Viscosimetro Ofite

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

Model 800 Viscometer

#130-10 - Model 800 Viscometer


#130-10-C - With Carrying Case and Power Supply
#130-10-L - With Retractable Legs (For Kits)

Instruction Manual
Updated 3/15/2011
Ver. 1.7
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A.
Tele: 832.320.7300 Fax: 713.880.9886 www.ofite.com
Copyright OFITE 2011

Table of
Contents

Introduction ......................................................................................2
Description .......................................................................................2
Components......................................................................................3
Specifications ...................................................................................5
Setup..................................................................................................8
Calibration.........................................................................................9
Operation ........................................................................................12
Calculations ....................................................................................14
Disassembly ...................................................................................14
Maintenance....................................................................................15

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Introduction

The Model 800 8-Speed Electronic Viscometer is designed and exclusively


manufactured by OFI Testing Equipment, Inc. It is extensively used worldwide in both the field and laboratory for the precise measurement of rheological properties of fluids.

Description

The OFITE Model 800 Viscometer determines the flow characteristics of oils
and drilling fluids in terms of shear rate and shear stress over various time
and temperature ranges at atmospheric pressure. Speeds are easily
changed with a control knob, and shear stress values are displayed on a
lighted magnified dial for ease of reading. An optional cup heater is used to
raise the temperature of the test sample.
The viscometers motor RPM is continuously monitored and automatically
adjusted by the OFITE Pulse-Power electronic speed regulator to maintain a
constant shear rate under varying input power and drilling fluid shear conditions. The eight precisely regulated test speeds (shear rates in RPM) are as
follows: 3 (Gel), 6, 30, 60, 100, 200, 300, and 600. A higher stirring speed is
also provided. Speeds may be changed with a control knob selection, without stopping the motor.
The Model 800 is suitable for both field and laboratory use and uses a motordriven electronic package to provide drilling fluid engineers with an extremely accurate and versatile tool. The Model 800 operates from a 12 Volt AC
power source, or standard rig power, either 115 or 230 Volt DC power, 50-60
Hz. The electronic regulator continuously monitors and automatically adjusts
the rotational speed to maintain a constant shear rate under varying fluid
shear conditions and input power variations that are commonly found on-site.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Components

#130-10 Model 800 8-Speed Viscometer


#130-10-2
Speed Reference PC Board
#130-10-3
Speed Control PC Board
#130-10-4
Encoder Pick-up PC Board
#130-10-5
Switch Plate
#130-10-6
Wiring Harness
#130-10-8
Encoder Disc
#130-10-14 Motor
#130-10-16 Belt
#130-10-17 Cover
#130-10-18 Lens for Cover
#130-10-19 Lamp
#130-10-22 Card Guide for PC Boards
#130-21
Cup, Stainless Steel
#132-50
Rotor, Drive
#132-56
Rotor, R1, 303 Stainless Steel
#132-57
Shield
#132-58
Bob
#132-59
Bearing Retainer
#132-71
Bearing
#132-74
Locking Ring, 1.653"
#132-75
Locking Ring, 12"
#132-76
Locking Ring, 2532"
#134-05
Shielded Bearing for Bob Shaft
#134-10
Torsion Spring Assy, F1
#134-39
Clamp Sleeve for Torsion Spring
#134-40
Shim, 58" 38"
#135-02
External Retainer Ring
#135-18
Socket Set Screw
#142-63
Locking Screw
#163-26
Small Clip
#170-21
Stand Support Rod
#170-44
Rubber Foot, 12"
#171-32
Midget Knob
#130-10-30 PC Power Supply
#152-37
AC Power Cord, 3-Conductor
Case:
#130-10-13 Carrying Case
#130-13-13-1 Custom Foam Insert

