Trombolyzer Com PDF

This document contains error codes and messages for various issues that may occur when using laboratory equipment. The error codes are related to problems such as full sample racks, incorrect calibration values, invalid test settings and missing reagents or samples. The messages provide instructions for resolving the issues, such as changing values, checking settings, or ensuring components are properly installed.

Uploaded by

JuanPablo Chavez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
239 views15 pages

Trombolyzer Com PDF

This document contains error codes and messages for various issues that may occur when using laboratory equipment. The error codes are related to problems such as full sample racks, incorrect calibration values, invalid test settings and missing reagents or samples. The messages provide instructions for resolving the issues, such as changing values, checking settings, or ensuring components are properly installed.

Uploaded by

JuanPablo Chavez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

155

SM_Compact_en_Feb2011_V01_sc
MP19 - Rack X is full, no free prep-positions
Sample prep: rack is full, scan of new samples is not possible.
MP20 - X value is too small (min. Y)
The calibration value X is too small. t must be greater than Y.
MP21 - X value is too great (max. Y)
The calibration value X is too great. t must be less than Y.
MP24 - X column: value is too small (min. Y)
Calibration table: the value for column X is too small. Please enter a higher value.
MP25 - X column: value is too great (max. Y)
Calibration table: the value for column X is too great. Please enter a smaller value.
MP28 - Reference value must be greater X
The start value in the fully automatic production of calibration curves must be greater than X.
MP29 - ISI < min. ISI value (min. X)
The S value entered in the Calibration curves menu item for manual and automatic production of
calibration curves is less than the value specified. Please correct your entry.
MP30 - ISI > max. ISI value (max. X)
The S value entered in the Calibration curves menu item for manual and automatic production of
calibration curves is greater than the value specified. Please correct your entry.
MP31 - Value is too small (min. X)
A value is too small (incubation, measurement time 1, measurement time , calibration). Change the
value. Service: message from applications sector.
MP32 - Value is too great (max. X)
A value is too great (incubation, measurement time 1, measurement time , calibration). Change the
value. Service: message from applications sector.
MP33 - Too many tests selected (max X)
Check reagent block setting in maintenance.
MP34 - Reagent block not defined
Select a reagent block with the correct number.
MP35 - This value must be empty: end of table is X value
When making manual entries into the calibration table, another value below the limit value (0) has been
entered. Please delete the value entered or change the line containing the limit value.
MP36 - Test X is not selected
Check test setting and reagent block in maintenance.
MP38 - Serial pipetting not possible
Check test setting in maintenance.
MP39 - add 1ml of 5% bleach in wash stn. + ENTER for start
When selecting the Probe cleaning menu item in Hardware, the wash station must be filled with
bleach before the second cleaning stage. Please fill with bleach and press e.
MP40 - Value not OK (X)
Check test setting in maintenance.
15
SM_Compact_en_Feb2011_V01_sc
MP41 - For predilution please pipette the buffer first
Check test setting in maintenance.
MP42 - Volume setting for predilution is not OK (X lines)
Check test setting in maintenance.
MP43 - Up volume of X must be greater 0
Check test setting in maintenance.
MP44 - Down volume of X must be 0
Check test setting in maintenance.
MP45 - Washing or cleaning in predil. X line not possible (must be empty)
Check test setting in maintenance.
MP46 - Only normal pipetting possible
Check test setting in maintenance.
MP47 - Incubation in predilution pipetting not possible (must be 0)
Check test setting in maintenance.
MP48 - For predilution, please pipette plasma in predil cuv.
Check test setting in maintenance.
MP49 - Down vol. for predil cuv. is too small (min. X)
Check test setting in maintenance.
MP50 - Down vol. for predil cuv. is too great (max. X)
Check test setting in maintenance.
MP51 - Down volume > up volume (X) not possible
Check test setting in maintenance.
MP52 - X. line: please pipett predil. plasma in cuv.
Check test setting in maintenance.
MP53 - Down volume of X must be greater 0
Check test setting in maintenance.
MP54 - Up volume of pr. plasma too great (X)
Check test setting in maintenance.
MP55 - Pipetting of predilution not possible
Check test setting in maintenance.
