Tbs1000 Manual
Tbs1000 Manual
Tbs1000 Manual
TBS1000 Series
Digital Storage Oscilloscopes
ZZZ
User Manual
*P077076000*
077-0760-00
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
TBS1000 Series
Digital Storage Oscilloscopes
ZZZ
User Manual
xx
Revision A
Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
OpenChoice is a registered trademark of Tektronix, Inc.
PictBridge is a registered trademark of the Standard of Camera & Imaging Products Association CIPA
DC-001-2003 Digital Photo Solutions for Imaging Devices.
TBS1000 Oscilloscopes
Warranty
Tektronix warrants that the product will be free from defects in materials and workmanship for a period of ve (5)
years from the date of original purchase from an authorized Tektronix distributor. If the product proves defective
during this warranty period, Tektronix, at its option, either will repair the defective product without charge for
parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from
this warranty. Parts, modules and replacement products used by Tektronix for warranty work may be new or
reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration
of the warranty period and make suitable arrangements for the performance of service. Customer shall be
responsible for packaging and shipping the defective product to the service center designated by Tektronix,
shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location within the country in which the Tektronix service center is
located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for
products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage
or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or
integrated with other products when the effect of such modication or integration increases the time or difculty
of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W19 03AUG12]
TPP0101 and TPP0201 Probes
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1)
year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its
option, either will repair the defective product without charge for parts and labor, or will provide a replacement
in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty
work may be new or reconditioned to like new performance. All replaced parts, modules and products become
the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of
the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible
for packaging and shipping the defective product to the service center designated by Tektronix, with shipping
charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within
the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage
or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or
integrated with other products when the effect of such modication or integration increases the time or difculty
of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE
AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 15AUG04]
Table of Contents
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Compliance Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
EMC Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Safety Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Help System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
Firmware Updates Through the Internet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiv
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xv
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Probe Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Voltage Probe Check Wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Manual Probe Compensation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Probe Attenuation Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Current Probe Scaling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Self Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operating Basics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Display Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Using the Menu System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Vertical Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Horizontal Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Trigger Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Menu and Control Buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Input Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Other Front-Panel Items. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Understanding Oscilloscope Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting Up the Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Triggering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Acquiring Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Scaling and Positioning Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Taking Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Application Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Taking Simple Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Using Autorange to Examine a Series of Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Taking Cursor Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Analyzing Signal Detail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
TBS1000 Series Oscilloscopes User Manual i
Table of Contents
Capturing a Single-Shot Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Measuring Propagation Delay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Triggering on a Specic Pulse Width. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Triggering on a Video Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Analyzing a Differential Communication Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Viewing Impedance Changes in a Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Data Logging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Limit Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Math FFT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Setting Up the Time-Domain Waveform. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Displaying the FFT Spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Selecting an FFT Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Magnifying and Positioning an FFT Spectrum. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Measuring an FFT Spectrum Using Cursors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
USB Flash Drive and Device Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
USB Flash Drive Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
File Management Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Saving and Recalling Files With a USB Flash Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Using the Save Function of the Print Front Panel Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
USB Device Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Installing the PC Communications Software on a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Connecting to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Connecting to a GPIB System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Command Entry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Connecting to a Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Printing a Screen Image. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Acquire. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Autorange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Autoset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Cursor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Default Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Horizontal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Math. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Probe Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Reference Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Save/Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
ii TBS1000 Series Oscilloscopes User Manual
Table of Contents
Trigger Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Vertical Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Appendix A: Specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Oscilloscope Specications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Appendix B: TPP0101 and TPP0201 Series 10X Passive Probes Information. . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Connecting the Probe to the Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Compensating the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Connecting the Probe to the Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Specications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Performance Graphs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Appendix C: Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Appendix D: Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
General Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
Appendix E: Default Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Appendix F: Font Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Index
TBS1000 Series Oscilloscopes User Manual iii
General Safety Summary
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
To Avoid Fire or Personal
Injury
Use proper power cord. Use only the power cord specied for this product and
certied for the country of use.
Connect and disconnect properly. Connect the probe output to the measurement
instrument before connecting the probe to the circuit under test. Connect the
probe reference lead to the circuit under test before connecting the probe input.
Disconnect the probe input and the probe reference lead from the circuit under test
before disconnecting the probe from the measurement instrument.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Power disconnect. The power switch disconnects the product from the power
source. See instructions for the location. Do not block the power switch; it must
remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
iv TBS1000 Series Oscilloscopes User Manual
General Safety Summary
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual's installation instructions for details
on installing the product so it has proper ventilation.
TBS1000 Series Oscilloscopes User Manual v
General Safety Summary
Terms in This Manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Symbols and Terms on the
Product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read
the marking.
WARNING indicates an injury hazard not immediately accessible as you
read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
vi TBS1000 Series Oscilloscopes User Manual
Compliance Information
This section lists the EMC (electromagnetic compliance), safety, and
environmental standards with which the instrument complies.
EMC Compliance
EC Declaration of
Conformity EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility.
Compliance was demonstrated to the following specications as listed in the
Ofcial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment
for measurement, control, and laboratory use.
1 2 3
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
4
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
5
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
6
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European Contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
Thisproduct isintendedfor useinnonresidential areasonly. Useinresidential areasmaycauseelectromagnetic
interference.
2
Emissionswhichexceedthelevels requiredby thisstandardmay occur whenthis equipment isconnectedtoa
test object.
