USM Go Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 194
At a glance
Powered by AI
The document outlines the general information, functions, and operation of the USM Go Ultrasonic Flaw Detector including powering on/off, using the keypad and display, and connecting ports.

To power on the instrument, press and hold the ON/OFF key for 2 seconds. To power off, press and hold the ON/OFF key again for 2 seconds.

The keypad is used to navigate menus and select options. It includes arrow keys, menu keys, and a joystick for positioning gates and cursors. Multi-key combinations also have special functions.

GE

Sensing & Inspection Technologies

Ultrasonic

USM Go
Operators Manual

P/N 1259801 Rev. 2


March 2010

USM Go
Ultrasonic Flaw Detector

Operators Manual
P/N 1259801 Rev. 2
March 2010

GESensingInspection.com
2009 General Electric Company. All rights reserved.
Technical content subject to change without notice.

[no content intended for this page - proceed to next page]

Contents

Chapter 1. General Information


1.1 Supplying Power to the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2 Powering the Instrument ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Using the Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3.1 Instrument Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3.2 Keypad Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3.3 Joystick Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.3.4 Multi-Key Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.4 Using the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.4.1 Accessing the Full-Screen A-Scan Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.4.2 Accessing the Acquire Mode Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.4.3 Accessing the Setup Mode Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1.5 Using the SD Slot, USB Connector & I/O Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.5.1 Removing the SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.5.2 Inserting the SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.5.3 Connecting the USB Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
1.5.4 Connecting the I/O Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

USM Go Operators Manual

iii

Contents

Chapter 2. Instrument Setup


2.1 Display Screen and Keypad Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.2 The Menu System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.3 Initial Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.3.1 Language, Units of Measurement, Date, and Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.3.2 Display Appearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2.3.3 Defining Function Toggle Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2.4 Installing a Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2.4.1 Connecting the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2.4.2 Configuring the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
2.4.3 Adjusting the Pulser Repetition Frequency (PRF) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
2.4.4 Setting the Pulser Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
2.4.5 Selecting the Pulser Type (OPTIONAL). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
2.4.6 Selecting the Pulser Energy or Width (OPTIONAL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2.4.7 Using the Phantom PRF Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
2.4.8 Selecting a Rectification Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
2.4.9 Setting the A-Scan REJECT Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

iv

USM Go Operators Manual

Contents

2.5 Adjusting the A-Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54


2.5.1 Setting the A-Scan Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
2.5.2 Setting the Display Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
2.6 Calibrating the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
2.6.1 Pre-Calibration Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
2.6.2 Using AUTOCAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
2.6.3 Checking the Calibration Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
2.7 Using the Calibration Reminder Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
2.8 Activating Instrument Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Chapter 3. Making Measurements
3.1 Configuring Gate A and Gate B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
3.1.1 Activating Automatic Gate Threshold (AGT). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
3.1.2 Positioning the Gates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3.1.3 Selecting the TOF-Detection Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
3.1.4 Setting Gate Alarms and Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

USM Go Operators Manual

Contents

3.2 Using Angle Beam Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76


3.2.1 Setting the Angle Beam Probe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
3.2.2 Indicating Leg with Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.3 Displaying Measured Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.4 Locking the Gain Toggle and the Joystick . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
3.5 Setting the Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
3.5.1 Changing the Gain-Adjustment Increment (dB STEP). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
3.5.2 Setting the User-Defined Gain Step (SETUP-GAIN-USER GAIN STEP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
3.6 Freezing the A-Scan Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
3.7 DAC/TCG Evaluation Mode (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
3.8 Using DAC Mode (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
3.8.1 Recording the DAC Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
3.9 Using TCG Mode (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
3.9.1 Generating the TCG Reference Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
3.9.2 Working with TCG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
3.10Adjusting DAC or TCG Display and Adding Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
3.10.1 Defining DAC or TCG Curve Offsets (DAC/TCG-OFFSETS-MODE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
3.10.2 Setting Transfer Correction (DAC/TCG-MAT ATTN-TRANSFER CORR) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93

vi

USM Go Operators Manual

Contents

3.11Editing and Deleting DAC and TCG Reference Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94


3.12Deleting a DAC Curve or TCG Reference Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
3.13DGS Evaluation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
3.13.1 Specifying a Probe and Preparing to Record the Reference Echo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
3.13.2 Record the Reference Echo that Defines the DGS Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
3.13.3 Display and Adjust the DGS Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
3.13.4 Evaluating Results in DGS Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
3.13.5 Locks and Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
3.13.6 Validity of the DGS Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
3.14dB REF Evaluation Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
3.15AWS D1.1 Weld Rating Evaluation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
3.16JISDAC Evaluation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
3.16.1 Recording the JISDAC Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
3.16.2 Working with JISDAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
3.16.3 Deleting a JISDAC Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Chapter 4. Data Sets & Reports
4.1 The FILES Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116

USM Go Operators Manual

vii

Contents

4.2 Working With Data Set Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117


4.2.1 Storing a New Data Set File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
4.2.2 Recalling a Data Set File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
4.2.3 Deleting a Data Set File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
4.2.4 Editing a Data Set File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
4.3 Creating a Memo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
4.4 Including a Memo in a Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
4.5 Creating a Report Header . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
4.6 Including a Header in a Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
4.7 Creating a Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
4.8 Storing a Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
4.9 Fast Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Chapter 5. Data Recorder Files
5.1 Naming the Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
5.1.1 File Selection Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
5.1.2 File Naming Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
5.2 Configuring the Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132

viii

USM Go Operators Manual

Contents

5.3 Creating the Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132


5.4 Viewing the Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
5.5 Using the Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
5.5.1 Adding Data and A-Scan Images to a Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
5.5.2 Previewing A-Scan Images in a Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
Appendix A. Specifications
A.1 LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
A.2 Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
A.3 Pulser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
A.4 Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
A.5 Gates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
A.6 Memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
A.7 Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
A.8 Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
A.9 USM Go Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Appendix B. Environmental Compliance
B.1 Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

USM Go Operators Manual

ix

Contents

B.2 Battery Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151


B.2.1 What do the Markings Mean? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
B.2.2 The Risks and Your Role in Reducing Them . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Appendix C. EN 12668 Specifications

USM Go Operators Manual

Preface

Safety Information
Before powering up or operating this instrument, the safety information in this section should be read carefully. This Operators Manual
should be stored in a safe place for reference.
IMPORTANT: This instrument is to be used only for testing materials in an industrial environment. Any use for medical applications or any

other purpose is not permitted.


IMPORTANT: This instrument is waterproof according to IP 67. It can be operated either with batteries or with the power supply unit. The
power supply unit meets the requirements of Electrical Safety Class II.

Batteries
For battery operation of this instrument, GEIT only recommends the use of a lithium-ion battery. You should only use the battery
recommended by GEIT for operation of this instrument. You can charge the lithium-ion battery either within the instrument itself or with the
external battery charger.
IMPORTANT: See Battery Disposal on page 151 for instructions on proper battery disposal procedures.

USM Go Operators Manual

xi

Preface

Important Ultrasonic Testing Guidelines


Please read the information in this section before using your instrument. It is important that you understand and observe this information to
avoid any operator errors that might lead to false test results. Such false results could result in personal injuries or property damage.

Using Ultrasonic Test Equipment


This Operators Manual contains essential information on how to operate your test equipment. In addition, there are a number of factors that
affect the test results, but a description of all these factors is beyond the scope of this manual. The three most important factors for safe and
reliable ultrasonic inspection are:

Operator training
Knowledge of special technical test requirements and limits
Choice of appropriate test equipment

xii

USM Go Operators Manual

Preface

Operator Training
The operation of an ultrasonic test device requires proper training in ultrasonic testing methods. Proper training comprises adequate
knowledge of:

The theory of sound propagation


The effects of sound velocity in the test material
The behavior of the sound waves at interfaces between different materials
The shape of the sound beam
The influence of sound attenuation in the test object and the influence of the surface quality of the test object
Lack of such knowledge could lead to false test results with unforeseeable consequences. You can contact GEIT or NDT societies or
organizations in your country (DGZfP in Germany; ASNT in the USA) for information on opportunities for training on ultrasonic
instruments that use time-of-flight measurements.
Accurate measurement results require a constant sound velocity in the test object. Steel test objects have only slight variations in sound
velocity, thus affecting only high precision measurements. Test objects made of other materials (e.g., nonferrous metals or plastics) may
have larger sound velocity variations, which could adversely affect the accuracy of the measurements.

USM Go Operators Manual

xiii

Preface

Test Object Material Effects


If the material of the test object is not homogeneous, the sound waves may propagate at different velocities in different parts of the test
object. An average sound velocity should then be used for the range calibration. This is achieved by using a reference block with a sound
velocity equal to the average sound velocity of the test object.
If substantial sound velocity variations are expected, then the instrument calibration should be adjusted to the actual sound velocity values at
shorter time intervals. Failure to do so may lead to false readings.

Test Object Temperature Effects


The sound velocity also varies as a function of the temperature of the test object. This can cause appreciable errors in measurements if the
instrument has been calibrated with a reference block at one temperature, and the instrument is then used on a test object at a different
temperature. Such measurement errors can be avoided either by ensuring that the reference block and the test object are at the same
temperature, or by using a correction factor obtained from published tables.

xiv

USM Go Operators Manual

Preface

Limited Warranty
For a period of two (2) years from the date of purchase, we warrant that the instrument will be free of any claim of ownership by third
parties, (ii) when new, be free from defects in material and workmanship and perform in accordance with the Products specifications under
normal use and service for the applicable warranty period following the date of sale. The second year of this warranty is valid only if the
instrument is calibrated to within the provided specifications, by us or one of our certified service providers after month twelve of ownership
but before month fourteen begins. The duration of the warranty may be extended or modified by explicit service contracts.
This limited warranty shall not apply to any problems arising from (i) failure to follow the product instructions or failure to perform
preventive maintenance, (ii) service, repair or modification by someone other than us or one of our authorized service representatives; or
(iii) external causes, such as accident, abuse, misuse, or problems with electrical power.
This warranty does not cover parts identified as wear-and-tear parts or lamps, transducers, tubes, accessories, or optional equipment not
manufactured by us, which items may be covered by separate manufacturers warranties.
Our obligation under this warranty is limited to the repair or replacement of components determined by us to be defective within the
warranty period at no cost to the original purchaser. Customer shall arrange for delivery to us in approved packing material. This warranty
extends to the original purchaser and cannot be assigned or transferred to any other party.
EXCEPT FOR THE WARRANTY SET ABOVE, WE EXPRESSLY DISCLAIM ALL WARRANTIES AND REPRESENTATIONS OF
ANY KIND WITH RESPECT TO OUR PRODUCTS, WHETHER EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, NON-INFRINGEMENT, TITLE AND ANY
WARRANTIES ARISING FROM COURSE OF PERFORMANCE, COURSE OF DEALING OR TRADE USAGE.
USM Go Operators Manual

xv

Preface

[no content intended for this page - proceed to next page]

xvi

USM Go Operators Manual

Chapter 1. General Information

Chapter 1. General Information


The USM Go is a portable ultrasonic flaw detector. In addition to its light-weight design, the USM Go includes a clean and simple user
interface and a large, easy-to-read color WVGA (800x480) display. When operating in Acquire Mode, the instrument provides ultrasonic
flaw detection and thickness measurements. In this mode, it is capable of storing A-Scans, operating parameters, and reports. Prior to using
the Acquire Mode, the instrument display and operating parameters must be configured by using the Setup Mode. The following specific
topics are discussed in this chapter:

Supplying power to the instrument


Powering the instrument ON and OFF
Using the keypad
Using the display

USM Go Operators Manual

Chapter 1. General Information

1.1 Supplying Power to the Instrument


Accessory Mounting Screw

Slot

1/4-Turn Screw (CW to close, CCW to open)


Power ON/OFF
Receive Probe
Connection

Pulse/Transmit Probe
Connection
Battery Charger
Connection
1/4-Turn Screw (CW to close, CCW to open)
Figure 1: Rear and Side Views of the USM Go Case
2

USM Go Operators Manual

Chapter 1. General Information

1.1 Supplying Power to the Instrument (cont.)


The USM Go can be powered in either of two ways (see Figure 1 on page 2):

A lithium battery pack that is installed in a compartment on the rear of the case, or
An external power adapter plugged into the connector on the side of the case
CAUTION!

Use only the GEIT Lithium battery pack in this instrument, and charge only this battery pack in the instrument or in
the provided GEIT charger/adapter.

To remove the battery compartment cover, loosen the two 1/4-turn screws and then lift up on the slot between the screws. The standard
GEIT Lithium battery pack is designed to provide maximum operating life between chargings.

The approximate level of remaining battery life is shown on the display (see Figure 2 on page 5) by the battery icon, , and the
approximate hours of charge indication below the icon. When a fully-charged battery pack is installed, the icon will appear as full, and
the icon will begin to empty as the battery life is depleted. When the battery indicator is down to one-quarter full, charge the battery pack
as soon as possible.
Note: The instrument automatically shuts OFF when the batteries are too weak for reliable operation. However, your settings are saved
and then restored when the instrument is turned back ON. When testing in remote locations, always carry a spare battery pack.

USM Go Operators Manual

Chapter 1. General Information

1.1 Supplying Power to the Instrument (cont.)


When the AC adapter is connected to the instrument, the icon in the upper right corner of the display indicates the percentage of full-charge
of the battery pack. When removing the battery pack to install a charged spare, the instrument will automatically turn OFF if the AC adapter
is not connected to the instrument. However, if the adapter is connected, the instrument will remain ON while you change battery packs.

1.2 Powering the Instrument ON and OFF


To power the USM Go ON or OFF, simply press the power button,
, on the side of the case (see Figure 1 on page 2). As soon as the
button press is recognized, you will hear the click of an internal relay. Then, after about 4 seconds, the display controller will be fully loaded
and the display will become visible.

1.3 Using the Keypad


The USM Go is designed to provide the user quick access to all of the instrument's functions. Its easy-to-use menu system allows any
function to be accessed with a minimum of effort. The functions typically used to collect ultrasonic data are located in the Acquire Mode
menu, while those used to configure the instrument are found in the Setup Mode menu.
See Figure 2 on page 5 for the locations of the front-panel components described in this chapter.

USM Go Operators Manual

Chapter 1. General Information

1.3 Using the Keypad (cont.)

Battery Indicator

Gain Toggle

Joystick

Function Toggle
Display Screen
Figure 2: Front Panel of the USM Go
USM Go Operators Manual

Chapter 1. General Information

1.3.1 Instrument Orientation


One of the innovative features of the USM Go is the user option to quickly and easily rotate the instrument 180 to accommodate either
right-hand or left-hand operation. During this process, the display image is also rotated to allow proper viewing. Figure 3 below shows the
instrument in both orientations.

Figure 3: USM Go in Left-Hand and Right-Hand Orientations


6

USM Go Operators Manual

Chapter 1. General Information

1.3.2 Keypad Components


The USM Go keypad includes the following items (see Figure 2 on page 5):

Center-press joystick (

): The joystick may be moved either left or right or up or down. In addition, the center of the joystick
may either be pressed or pressed and held.

Gain toggle (

): The two ends of the gain toggle act as separate buttons. One end is the Gain DOWN button, while the other end is
the Gain UP button. Either end of the gain toggle can either be pressed or pressed and held.

Function toggle (

): The two ends of the function toggle act as separate buttons. One end is the Function 1 button, while the other
end is the Function 2 button. Either end of the function toggle can either be pressed or pressed and held.

Note: The gain toggle is always at the top of the instrument, and the end of the function toggle closer to the display is always the
Function 1 button, regardless of the chosen instrument orientation.

USM Go Operators Manual

Chapter 1. General Information

1.3.3 Joystick Functions


The effects of the joystick actions described on the previous page are as follows:
Joystick Center Press:

When in Acquire Mode, a single press and release (


When in Setup Mode, a single press and release (

) toggles the A-Scan display between standard size and full screen mode.
) activates or deactivates a parameter for adjustment.

When in Acquire or Setup Mode, a 2-second press-and-hold (

) switches to the other mode.

Joystick Movement:

When in Acquire or Setup Mode, moving up/down (


When in Acquire or Setup Mode, moving left/right (

) scrolls between the available function options for the highlighted menu.
) highlights a menu option or adjusts the value of a selected parameter.

USM Go Operators Manual

Chapter 1. General Information

1.3.4 Multi-Key Functions


Note: All multi-key functions are defined with the instrument in the left-hand orientation (see the left side of Figure 3 on page 6).

Power button + Function 2 button + Gain DOWN button


Pressing and holding these three buttons simultaneously causes the instrument to initiate a software upgrade.
Note: A formatted SD card with a valid USM Go upgrade file in the root directory must be inserted prior to pressing these buttons.

Power button + Function 2 button + Gain UP button


Pressing and holding these three buttons simultaneously causes the instrument to ignore the last known setup and revert to the factory
default settings.
Important:

The last known setup will be overwritten and lost during this process.

Gain UP button + Gain DOWN button


Pressing and holding the two Gain Toggle buttons simultaneously activates the AUTO80 function for the current Evaluation Mode.

USM Go Operators Manual

Chapter 1. General Information

1.4 Using the Display


Typical displays for the USM Go Acquire Mode and Setup Mode menus are illustrated in Figure 4 below. See the following pages for
step-by-step instructions on accessing these menus.
Gain Window

Parameters

Gate Results

Battery Indicator

Menu Bar

Menu Title

Parameters

Grid Menu Bar

Figure 4: Acquire Mode (Left) and Setup Mode (Right) Displays


10

USM Go Operators Manual

Chapter 1. General Information

1.4.1 Accessing the Full-Screen A-Scan Display


When in Acquire Mode, a single press and release of the joystick (
) toggles the A-Scan display between standard-size and full-screen
display mode. When viewing the A-Scan in full-screen mode, the joystick can be used to change the parameters of Gate A, as follows:
Note: If the joystick lock function has been enabled (see Locking the Gain Toggle and the Joystick on page 82), any up-down and
left-right joystick movements will have no effect on the instrument.

Moving the joystick left or right (

) will change the Gate A range in that direction.

If AGT is OFF, moving the joystick up or down (

) will change the Gate A threshold, accordingly.

To disable full-screen Gate A control without effecting other instrument functions, activate the JOY CONTROL parameter as follows:

1. In the SETUP menu, activate the CONFIG1 submenu using the joystick (
2. Use the joystick (

).

) to select the JOY CONTROL function, and press the center of the joystick (

3. Use the joystick to select either the ON or OFF option, and press the center of the joystick (

USM Go Operators Manual

) to activate it.

) to deactivate the function.

11

Chapter 1. General Information

1.4.2 Accessing the Acquire Mode Menu


To access the Acquire Mode menu, refer to the left side of Figure 4 on page 10 while completing the following steps:

1. Move the joystick left or right (

) until the desired menu is highlighted on the menu bar. The function bar automatically shows the
parameters available in the highlighted menu.

2. Move the joystick up or down (

) to highlight the desired parameter.

3. While the desired parameter is highlighted, change its value by moving the joystick to the left or right (

) or by pressing either end

of the function toggle (


). For parameters with continuous values, the function toggle acts as a course adjustment, while the joystick
acts as a fine adjustment. For selecting a parameter value from a list, either the joystick or the function toggle may be used to step
through the list one item at a time.
Note: Pressing both ends of the function toggle simultaneously resets a highlighted parameter to its default value.

