Guidance For Industry: Coronary Drug-Eluting Stents - Nonclinical and Clinical Studies
Guidance For Industry: Coronary Drug-Eluting Stents - Nonclinical and Clinical Studies
Guidance For Industry: Coronary Drug-Eluting Stents - Nonclinical and Clinical Studies
DRAFT GUIDANCE
This guidance document is being distributed for comment purposes only.
Comments and suggestions regarding this draft document should be submitted within 120 days
of publication in the Federal Register of the notice announcing the availability of the draft
guidance. Submit comments to Dockets Management Branch (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. All comments should be
identified with the docket number listed in the notice of availability that publishes in the Federal
Register.
For questions regarding this draft document contact (CDRH) Ashley Boam at 240-276-4222 or
(CDER) Devi Kozeli at 301-796-2240.
or
TABLE OF CONTENTS
I. INTRODUCTION............................................................................................................. 1
II. BACKGROUND ............................................................................................................... 2
A. Regulatory Basis ............................................................................................................................ 2
B. Application Requirements............................................................................................................. 3
1. Product Classification ..................................................................................................................... 3
2. IDE Application Requirements ........................................................................................................ 3
3. IND Application Requirements ....................................................................................................... 4
4. PMA Application Requirements...................................................................................................... 4
5. Master Files ..................................................................................................................................... 4
6. Letters of Authorization (LOA) ........................................................................................................ 5
C. Least Burdensome Principles........................................................................................................ 6
III. PRODUCT DEVELOPMENT PATHWAYS FOR DRUG ELUTING STENTS ...... 6
A. The DES Development Pathway — Overview ............................................................................ 6
1. Drug Substance................................................................................................................................ 7
2. Finished DES ................................................................................................................................... 8
B. Factors Influencing Development: Prior Information on Components.................................... 8
1. Stent Platform .................................................................................................................................. 8
2. Delivery System................................................................................................................................ 8
3. Polymer/Carrier............................................................................................................................... 8
4. Drug Substance................................................................................................................................ 9
C. Factors Influencing Development: Local and Systemic Exposure ........................................... 9
IV. SYSTEMIC PHARMACOLOGY, TOXICOLOGY, AND SAFETY DATA FOR
THE DRUG SUBSTANCE ALONE ......................................................................................... 10
A. General Considerations ............................................................................................................... 10
B. Nonclinical Pharmacology and Toxicology................................................................................ 12
C. Clinical Pharmacology and Clinical Tolerance and Safety Information ................................ 13
1. Single IV Dose-Escalation Study ................................................................................................... 15
2. Multiple IV Dose-Escalation Study................................................................................................ 15
3. Mass Balance Study ....................................................................................................................... 15
4. In Vitro and In Vivo Metabolic Studies.......................................................................................... 15
5. Bioanalytical Methods ................................................................................................................... 16
V. CMC INFORMATION .................................................................................................. 16
A. CMC for the Drug Substance Component ................................................................................ 17
1. Physical and Chemical Characterization ...................................................................................... 17
2. Elucidation of Structure................................................................................................................. 17
3. Manufacturer ................................................................................................................................. 17
4. Manufacture and Control .............................................................................................................. 18
5. Specifications ................................................................................................................................. 18
6. Reference Standards ...................................................................................................................... 19
Contains Nonbinding Recommendations
Draft — Not for Implementation
This draft guidance, when finalized, will represent the Food and Drug Administration's (FDA's) current
thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind
FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the
applicable statutes and regulations. If you want to discuss an alternative approach, contact the FDA staff
responsible for implementing this guidance. If you cannot identify the appropriate FDA staff, call the
appropriate number listed on the title page of this guidance.
I. INTRODUCTION
The associated companion document provides additional information that may be useful, including
suggested contents of investigational and premarket approval applications; various examples (e.g.,
example of a DES clinical study summary, a commitment table, test article certification);
information on good animal husbandry, biocompatibility considerations, and issues related to U.S.
and OUS (outside the U.S.) studies; and labeling recommendations. The companion document is
intended to be used together with this guidance.
FDA's guidance documents, including this guidance, do not establish legally enforceable
responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should be
viewed only as recommendations, unless specific regulatory or statutory requirements are cited. The
1
This guidance has been prepared by a working group that included members of the Center for Devices and Radiological
Health (CDRH), Center for Drug Evaluation and Research (CDER), and Office of Combination Products (OCP) in the
Office of the Commissioner at the Food and Drug Administration.
2
For purposes of this guidance, sponsor refers to any person who takes the responsibility for and initiates a clinical
investigation; applicant refers to any person who submits an application, amendment, or supplement to obtain FDA
approval of a new medical product or any other person who owns an approved application. Sponsor is used primarily in
relation to investigational device exemption (IDE) applications and applicant is used primarily in relation to premarket
approval (PMA) submissions.
1
Contains Nonbinding Recommendations
Draft — Not for Implementation
use of the word should in Agency guidances means that something is suggested or recommended,
but not required.
II. BACKGROUND
Coronary stents are implantable devices that are placed percutaneously in one or more coronary
arteries to maintain patency. DESs incorporate a pharmacologically active agent (drug) that is
delivered at the site of stent deployment and is intended to reduce the incidence of restenosis due to
neointimal hyperplasia associated with bare metal stenting. In many cases, the drug is incorporated
into and released from a polymeric coating of sufficient capacity to accommodate the selected dose
and to modulate its delivery at the intended site of action and for the intended duration. The
chemical, physical, and mechanical attributes of the polymer coating system are important for stent
deployment, biocompatibility, and stability. To perform a regulatory assessment of a DES, FDA
would review data from a comprehensive evaluation of individual components (drug, polymer, and
stent), as well as from a comprehensive evaluation of the finished drug-device combination product.
After briefly discussing some general FDA jurisdictional considerations related to this drug-device
combination product, the guidance clarifies a number of issues related to the development of DESs
including the following:
• How to characterize the drug substance, including chemistry, nonclinical systemic and local
tissue pharmacology and toxicology, and how to evaluate the potential for and consequences
of systemic clinical exposure
• How to characterize the drug-device combination product, including the
chemical/physical/mechanical properties of the DES, the nonclinical local vascular and
regional myocardial toxicology, and the clinical performance of the drug-stent combination
• Regulatory considerations that are unique to DES combination products
We encourage sponsors and applicants to consult closely with FDA during development of a DES.
A. Regulatory Basis
DESs are combination products subject to section 503(g) of the Federal Food, Drug, and Cosmetic
Act (the Act) (21 U.S.C. 353(g)), because they are a combination of two different types of regulated
components (a device and a drug) that are physically and/or chemically combined and produced as a
single entity (21 CFR 3.2(e)(1)). A combination product is assigned to an Agency component, such
as the Center for Devices and Radiological Health (CDRH) or the Center for Drug Evaluation and
Research (CDER), for premarket review and regulation based on a determination of the product’s
primary mode of action.
In response to several requests for designation under 21 CFR 3.7, the Agency determined that for
current DESs where the device component maintains coronary artery patency and the drug
component augments the safety and/or effectiveness of the uncoated (bare) stent by preventing
2
Contains Nonbinding Recommendations
Draft — Not for Implementation
restenosis, the device mode of action is the primary mode of action.3 Therefore, the premarket
review and regulatory responsibility for these coronary DESs has been assigned to CDRH with
significant consultation from CDER.
B. Application Requirements
1. Product Classification
Coronary DESs, where the device component provides the primary mode of action, are regulated as
Class III devices that require the submission and approval of a premarket approval (PMA)
application prior to commercial marketing in the United States. To meet the standard for approval,
the PMA application must contain (or include by reference) valid scientific evidence to provide a
reasonable assurance of safety and effectiveness of the DES when used in accordance with its
labeled indication (21 U.S.C. 360c(a)(1)(C), 360c(a)(2)-(3)). Such evidence will usually consist of
nonclinical, animal, and human clinical testing.
FDA has determined that DESs pose a significant risk as defined in 21 CFR 812.3(m), and as such,
are not exempt from the requirement to submit an investigational device exemption (IDE)
application (21 CFR 812.2(b), 812.20(a)(1). When an IDE application is required, a sponsor must
not begin a clinical trial in humans in the United States until FDA has approved the application (21
CFR 812.20(a)(2), 812.42). Sponsors of such studies must comply with the following:
The companion document contains a listing of the elements FDA recommends be included in an
original IDE application.
FDA strongly encourages sponsors to use pre-submission interactions to obtain informal guidance
regarding product development prior to submission of an original IDE application. 5 FDA comments
provided to sponsors during the pre-submission process are informal input, intended to facilitate
open communication between the sponsor and the Agency. Pre-submission interactions for a DES
can be broad-based, or can focus on particular areas, such as engineering testing, CMC testing, or
3
See “Jurisdictional Update: Drug-Eluting Cardiovascular Stents,” http://www.fda.gov/oc/combination/stents.html.
This Jurisdictional Update discusses DESs for which the primary mode of action is the action of the device component in
maintaining vessel patency. However, a DES for which the primary mode of action is attributable to the drug component
would be assigned to CDER.
4
You should review the statutory definition of applicable clinical trial to determine if your trial must be registered
to comply with the law. See PL 110-85, Section 801(a), (adding new 42 U.S.C. 282(j)(1)(A)).
http://frwebgate.access.gpo.gov/cgi-bin/getdoc.cgi?dbname=110_cong_public_laws&docid=f:publ085.110.pdf
Information can be submitted to ClinicalTrials.gov using the Protocol Registration System (PRS). For more information
visit the PRS Information Page (http://prsinfo.clinicaltrials.gov).
5
FDA intends to develop guidance on pre-submissions.
3
Contains Nonbinding Recommendations
Draft — Not for Implementation
clinical protocols. Sponsors should clearly identify questions or particular items they would like to
have addressed as part of the pre-submission interaction. It may be appropriate to meet or hold pre-
submission discussions with Agency staff more than once, at different stages of the development
process.
Preclinical and clinical evaluation of the drug substance alone (e.g., not delivered via a stent) may be
appropriate to fully characterize potential toxicities (see Section IV. below). Human studies of an
investigational drug in the United States must be conducted under an IND application (21 CFR Part
312). The IND application should specify that the eventual intended use of the drug is to be in
combination with a stent. 6
To meet the standard for approval, a PMA application must provide reasonable assurance of the
safety and effectiveness of the finished DES (21 USC 360c(a)(1)(C)). See the companion document
for a list of the elements FDA recommends be included within an original PMA application.
Because of the extensive amount of nonclinical information that is typically needed (especially when
the drug component is a new molecular entity, or NME, that has never been the subject of a new
drug application) coupled with the relatively long primary endpoint timeline for a DES (e.g., 12
months or longer), applicants may wish to consider using the Modular PMA application program. 7
A modular PMA application is a compilation of discrete sections, or modules, submitted at different
times, as each is completed. Together the modules make up a complete application. The potential
advantage associated with the modular approach is that if any deficiencies in a particular section are
noted by FDA, the applicant may be able to resolve them earlier in the review process than would
occur with a traditional PMA application, where a complete application is submitted in a single
submission. 8
5. Master Files
Drug Master Files (DMFs) and Device Master Files (MAFs) permit the submission of proprietary
information to FDA so that parties other than the owners of that information may rely on it. With
the permission of the holder of that master file, a third party applicant may rely on the information in
that master file to support the third party’s application to FDA (e.g., IDE or PMA), even though the
contents of the master file remain proprietary to the holder of the master file (See 21 CFR 314.420,
814.3(d), 814.9(a)). The Agency will not review a DMF or MAF in support of a third party’s
application unless the third party applicant submits in its application a letter of authorization (LOA)
6
See the CDER guidance for industry Content and Format of Investigational New Drug Applications (INDs) for Phase
1 Studies of Drug.
7
See guidance for industry and FDA staff, Premarket Approval Application Modular Review.
8
Ibid.
4
Contains Nonbinding Recommendations
Draft — Not for Implementation
from the holder of the DMF or MAF, which authorizes FDA to refer to the master file in support of
that application. 9
As outlined in Section IV.C of the Guideline for Drug Master Files, each DMF should contain only
one type of information and all supporting data. If the DMF is administratively incomplete or
inadequate, it will be returned to the submitter with a letter of explanation from the Drug Master File
Staff, and it will not be assigned a DMF number. If you intend to submit a DMF that does not
conform to the Guideline for Drug Master Files, we recommend that you contact the appropriate
review division or Drug Master File Staff before making the submission.
We recommend that a sponsor intending to reference (or file) a DMF allow for sufficient time for the
Drug Master File Staff to administratively determine the adequacy of the DMF and assign a DMF
number before an IDE is submitted, given the 30-day review timeframe for IDE applications.
Additionally, sponsors who reference a DMF or MAF as a source of supportive data for an IDE or
PMA should clearly identify the specific volume and page number of the referenced information for
ease of review.
We have not issued guidance on the content of Device Master Files. In general, we will not accept a
submission as a MAF if it is not substantive in nature and does not contain information that may
reasonably be regarded as trade secret or confidential commercial information.
An LOA authorizes FDA, in its review of an application such as an IDE or PMA, to refer to
information contained in another regulatory submission such as an NDA, IND, ANDA, DMF, MAF,
IDE, or PMA. As part of its review of an IDE or PMA for a DES, FDA will review information
from a referenced file only when the IDE or PMA applicant submits an LOA from the holder of that
file, authorizing FDA to refer to the file in support of the IDE or PMA application. The extent of
access granted to the IDE or PMA applicant is typically a business arrangement between the
respective parties. An LOA may give the applicant the authority to rely on all of the information in a
regulatory file, or, if the right to reference is not totally inclusive, on only specific portions of the
file. A copy of the LOA should be included as part of the original IDE and subsequent PMA
applications, with the original LOA submitted to the DMF. (Please refer to Section V.A of the
Guideline for Drug Master Files for specific information to be included within an LOA.)
An LOA may grant FDA either the right to reference or the right to reference and discuss the
information included within one regulatory submission (e.g., NDA, IND, ANDA, DMF, MAF, IDE,
PMA) in support of another regulatory submission (e.g., IDE, PMA).
With a right to reference authorization letter, FDA will not discuss the contents of the referenced
submission with the third party applicant. In the event there are outstanding or unresolved issues
related to FDA’s review of the referenced submission, the Agency will inform the third party
applicant of the general nature of the outstanding issues that must be adequately addressed by the
9
See FDA guidance on Drug Master Files and the Introduction to Master Files for Devices for more information on the
submission of DMFs and MAFs
5
Contains Nonbinding Recommendations
Draft — Not for Implementation
referenced application holder, but will not identify the specific issues. Alternately, if the holder of
the referenced submission chooses not to address outstanding issues, the third party applicant could
potentially generate the requested data independently.
A right to reference and discuss authorization letter allows FDA to review the reference submission
as part of the third party’s application, and permits FDA to discuss information within the referenced
submission with the third party applicant. In the event that there are outstanding issues arising from
FDA’s review of the referenced submission that directly apply to the third party’s IDE or PMA, this
permission to discuss permits the Agency to discuss these issues directly with the IDE or PMA
applicant instead of requiring FDA to discuss specific issues solely with the holder of the referenced
submission.
The issues identified in this guidance document are issues we believe should be addressed before a
coronary DES can be marketed. In developing this guidance, we carefully considered the relevant
statutory criteria for Agency decision making. We believe that we have identified the least
burdensome approach to resolving the issues presented in the guidance. If, however, you believe
that there is a less burdensome way to address an issue, we recommend you follow the procedures
outlined in the guidance for industry A Suggested Approach to Resolving Least Burdensome Issues.
The development of a new DES calls for a thorough exploration of the safety of all of the relevant
components of the product intended for clinical use (e.g., stent, polymer/carrier, and drug), the
composite finished DES, and the delivery system. DES development can present numerous
challenges in that the action of the finished product (such as drug release profile) will affect the
evaluations to be conducted on the individual components, especially the drug substance. However,
testing of the finished product should be limited to in vitro and animal testing until sufficient safety
information is generated to support the introduction of the DES into humans under IDE.
The developmental process typically begins with selection of the drug, polymer or other carrier (if
applicable), and stent platform. The stent platform may be chosen for its previously demonstrated
performance, or it may be a new design developed specifically for use as a DES. In selection of the
polymer or other carrier, considerations will include the following:
6
Contains Nonbinding Recommendations
Draft — Not for Implementation
Whether previously studied or newly developed, the drug substance is intended to limit the growth
of excess neointimal hyperplasia after the injury caused by the stenting procedure without preventing
ultimate re-endothelialization of the stented artery. Selection of the drug dose, both total dose and
dose density, is critical. The amount of drug to be delivered should be carefully evaluated to ensure
that the lowest effective dose is chosen to minimize potential toxicities. Sponsors are encouraged to
consider dose-ranging studies of the DES in animals and possibly in humans to aid in identification
of an optimal dose.
1. Drug Substance
The drug substance should be carefully characterized through evaluation of its chemistry,
mechanism of action, and safety profile. In vitro and animal testing will reveal the types of toxicities
that may result from the drug and the exposure levels at which those toxicities occur. Animal
toxicology testing should establish the No Observed Adverse Effect Level (NOAEL), the highest
exposure at which no adverse effects occur.
Developmental animal studies of the DES are encouraged to provide an understanding of the local
and systemic exposure to the drug substance. Even if the amount of drug available systemically is
below the limit of detection of the assay used, the potential for toxicity may still exist. Therefore,
animal toxicology studies of the drug substance may be important to fully understand the potential
for adverse effects following stent implantation. If implantation of the DES results in significant
systemic exposure, data from human safety studies, specifically, single and multiple IV dose
escalation studies, should be provided (previously conducted or new). If implantation of the DES in
animals does not result in significant systemic exposure, data from human safety studies should not
generally be needed (see Section IV.B. on how to determine when systemic exposure is considered
to be significant).
When needed, these single and multiple IV dose escalation studies, conducted in healthy volunteers,
will provide critical safety information about the drug and its potential toxicities in humans. The
NOAEL determined in the animal studies described above should be used to select the starting dose.
These studies, in addition to metabolic studies, which are intended to describe the distribution,
metabolism, and excretion characteristics of the drug, should be performed prior to initiation of
human clinical studies of the DES under an IDE.
Information regarding the drug substance may be available to the IDE or PMA applicant through the
right to reference a third party’s IND or NDA. However, if the referenced submission does not
relate to intravenous or intra-arterial administration of the drug, as would be delivered by a coronary
DES, FDA may require that additional information related to intravascular safety be included in the
IDE and PMA applications. In some situations, particularly when the right of reference is not
available and a sponsor is relying on information in the public domain, additional studies (e.g., drug
interaction) may help the sponsor adequately support the safety of the drug, polymer, or stent
7
Contains Nonbinding Recommendations
Draft — Not for Implementation
component of a DES. FDA should be consulted on the need for additional studies in this situation
(See also Section IV. below).
2. Finished DES
The finished DES and its delivery system should be fully characterized. Characterization will
include engineering studies, biocompatibility evaluation, animal studies, and development of
complete chemistry, manufacturing and controls (CMC) information, including sterilization,
packaging, and shelf life/stability testing.
Evaluation of the finished DES in humans should include meaningful clinical information related to
stenting outcomes, as well as a systemic pharmacokinetic (PK) study. If significant systemic drug
exposure occurs as a result of DES implantation (see Section IV.B. below), a careful evaluation of
factors that may affect exposure, such as concomitant drugs and comorbidities (such as renal or
hepatic failure), should be carried out.
The clinical study program should include the pivotal trial(s) to support marketing approval,
extended follow-up of the patients in the pivotal trials following the primary endpoint evaluation,
and appropriate postapproval studies.
More specific recommendations regarding each of these development steps can be found in the
following sections of this document.
