C0110 00
C0110 00
C0110 00
1. Scope
1.1 These test methods cover physical testing of quicklime
and hydrated lime, and of limestone not otherwise covered in
ASTM standards.2
NOTE 1Quicklime and hydrated lime have a high affinity for moisture
and carbon dioxide. Caution should be taken to protect both hydrated and
quicklime during sampling, storage, and testing (see Practice C 50).
13
16
17
8
15
19
18
20
14
7
9
5
11
12
21
6
10
22
Copyright ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
C 110
E 11 Specification for Wire and Cloth Sieves for Testing
Purposes6
E 29 Practice for Using Significant Digits in Test Data to
Determine Conformance with Specifications6
E 691 Practice for Conducting an Interlaboratory Study to
Determine the Precision of a Test Method6
3. Terminology
3.1 DefinitionsUnless otherwise specified, for definitions
of terms used in this standard see Terminology C 51.
4. General Procedures
4.1 SamplingSamples of lime and limestone for chemical
analysis shall be taken and prepared in accordance with the
requirements of Practice C 50 applicable to the material to be
tested.
4.2 Calculation:
4.2.1 The calculations included in the individual procedures
sometimes assume that the exact weight specified has been
used. Accurately weighed samples which are approximately
but not exactly equal to the weight specified may be used
provided appropriate corrections are made in the calculation.
Unless otherwise stated, weights of all samples and residues
should be recorded to the nearest 0.0001 g.
4.2.2 In all mathematical operations on a set of observed
values, the equivalent of two more places of figures than in the
single observed values shall be retained. For example, if
observed values are read or determined to the nearest 0.1 mg,
carry numbers to the nearest 0.001 mg in calculation.
4.3 Rounding FiguresRounding of figures to the nearest
significant place required in the report should be done after the
calculations are completed, in order to keep the final results
free from calculation errors. The rounding procedure should
follow the principle outlined in Practice E 29.
5. Residue and Sieve Analysis
5.1 Significance and Use:
5.1.1 This test method determines the residue obtained from
slaking quicklime. Residue, in this case, is largely unreacted
material such as uncalcined limestone or dolomite, overburned
quicklime, or gross impurities, or a combination of these.
5.2 Apparatus:
5.2.1 The sieves used shall conform to the requirements of
Specification E 11. Preferably the sieves should have a 4-in.
depth.
5.2.2 If sieve calibrations are required, follow the method as
outlined in Test Method C 430.
5.2.3 Spray Nozzle conforming to the requirements of Test
Method C 430.7
5.2.4 Pressure Gage shall be 3-in. (75-mm) minimum
diameter, and shall be graduated in 1-psi (6.9 kPa) increments,
and shall have a maximum capacity of 30-psi (207 kPa). The
accuracy at 10 psi (69 kPa) shall be 60.25 psi (61.7 kPa).7
C 110
6.2.3 Base PlateThe base plate for supporting the ring
mold shall be of plate glass and about 100 mm square.
6.2.4 Mechanical Mixers .8
6.3 Standard Consistency Determination:
6.3.1 Mechanical Mixing Procedure Using the Vac-UMixerTo a measured amount of water contained in an
800-cm3 Vac-U-Mix bowl, add 300 g of hydrated lime and
hand mix for 10 s with a stiff spatula (Note 2). Cover putty to
prevent evaporation of water. After the applicable soaking
period, 30 min maximum for Type S, special hydrated lime,
and not less than 16 h nor more than 24 h for Type N, normal
hydrated lime, insert the paddle assembly and mix the putty for
30 s with the mechanical mixer. Remove the paddle assembly
and scrape down any putty adhering to it and to the sides of the
mixing bowl. Remix for 30 s and determine the consistency as
prescribed in 6.3. If the penetration is less than 15 mm, return
all of the material to the mixer bowl, add additional water, and
remix for 15 s. If the penetration is greater than 25 mm, repeat
the test.
consecutive sieves, depending upon the form of the specification for the use of the material under test.
5.6 Precision and Bias:
5.6.1 No precision data are available due to the limited use
of these test methods. Therefore, users are advised to develop
their own laboratory precision. No statement is being made
about the bias of these test methods.
6. Standard Consistency of Lime Putty
6.1 Significance and Use:
6.1.1 In order to measure certain physical properties of a
lime putty, such as plasticity, it is necessary to have a uniform
or standard consistency (viscosity), since the property measurement is affected by the consistency level.
