Introduction To The Mechatronic Engineering Laboratory Equipment
Introduction To The Mechatronic Engineering Laboratory Equipment
Introduction To The Mechatronic Engineering Laboratory Equipment
Intro - 1
Digital Multimeter
Function Generator
Oscilloscope
Components:
Resistors: one each of: 100 , 2 k, 5.1 k, 10 k, 100 k, 1 M and two 1 k resistors.
Introduction: Each workstation in the lab consists of the equipment shown in Figure 1.
Figure 1. Laboratory equipment in the Mechatronic Engineering Laboratory. The multimeter, arbitrary waveform
generator and oscilloscope are connected to the computer via GP-IB.
rev. 1.9.1
27JAN2014
Intro - 2
Study the operating instructions, and perform each exercise when it is presented. Make any
calculations at the time requested.
CHANNEL
red line
blue line
CHANNEL
red line
blue line
Figure 2. Solderless breadboard -- Top View. Wires and components can be simply inserted into the holes. DO
NOT insert anything larger than 22 AWG!
red line
blue line
CHANNEL
red line
blue line
CHANNEL
red line
blue line
Figure 3. Internal connections of the solderless breadboard -- Top View. The fifty holes in each of the two rows
between the red and blue lines are connected together internally. The remaining holes are connected together in
groups of five vertically.
Notice that the holes along the horizontal sets of red and blue lines are all connected along the
length of the lines. These horizontal strips are often called rails or busses, because they carry a
signal down the strip like a bus carries people down a road. It is conventional to route power to
the strips near the red lines and ground to the strips near the black lines. Follow this convention
when you build your circuits! Notice also how each set of five holes above and below the
horizontal channels are connected together. Thus anything plugged into one of the set of five
holes will be connected to anything else plugged into one of the remaining four holes of the set.
Figure 4 shows a circuit (actually that for Exercise 3) built on a breadboard. Power and common
ground are routed from the power supply to the solderless breadboard using red and black test
San Jos State University Dept. of Mechanical and Aerospace Engineering
rev. 1.9.1
27JAN2014
Intro - 3
leads respectively. These particular leads are called stacking banana-to-banana test leads. The
sockets they plug into at the power supply and on the breadboard are called binding posts.
Binding posts usually have a captive screwable barrel that can clamp on a wire, which is either
wrapped around the internal metal post or poked into a transverse hole drilled through the post.
Figure 4 shows two short lengths of red and black hookup wire that take power and ground
respectively from the binding posts on the breadboard to the power and ground busses on the
breadboard. Two other short lengths of red and black hookup wire (22-24 AWG and stripped of
insulation inch from each end) take power and ground respectively from the busses at the
center of the breadboard and route them down to the busses at the lower edge of the breadboard.
Look at the circuit schematic in Exercise 3, and see how the connections are made in Figure 4.
Figure 4. Circuit from Exercise 3 built on a solderless breadboard. Power and ground are routed from the red and
black binding posts to the two busses at the center of the breadboard by two short lengths of red and black hookup
wire (22-24 AWG and stripped of insulation inch from the end). Two other short lengths of red and black
hookup wire take power and ground respectively from the busses at the center of the breadboard and route them
down to the busses at the lower edge of the breadboard.
Pay attention to the feel of the connection when you plug a lead into one of the holes of the
breadboard. The sockets for each hole inside the breadboard are made from a U-shaped metal
conductor. Through wear (or by someone jamming too large a lead into one of the holes), the Uchannel can be bent open, which will result in a poor or intermittent electrical connection. (This is
one of the major disadvantages of using a solderless breadboard and something to keep in mind
when you are troubleshooting a circuit that is behaving strangely. You should feel some friction
when you slide a lead into one of the holes, and voltage readings should be stable even if you
wiggle the leads slightly.