12

USM Go Operators Manual

Chapter 1. General Information

1.4.3 Accessing the Setup Mode Menu


To access the Setup Mode menu, refer to the right side of Figure 4 on page 10 while completing the following steps:

1. Press and hold the center of the joystick (

) to toggle between Acquire Mode and Setup Mode.

2. Move the joystick left or right (

) until the desired menu is highlighted on the menu bar. The function bar automatically shows the
parameters available in the highlighted menu.

3. Move the joystick as required (


4. Press the center of the joystick (
the joystick (
of the joystick (

) to highlight the desired parameter.


) to activate the highlighted parameter. The parameter value can now be changed either by moving

) as a fine adjustment or by pressing either end of the function toggle (

) as a coarse adjustment. Press the center

) again to deactivate the highlighted parameter.

Note: Step 4 above applies to the setting of all parameters, except for specifying the actions associated with the function toggle keys. For
instructions on configuring these keys, see Defining Function Toggle Actions on page 37.

USM Go Operators Manual

13

Chapter 1. General Information

1.5 Using the SD Slot, USB Connector & I/O Connector


The USM Go uses a standard SD memory card for storing data set files and reports (see The FILES Menu on page 116) and for loading an
instrument software upgrade (see Activating Instrument Upgrades on page 63). The SD card slot is located in a compartment on the top
of the instrument, along with a USB connector and an I/O connector (see Figure 5 below).
USB Connector

Hinged Cover

SD Card Slot

I/O Connector

Figure 5: Top View of USM Go Case

14

USM Go Operators Manual

Chapter 1. General Information

1.5.1 Removing the SD Card


To remove the SD card from its slot, proceed as follows:

1. Access the SD card slot by pushing on the cover in the direction of the arrow and lifting the hinged cover.
2. Press down of the SD card with your finger and then remove your finger quickly. The SD card will be partially ejected, and you can
then grasp it by the edge and slide it completely out of the slot.

1.5.2 Inserting the SD Card


To insert the SD card into its slot, proceed as follows:
CAUTION!

Do not force the SD card into the slot if you feel significant resistance. If the card does not enter the slot freely, it is
probably oriented incorrectly.

1. Access the SD card slot by pushing on the cover in the direction of the arrow and lifting the hinged cover.
2. Orient the SD card so that the label side faces the rear of the instrument and the blank side with the row of gold-colored electrical
contacts faces the display side of the instrument. The edge of the card with the row of gold-colored electrical contacts must enter the
slot first.

3. Slide the card fully into the slot and push gently until the card seats into its socket. Then, close the SD card slot cover.

USM Go Operators Manual

15

Chapter 1. General Information

1.5.3 Connecting the USB Port


The connector closest to the hinge of the top compartment cover (see Figure 5 on page 14) is a Micro USB port. If you use a standard USB
cable to connect the USM Go to a PC (no special drivers are required), the installed SD card will be added to the list of active drives on the
PC. You can then perform all normal drive activities, such as copying and deleting files, on the USM Go SD card.
Important:

16

While the USM Go is connected to a PC via the USB port, the instrument will not accept any user input from the keypad.
Normal operation resumes as soon as the USB cable is disconnected.

USM Go Operators Manual

Chapter 1. General Information

1.5.4 Connecting the I/O Port


The connector furthest from the hinge of the top compartment cover (see Figure 5 on page 14) is an I/O port. This port serves a dual role:

Serial port pins - these are used strictly for factory service diagnostics.
Sync & Alarm pins - these signals are accessible to the user via a special optional cable.
To use this connector you must order the optional USM Go cable, which is available as P/N 022-510-032. The pin designations for
connecting the open end of this cable are listed in Table 1 below:
Table 1: I/O Connector Pin Assignments
Pin #
Color
Signal
1
Brown
+5V
2
Red
SAP
3
Orange
Alarm
4
Yellow
RS232 CTS
5
Green
RS232 TX
6
Blue
RS232 RX
7
Purple
GND

USM Go Operators Manual

17

Chapter 1. General Information

[no content intended for this page - proceed to next page]

18

USM Go Operators Manual

Chapter 2. Instrument Setup

Chapter 2. Instrument Setup


2.1 Display Screen and Keypad Features
The USM Go user interface has been designed for clarity and ease of use. Figure 6 below show the complete set of icons that may appear in
the icon area of the screen, and Figure 7 on page 20 shows the major components of the display and the keypad.
Freeze mode has been activated

TCG MODE is turned on

Trig settings indicate


Curved Surface I-D reflection

Reject function has been turned on

A DGS reference is stored

Trig settings indicate


Curved Surface O-D reflection

Pulser set for dual element probe

dB REF MODE is turned on

Trig settings indicate


Curved Surface Beam

Pulser set for single element probe

AGT MODE is active

Trig settings indicate


Flat Angle reflection

Battery level indicator


(shown 1/4 full)

SD Card installed
(flashes when saving)

Trig settings indicate


Flat Surface Beam

Figure 6: Display Screen Icons


USM Go Operators Manual

19

Chapter 2. Instrument Setup

2.1

Display Screen Features (cont.)


Gain Toggle

Gain Level

Reading 1

Reading 2

Reading 3

Reading 4

Battery Level

Icon Area
Large Reading

Joystick

Reading 6
Reading 5

Function 1
Function 2
Function Toggle

ACQUIRE Menu Bar

Figure 7: Display Screen Features


20

USM Go Operators Manual

Chapter 2. Instrument Setup

2.2 The Menu System


The USM Go menu system, as shown in Figure 8 on page 22, allows the operator to select and adjust various instrument features and
settings. It includes:

Acquire Menu: Consists of several submenus used to calibrate the instrument prior to a test, configure the instrument during a test,
select the pulser and receiver characteristics, and configure the gates.

SETUP Menu: Consists of several submenus used to configure the instrument prior to a test, including specifying the acquire mode and
the screen appearance, adjusting the A-Scan display, setting the alarms, and controlling other significant measurement parameters.
The information provided in this chapter describes each menu function and shows how to access each function through the menu system.
The EVALuation MODE Function, which is located in the EVAL submenu of the SETUP menu, determines which submenu for evaluating
results appears in the Acquire and SETUP menus (see the shaded cells in Figure 8 on page 22). See Figure 9 on page 23 for the available
Evaluation Mode options.

USM Go Operators Manual

21

Chapter 2. Instrument Setup

2.2

The Menu System (cont.)

T
R

B
TE
A

image in report
param in report

U
R
ES

LT
S
U
R
ES

cal reminder
cal reset
user gain step
dB step

PL
A
Y
IS
D
LS
ER

TU
P
R

date
time
orientation
joy control

mode
reading 5
reading 6
large

JISDAC

PU

gate a logic
gate b logic
b start mode
output select

AWS D1.1

ST
A

language
units
decimal
date format

reading 1
reading 2
reading 3
reading 4

LT
S2

R
EP
E

TE
M
O
D
G
A
A
V

filename
create
DR view

SE

DR

function 1
function 2
about
ascan fill

D
R

TU
P
SE
TU
P

CONFIG2

E
A
M
TR
IG
N
A
L

dBREF

serial number
code
confirm

DGS

EG
IO

DAC/TCG

C
O
D

CONFIG1

probe angle
thickness
x value
O-diameter*

TU
P2

Depends on
OPTIONAL
Evaluation Mode
Configuration

eval mode
color leg
magnify gate
AGT

SE

EV

EVAL

filename
action

FI
LE
N

R
/H
D
LM
O
D

M
EM

FILES

gate b start
gate b width
b threshold
tof mode

Press and hold


to switch
between menus

SETUP MENU
header edit
memo edit
memo in report
hdr in report

gate a start
gate a width
a threshold
tof mode

gate a start
s-ref1
s-ref2
record

TE

U
TO
C

SELECTED
EVALUATION
MODE

G
A

A
L

frequency
rectify
dual
reject

EC

EI

VE
R

voltage
width/energy
damping
prf mode

R
LS
E
PU

R
A

N
G
E

ACQUIRE MENU
range
probe delay
velocity
display delay

color
grid
ascan color
brightness
pulser type
prf mode
phantom prf

top
bottom
ADV direction

Figure 8: Acquire and Setup Menus


22

USM Go Operators Manual

Chapter 2. Instrument Setup

2.2

The Menu System (cont.)


EVALUATION MODE MENUS AVAILABLE

R
D
EC
O
R
A
A
TE
G

P
SE
TU
A
A
TE
G

P
SE
TU
A

gate A start
gate A width
A threshold
tof mode

JISDAC

test attn
transfer corr

gate A start
auto80
record
finish
gate A start
gate A width
A threshold
tof mode

SE
TU
P

JISDAC
bold line
delete curve
transfer corr

A
T

TT
N

ref atten
ampl correct

T
FF
SE
TS
O
A
G
A
TE

offset 1
offset 2
offset 3
offset 4

gate A start
gate A width
A threshold
tof mode

A indication
B reference
C attenuation
D D1.1 rating

gate A start
reference type
ref size
record ref

M
A

M
A
FF
SE
TS
O
FF
SE
T

S2

mode
offset

A
TE

probe name
xtal frequency
eff diameter
delay velocity

mode
reference
record
delete ref

AWS D1.1

delete ref

A
TT
N

transfer corr

dgs mode
dgs curve
probe #
probe name

P
SE
TU
B
PR
O
D
G
S
H
O
EC
R
EF
O
R
R

point
point pos
point gain
enter

tcg/dac mode
tcg display
dac type
delete curve

EF

gate A start
gate A width
A threshold
tof mode

dB REF

DGS

gate A start
auto80
record
finish

A
TT
N

ED
IT

SE
TU
P

A
TE

EC
O

R
D

DAC/TCG

Use the EVAL MODE function


to make one of these
five menus active.

offset 1
offset 2
offset 3
offset 4
gate A start
gate A width
A threshold
tof mode

Figure 9: Available Evaluation Mode Menus


USM Go Operators Manual

23

Chapter 2. Instrument Setup

2.3 Initial Setup


In this section, youll learn how to configure the USM Go display and operating features. Follow these procedures to turn the instrument
ON and make initial adjustments to the control settings. Because the instrument can be set to save the control settings when it is turned OFF
and restore them when it is turned back ON, you wont have to repeat these adjustments unless a change is required.
Power the instrument ON and note that the ACQUIRE menu is automatically activated. Activate the SETUP menu by pressing in and
holding the center of the joystick (

).

Note: The entire menu structure is shown in Figure 8 on page 22 and Figure 9 on page 23.

2.3.1 Language, Units of Measurement, Date, and Time


Use the procedures in this section to adjust the units of measurement, the date, the time, and the language that appears on the display screen
and the data output. The adjustments require access to the REGIONAL and STARTUP function groups. These are accessed from the
CONFIG1 submenu in the SETUP menu, as shown in Figure 10 on page 25.

24

USM Go Operators Manual

PROBE DELAY

Sh

WIDTH/ENERGY

RANGE

USM Go Operators Manual


Se

PULSER

ACQUIRE MENU
RECEIVER
UNITS

REGIONAL

DATE

STARTUP

GRID

Se

ts
th e

Se

VELOCITY
DAMPING
DUAL
DECIMAL
TIME
A-SCAN COLOR

DISPLAY DELAY
PRF MODE
REJECT
DATE FORMAT
ORIENTATION
BRIGHTNESS

ts

dis

ta t

sc
re

en

br
igh

od

or
ro
b

e)

dp

or

ss

dis
p

lay

oo

ch
e

ra
ti

sc
h

pe

int

on

on

fig

em
es
ur
a ti
on
me
s
s3
an
d4

ur

do

f fo

an

po

ike

ial

ro
b

de

he
p

mo

ug
ht

(sp

ter

hr
o

ma

ve
t

en
er
gy

im
al

oc

eo

t-h

tw
rs

of
co
lo

ne

on

rr

s
de
c

igh

th e

to

do

co
l or

an

gr
id
t

en

t-h

fo r

pla
y

e
inc
he

ag
or

gu

dis
pla

th e

dis

tn e

rf

en

lan

om
ma

l ef

cre

fo r

cre
co
lo

ys
an

pla

fo r
th e

ys

ion

or

a te
ef

di s
pl a

ASc
lay

the
Di
sp

jus

th e

dth
F

wi
oR

om

pla

ac
dis

or
he

td

rt

tim
rie
n

et

en
st

cre
nit

ys
du
rio
d

cu
rre
n

yo

the

pe

ye

di s
pl a

t fo

rre
nt

ts

pla

cu

Se

dis

th e

fo r
ma

ra

ss

od

ch
e

wa

cte
d

nd

se
le

ou

rig
h

av
em

att
a

an
d

th e

nt
b

to

the

th e

ing
or
l eha
no
ele
lf-w
f th
me
av
eA
e
nt
or
-S
ca
du
nf
al ro
ele
m
me
th e
nt
dis
pla
y

pla

th e

or
ti o

dis
he

th e

te

ei t

Se
ts

the

Ad

ts

da

ts

Se
ts

th e

COLOR

ts

Se

ts

lec

ts

ep

ea

nm

me

el

ua
re
w

ev
el

or

for

es

sc
re
en

f th

l ay

cit
y

lef
t

ag
el

he

h(
sq

ol t

lev

ca
tio
ob

tifi
pr

re
c

the

th e

Se
Se
ts

LANGUAGE

Se

es

om

tifi

ss

en

ate

ing

i dt

rv

i ns
tru

gr

th e

irin

u ts

th e

In p

rf

mp

rw

lse

ot

isp

me
o

dv
el o

l ti

ed

ve

f th

tra

un

nt

so

he

ng
eo

Sc
a

pu

lse

da

pu

th e

ra
or
t

th e

A-

pu
ts

ye
d

he

lse

es
t

Se
ts

ve

ts

th e

Se

di s
pl a

r in

sf

th e

a te

ts

ns

jus

ys
o

pu

tch

th e

Re
mo

RECTIFY

In d

ts

Ad
mp
e

Se
ts

FREQUENCY

Se

th e

sp
la

VOLTAGE

Ma

i fts

Di

RANGE

Co

Chapter 2. Instrument Setup

2.3.1 Language, Units of Measurement, Date, and Time (cont.)

DISPLAY

CONFIG1 MENU

Figure 10: ACQUIRE Menu (Left) and CONFIG1 Menu (Right)


25

Chapter 2. Instrument Setup

2.3.1a

Setting the Acquire Mode Language (SETUP-CONFIG1-LANGUAGE)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the LANGUAGE function, then press the center of the joystick (
) to activate the function. To
change the selected language, either move the joystick or press the function toggle. Youll note that the options available are English,
German, French, Spanish, Italian, Romanian, Polish, Czech, Russian, Japanese and Chinese. The default language is English.

3.

Press the center of the joystick (


the choice last selected.

26

). Several functions are displayed on the screen.

) to deactivate the function when complete. The display screen and report language are now set to

USM Go Operators Manual

Chapter 2. Instrument Setup

2.3.1b

Setting the Units of Measurement (SETUP-CONFIG1-UNITS)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled UNITS, then press the center of the joystick (
following options are available:

). Several functions are displayed on the screen.


) to activate the function. The

mm - default setting which displays values in millimeters


INCH - displays values in inches
3.

To change the units of measurement, either move the joystick or press the function toggle.

4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

27

Chapter 2. Instrument Setup

2.3.1c

Setting the Decimal Convention (SETUP-CONFIG1-DECIMAL)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled DECIMAL, then press the center of the joystick (
The following options are available:

). Several functions are displayed on the screen.


) to activate the function.

PERIOD - uses a period as a decimal point


COMMA - uses a comma as a decimal point
3.

To change the decimal convention, either move the joystick or press the function toggle.

4.

After making your choice, press the center of the joystick (

28

) to deactivate the function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.3.1d

Setting the Date and Time Formats (SETUP-CONFIG1-DATE FORMAT)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


function.

3.

To change the selected date and time format, either move the joystick or press the function toggle. Choose from the following date and
time formats:

). Several functions are displayed on the screen.

) to select the function titled DATE FORMAT, then press the center of the joystick (

) to activate the

Y-M-D date format and 12 or 24 hour time format


M/D/Y date format and 12 or 24 hour time format
D.M.Y date format and 12 or 24 hour time format
4.

After making your choice, press the center of the joystick (


) to deactivate the function. The date and time format shown on the
display screen and in the out reports are now set to the choice last selected.

USM Go Operators Manual

29

Chapter 2. Instrument Setup

2.3.1e

Setting the Date (SETUP-CONFIG1-DATE)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled DATE, then press the center of the joystick (
that the first character is highlighted.

3.

Move the joystick up or down (


other characters to be modified.

4.

After making your choice, press the center of the joystick (

30

). Several functions are displayed on the screen.


) to activate the function. Note

) to change the highlighted character. Then, move the joystick left or right (

) to select the

) to deactivate the function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.3.1f

Setting the Time (SETUP-CONFIG1-TIME)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled TIME, then press the center of the joystick (
that the first character is highlighted.

3.

Move the joystick up or down (


other characters to be modified.

4.

After setting the correct time, press the center of the joystick (

). Several functions are displayed on the screen.


) to activate the function. Note

) to change the highlighted character. Then, move the joystick left or right (

) to select the

) to deactivate the function.

Note: Once set, the internal clock maintains the current date and time.

USM Go Operators Manual

31

Chapter 2. Instrument Setup

2.3.1g

Setting Left-Hand or Right-Hand Orientation (SETUP-CONFIG1-ORIENTATION)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled ORIENTATION, then press the center of the joystick (
function. Select either RIGHT hand or LEFT hand control.

3.

To change the display screen orientation, either move the joystick or press the function toggle.

4.

After making your choice, press the center of the joystick (

32

). Several functions are displayed on the screen.


) to activate the

) to deactivate the function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.3.2 Display Appearance


Follow the procedures in this section to adjust the display appearance. The adjustments require access to the CONFIG1 submenu, which is
accessed from the SETUP menu (see Figure 10 on page 25).

2.3.2a

Setting the Display Color (SETUP-CONFIG1-COLOR)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled COLOR, then press the center of the joystick (
There are four preset color schemes.

3.

To change the display color scheme, either move the joystick or press the function toggle.

4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

). Several functions are displayed on the screen.


) to activate the function.

) to deactivate the function.

33

Chapter 2. Instrument Setup

2.3.2b

Selecting a Display Grid (SETUP-CONFIG1-GRID)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (

3.

To change the display grid type, either move the joystick or press the function toggle. The selected grid style will be shown in the
A-Scan window on the display screen, after you return to Acquire Mode.

4.

After making your choice, press the center of the joystick (

34

). Several functions are displayed on the screen.

) to select the function titled GRID, then press the center of the joystick (

) to activate the function.

) to deactivate the function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.3.2c

Setting the A-Scan Color (SETUP-CONFIG1-ASCAN COLOR)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled ASCAN COLOR, then press the center of the joystick (
function. There are six A-Scan color options.

3.

To change the A-Scan color, either move the joystick or press the function toggle.