6.2 Apparatus:
6.2.1 Modified Vicat ApparatusThe apparatus, constructed as shown in Fig. 1, shall consist of a bracket, A,
bearing a movable brass rod, B, 6.3 mm in diameter and of
suitable length to fit the Vicat bracket. A plunger, C, 12.5 mm
in diameter, made of aluminum tubing, shall be attached to the
lower end of the rod. The total weight of the rod with plunger
shall be 30 g. The lower end of the plunger shall be closed
without shoulders or curvature and the tube may be loaded with
shot to the specified weight. The total weight required may also
be obtained by means of a weight, D, screwed into the rod. The
rod can be held in any position by means of a screw, E, and has
a mark midway between the ends which moves under a scale,
F, graduated in millimetres, attached to the bracket, A.
6.2.2 MoldThe conical ring mold shall be made of a
noncorroding, nonabsorbent material, and shall have an inside
diameter of 70 mm at the base and 60 mm at the top, and a
height of 40 mm.
6.3.3 Consistency DeterminationTo determine consistency, place the mold with its larger end resting on the glass
base plate and fill with the lime putty. Then strike off the putty
flush with the top of the mold. Center the lime putty, confined
in the ring mold resting on the plate, under the rod of the
modified Vicat apparatus (Fig. 1). Bring the plunger end, C, in
contact with the surface of the lime putty and take an initial
reading. Release the rod and take the final reading 30 s after the
plunger is released. The lime putty is of standard consistency
when a penetration of 20 6 5 mm is obtained. Record both the
total amount of water required to bring the putty to standard
consistency and the actual penetration. Proceed with the
plasticity determination in accordance with 7.3.
6.4 Precision and Bias:
6.4.1 The precision and bias of this test method has not been
determined.
8
A Vac-U-Mixer or an N-50 Hobart Mixer, or equivalent, has been found
suitable for this purpose.
C 110
7.2.3 Absorption of Plasticimeter Base Plates:
7.2.3.1 Total AbsorptionPlasticimeter base plates when
immersed in water at room temperature for a period of 24 h
shall absorb not less than 40 g of water. Before making the
determination, dry the porcelain plates overnight in an oven at
temperatures of between 100 and 110C (212 and 230F) and
permit to cool to room temperature. Dry the plaster plates
overnight over calcium chloride at room temperature. After
immersion and before weighing, wipe off the excess water with
a damp cloth.
7.2.3.2 Rate of Absorption (Note 4)When tested over an
area 70 mm (234 in.) in diameter, the water absorbed shall be
in accordance with the following:
Time, min
Water Absorbed, mL
1
2
3
4
5
8 to 14
5 to 712
4 to 612
4 to 6
312 to 512
P 5 =F2 1 ~10T!2
(1)
where
P 5 plasticity figure,
F 5 scale reading at the end of the test, and
T 5 time in minutes from the time when the first portion of
paste was put in the mold to the end of the test.
9
Emley Plasticimeters are no longer manufactured. If test is required, contact the
chairman of Subcommittee C07.06 through ASTM Headquarters, 100 Barr Harbor
Drive, West Conshohocken, PA 19428.
C 110
7.5 Precision and Bias:
7.5.1 There are as yet insufficient analyzed data to permit
preparation of a precision and bias statement for this test
method. When data are collected and analyzed, precision and
bias statements will be proposed.
7g
16 g
14 g
8g
72 g
C 110
10.3 Consistency:
10.3.1 ApparatusThe flow table and mold used for the
measurement of consistency of the mortar shall conform to
Specification C 230.
10.3.2 ProcedureCarefully wipe dry the flow table top
and place the flow mold at the center. Immediately after
completing the mixing operation, fill the mold with mortar
gently pressed into place by the finger tips to ensure uniform
filling free of voids. Smooth off the mortar level with the top of
the mold by aid of a trowel, and remove the mold. Immediately
drop the table through a height of 13 mm (12 in.), 25 times in
15 s. The flow is the resulting increase in diameter of the
mortar mass, expressed as the percentage of the original
diameter. The mortar may be adjusted, if the flow is below
100 %, by additions of water until the flow is within the range
from 100 to 115 %. Make each adjustment by returning the
mortar to the original mixing bowl, add water, and then mix at
medium speed (2856 10 r/min) for 30 s. If the flow of the
original mortar is greater than 115 %, prepare a new batch.
10.4 Water Retention Test:
10.4.1 ApparatusThe equipment used to determine water
retention shall conform to the same apparatus used for the
water retention test in Specification C 91 (see Fig. 3).
