Instructions
Instructions
Instructions
www.tektronix.com
*P071056701*
071056701
Copyright © Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and
workmanship for a period of one (1) year from the date of shipment. If a product proves defective during this
warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor,
or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration
of the warranty period and make suitable arrangements for the performance of service. Customer shall be
responsible for packaging and shipping the defective product to the service center designated by Tektronix, with
shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a
location within the country in which the Tektronix service center is located. Customer shall be responsible for
paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any
damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been
modified or integrated with other products when the effect of such modification or integration increases the time
or difficulty of servicing the product.
List of Figures
Symbols and Terms Terms in this Manual. These terms may appear in this manual:
CAUTION
Refer to Manual
This document provides information on using and servicing the P6417 and
P6418 Logic Analyzer probes.
Related Documentation
Contacting Tektronix
Phone 1-800-833-9200*
* This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
This section provides a brief description of the P6417 and P6418 logic analyzer
probes, information on attaching and using color-coded probe labels, and
information on connecting the probes from the logic analyzer to the target
system.
Product Description
The P6417 and P6418 logic analyzer probes connect a Tektronix TLA series
logic analyzer to a target system. Each probe consists of 16 data channels and
one clock/data channel. You can connect the probes to the target system through
the podlet connectors or through the lead sets.
P6417 Probe
P6418 Probe
Table 1 lists the probe section and label color combinations. Refer to Table 1 and
to Figure 2 when you attach the probe labels.
Probe section Channels Label color Probe section Channels Label color
A3-A2 CK:0, A3:7-0, A2:7-0 Tan A1-A0 CK:1, A1:7-0, A0:7-0 Orange
D3-D2 QUAL:0, D3:7-0, D2:7-0 Blue D1-D0 CK:2, D1:7-0, D0:7-0 Yellow
C3-C2 CK:3, C3:7-0, C2:7-0 White C1-C0 QUAL:1, C1:7-0, C0:7-0 Gray
E3-E2 QUAL:3, E3:7-0, E2:7-0 Green E1-E0 QUAL:2, E1:7-0, E0:7-0 Violet
P6417 Labels Refer to Figure 2 and use the following steps to attach the probe labels:
1. Identify the module end of the probe cable.
2. From the sheet of labels, locate the color-coded label for the module end of
the probe cable.
NOTE. When you install the labels on the P6417 probe, make sure that you select
the correct label for module end of the probe. The labels that connect to both
ends of the probe look similar, but differ in information content. Refer to Figure
2 to ensure you have selected the correct label.
3. Attach the matching colored label to the area on the other end of the probe
cable.
P6418 Labels Refer to Figure 2 and use the following steps to attach the probe labels:
1. From the sheet of labels, locate the color-coded labels.
2. Install the correct label on the module end of the probe.
3. Locate the two labels for the podlet connectors, and attach a label on each
connector of the probe.
NOTE. Make sure that you attach the labels to the correct podlet connector. For
example, make sure that the A3 section at the podlet connector aligns with the
A3 section of the module end of the probe. The podlet connector adjacent to the
clock podlet has the higher-numbered channel group (if necessary, lay the probe
out on a flat surface so you can trace the channel groups from the module end of
the probe to the podlet connector end).
Module end
Clock connector
Podlet connector
P6417 Probe
Module end
Clock connector
Podlet connector
P6418 Probe
Match
color-coded labels
P6417
Attach optional
probe retainer
brackets
P6418
Refer to Figure 4 and connect the probe to the target system using one of the
following methods:
H Connect the probes directly to square pins on the target system (refer to
Probe Connectors on page 7 for information on probe dimensions).
H Connect the probes to the target system via the flying lead sets.
H Use the probe tip connectors with the flying lead sets for situations where
you cannot connect to square pins directly.
H If you use the optional Reduced Bias Voltage lead sets, you can reduce the
pull-up voltage that the probe delievers to the system under test. The pull-up
voltage reduces from 2.2 volts to 1.1 volts at the same time the DC input
resistance changes from 20 kΩ to 10 kΩ. These lead sets can also connect to
2 mm and 0.1-inch pin strips.
Note the location of the ground connections for the probe.
H The individual podlets have ground (GND) engraved on the podlet.
H Most podlet holder has ground engraved on one side of the holder and
numbers on the other side. For newer podlet holders, the ground side is
always opposite the side labeled 7------ 0 (the signal side).
H When you use the 8-channel lead sets, the ground lead is a single, black
connector. Make sure you connect the ground side of the 8-channel lead set
to the ground side of the 8-channel podlet holder.
