USB2000 Operating Instructions
USB2000 Operating Instructions
USB2000 Operating Instructions
Offices:
Trademarks
Microsoft, Windows, Windows 95, Windows 98, Windows ME, Windows NT, Windows 2000, Windows XP and Excel are either registered trademarks or trademarks of Microsoft Corporation.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The information provided is on an as is basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity with respect to any loss or damages arising from the information contained in this manual.
Table of Contents
1 USB2000 Introduction .....................................................................................................................1
Product Overview .................................................................................................................................................. 1
System Requirements ...................................................................................................................................................................................1 EEPROM Utilization ......................................................................................................................................................................................1 About OOIBase32 .........................................................................................................................................................................................2 Sampling System Overview ..........................................................................................................................................................................2 Modular Sampling Accessories .....................................................................................................................................................................2
USB2000 Specifications..................................................................................................................5
How the USB2000 Works ...................................................................................................................................... 5
USB2000 Component Table .........................................................................................................................................................................6
Troubleshooting ............................................................................................................................14
USB2000 Connected to PC Prior to OOIBase32 Installation .............................................................................. 14
Remove the Unknown Device from Windows Device Manager ..................................................................................................................14 Windows 98:................................................................................................................................................................................................15 Windows 2000/XP: ......................................................................................................................................................................................16
Troubleshooting the Serial Port Configuration..................................................................................................... 17 Older Version of OOIBase32 Installed ................................................................................................................ 17
Absorbance Experiments..................................................................................................................................... 20 Transmission Experiments .................................................................................................................................. 22 Reflection Experiments........................................................................................................................................ 24 Relative Irradiance Experiments.......................................................................................................................... 26 Time Acquisition Experiments ............................................................................................................................. 28
Appendix A:
Index ......................................................................................................................................................38
USB2000 Introduction
USB2000 Introduction
The following chapter contains introductory information about the USB2000 Spectrometer, shipment information, and spectrometer connectivity.
Product Overview
The USB2000 Spectrometer connects to a notebook or desktop PC via USB port or serial port. When connected to the USB port of a PC, the USB2000 draws power from the host PC, eliminating the need for an external power supply.
System Requirements
You can use the USB2000s USB connectivity with any PC that meets the following requirements: Windows 98/Me/2000/XP operating system (or Windows CE 2.11 or later for palm-sized PCs) Ocean Optics OOIBase32 software application installed and configured for use with the USB2000 (OOIPS2000 software if using a palm-sized PC). Consult the Configuring the USB2000 in OOIBase32 section of Chapter 3: Installing the USB2000 for specific configuration instructions.
Alternately, the USB2000 has a serial port for connecting to PCs, PLCs, and other devices that support the RS232 communication protocol. However, this connection method requires an external power supply to power the USB2000.
EEPROM Utilization
An EEPROM memory chip in each USB2000 contains wavelength calibration coefficients, linearity coefficients, and a serial number unique to each individual spectrometer. The OOIBase32 software application reads these values directly from the spectrometer, enabling the ability to hot-swap spectrometers between PCs without entering the spectrometer coefficients manually on each PC.
USB2000 Introduction
About OOIBase32
OOIBase32 is the latest generation of operating software for all Ocean Optics spectrometers and is available free to all customers. OOIBase32 is a user-customizable, advanced acquisition and display program that provides a real-time interface to a variety of signal-processing functions. With OOIBase32, you have the ability to perform spectroscopic measurements (such as absorbance, reflectance, and emission), control all system parameters, collect and display data in real time, and perform reference monitoring and time acquisition experiments. Note: When using a palm-sized PC, you will use the OOIPS2000 software instead of OOIBase32. The functionality of OOIPS2000 is similar to OOIBase32, but it is specifically for the palm-sized PC.
How Sampling Works The following list explains the function of Ocean Optics sampling components in the sampling process: 1. The user stores reference and dark measurements to correct for instrument response variables. 2. The light from the light source transmits through an optical fiber to the sample. 3. The light interacts with the sample. 4. Another optical fiber collects and transmits the result of the interaction to the spectrometer. 5. The spectrometer measures the amount of light and transforms the data collected by the spectrometer into digital information. 6. The spectrometer passes the sample information to OOIBase32. 7. OOIBase32 compares the sample to the reference measurement and displays processed spectral information.
USB2000 Introduction
Interface Options
The USB2000 has both USB and serial port connectors, enabling you to connect the spectrometer to a desktop or notebook PC via a USB port or to a desktop, notebook, or to a palm-sized PC via a serial port.
