Colposcope Model 955 Led: Owner'S Manual
Colposcope Model 955 Led: Owner'S Manual
Colposcope Model 955 Led: Owner'S Manual
PRECISION MICROSCOPES
PRECISION MICROSCOPES
We are here to serve you!
If you have any questions regarding Seilers products or services,
please feel free to contact us.
Toll-Free: 800.489.2282
Local: 314.968.2282
Fax: 314.968.3601
E-mail: [email protected]
www.seilerinst.com
TABLE OF CONTENTS
INDICATIONS FOR USE
TECHNICAL DATA
BINOCULARS
OBJECTIVE
COLPOSCOPE ASSEMBLY
COLPOSCOPE HANDLING
HOW TO FOCUS THE COLPOSCOPE
OPTICAL AND FIELD MAGNIFICATION TABLE
CLEANING
TROUBLESHOOTING TIPS
LED LIGHTSOURCE
OWNERS RECORD
NOTES
4
4
5
5
6
8
8
9
9
9
10
16
16
TECHNICAL DATA
A - Colposcope Head
Microscope Galilean System
Binoculars
F= 170 MM (optional: Inclined, F=135 mm)
Objective (focus)
300 mm (optional: 200 400 mm)
Eyepieces
Wide angle type 12,5x, adjustable
(-6 to +6 dioptries)
Magnification See table
Observation field
See table
Illumination field
See table
Interpupillary distance
Adjustable between 42 mm and 75 mm
Illumination System:
Source LED Illumination System
Filter Green filter
Focusing:
Macro Moving the pantographic arm
Micro
Manual, with rack and pinion
Total weight of the head
3.0 Kg
B- Colposcope stand
Horizontal reach
Angular rotation (arms)
Distance between the floor
and the objective
Load adjustment
Base
Total floor stand set weight
725 mm. - first arm 600 mm, second arm 110 mm.
340
635 mm to 1295 mm
4 to 7 kg
Base with five casters, two with brakes
34 kg
BINOCULARS
1
2
3
1. Binoculars body
2. Eyepieces with diopter adjustment
3. Eye Protectors
OBJECTIVE
The Colposcopes are factory shipped with the following objective:
F = 300 mm
The following optional objectives are available:
F = 200 mm
F = 250 mm
F = 300 MM
F = 400 mm
COLPOSCOPE ASSEMBLY
Step 3
Place Pantographic arm into
Lighthouse.
Step 2
Step 1
Step 4
Install Light Guide into bottom
cover of arm.
Step 5
Plunger
t
Step 6
Install Optic Pod into the
extension arm.
Step 7
Install Binocular Head onto the
Optic Pod and tighten thumb screw.
Remove these parts from the packing case, with maximum care:
a) Optical Head Support b) Binoculars
Introduce the spindle into the terminal. Screw the safety knob and lock the
spindle with the knob.
At first release the screw. Place the binoculars on the flange of the body
making sure the pin fits into the slit of the binoculars flange. Tighten the fixing screw and make sure that the binoculars
are firmly secured to the Colposcope.
The correct position of the Binoculars to the Colposcope
Align the guide-pin to the groove, adjust the base of the Binoculars to point the flange, and t ighten the set screw.
Finally check if the Binoculars are well secured and do not move.
The Binoculars Eyepieces
The eyepieces are mounted to the binocular housings. Make sure that they are well seated.
The Eye Protectors
The Eye Protectors are threaded to the Binoculars. The Eye Protectors protect the eyes, prevent the intake of lateral
lights and create a suitable darkroom to the observer.
Precautions
The Binoculars contain optical parts internally that cannot be cleaned without disassembly.
It is therefore advisable to keep the eyepieces inserted in the eye tubes at all times in order to keep impurities from
getting inside the housing. It is recommended that if the eyepieces are removed, a clean cloth be placed over the
openings.
COLPOSCOPE HANDLING
a- MAGNIFICATION SELECTOR
The magnification selector has 6 positions, engraved from 1 to 5. Note that the position 3 repeats. The position being
used is the one aligned to the black dot, engraved on the body. The magnification factors are found in the table
appended.
b- PANTOGRAPHIC ARM BALANCING
The Colposcope Model 955 is factory adjusted for balancing, using the regular optical head, and may only require
adjustment at the tension knob when accessories are added.
