LCMS Starting Procedure

- Turn on the vacuum pump, mass spectrometer components, computer, and software before starting procedures. Ensure connections are established between devices. - Perform manual and automatic purging of solvent lines and pumps if machine has been idle for 2+ days or daily before startup. Turn wheels to open/close lines and press purge buttons to flush systems. - Prepare samples using only HPLC/MS grade solvents. Centrifuge and filter to remove particles before injection. Start with dilute samples and increase concentration/volume as needed if signal is weak.

Uploaded by

mambojumb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
72 views7 pages

LCMS Starting Procedure

- Turn on the vacuum pump, mass spectrometer components, computer, and software before starting procedures. Ensure connections are established between devices. - Perform manual and automatic purging of solvent lines and pumps if machine has been idle for 2+ days or daily before startup. Turn wheels to open/close lines and press purge buttons to flush systems. - Prepare samples using only HPLC/MS grade solvents. Centrifuge and filter to remove particles before injection. Start with dilute samples and increase concentration/volume as needed if signal is weak.

Uploaded by

mambojumb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

Starting procedure: by Vicky

- is the vacuum pump running (should run all time, otherwise itll need 4-5 hours to have a adequate
vacuum for the MS) otherwise something has gone wrong- F.T
- Remove plug from ion inlet of mass spec making sure to hold needle at the end
(Figure 1)



Figure 1: Plug in ion inlet of mass spec


Turn on power buttons on all machine components (Figure 2) starting from the right
top side (red arrow) and finishing at the interface (blue arrow).


Figure 2: Turning on computer components from right to left


Turn on computer (Login: admin, password: admin)
Click on software icon labsolutions (login: admin, no password)
Click on LCMS 2020 (Figure 3) Make sure a beep is heard when software opens
(indicating a connection between LC-MS and software)

Figure 3: Labsolutions software LCMS 2020
Turn nebulizer gas on (Gas on icon Figure 3), make sure nebulizing gas says 1.5L
per min in right hand window (Figure 3)



Figure 3: Press NEB icon on, make sure gas reads 1.5L/min




Purge:
- Manual purge - Only necessary if machine has been idle for 2 days or more, you can check this by
visually checking solvent lines for air bubbles. Turn black wheel 90 anticlockwise on Pump A and B,
remove about 10 - 15 mL solvent with the syringe from both solvent lines A and B at the bottom of
the left tower. (Figure 4) NB: When finished purging manually and purging via pressing the
machines button (see auto purge instructions below) turn the black wheel back to the closed
position. (Figure 5), Otherwise your compound will go through the waste bin (happened to me) &
you wont get a spectrum of your compound as it would not have passed through the column. Please
do not tighten the tap too tight, Cedric said thats not good- F.T


Figure 4: Turn black wheels on Pump A and Pump B 90 to the left

Figure 5: Syringe out 10-15 mL solvent form line A and line B


- Auto Purge tubes/pumps - To be done every morning before start-up. (The nebulizer gas must be on
otherwise machine will not allow purging as the computer has been turned on. The typical flow
setting for the gas is 1.5 mL/min - F.T) usually its good to have drying gas on too when you are using
the MS. (Figure 6).
Turn black wheel 90 anticlockwise on pump A and pump B. Press purge and let run for
about 3-4 min. Then press purge again to stop it (unless it has stopped itself already).
(Default value is 8 mins but 3-4mins is enough). NB: When finished turn the black wheel
back to closed position.

Figure 6: Press purge on Pump A and Pump B

Purge the auto sampler (default 25 min, but 10 - 15 min should be enough): press purge on
front of machine and leave running for about 10-15 mins. NB: Make sure there is enough
rinsing solution (Solvent C) before purging. (Figure 7),

Figure 7: Press purge on front of autosampler
Note: Cedric said the pumps can withhold pressures up to 6000 psi. Usually the pump pressure is way
below 4000 psi and if it is at an instance > 4000psi then the pump is blocked. Usually, if solvent is not
flowing through the column yet, the pressure is about 200-300 psi but when the solvent is flowing
through the column the pressure is between ~300 and ~2500 psi depending on the solvent used and
column density - F.T.

Mass Spec: It is essential to remove the plug from ion inlet before using the MS (dont touch plug /
ion source plug at the tip), check that the vacuum light is green on MS (after closing the warning
window, which appears every time when the MS chamber is opened).
Turn Nebulizer gas on (if off - F.T), (Gas on icon). Click download. Click instrument icon to activate
pumps and column oven. Green lights will be seen on front of machine parts when activated.


Sample preparation
This is absolutely crucial. The instrument will block up if a sample is badly prepared
- use only HPLC/MS grade solvents to dissolve sample
- Any solvents that are miscible with the solvent used for the run can be used. E.g. MeOH - F.T.
- centrifuging and/or filtering the sample (pipette with cotton wool plug) to remove particles / solids
(the best way to filter solvents is to centrifuge first and then filter using the 0.2uM filters- F.T)

How much to inject?
If in doubt always start with a sample that is very dilute. If your signal is weak, you can increase the
concentration or injector volume. Standards (things that ionize well) can be analysed at concentrations
of XXX. As a guide: maximum concentration that should be employed is: 1mg / 10 mL
- minimum concentration: 10 pg/L ( 1 mg / 100 mL)