TSS70 130 PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

- 1 -

CTS GmbH
Lotzencker 21
72379 Hechingen
Tel.: 07471/98 50-0
Fax: 07471/98 50-23
www.cts-umweltsimulation.de

CTS - Thermal Shock Test Cabinet, Type TSS -70/130
Offer Nr. 180614A1





- 2 -



CTS - Thermal Shock Test Cabinet, Type TSS -70/130



Your benefits:

very low-noise level, only 58 dB(A)
user-friendly and convenient handling and programming with the colour CTS-Touch
panel or with the optional CID-Software on the PC
uniform operation due to identical operating panel for all cabinets from bench top to
walk-in chambers
use of refrigerants according to actual technologies
low power consumption, please see quotation text



Included in basic equipment

saving of energy by optimized control system (eco-mode)
1-chamber operation (temperature cycling in cold chamber)
Ethernet-Interface (incl. CID-network-version)
RS 232 - Interface
Temperature measuring on test equipment (movable PT 100)
Load for test equipment 50 kg

compressed air dryer for dehumidification to do up to 3000 cycles without
defrosting
operating panel with colour display LCD for easy handling menu-guided operation
defined keys for elementary functions of the installation such as start/stop,
fixed-value/automatic operation, interruption, light, etc.
error display in plain text
potential-free contact for malfunction signal
2 digital Out, potential-free, to switch test specimen ON/OFF
adjustable software-temperature-limiter min./max.
safety temperature limiter according DIN EN 14597
Door with window (460 x 300 mm)
1 x shelf, stainless steel (max. 5 kg), adjustable in height
1 x entry port 125 mm with cap
door lockable; one-hand operation with integrated safety function
to avoid unintentional closing
Test chamber light (LED Light) switched via operation panel
international standards are fulfilled (CE-Conformity)






- 3 -





Price overview:



Thermal shock test chamber, type TSS -70/130 56.580,50.- EUR
ex works CTS Hechingen Germany, unpacked



Options

CID Software 1.360,00.- EUR
Ethernet interface included
USB interface 215,00.- EUR
Air dryer system included
3/N, 380V 60Hz 3.350,00.- EUR
Additional shelf (wire mesh). 310,00.- EUR
max. load 5 kg / shelf. in total, max. 5 shelves possible
Increased shelf 960,00.- EUR
max. load 25 kg / shelf. in total, max. 5 shelves possible
(max. possible load for basket 50 kg)
Seaworthy package 1.000,00.- EUR


Note: After 10.000 cycles the lifting spindle have to be greased. This is a maintenance work
and is subject to a charge. It will not be covered by warranty.



- 4 -





Optimized control system to save energy costs (eco-mode)


Doing a shocktest the basket (incl. test equipment) is for a longer period in one temperature
zone (hot/cold chamber). During the whole dwell time, the other chamber has to hold the
temperature, that cost a lot of energy.


If the electrical load in the non-active chamber will be switched off (heating elements in the
hot chamber and fan and compressor in the cold chamber) the power consumption can be
reduced a lot, unless the dwell time is long enough to achieve afterwards the necessary test
temperature in the chamber.

As an example, the following cycle with the following conditions was used in a shocktest
chamber:

Temperature cold chamber -40 C (Pre-tempering -50 C)

Temperature hot chamber +85 C

Dwell time for each chamber 1 h

Weight of test specimen 10 kg steel


Result:

With an assumed operation time of 6000 hours per year, you have to calculate the following
consumption of energy:


Without optimized control system approx. 34.800 kWh.

At identical operation with
Optimized control system approx. 23.100 kWh.