Follow the convention of using red for power and black (or green) for grounds when you wire
jumpers on your breadboard. Try to also group and color-code other connections of the same type
to make it easy to follow them around the breadboard. It is best to keep jumpers and leads only as
long as needed (shorter is usually better) and neatly routed close to the breadboard. Doing so will
GREATLY help when you have to troubleshoot your circuit.
rev. 1.9.1
27JAN2014
Intro - 4
Figure 5. DC power supply. This is a triple-output power supply, which means it has three, independent outputs: 0
to 6 V, 0 to +20 V, and 0 to -20 V. The voltages for these outputs are referenced to COM (also called common)
ground. The terminal with the upside-down T symbol is earth ground, which we will not use.
This common connection is referred to as common ground. It is the point from which the output
voltages supplied by the +6, +20, and -20 terminals, are referenced. The COM output is to be
connected to circuit ground, i.e., the ground symbols in the schematics for the experiments.
There is also an equipment ground provided. This is the terminal marked with the upside-down Tshaped symbol. As a safety feature, each piece of equipment in the lab is connected to earth through
its 3-wire power cord. If for some reason a short circuit occurred to the case of the instrument, the
ground wire would carry the short-circuit current to the earth, instead of you! For this reason, the
equipment ground is more commonly referred to as earth ground. We will rarely use the DC supply
equipment ground.
Voltage Adjust
The +6V control knob sets the 0 to +6V output voltage.
The 20V control sets the 0 to +20V and the 0 to -20V outputs simultaneously. With the Tracking
Ratio control knob turned fully clockwise to its fixed position, the voltage of the negative supply
tracks the positive supply within 1%, giving balanced positive and negative supplies.
Turning the Tracking Ratio control knob counter-clockwise out of its fixed position allows you to set
the voltage of the -20V supply to a fixed fraction (less than unity) of the +20V supply. For example,
you can set the -20V supply to 1/3 of the +20V level, but not to 3 times the +20V level. Once this
ratio is set, the 20V control still controls both outputs and maintains a constant ratio between their
voltages.
rev. 1.9.1
27JAN2014
Intro - 5
Note: Before hooking up any circuit to the Power Supply, you should make sure that the
voltage is set to what you want to avoid any overpowering of your circuit.
Exercise 1. -- DC Supply
Turn on the power supply by pressing the white button labeled LINE. The red LED display will
show the voltage output and current drawn for the output selected by one of the three meter
pushbuttons.
Turn the Tracking Ratio control to its fixed position. Press the +20V meter button (to display
the +20V output) and adjust the 20V control knob to set the positive supply to +15V. Press the
-20V meter button. The voltmeter should read -15V. The positive and negative supplies are
balanced (15V).
Adjust the Tracking Ratio control until the negative supply reads -5V. The positive supply should
read +15 V. Check that it does. You now have a tracking ratio of 3:1 (+15V and -5V).
To see how tracking works, press the +20V meter button and readjust the 20 V control knob to
set the positive supply to +18V.
Finally, press the -20V meter button. The negative supply should read -6V, since the voltage ratio
of 3:1 was not changed.
(Note: In addition to describing in your lab report what you did in the Exercises and what
resulted during the lab session, also weave into your narrative the answers to the questions in
boldface type.)
Overload Indicators
An overload LED located next to the ammeter will light when the DC supply output is connected to a
load (circuit) that causes its current limit to be exceeded (0.5A for the 20V supplies, 2.5A for the
+6V supply). The cause of the overload is usually a short-circuit. If you see an overload light, turn
the power off immediately and double-check that your circuit is wired correctly.
rev. 1.9.1
27JAN2014
Intro - 6
-10.000,00 V DC
HI
HI
200V
Max
Trig
FUNCTION
Math
DC I
AC I
4W
Period
DC V
AC V
2W
Freq
Power
On/Off
MENU
Recall
Cont
RANGE / DIGITS
5
6
Off
On
Auto/
Man
CHOICES
LEVEL
ENTER
Input
V
1000V
Max
LO
dB
dBm
Null
Min
Max
LO
500Vpk
Max
Terminals
3A
RMS
Auto/Hold
Single
TRIG
Shift
LOCAL
Front
Rear
Fused on
Rear Panel
Figure 6. Digital Multimeter (DMM). The DMM is used to measure resistance, voltage, and current.
c.
rev. 1.9.1
27JAN2014
d.