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Optional:
#130-10-33 Power Cord for 12-Volt Cigarette Lighter Adapter
#132-56-C Rotor, R1 Closed-Cup, 316 Stainless Steel (For completion
and fracturing fluid testing.)
#132-48
Calibration Instrument
#132-80
Calibration Fluid, Certified, 100 cP, 16 oz
#900-275 Battery Pack; 5 - 19 Volt Battery Output; Includes Case, Bag,
AC Charger, and Plug Adapters
Additional
#132-58-1
#132-58-2
#132-58-3
#132-58-4
#132-58H

Bobs, Rotors, and Springs:


B2 Bob, 303 Stainless Steel
B3 Bob, 303 Stainless Steel
B4 Bob, 303 Stainless Steel
B5 Bob, 303 Stainless Steel
B1 Bob, Hastelloy

#132-56H

Rotor, R1, Hastelloy

#134-10-2
#134-10-3
#134-10-4
#134-10-5
#134-10-6

F0.2 Torsion
F0.5 Torsion
F2.0 Torsion
F3.0 Torsion
F4.0 Torsion

Spring Assembly
Spring Assembly
Spring Assembly
Spring Assembly
Spring Assembly

#130-10-SP One Year Spare Parts


#130-10-18 Lens for Cover
#130-10-61 Light Assembly
#130-45-1
Calibrating Instrument
#132-49
Calibration Check Kit
#132-59
Bearing Retainer
#132-71
Bearing, Qty: 2
#132-75
Locking Ring, "
#132-80
Calibration Fluid, 100 cP, 16 oz
#132-81
Calibration Fluid, 50 cP, 16 oz
#134-05-2
Bob Shaft Bearing, Shielded, Qty: 4
#134-10
Torsion Spring Assembly, F1
#135-02
External Retainer Ring, Qty: 2
#165-14-11
Fuse Holder

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Specifications

Instrument Geometry
Speed Accuracy (RPM)
Motor Speeds (RPM)
Readout
Heat Cup
Power Requirements
Weight (kg)
Dimensions (cm)
Shipping Weight (kg)
Shipping Dimensions (cm)

True Couette Coaxial Cylinder


.001
8 Fixed Speeds (600, 300, 200, 100, 60,
30, 6, and 3)
Direct Dial
Stainless Steel, 150 Watt, Max
Recommended Temp 190F, (88C)
97 - 250 Volts AC, 50/60 Hz
6.0
15.2 17.8 40.6
20.4
55.9 25.4 40.6

Range of Measurement for Model 800


Rotor - Bob
Rotor Radius, RR, (cm)
Bob Radius, RB, (cm)
Bob Height, L, (cm)
Shear Gap, (cm)
R Ratio, RB/RR

R1B1
1.8415
1.7245
3.8
0.117
0.9365

Max. Shear Stress, SSMAX,


(Dyne / cm2)
F 0.2
F 0.5
F 1.0
F 2.0
F 3.0
F 4.0
F 5.0
F 10.0

R1B2
1.8415
1.2276
3.8
0.6139
0.666

R1B3
1.8415
0.8622
3.8
0.9793
0.468

R1B4
1.8415
0.8622
1.9
0.9793
0.468

R1B1

R1B2

R1B3

R1B4

330
840
1680
3360
5040
6720
8400
16800

651
1657
3314
6629
9943
13257
16571
33143

1320
3359
6717
13435
20152
26870
33587
67175

2644
6730
13460
26921
40381
53841
67302
134603

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Shear Rate Range


Shear Rate Constant, KR,
(sec-1 per RPM)
Shear Rate, (sec-1 or 1/s)
3 RPM
6 RPM
30 RPM
60 RPM
100 RPM
200 RPM
300 RPM
600 RPM

R1B1

R1B2

R1B3

R1B4

1.7023

0.3770

0.2682

0.2682

5.11

1.13

0.80

0.80

10.21
51.07
102.14
170.23
340.46
510.69
1021.38

2.26
11.31
22.62
37.70
75.40
113.10
226.20

1.61
8.05
16.09
26.82
53.64
80.46
160.92

1.61
8.05
16.09
26.82
53.64
80.46
160.92

Viscosity Ranges<a>
(cP)