This section provides reference information and specifications for the P6417 and
P6418 probes.
Probe Connectors
Figure 5 shows the dimensions of the probe footprints for both the P6417 and
P6418 probes.
43.18 mm 40.64 mm
(1.700 in) (1.600 in)
Specifications
Table 2 lists the electrical and mechanical specifications for the P6417 and
P6418 probes. The electrical specifications apply when the probe is connected
between a compatible logic analyzer and a target system. All specifications apply
to both types of probes unless otherwise indicated.
Characteristic Description
Threshold accuracy ± 100 mV
Channel-to-channel-skew ≤ 1 ns
Input resistance (probe and lead sets)
Standard lead set 20 KΩ
Reduced bias lead set 10 KΩ
Input capacitance P6417 P6418
Data 2.0 pF typical 1.4 pF typical
Clock/Qual 2.0 pF typical 2.0 pF typical
Minimum slew rate 0.2 ns
Maximum operating signal 6.5 VP-P
- 3.5 V absolute input voltage minimum
6.5 V absolute input voltage maximum
Maximum nondestructive input ± 15 V
signal to probe
Minimum input pulse width 2 ns
signal
Delay from probe tip to mod- 7.33 ns
ule input connector
Probe length 1.8 m (6 ft)
Table 3 shows the environmental specifications for both probes. The probes are
designed to meet Tektronix standard 062-2847-00 class 5.
Characteristic Description
Temperature
Maximum operating +50° C (+122° F)
Minimum operating 0° C (+32° F)
Non-operating - 55° C to +75° C (-- 67° F to +167° F)
Humidity 10 to 95% relative humidity
Altitude
Operating 4.5 km (15,000 ft) maximum
Non-operating 15 km (50,000 ft) maximum
Electrostatic immunity The probe is not static sensitive
The P6417 and P6418 Probes do not require scheduled or periodic maintenance.
Use the procedures listed under Functional Check to check the basic functional-
ity of the probes.
To verify that the probes meet or exceed the performance requirements for
published specifications with a compatible logic analyzer module, refer to the
Performance Verification Procedures in the TLA7Nx, TLA7Px, & TLA7Qx Logic
Analyzer Service Manual..
Functional Check
The following procedure checks the basic operation of the probes by verifying
that the probes recognize signal activity at the probe tips.
LA module
J15
INT EXT
C2/C3/CK3 Adjustment/verification
J5 fixture Fixture supply
Channel/Group
J3
Clock channel
J26
J1 J2
3. Click the Set Thresholds button to display the Probe Threshold dialog box.
4. Adjust the threshold level to 700 mV for all channels.
5. Connect the acquisition probe to be tested to the C3/C2 channel group on the
LA module.
6. Refer to Figure 8 and connect the podlets of the acquisition probe to J1 and
J2 on the adjustment/verification fixture. Ensure that you connect the ground
side of the podlets to the ground side of the adjustment/verification fixture
connectors.
7. Connect the single clock (CK n) or the qualifier (Q n) channel to one of the
J3 CLK OUT connector pairs on the adjustment/verification fixture.
8. Return to the Setup window and click the Show Activity button to display
the Activity Monitor.
9. Verify that the Activity Monitor shows activity on all probe channels
connected to the test fixture.
Figure 9 shows an example of the Activity Monitor. Note the signal activity
for clock CK3 and data channels for the C3(7-0) and C2(7-0) groups. Also
note that there is no activity on the other groups because the probe podlets
are not connected to a signal source (the channels are all high).
Moving P6417 Probe Depending on your application, you can reposition probe podlets as shown in
Podlets Figure 10. When moving podlets in the podlet holder keep the ground side of the
podlets together.
Remove podlet
cables
Replacing Probe Podlets The podlet replacement procedure consists of identifying the faulty podlet and
replacing it with a new one. Refer to the Replaceable Parts section for informa-
tion on ordering the podlet replacement kit.
Equipment and Required Material. You will need the following equipment and
material:
H Replacement podlet
H Masking tape
H Small diagonal cutters
H Screwdriver with a #1 Phillips tip
NOTE. If you need to replace more than one podlet, replace the podlets one at a
time to avoid mixing them up.
3. Using the screwdriver with the #1 Phillips tip, remove the four screws on the
probe housing on the module end of probe cable (see Figure 11).
Top housing
Cable header
housing
Rubber comb
Bottom housing
Podlet holder
4. Remove the top and bottom probe housings and set them aside. Do not mix
the probe connector-end housings with the podlet connector housings; the
labels are different at each end of the probe.