1. Stent Platform
Stent platforms used in a DES may be chosen based on previously used bare metal stents or may be
developed expressly for use in the DES. If nonclinical testing has been performed on the platform as
a bare metal stent, much of this information may be incorporated by reference. Certain additional
testing on the finished DES, such as coating integrity and particulate matter evaluation, should also
be carried out. Additionally, the sponsor/applicant should consider whether the coating process or
other manufacturing steps will affect the stent integrity or corrosion resistance and repeat appropriate
bench testing (see Section VI.B.) as necessary.
2. Delivery System
Delivery system testing should be carried out as described in section VI.B. below. Evaluation of
aspects such as delivery and handling characteristics, when previously studied in conjunction with a
bare metal or other previously approved stent, can be incorporated by reference; however, delivery
system testing that incorporates the drug-eluting stent (e.g., deployment, balloon burst) should be
conducted using the intended DES and delivery system combination.
3. Polymer/Carrier
8
Contains Nonbinding Recommendations
Draft — Not for Implementation
As described in section V below, a full physicochemical description of any polymers used as drug
carriers should be provided either in the original application or by reference to DMFs, MAFs, or
other sources. Any change in the properties of the polymer due to the incorporation of the drug
substance within the polymer or the application of the polymer to the stent should be evaluated.
4. Drug Substance
An understanding of the systemic pharmacology and toxicology of the drug substance 10 and its
metabolism in the body is essential to guide the design of the clinical studies of the DES with respect
to monitoring for adverse events. Given this aim, testing should be performed prior to initiation of
an IDE for the DES.
The amount of new evidence needed to support the safety and effectiveness of a DES will be
determined by the amount of existing information about each of the components and, particularly,
the drug substance. For a DES using a studied drug, that is, a molecular entity that has been
previously approved or studied under IND (i.e., has an approved NDA or ANDA, or has undergone
human clinical studies under an active IND), the information on systemic use described below may
be available for the DES manufacturer to incorporate by reference. An unstudied drug that is a
molecular entity that has not been approved for use in humans or that does not have study
information available should undergo testing as described in Section IV below to develop this
information before human testing of the DES.
For any DES, the primary exposure to the drug substance will occur at the coronary artery wall
directly apposed to the stent and downstream in the stented vessel and myocardium. Exposure in the
rest of the body will be much lower. At first glance, this could suggest that evaluation of the
systemic toxicity of the drug substance alone should not be necessary and that the animal and
clinical testing of the finished DES should be sufficient to demonstrate preliminary safety of the
DES. However, several factors challenge this conclusion.
First, although the total dose of drug on a DES is almost always much lower than that given in a
systemic administration (e.g., orally or by injection), the exposure at the artery wall may be many
times higher than the blood levels achieved after an oral or injected dose. Therefore, the potential
toxicity at the coronary wall at the DES implantation site and within the coronary vascular bed and
myocardium distal to the DES implantation site should be studied. Animal studies of the finished
DES will be critical to this understanding, but as is typical of animal toxicology studies, it is also
important to assess the potential toxicity of exposure to higher doses than in the finished DES.
Animal studies of local doses well above those expected from a DES to examine the safety margin
over the doses that will be used in human DES implants should be completed.
Second, it has been our experience that in certain situations (i.e., multiple stents, major active
metabolites), systemic drug exposure from a stent, or stents, can cause systemic toxicities.
10
For the purpose of this guidance, drug substance is considered the active pharmacological agent.
9
Contains Nonbinding Recommendations
Draft — Not for Implementation
Therefore, it is crucial to have information gathered under acute and chronic conditions on the
systemic safety and toxicity profiles of the drug to be used in a DES system prior to initiating
clinical studies.
Furthermore, there is a greater need for information about the safety of the drug component prior to
beginning clinical studies of a DES because of the permanence of the DES. In addition, the planned
DES clinical trials may not explore the full range of clinical use likely to occur after marketing
approval, and there is a need to consider whether this more extensive use of permanent implants may
place patients at risk. As a result, an appropriate understanding should be gained of the safety of the
drug component prior to clinical studies with a DES.
In summary, a manufacturer of a new DES should establish preliminary evidence of the safety of the
DES prior to beginning human clinical trials (under an IDE, or under an IND if intravenous clinical
study of the drug substance alone is needed). A complete assessment of safety and effectiveness of
the DES should be submitted in the PMA application. Recommended testing to address issues
related to systemic pharmacology, toxicology, and safety of the drug substance follows. FDA
remains open to alternative methods to obtain this information as well to other considerations, such
as when the drug incorporated in the DES has known toxicities that may require modifications to the
recommendations below.
FDA believes that systemic pharmacology, toxicology, and safety data on a drug substance to be
incorporated in a stent are needed to fully understand the safety profile of the finished DES.
Nonclinical, and often clinical, studies should be performed as part of the effort to demonstrate the
safety of a DES.
A. General Considerations
A first step in characterizing a drug involves performing systemic nonclinical pharmacology and
toxicology studies of the drug substance using in vitro (cell culture) or in vivo (animal) models.
These nonclinical studies help provide an understanding of the metabolism of the drug, its
distribution and accumulation (e.g., in the regional myocardium or other important organs), and
whether the effects of the drug might be significantly affected by the presence of certain enzymes.
Animal testing will also help assess potential toxicities that cannot be identified during clinical trials
and will define the No Observed Adverse Event Level (NOAEL), which is used to determine the
starting dose for human safety studies (see Section IV.B.). In some cases, animal testing may
establish that an adequate factor of safety exists between the levels of drug exposure likely to be
reached in humans and the levels of exposure at which toxicities are seen in animal studies. In some
situations, when a sufficient safety margin exists, this testing may support the conclusion that human
intravenous safety studies would not be necessary to ensure safety of clinical systemic exposure. In
addition to determining the severity of the observed toxicities in animals and a careful definition of
the local, regional, and systemic adverse effects in animals, it is important to define the slope of the
10
Contains Nonbinding Recommendations
Draft — Not for Implementation
relationship between toxicity and exposure over a broad range of doses, extending to levels in excess
of the dose anticipated for use in humans.
• Determining when human safety studies are needed – PK parameters and the NOAEL
When deciding whether human intravenous safety studies also will be needed, one should first
consider what pharmacokinetic parameter—Cmax (maximum concentration) or AUC (area under
the curve describing concentration versus time) over some specific time—should be the basis of the
safety factor. If the parameter that best predicts toxicity is AUC (which is most likely the case), it is
important to base any comparisons on AUCs integrated over the same or nearly the same time
courses.
A second important consideration is identifying the preclinical toxicity that establishes the NOAEL.
Usually, this is based on testing in the most sensitive species and on the adverse effect seen at the
lowest dose.
When considering the relevance of a preclinical model for intravenous administration, the exposure
should, ideally, resemble the exposure from a DES. Release of drug from a DES can generally be
expected to follow two-phase kinetics—a first-order (or relatively fast) process with a time constant
on the order of hours and a zero-order (or very long time constant) process. the preclinical
intravenous exposure intended to match this would include infusion over several hours (first-order
phase) followed by a lower prolonged or repeated infusion (if the half-life in plasma is much less
than the release rate from a DES). 11 We recognize, however, that mimicking the time course of
release from the stent can greatly complicate the animal study. Furthermore, matching the DES
release should not be necessary when toxicity is likely to be mostly related to Cmax and the AUC
over the first several hours, and the safety margin related to this period is of greatest concern. In such
cases, preclinical assessment following a single bolus administration should be acceptable.
In such cases, preclinical assessment following a single bolus administration should be acceptable.
Another consideration for the relevance of a preclinical model is the possibility of species-specific
metabolism. If a metabolite is prominent in humans, but not in the animal, the resulting NOAEL
may not be pertinent to human exposure. If a sufficiently sensitive assay is available, it may be
appropriate to do a microdose study in humans 12 to confirm similar metabolism.
If the parameter that best predicts toxicity is AUC, it is important to base any comparisons on AUCs
integrated over the same or nearly the same time courses. Empirically, we recommend a comparison
based on AUC0-24h.
• Determining when human safety studies are needed – calculating the safety factor
Because multiple stents are commonly used in humans, the exposure parameter (generally,
11
The DES should initially be studied in an animal model to inform the design of the animal IV toxicology study.
12
See the CDER guidance for industry Content and Format of Investigational New Drug Applications (INDs) for Phase
1 Studies of Drug.
11
Contains Nonbinding Recommendations
Draft — Not for Implementation
AUC0-24h) measured from implantation of the DES in the animal model should be adjusted to reflect
the use of 120 mm of stented length as a likely maximum length to be encountered in common
clinical use. In a vast majority of cases, if the safety factor (ratio of the NOAEL AUC0-24h level in
the animal to the corresponding exposure AUC0-24h in humans) is a factor of 100 or more, DES
clinical studies can be initiated without a prior intravenous administration human safety study. This
conclusion is based on the observation that >100 fold increase in sensitivity to toxic effects in
humans versus animals is extremely unusual for drugs. See the following example.
The NOAEL for the most sensitive relevant toxicity (in the monkey) occurs at a
dose that produces AUC0-24h = 4500 ng-h/mL. If a single 40 mm DES in the mini-
pig produces AUC0-24h = 3 ng-h/mL; 120 mm of stent would be expected to yield
an AUC0-24h of 9 ng-h/ml, still just 1/500 of the NOAEL. Absent other factors, it
may be reasonable to conclude that no intravenous study in humans would be
necessary before the first DES implantation in humans.
For a previously studied drug, much of the information discussed below may be available for
incorporation in an IDE or PMA application through a right to reference or other means. However,
in some cases, gaps in the preexisting safety data may be identified. For example, for a drug that has
been developed for oral administration, additional nonclinical testing pertaining to the intravenous
route (e.g., hypersensitivity, hemocompatibility) may not have been performed and should be
conducted.
Where reference rights are unavailable, a sponsor may be able to use information in the public
domain (e.g., published literature) in support of an application. When a DES relies for approval on
data in a previously approved application for the drug substance to which the sponsor has an LOA,
or on literature in the public domain, the sponsor or applicant should demonstrate that the active
ingredient of the DES is the same as the active ingredient in the reference drug.
For an unstudied drug that has never been studied in humans, preclinical safety testing and
pharmacology studies should be conducted to fully characterize the drug-related effects, metabolites,
and toxicities of the drug administered intravenously (IV). Studies should be designed to describe
desired as well as off-target pharmacology and also potential drug toxicities; data from these studies
should be used to select safe starting doses for clinical trials. 13
The timing and types of studies that should be performed are described in International Conference
on Harmonisation (ICH) M3, Timing of Pre-clinical Studies in Relation to Clinical Trials.
Toxicology studies in two species, including one non-rodent species, should be designed to describe
13
See also Guidance for Industry Estimating the Maximum Safe Starting Dose in Initial Clinical Trials for Therapeutics
in Adult Healthy Volunteers at http://www.fda.gov/cder/guidance/5541fnl.htm.
12
Contains Nonbinding Recommendations
Draft — Not for Implementation
a maximum tolerated dose (MTD) and determine the NOAEL. The duration of these studies should,
at a minimum, span the length of time the DES is estimated to release drug in vivo. The minimum
duration should be two weeks for a DES without a polymer or other drug carrier, which could be
considered as a single IV dose drug study. The NOAEL from the IV studies should provide
significant safety multiples over the clinical systemic exposure from multiple DES implants.
Other recommended toxicology studies are designed to assess potential toxicities that may not be
monitorable in clinical studies. For example, tests for potential genetic toxicity (ICH S2A and S2B),
tests for reproductive toxicity (ICH S5), and safety pharmacology studies (ICH S7A and S7B).
Tests for the assessment of potential carcinogenicity are also described in the ICH guidances (S1A
and S1B). However, if drug exposure to the local tissue is shown to last less than six months,
carcinogenicity studies will generally not be required. Note that finished product biocompatibility
testing does not obviate the need for safety and pharmacology testing of the drug substance alone.
The decision tree provided in this section describes the clinical pharmacology (CP) studies that
should be considered for the assessment of the drug substance during the development of a DES.
The key focus of the tree is the initial determination about whether the drug is an unstudied drug,
about which little is known, or a previously studied drug, about which there already is a thorough
understanding and adequate information with an appropriate safety profile is referenced in the
application.
Human safety studies of the drug alone in healthy volunteers can provide critical information
regarding the tolerability, safety, and pharmacokinetics of a drug substance. Whether such studies
are needed will depend on the systemic exposure that will arise from the stent and how this
compares with the exposure seen in animal studies, specifically the NOAEL, of the most sensitive
species.
In general, for drugs that are well understood no additional clinical pharmacology studies are
warranted since all the factors that affect a drug’s safety and efficacy from a systemic point of view
will already have been well characterized. If a drug has been previously studied and the resulting
information is available, these studies need not be repeated. However, if the DES will incorporate a
total amount of drug higher than that used in previous studies of the drug alone or result in higher
sustained levels, additional information would be necessary to address the safety of the higher dose.
For an unstudied drug, the need for studies to elucidate the distribution, metabolism, and excretion of
the drug, and any intrinsic or extrinsic factors that could affect exposure should be carefully
assessed. Some of the metabolic information can be based on in vitro methods, notably the role of
CYP450 enzymes in metabolism; some can be obtained from studies on the DES. As already
mentioned, in some cases, human studies involving micro-doses may facilitate the assessment of the
drug’s pharmacokinetics.
13
Contains Nonbinding Recommendations
Draft — Not for Implementation
CLINICAL STUDIES
- Single IV dose tolerance study
- In vitro metabolic studies
Safety Safety
Concerns Concerns
YES NO
Significant systemic exposure may not have been observed in animal studies of the DES, in part
because the number of stents that can be implanted in an animal is limited. The potential for
multiple stent use in routine clinical practice should be considered when determining whether a
single IV dose escalation human study is needed to understand the systemic levels at which toxicities
are first observed. Absent other factors that increase concern, a separation between the NOAEL
established in the most sensitive animal species and the systemic exposure that could be reached of
two orders of magnitude could mitigate the need for human studies of systemic drug safety.
If human PK data (using the DES) are available from previously conducted studies outside the
United States, these data may provide a direct measure of systemic exposure (instead of the indirect
measure based on animal data on the DES) and further determine whether such a substantial
separation from toxicity causing concentrations exists. On the other hand, for DES where
appreciable systemic drug concentrations can reasonably be expected and for drugs with animal or
human toxicities that occur at only slightly above the anticipated human exposures, the full range of
studies to evaluate the consequences of systemic exposure to the drug would be warranted. Animal
14
Contains Nonbinding Recommendations
Draft — Not for Implementation
toxicology studies will then also serve to determine what is considered to constitute an initial safe
dose for human systemic drug safety studies.
The usual next steps in developing a DES that incorporates an unstudied drug would involve single
and multiple ascending dose studies. If the systemic exposure to the drug from a DES (or from
multiple DESs) is sufficiently low (i.e., a reasonable safety factor exists between the NOAEL and
the expected systemic exposure in man based on animal studies of the DES), such studies would
probably not be informative. 14 However, it should be noted that an adequate assessment of systemic
exposure from the DES in an animal model can only be made if the release characteristics of the
drug are well-characterized and have been shown to have minimal variation from stent to stent.
For unstudied drugs, testing to elucidate the distribution, metabolism and excretion characteristics of
the drug are essential in understanding the safety and efficacy profile of this new entity.
If a single IV dose-escalation study is indicated, the selected initial dose should be based on the
NOAEL information from the animal nonclinical studies. The drug should be given via intravenous
administration (if feasible). This study should be designed to collect information on the drug
substance’s tolerance, safety, and pharmacokinetics following administration of single doses and
escalating up to the maximum tolerated dose. The exposure should be engineered to resemble that
produced by the DES.
If the time course for release from a DES is long, data from a multiple IV dose- or from a continuous
infusion dose-escalation study to mimic the stent exposure should be provided.
We suggest that a mass-balance study be performed to define and assess the systemic exposure, the
disposition and pathways of elimination (including metabolism and excretion), and pharmacokinetic
measures or parameters of the drug substance administered intravenously.
The mass balance study should be based on the drug substance tagged with a radioactive label (i.e.,
14
C, 3H) to allow for sensitive monitoring of the distribution patterns of the tested drug after its
intravenous administration. Blood (plasma or serum as appropriate), urine, and fecal samples should
be collected and assayed for radioactive label. Other routes of elimination should be monitored as
appropriate. Both the parent drug substance and any metabolites present should be identified.
14
We note that single and multiple ascending dose studies are small and quite well monitored, and the insight into
human toxicity can be quite valuable.
15
Contains Nonbinding Recommendations
Draft — Not for Implementation
Since an integral part in understanding the safety of an unstudied drug is determining its metabolic
pathway and whether there is formation of any active/toxic metabolites, the Agency recommends
that a drug’s metabolism and metabolic pathway, as well as the activity of major metabolites, be
assessed relatively early in development of the DES.
In vitro metabolic studies designed to assess the P450 metabolizing enzymes of the drug as well as
to characterize the P450 isoenzymes that are inhibited or induced by the drug should be conducted so
that the clinical implications of interactions can be assessed later in the DES clinical studies.
In vitro metabolic studies can frequently serve as an adequate screening mechanism to assess the
contribution of cytochrome P450 on the metabolism of the drug, so that subsequent in vivo testing
will be unnecessary. In contrast, when positive findings of active or toxic metabolites arise in in
vitro metabolic studies, we recommend that drug interaction information be obtained from the
clinical trials using a drug interaction-population PK approach.
Information on the design and data analysis of the metabolic studies can be found in guidances In
Vivo Drug Metabolism/Drug Interaction Studies and Drug Metabolism/Drug Interaction Studies in
the Drug Development Process: Studies In Vitro.
5. Bioanalytical Methods
Validated bioanalytical methods should be used when evaluating the concentrations of the drug and
its metabolites in the clinical pharmacology and metabolic studies. Information on the validation of
assays can be found in the guidance Bioanalytical Method Validation.
V. CMC INFORMATION
This section provides guidance on the information to be submitted regarding the chemistry,
manufacturing, and controls (CMC) aspects of (1) the drug substance and (2) the finished product,
followed by the information needed for (3) the engineering evaluation. The information can be
provided in the submission, or incorporated by reference to another regulatory submission (e.g.,
DMF, NDA, ANDA, PMA, MAF) with copies of the LOA provided in the relevant section of the
IDE or PMA application. All of the topics described for the drug substance and finished product
should be included for both IDE and PMA submissions.
Because the product described in an initial IDE application will be permanently implanted into
patients with potentially life-threatening coronary artery disease, the CMC section should address all
of the items that would be provided in a PMA application. However, the level of detail and the
degree of documentation will differ in that the information for the IDE will focus more on patient
safety and product development and less on product and process controls.
In general, the information for the drug substance component is expected to be similar for both IDE
and PMA submissions. However, it is recognized that the finished product is still under
development at the time of the initial IDE submission. Consequently, clinical trials may be allowed
to proceed even though manufacturing processes are not fully optimized, analytical methods
16
Contains Nonbinding Recommendations
Draft — Not for Implementation
validation is incomplete, and the acceptance criteria for the finished product tests are still tentative,
provided all parameters that relate to safety are well characterized. The sponsor/applicant is strongly
encouraged to meet with the Agency before the initial IDE submission, during development and
before submitting a PMA application to discuss critical drug-related issues and the information
needed at various stages of development.