4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

). Several functions are displayed on the screen.


) to activate the

) to deactivate the function.

35

Chapter 2. Instrument Setup

2.3.2d

Setting the Display Brightness (SETUP-CONFIG1-BRIGHTNES)

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled BRIGHTNESS, then press the center of the joystick (
function. The available settings range from 1 to 10.

3.

To change the brightness level, either move the joystick or press the function toggle.

4.

After making your choice, press the center of the joystick (

36

). Several functions are displayed on the screen.


) to activate the

) to deactivate the function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.3.3 Defining Function Toggle Actions


The user can specify a desired action to occur when either end of the Function Toggle (
) is pressed or pressed-and-held. The
user-specified function action is ignored, however, whenever a parameter is selected and its value is being edited.

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (

). Several functions are displayed on the screen.

) to select either the function titled FUNCTION1, for the end of the toggle closer to the display screen, or the

function titled FUNCTION2, for the end of the toggle further from the display screen. Then, press the center of the joystick (
activate the function.

) to

Note: There are two values for each function. The upper parameter determines the action taken when then corresponding toggle end is
momentarily pressed. The lower parameter defines the action taken when the toggle end is pressed and held.
3.

To change the upper parameter, move the joystick left or right (

USM Go Operators Manual

).

37

Chapter 2. Instrument Setup

2.3.3 Defining Function Toggle Actions (cont.)


4.

To change the lower parameter, press the function toggle to scroll through the options. The available options include:

NONE no action is assigned.


FREEZE Freezes the A-Scan and displays the Freeze icon (see Figure 6 on page 19) in the status bar.
JOYSTICK LOCK Prevents up-down and left-right adjustments using the joystick (

) and displays the Lock icon (see Figure 6


on page 19) on the display screen. However, center-press operations with the joystick are not disabled.

COPY Performs the task specified by the ACTION function, which is located in the FILE submenu.
AUTO80 Adjusts the gain to place the Gate A triggering echo peak at 80% of full-screen height.
MAGNIFY GATE Zoom the A-Scan so that the displayed screen width matches the user-specified gate width.
HOME Selects the RANGE (i.e. HOME) menu in Evaluation Mode.
5.

38

After making your choice, press the center of the joystick (

) to deactivate the function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4 Installing a Probe


Follow the instructions in this section to install a probe on your USM Go.

2.4.1 Connecting the Probe

Receive
Probe

When connecting a probe to the instrument, the following steps must be taken:

Properly complete the physical connection of the probe to the instrument.


Properly configure the instrument to work with the connected probe.
The USM Go accepts either a single-element probe or a dual-element probe.
To install a single-element probe, connect the probe cable to either of the two ports on
the side of the instrument (see Figure 11 to the right). When a dual-element probe is
connected to the instrument, the key between and below the ports ensures proper
orientation of the transmit and receive connectors.

Transmit
Probe

Figure 11: Probe Connector Locations

USM Go Operators Manual

39

Chapter 2. Instrument Setup

2.4.2 Configuring the Instrument


Three instrument settings are directly dependent on the type of probe installed. These settings must be adjusted any time a probe of a
different type is installed, by following the instructions in the following sections.

2.4.2a

Selecting the Probe Type (RECEIVER-DUAL)

1.

In the ACQUIRE menu, activate the RECEIVER submenu using the joystick (

2.

Use the joystick (

3.

To change the probe type, move the joystick (


) or press the function toggle. Each available probe type is represented by an icon in
the Icon Bar, near the upper left corner of the display, whenever that probe type is highlighted. The following options are available:

).

) to select the function titled DUAL.

ON - Use for dual-element probes. If this option is selected, the Dual icon (see Figure 6 on page 19) is displayed.
OFF - Use for single-element probes. If this option is selected, the Single icon (see Figure 6 on page 19) is displayed.
4.

40

After completing your selection, move the joystick up or down (

) to navigate away from this function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.2b

Specifying the Probe Frequency (RECEIVER-FREQUENCY)

1.

In the ACQUIRE menu, activate the RECEIVER submenu using the joystick (

2.

Use the joystick (

3.

To change the specified frequency, move the joystick (

).

) to select the function titled FREQUENCY.


) or press the function toggle. The following options are available:

1, 2, 2.25, 4, 5, 10, 13, 15 MHz - Choose the frequency that matches the frequency of your probe.
BROADBAND - Select this option to use the built-in broadband filter.
4.

After completing your selection, move the joystick up or down (

USM Go Operators Manual

) to navigate away from this function.

41

Chapter 2. Instrument Setup

2.4.2c

Changing Damping Level to modify the Signal to Noise Ratio (PULSER-DAMPING)

1.

In the ACQUIRE menu, activate the PULSER submenu using the joystick (

2.

Use the joystick (

3.

To change the specified damping level and optimize the A-Scan signal appearance, move the joystick (
toggle. The following options are available:

).

) to select the function titled DAMPING.


) or press the function

50
1000
4.

42

After completing your selection, move the joystick up or down (

) to navigate away from this function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.3 Adjusting the Pulser Repetition Frequency (PRF)


The Pulser fires at a frequency which can be set either automatically or manually. To set the PRF mode and frequency level:
1.

In the ACQUIRE menu, activate the PULSER submenu using the joystick (

2.

Use the joystick (

3.

Press the function toggle to see the available choices:

).

) to select the function titled PRF MODE.

AUTO HIGH - The instrument calculates and sets the pulser firing rate at 75% of the maximum frequency possible, based on the
range and the material velocity.

AUTO MED - The instrument calculates and sets the pulser firing rate at 50% of the maximum frequency possible, based on the range
and the material velocity.

AUTO LOW - The instrument calculates and sets the pulser firing rate at 20% of the maximum frequency possible, based on the range
and the material velocity.

MANUAL - Allows the user to set the pulser frequency. However, unacceptable PRF settings will cause a display prompt to appear.

USM Go Operators Manual

43

Chapter 2. Instrument Setup

2.4.3 Adjusting the Pulser Repetition Frequency (PRF) (cont.)


4.

If PRF MODE is set to AUTO HIGH, AUTO MED or AUTO LOW, the automatically calculated value is displayed in the function
box. If you selected the MANUAL option, you may now adjust the PRF value by moving the joystick left or right (

).

Note: The PRF setting may be limited based on the user-selected pulser voltage setting. This feature acts to limit signal dissipation.

2.4.4 Setting the Pulser Voltage


The relative energy with which the pulser fires is adjusted by changing the VOLTAGE setting. To set the pulser voltage level:

1.

In the ACQUIRE menu, activate the PULSER submenu using the joystick (

2.

Use the joystick (

).

) to select the function titled VOLTAGE. Press the function toggle or move the joystick (

) to:

Set the voltage level to HIGH or LOW, for a standard PULSER TYPE setting of SPIKE, or
Specify the actual VOLTAGE value, for an optional PULSER TYPE setting of SQUARE
3.

44

After completing your selection, move the joystick up or down (

) to navigate away from this function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.5 Selecting the Pulser Type (OPTIONAL)


The standard pulser shape is a spike, and there is also an optional square pulser shape available. If the PULSER TYPE option is activated,
choose between the spike and square options as follows:

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (


function.

3.

Press the function toggle or move the joystick (


activate the function.

).

) to select the function titled PULSER TYPE. Then, press the center of the joystick (

) to activate the

) to select SPIKE or SQUARE. Then, press the center of the joystick (

) to

Note: This selection influences the VOLTAGE settings available and determines if a PULSER WIDTH function (SQUARE wave mode)
or a PULSER ENERGY function (SPIKE mode) is available.
4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

45

Chapter 2. Instrument Setup

2.4.6 Selecting the Pulser Energy or Width (OPTIONAL)


After activating the PULSER TYPE option, either SQUARE wave mode or SPIKE mode may be chosen (see the previous page).
Depending on your choice, proceed to the appropriate section below to specify either the pulser width (square wave mode) or the pulser
energy (spike mode).

2.4.6a

Selecting Pulser Width in Square Wave Mode

The pulser width generally varies from 30 to 500 nanoseconds. A recommended starting point from which the width setting can be adjusted
is found with the following equation:

1000
PULSE WIDTH (nanoseconds) = ------------ with f in MHz
2f
For example, if a 2 MHz probe is used, the equation becomes:

PULSE WIDTH (nanoseconds) = 1000


------------ = 250
22

46

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.6a

Selecting Pulser Width in Square Wave Mode (cont.)

To set the pulser width:


1.

In the ACQUIRE menu, activate the PULSER submenu using the joystick (

2.

Use the joystick (

) to select the function titled WIDTH, which is only available if the PULSER TYPE function is set to

SQUARE. Press the function toggle or move the joystick (


3.

) to set the width of the pulser.

After completing your selection, move the joystick up or down (

USM Go Operators Manual

).

) to navigate away from this function.

47

Chapter 2. Instrument Setup

2.4.6b

Selecting Pulser Energy in Spike Mode

To set the pulser energy:


1.

In the ACQUIRE menu, activate the PULSER submenu using the joystick (

2.

Use the joystick (

) to select the function titled ENERGY, which is only available if the PULSER TYPE function is set to

SPIKE. Press the function toggle or move the joystick (


3.

48

).

) to set the energy of the pulser to either HIGH or LOW.

After completing your selection, move the joystick up or down (

) to navigate away from this function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.7 Using the Phantom PRF Feature


When activated, this diagnostic feature varies the PRF to identify any wrap-around signals, which are phantom echoes caused by a PRF
setting that is too high. When this feature is activated, the time-based position of the phantom echoes varies while the true echoes remain
stationary on the display screen.

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (

3.

Press the function toggle or move the joystick (

4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

).

) to select the function titled PHANTOM PRF. Press the joystick (

) to activate this function.

) to choose either ON or OFF.


) to deactivate the function.

49

Chapter 2. Instrument Setup

2.4.8 Selecting a Rectification Mode


Rectification effects the orientation of the A-scan on the display screen. The A-scan represents the sound pulse (i.e. echo) that is returned to
the instrument from the material being tested. The series of echoes looks like the Radio Frequency (RF) signal shown in Figure 12 below.
Note that the RF signal has both a negative component below the axis and a positive component above the axis. In RF mode, Gate A and
Gate B can be positioned either above or below the axis, to be triggered by either a positive-heading echo or a negative-heading echo.

Figure 12: Typical RF and Rectified Signals

50

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.8 Selecting a Rectification Mode (cont.)


Positive Half Rectification means that only the upper (i.e. positive) half of the RF signal is displayed.
Negative Half Rectification means that only the bottom (i.e. negative) half of the RF signal is displayed (see Figure 12 on page 50). Note
that although only the negative half of the RF signal is displayed, it is shown in the same orientation as a positive component to simplify
viewing.

Full-Wave Rectification combines the positive- and negative-rectified signals together, and displays both of them in a positive
orientation (see Figure 12 on page 50).
Use the procedure on the next page to select a rectification mode.

USM Go Operators Manual

51

Chapter 2. Instrument Setup

2.4.8 Selecting a Rectification Mode (cont.)


1.

In the ACQUIRE menu, activate the RECEIVER submenu using the joystick (

2.

Use the joystick (

).

) to select the function titled RECTIFY. The following options are available:

NEG HALFWAVE - Shows the negative component of the RF signal but displays it in a positive orientation.
POS HALFWAVE - Shows the positive component of the RF signal.
FULLWAVE - Shows the positive and negative halves of the RF wave, but both are oriented in the positive direction.
RF - Shows the echo with no rectification.
3.

Press the function toggle or move the joystick (

4.

After completing your selection, move the joystick up or down (

52

) to select the desired rectification method.


) to navigate away from this function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.4.9 Setting the A-Scan REJECT Level


A portion of the A-Scan can be omitted from the display screen, by defining the percentage of the full-screen height you wish to omit. To set
a reject percentage:

1.

In the ACQUIRE menu, activate the RECEIVER submenu using the joystick (

2.

Use the joystick (

3.

To change the amount of the A-Scan you wish to omit from the display, as a percentage of the full-screen height, either move the
joystick (

).

) to select the function titled REJECT.

) or press the function toggle. You may omit a portion of the A-Scan up to 80% of the full-screen height.

Note: Whenever REJECT is set to a value greater then 0%, the reject icon (see Figure 6 on page 19) is displayed in the status bar.

USM Go Operators Manual

53

Chapter 2. Instrument Setup

2.5 Adjusting the A-Scan


To configure the USM Go A-Scan, follow the instructions in this section.

2.5.1 Setting the A-Scan Range


Calibration of the USM Go requires the use of two calibrated standards. These standards must be of different thicknesses, and they must be
made of the same material as the test piece. Prior to calibrating the instrument/probe combination, the the A-Scan display range, which is the
material thickness value represented by the full horizontal width of the screen, is normally set to a value equal to or slightly larger than the
thicker calibrated standard (see Figure 13 below).

Figure 13: Effects of the A-Scan Range Adjustment

54

USM Go Operators Manual

Chapter 2. Instrument Setup

2.5.1 Setting the A-Scan Range (cont.)


To set the A-Scan range:
1.

In the ACQUIRE menu, activate the RANGE submenu using the joystick (

2.

Use the joystick (


) to select the function titled RANGE, which has both coarse and fine adjustment modes. Coarse adjustments
are made with the function toggle, while fine adjustments are made with the joystick. When RANGE appears in all capital letters,
coarse adjustments are being made, while fine adjustments occur when range appears in all lower-case letters.

3.

Press the function toggle or move the joystick (

).

) to change the range setting. Values from 0.040 to 1100 in. are allowed.

Note: Pressing both buttons on the Function Toggle simultaneously resets the RANGE to the default value of 10.000 in.
4.

After completing your selection, move the joystick up or down (

USM Go Operators Manual

) to navigate away from this function.

55

Chapter 2. Instrument Setup

2.5.2 Setting the Display Delay


The display delay function shifts the displayed A-Scan to the left or the right in the viewing window. To set the display delay:
1.

In the ACQUIRE menu, activate the RANGE submenu using the joystick (

2.

Use the joystick (

3.

Press the function toggle or move the joystick (


echoes shift to the left or the right.

).

) to select the function titled DISPLAY DELAY.


) to change the display delay. As you change the value, you should see the displayed

Note: Pressing both buttons on the Function Toggle simultaneously resets the DELAY to the default value of 0.
4.

56

After completing your selection, move the joystick up or down (

) to navigate away from this function.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.6 Calibrating the Instrument


To calibrate the USM Go, follow the instructions in this section.

2.6.1 Pre-Calibration Check List


To improve the accuracy and quality of your calibration, be sure that the following steps have been taken before beginning the calibration:

Install the probe


Adjust the DUAL (RECEIVER) setting to match probe
Set the material type
Set the DISPLAY DELAY to 0 (recommended)
Set the PRF to AUTO LOW
Set TCG to OFF
Set REJECT to 0 (recommended).

USM Go Operators Manual

57

Chapter 2. Instrument Setup

2.6.2 Using AUTOCAL


Note: While following the instructions in this section, refer to Figure 14 on page 59.

1.

In the ACQUIRE menu, activate the AUTOCAL submenu using the joystick (

2.

Use the joystick (


) to select the function titled S-REF1. Then, press the function toggle or move the joystick (
) to change
the value to match the thickness of the thinner calibration standard. The S-REF1 function has both coarse and fine adjustment modes.
Coarse adjustments are made with the function toggle, while fine adjustments are made with the joystick.

3.

Use the joystick (


) to select the function titled S-REF2. Then, press the function toggle or move the joystick (
) to change
the value to match the thickness of the thicker calibration standard. The S-REF2 function has both coarse and fine adjustment modes.
Coarse adjustments are made with the function toggle, while fine adjustments are made with the joystick.

4.

Apply couplant and couple the probe to the thinner calibration standard. With the A START function highlighted, press the function

).

toggle or move the joystick (


) to shift the starting point of Gate A until Gate A lies over the echo corresponding to the thickness of
the thinner standard (see Figure 14 on page 59). Coarse adjustments are made with the function toggle, while fine adjustments are made
with the joystick.

58

USM Go Operators Manual

Chapter 2. Instrument Setup

2.6.2 Using AUTOCAL (cont.)

Figure 14: Auto Calibration Procedures


USM Go Operators Manual

59

Chapter 2. Instrument Setup

2.6.2 Using AUTOCAL (cont.)


Note: During the following steps, pressing both buttons on the Gain Toggle simultaneously activates the AUTO80 function.
5.

Use the joystick (

) to select the function titled RECORD. The value in the function box changes from OFF to S-REF1?.

While maintaining the signal in Gate A, move the joystick right or left (
now reads S-REF2?.
6.

) to record the reference echo. The value in the function box

Apply couplant and couple the probe to the thicker calibration standard. Then, use the joystick (

) to select the function titled

A START. Press the function toggle or move the joystick (


) to shift the starting point of Gate A until Gate A lies over the echo
corresponding to the thickness of the thicker standard (see Figure 14 on page 59). Coarse adjustments are made with the function
toggle, while fine adjustments are made with the joystick.
7.

Use the joystick (


left (

60

) to select the function titled RECORD. While maintaining the signal in Gate A, move the joystick right or

) to record the reference echo. The value in the function box now reads OFF.

USM Go Operators Manual

Chapter 2. Instrument Setup

2.6.3 Checking the Calibration Results


Following the calibration procedure, the calculated acoustical velocity and probe delay are displayed. To view these calculated values:

1.

In the ACQUIRE menu, activate the RANGE submenu using the joystick (

2.

Press the function toggle or move the joystick (

).

) to view the following selections:

PROBE DELAY - The adjustment made as a result of the AUTOCAL (zeroing) procedure is shown. This represents the time delay
caused by sound-wave traveling through the probe membrane, wear plate, or delay line.

VELOCITY - The calculated velocity after the calibration is shown. The material type is listed as custom.

USM Go Operators Manual

61

Chapter 2. Instrument Setup

2.7 Using the Calibration Reminder Alarm


The USM Go incorporates a timed alarm feature that causes an icon to appear on user-defined input intervals between 0.5 and 4.0 hours. To
use this alarm:

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (


) to select the function titled CAL REMINDER, then press the center of the joystick (
) to activate the
function. Input the intervals at which the reminder alarm will trigger (between 0.5 and 4.0 hours). Setting this value automatically resets
the alarm to trigger on the specified intervals. This function also allows the alarm to be disabled.

3.

Press the function toggle or move the joystick (

4.

Select and activate CAL RESET to acknowledge the triggered alarm and reset it to resume normal operation.

).

) to change the alarm interval.

Note: The selection of the PROBE DELAY or VELOCITY functions also resets the alarm.

62

USM Go Operators Manual

Chapter 2. Instrument Setup

2.8 Activating Instrument Upgrades


When provided, activation codes matched to your USM Go serial number can be input via the CODE submenu, which is located in the
CONFIG menu. This submenu also lists the serial number assigned to your instrument. To input an activation code:

1.

In the SETUP menu, activate the CONFIG1 submenu using the joystick (

2.

Use the joystick (


) to select the function titled CODE, and then press the center of the joystick (
Note that the first character of the current code value is highlighted.

3.

