Kane 940 Operating Manual
Kane 940 Operating Manual
Kane 940 Operating Manual
Kane International Limited Kane House, Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Tel: +44 (0) 1707 375550 Fax: +44 (0) 1707 393277 E-mail: [email protected] www.kane.co.uk Stock No: 18115 February 2006
CONTENTS
1. Page No: ANALYSER LAYOUT AND FEATURES............................................................ 4-7 1.1 1.2 1.3 1.4 2. 3. 4. Instrument features and keypad...........................................................................4 Instrument layout (Rear) .....................................................................................5 Standard Probe Configuration .............................................................................6 Analyser connections ..........................................................................................7
SAFETY WARNING ...................................................................................................8 FIRST TIME USE ........................................................................................................8 NORMAL START UP SEQUENCE...................................................................... 9-15 4.1 4.2 4.3 Every Time You Use The Analyser ....................................................................9 Automatic Calibration .........................................................................................9 Main Display Parameters ..................................................................................11 4.3.1 4 Page Mode.........................................................................................11 4.3.2 Line Scroll Mode..................................................................................12 4.3.3 8 Page Mode.........................................................................................12 Sampling the Flue Gas ......................................................................................13 Taking a Pressure Reading ................................................................................13 Regular Checks During Sampling.....................................................................14 Normal Shutdown Sequence .............................................................................14 Electromagnetic Compatibility .........................................................................15
MOVING THROUGH THE MENUS.................................................................. 16-22 5.1 5.2 Basic Operation .................................................................................................16 Menu Options and Settings ...............................................................................17 5.2.1 Main Menu ............................................................................................17 5.2.2 Select Menu ..........................................................................................17 5.2.3 Units Menu ............................................................................................19 5.2.4 Display Menu ........................................................................................20 5.2.5 Set-Up Menu .........................................................................................20
6.
PRINTING INFORMATION.....................................................................................23 6.1 Printing a Live Test..........................................................................................23 6.2 Standard Printout ................................................................................................23
7.
STORING AND RETRIEVING DATA .............................................................. 24-25 7.1 7.2 7.3 Storing a Live Test .........................................................................................24 Viewing and Printing a Stored Test................................................................24 Deleting Data.....................................................................................................25
-2-
8.
MAINTENANCE .......................................................................................................26 8.1 8.2 Emptying and Cleaning the in-line water trap ..................................................26 Changing the particle filter................................................................................26
9. 10. 11.
PROBLEM SOLVING ...............................................................................................27 ANNUAL RE-CALIBRATION .................................................................................27 PRODUCT SPECIFICATION ............................................................................. 28-29
APPENDICES: A. B. C. D. MAIN DISPLAY PARAMETERS ............................................................................30 COMBUSTION EFFICIENCY CALCULATION............................................... 32-33 CALCULATION OF FUEL DATA...........................................................................34 ELECTROMAGNETIC COMPATABILITY STATEMENT ...................................35
-3-
1. 1.1
Charger socket
Keypad
Flue Temperature Connector (underside) ON/OFF MENU Allows access to all menu functions PUMP Turns pump on and off ENTER Accepts a command ie enters a menu option UP Scrolls up through options ie Fuel DOWN Scrolls down through options STORE Enters data storage menu PRINT Prints current data
-4-
Options Label
SENSORS FITTED
Charger Socket
Exhaust Port
Serial Number
*NOTE! Do not cover exhaust port as this will severely affect analyser operation
-5-
-6-
Particle Filter
Neoprene Hose
-7-
2.
SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from the side of the instrument. This instrument must only be used in well ventilated locations. It must only be used by trained and competent persons after due consideration of all the potential hazards.
3.
Charge the battery for 12 hours, following this an overnight charge should be sufficient for an average 8 hour day. See Main Parameter displays for Battery Indicator. The KANE940 has a rechargeable lead acid battery which uses a different charger than other Kane analysers. Ensure the correct charger is used or damage may occur to the instrument. Check that you have all the items you have ordered. Take time to read this manual fully. When using the analyser for the first time you will need to choose from:Language selection Calibration countdown time CO gas alarm NOx percentage for calculation Time and Date Printed header name and telephone number The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.
