Filled System Temperature Recorders & Recorder Controllers: Specification
Filled System Temperature Recorders & Recorder Controllers: Specification
RECORDER CONTROLLERS
SPECIFICATION
ACCURACY +/-1% FSD
AMBIENT -20C to +50C
TEMPERATURE
MEASURING Rototherm thermal bourdon tube temperature
ELEMENT compensated.
CAPILLARY Microbore stainless steel tube with 3mm
diameter stainless steel cover (type C1) as
standard. Other capillary types are available.
BULB AND STEM Stainless steel BS970 316 S16
Standard diameter is 12.7mm.
FITTINGS Compression gland (adjustable).
Stainless steel BS970 316 S16 (1/2 BSP is
standard). Other BSP, NPT and API are avail
able on request. (suitable for pressures up to
3.5 bar)
CHART DRIVE Synchronous electric, battery or mechanical
spring wound. 24 hour or 7 day. (other rotations
on request)
CHARTS Disc 240mm (9.5 inches) diameter.
Standard dividing 40, 50, 60, 70 or 80 linear
divisions.
INKING SYSTEM Disposable fibre tipped pens. Single pen : Red.
2 pen : red & blue. 3 pen : red, blue and
green.
CASE Glass filled polyester resin. protection rated to
IP55
WINDOW Acrylic
POWER SUPPLY 200 to 250 volts, 50Hz (60Hz available);
100 to 110 volts, 50Hz (60Hz available);
Instruments without electrical control require no
power supply provided that a clockwork chart
drive is fitted. Battery operated chart drives
require 1.5 volt type C cell.
ELECTRICAL ALARM AND CONTROL
Contact rating: 0.1A at 250V AC (non inductive)
(when fitted) 0.01A at 250V DC (non-inductive)
Relay contact 10A at 250V 50Hz
rating
Contact adjustment 2% to 100%
range
MOUNTING Suitable for surface or panel mounting. (Pipe
mounting and portable options available)
WEIGHT Single pen : 6.5 kg (14.25lb)
Two pen : 7.0 kg (15.5lb)
Three pen : 7.5 kg (16.75lb)
The above weights include 3 metres of capillary. For each
additional 3 metres add 0.25kg (0.5lb)
The filled system range of temperature circular chart recorders
uses the well proven and reliable Rototherm stainless steel
thermal system.
These reliable and accurate instruments are suitable for most
applications where there is a requirement for low cost tempera-
ture recording.
Up to 3 pens can be fitted with three temperature systems or
combination of temperature, pressure or flow sensing systems.
FEATURES
1, 2 or 3 pen temperature recording and controlling at low cost
Uses proven and reliable Rototherm filled temperature systems
Simple installation and maintenance requirement
Portable stand or pipe mounting available
Fully mechanical versions available ble
Electrical, battery or mechanical chart drive options
Electrical control options
Pneumatic control options
Pens
Fitted to the pen arms are disposable fibre tipped pens which are
readily replaced without mess. Single pen recorders and the first
pen of the two and three pen recorders trace in red ink, the second
pen in blue, the third in green.
Control Options
Full pneumatic control is available on this range of recorders - for
full details please refer to the Pneumatic recorder controller data
sheet.
Electric Control / Alarm - One and two pen recorders in this range
may be specified with electrical control options.
HISPACONTROL S.L.
P Delicias 65 Bis
28045 Madrid
Tel. 915.308.552
Fax. 914.673.170
E-mail: [email protected]
Web site: www.hispacontrol.com
BS EN
ISO 9001
Registration
no: FM11958
This literature is for general guidance onl y. It does not constitute recommendations, representations or advice, nor is it par t of any contract. The British Rototherm Co. Ltd. 2000
We reserve the right to amend specifications without notice.
1 pen RTO RTO 10
2 pen RTO RTO 20
3 pen RTO RTO 30
MOUNTING
Wall/Panel 4
Portable 5
Pipe 6 XXX
CHART DRIVE
Electric 240V 50Hz A XXX
Electric 110V 50Hz B
Electric other* C D
Spring Wound D E
Battery E F
CHART ROTATION G
24 hour 1
7 day 2 A
Other* 3 B
CONTROLS / ALARMS C
No control
1st pen High & Low A 03
2nd pen High & Low B 06
1st & 2nd pen High & Low C 09
FILLING MEDIUM 12
Mercury M XX
Non- Toxic Filsafe F
BULB OPTION C1
Type 301 1 C2
Type 302 2 C3
Type 303 3 C4
Other* 4 C5
Notes (a) Where Other* is specified please add note to order code stating exact requirements
(b) Should a combination of Pressure & Temperature be required please refer to the pressure recorder data sheet.
(c) When specifying di fferent ranges for 2 or 3 pen recorders, care should be taken to use standard/compatible charts ranges.
RTO SERIES ORDER CODE
DIMENSIONS
336
288
121
87
19 max
228.6
3
7
1
.
5
3
5
6
4
7
5
3
9
6
FREE STANDING
PORTABLE VERSION
all dimensions in mm
BULB TYPES
RANGE (specify C or F)
IMMERSION LENGTH L
(Length in mm)
GLAND MATERIAL
No gland
Mild steel
Stainless steel
Other*
FITTINGS/GLAND
No fitting
Gland 1/2 BSP
Other*
CAPILLARY LENGTH
3 metres (standard)
6 metres
9 metres
12 metres
Other* (enter length in metres)
CAPILLARY
Stainless steel
Copper
Flexible stainless armour
Flexible stainless armour over stainless
Other*
STANDARD TEMPERATURE RANGES
Range
-40 to +40C
-30 to +30C
-30 to +50C
-25 to +25C
-25 to +30C
0 to 40C
0 to 50C
0 to 60C
0 to 100C
0 to 120C
Range
0 to 160C
0 to 200C
0 to 300C
0 to 400C
20 to 120C
50 to 150C
50 to 250C
100 to 400C
100 to 500C
System Filling
f
f, m
f, m
f, m
f, m
f, m
f, m
f, m
f, m
f, m
System Filling
f
f, m
f, m
f, m
f, m
f, m
f, m
f, m
f, m
key f = non toxic filsafe m = mercury
Type 301
This bulb is used when no fitting is required. The bulb may be held
in place by a bracket or clip (not supplied by Rototherm) or may be left free.
If a fitting is not requested at the time of ordering this type of bulb will usually be
given.
Type 302
This bulb has a fitting which slides along the capillary and is clamped into
the required position. Bulb length cannot be specified.
Type 303
The fitting on this bulb is fitted on the actual bulb and not to the capillar y.
When ordering this type of bulb please specify the insertion length "L"of
the bulb - i.e. is the bulb to be fitted into a thermowell? - if yes, what is
the insertion length of the thermowell?
this length is adjustable
length L