Use the joystick (


) to change the highlighted character to match the value provided by GEIT. Then, move the joystick (
select the next character, and continue modifying the values until they all match the code provided by GEIT.

).
) to activate the function.

) to

Note: Holding either end of the Function Toggle down for three seconds resets all values in the code to 0.

USM Go Operators Manual

63

Chapter 2. Instrument Setup

2.8

Activating Instrument Upgrades (cont.)

4.

After completing the code entry, press the center of the joystick (

5.

Use the joystick (


the new code is correct.

6.

Be sure to follow on-screen instructions to successfully complete the activation.

64

) to deactivate the function.

) to select the function titled CONFIRM, and then press the center of the joystick (

) to acknowledge that

USM Go Operators Manual

Chapter 3. Making Measurements

Chapter 3. Making Measurements


This chapter explains how to configure the USM Go flaw detection and thickness measurement capabilities. It then explains how to make
ultrasonic measurements.

3.1

Configuring Gate A and Gate B

Setting the position and the characteristics of Gate A and Gate B is the first step in configuring the instrument for flaw detection or material
thickness measurements. The GATE A and GATE B submenus control the location of Gate A and Gate B, while the EVAL and CONFIG2
submenus in the SETUP menu control the operating characteristics for the gates.

3.1.1

Activating Automatic Gate Threshold (AGT)

IMPORTANT: The Automatic Gate Threshold (AGT) function is available for Gate A only.
The AGT function automatically programs the Gate A threshold level to a percentage of the peak amplitude detected in the gate. The
threshold percentage scale is adjustable in 1% steps by the user from 5% to 95% of the detected peak amplitude. The calculated threshold
value programmed into the hardware ranges from 5% to 95% of the full-screen height.
Note: An icon is displayed in the Symbol Area when the AGT mode is activated.

USM Go Operators Manual

65

Chapter 3. Making Measurements

3.1.1

Activating Automatic Gate Threshold (AGT) (cont.)

To configure the Automatic Gate Threshold (AGT) function, follow this procedure:
1.

In the SETUP menu, activate the EVAL submenu using the joystick (

2.

Use the joystick (

).

) to select the function titled AGT. Use the joystick (

) to select one of the following options:

OFF (both RF and rectified modes) - Disables the AGT function.


ON (rectified modes only) - Activates the AGT function. The Gate A threshold value is scaled to a percentage of the rectified peak
amplitude value detected in Gate A.

ON (-) (RF mode only) - Activates the AGT function. The Gate A threshold value is scaled to a percentage of the negative peak amplitude
value detected in Gate A.

ON (+) (RF mode only) - Activates the AGT function. The Gate A threshold value is scaled to a percentage of the positive peak amplitude
value detected in Gate A.
Note: The AGT sign mode (+ or -) is preserved when switching between RF and rectified A-Scan display modes.

66

USM Go Operators Manual

Chapter 3. Making Measurements

3.1.2

Positioning the Gates

Use the procedures in this section to set the vertical and horizontal position of Gate A and Gate B. Remember that gate position has the
following effects on instrument performance:

A-Scan echoes on the right side of the display screen represent features that occur at a greater depth from the test-material surface than
those on the left side of the display screen. Therefore, moving a gate to the right means that the gate is evaluating a deeper portion of the
test material.

A wider gate setting means that the gate is evaluating a thicker portion of the test material.
Increasing the vertical height, called the threshold, of a gate means that only reflected signals of sufficiently large amplitude will cross
the gate.
To configure the gate parameters, complete the following steps and then proceed to the appropriate section:
1.

In the ACQUIRE menu, activate the Gate A or Gate B submenu using the joystick (

2.

Choose the desired function from the available options.

USM Go Operators Manual

).

67

Chapter 3. Making Measurements

3.1.2a

Setting the Gate Starting Point (GATE A or GATE B-GATE START)

1.

In the ACQUIRE menu, activate the GATE A or GATE B submenu using the joystick (

2.

Use the joystick (

3.

To change the gate starting point, either move the joystick (

).

) to select the GATE A START (or GATE B START) function.


) or press the function toggle.

Increasing and decreasing the value of the starting point moves the gate to the right and left, respectively. The gate starting point remains as
set here, even when width adjustments are made.
Note: The GATE A START and GATE B START functions have both coarse and fine adjustment modes. Coarse adjustments are made
with the function toggle, while fine adjustments are made with the joystick. When the function name appears in all capital letters,
coarse adjustments are being made, while fine adjustments occur when the function name appears in all lower-case letters.
4.

68

After completing your selection, move the joystick up or down (

) to navigate away from this function.

USM Go Operators Manual

Chapter 3. Making Measurements

3.1.2b

Adjusting the Gate Width (GATE A WIDTH or GATE B-GATE WIDTH)

1.

In the ACQUIRE menu, activate the GATE A or GATE B submenu using the joystick (

2.

Use the joystick (

3.

To change the gate width, either move the joystick (

).

) to select the GATE A WIDTH (or GATE B WIDTH) function.


) or press the function toggle.

Note: This function has both coarse and fine adjustment modes. Coarse adjustments are made with the function toggle, while fine
adjustments are made with the joystick.
4.

After completing your selection, move the joystick up or down (

USM Go Operators Manual

) to navigate away from this function.

69

Chapter 3. Making Measurements

3.1.2c

Setting the Gate Threshold (Vertical Position) (A THRESHOLD or B THRESHOLD)

1.

In the ACQUIRE menu, activate the GATE A or GATE B submenu using the joystick (

2.

Use the joystick (

).

) to select the A THRESHOLD (or B THRESHOLD) function.

IMPORTANT: If the AGT function has been activated (see Activating Automatic Gate Threshold (AGT) on page 65), the A THRESHOLD
menu parameter is replaced with the AGT Scale Value. The scale value is displayed in the form of AGT = ss%, where ss
represents the AGT scale value. The user can modify the AGT scale value using the joystick and/or the function keys.
3.

To change the gate height, either move the joystick (


threshold moves the gate up or down, respectively.

4.

After completing your selection, move the joystick up or down (

70

) or press the function toggle. Increasing or decreasing the value of the

) to navigate away from this function.

USM Go Operators Manual

Chapter 3. Making Measurements

3.1.3

Selecting the TOF-Detection Method

A-Scan signals crossing either Gate A or Gate B are evaluated for the purposes of flaw detection and material thickness evaluation. When
the signal crosses Gate A or Gate B, either the gate-crossing point (i.e. the flank) of the signal, or the maximum point (i.e. the peak) of the
signal in that specific gate is used for evaluation purposes. The TOF MODE function allows the user to specify which A-Scan feature
(FLANK or PEAK or JFLANK) is used to evaluate the signal in each gate.

1.

In the ACQUIRE menu, activate the GATE A or GATE B submenu using the joystick (

2.

Use the joystick (

3.

To change the gate detection mode, either move the joystick (

).

) to select the TOF MODE function.


) or press the function toggle. The following options are available:

PEAK - All detection, whether time-based and peak height, is based on the HIGHEST triggering echo in the gate.
FLANK - The time-based triggering point is the first flank to cross the gate, and amplitude-based results are based on the HIGHEST peak
of any echo crossing the gate. This is not necessarily the peak of the same echo whose flank triggered the gate.

JFLANK - The time-based triggering point is the first flank to cross the gate, and the amplitude-based results are based on the peak of this
first echo to cross the gate. This is not necessarily the highest echo in the gate.

USM Go Operators Manual

71

Chapter 3. Making Measurements

3.1.3
4.

Selecting the TOF-Detection Method (cont.)

After completing your selection, move the joystick up or down (

) to navigate away from this function.

Note: The detection method chosen is indicated by a small icon. This icon is displayed in the display box containing the measured reading,
and in the options offered in the READING 1 through 6 and LARGE function boxes.

3.1.4

Setting Gate Alarms and Outputs

An alarm can be set for either one or both of the gates. This is accomplished by configuring either Reading Box 4 (in LARGE mode) or the
single Large Reading Box (in SMALL mode) to act as a virtual LED that mimics an indicator light. When so configured, the virtual LED
reading box is green when there is no fault condition and red when the alarm has been triggered.

72

USM Go Operators Manual

Chapter 3. Making Measurements

3.1.4a

Defining Gate-Alarm Logic (SETUP-CONFIG2-GATE A or B LOGIC)

Each gate alarm can be triggered by one of two circumstances: either when an A-Scan echo crosses the gate or when no A-Scan echo crosses
the gate. Use the following procedure to specify the gate LOGIC settings:

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (


the function.

3.

To choose the gate-alarm triggering logic, move the joystick up or down (

).

) to select the function titled GATE A (or GATE B) LOGIC. Press the center of the joystick (

) to activate

). The following options are available:

POSITIVE - An A-Scan signal crosses the gate


NEGATIVE - No A-Scan signal crosses the gate
OFF - No alarm is connected to the selected gate
Note: Gate A and Gate B can be configured so that either gate triggers the alarm.
4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

73

Chapter 3. Making Measurements

3.1.4b

Assigning Alarm Output Indicator Lights (SETUP-CONFIG2-OUTPUT SELECT)

As an option, one reading box may be configured as a virtual LED (see Setting Gate Alarms and Outputs on page 72), which is green
when there is no fault condition and red when the alarm has been triggered. This virtual LED corresponds to an OUTPUT, which is in-turn
assigned to a gate alarm. When an alarm is triggered, the virtual LED is illuminated (except when the GATE LOGIC is set to OFF). Use the
following procedure to indicate which gate activates the virtual LED:

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (


function.

3.

To configure the gate-alarm virtual LED operation, move the joystick up or down (

).

) to select the function titled OUTPUT SELECT. Press the center of the joystick (

) to activate the

). The following options are available:

A(+), A(-) - the virtual LED indicates the Gate A alarm is triggered (see previous page for description of + and - alarms)
B(+), B(-) - the virtual LED indicates the B Gate alarm is triggered (see previous page for description of + and - alarms)
A or B(+), A or B (-) - the virtual LED indicates either Gate alarm is triggered (see previous page for description of + and - alarms)
4.

74

After making your choice, press the center of the joystick (

) to deactivate the function.

USM Go Operators Manual

Chapter 3. Making Measurements

3.1.4c

Select Gate to Be Magnified When Function Toggle Is Pressed (SETUP-EVAL-MAGNIFY GATE)

The user can indicate the action to be taken when either end of the Function Toggle is pressed. One option magnifies the A-Scan display, so
the assigned gate spans the entire displayed range. To specify the gate that should be magnified on demand:

1.

In the SETUP menu, activate the EVAL submenu using the joystick (

2.

Use the joystick (


function.

3.

To select Gate A or the Gate B, move the joystick up or down (

).

) to select the function titled MAGNIFY GATE. Press the center of the joystick (

) to activate the

).

Note: Pressing the assigned end of Function Toggle, when FUNCTION1 or FUNCTION 2 is set to MAGNIFY GATE, magnifies the
display so the selected gate spans the entire displayed range.
4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

75

Chapter 3. Making Measurements

Using Angle Beam Probes

3.2

When connecting an angle beam probe to the instrument, adjustments must be made for probe characteristics as well as test piece geometry.
These adjustments include:

Probe angle
Probe X value = the distance from the probe Beam Index Point (BIP) to the front edge of its wedge
Test piece thickness
O-Diameter
3.2.1

Setting the Angle Beam Probe Parameters

To configure the instrument for an angle-beam probe, follow this procedure:


1.

In the SETUP menu, activate the EVAL submenu using the joystick (

2.

Use the joystick (


) to select the function titled PROBE ANGLE. Input the angle for the installed probe. Surface evaluation is
performed by installing the correct probe/wedge and setting this value to > 89.

).

Note: This function has both coarse and fine adjustment modes. Coarse adjustments are made with the function toggle, while fine
adjustments are made with the joystick.
76

USM Go Operators Manual

Chapter 3. Making Measurements

3.2.1

Setting the Angle Beam Probe Parameters (cont.)

3.

Use the joystick (


) to select the THICKNESS function, and input the thickness of the test piece. The thickness of a solid rod
should be input as 1/2 of its diameter.

4.

Use the joystick (


) to select the X VALUE function, and input the user-determined X value for the probe. This compensates for
the distance from the BIP to the front of the probe wedge.

5.

Use the joystick (


) to select the O DIAMETER function, and input the outside diameter (from 50 to 2000 mm) of a curved test
piece. When set to FLAT, curvature correction is not applied.

Note: Based on the values input for the PROBE ANGLE, THICKNESS, X VALUE, and O DIAMETER parameters, the instrument
operates in one of five modes. Each of these modes is represented by an icon in the icon area at the upper left corner of the screen.
See Figure 6 on page 19 for an illustration of these icons.

USM Go Operators Manual

77

Chapter 3. Making Measurements

3.2.2

Indicating Leg with Color

The leg in which a reflector is encountered, can be visually indicated on the instrument display using color. Setting the COLOR LEG
function, located in the EVAL submenu of the SETUP menu, to ON causes each ultrasonic time region to be displayed in a unique color.

3.3

Displaying Measured Results

The instrument is capable of displaying up to seven measured readings at one time. The displayed readings are selected using the EVAL
submenu located in the SETUP menu. The parameters available for display, which depends on the instrument configuration and the
evaluation mode, include the following:

A%A - Amplitude, as a % of full-screen height, of the highest echo falling within the width of Gate A
A%B - Amplitude, as a % of full-screen height, of the highest echo falling within the width of Gate B
Note: When S, D, P, or R readings are displayed, the Gate-Detection Mode setting for the referenced gate (A or B) is indicated by a
^ for Peak Mode or a / for Flank Mode.

SA - Sound-Path distance or duration, represented by the highest echo or the first flank to cross Gate A
SB - Sound-Path distance or duration, represented by the highest echo or the first flank to cross Gate B
SBA - Sound-Path distance or duration, from the highest echo or the first flank in Gate A to the echo in Gate B. The reading is only
available if Gate A and Gate B are set to ON.

78

USM Go Operators Manual

Chapter 3. Making Measurements

3.3

Displaying Measured Results (cont.)

DA - Material-thickness depth from the probe-contacted side of the test-piece surface to the reflector, represented by the Gate A echo.
DB - Material-thickness depth from the probe-contacted side of the test-piece surface to the reflector, represented by the Gate B echo.
LA - Leg number of the reflector, represented by the Gate A echo.
LB - Leg number of the reflector, represented by the Gate B echo.
PA - Projection distance from the probe BIP to the reflector, represented by the Gate A echo.
PB - Projection distance from the probe BIP to the reflector, represented by the Gate B echo.
RA - Projection distance from the probe BIP to the reflector, represented by the Gate A echo, minus the current X-VALUE.
RB - Projection distance from the probe BIP to the reflector, represented by the Gate B echo, minus the current X-VALUE.
A%rA - Amplitude of the signal crossing Gate A, as a percentage of the active EVAL MODE reference amplitude.
A%rB - Amplitude of the signal crossing Gate B, as a percentage of the active EVAL MODE reference amplitude.
dBrA - dB equivalent height difference between the signal crossing Gate A and the active EVAL MODE reference height.
dBrB - dB equivalent height difference between the signal crossing Gate B and the active EVAL MODE reference height.

USM Go Operators Manual

79

Chapter 3. Making Measurements

3.3

Displaying Measured Results (cont.)

Note: Reference results (identified with r) behavior based on EVAL MODE are as follows:

DAC - % Amplitude or dB compared to corresponding DAC curve point


TCG - % Amplitude or dB compared to TCG reference level
dB REF - % Amplitude or dB compared to reference level
DGS - % Amplitude or dB compared to curve/size selected
JISDAC - % Amplitude or dB compared to JIS LINE (H, M, or L) set to BOLD
NONE - % Amplitude or dB compared to gate threshold height
CLS - JIS CLASS (I, II, III, or IV). Available only when JISDAC evaluation mode is active.
ERS - Evaluates the reflected echo in DGS Mode, and calculates the Equivalent Reflector Size
GT - DGS test gain, which initializes the DGS curve maximum height at 80% FSH.
GR - DGS reference gain, which represents the instrument gain at which the reference echo peak reaches 80% FSH.
VIRTUAL LED - Reading box color changes from green to red when a gate alarm is triggered.
OFF - No reading is displayed in the reading box.
80

USM Go Operators Manual

Chapter 3. Making Measurements

3.3

Displaying Measured Results (cont.)

The measured readings can be displayed at the top of the display screen either in six small reading boxes and one large reading box or in
four large reading boxes. To set the reading box configuration:

1.

In the SETUP menu, activate the RESULTS2 submenu using the joystick (

2.

Determine the configuration of the reading boxes by setting the MODE function to either LARGE (four parameters displayed) or
SMALL (six parameters displayed in small boxes and a seventh parameter displayed in a large box).

3.

In the RESULTS and RESULTS2 submenus, access and set the READING 1 through READING 6 functions to select the desired
result to be displayed. When time or thickness readings are displayed, the detection method selected for that gate is indicated with a
^ (representing PEAK) or / (representing FLANK or JFLANK).

).

Note: Under certain conditions, while TCG reference points are being recorded, two measurement results boxes are automatically set (if
not so configured already) to display SA and A%A values.

USM Go Operators Manual

81

Chapter 3. Making Measurements

3.4

Locking the Gain Toggle and the Joystick

The gain toggle can be locked so that pressing it has no effect on the instrument.

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (

3.

Move the joystick left or right (


toggle is now locked.

4.

To unlock the gain toggle, change the setting of dB STEP to any value other than LOCK.

5.

After making your choice, press the center of the joystick (

).

) to select the function titled dB STEP. Press the center of the joystick (

) to activate the function.

) to change the value of dB STEP until the word LOCK appears as the dB STEP value. The gain

) to deactivate the function.

Joystick operation is disabled by assigning either FUNCTION1 or FUNCTION2 to a value of JOYSTICK LOCK, then pressing that end of
the Function Toggle. When the joystick is locked, the Lock icon (see Figure 6 on page 19) appears on the display. To unlock the joystick,
simply press the assigned end of the Function Toggle a second time.
IMPORTANT: When the joystick is locked, the center-press function still works.

82

USM Go Operators Manual

Chapter 3. Making Measurements

3.5

Setting the Gain

Instrument gain, which increases and decreases the height of a displayed A-Scan, is adjusted with the Gain Toggle. The instrument gain can
be adjusted while in any menu location, except when the dB STEP feature is set to LOCK.
Note: Pressing both buttons on the Gain Toggle simultaneously activates the AUTO80 function.

3.5.1

Changing the Gain-Adjustment Increment (dB STEP)

When adjusting the A-Scan gain, each press of the Gain Toggle increases or decreases the gain level by a dB increment equal to the
dB STEP value. Several values can be specified for dB STEP, including: a user-specified value known as the USER GAIN STEP. There
is also a Gain Toggle adjustment LOCK which prevents any gain adjustment by the user. To select one of the existing dB STEP values:

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (

3.

Move the joystick left or right (


) to change the value of dB STEP. Available increments include:
0.2 dB, 0.6 dB, 1.0 dB, 2.0 dB, 6 dB, a user-defined Gain Step (if one has been defined), and LOCK.

4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

).