R1B1

R1B2

R1B3

R1B4

Minimum Viscosity<b>
@600 RPM

0.5<c>

4.5

12.7

25

Maximum Viscosity<d>
10,000,000 89,000,000 255,000,000 500,000,000
@0.01 RPM
<a> Computed for standard Torsion Spring (F 1.0). For other torsion springs, multiply by F factor
<b> Lower viscosities can be measured by the Model 900, however one must take into account the
effect of bearing drag, Taylor vortices, zero offset, etc. when looking at the expected accuracy of
the reading
<c> For practical purposes the minimum viscosity is limited to 0.5 cP due to Taylor Vortices
<d> Maximum viscosity is based on Maximum Shear Stress and Minimum shear rate (RPM). However,
due to practical and physical limitations, it may be difficult to take these measurements.

Viscosity Conversions
To convert from units on left side to units on top, multiply by factor @ intercept.
Centipoise
Poise (P) g/(cm*s) (mN*s)m2 mPa*s
(cP)
Centipoise
(cP)
Poise (P)
g/(cm*s)
(mN*s)m2
mPa*s
(lb*s)
100 ft2

(lb*s)
100 ft2

0.01

0.01

0.002088

100
100
1
1

1
1
0.01
0.01

1
1
0.01
0.01

100
100
1
1

100
100
1
1

0.2088
0.2088
0.002088
0.002088

478.93

4.789

4.789

478.93

478.93

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Shear Stress Conversions


To convert from units on left side to units on top, multiply by factor @ intercept.
Dyne/cm2

Pa

lb/100ft2

lb/ft2

DR

Dyne/cm2

0.1

0.2084

0.002084

0.1957

Pa

10

2.084

0.02084

1.957

4.788

0.4788

0.01

0.939

lb/ft

478.8

47.88

100

93.9

DR

5.107

0.5107

1.065

0.01065

lb/100ft
2

What Bob & Spring Should I Use?


There is often confusion or misunderstanding about what a viscometer can
actually measure. For example, a viscometer with an R1B1 F1 combination can measure water fairly well at 100 RPM and higher, but at 3 RPM,
the readings would be shaky at best. While on the other hand, a linear fluid
with a viscosity of 15000, could not get past 6 RPM with the same combination.
To estimate which spring might be best, use the formula below to calculate
a Minimum Spring factor, where one establishes the maximum RPM the
fluid is going to be tested at, as well as what the expected Apparent
Viscosity of the fluid at that RPM. If the Factor comes out as .87, then an
F 1.0 spring should be used. If it comes out as .16, then an F 0.2 spring
would be best. To cover all ranges, it may be necessary to use more than
one spring.
Minimum Spring Factor (F) =

RPM(max) * AV(max)
BOB(F)*90000
Bob (F)
R1B1
1
R1B2
8.9
R1B3
25.4
R1B4
50.7

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Setup

Note

1. Connect the instrument to a power source. This viscometer will operate equally well at 12 Volts, 115 Volts, or 230 Volts, 50 or 60 Hz.
For longer-lasting, continuous operation, it is strongly recommended that the Model 800 Viscometer only be used with the
OFITE Power Supply (#130-10-30). This particular power supply
has the filtering range to assure true frequency under varying
input power conditions.
2. With the instrument turned off, place the splash guard onto the bob
shaft with the short tube end pointed up towards the bearings. Push up.
3. Screw on the appropriate bob with the tapered end up towards the
splash guard.
4. Place the R1 rotor onto the drive rotor over the bob. The threads
assure the rotor will attach evenly and uniformly each and every time.
5. The power switch is located on the back panel. Turn the unit on.
6. Place a properly prepared sample of test fluid in a sample cup and
immerse the rotor exactly to the fill line on the rotor by raising the platform. Tighten the lock nut on the platform.

!


Important

If using a thermostatically temperature-controlled viscosity cup,


the maximum operating temperature is 190F (93C). Liquid
trapped inside a hollow bob may vaporize when immersed in very
high temperature fluids and cause the bob to explode.

