Octopus 3000 Manual V 22
Octopus 3000 Manual V 22
Octopus 3000 Manual V 22
Aquadyne
Version 2.2
First Edition
JULY 1998
Copyright
AQUADYNE Computer Corp., 19941998. All rights reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system or translated into any language in any form by any means without the written permission of AQUADYNE Computer Corp.
Credits Designed and edited by Kirkpatrick Price & Associates Technical Review by Matt Kim, Patrick Johnston, Rick Kriss, Christophe Dugas, Rod Warlick, and Dean McDaniel.
Part Number: 75-050705 AQUADYNE Computer Corporation 7343 "P" Ronson Road San Diego, California 92111 USA
Sales: (619) 569-2082 Tech. Support: (619) 569-2082 x250 Fax: (619) 495-3119 E-Mail: [email protected] [email protected] Web: http://www.aquadyne.com Printed in USA
Table of Contents
1. INTRODUCTION
How the Octopus Works ............................................................... 1-3 User Guide Features ................................................................... 1-4 Quick Reference Menu Map ......................................................... 1-4 Monitoring Capability .................................................................. 1-4 Flash Memory Architecture ......................................................... 1-5 Status Lights .............................................................................. 1-5 Infrared Printer Port .................................................................... 1-6 Powering the Octopus .................................................................. 1-6 Control Capability ....................................................................... 1-7
Table of Contents
Initial Software Setup .................................................................. 3-4 Step 9. Set the System Clock ....................................................... 3-4 Step 10. Enter the Protective Password ........................................ 3-6 Step 11. Set the Noise Filter Control ............................................ 3-6 Step 12. Select the Conductivity Temp Compensation Settings ..... 3-7 Step 13. Select a Conductivity Range ........................................... 3-7 Step 14. Select a Temperature Scale. ........................................... 3-8 Step 15. Enter Setpoint and Range Values ................................... 3-8 Step 16. Entering Alarm Range Values ....................................... 3-12 Step 17. Set up Digital Input Alarm ........................................... 3-15 Step 18. Set up the Lighting System .......................................... 3-15 Step 19. Set up the WaveMaker System ..................................... 3-16 Step 20. Set Up System Hold Control ......................................... 3-16 Step 21. Assign System Control Addresses ................................. 3-17 Step 22. Set Up the Communication Software ............................ 3-18 Step 23. Enable the Octopus System Control ............................. 3-21 Step 24. Test the System Configuration ..................................... 3-22
ii
Table of Contents
5. SETTING UP AQUAWEB FOR WINDOWS 95
AquaWeb - Introduction ............................................................... 5-2 Step1-Configuring Windows Com port setting to support AquaWeb 5-4 Step 2 - Installing AquaWeb on your PC ....................................... 5-5 Step 3 - Starting and configuring AquaWeb .................................. 5-7
7. TROUBLESHOOTING
Probe Troubleshooting ................................................................. 7-2 Peripheral Equipment Response Troubleshooting ......................... 7-5 Modem/pager Troubleshooting .................................................... 7-6 AquaWeb Troubleshooting ........................................................... 7-7
APPENDICES
Appendix A: Menu Maps Appendix B: Understanding Water Quality Appendix C: Conversion Tables Appendix D: Product Warranty
This manual is available in an Adobe pdf format from our web page at http://www.aquadyne.com. For the latest information, electronic updates to this documentation and other technical information, we recommend that you check with us regularly. Aquadyne Computer Corp. 7343 P Ronson Road San Diego, California 92111 USA
iii
Chapter
1
Introduction
Introduction
1-2
CHAPTER 1: INTRODUCTION
Introduction
HOW THE OCTOPUS 3000 WORKS
Congratulations on your purchase of the most sophisticated, yet easy-to-use aquatic environment controller on the market today. Designed with a powerful INTEL microprocessor, your Octopus 3000 controller includes the following features: Continuous monitoring of conductivity, water temperature, pH and oxidation reduction potential (ORP). H2O/S software with menu-driven programming environment. Aux. Input supports power failure, water level or pressure alarming. FLASH memory that makes upgrading your Octopus 3000 with new H2O/S software a simple task. Local audible alarm or optional remote dial-up pager alarm capability. Parameter control capability for management of peripheral equipment such as pumps, and lighting. Powerful datalogging capability that holds up to 7-days of probe data. Operates on 9V AC or on an optional 9V DC battery backup. optional X-10 support for up to 255 control outputs. An infrared Data Port that supports an optional hand-held HP printer for data reporting and graphical analysis. Single communication port supporting both direct connection to a PC or remote modem communication. AquaWeb Remote AccessOur latest software product that allows you to communicate with your Octopus 3000, perform advanced graphical analysis of current and historical datafrom any Windows 95 PC. The standard version of AquaWeb is available for no cost through our WEB site at: HTTP://www.aquadyne.com Single User and Multi-User versions of AquaWeb are available for additional cost through your dealer. Until now, performing even the simplest functions with a controller required technical or electronic expertise. The Octopus 3000 controller makes maintaining the optimum aquatic environment easy. The Octopus 3000 is constructed with only the highest quality components. The heart of the Octopus 3000 is an INTEL microprocessor which manages all of the activities of the controller. All information gathered by the controller is processed digitally. The Octopus 3000 Architecture includes Digital Signal
Note
The helpful Notes logo highlights items which will save you time and effort, and increase your understanding of the Octopus operation. The CAUTION logo highlights items which are critical to safety or which may cause the Octopus to malfunction or become damaged.