5. Identify the faulty probe channel at the module end of the cable.
6. Refer to Figure 12 and use a small, pointed tool, such as a straightened paper
clip or a mechanical pencil, to release the faulty cable from the cable header
housing. Discard the small plastic retaining snap.
Paper clip
Cable header housing
Retaining snap
Podlet cable
Figure 12: Removing the faulty channel from the probe cable header housing
7. Identify the faulty probe channel at the podlet end of the probe.
CAUTION. Be sure that you have identified the correct podlet cable before you cut
the cable.
9. Remove the color-coded ring and set it aside; you will need to install it on
the replacement cable later in this procedure.
10. Refer to Figure 13 and bend the rubber comb to remove the faulty podlet
cable from the comb.
11. Attach the module end of the replacement podlet cable to the cut end of the
faulty podlet cable in the following manner:
a. Lay the two cables in parallel with a four-inch (10 cm) to six-inch
(15 cm) overlap.
b. Wrap the overlapping portion of the two podlet cables tightly with
masking tape to temporarily splice them together.
12. Pull the faulty cable from the module end through the mesh sleeve so that
the replacement cable is drawn through the mesh sleeve. Stop pulling when
the replacement cable is located at the proper position in the cable wire
bundle.
13. Remove the faulty podlet cable from the podlet holder and replace it with the
new cable as shown in Figure 14.
Pull open
14. Take the color-coded ring that you removed in step 9 on page 16 and slide it
onto the installation tool. (The installation tool is included in the podlet
replacement kit.)
CAUTION. Do not damage the cable with the installation tool when installing the
color-coded ring.
15. Refer to Figure 15 and place the installation tool over the replacement podlet
cable as shown. Slide the color-coded ring down the installation tool onto the
replacement cable. Position the ring so that it is in the same relative position
as the other color-coded rings (approximately 0.5-inch from the podlet).
Installation tool
Replacement
cable
16. Push the new podlet cable into the rubber comb.
17. Unwrap and remove the masking tape from the splice of the two cables and
discard the faulty cable.
18. Remove the protective 1x2 housing from the new podlet cable. Use the same
procedure as in Step 6 on page 16 to remove the plastic retaining snap.
Discard the retaining snap and the protective housing.
19. Insert the new podlet cable partially into the cable header housing. Place a
new retaining snap (included with the kit) on the podlet cable as shown in
Figure 16 and fully insert the cable and retaining snap into the cable header
housing. Tug lightly on the podlet cable to ensure the retaining snap is in
place.
Podlet cable
20. Reinstall the top and bottom probe housings on the module end of the probe
cable in the following manner (refer to Figure 11 on page 15, if necessary):
a. Line up the two tabs on the probe header with the two embedded slots in
the bottom (deep) probe housing.
NOTE. It may be necessary to push or pull the cables in the sleeve so that the
probe header lines up properly with the slots in the bottom probe housing.
b. Place the two metallic guides on the probe sleeve in the slots in the
bottom probe housing.
c. Line up the three tabs of the probe header with the three embedded slots
in the top probe housing.
Repackaging Instructions
If at all possible, use the original packaging to ship or store the probes. If the
original packaging is not available, use a corrugated cardboard shipping carton.
Add cushioning material to prevent the probes from moving around in the
shipping container.
Enclose the following information when shipping the probe to a Tektronix
Service Center:
H The owner’s address
Mfr. Code to Manufacturer The table titled Manufacturers Cross Index shows codes, names, and addresses
Cross Index of manufacturers or vendors of components listed in the parts list.
Mfr.
code Manufacturer Address City, state, zip code
060D9 TENSOLITE COMPANY PRECISION HARNESS AND ASSEMBLY VANCOUVER, WA 98661
3000 COLUMBIA HOUSE BLVD, #120
23633 RICHEY ELECTRONICS INC 7441 LINCOLN WAY GARDEN GROVE, CA 92641
53387 3M COMPANY 3M AUSTIN CENTER AUSTIN, TX 78769-- 2963
INTERCONNECT SOLUTIONS DIVISION
6801 RIVERPLACE BLVD
0GV90 GLOBTEK INC 186 VETERANS DRIVE NORTHVALE, NJ 07647-- 2303
0KB01 STAUFFER SUPPLY CO 810 SE SHERMAN PORTLAND, OR 97214-- 4657
TK0JL CHROMA ATE INC 43 WU-- CHUAN ROAD WU-- KU, TAIPEI HSIEN, TAIWAN CN
WU-- KU INDUSTRIAL PARK
80009 TEKTRONIX INC 14150 SW KARL BRAUN DR BEAVERTON, OR 97077-- 0001
PO BOX 500
1
2
3
Fig. &
index Tektronix part Serial no. Serial no. Mfr.