-8-
4. 4.1
4.2
Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic sensors (if fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %. After switch-on the analyser will briefly display header information :-
-9-
Once the time has reached zero an audible beep will be heard and will show the selected fuel on the following display:NATURAL GAS
PRESS
Press
-MENU- KEY
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN DISPLAY of the analyser:-
Use
and
- 10 -
4.3
Main Displays
The main display can be changed to show either 4 or 8 parameters at one time. Two options are available when 4 parameters are selected. 4 Page Mode displays 4 lines of data in set format, each page is predefined. Line scroll mode allows you to customise the display to show the data you require. 8 Page Mode displays 8 parameters on 4 lines in set format, the bottom two can be changed. Changing between the different modes is detailed in Display Menu Section 5.2.4.
NATURAL GAS DATE . . . 07-08-96 TIME . . . 12:31:35 BATTERY % . . 54 NETT C . . . 0.0 O2 % . . . 20.9 CO ppm . . 0000 EFF (G) % . . . 0.0 CO2 % ... 0.0 FLUE C . . . 0.0 INLT NOT FITTED AMBIENT C . . . 21 CO/CO2 R . . 0.0001 P INDEX % . . . 0.01 XAIR % ... 0.0 Prs mbar 0.00 This screen only displayed on an analyser fitted with an NO sensor NO ppm NOx ppm NOx calc% O2 ref % . . 0000 . . 0000 ... 5 ... 3.0
- 11 -
C ... 0.0 % . . . 20.9 ppm . . 0000 % ... 0.0 % . . . 20.9 ppm . . 0000 % ... 0.0 % ... 0.0
Select next parameter. Repeat above until display reads desired data 4.3.3 8 Page Mode
Displays 8 parameters on the screen at one time. Symbols used in this mode are different to those used in 4 page and line scroll modes and are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
O2 CO PI
CO2 Eff T Tf
:-:-: 0C : 21C
The bottom line of the display can be changed to display other parameters. Use the and keys to change this line.
- 12 -
4.4
Once the automatic calibration procedure has been completed and the specific fuel has been selected (See SELECT menu) the probe can be inserted into the desired sampling point. It is recommended that the sampling point be located at least two flue diameters downstream of any bend and that the probe tip is in the centre of the flue. With balanced flues and other domestic units the probe should be positioned far enough into the flue so that no air can back flush into the probe. This will be indicated by a low oxygen reading and/or a low Poison Index reading.
The probe depth stop cone provided with the instrument allows the probe to be used in holes whose diameters range from 8 mm to 21 mm (5/16 to 13/16 inch). The standard probe is rated at 650C. Temperatures of up to 1100C (2012F) can be accommodated using an optional high temperature probe. TIP To conserve battery power, switch off the pump when you are not taking a measurement. Use the key to turn ON and OFF the pump.
4.5
With the optional pressure module fitted a flue draught measurement can be made at any time. Connect the standard probe to the pressure sensor inlet and the probe in the flue. The pressure reading will be displayed :-
CO/CO2 R . . 0.0001 P INDEX % . . . 0.01 XAIR % ... 0.0 Prs mbar 0.00
To perform a combustion test and display draught pressure at the same time a special probe is required. Contact Kane International or Authorised Distributor for details.
- 13 -
4.6
Care must be taken at all times not to exceed the analysers operating specifications, in particular ensure the following : Do not exceed the maximum temperature of the flue probe. The analyser internal temperature does not exceed normal operating range, typically 040C. DO NOT PLACE THE INSTRUMENT ON A HOT SURFACE. The water trap is vertical at all times. Water condenses in the probe line and can quickly fill the water trap when the probe is moved. Take care and watch the water trap closely. The in-line particle filter is clean and does not become blocked.
