Drug Delivery Inhalation Bioequivalence
Drug Delivery Inhalation Bioequivalence
Drug Delivery Inhalation Bioequivalence
Introduction
2. Current status of regulatory
guidance
3. Interpreting in vitro
equivalence data
4. Interpretation of clinical
bioequivalence studies
5. Integrating guidance with
experience
6. Expert opinion
Review
Establishing bioequivalence for
inhaled drugs; weighing the
evidence
Peter T Daley-Yates
v
s
B
u
d
e
s
o
n
i
d
e
F
l
e
x
h
a
l
e
r
1
,
2
G
o
o
d
m
a
t
c
h
E
q
u
i
v
a
l
e
n
t
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
N
o
t
e
q
u
i
v
a
l
e
n
t
O
p
e
n
-
l
a
b
e
l
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
3
7
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
p
l
a
c
e
b
o
-
c
o
n
t
r
o
l
l
e
d
,
p
a
r
a
l
l
e
l
g
r
o
u
p
,
1
2
-
w
e
e
k
;
n
=
6
2
1
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
p
l
a
c
e
b
o
-
c
o
n
t
r
o
l
l
e
d
,
p
a
r
a
l
l
e
l
g
r
o
u
p
,
1
2
-
w
e
e
k
;
n
=
5
1
6
B
u
d
e
s
o
n
i
d
e
T
u
r
b
u
h
a
l
e
r
v
s
B
u
d
e
s
o
n
i
d
e
E
a
s
y
h
a
l
e
r
3
,
4
S
m
a
l
l
d
i
f
f
e
r
e
n
c
e
s
E
q
u
i
v
a
l
e
n
t
(
c
h
a
r
c
o
a
l
b
l
o
c
k
)
E
q
u
i
v
a
l
e
n
t
O
p
e
n
-
l
a
b
e
l
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
3
3
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
c
o
n
t
r
o
l
,
p
a
r
a
l
l
e
l
g
r
o
u
p
,
1
2
-
w
e
e
k
;
n
=
1
6
1
B
D
P
I
n
n
o
v
a
t
o
r
C
F
C
M
D
I
v
s
B
D
P
G
e
n
e
r
i
c
C
F
C
M
D
I
5
G
o
o
d
m
a
t
c
h
E
q
u
i
v
a
l
e
n
t
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
E
q
u
i
v
a
l
e
n
t
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
5
1
F
P
/
S
a
l
m
e
t
e
r
o
l
I
n
n
o
v
a
t
o
r
p
M
D
I
v
s
F
P
/
S
a
l
m
e
t
e
r
o
l
G
e
n
e
r
i
c
p
M
D
I
6
S
i
m
i
l
a
r
E
q
u
i
v
a
l
e
n
t
f
o
r
F
P
;
n
o
t
e
q
u
i
v
a
l
e
n
t
f
o
r
s
a
l
m
e
t
e
r
o
l
N
D
O
p
e
n
,
r
a
n
d
o
m
i
s
e
d
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
(
c
h
a
r
c
o
a
l
b
l
o
c
k
)
;
n
=
3
1
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
3
1
F
P
D
i
s
k
u
s
v
s
F
P
D
i
s
k
h
a
l
e
r
7
-
-
9
S
m
a
l
l
d
i
f
f
e
r
e
n
c
e
s
N
o
t
e
q
u
i
v
a
l
e
n
t
i
n
h
e
a
l
t
h
y
.
C
l
o
s
e
m
a
t
c
h
i
n
a
s
t
h
m
a
t
i
c
s
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
E
q
u
i
v
a
l
e
n
t
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
p
l
a
c
e
b
o
-
c
o
n
t
r
o
l
l
e
d
,
c
r
o
s
s
o
v
e
r
,
4
-
d
a
y
d
o
s
i
n
g
;
n
=
2
1
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
1
2
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
1
2
O
p
e
n
-
l
a
b
e
l
,
n
o
n
r
a
n
d
o
m
i
s
e
d
,
u
n
c
o
n
t
r
o
l
l
e
d
,
7
.
