Respironics BiPAP Auto Bi Flex Provider Manual NOT Mseries

Download as pdf or txt
Download as pdf or txt
You are on page 1of 72

with

Bi-Flex

PROVIDER MANUAL

This BiPAP system is covered by one or more of the following patents: US Patent Nos.
5,148,802; 5,313,937; 5,433,193; 5,632,269; 5,803,065; 6,029,664; 6,305,374;
6,539,940; 5,239,995; Re 35,295; 5,492,113; 5,551,418; 5,904,141; 5,970,975; and
6,426,689.

2005 Respironics, Inc. All rights reserved.

TABLE

OF

CONTENTS

Chapter 1: Package Contents ............................................................................................1-1


Chapter 2: Warnings and Cautions ...................................................................................2-1
2.1 Warnings .......................................................................................................2-1
2.2 Cautions ........................................................................................................2-3
2.3 Intended Use .................................................................................................2-3
2.4 Contraindications ..........................................................................................2-4
Chapter 3: Introduction.....................................................................................................3-1
3.1 Overview ......................................................................................................3-1
3.2 Therapies .......................................................................................................3-2
3.2.1 Bi-Flex Comfort Feature .....................................................................3-3
3.3 Ramp ...........................................................................................................3-4
3.4 Event Denitions .........................................................................................3-5
3.5 Auto-On and Auto-O Features ....................................................................3-6
3.5.1 Auto-On ............................................................................................3-6
3.5.2 Auto-O ............................................................................................3-6
3.6 Access Levels .................................................................................................3-6
3.6.1 Provider Mode Access Level (Setup) ...................................................3-6
3.6.2 User Mode Access Level ....................................................................3-7
3.7 Denitions, Acronyms, and Abbreviations ....................................................3-8
3.8 Symbol Key .................................................................................................3-9
3.9 How to Contact Respironics........................................................................3-10
Chapter 4: Controls and Displays ......................................................................................4-1
4.1 Controls and Displays ...................................................................................4-1
4.1.1 Display Screen.....................................................................................4-1
4.1.2 Control Buttons ..................................................................................4-3
4.1.3 Audible Alerts and Indicators .............................................................4-4
4.2 Navigating the Screens ..................................................................................4-4
4.2.1 LED Backlight for Buttons .................................................................4-5
4.3 Patient Circuit Connection ..........................................................................4-5
4.4 Rear Panel ....................................................................................................4-5
4.5 SmartCard ....................................................................................................4-6
Chapter 5: Setup ................................................................................................................5-1
5.1 Preparing the Device .....................................................................................5-1
5.1.1 Installing the Air Filters .......................................................................5-1
5.1.2 Assembling the Patient Circuit ............................................................5-2
BiPAP Auto with Bi-Flex Provider Manual

5.1.3 Supplying Power to the Device ............................................................5-2


5.1.4 Startup ................................................................................................5-4
5.1.5 Entering Provider Mode .....................................................................5-6
5.2 Connecting the Patient ..................................................................................5-6
5.3 Setting up the SmartCard ..............................................................................5-6
5.3.1 Downloading Data .............................................................................. 5-6
5.3.2 Programming a SmartCard ..................................................................5-7
5.3.3 Changing Settings Using a SmartCard ................................................5-7
Chapter 6: Changing Settings ............................................................................................6-1
6.1 Changing Settings in the Provider Mode .......................................................6-1
6.1.1 Provider Mode Navigation .................................................................6-1
6.1.1.1 Changing the Provider Mode Settings .......................................6-2
6.2 Changing Settings in User Mode ...................................................................6-9
Chapter 7: Alerts ...............................................................................................................7-1
7.1 Introduction .................................................................................................7-1
7.1.1 Overview of Alert Behavior ................................................................7-1
7.1.1.1 Alert Sounds Behavior ..............................................................7-1
7.1.1.2 Display Behavior .......................................................................7-2
7.2 Alerts ...........................................................................................................7-2
7.2.1 System Error Alert .............................................................................7-2
7.2.2 Card Error Alert .................................................................................7-3
7.3 Patient Disconnect Alert ...............................................................................7-3
7.4 Prescription Complete Screen .......................................................................7-4
7.5 Alert Summary Table ....................................................................................7-5
Chapter 8: Cleaning and Maintenance...............................................................................8-1
8.1 Cleaning the Device .....................................................................................8-1
8.2 Cleaning or Replacing the Inlet Filters ...........................................................8-1
8.3 Cleaning and Disinfection for Multiple Users ...............................................8-3
8.4 Maintenance .................................................................................................8-3
Chapter 9: Adding Supplemental Oxygen .........................................................................9-1
9.1 Adding Supplemental Oxygen .....................................................................9-2
9.2 Supplemental Oxygen Concentrations ........................................................9-2
Chapter 10: Circuits and Accessories ...............................................................................10-1
10.1 Circuit Congurations .............................................................................10-1
10.2 Circuits and Accessories ............................................................................10-1
10.3 Masks, Exhalation Ports, and Related Accessories ....................................10-2
ii

BiPAP Auto with Bi-Flex Provider Manual

10.4 Humidiers .............................................................................................10-2


10.5 Software ..................................................................................................10-2
Chapter 11: Specications ..............................................................................................11-1
Environmental...................................................................................................11-1
Physical .............................................................................................................11-1
Electrical ...........................................................................................................11-1
Pressure .............................................................................................................11-2
Control Accuracy .............................................................................................. 11-2
Disposal ............................................................................................................11-2
Pressure Drop Versus Flow for Patient Circuits ................................................11-3
Appendix A EMC Information ......................................................................................... A-1
Guidance and Manufacturers Declaration - Electromagnetic Emissions ............. A-1
Guidance
uidance and Manufacturers Declaration - Electromagnetic Immunity ............. A-2
Guidance
uidance and Manufacturers Declaration - Electromagnetic Immunity ............. A-3
Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and This Device ............................................................................... A-4

BiPAP Auto with Bi-Flex Provider Manual

iii

iv

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 1: PACKAGE CONTENTS


NOTE:

Always use these instructions along with the User Manual when assembling
or adjusting this equipment.

Encore Pro SmartCard

Power Cord

BiPAP Auto w/ Bi-Flex

Filter Cap

Disposable Ultrane Filter


Reusable Gray
Foam Filters

User Manual

Flexible Tubing
6 ft. (1.83 m) X 22 mm i.d.

BiPAP Auto with Bi-Flex Provider Manual

External AC Power Supply

1-1

1-2

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 2: WARNINGS
CAUTION!

2.1

AND

CAUTIONS

US federal law restricts this device to sale by or on the order of a physician.

WARNINGS

WARNING:

Indicates the possibility of injury to the patient or the operator.

This manual serves as a reference. The instructions in this manual are not intended to supersede
the health care professionals instructions regarding the use of the device.

The operator should read and understand this entire manual before using the device.

The device should be used only with masks and connectors recommended by Respironics or with
those recommended by the health care professional or respiratory therapist. See Chapter 10 for
approved patient circuits. A mask should not be used unless the device is turned on and operating
properly. The exhalation port(s) associated with the mask should never be blocked.
Explanation of the Warning: The device is intended to be used with special masks or connectors
that have exhalation ports to allow continuous ow of air out of the mask. When the device is
turned on and functioning properly, new air from the device ushes the exhaled air out through
the mask exhalation port. However, when the device is not operating, enough fresh air will not
be provided through the mask, and exhaled air may be rebreathed. Rebreathing of exhaled air for
longer than several minutes can in some circumstances lead to suocation.

In the event of a power or device failure, audible and visual alarm signals will activate. The device
must be disconnected from the patient immediately. As is the case with most therapy devices with
passive exhalation ports, when power is lost, sucient air will not be provided through the circuit,
and exhaled air may be rebreathed.

At low EPAP pressures, the ow through the exhalation port may be inadequate to clear all exhaled gas from the tubing. Some rebreathing may occur.

If oxygen is used with the device, the oxygen ow must be turned o when the device is not operating.
Explanation of the Warning: When the device is not in operation and the oxygen ow is left on,
oxygen delivered into the tubing may accumulate within the devices enclosure. Oxygen accumulated in the device enclosure will create a risk of re.

Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an
open ame.

When using oxygen with this system, a Respironics Pressure Valve (Part Number 302418) must
be placed in-line with the patient circuit.

For proper use, the external AC power supply must be placed feet down, in the upright position.

BiPAP Auto with Bi-Flex Provider Manual

2-1

Operation of the device may be adversely aected by:


Electromagnetic elds exceeding the level of 10 V/m in the test conditions of EN
60601-1-2
Operation of high frequency (diathermy) equipment
Debrillators, or short wave therapy equipment
Radiation (e.g., x-ray, CT)
Magnetic elds (e.g., MRI)

2-2

Do not use the device in the presence of a ammable anaesthetic mixture in combination with
oxygen or air, or in the presence of nitrous oxide.

Do not use the device at room temperatures above 95 F (35 C). If the device is used at room
temperatures above 95 F (35 C), the temperature of the airow may exceed 105 F (41 C),
which could cause thermal irritation or injury to the patients airway.

Do not operate the device in direct sunlight or near a heating appliance because these conditions
can increase the temperature of the airow delivered to the patient.

To reduce the risk of contamination, you may place a bacteria lter (Part Number 342077) inline between the device and the patient.

The device does not have an alarm to detect occlusion of the exhalation port. Before each use,
inspect the patient circuit to verify that the port is not occluded. Occlusion or partial occlusion
can reduce airow and result in rebreathing of exhaled air.

Do not use antistatic or electrically conductive hoses or tubing with the device.

When the device is used with a humidier, position the humidier so that the water level in the
humidier is lower than the patient, and the humidier is on the same level or lower than the
device.

If you detect any unexplained changes in the performance of the device, if the device and/or the
power supply is dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is
broken, seek the assistance of Respironics or an authorized service center.

Repairs and adjustments must be performed by Respironics or an authorized service center. Service done by inexperienced or unqualied personnel, or installation of unauthorized parts could
cause injury, invalidate the warranty, or result in costly damage.

Electrical cords, cables, and the power supply device should be periodically inspected for damage
or signs of wear. Replace any damaged parts before using.

To avoid electrical shock, unplug the device before cleaning it.