4.7
DO THIS EVERY TIME YOU USE THE ANALYSER Remove the probe from the flue - TAKE CARE ! THE PROBE WILL BE HOT - and allow it to cool naturally. Do not immerse the probe in water as this will be drawn into the analyser and damage the pump and sensors. Once the probe is removed from the flue press to switch off. and the analyser will count down from 30
OFF
30
MENU TO ESCAPE
If you have not finished but press operation and not switch OFF. by mistake, you can press to return to normal
- 14 -
4.8
Electromagnetic Compatibility
The European Council Directive 89/336/EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and has an adequate level of immunity to enable it to be operated as intended. The specific standards applicable to this product are detailed in the appendices. Since there are many electrical products in use that pre-date this Directive and may emit electromagnetic radiation in excess of the standards defined in the Directive there may be occasions where it would be appropriate to check the analyser prior to use. The following procedure should be adopted: Go through the normal start up sequence in the location where the equipment is to be used. Switch on all localised electrical equipment that might be capable of causing interference. Check that all readings are as expected. (A level of disturbance in the readings is acceptable). If not adjust the position of the instrument to minimise interference or switch off, if possible, the offending equipment for the duration of the test. N.B. Maximum cable lengths must be less than 3 metres. At the time of writing this manual (November 2005) Kane International Ltd is not aware of any field based situation where such interference has ever occurred and this advice is only given to satisfy the requirements of the Directive.
- 15 -
5. 5.1
MOVING THROUGH THE MENUS Basic Operation NETT C O2 % CO ppm EFF (G) % . . . . 0.0 . . . 20.9 . . . 0000 ... 0.0
MAIN MENU Press to access the MAIN MENU 1. SELECT 2. UNITS 3. DISPLAY 4. SETUP
MAIN MENU Press and to move cursor up and down 1. SELECT 2. UNITS 3. DISPLAY 4. SETUP
MAIN MENU Press to access selected Menu 1. SELECT 2. UNITS FUEL O2 Ref SMOKE RESET FUEL O2 Ref SMOKE RESET FUEL O2 Ref SMOKE RESET 3. DISPLAY 4. SETUP
: LIGHT OIL : OFF : OFF : NO : NATURAL GAS : OFF : OFF : NO : LIGHT OIL : OFF : OFF : NO MAIN MENU
Press to enter value and move to next parameter Press to save settings and return to the MAIN MENU Press
1. SELECT 2. UNITS
3. DISPLAY 4. SETUP
5.2
3. DISPLAY 4. SETUP
The
and
parameters to be changed. TIP Holding down one of these keys scrolls through the data quicker.
5.2.2 Select Menu FUEL O2 Ref SMOKE RESET : NATURAL GAS : OFF : OFF : NO
This menu allows selections to be made for the parameters detailed below. FUEL : Select the fuel being used by the boiler from either a standard fuel stored in the analyser or by entering the user fuel. Once the correct fuel has been selected press to view the fuel constants.
NATURAL GAS K1g : 0.350 K1n : 0.390 K_2 : 11.89 K_3 : 9.83 K_4 : 32 O2r : 3.0
- 17 -
Calculation of fuel constants are detailed in the Appendix. Fuel constants will have to be calculated before a user fuel can be entered. To enter the user fuel select User Fuel and Press
USER FUEL K1g : 0.000 K1n : 0.000 K_2 : 0.00 K_3 : 0.00 K_4 : O2r : 00 00
Use and to select the correct value.
Use to move to the next parameter, repeat above until all parameters are correct. Press to return to SELECT menu.
O2 Ref :
Toxic gas measurements can be referenced to defined oxygen levels. Reference values can be set from 1-20%, to AUTO or more normally to the default value - OFF. Setting to AUTO uses the figure in the FUEL constants data. Oxygen referencing is required by some regulations such as TA-LUFT. If a reference value is selected then toxic gas measurements will be displayed with the symbol (n) attached to the reading. i.e. CO(n) What does Oxygen reference mean ? If 3 % O2 reference is selected and 5 % O2 is measured in the flue then toxic gas values will be recalculated as if 3 % were measured. The equation for referencing is detailed in the Appendix. Oxygen referencing prevents false readings being submitted, e.g. allowing more air into the boiler will increase the oxygen level in the flue and hence dilute any toxic gas reading. Oxygen referencing gives readings as if they were undiluted.
SMOKE: RESET :
Allows the user to enter a smoke test number from 0-9. This value will be printed on the standard printout. Default value is OFF. Allows the user set the Oxygen to 20.9% and zero the toxic sensors without turning the analyser off.
- 18 -
WARNING :
The sensors must only be reset if you are sure they have been sampling fresh air for at least 3 minutes. Errors in measurement will occur if the sensors are reset during or just after sampling.
