3B-Using CLSI Guidelines To
3B-Using CLSI Guidelines To
Correlation coefficient, r
Responds to random error.
Value depends on the range of data.
Does not estimate analytical bias or random error between
methods.
Merely presents the relationship of the range of the data to the
scatter of the data between methods.
Therefore, the correlation coefficient should NOT be used to
judge acceptability of analytical methods in method comparison
studies.
Linear regression statistics
Subject to certain limitations:
Data must be linear
Outliers must be carefully examined
Range of data must be wide:
a. r > 0.99 (Waakers et al.)
b. r > 0.975 (CLSI EP9-A)
Recommendations for Method
Comparison
Summary
Present graph of data
Present slope, y-intercept, and S
y/x
Present mean and standard deviation of X data
Present correlation coefficient ONLY to show that least
squares regression is applicable; if not, use Deming or
Passing-Bablock regression statistics
Performance Characteristics:
Linearity
Linearity Measuring Interval or Analytical Measurement Range (AMR)
A linearity study is used to establish or verify the measuring interval for a
measurement method.
Measuring Interval: the interval between lower and upper numerical
values for which a method can produce quantitative results suitable for
the intended clinical use.
The measuring interval is verified by demonstrating a linear relationship
between the measured and expected concentration relationships.
CLSI Guideline for Linearity Measuring Interval
EP06: procedures to verify or establish the linear measuring interval of a
measurement procedure.
Performance Characteristics:
LoD/LoQ
Limit of Detection (LoD) & Limit of Quantitation (LoQ)
(sometimes referred to as Analytical Sensitivity)
LoD: the lowest amount of analyte (measurand) in a sample that can be
detected with a stated probability.
LoQ: the lowest amount of analyte (measurand) in a sample that can be
quantified with acceptable precision and bias under stated experimental
conditions.
Usually, laboratories review and accept the manufacturers claims for LoD
and LoQ.
But these characteristics can be tested by laboratories using:
CLSI Guideline for LoD and LoQ
EP17: procedures for verifying or establishing the LoD and the LoQ
Performance Characteristics:
Interference
Interference: an artifactual increase or decrease in the apparent quantity
of an analyte due to the presence of a substance that reacts
nonspecifically with the measuring system.
Most manufacturers evaluate a large number of substances known or
suspected to be potential interferents. They report this information in the
Instructions For Use (IFU).
It is not practical for most clinical laboratories to repeat such an
investigation and inspection of the manufacturers information is
frequently sufficient.
But these characteristics can be tested by laboratories using:
CLSI Guideline for Interference
EP7: procedures for testing constant error due to interference
1. See CLSI EP7-A2
2. What to test:
Literature review
Always test hemolysis, lipemia, bilirubin
Tube additives
3. Concentrations to test:
Interferent: highest compatible with life
Analyte: at medical decision levels
4. Volume of interferent <10% of sample
5. Replicates: Based on Effect / S
tm
(see EP7)
6. Validate technique with current method
Interference Experiment:
Factors
Interference Experiment: N=?
Number of Measurements / Replicates:
at least several samples per interferent
at least duplicates per sample
EP7 lists a table of N as a function of bias/s
tm
(E
A,I
/S
tm
), with which
one can determine how many replicates are necessary to reach
95% probability of observing a certain magnitude of error:
E
A,I
/S
tm
No. Replicates E
A,I
/S
tm
No. Replicates
0.8 41 1.5 12
1.0 26 1.6 10
1.1 22 1.8 8
1.2 18 2.0 7
1.3 16 2.5 6
1.4 14 3.0 3
Performance Characteristics:
Carryover
Carryover: the discrete amount of reagent or analyte carried by the
measuring system from one test into subsequent test(s), thereby
erroneously affecting test results.
Reagent carryover among different measurement procedures on
multichannel automated analyzers is an evaluation that is usually
conducted by measuring system manufacturers.
But this characteristic can be tested by laboratories using:
CLSI Guideline for Carryover
EP10: includes an assessment of sample carryover along with other
parameters.
NOTE: EP10 is intended to determine if a device has unacceptable
performance. It is recognized in the CAP Chemistry Checklist as an
acceptable way to measure carryover.
Performance Characteristics:
Reference Intervals
Reference Interval: interpretive information for laboratory test results that
is frequently provided as the central 95% interval of results for a group of
well-defined reference individuals.
Laboratories can produce reference intervals in a variety of ways,
including testing procedures found in
CLSI Guideline for Reference Intervals or Decision Value
C28: procedures for establishing a reference interval or verifying the
suitability of a manufacturer-proposed reference interval
Transference of established reference intervals to an individual laboratory or
a new method may be accomplished in a variety of ways:
1. Subjective assessment by a responsible individual;
the Medical Director (sometimes called by divine judgment)
2. Donor testing
a. Verify with ~ 20 donor samples
b. Validate/Estimate using ~ 60 donor samples
c. Establish using ~ 120 donor samples
3. Calculation
use regression statistics from a comparison of methods study to calculate
reference limits for the new method (Y) that correspond to the reference
interval limits of the former method (X).
Reference Interval Determination
Y = a + b X
CLSI Makes Life Easier with
StatisPro
In October 2010, CLSI released
StatisPro software:
Direct, faithful implementation of
CLSI Evaluation Protocol Guidelines
Study Advisor step-by-step help for
each study
Four steps to complete a study:
Definition, Data Input, Analysis, and
Signoff
StatisPro Pick a Study Type
StatisPro Study Design
Performance Claim
to be Verified
Study Goal
Identifying
Information
Details of
the Study
Description of
Materials Used
StatisPro Data Entry
Copying and Pasting from any spreadsheet application or Windows
application with clipboard support is easy.
StatisPro Analysis
1 - Inspect group: Evaluate the data visually using various plots and tables.
You can choose to show or hide excluded observations.
2 - Outliers group: Select an observation to exclude from the calculations.
3 - Study-specific group: Select commands that continue to evaluate the
data and reach a study conclusion.
4 - Sign Off group: Add any comments, your name, and a signature line to
the study report so it is ready for a handwritten signature when printed.
StatisPro Study Advisor
StatisPro Demonstration
Demonstrate EP15 (method comparison) and EP06 (linearity).
User Experience with StatisPro
StatisPro is useful when introducing new methods into your laboratory.
StatisPro is useful when performing six-month linearity or calibration verification
studies.
By using StatisPro:
You are demonstrating compliance with regulatory and accreditation bodies.
You are ensuring that your laboratory delivers accurate results.
Thank You
Questions?