10.4.2 Procedure:
10.4.2.1 Adjust the mercury relief column to maintain a
vacuum of 50.8 mm as measured on the manometer. Seat the
perforated dish on the greased gasket of the funnel. Place a
wetted filter paper in the bottom of the dish. Turn the stopcock
to apply the vacuum to the funnel and check the apparatus for
C 110
23 6 1.7C (73.46 3F) and mix thoroughly by alternately
inverting and righting the cylinder slowly for a period of 2 min.
Allow the graduate and contents to stand at 23 6 1.7C for 30
min and then dilute to the 100-mL mark with CO2-free distilled
water at 23 6 1.7C. Mix contents again thoroughly as before
and allow to stand undisturbed at 23 6 1.7C for 24 h.
11.3 Report:
11.3.1 Report the sedimentation height in millilitres after 14,
12, 34, 1, 2, 4, and 24 h, reading the bottom of the meniscus.
(2)
where:
A 5 flow after suction, and
B 5 flow immediately after mixing.
10.5 Precision and Bias:
10.5.1 No precision data are available due to the limited use
of this test method. Therefore, users are advised to develop
their own laboratory precision.
11. Settling Rate of Hydrated Lime
11.1 Significance and Use:
11.1.1 This test method provides a measure of the rate of
settling of a hydrated lime slurry, a form in which this material
is frequently used. In some applications a slow settling slurry
is desirable; in others, fast settling is preferred.
11.2 Procedure:
11.2.1 Place 10.0 g of lime hydrate in a 100-mL glassstoppered graduated cylinder (internal diameter about 24 mm).
Wet with 50 mL of carbon dioxide (CO2) free distilled water at
10
11
A Fisher 14-498, or equivalent, has been found suitable for this purpose.
A Fisher 10-197, or equivalent, has been found suitable for this purpose.
C 110
Temperature of water, C
Quantity of water, mL
Quantity of quicklime, g
Material to Be Tested
Dolomitic
High Calcium
40
25A
400
400
120
100
A
Initial temperature of 40C may be used, provided the report of results states
the initial temperature.
C 110
TABLE 3 Unit Weights and Apparent Specific Gravities
Materials
Portland cement
Hydrated lime
Blended Ottawa sand
Results in C Rise
High Calcium
High Calcium
Dolomitic
Dolomitic
12
11
10
9
Time
Range Tested
30 s
3 min
30 s
3 min
12.344.4
32.156.1
3.612.0
21.236.4
1.56
1.72
1.38
1.62
4.21
4.72
2.84
3.72
1,504 (94)
800 (50)
1,280 (80)
3.15
2.30
2.65
Proportions
by Volume
M
S
N
O
Lime/Sand
1:14:3
1:12:412
1:1:6
1:2:9
1:3
Portland
Cement, g
Hydrated
Lime, g
Blended
Ottawa
Sand, g
564
376
282
188
0
75
100
150
200
300
1,440
1,440
1,440
1,440
1,400
13.4 Procedure:
13.4.1 If the mortar has the correct flow, use a separate
portion of the mortar for the determination of entrained air.
Determine the weight of 400 mL of mortar in accordance with
Test Method C 185.
13.5 Calculation:
13.5.1 Calculate the air content of the mortar and report it to
the nearest 0.1 % as follows:
D 5 ~W1 1 W2 1 W3 1 Vw!/
@~W1/S1! 1 ~W2/S2! 1 ~W3/S3! 1 Vw# A
5 100 2 ~Wm/4D!
where:
D
5
W1 5
W2 5
W3 5
Vw 5
S1 5
S2 5
S3 5
A
5
Wm 5
(3)
Specific
Gravity
A
The unit weight values listed for cementitious materials are assumed values
commonly used in construction practice.
Material
Unit weight,
kg/m3(lb/ft3)A
C 110
(ASTM 152H), variations of as much as 1.0-scale divisions
between two similar hydrometers have been noted. The correction factor brings all four into line with one another. It is not
necessary to repeat this calibration unless changing to a
different hydrometer.
14.4.3 Weigh 40 g of sample.
14.4.4 Add approximately 300 mL of distilled water to the
mixer, 30 mL of the particle-dispersing solution, followed by
40 g of unknown sample. Cover. Agitate for exactly 2 min at
high speed.