The following items should be included for the drug substance in both the IDE and PMA
submissions. When submitting an IND (e.g., when the drug substance is an unstudied drug and
human safety studies will be conducted in the United States), guidance on Phase 1 (CMC section)
should be carefully consulted. 16
The chemical structure of the drug substance (including stereochemistry), molecular formula, and
molecular weight should be provided. All appropriate names or designations for the drug substance
should be listed (e.g., USAN, Chemical Abstracts, IUPAC, code number). The physicochemical
properties of the drug substance should be described and should include, but not be limited to,
information on the following, as appropriate:
2. Elucidation of Structure
The chemical structure of the drug substance should be confirmed using physical and chemical
techniques, such as elemental analysis, mass spectrometry (MS), nuclear magnetic resonance (NMR)
spectroscopy, ultraviolet (UV) spectroscopy, infrared (IR) spectroscopy, X-ray crystallography, and
other tests (e.g., functional group analysis, derivatization, complex formation).
3. Manufacturer
15
See the CDER guidance Submitting Supporting Documentation in Drug Applications for the Manufacture of Drug
Substances. Another drug substance guidance is forthcoming that, once finalized, will supersede this guidance.
16
See the CDER guidance Content and Format of Investigational New Drug Applications (INDs) for Phase 1 Studies of
Drug.
17
Contains Nonbinding Recommendations
Draft — Not for Implementation
The name, address, and manufacturing responsibility should be provided for each facility (including
contract manufacturers and testing laboratories) that will be involved in the manufacturing or testing
of the drug substance. The addresses should be those of the locations where the relevant
manufacturing or testing operation will be performed. Registration numbers (i.e., CFN, FEI
numbers) should be provided to facilitate CGMP inspections.
The description of the manufacturing process should include a flow diagram and a narrative of the
processes and process controls that will be used to manufacture the drug substance. The flow
diagram should include each manufacturing step with chemical structure, solvents, reagents,
auxiliary materials, critical operating parameters, and expected yield. A narrative description of the
sequence of manufacturing steps and the scale of production should be provided in more detail than
that given in the flow diagram.
Process controls used to monitor and adjust the manufacturing process should be provided and
include in-process tests and acceptance criteria. These controls should ensure that intermediates and
drug substance will conform to their established specifications.
Specifications, certificates of analysis, and quality or grade of the starting materials, reagents,
solvents, and auxiliary materials that will be used to manufacture the drug substance (including
deriving it from a biological source) should be provided. When appropriate, specific tests and
acceptance criteria to control microbial contamination in materials derived from biological sources
should be included in the specifications.
5. Specifications
Specifications are established to control the quality of the drug substance and should focus on those
characteristics necessary to ensure the safety and efficacy of the finished product. The specifications
should include all tests, analytical procedures, and associated acceptance criteria to which each batch
of a drug substance will conform over its retest period/shelf-life. 17 Acceptance criteria are numerical
limits, ranges, or other measures for the tests described. We recommend that the information be
presented in tabular form.
Analytical procedures, including validation information, for each of the tests proposed in the
specification should be described in detail. If the analytical procedure is in the current version of the
United States Pharmacopeia (USP) or other FDA-recognized standard reference (e.g., AOAC
International Book of Methods), details need not be provided. Analytical procedures should be
validated to demonstrate that the methods are suitable for their intended use. Validation should
include experimental data (e.g., representative chromatograms with peak identification). 18
17
See ICH Guidance Q6A Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and Drug
Products: Chemical Substances.
18
See ICH Guidances Q2A Text on Validation of Analytical Procedures and Q2B Validation of Analytical Procedures:
Methodology.
18
Contains Nonbinding Recommendations
Draft — Not for Implementation
6. Reference Standards
Information on the reference standards or reference materials used for testing the drug substance
should be provided. A reference standard obtained from an official source should be identified. A
reference standard not from an official source should be appropriately characterized. A list of any
available reference standards for impurities should be included.
7. Container/Closure System
A description of the container closure system for the drug substance should be provided, including
the identity of materials of construction for each primary packaging component and specifications.
8. Stability
Stability data should be generated in accordance with ICH guidances. 19 The studies conducted,
protocols used, and the results of the studies should be summarized. The discussion should include
(1) a summary of stability batches tested, storage conditions used, attributes tested, acceptance
criteria, test schedule, and analysis of all available data (including a summary of the statistical
analysis if performed) and (2) conclusions regarding the storage conditions and retest or expiration
dating period, as appropriate. Data regarding stability under stressed (e.g., pH extremes, oxidation,
heat, light) conditions should also be provided. We recommend that the results of stability studies be
presented in tabular form.
For the purpose of this section, the phrase finished product refers to a packaged and sterilized DES
that contains all the materials (e.g., drug and polymer coating materials) applied to or incorporated
within a bare metallic stent substrate and the stent delivery system. The following sections discuss
the information on the finished product that should be submitted in support of an IDE or PMA
19
See ICH guidance Q1A(R2) Stability Testing of New Drug Substances and Products.
19
Contains Nonbinding Recommendations
Draft — Not for Implementation
A detailed description of the finished DES should be provided and should include the proprietary
name, model numbers, stent sizes, product code, and intended use. Detailed engineering drawings
should also be provided. In addition to a detailed written description, a cross-sectional schematic of
the stent platform, coating layers (e.g., primer layer, polymer/drug layer, drug-free polymer topcoat)
and stent delivery system should also be included that pictorially depicts the coating and drug
distribution across the stent geometry (e.g., length, circumference, strut sides, adluminal, abluminal).
The schematic should also include a description of the drug release mechanism. The total drug
content (µg/stent) and drug dose density (µg/mm2) should also be provided for each stent size.
2. Product Development
This section should contain information on the development studies conducted to establish that the
components of the finished DES, the formulation, manufacturing process and controls, and
packaging system are appropriate for the purpose specified in the application. The studies included
in this section can be distinguished from controls used for routine batch release. Additionally, this
section should identify and describe the formulation and process attributes, including critical
parameters that can influence batch reproducibility, product performance, and quality. Development
reports allow the Agency to understand critical variables and focus attention on high-risk aspects of
a product and process.
a. Components of the Finished DES Product
• Drug Substance
20
See the CDER guidance for industry Submitting Documentation for the Manufacturing of and Controls for Drug
Products (1987). Another drug product guidance is forthcoming that will supersede the 1987 guidance.
20
Contains Nonbinding Recommendations
Draft — Not for Implementation
• Excipients
The choice of excipients (e.g. polymer carriers), their concentrations, and the
characteristics that can influence the finished product performance or
manufacturability should be discussed. The applicant should demonstrate an
understanding of the effects of excipient variability on the critical quality attributes of
the finished product. Since organic solvents are usually employed to dissolve both
the drug substance and polymer carrier to form a coating solution, the rationale for
choice of solvent should be provided. The ability of functional excipients (e.g.
antioxidants) to perform throughout the intended shelf life of the DES should also be
discussed.
The design of and the rationale for the selection of the key elements of the stent
substrate 21 (e.g., materials, surface characteristics and area, cell structure, engineering
performance), which can influence the performance and manufacturability of the
finished DES, should be discussed. The applicant should also describe the
components and design elements of the stent delivery systems used for stent
deployment in the coronary vasculature.
b. Formulation Development
Since a DES is formulated to provide extended release of the drug substance, a description of
the drug release mechanism (e.g. erodible polymer matrix, diffusion) should be provided.
The development of target release rates of the drug from the polymer matrix should be
discussed. The applicant should provide a scientific rationale for the selection of the final
formulation by evaluating appropriate models for drug release. The applicant should show
how the formulation and product construction were chosen, incorporating the principles of
modern pharmaceutical development practices, Quality System regulations, and/or Design
Control requirements as appropriate. 22, 23 , 24
The selection of the manufacturing process with emphasis on understanding its critical
aspects should be described. Manufacturing process development generally starts with the
identification of critical quality attributes of the finished product, which are necessary for its
desired performance. Manufacturing process options in conjunction with appropriate control
21
See Guidance for Industry and FDA staff on Non-Clinical Tests and Recommended Labeling for Intravascular Stents
and Associated Delivery Systems.
22
See also the CDER guidance for industry PAT — A Framework for Innovative Pharmaceutical Development,
Manufacturing, and Quality Assurance.
23
See ICH Guidance Q8 Pharmaceutical Development.
24
See 21 CFR 820.30 for more detailed Design Control requirements.
21
Contains Nonbinding Recommendations
Draft — Not for Implementation
strategies that can reliably result in finished product with critical quality attributes within
acceptable ranges should be considered. Critical process parameters that should be
controlled or monitored to ensure batch-to-batch reproducibility and to minimize intra-batch
variability should also be discussed. This approach demonstrates knowledge and
understanding of the product and associated processes, which in turn provides greater
assurance of product quality. The benefits of having an efficient and reliable process, with
reduced reliance on end-product testing, include enhanced manufacturing efficiency and a
reduced risk of producing a poor quality product. These concepts, when implemented, would
be a significant advantage to stent manufacturers who typically produce small batch sizes.
Operations using process analytical technologies (PAT) 25 that measure an endpoint
indicating the manufacturing process (e.g., coating) is under control are preferable to a
measurement of a quality attribute on representative samples. Generally, this allows for
adjustments to process parameters to mitigate anticipated variation in raw materials,
equipment, environment, or other conditions.
The applicant should describe how the packaging system was selected and designed to
provide protection and maintain sterility throughout the shelf life of the finished product.
The suitability of the packaging system should be demonstrated with respect to protection
from moisture, oxidation, and light, and compatibility of materials with all components of the
finished product.
The morphology of the solid drug-polymer carrier system in the finished product should be
described (i.e., dispersed drug phase, continuous separate drug phase, reservoirs). Micrographs of
the surface and full thickness cross-section of the coating should be provided. The micrographs will
aid in gaining an understanding of the drug release process, which may have implications for coating
durability and particulate matter formation.
A detailed description of the physical and chemical tests performed to characterize the finished
product should be provided. The physical, chemical, and mechanical characteristics of a DES are
critical to ensure finished product quality and performance. Physical and chemical characterization
of a DES should include tests for surface coat composition, coating/carrier thickness and uniformity,
and coating/carrier erodability as applicable. These tests are useful for characterization and may be
provided as one-time tests—not to be confused with routine control and release testing.
Note: These tests are a subset of testing recommendations provided in Section VI.C of this guidance
for the mechanical/engineering performance tests for the finished DES.
25
See 21 CFR 820.30 for more detailed Design Control requirements.
22
Contains Nonbinding Recommendations
Draft — Not for Implementation
A qualitative and quantitative list of drug substance(s) and excipients making up the finished product
should be provided. We recommend including a detailed components and composition table per unit
and per batch for each stent configuration to be marketed. Ingredients used in the manufacture of the
finished product, regardless of whether or not they appear in the finished product, such as solvents,
should be identified. Ingredients of human or animal origin should also be identified and their use
supported with appropriate safety information.
a. Component Function
The function (i.e., role) of each ingredient in the formulation should be described.
Ingredients that are used in the manufacture but are not intended to be part of the finished
product (e.g. solvents) should be identified as processing agents.
b. Component Controls
The applicant should identify all component tests that the finished product manufacturer will
routinely perform as well as test results that will be accepted from the excipient and drug
substance manufacturer (Certificate of Analysis, COA). At a minimum, the finished product
manufacturer must perform an appropriate component identification test (21 CFR
211.84(d)(2)).
(ii) Excipients
23
Contains Nonbinding Recommendations
Draft — Not for Implementation
Since most DESs use a polymer matrix as a carrier or barrier for the drug release,
special attention should be paid to this component. In addition to the items listed
above, the following information should also be included for the polymer:
Many of these items should be tested on a routine basis as part of the polymer
specifications and adequate justification should be provided for any exclusions.
The following detailed information for each component used in the fabrication of the
stent substrate and its delivery catheter system should be provided:
24
Contains Nonbinding Recommendations
Draft — Not for Implementation
5. Manufacturer
The name, address, and manufacturing responsibility should be provided for each facility (including
contract manufacturers and testing laboratories) that will be involved in the manufacturing or testing
of the finished product. 26 Addresses should be provided for the locations where the relevant
manufacturing or testing operation will be performed. Registration numbers (i.e., CFN, FEI
numbers) should be provided to facilitate GMP inspections. This information may be submitted in
the Manufacturing -- Quality System (QS) Regulation and Current Good Manufacturing Practice
(CGMP) Regulations section (see Section VII.A. below) and incorporated by reference or
reproduced here for ease of review.
A complete description of the manufacturing process and controls (or a reference to this information)
should be provided within this section of an application to provide a thorough understanding of the
critical attributes that should be assessed at final product release and to assess the potential impact of
changes made in the manufacturing procedures used during the course of product development. A
discussion of any differences between the manufacturing process to be used for the marketed product
and any used to produce batches for clinical efficacy and/or primary stability studies should be
addressed in the PMA application. This should include an evaluation of how the differences will not
adversely affect the performance of the product. (See also Section VII.A below.)
a. Flow Diagram
A flow diagram (or series of flow diagrams) should be provided that includes all the steps in
the manufacturing process for the finished DES. The diagram should include the following:
• Steps where materials enter the process (e.g., catheters, stents, polymers)
• Critical processing steps that may have an influence on the chemical or physical
properties of the stent, polymer, or drug (e.g., application of coating, including any
primers or coupling agents, use of oxygen scavengers or antioxidants, crimping of stent
onto catheter, heat sets, use of sheath protectors)
• In-process testing (identify method) and the manufacturing step where it is performed
• Sterilization (identify method) and packaging steps
• Any end-process (reliability) testing conducted prior to product release
• Differentiation of manual versus automated processes
• Depiction of differences in manufacturing processes for the catheters (e.g., Over-The-
Wire versus Rapid eXchange)
26
A statement should be provided that ruminant-derived materials from bovine spongiform encephalopathy (BSE)
countries as defined by the U.S. Department of Agriculture (9 CFR 94.11) are not used or manipulated in the same
facility.
25
Contains Nonbinding Recommendations
Draft — Not for Implementation
c. Process Controls
Controls used to monitor the manufacturing process should be described, including operating
parameters, environmental controls, and process/in-process tests. A description of critical
process controls (as justified in section V.B.2.c. Manufacturing Process Development)
should include tests, analytical procedures, limits (ranges), or other acceptance criteria.
In some cases, results from in-process controls can be used in lieu of finished product testing.
This approach, however, should be supported with data that demonstrate a clear relationship
between in-process testing and the critical quality attributes of the finished product.
d. Sterilization Process
The sponsor should clearly identify the method of sterilization (e.g., ethylene oxide, E-beam
radiation, gamma) along with the specific parameters (e.g., concentrations, humidity, time,
and temperatures) and an assessment of its effect on the finished product. The assessment
should address the effects on such elements as coating integrity, drug substance, and polymer
carrier stability.
See Section VI.C for engineering test methods to evaluate the effect of sterilization on the
coating characteristics.
7. Packaging System
A description and the following information on each component of the primary packaging system for
the finished product should be provided:
• Supplier/manufacturer
• Composition
• Quality/grade of materials
• Schematic drawing including dimensions, tolerances, etc.
• Specifications
26
Contains Nonbinding Recommendations
Draft — Not for Implementation
The same type of information should be provided for functional secondary packaging components as
well. For nonfunctional secondary packaging components (e.g., those that do not provide additional
protection), only a brief description is necessary.
Regulatory specifications should be provided for the finished product; these specifications apply to
every batch at release and throughout shelf-life. A specification consists of a list of tests, references
to analytical procedures, and appropriate acceptance criteria that are numerical limits, ranges, or
other criteria for the tests described. An example of a regulatory specification table is provided in
Appendix A. Finished product specifications should focus on those characteristics found to be
useful in ensuring product quality as it relates to safety and efficacy. Testing should be performed on
every batch of the finished product after packaging and sterilization. All testing should be
performed on expanded stents, unless otherwise justified. To ensure that the regulatory specifications
are met throughout the shelf life, tighter acceptance criteria may be established for product release.
When product knowledge and process understanding have been demonstrated in the application, and
relevant in-process control strategies are being implemented routinely, it may be possible to use in-
process tests in lieu of traditional off-line end-product testing. In addition, PAT, if applied, can serve
as a basis for real-time release of the finished product to demonstrate that each batch conforms to
established regulatory attributes. It should be emphasized that any alternate proposals to end-
product testing should be discussed with the Agency during development and regulatory approval
obtained before implementation.
The analytical procedures and their validation 27 should be described in detail for each test listed in
the specifications. Acceptance criteria should be primarily based on consideration of safety,
efficacy, manufacturability, and stability. The justification for the acceptance criteria can be based
upon batch analysis data for all relevant batches (e.g., nonclinical, clinical, and primary stability
batches). Ideally, the data should be representative of batches of finished product manufactured
using different lots of drug substance, polymer, and coating solution. The sampling plan should be
described. The batch analysis reports should include:
A batch is defined as a quantity of DES produced according to a single manufacturing order during
the same cycle of manufacture. A batch should be made with only one lot of coating solution.
Combining stents having different expanded diameters into one batch would only be appropriate
27
See ICH guidances Q2A Text on Validation of Analytical Procedures and Q2B Validation of Analytical Procedures:
Methodology.
27
Contains Nonbinding Recommendations
Draft — Not for Implementation
when the stents originated from the same diameter tubing, have the same design/platform, and only
differ in the balloon diameter to be used. Combining stents of different lengths into one batch is
discouraged.
Because DES batch sizes are typically small and end-product testing consumes a large quantity of
test samples, the applicant may consider any of the following alternative approaches:
• Using in-process testing as a substitute for some release tests (e.g. residual solvents). In
these cases, the tests should still be listed in the finished product specifications with
appropriate notation.
• Using the same test samples for several release tests (e.g. identification, assay, and content
uniformity).
• Using a smaller number of samples than recommended by USP for certain tests (e.g. content
uniformity) with tighter acceptance criteria.
• Using quality by design principles, which rely less on end-product testing and more on
building quality into the product and process design.
General tests that are expected to be included in the specifications for a finished DES are listed
below. A tabular format similar to the example shown in the Appendix A is recommended for
presentation of the specifications.
a. Appearance
b. Identification
Identification testing to establish the identity of the drug substance in the finished product
should be specific (e.g., infrared spectroscopy or a chromatographic method in combination
with an additional test such as UV diode array or MS) and able to discriminate between
compounds of closely related structure that are likely to be present. Identification solely by a
single chromatographic retention time, for example, is not regarded as being specific.
However, the use of two chromatographic procedures, where the separation is based on
different principles, or a combination of tests into a single procedure, such as HPLC/UV
diode array, HPLC/MS, or GC/MS, is generally appropriate.
c. Assay
A specific, stability-indicating assay to determine content should be included for all drug
substances in the finished product. In many cases, it is possible to employ the same
procedure (e.g., HPLC) for assay of the drug substance and quantitation of impurities.
28
Contains Nonbinding Recommendations
Draft — Not for Implementation
When use of a nonspecific assay can be justified, other supporting analytical procedures
should be used to achieve overall specificity. When the assay is not stability indicating, a
separate impurity assay can be employed. A specific procedure should be used when there is
evidence of inactive ingredient interference with the nonspecific assay.
Any impurities, degradation products, and/or residual solvents are included in this category.
We recommend sponsors refer to the ICH Q3B guidance covering finished product
impurities. Appropriate stability-indicating analytical methodology should be used to
monitor degradation products and acceptance limits should be defined for individual
specified degradation products, both identified and unidentified, unspecified degradation
products, as well as total degradation products.
e. Content Uniformity
This test assesses drug content variation from stent to stent within a batch and is to be
distinguished from uniformity along an individual stent length. The latter is typically a one-
time test to establish coating uniformity. The method and limits established in USP <905>
Uniformity of Dosage Units are considered appropriate for determining content uniformity
within DES batches.
f. Drug Release
The specification should include a test for in vitro drug release. The test should be performed
over a sufficient period of time and include a sufficient number of time points to correlate to
in vivo release. The test is generally used as a quality control tool and should be
discriminatory. The results should ideally be reported as percent of label claim released per
unit time. See section VI. E. for additional details regarding in vitro elution testing.