Drive Rotor
Splash Guard
Bob

R1 Rotor

Platform

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Calibration

The Model 800 viscometer can lose calibration while in service if the bob
shaft bearings become contaminated or if the bob shaft itself is bent. If the
dial does not read zero when it should or if there is excessive dial deflection when the main shaft is turning, this may indicate that the bob shaft
bearings are sticking. If the spring appears to be non-linear, the bob shaft
may be bent. Your viscometer will require servicing if it exhibits any of
these symptoms. Periodic calibration checks are recommended to ensure
that your instrument is in proper working condition at all times.
The OFITE Calibrating Instrument (Part No. 130-45-1) and the Calibration
Check Kit (Part No. 132-49) are available to facilitate calibration. The
Calibrating Instrument uses a dead-weight method of calibrating, while the
Calibration Check Kit is fluid based. Both are provided in the Model 800
spare parts kit (Part No. 130-10-SP).

132-49 Calibration Check Kit


130-45-1 Calibrating Instrument

Equipment:
#130-45-1 Calibrating Instrument for 8-Speed Viscometer (Requires a
trained technician)
#132-49

Calibration Check Kit

Calibration Fluids, Certified:


#132-84 Calibration Fluid, 20 CP, 16 oz (500 mL)
#132-84-5 Calibration Fluid, 20 CP, 5 gal (18.927 L)
#132-81 Calibration Fluid, 50 CP, 16 oz (500 mL)
#132-81-5 Calibration Fluid, 50 CP, 5 gal (18.927 L)
#132-80 Calibration Fluid, 100 CP, 16 oz (500 mL)
#132-80-5 Calibration Fluid, 100 CP, 5 gal (18.927 L)
#132-83 Calibration Fluid, 200 CP, 16 oz (500 mL)
#132-83-1 Calibration Fluid, 200 CP, 1 gal (3.785 L)
#132-83-5 Calibration Fluid, 200 CP, 5 gal (18.927 L)
#132-85 Calibration Fluid, 350 CP, 16 oz (500 mL)
#132-82 Calibration Fluid, 500 CP, 16 oz (500 mL)
#132-82-5 Calibration Fluid, 500 CP, 5 gal (18.927 L)

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

Procedure:
1. Choose the certified calibration fluid using the temperature-viscosity
chart supplied with the calibration fluid to cover the viscosity range of
interest. Make sure the lot number on the chart matches the lot number on the fluid container. Each lot of standard fluid is individually certified. The viscosity will normally vary slightly from lot to lot.
2. Clean and dry the viscometer bob, rotor, and cup. Place the viscometer and the calibration fluid side-by-side on the counter top in a room
with a reasonably constant temperature [< 5F (< 2.7C)]. Allow the
viscometer and the fluid to stand at least two hours to equilibrate.
3. Operate the viscometer in air for two to four minutes to loosen up the
bearings and gears. Observe the rotor for excessive wobbling and
replace if necessary.

!


4. Fill the cup to the scribed line with calibration fluid and place it on the
viscometer stage. Move the stage upward until the fluid level is to the
fill line on the rotor.
Never immerse the rotor in fluids above the fill line.

Important

Fill Line

R1 Rotor
5. Place a thermometer capable of 0.2F (0.1C) into the fluid and hold
or tape it in place to prevent breakage. Operate the viscometer at a
low speed setting until the thermometer reading becomes stable to
within 0.2F (0.1C) per 30 seconds. Note and record the temperature reading.
6. Once the temperature has stabilized, operate the viscometer at 600
RPM and then at 300 RPM. Note and record the dial readings to the
nearest 0.5 dial unit.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

10

7. Using the temperature-viscosity chart supplied with the calibration fluid,


determine the certified viscosity to the nearest 0.5 centipoise.
a. Compare the 300 RPM reading to standard viscosity and record the
deviation plus or minus.
b. Divide the 600 RPM dial reading by 1.98, compare that to standard
viscosity, and record the deviation plus or minus.
8. Deviations exceeding 1.5 dial units are not acceptable. If the deviation
exceeds this tolerance, the viscometer will require adjustments or calibration by a qualified technician.
9. Record the viscometer serial number, date, and deviation. Mark on the
viscometer the date of calibration and a general indication of the calibration check status.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