Eyepieces
Binocular Head
Thumb Screw
Magnification Turret
Tension Adjustment Knob
Objective Lens
Handle
Fine Focus Knob
Posi(on
of
BuNon
1(0.4)
2(0.6)
3(1.0)
10X/18
4(1.6)
5(2.5)
1(0.4)
2(0.6)
12.5X/16 3(1.0)
4(1.6)
5(2.5)
1(0.4)
2(0.6)
3(1.0)
16X/16
4(1.6)
5(2.5)
Illuminated
Field
Objec(ve
Lenses
175
200
250
300
400
Magnica(on
Field
Magnica(on
Field
Magnica(on
Field
Magnica(on
Field
Magnica(on
Field
3.89 46.32
3.40
52.94
2.72
66.18
2.27
79.41
1.70
105.88
5.83 30.88
5.10
35.29
4.08
44.12
3.4
52.94
2.55
70.59
9.71 18.53
8.50
21.18
6.80
26.47
5.67
31.76
4.25
42.35
15.54 11.58
13.60
13.24
10.88
16.54
9.07
19.85
6.80
26.47
24.29
7.41
21.25
8.47
17.00
10.59
14.17
12.71
10.63
16.94
4.86 41.18
4.25
47.06
3.40
58.82
2.83
70.59
2.13
94.12
7.29 27.45
6.38
31.37
5.10
39.22
4.25
47.06
3.19
62.75
12.14 16.47
10.63
18.82
8.50
23.53
7.08
28.24
5.31
37.65
19.43 10.29
17.00
11.76
13.60
14.71
11.33
17.65
8.50
23.53
30.36
6.59
26.56
7.53
21.25
9.41
17.71
11.29
13.28
15.06
6.22 41.18
5.44
47.06
4.35
58.82
3.63
70.59
2.72
94.12
9.33 27.45
8.16
31.37
6.53
39.22
5.44
47.06
4.08
62.75
15.54 16.47
13.60
18.82
10.88
23.53
9.07
28.24
6.80
37.65
24.87 10.29
21.76
11.76
17.41
14.71
14.51
17.65
10.88
23.53
38.86
6.59
34.00
7.53
27.20
9.41
22.67
11.29
17.00
15.06
65
72
90
108
144
CLEANING
The Objective, pointed to the field under clinical or surgical treatment, is exposed to blood and other bodily fluids. Spots
on the objective dim the passage of the light. Stains blur the light, reducing the luminosity of the optical observation.
Cotton padding wetted in alcohol rubbed softly, in CIRCULAR MOVEMENTS over the optical surface, is enough to clean
the stains, except those of clotted blood. Clotted blood shall be cleaned with cotton and hydrogen peroxide.
If the eyepieces are too dirty, change the cotton at each circular friction motion to avoid spreading out the dirt.
On the other hand, there is an effective protector for the objective, fitted under pressure at the outer edge of the lens
which protects the objective against splatter and contamination.
To have the eyepieces and objective well cleaned for clinical use, it is sufficient to use cotton.
The metallic parts (chrome-plated or painted) can be cleaned with cotton and alcohol.
TROUBLESHOOTING TIPS
ELECTRIC CONNECTIONS
Ensure that none of the following are present:
Broken wires inside the cable.
Loose wires at the connector.
Broken wires at the fixation, inside the connector.
Oxidized connectors terminals.
LED LIGHTSOURCE
1. INTRODUCTION
Congratulations on the purchase of your new SSL-2000 Light Source! This user-friendly LED light source is a high
efficiency light source utilizing state of-the-art superior illumination technology. It offers a variety of features such as:
6500 K daylight brightness for good color definition
Quiet operation
Long life, 50,000 hrs.
Compact and light weight
Pulse-width Modulation (PWM) Electrical Dimming
This section of the manual will help you to install the device and optimally integrate it with other components of
your system. It will also instruct you how to operate the LED Light Source and how to keep it clean. It will give you
maintenance and service guidelines as well as recommendations for best performance results.
1.1 INDICATIONS FOR USE: The LED Light Source is used to illuminate the site of surgery during minimally invasive
surgical procedures. The light is transmitted from the light source through a fiber optic cable and a scope.
2. WARNINGS
To prevent fire or electric shock, do not open or expose the LED light source unit to rain or moisture. Refer all servicing to
qualified personnel only.
Not suitable for use in presence of flammable anesthetic mixture with air or with oxygen or nitrate oxide.
To prevent any potential electro-magnetic interference, do not use any kind of cellular phone near the
light source.
This product should be used only with type BF endoscopic instruments which have been certified
according to IEC 601-1 for medical equipment and IEC 601-2-18 for endoscopic equipment.
Caution: This product is not provided as sterile.