A test procedure and acceptance criterion for evaluation of sterility testing and package
integrity should be included. When test methods differ significantly from compendial test
methods, a demonstration of the equivalency to the compendial method should be provided.
Parametric release can be proposed when appropriate data are generated during development
and validation.
The tests and methods demonstrating the integrity of the microbiological barrier of the
packaging system should be well defined and scientifically justified. Sufficiently sensitive
packaging integrity testing may reduce the need for end product sterility testing.
h. Endotoxins
A test procedure and acceptance criteria for endotoxins, using a procedure such as the
Limulus Amoebocyte Lysate (LAL) test, should be included in the specification.
29
Contains Nonbinding Recommendations
Draft — Not for Implementation
Note: All blood-contacting cardiovascular devices and combination products should be non-
pyrogenic regardless of whether any claims regarding their non-pyrogenic status are made in
the labeling. Pyrogenicity testing is used to help define limits to protect patients from the
risk of febrile reaction. Pyrogenic responses to gram-negative bacterial endotoxins can be
tested using standard methods such as the USP Bacterial Endotoxins Test (<85>) using LAL.
Pyrogenic responses to leachables over the implant life can be tested using a material-
mediated pyrogenicity test. See the companion document (Section titled “General
Biocompatibility Considerations”) for additional specifics on materials-mediated
pyrogenicity testing.
This test evaluates the presence of sub-visible particulate matter. Particulate matter may
include particles shed from the formulation components as well as extraneous particles from
the stent platform, stent delivery system, packaging, and environmental factors. Appropriate
testing and acceptance criteria should be established for particulate matter. See section VI.B
for analytical procedures for characterizing particulate matter.
j. Additional Testing
Additional testing of the finished DES may be necessary to address unique characteristics of
an individual DES. Examples include tests for polymer molecular weight, residual
monomers, catalysts, or other additives.
9. Stability
Stability testing is performed to support the establishment of a shelf life or expiration dating period
for a DES (See also Section VII.C below). Stability studies should also be conducted during
investigational phases to support product stability for the duration of clinical trials.
A stability protocol should be provided that includes storage conditions, time points, test parameters,
analytical methods, and acceptance criteria. The formal stability protocol can include an appropriate
matrixing and bracketing design. At a minimum, the protocol design should include the extremes (in
terms of both stent dimensions and total drug load) as well as an intermediate size to provide
assurance of consistent behavior across the entire proposed matrix of DES sizes to be
commercialized. 28 If there are design differences (e.g., multiple stent platforms) within the proposed
DES matrix, the sponsor should bracket each design or provide a scientific rationale to support the
applicability of the sizes that are tested for the entire product matrix. We recommend that stability
testing include samples from a minimum of three finished product batches for each size tested.
28
See ICH guidance Q1D Bracketing and Matrixing Designs for Stability Testing of New Drug Substances and
Products.
30
Contains Nonbinding Recommendations
Draft — Not for Implementation
Stability testing should be conducted under ICH recommended conditions at room temperature
(25oC/60% RH or 30oC/65% RH) and accelerated conditions (40°C/75% RH). 29 If long-term testing
is conducted at 25oC/60% RH and a significant change as described in ICH Q1A(R2) is observed in
the results obtained for a DES tested under accelerated conditions, additional testing using
intermediate conditions (30oC/65% RH) should be conducted and evaluated against significant
change criteria.
For each set of stability data provided, the sponsor should identify the packaging system, the batch
number and scale, manufacturing date and site, the manufacturing process and formulation. For ease
of review, the Agency recommends that all stability information be provided in tabular format. See
Appendix A for an example of a stability table.
In general, the following tests should be performed at each of the preselected stability time points on
a minimum of three finished product batches to generate the primary stability data used to support an
expiration date:
• Appearance
• Assay/drug content
• Impurities/degradation products
• In vitro drug release
• Particulate matter 30
In addition, some tests, such as sterility, and package integrity, should be performed at release,
annually, and at expiry.
If different finished product manufacturing sites will be used, appropriate release/stability data to
ensure the consistency and equivalency of the finished product should be generated. Generally real-
time, room temperature data should be used to establish a DES shelf life. However, based on the
quality of the data (e.g., accelerated, long-term testing) provided by the applicant, a reasonable
extrapolation of data may be considered to assign the shelf life. It is recommended that simulated
transportation/shipping studies also be conducted as a one-time test to support excursions that may
occur during distribution of a DES.
10. Labeling
Detailed guidance on labeling and examples of text that can be used are included in the stand-alone
companion document. CMC information should appear in the Description sections of the label.
An Environmental Assessment or request for a waiver (with justification) should be submitted (21
CFR 814.20(b)(11)).
29
See ICH guidance Q1A(R2) Stability Testing of New Drug Substances and Products.
30
See section VI. B for test method considerations for particulate matter testing as part of the stability protocol.
31
Contains Nonbinding Recommendations
Draft — Not for Implementation
A. Summary Tables
FDA recommends that a master table be compiled to summarize all mechanical performance,
animal, and clinical testing that has been conducted in support of the DES to either be tested
clinically (under the IDE) or commercialized (for the PMA application) in the United States. An
example of the parameters to be captured in tabular format as part of the master table has been
included in the Companion Document to this guidance. The master table should be provided and
updated, as necessary, for both IDE and PMA applications. To enable the integration of the master
table into the regulatory submission, the sponsor/applicant may decide to divide the table into more
discrete units (e.g., separate tables for engineering, PK, pharmacology/toxicity studies for the drug
substance, and animal studies in support of the DES). This table, or set of tables, will greatly aid in
the sponsor’s and the Agency's assessment of whether sufficient supportive acute and chronic safety
and/or effectiveness data have been provided for the proposed DES as part of both the IDE and PMA
reviews.
Also for ease of review, FDA recommends that a one-page summary of significant trial design
parameters for each clinical study conducted in support of either the IDE and/or PMA applications
be provided. The companion document includes more details regarding this recommendation.
In the event that the DES evaluated in nonclinical or clinical studies differs from the DES that is
intended for commercialization, the sponsor/applicant should provide an appropriate justification for
the applicability of testing provided. This justification, which can include additional limited testing,
can be referred to as a bridging document. FDA will assess the significance of any such differences
when determining whether sufficient information has been provided to support initiation of a clinical
study (IDE) or whether valid scientific evidence has been submitted to provide reasonable assurance
of safety and effectiveness for a PMA application.
B. Engineering Evaluation
The battery of tests and content and format of test data outlined in FDA’s guidance document on
bare metal intravascular stents and their associated delivery systems 31 are relevant for this guidance
and for DES development. FDA recommends that sponsors complete all tests outlined in that
guidance on the finished DES intended for commercialization. Additionally, for those tests that
evaluate characteristics that could be affected by the addition of the drug and/or drug coating,
sponsors should compare those results with the performance characteristics of the bare metal stent
system in a side-by-side fashion. If a test article other than the finished, sterilized DES (e.g., bare
metal stent, prototype, coupon) is used for a specific test, a scientific rationale should be provided
for the applicability of the test article.
31
See guidance for industry and FDA staff on Non-Clinical Tests and Recommended Labeling for Intravascular Stents
and Associated Delivery Systems.
32
Contains Nonbinding Recommendations
Draft — Not for Implementation
FDA recommends that the final, finished DES be evaluated to determine the initial performance
characteristics of the DES. However, if there are any differences between DES tested for initial
characterization, clinical builds (DES used in the human studies) and the DES sought to be
commercialized (due to scale up of the manufacturing process), the changes should be clearly
documented and, as a part of the PMA submission, appropriate additional testing should be
conducted or a scientific rationale provided to demonstrate that these modifications will not affect
the safety and effectiveness of the DES.
A thorough description of the entire manufacturing process should be provided for review. This
description should clearly indicate whether any modifications have been made to the native stent
platform (e.g., texturizing of the stent surface, use of coupling agents, polishing) to facilitate coating
deposition/adhesion onto the stent substrate. The potential effect of additional processing steps on
the durability of the stent substrate as well as the coating should be evaluated.
Since unintended delamination or premature dissolution of a DES coating may influence its clinical
performance and/or mechanical integrity, additional evaluations and suggested modifications to the
battery of traditional engineering testing as outlined in the guidance document referenced above
should be taken into consideration for a DES.
• Test protocols
In addition to the test data (summaries are not typically sufficient), detailed test protocols, which
include the loading parameters, test conditions, samples tested, acceptance criteria, and conclusions
drawn for each of the tests performed on finished, sterilized product, should be provided for FDA
review. A brief description of the derivation or development of the test method, or identification of
other applications in which the method has been previously used should be included.
Test protocols should assess the worst-case conditions that the DES is likely to experience in clinical
practice. Both device configuration and physiologic conditions can affect the performance of a DES.
Extreme device dimensions, tolerances, sizes, and any other important device parameters should be
evaluated. We also recommend that the outer limits of physiologic variables, such as blood pressure,
vascular compliance, and anatomic types, be examined. All test conditions should be clearly stated
in the test protocol and supported with references to applicable literature, standards, or both.
Occasionally, the worst performing combination of device configuration and physiologic conditions
occurs in the mid-range of the relevant variables. This should be considered when developing
protocols to ensure that the worst performing combination has been evaluated.
The term coating may refer to the drug carrier (usually polymeric, but not limited to such), the drug
itself if it is solely coated onto the stent platform, any other coating, or the drug carrier even if it is
incorporated onto the stent in a geometry other than a coating.
1. Coating Characterization
33
Contains Nonbinding Recommendations
Draft — Not for Implementation
As part of the overall coating characterization of a finished DES, the sponsor should conduct
additional studies on a one-time basis as part of the product assessment to establish an understanding
of their DES system as well as appropriate baseline data. FDA believes that adequate baseline
characterization of a DES may help the sponsor identify potential coating integrity concerns earlier
rather than later in the development process. It should be noted that the tests recommended to
characterize the coating and to assess acute and chronic coating integrity are not typically considered
quality control (QC) tests; however, tests for particulate matter recommended in Section VI.B.3.iii
are suggested as part of the QC assessment as described.
Specifically, testing should be provided to address each of the following issues as part of
characterization studies:
• Coating thickness and uniformity along the stent length (both abluminal and adluminal
surfaces, if relevant), circumferentially, and along the sides of the struts.
• Adhesion of the coating to the stent substrate. We recommend a quantitative characterization
of the adhesion strength. If the coating consists of multiple layers (e.g., primers), we
recommend that a quantitative test be performed to determine the cohesive strength between
the layers.
• Chemical identification of particles recovered as part of particulate matter testing (see
Section VI.D.3 below)
2. Coating Integrity
The acute and chronic integrity of coating on the stent substrate should be assessed to provide
reasonable assurance that the coating is able to sustain its integrity according to its design
specifications. The Agency requests that the sponsor qualitatively and quantitatively determine
whether subjecting a DES system to expansion, deployment, and repetitive cycling modalities as
experienced in the clinical setting will influence the ability of the coating to interact appropriately
with the stent substrate. Part of this evaluation will entail determining whether there are areas where
the coating has not been adequately deposited onto the substrate (e.g., defects such as bare spots or
webbing due to manufacturing) versus areas in which the coating may have physically dislodged
(e.g., delaminated) from the substrate due to being subjected to mechanical forces.
As part of this testing, it is recommended that a sampling plan be implemented to examine multiple
lots of DES as well as comparing regions of high stress/strain versus low stress/strain areas to assess
both inter- and intra-lot variability. A sufficient number of images should be provided so that FDA
can make an assessment of consistency.
Furthermore, FDA recommends that coating integrity be evaluated by testing under certain
conditions before and after aging (at a minimum, the product should be aged to the requested shelf
life). These samples do not need to be real-time aged, but can be subjected to accelerated aging
conditions.
For this section of the guidance, acute refers to any time up through expansion and deployment of
the DES, whereas chronic refers to any time after assessment of the initial stent deployment in a
simulated vessel throughout the lifetime of the implant.
34
Contains Nonbinding Recommendations
Draft — Not for Implementation
Acute coating integrity of a DES should be assessed via some visualization method (e.g., scanning
electron microscope). The stents used for this characterization should be representative of the
finished product, subjected to all manufacturing processes, including sterilization. A visual
assessment of the coating integrity on all appropriate surfaces of the DES after expansion in air to
nominal diameter with characteristics appropriately quantified (e.g., continuity, voids) is strongly
recommended to establish a baseline for comparison to coating characteristics after testing
performed under other conditions.
Further visual characterization of the coating should be performed after deployment of the DES to
the maximum diameter as described in the Instructions for Use. If overexpansion of the DES (post-
dilatation) is to be allowed, this should be taken into consideration as part of this testing. It is
recommended that deployment be simulated in an in vitro model intended to mimic in vivo
physiologic and anatomic conditions (e.g., tortuous path, aqueous environment). The stent should be
in direct contact with the simulated vessel without the use of other coatings, lubricants, sheaths, or
protective wraps between the stent and simulated vessel. The rationale for the final model selected
should be provided.
Ideally, the coating should not significantly change in configuration or prematurely delaminate from
the stent substrate upon expansion or deployment.
Chronic coating integrity or, for a degradable polymer system, the loss of coating integrity over time,
can be assessed by performing accelerated durability testing in a simulated in vivo environment. It
is highly recommended that the visual integrity of a DES after 30 and 400 million cycles of fatigue
testing (representing approximately 1 and 10 years of equivalent implant time) be compared to
baseline data in a side-by-side fashion. For degradable polymer systems, timepoints for evaluation
may be specific to the expected degradation profile. A detailed fatigue test protocol, clearly
describing the test equipment, aqueous environment, frequency, loading parameters, and mounting
of samples should be provided with the results from these tests.
The sponsor should consider the following when designing tests to appropriately demonstrate the
chronic coating integrity of a DES:
1. The sponsor should clearly indicate whether the sample consists of single or multiple stents
along with a justification supporting test methods testing multiple samples. Since there is a
reasonable expectation that stents will be overlapped during some clinical procedures,
accelerated durability testing should be performed on multiple stents in an overlapped
configuration.
2. We recommend that testing be conducted with stents in a bent configuration, with a clinically
relevant radius of curvature.
35
Contains Nonbinding Recommendations
Draft — Not for Implementation
3. If a product’s drug elution is completed in a short time relative to the intended lifetime of the
product, coating integrity test samples should be pre-eluted for a worst-case evaluation. This
is a particularly important consideration for those coatings that become porous over time
because of drug elution.
4. At a minimum, we recommend that these additional tests be performed on the finished DES
for the worst-case product sizes for each stent design to demonstrate that the acute and
chronic integrity of the coating has not adversely affected the characteristics of the DES
system.
5. This testing can be combined with fatigue testing intended to evaluate integrity of the stent
platform, if the apparatus can accommodate both tests.
Refer to the section immediately below for additional issues related to characterization of the coating
integrity of a DES.
FDA recommends measurement of particulate matter generated by breakdown of the coating or from
the stent platform, stent delivery system, and product packaging both at release and after aging.
Particulate matter testing serves multiple purposes: (1) it provides an indirect evaluation of the
coating integrity of the finished product and (2) it establishes the number of particles that can
potentially be introduced systemically using the stent system. FDA believes that the main purpose in
particulate matter testing for DESs is to provide a level of assurance of patient safety in terms of
total particulate matter introduced into the bloodstream. Therefore, since the concern applies to the
total number of particles released into the bloodstream, the test should apply to the entire stent
delivery system, not just the stent.
a. Testing Considerations
The sponsor should consider the following when designing tests to appropriately determine
the number, size and/or type of particles for a DES system when subjected to the conditions
described in b-d below.
36
Contains Nonbinding Recommendations
Draft — Not for Implementation
3. The number of samples (a stent, not a strut or portion of a stent) used, the stent size,
and the stent lot should be specified for each test. The selection of the samples
should be scientifically justified.
37
Contains Nonbinding Recommendations
Draft — Not for Implementation
LENGTH (MM)
8 11 15 18 21 24 27
2.5 X X
Diameter 3.0 X
(mm) 3.5
4.0 X X
6. For each test performed, a robust number of stents from multiple stent lots (minimum
of 3 batches) should be evaluated.
Note: In the event that an accessory device (e.g., embolic protection, atherectomy) is
intended to be used in conjunction with a DES, the sponsor should provide appropriate
supportive engineering performance test data to ensure that the integrity of the coating is
maintained. We recommend that sponsors contact appropriate FDA staff to discuss
engineering testing recommendations.
b. Characterization
For the purposes of characterization of the finished, sterilized DES, particulate matter testing
should be performed and particles collected and appropriately measured for several different
test cases:
Such testing should involve expansion of the stent to its nominal diameter in a beaker of
solution. If the stent is not a balloon-deployed stent and is self-expanding, this condition and
the over-expansion condition described below may be equivalent and combined into one test
condition.
38
Contains Nonbinding Recommendations
Draft — Not for Implementation
This testing should involve expansion of the stent to the maximum diameter allowed, as
described in the post-dilatation limits in the IFU in a beaker of solution.
This testing should be performed with use of an in vitro model as described in section B.2
(acute coating integrity) above. Note that physiologically relevant worst-case conditions
should be applied. To ensure measurement of the total number of particles that could be
potentially introduced into the bloodstream, the stent delivery system should be inserted into
the text fixture to the point at which it would be inserted in clinical use.
• Fatigue/durability testing
This testing should be performed with use of a test fixture as described in section B.2
(chronic coating integrity) above. Note that physiologically relevant worst-case conditions
should be applied. This should include multiple stents placed in an overlapped and bent
configuration. It is recommended that particulate matter generation be measured at multiple
time points, rather than at t=0 and 400 million cycles. One advantage of this approach is that
a pattern/trend of particulate matter generation can be described (e.g., plateaus, monotonic
increases). Depending on this trend, the sponsor may be able to determine the appropriate
number of fatigue cycles (which may be significantly less than 400 million) necessary to
demonstrate that the coating will not unintentionally break apart or, for a degradable polymer
system, to quantify the particulate matter generation associated with the degradation of the
polymer.
c. Quality Control
If the amount of particulate matter recovered from over-expansion testing and simulated use
testing is substantially similar, either test may be used for quality control testing. However,
if these two test conditions resulted in different amounts of particulate matter, the more
challenging test, the simulated use condition, should be performed for quality control
purposes. In either case, the test should be performed on every batch of product
manufactured as part of batch release (see Section V.B.8 above for other parameters to be
measured for batch release).
d. Stability
For stability testing, we recommend that aged samples be evaluated using the simulated use
test condition. If the over-expansion condition is used for quality control purposes,
additional testing using the simulated use condition should be performed on stability batches
at t=0. It is highly recommended that particulate matter generation over time be evaluated at
each time point in the stability protocol (instead of only at t=0 and t=proposed expiration
date). In the event that the particle counts continually increase with aging or fail to meet the
39
Contains Nonbinding Recommendations
Draft — Not for Implementation
acceptance criteria at the proposed expiration date, additional data will be available to
support a shorter expiration date for the DES.
If the underlying stent substrate of the DES is metallic, FDA recommends that the sponsor evaluate
the effects of cracked or delaminated coatings on corrosion resistance. We recommend that
corrosion testing be performed after intentionally creating a defect in the coating, which exposes the
base stent substrate. We recommend testing according to the methods described in ASTM F746 32
or an equivalent method. The sponsor can modify the method by incorporating the experimental
setup described in ASTM F2129. 33
Additionally, since there is a reasonable expectation of stent overlap during clinical procedures, the
potential for fretting corrosion between two DESs should also be addressed. The sponsor should
ensure that micromotion between strut elements is actually occurring. We recommend that the
sponsor incorporate examination of samples for fretting corrosion as part of fatigue/durability
testing. A scientific rationale for the number of samples evaluated for fretting corrosion should be
provided.