Pins of connectors identied with the ESD warning symbol should not be touched. Connections should not be made to these connectors unless ESD precautionary procedures are used.
Precautionary procedures include methods to prevent build-up of electrostatic discharge (e.g., air
conditioning, humidication, conductive oor coverings, non-synthetic clothing), discharging
ones body to the frame of the equipment or system or to earth or a large metal object, and bonding oneself by means of a wrist strap to the equipment or system or to earth.

Verify the operation of the Patient Disconnect alert with any changes in the patient circuit.

Verify that the Patient Disconnect alert is active if required for medical reasons.

BiPAP Auto with Bi-Flex Provider Manual

2.2

CAUTIONS

CAUTION:

Indicates the possibility of damage to the device.

The device may only be operated at temperatures between 41 F and 95 F.

Do not immerse the device or allow any liquid to enter the enclosure or the inlet lter.

Condensation may damage the device. Always allow the device to reach room temperature before
use.

Use the AC cable retainer to keep the AC power cord from being unintentionally disconnected.

NOTE:

2.3

Additional warnings, cautions, and notes are located throughout this manual.

INTENDED USE

The BiPAP Auto with Bi-Flex system delivers positive airway pressure therapy for the treatment of
adult Obstructive Sleep Apnea (OSA) only.
The BiPAP Auto with Bi-Flex is to be used only on the instruction of a licensed health care professional. The system can deliver Bi-level therapy or Auto Bi-level therapy, both with and without
Bi-Flex. Your home care provider will make the correct pressure settings according to your health care
professionals prescription.
When set in the Auto Bi-level therapy mode, the system will monitor patient breathing as they sleep
and automatically adjust the pressure to meet patient needs. When in the Bi-level therapy, the system
will deliver a continuous, set pressure during the night.
Several accessories are available to make OSA treatment with the BiPAP Auto with Bi-Flex system
as convenient and comfortable as possible. To ensure that patients receive the safe, eective therapy
prescribed for them, use only Respironics accessories.
IMPORTANT! Read and understand the entire provider manual before operating
this system.

BiPAP Auto with Bi-Flex Provider Manual

2-3

2.4

CONTRAINDICATIONS

When assessing the relative risks and benets of using this equipment, the clinician should understand
that this device can deliver pressures of up to 25 cm H2O. Also, in the unlikely event of certain fault
conditions, a maximum pressure of 35 cm H2O is possible. Studies have shown that the following pre-existing conditions may contraindicate the use of positive airway pressure therapy for some
patients:

Bullous lung disease

Bypassed upper airway

Pneumothorax

Pathologically low blood pressure

Pneumocephalus has been reported in a patient using nasal Continuous Positive Airway Pressure.
Caution should be used when prescribing CPAP for susceptible patients such as those with cerebral spinal uid (CSF) leaks, abnormalities of the cribriform plate, prior history of head trauma,
and/or pneumocephalus. (Chest 1989; 96:1425-1426)

The use of positive airway pressure therapy may be temporarily contraindicated if the patient exhibits
signs of a sinus or middle ear infection. This therapy is not for use with patients whose upper airways
are by-passed. Should your patient have any of these conditions, a physician will determine if Bi-level
therapy is appropriate.

2-4

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 3: INTRODUCTION
3.1

OVERVIEW

WARNING:

The device can operate on AC or DC power. The DC power option is


not intended as a battery backup during use of AC power.

CAUTION:

When DC power is obtained from a vehicle battery, the device should


not be used while the vehicles engine is running. Damage to the device or the vehicle may occur.

FIGURE 31 BIPAP AUTO

WITH

BI-FLEX DEVICE

The device, shown in Figure 3-1, is a low-pressure, electrically driven sleep apnea system with electronic pressure control. The devices pressure controls are adjusted to deliver pressure support to the
patient.
The device is intended to augment patient breathing by supplying pressurized air through a patient
circuit. It senses the patients breathing eort by monitoring airow in the patient circuit and adjusts
its output to assist in inhalation and exhalation. This assistance is provided by the administration of
two levels of positive pressure. During exhalation, pressure is variably positive or near ambient. During
inspiration, pressure is variably positive and always equal to or higher than the expiratory level.

BiPAP Auto with Bi-Flex Provider Manual

3-1

3.2

THERAPIES

The BiPAP Auto with Bi-Flex device has the following therapies:
Bi-level: Provides one level of output pressure during EPAP (Expiratory Positive Airway Pressure) and a
second higher level during IPAP (Inspiratory Positive Airway Pressure). All breaths are patient-triggered,
patient-cycled, and pressure controlled.
Bi-level with Bi-Flex: Bi-level therapy with pressure relief upon exhalation to improve patient comfort
based on patient needs.
Auto Bi-level: Delivers spontaneous Bi-level therapy with automatically adjusting EPAP and IPAP levels
that adjust to meet the patients needs.
Auto Bi-level with Bi-Flex: Auto Bi-level therapy with pressure relief upon exhalation to improve patient
comfort based on patient needs.
The device triggers to Inspiratory Positive Airway Pressure (IPAP) in response to spontaneous inspiratory eort and cycle to Expiratory Positive Airway Pressure (EPAP) during exhalation. Figure 3-2
illustrates the trigger and cycle concepts.

FIGURE 3-2 TRIGGERING

AND

CYCLING

The level of pressure support (PS) delivered is determined by the dierence between the IPAP and
EPAP settings (PS = IPAP - EPAP).
When the device is in Provider access mode, you can congure the following settings:

3-2

Therapy mode setting (bPAP for Bi-level mode, bFLE for Bi-level with Bi-Flex mode, AbPAP for
Auto Bi-level mode, and AbFLE for Auto Bi-level with Bi-Flex mode)

IPAP setting

EPAP setting

Min EPAP setting

Max IPAP setting

Max PS setting

Flex setting

Rise time setting

Ramp length setting

Ramp start pressure setting

Split night time setting

Patient disconnect alert (enable/disable)


BiPAP Auto with Bi-Flex Provider Manual

Erase therapy hours

LED backlight setting (enable/disable)

Humidier heat setting (for Respironics REMstar Heated Humidier only)

3.2.1 BI-FLEX COMFORT FEATURE


The BiPAP Auto device consists of a special comfort feature called Bi-Flex. When the device is in Auto
Bi-level with Bi-Flex or Bi-level with Bi-Flex mode, the Bi-Flex attribute adjusts therapy by inserting a
small amount of pressure relief during the latter stages of inspiration and during active exhalation (the
beginning part of exhalation). In the following diagram, the bold lines represent Bi-Flex in comparison
to the dashed line representing normal BiPAP therapy. Bi-Flex levels of 1, 2, or 3 progressively reect
increased pressure relief that will take place at the end of inspiration and at the beginning of expiration.
Note:

The patient also has access to this setting, if Bi-Flex is enabled.

FIGURE 3-3 BI-FLEX

IN

COMPARISON

BiPAP Auto with Bi-Flex Provider Manual

TO

TRADITIONAL BI-LEVEL THERAPY

3-3

3.3

RAMP

The device is equipped with a linear ramp function. The Ramp feature will reduce the pressure and
then gradually increase (ramp) the pressure to the prescription pressure setting so patients can fall
asleep more comfortably. Figure 34 illustrates how the devices ramp function works.
Airflow
Turned On
Ramp Button
Pressed

IPAP
Pressure
CPAP/EPAP
Pressure

lta

de

Minimum
Ramp
Pressure
0 cm H2O
Minutes

Ramp Time

Linear Ramp
* The delta is defined as the lesser of 2 cm H2O
and the difference between IPAP and EPAP
pressure settings.

FIGURE 34 THE RAMP FUNCTION


NOTE:

3-4

When the device is in Auto Bi-level or Auto Bi-level with Bi-Flex mode, pressing the Ramp button provides pressure relief by lowering the device output
pressure to the Minimum Auto pressure setting and restoring the Auto Bi-level
algorithm at the new pressure, as long as the device is not congured for split
night therapy or the preset split night Bi-level time period has expired during
split night therapy.

BiPAP Auto with Bi-Flex Provider Manual

3.4

EVENT DEFINITIONS

The BiPAP Auto monitors breathing and detects apneas and hypopneas:
EVENT

DEFINITION

Apnea Detection

An apnea is indicated if there is an 80% reduction in airow for 10


seconds compared to the average airow over an extended period of
several minutes or if there is no airow detected for 10 seconds.

Flow Limitation
Detection

The ow limitation algorithm analyzes the peak and shape of the inspiratory airow waveform. The algorithm looks for relative changes
in the peak, atness, roundness, or skewness (shape) of the inspiratory portion of the airow waveform. These changes are observed
both over a short period of time (groups of 4 breaths) and over a long
period of time (several minutes). Statistical measures are used to
help minimize false event detection while allowing the device to be
sensitive to even small changes. Also, these measures are disabled
during unstable breathing conditions such as snore, apneas, hypopneas, high leak, and variable breathing.
NOTE:

Hypopnea Detection

Snore Detection

Flow Limitation Detection is only active


during the Auto Bi-level and Auto Bi-level
with Bi-Flex modes.

A hypopnea is indicated if there is approximately a 40% reduction


in airow for a duration of between 10 and 60 seconds, compared
to the average airow over an extended period of several minutes.
Following a reduction, the device must see two recovery breaths in
order to label the event as a potential hypopnea. For the event to be
determined to be a hypopnea, additional criteria are evaluated:
Are the pressures less than or equal to 8 cm H2O

Is ow limitation present in the valley of the hypopnea

Is the hypopnea terminated with breaths having large


tidal volumes

Has a reduction in the energy content of the airow


signal been observed

The pressure signal is only evaluated for the possibility of snoring


when the patient is inhaling. Snoring-related pressure vibrations
detected on exhalation are often linked to mask vibrations. Snore
detection is disabled at IPAP pressure setpoints greater than 20 cm
H2O.

BiPAP Auto with Bi-Flex Provider Manual

3-5

3.5

AUTO-ON

AND

AUTO-OFF FEATURES

The device includes automatic pressure activation (Auto-On) and automatic pressure cessation (AutoO) features.

3.5.1 AUTO-ON
With the Auto-On feature, the device automatically transitions from the Standby state to the Operate
state when the patient begins breathing on the device (after 3 consecutive breaths).