Intro - 7
Note: Resistor values are not exact. They have a tolerance on their nominal value, such as
1%, 5%, etc. When you measure their value, double-check that it is within the
specified tolerance, which can be deciphered by reading the associated color band
on the resistor.
4W Sense/
Ratio Ref
Input
V
HI
HI
200V
Max
1000V
Max
LO
LO
500Vpk
Max
Terminals
unknown
resistor
3A
RMS
I
Front
Rear
Fused on
Rear Panel
Figure 7. Resistance measurement (two-wire). The probe leads are connected to the rightmost HI and LO terminals
on the DMM.
Connect leads to the HI and LO terminals marked 1000 V Max (see Figure 8).
Voltmeters are connected in parallel with the element. We do not need to modify the
circuit when making voltage measurements. Since voltages are measured across an
element, we simply place the leads on two points across the circuit element to be tested.
Select DC V or AC V. In the AC volts function, the meter removes the DC component
and measures the RMS value of the AC component, only.
Leave the meter on autoranging or select manual ranges of 100 mV, 1 V, 100 V, or 1000
V (750 V AC).
b.
c.
4W Sense/
Ratio Ref
Input
V
HI
HI
200V
Max
1000V
Max
LO
LO
500Vpk
Max
Terminals
3A
RMS
I
Front
Rear
Fused on
Rear Panel
Figure 8. Voltage measurement with the DMM. Note the location of where the leads attach to the DMM.
rev. 1.9.1
27JAN2014
Intro - 8
connect the ammeter in series at the break. Note the lead polarity with respect to the
current direction.
b. Select DC I or AC I (using the shift key). In the AC current function, the meter removes
the DC component and measures the RMS value of the AC component, only.
c. Leave the meter on autoranging or select manual ranges of 10 mA (DC only), 100 mA
(DC only), 1 A, or 3 A.
d. Note that current must go through the meter in order to complete the electrical circuit.
4W Sense/
Ratio Ref
Input
V
HI
HI
200V
Max
1000V
Max
LO
LO
500Vpk
Max
Terminals
3A
RMS
I
Front
Rear
Fused on
Rear Panel
Figure 9. Current measurement using the DMM. Note where the probes connect to the DMM (it is different than for
voltage measurement). Note that to make a current measurement, you must break the circuit in the branch where the
current flows and route it through the DMM. Through which terminal does current flow into the DMM?
rev. 1.9.1
27JAN2014
Intro - 9
measur ed - nominal
x 100%
nominal
Vs
+
V
R 1 = 10 k
R2 =2 k
+
V2
Ib
R3 =1 k
R 4 = 5.1 k
VS = 10 VDC
+
R2
V4
R1 = 10 k
R2 = 2 k
R3 = 1 k
R4 = 5.1 k
R4
V s = 10 Vdc
Binding
posts
rev. 1.9.1
27JAN2014
Intro - 10
THE HP 33120A
FUNCTION GENERATOR / ARBITRARY WAVEFORM GENERATOR
The function generator can output five standard waveforms including sine, square, triangle,
ramp, and noise. Each of these waveforms has zero average value, which means they are
symmetrical above and below the 0 V reference line. You can also select one of five predefined
arbitrary waveforms or download your own custom waveforms. You can internally modulate any
of the standard waveforms (including arbitrary) using AM, FM, FSK, or burst modulation.
Linear or logarithmic frequency sweeping is available for any of the standard waveforms (except
noise) and arbitrary waveforms. Figure 12 shows the front panel of the function generator.