Computer Interface Desktop or Notebook PC via USB Port Desktop or Notebook PC via Serial Port Palm-sized PC via Serial Port Operating System Requirements Windows 98/Me/ 2000/XP Any 32-bit Windows operating system Windows CE 2.11 or higher Part Needed USB-CBL-1 (included) USB-ADP-PC (not included) USB-ADP-H (not included) Description of Part Cable that connects from USB port on USB2000 to USB port on desktop or notebook PC Adapter block that enables connection from serial port on USB2000 to serial port on desktop or notebook PC; comes with 5 VDC power supply (required when connecting to serial port) Adapter block that enables connection (with standard 9-pin serial cable) from serial port on USB2000 to serial port on palmsized PC; comes with 5 VDC power supply (required when connecting to serial port)
Shipment Components
The following information and documentation ships with the USB2000 Spectrometer:
Packing List
The packing list is inside a plastic bag attached to the outside of the shipment box (the invoice arrives separately). The packing slip lists all items in the order, including customized components in the spectrometer (such as the grating, detector collection lens, and slit). The packing list also includes the shipping and billing addresses, as well as any items on back order.
USB2000 Introduction
Other Documentation
You can find detailed instructions for the OOIBase32 Spectrometer Operating Software at http://www.oceanoptics.com/technical/operatinginstructions.asp. Note: You can find instructions for the OOIPS2000 Operating Software for the handheld PC at the following location: http://www.oceanoptics.com/products/ooips2000.asp.
To find operating instructions on some of Ocean Optics most popular spectroscopic accessories (including light sources, sampling chambers, sampling optics, and probes), consult the operating instructions for Ocean Optics products from the Software and Technical Resources CD that ships with the product, or visit our website at http://www.oceanoptics.com/technical/operatinginstructions.asp.
USB2000 Specifications
USB2000 Specifications
This chapter contains information on spectrometer operation, specifications, and system compatibility. It also includes accessory connector pinout diagrams and pin-specific information.
The USB Component Table on the following page explains the function of each numbered component in the USB2000 Spectrometer diagram (Figure 2-1).
USB2000 Specifications
Slit*
Filter*
Collimating Mirror
Grating*
Focusing Mirror
USB2000 Specifications
USB2000 Specifications
The following sections provide specification information for the CCD Detector in the USB2000, as well as the USB2000 Spectrometer itself.
Operating systems:
USB2000 Specifications
System Compatibility
The following sections provide information on hardware and software requirements for the USB2000: Compatibility for Desktop or Notebook PCs To use the USB2000, you must have a PC that meets the following minimum requirements: IBM-compatible PC with Pentium (or higher) processor 32 MB RAM OOIBase32 Spectrometer Operating Software Windows 98/Me/2000/XP operating system (when connecting the USB2000 to a PC via USB port) or Any 32-bit version of Windows (when connecting the USB2000 to a PC via serial port) Compatibility for Palm-sized PCs In order to use the USB2000 with your palm-sized PC, the computer must meet the following requirements: Palm-sized PC running Windows CE 2.11 or higher 32 MB OOIPS2000 Spectrometer Operating Software Serial port connectivity
10-pin Connector
USB2000 Specifications
If you are customizing your USB2000 Spectrometer system or configuring an External Triggering configuration, you may need these part numbers to complete your setup.
3
Note:
This chapter contains instructions in parallel for connecting the USB2000 via both USB and serial modes.
USB2000 Installation
USB Mode To connect the USB2000 to a PC via the USB port, the PC must be running the Windows 98/ME/2000/XP operating system. Note: The USB port on a PC can power up to five USB2000 spectrometer channels. Systems with more than five channels require a powered USB hub. Serial Port Mode To use the serial port capacity of the USB2000 Spectrometer, the PC must be running a 32-bit version of the Windows operating system (or Windows CE 2.11 or higher for palm-sized PCs). Follow the steps below to connect the USB2000 to the PC via serial port: 1. Connect the serial cable adapter block to the USB2000. 2. Connect one end of the 9-pin serial cable to the adapter block on the USB2000, and then connect the other end to a serial port on the PC. 3. Note the number of the serial port (COM Port) to which you connected the USB2000 (some PCs may not have numbered ports; palmsized PCs typically have only one serial port). 4. Plug the 5 VDC external power supply into an outlet and connect it to the USB2000. Note: Connecting the spectrometer to the PCs serial port requires that you manually enter the calibration coefficients from the Wavelength Calibration Data Sheet into OOIBase32 software (select Spectrometer | Configure | Wavelength Calibration tab). See the OOIBase32 documentation for more information.