If a stent contains more than one type of metal, such as a laminate, we recommend that the resistance
of the stent to galvanic corrosion be demonstrated. If stents of different materials will be overlapped
during clinical procedures and the contacting or overlapping stents may be made of different
materials, we recommend that the potential for galvanic corrosion between stents be addressed. We
recommend testing according to the methods described in ASTM G71, 34 or an equivalent method.
Sponsors can modify the method by incorporating the experimental setup described in ASTM
F2129.
5. Degradable coatings
If a DES has a degradable polymer carrier, the environments for the experimental tests described
above should be carefully taken into consideration since they may affect the interpretation of the
results. Therefore, we recommend that a full characterization be performed of the degradation
profile (both in vitro and in vivo) of the biodegradable polymer carriers. The resulting information
should be used to design the test environment for the evaluations described above, as well as to
assess the appropriate timelines for additional nonclinical studies (e.g., supportive animal studies,
elution characteristics).
The durability of the degradable coating becomes important near the end of the coating lifetime
when degradation has weakened the coating. We therefore recommend that particulate matter
testing be conducted in fatigue testing for the life of the coating. The trend or pattern of particulate
matter generation as the coating degrades should be described. It may also be instructive to observe
32
ASTM F746 Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials.
33
ASTM F2129 Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to
Determine the Corrosion Susceptibility of Small Implant Devices.
34
ASTM G71 Standard Guide for Conducting and Evaluating Galvanic Corrosion Tests in Electrolytes.
40
Contains Nonbinding Recommendations
Draft — Not for Implementation
the coating via visual/microscopic methods near the end of the coating lifetime to characterize the
pattern of degradation to understand the potential for increased particulate matter generation (e.g.,
Does the degradation occur preferentially at the surface or stent interface once some interface has
been exposed? Is the degradation patchy?).
Shelf life/stability characterization becomes very important for degradable/resorbable polymers. For
example, exposure to humidity may begin the degradation process and therefore not only reduce the
shelf life, but increase the elution at early stages of the product and decrease the effective lifetime of
the coating.
It is also very important to characterize the effects of the sterilization processes on the coating,
because many processes (e.g., irradiation) reduce the molecular weight of the polymers, which may
allow an increase of elution at early stages of the product and reduce the effective lifetime of the
coating.
C. Biocompatibility
Biocompatibility testing should be conducted in accordance with ISO 10993. 35 For certain tests,
evaluation of the stent should be carried out separately from the delivery system. For additional
considerations related to biocompatibility testing, refer to the companion document.
Prior to undertaking GLP animal safety studies, pilot DES animal studies should be conducted to
evaluate the degree of systemic exposure, local vascular and regional myocardial levels of the drug
component of the stent. This information can be discussed with FDA and will inform the need for,
and extent of, separate studies or data on systemic clinical pharmacology.
DES nonclinical in vivo safety studies conducted in appropriate validated healthy animal models are
intended to assess handling characteristics (delivery and deployment), the biological response to the
DES, drug effects, and stent-related pathology. In addition, these studies are used to identify
potential clinically relevant major adverse events that should be considered prior to beginning
human clinical trials or that may influence clinical study design. The design of these studies should
also evaluate stents that incorporate a safety margin over the highest drug dosage and greatest
polymer concentration intended to be evaluated in the IDE clinical study as well as for all reasonably
anticipated intended clinical uses of the DES.
Animal studies should compare combinations of the stent components (i.e., bare stent, and stent +
polymer + drug) in both nonoverlapping and overlapping configurations. The sponsor should clearly
identify any differences (e.g., stent design differences, polymer thickness, drug amounts) between
the DES used for nonclinical studies and the proposed IDE study.
Studies of stent + polymer (without drug) should be performed if safety concerns are observed with
the finished DES product so as to help identify whether pathologic changes are more likely due to
35
ISO 10993-1 Biological evaluation of medical devices—Part 1: Evaluation and testing.
41
Contains Nonbinding Recommendations
Draft — Not for Implementation
the drug or the coating. The stent + polymer sample should include both biodegradable and non-
biodegradable polymer carriers as well as the primer layer.
Demonstration of probable product safety is currently considered to be the primary purpose of the
nonclinical animal studies. Demonstrating potential product efficacy (i.e., inhibition of neointimal
hyperplasia) is an important secondary endpoint. However, for any given drug-device combination,
the potential efficacy observed during animal studies should be appropriate to balance any potential
safety concerns that were observed during the same studies. Also, it is reasonable to presume that
the demonstration of the potential efficacy of a new DES in an animal model may assume increasing
importance over time if multiple DESs are approved for clinical use.
Refer to the companion document for general recommendations regarding good animal husbandry.
Because of the similarities in the size, anatomic distribution, and time-dependent progression neointimal
growth within stents in human coronary arteries, the swine model has historically been relied on for
testing of intracoronary devices. However, because of inherent differences between animal and human
vascular responses to stent implantation, animal testing is primarily focused on the evaluation of safety,
rather than sustained long-term efficacy. Small animal models (e.g., rabbit iliac artery) can provide
complimentary data on optimal dose finding and DES mechanism of action.
Currently, there is no animal model that can both (1) replicate the heterogeneity of human
atherosclerotic coronary disease and (2) accommodate the sizes of catheters and stents used in
humans. Due to potential experimental complexity and in the absence of studies demonstrating
predictive capabilities, atherosclerotic animal models to test the safety and performance of these
products have not been routinely requested. However, although advanced stenotic atherosclerotic
lesions in animals may not be available, sponsors may consider DES implantation in modifications
of normal vessels (e.g., intimal lipid/inflammatory cell-rich or fibrotic lesions) to test device
performance in vascular environments that may be relevant to human use.
Unless there is a specific reason to do otherwise, the stent should be implanted in an artery that has no
prior injury. Antiplatelet therapy should be administered based on the current clinical standard of care
and that to be used during the clinical study.
The Agency recommends the use of, at minimum, general animal study guidelines, necropsy, and
arterial histopathology methods, including those described below. The study findings from each stent
42
Contains Nonbinding Recommendations
Draft — Not for Implementation
type (i.e., bare stent, stent + polymer + drug, and if indicated, stent + polymer) should be compared.
We recommend the following.
43
Contains Nonbinding Recommendations
Draft — Not for Implementation
arterial sections should be included as an appendix in the final pathology report. To ease
review, we recommend providing all gross photographs, radiographs, and photomicrographs
in electronic format.
• Histomorphometric evaluation of sections is essential for the assessment of DES biological
response and safety. These measurements should minimally include the following:
neointimal area, neointima thickness at each strut site, medial area, internal and external
lamina area, lumen area and percent area stenosis. Measurements should be performed on
each stent section (proximal, middle, and distal), and a mean measurement for each
parameter for the entire stent should be reported. From these data, the percentage of the stent
narrowed by neointimal tissue (percent stent stenosis) can be calculated. A mean injury
score for each stent should be determined.
The non-stented adjacent arterial sections (5 mm proximally and distally) should undergo
comprehensive histologic evaluation including an assessment of arterial injury, neointimal
thickening, inflammation, and thrombus deposition.
Following DES implantation, any sudden or unscheduled animal deaths should be vigorously
investigated for cause. In such cases, a thorough necropsy should be conducted, including
evaluating all stented arteries and specifying the cause of death. Any clinical problems (e.g., fever,
allergy, evidence of renal or hepatic dysfunction) should also be recorded. We recommend that
complete data on thrombus, myocardial infarction, aneurysm, and perforation be collected and
included with the pathology report within the IDE submission.
3. Study Duration
Animal studies designed to assess biological response and safety of the final clinical version of the
DES should be conducted prior to first in human use. At a minimum, 1- and 6-month studies are
suggested; 3-month animal data are optional, and depending on the results, may be sufficient to
begin a clinical feasibility trial.
In view of the mechanism of action of most DESs, longer term follow-up studies (e.g., beyond 6
months) are likely to be necessary to assess (1) chronic inflammatory reactions, (2) delayed or
incomplete endothelialization, (3) late stent thrombosis and restenosis, and (4) chronic biological
responses to the surface polymer after complete drug elution and, in the case where a biodegradable
polymer is used that takes longer than 6 months to fully degrade.
In nonclinical studies at all time points, histology should be carefully evaluated for polymer
delamination from the stent.
44
Contains Nonbinding Recommendations
Draft — Not for Implementation
Note: Given the differences in injury and healing responses between the animal models and humans,
in addition to inherent variability between the designs of different DES systems, a definitive long-
term follow-up time point for animal model studies to assess late effects cannot be explicitly
recommended.
4. Biological Response
We recommend that a three-way comparison of the histopathological findings for the bare metal
stent, polymer-only stent (if indicated), and the polymer-drug stent combination be conducted at
appropriate time points, minimally to include 1 and 6 months. We recommend that at least six to
eight samples of each of the stent types be evaluated with a minimum of three to four animals per
time point. We recommend enrollment of extra animals in anticipation of possible early animal
deaths.
a. Histopathology Endpoints Assessing Drug Effects
Study endpoints should focus on the characterization of localized drug effects within the vessel
wall of the stented vessel as well as immediately proximal and distal to the stented vessel segment
(i.e., to observe any potential edge effects). Evidence of DES-related drug effects and pathology
includes factors such as mural thrombus formation, fibrin deposition, inflammation (strut
associated; neointima, media, adventitia), granulomas, neointimal smooth muscle density, medial
necrosis and thinning, dystrophic calcification, endothelialization, vessel wall hemorrhage, and
neoangiogenesis. We recommend that a scoring system be used to record the incidence and
severity reported by stent segment region (i.e., proximal, mid, distal).
It is important to evaluate whether a drug produces pathology in the tissue downstream from
the stent. Using the highest total drug dosage proposed for clinical use, a thorough gross and
histopathology evaluation of multiple anatomic regional sections of myocardium perfused by
the stented artery should be conducted to identify stent-related cardiac pathology (e.g.,
infarcts, thromboembolic material, myocardial necrosis and fibrosis).
In addition, the drug effects immediately proximal and distal to the stented segment of the
vessel (referred to as an edge effect) should be assessed. Using similar histopathology and
histomorphometric endpoints as described above (VI.C.2 and 4a), the findings should be
compared to the stent segment of the vessel.
If long stents are evaluated separately (refer to section VI.F.7), this evaluation should be
completed both for standard length stents and for long stents.
The objective of studies of stents with higher drug and polymer dosages than will be applied to the
clinical or to-be-commercialized version of the stent is to establish a safety margin over and above the
dose intended for clinical use. These studies can reveal whether adverse effects are observed at higher
dosages, and at what dosage the effects are observed. The following drug formulation characteristics
45
Contains Nonbinding Recommendations
Draft — Not for Implementation
• Dose density
• Total dose loaded
• Coating thickness
• Amount of drug delivered to the tissues
• Residual amount of drug on the stent
• Release rate
In animal studies intended to establish a safety margin, the dose density, amount of drug or polymer
loaded, and number of stents should be designed to justify a margin of safety over the proposed clinical
trial dose. In addition, drug release characteristics should be analyzed in relation to local tissue drug
concentration, vascular biological responses and local toxicity. The release rate is important because it
directly correlates with the local vascular toxicity. Additional animal studies should be carried out to
evaluate the safety of stents containing higher dosages of drug and polymer (i.e., a three- to ten-fold
margin over the intended drug dosage density of the final product) to evaluate whether the DES has an
appropriate local, regional, and (possibly) systemic safety margin with regard to drug dosage density. If
loading high drug concentrations onto the stent is technically difficult or significantly alters the
degradation profile for a degradable carrier, the Agency recommends evaluating regions of overlapped
stents to theoretically support safety margins. Evaluation of over-dosage stents should include the
longest, largest diameter stent, and if multiple stents are routinely used, the combined drug density of the
highest number of, and the longest, stents allowed in the planned human study.
6. Overlapping Stents
Since overlapping stents are commonly implanted in current clinical practice, animal studies should
be undertaken to evaluate the safety of overlapping DESs and provided as part of the IDE
submission. Stents overlapping by a minimum of 4 mm should be evaluated at 1 and 6 months
(optionally at 3 months), in a minimum of six stents per stent type. Histopathology sections should
be obtained from both overlapped and non-overlapped regions. Histopathology and
histomorphometric endpoints should be reported and compared by stent segment (i.e., proximal,
overlapped, distal stent).
Due to the likely possibility that multiple overlapping stents will be used, FDA recommends that
animal testing on overlapping stents be provided as part of the PMA submission whether or not
testing is included within the clinical study to provide a preliminary assurance of safety.
7. Long Stents
A separate evaluation should be completed for the longest stent model if a long DES (i.e., >30 mm)
is to be marketed. Evaluation of angiography and histopathology is particularly important to
characterize the biological and drug response along the full length of the stent. Histopathology
sections should cut at approximately 10 mm intervals, plus one section 5 mm proximally and distally
beyond each end of the stent. The Agency will not routinely request comparisons to long stent
46
Contains Nonbinding Recommendations
Draft — Not for Implementation
controls. Results of the long DES may be compared to those observed for standard-length control
stents and DES.
This section provides suggestions on elements to consider in the assessment of the clinical
pharmacokinetics of a DES and on the evaluation of both in vivo and in vitro release characteristics
of the drug from a DES.
The evaluation of the pharmacokinetics (PK) of a DES can be accomplished in one of the
trials of patients implanted with the DES. The sponsor should provide a detailed protocol
describing the design of the PK study. The in vivo drug release kinetic information
generated during the animal studies could be useful in designing the human PK study (i.e.,
appropriate PK sampling times, length of PK study).
To obtain PK information at the highest possible drug exposure, it is recommended that the
PK evaluation occur in a trial including patients receiving multiple and overlapping stents.
The measures or parameters for the drug should include area under the plasma concentration
versus time curve (AUC), peak plasma concentration (Cmax), time to peak plasma
concentration (Tmax), elimination half-life (T1/2), and total clearance (Clt). If there are major
metabolites associated with the therapeutic or toxic effects of the drug, they should also be
determined.
b. Population-PK
A population PK-sparse sampling approach can also be used for the collection of clinical PK
data for the DES from patients enrolled in the clinical trials. See CDER’s guidance for
industry Population Pharmacokinetics.
c. Bio-Analytical Methods
The evaluation of the samples collected during the PK study should be evaluated for drug
content using properly validated analytical methods. Additional information on validation of
methods can be found in CDER’s guidance for industry Bioanalytical Method Validation.
The in vivo drug release information generated in the animal studies can be very useful (1) in the
design of the in vivo human PK assessment conducted as part of the clinical program (i.e.,
47
Contains Nonbinding Recommendations
Draft — Not for Implementation
appropriate PK sampling times, length of PK study), (2) in the development of in vitro release
methodology that mimics the in vivo drug release, and (3) in the development of an in vivo-in
vitro correlation (IVIVC).
The in vivo release of a drug can be divided into two types. First, the release can be directly
measured using the amount of drug remaining in explanted stents with respect to time until
complete drug elution profile is obtained. The release can also be measured using the blood
and/or tissue concentration data. The in vivo release profile generated using the first method
represents drug release from the stent to the surrounding tissues and systemic circulation
while that generated using the second method represents drug released from the stent and the
surrounding tissue into the systemic circulation.
The in vivo local and systemic drug kinetics of the DES to be used in the IDE clinical studies
and submitted in the PMA application for marketing approval (if there are modifications)
should be thoroughly characterized in an appropriate animal model. The release of drug
from the stent should be evaluated at specified time intervals covering the complete drug
elution profile (immediately after implantation until the drug is completely eluted from the
stent). Drug concentrations should be assessed in the blood, in arterial tissue, and in
myocardial tissue proximal and distal to the stent, as well as in remote tissue, such as the
liver, lung, and kidney. In the tissue surrounding the stent, the drug should be evaluated until
there are no longer detectable levels.
Assessments should include whether the drug’s concentration is uniform along the stent
length or preferentially distributed at either end. Evaluations should compare the terminal
elimination t1/2 of drug from stent to the true elimination t1/2 obtained after IV administration.
If drug release from the stent is slower than the elimination process (flip-flop phenomenon),
the rate limiting step is the release of drug from the stent.
In vitro release testing is a powerful and useful tool for obtaining data related to a product’s
quality and, potentially, its clinical performance. The Agency considers the development of
acceptable, discriminating in vitro elution methodology and specifications as critical for the
adequate characterization of a DES product tested clinically as well as to validate consistency
in the commercially manufactured product. Because this testing serves multiple important
purposes, including use in DES characterization, batch release, and stability testing, the in
vitro elution method for the testing of the release of drug from the DES should be developed
and validated as early in the development process as possible and definitely prior to
submission of the PMA application.
The in vitro drug release/elution kinetics should be evaluated under appropriate conditions
based on the mechanism of drug release and to emulate hydrodynamic considerations of stent
deployment. In vitro drug release kinetics characterization should provide valuable insight
on the time course of drug release and on the drug remaining on the stent. The relative
48
Contains Nonbinding Recommendations
Draft — Not for Implementation
solubility of the drug also determines the relative kinetics such that a more lipophilic drug
exhibits a longer time of elution. We recommend that the in vitro release profile generated
with the chosen method mimic the in vivo elution behavior of the drug from the DES. If this
is not possible (e.g., the in vivo release is limited), the in vitro method should be optimized
for its ability to detect manufacturing lots outside the boundaries established in the clinical
trials.
A detailed description of the optimal in vitro elution methodology and the developmental
parameters (i.e., equipment/apparatus, in vitro release media, agitation/speed, temperature,
pH, assay) that were used to identify this method as most appropriate should be submitted to
the Agency in the IDE. Also, the method validation information showing that the chosen
method is able to detect manufacturing changes (under meaningful testing) that may have an
effect on the release of the drug should be submitted. Validation studies are important for
identifying critical formulation and manufacturing variables during development,
establishing relevant controls for manufacturing, and developing a relevant stability
indicating test method for final product testing. An in vitro test method based on mechanism
of drug release can also be a valuable tool for ensuring unchanged performance of
manufactured lots.
The elution profile should be complete and cover at least 80 percent of drug release of the
label amount or whenever a plateau is reached. We recommend use of at least six samples
per testing variable. The elution data (individual, mean, profiles) should be reported as the
cumulative percentage of drug eluted with time (the percentage is based on the product’s
label claim).
In vitro drug release kinetics should be reproducible between stents within a lot and between
manufacturing lots and should be stability-indicating. The chosen method should be
discriminatory and sensitive enough to reject lots that would have less than acceptable
clinical performance.
For the setting of the drug release/elution acceptance criteria, the following points should be
considered:
• The in vitro elution specifications should encompass the timeframe over which at
least 80 percent of the drug is eluted or where the plateau of drug elution is reached if
incomplete elution is occurring.
• Data from lots used in the clinical trials and stability studies, and also on to-be-
marketed batches, should be used.
• The establishment of at least three sampling times covering the initial, middle, and
terminal phases of the complete elution profile data should be selected. The
acceptance criteria ranges should be based on the overall elution data generated at
these times.
• Acceptance criteria should be set in a way to ensure consistent performance from lot
to lot.
49
Contains Nonbinding Recommendations
Draft — Not for Implementation
• The chosen acceptance criteria should not allow the release of any lots with elution
profiles outside those that were tested clinically.
The applicant should note that an agreed upon in vitro elution test (i.e., specifications and
acceptance criteria) is critical as a quality control (QC) tool during the stability program and
establishment of the DES shelf life and is part of the QC tests performed for the release of
DES batches.