) to select the function titled dB STEP. Press the center of the joystick (

) to activate the function.

) to deactivate the function.


83

Chapter 3. Making Measurements

3.5.2

Setting the User-Defined Gain Step (SETUP-GAIN-USER GAIN STEP)

When adjusting the A-Scan gain, each press of the Gain Toggle increases or decreases the gain level by the amount of the dB STEP. To
enter a user-specified gain step, known as USER GAIN STEP, complete the following steps:

1.

In the SETUP menu, activate the CONFIG2 submenu using the joystick (

2.

Use the joystick (


function.

3.

Move the joystick left or right (


dB STEP function is activated.

4.

After making your choice, press the center of the joystick (

84

).

) to select the function titled USER GAIN STEP. Press the center of the joystick (

) to activate the

) to change the value of USER GAIN STEP. This value is now available as an option whenever the

) to deactivate the function.

USM Go Operators Manual

Chapter 3. Making Measurements

3.6

Freezing the A-Scan Display

By using the FUNCTION1 or FUNCTION2 functions in the SETUP menu, you can program one end of the Function Toggle to FREEZE
the display. Then, whenever an A-Scan is active, pressing that end of the Function Toggle freezes the A-Scan display. The active A-Scan
will remain as it appeared when the toggle was pressed and the display will remain frozen until it is pressed again. While the display is
frozen, the displayed readings are based on the frozen echoes.

3.7

DAC/TCG Evaluation Mode (Optional)

Note: Optional evaluation features, such as DAC/TCG, are selected via the EVAL MODE function in the EVAL menu. The functions for
the selected evaluation mode then appears in the ACQUIRE menu.
Conventional measurements can be made using the Time Corrected Gain (TCG) and Distance Amplitude Correction (DAC) modes. The
functions for these modes are accessed through the DAC/TCG menu on the ACQUIRE menu bar. The DAC and TCG modes operate based
on a set of user-recorded data points. These points are recorded from the DAC/TCG menu as described below.
The TCG mode displays reflectors of equal size at equal A-Scan amplitudes, regardless of the depth of the reflector in the test material. This
is accomplished by adjusting the gain at different locations in the A-scan display, corresponding to different material depths, to compensate
for signal loss (or variation) due to attenuation, beam spread, or other factors.
The DAC mode displays all echoes at their true amplitude, without depth compensation. However, when operating in DAC mode, a
Distance Amplitude Correction curve is superimposed on the A-Scan display. The curve represents the constant reflector size at varying
material depth.
USM Go Operators Manual

85

Chapter 3. Making Measurements

3.8

Using DAC Mode (Optional)

When displayed, the DAC curve visually represents a line of constant reflector peaks over a range of material depths. Remember that in
DAC mode, the only deviation from traditional display and operation is the appearance of the DAC curve. All A-Scan echoes are displayed
at their non-compensated height. A DAC curve can be based on up to 16 data points (material depths).
A DAC curve is programmed using a series of same-reflector echoes at various depths, covering the range of depths to be inspected in the
test material. Because near field and beam spread vary according to transducer size and frequency, and materials vary in attenuation and
velocity, DAC must be programmed differently for different applications.
The minimum time difference between successive points is 60 ns. This corresponds to a distance of 0.18 mm (0.007 in.) at a velocity of
5900 m/s (.2323 in./s). Successive data points do not have to progressively decrease in amplitude. That is, the DAC/TCG curve does not
have to have a constantly descending slope.

86

USM Go Operators Manual

Chapter 3. Making Measurements

3.8.1

Recording the DAC Curve

DAC Curve points are typically recorded from a standard with equally-sized reflectors (holes) located at various material depths. The
primary echo from each of these points, for a total of up to 16 echoes, are recorded. When DAC mode is active, the instrument displays a
line that represents echo peaks for constant reflectors at varying material depth. This line is drawn as either a curved, linear in gain, line
based on a series of collected DAC points or a series of straight line segments that join these DAC points. Only one DAC curve can be
stored at a time. To program the DAC curve proceed as follows:

1.

In the ACQUIRE menu, activate the DAC/TCG submenu using the joystick (

2.

Couple the probe to the first reference point and use the GATE A START function to adjust Gate A so that it is broken by the primary
echo. If necessary, use the Gain Toggle to adjust the gain so that the echo crosses Gate A, and the highest peak in Gate A is at
approximately 80% of full-screen height. The highest peak must not be higher than 100% full-screen height.

).

Note: The AUTO 80 function allows for automatic application of gain to set the Gate A triggering echo to 80% of full screen height.
3.

While Gate A is lined up over the first reference echo, use the joystick (

) to select the function titled RECORD. Press the

center of the joystick (


) to activate the function. When the value of the RECORD function changes from 0 to 1, you have recorded
the first DAC Curve point, which is then treated as the reference echo. The amplitude value at which this point is recorded becomes the
reference amplitude value. However, the reference amplitude is updated if a new point is recorded with an earlier TOF.

USM Go Operators Manual

87

Chapter 3. Making Measurements

3.8.1

Recording the DAC Curve (cont.)

Note: If the Gate Detection Mode is set to PEAK, after the first DAC reference point is stored, two measurement result boxes are
automatically set, if not already configured, to display SA and A%A values.
4.

Repeat steps 2 and 3 to record additional DAC Curve points, up to a maximum of 16 points. A minimum of two DAC Curve points are
required to define the curve.

5.

After entering all of your points, press the center of the joystick (
the amplitude of the earliest TOF as the reference amplitude.

6.

In the SETUP menu, activate the DAC TYPE function using the joystick (

) to deactivate the function. The TCG/DAC data is stored, using

). Then, select one of the following options:

STRAIGHT: The DAC curve is displayed as a series of straight line segments that join the stored DAC points.
CURVED: The DAC curve is displayed as a curved line that is linear in gain, based on the stored DAC points.
Note: DAC curve points and status are stored with a data set. When recalled, the curve status is the same as when it was stored.

88

USM Go Operators Manual

Chapter 3. Making Measurements

3.8.1

Working with DAC Mode

In DAC mode the instrument uses the user-input reference points to create a curve representing the amplitudes of echoes representing
same-size reflectors at varying material depth. The recorded point data is stored until replaced or edited. To create a DAC curve and operate
in DAC mode:
1.

With the DAC/TCG menu accessed, select the SETUP submenu.

2.

Use the TCG/DAC MODE menu to activate the DAC function. The DAC Curve appears whenever operating in DAC mode.

3.

The DAC curve can appear either as a series of straight line segments joining the stored DAC points or as a curved (linear in gain) line
based on these points. Access the DAC TYPE function in the SETUP submenu to display the desired DAC curve type.

Note: To operate in TCG/DAC mode, a DAC curve must first be generated. After the DAC curve has been generated, the TCG mode can
be accessed by setting the TCG/DAC function value to TCG.
Note: When rectification is set to RF, reference lines are not displayed.

USM Go Operators Manual

89

Chapter 3. Making Measurements

3.9

Using TCG Mode (Optional)

When the optional TCG mode is in use, echoes from equally sized reflectors are shown at the same height on the A-Scan display. Before
using the TCG mode do the following:
1.

Make sure that the instrument/probe combination has been properly calibrated and that all instrument settings (PULSER,
RECEIVER, etc.) have been properly configured. Changing these settings after the TCG reference points have been entered will affect
the accuracy of your measurements.

2.

Up to 16 TCG points may be recorded, but a minimum of two points are required to define the curve. This process allows the
instrument to calculate and compensate for the effect of material depth on reflector echo height. Successive data points do not have to
progressively decrease in amplitude. That is, the DAC/TCG curve does not have to have a constantly descending slope.

3.9.1

Generating the TCG Reference Curve

TCG reference points are derived from the points used to create the DAC curve. Points are typically taken from a standard with equally
sized reflectors (holes) located at various material depths. The primary echo from each of these points, for a total of up to 16 echoes, are
recorded. When TCG is active, the instrument compensates for different material thickness by applying a varying gain level to echoes at
material depths other than the baseline depth.
Note: TCG reference points, curve, and status are stored with a data set. When recalled, curve status is the same as when it was stored.

90

USM Go Operators Manual

Chapter 3. Making Measurements

3.9.2

Working with TCG

In TCG mode the instrument uses the recorded reference points to calculate an amount of gain correction required to display each echo from
same-size reflectors at the same amplitude. The recorded reference point data is stored until replaced or edited. To use the stored reference
points and operate in TCG mode:
1.

Enter the SETUP menu, and then select the TCG/DAC MODE function.

2.

Activate the TCG/DAC MODE function, and select the TCG option.

3.

Select the TCG DISPLAY function, and turn this feature ON or OFF.

Note: The TCG curve begins at the first reference point recorded. It then proceeds horizontally from the amplitude of this first reference
point to the depth (i.e. time position) of the last reference point recorded.
Note: When rectification is set to RF, reference lines are not displayed.

USM Go Operators Manual

91

Chapter 3. Making Measurements

3.10 Adjusting DAC or TCG Display and Adding Offsets


After the DAC or TCG curves are displayed, the addition of guidelines offset from the reference line by a fixed or variable dB value
provides enhanced evaluating capabilities. Similarly, the TRANSFER CORR function applies dB compensation for the difference in
coupling conditions between the known standard and the test piece.

3.10.1

Defining DAC or TCG Curve Offsets (DAC/TCG-OFFSETS-MODE)

To add a series of guidelines, offset from the DAC or TCG curve by a definable dB increment (but not exceeding 24 dB from the reference
curve):
1.

Access the OFFSETS submenu in the DAC/TCG menu.

2.

Set the MODE function to FIXED for evenly spaced offset lines or to CUSTOM for user-controlled offset line placement. Set this
function to OFF to remove offset lines from the display.

3.

If FIXED offset MODE is selected, set the OFFSET function to the dB increment at which offset lines are drawn above and below the
DAC or TCG line.

4.

If CUSTOM offset MODE is selected, access the OFFFSETS2 submenu and input the positive or negative offset at which each of up
to four guidelines are drawn.

92

USM Go Operators Manual

Chapter 3. Making Measurements

3.10.2 Setting Transfer Correction (DAC/TCG-MAT ATTN-TRANSFER CORR)


To adjust the dB compensation for differences in ultrasonic coupling between the standard and the test piece:
1.

Access the MAT ATTN submenu in the DAC/TCG menu.

2.

Adjust the TRANSFER CORR function as required to compensate for coupling differences.

Note: When TRANSFER CORR is set to any value other than 0, the displayed gain value appears in a contrasting color, indicating that
the amount of applied gain differs from the listed instrument gain.

USM Go Operators Manual

93

Chapter 3. Making Measurements

3.11 Editing and Deleting DAC and TCG Reference Points


After your reference points are recorded, they may be individually deleted, their values may be manually adjusted, or new points may be
manually entered (as long as the total number does not exceed 16 points). To edit points or input additional points:
1.

With the DAC/TCG menu accessed, select the EDIT submenu.

2.

Activate the POINT function and select the desired point number (or NEW if adding a new point).

3.

Activate the POINT POS. function and adjust (or input for NEW points) the horizontal position of the point.

4.

Activate the POINT GAIN function to adjust (or input for NEW points) the applied gain (i.e. the vertical position) of the point. Note
that this adjustment applies regardless of whether the point is used for TCG reference or a DAC curve.

5.

Activate the ENTER function to confirm the adjusted point values (or the position of a new point). The DAC curve or the TCG
reference function is adjusted to match the edited input.

6.

To delete any single DAC point, press and hold the center of the joystick for the POINT, POINT POS, or POINT GAIN function while
the point to be deleted is active. When prompted, confirm the deletion. Note that the first point cannot be deleted, and that all DAC
curves must contain at least two points. Therefore, if a curve has only two points stored, neither can be deleted.

94

USM Go Operators Manual

Chapter 3. Making Measurements

3.12 Deleting a DAC Curve or TCG Reference Points


To delete a stored DAC curve or stored TCG reference points:
1.

With the DAC/TCG menu activated, select the SETUP submenu.

2.

Activate the DELETE CURVE function.

3.

Activate the DELETE CURVE function a second time, then confirm your selection.

4.

The statement in the function box changes to TCG/DAC MODE OFF.

3.13 DGS Evaluation Mode


Note: Evaluation features, such as DGS mode, are first selected by the user via the EVAL MODE function in the EVAL menu. The menu
for the selected evaluation mode is then available in the ACQUIRE menu bar.
The USM Go Distance Gain Sizing (DGS) feature is accessed through the DGS menu, which is located in the ACQUIRE menu bar. DGS
mode allows the user to use a particular probe to compare a reflector in a test piece with a known standard reflector. The DGS mode relies
on a reference curve based on a recorded reference point. The procedure for recording a reference point using the DGS Menu is described on
the next page.

USM Go Operators Manual

95

Chapter 3. Making Measurements

3.13

DGS Evaluation Mode (cont.)

Using the DGS mode (Distance Gain Size), you can compare the reflecting power of a natural flaw in the test object with that of a
theoretical flaw (e.g. a circular, disk-shaped equivalent reflector) at the same depth.
CAUTION!

You are comparing the reflecting power of a natural flaw with that of a theoretical flaw. No definitive conclusions
may be drawn about the natural flaw (roughness, inclined position, etc.).

The DGS diagram forms the basis for this comparison of the reflecting power. This diagram consists of a set of curves showing the
connection of three influencing variables:

Distance (D) between the probe and circular, disk-shaped equivalent reflector
Difference in gain (G) between various large circular, disk-shaped equivalent reflectors and an infinitely large backwall
Size (S) of the circular, disk-shaped equivalent reflector. The influencing variable S remains constant for one curve in a set
The advantage of the DGS method lies in the fact that you can carry out reproducible evaluations of small discontinuities. This
reproducibility is most important whenever you carry out an acceptance test.

96

USM Go Operators Manual

Chapter 3. Making Measurements

3.13

DGS Evaluation Mode (cont.)

In addition to the influencing variables already mentioned, these other factors affect the curve shape:

sound attenuation
transfer losses
amplitude correction value
probe
Also, the following probe parameters affect the curve shape:

element or crystal diameter


frequency
delay length
delay velocity
You can adjust these parameters in such a way that you can use the DGS method with many different probes and on various materials.

USM Go Operators Manual

97

Chapter 3. Making Measurements

3.13

DGS Evaluation Mode (cont.)

When the DGS mode is in use, echoes from equally-sized reflectors located at varying depths appear to lie along the DGS Reference Curve.
When operating in DGS mode, the DGS Reference Curve is shown on the display screen. Before using the DGS mode, do the following:

Calibrate the instrument/probe combination


Make all required instrument settings related to the pulser, receiver, and material velocity functions
The instrument prevents changes to certain settings after the DGS Reference Echo has been recorded and DGS is turned ON.

3.13.1 Specifying a Probe and Preparing to Record the Reference Echo


Before using the DGS mode to evaluate reflectors in test pieces, the characteristics of the attached probe must be specified, certain
characteristics of the reference standard must be input, and a reference echo must be stored. To specify the probe characteristics:
1.

Select the DGS menu on the ACQUIRE menu bar.

2.

Select the SETUP submenu, then the PROBE # function. Once activated, this function allows you to choose from a list of available
probe types . These are probes for which probe characteristics are already stored in the instrument, with the exception of the
user-defined probe (PROBE # 0).

98

USM Go Operators Manual

Chapter 3. Making Measurements

3.13.1 Specifying a Probe and Preparing to Record the Reference Echo (cont.)
3.

If the user-defined probe type (PROBE #0) is selected, you can then select the PROBE NAME function. Then use the Gain Toggle
and the Function Toggle to input a new name. Note that selecting any PROBE # value other than 0 will prevent you from modifying
the PROBE NAME or any of the other settings described in this section.

4.

If the user-defined probe type (PROBE #0) is selected, you must then select the DGS PROBE submenu and input the characteristics
for the connected probe:

XTAL FREQUENCY - The frequency rating of the probe


EFF. DIAMETER - The effective diameter rating of the probe element
DELAY VELOCITY - User-determined delay-line velocity
Note: These characteristics cannot be changed for any probes other than PROBE #0.

USM Go Operators Manual

99

Chapter 3. Making Measurements

3.13.2 Record the Reference Echo that Defines the DGS Curve
Prior to generating the DGS curve, a test standard with a known reflector must be used to define a reference point. Acceptable test standards
include these reference types:

BW - Backwall echo with reference defect size defined as infinity


SDH - Side-drilled hole with a reference defect size defined as the diameter of the hole
FBH - Flat-bottom hole with a reference defect size equal to the facial diameter of the hole
Follow these steps to record a reference echo:
1.

Select the REF ECHO submenu, then the REFERENCE TYPE function. Once activated, this function allows you to select one of the
three reference types described above, and to specify the size of the known standard reference flaw.

2.

Couple the probe to the known standard, capture the reference flaw so that its reflected echo is displayed on the instrument A-Scan,
and adjust the Gate A starting point to ensure that the resulting echo triggers the gate.

3.

Adjust the Gain Toggle until the reference flaw A-Scan peak measures 80% of FSH (A%A = 80%).

4.

With the probe coupled to the standard, and the reference flaw echo captured by Gate A, use the RECORD REF function to store a
DGS reference echo.

100

USM Go Operators Manual

Chapter 3. Making Measurements

3.13.2 Record the Reference Echo that Defines the DGS Curve (cont.)
Note: The AUTO 80 function automatically sets the Gate A triggering echo to 80% of full screen height.
Note: Only one DGS reference echo can be stored at a time. To delete the currently stored reference, access the REF CORR submenu,
select DELETE REF, and follow the on-screen prompts.
Note: When a DGS reference echo is stored, two measurement result boxes are automatically set, if not already configured, to display SA
and A%A values.
IMPORTANT: The following two adjustments should be made prior to recording a reference echo. Changing these values after the DGS
curve is generated will cause the curve to be deleted.

REF ATTEN (found in the REF CORR submenu) - Specify a sound attenuation value, in dB per inch or mm of material thickness, for the
material from which the known standard is made.

AMPL CORRECT (found in the REF CORR submenu) - Correction required when using an angle-beam type probe. This value is
specified on the probe data sheet.

USM Go Operators Manual

101

Chapter 3. Making Measurements

3.13.3 Display and Adjust the DGS Curve


After a reference echo has been recorded, the DGS curve is displayed simply by selecting the SETUP submenu in the DGS menu, then
setting the DGS MODE function to ON. Note that switching this value to OFF does not delete the curve - it simply removes the curve from
the display and disables the DGS mode. After the DGS curve is displayed, it can be adjusted using one of these three functions:

TEST ATTEN (found in the MAT ATTN submenu) - Specify a sound attenuation value, in dB per inch or mm of material thickness, for the
material from which the test piece is made.

TRANSFER CORR. (found in the MAT ATTN Submenu) - dB compensation for difference in coupling conditions between the known
standard and the test piece. Setting this to values other than zero causes a T to be added to the DGS icon and the displayed gain value
to appear in a contrasting color, indicating that the amount of applied gain differs from the listed instrument gain.

DGS CURVE (found in the SETUP submenu) - Positions the probe DGS curve based on the size of the reflector (flaw) being tested. The
setting usually depends on the largest acceptable flaw size.