5
d
a
y
;
n
=
2
4
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
6
-
w
e
e
k
;
n
=
2
3
2
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
1
2
-
w
e
e
k
;
n
=
2
1
2
F
P
H
F
A
M
D
I
v
s
F
P
C
F
C
M
D
I
1
0
G
o
o
d
m
a
t
c
h
N
o
t
e
q
u
i
v
a
l
e
n
t
a
t
a
l
l
d
o
s
e
s
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
E
q
u
i
v
a
l
e
n
t
O
p
e
n
-
l
a
b
e
l
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
2
3
S
a
l
m
/
F
P
D
i
s
k
u
s
v
s
S
a
l
m
/
F
P
R
P
I
D
1
1
G
o
o
d
m
a
t
c
h
N
o
t
e
q
u
i
v
a
l
e
n
t
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
E
q
u
i
v
a
l
e
n
t
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
c
o
n
t
r
o
l
,
c
r
o
s
s
o
v
e
r
,
1
4
d
a
y
;
n
=
2
2
B
e
c
l
o
m
e
t
a
s
o
n
e
D
P
C
F
C
M
D
I
v
s
B
e
c
l
o
m
e
t
a
s
o
n
e
D
P
H
F
A
M
D
I
5
B
i
g
d
i
f
f
e
r
e
n
c
e
s
N
o
t
e
q
u
i
v
a
l
e
n
t
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
N
o
t
e
q
u
i
v
a
l
e
n
t
D
o
u
b
l
e
-
b
l
i
n
d
,
r
a
n
d
o
m
i
s
e
d
,
c
r
o
s
s
o
v
e
r
,
s
i
n
g
l
e
d
o
s
e
;
n
=
5
1
F
l
u
n
i
s
o
l
i
d
e
C
F
C
M
D
I
v
s
F
l
u
n
i
s
o
l
i
d
e
H
F
A
M
D
I
1
2
B
i
g
d
i
f
f
e
r
e
n
c
e
s
N
o
t
e
q
u
i
v
a
l
e
n
t
(
n
o
c
h
a
r
c
o
a
l
b
l
o
c
k
)
N
o
t
e
q
u
i
v
a
l
e
n
t
O
p
e
n
-
l
a
b
e
l
,
r
a
n
d
o
m
i
s
e
d
,
a
c
t
i
v
e
-
c
o
n
t
r
o
l
,
p
a
r
a
l
l
e
l
g
r
o
u
p
,
1
3
.
5
d
a
y
s
;
n
=
3
1
1
P
e
r
s
s
o
n
2
0
0
8
;
2
K
e
r
w
i
n
2
0
0
8
;
3
L
a
h
e
l
m
a
2
0
0
4
;
4
T
u
k
i
a
i
n
e
n
2
0
0
2
;
5
D
a
l
e
y
-
Y
a
t
e
s
1
9
9
9
;
6
C
l
e
a
r
i
e
2
0
1
0
;
7
F
a
l
c
o
z
2
0
0
0
;
8
M
a
c
k
i
e
2
0
0
0
;
9
K
u
n
k
a
2
0
0
0
;
1
0
D
a
l
e
y
-
Y
a
t
e
s
2
0
0
9
;
1
1
N
o
l
t
i
n
g
a
2
0
0
1
.
Bioequivalence of inhaled drugs
1302 Expert Opin. Drug Deliv. (2011) 8(10)
E
x
p
e
r
t
O
p
i
n
.
D
r
u
g
D
e
l
i
v
.
D
o
w
n
l
o
a
d
e
d
f
r
o
m
i
n
f
o
r
m
a
h
e
a
l
t
h
c
a
r
e
.
c
o
m
b
y
N
o
v
a
r
t
i
s
P
h
a
r
m
a
o
n
0
2
/
0
3
/
1
2
F
o
r
p
e
r
s
o
n
a
l
u
s
e
o
n
l
y
.
in vivo the oropharyngeal cross section can vary with device
type, device resistance, tongue position and patient effort [19].
