DiaDENS-PCM Operations Manual
DiaDENS-PCM Operations Manual
DiaDENS-PCM Operations Manual
com
Электростимулятор с внутренними
и выносными электродами портативный
для стимуляции БАТ и БАЗ «ДиаДЭНСПКМ»
ДиаДЭНСПКМ
Руководство по эксплуатации
РЦ АРТ 04.003.703 РЭ
ТУ 944400235266303-2005
DiaDENS-PCM
Operations Manual
RC ART 04.0-03.7-03 RE
TU 9444-002-35266303-2005
RC АRТ 04.0-03.7-03 RE
TU 9444-002-35266303-2005
EN
www.getdenas.com
CONTENTS
Certificate of Acceptance...............................................137
Coupon for Warranty Repair.............................................139
48
EN
49
EN
www.getdenas.com
PART 1
TECHNICAL PASSPORT
1. Safety regulations
50
EN
Utilization:
All packaging materials are not environmentally
harmful, it may be used repeatedly.
51
EN
www.getdenas.com
52
EN
2. FUNCTION
The DiaDENS-PСM apparatus is used for electrostimulation
of biologically active points and zones (BAP and BAZ). The ap-
paratus has built-in electrodes as well as a slot for connection
of external electrodes — DENS-APPLICATOR*.
The DiaDENS-PСM apparatus is intended for individual ap-
plication in patient care institutions and in life conditions in com-
pliance with the recommendations of the attending doctor.
53
EN
www.getdenas.com
3. SPECIFICATIONS
3.1. Electric impulse (fig. 1) parameters must have the fol-
lowing output parameters:
3.1.1. Minimal impulse parameters:
— duration of the positive impulse, uS, not more than........5
— amplitude of the positive impulse, V, not more than......10
— amplitude of the negative impulse, V, not more than.....10
3.1.2. Maximal impulse parameters:
— duration of the positive impulse, uS.....................500±70
— amplitude of the positive impulse, V.......................30±10
— amplitude of the negative impulse, V:
without load..........................................................350±70
with load (20±5%) kOhm........................................300±70
3.1.3. Minimum load resistance under which the parameters
of the impulse keep, Rmin.......................................500 Ohm
Figure 1. Impulse form
impulse duration
54
EN
55
EN
www.getdenas.com
56
EN
57
EN
www.getdenas.com
4. COMPLETE SET
The complete set of delivery of the DiaDENS-PСM apparatus
should correspond to Table 1.
Table 1
Name Quantity
1
Portable electrostimulator DiaDENS-PСM
Package 1
Power source 2
Holder lace 1
58
EN
5. APPARATUS ARRANGEMENT
Figure 2. DiaDENS-PСM (front)
body
keyboard
power On/Off
59
EN
www.getdenas.com
built-in electrodes
body
cover of power
source replace-
ment
60
EN
Table 2.
Op- Button
eration
Left Right Central Up Down
modes
of appa-
ratus
Switch Decrease
on the of timer`s
When
timer and values
working Decrease Increase Switch
increase and
in elec- of influ- of influ- to ap-
timer val- switch off
trostim- ence`s ence`s paratus
ues by 30 timer if its
ulation power* power menu value is
seconds
modes equals to
with every
zero**
touch**
When Return Move
Confirm Confirm Move up
working to pre- down
the ac- the ac- in this
in ap- vious in this
tion se- tion se- menu`s
paratus menu`s menu`s
lected lected level
menu level level
* Do not used for “SCREENING” mode
** Is used only for “THERAPY” (manual setting) mode
61
EN
www.getdenas.com
Time of re-
Fixed period of minder
the procedure (if
TIMER function Quantity of
is switched on) reminders
Current power
level
62
EN
6. GENERAL RECOMMENDATIONS
6.4. Timer
Choose necessary mode (available in “THERAPY” (manual
setting) mode), the display will show a “STAND-BY” message.
Set the timer if necessary: switch on the timer by pressing the
up button , with every subsequent pressing the timer, value
will increase by 30 seconds, by every pressing of the down but-
ton timer value will be reduced by 30 seconds. When the
timer reaches zero value, it will be switched off.
During the procedure with the timer switched on, timer
countdown shows the time left until the end of the procedure;
with the timer switched off – the timer shows the time passed
since beginning of the procedure.
After the end of the procedure with the timer switched on you
will hear a sound signal. The timer will automatically switch to
the value set before if electrodes have no contact with the skin
surface.
63
EN
www.getdenas.com
6.5. Power
When operating any of the electrostimulation modes (“EX-
PRESS” (express-therapy), “THERAPY” (manual setting),
“MED”), you can change power of influence: by pressing the
left button (power reduction) and the right button (power
increase).