102

USM Go Operators Manual

Chapter 3. Making Measurements

3.13.4

Evaluating Results in DGS Mode

After the DGS curve is recorded and displayed, by turning DGS mode ON, echoes are automatically compared to the recorded reference.
There are three ways this comparison can be made and two additional DGS-related results that can be displayed:

A%rA - Amplitude of the signal crossing Gate A as a percentage of the corresponding DGS curve amplitude.
dBrA - dB equivalent height difference between the signal crossing Gate A and the corresponding DGS curve height.
ERS - Evaluates the reflected echo and calculates the Equivalent Reflector Size.
Gt - DGS test gain, which initializes the DGS curve maximum height at 80% FSH.
Gr - DGS reference gain, which represents the instrument gain at which the reference echo peak reaches 80% FSH.
To aid in interpreting the DGS mode display, the instrument displays up to four curves representing fixed-gain offsets from the DGS curve.
These curves are enabled and positioned, by specifying the dB-equivalent amount they are offset above or below the DGS curve, by
accessing the OFFSETS submenu in the DGS menu.

USM Go Operators Manual

103

Chapter 3. Making Measurements

3.13.5 Locks and Error Messages


As long as a valid reference echo is stored, no functions can be changed which could cause an incorrect DGS evaluation. If an attempt is
made to change such a function, the following error message appears:

Function Locked - DGS Reference has been recorded


Likewise, the DGS evaluation must be switched OFF and the reference echo deleted when selecting a new probe (e.g. for a new test
application).

3.13.6

Validity of the DGS Method

Echo amplitude evaluations according to the DGS method are only reliable and reproducible in cases when:

The reference echo is received from the test object if possible. If this is not possible, it should be verified that the reference block is made
of the same material as the test object.

The evaluation is carried out using the same probe which was also used for recording the reference echo. Another probe of the same type
can be used only after recording a new reference echo.

Echo amplitudes for reflector distances smaller than half of the probe near-field length are subject to extreme variation, for physical
reasons due to interference phenomena effecting the area. Thus, evaluation results may fluctuate more than the usually permissible
2 dB. An evaluation according to the DGS method is possible, but it is not recommended in such a case.

104

USM Go Operators Manual

Chapter 3. Making Measurements

3.14 dB REF Evaluation Mode


Note: Evaluation modes, such as dB REF, are selected via the EVAL MODE function in the EVAL menu. The selected evaluation mode
menu then appears in the ACQUIRE menu bar.
When dB REF is activated, the amplitude of the highest echo in A-Gate becomes the reference echo against which subsequent echo
amplitudes are evaluated. At the time of the dB REF activation, the gain setting also becomes a reference against which subsequent gain
values are compared.
1.

Access the SETUP submenu in the dB REF menu.

2.

Locate the reference echo in Gate A and adjust the gain as required. Then, use the RECORD function to record the reference echo. The
function box now indicates that a reference has been stored. Also note the following:

The instrument gain and the reference echo amplitude are now shown in the function box titled REFERENCE.
The instrument automatically switches the dB REF MODE function to ON, as indicated on the display.
Setting MODE to OFF allows you to disable the dB REF mode without deleting the reference echo.
Use DELETE REF function to delete the current reference.

USM Go Operators Manual

105

Chapter 3. Making Measurements

3.14

dB REF Evaluation Mode (cont.)

IMPORTANT: Remember that the highest echo in Gate A and the GAIN setting, when dB REF is selected, become the reference amplitude
and gain values for as long as dB REF is activated.
Note: The reference echo amplitude must not exceed 100% of full-screen height.
After dB REF has been activated, the Gain-Display Window lists both the Reference Gain and Incremental Gain levels.
The Reference Gain remains constant throughout the dB REF session while the Incremental Gain value changes with the Gain Toggle.
After dB REF has been activated, any amplitude measurements are stated in relation to the reference echo amplitude. Available amplitude
readings when operating in dB REF mode are:

dBrA - dB difference between the reference echo and the highest echo to cross Gate A.
A%rA - Amplitude of the signal crossing Gate A, as a percentage of the reference amplitude.
dBrB - dB difference between the reference echo and the highest echo to cross Gate B.
A%rB - Amplitude of the signal crossing Gate B, as a percentage of the reference amplitude.

106

USM Go Operators Manual

Chapter 3. Making Measurements

3.15 AWS D1.1 Weld Rating Evaluation Mode


Note: Evaluating modes, such as AWS D1.1 / D1.5, are selected in the EVAL MODE function in the EVAL menu. The selected evaluation
mode menu then appears in the ACQUIRE menu bar.
This mode allows the analysis of welds according to AWS specifications D1.1 or D1.5 and provides a D1.1 or D1.5 rating. The AWS D1.1
mode is accessed via the ACQUIRE menu. The mode utilizes four AWS-specified variables, including:

A INDICATION - Gain (in dB) required to position an A-Scan echoes peak from the measured reflector) at an amplitude equal to the
reference amplitude (between 10 and 90% of full-screen height)

B REFERENCE - Gain (in dB) required to position an A-Scan echo peak from the reference reflector at the user-selected amplitude
(between 10 and 90% of FSH).

C ATTENUATION - Determined by subtracting 1 inch from the sound-path distance to the discontinuity, using the equation (inch units):
C = (SA1) 2. This compensates for sound loss from material attenuation along the sound path to the discontinuity.

D D1.1 RATING - Calculated based on the AWS formula: D = A-B-C.


Note: The dB value of A-Indication is automatically adjusted to match the amplitude of the B-Ref upon performing the dB rating
calculation.

USM Go Operators Manual

107

Chapter 3. Making Measurements

3.15

AWS D1.1 Weld Rating Evaluation Mode (cont.)

Before activating the AWS D1.1/D1.5 weld rating mode, be sure that all instrument settings are properly adjusted for the specific
measurement application. Then access the AWS D1.1 submenu in the ACQUIRE menu, and follow this procedure:
1.

Apply couplant and couple the probe to a suitable reference test standard.

2.

Ensure that Gate A is positioned over the desired echo. Then, adjust the gain until the peak of the desired reference echo reaches the
desired amplitude. Note that if the echo peak amplitude (A%A) does not fall between 10% and 90% of FSH, the point will not be
accepted.

3.

Use the B REFERENCE function to define the reference dB level.

4.

To evaluate a reflector in a test piece, couple the probe to the test piece. Now adjust the Gate A position, if required, so that it is over the
desired echo.

5.

Adjust the gain until the peak of the test-piece echo reaches the desired screen height.

Note: The dB value recorded for A-Indication is automatically adjusted based on the amplitude difference between it and the value
recorded for B-REF. It will usually not match the instrument dB gain setting.
6.

108

With the AWS D1.1 SETUP submenu displayed, note that the A, C, and D parameters automatically update to match the Gate A
triggering echo. Use the A INDICATION function to HOLD the current parameters. Note that the displayed RESULTS continue to
update while the A, C, and D parameters are held.

USM Go Operators Manual

Chapter 3. Making Measurements

3.16 JISDAC Evaluation Mode


Note: Evaluation modes, such as JISDAC, are selected in the EVAL MODE function in the EVAL menu. The selected evaluation mode
menu then appears in the ACQUIRE menu bar. While operating in JISDAC mode, the Gate A TOF MODE must be set to PEAK or
JFLANK.
Conventional measurements can be made using the JIS Distance Amplitude Correction (JISDAC) mode. This mode operates based on a set
of user-recorded data points. These points are recorded from the JIS/DAC menu, as described below.
The JISDAC mode displays all echoes at their true amplitude. However, when operating in JISDAC mode, a Distance Amplitude
Correction curve is superimposed on the A-Scan display. The curve represents constant reflector size at varying material depth.
When displayed, the JISDAC curve visually represents a line of constant reflector peaks over a range of material depths. Remember that in
JISDAC mode, the only deviation from traditional display and operation is the appearance of the JISDAC curve and a series of OFFSET
curves. All A-Scan echoes are displayed at their non-compensated height. A JISDAC curve can be based on up to 15 data points
(i.e. material depths). A JISDAC curve is programmed using a series of same-reflector echoes at various depths, covering the range of
depths to be inspected in the test material. Because near field and beam spread vary according to transducer size and frequency, and
materials vary in attenuation and velocity, JISDAC must be programmed differently for different applications.
The dynamic range of the JISDAC function is 60 dB, and the maximum curve slope is 12 dB per microsecond. Successive data points do
not have to progressively decrease in amplitude. That is, the JISDAC curve does not have to have a constantly descending slope.

USM Go Operators Manual

109

Chapter 3. Making Measurements

3.16.1 Recording the JISDAC Curve


JISDAC Curve points are typically taken from a standard with equally-sized reflectors (holes) located at various material depths. The
primary echo from each of these points, for a total of up to 15 echoes, are recorded. When JISDAC is active, the instrument displays a line
that represents echo peaks for constant reflectors at varying material depth. This line is drawn as either a curved (linear in gain) line based
on a series of collected JISDAC points or a series of straight line segments that join these JISDAC points. Only one JISDAC curve can be
stored at a time. To program the JISDAC Curve:
1.

Access the JISDAC menu, and select the RECORD submenu .

2.

Couple the probe to the first reference point and adjust Gate A so that it is broken by the primary echo. If necessary, use the Gain Toggle
to adjust the gain so that the echo crosses Gate A and the highest peak in Gate A is at approximately 80% of full-screen height. The
highest peak must not be higher than 100% full-screen height.

Note: The AUTO 80 function automatically adjusts the gain to set the Gate A triggering echo to 80% of full screen height. If desired, select
the AUTO 80 function in the RECORD submenu.
3.

110

While the Gate is lined up over the first reference echo, activate the RECORD function. When the value of the RECORD function
changes from 0 to 1, you have recorded the first JISDAC Curve point. Note that the first JISDAC point is treated as the reference echo.
The amplitude value at which this point is recorded becomes the reference amplitude value.

USM Go Operators Manual

Chapter 3. Making Measurements

3.16.1

Recording the JISDAC Curve (cont.)

Note: When the first JISDAC reference point is stored, two measurement result boxes will be automatically set (if not already configured) to
display SA and A%A values.
4.

Continue to record additional curve points, up to a maximum of 15 points (note that at least two JISDAC Curve points are required).

5.

When all of your points have been entered, select the FINISH function.

3.16.2

Working with JISDAC

In JISDAC mode, the instrument uses the user-input reference points to create a curve representing the amplitudes of echoes representing
same-size reflectors at varying material depth. To turn JISDAC evaluation mode ON or OFF:
1.

With the JISDAC menu accessed, select the SETUP submenu.

2.

Use the JISDAC function to select ON or OFF. The JISDAC Curve is displayed whenever the feature is turned ON.

USM Go Operators Manual

111

Chapter 3. Making Measurements

3.16.2a Interpreting JISDAC Lines and Classes


The JISDAC display contains a recorded reference line and five fixed offset lines. Three of the lines are labeled as follows:

H Line Connects the recorded reference points


M Line Fixed offset positioned 6 dB below the H line
L Line Fixed offset positioned 12 dB below the H line
Any one of these three lines can be identified as the reference from which measurements are based (see the BOLD LINE function below). In
addition, the remaining three offset lines are drawn at 6, 12, and 18 dB above the H line.
The regions of the A-Scan display bordered by the H, M, and L lines are identified with JIS Class numbers. These identifying numbers can
also be displayed as results. CLS reports the region that contains the peak of the Gate A triggering echo.

112

USM Go Operators Manual

Chapter 3. Making Measurements

3.16.2b Defining a BOLD LINE Measurement Reference (JISDAC-SETUP-BOLD LINE)


The user-defined measurement reference is selected as follows:
1.

Access the SETUP submenu in the JISDAC menu.

2.

Use the BOLD LINE function to select the H, M, or L line.

3.

Note that the selected line is now shown in bold on the A-Scan display. This line now serves as the reference level from which all
reference measurements (such as A%rA or dBrB) are made.

3.16.2c Setting TRANSFER CORRection (JISDAC-MAT ATTN-TRANSFER CORR)


The TRANSFER CORR function applies dB compensation for the difference in coupling conditions between the known standard and the
test piece.
1.

Access the MAT ATTN submenu in the JISDAC menu.

2.

Adjust the TRANSFER CORR function as required to compensate for coupling differences.

Note: When TRANSFER CORR is set to any value other than 0, the displayed gain value appears in a contrasting color, indicating that
the amount of applied gain differs from the listed instrument gain.

USM Go Operators Manual

113

Chapter 3. Making Measurements

3.16.3

Deleting a JISDAC Curve

To delete a stored JISDAC curve:

2.

With the JISDAC menu activated, select the SETUP submenu.


[no content intended for this page - proceed to next page]
Activate the DELETE CURVE function, and then confirm your selection.

3.

The entry in the JISDAC function box changes to OFF.

1.

114

USM Go Operators Manual

USM Go Operators Manual


HDR IN REPORT

MEMO/HDR
d

se
d

FILENAME

ACTION

FILENAME
U

se

MEMO IN REPORT
d

PARAM IN REPORT
se

to

nt

th
e

te

se

la
y

in

or
ta
s

et
er
s

or
t

re

po
r

file

im
ag
e

an

in

rt

file

m
e

se

in

ta

na

da

de
d

a
re
po

in

in
cl
u

ex
is
tin

po
r

in

re

an

ed

re
p

am

re
p

ar

or

tp

in

se
le
ct

ed

lu
d

ed

ti

lu
d

th
a

in
c

er

ti
s

in
c

th
a

he
ad

lu
d

tf
ile

or
t

in
c

tru
m
en

ta

di
sp

in
s

da

an

e
ASc

ac
tiv

ri
s

is

m
em

di
ta

m
em
o

or
e

na
m
e

ad
e

ed
it

he

fil
e
et
e

ne
w
rd
el

or

cr
ea

ex
is
tin

is
tin
g

if
an

cu
rre

st
or
e

ca
ll
o

to

at
e

cr
e

se

ex

te

cr
ea

th
e

to
st
or
e

to

ec
ify

if
a

sp

re

to

e,

se
d

st
or

to

to

ec
ifi
y

se
d

sp

to

se
d

MEMO EDIT
se
d

HEADER EDIT

tf

or

or
t
ile

re
p

re
p

or
t

Chapter 4. Data Sets & Reports

Chapter 4. Data Sets & Reports

The USM Go can store data set files and generate reports. To perform these functions, see Figure 15 below and proceed with this chapter.

IMAGE IN REPORT

REPORT

Figure 15: The FILES Menu


115

Chapter 4. Data Sets & Reports

4.1

The FILES Menu

To access the FILES menu and select the desired function, refer to Figure 15 on page 115 and complete the following steps:
1.

In the SETUP menu, activate the FILES submenu using the joystick (

2.

Use the joystick (

3.

Move the joystick left or right (

).

) to select the function titled ACTION. Press the center of the joystick (

) to activate the function.

) to select one of the following actions:

STORE DATASET - see Storing a New Data Set File on page 117
RECALL DATASET - see Recalling a Data Set File on page 119
DELETE DATASET - see Deleting a Data Set File on page 120
STORE REPORT - see Storing a Report on page 127
FAST REPORT - see Fast Report on page 128
4.

116

After making your choice, press the center of the joystick (

) to deactivate the function.

USM Go Operators Manual

Chapter 4. Data Sets & Reports

4.2

Working With Data Set Files

The current instrument settings, which includes most of the functional settings, can be stored as a data set file. When a stored data set file is
later recalled, all active function settings are modified to match those contained in the data set file. Also, if the A-Scan was stored in the data
set file, it is displayed and frozen on the display screen. After a data set is recalled, the newly active functional settings may then be
modified if desired. To perform data set operations:
After a data set file is recalled, the instrument settings may then be modified, but these changes only affect the stored data set file if the new
setting are stored with the same file name as the original data set file.
Note: The USM Go data set files can be recognized by the .UGO extension on the filenames.

4.2.1

Storing a New Data Set File

After selecting the STORE DATASET action (see The FILES Menu on page 116), proceed to the appropriate section on the following
page to either create a new file name or select an existing file name.

USM Go Operators Manual

117

Chapter 4. Data Sets & Reports

4.2.1a

File Selection Mode


) to select the function titled FILENAME. Press the center of the joystick once (

1.

Use the joystick (


in File Selection mode.

2.

Move the joystick up or down (

3.

Press the center of the joystick (

4.2.1b

2.

Move the joystick up or down (

118

) to deactivate the function.

) to select the function titled FILENAME. Press the center of the joystick twice (

Use the joystick (


in File Naming mode.

3.

) to select the desired file name from the list of SD card files.

File Naming Mode

1.

) to activate the function

) to activate the function

) to choose the first character in the desired file name. Then, move the joystick left or right

) to move the cursor to another character position in the text string. Repeat the process until you have entered the entire file name.

Press the center of the joystick (

) to deactivate the function.

USM Go Operators Manual

Chapter 4. Data Sets & Reports

4.2.2

Recalling a Data Set File

After selecting the RECALL DATASET action (see The FILES Menu on page 116), continue as follows:

) to select the function titled FILENAME. Press the center of the joystick (

1.

Use the joystick (


File Selection mode.

2.

After the list of data set files available on the SD card is displayed, move the joystick up or down (
set file name from the list.

3.

After selecting the desired data set file, press the center of the joystick (

USM Go Operators Manual

) to activate the function in

) to select the desired data

) to deactivate the function.

119

Chapter 4. Data Sets & Reports

4.2.3

Deleting a Data Set File

After selecting the DELETE DATASET action (see The FILES Menu on page 116), continue as follows:

) to select the function titled FILENAME. Press the center of the joystick (

1.

Use the joystick (

2.

Move the joystick up or down (


(

3.

120

) to activate the function.

) to choose the first character in the desired filename. Then, move the joystick left or right

) to move the cursor to a different character position in the text string.

Repeat step 2 until you have entered the entire file name. Then, press the center of the joystick (

) to deactivate the function.

USM Go Operators Manual

Chapter 4. Data Sets & Reports

4.2.4

Editing a Data Set File

Editing an existing data set file, requires a simple combination of the procedures described in the previous sections:
1.

Recall the data set file that you wish to edit (see Recalling a Data Set File on page 119).

2.

Using the methods described in previous chapters, modify the active functional settings as desired.

3.

Store the modified settings as a data set file (see Storing a New Data Set File on page 117). When choosing a filename, overwrite the
original data set file by choosing the same filename for the modified settings.

USM Go Operators Manual

121

Chapter 4. Data Sets & Reports

4.3

Creating a Memo

Memos may be attached to Data Set Files at the time the files are stored or to Reports at the time they are generated. After a file is stored,
the attached memo may be modified. To create or edit a memo:

1.

In the SETUP menu, activate the FILES submenu using the joystick (

2.

Use the joystick (

3.

A text area for the memo is now available on the display, with the cursor positioned for the first character. Move the joystick up or down

).

) to select the function titled MEMO EDIT. Press the center of the joystick (

(
) to enter the first character in the memo. Then, move the joystick left or right (
character position in the text string.
4.