Attempts to design induction ports that better reflect
and/or mimic physiological reality have included changing
the dimensions and angles in the port, using sticky surfaces
to reflect those present in vivo and using casts from
cadavers [20-22]. Magnetic resonance imaging data have also
been used to determine the retention effect of the oropharyn-
geal airspace during the administration of drug aerosols as
well as to generate mouth--throat models that can be used
with patient relevant flow profiles to evaluate emitted
APSD [23,24].
Deposition within the cascade impactor occurs through
impaction that differs from the in vivo situation where depo-
sition occurs through impaction, sedimentation and diffusion,
the extent of which is dependent on particle size and the inha-
lation manoeuvre of the patient. The cascade impaction test is
conducted at a fixed flow rate so as to allow the aerosol to be
characterised into predefined size ranges, this differs from the
in vivo situation where the inhalation profile is dynamic with
varying flow and acceleration rate.
In an MDI, drug delivery is ballistic in nature being driven
by the propellant present in the formulation, this means that
the patient has to carefully coordinate their inhalation
manoeuvre with actuation of the device; any differences
between MDIs that may impact this coordination would
not be detected by the cascade impaction test. In the case of
the DPI, aerosolisation is driven by the patients inhalation
manoeuvre. This means that, when comparing DPIs, the
resistance of the device itself must be considered as few devices
have the same airflow resistance [25], thus affecting the flow
rates achieved by patients.
As DPI performance is often a function of flow rate, which
can differ from device to device [26], it is necessary to compare
in vitro product performance across a range of flow rates.
Several different approaches have been adopted to address
these issues. Flow rate dependency can be studied using a
range of fixed flow rates such as 30, 60 and 90 l/min; how-
ever, this approach results in different pressure drops across
the device for devices of differing resistivity [27]. An alternative
approach is to compare devices using the same pressure drop
[typically 4 kPa (European Pharmacopedia)], but this means
that the devices operate at different flow rates. Therefore,
the stage cutoffs for the two test conditions will differ. Conse-
quently, attributes such as fine particle dose (< 5 m) can only
be derived by interpolation, and comparison of full cascade
profiles is not straightforward. Using the same pressure drop
with different flow rates may still not allow the comparison
of devices from the perspective of comparing devices with
the same patient inspiratory effort [28]. Cascade impactors
are limited in the range of flow rates at which they can operate
to ensure that stage cutoffs are in the appropriate size
range [29]. This means that the flow rates at which they are
operated may not be the same as those actually achieved by
some patient populations using the device in clinical practice.
In addition to a fixed flow rate, cascade impactors also use
fixed high acceleration rates and airflow profiles that do not
Box 2. Key regulatory requirements for pharmacokinetic studies of orally inhaled drugs in the
European Medicines Agency guideline.
The demonstration of bioequivalence for orally inhaled drugs requires that standard criteria be fulfilled, that is, 90% confidence
intervals for the log-transformed test/reference C
max
and AUC
(0--t)
ratios should lie within 80 -- 125% Tighter limits for AUC and
possibly C
max
may be appropriate for drugs with a narrow therapeutic index. Widened limits for C
max
may be acceptable for
highly variable products
Bioequivalence should be confirmed for partial AUC as a measure of early exposure where a rapid onset of effect is important
Both pulmonary deposition and total systemic exposure should be assessed, unless drug absorption via the oral route is very low
such that pulmonary and systemic bioavailabilities are essentially the same
Total systemic exposure may be acceptable as a surrogate of systemic safety
Dose selection should be based on pharmacokinetic linearity/nonlinearity
Both urinary or plasma pharmacokinetic studies are acceptable in adults, whereas in children only the latter are advocated.