3
ToensurecompliancewiththeEMCstandardslistedhere, highqualityshieldedinterfacecablesshouldbeused.
4
Theinstrument will exhibit 1.0divisionwaveformdisplacement and 2.0divisionincreaseinpeak-to-peak
noisewhensubjectedtoradiatedinterferenceper IEC61000-4-3.
5
Theinstrument will exhibit 0.5divisionwaveformdisplacement and 1.0divisionincreaseinpeak-to-peak
noisewhensubjectedtoconductedinterferenceper IEC61000-4-6.
6
PerformanceCriterionCappliedat the70%/25cycleVoltage-Dipandthe0%/250cycleVoltage-Interruptiontest
levels(IEC61000-4-11). If theinstrument powersdownuponavoltagedipor interruption, it will takelonger than
tenseconds toreturntotheprevious operatingstate.
TBS1000 Series Oscilloscopes User Manual vii
Compliance Information
Australia / New Zealand
Declaration of
Conformity EMC
Complies with the EMC provision of the Radiocommunications Act per the
following standard, in accordance with ACMA:
CISPR 11:2003. Radiated and Conducted Emissions, Group 1, Class A, in
accordance with EN 61326-1:2006 and EN 61326-2-1:2006.
viii TBS1000 Series Oscilloscopes User Manual
Compliance Information
Safety Compliance
EC Declaration of
Conformity Low Voltage
Compliance was demonstrated to the following specication as listed in the
Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-1: 2001. Safety requirements for electrical equipment for
measurement control and laboratory use.
U.S. Nationally Recognized
Testing Laboratory Listing
UL 61010-1:2004, 2
nd
Edition. Standard for electrical measuring and test
equipment.
Canadian Certication
CAN/CSA-C22.2 No. 61010-1:2004. Safety requirements for electrical
equipment for measurement, control, and laboratory use. Part 1.
Additional Compliances
IEC 61010-1: 2001. Safety requirements for electrical equipment for
measurement, control, and laboratory use.
Equipment Type
Test and measuring equipment.
Safety Class
Class 1 grounded product.
Pollution Degree
Description
A measure of the contaminants that could occur in the environment around
and within a product. Typically the internal environment inside a product is
considered to be the same as the external. Products should be used only in the
environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs.
Products in this category are generally encapsulated, hermetically sealed, or
located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs.
Occasionally a temporary conductivity that is caused by condensation must
be expected. This location is a typical ofce/home environment. Temporary
condensation occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution
that becomes conductive due to condensation. These are sheltered locations
where neither temperature nor humidity is controlled. The area is protected
from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through
conductive dust, rain, or snow. Typical outdoor locations.
TBS1000 Series Oscilloscopes User Manual ix
Compliance Information
Installation (Overvoltage)
Category Descriptions
Terminals on this product may have different installation (overvoltage) category
designations. The installation categories are:
Measurement Category IV. For measurements performed at the source of
low-voltage installation.
Measurement Category III. For measurements performed in the building
installation.
Measurement Category II. For measurements performed on circuits directly
connected to the low-voltage installation.
Measurement Category I. For measurements performed on circuits not
directly connected to MAINS.
Overvoltage Category
Overvoltage Category II (as dened in IEC 61010-1).
x TBS1000 Series Oscilloscopes User Manual
Compliance Information
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life
Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of this equipment required the extraction and
use of natural resources. The equipment may contain substances that could be
harmful to the environment or human health if improperly handled at the products
end of life. In order to avoid release of such substances into the environment and
to reduce the use of natural resources, we encourage you to recycle this product
in an appropriate system that will ensure that most of the materials are reused or
recycled appropriately.
This symbol indicates that this product complies withtheapplicableEuropean
Union requirements according to Directives 2002/96/ECand 2006/66/EC
on waste electrical and electronic equipment (WEEE) and batteries. For
informationabout recyclingoptions, check theSupport/Servicesectionof the
Tektronix Website(www.tektronix.com).
Restriction of Hazardous
Substances
This product is classied as an industrial monitoring and control instrument,
and is not required to comply with the substance restrictions of the recast RoHS
Directive 2011/65/EU until July 22, 2017.
TBS1000 Series Oscilloscopes User Manual xi
Compliance Information
xii TBS1000 Series Oscilloscopes User Manual
Preface
Preface
This manual contains operating information for the TBS1000 Series Digital
Storage Oscilloscopes.
Help System
The oscilloscope has a Help system with topics that cover all the features of the
oscilloscope. You can use the Help system to display several kinds of information:
General information about understanding and using the oscilloscope, such
as Using the Menu System.
Information about specic menus and controls, such as the Vertical Position
Control.
Advice about problems you may face while using an oscilloscope, such as
Reducing Noise.
The Help system provides several ways to nd the information you need:
context-sensitive help, hyperlinks, and an index.
Context-Sensitive Help
The oscilloscope displays information about the last menu displayed on the
screen when you push the Help front-panel button. When viewing help topics,
an LED lights next to the multipurpose knob to indicate that the knob is active.
If the topic uses more than one page, turn the multipurpose knob to move from
page to page within the topic.
Hyperlinks
Most of the help topics contain phrases marked with angle brackets, such as
<Autoset>. These are links to other topics. Turn the multipurpose knob to move
the highlight from one link to another. Push the Show Topic option button to
display the topic corresponding to the highlighted link. Push the Back option
button to return to the previous topic.
