Series 5400 Humidity/Temperature Operating Systems: °F (0-55°C) - Use With ETS or Other Controllers and
Series 5400 Humidity/Temperature Operating Systems: °F (0-55°C) - Use With ETS or Other Controllers and
OPERATING SYSTEMS
Series 5400
Humidity
1 Electric Heaters
1 Use with new chambers or
retrofit existing chambers
1 Custom systems available
Temperature
Description:
All ETS Operating Systems require 115 or 230 VAC, 50/60 Hz power. Voltage requirements
must be specified at time of order.
Dehumidifiers: Available in closed and open loop
configurations. Closed loop systems use
desiccant/pump technology with a pump drawing
air from the enclosure, passing it through a drying
column containing indicating calcium sulfate
(CaSO4) then returning dry air to the chamber.
The process is repeated until the desired humidity
level is achieved.
Model 5461
Model 5471
Model 5478
Model 5462
Model 5472
Gas cooling is not recommended for high humidity applications since the cold gas will dry the
environment and may overwhelm the capacity of the humidifier.
Heaters: Electric heaters are supplied as single units or cascaded to achieve heating
capacities from 125 to 1000 Watts. Integrated heating systems include circulating fans,
overload protection, power on indication plus auxiliary weather-tight North American AC
outlets housed in aluminum or stainless steel enclosures.
Model 5474
Model 5474L
Specifications:
Dehumidifiers:
Model 5461: Closed loop, Desiccant/Pump
Drying medium: Renewable, indicating CaSO4
Capacity: 1 lb (.45kg)
Outlet Dew Pt.: -100F @ 100F Sat. air in
Pump/ Output flow rate: Reciprcating/0.21 SCFM
Dimensions: Approx. 12x12x12 (30x30x30cm)
Weight: 3.25 lbs (1.5kg)
Humidifiers:
Model 5462: Open Loop, Single-Head Ultrasonic
Air source: Fan, 1.0 (25mm) ID tube
Water: Distilled or deionized
Capacity: 0.5 Gal (1.9 l)
Max RH: 100% @ 72F (22C)
Dimensions: 11Wx6Dx9H (28x15x23cm)
Weight: 3 lbs (1.4kg)
Output: < 0.5 cfm
Coolers:
Model 5473 Thermoelectric
Model 5475 Refrigeration
Cooling capacity: 100, 150, 200, 300 & 400 Watts
Cooling capacity: 700 Watts
Cooling requirements: Refer to Table below
Cooling requirements: Refer to Table below
Power Req.:6A@115/3.5A@230VAC per 200W cooling Power Req.: 8A
Model 5463 Liquid CO2
Model 5466 Liquid Nitrogen
Source: Liquid CO2 tank with dip tube (user supplied)
Source: Liquid Ni Tank, 50psi min (user supplied)
Cooling: To 25F (-4C) in 13 ft3 acrylic chamber
Cooling: To 25F (-4C) in 13 ft3 acrylic chamber
Control: Solenoid valve with pressure regulator
Control: Cryogenic hose & solenoid valve
Cooling requirement Table for non-insulated acrylic chambers
Area (in2)
T - C
5
10
15
20
2000
60
125
200
275
3000
80
175
275
400
4000
100
225
375
500
5000
125
275
450
650
6000
150
325
525
750
7000
175
375
625
900
8000
200
425
700
1000
Heaters:
Model 5474
Capacity: 2x250 Watt Electric strip
Housing: Stainless steel, 16x13x3.125 (41x33x8cm)
Fans: 1x110 cfm
Thermal Protection: 135F (55C)
Aux AC Outlet: Dual North American GFI
Model 5474-L
Capacity: 2x250 Watt Electric strip
Housing: Aluminum, 24x14x4 (60x36x10cm)
Fans: 2x110 cfm
Thermal Protection: 135F (55C)
Aux AC Outlet: Dual North American GFI
12/07