Caution: All devices connecting to the LED Light Source must be classified as medical equipment. When
additional information processing equipment is connected to the LED Light Source, the operator must
determine that all equipment complies with the appropriate end-product standards (such as IEC 60950
or IEC 60065 and the Standard for Medical System, IEC 60601-1-1).
Caution: The LED Light source can cause permanent eye damage if viewed directly with unprotected eye.
To reduce the chance of eye damage, set the intensity control always to the minimum level and plug the
fiber optic cable into the unit before turning on the power.
EQUIPMENT CONNECTION
Caution: The fiber optic cable must be a NON-CONDUCTIVE CABLE. It should not have conductive
shielding or any other conductive connection between the patient and equipment. Such connection will
impair safety of the equipment. It must be rinsed free of soaking/disinfectant solution and dried before
plugging into the LED light source receptacle. Ensure the optical surface is clean before engaging into the
light source.
10
3. SPECIFICATIONS
Item
Light Source Type
Power
Color Temperature
Specification
LED (Light Emitting Diode)
50 Watt
Led life
Light guide adapter
Brightness control
Input voltage
Rated Power
Regulatory Approvals
6500 K
Equipment Class
Mode of Operation
Water Resistant
Operating Environment
Temperature
Relative Humidity
Air Pressure
Storage Environment
Temperature
Relative Humidity
Air Pressure
Dimensions
Weight
No.
1
2
3
Name
Power switch
Light Guide Adapter
Intensity Control
Function
Turns the light source on and off. Fan will operate
Electronically controls the light output
11
2
3
1
Figure 2. LED Light Source Rear Panel
No.
1
2
3
4
Name
AC main inlet
Fuse
Grounding Connector
Venting grids
Function
Accepts AC power cord
2A, 250V (1/4 x 1 ) time delay
For potential equalization
Allows adequate ventilation and cooling of the light source
12
5. INSTALLATION
5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE: Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable gases.
DO NOT block the venting grids of the LED light source.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
CAUTION: Use only cords provided with the light source.
Plug the AC power cord into a wall outlet using the three-prong plug supplied with the unit.
CAUTION: To prevent electric shock, connect power cords of peripheral equipment through medical isolation
transformers.
NOTE: When using medical isolation transformer, be sure to check the transformer power ratings.
Make sure that the power cord is connected to the main power with three-prong plug (USA use UL2601-1 rated
isolation transformers and/or power strips only).
CAUTION: When used in clinical or residential areas near radio or TV units, this equipment may be subjected to radio
interference. To avoid adverse electromagnetic effects, DO NOT operate this equipment near RF energy equipment.
5.2 CONNECTING THE FIBER OPTIC CABLE
Turn intensity control to the minimum setting. Connect the light cable to the instrument and then plug the light cable
endfitting into the port on the front panel.
6. OPERATION
6.1 POWERING UP THE LED LIGHT SOURCE
To operate the LED light source: Turn on the power switch. The indicator on power switch will light.
6.2 LIGHT BRIGHTNESS DISPLAY AND CONTROL
The light brightness adjustment enables the user to obtain a brighter or darker illumination of the object of observation.
Adjust the light intensity by turning the intensity control knob.
7. CLEANING
NOTE: Always disconnect the power cord before cleaning the system.
The unit can be cleaned with any cleaning agent, which is used for external cleaning of electric equipment, according to
instructions given by the manufacturer of the cleaning solution.
Do not allow excessive moisture or liquids to come in direct contact with the unit.
Do not use cleaning agents that are not permitted for use with plastics, i.e., ammonia, acetone, salty acids (HCl), etc.
Do not allow cleaning agents or liquids to enter the unit outlets.
8. DISINFECTION
8.1 DISINFECTING THE UNIT
NOTE: Always disconnect the power cord before cleaning the system.
Use any disinfectant agents which are commonly applied while disinfecting surfaces of electric medical equipment. Such
13
Problem
The power indicator (refer to 4.1)
is not lit.
The power indicator is lit, but the
lamp will not ignite.
14
Solution
A. Check that the AC power cord is properly connected.
B. Check the unit fuses. If necessary, replace.
Turn intensity control know clockwise to increase light output intensity.
OWNERS RECORD
Model Number
Serial Number
Date Purchased
Location Purchased
NOTES
15
PRECISION MICROSCOPES
3433 Tree Court Industrial Blvd.
St. Louis, Missouri 63122
Toll Free: (800) 489-2282
Local: (314) 968-2282
Email: [email protected]
Web: www.seilerinst.com