Paint Application Standard No. 2: SSPC: The Society For Protective Coatings
Paint Application Standard No. 2: SSPC: The Society For Protective Coatings
Paint Application Standard No. 2: SSPC: The Society For Protective Coatings
June 1, 1996
2.1 DEFINITIONS:
3. Reference Standards
3.1 The documents and standards referenced in this
standard are listed in Section 3.4 and form a part of this
standard.
3.2 The latest issue, revision or amendment of the
referenced documents in effect on the date of invitation to
bid shall govern unless otherwise specified.
3.3 If there is a conflict between the requirements of
any of the cited documents and this standard, the requirements of this standard shall prevail.
3.4 NATIONAL INSTITUTE OF STANDARDS AND
TECHNOLOGY (NIST) STANDARD REFERENCE MATERIALS. (See Section 8.15.)
502
SSPC-PA 2
June 1, 1996
4.1 GENERAL
4.1.1 Access to Bare Substrate: All gages are affected to some degree by substrate conditions such as
roughness, shape, thickness and composition. To correct
for this effect, access to the uncoated substrate is recommended. Another option is to use separate uncoated reference panels with similar roughness, shape, thickness and
composition. (See Sections 8.3 to 8.9.) These would be
used as the bare substrate in the procedures of Sections 4.2
and 4.3. Reference panels shall be of sufficient size to
preclude edge effects. (See Section 8.7.)
Measurements on the bare substrate can be taken
before areas during painting. If the coating has already
been applied to the entire surface, small areas of coating
may be removed and later patched. Do not allow the
removal process to alter the condition of the substrate. Paint
strippers should be used in order to retain the profile.
4.2.3 When the gage no longer agrees with the standard, check the probe for cleanliness. If dirty, clean as
described in Section 8.5.1. If the gage still does not agree
with the standard, the gage is in need of repair or replacement. Some gages can be adjusted to read accurately in a
given range. Adjust the gage to read correctly on a given
standard. Then check the gage on nonlinear scales and any
adjusting feature is linear in nature. Therefore, only a given
segment of the scale can be accurate after adjustment.
4.3 CALIBRATION, VERIFICATION AND MEASUREMENT: TYPE 2CONSTANT PRESSURE PROBE GAGES
4.2.1 For Type 1 gages, use test blocks bearing calibrated nonmagnetic coatings that are traceable to a suitable
national standard. (See Section 8.15.) The standards must
be large enough to exceed the critical mass of steel needed
to satisfy the magnetic field of the Type 1 (pull-off) magnets.
Shims of plastic or of non magnetic metals which are
acceptable for calibration of Type 2 (constant pressure
probe) gages should not be used for calibration of the Type
1 gages. (See Section 8.1.1.) If the manufacturers instructions are in conflict with this standard by allowing the use of
plastic or other nonmagnetic shims for the calibration of a
Type 1 gage, the contracting parties must both be notified of
this fact and agree on a method of calibration.
If proprietary thickness standards are to be used, agreement between contracting parties should be reached prior to
starting the job.
4.3.1 Different manufacturers of Type 2 (constant pressure probe) gages follow different methods of calibration or
adjustment. Calibrate the gage according to manufacturers
instructions.
4.3.2 With a properly calibrated gage, measure the dry
coating as specified. (See Section 4.1.2.)
4.3.3 Verify the calibration of the gage at the beginning
and the end of each work shift with one or more of the
standards. (See Appendix 2). During the work shift, if the
gage is dropped or suspected of giving erroneous readings,
its calibration should be rechecked. If deemed appropriate
by the contracting parties, initial agreement can be reached
on the details and frequency of verification or calibration.
Record the calibration data and the method used to verify
the calibration. If the gage is found to be out of calibration
at the end of the works shift, all measurements made since
the last calibration are suspect.
503
SSPC-PA 2
June 1, 1996
6. Accuracy
7. Disclaimer
5.1.4 If the dry film thickness for any 10 m2 (100 ft2) area
(see Sections 5.1.2 and 5.1.3) is not in compliance with the
requirements of Sections 5.2.1 and 5.2.2, then additional
measurements must be made to isolate the nonconforming
area.
8. Notes
Notes are not a requirement of this standard.
504
SSPC-PA 2
June 1, 1996
8.2 REPEATABILITY: Magnetic gages are necessarily sensitive to very small irregularities of the coating surface
or of the steel surface directly below the probe center.
Repeated gage readings on a rough surface, even at points
very close together, frequently differ considerably, particularly for thin films over a rough surface with a high profile.
505
SSPC-PA 2
June 1, 1996
Spot
Spot
Spot
Spot
Spot
Avg.
1
2
3
4
5
64 m 2.5 mils
76
3.0
53
2.1
76
3.0
58
2.3
66 m 2.6 mils
8.15 Polished metal calibration standards are manufactured by the National Institute of Standards and Technology
(NIST). The chrome plated panels are flat smooth steel 2.86
x 2.86 cm (1.125 x 1.125 in) in size. Examples of some NIST
standards are:
Certified Coating Thickness Calibration Standards
Nonmagnetic Coating on Steel
SRM 1358 Set of 3 80, 225, 1000 m (3, 9, 40 mil)
SRM 1359 Set of 4 48, 140, 505, 800 m (2, 5.5, 20,
31 mil)
SRM 1362a Set of 4 40, 80, 140, 205 m (1.6, 3, 5.5,
8 mil)
SRM 1331a to 1339a Single standards from 3 m
(0.1 mil) to 62 m (2.4 mil)
8.16 CORRECTING LOW OR HIGH THICKNESS: The
506
SSPC-PA 2
June 1, 1996
FIGURE A1
Part "A" of Structure (Area Approx. 10 m2 [100 ft2])
507
SSPC-PA 2
June 1, 1996
508