Index
Push the front-panel Help button, then push the Index option button. Push the
Page Up or Page Down option buttons until you nd the index page that contains
the topic you want to view. Turn the multipurpose knob to highlight a help topic.
Push the Show Topic option button to display the topic.
NOTE. Push the Exit option button or any menu button to remove the Help text
from the screen and return to displaying waveforms.
TBS1000 Series Oscilloscopes User Manual xiii
Preface
Firmware Updates Through the Internet
If a newer version of rmware becomes available, you can use the Internet and
a USB ash drive to update your oscilloscope. If you do not have access to the
Internet, contact Tektronix for information on update procedures.
To update the rmware from the Internet, follow these steps:
1. Push the Utility System Status option, and write down the rmware
version number of the oscilloscope.
2. From your computer, access the www.tektronix.com web site and check if a
newer version of oscilloscope rmware is available.
3. If there is a newer version of rmware, download the rmware le from the
web page.
You may need to unzip the downloaded le.
4. Copy the rmware le to the root folder of a USB ash drive.
5. Insert the USB ash drive into the USB Flash Drive port on the front of the
oscilloscope.
6. From your oscilloscope, push the Utility File Utilities - more - page 2
of 2 Update Firmware option button.
It takes several minutes to update the rmware.
Your oscilloscope will prompt you to press a button when the rmware update is
complete. You must not remove the USB ash drive, or power off the oscilloscope
until the rmware update is complete.
xiv TBS1000 Series Oscilloscopes User Manual
Preface
Conventions
This manual uses the following conventions:
Menu options appear with the rst letter of each word in upper case. For
example: Peak Detect, Window Zone.
Multipurpose knob
Front-panel buttons and knob labels First
letter of each word in upper case
\
\
Optionbuttons First letter of eachwordonscreenis upper case
NOTE. Option buttons may also be called screen buttons, side-menu buttons,
bezel buttons, or soft keys.
The delimiter separates a series of button pushes. For example, Utility
Options Set Date and Time means that you push the Utility front-panel
button, then push the Options option button, and then push the Set Date and
Time option button. Multiple pushes of an option button may be required
to select the desired option.
TBS1000 Series Oscilloscopes User Manual xv
Preface
xvi TBS1000 Series Oscilloscopes User Manual
Getting Started
TBS1000 Series Digital Storage Oscilloscopes are small, lightweight, benchtop
instruments, which you can use to take ground-referenced measurements.
This chapter describes how to do the following tasks:
Install your product
Perform a brief functional check
Perform a probe check and compensate probes
Match your probe attenuation factor
Use the self calibration routine
NOTE. You can select a language to display on the screen when you power on the
oscilloscope. At any time, you can also access the Utility Language option to
select a language.
General Features
The next table and list describe the general features.
Model Channels Bandwidth Sample rate Display
TBS1022 2 25MHz 500MS/s Color
TBS1042 2 40MHz 500MS/s Color
TBS1062 2 60MHz 1GS/s Color
TBS1102 2 100MHz 1GS/s Color
TBS1152 2 150MHz 1GS/s Color
Context-sensitive help system
Color LCD display
Selectable 20 MHz bandwidth limit
2500 point record length for each channel
Autoset
Autoranging
Probe Check Wizard
Setup and waveform storage
USB Flash Drive port for le storage
Direct printing to any PictBridge compatible printer
TBS1000 Series Oscilloscopes User Manual 1
Getting Started
PC communications through the USB Device port with OpenChoice PC
Communications software
Connect to a GPIB controller through an optional TEK-USB-488 adapter
Cursors with readouts
Trigger frequency readout
Sixteen automatic measurements
Waveform averaging and peak detection
Dual time base
Math functions: +, -, and operations
Math Fast Fourier Transform (FFT)
Pulse Width trigger capability
Video trigger capability with line-selectable triggering
External trigger
Variable persistence display
User interface and help topics in ten languages
Installation
Power Cord
Use only the power cord provided with your oscilloscope. Appendix B:
Accessories lists the standard and the optional accessories.
Power Source
Use a power source that delivers 90 to 264 VAC
RMS
, 45 to 66 Hz. If you have a
400 Hz power source, it must deliver 90 to 132 VAC
RMS
, 360 to 440 Hz.
Security Loop
Use a standard laptop computer security lock, or thread a security cable through
the built-in cable channel to secure your oscilloscope to your location.
2 TBS1000 Series Oscilloscopes User Manual
Getting Started
Security cablechannel Security lock hole Power cord
Ventilation
NOTE. The oscilloscope cools by convection. Keep two inches clear on the sides
and top of the product to allow adequate air ow.
TBS1000 Series Oscilloscopes User Manual 3
Getting Started
Functional Check
Perform this functional check to verify that your oscilloscope is operating
correctly.
ON/OFFbutton
1. Power ontheoscilloscope.
PushtheDefault Setup button.
The default Probeoptionattenuationsettingis
10X.
Default Setup button
PROBECOMP
2. Connect theTPP0101/TP0201probetochannel
1ontheoscilloscope. Todothis, aligntheslot
intheprobeconnector withthekey ontheCH1
BNC, pushtoconnect, andtwist totheright to
lock theprobeinplace.
Connect theprobetipandreferenceleadtothe
PROBECOMPterminals.