11

Operation

Measuring Viscosity:
1. Mix the sample on the STIR setting for 10 seconds while heating or
cooling the fluid. Monitor the temperature with a thermometer.
Continue to mix until the sample reaches the target temperature.
2. Rotate the knob to one of the speed settings. When the dial reading
stabilizes, record the reading and the temperature. Repeat this step for
any other speeds that your test requires.


Tip

Always start with the higher RPM and work your way down to the lowest RPM. For example, if you need readings at 100 RPM, 200 RPM,
and 300 RPM, record the measurement at 300 RPM first, then 200
RPM, then 100 RPM, then gel strengths (if necessary).
Measuring Gel Strength:
1. Mix the sample on the STIR setting for 10 seconds.
2. Rotate the knob to GEL and immediately shut off the power.
3. As soon as the rotor stops rotating, wait 10 seconds and turn the power
back on while looking at the dial. Record the maximum dial deflection
before the gel breaks. This is the 10 second gel strength.

Note

When measuring gel properties, the dial does not have to return to zero
during the quiescent period. Therefore, it should not be forced back to the
zero setting if it does not freely do so. When determining the maximum
dial deflection, no allowance needs to be made if the dial did not start at
zero. Only the maximum dial deflection is of interest. Rheological properties and characteristics of the sample will determine if the dial returns to
zero during the quiescent periods of the gel measurements.
4. For the 10 minute gel strength, re-stir the fluid and wait 10 minutes
before recording the maximum dial deflection.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

12

API Testing
1. Mix the sample on the STIR setting for 10 seconds.
2. Set the speed to 600 RPM. Wait for the reading to stabilize and then
record the dial reading and temperature.
3. Set the speed to 300 RPM. Wait for the reading to stabilize and then
record the dial reading and temperature.
4. Stir the sample again for 10 seconds.
5. Set the speed to GEL and immediately shut off the power.
6. As soon as the rotor stops rotating, wait 10 seconds and turn the power
on while looking at the dial. Record the maximum dial deflection before
the gel breaks. This is the 10 second gel strength.
7. For the 10 minute gel strength, re-stir the fluid and wait 10 minutes before
recording the maximum dial deflection.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

13

Calculations

Disassembly

Plastic Viscosity (PV), cP

= 600 RPM reading - 300 RPM reading

Yield Point (YP), lb./100 ft2

= 300 RPM reading - Plastic Viscosity (PV)

Apparent Viscosity (AV), cP

= 600 RPM reading / 2

Gel Strength, 10 second, lb/100 ft2

= the maximum dial deflection after 10 sec.

Gel Strength, 10 minute, lb/100 ft2

= the maximum dial deflection after 10 min.

Clean the viscometer after every test.


1. Remove the R1 rotor from the drive rotor.
2. Remove the bob.
3. Once the bob is removed, remove the splash guard and wipe down the
bob shaft. Clean all removed parts with soap and water and dry them
thoroughly.

!


Important

Keep the instrument upright at all times, especially when cleaning


so that water does not get into the bearings.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

14

Maintenance

From time to time, the bearings will need to be changed. Complete the following to determine if it is time to perform this maintenance procedure:
1. The viscometer should have a zero dial reading when placed in an
upright position with the rotor not immersed in fluid prior to running tests.
2. With the instrument in this position, rotate the rotor at 600 RPM. The dial
reading in air should not exceed one.
3. Place water in a suitable container and immerse the rotor to the fill line.
4. Rotate the rotor at 600 RPM. The dial reading in water should be
between 1.5 and 3.0.
5. At 300 RPM, the dial reading in water should be between 0.5 and 2.0.
If the viscometer fails to pass any one of the above tests, the bearings are
bad and should be replaced by a qualified instrument technician.