3.5.2 AUTO-OFF
Setting the Patient Disconnect setting to 1 enables the Auto-O feature (in addition to enabling the
Patient Disconnect alert). When Auto-O is enabled, the device automatically transitions from the
Operate state to the Standby state when the patient removes the mask from the airway. The device will
transition to standby within one and a half to two minutes following removal of the mask.

3.6

ACCESS LEVELS

There are two levels of access for the device: Provider Mode and User Mode.

3.6.1 PROVIDER MODE ACCESS LEVEL (SETUP)


The Provider mode unlocks additional parameters that are not available to the patient. Providers can
temporarily access these parameters via the user interface by completing the following steps:
1.

Press the Left and Right user buttons down simultaneously.

2.

With the user buttons pressed down, plug in the device to power up the unit.

3.

Release the Left and Right user buttons when you see the word SETUP appear in the top right
corner of the display. This indicates that you are now in Provider mode.

NOTE:

If the device is already on, you must rst turn the device off and unplug it
before entering Provider mode.
Left User
Button

Right User
Button

Humidifier
Button

Ramp Button

Start/Stop Button

FIGURE 3-5 DEVICE CONTROL PANEL


3-6

BiPAP Auto with Bi-Flex Provider Manual

LED Backlight setting

Maximum IPAP and minimum EPAP auto pressure

Split night BiPAP time

Maximum PS

The following is also true in Provider mode:

The IPAP and EPAP setting screens only display if the device is in Bi-level or Bi-level with Bi-Flex
mode.

The Rise Time screen is only displayed if the device is in Bi-level mode and IPAP is not equal to
EPAP, or if the device is in Auto Bi-level mode.

The Ramp Start Pressure screen is only displayed if the device is in Bi-level or Bi-level with Bi-Flex
mode and the Ramp length setting is between 5 and 45 minutes.

The Flex Setting screen is only displayed if the device is in Auto Bi-level with Bi-Flex or Bi-level
with Bi-Flex mode.

The Split Night Time screen only displays if the device is in Auto Bi-level or Auto Bi-level with
Bi-Flex mode.

The Max IPAP, Min EPAP, and Max PS screens only display if the device is in Auto Bi-level or
Auto Bi-level with Bi-Flex mode.

3.6.2 USER MODE ACCESS LEVEL


The device defaults to User mode upon startup. Users can access user mode settings by pressing and
holding the Ramp button while the device is in standby.
NOTE:

If you temporarily set the device to Provider mode by pressing the Left and
Right User buttons, the unit will return to User mode when any of the following occurs:

The Start/Stop button is pressed.

Any user interface button is not pressed for more than 60 seconds.

The device is unplugged and then powered up again.

The following settings can be modied in User mode when the user presses and holds the Ramp button for several seconds:

Flex setting (when Bi-Flex is enabled by the Provider)

Rise time setting (when in Bi-level or Auto Bi-level mode)

Ramp start pressure setting (when ramp enabled by Provider)

LED backlight for control buttons (enable/disable)

Patient disconnect alert (enable/disable)

Answers to FOSQ test questions (when the SmartCard is inserted)

Additionally, the Humidier heat setting can be modied by pressing and holding the Heat button.

BiPAP Auto with Bi-Flex Provider Manual

3-7

The following is also true in User mode:

The Flex Setting screen is only displayed if the device is in Auto Bi-level with Bi-Flex or Bi-level
with Bi-Flex mode.

The Rise Time screen is only displayed if the device is in Bi-level mode and IPAP is not equal to
EPAP, or if the device is in Auto Bi-level mode.

The Ramp Start Pressure Setting screen is only displayed if the device is in Bi-level or Bi-level with
Bi-Flex mode and the Ramp length is between 5 and 45 minutes.

The FOSQ screen is only displayed if the SmartCard is inserted prior to entering the user mode
setting screens.

3.7

DEFINITIONS, ACRONYMS,

AND

ABBREVIATIONS

The following terms appear in this manual:


TERM/

DEFINITION

ACRONYM

3-8

Apnea

A condition marked by the cessation of spontaneous breathing.

Auto-OffThis feature

When enabled through the Patient Disconnect setting, causes the device to automatically transition from the Operate state to the Standby
state whenever the mask is removed from the airway.

Auto-On

The device automatically transitions from the Standby state to the


Operate state when you begin breathing (3 consecutive breaths) on
the device.

Bi-Flex

A therapy feature that establishes a level of pressure relief taking


place at the end of inhalation and at the start of exhalation.

EPAP

Expiratory Positive Airway Pressure

High Priority Alert

Alert signal indicating a condition that requires immediate attention.

IPAP

Inspiratory Positive Airway Pressure

Low Priority Alert

Alert signal indicating an informational message.

Medium Priority Alert

Alert signal indicating a condition that requires operator awareness.

Operate State

The state of the device when the unit and the airow are both on.

Standby State

The state of the device when the unit is on, but the airow is off.

OSA

Obstructive Sleep Apnea

Ramp

A feature that may increase patient comfort when therapy is started.


The ramp feature reduces the pressure and then gradually increases
(ramps) the pressure to the prescription setting, so you can fall asleep
more comfortably.

Rise Time

The time it takes for the device to change from EPAP to IPAP. You can
adjust this time for your comfort.
BiPAP Auto with Bi-Flex Provider Manual

3.8

SYMBOL KEY

The following symbols appear on the device label:


SYMBOL

DESCRIPTION
Consult accompanying instructions for use.
DC Power
Type BF Applied Part
Class II (Double Insulated)

IPX1

Drip Proof Equipment


Electrostatic Discharge
European Declaration of Conformity

0123

Canadian/US Certication
Notied Body Approval for Standards Compliance
TUV Safety Standard Compliance
UL Recognized for Canada and the United States
No User Serviceable Parts

BiPAP Auto with Bi-Flex Provider Manual

3-9

3.9

HOW

TO

CONTACT RESPIRONICS

To have your unit serviced, contact your home care provider. If you need to contact Respironics
directly, call 1-800-345-6443 or use the following address:

3-10

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 4: CONTROLS

AND

DISPLAYS

This chapter describes the device control panel and displays, patient circuit connections, and rear panel
connections.

4.1

CONTROLS

AND

DISPLAYS
Display
Screen

Heated
Humidifier
Button

Start/Stop
Button
Ramp
Button

FIGURE 4-1 CONTROL PANEL


Figure 4-1 illustrates the device control panel, which includes:

A display screen where all device settings appear

Control buttons

4.1.1 DISPLAY SCREEN


The display screen shows operating parameters, instructions, and messages. Figure 4-2 shows the
devices display screen.

FIGURE 42 DISPLAY SCREEN


The information shown on the display screen is dened as follows:
TEXT

DESCRIPTION

ALERT

Indicates that the device requires user attention as


indicated on the screen.

AUTO

Indicates that Auto mode is active.

CARD

Indicates that a SmartCard is inserted and detected.

BiPAP Auto with Bi-Flex Provider Manual

4-1

TEXT

4-2

DESCRIPTION

cm H2O

Indicates that the alphanumeric digits are displaying


a pressure value.

EPAP

Indicates that the EPAP pressure or Max EPAP pressure is being displayed.

Erase

Indicates that the user may clear the Therapy Session


Counter.

FLEX

Indicates that a Bi-Flex comfort setting is being displayed or that Bi-Flex is active.

FOSQ

Indicates that the user may begin the FOSQ test or


that the FOSQ test is active.

HEAT

Indicates that the humidier is turned on and/or its


setting is displayed.

HOURS

Indicates that the Therapy Time is being displayed.

IPAP

Indicates that the IPAP pressure or Max IPAP pressure


is being displayed.

Light

Indicates that the control panel LED backlight setting


is being displayed or is active.

Max IPAP/Max PS

Indicates that the Max IPAP setting or Max PS setting


is being displayed.

Min EPAP

Indicates that the Min EPAP setting is being displayed.

NIGHTS

Indicates that the session counter is being displayed.

PATIENT

Indicates that a Patient Disconnect alert is active.

PS

Indicates that the Pressure Support setting is being


displayed.

RAMP

Indicates that the ramp function is in progress.

RAMP START

Indicates that the ramp starting pressure is being


diplayed.

RISE

Indicates that a rise time setting is being displayed.

Setup

Indicates that the device is in Provider mode and not


in User mode.

BiPAP Auto with Bi-Flex Provider Manual

4.1.2 CONTROL BUTTONS


The device control buttons, shown in Figure 4-3, are dened below.

Display
Screen

Heated
Humidifier
Button

User
Buttons

Start/Stop
Button

Ramp
Button

FIGURE 43 CONTROL BUTTONS


BUTTON

DESCRIPTION

START/STOP

This button starts or stops the units airow. Press the button in
to turn the airow on and put the device in the Operate state.
When the button is turned off, the device is in the Standby state.
When in Standby, any ramp in progress is terminated, the alerts
are reset (except for the System Errors alert), and the humidier is turned off. This button is also used to exit the parameter
screens.

HEAT

When the optional REMstar Heated Humidier is prescribed,


this button controls the humidiers heater plate setting. Follow the instructions provided with the humidier. You can also
use this button to adjust the settings shown in the user menu
screens.

RAMP

When the airow is turned on and the ramp function is enabled,


this button lowers the airow pressure, allowing you to fall
asleep more easily. You can also use this button to adjust the settings shown in the user menu screens.

USER

The left and right user buttons allow you to navigate the display
screens.

NOTE:

Additionally, you can press any of the above buttons to clear the patient disconnect alert or to silence the high priority alert.

BiPAP Auto with Bi-Flex Provider Manual

4-3

4.1.3 AUDIBLE ALERTS

AND INDICATORS

The device provides audible and visual indicators of alert conditions. These indicators occur in the
following situations:

Power On All visual indicators appear momentarily, and the audible indicator briey sounds to
signify that all indicators are functioning properly.

Continuous alerts An alert sounds continuously.

High priority alerts An alert consisting of three beeps, a pause, and then two more beeps sounds
several times at intervals for a high priority alert.

Medium priority alerts An alert sounds repeatedly, with a brief interval between alerts for a
medium priority alert.

Low priority system alerts An alert sounds repeatedly, with a longer interval between alerts for a
low priority alert.

Provider mode An alert sounds when the provider mode is accessed using the button sequence
described in Section 3.6.1.

SmartCard activity An alert sounds once when the SmartCard is inserted or removed.

SmartCard error A medium priority alert sounds and the CARD icon and an error code appear
on the display when a card error occurs.