To obtain an in vitro-in vivo relationship, two sets of data should be collected. The first set
contains the in vitro data, usually drug release data from an elution test, and most often takes
the form of percentage of drug released as a function of time. The second data set contains
the in vivo data. For a DES, the in vivo release of a drug can be assessed by determining the
blood-drug concentration data and also by measuring the amount of drug remaining to be
released from the recovered stents. Although data from either or both methods can be used
in the development of an IVIVC, for a DES, the systemic drug levels might be very low or
below quantitation limit. Thus it becomes more feasible in constructing the IVIVC model to
use the in vivo release data from the explanted stents. A model that integrates both (i.e.,
mechanism of drug release and systemic drug concentration) may provide a means for
developing a physiologically based PK model for predicting drug disposition and for
establishing relevant mechanism based IVIVC.
Once the in vitro and in vivo data sets are available, a mathematical model describing the
relationship between the in vitro and in vivo data sets should be developed. One mechanism
for determining whether a correlation exists between the in vitro release kinetics and the in
vivo tissue uptake is to plot the amount of drug released in vitro versus the amount released
in vivo at the same time points to see whether a point-to-point relationship exists (level A
correlation). When trying to develop such a relationship, the in vivo data set is fixed. Once
this information is generated, it establishes the relevant performance of the DES product. On
50
Contains Nonbinding Recommendations
Draft — Not for Implementation
the other hand, the in vitro release profile may be modified through changes in the release
test conditions to obtain a consistent relationship between the percentage of drug released in
vitro and the fraction of drug released in vivo.
A PMA must include a complete description of the methods, facilities, and controls in sufficient
detail that FDA can make a knowledgeable assessment of the quality control used in producing the
finished DES (see 21 CFR 814.50). Although particular aspects of the manufacturing of the finished
DES are addressed in Section V.B., Chemistry, Manufacturing, and Controls, a full description of
the manufacturing methods, facilities, and controls must be provided at the time of the PMA
submission (see 21 U.S.C. 515(c)(1)(C)).
A drug-device combination product must meet current good manufacturing practice requirements for
both the drug and device constituent parts of the combination product (e.g., 21 CFR 210/211 for
drugs, 21 CFR 820 for devices). For a discussion of the Agency’s current thinking on how to apply
these manufacturing requirements for a combination product, you may wish to refer to the draft
guidance for industry Current Good Manufacturing Practice for Combination Products, issued by
the agency in September 2004. 36 The draft guidance describes a quality management framework for
combination products that, if properly implemented, would give manufacturers the flexibility to
select either the CGMP regulations (21 CFR 210/211) or the Quality System regulation (21 CFR
820) as their umbrella manufacturing operating system, provided their current good manufacturing
practice operating system incorporates key specific provisions pertaining to the other part of their
combination product. 37 Under such an approach, if the Quality System (QS) regulation (21 CFR
820) is chosen as the umbrella set of regulations for the manufacturing operative system for a DES
product, complete manufacturing and quality control information for the DES product would be
provided pursuant to the QS regulation (see 21 CFR 814.20(4)), 38 incorporating key, specific
provisions from the drug CGMP regulations (21 CFR 211). Likewise, if the CGMP regulation is
chosen as the umbrella manufacturing operating system, complete manufacturing and quality control
information should be provided for the DES product pursuant to the CGMP regulations (21 CFR
36
See http://www.fda.gov/oc/combination/OCLove1dft.html.
37
The Agency has since announced its intent to issue a Proposed Rule on Current Good Manufacturing Practice for
Combination Products (72 Fed. Reg. No, 236 (2007), available at
www.RegInfo.gov/public/do/eAgendaViewRule?ruleID=279375.
38
See, e.g., guidance for industry Quality System Information for Certain Premarket Application Reviews,
www.fda.gov/cdrh/comp/guidance/1140.pdf, for more information.
51
Contains Nonbinding Recommendations
Draft — Not for Implementation
Parts 210 and 211), incorporating key, specific provisions from the device QS regulation (21 CFR
820).
B. Sterilization
The PMA application should identify the sterilization method and include the validation for the
sterilization method and the sterility assurance level (SAL) achieved. In general, sterile devices
would meet an SAL of 10-6, unless there is a substantial scientific justification provided for not being
able to achieve this level and for why patients would not be at increased risk. Sterilization validation
should be carried out in accordance with a recognized standard or equivalent method. 39
C. Package Integrity
Package integrity testing should be performed to demonstrate the ability of the package to maintain
the sterility of the product contained within it. Package integrity testing generally consists of a
whole package physical integrity test in conjunction with a seal integrity test. Some methods for
package integrity testing may be found in ISO 11607.
Additionally, appropriate testing should be conducted to evaluate the ability of the packaging to
withstand forces generated during shipping and distribution from the manufacturer to the end user.
Test methods such as those described in ISO 2248 and ISO 8318 40 may be appropriate.
In addition to the tests recommended to demonstrate stability of the DES discussed above (see
Section V.B.9), testing should also be performed to demonstrate that the functionality of the stent
and delivery system (i.e., mechanical performance), the coating integrity, and the package integrity
have not degraded over the requested shelf life. Testing should be performed on a finished,
sterilized DES product that has been manufactured and packaged in the same manner as intended to
be commercialized. Due to the presence of the polymer and drug components accelerated aging is
not appropriate for stability testing as described in Section V.B.9 above; however, testing to
establish the continued functionality of the stent and delivery system may be conducted using
samples subjected to accelerated aging. For certain tests, such as coating integrity, accelerated aging
conditions can have a significant detrimental impact on the DES such that real-time aging should be
considered.
39
FDA recognizes the following standards for steam, ethylene oxide, and radiation sterilization, respectively: ISO
11134, ISO 11135, and ISO 11137 (see guidance for industry Recognition and Use of Consensus Standards,
http://www.fda.gov/cdrh/ost/guidance/321.html.
40
ISO 2248 Packaging – Complete, filled transport packages – Vertical impact test by dropping; ISO 8318 Packaging
— Complete, filled transport packages and unit loads — Sinusoidal vibration tests using a variable frequency
52
Contains Nonbinding Recommendations
Draft — Not for Implementation
A. General Considerations
Clinical trials of a new DES should not begin until the sponsor demonstrates that there is reason to
believe that risks to subjects are outweighed by the anticipated benefits to the subjects and the
importance of the knowledge to be gained. Depending on the amount of available information, a
feasibility study may be recommended to allow the collection of initial data in human subjects. If
feasibility (sometimes referred to as “first in human”) data are available from studies undertaken
outside the United States (OUS), additional data collection in a feasibility study in the United States
may not be necessary. However, the quality, applicability, and duration of such OUS feasibility
studies will be critical to assess whether these data can be considered directly or indirectly applicable
to the DES intended for clinical use in the United States. Such information should be reported in the
Report of Prior Investigation section of an IDE. The companion document includes an example of a
one-page summary that may be used for ease of review.
FDA encourages study sponsors to use the pre-submission process 41 to gain informal feedback on
proposed clinical protocols for DES, including feasibility or pivotal studies. Additionally, although
FDA generally does not regulate device clinical studies performed outside of the United States, we
are willing to provide informal feedback on clinical protocols for OUS studies that are planned to
support either an IDE or PMA application.
FDA believes that a clinical protocol for a coronary DES should include the following elements:
41
See guidance on IDE Policies and Procedures, http://www.fda.gov/cdrh/ode/idepolcy.pdf.
42
You should review the statutory definition of applicable clinical trial to determine if your trial must be registered
to comply with the law. See PL 110-85, Section 801(a), (adding new 42 U.S.C. 282(j)(1)(A)).
http://frwebgate.access.gpo.gov/cgi-bin/getdoc.cgi?dbname=110_cong_public_laws&docid=f:publ085.110.pdf
53
Contains Nonbinding Recommendations
Draft — Not for Implementation
B. Intended Use
The sponsor should identify, as clearly and precisely as possible, the intended use of the DES. The
specific indications should include the following:
The intended use determines the objectives of the clinical trial, which are generally to demonstrate
the safety (i.e., associated morbidity and mortality) and effectiveness (i.e., associated patient benefit)
of the product for a defined clinical benefit in a target population under specific conditions of use. 43
Following the approval of the first two coronary DES, data were collected that suggested a small but
significant increase in the rate of stent thrombosis associated with DES as compared to bare metal
stents, occurring after the first year of implantation. FDA convened an Advisory Panel meeting on
December 7 and 8, 2006, in an effort to fully characterize the risks, timing, and incidence of DES
thrombosis. Three topics were discussed by the experts on the panel, DES manufacturers, and
clinical investigators: (1) the rates of stent thrombosis and associated clinical sequelae (death and
MI) when DES are used in accordance with their labeled indications; (2) the rates of stent
thrombosis and associated clinical sequelae (death and MI) when DES are used in a broader, more
complex population of patients and lesions; and (3) the optimal duration of dual antiplatelet therapy
in patients who receive DES. More specific information about the meeting and the conclusions
reached are available on FDA’s Web site. 44
Information can be submitted to ClinicalTrials.gov using the Protocol Registration System (PRS). For more information
visit the PRS Information Page (http://prsinfo.clinicaltrials.gov).
43
Although indications are commonly refined over time as clinical data from feasibility studies are analyzed, at the
pivotal trial stage of product development, the intended use and indications should be in reasonably sharp focus.
44
FDA statements available at http://www.fda.gov/cdrh/news/091406.html and
http://www.fda.gov/cdrh/news/010407.html. Panel summary and transcript available at
http://www.fda.gov/ohrms/dockets/ac/cdrh06.html#circulatory.
54
Contains Nonbinding Recommendations
Draft — Not for Implementation
As an outcome of that panel meeting, FDA recommends that all DES clinical programs address the
following questions as part of the information provided to demonstrate a reasonable assurance of
safety and effectiveness:
1. The rates of critical clinical endpoints related to safety and effectiveness, such as death,
myocardial infarction, and need for revascularization should be determined.
2. The rate of death and myocardial infarction (MI) should be determined. Not only are these
critical safety endpoints, but adequate precision around the rates of death and MI is needed to
understand the impact of stent thrombosis on the overall safety and effectiveness profile of a
DES.
3. The rate of stent thrombosis over time should be addressed. For example, the rate of stent
thrombosis up to and after 1 year should be determined, including whether the rate increases,
decreases, or plateaus over time. Analyses should be presented for both patients receiving the
DES within the labeled indication and patients representing broader use of the product.
Clinical resistance to antiplatelet therapy (resistance to aspirin, clopidogrel, or both) may emerge as
an important risk factor for stent thrombosis. Evaluation of responsiveness resistance to antiplatelet
therapy may be a future recommended test. FDA is open to different approaches and trial designs to
address these critical questions. Suggested approaches are discussed in the sections to follow.
D. Study Designs
Randomized controlled trials (RCTs) are the most appropriate trial design for a new DES, although
for certain additional indications beyond initial approval (e.g., additional stent diameters, lengths or
certain lesion types), other trial designs may be appropriate. Both superiority and noninferiority
RCTs can be used to support the safety and effectiveness of a DES.
1. Superiority Study
For a DES, an RCT study design could compare a DES, as the investigational device, to a bare metal
stent, as the control arm. However, the choice of control in a superiority design is not limited to a
bare metal stent. A sponsor may choose to evaluate the superiority of an investigational DES to an
active DES control (i.e., an FDA approved DES). The investigational DES should be shown to be
superior to the preselected control by a margin agreed to be clinically significant by the clinical
55
Contains Nonbinding Recommendations
Draft — Not for Implementation
community and FDA. In a bare metal control trial, it may also be useful to include a third arm,
another DES; this enables assurance of comparability to other DESs.
2. Noninferiority Study
The noninferiority, or equivalence, approach to study design has been used increasingly in clinical
trial settings where a placebo or previous standard of care as control is either unavailable or
unacceptable for logistical or ethical reasons. In this design, patients are randomized to
investigational DES or active DES control, as above, but the study hypothesis is noninferiority, not
superiority.
A noninferiority clinical trial usually refers to a study designed to show that an investigational
device is as effective, or almost as effective, as an approved device or a standard of care (active
control), from which it is then inferred that the investigational device is effective. In fact, the study
actually demonstrates that the investigational device is not inferior to the control by more than a
prespecified noninferiority margin delta. The margin delta used would be the largest acceptable
reduction in therapeutic response with the investigational device (i.e., the maximum tolerable
treatment difference such that the new device would still be considered sufficiently effective).
Before a noninferiority margin can be chosen, the treatment effect size for the active control device,
compared to the previous standard of care (BSM, in the case of DES), should be established based
on historical evidence of safety and effectiveness from controlled clinical trials. Subsequently, the
noninferiority margin for a new trial can be chosen based on clinical judgment regarding the
proportion of the initial effect size that should be maintained in the new comparison. It is also
critical to consider whether there is reason to believe that past examples of safety and effectiveness
would still be applicable to the current study (the constancy assumption). We recommend that
sponsors discuss selection of an appropriate noninferiority margin with FDA as the clinical study is
being designed.
To investigate whether the investigational device is noninferior to the control, the appropriate null
hypothesis is that the control is better than the investigational device by at least the noninferiority
margin. The alternative hypothesis is that the investigational device is not worse than the control by
the noninferiority margin. These two hypotheses are the essence of how FDA views noninferiority
trials.
Although the noninferiority trial design is a strategy that could be used when a placebo-controlled
study cannot be conducted, there are some limitations to the noninferiority study design that should
be considered prior to adopting this approach. When a noninferiority study includes as a control a
DES that has not been directly compared to a BMS, the potential exists for a downward drift in the
true difference in safety and effectiveness between the investigational DES and a BMS. After serial
noninferiority studies, this so-called outcome drift could lead to a situation in which the
investigational DES could be found noninferior to the latest noninferior DES, but no longer superior
to a BMS, if such a direct comparison were made.
The quantification of delta should be clinically relevant and statistically feasible and should be
established through cogent discussion and agreement between the sponsor and the Agency. The
quantity needs to be sufficiently small so that, from a clinical point of view, the investigational
56
Contains Nonbinding Recommendations
Draft — Not for Implementation
device can still be considered to be noninferior to the control as long as the advantage of the control
over the investigational device is smaller than delta. Additionally, the delta should not be so large,
that in a direct comparison with the previous standard of care (in this case, bare metal stents), the
new treatment could be noninferior to the active control, but no longer superior to a bare metal stent
(so-called outcome drift 45 ). To investigate whether the investigational device is noninferior to the
control, the appropriate null hypothesis is that the control is better than the investigational device by
at least delta, against the alternative hypothesis that the investigational device is not worse than the
control by delta. These two hypotheses are the essence of how FDA views noninferiority trials.
Although the non-inferiority trial design is a strategy that could be used when a placebo-controlled
study cannot be conducted, there are some limitations to the noninferiority study design that should
be considered prior to adopting this approach. For example, selection of an appropriate delta value,
while ideally based on prior data and expectations of performance, should be determined by what is
a clinically meaningful definition of a delta, agreed to by the clinical community and FDA. In
addition, the trial design and analysis plan should take into consideration the potential for outcome
drift.
3. Endpoints for DES Trials
Based on the definition of effectiveness (21 CFR 860.7), the most direct method of providing valid
scientific evidence of effectiveness is to select an appropriate clinical outcome and design a study to
evaluate a statistically significant and clinically meaningful treatment effect.
FDA recommends that definitions for outcomes of interest (death, MI, Target Lesion
Revascularization (TLR), Target Vessel Revascularization (TVR), stent thrombosis) be standardized
in the protocol. One potential set of definitions can be found in Cutlip et al., 46 although alternate
definitions may be proposed with a clinical justification.
Historically, the conventional intracoronary device study endpoint has typically been a
composite endpoint (e.g., target vessel failure (TVF), which is a composite of death, nonfatal
myocardial infarction (MI), and target vessel revascularization (TVR) after an index stenting
procedure). The paper by Cutlip et al. referenced above recommends the use of a patient-
oriented composite including all death, MI, and TVR and a device-oriented composite
including cardiac death, target vessel MI, and TLR. We recommend the use of the device-
oriented composite as a primary clinical endpoint. Other endpoints may be appropriate for
specific studies; a clinical justification should be provided for the endpoint selected.
Although a composite may not be the ideal primary endpoint, because the components have
different weights, the use of such a composite allows for trials of reasonable sample size to
be conducted. For example, a trial seeking to evaluate mortality would need tens of
45
Outcome drift can occur when successive generations of inferior devices are found to be non-inferior to the previous
generation as an active control, but might be inferior if tested against the original placebo treatment.
46
Cutlip et al., on behalf of the Academic Research Consortium. Circulation 2007:115;2344-2351. Clinical endpoints in
coronary stent trials: a case for standardized definitions.
57
Contains Nonbinding Recommendations
Draft — Not for Implementation
The initial DES approvals were based on a primary endpoint assessment at 9 months post-
implant. FDA currently believes that a 12-month primary endpoint, with a substantial
proportion of patients having 2-year data at the time of marketing application submission, is
critical to assess the potential for important adverse events such as stent thrombosis (and
related deaths and MIs) that may occur after 9 months. Patients in all trials to be used to
support approval of a PMA application should be consented at the time of enrollment for
follow-up to 5 years.
Imaging-derived measures of restenosis, such as percent diameter stenosis and late lumen
loss, are potentially powerful effectiveness endpoints. Such outcome measures have the
advantage of providing quantitative data for the comparison of specific parameters of stent
performance, such as suppression of neointimal hyperplasia. Furthermore, they can provide
additional effectiveness data, even in patients who have not developed a major clinical
adverse event, and consequently have the potential to increase the sensitivity of outcome
measures between treatments. Imaging endpoints are commonly measured as continuous
variables and this powerful discriminatory advantage can be apparent with sample sizes
considerably smaller than typically needed for clinical endpoints. However, the use of these
potential imaging measures as primary endpoints does not preclude the need for evidence of
safety through evaluation of a clinical endpoint, such as death, MI, and/or TLR, either
individually or as a composite.
FDA believes that use of an imaging endpoint as the sole primary effectiveness endpoint in
pivotal DES trials is currently acceptable only for certain second-generation DESs, such as
iterative modifications from currently approved DESs and/or indication expansion, in
specific patient populations or in specific vessel or lesion types. For a novel DES, clinical
studies performed to support regulatory approval should include at least one study of
sufficient size that has as its primary endpoint a clinical endpoint and is appropriately
powered for statistical demonstration of superiority or non-inferiority against an appropriate
control. See Section VIII.D for more discussion of next-generation DESs.
58
Contains Nonbinding Recommendations
Draft — Not for Implementation
one potential method. If co-primary endpoints are proposed for the trial, the selection of the
noninferiority margin for the clinical endpoint may be less conservative than when used as a
stand-alone endpoint, reflecting the fact that additional information from another parameter
(such as angiograph) is being evaluated. When using co-primary endpoints, FDA
recommends that adequate adjustments for correlation between the endpoints and
preservation of type I error be carefully considered. Study success using co-primary
endpoints is typically defined as meeting both endpoints. Appropriate definitions for
superiority and for selection of noinferiority margins should be discussed with the Agency
when the use of multiple endpoints is contemplated.
d. Secondary Endpoints
Separate from the primary endpoint chosen for effectiveness, we recommend collecting
additional vessel imaging information to evaluate healing and remodeling of the arterial wall,
including parameters such as stent apposition, aneurysm formation, edge effects, and
quantification of intimal proliferation, especially at the proximal and distal borders of an
implanted DES. Quantitative coronary angiographic (QCA) analyses should report stent,
lesion, and analysis segment parameters to assess the importance of any edge effects caused
by the drug. The angiographic analysis should also include review and analysis for stent
fracture; use of a grading system such as that described by Rocha-Singh et al., 47 may be
helpful for reporting the incidence and type of fracture, if observed. Side branch occlusion,
when observed, should also be reported.