64
EN
65
EN
www.getdenas.com
66
EN
67
EN
www.getdenas.com
68
EN
69
EN
www.getdenas.com
70
EN
8. TECHNICAL MAINTENANCE
8.1. Daily technical maintenance should include the follow-
ing:
— external examination of the apparatus;
— disinfection of electrodes.
Use standard disinfection means (e.g. 70% alcoholic solu-
tion) and soft napkins to clean the electrodes.
8.2. Check of serviceability in accordance with instructions
in Part 6.
8.3. If the apparatus is supposed not to be used for a long
period, remove the power source* from its compartment (Part.
10).
8.4. The battery symbol flashes on the display and there is a
pulsing sound signal. Change the battery.
Attention! If power source are absent more than 10 sec-
onds all individual settings of the user and set reminders will
be deleted.
71
EN
www.getdenas.com
* Set only those power sources that are provided for this de-
vice – type LR6/АА, voltage rating1,5 V.
72
EN
No contact be-
tween the appa-
When using an exter- ratus and an ex- Check contact of
nal electrode the ap- ternal therapeutic the slot (Fig. 2)
paratus is constantly electrode
in the STAND-BY
mode Treat skin with a
Skin is dry tampon moistened
with water
Wrong setting of
Set current date
The apparatus does the current time
and time
not switch on the time and date
set in the CALENDAR Power sources
function Change power
are discharged or
sources
absent
73
EN
www.getdenas.com
74
EN
Manufacturer’s address:
JSC “RC ART”, 620146 Russia,
Ekaterinburg, Postovskogo St., 15
Phone: +7 (343) 267-23-30
http://www.denascorp.ru,
e-mail: [email protected]
75
EN
www.getdenas.com
PART 2
USER’S INSTRUCTIONS
76
EN
77
EN
www.getdenas.com
3. CONDITIONS OF TREATMENT
You do not need any special conditions for treatment with
DENS. Treatment procedures can be carried out both individu-
ally and with help of an operator who will carry out treatment on
those zones and points, which you cannot reach yourself.
Electrotherapy is taken in a comfortable sitting or lying posi-
tion. After the treatment procedure, the patient should relax for
10-15 minutes.
The present manual describes treatment plans with most
some widespread disease states. More detailed information
about DENS opportunities is provided in the Manual on Dy-
namic Electroneurostimulation with apparatuses DiaDENS-T
and DiaDENS-DT*.
Attention! Treat the apparatus electrodes with a standard
disinfection means (e.g. 70% alcoholic solution) after each
procedure. The electrodes of the apparatus should be kept
dry.
78
EN
5. METHODS OF APPLICATION
Dynamic electroneurostimulation with DiaDENS-PCM appa-
ratus is applied using three methods: stable, labile and labile-
stable.
Stable method of application (fixed electrodes) is used for
treating small zones and on places with mutated skin (rash, ab-
scesses, burns, postoperative and post burn scars, edemas,
large birthmarks and so on).
With labile method of application, the built-in electrodes
are moved smoothly within the application zone without taking
them off the skin at 0,5-2-3 cm/sec. Movements are rectilinear,
spiral, circular and other depending on the size and shape of
the zone treated and on uninjured skin.
Labile-stabile method is a combination of both variants of
treatment when electrodes are moved on skin with fixation on
some places (e.g. on the zone of maximal painfulness).
79
EN
www.getdenas.com
80
EN
81
EN
www.getdenas.com
Fre-
quency, Pathology*
Hz
Autoimmune diseases, tachycardia, weakness in
1,2
knee joints
1,6 Arthritis-arthrosis
82
EN
83
EN
www.getdenas.com
84
EN
85
EN
www.getdenas.com
7. MED (PROPHYLAXIS)
MENU → MED
The MED* mode (PROPHYLAXIS) is
applied in cases of expected intensive
physical and mental work, physical and
86
EN
87
EN
www.getdenas.com
8. SCREENING
«SCREENING» mode allows selecting the most optimal
zones for impact to enforce system effect of DENS. The mode
allows determine the latent (hidden problem zones) trigger
zones by estimation of the skin electrical resistance increase
to nearby areas of the chosen zone.
Work procedure:
МЕNU → SCREENING
Put the electrodes of the apparatus to the chosen skin zone.
Frequency (10 Hz) and power of impact is automatically set by
the apparatus.
ATTENTION! During work in «SCREENING» mode the elec-
trodes must be set in “stable” way that is you may not move the
electrodes during work with the apparatus.
88
EN
9. Special EMC-Information:
9.1. The use of accessories, transducers, cables and cable
length other than those specified, with the exception of trans-
ducers and cables sold by JSC RC ART as replacement parts
for internal components, may result in increased emission
and/or decreased immunity of the Portable electrostimulator
DiaDENS-PCM.
89
EN
www.getdenas.com
90