The BeneCheck PLUS Multi-Monitoring System includes glucose, uric acid, and
total cholesterol tests which are all calibrated by J&J DT60 Analyzer (Kodak
EKTACHEM DT60 Analyzer) using venous plasma. The DT60 Analyzer is calibrated
by VITROS DT calibrators which were referenced to the following standard and
method :
Test
Glucose
Uric Acid
Total Cholesterol
Standard
NIST SRM 917b
NIST SRM 913a
NIST SRM 911b
Method
AACC/CDC (Hexokinase/G6PDH)
Uricase/UV
Abell/Kendall
Please read the entire USERS MANUAL carefully before using the BeneCheck PLUS
Multi-Monitoring system to test your blood glucose, uric acid or total cholesterol
levels. This USERS MANUAL guides you through the operation procedures with
illustrated directions and help you find the required information.
WARNING :
Please carefully read the instructions in this users manual and become familiar
with the test procedures before using the system to test blood glucose, uric acid
or total cholesterol levels. Users should consult with a healthcare professional
before going ahead important medical decision.
2
Description
PD-G001-1
PD-A041
PD-D001
PD-E001
PD-A020
PD-F004
PP-E018
PP-E019
PP-E020
PP-E021
PP-E023
PD-A006
PD-A019
PD-A015
Quantity
1
1
1
1
1
1
1
1
1
1
1
1
1
1
3
Item No.
PD-A018
Description
Quantity
Lancets (pack)
1
(Manufacturers of Lancing Device and Lancet : For detail
information, please refer to inside package. )
Optional : ( not included in the standard kit package, please contact your
local distributor for ordering. )
BeneCheck PLUS Glucose Test Strip
BeneCheck PLUS Uric Acid Test Strip
BeneCheck PLUS Total Cholesterol Test Strip
BKP-G-S001
BKP-U-S001
BKP-C-S001
Use by
Keep dry
Read instructions
Caution, consult
accompanying documents
Operation temperature
limitation
EU Representative
Storage temperature
limitation
Catalogue number
(Product number)
CE certification
In-vitro diagnostic
Manufactured by
Lot number
5
Left M Button :
For memory mode operation.
Battery Cover :
For covering the battery.
Battery Socket :
For holding a 3V-lithium
battery
BeneCheck
Tag :
Display information
about the meter.
Battery :
3V lithium battery
(CR2032)
Test Strip
Glucose & Uric Acid Strip
Electronic Contact Bar :
Insert Electronic Contact Bar of the
test strip into the Test Port of the
meter to activate it.
BeneCheck
Sample Inlet :
The inlet where the whole blood or control solution is drawn
in. Touching a blood sample or control solution will result in
a capillary action automatically
Reaction Zone :
Reaction zone of
Total Cholesterol
Sample Inlet :
The inlet where the whole blood or control solution is
drawn in. When a blood sample or control solution touches
this end, the sample will be sucked into the reaction zone
automatically. The reaction zone of total cholesterol strip
full of control solution or a blood sample at least 10 L is
essential.
9
Contact Bar :
Insert this end into
the Test Port of the
meter.
Code Number
Test Mode
Test Mode :
GLUC - Glucose
UA - Uric Acid
CHOL - Total Cholesterol
10
1.
Hold the meter with one hand, slightly pull right the back
cover to remove the back cover.
2.
3.
12
13
After you finish setting the meter, press the right Set button, and the the meter
will show OFF on the screen and power-off.
Note :
Please refer to Chapter 5 for the measurement units setting procedure.
14
1.
Check the code number and test mode on the code strip is
the same as the code number and test mode on the glucose
test strip vial, and check the test mode of the strip.
15
2.
3.
4.
16
Make sure to check that the code number and test mode on
the code strip are the same as the code number and test
mode on the uric acid test strip vial.
Remove the code strip. should appear on the screen,
indicating that the meter has been successfully coded to uric
acid test mode.
5.
Make sure to check the code number and test mode on the
code strip is the same as the code number and test mode on
the total cholesterol test strip vial.
Remove the code strip. should appear on the screen,
indicating that the meter has been successfully coded to
total cholesterol test mode.
Cautions :
If E-E appears on the Screen during coding procedure, there may
be a problem with the code strip or a wrong category code strip
may be misused. Perform the coding procedures again, and if E-E
continues to appear, please contact the authorized distributor for
service.
17
18
Directions :
1.
Insert a test strip into test port, the meter will turn on
automatically.
2.
3.
4.
5.
7 5
1 .6
-1 8 .3
7 4
20
Note :
Please read this users manual and instructions for control solution test
carefully before use.
