Manual Hardw e
Manual Hardw e
Manual Hardw e
1. Product Overview
1.1 Display Large LCD display for measured values and status
Time (except for TN2) Alarm (will stay lit until the alarm is reset) Upper / Lower limit value
Sensor number Alarm indicator for sensor 1.. 4 Figure 1 Communication Battery low ( Hints) Normal operation + Wait = IdLE ( Statuscodes)
F F
Figure 2
1.3 Battery
3.6 Volt lithium battery with extended temperature range. For more information about the operating range of the battery and the battery lifetime see chapter 2, Application and Safety Regulations.
DB15 connector
Battery case
Position of internal temperature sensor (TN2 & TN3-P only) D-EZ-6002Bd 8. 2000
English
3. Mounting Instructions
As accessories ELPRO provides a protective housing made of shock proof plastic material with IP66, Part No. 2350-xx and 3 differenterent brackets for simple fixation of dataloggers. 3.1 Fixation Elements Simple fixation bracket Part No. 2804-A
86 for M3
Figure 4 93
57
DB15 connector Part. No. 3032-xx 3215-xx Figure 5 Padlock to prevent unauthorized removal of the logger
for M4 / M5
93
Mounting bracket with screw terminals and cover Part No. TNx: 2804-C 2804-CR 2804-CRA THx: 2805-CR 2805-CRA
57
xxxx-CR with additional RS232 connector xxxx-CRA with additional RS232 connector and acoustic alarm
Seite - 12 Getting Started ECOLOG D-EZ-6002Bd 8. 2000
3.2 Pin Assignment and Connection Diagram Wire the DB15 connector for the ECOLOG as shown below:
English
TN2
DB15 connector 8 7 6 5 4 3 2 1 GND TXD NTC2 +Ref Res. Res. +Ref. NTC1 15 14 13 12 11 10 9 RXD Busy Res. Start/Inpos Ubat. Alarm Reset Res.
Ubat Start/Inpos
Start/Inpos Connector
TN3-P
DB15 connector 8 7 6 5 4 3 2 1 GND TXD NTC2 +Ref Res. NTC3 +Ref. NTC1 15 14 13 12 11 10 9 RXD Busy Res. Start/Inpos Ubat. Alarm Reset Alarm
Ubat Start/Inpos
Start/Inpos Connector
TN4 TN4-L
DB15 connector 8 7 6 5 4 3 2 1 GND TXD NTC2 +Ref NTC4 NTC3 +Ref. NTC1 15 14 13 12 11 10 9 RXD Busy Res. DigitIn1 Ubat. DigitIn2 Alarm
Connection diagram NTC +Ref Ubat DigitIn Alarm GND NTC - Sensor
Contact Input
+
0..50VDC max 200mA Alarm Output
PT100 - Sensor
Contact Input
+
0..50VDC max 200mA Alarm Output
TH1 TH2
DB15 connector 8 7 6 5 4 3 2 1 GND TXD +Ref. A1 A2 B1,2 D2 C1,2 15 14 13 12 11 10 9 RXD Busy NTC1 NTC2 D1 DigitIn Alarm
Contact Input
+
0..50VDC max 200mA Alarm Output
Seite - 13 8. 2000
4. Operation
4.1 Main Menu
Start (only appears when the logger is in external start/ stop mode and is waiting for the external start signal) Resetting the alarm display (only appears when an alarm display exists) No function (exit menu) THx: Time Stamp marked as D2 TNx: No function (exit menu) TPx: No function (exit menu) Printing the short protocol Scrolling the measured values
Submenu A
Submenu B Prints: Backwards starting with displayed measured value Stop: Press menu button
Submenu C
1x backward
1x forward
Step size: 10 measurements 10x backward Step size: 100 measurements 100x backward Print data and exit menu
10x forward
100x forward
backward
forward
Corresponds to submenu A
Year minus Month Day Hour Minute Load settings and exit menu plus
minus
plus
Seite - 15 8. 2000
English
By activating this check box, the alarm function is switched on. Data entry field for lower and uper limit values. The switching hysteresis is used to avoid uncontrolled on/off switching of the contact. To move the contact, the measurement value has to differ from the limit value by the value defined in this field. Alarm Delay Time - normal To close the alarm contact, the duration of a measurement value, outside the limit values, has to be longer than the delay time defined in the "normal" data entry field. - defrost on If the "check defrost" function is selected and the defrost contact D1 is closed, then the duration of a measurement value, outside the limit values, has to be longer than the delay time defined in the "defrost on" data entry field. Alarm Output - no selection This mode is used for all applications where the alarm contact is used to control any external devices like a flash light or a telephon dailing unit. The contact will be closed as long as a limit value is exeded. - using CRA bracket This mode is used in combination with a 280x-CRA bracket only. - self-sustaining This mode is used for all applications where the alarm contact is used to control any external devices like a flash light or a telephon dailing unit. The contact will be closed till the alarm is reseted by the user . Close / Download / Print Status These boutons are used to program the logger, to printout the current status of the logger and to close the Setup of Alarm Parameters window. Seite - 16 Getting Started ECOLOG D-EZ-6002Bd 8. 2000
All dataloggers except TN2 and TN3-P are able to printout complete measurement value tables on a protocol printer. Customer Module Tag
Ld. Prn b.F. CAL.E. C.F. Err HHHH L.C. Ld.F. n.c. S.C. O.F. U.F.
8. Hints
TNx TH1 & TH2 TP1 & TP2 Battery low Software CD-ROM Battery life-span approx. 2 years Please do not calibrate with 2 connected rH/T sensors Battery life-span approx. 2 years Battery life-span approx. 1.5 years, depending on measurement interval and resolution. This indicator (see page 10) is going to be switched on about 6 month before the battery life ends. Please change the battery at your next possibility. ECOLOG data loggers require an evaluation software of version 2.05 or higher. For more product information please see elproLOG WIN CD-ROM.
9. Measuring Ranges
Temperature TNx and THx with sensor Range -20C.. +50C -50C.. +90C +90C.. +140C Range -100C.. +400C -200C.. -100C +400C.. +500C +500C.. +550C Range 0% rH .. 100% rH Resolution 0.1C 0.2C 0.2C Accuracy 0.2C 0.4C 1C
Resolution max. 0.1C 0.2C 0.1C 0.2C Resolution 0.2% rH Accuracy At room temperature, 23C: 1.5%rH Hysteresis 10-90-10%rH: <1%rH Temp. coefficient: 0.05% rH/C D-EZ-6002Bd 8. 2000