.ca.gov/hq/esc/ctms/pdf/lab_safety_manual.pdf
End of Text
California Test 670 has 11 pages
-6-
Figure 1a. Measuring device with dial indicators used for measuring slip.
Figure 1b. Close up of hex socket bolts used to hold the vertical
bars during the set up of the measuring device.
-7-
Figure 2a. Close up of the removal of the hex socket bolts before testing.
Figure 2b. Measuring device ready for testing with bottom hex socket bolts removed.
-8-
Date Received:
Material
TL-101 No.
Lot No.
Sample No.:
Contract No.
Bar Size:
Manufacturer:
Splice Type:
Service or Ultimate?
Date Tested:
Lab Technician:
Testing Machine
Sampler/Inspector:
Test Results
Reported To:
Sample No. 1
Sample No. 2
-9-
Sample No. 3
Sample No. 4
Sample No. 2
Sample No. 3
Sample No. 4
n/a
n/a
n/a
n/a
n/a
Side B:
Initial gage length (Side B)
Final gage length (Side B)
Strain (Side B)
B. Hoops:
Outside hoop diameter
Bar nominal diameter
Hoop diameter (D) - (bar center to bar
center)
Side A:
Initial gage length (Lo), Side A:
Final gage length (Lf), Side A:
Correction factor
Corrected final gage length (Ln), Side A
Strain (Side A)
Side B:
Initial gage length (Lo), Side B:
Final gage length (Lf), Side B:
Correction factor
Corrected final gage length (Ln), Side B
Strain (Side B)
CYCLICAL TESTING:
Specimen fractured?
FATIGUE TESTING:
Specimen fractured?
n/a
[ ] Samples pass.
[ ] Samples fail, because
- 10 -
Figure 4. Digital calipers used to measure initial gage length to the nearest 0.001.
- 11 -