122

Repeat step 3 until you have entered the entire memo. Then, press the center of the joystick (

) to activate the function.

) to move the cursor to a different

) to deactivate the function.

USM Go Operators Manual

Chapter 4. Data Sets & Reports

4.4

Including a Memo in a Report

After creating a memo, you can specify whether that memo is included in a report. To add or remove a memo from a report:

1.

In the SETUP menu, activate the FILES submenu using the joystick (

2.

Use the joystick (


function.

3.

Move the joystick left or right (

).

) to select the function titled MEMO IN REPORT. Press the center of the joystick (

) to activate the

) to select one of the following options:

YES - the memo is included with a report


NO - the memo is not included with a report
4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

123

Chapter 4. Data Sets & Reports

4.5

Creating a Report Header

Report Headers are attached to data set files at the time the files are stored. After a file is stored, the attached Report Header may be edited.
To create or edit a Report Header:
1.

In the SETUP menu, activate the FILES submenu using the joystick (

2.

Use the joystick (

3.

A text area for the header is now available on the display, with the cursor positioned for the first character. Move the joystick up or

).

) to select the function titled HEADER EDIT. Press the center of the joystick (

down (
) to enter the first character in the header. Then, move the joystick left or right (
character position in the text string.
4.

124

) to activate the function.

) to move the cursor to a different

Repeat step 3 until you have entered the entire header (nine characters maximum). Then, press the center of the joystick (
deactivate the function.

) to

USM Go Operators Manual

Chapter 4. Data Sets & Reports

4.6

Including a Header in a Report

After creating a header, you can specify whether that header is included in a report. To add or remove a header from a report:

1.

In the SETUP menu, activate the FILES submenu using the joystick (

2.

Use the joystick (


function.

3.

Move the joystick left or right (

).

) to select the function titled HDR IN REPORT. Press the center of the joystick (

) to activate the

) to select one of the following options:

YES - the header is included with a report


NO - the header is not included with a report
4.

After making your choice, press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

125

Chapter 4. Data Sets & Reports

4.7

Creating a Report

IMPORTANT: Before proceeding be sure an SD card is installed in the instrument (see Using the SD Slot, USB Connector & I/O
Connector on page 14).
A report, with contents specified by the user, can be generated and stored on the USM Go SD card. The following features of the active data
set file can either be omitted from or included in the report:

Header (use the HDR IN REPORT function)


Memo (use the MEMO IN REPORT function)
A-Scan Image (use the IMAGE IN REPORT function)
Instrument parameters (use the PARAM IN REPORT function)
After configuring the report as desired, proceed to the next page to store the report.

126

USM Go Operators Manual

Chapter 4. Data Sets & Reports

4.8

Storing a Report

To store the report specified in the previous section:


Note: If you have chosen to store the A-Scan image with the report, the image will be saved as a JPG file.
) to select the function titled FILENAME. Press the center of the joystick (

1.

Use the joystick (

2.

Move the joystick up or down (


(

) to activate the function.

) to choose the first character in the desired filename. Then, move the joystick left or right

) to move the cursor to a different character position in the text string.

3.

Repeat step 2 until you have entered the entire file name. Then, press the center of the joystick (

4.

If you have not already done so, assign the COPY function to one of the Function Toggle buttons (see Defining Function Toggle
Actions on page 37). Your report can now be saved by pressing the assigned toggle button, whenever you are in Acquisition Mode and
the cursor is in the menu bar or the full-screen display is active.

USM Go Operators Manual

) to deactivate the function.

127

Chapter 4. Data Sets & Reports

4.9

Fast Report

The Fast Report option performs the same basic function as the Store Report option. The difference is that, if you save the A-Scan screen
image along with the report, the image is saved as a bitmap file. This permits the save operation to be completed significantly faster, but you
should be aware that some SD card printers may not recognize the bitmap file. To use the Fast Report option:

) to select the function titled FILENAME. Press the center of the joystick (

1.

Use the joystick (

2.

Move the joystick up or down (


(

) to activate the function.

) to choose the first character in the desired filename. Then, move the joystick left or right

) to move the cursor to a different character position in the text string.

3.

Repeat step 2 until you have entered the entire file name. Then, press the center of the joystick (

4.

If you have not already done so, assign the COPY function to one of the Function Toggle buttons (see Defining Function Toggle
Actions on page 37). Your report can now be saved by pressing the assigned toggle button, whenever you are in Acquisition Mode and
the cursor is in the menu bar or the full-screen display is active.

128

) to deactivate the function.

USM Go Operators Manual

Chapter 5. Data Recorder Files

Chapter 5. Data Recorder Files


The Data Recorder (DR) menu functions are shown in Figure 16 below.

Figure 16: The DR Menu


USM Go Operators Manual

129

Chapter 5. Data Recorder Files

5.1

Naming the Data Recorder File

In the SETUP menu, use the joystick (


) to activate the DR menu (see Figure 16 on page 129). Then, proceed to the appropriate section
below to either create a new file name or select an existing file name.

5.1.1
1.

File Selection Mode

Use the joystick (


(

) to select the function titled FILENAME in the DR SETUP submenu. Press the center of the joystick once

) to activate the function in File Selection mode.

2.

Move the joystick up or down (

3.

Press the center of the joystick (

130

) to select the desired file name from the list of SD card files.
) to deactivate the function.

USM Go Operators Manual

Chapter 5. Data Recorder Files

5.1.2
1.

Use the joystick (


(

2.

) to select the function titled FILENAME in the DR SETUP submenu. Press the center of the joystick twice

) to activate the function in File Naming mode.

Move the joystick up or down (


(

3.

File Naming Mode

) to choose the first character in the desired file name. Then, move the joystick left or right

) to move the cursor to another character position in the text string. Repeat the process until you have entered the entire file name.

Press the center of the joystick (

USM Go Operators Manual

) to deactivate the function.

131

Chapter 5. Data Recorder Files

5.2

Configuring the Data Recorder File

After entering a name for the data recorder file, the following parameters must be specified:

TOP - the coordinates of the first cell in the file, using a row number and an column letter (e.g. 1A).
BOTTOM - the coordinates of the last cell in the file, using a row number and an column letter (e.g. 10B).
ADV DIRECTION - specifies the direction (DOWN or RIGHT) in which the DR advances after each thickness reading is recorded.
Use the usual joystick methods to select each of the above functions and program the desired values.

5.3

Creating the Data Recorder File

1.

After naming and configuring the DR file, use the joystick (

2.

Press the center of the joystick (

) to select the function titled CREATE.

) to create and activate a DR file that uses your previous selections.

IMPORTANT: After the DR file is created, the TOP and BOTTOM parameters cannot be changed. In fact, these functions are replaced on the
menu with listings of the NUM OF COLS and NUM OF ROWS, respectively.

132

USM Go Operators Manual

Chapter 5. Data Recorder Files

5.4

Viewing the Data Recorder File

To view the DR file that you have just activated, you must complete the following steps:
1.

In the SETUP menu, activate the DR submenu using the joystick (

2.

Use the joystick (

3.

Move the joystick left or right (

4.

Press and hold the center of the joystick (

5.

When in Acquire Mode, press the center of the joystick (

).

) to select the function titled DR VIEW. Then, press the center of the joystick (
) to select the ON option. Then, press the center of the joystick (

) to activate the function.

) to deactivate the function.

) to switch to the Acquire Menu.


) to enter the full-screen display mode.

As shown in Figure 17 on page 134, the DR file is displayed on the right side of the display screen.

USM Go Operators Manual

133

Chapter 5. Data Recorder Files

5.4

Viewing the Data Recorder File (cont.)

Figure 17: The DR File Display

134

USM Go Operators Manual

Chapter 5. Data Recorder Files

5.5
5.5.1

Using the Data Recorder File


Adding Data and A-Scan Images to a Cell

When the DR file is displayed (see Figure 17 on page 134), the name of the DR file is shown at the top of the grid and the currently selected
cell is highlighted. To add data or data plus an A-Scan image to a cell, proceed as follows:
IMPORTANT: You must delete any existing data from a cell before using the two SEND functions described below.

Move the joystick as required (

) to highlight any desired cell in the grid.

Use the Function 1 end of the function toggle (

) to send the current thickness data to the highlighted cell.

Use the Function 2 end of the function toggle (

) to send the current thickness data and the A-Scan image to the highlighted cell.

Note: A flag in the upper left corner of a cell (see Figure 17 on page 134) indicates that an A-Scan image is attached to the data.

Press both ends of the function toggle (

) simultaneously to delete the contents of the highlighted cell.

Note: In DR mode, the function keys behave as described above, regardless of any previous user-defined actions. However, if the
user-defined HOLD functions have been assigned to these keys, those functions are still available.

USM Go Operators Manual

135

Chapter 5. Data Recorder Files

5.5.2

Previewing A-Scan Images in a Cell

When you select a cell in the DR file that has an A-Scan image attached, the attached image is loaded and the instrument displays a frozen
preview of the saved image with the corresponding gate positions (see Figure 18 on page 137).
While a frozen A-Scan image is displayed, no new data is taken until you resume live data input by one of the following actions:

Delete both the stored data and the attached A-Scan image from selected the cell
Navigate to an empty cell
As soon as one of the above actions is taken, the instrument unfreezes and resumes taking data in the manner specified by the current setup
configuration.

136

USM Go Operators Manual

Chapter 5. Data Recorder Files

5.5.2

Previewing A-Scan Images in a Cell (cont.)

Figure 18: Frozen A-Scan Image Attached to Selected Cell

USM Go Operators Manual

137

Chapter 5. Data Recorder Files

[no content intended for this page - proceed to next page]

138

USM Go Operators Manual

Appendix A. Specifications
Note: All instrument specifications listed in this appendix are subject to change without prior notice. Also, see EN 12668 Specifications
on page 153.

A.1 LCD Display


Active Area:

108.0 mm (W) x 64.8 mm (H), 5.0 Diagonal

Size:

5.0

Resolution:

800 (W) x 480 (H) pixel

Contrast Ratio:

Brightness:

200 cd/m2

USM Go Operators Manual

139

A.2 Connectors
Probe Connectors:

2, LEMO-00

UT Output Connector:

SAP output, with an added ALARM pin

USB Interface:

Micro USB connector

SD-Card Connector:

Full-size SD card slot, accommodates all standard SD cards

Range:

14,016 mm (552) for longitudinal wave in steel

Display Delay:

-15 s to 3500 s

Probe Delay:

0 to 1000 s

Sound Velocity:

1000 to 16,000 m/s

PRF:

Manually adjusted from 5 to 2000 Hz or automatically optimized,


3 automatic adjustment modes: Auto Low, Auto Med, Auto High

140

USM Go Operators Manual

A.3 Pulser
Note: All pulser measurements were taken according to EN12668 specifications.
Pulser Mode:

Standard: simulated spike


Optional: uni-polar square wave via software control

Pulser Voltage (SQ Mode):

120 V to 300 V, with a 10 V step and tolerance of 10%

Pulser Falling/Rising Time:

10 ns maximum

Pulser Width (SQ Mode):

30 ns to 500 ns, with a 20 ns step and a tolerance of 10%

Pulser Amplitude (SQ Mode):

Low: 120 V
High: 300 V

Pulser Energy (Spike Mode):

Low: 30 nS
High: 100 nS

Damping:

50 ohms or 1000 ohms

USM Go Operators Manual

141

A.4 Receiver
Digital Gain:

Dynamic range of 110 dB, with a 0.2 dB step

Analog Bandwidth:

0.9 to 20 MHz

Equivalent Input Noise:

<80 nV/ Hz , across the full bandwidth

Recovery Time:

Target of <1 s (no EN12668 specification is given for this parameter)

Input Linearity:

2% by method E317, for output data on all 4 ADCs

Filters:

Broad Band: 0.5 to 17.2 MHz


1, 2 and 2.25 MHz: 0.9 to 5.2 MHz
4, 5 MHz: 2.4 to 7.8 MHz
10 MHz: 6.3 to 13.5 MHz
13, 15 MHz: 10.5 to 17.3 MHz

142

USM Go Operators Manual

A.5 Gates
Independent Gates:

2 Gates (A and B), Gate B supports triggering by Gate A

Rectification:

POS (positive)
NEG (negative)
FW (full wave)
RF

Measurement:

Peak
Flank
J-Flank

A.6 Memory
Capacity:

2 GB, SD card

Data Sets:
Reports:

USM Go Operators Manual

jpg or bmp A-Scan images

143

A.7 Environmental
Battery:

Life: 6 hr per full charge


Charging (standard): Internal
Charging (optional): External adapter
Level: proportional battery gauge

Battery Charger:

Universal AC input (100 to 240 VAC, 50-60 Hz); meets CCC, CE, UL, CSA and PSE requirements

Size:

175 mm (W) x 111 mm (H) x 50 mm (D)

Weight:

1 kg with battery

Languages:

English, German, French, Spanish, Chinese, Japanese

144

USM Go Operators Manual

A.8 Protection
Damp Heat & Humidity (Storage):

10 Cycles: 10 hr at +60C down to +30C, 10 hr at +30C up to +60C,


Transition within 2 hr (507.4)

Temperature Shock (Storage):

3 Cycles: 4 hr at -20C up to +60C, 4 hr at +60C,


Transition within 5 minutes (503.4 Procedure II)

Vibration:

General Exposure: 1 hr each axis, 514.5-5 Procedure I, Annex C, Figure 6

Shock:

6 Cycles each axis, 15 g, 11 ms, half-sine (516.5 Procedure I)

Loose Cargo (in Shipping Container):

514.5 Procedure II

Transit Drop (Packaged for Shipment): 26 Drops, 516.5 Procedure IV


Dust Proof/Dripping Water Proof:

As per the IEC 529 specification for IP67 classification

Operating Temperature Range:

0 to 55C

Storage Temperature Range:

-20 to +60C, 24 hr with battery

USM Go Operators Manual

145

A.8 Protection (cont.)


Compliance:

EMC/EMI:
Ultrasound:

EN 55011
EN 61000-6-2:2001
EN 12668
ASTM E1324
E317
ANSI/NCSL Z 540-1-1994
MIL STD 45662A
MIL STD 2154

A.9 USM Go Options


AWS Option:

AWS sizing tool, according to AWS D1.1 structural welding code

DAC Option:

DAC sizing tool, 16 points,


Complaint with: EN 1712 - EN 1713 - EN1714
ASTM E164
ASME & ASME III
JIS Z3060
TCG: 120 dB Dynamic
110 dB/s slope

146

USM Go Operators Manual

A.9 USM Go Options (cont.)


DGS Option:

DGS sizing tool,


Complaint with: EN 1712 - EN 1713 - EN1714
ASTM E164

Embedded Data Logger Option:

Grid file creation

Square Wave Pulser Option:

Allows fine-tuning of pulser parameters, Voltage adjustment from 120 to 300 V per 10 V steps,
pulse width adjustment from 30 to 500 ns per 10 ns steps

Phantom Option:

Phantom PRF will help to identify a ghost echo caused by multiple reflections in low-attenuation
materials

USM Go Operators Manual

147

[no content intended for this page - proceed to next page]

148

USM Go Operators Manual

Appendix B. Environmental Compliance

Appendix B. Environmental Compliance


This appendix contains information on the following topics:

WEEE Directive (see Section B.1 on page 150)


Battery disposal (see Section B.2 on page 151)

USM Go Operators Manual

149

Appendix B. Environmental Compliance

B.1 Waste Electrical and Electronic Equipment (WEEE) Directive


GE Sensing & Inspection Technologies is an active participant in Europes Waste Electrical and Electronic Equipment (WEEE) take-back
initiative, directive 2002/96/EC.
The equipment that you bought has required the extraction and use of natural resources for its production. It may
contain hazardous substances that could impact health and the environment.
In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural
resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of
the materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration.
Visit www.ge.com/inspectiontechnologies for take-back instructions and more information about this initiative.

150

USM Go Operators Manual

Appendix B. Environmental Compliance

B.2 Battery Disposal


This product contains a battery that cannot be disposed of as unsorted municipal waste in the European Union. See
the product documentation for specific battery information. The battery is marked with this symbol, which may
include lettering to indicate cadmium (Cd), lead (Pb), or mercury (Hg). For proper recycling return the battery to your
supplier or to a designated collection point.

B.2.1 What do the Markings Mean?


Batteries and accumulators must be marked (either on the battery or accumulator or on its packaging, depending on size) with the
separate collection symbol. In addition, the marking must include the chemical symbols of specific levels of toxic metals as follows:

Cadmium (Cd) over 0.002%


Lead (Pb) over 0.004%
Mercury (Hg) over 0.0005%

USM Go Operators Manual

151

Appendix B. Environmental Compliance

B.2.2 The Risks and Your Role in Reducing Them


Your participation is an important part of the effort to minimize the impact of batteries and accumulators on the environment and on human
health. For proper recycling you can return this product or the batteries or accumulators it contains to your supplier or to a designated
collection point.
Some batteries or accumulators contain toxic metals that pose serious risks to human health and to the environment. When required, the
product marking includes chemical symbols that indicate the presence toxic metals: Pb for lead, Hg for mercury, and Cd for cadmium.

Cadmium poisoning can result in cancer of the lungs and prostate gland. Chronic effects include kidney damage, pulmonary
emphysema, and bone diseases such as osteomalcia and osteoporosis. Cadmium may also cause anemia, discoloration of the teeth, and
loss of smell (anosmia).

Lead is poisonous in all forms. It accumulates in the body, so each exposure is significant. Ingestion and inhalation of lead can cause
severe damage to human health. Risks include brain damage, convulsions, malnutrition, and sterility.

Mercury creates hazardous vapors at room temperature. Exposure to high concentrations of mercury vapor can cause a variety of severe
symptoms. Risks include chronic inflammation of mouth and gums, personality change, nervousness, fever, and rashes.

152

USM Go Operators Manual

Appendix C. EN 12668 Specifications

Appendix C. EN 12668 Specifications


The EN 12668 specifications for the USM Go are listed in Table 2 below.
Table 2: EN 12668-1:2000 Specifications
Paragraph Parameter
8.2

9.3.2

9.3.3

9.3.4

LSL

Typ.

USL

Units Test Conditions

Timebase Stability Against Temperature

-1

%FS/C After 30 minute warm-up

Amplitude Stability Against Temperature

-5

%FS/C After 30 minute warm-up

Timebase Stability After Warm Up

-1

%FS

After 30 minute warm-up

Amplitude Stability After Warm Up

-2

%FS

After 30 minute warm-up

Timebase Display Jitter

-1

%FS

After 30 minute warm-up

Amplitude Display Jitter

-2

%FS

After 30 minute warm-up

Timebase Stability Against Voltage Variation

-1

%FS

Amplitude Stability Against Voltage Variation

-2

%FS

8.3.2

Pulse Repetition Frequency Error

-20

20

% Err

8.3.3

Pulser Output Impedance

8.3.4

*Pulser Spectrum

USM Go Operators Manual

<50

Ohm

See Chart

153

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.4.2

154

Loaded Pulse Voltage

LSL

Typ.