Patient Disconnect A low priority alert sounds and the ALERT and PATIENT icons ash on
the display.

Conrmation An alert sounds once when the humidier setting screen is entered and when the
humidier is turned on.

4.2

NAVIGATING

THE

SCREENS

Note the following when navigating the Provider or User mode screens:

4-4

The Left and Right User buttons allow you to go to the previous or next setting, respectively.

The Heat and Ramp buttons operate as up and down buttons to adjust the settings. Pressing and
holding the Heat or Ramp button down for at least 2 seconds will change the settings at a faster
rate.

The Start/Stop button allows you to exit a provider or user mode screen.

The alphanumeric digits and icons ash to indicate setting adjustment.

BiPAP Auto with Bi-Flex Provider Manual

4.2.1 LED BACKLIGHT

FOR

BUTTONS

The Start/Stop, Ramp, and Heat buttons can be lit by an LED backlight. The LED backlight is on
when the device is in the Standby state. When the device is in the Operate state, the LED backlight is
lit according to the setting in the LED backlight setting. The LED backlight may ash to indicate an
alert condition.

CONTROL PAD INACTIVITY


Some screens have timeout periods. The screens timer starts when the screen is initially displayed, and
is restarted whenever a button is pressed. When a screen times out, the Monitoring/Standby screen is
displayed.

4.3

PATIENT CIRCUIT CONNECTION

The patient circuit is connected to the breathing circuit connection shown in Figure 44. The breathing circuit connection accepts a bacteria lter or a tubing connector for reusable or disposable tubing.
Breathing Circuit Connection

FIGURE 44 BREATHING CIRCUIT CONNECTION

4.4

REAR PANEL

Figure 45 shows the rear panel of the BiPAP Auto with Bi-Flex.

FIGURE 45 REAR PANEL


NOTE:

The SmartCard Connector is located on the side of the BiPAP Auto with BiFlex unit.

BiPAP Auto with Bi-Flex Provider Manual

4-5

The BiPAP Auto with Bi-Flex rear panel contains the following:

A power inlet for connecting either the external AC power supply or a DC power adapter.

The lter cap that is removed to inspect the inlet air lters.

4.5

SMARTCARD

The BiPAP Auto with Bi-Flex is delivered with the SmartCard installed. The SmartCard is a plastic
card similar in size and shape to a normal credit card. However, instead of holding information on a
magnetic strip, it holds it in a small silicon chip embedded in the card. When installed into a BiPAP
Auto with Bi-Flex device, the SmartCard records the following data:

Date

Time

Leak

Pressure

Apnea/Hypopnea events

Non-Responsive Apnea/Hypopnea events (Auto Bi-level & Auto Bi-level with Bi-Flex modes)

Snoring

Duration of each use (minimum storage capacity: six months)

Flow Limitation (Auto Bi-level & Auto Bi-level with Bi-Flex modes)

When capacity is reached, the oldest data is overwritten. Using the Respironics SmartCard reader/writer and the Encore Pro software, you can download and view the usage data. Follow the instructions
included with the Encore Pro software to download the data.
NOTE:

If the card is not installed, the device usage will not be recorded. When a
SmartCard is installed, the word CARD appears in the upper right corner of
the display screen.

The SmartCard can also be programmed with the patients prescription by using applicable Respironics software and the SmartCard reader/writer (available from Respironics). When the programmed
SmartCard is inserted into the BiPAP Auto with Bi-Flex, the prescription settings are automatically
transferred into the device and erased from the SmartCard. The SmartCard will then start to collect
patient data.

4-6

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 5: SETUP
5.1

PREPARING

THE

DEVICE

This section contains information on:

Installing the air lters

Assembling the patient circuit

Supplying power to the device

Startup

Entering Provider mode

5.1.1 INSTALLING

THE

AIR FILTERS

The device uses one or two removable lters at the air inlet. You must install the gray foam lter before
operating the device. The foam lter is washable and reusable. The disposable white ultra-ne lter is
optional.
CAUTION:

A properly installed, undamaged foam lter is required for proper operation.

Reusable Gray
Foam Filter
(required)
Filter
Cap
Disposable Ultra-fine
Filter (optional)

FIGURE 5-1 INSTALLING

THE

AIR FILTERS

To install the air lter, complete the following steps:


1.

Place the gray foam lter on top of the ultra-ne lter (if using the ultra-ne lter).

2.

Slide the lters into the air inlet at the rear of the device, and push them down into the recess as
shown in Figure 5-1.

3.

Place the bottom of the lter cap into the bottom of the air inlet opening. Position the cap so that
the small opening on the cap is facing down.

4.

Insert the caps bottom tabs into the opening below the lter area. Snap the cap into place.

See Chapter 8 to clean or replace the lters.


NOTE:

The lter cap should be installed with the air inlet opening at the bottom.

BiPAP Auto with Bi-Flex Provider Manual

5-1

5.1.2 ASSEMBLING
WARNING:

THE

PATIENT CIRCUIT

The exhalation device (e.g., the Whisper Swivel II) or exhalation port
(on masks with an integrated exhalation port) is designed to exhaust
CO2 from the patient circuit. Do not block or seal the ports on the
exhalation device.

1.

Assemble the patient circuit according to the congurations described in Chapter 10.

2.

If required, connect a bacteria lter to the patient interface port (shown in Figure 5-2), and connect the patient tubing to the outlet of the bacteria lter.

If the bacteria lter is not required, connect the patient tubing directly to the breathing circuit
connection.

If you are using a humidier, connect the inlet to the bacteria lter outlet or to the breathing
circuit connection.

A completed assembly appears in Figure 5-2.


NOTE:

A bacteria lter should be used to prevent contamination if the device is


intended for use by multiple patients.
Patient Interface

Exhalation Port

Breathing
Circuit
Connection

FIGURE 5-2 AN EXAMPLE

5.1.3 SUPPLYING POWER

5-2

TO THE

OF A

TYPICAL CIRCUIT

DEVICE

WARNING:

The device can operate on AC or DC power. The DC power option is


not intended as a battery backup when using AC power.

WARNING:

Route the wires to avoid tripping.

BiPAP Auto with Bi-Flex Provider Manual

AC OPERATION
WARNING:

For proper use, the external AC power supply must be placed feet
down, in the upright position, as shown in Figure 53.

1.

Plug the pronged end of the AC power supplys cord into an electrical outlet that is not controlled
by a wall switch.

2.

The external AC power supply features a cord retainer to provide strain relief for the AC power
cord. Wrap the cord around the AC power supplys cord retainer, using the wire tie supplied with
your power supply.

3.

Leaving a small amount of slack in the cord, connect the cord on the other side of the power
supply to the power inlet on the device. The power cord has a locking connector. To properly plug
the cord in:

4.

a.

Pull the locking mechanism back.

b.

Push the connector into place.

c.

Release the lock.

Ensure that all connections are secure.

FIGURE 5-3 USING


NOTE:

THE

EXTERNAL POWER SUPPLY

If you need to disconnect the power cord from the device, slide the locking
connector back and then remove the power cord.

DC OPERATION
You can operate the device on DC power by using the Respironics DC power adapter accessory (when
available). See the DC power adapter instructions for information on how to operate the device using
DC power.
CAUTION:

When DC power is obtained from a vehicle battery, the device should


not be used while the vehicles engine is running. Damage to the vehicle or the device may occur.

CAUTION:

Only use the Respironics DC power adapter available from your home
care provider. Use of any other system may cause damage to the device or the vehicle.

BiPAP Auto with Bi-Flex Provider Manual

5-3

5.1.4 STARTUP
When the power cord is plugged into an AC or DC power source, the device sounds a conrmation
alarm, and the control panel buttons light up.
NOTE:

1.

If the alarm does not sound or the buttons do not light up, the device requires
servicing. Additionally, if any of the alphanumeric digits shown in Figure 54
do not display on the Self Test screen, the device requires servicing.

The rst screen to appear is the Self Test screen:

FIGURE 5-4 SELF TEST SCREEN


2.

The next screen displays the software version:

FIGURE 5-5 SOFTWARE VERSION SCREEN


3.

The Blower Hours Screen then appears, which displays the Blower Hours Time Meter:

FIGURE 5-6 BLOWER HOURS SCREEN

5-4

NOTE:

The control panel is inactive during these rst three screens.

NOTE:

Each of the rst three screens appears for approximately 1-3 seconds.
BiPAP Auto with Bi-Flex Provider Manual

4.

The next screen to appear is the Standby screen:

FIGURE 5-7 STANDBY SCREEN


The Standby screen appears when the device is in the Standby state and displays the Therapy
Time Meter. Pressing the Start/Stop button puts the device in the Operate state. The Monitoring screen then appears, displaying the current pressure setting:

FIGURE 5-8 MONITORING SCREEN


Both the Monitoring and the Standby screens display the PATIENT, FLEX, and
LIGHT icons if these features are enabled. Additionally, the CARD icon displays if a
SmartCard is inserted, and the AUTO icon displays if the device is in Auto Bi-level
or Auto Bi-level with Bi-Flex mode. The HEAT icon also displays if the humidier is
turned on. The RAMP icon is displayed in Monitoring Screen if ramp is active.
5.

When in the Standby or Monitoring screens, you can modify the humidier setting
by pressing and holding the Heat button until the screen below appears (Figure
59).

FIGURE 59 HUMIDIFIER SETTING SCREEN


You can increase or decrease the humidier setting from 1 to 5 in increments of 1. The
setting changes immediately as you adjust it. When the START/STOP button is pressed
to exit this screen, the Operate/Standby state does not change.
BiPAP Auto with Bi-Flex Provider Manual

5-5

5.1.5 ENTERING PROVIDER MODE


Enter Provider mode by completing the following steps:
1.

Unplug the device if it is already plugged in.

2.

Hold down the Left and Right User buttons simultaneously.

3.

Plug the device in. The SETUP icon appears in the upper right corner of the screen to indicate
that the device is in Provider mode. Release the Left and Right User buttons.

IMPORTANT: Prescribed therapy settings can only be set in Provider mode. To


prevent patients from tampering with the settings, do not show them
how to access Provider mode.

5.2
NOTE:

CONNECTING

THE

PATIENT

Before connecting the patient to the device, check the integrity of the patient
circuit, exhalation port, and alarms.

1.