3. PushtheAutoSet button. Withinafewseconds,
youshouldseeasquarewaveinthedisplay of
about 5Vpeak-to-peak at 1kHz.
Pushthe1 channel 1menubuttononthefront
panel twice to remove channel 1, push the 2
channel 2menubuttontodisplaychannel 2, and
repeat steps 2 and 3. For 4-channel models,
repeat for 3 and4.
4 TBS1000 Series Oscilloscopes User Manual
Getting Started
Probe Safety
Check and observe probe ratings before using probes.
A guard around the TPP0101/TPP0201 probe body provides a nger barrier for
protection from electric shock.
Finger guard
WARNING. To avoid electric shock when using the probe, keep ngers behind
the guard on the probe body.
To avoid electric shock while using the probe, do not touch metallic portions of
the probe head while it is connected to a voltage source.
Connect the probe to the oscilloscope, and connect the ground terminal to ground
before you take any measurements.
Voltage Probe Check Wizard
You can use the Probe Check Wizard to verify that a voltage probe is operating
properly. The wizard does not support current probes.
The wizard helps you adjust the compensation for voltage probes (usually with a
screw on the probe body or probe connector) and set the factor for the Attenuation
option for each channel, such as in the 1 Probe Voltage Attenuation
option.
TBS1000 Series Oscilloscopes User Manual 5
Getting Started
You should use the Probe Check Wizard each time you connect a voltage probe to
an input channel.
To use the Probe Check Wizard, push the PROBE CHECK button. If the voltage
probe is connected properly, compensated properly, and the Attenuation option in
the oscilloscope Vertical menu is set to match the probe, the oscilloscope displays
a PASSED message at the bottom of the screen. Otherwise, the oscilloscope
displays directions on the screen to guide you in correcting these problems.
NOTE. The Probe Check Wizard is useful for 1X, 10X, 20X, 50X, and 100X
probes. It is not useful for 500X or 1000X probes, or for probes connected to
the Ext Trig BNC.
NOTE. When the process is complete, the Probe Check Wizard restores the
oscilloscope settings (other than the Probe option) to what they were before you
pushed the PROBE CHECK button.
To compensate a probe that you plan to use with the Ext Trig input, follow these
steps:
1. Connect the probe to any input channel BNC, such as to channel 1.
2. Push the PROBE CHECK button and follow the directions on the screen.
3. After you verify that the probe functions and is compensated properly, connect
the probe to the Ext Trig BNC.
6 TBS1000 Series Oscilloscopes User Manual
Getting Started
Manual Probe Compensation
As an alternative method to the Probe Check Wizard, you can manually perform
this adjustment to match your probe to the input channel.
PROBECOMP AutoSet
button
1. Push the 1 Probe Voltage
Attenuation option and select 10X. Connect
the TPP0101/TPP0201 probeto channel 1on
theoscilloscope. If youusetheprobehook-tip,
ensureaproper connectionbyrmlyinsertingthe
tipontothe probe.
2. Attach the probe tip to the PROBECOMP
~5V@1kHz terminal andthereferenceleadto
thePROBECOMPchassisterminal. Display the
channel, andthenpushtheAutoSet button.
Overcompensated
Undercompensated
Compensatedcorrectly
3. Check theshapeof thedisplayedwaveform.
4. If necessary, adjust your probe.
Repeat as necessary.
Probe Attenuation Setting
Probes are available with various attenuation factors which affect the vertical
scale of the signal. The Probe Check Wizard veries that the attenuation factor
in the oscilloscope matches the probe.
As an alternative method to Probe Check, you can manually select the factor
that matches the attenuation of your probe. For example, to match a probe set
to 10X connected to CH 1, push the 1 Probe Voltage Attenuation
option, and select 10X.
NOTE. The default setting for the Attenuation option is 10X.
TBS1000 Series Oscilloscopes User Manual 7
Getting Started
If you change the Attenuation switch on a P2220 probe, you also need to change
the oscilloscope Attenuation option to match. Switch settings are 1X and 10X.
Attenuation switch
NOTE. When the Attenuation switch is set to 1X, the P2220 probe limits the
bandwidth of the oscilloscope to 6 MHz. To use the full bandwidth of the
oscilloscope, be sure to set the switch to 10X.
Current Probe Scaling
Current probes provide a voltage signal proportional to the current. You need to
set the oscilloscope to match the scale of your current probe. The default scale is
10 A/V.
For example, to set the scale for a current probe connected to CH 1, push the 1
Probe Current Scale option, and select an appropriate value.
Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for
maximum measurement accuracy. You can run the routine at any time but you
should always run the routine if the ambient temperature changes by 5 C (9 F)
or more. The routine takes about two minutes.
For accurate calibration, power on the oscilloscope and wait twenty minutes to
ensure it is warmed up.
To compensate the signal path, disconnect any probes or cables from the input
connectors. Then, access the Utility Do Self Cal option, and follow the
directions on the screen.
8 TBS1000 Series Oscilloscopes User Manual
Operating Basics
The front panel is divided into easy-to-use functional areas. This chapter provides
you with a quick overview of the controls and the information displayed on the
screen.
Display Area
In addition to displaying waveforms, the display is lled with many details about
the waveform and the oscilloscope control settings.
NOTE. For details on displaying the FFT function, (See page 57, Displaying
the FFT Spectrum.)