14.4.5 Transfer the slurry quantitatively to the 1000-mL
sedimentation cylinder. Make up to approximately 3.2 mm (18
in.) above the mark since it must be read from the top (as the
bottom of the meniscus is not visible) and this will approximate
the 1000-mL calibration of the cylinder. Cylinder temperature
can be adjusted to 20C by running cool water on the outside
of the cylinder and stirring with a thermometer until 20C is
reached. Cap with the rubber stopper. Mix well by inverting the
cylinder 15 or more times. Remove the stopper and put the
cylinder in a water bath that has been previously adjusted to as
close to 20C as is possible. Start the stop watch and note the
time on the clock. At exactly 412 min after start, carefully insert
the hydrometer to the approximate point where the reading is
to be made. Take the reading at exactly 5 min. Record the
reading and temperature (Note 7). Remove the hydrometer and
wash clean of any slurry. Cover the cylinder with the watch
glass.
equivalent spherical diameter of the individual particles expressed in micrometres, using the principle of sedimentation
and Stokes law for particle size determination. It is intended
for use with pulverized limestones with not more than 0.5 %
residue on a 45-m (No. 325) sieve.
14.2 Apparatus:
14.2.1 Soil Hydrometer, ASTM 152H.12
14.2.2 Sedimentation Cylinder, ASTM, 1000-mL capacity.
14.2.3 Rubber Stopper, Size 12.
14.2.4 Thermometer, 0 to 105C.
14.2.5 Stop Watch.
14.2.6 Regular Clock or Watch.
14.2.7 Mixer .13
14.2.8 Water Bath.
14.2.9 Balance .14
14.2.10 Watch Glass.
14.2.11 Graph Paper ,15 3 cycles 3 70 divisions.
14.2.12 Sieve, 45-m (No. 325), stainless steel cloth, brass
frame, 8-in. diameter.
14.2.13 Sieve, 500-mesh, stainless steel cloth, brass frame,
4-in. diameter, 5-in. tall frame.
14.3 Reagents:
14.3.1 Particle-Dispersing Agent 16 (30 mL of 25 % solution is diluted up to 400 mL with distilled water).
14.4 Procedure:
14.4.1 Determine meniscus correction by inserting the hydrometer in the sedimentation cylinder filled to mark with
distilled water. Record the reading at the top of the meniscus
and at the bottom of the meniscus. The difference between the
two readings is the meniscus correction. For example, in Fig. 6,
the correction for the hydrometer used is 1.2. This reading is
added to each R to obtain Rr.
14.4.2 Calibrate the hydrometer by adding 30 mL of the
particle-dispersing solution to the sedimentation cylinder, then
bringing up to the mark with distilled water at 27C. Mix
thoroughly and take a hydrometer reading (read at the top of
the meniscus). Repeat after cooling the cylinder to 17C and
adjusting the meniscus so it is on the mark. Assume a
straight-line relationship and draw a line that gives the composite correction factor. This factor is the difference between
the reading and zero. These are the corrections entered in Table
5 and should be determined for each hydrometer. Four factors
are compensated for in the correction factor: (1) Temperature:
Hydrometers and cylinders are calibrated at 20C; variations
from this temperature produce inaccuracy in the hydrometer
reading; (2) Specific gravity: Addition of dispersant changes
the specific gravity of the solution; (3) Meniscus correction:
Hydrometers are graduated to read at the bottom of the
meniscus but opaque calcium carbonate solutions require
readings at the top of the meniscus; and (4) Hydrometers: In
spite of the supposed similarity in volume of the hydrometers
NOTE 7Temperature must be taken inside the cylinder and not in the
water bath.
12
10
C 110
Correction Factor
17
18
19
20
21
22
23
24
25
26
27
+1.90
+1.52
+1.14
+0.76
+0.39
0.00
0.38
0.76
1.14
1.52
1.90
C 110
TABLE 6 Effective Depth, L, for Hydrometer 152H
Rr
L, cm
Rr
L, cm
0
1
2
3
4
5
6
16.3
16.1
16.0
15.8
15.6
15.5
15.3
31
32
33
34
35
36
37
11.2
11.1
10.9
10.7
10.6
10.4
10.2
7
8
9
10
11
12
15.2
15.0
14.8
14.7
14.5
14.3
38
39
40
41
42
43
10.1
9.9
9.7
9.6
9.4
9.2
13
14
15
16
17
18
19
20
21
14.2
14.0
13.8
13.7
13.5
13.3
13.2
13.0
12.9
44
45
46
47
48
49
50
51
52
9.1
8.9
8.8
8.6
8.4
8.3
8.1
7.9
7.8
22
23
24
25
26
27
28
29
30
12.7
12.5
12.4
12.2
12.0
11.9
11.7
11.5
11.4
53
54
55
56
57
58
59
60
7.6
7.4
7.3
7.1
7.0
6.8
6.6
6.5
17
The Hunter D25A-Tristimulus colorimeter, or equivalent, has been found
suitable for this purpose.
18
The Bausch and Lomb or Carl Zeiss Dry Powder Press Assembly, or
equivalent, has been found satisfactory for this purpose.