MP56 - Chromogenic coagulation X not possible
Check test setting in maintenance.
MP57 - Following test X is not possible
Check test setting in maintenance.
MP58 - Following test X is not selected
Check test setting in maintenance.
MP60 - No test found with reagent X
Menu Run Control.
157
SM_Compact_en_Feb2011_V01_sc
MP61 - Please read values for reagent X
Menu Run Control.
MP62 - Rotor X is not in working station (Y)
Before scanning the tubes in the sample preparation, rotor X is missing in the device (error code Y).
Please put the rotor in the corresponding position.
MP63 - Rotor X home position not found (Y)
The rotor X Home position is not reached before the samples are scanned (error code Y). Please
check whether the rotor is meshing correctly with the toothed gear or whether the bar code in the
Home position is dirty.
MP64 - No patients in rotor X (Y)
When scanning the samples, no samples were found in rotor X (error code Y). Please check if the bar
codes on the small plasma tubes are aligned correctly.
MP65 - Patient bar code in prep. X is not OK (Y)
Sample X was scanned in previous scan. The actual scan process of sample X ends with an error (error
code Y).
MP68 - Rotor: parameter error X
Start software with protocol. Check via adjusting rotor.
MP70 - Test X not possible: has self depending tests
Calibration: test X can not depend on another test, because test X is main test. Check test setting in
maintenance.
MP71 - Test X is depending on test Y
Calibration: Test X is depending on test Y, additional dependency on another test is not possible. Check
test setting in maintenance.
MP72 - Calibration not possible: depending on test X
The test does not have its own calibration curve, because it is linked to the calibration curve of another
test. Change tests.
MP73 - Test X depending on test Y not possible
Test X can not depend on test Y, because it is the same test. Check test setting in maintenance.
MP74 - Depending tests for calculation type X not possible
Dependency tests are not allowed for the requested calculation type. Check test setting in maintenance.
MP75 - Pipetting liquid X into position Y not possible
Liquid X can not be pipetted into destination position Y. Check test setting in maintenance.
MP76 - Coagulation type of derived test X not OK
Only some coagulation types are allowed for derived tests. Check test setting in maintenance.
MP77 - Too many records selected (max. X)
The maximum amount of data possible for printing or sending has been exceeded (maximum X).
MP78 - Wash sequence needed
After every pipetting of a reagent or specimen, a washing cycle must follow.
MP79 - Wash sequence X in last line of table not possible (must be without C)
Check test setting in maintenance.
158
SM_Compact_en_Feb2011_V01_sc
MP80 - Please put X mm cup on the wash station and press <ENTER>
Serves as a daily check of the needle (see daily routine).
MP81 - Sample in prep. X not found
After acquiring the specimen and during the pipetting, it was detected that the sample tube at position X
was removed (Reflector foil was detected).
MP83 - Dilution X not possible: volume setting of plasma without dilution
Fully automatic calibration: dilutions :1, 3:1, :1 not possible. Requires plasma dilution in volume
setting. Check plasma pipetting in volume table or use other dilutions.
MP84 - USB stick could not be mounted
The USB stick can not be recognized by the system. Remove the USB stick and insert it again.
MP85 - No test on USB stick found
No test was found on the USB stick that could be copied to the system.
MP86 - Device not OK (X)
Accessing Linux device X (e.g. USB stick) not possible. Check device names in profile.local.
MP87 Error from script X
nternal error from copy script. Check device names in profile.local. Reinstall software.
MP88 - No USB stick found (Y)
No USB stick found.
MP89 - Too many USB sticks found (Y)
Too many USB sticks (count Y) found. Remove all USB sticks and external disks from PC and try again.
MP90 - X position difference too large
Hardware/Cuvette Rack Adjust: the tolerance of position X differences is too large. Check via
Adjusting/Sampler positions.
MP93 - X column: not all values ascending
The values entered in the column X (left/right) of the calibration table must be entered in ascending
order to be able to manually produce calibration curves. Please correct this value.
MP94 - X column: two values are equal
Two equal values have been entered in the column X (left/right) of the calibration table in order to
manually produce calibration curves. The values must be different. Please correct this value.
MP95 - X column: not all values descending
The values entered in the column X (left/right) of the calibration table must be entered in descending
order to be able to manually produce calibration curves. Please correct this value.