Turn the devices airow on by pressing the Start/Stop button.

2.

If oxygen is being used, turn on the oxygen ow. Make sure you place the Respironics Pressure
Valve (Part Number 302418) in-line with the patient circuit.

WARNING:
Always turn the airow on before turning on the oxygen, and always
turn the oxygen off before turning off the airow.
3.

Place the mask on the patient.

5.3

SETTING

UP THE

SMARTCARD

This section provides information on:

Downloading data

Programming a SmartCard

Changing Settings using a SmartCard

5.3.1 DOWNLOADING DATA


You can download data from the SmartCard by completing the following steps:
1.

Connect a Respironics SmartCard reader/writer directly to a Windows-compatible computer following the instructions included with the reader/writer. Remove the SmartCard from the device
and insert it into the reader/writer.

2.

Follow the instructions included with your Encore Pro software to download the data.

WARNING:
5-6

Any IEC 60950 device must be connected through the 7-pin mini-din
connector with a Respironics-supplied isolation cable.
BiPAP Auto with Bi-Flex Provider Manual

5.3.2 PROGRAMMING

SMARTCARD

1.

Connect a Respironics SmartCard reader/writer directly to a Windows-compatible computer following the instructions included with the reader/writer. Remove the SmartCard from the device
and insert it into the reader/writer.

2.

Follow the instructions included with your Encore Pro software to program the SmartCard.

3.

Remove the SmartCard from the reader/writer. If desired, write the patients name on the front of
the card.

5.3.3 CHANGING SETTINGS USING

SMARTCARD

To change the settings in the device using a programmed SmartCard:


1.

Make sure the device is plugged in. Insert the programmed SmartCard into the slot on the left
side of the device (logo side facing up). When the Monitoring or Standby screen displays, the
word CARD appears in the upper right corner indicating that the card is inserted correctly.

2.

Turn the airow on to verify the new prescription setting. The card can now be removed or you
can leave the card in the device to record device usage. Once the prescription settings have been
transferred to the device, they will be deleted from the SmartCard.

BiPAP Auto with Bi-Flex Provider Manual

5-7

5-8

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 6: CHANGING SETTINGS


This chapter describes the settings that can be changed when the device is in the Provider and User
modes.

6.1

CHANGING SETTINGS

IN THE

PROVIDER MODE

Accessing the Provider mode setup level unlocks additional settings that cannot be changed while in
User mode.

6.1.1 PROVIDER MODE NAVIGATION


Figure 61 shows how to navigate the Provider mode screens using the Left and Right User buttons.
The parameter icon and setting will ash.
Mode Setting Screen

Right User
Button

Left User
Button

Rise Time Setting Screen


Right User
Button

Only displayed if the device


is in Bi-level mode and IPAP does not
equal EP
EPAP, or if the device is in
Auto Bi-level mode.

Left User
Button

IP
IPAP
Setting Screen

Right User
Button

Only displayed if the device


is in Bi-level or Bi-level with
Bi-Flex mode.
Right User
Button

Ramp Length Setting Screen

Left User
Button

Only displayed if the device is in


Bi-level or Bi-level with Bi-Flex mode.

EP
EPAP
Setting Screen
Only displayed if the device
is in Bi-level or Bi-level with
Bi-Flex mode.

Right User
Button

Right User
Button

Ramp Start Pressure


Setting Screen

Right User
Button

Only displayed if the device


is in Auto Bi-level or Auto
Bi-level with Bi-Flex mode.

Only displayed if the device


is in Auto Bi-level or Auto
Bi-level with Bi-Flex mode.

Right User
Button

Left User
Button

LED Backlight
Setting Screen

Left User
Button

Flex Setting Screen

Right User
Button

BiPAP Auto with Bi-Flex Provider Manual

Left User
Button

Reset Session Counter


Setting Screen

Left User
Button

FIGURE 61 NAVIGATING

Only displayed if the device is in


Auto Bi-level or Auto Bi-level with
Bi-Flex mode.

Patient Disconnect Alert


Setting Screen

Only displayed if the device


is in Auto Bi-level or Auto
Bi-level with Bi-Flex mode.

Right User
Button

Left User
Button

Left User
Button

Max PS Setting Screen

Only displayed if the device


is in Bi-level with Bi-Flex or
Auto Bi-level with Bi-Flex
mode.

Only displayed if the device is in


Bi-level or Bi-level with Bi-Flex
mode and the Ramp Length Setting
is between 5 and 45 minutes.

Split Night Time


Setting Screen

Right User
Button

Right User
Button

Left User
Button

Left User
Button

Max IP
IPAP Setting Screen

Right User
Button

Left User
Button

Left User
Button

Min EP
EPAP Setting Screen

Right User
Button

Left User
Button

THE

PROVIDER MODE SCREENS


6-1

6.1.1.1
1.

CHANGING

THE

PROVIDER MODE SETTINGS

Mode Setting Screen

For BiPAP Auto with Bi-Flex devices, the Mode Setting screen appears after the Software Version
screen when you rst enter Provider mode. This screen, shown in Figure 62, allows you to select
the therapy mode. The available modes are: Bi-level, Bi-level with Bi-Flex, Auto Bi-level, and Auto
Bi-level with Bi-ex (bPAP, bFLE, AbFLE, and AbPAP on the display).
To change the selection, press the Heat or Ramp buttons until the correct setting appears. The icons
ash when selected to indicate the current therapy.

FIGURE 62 MODE SETTING SCREEN


Press the Right User button to access the Pressure Setting screens.
2.

Bi-level and Bi-level with Bi-Flex Pressure Settings

NOTE:

These screens are only displayed if the device is in Bi-level or Bi-level with BiFlex mode.

NOTE:

IPAP cannot be set lower than EPAP, and EPAP cannot be set higher than
IPAP.

a.

IPAP Setting Screen


The IPAP Setting screen is shown in Figure 63. It allows you to modify the IPAP setting.

FIGURE 63 IPAP SETTING SCREEN


Increase or decrease the IPAP pressure by pressing the Heat and Ramp buttons until the correct pressure appears. You can adjust the pressure in 0.5 cm H2O increments. The setting can be adjusted from
4 to 25 cm H2O.
Press the Right User button to access the EPAP pressure setting.
6-2

BiPAP Auto with Bi-Flex Provider Manual

b.

EPAP Setting Screen


The EPAP Setting screen is shown in gure 6-4. It allows you to modify the EPAP setting.

FIGURE 64 EPAP SETTING SCREEN


Increase or decrease the EPAP pressure by pressing the Heat and Ramp buttons until the correct pressure appears. You can adjust the pressure in 0.5 cm H2O increments. The setting can
be adjusted from 4 to 25 cm H2O.
WARNING:

High EPAP pressures can cause the patient discomfort. Carefully evaluate the patient if you set the EPAP level above 15 cm H2O.

If the EPAP is set to less than the ramp start pressure, the ramp start pressure automatically
sets to the EPAP.
Press the Right User button to access the Flex setting (if in Bi-level with Bi-Flex mode) or the
Rise Time setting (if in Bi-level mode).
3.

Auto Bi-level and Auto Bi-level with Bi-Flex Pressure Settings

NOTE:

These screens are only displayed if the device is in Bi-level or Bi-level with BiFlex mode.

NOTE:

The settings for Min EPAP, Max IPAP, and Max PS are highly interdependent.
Consider the range of allowable pressures as bounded at the top by the Max
IPAP setting and at the bottom by the Min EPAP setting. While setting up one
parameter, another may change according to the following set of rules:

The device will not allow a Max PS setting outside the range of 3.0 to
8.0 cm H2O,
O, and the device will prohibit other settings to ensure that
this rule is not violated.

Within the range of 3.0 to 8.0 cm H2O,


O, any setting can be chosen for
Max PS, but it is further subject to the following rule: The Max PS setting cannot exceed Max IPAP setting minus Min EPAP setting.

If the Min EPAP setting is incremented greater than Max IPAP setting
minus Max PS setting, it will force the Max PS to decrease. The Max PS
setting will only decrease down to the minimum setting of 3.0
cm H2O,
O, so Min EPAP setting cannot be incremented greater than Max
IPAP setting minus 3.0.

If the Max IPAP setting is decremented less than Min EPAP setting plus
Max PS setting, it will force the Max PS to decrease. The Max PS setting
will only decrease down to the minimum setting of 3.0 cm H2O, so
Max IPAP setting cannot be decremented less than Min EPAP setting
plus 3.0.
BiPAP Auto with Bi-Flex Provider Manual
6-3

a.

Min EPAP Setting Screen


This screen (shown in Figure 6-5) allows you to modify the Auto Min EPAP setting. You can
adjust the Min EPAP setting in 0.5 cm H2O increments by pressing the Heat and Ramp buttons. The setting can be adjusted from 4 to 22 cm H2O.

FIGURE 65 MIN EPAP SETTING SCREEN


Press the Right User button to access the Max IPAP setting.
b.

Max IPAP Setting Screen


This screen (shown in Figure 6-6) allows you to modify the Auto Max IPAP setting. You can
adjust the Max IPAP setting in 0.5 cm H2O increments by pressing the Heat and Ramp buttons. The setting can be adjusted from 7 to 25 cm H2O.

FIGURE 66 MAX IPAP SETTING SCREEN


Press the Right User button to access the Max PS setting.
c.

Max PS Setting Screen


This screen (shown in Figure 6-7) allows you to modify the Auto Max PS setting. You can
adjust the Max PS setting from 3.0 to 8.0 cm H2O. Press the Heat and Ramp keys to increase
or decrease the setting in 0.5 cm H2O increments.

FIGURE 67 MAX PS SETTING SCREEN


Press the Right User button to access the Flex setting (if in Auto Bi-level with Bi-Flex mode)
or the Rise Time setting (if in Auto Bi-level mode).
6-4

BiPAP Auto with Bi-Flex Provider Manual

4.

Flex Setting Screen


The Flex Setting screen, shown in Figure 68, allows you to adjust the level of air pressure
relief that the patient feels when he or she exhales during therapy. It is only displayed when
the device is in Bi-level with Bi-Flex or Auto Bi-level with Bi-Flex mode.
To change the selection, press the Heat or Ramp buttons until the correct setting appears.
The settings range from 1-3. It is recommended that you start with the minimum setting of
1, which provides the least relief. Levels 2 and 3 progressively reect increased pressure relief.

