Agilent Models 6690A-6692A System DC Power Supply: Data Sheet

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Agilent Models 6690A-6692A

System DC Power Supply


Data Sheet
Low ripple & noise
Fast up-and-down programming
High accuracy current programming and read back
Industry standard SCPI programming commands
Analog programming
Analog monitoring
Parallel connection for more current output
Full protection from overcurrent, overvoltage overtemperature
Remote sense
Electronic calibration
Minimize your testing downtime with this reliable,
high power DC supply
2
The 6690A Series is DC
power you can depend on.
At the 6600 watt power level, its
not easy to design a reliable DC
power supply. Agilent Technologies
builds in ruggedness and durability,
to meet even the grueling needs of
continuous stress testing in harsh
environments. During the develop-
ment phase, Agilent power supplies
undergo a battery of environmental
tests such as 8-days temperature
profile. Other tests include humidity,
altitude, shock and vibration, ESD,
AC line tests, EMC and RFI. The
power supplies are designed with
built-in margin so that they can meet
their specifications over time, under
all conditions, and withstand peak
stress.
Protect your DUT
At these high power levels, you
want to protect your expensive DUT.
Whether your DUT is a rack of data
storage control units, or a base
station ready for deployment, it
is quite valuable and well worth
recovery. The 6690 Series has a wide
range of protection features available
to protect your DUT from damaging
conditions.
Overvoltage protection
Overcurrent protection
Overtemperature protection
Programmable shut-down
Error status reporting
DFI-RI
External TTL shut-down input
Linked power supply shutdown
Easy GPIB programming
All Agilent system power supplies
have SCPI (Standard Commands for
Programmable Instruments) based
command sets. This means that
the same function would have the
same command for any instrument.
For example, to program the output
voltage on two power supplies, the
command would be the same, and
to measure the output voltage the
command would be the same for the
Agilent system power supply as any
SCPI voltmeter. Using SCPI instru-
ments makes your software simpler
and quicker to design.
Easy system configuration
and enhancement
The 6600 watt DC power supplies
of the Agilent 6690 Series are the
same size as the 5000 watt power
supplies of the Agilent 6680 Series.
This means that system power can be
upgraded without needing more rack
space. All programming commands
and features are 100% compatible
between both series. Also, up to
three same model 6690 Series power
supplies can be connected in parallel,
to provide additional power as your
needs expand.
Manual power supply control
Non-automated testing in R&D,
or power for the repair bench in
manufacturing, are no problem for
the 6690 Series. The front panel has
everything that is needed to control
both the power supply output and the
associated protection features. Both
the output voltage and current can be
easily monitored on the front panel
meters.
Analog programming
and monitoring
Analog signals can be used to program
the output voltage and current. This
allows custom waveforms to be
generated, and is also useful for
process control applications. The
output current and voltage can be
monitored via analog signals on
the rear panel, for custom interface
applications, and process control
implementations.
You Have Better Things to Worry About than DC Power.
Will failure of your DC power supply delay your testing?
Will failure of your DC power supply damage your DUT?
Will failure of your DC power supply keep you from meeting your customers needs?
3
Parameter Agilent model number
6690A 6691A 6692A
Output ratings
Voltage 0-15 V 0-30 V 0-60 V
Current*
*Derated linearly 1%/C from 40 to 55 C
0-440 A 0-220 A 0-110 A
Programming accuracy (@ 25 5 C)
Voltage 0.04% + 15 mV 30 mV 60 mV
Current 0. 1% + 230 mA 125 mA 65 mA
Ripple & noise
(from 20 Hz to 20 MHz with outputs
ungrounded, or with either output
terminal grounded)

Constant Voltage rms 2.5 mV 2.5 mV 1.5 mV
Constant Voltage p-p 15 mV 25 mV 25 mV
Constant Current**
**With load inductance > 5 H
rms 200 mA 50 mA 30 mA
Readback accuracy
(from front panel or over GPIB with
respect to actual output @ 25 5 C)

Voltage 0.05% + 22.5 mV 45 mV 90 mV
Current 0.1% + 300 mA 165 mA 80 mA
Load regulation
(change in output voltage or current
for any load change within ratings)

Voltage 0.002% + 650 V 1.1 mV 2.2 mV
Current 0.005% + 40 mA 17 mA 9 mA
Line regulation
(change in output voltage or current
for any line change within ratings)
Voltage 0.002% + 650 V 650 V 650 V
Current 0.005% + 40 mA 17 mA 9 mA
Transient response time
(for the output voltage to recover to within 150 mV
following any step change from 100% to 50% or
50% to 100% of the rated output current): < 900 s

