Which Buffer Should I Select
Which Buffer Should I Select
Which Buffer Should I Select
specify when to measure pH a good recent example specifies that the pH should be measured before
adding cyclodextrin to the buffer5 sulphated cyclodextrins are acidic and will lower the pH when
added. Heat is usually generated when acids and bases are mixed. Buffer solutions should be allowed to
reach room temperature (after mixing or removing from a fridge) before measuring pH as pH is
temperature dependent.
Incorrect use/care of pH meter: Manufacturers of modern pH meters have made operation so simple
that successful operation is often taken for granted, but it is not quite as easy as it looks. Electrodes
must be clean and properly filled. Calibration buffers must be fresh and span the pH range of interest.
Temperature is an important factor; not only does the pH of most buffered systems change with
temperature, the response of a pH electrode has a temperature dependence of its own. The
temperature setting on a pH meter is used to adjust the temperature of the electrode during the
measurement, not to correct the temperature where the buffer solution will be used. Specific volumes
of specific concentrations of acids and bases can be mixed to make pH values, for example, phosphoric
acid and triethanolamine, which avoids use of a pH meter.3
"The best way to use a pH meter is to avoid its use it is more reliable to mix measured volumes of
acids and bases together to generate a specific pH rather than rely on a pH meter."
Buffer not used within its buffering range: A buffer is only a buffer within 1 pH unit of its pKavalue;
working within 0.5 pH units of pKa is better for critical work. A buffer is needed not only to provide an
appropriate pH for the desired process but also to maintain that pH in the face of outside influences
such as temperature changes, reaction product formation or factors entering from the environment
(i.e., gases such as CO2).
However, it has been found during literature investigations that many methods cited6,7 do not use the
buffer within their adequate range, for example, phosphate pH5 is an electrolyte solution not a
buffer.6 As a result, in some instances, low capacity buffers used require frequent buffer replenishment
because they are quickly affected by "buffer depletion".
Inconsistent reagent grades or descriptions: The grade and source of reagents should be specified
because this affects the purity (and cost) of the reagents such as cyclodextrins and sodium dodecyl
sulphate. Variations in purity will lead to alterations in ionic strength, current, separation selectivity and,
hence, lack of method robustness.
Not specifying shelf-life and storage conditions of buffers: Shelf-life for buffers cannot be assumed.
They must be experimentally assessed and proven before writing into methods. For example, materials
such as pthalates and metal ions can leach from containers, microbial growth can occur in buffers
containing carbohydrate sources such as cyclodextrins or carbon dioxide can be absorbed from the
atmosphere.
Insufficiently described buffer filtration procedure: Generally all electrolytes (and sample) should be
filtered through at least a 0.45 m filter to remove particulates, which would appear as apparent noise
on the detector baseline. The buffer filtration procedures can result in removal, or addition, of
components to the buffer. This may be a result of adsorption onto the filter or extraction from the filter.
The filtration process should be assessed, validated and stated in the method. Failure to do this can
result in non-robustness. For example, it was observed that up to 15% of the sodium dodecyl sulphate
(SDS) was being retained using certain filter types when filtering small volumes.8
"The exact filtration process should be described in the method including the volumes filtered and the
filter type used."