ME range: (m-e-msg.txt)
ME02 - Key permitted only in main-menu (press ESC)
The key just pressed is not permitted in the current menu. Press the key to go back until you reach
the main menu. You can then re-select the key you pressed previously.
ME03 - Test not available
A test for which there are no abbreviations has been entered in the sample prep menu item.
15
SM_Compact_en_Feb2011_V01_sc
ME04 - Menu entrance not permitted, check calibration data via Manual, Curve
Selecting of Calibration, Curve not possible. Create Calibration via Calibration, Manual, Curve.
ME07 - LED test not possible
The LED cannot be tested during the current routine. Press nand wait until the current pipetting process
is complete. You can then test the LEDs in the Hardware menu.
ME12 - Mixing ball is missing
There is a ball missing in the cuvette rack. The test is repeated automatically.
ME14 - Probe is not in wash position
When exiting Hardware, the needle must first be moved into the wash position. Before exiting
Hardware, you should, therefore, first select the wash station item. Only then should you exit
Hardware using the key.
ME15 - Syringe is not in home position
Before exiting Hardware, after changing the syringe, this has to first be moved into the Home
position. Select the Home position item. You can now exit Hardware using the key.
ME18 - Too many tests selected
n sample prep too many tests are selected for one sample. Delete one or more tests.
ME20 - Prime pumps and probe clean not possible
The Prime Pumps menu item cannot be activated while the system is processing the routine. Please
wait until the system has completed the routine and then restart Prime Pumps.
ME21 - Hardware test not possible
The programs of the Hardware menu item cannot be selected during the current routine. Press n and
wait until the current pipetting process is complete. You can then run the Hardware programs.
ME23 - Volume table for single test is empty
Check test setting in maintenance.
ME25 - Position 7 for buffers only
Check reagent block setting in maintenance.
ME26 - LED Test not finished
The Hardware menu item cannot be exited if the program for the LED test has not been completed.
Please wait until this is the case.
ME27 - Control-plasma not defined
A quality control is integrated in the normal routine. This message is displayed if the control plasma has
not been defined. Change the name if you have incorrectly typed it, or delete the request.
ME28 - Test not defined for this control plasma
A quality check is integrated in the normal routine. The test selected is not defined for this check.
ME29 - Values are not OK
The QC limits in the check values for high, average and low are not correct, e.g. the average value is
less than the value for low.
ME30 - No Curve for this calculation type
Calibration: this calculation type works without calibration curve.
ME31 - Clear track not finished
The Hardware menu item cannot be exited if the process of ejecting the rack has not been completed.
10
SM_Compact_en_Feb2011_V01_sc
ME33 - In the first three lines must be a right value
When manually producing the calibration curve, the values for the first three calibration points must be
entered in the calibration table. Please enter the missing values.
ME34 - In the first two lines must be a right value
When manually producing the calibration curve, the values for the first two calibration points must be
entered in the calibration table (chromogenic tests). Please enter the missing values.
ME43 - Normal < min. calibration value
The normal time entered in the calibration table to manually and automatically produce calibration
curves is too low. Please correct your entry.
ME44 - Normal > max. calibration value
The normal time entered in the calibration table to manually and automatically produce calibration
curves is too high. Please correct your entry.
ME46 - Only one standard print possible
Maintenance: Both types of standard print have been selected from the Print menu. Please select just
one type of standard print.
ME50 - Press <ENTER> to end the probe clean
The Probe cleaning menu item in Hardware must be quit after a sufficient cleaning time by pressing
e.
ME52 - Volume table for duplicate test is empty
Check test setting in maintenance.
ME55 - Please read first control plasma
This message appears when scanning control plasma data using a hand-held scanner.
ME56 - Please read first name of reagent
This message appears when scanning reagent data using a hand-held scanner.
ME57 - Please scan <END> for next line
This message appears when scanning reagent data using a hand-held scanner.
ME58 - Lot number missing
This message appears when scanning reagent data using a hand-held scanner
ME59 - To register the patients please press <F3> first