19
Registered trademark.
12
C 110
TABLE 7 D as Related to
=L / T
=L / T at 20C
=L / T
D, m
TABLE 8 DD from
=L / T
D, m
=L / T
DD
0.05
0.10
0.15
0.20
0.25
0.7
1.4
2.0
2.7
3.4
1.05
1.10
1.15
1.20
1.25
14.0
14.7
15.4
16.0
16.7
2.0
1.9
1.8
1.7
1.6
0.32
0.31
0.29
0.27
0.26
0.30
0.35
0.40
0.45
0.50
4.0
4.7
5.4
6.0
6.7
1.30
1.35
1.40
1.45
1.50
17.4
18.0
18.7
19.4
20.1
1.5
1.4
1.3
1.2
1.1
0.24
0.23
0.21
0.19
0.18
0.55
0.60
0.65
0.70
0.75
7.4
8.0
8.7
9.4
10.0
1.55
1.60
1.65
1.70
1.75
20.8
21.4
22.1
22.8
23.4
1.0
0.9
0.8
0.7
0.6
0.16
0.15
0.13
0.11
0.10
0.80
0.85
0.90
0.95
1.00
10.7
11.4
12.0
12.7
13.4
1.80
1.85
24.1
24.8
0.5
0.4
0.3
0.2
0.1
0.08
0.07
0.05
0.03
0.02
16.2 Apparatus:
16.2.1 Flour SifterA 114 to 127 mm (412 to 5 in.)
kitchen-type flour sifter of either the squeeze handle type or the
hand crank type. It shall be able to hold at least 300 g of
hydrated lime or 500 g of limestone or quicklime. The wire
mesh openings should be between 0.8 and 1.5 mm.
C 110
TABLE 9 Values for P as Related to Rc, Using a 5 0.988
and W 5 40
16.3.4 Weigh the cup and sample to the nearest 0.1 g and
determine the weight of the sample by difference.
16.3.5 The loose density of the material is calculated and
reported as grams per cubic centimetre, or as pounds per cubic
foot.
16.4 Calculation:
16.4.1 Calculate the loose density as follows:
Rc
Rc
Rc
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
7.0
7.5
8.0
8.5
9.0
9.5
10.0
10.5
11.0
0.0
1.5
2.5
3.5
5.0
6.0
7.5
8.5
10.0
11.0
12.5
13.5
15.0
16.0
17.5
18.5
20.0
21.0
22.5
23.5
25.0
26.0
27.5
11.5
12.0
12.5
13.0
13.5
14.0
14.5
15.0
15.5
16.0
16.5
17.0
17.5
18.0
18.5
19.0
19.5
20.0
20.5
21.0
21.5
22.0
22.5
28.5
30.0
31.0
32.0
33.5
35.0
36.0
37.5
38.5
39.5
40.5
42.0
43.0
44.5
45.5
47.0
48.0
49.5
50.5
52.0
53.0
54.5
56.5
23.0
23.5
24.0
24.5
25.0
25.5
26.0
26.5
27.0
27.5
28.0
28.5
29.0
29.5
30.0
30.5
31.0
31.5
32.0
32.5
33.0
33.5
34.0
57.0
58.0
59.0
60.0
62.0
63.0
64.0
65.0
66.5
67.5
69.0
70.0
71.5
72.5
74.0
75.0
76.5
77.5
79.0
80.0
81.5
82.5
84.0
34.5
35.0
35.5
36.0
36.5
37.0
37.5
38.0
38.5
39.0
39.5
40.0
40.5
85.0
86.5
87.5
89.0
90.0
91.5
92.5
94.0
95.0
96.5
97.5
99.0
100.0
D 5 W/V
(4)
where:
D 5 loose density,
W 5 weight of sample, g, and
V 5 volume of cup, cm3.
16.4.2 For reporting as pounds per cubic foot, multiply
grams per cubic centimetre by 62.43.
16.5 Precision and Bias:
16.5.1 Single Operator PrecisionThe single operator
standard deviation has been found to be 0.4 lb/ft3(Note 10).
Therefore, results of two properly conducted tests by the same
operator on the same material should not differ by more than
1.13 lb/ft3.
16.5.2 Multilaboratory PrecisionThe multilaboratory
standard deviation has been found to be 1.26 lb/ft3(Note 11).
Therefore, results of two properly conducted tests from two
different laboratories on samples of the same material should
not differ by more than 3.6 lb/ft3.
NOTE 11These numbers represent, respectively, the (1s) and (d2s)
limits as described in Practice C 670.
