GC Specs
GC Specs
GC Specs
UNSPSC Code:
ECRI Code:
1 Description of Function
Sl
Name
1.1
The instrument is used to detect the presence of organic compounds like Volatiles, aldehydes, free nicotine, PAHs, flavours etc.
and. to measure their concentrations.
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
2 Operational Requirements
Sl
Name
2.1
Microprocessor controlled Gas Chromatograph with Electronic Pneumatic Controls, appropriate detectors, Injection Ports,
Columns, Accessories and spares along with Data handling system is required
2.2
Inlets: Should be used in gas saver mode to reduce the gas consumption.
2.3
Chromatography Data processing system which is based on Microsoft Windows operating system for instrument control, data
acquisition, data analysis, quantitation, automation & customization with online and offline sessions provided System should have
capability of locking the retention time so that same Retention time can be reproduced from system to system and the method
should be electronically transferred.
2.4
A microprocessor based system should be supplied with mass spectroscopic quadrupole detector, columns, injectors, gas purifiers
etc for the selected applications
3 Technical Specifications
Sl
Name
3.1
GENERAL
FEATURES
Technical
Specs
quoted by
bidder
Vacuum Pumps: (nitrogen) air- cooled turbomolecular pump with minimum pumping of speed 200 liter per second
with 2.5 to 2.9 m3/hr (50-60 Hz) rotary vanes.
Bidders
Deviation
if any
SENSITIVITY
Test
EI Full scan
Positive CI full scan
Negative CI full scan
PERFORMANCE
o
o
o
o
o
Amount
1pg of OctaFluoroNaphtalene(OFN)
100 pg of Benophenone
1 pg OFN
Detection Limit(S/N)
180:1 RMS
300:1 RMS
450:1 RMS
Possible to program 20 temp ramps / 21 plateaus with ramp upto 30C per min in 0.1C steps
INLETS:
Electronic Pneumatic Contol
Programmable Temperature Vaporizer
Split /Splitless capillary (S/SL):
Electronic Pressure / flow control
Pressure setting range:0-100 psi
Total flow setting range: 0 to 200ml/min. for N2 0 to 100 ml & for H2 or He 0 to 100 ml.
Autosampler
Injection speed
Program modes
Sample positions
Vial size
Waste and wash vials
Waste and wash vial size
Syringe size
Sampling volume
Viscosity settings
Maximum injections per
vial
Maximum solvent
postwashes
Maximum sample pumps
Maximum sample
prewashes
Minimum sample
Sample pre-rinse
Upgradable with head space sampler &/or purge & trap system in future
4 System Configuration Accessories, spares and consumables
Sl
Name
4.1
ACCESSORIES:
1.Micro Syringes 5 l, 10 l and 25 l (3 each)
2.GAS CYLINDERS: The equipment should be provided with oxygen, nitrogen, hydrogen and air cylinders along with regulators,
connectivity copper pipes and filters through local supplies.
3. Flow meter for normal gas flow measurement.
Column
60 m X 0.32 mm X 1 m DB-5MS column (or equivalent) with 1m X 0.25 mm deactivated fused silica transfer line
4.2
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
5 Environmental factors
Sl
Name
5.
1. The unit shall be capable of being stored continuously in ambient temperature in the range of 0-50 deg C and relative humidity
in the range of 15-90%.
2. The unit shall be capable of operating continuously in ambient temperature in the range of 10- 42 deg C and relative humidity
in the range of 15-90%
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
6 Power Supply
Sl
Name
6.1
1. Power input to be 220-240VAC, 50Hz fitted with BIS plug (ISI mark).
2. UPS of suitable rating with voltage regulation ,spike protection and 60 minutes back up.
Sl
Name
7.
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
Technical
Specs quoted
by bidder
Bidders
Deviation if
any
8 Documentation
Sl
Name
8.1