TBS1000 Series Oscilloscopes User Manual 9
Operating Basics
1. Icon display shows acquisition mode.
Sample mode
Peak detect mode
Average mode
10 TBS1000 Series Oscilloscopes User Manual
Operating Basics
2. Trigger status indicates the following:
Theoscilloscopeis acquiringpretrigger data. All triggers are
ignored in this state.
All pretrigger datahas beenacquiredandtheoscilloscopeis
ready to accept atrigger.
The oscilloscope has seen a trigger and is acquiring the
posttrigger data.
Theoscilloscopehas stoppedacquiringwaveformdata.
TheoscilloscopehascompletedaSingleSequenceacquisition.
Theoscilloscopeis inautomodeandis acquiringwaveforms in
the absence of triggers.
Theoscilloscopeis acquiringanddisplayingwaveformdata
continuously inscanmode.
3. Marker shows horizontal trigger position. Turn the Horizontal Position knob
to adjust the position of the marker.
4. Readout shows the time at the center graticule. The trigger time is zero.
5. Marker shows Edge or Pulse Width trigger level.
6. On-screen markers show the ground reference points of the displayed
waveforms. If there is no marker, the channel is not displayed.
7. An arrow icon indicates that the waveform is inverted.
8. Readouts show the vertical scale factors of the channels.
9. A B
W
icon indicates that the channel is bandwidth limited.
10. Readout shows main time base setting.
11. Readout shows window time base setting if it is in use.
12. Readout shows trigger source used for triggering.
13. Icon shows selected trigger type as follows:
Edgetrigger for therisingedge.
Edgetrigger for thefallingedge.
Videotrigger for linesync.
Videotrigger for eldsync.
PulseWidthtrigger, positivepolarity.
PulseWidthtrigger, negativepolarity.
TBS1000 Series Oscilloscopes User Manual 11
Operating Basics
14. Readout shows Edge or Pulse Width trigger level.
15. Display area shows helpful messages; some messages display for only three
seconds.
If you recall a saved waveform, readout shows information about the reference
waveform, such as RefA 1.00V 500s.
16. Readout shows date and time.
17. Readout shows trigger frequency.
Message Area
The oscilloscope displays a message area (item number 15 in the previous gure)
at the bottom of the screen that conveys the following types of helpful information:
Directions to access another menu, such as when you push the Trig Menu
button:
For TRI GGER HOLDOFF, go t o HORI ZONTAL MENU
Suggestion of what you might want to do next, such as when you push the
Measure button:
Push an opt i on but t on t o change i t s measur ement
Information about the action the oscilloscope performed, such as when you
push the Default Setup button:
Def aul t set up r ecal l ed
Information about the waveform, such as when you push the AutoSet button:
Squar e wave or pul se det ect ed on CH1
Using the Menu System
The user interface of the oscilloscopes was designed for easy access to specialized
functions through the menu structure.
When you push a front-panel button, the oscilloscope displays the corresponding
menu on the right side of the screen. The menu shows the options that are available
when you push the unlabeled option buttons directly to the right of the screen.
The oscilloscope uses several methods to display menu options:
Page (Submenu) Selection: For some menus, you can use the top option
button to choose two or three submenus. Each time you push the top button,
the options change. For example, when you push the top button in the Trigger
Menu, the oscilloscope cycles through the Edge, Video, and Pulse Width
trigger submenus.
Circular List: The oscilloscope sets the parameter to a different value each
time you push the option button. For example, you can push the 1 (channel
12 TBS1000 Series Oscilloscopes User Manual
Operating Basics
1 menu) button and then push the top option button to cycle through the
Vertical (channel) Coupling options.
In some lists, you can use the multipurpose knob to select an option. A hint
line tells you when the multipurpose knob can be used, and an LED by the
multipurpose knob lights when the knob is active. (See page 15, Menu and
Control Buttons.)
Action: The oscilloscope displays the type of action that will immediately
occur when you push an Action option button. For example, when the Help
Index is visible, and you push the Page Down option button, the oscilloscope
immediately displays the next page of index entries.
Radio: The oscilloscope uses a different button for each option. The
currently-selected option is highlighted. For example, the oscilloscope
displays various acquisition mode options when you push the Acquire Menu
button. To select an option, push the corresponding button.
PageSelection Circular List Action Radio
TRIGGER CH1 HELP ACQUIRE
Type
Edge
Coupling
DC
Page
Up
Sample
or or
Page
Down
Peak Detect
TRIGGER CH1
Average
Type
Video
Coupling
AC
or or
TRIGGER CH1
Type
Pulse
Coupling
Ground
Vertical Controls
TBS1000 Series Oscilloscopes User Manual 13
Operating Basics
Position (1 and 2). Positions a waveform vertically.
1 & 2 Menu. Displays the Vertical menu selections and toggles the display of the
channel waveform on and off.
Scale (1 & 2). Selects vertical scale factors.
Math. Displays waveform math operations menu and toggles the display of the
math waveform on and off.
Horizontal Controls
Position. Adjusts the horizontal position of all channel and math waveforms.
The resolution of this control varies with the time base setting. (See page 86,
Window Zone.)
NOTE. To make a large adjustment to the horizontal position, turn the Horizontal
Scale knob to a larger value, change the horizontal position, and then turn the
Horizontal Scale knob back to the previous value.