Follow the steps below to connect the USB2000 to a PC via the USB port: 1. Install OOIBase32 on the destination PC. 2. Locate the USB cable (USB-CBL-1) provided with the USB2000. 3. Insert the square end of the cable into the side of the USB2000. 4. Insert the rectangular end of the cable into the USB port of the PC. If you installed OOIBase32 prior to connecting the USB2000, the Add New Hardware Wizard appears and installs the USB2000 drivers. If the drivers do not successfully install (or if you connected the USB2000 to the PC before installing OOIBase32), consult Chapter 4: Troubleshooting.
10
The following sections contain instructions on initially configuring the USB2000 the first time you start OOIBase32: Operator and Serial Number Dialog Box The Operator and Serial Number screen prompts you to enter a user name and software serial number into OOIBase. Some data files created by OOIBase32 during sampling procedures use this information in the file headers. Default Spectrometer Configuration File The Default Spectrometer Configuration File screen prompts you to select a spectrometer configuration (.SPEC) file for use with the USB2000. The unique serial number of the USB2000 precedes the file extension (for example, USB2J0162.SPEC). Navigate to the OOIBase32 installation directory and select the spectrometer configuration file. Configure Hardware Screen The Configure Hardware screen prompts you to enter spectrometer-specific information into OOIBase32 the first time you run the program. Typically, you will only enter this information the first time you run OOIBase32. However, you can alter the hardware configuration at any time using the Spectrometer Configuration screen. Select Spectrometer | Configuration from the OOIBase32 menu bar to access the Spectrometer Configuration screen. Follow the steps on the next page to configure the USB2000: Note: You do not need to configure the spectrometer hardware in the OOIPS2000 palm-sized PC operating software. Most palm-sized PCs have only one serial connector and do not have USB ports. Thus, OOIPS2000 will communicate with the spectrometer via this port at a fixed baud rate. There is no way to customize the USB2000 configuration with OOIPS2000.
11
USB Mode 1. Specify USB2000 in the Spectrometer Type drop-down menu. 2. Specify USB2000 in the A/D Converter Type drop-down menu. 3. Specify the serial number of the USB2000 under the USB Serial Number drop-down menu. Note: The system pre-fills this drop-down menu with the serial numbers of all discovered USB2000 Spectrometers.
Serial Port Mode 1. Select the USB2000 option from the Spectrometer Type drop-down menu. 2. Select the Serial (RS-232) A/D option from the A/D Converter Type drop-down menu. This selection enables serial-specific options in the lower portion of the Configure Hardware screen. 3. Select the COM port that the USB2000 is connected to in the Serial Port drop-down menu. Consult Chapter 4: Troubleshooting for information on identifying serial ports. 4. Select the speed at which the USB2000 will operate from the Baud Rate drop-down menu (115,200 baud is recommended). 5. Specify the pixel resolution (from 1 to 500) in the Pixel Resolution box. This value specifies that every nth pixel of the spectrometer will transmit from the USB2000 to the PC. Note: You can sacrifice pixel resolution to gain speed. The transfer of one complete spectra requires ~0.3 seconds at 115,200 baud.
6. Enable the Compress Data function to minimize the amount of data transferred over the RS-232 connection. The transmission of spectral data over the serial port is a relatively slow process. Enabling this function ensures that the USB2000 compresses every scan that it transmits. This greatly increases the data transfer speed of the USB2000. 7. Click the OK button to complete setup.
The spectrometer should now be able to acquire data and respond to light. Exit and restart OOIBase32 to save configuration data to disk.
12
Spectrometer Configuration Screen The Spectrometer Configuration screen prompts you to configure specific channel-level spectrometer information, if necessary. Select Spectrometer | Configure | Wavelength Calibration tab from the menu and set system parameters. If you have connected your spectrometer to the PCs USB port, OOIBase32 pre-fills the coefficients for the USB2000 from information on a memory chip in the spectrometer. Otherwise, you must manually type the coefficients as they are printed on the Wavelength Calibration Data Sheet that accompanied your spectrometer. Verify that the calibration coefficients match the coefficients from the Wavelength Calibration Data Sheet. If necessary, modify these values using the USB Programmer utility. Additionally, ensure that you select both the Master and Channel Enabled boxes. In the A/D Interface tab, enter the same values as in the Configure Hardware dialog box. OOIBase32 stores this information for future use once you close the program. Note: For information on using the USB2000 with the OOIPS2000 Operating Software for the palm-sized PC, visit http://www.oceanoptics.com/technical/palmspec.pdf.
13
Troubleshooting
Troubleshooting
The following sections contain information on troubleshooting issues you may encounter when using the USB2000 Spectrometer. Note: For issues encountered when using a palm-sized PC, consult the OOIPS2000 manual.