During the routine, you can place new samples in the rotor once the current rack has been fully
pipetted.
ME60 - SCAN permitted only in main- or plasma-prep menu
The SCAN keys on the system are only permitted in the main or routine menu.
ME61 - LIS communication error (already in work, or ready)
The host computer sends tests for a sample. The actual sample has been processed or is just processing.
Reactivate sample with q or wait until sample processing is finished.
ME62 - Calculation type only duplicate test possible
Calculation type Rat1: needs duplicate testing. Check test setting via maintenance.
ME63 Please put test cup on the washstation and press <ENTER>
Message from Hardware/Probe check.
11
SM_Compact_en_Feb2011_V01_sc
ME64 - Please press <ENTER> for end of probe check
The needle check is hereby ended.
ME65 - Probe check is not finished
Press o/ALARM OFF; follow the instructions in the message box.
ME66 - Please remove test cup from washstation and press <F4>
Now, take the container off the wash station to check the volume and press o/ALARM OFF.
ME69 - Please wait
Please wait until the process has been completed.
ME71 - Error: Copy not OK
An error occurred during copying from or to floppy disk/USB stick.
ME72 - First unit is not OK
The conversion unit for this test is not OK. Check test setting via maintenance.
ME73 - Success
Message appears after copying successfully.
ME 74 - Second parameter must be empty
Check test setting in maintenance.
ME75 - Second parameter can be NorISI, CurISI or empty
Check test setting in maintenance.
ME76 - Second unit is not OK
Check test setting in maintenance.
ME77 - Second print format is not OK
Check test setting in maintenance.
ME78 - Second unit must be empty
Check test setting in maintenance.
ME79 - Second print format must be empty
Check test setting in maintenance.
ME80 - LIS is switched off in system parameters
When sending to the EDP with p it is ascertained that the communication is turned off.
ME85 - Printer is switched off in system parameters
When attempting to print, it was ascertained that the printer is switched off in the system parameters.
ME86 - Please wait: copying
Message during copying data.
ME89 - CurISI can not be first parameter
Check test setting in maintenance.
ME93 - Attention: you are working without checksum
Check system parameter setting in maintenance. Ensure that a checksum on the host connection is not
necessary.
1
SM_Compact_en_Feb2011_V01_sc
ME94 - Error: Thrombolyzer device and host device equal not possible
Check system parameter setting in maintenance. Use different devices for system and host.
ME95 - Attention: probe drives to manual clean position, dont grab into work area!
Hardware/Probe Clean manually: be careful!
ME96 - Attention: wipe probe with alcohol only from top to bottom!
Hardware/Probe Clean manually.
Messages, ST range: (m-e-zust.txt)
ST01 - System is not yet ready
This message indicates that the system cannot be started yet. You must wait for the cuvette rack to eject
after the system has been restarted.
ST02 - System ready
The routine can be started.
ST03 - System is working
The system is working. Details are also provided for the area in which it is working.
ST04 - System stopped with error, restart with F4
Check the incident before you press o/ALARM OFF; call service if necessary.
ST05 - System has finished calibration. See curve
The system has completed the measurements for automatic calibration, for fully automatic calibration or
for recalibration. Enter the Calibration menu item and look at the curve.
ST06 - Stopped (arm is still working) wait for ready
You have pressed nto stop the routine process after pipetting the current rack to e.g. scan in new
samples. Press o/ALARM OFF to restart the routine.
ST07 - Calibration finished. Failed!
The calibration curve produced is incorrect and must be repeated. Check the calibration process.
ST08 - Pipetting stop is active
The pipetting process has been stopped using the stop button on the system.
You have to press the stop button again to restart the process.
1) Only press the stop button if the pipetting process is to be stopped immediately.
) Only briefly press the stop button to e.g. replace a reagent!
Caution: delays caused by the stop button being pressed for too long (>1 minute) may have an impact
on the measurement results!
ST09 - Wait for scanning (arm is still working)
You have pressed the scan button to undertake a scanning process after pipetting the current rack.
Scanning process is started automatically after pipetting the current rack.
ST10 - (Routine)
Status message in (....): This process is started once mis pressed.
ST11 - (Calibration)
Status message in (....): This process is started once mis pressed.
ST12 - (Control)
Status message in (....): This process is started once mis pressed.