Horiz. Displays the Horizontal Menu.
Set to Zero. Sets the horizontal position to zero.
Scale. Selects the horizontal time/division (scale factor) for the main or the
window time base. When Window Zone is enabled, it changes the width of the
window zone by changing the window time base. (See page 86, Window Zone.)
14 TBS1000 Series Oscilloscopes User Manual
Operating Basics
Trigger Controls
Level. When you use an Edge or Pulse trigger, the Level knob sets the amplitude
level that the signal must cross to acquire a waveform.
Trig Menu. Displays the Trigger Menu.
Set To 50%. The trigger level is set to the vertical midpoint between the peaks of
the trigger signal.
Force Trig. Completes an acquisition regardless of an adequate trigger signal.
This button has no effect if the acquisition is already stopped.
Trig View. Displays the trigger waveform in place of the channel waveform while
you hold down the Trig View button. Use this to see how the trigger settings
affect the trigger signal, such as trigger coupling.
Menu and Control Buttons
Multipurpose knob
TBS1000 Series Oscilloscopes User Manual 15
Operating Basics
Refer to the Reference chapter for detailed information on the menu and button
controls.
Multipurpose Knob. The function is determined by the displayed menu or selected
menu option. When active, the adjacent LED lights. The next table lists the
functions.
Active menu or option Knob function Description
Cursor Cursor 1or Cursor 2 Positions theselectedcursor
Help Scroll Selects entries in the Index; selects
links in a topic; displays the next or
previous pagefor atopic
Horizontal Holdoff Setstheamount of timebeforeanother
trigger event can beaccepted; (See
page100, Holdoff.)
Position Positions theMathwaveform Math
Vertical Scale Changes the scale of the Math
waveform
Measure Type Selects the type of automatic
measurement for eachsource
Action Sets thetransactionas saveor recall
for setup les, waveformles, and
screenimages
Save/Recall
Fileselection Selectssetup, waveform, or imageles
tosave, or selects setupor waveform
les to recall
Source Selects thesourcewhentheTrigger
Typeoptionis set toEdge
Videolinenumber Sets theoscilloscopetoaspecic line
number whentheTrigger Typeoption
is set toVideoandtheSync optionis
set toLineNumber
Trigger
Pulsewidth Sets thewidthof thepulsewhenthe
Trigger Typeoptionisset toPulse
Fileselection Selects lestorenameor delete; (See
page 102, File Utilities for the USB
Flash Drive.)
Utility FileUtilities
Nameentry Renames the le or folder; (See
page103, Rename File or Folder.)
Utility Options
GPIBSetup Address
Valueentry Sets the GPIB address for the
TEK-USB-488adapter
Utility Options Set
DateandTime
Valueentry Sets thevaluefor thedateandtime;
(Seepage101, Setting the Date and
Time.)
16 TBS1000 Series Oscilloscopes User Manual
Operating Basics
Active menu or option Knob function Description
Vertical Probe
Voltage Attenuation
Valueentry For achannel menu(suchas theCH
1menu), sets theattenuationfactor in
theoscilloscope
Vertical Probe
Current Scale
Valueentry For a of channel menu (such as the
CH1 menu), sets the scale in the
oscilloscope
AutoRange. Displays the Autorange Menu, and activates or deactivates the
autoranging function. When autoranging is active, the adjacent LED lights.
Save/Recall. Displays the Save/Recall Menu for setups and waveforms.
Measure. Displays the automated measurements menu.
Acquire. Displays the Acquire Menu.
Ref. Displays the Reference Menu to quickly display and hide reference
waveforms stored in the oscilloscope non-volatile memory.
Utility. Displays the Utility Menu.
Cursor. Displays the Cursor Menu. Cursors remain visible (unless the Type
option is set to Off) after you leave the Cursor Menu but are not adjustable.
Display. Displays the Display Menu.
Help. Displays the Help Menu.
Default Setup. Recalls the factory setup.
AutoSet. Automatically sets the oscilloscope controls to produce a usable display
of the input signals.
Single. (Single sequence) Acquires a single waveform and then stops.
Run/Stop. Continuously acquires waveforms or stops the acquisition.
Starts the print operation to a PictBridge compatible printer, or performs the
Save function to the USB ash drive.
Save. An LED indicates when the print button is congured to save data to the
USB ash drive.
TBS1000 Series Oscilloscopes User Manual 17
Operating Basics
Input Connectors
1 & 2. Input connectors for waveform display.
Ext Trig. Input connector for an external trigger source. Use the Trigger Menu to
select the Ext, or Ext/5 trigger source. Push and hold the Trig View button to see
how the trigger settings affect the trigger signal, such as trigger coupling.
Other Front-Panel Items
USB Flash Drive port
USB Flash Drive Port. Insert a USB ash drive for data storage or retrieval. The
oscilloscope displays a clock symbol to indicate when the ash drive is active.
After a le is saved or retrieved, the oscilloscope removes the clock, and displays
a hint line to notify you that the save or recall operation is complete.
For ash drives with an LED, the LED blinks when saving data to or retrieving
data from the drive. Wait until the LED stops to remove the drive.
PROBE COMP. Probe compensation output and chassis reference. Use to
electrically match a voltage probe to the oscilloscope input circuit. (See page 5,
Voltage Probe Check Wizard.) (See page 7, Manual Probe Compensation.)