FIGURE 68 FLEX SETTING SCREEN


Press the Right User button to access the Ramp Length setting (if in Bi-level with Bi-Flex
mode) or the Split Night Time setting (if in Auto Bi-level with Bi-Flex mode).
5.

Rise Time Setting Screen


Rise Time Setting screen is shown in Figure 69. Rise time is the time it takes for the device
to change from EPAP to IPAP. This screen allows you to adjust the rise time so you can nd
the most comfortable setting for the patient. The Rise Time Setting screen only displays if the
device is in Bi-level mode and IPAP is not equal to EPAP, or if the device is in Auto Bi-level
mode.

FIGURE 69 RISE TIME SETTING SCREEN


Increase or decrease the rise time setting from 0 to 3 by pressing the Heat and Ramp buttons
until you nd the right setting. The rise time of 0 to 3 corresponds to tenths of a second (e.g.,
a setting of 0 equals 0.1 second rise time, 1 = 0.2 seconds, etc.).
Press the Right User button to access the Ramp Length setting (if in Bi-level mode) or the
Split Night Time setting (if in Auto Bi-level mode).

BiPAP Auto with Bi-Flex Provider Manual

6-5

6.

Ramp Length Setting Screen


The Ramp Length Setting screen, shown in Figure 610, allows you to change the ramp time.

FIGURE 610 RAMP LENGTH SETTING SCREEN


NOTE:

The Ramp length setting screen only displays if the device is in Bi-level or Bilevel with Bi-Flex mode.

To change the ramp time, press the Heat and Ramp buttons until the correct time appears. The
setting increases or decreases from 0 to 45 minutes in 5 minute increments. If you do not want
ramp, set the time to zero.
Press the Right User button to access the ramp start pressure setting (if ramp length is between 5
and 45 minutes) or the Patient Disconnect setting (if ramp length is set to 0).
7.

Ramp Start Pressure Setting Screen


The Ramp Start Pressure Setting screen is shown in Figure 611.

NOTE:

This screen only displays if the device is in Bi-level or Bi-level with Bi-Flex
mode and the ramp length setting is greater than zero.

FIGURE 611 RAMP START PRESSURE SETTING SCREEN


To change the ramp starting pressure, press the Heat and Ramp buttons until the correct pressure
appears. The setting increases or decreases in 0.5 cm H2O increments. The user can adjust the
setting from 4 cm H2O to the EPAP pressure setting.
Press the Right User button to access the Patient Disconnect setting.

6-6

BiPAP Auto with Bi-Flex Provider Manual

8.

Split Night Time Setting Screen


The Split Night Time Setting screen is shown in Figure 612. This screen allows you to modify
the setting for Split Night Time, which indicates the amount of time spent in Bi-level or Bi-level
with Bi-Flex mode before transitioning to Auto Bi-level or Auto Bi-level with Bi-Flex mode. It
can be set to 0 (o) and from 2 to 4 hours in 1 hour increments when the Heat or Ramp button
is pressed.

FIGURE 612 SPLIT NIGHT TIME SETTING SCREEN


NOTE:

This screen is only displayed if the device is in the Auto Bi-level or Auto Bilevel with Bi-Flex mode.

Press the Right User button to access the Patient Disconnect setting.
9.

Patient Disconnect Setting Screen

The Patient Disconnect Setting screen is shown in Figure 613.

FIGURE 613 PATIENT DISCONNECT SETTING SCREEN


This setting enables or disables the audible alert (a beeping sound) when a large, continuous air
leak (such as mask removal) has been detected in the circuit. Press the Heat and Ramp buttons to
select the desired setting to enable or disable the alert. A setting of 0 disables the Patient Disconnect alert, while a setting of 1 enables the alert.
NOTE:

Setting the Patient Disconnect setting to 1 also enables the Auto-Off feature,
which causes the unit to automatically change from the Operate state to the
Standby state whenever the mask is removed from the patients airway.

Press the Right User button to access the LED Backlight Setting screen.

BiPAP Auto with Bi-Flex Provider Manual

6-7

10. LED Backlight Setting Screen


The LED Backlight Setting screen is shown in Figure 614. This setting allows you to have the
control panel lights behind the buttons turned on or o while the airow is turned on and the
device is in the Operate state.
NOTE:

The lights will always be on when the airow is off and the device is in
Standby.

FIGURE 614 LED BACKLIGHT SETTING SCREEN


To change the LED backlight setting, press the Ramp or Heat button until the correct setting appears. 1 means the light is on, while 0 means the light is o.
Press the Right User button to access the Reset Session Counter screen.
14. Reset Session Counter Screen
The Reset Session Counter screen is shown in Figure 615.

FIGURE 615 RESET SESSION COUNTER SETTING SCREEN


This screen displays the number of hours that the device delivered therapy to the patient. The
screen allows you to reset the session counter. Resetting the session counter also resets the Therapy
Time Meter.
To erase the totals and go back to zero, press and hold the Ramp or Heat button. The ERASE
icon displays on the screen. Hold the button down until the time changes to zero and the ERASE
icon disappears. The NIGHTS icon also displays on this screen.

6-8

BiPAP Auto with Bi-Flex Provider Manual

6.2

CHANGING SETTINGS

IN

USER MODE

With the device in User mode, the patient is restricted to viewing the current pressure setting.
The patient can also view and change the following settings in User mode by pressing and holding the
Heat or Ramp button while the device is in Standby:

Session counter (cannot be changed in User Mode)

Flex setting (when Bi-Flex is enabled at the Provider level)

FOSQ Questionnaire

Rise time (when in Bi-level mode and IPAP is not equal to EPAP or when in Auto Bi-level mode)

Ramp start pressure (when in Bi-level or Bi-level with Bi-Flex mode and ramp length is set between 5 and 45 minutes)

Patient disconnect

LED backlight

Additionally, the patient can change the Humidier (heat) setting by pressing and holding the Heat
button to access the Humidier Setting screen.
See the BiPAP Auto with Bi-Flex User Manual for additional information.

BiPAP Auto with Bi-Flex Provider Manual

6-9

6-10

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 7: ALERTS
This chapter describes the device alerts, how to set them, and what corrective actions to take for the
alert conditions.

7.1

INTRODUCTION

The device provides three alert levels: high, medium, and low priority.
ALERT

DESCRIPTION

High Priority

These alerts require immediate operator


response. The alert signal consists of a high
priority sound. The display has the message
ALERT at the top of the screen.

Medium Priority

These alerts require prompt operator response.


The alert signal consists of a medium priority
sound. The display has the message ALERT at
the top of the screen.

Low Priority

These alerts require operator awareness. The


alert signal consists of a low priority sound.
The display has the message ALERT at the top
of the screen.

Some audible alerts are self-cancellable. This means that the alert sound stops when the cause of the
alert is corrected. See section 7.5 for detailed descriptions of the alert indicators and sounds.

7.1.1 OVERVIEW

OF

ALERT BEHAVIOR

Alert conditions are signalled by the device in two ways: a sound and a display message. Each signal
type behaves dierently depending on the type of alert.

7.1.1.1
1.

ALERT SOUNDS BEHAVIOR

High Priority Sounds


There are two possible high priority sounds:

High Priority The sound repeats a pattern of three beeps followed by a pause and then
two more beeps until a button is pressed. This pattern is indicated in Section 7.5 as

Continuous An audible alert sounds continuously. This pattern is indicated in Section


7.5 as

BiPAP Auto with Bi-Flex Provider Manual

7-1

2.

Medium Priority Sound


The medium priority sound repeats a pattern of beeps with a short interval between each beep
until the cause of the alert is corrected.
This pattern is indicated in Section 7.5 as

3.

Low Priority Sound


The low priority sound repeats a pattern of beeps with a longer interval between each beep until a
button is pressed or until the cause of the alert is corrected.
This pattern is indicated in Section 7.5 as

7.1.1.2

DISPLAY BEHAVIOR

For high, medium, and low priority alerts, the display shows ALERT and the name of the alert.

7.2

ALERTS

The device has system, card, and patient alerts.

7.2.1 SYSTEM ERROR ALERT


The System Error alert is a high priority alert. It indicates that there is a problem with the device.
The System Error screen is shown in Figure 71.

FIGURE 71 SYSTEM ERROR SCREEN


A three digit error code displays on the screen, indicating the type of system error (e.g., error code 57
displays as E57). When a system error occurs, the system error audible alert sounds and the blower
and humidier turn o. Pressing any button while the System Error alert is active shuts o the audible
alarm.

7-2

BiPAP Auto with Bi-Flex Provider Manual

7.2.2 CARD ERROR ALERT


The Card Error alert is a medium priority alert. It indicates that a problem exists with the card inserted
in the SmartCard connectivity slot. Removing the SmartCard automatically resets this alert.
The Card Error screen is shown in Figure 72.

FIGURE 72 CARD ERROR SCREEN


A three digit error code displays on the screen, indicating the type of card error (e.g., error code 57 displays as C57). The SmartCard must be removed to exit this screen and return to the previous screen.

7.3

PATIENT DISCONNECT ALERT

Figure 73 shows the Patient Disconnect alert screen.

FIGURE 73 PATIENT DISCONNECT ALERT SCREEN


The Patient Disconnect alert is a low priority alarm. If the Patient Disconnect alert is enabled, the alert
occurs when the patient is disconnected from the device. See Chapter 6 for information on how to
enable or disable the Patient Disconnect alert.
When a patient disconnect alert occurs, the PATIENT and ALERT icons ash on the display. Press
any button to shut o the alert.

BiPAP Auto with Bi-Flex Provider Manual

7-3

7.4

PRESCRIPTION COMPLETE SCREEN

Figure 74 shows the Prescription Complete screen.

FIGURE 74 PRESCRIPTION COMPLETE SCREEN


If a prescription-only SmartCard is inserted, the CARD, ALERT
ALERT, and cm H2O icons ash and a
Card Error audible alert sounds when the prescription has been successfully written to the device. The
SmartCard must be removed to exit this screen and return to the previous screen.

7-4

BiPAP Auto with Bi-Flex Provider Manual

7.5 ALERT SUMMARY TABLE


The following table summarizes the alert information.
Alert

Display
Message

System
Error

ALERT
T icon
flashes and
system error
code ("Exx")
displays

Card Error

CARD icon
flashes and
card error
code ("Cxx")
displays

Patient
Disconnect

ALERT
T and
PATIENT
icons flash

Audible
Indicator

Device Action

Possible Cause

Shuts down and


blower cannot
be restarted.