PRESS. : EFF. :
- 19 -
CONTRAST :The contrast is set to a DEFAULT value or can be adjusted LIGHTER or DARKER. Use the and keys to adjust.
ZERO :
Allows setting of the Autocalibration time in minutes. Care must be taken when changing this parameter as sensors may drift from zero if too short a time is used. Kane International advise 3 minute countdown. Allows an alarm level to be set on for the CO reading. This is set as a default at 1000 ppm.
- 20 -
CO ALARM :
Once an alarm has been exceeded the display will flash every two minutes warning the user of an alarm state and display the gas concentration. A similar display will be shown during a RECHARGE BATTERY and PUMP OFF alarms.
-- -- -- -- -- -- -- -- -- -- -- -- -- -CO ALARM 1010 ppm -- -- -- -- -- -- -- -- -- -- -- -- -- -NO REF: Displayed on the Nitric Oxide unit only. Allows the percentage P in the following calculation to be set. The default value set is 5%. Note the percentage allows for NO2 in a typical boiler. NOx = NO + P% NO CALENDAR: Allows the user to change the date and time. (24 hour clock). The following screen will be shown once the parameter is entered: OFF 1-9 %
hh 13 dd 03
: mm : ss : 53 : 26 : mm : yy : 01 : 96
dd : mm : yy yy : mm : dd mm : dd : yy
To change the time position the cursor on Time and press the left of the 13 as shown below:
hh 13 dd 03
: mm : ss : 53 : 26 : mm : yy : 01 : 96
Once the correct hour is set press to move to the next parameter, the cursor will move to the left of minutes (53). Move to each parameter until the correct time is set. Pressing after setting the seconds will return the cursor to the left of the screen. Format and Date are set in a similar manner.
- 21 -
Header :
Allows two lines of 20 characters to be programmed into the analyser. The header appears on the top of the standard printout. This can be used to print your company name and/or phone number.
- 22 -
6.
PRINTING INFORMATION
Supplied as accessories for the KANE940 are an infra-red thermal printer or a dot matrix serial printer. Read the manual supplied with each printer prior to operation. Connections to the KANE940 are detailed below : Infra-red thermal printer - this does not require a cable to transmit the data but uses an infra-red (IR) link similar to a TV remote control. The IR emitter is positioned on the top of the KANE940 and the bottom of the printer. Ensure they are pointing at each other and within 300 mm, with no obstructions in the way. Data may be lost if transmission is interrupted. Keep the KANE940 pointing at the printer until the printout has finished. Dot matrix serial printer - requires the supplied serial cable to transmit data. Connect the cable to the 8 pin DIN socket on the top of the KANE940 and the 25 pin D-connector on the printer. Data can either be printed from a live test or from stored data. Printing of stored data is detailed in STORING AND RETREIVING DATA.
6.1
During a combustion test the KANE940 will print data on request. With the analyser showing the MAIN DISPLAY press and current data will be sent to the printer. The display will show the following until data transmission is complete.
Printing
6.2 Standard Printout
The standard printout is shown below :-
- 23 -
7.
The KANE940 can store up to 100 combustion tests. Once stored, the data can be viewed on the display or downloaded to a PC or printer.
7.1
While performing a test and viewing the data on the MAIN display access the STORE menu as follows :-
Mode:
Select from the following : STORE - Allows data to be stored in memory. VIEW / PRINT - Stored data can be viewed or printed. DELETE - Clears all data in memory.
Location:
Automatically allocates a location in the memory of the instrument for the next test. On the display shown above the next location will be 3. . The current readings will be stored in
To store a test set MODE to STORE and press the analysers memory. Tip :
Make a note of the location number for your particular test as it may be useful when downloading or printing.
7.2
Multiple tests can be printed easily with the KANE940. Select PRINT under MODE in the STORE menu. This feature is in addition to the VIEW/PRINT, STORE and DELETE options.
The cursor will move to the first number, use the start printing. Press
to move the cursor to the second number, select the last location to print.
