Sanitary Dual Tube Coil: High Efficiency Tube-In-Tube Heat Exchanger

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Continuous tube-in-tube

Interior surface finish of R


a
< 20
Fully drainable inner and outer coil
Spiral wound for maximum counterflow ef ficiency
Highly resistant to thermal and hydraulic shock
The Sentry DTC-6 dual tube coil is specifically
designed to cool or heat samples that require a high
level of purit y. The design provides the benefits of a
compact high ef ficiency helical coil heat exchanger
while maintaining a high level of surface finish inside
the sample tube. This is achieved by using a highly
polished (better than R
a
10) tube as the material for
creating the helical coil. Af ter specially controlled
coiling, the ID surface maintains a R
a
< 20. Tri-clover


sample connections are orbitally welded to the sample
tube and electropolished. Contact factory for other
sanitary dual tube coil configurations.
All other features of the DTC-6 are similar to other
Sentry dual tube coils including our well-known qualit y
workmanship and ef ficiency.
Sentry of fers a variet y of dual tube coils with a choice
of heat transfer area, materials of construction and
shell design, as well as, a variet y of other sample
conditioning products that help assure the most
representative sample possible.
www.sentry-equip.com
17.2 Rev. 3 08/12
S
A
N
I
T
A
R
Y

D
U
A
L

T
U
B
E

C
O
I
L
SANITARY DUAL TUBE COIL
Description Features
High Efficiency Tube-in-Tube
Heat Exchanger
Full counter flow heat exchangers
suited for a variety of applications
Pressure Rating
Max. 150 psi @450F (10 Bar @ 232C) inner tube*
1100 psi # 200F (82 Bar @ 93C) outer tube
End Connections 1/2" Tri-clamp
Wetted Materials 0.375" x .035" wall, SA249 316SS, polished to R
a
< 20
Heating Surface 1.9 f t
2
(0.18 M
2
)
Weight 16 lbs (7.3 kgs)
Model Number DTC-SSG/SSC-6-4-1
Part Number 7-02672F
*inner tube ratings determined by tri-clamp connector and elastomer material used.
SPECIFICATIONS
TECHNICAL DATA
COOLANT OUT
COOLANT IN
FLUID OUT
FLUID IN
9/32" [7] DIA. HOLES
4 1/4" [108]
1 5/8" [41]
7" [178] 7" [178]
1" [25]
8 1/2" [216]
2" [51]
9" [229] 13" [330]
Sentry Equipment Corp 2009 17.2 Rev. 3 08/12
5
10
15
20
25
30
35
40
45
50
A
p
p
r
o
a
c
h

T
e
m
p
e
r
a
t
u
r
e

(
o
C
)
4
3
2
1
P
r
e
s
s
u
r
e

D
r
o
p

(
P
S
I
)
1

P
S
I

=

0
.
0
6
9

B
a
r
0
10
20
30
40
50
60
70
80
90
A
p
p
r
o
a
c
h

T
e
m
p
e
r
a
t
u
r
e

(
o
F
)
1600 1800 2000 2200 2400 2600
CCM
0.4 0.5 0.6 0.7
GPM
200 220 240 260 280 300 320 340 360
Lbm/Hr
Model DTC-6
Surface Area 1.90 Ft.
2
[0.18 m
2
]
5 GPM [18.9 L/min] Cooling Water 95
o
F [35
o
C]
18 PSI [1.2 Bar] Pressure Drop
Inlet Tem
p. 300
o
F (148.9
o
C)
4
0
0
o
F
(2
0
4
.4
o
C
)
5
0
0
o
F


(
2
6
0
.
0
o
C
)
P
r
e
s
s
u
r
e


D
r
o
p

(
P
S
I
)
Flow Rate
Performance data shows approach temperature
(outer tube inlet + approach temp. = inner tube outlet temp.)
for three inner tube inlet temperatures and various flow rates.
Data based on thermal properties of water. Cooling water 95F
(35C) @ 5 GPM (18.9 L/min). Pressure drop <18 psi (1.2 bar).
Consult factory for other fluids flow rates and pressure drops.
For further information, contact:
It is solely the responsibility of the end-user, through its own analysis and testing, to select products and materials suitable for their specific
application requirements, ensure they are properly installed, safely applied, properly maintained, and limit their use to their intended purpose.
Improper selection, installation, or use may result in personal injury or property damage.
SENTRY EQUIPMENT CORP
966 Blue Ribbon Circle North
Oconomowoc, WI 53066 USA
Tel: 262-567-7256 Fax: 262-567-4523
E-mail:
[email protected]
Website:
www.sentry-equip.com

You might also like