3 Equipment Safety Features Diathermy
3 Equipment Safety Features Diathermy
3 Equipment Safety Features Diathermy
Classification of devices
! ! ! Classification by class: Class I, II, III Classification by type: Type B,BF,CF Classification by suitability for use in flammable atmosphere: AP and APG Protection against ingress of water: IPX classification
! Class II
! Double insulation
! Class III
! Safety Electrical Low Voltage (S.E.L.V) no > 25V for A.C and 60V for D.C
Classification by type
! All diagnostic,therapeutic,life supporting or patient handling equipment must comply with type B, BF or CF specifications. ! Classified on the basis of permissible leakage currents.
Leakage current
L N E
The current that does not return back via the neutral is the leakage current
Type B
Type BF
Type CF
AP
AP G
Class II
Anaesthetic Proof
Anaesthetic Proof G
Type - B
! May have applied parts
! Maximum permissible leakage currents !below the thresholds for skin sensation (500A),! unlikely to cause any pain or injury ! e.g; thermometers, manometers, gas analysers and ventilator alarms
Type - BF
Type-F isolated circuits, suitable for applications where surface electrodes are used, diminishing the safety factor provided by normal skin resistance Maximum permissible leakage currents (500A), similar to those of type-B e.g Forced Air warmers, Blood warmers, Nerve stimulators etc
Type - CF
Most stringent protection against shock hazard Intended for applications where an intra-cardiac connections is likely Maximum permissible patient leakage current set at 1/10 (<50 A) of that for type B or BF devices e.g ECG, EMG, EEG , pacemakers and devices likely to be connected to cardiac catheters (such as syringe drivers and pressure transducers)
Zones of Risk
AP
5-25 cms
AP G
0-5 cms
AP
Standards met by prevention of static sparks, restriction of temperatures (<200oC) and energy levels, and by fireproof enclosures
AP G
IPX Classification
Summary
! ! ! ! Macro and micro shock Safety features in prevention of shock How the equipment are made safe. Safety classification of equipment
Surgical Diathermy
Electrical Hazards
Diathermy
! High frq current:
! 300KHz 3 MHz
Diathermy
! Cutting
! High Frq oscillation ! Sine wave pattern
! Coagulation
! Bursts of Lower Frq ! Damped , pulsed sine wave pattern
Diathermy waveforms
Time
Cutting
Time
coagulation
Time
Blended
Power settings
Power = Current 2 x Resistance ( R ) P= I2 X R
Medium Power
Cutting 30 to 100 Watts Coagulation 30 to 70 Watts ! ! ! ! ! ! Gen Surgery Head and neck (ENT) Laparotomy Orthopaedics Surgery (Major) Thoracic Surgery Vascular surgery
High Power
Cutting > 100 Watts Coagulation > 70 Watts ! Ablative cancer surgery, mastectomies cut 180 300 watts, coag 70 120 ! Thoracotomy ( heavy fulguration 70 120 watts) ! Transurethral resections ( cut 100 170 watts, coag 70 120 watts)
Active electrode
Current density
High Density
Low Density
Active Electrode
Diathermy machine
Bipolar
Active Electrode
Diathermy machine
Monopolar
Electrode
! ! Current flow: between Active electrode & indifferent electrode ( Plate ) Plate electrode connected in the machine in various ways.
1.! Directly to the earth 2.! Isolation Capacitor 3.! Isolated Circuit
Earthed Plate
Active Electrode
Diathermy machine
Diathermy machine
120/80 mmHg
Diathermy machine
Isolating Capacitor
0.01 F ! Reactance At diathermy frq!20! At 50 to 60 Hz !300,000!
Isolation capacitor
Diathermy machine
Diathermy machine
Diathermy machine
Isolated Diathermy
Floating Diathermy
Active Electrode
Isolated diathermy
Active Electrode
Isolated diathermy
Active Electrode
Isolated diathermy
Plate Monitor
Pacemaker interference
Depends on
! The type of pacing electrodes (i.e., unipolar or bipolar) in the patient ! How well the pacemaker circuitry is shielded and ! The strength and the proximity of the discharge from the electrosurgical unit.
EFFECTS OF CAPACITANCE DUE TO HIGH FREQUENCY CURRENT ( DIATHERMY UNIT ITSELF, RF CURRENT GENERATOR)
Direct effects
PACEMAKER UNIT
DIATHERMY UNIT
Capacitive effects
Diathermy noise ( RF )
! Inhibit or turn off any pacemaker " Complete Heart Block and no pacing ! Disrupt the pacing program in a programmable pacemaker " Severe tachycardia