- 24 -
NOTE While the display above is shown (i.e. the instrument is printing a test) the keypad is disabled. To exit from printing wait until the current test has finished and the display below is shown:
Press to exit the print routine. The instrument will return to main display
7.3
Deleting Data
to obtain the STORE MENU (as above) :-
Press
- 25 -
8. 8.1
The in-line water trap should be checked and emptied on a regular basis. Water vapour will condense and gather in the probe line. This may move suddenly to the trap when the probe is moved. Care should be taken at all times. Emptying of the water trap is detailed below :-
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a suitable drain, care must be taken as it could be acidic. If condensate spills onto the skin or clothing, clean off immediately using fresh water, seek medical advice if problems occur.
8.2
This is a very important part of the analyser and should be changed regularly. It prevents dust and dirty particles entering the pump and sensors and hence causing damage. The filter MUST be changed when it is discoloured.
Remove the end cap from the in-line filter housing. Carefully remove the paper filter element and dispose of it. Clean the inside of the filter housing with a suitable soft cloth. Insert a new filter element onto the spigot in the filter housing and carefully replace the end cap.
- 26 -
9.
PROBLEM SOLVING
The following is a list of problems that may occur on the instrument through its operating life. If the cause of the fault is not easy to identify then we advise you contact Kane International Service Department or an International Distributor for expert advice. Fault symptom Oxygen too high CO2 too low Oxygen Error (FAULT) Toxic sensor Error (FAULT) Causes Air leaking into probe, tubing, water trap, connectors or internal to instrument. Oxygen cell needs replacing. Calibration time set too short and instrument not allowed to stabilise Instrument has been stored in a cold environment and is not at normal working temperature. Oxygen cell or toxic sensors needs replacing. Battery exhausted. AC charger not giving correct output. Fuse blown in charger plug. Particle filter blocked. Probe or tubing blocked. Pump not working or damaged with contaminents. Probe connected to pressure connector. Temperature plug reversed in socket. Faulty connection or break in cable or plug. Battery below alarm level. Ambient temperature above 50C. Battery quickly discharging and is faulty. Fault has occured on the instrument electronics and requires resetting. Contact Kane International or Distibutor.
Analyser not holding charge Analyser not charging Analyser does not respond to flue gas
Flue temperature readings erratic Analyser automatically switches off in operation. Display shows dark lines and no reponse from ON/OFF key.
10.
ANNUAL RE-CALIBRATION
Whilst the sensors have an expected life of more than two years in normal use it is recommended that the analyser is re-calibrated at least annually. This is so that long term drift on the sensors and electronics can be eliminated. Local regulation may require more frequent re-calibration and users should check with appropriate authorities to ensure they comply with relevant guidelines.
- 27 -
11.
PRODUCT SPECIFICATION
Parameter Temp Measurement Flue Temperature with probe Inlet Temperature Pressure Gas Measurement *1 Oxygen Carbon Monoxide
(standard: H compensated)
Resolution
Accuracy
Range
0.1% 1ppm
+0.2% +20ppm <400ppm +5%<5000ppm +10%>5000ppm +5% reading from 0.1% to 10% +5ppm<100ppm +5%>100ppm +2ppm<30ppm +5ppm>30ppm +5ppm<100ppm +10ppm<500ppm +5%>500ppm +5ppm<100ppm +5%>100ppm
0-21% 0-10,000ppm
Carbon Monoxide
(high range)
Nitric Oxide
(standard)
Nitric Oxide
(low range)
Nitrogen Dioxide
Sulphur Dioxide Calculations Carbon Dioxide*2 Losses Efficiency Excess Air Temp (Nett) CO/CO2 ratio Poison Index
1ppm
0-5000ppm
+0.3% reading +1.0% reading +1.0% reading +0.2% +2C +0.3% reading +0.0001 +0.01
Pre-programmed Fuels
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil, Propane, Butane, Anthracite, Coke, Coal, Kinsale Gas
Dimensions Weight Handset Probe Ambient operating range Power supply (battery charger) Battery Life