16.3 Procedure:
16.3.1 Weigh the empty density cup to the nearest 0.1 g on
a balance. Place the tared cup on a solid table with a suitable
mat inserted underneath the cup to collect excess sample
spilling over the cup. Fill the flour sifter with more than enough
material to fill the density cup. Start the clock and the sifting
device to facilitate the flow of powder into the cup. Overflow
the cup until there is a cone of excess material.
16.3.2 After 3 min, carefully remove the excess powder by
passing the edge of a spatula blade parallel with, and in contact
with, the top of the cup. Move the spatula smoothly and keep
it level at all times to prevent packing or pulling the sample out
of the cup.
16.3.3 After the cup is level, lightly tap it with the edge of
the spatula to settle the powder. Wipe the outside of the cup
with a lintless cloth or paper towel. Avoid spilling the sample
while transferring the cup to the balance for weighing.
14
C 110
D 5 W/V
(5)
where:
D 5 packed density,
W 5 weight of sample, g, and
V 5 final volume of sample, cm3.
17.4.2 For reporting as pounds per cubic foot, multiply
grams per cubic centimetre by 62.43.
17.5 Precision and Bias:
17.5.1 Single Operator PrecisionThe single operator
standard deviation has been found to be 0.5 lb/ft3(Note 11).
Therefore, results of two properly conducted tests by the same
operator on the same material should not differ by more than
1.4 lb/ft3.
17.5.2 Multilaboratory PrecisionThe multilaboratory
standard deviation has been found to be 1.7 lb/ft3(Note 11).
Therefore, results of two properly conducted tests from two
different laboratories on samples of the same material should
not differ by more than 4.8 lb/ft3.
17.5.3 The above precision statements are based on a
multilaboratory testing program for determination of the
packed bulk density of hydrated lime. No statement is made
regarding the precision of this method as it relates to other
materials. Due to a lack of a recognized industry standard, the
bias of this test method has not been determined.
19.2 Apparatus:
19.2.1 An Enclosed Device, capable of creating a vacuum
on the backside of a sieve causing a rotating slit nozzle to
supply an air stream perpendicular to the bottom of the sieve.
The purpose is to suspend all material on the sieve by the air
stream on a rotating basis.20
19.2.2 Balance, suitable for weighing accurately to 0.01 g.
NOTE 13Selection of balance with regard to accuracy is dependent on
the sample size chosen and residue retained and must be consistent with
the accuracy required. Therefore, a balance weighing accurately to 0.001
g may be desired.
NOTE 15Static electrical charges can often develop on the cover (if it
is made of plastic) causing it to hold a heavy film of the material being
20
The apparatus describes commercially available units sold by the Alpine
American Corporation of Natick, Massachusetts. Although the description of the
apparatus is directed toward this commercially available equipment, it does not
restrict the use of other equivalent equipment which may be available or may be
constructed, as long as it follows the general principles outlined under the summary
of this test method.
15
C 110
20.4.8 Riffle Sample Splitter ,26 open pan, 12.7 mm (12 in.)
chute width.
20.4.9 Stopwatch.
20.5 Reagents and Materials:
20.5.1 Milling Solution, a 0.1 % solution of acrylate based
dispersant.27 The dispersant chosen should not increase the
solubility of limestone in water.28
20.6 Sampling:
20.6.1 Sample in accordance with Practice D 75.
20.6.2 Reduce the sample in accordance with Practice
C 702 and prepare by sieving out the material that passes a
850-m (No. 20) sieve29 and is retained on a 425-m (No. 40)
sieve.29
20.7 Procedure:
20.7.1 Weigh seven grinding media, make adjustments (by
substitutions or filing) to bring total weight to 160 g 6 1 g.
20.7.2 If the jar mill has provision for automatic shut-off, set
it for 5000 revolutions, otherwise determine the mill r/min by
counting the revolutions in an accurately timed period (using
stopwatch) and then calculate the exact time required for 5000
revolutions.
20.7.3 Weigh out 20 6 0.01 g of dried 20 by 40 mesh
limestone. Record actual weight as W1.
20.7.4 Add 180 mL of milling solution to clean and empty
mill jar.
20.7.5 Add the seven grinding media and quantitatively
transfer the limestone sample to the mill jar and secure the top.
20.7.6 Place the mill jar on the mill rollers and operate the
mill for the exact time required to make 5000 revolutions.
20.7.7 Quantitatively transfer the limestone slurry from the
jar mill by rinsing the entire contents onto a coarse sieve (for
example, 3.35 mm (No. 6)) and an underlying 75-m (No. 200)
sieve. Rinse the media and coarse sieve and separate the sieves.