Bob Shaft Bearing and Torsion Spring Replacement:


Tools Required:
Allen Wrench, 116"
Torsion Spring Removal Tool (#134-10-1)
Retainer Ring Pliers (small)
External Retainer Ring Pliers
1. Unplug the power supply and remove the rotor and bob.
2. With the appropriate allen wrench, loosen the set screw in the speed control knob. Remove the speed control knob from the unit.
3. Remove all four flat head screws on the outer housing cover and gently
lift the cover upward.
4. Use an allen wrench to loosen the set screw (one turn counter-clockwise)
from the aluminum bushing located inside the torsion spring. The set
screw is located at the top of the clamp sleeve used for the torsion
spring.
5. Loosen the two screws and remove the stop block.
6. Loosen the top set screw (one turn counter-clockwise) on the dial.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

15

!


7. Gently lift up on the torsion spring to remove it from the clamp sleeve
and dial.
Do not stretch the torsion spring.

Important

If you are replacing only the torsion spring, discard the old spring, and
skip to step 20 to complete the process.

Set Screws (Step 5)


Clamp Sleeve
Set Screw (Step 4)
Stop Block
Bridge
Torsion Spring
Set Screw (Step 6)
Dial
Brass Bearing Mount

!


8. Remove the four screws from the bridge and lay it gently to the right
side.
Do not touch the encoder that encircles the flywheel.

Important

9. Loosen the set screw (one turn counter-clockwise) on the bottom of the
dial and pull down on the bob shaft to remove them both.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

16

!


10. Remove the two screws on the brass bearing mount (upper bearing).
Hold the upper brass bearing mount by hand for easier access to the
retainer ring and bearing.
Be careful not to damage the pins.

Important
Set Screw

Retainer Clip

Upper Bearing
Brass Bearing Mount

Set Screw

11. The upper bearing is located inside the brass bearing mount. Use retaining ring pliers to remove the retainer ring. Push the bearing upwards to
remove the upper bearing from the brass bearing mount.
12. Use surgical gloves to insert a new upper bearing into the brass bearing
mount and replace the retainer ring with the retaining ring pliers.

!


Important

Do not touch the bearing with bare hands. Dirt, oils, and other contaminants can damage the bearing.
13. Replace the brass bearing mount back onto the two pins and tighten the
two screws.
14. Remove the Splash Guard, loosen the set screw on the lower brass bearing shield (one turn counter-clockwise), and slide the lower bearing down
the bob shaft to remove all three components.
15. Use surgical gloves to apply a new lower bearing onto the bob shaft.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

17

16. Install the lower brass bearing shield and tighten the set screw. Slide the
splash guard back up the bob shaft until in touches the lower brass bearing shield.

The bearing should be able to free spin for up to two minutes. If it


does not, replace it with a new bearing.

Note

Torsion Spring

Brass Bearing Mount

Bob Shaft
Bearing Shield
Lower Bearing
Splash Guard

17. Slide the bob shaft up through the upper brass bearing mount. Turn the
bob shaft in the direction that allows the Flat located at the top of the
bob shaft to be positioned towards you.

Note

18. Align the bottom set screw on the dial with the Flat on the bob shaft and
tighten firmly. Make sure the set screws are perpendicular to the tip of
the bob shaft.
The bob shaft should rotate smoothly without any noise.
19. Replace the bridge being careful not to touch, or bend the encoder.
Tighten the four screws.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

18

20. Gently slide the torsion spring through the clamp sleeve and into the dial.
Make sure the bottom of the torsion spring is seated properly inside the
dial. Tighten the set screw located at the top of the dial to secure the torsion spring.
21. Replace the stop block and the two screws. Make sure that the zero dial
reading is aligned with the fill on the stop block.
22. Press down on the torsion spring until it is flush with the clamp sleeve
and tighten the set screw.
23. Replace the bob and rotor.
24. Re-attach the main housing cover and re-calibrate.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

19

You might also like