Specifications
4
DC floating voltage: Output terminal
can be floated up to 60 VDC from
chassis ground
Remote sensing: Up to half the rated
output voltage can be dropped in
each load lead. The drop in the load
leads subtracts from the voltage
available at the load.
Command processing time: Average
time required for the output voltage
to begin to change following receipt
of digital data is 20 ms for power sup-
plies connected directly to the GPIB.
Modulation: (analog programming of
output voltage and current):
Input signal: 0 to 5 V for voltage,
and 0 to +5 V for current.
Input impedance: 30 k or greater.
AC input (47 to 63 Hz): 180 to 235 VAC
(line-to-line 3 phase) 36 Arms maxi-
mum worst case, 28 Arms nominal;
360 to 440 VACX, 18 Arms maximum
worst case, 14 Arms nominal.
(Maximum line current includes 5%
unbalanced phase voltage condition).
CD ships with: IVI and VXI Plug n
Play drivers, Operating, Programming,
Service and Quick Start Guides.
Input power: 9000 VA and 7950 W
maximum; 175 W at no load.
Size: 425.5 W x 221.5 H x 674.7 mm D
(16.75 x 8.75 x 25.56 in).
Ordering information
Option 208 180 to 235 VAC, 3 phase,
47 to 63 Hz
Option 400 360 to 440 VAC, 3 phase,
47 to 63 Hz
Option 602 Two bus bar spacers for
paralleling power supplies
(p/n 5060-3514)
Option 1CM028A Rackmount flange
kit 88.1 mm H (3U) and 132.6 mm
H (2U) - 4 brackets (5U total)
Option 1CP014A Double rackmount
flange and Handle Kit 88.1 mm
H (2U) and 132.6 mm H (3U)
Accessories
5065-6935 Replacement fuse kit for
360-440 VAC line.
5065-6934 Replacement fuse kit for
180-235 VAC line.
E3663AC Support rails for Agilent
rack cabinets.
5080-2148 Serial link cable 2 m (6.6 ft.)
Specifications
(at 0 to 55 C unless otherwise specified)
6690A 6691A 6692A
Supplemental Characteristics
(Non-warranted characteristics determined by design that are useful in applying this product)
Ripple and noise constant current mode
from 20 Hz to 20 MHz
rms 200 mA 50 mA 30 mA
Average programming resolution
Voltage 4.1 mV 8.1 mV 16 mV
Current 118.5 mA 59 mA 30 mA
OVP 90 mV 170 mV 330 mV
Output voltage programming
response time (excludes
command-processing time)
Full-load programming rise or fall time
(10 to 90% or 90 to 10%, resistive load)
45 ms 60 ms 100 ms
Output common-mode
noise current rms
(to signal-ground peak-to-peak
binding post)

3 mA
20 mA

3.5 mA
20 mA

4 mA
25 mA
Agilent models: 6690A, 6691A, 6692A
+

Side
Output
buss bars
Top
Buss bar detail
Rear
12.7 mm
0.5"
28.6 mm
1.13"
30.2 mm
1.19"
65.1 mm
2.56"
27.8 mm
1.09"
9.5 mm
0.38"
12.7 mm
0.5"
674.7 mm
26.56"
667.3 mm
26.27"
574 mm
22.6"
221.5 mm
8.75"
425.5 mm
16.75"
3/8" Diameter (6)
3 3/4"
Supplemental characteristics
For more information on Agilent
Technologies products, applications or
services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada (877) 894 4414
Brazil (11) 4197 3500
Mexico 01800 5064 800
United States (800) 829 4444
Asia Pacific
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 375 8100
Europe & Middle East
Belgium 32 (0) 2 404 93 40
Denmark 45 70 13 15 15
Finland 358 (0) 10 855 2100
France 0825 010 700*
*0.125 /minute
Germany 49 (0) 7031 464 6333
Ireland 1890 924 204
Israel 972-3-9288-504/544
Italy 39 02 92 60 8484
Netherlands 31 (0) 20 547 2111
Spain 34 (91) 631 3300
Sweden 0200-88 22 55
United Kingdom 44 (0) 131 452 0200
For other unlisted countries:
www.agilent.com/find/contactus
Revised: June 8, 2011
Product specifications and descriptions
in this document subject to change
without notice.
Agilent Technologies, Inc. 2011
Published in USA, December 5, 2011
5988-3149EN
www.agilent.com
www.agilent.com/find/power
Agilent Advantage Services is committed
to your success throughout your equip-
ments lifetime. To keep you competitive,
we continually invest in tools and
processes that speed up calibration and
repair and reduce your cost of ownership.
You can also use Infoline Web Services
to manage equipment and services more
effectively. By sharing our measurement
and service expertise, we help you create
the products that change our world.
www.agilent.com/find/advantageservices
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilents
measurement expertise and product
breadth, combined with channel
partner convenience.

You might also like