18 TBS1000 Series Oscilloscopes User Manual
Understanding Oscilloscope Functions
This chapter contains general information that you need to understand before
you use an oscilloscope. To use your oscilloscope effectively, you need to learn
about the following functions:
Setting up the oscilloscope
Triggering
Acquiring signals (waveforms)
Scaling and positioning waveforms
Measuring waveforms
The next gure shows a block diagram of the various functions of the oscilloscope
and their relationships to each other.
Setting Up the Oscilloscope
You should become familiar with several functions that you may use often when
operating your oscilloscope: Autoset, Autorange, saving a setup, and recalling
a setup.
Using Autoset
Each time you push the AutoSet button, the Autoset function obtains a stable
waveform display for you. It automatically adjusts the vertical scale, horizontal
scale and trigger settings. Autoset also displays several automatic measurements
in the graticule area, depending on the signal type.
Using Autorange
Autorange is a continuous function that you can enable or disable. The function
adjusts setup values to track a signal when the signal exhibits large changes or
when you physically move the probe to a different point.
TBS1000 Series Oscilloscopes User Manual 19
Understanding Oscilloscope Functions
Saving a Setup
The oscilloscope saves the current setup if you wait ve seconds after the last
change before you power off the oscilloscope. The oscilloscope recalls this setup
the next time you apply power.
You can use the Save/Recall Menu to save up to ten different setups.
You can also save setups to a USB ash drive. The oscilloscope accommodates
a USB ash drive for removable data storage and retrieval. (See page 63, USB
Flash Drive Port.)
Recalling a Setup
The oscilloscope can recall the last setup before the oscilloscope was powered off,
any saved setups, or the default setup. (See page 90, Save/Recall.)
Default Setup
The oscilloscope is set up for normal operation when it is shipped from the
factory. This is the default setup. To recall this setup, push the Default Setup
button. To view the default settings, refer to Appendix D: Default Setup.
Triggering
The trigger determines when the oscilloscope starts to acquire data and to display
a waveform. When a trigger is set up properly, the oscilloscope converts unstable
displays or blank screens into meaningful waveforms.
Triggeredwaveform Untriggered waveforms
For oscilloscope-specic descriptions, refer to the Operating Basics chapter. (See
page 15, Trigger Controls.) Refer also to the Reference chapter. (See page 95,
Trigger Controls.)
When you push the Run/Stop or Single button to start an acquisition, the
oscilloscope goes through the following steps:
1. Acquires enough data to ll the portion of the waveform record to the left of
the trigger point. This is called the pretrigger.
2. Continues to acquire data while waiting for the trigger condition to occur.
3. Detects the trigger condition.
20 TBS1000 Series Oscilloscopes User Manual
Understanding Oscilloscope Functions
4. Continues to acquire data until the waveform record is full.
5. Displays the newly-acquired waveform.
NOTE. For Edge and Pulse triggers, the oscilloscope counts the rate at which
trigger events occur to determine trigger frequency. The oscilloscope displays
the frequency in the lower right corner of the screen.
Source
You can use the Trigger Source options to select the signal that the oscilloscope
uses as a trigger. The source can be the AC power line (available only with Edge
triggers), or any signal connected to a channel BNC or to the Ext Trig BNC.
Types
The oscilloscope provides three types of triggers: Edge, Video, and Pulse Width.
Modes
You can select the Auto or the Normal trigger mode to dene how the oscilloscope
acquires data when it does not detect a trigger condition. (See page 95, Mode
Options.)
To perform a single sequence acquisition, push the Single button.
Coupling
You can use the Trigger Coupling option to determine which part of the signal
will pass to the trigger circuit. This can help you attain a stable display of the
waveform.
To use trigger coupling, push the Trig Menu button, select an Edge or Pulse
trigger, and select a Coupling option.
NOTE. Trigger coupling affects only the signal passed to the trigger system. It
does not affect the bandwidth or coupling of the signal displayed on the screen.
To view the conditioned signal being passed to the trigger circuit, push and hold
down the Trig View button.
Position
The horizontal position control establishes the time between the trigger and the
screen center. Refer to Horizontal Scale and Position; Pretrigger Information
for information on how to use this control to position the trigger. (See page 23,
Horizontal Scale and Position; Pretrigger Information.)
Slope and Level
The Slope and Level controls help to dene the trigger. The Slope option (Edge
trigger type only) determines whether the oscilloscope nds the trigger point on
the rising or the falling edge of a signal. The Trigger Level knob controls where
on the edge the trigger point occurs.
TBS1000 Series Oscilloscopes User Manual 21
Understanding Oscilloscope Functions
Rising edge Falling edge
Trigger level canbeadjusted
vertically
Trigger can be rising or falling
Acquiring Signals
When you acquire a signal, the oscilloscope converts it into a digital form and
displays a waveform. The acquisition mode denes how the signal is digitized,
and the time base setting affects the time span and level of detail in the acquisition.
Acquisition Modes
There are three acquisition modes: Sample, Peak Detect, and Average.
Sample. In this acquisition mode, the oscilloscope samples the signal in evenly
spaced intervals to construct the waveform. This mode accurately represents
signals most of the time.
However, this mode does not acquire rapid variations in the signal that may occur
between samples. This can result in aliasing, and may cause narrow pulses to be
missed. In these cases, you should use the Peak Detect mode to acquire data.