Where urinary pharmacokinetic studies are undertaken, plasma C
max
should be estimated, if feasible, alongside the urinary data
Pharmacokinetic data for parent compounds/pro-drugs should be presented alongside that of active metabolites, assuming
pharmacokinetics of the former are linear and plasma concentrations easily measurable, as C
max
for parent compounds is more
sensitive to detect differences between products
For pressurised metered dose inhalers, pharmacokinetic data comparing test and reference products in conjunction with spacers
should be provided unless comparative in vitro spacer data satisfy stringent criteria for bioequivalence
For dry powder inhalers, the relevance of differences in intrinsic device resistance should be considered with respect to children
AUC
(0--t)
: Area under the plasma concentration--time curve from time zero to infinity; C
max
: Maximum plasma concentration.
Box 3. Key components of the aggregate
weight of evidence approach described by the
FDA [15].
Similarity of formulation
Similarity of device design
Comparative in vitro tests
Comparative systematic exposure studies
Pharmacodynamic or clinical end point studies
Daley-Yates & Parkins
Expert Opin. Drug Deliv. (2011) 8(10) 1303
E
x
p
e
r
t
O
p
i
n
.
D
r
u
g
D
e
l
i
v
.
D
o
w
n
l
o
a
d
e
d
f
r
o
m
i
n
f
o
r
m
a
h
e
a
l
t
h
c
a
r
e
.
c
o
m
b
y
N
o
v
a
r
t
i
s
P
h
a
r
m
a
o
n
0
2
/
0
3
/
1
2
F
o
r
p
e
r
s
o
n
a
l
u
s
e
o
n
l
y
.
reflect patient profiles [30]. This is relevant when comparing
DPIs, as the performance of a device may be influenced
by the acceleration rate used in testing [31]. In contrast,
others have concluded that constant flow rates could be used
providing that the rates selected reflected those expected
in vivo [32].
The differences discussed might explain some discrepancies
between clinical observations and in vitro data. For example, a
comparison of two DPIs raised concerns about the predict-
ability of in vitro FPM data [33]: the reservoir powder inhala-
tion device (RPID) and Diskus
(GlaxoSmithKline, Ware,
UK) inhalers administered the same lactose blend of salme-
terol and fluticasone propionate. The performances of the
inhalers were comparable over a range of flow rates. In vitro
FPM data were similar for the two devices, but pharmacoki-
netic data showed that the devices could not be considered
bioequivalent in terms of fluticasone propionate AUC
[RPID:Diskus estimated ratio: 2.00; 90% confidence interval
(CI): 1.56 -- 2.55] and salmeterol C
max
(ratio: 1.92; 90% CI:
1.64 -- 2.25). However, data showed the two inhalers to be
bioequivalent in terms of mean morning peak expiratory
flow in a clinical efficacy study. There was clearly a lack of
association both between in vitro FPM and in vivo pharmaco-
kinetic data and between the pharmacokinetic data and clini-
cal efficacy data. The key point to takeaway from this
observation is that the difference in pharmacokinetic profile
between the two inhalers could not be predicted from the
in vitro data alone.
4. Interpretation of clinical bioequivalence
studies
There are relatively few well-documented examples of inhaled
product bioequivalence assessments that contain data from
in vitro profiling, pharmacokinetics, lung deposition and clinical
efficacy findings. Some of the best documented examples were
summarised recently [34]. These studies included comparisons
of generic and innovator products and studies assessing the effect
of changes made to propellant or inhaler type. In most examples,
in vitro, pharmacokinetic and clinical efficacy data were not in
agreement. In general, in vitro data did not predict correctly
the findings of pharmacokinetic and/or clinical efficacy studies
(Table 1), and the pharmacodynamic data were less discriminat-
ing than pharmacokinetic assessments. Of the examples shown
in Table 1, three are studies where bioequivalence was concluded
from pharmacokinetic data. However, it would appear that,
when other data are considered more closely, these conclusions
may be unreliable. For example, a comparison of budesonide
administered using either a Flexhaler (AstraZeneca,
Soldertalje, Sweden) (budesonide 180 g) or a Turbuhaler