Device failure.

Operates

Operates

A problem exists
with the Smart
Card inserted in
the SmartCard
connectivity slot.
The card may be
inserted upside
down or
backwards.

Patient circuit is
disconnected or
has a large leak.

Patient Action

Provider Action

Press any button


to silence
the alarm.
Remove power
from the device.
Restore power. If
the alarm
continues to occur,
contact your home
care provider.

Have the device


serviced by either
an authorized
service
representative or
Respironics.

Confirm
that the card is
properly inserted.

Confirm that the


card is properly
inserted.

If the alarm
If the alarm
continues to
continues, replace
occur, remove the the SmartCard.
SmartCard from
the device and
contact your home
care provider.

Press any
button to clear
the alarm.

Reconnect the
patient circuit or
fix the leak.

Reconnect the
patient circuit or
correct the leak.

If the alarm
continues, contact
an authorized
service
If the alarm
representative or
continues, contact Respironics to
your home care
have the device
provider.
serviced.

Prescription
Complete

ALERT,
T, CARD,
T
and cm H2O
icons flash

Operates

Prescription
SmartCard
inserted into
device.

Remove the
SmartCard to
exit this screen
and return to the
previous screen.

Remove the
SmartCard to
exit this screen
and return to the
previous screen.

Audible alert
sounds when
prescription has
been successfully
written to the
device.

BiPAP Auto with Bi-Flex Provider Manual

7-5

7-6

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 8: CLEANING

AND

MAINTENANCE

This chapter describes how to clean the device and its lters and how to maintain the unit.

8.1

CLEANING

THE

DEVICE

WARNING:

To avoid electrical shock, always unplug the device power cord from
the AC (wall outlet) or DC power source before cleaning the device.

CAUTION:

Do not immerse the device in liquid or allow any liquid to enter the
enclosure, inlet lter, or any opening.

WARNING:

If you are using the device on multiple users, discard and replace the
bacteria lter each time the device is used on a different person.

Unplug the device and clean the front panel and exterior of the enclosure as needed using a cloth
dampened with water and a mild detergent. Allow the device to dry completely before plugging in the
power cord.
The mask and tubing should be cleaned daily. For details on cleaning your mask and accessories, refer
to the cleaning instructions packaged with the accessories.
CAUTION:

8.2

Dirty inlet lters may cause high operating temperatures that may affect
device performance. Regularly examine the inlet lters as needed for
integrity and cleanliness.

CLEANING

OR

REPLACING

THE INLET

FILTERS

The device uses two removable lters at the air inlet. The gray foam lter is washable and reusable. The
optional white ultra-ne lter is disposable. Under normal usage, clean the gray foam lter at least
once every two weeks and replace it with a new one every six months.
1.

If the device is operating, stop the airow by pressing the Start/Stop button. Disconnect the
device from the power source.

2.

Remove the lter cover by gently pressing down on the two latches on top of the lter cover (Figure 81).

FIGURE 81 REMOVING
BiPAP Auto with Bi-Flex Provider Manual

THE

FILTER COVER
8-1

3.

Remove the lters from the enclosure by gently pulling around the edges of the lters. The top
lter is the reusable gray foam lter. The bottom lter is the optional disposable white ultra-ne
lter (Figure 82).

Reusable Gray
Foam Filter

Disposable Ultra-fine
Filter
FIGURE 82 REMOVING

FILTERS

4.

Examine the lters regularly for cleanliness and integrity.

5.

If needed, wash the gray foam lter in warm water with a mild detergent. Rinse thoroughly to
remove all detergent residue. Allow the lter to dry completely before reinstalling it. If the foam
lter is torn, replace it. (Only Respironics-supplied lters should be used as replacement lters.)

6.

If the ultra-ne lter is dirty or torn, replace it.

7.

Reinstall the lters, with the ultra-ne lter on the bottom. Slide the lters into the air inlet at the
rear of the device and push them down into the recess.

CAUTION:

8.

8-2

THE

Never install a wet lter into the device. It is recommended that you
clean the lter in the morning and alternate using the two foam lters
provided with the system to ensure sufcient drying time for the
cleaned lter.

Reinstall the lter cover.

BiPAP Auto with Bi-Flex Provider Manual

8.3

CLEANING

AND

DISINFECTION

FOR

MULTIPLE USERS

If you are using the device on multiple users, complete the following steps to clean and disinfect the
unit before each new user.
1.

Unplug the device before disinfecting.

2.

Disinfect the outside of the unit only. Use a cloth with one of the following cleaning agents to
clean the exterior of the device: hydrogen peroxide, 3%; 100% isopropyl alcohol; vinegar, 5%
acidity; water; or household chlorine bleach, 5.25% sodium hypochloride, 1 to 5 part reduction
with water.

3.

Allow the device to dry completely before plugging in the power cord.

WARNING:

Do not immerse the device in liquid or allow any liquid to enter the
enclosure, inlet lter, or any opening. This may result in equipment
damage.

WARNING:

If you are using the device on multiple users, discard and replace the
bacteria lter every time the device is used on a different person.

8.4

MAINTENANCE

WARNING:
Electrical cords or cables should be periodically inspected for damage
or signs of wear.
See the BiPAP Auto with Bi-Flex Service Manual for recommended periodic maintenance.

BiPAP Auto with Bi-Flex Provider Manual

8-3

8-4

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 9: ADDING SUPPLEMENTAL OXYGEN


Oxygen may be added at the mask connection. Please note the warnings listed below when using
oxygen with the device.
WARNING:

The oxygen supply must comply with local regulations for oxygen use.

WARNING:

When using oxygen with this system, a Respironics Pressure Valve (Part
number 302418) must be placed in-line with the patient circuit. Place
the valve in-line with the patient circuit and connect the oxygen as
shown here:
To Controlled
Oxygen Source

O2 Enrichment
Attachment

FIGURE 91 USING

THE

RESPIRONICS PRESSURE VALVE

WARNING:

Oxygen should be administered only on the order of a physician.

WARNING:

Supplemental oxygen should not be added to the breathing circuit by


placing the source where the gas will be entrained through the inlet
lter on the rear of the device.

WARNING:

Continuous patient monitoring is recommended while administering


oxygen. Patient monitoring should consist of, at a minimum, patient
observation and pulse oximetry. Arterial blood gas measurements
should be used when necessary.

WARNING:

If administering xed-ow supplemental oxygen, the oxygen concentration may not be constant. The inspired oxygen concentration will
vary, depending on the IPAP and EPAP settings, patient breathing pattern, and the leak rate. Substantial leaks around the mask may reduce
the inspired oxygen concentration to less than the expected concentrations shown in section 9.2. Appropriate patient monitoring should be
implemented.

WARNING:

When using oxygen with this system, turn the device on before turning the oxygen on. Turn the oxygen off before turning the device off.
This will prevent oxygen accumulation in the device.

BiPAP Auto with Bi-Flex Provider Manual

9-1

WARNING:
Oxygen accelerates res. Keep the device and the oxygen container
away from heat, open ames, any oily substance, or other sources of ignition.
Do not smoke in the area near the device or the oxygen.

9.1

ADDING SUPPLEMENTAL OXYGEN

The delivered oxygen concentration varies with changes in ow in the circuit. The following may have
an impact on oxygen concentration:

Pressure settings

Patient Tidal Volume

Peak Inspiratory Flow

I:E Ratio

Respiratory rate

Circuit leak rate

Oxygen ow rate

To add oxygen to the circuit, the oxygen supply must comply with the local regulations for medical oxygen. The oxygen ow into the oxygen valve cannot exceed 15 L/min and the pressure cannot
exceed 50 psi.

9.2

SUPPLEMENTAL OXYGEN CONCENTRATIONS

Figures 92 and 93 illustrate the potential range of oxygen concentration available to the patient at
a given tidal volume, supplemental oxygen ow, and pressure setting. These gures represent bench
test results without inadvertent mask leaks when oxygen is administered at the mask. Substantial leaks
around the mask may reduce the expected oxygen concentration to below the levels shown in Figures
92 and 93. This guideline may be used as a starting point for initiating oxygen therapy. Oxygen
ow should be gradually adjusted until the patients oxygen needs are adequately met.

9-2

BiPAP Auto with Bi-Flex Provider Manual

FIGURE 92 OXYGEN CONCENTRATION

FOR

600

ML

TIDAL VOLUME

90

Oxygen Concentration (%)

80
70
60
50
40
30
20
10
0

10/5

20/5

30/5

IPAP/EPAP
IPAP/EP
AP/EPAP Settings
1000 ml Tidal
idal V
Volume. Respiratory Rate: 20 BPM

Supplemental

FIGURE 93 OXYGEN CONCENTRATION

BiPAP Auto with Bi-Flex Provider Manual

FOR

1000

ML

TIDAL VOLUME

9-3

9-4

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 10: CIRCUITS

AND

ACCESSORIES

This chapter details the Respironics-approved breathing circuit and accessories.


WARNING:

Refer to each accessorys instruction sheets for the applicable warnings,


cautions, and notes.

10.1 CIRCUIT CONFIGURATIONS


WARNING:

Do not connect any equipment to the device unless recommended by


Respironics or the health care professional. Verify that an exhalation
port is present to exhaust CO2 from the circuit. If circuit accessories
other than those recommended by Respironics are connected to the
device, then pressures must be veried. Use of these accessories may
alter the pressure received, reducing the effectiveness of treatment.

The device is intended for use with Respironics-approved patient circuits.


Typical components are:

Bacteria lter (optional)

22 mm reusable circuit tubing

Exhalation device

Respironics patient interface (e.g., mask)

Respironics Pressure Valve (Part Number 302418), if adding supplemental oxygen

Humidier (optional)

Additional accessories may be added to the circuit to meet specic needs.

10.2 CIRCUITS
1.

2.

3.

AND

ACCESSORIES

Reusable or Disposable Circuit

Reusable smooth inner lumen circuit tubing and exhalation port

Disposable smooth inner lumen circuit tubing and exhalation port

Circuit Accessories

6 disposable circuit tubing

18 disposable circuit tubing

72 disposable circuit tubing

O2 enrichment attachment

Bacteria lter

Device Accessories

DC power adapter, when available

Respironics Communication cable

BiPAP Auto with Bi-Flex Provider Manual

10-1

10.3 MASKS, EXHALATION PORTS,


1.