Windows 2000/XP: Click Start | Settings | Control Panel | System. Select the Hardware tab. Click on the Device Manager button.
2. Locate the Other Devices option and expand the Other Devices selection by clicking on the "+" sign to the immediate left. Note: Improperly installed USB devices can also appear under the Universal Serial Bus Controller option. Be sure to check this location if you cannot locate the unknown device.
14
Troubleshooting
3. Locate the unknown device (marked with a large question mark). Right-click on the Unknown Device listing and select the Uninstall or Remove option. 4. Click the OK button to continue. A warning box appears confirming the removal of the Unknown Device. Click the OK button to confirm the device removal. 5. Disconnect the USB2000 from your computer. 6. Locate the section in this chapter that is appropriate to your operating system and perform the steps in the Remove Improperly Installed Files section.
Windows 98:
Remove Improperly Installed Files: 1. Open Windows Explorer. 2. Navigate to the Windows | INF directory. If the INF directory is not visible, you will need to disable the Hide System Files and Folders option on in Windows Folder Options. Note: If the INF directory is not visible, you will need to disable the Hide System Files and Folders and Hide File Extensions for Known File Types options in Windows Folder Options. You can access Windows Folder Options from Windows Explorer, under the View | Options menu selection. 3. Delete the OOI_USB.INF file in the INF directory. 4. Navigate to the Windows | System32 | Drivers directory. 5. Delete the EZUSB.SYS file. 6. Reinstall your Ocean Optics application and reboot the system when prompted. 7. Plug in the USB device. The system will now be able to locate and install the correct drivers for the USB device.
15
Troubleshooting
Windows 2000/XP:
Remove Improperly Installed Files: 1. Open Windows Explorer. 2. Navigate to the Windows | INF directory. If the INF directory is not visible, you will need to disable the Hide System Files and Folders option on in Windows Folder Options. Note: If the INF directory is not visible, you will need to disable the Hide System Files and Folders and Hide File Extensions for Known File Types options in Windows Folder Options. You can access Windows Folder Options from Windows Explorer, under the Tools | Folder Options menu selection. 3. Delete the OOI_USB.INF and OOI_USB.PNF files in the INF directory. 4. Navigate to the Windows | System32 | Drivers directory. 5. Delete the EZUSB.SYS file. 6. Reinstall your Ocean Optics application and reboot the system when prompted. 7. Plug in the USB device. The system will now be able to locate and install the correct drivers for the USB device.
16
Troubleshooting
Windows 2000/XP
Windows NT
3. Double-click on the Ports (COM & LPT) option to display COM port numbers. Ensure that no warning icon appears next to the USB2000s COM port. 4. Verify that the COM port to which the USB2000 is interfaced is active. If the ports on the PC are not labeled and you do not know the COM port number, use trial-and-error to find the correct COM port. Open OOIBase32 and view the displayed graph. If the correct COM port is selected, you will see a dynamic trace responding to light near the bottom of the graph. If the correct COM port is not selected, you will see a straight line at zero counts. 5. Disable virus protection to ensure timely and complete data transfer (optional some computers require this step).
17
Sample Experiments
Sample Experiments
The following sections contain information on conducting sample experiments using the USB2000 and OOIBase32. For information on experiments with OOIPS2000, consult the OOIPS2000 Operating Instructions.
The type of measurement you will take determines the configuration of the sampling optics for your system. Furthermore, your choice of reference and data analysis determines how the OOIBase32 presents the results. Note: For each measurement, you must first take a reference and dark spectrum. After you take a reference and a dark spectrum, you can take as many measurement scans as needed. However, if you change any sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must store a new dark and reference spectrum.
18
Sample Experiments
Application Tips If the signal you collect is saturating the spectrometer (intensity greater than 4000 counts), you can decrease the light level on scale in scope mode by: Decreasing the integration time Attenuating the light going into the spectrometer Using a smaller diameter fiber Using a neutral density filter with the correct optical density
If the signal you collect has too little light, you can increase the light level on scale in scope mode by: Increasing the integration time Using a larger diameter fiber Removing any optical filters
19
Sample Experiments
Absorbance Experiments
Absorbance spectra are a measure of how much light a sample absorbs. For most samples, absorbance is linearly related to the concentration of the substance. OOIBase32 calculates absorbance (A) using the following equation
A = - log10
( R -D )
S - D
where S is the sample intensity at wavelength , D is the dark intensity at wavelength , R is the reference intensity at wavelength .
Figure 6-1: Typical absorbance setup. The light source (far right) sends light via an input fiber into a cuvette in a cuvette holder (bottom center). The light interacts with the sample. The output fiber carries light from the sample to the spectrometer (top center), which is connected to the PC (far left).