In the ME 106 lab, we will only be interested in the standard and arbitrary waveforms. We will
not use the modulation capability of the generator.
At power-on, the function generator outputs a sine wave at 1 kHz with an amplitude of 100
mV peak-to-peak and a DC offset of 0 V (into a 50 termination).
15.000,000 MHz
FUNCTION / MODULATION
AM
Power
2
Freq
Off
On
FM
Freq
FSK
3
AM/FM
Burst
4
Sweep
Arb List
Noise
Arb
+
-
Level
% Duty
Internal
Store
Cancel
Ampl
Offset
Single
Recall
Enter
Number
MODIFY
TRIG
MENU
On/Off
Enter
Shift
STATE
LOCAL
MHz
m Vpp
SYNC
kHz
m Vrms
Hz
dBm
Back Space
Recall Menu
OUTPUT
42V
Max
Shift
rev. 1.9.1
27JAN2014
Intro - 11
the keys that have the desired numbers printed next to them. For example, to select the
number 10, press Enter Number, then press the key with the 1 next to it, then press the
key with the 0 next to it, then press the Enter button (above the Shift key) to commit the
value you entered.
If you accidentally press Enter Number, just press Shift, then press Cancel to turn off the
Num annunciator.
If the number requires units, press the appropriate arrow key. If the number does not require
units, press the Enter key.
2. If one of the displayed numbers is flashing you can use the arrow keys to edit the flashing
digit.
In this case, you are merely changing the value of the displayed number, and you do not need
to enter the units.
3. You can also change the value of the flashing digit by turning the knob.
Turning the knob clockwise will increment the flashing digit. (It will also cause digits to the
left of the flashing digit to change if the knob is turned far enough.) Turning the knob counter
clockwise will decrement the flashing digit. (It will also cause digits to the right of the
flashing digit to change if the knob is turned far enough.) You can also use the knob along
with the left/right arrow keys to modify the displayed number.
The Front-Panel Menu
The front-panel menu offers access to many more features and capabilities of this generator. Most
of these features will not concern us in the ME 106 Lab. Only one of these features is of interest
at this time. That feature is output termination.
Output Termination
The function generator has a 50 output impedance on the OUTPUT terminal. You can specify
whether the load you are connecting is a 50 load or an open circuit (HIGH Z). (Those are the
only two choices.)
To select the output termination
1. Press the Shift key and then the Menu On/Off key to turn on the menu.
2. Press the > key three times to move across to the D: SYS MENU choice.
3. Press the key to get to the OUT TERM command.
4. Press the key a second time to get the place where you can change the expected output
impedance value (either 50 or HIGH Z). Note that the default value is 50 , which is
also the value selected when the function generator is first powered on. For ALL the
measurements you will do in ME 106, you should select HIGH Z.
5. Use the < and > keys to select between the 50 and HIGH Z choices.
6. Press Enter. This saves the change and turns off the menu.
The section on mismatched loads below discusses what happens when you do not terminate
into a 50 load or an open circuit.
San Jos State University Dept. of Mechanical and Aerospace Engineering
rev. 1.9.1
27JAN2014
Intro - 12
1 2 V
0.00s
Time
Cursors
Run
Stop
Autostore
1 RUN
500 s
HORIZONTAL
Save / Recall
Trace
Setup
Autoscale
Display
Erase
TRIGGER
Source
Delay
Print
Utility
Level
Main
Delayed
Mode
VERTICAL
Volts/Div
Volts/Div
Time/Div
TV
Holdoff
Slope
Coupling
5V
2 mV
Ext
2 mV
5s
5 ns
Position
5V
Position
Line
1X
2Y
External Trigger
Line
O
Figure 13. The oscilloscope. This device is used to measure dynamic signals, i.e., signals that change relatively
quickly.
Probes Attention!
