Assembly Instructions For Cable Gland: 755 32R4: Connection Solutions
Assembly Instructions For Cable Gland: 755 32R4: Connection Solutions
Assembly Instructions For Cable Gland: 755 32R4: Connection Solutions
International
Assembly Instructions AI 319 Issue L - 11/08 Certification Details
Class 1 Zone 2 AExd IIC & AExe II See Schedule of Limitations : AExe Class 1 Div 2 ABCD, Class II Div 2 FG & Class III Listing No. E84940
4 5
Armour Clamping
6
1. 2. 3. 4. 5. 6. Backnut Middle Nut Armour Clamping Ring Armour Spigot Brass Pot Entry (with captive deluge seal), if fitted
Cable Preparation
Note: Armour cable acceptance sizes are marked on the clamping ring.
II
Braid
4 3
I
3
4 3 2 1
2 1 1
Shroud Option A Strip Cable to suit equipment as shown above and expose the armour 'I' removing all cable fillers. 'I' = 13/16" (20mm) for cable gland sizes Os to A 'I' = 1" (25mm) for cable gland sizes B to C2 'I' = 1 1/4" (32mm) for cable gland sizes D to H 'II' to suit equipment. If required, fit shroud. See Notes re. Drain Wires.
B Push the cable through the armour spigot . Spread armour over the spigot until the end of the armour is up against the shoulder of the spigot. Position the clamping ring .
C Remove the brass pot from the entry . Place the entry over the armour spigot Move the sub-assembly and up to meet the entry . Note: If the equipment has a threaded entry it may be advisable to screw the entry component into the equipment to prevent twisting of the cable after step D
Connection Solutions
Hawke International is a division of Hubbell Ltd. Registered No. 669157 in England. Registered Office: Mitre House, 160 Aldersgate Street, London EC1A 4DD. A member of the Hubbell Group of Companies
Hawke International UK Office Oxford Street West, Ashton-Under-Lyne, Lancashire OL7 0NA. UK
www.ehawke.com
AI 319 - Issue L / Page 1 of 4 Sales (UK): 0870 606 0105 Sales: +44 (0) 161 830 6698 Technical (UK): 0870 458 9358 Technical: +44 (0) 161 830 6697 Fax: +44 (0) 161 830 6648 E-mail: [email protected]
6
Compound
4 3 2 2 2 1 4
D Unless already screwed into the equipment hold the entry in position with a spanner/wrench to prevent rotation. Hand tighten the middle nut onto the entry and turn a further to of a turn with a spanner/wrench.
E Unscrew the middle nut and visually inspect that the armour has been successfully clamped between the armour spigot and the clamping ring . If braid is not clamped repeat assembly.
F Remove the entry spread out the cable cores and the individual strands of uninsulated conductors for the compound packing. Pack the compound between the cores and strands shown. See notes below and Fig. 7 for compound preparation.
EPOXY COMPOUND PREPARATION When handling this material, the gloves supplied must be worn. The epoxy compound is supplied in the form of a two part package. These should be mixed into the ratio of 1:1 until both colours have blended into one, without any streaks. Rolling and folding is the most satisfactory method of obtaining an even blend. Once mixed, the compound must be used within 30 minutes. After this time it will begin to stiffen. The compound should be kept at an ambient temperature of no less than 20C prior to using. At lower temperatures it becomes difficult to mix. Should any compound come into contact with the skin it should be cleaned off with skin cleaner and not allowed to dry on the skin. Only compound for immediate terminations should be mixed. The mixing and installation of the compound at an ambient temperature below 5C is not recommended due to extended curing periods.
Tape
6 5
Compound Remove surplus compound
35mm
5
Remove surplus compound
G With all gaps and voids filled, bring the conductors back together and pack more compound around the outside of the conductors. Tape the conductors together to prevent disturbance of the compound seal. Pass the brass pot over the armour spigot and remove any surplus compound from the top of brass pot and the joint face as indicated.
H Replace the entry over the brass pot ensuring that compound does not cover the end of the brass pot .
I Locate and hand tighten the sub-assembly and onto the entry . Unscrew the sub-assembly and from the entry then remove any surplus compound from the gland components. Again locate and hand tighten the sub-assembly and onto the entry .
IMPORTANT: The conductors must not be moved for a minimum of four hours.
2 2 1 1
J Allow the compound to cure. (See Fig. 7 for Curing Times). Untighten the sub-assembly and from the entry . Check that the compound has cured.
K Hand tighten the sub-assembly and onto the entry and add half to three quarters of a turn with a spanner/wrench. Tighten the backnut to form a seal around the cable, then tighten a further full turn using a wrench/spanner. Ensure that the middle nut does not rotate when tightening the backnut. Ensure that the deluge seal is pulled down into position, if fitted. Locate the shroud over the cable gland, if applicable.
The following instructions are the approved methods of passing drain wires etc. through the compound barrier and should be followed if permitted by cable installation specifications.
Armour
Fig. 1
Fig. 2
Armour
Outer Sheath
Fig. 3
10mm
Fig. 4
Fig. 5
Sleeving
Shore Hardness