Absorbance is also proportional to the concentration of the substance interacting with the light (this is known as Beers Law). Common absorption applications include the quantification of chemical concentrations in aqueous or gaseous samples. Follow the steps below to take an absorbance measurement using OOIBase32: 1. Place OOIBase32 in scope mode by clicking the scope mode icon on the toolbar or selecting Spectrum | Scope Mode from the menu bar. 2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500 counts. If necessary, adjust the integration time until the intensity is approximately 3500 counts.
(Continued)
20
Sample Experiments
3. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a reference spectrum before measuring absorbance. Note: Do not put the sample itself in the path when taking a reference spectrum, only the solvent.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the menu bar to store the reference. This command merely stores a reference spectrum in memory. You must select File | Save | Reference from the menu bar to permanently save the spectrum to disk. 4. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to permanently save the spectrum to disk. Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your light source to store a dark spectrum, allow enough time for the lamp to warm up again before continuing your experiment. After the lamp warms up again, store a new reference (Step 3).
You must take a dark spectrum before measuring absorbance. 5. Put the sample in place and ensure that the light path is clear. Then, take an absorbance measurement by clicking on the Absorbance Mode icon on the toolbar or selecting Spectrum | Absorbance Mode from the menu. To permanently save the spectrum to disk, click the Save icon on the toolbar or select File | Save | Processed from the menu bar. Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must store a new dark and reference spectrum.
21
Sample Experiments
Transmission Experiments
Transmission is the percentage of energy passing through a sample relative to the amount that passes through the reference. Transmission Mode also displays the portion of light reflected from a sample, since transmission and reflection measurements use the same mathematical calculations. The transmission is expressed as a percentage (%T) relative to a standard substance (such as air). OOIBase32 calculates %T (or %R) with the following equation
%T =
S - D R - D
x 100%
where S is the sample intensity at wavelength , D is the dark intensity at wavelength , R is the reference intensity at wavelength .
Figure 6-2: Typical transmission setup. The light source (far right) sends light via the input leg of a transmission probe into a container (bottom center). The light interacts with the sample. The output leg of the transmission probe carries the information to the spectrometer (top center), which transmits the information to the PC (far left).
Common transmission applications include measuring light through solutions, optical filters, optical coatings, and other optical elements (such as lenses and fibers). Perform the following steps to take a transmission measurement using OOIBase32: 1. Place OOIBase32 in scope mode by clicking the Scope Mode icon on the toolbar or by selecting Spectrum | Scope Mode from the menu bar. 2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500 counts. If necessary, adjust the integration time until the intensity is approximately 3500 counts.
(Continued)
22
Sample Experiments
6. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a reference spectrum before measuring transmission. Note: Do not put the sample itself in the path when taking a reference spectrum, only the solvent.
Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the menu bar to store the reference. This command merely stores a reference spectrum in memory. You must select File | Save | Reference from the menu bar to permanently save the spectrum to disk. 3. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to permanently save the spectrum to disk. Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your light source to store a dark spectrum, allow enough time for the lamp to warm up again before continuing your experiment.
You must take a dark spectrum before measuring transmission. 4. Put the sample in place and verify that the light path is clear. Then, take a transmission measurement by clicking the Transmission Mode icon on the toolbar or selecting Spectrum | Transmission Mode from the menu bar. To save the spectrum to disk, click the Save icon on the toolbar or select File | Save | Processed from the menu bar. Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must store a new dark and reference spectrum.
23
Sample Experiments
Reflection Experiments
Reflection is the return of radiation by a surface, without a change in wavelength. Reflection can be: Specular (the angle of incidence is equal to the angle of reflection) Diffuse (the angle of incidence is not equal to the angle of reflection)
Every surface returns both specular and diffuse reflections. Some surfaces may return mostly specular reflection, while others may return mostly diffuse reflection. Specular reflection increases proportionately with the amount of gloss on a surface. Reflection is expressed as a percentage (%R) relative to the reflection from a standard reference substance
%R =
S - D R - D
x 100%
where S is the sample intensity at wavelength , D is the dark intensity at wavelength , R is the reference intensity at wavelength .
Figure 6-3: Typical reflection setup. A light source (far right) sends light via the input leg of a reflection probe onto a sample (bottom center). A reflection probe holder holds the probe in either a 90 or 45-degree angle from the surface. The output leg of the reflection probe carries light from the sample to the spectrometer (top center), which is connected to the PC (far left).
Common reflection applications include measuring the properties of mirrors and coatings. Other applications include measuring the visual properties of the color in paints, plastics, and food products.