20.7.8 Wet sieve the sample remaining on the 75-m (No.
200) sieve to remove the finer material.
20.7.9 Dry and weigh the residue from the 75-m (No. 200)
sieve and record as W2 (to the nearest 0.01 g).
20.8 Calculation:
20.8.1 Calculate the grindability index (GI) as follows:
sieved. If tapping will not loosen the material, a static face sheet21 may be
used to wipe the cover surface before starting the test.
(6)
where:
S 5 sample weight, g, and
R 5 weight of sieve residue, g.
19.5 Precision and Bias:
19.5.1 There are as yet insufficient analyzed data to permit
preparation of a precision and bias statement for this test
method.
20. Limestone Grindability Determination by the
Laboratory Ball Mill Method
20.1 Scope:
20.1.1 This test method is used to determine the relative
grindability or ease of pulverization of limestones of differing
hardness and to report this as a grindability index.
20.1.2 This test method is applicable to all types of limestone.
20.2 Summary of Test Method:
20.2.1 Limestone of a specified size range is wet ground in
a ball mill therein receiving a specified amount of grinding
energy. The amount of minus 75-m (200-mesh) limestone
produced is measured by wet sieving and reported as the
percent passing 75-m (200-mesh) after 5000 revolutions. This
is the grindability index.
20.3 Significance and Use:
20.3.1 This test method is useful for comparison and acceptance testing of limestone for applications where fine ground
limestone is desired.
20.4 Apparatus:
20.4.1 Jar Mill ,22 operated at 110 6 10 r/min.
20.4.2 Mill Jar ,23 ceramic 14 cm (512in.) diameter by 16.2
cm (634in.) high.
20.4.3 Grinding Media, 160 6 1 g total, consisting of seven
ceramic 21 by 21 mm (1316 by 1316 in.) cylindrical grinding
media24 (about 23 g each).
20.4.4 The sieves used shall conform to the requirements of
Specification E 11.
20.4.5 Weights and weighing devices, shall conform to the
requirements of Specification C 1005.
20.4.6 Drying Oven, capable of maintaining 100C.
20.4.7 A Chipmunk Crusher ,25 capable of breaking large
rocks to less than 6.35 mm (14 in.).
(7)
21
Commercially available.
U.S. Stoneware No. 753-RM rotating mill, or equivalent, has been found
satisfactory for this purpose.
23
U.S. Stoneware mill jar No. 774-B-00, or equivalent, has been found
satisfactory for this purpose.
24
U.S. Stoneware burundum 1316in. grinding media, or equivalent, has been
found satisfactory for this purpose.
25
Bico-Braun Crusher (jaw, chipmunk) Model No. 241-34, or equivalent, has
been found satisfactory for this purpose.
22
26
Curtis Matheson Scientific Model No. 095-851, or equivalent, has been found
satisfactory for this purpose.
27
Alcosperse 149 from Alco Chemical Corp., or equivalent, has been found
satisfactory for this purpose.
28
Distilled or deionized water should be used for milling solution or solubility
tests.
29
U.S. Standard Sieves 6, 20, 40 and 200 mesh sizes; 20.3 cm (8 in.) diameter
by 5 cm (2 in.) depth with stainless steel wire cloth.
16
C 110
inclined position (Note 17), or gently whirl it in a horizontal
circle as to free the lime from air until no further air bubbles
rise to the surface of the liquid. If a proper amount of hydrated
lime has been added, the level of the liquid will be in its final
position at some point of the upper series of graduations. Take
the final reading after the flask has been immersed in the water
bath in accordance with 22.3.4.
21.4.4 Immerse the flask in a constant-temperature water
bath for sufficient periods of time in order to avoid flask
temperature variations greater than 0.2C between the intial
and final readings.
21.5 Calculation:
21.5.1 The difference between the first and the final readings
represents the volume of liquid displaced by the mass of
hydrated lime used in the test.
21.5.2 Calculate the hydrated lime density, p, as follows:
p~Mg/m3 ! 5 p~g/cm3! 5
(8)
21.3 Reagents:
21.3.1 Ethyl Alcohol (Abolute), free of water, shall be used
in the density determination. The ethanol used should not have
been denatured with methanol.
21.4 Procedure:
21.4.1 Determine the specific gravity of hydrated lime on
the material as received, unless otherwise specified.
21.4.2 Fill the flask (Note 17) with denatured ethyl alcohol
to a point on the stem between the 0 and 1-mL mark. Dry the
inside of the flask above the level of the liquid, if necessary,
after pouring. Record the first reading after the flask has been
immersed in the water bath (Note 18) in accordance with
22.3.4.