1kg 220mm x 55mm x120mm L 420mm x Dia 8mm with stainless steel shaft, type K thermocouple and 3m hose 0oC to 45oC/ 10% to 90% RH non condensing Input 110Vac / 220Vac nominal Output: 12Vac off load >6 hours from full charge
*1 *2
- 29 -
APPENDICES
A - Main Display Parameters
The parameters and their meanings are detailed as follows : DATE : TIME : BATTERY : (BAT) NETT : (T) Analyser date. See Set-Up menu section 5.2.5 to change. Analyser time. Use Set-Up menu section 5.2.5 to change. Displays the battery level from 0-100%. The analyser will flash RECHARGE BATTERY at less than 10 % of charge. With the charger connected the display shows AC ON. Nett temperature calculated by deducting the internal AMBIENT temperature from the measured FLUE temperature. Displays in either C (C) or F (F) and will display NOT FITTED (N/F) if flue probe is not connected. If an external INLET probe is used then INLET is deducted from FLUE. O2 : CO : Oxygen reading in percentage %. Carbon Monoxide reading indicated in ppm or mg/m3. If the figures are referenced to oxygen then the display will show CO(n). See SELECT menu 5.2.2 for oxygen reference. The display will read O2 > 20% if referenced values selected and instrument is in clean ambient air. Combustion Efficiency calculation displayed in percentage. Gross G or Net N can be set see SELECT menu 5.2.3. The calculation is determined by fuel type see Appendix B for calculation. The efficiency is displayed during a combustion test, - - is displayed while in fresh air. Carbon Dioxide calculation determined by the type of fuel. This only shows a reading when a combustion test is being carried out. - - is displayed while in fresh air. and
EFF (G) :
CO2 :
FLUE : Temperature measured by flue gas probe in Centigrade or (Tf) Fahrenheit. Will show ambient temperature after fresh air calibration NOT FITTED (N/F) or FAULT (FLT) if probe disconnected. INLET :
Temperature measured by the optional inlet air probe. This probe is (Ti) plugged into the instrument through the RS232 socket. This figure is used to calculate the NET temperature instead of AMBIENT when fitted. Temperature measured by the internal sensor, used in the NET
- 30 -
AMBIENT :
(Ta) CO/CO2 R :
temperature calculation if an INLET probe is not fitted. The CO/CO2 ratio, is the ratio of measured CO divided by calculated CO2. It gives an indication of the following : How good a gas sample the instrument is reading. How clean the boiler is running.
For example : A new or clean domestic boiler will display a ratio of less than 0.004, a unit in need of cleaning 0.004-0.008 and a unit in need of major overhaul will show greater than 0.008. This only shows a reading when a combustion test is being carried out. - - is displayed while in clean ambient air. P INDEX : The CO/CO2 ratio expressed as a percentage %, called the Poison (PI) Index i.e. P INDEX % = 100 x CO/CO2. - - is displayed while in clean ambient air. in
XAIR % : Excess air calculated from the measured oxygen and type of fuel ( ) used. During a combustion test O2 > 20% will be displayed while clean ambient air. Prs : NO:
Flue draught pressure reading. Displayed when pressure sensor fitted. See UNITS menu 5.2.3. for scales. Nitric Oxide reading in ppm or mg/m3. Displayed when Nitric Oxide sensor fitted. Also displayed as NO (n) when referenced to oxygen. The display will read O2 > 20% if referenced values selected and instrument is in clean ambient air. Calculated total Nitric oxides displayed in ppm or mg/m3. Where NOx = NO + P%NO, note P can be set from 0-9%, default = 5%. See SELECT menu 5.2.2. Also displayed as NOx (n) referenced to oxygen. The display will read O2 > 20% if referenced values are selected and instrument is sampling clean ambient air. Sulphur Dioxide reading in ppm or mg/m3. Displayed when Sulphur Dioxide sensor fitted. Also displayed as SO2 (n) referenced to oxygen. The display will read O2 > 20% if referenced values selected and instrument is in clean ambient air. Toxic gas measurements can be referenced to defined oxygen levels. See SELECT menu 5.2.2 for details.
NOx :
SO2 :
O2 ref % : (O2r)
- 31 -
B.