24
Sample Experiments
Perform the following steps to take reflection measurements using OOIBase32: 1. Place OOIBase32 is in scope mode by clicking the Scope Mode icon on the toolbar, or by selecting Spectrum | Scope Mode from the menu bar. 2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500 counts. 3. Take a reference spectrum with the WS-1 Diffuse Reflectance Standard or the STAN-SSH Highreflectivity Reference Standard. You must take a reference spectrum before measuring reflection. Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the menu bar to store the reference. This command merely stores a reference spectrum in memory. You must select File | Save | Reference from the menu bar to permanently save the spectrum to disk. 4. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to permanently save the spectrum to disk. Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your light source to store a dark spectrum, allow enough time for the lamp to warm up again before continuing your experiment.
You must take a dark spectrum before measuring transmission. 5. Put the sample is in place and ensure that the light path is clear. Then, take a reflection measurement by clicking on the Transmission Mode icon on the toolbar or selecting Spectrum | Transmission Mode from the menu bar (since the mathematical calculations used to calculate transmission and reflection are identical). To save the spectrum to disk, click the Save icon on the toolbar or select File | Save | Processed from the menu bar. Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must store a new dark and reference spectrum.
25
Sample Experiments
I = B
( R -D )
S - D
where B is the relative energy of the reference (calculated from the color temperature) at wavelength , S is the sample intensity at wavelength , D is the dark intensity at wavelength , R is the reference intensity at wavelength .
Figure 6-4: Typical Relative Irradiance Setup. A light source with a known color temperature (such as the LS-1or LS-1-LL - lower right) is used to take a reference spectrum. The light to measure (lower left) is accumulated through a CC-3 Cosine Corrector (or FOIS integrating sphere) into an input fiber, which carries the light information to the spectrometer. The spectrometer then transmits the information to the PC, which compares the measured spectra against the reference spectrum, thus removing wavelength-dependent instrument response from the measurement.
Common applications include characterizing the light output of LEDs, incandescent lamps, and other radiant energy sources such as sunlight. Relative irradiance measurements also include fluorescence measurements, which measure the energy given off by materials that have been excited by light at shorter wavelengths.
26
Sample Experiments
Perform the following steps to take a relative irradiance measurement using OOIBase32: 1. Place OOIBase32 is in scope mode by clicking the Scope Mode icon on the toolbar, or by selecting Spectrum | Scope Mode from the menu bar. 2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about 3500 counts. Note: The light source must be a blackbody of known color temperature.
3. In the Reference Color Temperature dialog box, enter the color temperature of the light source (in Kelvin) and click the OK button. 4. Take a reference spectrum using a light source with a black body of a known color temperature, such as the LS-1. Click the Store Reference spectrum icon on the toolbar or select Spectrum | Store Reference from the menu bar to store the reference. This command merely stores a reference spectrum in memory. You must select File | Save | Reference from the menu bar to permanently save the spectrum to disk. 5. Block the light path to the spectrometer. Then, take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum | Store Dark from the menu bar. This command merely stores a dark spectrum in memory. You must select File | Save | Dark from the menu to permanently save the spectrum to disk. Note: If possible, do not turn off the light source when taking a dark spectrum. If you must turn off your light source to store a dark spectrum, allow enough time for the lamp to warm up again before continuing your experiment.
You must take a dark spectrum before measuring relative irradiance. 6. Position the fiber at the light source you wish to measure. Then, choose the Irradiance mode icon on the toolbar or select Spectrum | Relative Irradiance Mode from the menu bar. 7. Click the Save icon on the toolbar or select File | Save | Processed from the menu bar to save the spectrum to disk. Note: If you change any sampling variable (integration time, averaging, smoothing, angle, temperature, fiber size, etc.), you must store a new dark and reference spectrum.
27
Sample Experiments
d. Specify the analysis wavelength in the Wavelength (nm) box. e. Specify the number of pixels around the analysis wavelength to average in the Bandwidth (pixels) box. f. Select a multiplicative factor to apply to the data before plotting or storing. Then, select an additive constant or offset to apply to the data. OOIBase32 applies the additive constant or offset after applying the factor but before plotting or storing data. The equation for the Factor and Offset functions is: Results = (Factor * Data) + Offset 2. Configure a time acquisition process for the second single wavelength (if desired). Select the Channel B page and repeat Steps 1-3 for Channel B. To configure a time acquisition process for the third, fourth, fifth, and sixth single wavelengths, select the Channel C, Channel D, Channel E, and Channel F pages, respectively, and set the necessary parameters.