number number effective discont’d Qty Name & description code Mfr. part number
P6417 STANDARD ACCESSORIES
- 010-- 6417-- 10 1 PROBE SET:17 CH PROBE 80009 010-- 6417-- 10
- 071-- 0567-- XX 1 MANUAL,TECH:INSTRUCTIONS, P6417 & P6418 80009 071-- 0567-- XX
- 334-- 9239-- 00 1 MARKER,IDENT:PROBE,METIS, EIGHT SETS OF TWO 80009 334-- 9239-- 00
17-- 1 196-- 3432-- 00 1 CA ASSY,SP:DESCRETE,1 CH LEAD SET,SDI,2,22 AWG,5.0 53387 98-- 0300-- 7600-- 8
L,1 X 2,0.278 MLG,PIN HSG X 131-- 1998-- 00,W
-2 196-- 3431-- XX 2 CA ASSY,SP:DESCRETE,8 CH LEAD SET,SDI,16,22 AWG,5.0 53387 98-- 0300-- 7601-- 6
L,2X8,MALE,0.278 MLG PIN HSG X 2X8,ULTR
-3 206-- 0364-- XX 20 TIP,PROBE:MICROCKT TEST,0.05 CTR 80009 206-- 0364-- XX
-2 196-- 3431-- XX 2 CA ASSY,SP:DESCRETE,8 CH LEAD SET,SDI,16,22 AWG,5.0 53387 98-- 0300-- 7601-- 6
L,2X8,MALE,0.278 MLG PIN HSG X 2X8,ULTR
-3 206-- 0364-- XX 20 TIP,PROBE:MICROCKT TEST,0.05 CTR 80009 206-- 0364-- XX
6
7
Fig. &
index Tektronix part Serial no. Serial no. Mfr.
number number effective discont’d Qty Name & description code Mfr. part number
1 P6417 PODLET REPLACEMENT KIT (020- 2200- XX)
18-- 2 10 PODLET LATCHES
-4 1 PODLET (INCLUDING CABLE)
-5 2 8-CHANNEL PODLET HOLDERS
-7 INSTALLATION TOOL & PODLET COLOR-CODED BANDS
1 P6417 PROBE REPAIR KIT (020- 2196- XX)
18-- 1 2 PROBE CASE HALVES WITH SCREWS
-3 1 PROBE SLEEVE (CABLES NOT INCLUDED)
-5 2 8-CHANNEL PODLET HOLDERS
-6 1 PODLET CHANNEL COMB
-7 INSTALLATION TOOL & PODLET COLOR-CODED BANDS
-8 1 2X17 HEADER CONNECTOR
Fig. &
index Tektronix part Serial no. Serial no. Mfr.
number number effective discont’d Qty Name & description code Mfr. part number
P6417 & P6418 OPTIONAL ACCESSORIES
- 071-- 0864-- XX 1 MANUAL,TECH:SERVICE,TLA7NX/TLA7PX/TLA7QX SERIES 80009 071-- 0864-- XX
LOGIC ANALYZER MODULE
- 071-- 0728-- XX 1 MANUAL,TECH:SERVICE,TLA600 SERIES LOGIC ANALYZER 80009 071-- 0728-- XX
MODULE
-2 196-- 3477-- 00 2 LEAD SET,ELEC:8 CHANNEL ASSY, REDUCED BIAS, 060D9 196-- 3477-- 00
P6417/18, 2MM ≤ SPACINGS
-1 196-- 3479-- 00 1 LEAD SET,ELEC:1 CHANNEL ASSY, P6417/18, 2MM ≤ 060D9 196-- 3479-- 00
SPACINGS
-2 196-- 3476-- 00 1 LEAD SET,ELEC:8 CHANNEL ASSY, P6417/18, 2MM≤ 060D9 196-- 3476-- 00
SPACINGS
19-- 2 016-- 1784-- 00 1 CABLE MANAGERS; KIT OF FIVE (P6418 ONLY) 80009 016-- 1784-- 00
-3 407-- 4435-- 01 1 BRACKET,SUPPORT1.2MM THK CRS,LOGIC ANALYZER TK0JL 407-- 4435-- 01
PROBE,SATIN NICKEL PL
-4 211-- 0292-- 00 2 SCR,ASSEM WSHR 4-- 40 X 0.29,PNH,BRS NI PL,POZ 0KB01 211-- 0292-- 00
MACHINE (P6418 ONLY)