Oscilloscope leads are called probes. The low side of the probe (the alligator clip) is usually tied
to the scope case, which is connected to earth ground through its 3-wire power cord.
If you have connected the equipment ground from any other piece of equipment (i.e. the power
supply) to your circuit, then the low side of the probe must be connected to the same point in the
circuit as the equipment ground.
Most of the circuits built in this lab, however, will have a common ground as discussed in the
Triple Output Power Supply section earlier. When using common ground, the alligator clip can be
connected anywhere in the circuit. The only restriction in this case is that if you use two probes,
both alligator clips MUST be connected to the SAME point in the circuit.
Probes are labeled with their attenuation factor either 100X, 10X or 1X. A 10X probe reduces
the amplitude of the input signal to 1/10th of its actual value. A 100X probe reduces the amplitude
of the input signal to 1/100th of its actual value. A 1X probe does not attenuate the signal.
To set the probe attenuation factor for the probe connected to channel 1:
1. Press the 1 key (it is a grey front-panel key).
San Jos State University Dept. of Mechanical and Aerospace Engineering
rev. 1.9.1
27JAN2014
Intro - 13
2. Next, toggle the Probe softkey to change the attenuation factor to 10 (since we have
10X probes in this lab).
To set the probe attenuation factor for the probe connected to channel 2:
1. Press the 2 key, then toggle the Probe softkey to set the correct attenuation factor.
Note: if you do not correctly set the attenuation factor in the oscilloscope, the oscilloscope
may only display 1/10th of the actual reading.
Exercises
For exercises 5 through 11, you will be connecting the function generator directly to the
oscilloscope. Since the oscilloscope has a very high input impedance (approximately 1 M by
itself and 10 M with a 10x probe), you will need to set the output termination of your
function generator to HIGH Z to match the impedance of the scope (see page 9).
After setting the HIGH Z termination, set the function generator to output an 8 V peak-to-peak
500 Hz triangular wave with an offset of 2 VDC. To do this:
1. Select the triangular waveshape.
2. To set the output frequency to 500 Hz:
a. Press the Freq key to enable the frequency modify mode.
b. Press the Enter Number key to enable the number mode. Notice that the Num
annunciator turns on and ENTER NUM flashes on the display.
c. Press 5, then press 0 twice.
d. Press > to set the units to Hz.
3. To set the output amplitude to 8 VP-P:
a. Press the Ampl key to enable the amplitude modify mode.
b. Press the Enter Number key to enable the number mode.
c. Press 8.
d. Press the to set the units to VP-P.
Note that if you had wanted an amplitude of 10 mVP-P you would
need to press Shift before pressing since m is in blue.
4. To set the offset to 2 VDC:
a. Press the Offset key to enable the offset modify mode.
b. Press the Enter Number key to enable the number mode.
c. Press 2.
d. Press Enter. Since it is understood that the offset voltage is a DC voltage, no units are
needed. The display will read VDC. Note also that once you have set an offset voltage,
the word offset is displayed (even if you go back and display the amplitude or
frequency again).
Exercise 5. -- Displaying a signal automatically.
a. Connect the output of the function generator to the input channel 1 of the oscilloscope as
shown in Figure 14.
rev. 1.9.1
27JAN2014
Intro - 14
FUNCTION
GENERATOR
OSCILLOSCOPE
SYNC
1X
2Y
OUTPUT
42V
Max
red clip
black clip
probe
alligator clip
rev. 1.9.1
27JAN2014
Intro - 15
keys cause the status line to change. Notice, also, which softkeys cause a change in the
display and/or the position of the ground reference. Notice that when you change the input
coupling from DC to AC, the signal moves vertically so that it is symmetrical above and
below the reference line. This means that the DC component has been removed.
d. Channels 1 and 2 have a vernier softkey that allows the Volt/Div knob to change the vertical
step size in smaller increments. These smaller increments are calibrated, which results in
accurate measurements even with the vernier turned on.
e. To turn the channel off, either press the 1 key a second time or press the left-most softkey.