(Continued)
28
Sample Experiments
3. Configure a time acquisition process for a combination of two time channels (if desired) by selecting Combination 1. Perform the steps below to configure a combination: g. Select Enabled to set the time acquisition calculation for the wavelength. h. Enable Plotted to see a real-time graph of the acquired data in a spectral window. i. j. k. l. Specify Time Channel A through F for the First Channel. Select the mathematical operation to produce the data for Combination 1. Specify Time Channel A through F for the Second Channel. Select a multiplicative factor to apply to the data before plotting or storing. Then, select an additive constant or offset to apply to the data. OOIBase32 applies the additive constant or offset after applying the factor but before plotting or storing data. The equation for the Factor and Offset functions is: Results = (Factor * Data) + Offset 4. Configure a time acquisition process for the Combination 2 page, if desired. This page is virtually identical to the Combination 1 page, with the exception that you can choose Combination 1 for the first or second channel in Combination 2. 5. Click the Apply button to apply the changes, and then click the OK button to close the Time Acquisition Channel Configuration screen. Proceed to the Configuring the Configure Acquisition Screen section below. Configuring the Time Acquisition Configuration Screen 1. Select Time Acquisition | Configure | Configure Acquisition from the menu bar to open the Time Acquisition Configuration screen. 2. Enable Stream Data to Disk to save time acquisition data. 3. Enter a value in the Write Data to Disk Every X Acquisitions box to set the frequency for data saves. OOIBase32 saves data more frequently if the number is smaller, or less frequently if the number is larger. Entering a large number enhances the performance of the time acquisition process. Note: At specified time intervals, OOIBase32 stores data into time acquisition channels or combination channels. OOIBase32 can plot the data in a spectral window, or stream the data to disk, or both. OOIBase32 can display up to 2048 acquisitions in a spectral window. If OOIBase32 collects more than 2048 acquisitions, it only displays the last 2048. To store more than 2048 acquisitions, you must stream the data to disk. Writing data to the disk is a slow process (relative to the speed of some spectral acquisitions) and causes a decrease in system performance. However, writing data to disk more frequently gives a larger margin of safety. 4. Enable Show Values in Status Bar to see the time acquisition values in the status bar. These values replace the cursor values. 5. Name the Stream Filename for the time acquisition process. Clicking on the ellipsis to the right of this box opens a file save dialog box, allowing you to navigate to a designated folder. (Continued)
USB2000 Operating Instructions 29
Sample Experiments
6. Enable Save Every Acquisition to store data for every spectral acquisition during a time acquisition process (optional). Note: OOIBase32 has options to either store data for each acquisition, or to collect data only after a specified delay. Several factors affect the minimum time acquisition frequency, including integration time, number of spectrometer channels, samples averaged, and computer speed. If you instruct OOIBase32 to store data every 100 milliseconds, the delay between data acquisitions will be 100 milliseconds or more, depending on your experimental configuration. OOIBase32 spends a large amount of time calculating, rendering, and displaying the spectra in a spectral window. You can suspend the graph display, which greatly improves the performance of OOIBase32.
7. Enter an Initial Delay to set the delay preceding the time acquisition process. Keep in mind that the delay countdown does not begin until you start the time acquisition process. Be sure to select Hours, Minutes, Seconds, or Milliseconds immediately to the right of the initial delay entry. 8. Enter a value to set the Frequency of the data collected in a time acquisition process. OOIBase32 stamps data from a time acquisition with a time accurate to one millisecond. Be sure to select Hours, Minutes, Seconds, or Milliseconds immediately to the right of the frequency entry. You can enable the Save Every Acquisition box to store the acquisitions that occur at this frequency. See Step 6 for more information. 9. Enter a value to set the Duration for the entire time acquisition process. Be sure to select Hours, Minutes, Seconds, or Milliseconds to the right of the duration entry. Click the OK button to close the Time Acquisition Configuration dialog box. Then, enable Continue Until Manually Stopped, which instructs OOIBase32 to store data until you manually stop the acquisition process (optional).
30
Appendix A
Appendix A:
This Appendix describes how to calibrate the wavelength of your spectrometer. Though each spectrometer is calibrated before it leaves Ocean Optics, the wavelength for all spectrometers will drift slightly as a function of time and environmental conditions. Ocean Optics recommends periodically recalibrating the USB2000.
p = I + C1 p + C2 p2 + C3 p3
...where is the wavelength of pixel p, I is the wavelength of pixel 0, C1 is the first coefficient (nm/pixel), C2 is the second coefficient (nm/pixel2), and C3 is the third coefficient (nm/pixel3). You will be calculating the value for I and the three Cs.
A USB2000 spectrometer An optical fiber (for spectrometers without a built-in slit, a 50-m fiber works best) A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-order linear regressions If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak and AnalysisTookPak-VBA.