(See page 23, Time Domain Aliasing.)
Peak Detect. In this acquisition mode, the oscilloscope nds the highest and
lowest values of the input signal over each sample interval and uses these values
to display the waveform. In this way, the oscilloscope can acquire and display
narrow pulses, which may have otherwise been missed in Sample mode. Noise
will appear to be higher in this mode.
Average. In this acquisition mode, the oscilloscope acquires several waveforms,
averages them, and displays the resulting waveform. You can use this mode to
reduce random noise.
Time Base
The oscilloscope digitizes waveforms by acquiring the value of an input signal
at discrete points. The time base allows you to control how often the values are
digitized.
To adjust the time base to a horizontal scale that suits your purpose, use the
Horizontal Scale knob.
22 TBS1000 Series Oscilloscopes User Manual
Understanding Oscilloscope Functions
Scaling and Positioning Waveforms
You can change the display of waveforms by adjusting the scale and position.
When you change the scale, the waveform display will increase or decrease in size.
When you change the position, the waveform will move up, down, right, or left.
The channel indicator (located on the left of the graticule) identies each
waveform on the display. The indicator points to the ground reference level of
the waveform record.
You can view the display area and readouts. (See page 9, Display Area.)
Vertical Scale and Position
You can change the vertical position of waveforms by moving them up or down in
the display. To compare data, you can align a waveform above another or you can
align waveforms on top of each other.
You can change the vertical scale of a waveform. The waveform display will
contract or expand relative to the ground reference level.
For oscilloscope-specic descriptions, refer to the Operating Basics chapter. (See
page 13, Vertical Controls.) Refer also to the Reference chapter. (See page 103,
Vertical Controls.)
Horizontal Scale and
Position; Pretrigger
Information
You can adjust the Horizontal Position control to view waveform data before
the trigger, after the trigger, or some of each. When you change the horizontal
position of a waveform, you are actually changing the time between the trigger
and the center of the display. (This appears to move the waveform to the right
or left on the display.)
For example, if you want to nd the cause of a glitch in your test circuit, you
might trigger on the glitch and make the pretrigger period large enough to capture
data before the glitch. You can then analyze the pretrigger data and perhaps nd
the cause of the glitch.
You change the horizontal scale of all the waveforms by turning the Horizontal
Scale knob. For example, you might want to see just one cycle of a waveform to
measure the overshoot on its rising edge.
The oscilloscope shows the horizontal scale as time per division in the scale
readout. Since all active waveforms use the same time base, the oscilloscope only
displays one value for all the active channels, except when you use Window
Zone. Refer to Window Zone for information on how to use the window function.
(See page 86, Window Zone.)
For oscilloscope-specic descriptions, refer to the Operating Basics chapter. (See
page 14, Position.) Refer also to the Reference chapter. (See page 85, Horizontal.)
Time Domain Aliasing. Aliasing occurs when the oscilloscope does not sample the
signal fast enough to construct an accurate waveform record. When this happens,
TBS1000 Series Oscilloscopes User Manual 23
Understanding Oscilloscope Functions
the oscilloscope displays a waveform with a frequency lower than the actual input
waveform, or triggers and displays an unstable waveform.
Actual high-frequency waveform
Apparent low-frequency
waveformdue to aliasing
Sample points
The oscilloscope accurately represents signals, but is limited by the probe
bandwidth, the oscilloscope bandwidth, and the sample rate. To avoid aliasing,
the oscilloscope must sample the signal more than twice as fast as the highest
frequency component of the signal.
The highest frequency that the oscilloscope sampling rate can theoretically
represent is the Nyquist frequency. The sample rate is called the Nyquist rate, and
is twice the Nyquist frequency.
The oscilloscope maximum sample rates are at least ten times the bandwidth.
These high sample rates help reduce the possibility of aliasing.
There are several ways to check for aliasing:
Turn the horizontal Scale knob to change the horizontal scale. If the shape of
the waveform changes drastically, you may have aliasing.
Select the Peak Detect acquisition mode. (See page 22, Peak Detect.) This
mode samples the highest and lowest values so that the oscilloscope can
detect faster signals. If the shape of the waveform changes drastically, you
may have aliasing.
If the trigger frequency is faster than the display information, you may have
aliasing or a waveform that crosses the trigger level multiple times. Examining
the waveform allows you to identify whether the shape of the signal is going
to allow a single trigger crossing per cycle at the selected trigger level.
If multiple triggers are likely to occur, select a trigger level that will generate
only a single trigger per cycle. If the trigger frequency is still faster than the
display indicates, you may have aliasing.
If the trigger frequency is slower, this test is not useful.
If the signal you are viewing is also the trigger source, use the graticule or the
cursors to estimate the frequency of the displayed waveform. Compare this to
the Trigger Frequency readout in the lower right corner of the screen. If they
differ by a large amount, you may have aliasing.
The next table lists the time base settings that you can use to avoid aliasing at
various frequencies and the respective sample rate. At the fastest horizontal scale
setting, aliasing is not likely to occur due to the bandwidth limitations of the
oscilloscope input ampliers.
24 TBS1000 Series Oscilloscopes User Manual
Understanding Oscilloscope Functions
Settings to avoid aliasing in Sample mode
Time base Samples per second Maximum
5.0to250.0ns 1 GS/s 150.0MHz