Exercise 7. -- Some of the horizontal (time base) controls
a. Turn the Time/Div knob. Notice the change it makes to the status line and to the displayed
waveform. The Time/Div knob changes the sweep speed form 2 ns to 5 s in a 1-2-5 step
sequence, and the value is displayed in the status line.
b. Pressing the Main/Delayed key brings up a softkey menu with six softkey choices. We will
not discuss their functions at this time. Nor will we discuss the Delay knob.
Exercise 8. -- Some of the trigger controls.
a. Turn the trigger Level knob and notice the changes it makes to the display. In particular,
notice that the waveform moves horizontally as the level changes. Also, notice that a voltage
value is displayed (since, the trigger level is a voltage level). What happens if you move the
trigger level above or below the displayed waveform? Try this for channel 1 input coupling set
to AC and for channel 1 input coupling set to DC.
b. Pressing the Source key brings up a softkey menu. Toggle each of the softkeys and notice that
each key causes the status line to change. Notice, also, what happens to the displayed
waveform. Reset the source to 1 before going on to part c. We will not discuss these menus at
this time
Exercise 9. -- Making frequency measurements automatically.
a. Press the Time key.
b. Toggle the Source softkey to select a channel for the frequency measurement (channel 1 in
this case).
c. Press the Freq softkey. The oscilloscope automatically measures the frequency and displays
the result on the lower line of the display. The number in parentheses after the word Freq is
the number of the channel that the oscilloscope used for the measurement. The oscilloscope
retains in memory and displays the three most current measurement results. If you make a
fourth measurement, the left-most result is dropped. Try pressing other Time measurement
softkeys to observe their behavior.
Exercise 10. -- Making voltage measurements automatically.
You can measure the following voltage parameters automatically with the oscilloscope: peak-topeak, average, RMS, maximum, minimum, top, and base. Figure 15 and Figure 16, below, show
pulses with some of the voltage measurement points labeled.
rev. 1.9.1
27JAN2014
Intro - 16
Vmax
Vtop
Vbase
Vmin
Vmax = Vtop
Vmin = Vbase
Figure 16. Pulse where top and bottom are not well-defined.
(VACRMS )2 + (VDC )2
What values do you measure? Set the input coupling back to DC before you begin step d.
d. Press the Next Menu softkey.
The Show Meas softkey displays the horizontal and vertical cursors that show where the
measurement was taken on the signal.
The Voltage Measurement softkeys give you the choice of VMAX, VMIN, VTOP, and VBASE.
The Previous Menu softkey return to the previous softkey menu.
e. Press VMAX, press VMIN, then press Show Meas.
San Jos State University Dept. of Mechanical and Aerospace Engineering
rev. 1.9.1
27JAN2014
Intro - 17
What happens when the source and load impedances do not match?
The voltage applied to the load will not match the voltage value displayed on
the function generator (be it amplitude or offset voltage).
Recall from the section on Output Termination that the function generator allows you to choose
an output termination of either 50 or HIGH Z. In exercises 5 through 10, we chose the HIGH Z
output termination, because the input impedance of the oscilloscope with a 10x probe is
approximately 10 M (practically an open circuit). In exercise 11, we chose the 50 output
termination and found a large discrepancy between the voltage displayed on the function
generator and the voltage supplied to the 'scope.
The function generator can be modeled (see Figure 17) as an ideal voltage source, VGEN, in
series with the generator's 50 output impedance. To simplify the discussion, we will assume
this impedance is purely resistive and label it as RGEN.
FUNCTION GENERATOR
R gen = 50
Vgen
The function generator is designed so that the voltage you set on the display is equal to
VGEN if HIGH Z is chosen, or is one-half VGEN if 50 is chosen.