Note:
31
Appendix A
True Wavelength (nm) 253.65 296.73 302.15 313.16 334.15 365.02 404.66 407.78 435.84 546.07 576.96 579.07 696.54 706.72 727.29 738.40 751.47
Pixel # 175 296 312 342 402 490 604 613 694 1022 1116 1122 1491 1523 1590 1627 1669
Pixel #
Predicted Wavelength 253.56 296.72 302.40 313.02 334.19 365.05 404.67 407.78 435.65 546.13 577.05 579.01 696.70 706.62 727.24 738.53 751.27
Difference 0.09 0.01 -0.25 0.13 -0.05 -0.04 -0.01 0.00 0.19 -0.06 -0.09 0.06 -0.15 0.10 0.06 -0.13 0.19
5359375 25934336 30371328 40001688 64964808 117649000 220348864 230346397 334255384 1067462648 1389928896 1412467848 3314613771 3532642667 4019679000 4306878883 4649101309
116964 161604 240100 364816 375769 481636 1044484 1245456 1258884 2223081 2319529 2528100 2647129 2785561
(Continued)
32
Appendix A
5. Use the spreadsheet or calculator to calculate the wavelength calibration coefficients. In the spreadsheet program, find the functions to perform linear regressions. If using Quattro Pro, look under Tools | Advanced Math If using Excel, look under Analysis ToolPak
6. Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number squared, and the pixel number cubed as the independent variables (X). After executing the regression, you will obtain an output similar to the one shown below. Regression Statistics
Multiple R R Square Adjusted R Square Standard Error Observations 0.999999831 0.999999663 0.999999607 0.125540214 22
Intercept
Coefficients
Intercept X Variable 1 X Variable 2 X Variable 3 190.473993 0.36263983 -1.174416E-05 -2.523787E-09
Third coefficient
Standard Error
0.369047536 0.001684745 8.35279E-07 2.656608E-10 First coefficient
Second coefficient
The figure above notes the numbers of importance. 7. Record the Intercept, as well as the First, Second, and Third Coefficients. Additionally, look at the value for R squared. It should be very close to 1. If not, you have most likely assigned one of your wavelengths incorrectly. Keep these values at hand.
33
Appendix A
34
Appendix A
To Save Wavelength Calibration Coefficients Using the Serial Mode 1. Open the OOIBase32 application. 2. Select Spectrometer | Configure from the OOIBase32 menu bar. The Configure Spectrometer screen appears. 3. Select the Wavelength Calibration tab to update the wavelength coefficients within OOIBase32. 4. Enter in the new values acquired from Steps 5 and 6 of the Calibrating the Wavelength of the Spectrometer section in this Appendix. 5. Click the OK button to save the information in OOIBase32.
35
Appendix B
36
Appendix B
8. Enable Single Strobe output by checking the Strobe Enable check box (See Figure B-1 on the following page). The strobe signal is a rising edge trigger signal that is TTL High for the entire delay period. See Figure B-2 for a timing diagram. The USB2000-FLG is now configured for operation in Variable Delay Mode.
Figure B-1: The Acquisition Parameters Toolbar. Note the location of the Strobe Freq (entered in s) and Strobe/Lamp Enable options.
37
Index
A
Accessories, 13 Adobe Acrobat Reader, 4 AgPlus Mirrors, 36 OOIBase32, 11 Overview, 1
P C
Calibrating, 31 CCD Detector, 6 Compatibility, 8 Configure Hardware, 11 Packing List, 3 Passwords, 4 Pin Definitions, 9 Pinout, 8 Power, 7
D
Data transfer rate, 7 Default Spectrometer Configuration File, 11 Detector Collection Lens, 6 Dimensions, 7 Dynamic range, 7 range, 7 resolution, 7
S
Slit, 6 SMA Connector, 6 Software and Resources Library CD, 4 Specifications, 5
E
EEPROM, 1 Experiments, 18
Detector, 7 Spectrometer, 7
Spectroscopic Accessories, 13
T
Time Acquisition, 28 Timing Diagram, 37 Troubleshooting, 14
U
Upgrades, 4 USB2000-FLG, 36 USB-ADP-H, 3 USB-ADP-PC, 3 USB-CBL-1, 3
G
Grating, 6
I
Installed Filter, 6 Installing, 10 Variable Delay Mode, 36
L
L2 Detector Collection Lens, 6 Lens, 6
W
Wavelength Calibration, 31 Wavelength Calibration Data File, 3 Wavelength Calibration Data Sheet, 3 Weight, 7
M
memory chip, 1 Mirror, 6
38