13
SM_Compact_en_Feb2011_V01_sc
ST13 - (Hardware)
Status message in (....): Always exit the hardware so the needle is in the wash station.
ST14 - (Prime pumps)
Status message in (....): currently underway.
ST15 - (Probe clean) press <ENTER> to continue
Press eto end cleaning of the needle (recommended interval of 10-0 minutes).
ST16 - Distilled reservoir level low - refill
There is not enough water in the wash tank. You can only start the machine once you have topped up
the water. f there is already sufficient water, the sensor may be defective.
ST17 - Temperature in pipetting station not OK
The warm-up phase can last up to one hour depending on the environmental temperature/conditions.
ST18 - Temperature in incubation station not OK
The warm-up phase can last up to one hour depending on the environmental temperature/conditions.
ST19 - Temperature in measuring station not OK
The warm-up phase can last up to one hour depending on the environmental temperature/conditions.
ST20 - Rack at position X not OK
A problem has occurred when transporting the cuvette rack inside the incubation (e.g. NCU3). Quit the
software and follow the instructions provided in the warning window after the restart.
ST21 - Plasma bloc in working station not present
The plasma rack or rotor is not in the system during routine.
ST24 - Reagent block not present
During the routine, the reagent insert is not in the device.
ST25 - Temperature in reagent station not OK
Service message: in case reagent cooling fails. Check environmental temperature.
ST26 - Stopped with F3 (arm ready: restart F4)
The system has been stopped using nand can be restarted using o/ALARM OFF.
ST27 - Place pre.buf.: pos. 7, plasma: c1, cups: X
This message appears if the fully automatic calibration is to be started. Place the reference plasma in
the reagent block in the Control 1 position, corresponding empty containers (e.g. Hitachi cups) in the
specified positions X (e.g. X..X) in the rotor or plasma rack.
ST32 - Plasma rack not present (arm stopped)
The plasma rack is missing or is not completely inserted.
ST33 - Reagent block not present (arm stopped)
The reagent block is missing or is not completely inserted.
ST34 - Scanning in progress
Wait until the scanning process is complete.
ST35 - System is not yet ready, ejecting strips
Normal status message once the routine software has been started. Ready appears after about 0
seconds.
1
SM_Compact_en_Feb2011_V01_sc
ST36 - > X minutes
Gives the prospective rest time in minutes for the currently started specimens.
ST37 - Host offline
Data transfer to the Host (EDV, LS) is not possible at the moment. Reactivate with bp. Check
system parameter settings via maintenance.
ST38 - Printer offline
Reactivate with bi. Check system parameter settings via maintenance. Check via icon
configure printer.
ST39 - Waste container nearly full
The container for the waste is almost full and must be emptied at the next opportunity.
ST40 - Waste container full
The container for the waste is full and must be emptied.
ST41 - Waste container not present
Please insert an empty waste container into the device.
ST42 - System ready (F3, for start press first F4)
nis activated. Press o/ALARM OFF to restart.
ST43 - Place predil buffer at pos.7 and press <ENTER>
This message appears if recalibration is to be started. Recalibration needs buffer only.
15
SM_Compact_en_Feb2011_V01_sc
Host-
Computer
Compact PC
USB
Compact
Host communication
Introduction
Subject of documentation
The following documentation describes the communication between a host computer and a
Compact.
The host computer is connected to the Compact computer either serially or by means of the laboratory
network (LAN).
The serial connection is described in this document.
Overview
The hardware is modular. t consists of the Compact PC and the Compact. The following diagram
illustrates the setup with a serial connection.
serial
or
network
Serial connection for host to Compact PC
The host computer is connected to the Compact PC by means of serial port.
Network connection (LAN) host to Compact PC
The host computer is connected to the Compact PC by means of the laboratory network (LAN,
Ethernet). This type of connection is described in a separate document.
Serial communication
There are two different connection types for communication between the PC and the host computer:
unidirectional and bidirectional. Unidirectional enables solely the transmission of measuring results to the
host computer. Bidirectional permits the host computer to request the tests for the samples ready to
be measured. The interface parameters Baud rate, data bits, stop bits and parity can be changed in the
maintenance program.
1
SM_Compact_en_Feb2011_V01_sc
Host Computer
Validation
Results
Database
Measurement
Plasma prep.
1