The secondary endpoints will, in most cases, not be descriptive and exploratory, not leading
to additional claims. If a formal comparison of treatment arms for a secondary endpoint is
desired, formal null and alternative hypotheses should be developed and pre-specified in the
protocol. If no pre-specified hypotheses are included in the protocol, p-values for such
comparisons will not be appropriate and should not be presented in labeling. If analyses
beyond descriptive statistics are planned for secondary endpoints, appropriate steps should be
taken to adjust for multiple comparisons and to preserve Type I error. Sponsors with studies
ongoing prior to the issuance of this guidance should discuss with FDA an appropriate
approach for presentation of such analyses in the labeling.
When a DES incorporates an unstudied drug, the data from a sufficient number of patients
exposed to the new DES should be collected for submission in the PMA. The number of
patients should be large enough to enable the detection with adequate precision of low
frequency adverse events (i.e., those occurring at a rate of 1 percent or less) that may be
associated with the unstudied drug. A single study or multiple studies (both randomized
trials and single-arm registry studies) can be used to complete this population.
47
Rocha-Singh, et al, Performance goals and endpoint assessments for clinical trials of femoropopliteal bare nitinol
stents in patients with symptomatic peripheral arterial disease. Catheter Cardiovasc Interv 2007;69(6):910-919
59
Contains Nonbinding Recommendations
Draft — Not for Implementation
Also, certain additional safety data beyond what are typically collected in a stent trial should
be obtained and provided in the PMA to allow for analysis of potential drug-related adverse
events. The specific safety data to be collected will generally be specific to the drug
incorporated on the stent; however, the following are examples of typically requested
information:
• Liver enzyme values pre- and post-procedure and at appropriate follow-up
intervals
• Hypersensitivity reactions (definition should be pre-specified) including
symptoms, signs, and relevant laboratory values, treatment, and clinical course
• White blood cell counts to document the incidence of leukopenia
• EKG parameters
• EKG changes, particularly QT intervals
• Concomitant medications
Sponsors with such DES are encouraged to meet with FDA prior to beginning clinical trials
to ensure that case report forms capture appropriate cardiac and non-cardiac safety
information.
In a randomized controlled trial, the use of study blinding, or masking, further reinforces the
integrity of the random allocation of patient assignment and assessment of treatment effect.
In a superiority RCT study design using a DES and its corresponding bare metal stent, a
triple-blinded (i.e., patient, physician and monitoring committee are all blinded) study design
is logistically possible because of the physically similar appearance of the DES and bare
metal stents. However, for some medical devices, designing a double-blinded (i.e., patient
and physician are blinded to treatment assignment) or triple-blinded RCT can be impractical
and logistically impossible because of the physical characteristics and/or the mode of action
of the product (e.g., a DES versus coronary artery bypass grafting (CABG)). For
noninferiority study designs that are evaluating a DES with different platforms, the DES
might have different physical characteristics (e.g., radiologically and/or visually different in
appearance), making such study blinding logistically difficult to implement. Because certain
individuals involved in stent handling/implantation at the time of the index procedure will
have knowledge of treatment assignment.
Nonetheless, because there is a potential for considerable investigator and/or patient bias
introduced by knowledge of treatment assignment, possibly confounding study outcomes and
diminishing the scientific validity of the study, the study design should incorporate blinding
to the maximum extent possible, maintaining the blind for patients (single-blind), follow-up
study investigators, and study staff to minimize the potential for bias and confounding. In
addition, increasing the objectivity of study parameters as much as possible and including
special analytical methods to evaluate for the potential influence of bias in study outcome are
potential ways to maximize the scientific validity of study design.
60
Contains Nonbinding Recommendations
Draft — Not for Implementation
Many of the novel technologies employed in a DES have never been used previously in the
same combinations or anatomic locations in human beings. This fact raises new questions of
safety for participants in investigational DES trials. Given that most DESs under
development are intended to be permanent implants and that safe and reliable retrieval of
deployed stents is generally not possible, a heightened and constant vigilance during the
conduct of a DES trial is necessary. With this in mind, FDA strongly recommends the use of
data monitoring committees (DMC, also called data safety monitoring boards, or DSMBs)
for DES studies to keep track of and evaluate significant adverse events, including stent
thrombosis, in real time (i.e., as the study enrollment progresses). 48 Sponsors are responsible
for ensuring proper monitoring of the investigations (21 CFR 812.40), and must select
monitors qualified by training and experience to monitor the investigational study (21 CFR
812.43(d)). Before the study begins, the DMC/DSMB charter should have an adequate
monitoring plan (e.g., number of predetermined meetings, timing of reports, appropriate
stopping rules, correspondence to FDA as appropriate) in place to adequately ensure that
patients are not subjected to undue risk. For sponsors conducting multiple trials with the
same investigational DES, FDA recommends that sponsors as part of their obligation to
monitor the studies, use the same DMC/DSMB for both studies or have a super-DMC/DSMB
that communicates with the DMC for each trial be considered. If this is not possible, the
sponsor should ensure that the DMCs/DSMBs for each of the studies communicate
frequently and regularly exchange safety information and ensure that all members of the
committee are apprised of the global safety data for the investigational DES.
FDA strongly recommends that interpretation of data from tests such as angiograms, IVUS,
and ECGs be performed by independent core labs and that blinded adjudication of clinical
events be conducted by a clinical events committee (CEC Clinical adjudication committees
should be independent of core lab analysis centers to avoid potential bias. .
The proposed protocol should include a comprehensive statistical analysis plan with prospectively
defined methods to address the following:
• Study hypotheses
• Sample size calculation
• Blinding
• Number of proposed study centers
• Study success criteria
• Effectiveness patient populations (e.g., intent-to-treat, evaluable)
• Pooling of data
• Covariate adjustments
48
Guidance for clinical trial sponsors on Establishment and Operation of Clinical Trial Data Monitoring Committees,
March 2006.
61
Contains Nonbinding Recommendations
Draft — Not for Implementation
• Stratification
• Protocol deviations
• Handling drop-outs and methods to address missing data
• Analysis plan and statistical methods
• Data auditing
1. Analysis Cohorts
The intention-to-treat population, which is defined as the cohort of all patients randomly assigned to
treatment in an RCT design, is usually the preferred population for superiority studies. Intention-to-
treat analysis allows for the evaluation of all patients who enroll in the study, even though some may
not complete the study (e.g., patients who are, for any reason, lost to follow-up, drop-outs, or
terminated by investigator). In an RCT design, the intention-to-treat principle means that any
comparison of the treatments is based on comparison of the outcome results of all patients in the
treatment groups to which they were randomly assigned. Within the protocol, the sponsor should
prospectively specify the analysis plans that will account for patients who do not complete the study.
The sponsor should also present analysis of the per protocol patient cohort (i.e., patients who enter
and complete the study according to protocol) and the as-treated patient cohort (recognizing such
analyses are subject to bias).
Comparison of outcomes on the basis of intention-to-treat, per protocol, and as-treated patients
allows assessment of outcome robustness. Analysis details should be prospectively agreed to by the
sponsor and FDA.
Pivotal studies of DES should be conducted at multiple investigational sites. Additionally, there can
be advantages to conducting multiple clinical studies of the same DES. Potential advantages to
combining data from different studies include having the ability to evaluate DES performance across
a broader population than can be achieved by one study and could increase generalizability of study
results because of wider demographic and geographic inclusion. Furthermore, demonstration of
comparable DES performance across different investigational sites and studies can permit more
robust conclusion of product safety and efficacy. However, when planning to conduct clinical
studies at multiple investigational centers, or in centers OUS (outside the United States), an analysis
of poolability of data should be included in the prospective analysis plan.
When FDA considers foreign data as supportive evidence for U.S. product approval, a key
consideration in assessing the applicability of OUS studies in support of product safety and
effectiveness is to evaluate the generalizability of the OUS studies to the patient population and to
medical practice in the United States. Factors that FDA considers include, for example,
62
Contains Nonbinding Recommendations
Draft — Not for Implementation
These factors have the potential to affect DES performance in terms of both safety and effectiveness.
Some examples of key factors that should be addressed when considering the poolability of results
and extrapolating study results to those expected in the United States can be found in the stand alone
companion document. 49
Whether studies have been conducted solely in the United States or both in or out of the United
States, statistical analysis should examine the homogeneity of demographic and procedural
covariates across centers and geographical regions. Evaluation of interactions between treatment
and region is recommended. Furthermore, outcome comparability should be examined after
adjustment for covariate differences, using multivariate regression modeling and propensity scoring
methodology. In addition, sensitivity analysis should be performed to verify the robustness of any
statistical modeling using pooled data.
FDA is willing to comment informally on OUS study protocols through the pre-submission process.
Such comments may increase the likelihood that these data can be used to support a PMA
application.
Optimal duration of antiplatelet therapy and use of glycoprotein IIb/IIIa inhibitors and direct
thrombin inhibitor treatments in DES patients are currently unclear and may significantly affect
clinical outcomes. Consequently, to minimize confounding variables in the interpretation of the
study results, a uniform regimen of intra- and postprocedure concomitant medications should be
used. Careful consideration should be given to the optimal dosage and duration of antiplatelet
therapy for DES postimplantation, given the delay in endothelialization within DES compared to that
of bare metal stents and subsequent concerns regarding stent thrombosis due to premature
discontinuation of antiplatelet therapy.
At the December 2006 Circulatory System Devices Advisory Panel meeting on DES thrombosis ,
the Panel recommended that the labeling for the two approved DES include reference to the
AHA/ACC/SCAI practice guidelines. FDA agreed with this recommendation and both approved
DES Instructions for Use include this information. For this reason, for trials that use the CYPHER
stent or TAXUS stent as the control DES, we currently recommend that the prescribed antiplatelet
therapy follow the AHA/ACC/SCAI guidelines 50 ; that is, patients should receive aspirin and a
minimum of 3 (CYPHER) or 6 months (TAXUS) of clopidogrel with therapy extended to 12 months
in patients at a low risk of bleeding. Despite the desire to have administration and use of dual
antiplatelet therapy, circumstances will cause some patients to have different regimens, and FDA is
particularly interested in how differences in duration affect patient outcome. Therefore, patients
should be carefully monitored and case report forms should be designed to capture compliance with
prescribed antiplatelet therapy and significant bleeding complications over the course of the trial.
49
See draft guidance for industry and FDA staff on Coronary Drug-eluting Stents – Nonclinical and Clinical Studies:
Companion Document,” published together with this document.
50
Available at http://www.acc.org/qualityandscience/clinical/guidelines/percutaneous/update/index.pdf
63
Contains Nonbinding Recommendations
Draft — Not for Implementation
Eventual product labeling should include both the prescribed antiplatelet therapy and patient
compliance with that therapy as experienced in the clinical trials and should clearly specify the risks
of premature antiplatelet medication discontinuation.
Although nonclinical and clinical testing of DESs provide invaluable information on the short-term
safety and effectiveness of these products in a select patient population, such as that typically found
in the clinical trial setting, much information on the performance and safety profile of a DES can be
obtained only when the product moves into the larger, more diverse patient population after
marketing.
For purposes of regulatory approval, the current primary endpoint data for DES studies should be
collected over a period of approximately 12 months after implantation of the DES. However, DES
study length should be viewed in terms of the entire follow-up, which should extend through a 5-
year clinical follow-up period. Although the 12-month postimplantation endpoint might be
acceptable for a PMA submission, the study is not considered complete until study patients have
completed their long-term clinical follow-up as described in the protocol. At a minimum, this would
include annual follow-up telephone evaluations and, preferably, annual study visits, for five years in
a significant cohort of patients enrolled in the pivotal, feasibility, and/or any additional clinical
studies conducted to support product approval. During the long-term follow-up phase, the
occurrence and sequelae of late phenomena, such as incomplete stent apposition, late stent
thrombosis, and polymer compatibility issues, are important parameters that should be evaluated.
The actual duration of dual antiplatelet therapy and any interruptions should be captured as well (see
Section C above for objectives related to antiplatelet therapy).
At the time of PMA submission, all available long-term follow-up from the pivotal and
supplementary clinical studies should be provided to demonstrate the chronic performance of the
DES. Additionally, as part of the PMA review, the applicant is also required to submit a
bibliography of all published reports and other information relevant to an evaluation of the safety
and effectiveness of the device (see 21 CFR 814.20(b)(8)).
During the PMA review, a three-month update of any additional clinical data must be submitted
(21 CFR 814.20(e)). The applicant must submit new information learned about the device from
ongoing or completed studies that may reasonably impact an evaluation of the safety and
effectiveness of the product or that may reasonably affect the draft labeling. Note that when
reasonably limited in scope, this update would be considered a minor amendment to the PMA.
Additional (i.e., later) endpoint evaluations, a significant increase in the number of evaluable
patients, or new analyses may be considered a major amendment requiring significant review. In
addition, as a condition of approval for a PMA application, applicants are required to submit updated
clinical reports to the Agency (§ 814.82 and 814.84)
64
Contains Nonbinding Recommendations
Draft — Not for Implementation
follow-up to ensure that appropriate information is being collected and to minimize bias since
treatment assignment may be known upon disclosure of the primary endpoint results.
A. Postapproval Studies
Therefore, in addition to postapproval follow-up of clinical outcomes from the patients enrolled in
the preapproval clinical trials, the Agency will generally require the collection of additional
postapproval data for a DES (§ 814.82(a)(2)). Serious but rare DES-related adverse events that
might only be identified in a postapproval period include late stent thrombosis, drug interactions,
unforeseen complications of multivessel or overlapping stent placement, and experience with a DES
in different patient demographic subsets not adequately represented in preapproval studies (i.e., real
world use). A proposed postapproval study protocol should be included in the PMA application.
The postapproval study should have two primary goals: assessment of the rate of stent thrombosis
and assessment of the rate of cardiac death plus MI. As discussed above, the postapproval data
collected on currently approved DESs have signaled a potential increase in late stent thrombosis
after one year compared to bare metal stents. However, it is not known if this rate plateaus or
continues to increase over time, nor is the impact of stent thrombosis on rates of cardiac death and
MI completely understood. Therefore, one primary endpoint of the postapproval study should be
the rate of stent thrombosis after one year. As stent thrombosis is closely associated with cardiac
death and MI, a second primary endpoint of the postapproval study should be a comparison of the
rate of cardiac death and MI between the new DES and the control stent used in the pivotal study. To
gain a better understanding of these risks in the setting of actual clinical use of the product, FDA
recommends that postapproval data be collected on a series of patients who are consecutively
enrolled to avoid the introduction of selection bias.
A sufficient number of patients should be enrolled to confirm that the upper bound of the one-sided
95 percent confidence interval around the observed rate of stent thrombosis between 12 and 24
months, 24 and 36 months, 36 and 48 months, etc. is ≤ 1 percent with at least 80% probability for
patients treated in accordance with the labeled indication. The total study sample size should be
sufficient to ensure a sufficient number of patients treated in accordance with the labeled indication
are available for analysis.
To evaluate the rate of cardiac death and MI, we suggest that the cohort of patients treated in
accordance with the labeled indications be pooled with the preapproval pivotal trial to reach a
sample size sufficiently large to provide adequate power to compare the rates of cardiac death and
target vessel MI for the new DES and the control stent used in the pivotal study and to rule out an
increased risk. This cohort of postapproval patients may be in a single-arm or randomized study,
and data pooling may be approached from either a frequentist or Bayesian perspective.
65
Contains Nonbinding Recommendations
Draft — Not for Implementation
Additionally, postapproval studies to date have demonstrated that routine clinical use of DESs
typically includes the treatment of patients outside of the labeling indications, including higher risk
patient and lesion subsets. Based on this previous experience, FDA recognizes that a postapproval
study of consecutively enrolled patients is likely to include patients representing a broader use of the
product and recommends that data from such patients be analyzed separately to better understand
whether significant safety issues exist in the treatment of these patients.
All patients should be consented for five years of follow-up. If stent thrombosis rates are
demonstrated to plateau or decrease in prior years, shorter follow-up may be sufficient.
Alternatively, if stent thrombosis rates continue to increase, longer term follow-up or specific
labeling changes may be appropriate.
The statistical plan should include planned descriptive statistics on certain subgroups of interest
including:
Demographics
• Age (age < 65 years; age ≥ 65 years)
• Sex (male, female)
• Race and ethnicity
Patient characteristics
• Patients with diabetes, further characterized as insulin-requiring or noninsulin-requiring
• Patients with renal insufficiency, further characterized as creatinine clearance (rCl) using the
Cockcroft-Gault equation (CrCl > 60 mL/min, CrCl ≥ 30 and ≤ 60 mL/min, CrCl < 30
mL/min)
• Degrees of left ventricular (LV) dysfunction (ejection fraction < 30%, 30-40%, > 40%)
• Patients with 3 vessel disease
66
Contains Nonbinding Recommendations
Draft — Not for Implementation
• Patients with 2 vessel disease including proximal left anterior descending coronary artery
disease
Lesion characteristics
• Lesions in the setting of acute ST elevation myocardial infarction (STEMI)
• Percutaneous coronary interventions within 36 hours of non-STEMI ACS
• Lesion length (≤ 20 mm, 21-30 mm, 31-40 mm, > 40 mm)
• Vessel diameter (2.0 - ≤ 2.5 mm; 2.6 – 2.9 mm; 3.0 - ≤ 3.5 mm, and > 3.5 mm)
• Ostial lesions
• Bifurcation lesions
• Trifurcation lesions (i.e., left main coronary artery, left circumflex coronary artery, left
anterior descending artery, and ramus intermedius)
• Thrombus-containing lesions
• Lesions with residual dissection post stenting
• Left main coronary artery (LMCA) lesions
• Include whether disease was ostial, mid, or terminal and whether or not it involved the
ostial LAD +/- LCFX
• Chronic total occlusions (CTO)
• Saphenous vein grafts (SVGs)
• Arterial grafts (internal mammary artery, radial artery, gastroepiploic artery)
• Post-brachytherapy
• Instent restenosis (ISR) (BMS)
• Instent restenosis (ISR) (DES)
• Overlapping BMS
• Overlapping DES
• Overlapping BMS and DES
• Non-overlapped multiple stents (in the same vessel or in different vessels)
• Intravascular ultrasound guidance for initial stent deployment
Case report forms should capture patient compliance with prescribed antiplatelet therapy and
significant bleeding complications.
For patients who experience stent thrombosis, in addition to the above characteristics, the following
additional information should be reported:
67
Contains Nonbinding Recommendations
Draft — Not for Implementation
• Any discontinuation of Plavix and/or aspirin and whether or not there was premature
discontinuation of these medications
Sponsors should contact the CDRH review division for more information on the use of these studies
to support additional indications. For more information on postapproval studies, see the CDRH
guidance for industry and FDA staff on Procedures for Handling Post-Approval Studies Imposed by
PMA Order.
Because a DES is regulated under the device provisions of the Act, the adverse event and device
defect reporting requirements for devices are applicable. 51 The medical device reporting (MDR)
requirements mandate that manufacturers report to the Agency (1) all device-related deaths and
serious injuries and (2) all malfunctions of the device or similar device that would be likely to cause
or contribute to a death or serious injury if the malfunction were to recur (21 CFR 803.3).
51
Each constituent part of a combination product is governed by a different set of postmarket reporting requirements
(for drugs, 21 CFR Parts 310 and 314, and for devices 21 CFR Part 803). This is the case for a DES product. The
Agency has announced its intention to issue a Proposed Rule, Postmarket Safety Reporting for Combination Products
that would clarify the postmarketing safety reporting requirements for combination products (72 Fed. Reg. No. 82,
22515 (2007). The proposed rule would provide a framework for the reporting of adverse events for combination
products and specify the circumstances in which following one set of postmarket safety reporting regulations (e.g., 21
CFR 803) generally would meet the requirements of another set and the circumstances in which these requirements
would be supplemented with specific reporting provisions applicable to the constituent part of the combination product.