USL

-140

-156

-172

Units Test Conditions


V

Damping = 50, Voltage = Low, Energy = Low, RepRate = 15

-139

-154

-169

Damping = 50, Voltage = Low, Energy = Low, RepRate = 2000

-171

-190

-209

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 15

-171

-190

-209

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 2000

-103

-114

-125

Damping = 50, Voltage = Low, Energy = High, RepRate = 15

-102

-113

-124

Damping = 50, Voltage = Low, Energy = High, RepRate = 2000

-115

-128

-141

Damping =1000, Voltage = Low, Energy = High, RepRate =15

-115

-128

-141

Damping =1000, Voltage = Low, Energy = High, RepRate =2000

-167

-186

-205

Damping =50, Voltage = High, Energy = Low, RepRate =15

-167

-185

-204

Damping =50, Voltage = High, Energy = Low, RepRate =200

-209

-232

-255

Damping =1000, Voltage = High, Energy = Low, RepRate =15

-209

-232

-255

Damping =1000, Voltage = High, Energy = Low, RepRate =2000

-185

-206

-227

Damping =50, Voltage = High, Energy = High, RepRate =1

-185

-205

-226

Damping =50, Voltage = High, Energy = High, RepRate =2000

-211

-234

-257

Damping =1000, Voltage = High, Energy = High, RepRate =15

-211

-234

-257

Damping =1000, Voltage = High, Energy = High, RepRate =2000

USM Go Operators Manual

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.4.2

Pulse Rise Time

USM Go Operators Manual

LSL

Typ.

USL

Units Test Conditions

10

nSec

10

nSec

Damping = 50, Voltage = Low, Energy = Low, RepRate = 2000

10

nSe

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 15

10

nSec

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 2000

10

nSec

Damping = 50, Voltage = Low, Energy = High, RepRate = 1

10

nSec

Damping = 50, Voltage = Low, Energy = High, RepRate = 2000

10

nSec

Damping =1000, Voltage = Low, Energy = High, RepRate =15

10

nSec

Damping =1000, Voltage = Low, Energy = High, RepRate =2000

10

nSec

Damping =50, Voltage = High, Energy = Low, RepRate =15

10

nSec

Damping =50, Voltage = High, Energy = Low, RepRate =2000

10

nSec

Damping =1000, Voltage = High, Energy = Low, RepRate =15

Damping = 50, Voltage = Low, Energy = Low, RepRate = 15

10

nSec

Damping =1000, Voltage = High, Energy = Low, RepRate =2000

10

nSec

Damping =50, Voltage = High, Energy = High, RepRate =15

10

nSec

Damping =50, Voltage = High, Energy = High, RepRate =2000

10

nSec

Damping =1000, Voltage = High, Energy = High, RepRate =15

10

nSec

Damping =1000, Voltage = High, Energy = High, RepRate =2000

155

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.4.2

156

Pulse Duration

LSL

Typ.

USL

Units Test Conditions

16

18

20

nSec

Damping = 50, Voltage = Low, Energy = Low, RepRate = 15

16

18

20

nSec

Damping = 50, Voltage = Low, Energy = Low, RepRate = 2000

27

30

33

nSec

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 15

28

31

34

nSec

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 2000

57

63

69

nSec

Damping = 50, Voltage = Low, Energy = High, RepRate = 15

57

63

69

nSec

Damping = 50, Voltage = Low, Energy = High, RepRate = 2000

93

103

113

nSec

Damping =1000, Voltage = Low, Energy = High, RepRate =1

94

104

114

nSec

Damping =1000, Voltage = Low, Energy = High, RepRate =2000

16

18

20

nSec

Damping =50, Voltage = High, Energy = Low, RepRate =1

16

18

20

nSec

Damping =50, Voltage = High, Energy = Low, RepRate =200

28

31

34

nSec

Damping =1000, Voltage = High, Energy = Low, RepRate =15

28

31

34

nSec

Damping =1000, Voltage = High, Energy = Low, RepRate =2000

57

63

69

nSec

Damping =50, Voltage = High, Energy = High, RepRate =1

57

63

69

nSec

Damping =50, Voltage = High, Energy = High, RepRate =2000

94

104

114

nSec

Damping =1000, Voltage = High, Energy = High, RepRate =15

94

104

114

nSec

Damping =1000, Voltage = High, Energy = High, RepRate =2000

USM Go Operators Manual

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.4.2

Pulse Reverberation

USM Go Operators Manual

LSL

Typ.

USL

Units Test Conditions

nSec

Damping = 50, Voltage = Low, Energy = Low, RepRate = 15

nSec

Damping = 50, Voltage = Low, Energy = Low, RepRate = 2000

nSec

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 15

nSec

Damping = 1000, Voltage = Low, Energy = Low, RepRate = 2000

nSec

Damping = 50, Voltage = Low, Energy = High, RepRate = 15

nSec

Damping = 50, Voltage = Low, Energy = High, RepRate = 2000

nSec

Damping =1000, Voltage = Low, Energy = High, RepRate =15

nSec

Damping =1000, Voltage = Low, Energy = High, RepRate =2000

nSec

Damping =50, Voltage = High, Energy = Low, RepRate =15

nSec

Damping =50, Voltage = High, Energy = Low, RepRate =2000

nSec

Damping =1000, Voltage = High, Energy = Low, RepRate =15

nSec

Damping =1000, Voltage = High, Energy = Low, RepRate =2000

nSec

Damping =50, Voltage = High, Energy = High, RepRate =15

157

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.4.2

LSL

Pulse Reverberation

8.4.2

Crosstalk Pulser to Receiver

8.4.3

Dead Time After Transmitter Pulse

8.4.4

Dynamic Range

8.4.5

Receiver Input Impedance

80

100

Typ.

USL

nSec

Damping =50, Voltage = High, Energy = High, RepRate =2000

nSec

Damping =1000, Voltage = High, Energy = High, RepRate =15

nSec

Damping =1000, Voltage = High, Energy = High, RepRate =2000

80

dB

10

uSec

Measured in worst case frequency setting

dB

Measured in worst case frequency setting

100
950
0

8.4.6

158

Distance Amplitude Correction

-1.5

Units Test Conditions

Ohm

(Rmax gain - Rmin gain)/Rmax gain

0.1

40

Ohm

1.03

nF

0.15

1.5

Real Impedance at 4MHz

Imaginary Impedance at 4MHz


Input Capacitance
(Cmax gain - Cmin gain)/Cmax gain

dB

Maximum error between TCG curve and actual TCG correction

USM Go Operators Manual

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.5.2

Amplifier Frequency Response

USM Go Operators Manual

LSL

Typ.

USL

Units Test Conditions

0.72

0.76

0.80

MHz

Center Frequency (Geometric Mean), 1, 2 or 2.25 MHz selected

1.78

1.87

2.06

MHz

Bandwidth, 1, 2 or 2.25 MHz selected

4.51

4.75

4.99

MHz

Center Frequency (Geometric Mean), 4 or 5 MHz selected

4.56

5.07

5.58

MHz

Bandwidth, 4 or 5 MHz selected

8.79

9.25

9.71

MHz

Center Frequency (Geometric Mean), 10 MHz selected

6.17

6.85

7.54

MHz

Bandwidth, 10 MHz selected

12.25

12.89

13.53

MHz

Center Frequency (Geometric Mean), 13 or 15 MHz selected

5.02

5.58

6.14

MHz

Bandwidth, 13 or 15 MHz selected

3.84

4.27

4.69

MHz

Center Frequency (Geometric Mean), BroadBand selected

14.54

16.16

17.78

MHz

Bandwidth, BroadBand selected

159

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.5.3

9.5.4

160

Equivalent Input Noise

Accuracy of Calibrated Attenuator

LSL

Typ.

USL

Units Test Conditions

50

80

nV/
1, 2 or 2.25 MHz selected
sqrt Hz

37

80

nV/
4 or 5 MHz selected
sqrt Hz

33

80

nV/
10 MHz selected
sqrt Hz

40

80

nV/
13 or 15 MHz selected
sqrt Hz

45

80

nV/
BroadBand selected
sqrt Hz

-1

0.7

dB

Cumulative Error over 20dB Range

-2

0.7

dB

Cumulative Error over 60dB Range

USM Go Operators Manual

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
9.5.5

Linearity of Vertical Display

LSL

Typ.

USL

Units Test Conditions

-2

0.6

%FSH 1, 2 or 2,25 MHz selected

-2

0.6

%FSH 4 or 5 MHz selected

-2

-0.9

%FSH 10 MHz selected

-2

-1.25

%FSH 13 or 15 MHz selected

-2

0.18

%FSH BroadBand selected

8.7.2

Linearity of Timebase

0.03

0.5

%FSW

8.7.3

Digitization Sampling Error

-5

-4

%FSH

Dimensions

USM Go Operators Manual

17.1

CM

Height

28.2

CM

Width

15.9

CM

Depth

3.8

KG

Weight (With Battery)

161

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
8.7.3

LSL

Battery Life

Operating Temperature Range

Typ.

USL

Rectification Modes

Units Test Conditions


Hr

Instrument will automatically shut down when batteries are too


low for reliable operation

50

DegC

2000

Hz

Continuously Variable

CM

Diagonal

FW
HWP
HWN
RF

Pulse Repetition Frequency


Display Dimensions

15
16.5
640 x 480

Number of Pixels to Display an A-Scan


A-Scan Graticule Major Divisions

162

Pixels

512
None, 5 or 10

Vertical and Horizontal, User selectable

A-Scan Graticule Minor Horizontal Divisions

50

Displayed along the baseline

A-Scan Graticule Minor Vertical Divisions

50

Displayed on the center vertical

USM Go Operators Manual

Appendix C. EN 12668 Specifications

Table 2: EN 12668-1:2000 Specifications (cont.)


Paragraph Parameter
8.7.3

Velocity Range

Display Delay Range

LSL

Typ.

USL

Units Test Conditions

0.0098

0.6299

in/uS

250

16000

M/S

-15

3500

uS

TCG Range

40

dB

TCG Maximum Slope

dB/uS

TCG Maximum Number of Points

15

USM Go Operators Manual

163

Appendix C. EN 12668 Specifications

164

USM Go Operators Manual

Index

A
Acquire Menu
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Acquire Mode
Accessing the Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Alarms
Calibration Reminder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Defining Logic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Indicator Light, Assigning . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Amplitude Readings in dB REF Mode . . . . . . . . . . . . . . . . . . . 106
Angle Beam Probe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 76

USM Go Operators Manual

A-Scan
Flank/Peak of Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Freezing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Full-Screen Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Previewing Saved DR File Image . . . . . . . . . . . . . . . . . . . . 136
Saving in Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Saving to DR File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Selecting a Rectification Mode . . . . . . . . . . . . . . . . . . . . . . . 50
Setting the Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Setting the Display Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Setting the Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Setting the Reject Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Autocal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Automatic Gate Threshold (AGT) . . . . . . . . . . . . . . . . . . . . . 65, 70
AWS D1.1 Mode
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108

165

Index

B
Batteries
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Level Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Brightness, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Checking Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Preliminary Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Reminder Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Using Autocal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Case, Rear and Side Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Color
Setting for A-Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Setting for Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Setting for Indicating Leg . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Config1 Menu Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

166

Connectors
I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
USB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
COPY Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Creating
Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Memo. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Customer Support Centers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

D
DAC Mode
Deleting a DAC Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Recording the DAC Curve . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Working With . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

USM Go Operators Manual

Index

DAC/TCG Mode
Adding Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Adding/Deleting Reference Points . . . . . . . . . . . . . . . . . . . . 94
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Setting Transfer Correction . . . . . . . . . . . . . . . . . . . . . . . . . 93
Damping Level, Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Data Recorder Files
Creating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Naming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Viewing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Data Set Files
Deleting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Editing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Recalling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Storing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Working With . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Date of Publication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Date, Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

USM Go Operators Manual

dB REF Mode
Amplitude Readings Available . . . . . . . . . . . . . . . . . . . . . . 106
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
dB Step Gain Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Decimal Convention, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Delay, Setting for Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Deleting Data Set Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
DGS Mode
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Display & Adjust Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Evaluating Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Influencing Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Locks & Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Recording the Reference Echo . . . . . . . . . . . . . . . . . . . . . . 100
Specifying a Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Validity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104

167

Index

Display
Appearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Freezing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Measured Results Configuration . . . . . . . . . . . . . . . . . . . . . 81
Measured Results Parameters . . . . . . . . . . . . . . . . . . . . . . . . 78
Screen Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Screen Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Selecting the Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Setting the A-Scan Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Setting the Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Setting the Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Disposal
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Electronic Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

E
Editing Data Set Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
EN 12668 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Energy, Setting for Pulser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

168

Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 144


Error Messages in DGS Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Evaluation Mode
AWS D1.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
DAC/TCG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
dB REF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
DGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
JISDAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Options Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
TCG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90

F
Fast Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Features, Display Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
FILES Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Flank, A-Scan Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Freezing the A-Scan Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Frequency, Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Full-Screen A-Scan Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

USM Go Operators Manual

Index

Function Toggle
COPY Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Defining the Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Magnify Gate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

G
Gain
Changing the dB Step Value. . . . . . . . . . . . . . . . . . . . . . . . . 83
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
User-Defined Step Value . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Gain Toggle
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Gate A
see Gates
Gate Alarms
see Alarms
Gate B
see Gates

USM Go Operators Manual

Gates
Adjusting in Full-Screen Mode . . . . . . . . . . . . . . . . . . . . . . . 11
Adjusting the Width. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Automatic Threshold (AGT) . . . . . . . . . . . . . . . . . . . . . . 65, 70
Configuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Magnify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Setting the Starting Point. . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Setting the Threshold. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Grid, Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

H
HDR IN REPORT Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Header
Creating for Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Including in Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125

I
I/O Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

169

Index

Icons
Battery Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Complete Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
IMAGE IN REPORT Function . . . . . . . . . . . . . . . . . . . . . . . . . 126
Indicating Leg Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Indicator Light, Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

J
JFlank, A-Scan Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
JISDAC Mode
Defining a Bold Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Deleting a JISDAC Curve. . . . . . . . . . . . . . . . . . . . . . . . . . 114
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Interpreting Lines and Classes . . . . . . . . . . . . . . . . . . . . . . 112
Recording the JISDAC Curve. . . . . . . . . . . . . . . . . . . . . . . 110
Setting the Transfer Correction. . . . . . . . . . . . . . . . . . . . . . 113
Working With . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111

170

Joystick
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Joy Control parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

K
Keypad
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 7
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

L
Language, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Locking the Gain Toggle or Joystick . . . . . . . . . . . . . . . . . . . . . . 82
Locks in DGS Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104

M
Magnify Gate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Measured Results
Available Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Configuration of Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Evaluating in DGS Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 103

USM Go Operators Manual

Index

Memo
Creating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Including in Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
MEMO IN REPORT Function . . . . . . . . . . . . . . . . . . . . . . . . . 126
Menu System
Acquire Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Evaluation Mode Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Multi-Key Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

N
Naming Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . 130

O
Offsets, Adding to DAC/TCG Curve. . . . . . . . . . . . . . . . . . . . . . 92
Operator Training. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Options, Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Orientation
Instrument Photos . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

USM Go Operators Manual

P
PARAM IN REPORT Function . . . . . . . . . . . . . . . . . . . . . . . . . 126
Part Number, Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Peak, A-Scan Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Phantom PRF Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Positioning the Gates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Power
Battery Pack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
External Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Powering ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Previewing Saved A-Scan Image . . . . . . . . . . . . . . . . . . . . . . . . 136
PRF
Modes Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Selecting the Pulser Energy (Spike Mode) . . . . . . . . . . . . . . 48
Selecting the Pulser Type . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Selecting the Pulser Width (Square Wave Mode) . . . . . . . . . 46
Setting the Pulser Frequency . . . . . . . . . . . . . . . . . . . . . . . . . 43
Setting the Pulser Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Using the Phantom Feature . . . . . . . . . . . . . . . . . . . . . . . . . . 49

171

Index

Probe
Angle Beam Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Changing the Damping Level . . . . . . . . . . . . . . . . . . . . . . . . 42
Configuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Selecting the Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Specifying for DGS Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Specifying the Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Protection Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Pulser
Frequency Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Selecting the Energy (Spike Mode) . . . . . . . . . . . . . . . . . . . 48
Selecting the Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Selecting the Width (Square Wave Mode) . . . . . . . . . . . . . . 46
Setting the Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Pulser Repetition Frequency
see PRF

R
Range, Setting for A-Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

172

Recalling Data Set Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119


Receiver Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Rectification Mode, Selecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Reference Echo, Recording for DGS Curve. . . . . . . . . . . . . . . . 100
Reference Points, Adding/Deleting for DAC/TCG . . . . . . . . . . . 94
Reject Level, Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Report
Creating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Creating Header . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Fast Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Including Header . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Including Memo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Storing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127

S
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xi
Setup Mode
Accessing the Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Menu Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Setup, Initial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

USM Go Operators Manual

Index

Software Upgrades . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Specifications
Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
EN 12668 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Gates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Pulser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
USM Go . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
USM Go Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Spike Mode, Pulser Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Square Wave Mode, Pulser Type. . . . . . . . . . . . . . . . . . . . . . . . . 45
Starting Point, Setting for Gates . . . . . . . . . . . . . . . . . . . . . . . . . 68
Storing
Data Set Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127

USM Go Operators Manual

T
TCG Mode
Deleting Reference Points . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Generating the Reference Curve . . . . . . . . . . . . . . . . . . . . . . 90
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Working With . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Test Object
Material Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Temperature Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xiv
Threshold, Setting for Gates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Time, Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
TOF-Detection Method, Selecting . . . . . . . . . . . . . . . . . . . . . . . . 71
Transfer Correction
Setting for DAC/TCG Mode . . . . . . . . . . . . . . . . . . . . . . . . . 93
Setting for JISDAC Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 113

U
Ultrasonic Testing Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . xii
Units of Measure, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Upgrades, Instrument Software . . . . . . . . . . . . . . . . . . . . . . . . . . 63
USB Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

173

Index

Using Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135

V
Validity of DGS Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Viewing
Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

W
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Waste Disposal
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Electronic Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
WEEE Directive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Width, Adjusting for Gates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

174

USM Go Operators Manual

Customer Support Centers


North & South America
50 Industrial Park Road
Lewistown, PA 17044
U.S.A.
Tel: 866 243 2638 (toll-free)
717 242 0327
United Kingdom/Ireland
892 Charter Avenue Canley
Coventry CV4 8AF
England
Tel: +44 845 130 3925

France
68, Chemin des Ormeaux
Limonest 69760
France
Tel: +33 47 217 9216
Germany
Robert Bosch Str.
50354 Hrth
Germany
Tel: +49 2233 601 0
Spain
San Maximo,31, Planta 4A, Nave 6
Madrid 28041
Spain
Tel: +34 195 005 990

E-mail: [email protected]
www.geinspectiontechnologies.com

P/N 1259801 Rev 2

2009 General Electric Company. All rights reserved.


Technical content subject to change without notice.

China
5F, Building 1, No. 1 Huatuo Road,
Zhangjiang High-Tech Park,
Shanghai 201203
China
Tel: +86 800 915 9966 (toll-free)
+86 (0)21 3877 7888
Japan
7F Medie Corp Bldg. 8
2-4-14-Kichijoji Honcho,
Musashino-shi
Tokyo 180-0004
Japan
Tel: +81 442 67 7067

You might also like