The efficiency calculation is based upon British Standard BS845. This identifies three sources of loss associated with fuel burning: Losses due to flue gasses: Dry Flue gas loss, Moisture and hydrogen Sensible heat of water vapour Unburned gas Combustible in ash Combustible in riddlings Combustible in dust radiation convection conduction other unmeasured losses
Other losses:
Net efficiency calculations assume that the energy contained in the water vapour (formed as a product of combustion and from wet fuel) is recovered and the wet loss term is zero. Gross efficiency calculations assume that the energy contained in the water vapour is not recovered. Since the fuel air mixture is never consistent there is the possibility of unburned/partially unburned fuel passing through the flue. This is represented by the unburned carbon loss. Losses due to combustible matter in ashes, riddlings, dust and grit, radiation, convection and conduction are not included. Efficiency Calculation: Known Data - Fuel: Qgr = Gross Calorific Value (kJ/kg) Qnet = Net Calorific Value (kJ/kg) K1 = Constant based on Gross or Net Calorific Value: K1g = ( 255 x %Carbon in fuel )/Qgr K1n = ( 255 x %Carbon in fuel )/Qnet K2 = % max theoretical CO2 (dry basis) K3 = % Wet Loss H2 = % Hydrogen H2O = % Water Tf = Flue Temperature Ti = Inlet Temperature O2m = % Oxygen in flue gas O2r = Oxygen reference %
Measured Data:
- 32 -
Calculated data:
Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unburned carbon loss % Efficiency = Flue Temperature - Inlet Temperature
Tnet
Dry flue gas loss % = 20.9 x K1 x (Tnet) / K2 x (20.9 - O2m) Wet loss % simplified Wet loss % Where K3 Net Efficiency % = 9 x H2 + H2O / Qgr x [2488 + 2.1Tf - 4.2 Ti] = [(9 x H2 + H2O) / Qgr] x 2425 x [1 + 0.001 Tnet] = K3(1+0.001xTnet) = [(9 x H2 + H2O) / Qgr] x 2425 = 100 - dry flue gas losses = 100 - 20.9 x K1n x (Tnet) / K2 x (20.9 - O2m) Gross Efficiency % = 100 - {dry flue gas losses + wet losses} = 100 {[20.9 x K1g x (Tnet) / K2 x (20.9 - O2m)]+ [K3 x (1 + 0.001 x Tnett)]} Excess Air CO2% Unburned fuel Loss % Where K4 = [(20.9% / (20.9% - 02m%)) 1] x 100% = [(20.9 - O2m) x K2 / 20.9] = K4 x CO / ( CO + CO2 ) = = = = = = 70 for coke 65 for anthracite 63 for Bituminous coal 62 for coal tar fuel 48 for liquid petroleum fuel 32 for natural gas Note: CO scaled in %
The formula for K4 is based on the gross calorific value Qgr. To obtain the loss based on net calorific value multiply by Qgr/Qnet. Since this loss is usually small this conversion has been ignored. This loss is subtracted from the efficiency. Oxygen Reference CO(n) = CO x (20.9 - O2r) (20.9 - O2m)
- 33 -
C.
For any fuel not specified by Kane International the net calorific value, gross calorific value and composition should be obtained from the fuel supplier. The following fuel data has been calculated with reference to the efficiency calculation. Example 1: Chemical composition: C H2 H2O Qnet Qg Max CO2 25% 3% 50% 8.35 MJ/kg 9.3 MJ/kg 20.4%
K1n
= (255 x % carbon in fuel) / Qnet (kJ/Kg) = (255 x 25) / 8350 = 0.763 = (255 x % carbon in fuel) / Qg (kJ/Kg) = (255 x 25) / 9300 = 0.685 = Max % CO2 = 20.40 = Wet Loss = [(9 x %H2 + %H2O) / 9300] x 2425 = [(9 x 3 + 50) / 9300] x 2425 = (77 / 9300) x 2425 = 20.08
K1g
K2 K3
K4
NATURAL GAS K1n : 0.685 K1g : 0.763 K_3 : 20.08 K_2 : 20.4 O2r : 8.0 K_4 : 65
* Assumed values in the absence of supplied data. See previous appendix for other fuels.
- 34 -
D.
This product has been tested for compliance with the following generic standards: EN 61000-6-3 EN 61000-6-1 and is certified to be compliant Specification EC/EMC/KI/KANE940 details the specific test configuration, performance and conditions of use.
Please Note:
Batteries used in this instrument should be disposed of in accordance with current legislation and local guidelines.
- 35 -