- 5 -



Technical data


Hot chamber:

Temperature range +50 C to +200 C

Temperature fluctuation 1 K temporally
under stabilized condition 2 K in space

Temperature change rate 6 K/min from ambient temperature to +180C
without test equipment
in average

Cold chamber:

Temperature range -80 C to +100 C

Temperature fluctuation 1 K temporally
In stabilized condition 2 K in space

Cooling change rate 4 K/min from +100C to -50C
without test equipment
in average

Heating change rate 2,5 K/min from -70C to +100C
without test equipment
in average


Performance data

Temperature change rate up to 100 K/min
for shock-test

Adaption time
with 10 kg Alu approx. 18 K/min in temp. range 65 C to
+200 C means <15 min

Time for changing < 10 sec.

Max. load for basket 50 kg
(shared equally on min 4-5 shelves)

Fulfilled test norms MIL-STD-810 E, Method 503
MIL-STD-883 E, Method 1010.7
Var. A,B,C,D,F
IEC 60068-2-14, Test Na
DIN 40 046, page 14, test Na


All figures are values which have been obtained at an ambient temperature of +25 C
and nominal voltage 400 V, with / without test specimens, without radiation and without
accessories.
Options can influence the technical data.



- 6 -



Dimensions:

Test space capacity 130 litre

Test space 480 x 660 x 410 mm (W x D x H)

Overall dimensions see enclosed drawing

Weight approx. 950 kg net


Data for installation and operation:


Nominal voltage 400 V +6/-10 %, 3/N, 50 Hz
Nominal power 10 kW
Nominal current 17,5 A

Connection CEE 32 A

Fuse (on site) 32A time-lag

Protection class IP 22

Power consumption:

Consumption in average at a 7,7 kWh
test cycle from -40C to +120C
with 10kg Alu
and 30min dwell time

Consumption after 16,5 KW
changing the basket

Cooling water: water pressure 2,5 bar to 6 bar over pressure
losses in the systems appr. min. 1,5 bar
max. 3 bar
Inlet R 1
Outlet R 1
Cooling water temperature max. +20 C
Cooling water consumption max. 1,5 m
3
/h
at t = 10 K

Compressed air: Inlet, coupling with hose connection 9,
Pressure: 7 - 10 bar
Consumption: max. 1,5 Nm3 / h
Max. Inlet temperature: 35C
Compressed air quality classes
acc. to ISO 8573-1
- particle: class 3
- water content: class 6
- oil content: class 3

Drain for condensate: pressureless R




- 7 -



Design:


Test space and basket stainless steel V2A, grade 1.4301

Wire mesh shelf Mesh width 8 mm,
Size 465 x 640 mm
max. load 5 kg / shelf (standard)
max. 5 shelves

Entry port stainless steel tube 125 mm with cap

Basket drive via electromotor and spindle

safety devices mechanical and electrical locking of the
basket when the door is opened.


Heating/Cooling heating by electrical resistant heaters, made of
stainless steel,
mechanical cooling by direct evaporation in
laminated heat exchanger

Illumination powerful and bright LED light (approx. 2400
Lumen) installed on the door.

Casing galvanized steel, resistant powder coating

Colour casing RAL 9006, white-aluminium
decorative plates RAL 9007, grey-aluminium

Door with window door can be opened completely
in the hot chamber hinged on left-hand side,
one hand operation, lockable
size of window 460 x 300 mm

Refrigerating unit water cooled, low noise refrigeration unit,
continuous performance adjustment via
electronic monitoring and control system,
environmental friendly refrigerants R 404 A / R
23

Heat dissipation to the max approx. 1,5 kW
installation room

Sound pressure level in accordance 60 dB(A), measured in a distance of 1 metre
to DIN EN ISO 3744 from front




- 8 -



Control - 32-bit controller
- operating panel ergonomically integrated in the
front
- digital display of temperature, set value and
actual value in colour
- numerical input of set values
- error messages in clear text with report function
- compatible with all CID software versions
- external control by customer via ASCII protocol,
identical for all CTS equipment
- 3 analog channels visible at the same time, i.e.
temperature hot chamber, temperature cold
chamber, PT 100 sensor for basket,
T, several movable PT 100, etc. scrollable

Functions - menu guided
- error report
- graphic curve progressions
- easy input of test programs
- graphic program preview

Measurement sensor temperature sensor PT 100, DIN 43 760,
class A

Malfunction signal supplied to potential-free contact
max. load 30 V - DC, 2A

Touch-Panel:






























- 9 -




Data interfaces Networking of the CTS cabinets via ETHERNET
interface with customer network (incl. CID-
network-version) or RS 232 - interface for
communication via a pc.