RELATED ACCESSORIES

Masks

2.

AND

Respironics mask with built-in exhalation port or Respironics mask with separate exhalation
device

Accessories

Disposable headgear

Reusable headgear

Chin strap

10.4 HUMIDIFIERS
1.

Respironics REMstar Heated humidier

2.

Respironics REMstar Integrated humidier

When using other humidiers, verify that the delivered pressure is correct and that proper therapy is
being delivered.
NOTE:

Refer to the humidiers instructions for information on how to set up the


device with the humidier.

10.5 SOFTWARE
Respironics Encore Pro Data Management software for reading compliance data and setting up a
prescription via a SmartCard.

10-2

BiPAP Auto with Bi-Flex Provider Manual

CHAPTER 11: SPECIFICATIONS


ENVIRONMENTAL
Operating

Storage

Temperature

41 F to 95 F

-4 F to 140 F

Relative Humidity

15 to 95%
(non-condensing)

15 to 95%
(non-condensing)

Atmospheric Pressure

83 to 102kPa

(5600 feet to sea level)

PHYSICAL
Dimensions:

9.75 L x 6.625 W x 4.4 H

Weight: Less than 7 lbs.

ELECTRICAL
AC Voltage Source:

100 to 240 V, 50/60 Hz

DC Voltage Source:

24 V (when operated with the external DC


power adapter accessory)

AC Current:

1.25 A maximum

DC Current:

3.0 A maximum

Protection against electric shock:

Class II

Degree of protection against electric shock:

Type BF Applied Part

Degree of protection against harmful ingress of water: Device: Ordinary Equipment, IPX0
AC Power Supply (Reorder number
1012832): Drip Proof, IPX1
DC Power Adapter (when available):
Drip Proof, IPX1
Modes of Operation:

Continuous

Electromagnetic Compatibility:

The device meets the requirements of


EN 60601-1-2, second edition (2001).

Fuses:

There are no user-replaceable fuses.

BiPAP Auto with Bi-Flex Provider Manual

11-1

PRESSURE
Output:

4 to 25 cm H2O

Pressure Stability:

4.0 to 10.0 cm H2O (0.5 cm H2O)


>10.0 to 25.0 cm H2O (1.0 cm H2O)

Measured in accordance with EN ISO 17510-1 @ 6.6, 13.2, and 20 cm H2O @ 500 ml with BPM
set to 10, 15 & 20 BPM @ 23 C (2.0 C), 50% RH (5%), and an atmospheric pressure of 101.54
kPascals.
Sound Pressure Level:

<45 dB(A)

Measured in accordance with EN ISO 17510-1 @ 10 cm H2O at the patient circuit. This measurement applies to the BiPAP Auto with Bi-Flex with or without the optional REMstar Heated Humidier.
Maximum Flow: <35 LPM
Measured in accordance with EN ISO 17510-1 @ 8, 16.5, and 25 cm H2O @ 500 ml with BPM set
to 10, 15 & 20 BPM @ 23 C (2.0 C), 50% RH (5%), and an atmospheric pressure of 101.54
kPascals.

CONTROL ACCURACY
Parameter

Range

Accuracy

IPAP

4 to 25 cm H2O

1.5 cm H2O from static setpoint*/****

EPAP

4 to 25 cm H2O

1.5 cm H2O from static setpoint*/****

Ramp Duration

0 to 45 minutes

10% of setting

Rise Time

0 to 3**

25%***

* Dynamic pressure accuracy is measured at the patient end of the circuit with a Whisper Swivel II and
varying ow conditions.
** The range of values correspond to tenths of seconds (0 to 3 corresponds to 0.1 to 0.4 seconds).
*** Measured at the patient end of circuit with a Whisper Swivel II exhalation device and no patient
ow.
**** Measured in accordance with EN ISO 17510-1 @ 4, 8, 16.5, and 25 cm H2O @ 500 ml with
BPM set to 10, 15 & 20 BPM @ 23 C (2.0 C), 50% RH (5%), and an atmospheric pressure of
101.54 kPascals.

DISPOSAL
When necessary, dispose of the device and accessories in accordance with local regulations.

11-2

BiPAP Auto with Bi-Flex Provider Manual

PRESSURE DROP VERSUS FLOW

FOR

PATIENT CIRCUITS

The device automatically compensates for pressure drops associated with a 6-foot smooth bore tube.
Additional pressure drop will occur when restrictive elements are added to the patient circuit. The
following graph shows the additional pressure drop when adding: 1. a bacteria lter; 2. a bacteria lter
and a Respironics humidier.
NOTE:

Always use a manometer to verify patient mask pressure.

FIGURE 111 PRESSURE DROP VERSUS FLOW

BiPAP Auto with Bi-Flex Provider Manual

FOR

PATIENT CIRCUITS

11-3

11-4

BiPAP Auto with Bi-Flex Provider Manual

APPENDIX A
GUIDANCE

EMC INFORMATION
AND

MANUFACTURERS DECLARATION - ELECTROMAGNETIC EMISSIONS

This device is intended for use in the electromagnetic environment specied below. The user of this
device should make sure it is used in such an environment.
EMISSIONS TEST

COMPLIANCE

ELECTROMAGNETIC ENVIRONMENT
- GUIDANCE

RF emissions
CISPR 11

Group 1

The device uses RF energy only for


its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.

RF emissions
CISPR 11

Class B

Harmonic emissions
IEC 61000-3-2

Class A

Voltage uctuations/Flicker
emissions
IEC 61000-3-3

Complies

The device is suitable for use in all


establishments, including domestic
establishments and those directly
connected to the public low-voltage power supply network.

BiPAP Auto with Bi-Flex Provider Manual

A-1

GUIDANCE

AND

MANUFACTURERS DECLARATION - ELECTROMAGNETIC IMMUNITY

This device is intended for use in the electromagnetic environment specied below. The user of
this device should make sure it is used in such an environment.
IMMUNITY TEST

Electrostatic
Discharge (ESD)

IEC 60601 TEST


LEVEL

COMPLIANCE LEVEL

+6 kV contact

+6 kV contact

+8 kV air

+8 kV air

Electrical fast
Transient/burst

+2 kV for power
supply lines

+2 kV for supply
mains

IEC 61000-4-4

+1 kV for input-output lines

+1 kV for input/output lines

Surge
IEC 61000-4-5

+1 kV differential
mode

+1 kV differential
mode

+2 kV common
mode

+2 kV for common
mode

<5% UT
(>95% dip in UT) for
0.5 cycle
40% UT
(60% dip in UT) for
5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip
in UT) for 5 sec

<5% UT
(>95% dip in UT) for
0.5 cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip
in UT) for 5 sec

IEC 61000-4-2

Voltage dips, short


interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11

ELECTROMAGNETIC
ENVIRONMENT GUIDANCE
Floors should be
wood, concrete or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical home or hospital environment.
Mains power quality
should be that of a
typical home or hospital environment.
Mains power quality
should be that of a
typical home or hospital environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

A-2

BiPAP Auto with Bi-Flex Provider Manual

GUIDANCE

AND

MANUFACTURERS DECLARATION - ELECTROMAGNETIC IMMUNITY

This device is intended for use in the electromagnetic environment specied below. The user of this
device should make sure it is used in such an environment.
IMMUNITY TEST
Power frequency (50/60
Hz) magnetic eld

IEC 60601
TEST LEVEL

COMPLIANCE
LEVEL

+6 kV contact

+6 kV contact

+8 kV air

+8 kV air

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80
MHz

3 Vrms

Radiated RF
IEC 61000-4-3

3 V/m
80 MHz to 2.5
GHz

3 V/m

IEC 61000-4-8

ELECTROMAGNETIC ENVIRONMENT GUIDANCE


Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic material, the relative humidity should be at least
30%.
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P
d = 2.3 P

80 MHz to 800 MHz


800 MHz to 2.5 GHz

where P is the maximum output power rating


of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site surveya, should be less than the compliance level
in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects, and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the device is
used exceeds the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, the eld strengths should be less than 3 V/m.

BiPAP Auto with Bi-Flex Provider Manual

A-3

RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE


COMMUNICATIONS EQUIPMENT AND THIS DEVICE

AND

MOBILE RF

The device is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of this device can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and this device as recommended below, according to the maximum output power of the
communications equipment.
RATED MAXIMUM
POWER OUTPUT OF
TRANSMITTER
W

SEPARATION DISTANCE ACCORDING

TO

FREQUENCY

OF

TRANSMITTER

150 KHZ TO 80
MHZ
d = 1.2 P

80 MHZ TO 800
MHZ
d = 1.2 P

800 MHZ TO 2.5


GHZ
d = 2.3 P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reection from structures, objects, and people.

A-4

BiPAP Auto with Bi-Flex Provider Manual

LIMITED WARRANTY
Respironics, Inc. warrants that the device shall be free from defects of workmanship and
materials and will perform in accordance with the product specications for a period of
two (2) years from the date of sale by Respironics, Inc. to the dealer. If the product fails to
perform in accordance with the product specications, Respironics, Inc. will repair or replace
at its option the defective material or part. Respironics, Inc. will pay customary freight
charges from Respironics, Inc. to the dealer location only. This warranty does not cover damage caused by accident, misuse, abuse, alteration, and other defects not related to material or
workmanship.
Respironics, Inc. disclaims all liability for economic loss, loss of prots, overhead, or consequential damages which may be claimed to arise from any sale or use of this product. Some
states do not allow the exclusion or limitation of incidental or consequential damages, so the
above limitation or exclusion may not apply to you.
This warranty is given in lieu of all other express warranties. In addition, any implied warranties including any warranty of merchantability or tness for the particular purpose are
limited to two years. Some states do not allow limitations on how long an implied warranty
lasts, so the above limitation may not apply to you. This warranty gives you specic legal
rights, and you may also have other rights which vary from state to state.
To exercise your rights under this warranty, contact your local authorized Respironics, Inc.
dealer or contact Respironics, Inc. at:
1001 Murry Ridge Lane
Murrysville, Pennsylvania 15668-8550
1-724-387-4000

1023859
HM 7/6/05

You might also like