68
Contains Nonbinding Recommendations
Draft — Not for Implementation
69
Contains Nonbinding Recommendations
Draft — Not for Implementation
• The manufacturer takes or would be required to take action under section 518 or 519(f) of the
Act as a result of the malfunction of the device or other similar devices.
For more information see the Medical Device Reporting (MDR) Web site at:
http://www.fda.gov/cdrh/mdr/, and you may direct questions regarding MDRs to the Reporting
Systems Monitoring Branch at 240-276-3464.
Adverse events reported through MDR are shared with CDER so that drug-related aspects of
postapproval adverse events reported to CDRH can be evaluated.
C. Peri-Approval Studies
FDA has typically required postapproval studies for DESs. However, when the postapproval study
protocol was approved only at the time of the PMA approval, FDA found that there were significant
delays in beginning enrollment in the study due to delays in awaiting IRB review and approval.
There was also confusion on the part of some IRBs regarding the rationale for an additional study of
an approved product. The delays in enrollment and data collection in this scenario meant that an
important source of postmarket data was unavailable to the manufacturer and to FDA for multiple
months following PMA approval.
To minimize this delay, FDA has encouraged PMA applicants to submit the postapproval study
protocol earlier in the PMA review process. If FDA has reached the conclusion that the PMA will
be approved (e.g., only minor issues such as labeling are pending), the postapproval study protocol
can be approved in advance of the PMA approval. A protocol for such a peri-approval study can be
submitted as an IDE supplement. Upon IDE approval, the study can begin enrolling under the IDE
with a prespecified patient limit, with the remainder of patients enrolled after PMA application
approval. Consequently, the peri-approval study does not obviate the need for the collection of
information after the initiation of marketing. The IDE approval does, however, enable a sponsor to
ensure that IRB review/approvals are in place and selected sites are eligible for active enrollment of
patients at the time of PMA application approval.
FDA strongly encourages sponsors to select a broad cross-sectional distribution of institutions (e.g.,
geographic location, private versus public versus academic hospitals, volume of procedures) to
address generalizability of the study findings. The main impetus for the peri-approval approach has
been to facilitate the enrollment of patients and streamline completion of the study so that both the
FDA and the applicant can assess patient safety in a real-world scenario in a timely manner to
support the total product life cycle of the DES.
70
Contains Nonbinding Recommendations
Draft — Not for Implementation
DES candidates may employ a range of new and old technologies, making classification of a next-
generation DES dependent on the specific components and/or modifications to the product. Unlike
second-generation bare metal stents, in which modifications in a product line were limited to either
the stent substrate (e.g., geometry, such as strut thickness, cell configuration, material), or delivery
catheter, for DES, manufacturers should carefully consider that planned modifications to the stent
substrate or polymer carrier may have unintended or unanticipated effects on other product
performance parameters (e.g., changes in drug density, total drug load, elution kinetics) and on the
overall safety and effectiveness of the finished product. Additionally, if a sponsor wants to make a
manufacturing change in the coating process, depending on the change, it may be necessary to
perform additional studies to ensure safety and/or effectiveness for the modified product if the rate
and/or extent of drug elution is materially affected.
Some examples of questions for the sponsor or applicant to address regarding design modifications
to a DES that may affect rate and/or extent of drug elution include, but are not limited to, the
following:
• Is this a first generation DES, a combination of new and old technologies, or essentially a
design iteration?
If the answer to “is this a first generation DES?” is no, some additional questions to address
include:
• Which components of the DES system have stayed the same and/or which have been
changed? Be sure to consider both intentional and unintentional changes that may have
occurred.
• If the stent substrate has changed, what specifically has been altered (e.g., stent substrate
material only (from 316L to CoCr); geometry elements, such as strut thickness, which can
lead to differences in surface area; and/or a change in the drug density and/or drug content)?
• Has the delivery catheter been modified (e.g., distal tip or other elements)?
• Is the drug formulation the same or different (e.g., change in polymer/drug ratio, increased or
decreased drug content)?
• Have any of these modifications resulted in alterations to the release kinetics (e.g., amount or
significant modifications in profile)?
• Have there been any modifications in any critical manufacturing parameters (e.g., coating
application, new sources of heat or humidity, sterilization method)?
• Does the new product still meet the original product specifications?
• How robust are the in vitro test methods and quality control specifications used to assess
product variability to ensure product quality and consistency?
The significance of the changes in a DES system for a second generation DES will directly influence
the amount of additional nonclinical and/or clinical testing needed to support the safety and efficacy
71
Contains Nonbinding Recommendations
Draft — Not for Implementation
of a modified DES. FDA encourages sponsors and applicants to discuss with the Agency proposed
changes to their DES and appropriate testing to validate those changes.
X. COMPANION DOCUMENT
To facilitate the use of this guidance, a stand alone companion document is available to be used
together with this guidance. It is posted with this guidance on the FDA Web site. The companion
document contains the following:
72
Contains Nonbinding Recommendations
Draft — Not for Implementation
APPENDIX A
Below is an example of a regulatory specification table for the finished DES product.
Shelf Life :
NMT 3500 particles ≥ 10 μm
NMT 300 particles ≥ 25 μm
Endotoxins NMT 0.5 EU/mL LAL (USP <85>)
Sterility or package Pass USP <71>
integrity
Drug Release 10% - 20% 2 hours USP <724>
20% - 50% 4 hours
40% - 70% 8 hours
> 80% 24 hours
1
In the table above, all numerical limits and the time points in the drug release test are for illustrative purposes only.
2
Relative retention time
3
Example of an attribute for which tighter release limits are assigned in order to maintain a safety margin so that
the product remains within the approved shelf life acceptance criteria for that attribute.
73
Contains Nonbinding Recommendations
Draft — Not for Implementation
74
Contains Nonbinding Recommendations
Draft — Not for Implementation
GLOSSARY OF TERMS
Acceptance criteria: Numerical limits, ranges, or other suitable measures for acceptance of
results of analytical procedures (see ICH guidance Q6A)
Acute: Refers to any time up through expansion and deployment of the DES
Chronic refers to any time after assessment of the initial stent deployment in a simulated vessel
throughout the lifetime of the implant.
Adhesion: The degree of attachment between two different surfaces, such as a coating or film
and the underlying material.
Area under curve (AUC): PK parameter, area under the blood concentration-time curve
Balloon expandable stent: A stent that is expanded by a balloon. The diameter of the stent
increases as the balloon diameter increases. The stent remains expanded after deflation of the
balloon.
Bare metal stent (BMS): An intravascular stent that is not coated with either a polymer or drug.
Traditional materials for BMSs include 316L stainless steel and cobalt chromium alloy.
Batch: A specific quantity of a drug or other material that is intended to have uniform character
and quality, within specified acceptance criteria, and is produced according to a single
manufacturing order during the same cycle of manufacture (21 CFR 210.3(b)(2)). See also “lot.”
Bias (statistical and operational): The systematic tendency of any factors associated with the
design, conduct, analysis, and evaluation of the results of a clinical trial to make the estimate of a
treatment effect deviate from its true value. Bias introduced through deviations in conduct is
referred to as operational bias. The other sources of bias listed above are referred to as
statistical bias. 52
Clinical batch: Batch used to support the efficacy, safety, bioavailability, or bioequivalence of a
product
Coating: The drug carrier (usually polymeric, but not limited to such), the drug itself if it is
solely coated onto the stent platform, any other coating, or the drug carrier even if it is
incorporated onto the stent in a geometry other than a coating.
52
ICH Guidance E9 Statistical Principles for Clinical Trials
75
Contains Nonbinding Recommendations
Draft — Not for Implementation
Combination product: A product (defined in further detail in 21 CFR 3.2(e)) comprised of two
or more different types of regulated entities (i.e., drug-device, drug-biologic, device-biologic, or
drug-device-biologic products).
Component: For a drug: Any ingredient intended for use in the manufacture of a product,
including those that may not appear in such product (21 CFR 210.3(b)(3)).
Component: For a device: any raw material, substance, piece, part, software, firmware,
labeling, or assembly which is intended to be included as part of the finished, packaged, and
labeled device (21 CFR 820.3(c)).
Device history record: (DHR) a compilation of records containing the production history of a
finished device (21 CFR 820.3(i))
Double-blinded: A double-blind trial is one in which neither the subject nor any of the
investigators or sponsor staff involved in the treatment or clinical evaluation of the subjects are
aware of the treatment received. This includes anyone determining subject eligibility, evaluating
endpoints, or assessing compliance with the protocol; blinding is maintained throughout the
conduct of the trial.53
Blinding, or masking, is intended to limit the occurrence of conscious and unconscious bias in
the conduct and interpretation of a clinical trial arising from the influence that the knowledge of
treatment may have on the recruitment and allocation of subjects, their subsequent care, the
attitudes of subjects to the treatments, the assessment of end-points, the handling of withdrawals,
the exclusion of data from analysis, and so on. 54
Drug-eluting stent (DES): A combination product consisting of both drug and device
components. The device component consists of an intravascular stent platform that is used not
only for radial support, but also as a vehicle for the delivery of an active pharmaceutical agent or
drug. The drug component is commonly incorporated and released from a polymeric carrier,
either a single polymer or a combination of polymers, which is physically or chemically adherent
to the stent substrate. The purpose of the polymer carrier is to allow for adequate deposition of
the drug onto the stent surface as well as to influence the release kinetics of the drug from the
53
ICH Guidance E9 Statistical Principles for Clinical Trials
54
ICH Guidance E9 Statistical Principles for Clinical Trials
76
Contains Nonbinding Recommendations
Draft — Not for Implementation
stent surface. The DES is mounted onto a stent delivery system to deliver the stent to its final
intended location in the vasculature.
Drug product: A finished dosage form, for example, tablet, capsule, or solution, that contains a
drug substance, generally, but not necessarily, in association with one or more other ingredients
(21 CFR 314.3(b)).
Established name: The designated FDA official name, the compendial name, the USAN
Council name, or the common or usual name (section 502(e)(3) of the Act and 21 CFR 299.4).
Ordinarily, the established name of a drug will be the compendial name. However, FDA may
designate an established name in cases where a monograph does not exist (see the CDER Data
Standards Manual).
Excipient: Any component other than the drug substance(s) present in the finished product.
Extended release: Products that are formulated to make the drug available over an extended
period after implantation.
Formulation: The qualitative and quantitative composition of the finished product. This is
often called the composition statement.
Four corners: Refers to a 2 x 2 factorial of the largest and smallest diameters and
lengths for each stent design.
Generalizability, generalization: The extent to which the findings of a clinical trial can be
reliably extrapolated from the subjects who participated in the trial to a broader patient
population and a broader range of clinical settings. 55
Glass transition temperature (Tg): The temperature at which a polymer changes from glassy
to elastomeric behavior.
Independent data monitoring committee (IDMC) (data and safety monitoring board,
monitoring committee, data monitoring committee): An independent data monitoring
committee that may be established by the sponsor to assess at intervals the progress of a clinical
55
ICH Guidance E9 Statistical Principles for Clinical Trials
77
Contains Nonbinding Recommendations
Draft — Not for Implementation
trial, the safety data, and the critical efficacy endpoints, and to recommend to the sponsor
whether to continue, modify, or stop a trial. 56
Intention-to-treat principle: The principle that asserts that the effect of a treatment policy can
be best assessed by evaluating on the basis of the intention to treat a subject (i.e., the planned
treatment regimen) rather than the actual treatment given (e.g., results from a patient who
discontinues a treatment are counted in the treatment group). It has the consequence that subjects
allocated to a treatment group should be followed up, assessed, and analyzed as members of that
group irrespective of their compliance with the planned course of treatment. 57
Intravascular stent: For this guidance, an intravascular stent is a synthetic tubular structure
intended for permanent implantation in the native coronary vasculature. The stent is designed to
provide mechanical radial support after deployment; this support is meant to enhance vessel
patency over the life of the stent. Once the stent reaches the intended location, it is expanded by a
balloon or self-expanding mechanism.
Lot: Or batch means one or more components or finished devices that consist of a single type,
model, class, size, composition, or software version that are manufactured under essentially the
same conditions and that are intended to have uniform characteristics and quality within
specified limits (21 CFR 820.3(m)). (Note that a similar definition is provided within the CGMP
regulations: A batch, or a specific identified portion of a batch, having uniform character and
quality within specified acceptance criteria. In the case of a product produced by continuous
process, it is a specific identified amount produced in a unit of time or quantity in a manner that
ensures its having uniform character and quality within specified acceptance criteria (21 CFR
210.3(b)(10)).)
Master file: A reference source submitted to FDA, which may include drug master files (DMF),
device master files (MAF), etc. A master file may contain detailed information on a specific
manufacturing facility, process, methodology, or component used in the manufacture,
processing, or packaging of a drug (21 CFR 314.420) or a medical device (21 CFR 814).
56
ICH Guidance E9 Statistical Principles for Clinical Trials
57
ICH Guidance E9 Statistical Principles for Clinical Trials
78
Contains Nonbinding Recommendations
Draft — Not for Implementation
Master production record: A record containing the method of manufacture of the product,
including, in part, the master formula of defined size, complete manufacturing and control
instructions, in-process tests and acceptance criteria, equipment and operating parameters, yield
and yield reconciliation calculations, and provisions for packaging and labeling (see 21 CFR
211.186(b)) See also “Device history record.”
Molecular weight (MW) (of a polymer): Weight of an average polymer molecule. The two
most popular expressions of molecular weight of polymers are number-average molecular
weight (Mn) and weight-average molecular weight (Mw). Mn is the total weight of all the
polymer molecules in a sample, divided by the total number of polymer molecules in a sample.
This number represents the average weight of a chain, Mi, weighted according to number
fraction of each component i. Mw is the average molecular weight of a chain, Mi, weighted
according to weight fractions of each component i.
No Observed Adverse Effect Level (NOAELNOAEL means the highest dose level that does
not produce a significant increase in adverse effects. The NOAEL can serve as the starting point
for determining a reasonably safe starting dose of a new drug in healthy human volunteers.
Studies to determine the NOAEL by examining at least two different species are needed to
identify the starting dose for intravenous human studies (see guidance for industry Estimating the
Maximum Starting Dose in Initial Clinical Trials for Therapeutics in Adult Health Volunteers).
The duration of an animal study is determined by the duration of drug elution from the stent.
The minimum duration should be 2 weeks for a nonpolymerized drug, which is considered a
single dose. See the guidance for industry Single Dose Acute Toxicity Testing for
Pharmaceuticals and M3 Nonclinical Safety Studies for the Conduct of Human Clinical Trials
for Pharmaceuticals, for more information. 58
Noninferiority trial: A trial with the primary objective of showing that the response to the
investigational product is inferior to a comparative agent by more than a defined amount (the
noninferiority margin).
Novel excipient: An ingredient used for the first time in a human drug or combination product
in the United States or in a new route of administration.
Packaging system: The sum of packaging components that together contain and protect the
product. This includes primary packaging components and secondary packaging components, if
the latter are intended to provide additional protection to a DES.
Partition coefficient: The ratio of the concentration of a chemical species in one environment to
its concentration in another environment.
58
In December 2002, the Agency issued a draft guidance for industry and reviewers Estimating the Safe Starting
Dose in Clinical Trials for Therapeutics in Adult Healthy Volunteers. Once finalized, it will represent the Agency's
current thinking on this topic.
79
Contains Nonbinding Recommendations
Draft — Not for Implementation
Per protocol set (valid cases, efficacy sample, evaluable subjects sample): The set of data
generated by the subset of subjects who complied with the protocol sufficiently to ensure that
these data would be likely to exhibit the effects of treatment according to the underlying
scientific model. Compliance covers such considerations as exposure to treatment, availability
of measurements, and absence of major protocol violations.59
Pharmacodynamics: The study of the biochemical and physiological effects of drugs (and/or
metabolites) on the body and the mechanisms of drug action, including the characterization of
the relationship between the drug exposure and pharmacologic effects (efficacious and toxic),
and the factors influencing such relationships. Often, the time course of these effects is also
described.
Primary stability data: Data on the finished product stored in the proposed package for
marketing under storage conditions that support the proposed shelf life
Quality: The suitability of a DES for its intended use. This term includes such attributes as the
identity, content, purity, and potency.
Specification: The quality standard (i.e., tests, analytical procedures, and acceptance criteria)
provided in an application to confirm the quality of drug substances, products, intermediates, raw
materials, reagents and other components including packaging system, and in-process materials.
A specification sheet includes the list of tests, references to analytical procedures, and
acceptance criteria.
Statistical analysis plan: A statistical analysis plan is a document that contains a more technical
and detailed elaboration of the principal features of the analysis described in the protocol, and
includes detailed procedures for executing the statistical analysis of the primary and secondary
variables and other data. 60
Stent platform: The component of the DES that provides mechanical structural support when
deployed in a vessel and is usually metallic and either balloon expandable or self-expanding.
Stent delivery system: A stent delivery system delivers a stent through the vasculature to its
intended target site and then deploys the stent. A stent delivery system for a balloon expandable
stent consists of a balloon catheter. Self-expanding stent delivery systems may or may not
include a balloon.
59
ICH Guidance E9 Statistical Principles for Clinical Trials
60
ICH Guidance E9 Statistical Principles for Clinical Trials
80
Contains Nonbinding Recommendations
Draft — Not for Implementation
Studied drug: a molecular entity that has been previously approved or studied under IND (i.e.,
has an approved NDA or ANDA, or has undergone human clinical studies under IND)
Superiority trial: A trial with the primary objective of showing that the response to the
investigational product is superior to a comparative agent (active or placebo control). 61
United States Pharmacopeia (USP): The United States Pharmacopeia (USP) is the official
public standards-setting authority for all prescription and over-the-counter medicines, dietary
supplements, and other healthcare products manufactured and sold in the United States.
Unspecified degradation product: A degradation product that is not included in the list of
specified degradation products
Unstudied drug: a molecular entity that has not been approved for use in humans, or that does
not have human clinical study information available
61
ICH Guidance E9 Statistical Principles for Clinical Trials
81
Contains Nonbinding Recommendations
Draft — Not for Implementation
BIBLIOGRAPHY
The following documents have either been referenced in this guidance or will be of interest to
DES applicants and sponsors. They are grouped by document type and listed in alphabetical
order.
82
Contains Nonbinding Recommendations
Draft — Not for Implementation
Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery
Systems
PAT – A Framework for Innovative Pharmaceutical Development, Manufacturing, and Quality
Assurance
Premarket Approval Application Modular Review
Single Dose Acute Toxicology Testing for Pharmaceuticals
Submitting Supporting Documentation in Drug Applications for the Manufacture of Drug
Substances
Submitting Documentation for the Manufacturing of and Controls for Drug Products
2248 Packaging – Complete, filled transport packages – Vertical impact test by dropping
8318 Packaging — Complete, filled transport packages and unit loads — Sinusoidal vibration
tests using a variable frequency
11607 Packaging for terminally sterilized medical devices —Part 1: Requirements for materials,
sterile barrier systems and packaging systems
83
Contains Nonbinding Recommendations
Draft — Not for Implementation
F746 Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant
Materials
F2129 Standard Test Method for Conducting Cyclic Potentiodynamic Polarization
Measurements to Determine the Corrosion
G71 Standard Guide for Conducting and Evaluating Galvanic Corrosion Tests in Electrolytes
Susceptibility of Small Implant Devices
84