A passive load can be modeled as an impedance. Again, for simplicity, we will assume this load
is purely resistive and label it as RLOAD. As Figure 18 shows, the load is connected in series with
the generator impedance. Therefore, the voltage supplied to the load, VS, is (by voltage division)
some fraction of VGEN.
rev. 1.9.1
27JAN2014
Intro - 18
FUNCTION GENERATOR
Rgen = 50
+
Vgen
Vs
Rload
VS =
R LOAD
VGEN
R GEN + R LOAD
In exercises 5 through 10, we selected HIGH Z termination, and our load (the oscilloscope) was
10 M. The voltage set on the function generator display was 8 VP-P plus 2 VDC offset. The
voltage supplied to our load was
VS =
10M
10M
8VP-P +
2VDC 8VP-P + 2VDC
50 + 10M
50 +10M
The voltage supplied matched the voltage displayed on the function generator.
In exercise 11, we chose an output termination of 50 When this was done, the function
generator changed the displayed voltage value (NOT the actual voltage it output) to be 4 VP-P
with a 1 VDC offset. The actual signal output did not change from 8 VP-P with a 2 VDC offset. (You
can see this on the scope by going through the steps of changing the output termination and
watching the scope trace. The voltage values displayed by the function generator are halved, but
there is no change in the voltage that the scope measures!)
If we actually had a 50 load (instead of the scopes 10 M, and 50 output termination
were selected on the function generator, then the voltage displayed on the generator would match
the actual voltage being supplied to the load.
When connecting the function generator to a circuit, the impedance the function generator sees is
usually much smaller than 10 M, and is also usually much larger than 50 . Therefore, we will
usually see a discrepancy between the voltage supplied to the load and the voltage displayed
on the function generator. Keep this in mind!
For example, if a circuit has an impedance of 250 and you set the function generator to 5 V
(either amplitude or offset), then from Figure 19, the voltage supplied to the load will be
VS
250 // 10 M
250
5V
5V 4.17V
50 250 // 10 M
50 250
rev. 1.9.1
27JAN2014
Intro - 19
Since there will usually be a discrepancy between the voltage displayed on the generator and the
voltage supplied to the load, it is best to set the voltage to the desired value by placing the 'scope
in parallel with the load and measuring the voltage supplied. (See Figure 19.)
LOAD = Rckt
FUNCTION GENERATOR
Rgen = 50
// Rand'scope
10x probe
+
Vgen
Vs
Rckt = 250
R
= 10M
and
10x probe
'scope
Figure 19. Circuit schematic showing 'scope used to measure supply voltage. The input impedance of the scope by
itself is about 1 M and about 10 M with a 10x probe.
With the scope placed in parallel to the circuit, the total load the generator sees is the parallel
combination of RCKT and RLOAD, or 249.9 . Since the scope impedance is so high, it is
effectively an open circuit and has little effect on the voltage being supplied to the circuit.
Exercise 12. -- AC Measurements
The circuit used in Exercise 3 is shown again in Figure 20. This time, you will use your function
generator to supply a 5 V peak-to-peak sinusoid with 2 VDC offset.
Set this voltage using your oscilloscope to measure the peak-to-peak value and the average
value. Record the values measured with your oscilloscope and the values displayed on your
function generator. State whether you have selected HIGH Z or 50 termination of your
function generator.
Is
Ia
+
V1
Vs
R1
R3
V3
R 1 = 10 k
R2 = 2 k
V2
Ib
R3 = 1 k
R 4 = 5.1 k
+
R2
V4
R4
V s = 5Vp-p + 2Vdc
sinusoid
rev. 1.9.1
27JAN2014
Intro - 20
With your DMM connected as an ammeter, measure and record each current indicated in Figure
20. You will need to measure both AC IRMS and DC I. Current cannot be measured with the
oscilloscope with the probes that we have. There are special scope probes called, current
probes that can be used to make current measurements with a scope.
rev. 1.9.1
27JAN2014