NOTE 17It is advisable to use a rubber pad on the table top when
filling or rolling the flask.
NOTE 18Before the hydrated lime has been added to the flask, a loose
fitting lead-ring weight around the stem of the flask will be helpful in
holding the flask in an upright position in the water bath, or the flask may
be held in the water bath by a buret clamp.
C 110
should be checked. It could, also, be an indication of degradation of the material.
22.3 Apparatus and Materials:
22.3.1 BalanceA balance or scale readable and accurate
to 0.1 g or 0.1 % of the test load, whichever is greater, at any
point within the range of use.
22.3.2 SievesThe sieves shall be mounted on substantial
frames, either full height or extended, and constructed in a
manner that will prevent loss of material during sieving. The
sieves shall be of equal diameter and shape to allow for
nesting. The sieves shall conform to Specification E 11.
22.3.3 Sieve PanA pan of the same diameter as the sieves.
22.3.4 ContainerA pan or vessel of a size sufficient to
contain the sample covered with water and to permit vigorous
agitation without loss of any part of the sample or water.
22.3.5 OvenAn oven of sufficient size, capable of maintaining a uniform temperature of 110 6 5C.
22.4 Sampling:
22.4.1 Since sampling is equally as important as testing, it is
critical that samples be obtained that are representative of the
materials being investigated. Therefore, sampling must be
conducted in as conscientious a manner as possible. Incremental sampling should be employed, combining the increments
into one bulk sample.
22.4.2 Use the procedures described in Practice C 50, or
Practice D 75.
22.4.3 To reduce the bulk sample to the amount needed for
testing, the use of mechanical splitters or the process of
quartering must be used. At no time can a random lot be drawn
from the bulk sample for testing.
22.5 Procedure:
22.5.1 Dry the test sample to constant mass at a temperature
of 110 6 5C. Determine the mass to the nearest 0.1 % of the
mass of the test sample. The amount of sample required for this
test shall be between 200 to 400 g. The test sample shall be the
end result of the reduction. Reduction to an exact predetermined mass shall not be permitted.
22.5.2 After drying and determining the mass, place the test
sample in the container and add sufficient water to cover it.
Agitate the sample with sufficient vigor to result in complete
separation from the coarser particles of all particles finer than
the 75 micrometre sieve from the coarser particles, and to bring
the fine material into suspension. Immediately pour the wash
water containing the suspended and dissolved solids over the
nested sieves (see Note 21), arranged with the coarser sieve on
top. Take care to avoid, as much as feasible, the decantation of
coarser particles of the sample.
NOTE 24A twelve inch screen can be used, but the amount of material
retained on a sieve is limited to 6 kg/m2.
22.5.7 Record the weight of material from each size increment and the sieve pan by weighing on a balance to the nearest
0.1 g (see Note 25). Total the retained masses of all the size
increments and the sieve pan from the dry sieve test. If this
amount differs from the dry mass of the sample after washing
by more than 0.3 %, the results should not be used for
acceptance purposes.
NOTE 25A soft bristle can be used to help remove material from
sieves, but at no time can material be forced through a sieve.
22.6 Calculation:
22.6.1 Calculate the amount of a material passing a 75
micrometre sieve by washing as follows:
A 5@~B C! / B# 3 100
(9)
where:
A 5 percentage of material finer than a 75 micrometre sieve
by washing,
B 5 original dry mass of sample, g, and
C 5 dry mass of sample after washing, g.
22.6.2 Calculation of amount of material passing individual
sieves follows:
22.6.2.1 Add weight of material finer than a 75 micrometre
sieve by washing (B C, in 22.6.1) to the weight of material
finer than a 75 micrometre sieve obtained in the dry screening,
if washed.
22.6.2.2 Calculate the percent retained on each sieve and the
pan as follows:
A 5~B / C! 3 100
18
(10)
C 110
22.7 Precision and Bias:
22.7.1 The precision and bias of this method has not been
determined by this subcommittee at this time. For precision
and bias data from similar methods refer to Test Method C 117
and Test Method C 136.
where:
A 5 percentage of material retained on each sieve (pan),
B 5 dry mass of sample, g, retained on each sieve (pan),
and
C 5 original dry mass of sample, g.
22.6.2.3 Calculate the cumulative percent retained for each
screen by adding the percentage retained on that sieve and all
sieves of larger diameter opening.
22.6.2.4 For each screen, subtract the cumulative percent
retained from 100 to obtain the percent passing.
23. Keywords
23.1 apparent density; dry brightness; dry screening; fineness; lime putty; limestone grindability; particle size; slaking
rate
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at
610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website (www.astm.org).
19