Unidirectional transfer
The connection between the Compact PC and the host computer appears as follows:
Plasma Prep. / Measurement => Database
The values measured by the Compact are stored in the result database
Database / Results => Host computer
17
SM_Compact_en_Feb2011_V01_sc
The test results of a given patient or quality control are transmitted to the host computer as they are in the
database.
There are two methods for transmitting results to the host computer:
1. Method: The test results are always immediately transmitted to the host computer and
simultaneously entered in the PC database.
. Method: The test results are frst entered in the database and can be transmitted by the user to
the host computer by means of the menu.
The methods for data transmission are set in the maintenance program under menu item Defne System
Parameter - Validation.
Validation - No Method 1
"Validation" - "Yes" Method
Two different transmission methods can be set in the Maintenance program.
The following two methods are after: 1: each patient
: each patient and each Quality Control

Protocol
When moving data, the following ASC control codes are added: <STX>, <ETX>, <RS>, <GS>.
These codes indicate the beginning and the end of data transmissions, results and error control.
<STX> indicates the beginning of data transmission
<ETX> indicates the end of data transmission
<RS> a data block follows this in ASC code
<GS> a checksum follows this in ASC code
A <CR><LF> is added to each control character and data record.
Example:
<STX><CR><LF>
<RS>[Data record]<CR><LF>
<GS>[Checksum]<CR><LF>
<ETX><CR><LF>
Checksum
The checksum consists of two ASC codes which are transmitted after ASC code <GS>. t is
calculated by adding together the 8-bit binary values of all codes, starting with <RS> and ending with
<GS>.
Example:
<STX><CR><LF>
<RS>PATENT<CR><LF>
<GS>7<CR><LF>
<ETX><CR><LF>
18
SM_Compact_en_Feb2011_V01_sc
The following codes are used for calculating:
<RS>PATENT<CR><LF>
<GS>
Code 8-bit binary value
<RS> 0001 1110
P 0101 0000
A 0100 0001
T 0101 0100
0100 1001
E 0100 0101
N 0100 1110
T 0101 0100
<CR> 0000 1101
<LF> 0000 1010
<GS> 0001 1101
8-bit binary value of the checksum: 0110 0111
ASC-Code of the check sum: 7
Record block construction for transmitting test results
The data record has a fixed length. Each data record begins with <RS>, followed by the individual
data which is separated by so-called data delimiters, and ends with <CR><LF>.
This data record for moving test results is constructed as follows:
- <RS> ( 1 Character )
- Thrombo ID ( 20 Characters) optional
- Data delimiter ( 1 Character )
- ID ( 1 Character )
- Data delimiter ( 1 Character )
- Prep. No. ( 2 Characters) optional
- Data delimiter ( 1 Character )
- ID No. ( 30 Characters)
- Data delimiter ( 1 Character )
- Test ( 2 Characters)
- Data delimiter ( 1 Character )
- Calculation type ( 1 Character )
- Data delimiter ( 1 Character )
- Single / Double ( 1 Character ) optional
- Data delimiter ( 1 Character )
- Error Flag 1 ( 3 Characters) optional
- Data delimiter ( 1 Character )
- Error Flag 2 ( 3 Characters)
1
SM_Compact_en_Feb2011_V01_sc
- Data delimiter ( 1 Character )
- Measuring time ( 6 Characters)
- Data delimiter ( 1 Character )
- Measuring value ( 7 Characters)
- Data delimiter ( 1 Character )
- INR value ( 4 Characters)
- <CR><LF> ( 2 Characters)
A vertical stroke, | (ASC code 7C) is used as a data delimiter between the individual data.
- Thrombo D is the name of the Compact as entered in the Maintenance program under Defin
SystemParameter. (f not 0 characters long, fill out with blanks). This field can be transmitted as an
option.
- D indicates the type of data record (1 character long).
- Prep. No. indicates the position of the plasma tube in the sample tray ( characters long). This can
also be set as an option.
- D consists of the patients name or D No. which is filled out with blanks if not 30 characters long.
- Test consists of the number (0-3) of the coagulation test. Enter a blank before all single digit
numbers.
- Calculation type consists of a number corresponding to the type of measuring value calculation. This
field always contains a 0.
- Single / Double indicates whether a test is to be carried out as a single or double determination
(1 character long, the 0 is set for single determination and the 1 for double determination). The
transmission can be set as an option.
- Error flag 1 and error flag transmit a number from the error flag list (the transmission of the error
flag can be selected by maintenance: no error flag, one error flag or both error flags are transmitted).
Each error flag is transmitted with three characters.
Should both error flags be transmitted, the second error flag relates to the measurement repetition.
Should the measurement not be repeated, the second error flag is at 1 (meaning no flag has been
set).
- Measuring time contains the measured coagulation time ( characters long).
- Measuring value consists of the determined measuring value (7 characters long).
- NR value consists of the calculated NR value or entry 0.0 in the event that no NR value is to be
calculated for the test ( characters long).