As an option the cabinets can be cross-linked via
RS 485 - interfaces. This will require installation
of cabling by CTS. This method is necessary if
long distance cabling cannot be avoided and our
optional cross-linkage via ETHERNET is not
utilised.

Optional USB port instead of RS 232.










Note:

Company-specific standards and norms are not considered.

Clarifications with public authorities are on site. Costs are not included in the quotation.


The installation cannot be used for tests on explosive, corrosive, toxic or easy inflammable
materials or with specimens generating or releasing such materials. This applies particularly
to all tests with liquids that boil easily, fuel, hydraulic fluids, lubricants, etc.
In this case, please pay attention to the information of the material safety data sheet.

The operator has to check the material compatibility of the materials fitted in the testing room
(stainless steel 1.4301, non-ferrous metals, aluminium, silicone) to the materials/gases
which might be emitted by the test material. This can form acids or bases when exposed to
humidity. The leaking materials/gases can lead to extensive damage of the equipment.

CTS GmbH will not take responsibility (no warranty) for damages resulting of emission from
toxic or aggressive substances from test specimen.



- 10 -



The technical design of the unit meets on the basic of its conception and of
its construction the relevant basic security and health requirements by the
EC directive:


EC Directives and National Legislation

EC Directive for machines 2006/42/EC
EC Directive on Low Voltage 2006/95/EC
EC Directive on EMC 2004/108/EC
EC Directive for pressure devices 97/23 EC


Mechanical Norms

DIN EN ISO 13857 (issue 06/2008)
DIN EN ISO 378 1,2,3,4 (issue 06/2008)
DIN EN 563 1,2,3,4 (issue 01/2000)
DIN EN ISO 12100 (issue 03/2011)
AD2000 (issue 10/2000)


Electrical Norms

DIN EN ISO 13849-1:2008-12
DIN EN ISO 13849-2:2013-02
DIN EN 61000-6-2: 2006-03
DIN EN 61000-6-3: 2011-09
DIN EN 61010-1: 2011-07
DIN EN 61010-2-010: 2004-06
DIN VDE 0100-410: 2007-06
BGV A3



- 11 -



CTS-Chamber, TSS-70/130




1 Entry port 125 mm
2 Control panel
3 Window 460x300
4 Electr. Connection, cabel
length 5 mtr.
5 Compressed air inlet,
coupling with hose
connection 9, Area for
cooling air outlet, both sites
7 Drain for test room R
9 Cooling water outlet R 1,
10 Cooling water inlet R 1,
11 Main switch
12 Control unit
13 Operating temperature
limiter, Option
14 Cable entry for interface
parts

* clearance width/high
+ entrance clearance




- 12 -




General conditions of offer:

This quotation is based on our general Terms and Conditions for sales, supply and services.
You can have a look for our Terms and Conditions on our website www.cts-
umweltsimulation.de. If you require a copy, please contact us.

Delivery period approx. 14-16 weeks or according
to agreement, ex CTS (unpacked).
Please recheck binding delivery terms when
placing firm order.

Value added tax is not included in the prices

Payment 100% with letter of credit with order confirmation

Site services Power supply
Cooling water connection for water-cooled units
Distilled, fully demineralized or adequately
treated water for test units with humidification
system.

Reservations We reserve the right of changes in design
resulting from technical progress and therefore
changes in supply and consumption data. Verbal
agreements leading to a modification of the
quotation will only become valid if they are
confirmed by us in written form.


Validity of quotation 3 months


Warranty 12 months (after delivery ex works)

You might also like