Do not reuse the test strips and always close the cap of test strip vial tightly
after used.
The first three drops of control solution must be discarded and the vial should
be shaken well before use when the vial is newly opened.
Newly opened vials of control solution and test strips must be marked with an
opening date.
Do not use BeneCheck PLUS Multi-Monitoring System to do any test until you
can get the control test result to fall within the control range listed on the test
strip vial.
The test results fall outside the control range, please reconfirm and repeat the
test, if it still fall outside the range, please contact an authorized distributor
for help and service.
Uric acid control solutions and total cholesterol control solutions are not
included. Please contact an authorized distributor for ordering.
21
22
1.
2.
Fully insert the contact bar of the test strip into the test port
of the meter. The meter will turn on automatically with test
mode display and code number appear on the Screen.
3.
Check the test mode and code number on the screen are
the same as the one on the test strip vial. Icon GLUC, UA
and CHOL indicate measuring mode of glucose, uric acid
and total cholesterol respectively. If not, remove the test
strip and recoding the meter is necessary as described in
Section2.3.
23
Note :
One measurement must be performed within 3 minutes, otherwise the meter
turn off automatically. Please, remove and re-insert the test strip to do the
measurement again.
Do not reuse test strips. Warning Error message E-U will display on the
screen, if a used test strip is inserted.
If the wrong end of the test strip is inserted, or if the test strip is inserted
backwards the meter will not activate.
Warning :
Lancet, test strip cannot be reused.
Please always use certified lancet to ensure safety.
3.2 Testing procedure for Blood Glucose
Directions :
24
1.
2.
3.
Note :
Be sure to wipe off the first drop of blood to avoid body fluid contamination.
25
Note :
The BeneCheck PLUS Multi-Monitoring System includes glucose, uric acid and
total cholesterol test that are all plasma calibrated. The user does not have to
convert it into plasma result.
Use the strips before the expired date, the strips can only be used within 3
months remaining shelf life after the first opening of the vial.
Apply the blood sample onto the correct position of the strip precisely,
follow the directions to familiar with correct operation.
Prohibited applying the blood sample to the test strip twice, it will cause
inaccurate test results.
Special Messages :
Messages
Mode
Glucose
Uric Acid
Total Cholesterol
Lo
< 20 mg/dL (1.11 mmol/L)
<
3 mg/dL (0.18 mmol/L)
< 100 mg/dL (2.59 mmol/L)
Hi
> 600 mg/dL (33.3 mmol/L)
> 20 mg/dL (1.19 mmol/L)
> 400 mg/dL (10.35 mmol/L)
27
28
4. Followed by 14, 21 and 28-day average when press the left M button
repeatedly.
5. After the 28day-average, the latest stored result M01 appears on the screen.
The memory number increases progressively each time the button is pressed
untill M360.
6. Press the left M button for over 3 seconds and the meter will turn off after
a long beep sound. Meter will also automatically shut down after 3 minutes
without any operation.
Note :
If 360 test results are already stored and a new one is added, the oldest test
result is automatically overwrited. The newest result will be shown in M01.
30
31
3. Click the right "Set" button for six times, and a blinking "mg/dL" will display on
the screen.
4. Pressing the M button will switch the unit of measure between "mg/dL" and
"mmol/L" (See Fig.3&4)
5. Click the right Set button once again the meter will turn off.
Note :
Before each test, please make sure correct unit of the measure is set up.
32
Store the meter, code strip, test strips and control solution at 10-30C.
Do not store test strips in high humidity enviroment or dispose under direct sunlight.
Do not freeze or refrigerate the meter and strips.
Keep the meter clean by wiping the exterior appearance of the meter with a
tissues or lint-free cloth.
Handle the strip with clean and dry hands; lancing site for the test should be
cleaned and dried thoroughly.
33
mode icon
Temperature lcon
Date lcon
Time lcon
DAY
Average lcon
Unit lcon
Cause
Solution
E-0
E-b
Err.3
E-E
35
36
Message
Cause
Solution
E-t &
Temperature
Icon Sign
E-U
It could be caused by
a used test strip or a
temporary or permanent
electronic problem.
Chapter 9 Specifications
System Accuracy
38
Chapter 10 Limitations
Please observe the following limitations for obtaining accurate results of
BeneCheck PLUS Multi-Monitoring System :
1. For In-vitro diagnostic use only (External use only).
2. BeneCheck PLUS Test Strip is designed for one-time use only. Do not reuse.
3. Do not perform the test while a mobile phone is used beside; prevent the
electronic signal interference.