!
CAUTION
CHAPTER 1: INTRODUCTION
1-3
Introduction
Processing (DSP) hardware and software that provides accurate and repeatable readings over the life of the Octopus 3000. In addition, the microprocessor design provides unsurpassed flexibility, as reflected by the variety of software options that are available for the Octopus 3000. mation before proceeding with Chapter 2, Installation and Software Programming. This User Guide includes several quick reference features, including Notes, Cautions and a Quick Reference Menu Map.
AQUADYNE is committed to building products that meet world class standards of quality and workmanship. If you have any questions regarding the use of this product, AQUADYNE provides hotline service through the Technical Support Group. They can be reached between 9:00 AM and 5:00 PM Pacific Standard Time at (619) 569-2082 x 250 or via the Internet at: [email protected] The primary source for technical and other useful information about the Octopus 3000, including the latest release of this User Guide in Adobe PDF format, can be found on our Web Page: http://www.aquadyne.com. Our Web Page contains current software releases and always reflects the latest information about our products, so we suggest that you visit it frequently.
MONITORING CAPABILITY
The Standard Octopus 3000 software provides for monitoring, control, and alarm capabilities. The four water quality parameters monitored are: Conductivity Temperature pH ORP
In addition to the four parameters the Octopus 3000 also provides an aux. input which can be configured to provide a digital input for power failure detection, or water level alarming. Probes connected to the bottom of the Octopus 3000 gather information and relay it to the Octopus 3000 for processing and display on the screen.
1-4
CHAPTER 1: INTRODUCTION
Introduction
The Octopus 3000 is programmed using a membrane keypad on the face of the unit. Directional arrows on the keypad are used to move horizontally across a given menu level. The Enter key is used to move down the menu map or to make a selection. Once in the submenus, if there is no activity for ten minutes in most menus, the Octopus LCD display window will return to the Current Status Display. Settings entered up to that time will be saved. When not in programming mode, current status information gathered by the probes is shown on the Octopus LCD Display Window.
Conductivity
Solid Green. (System is dosing to lower conductivity value.) Flashing Green. Dwell State. (Will read conductivity again after dwell time is complete.) Solid Red. (System is Dosing to raise conductivity value.) Flashing Red. Dwell State. (Will read conductivity again after dwell time is complete.)
Temperature
Solid Green. Chiller On. (Current temperature is too high.) Flashing Green. Chiller Dwell State. (Dwell time to prevent compressor burn out.) Solid Red. Heater On. (Current temperature too low.)
pH
Solid Green. (CO2 is Dosing to lower pH.) Flashing Green. Dwell State. (Will read pH again after dwell time is complete.) Solid Red. (pH buffer is dosing to raise pH.) Flashing Red. Dwell State. (Will read pH again after dwell time is complete.)
CHAPTER 1: INTRODUCTION
1-5
Introduction
light lit red indicates that the system detected a low condition, a green light indicates a high condition. A summary of Parameter Status Lights is given in Figure 1.2.
Flashing Green/Orange. Sending commands through the power line to the Satellite Modules. Flashing Red. Controller hardware failure. Unplug the Octopus controller and call Aquadyne customer service.
The backup battery-pack is an inexpensive means of providing about 45 minutes of backup power during a power failure. While operating from the battery pack, the Octopus
!
CAUTION
UPS power supplies have built in noise and surge suppression and WILL block X-10 control signals. Do not plug the X-10 power line interface module into a power outlet that is powered by the UPS, unless all the X-10 control modules are also powered by the UPS as well.
1-6
CHAPTER 1: INTRODUCTION
Introduction
Your Octopus is a precision laboratory instrument that could be damaged by unexpected power surges or fluctuations. The use of a quality surge protection device is highly recommended, and is required to preserve your warranty. However, DO NOT USE a surge suppressor between an Octopus power line interface and a control module. will continue to operate normally until the battery is exhausted. If a UPS system is used as a backup power source, it may be purchased from Aquadyne. A UPS backup is a good idea if you require more than 45 minutes of backup power, or if you wish to backup other critical peripheral or communications equipment such as a pump, a bio-filter or a modem for example. Note however, that the duration of backup of the UPS is reduced by the addition of such equipment. The proper and reliable functioning of the Octopus and the X-10 control system is dependent on the proper installation of the Octopus, the X-10 Power Line Interface, and each Control Module. Other electrical devices, such as motors, ballasts, and dimmer switches produce electrical interference that may affect X-10 signaling. Refer to Chapter 5, Peripheral Device Troubleshooting, for further details.
!
CAUTION
!
CAUTION
The Octopus 3000 control system has six basic components, as shown in Figure 1.4: The Octopus 3000 Controller Wall Mount Power Supply
Probes The X-10 Power Line Interface One or More Control Modules A Modem
Control Capability
H2O/S software is used to program the Octopus 3000 and give it the capability for parameter control as well as monitoring. With the purchase of the optional X-10 power line interface and one or more control modules, a wide variety of monitoring and control features are automatically activated.
(Note: The modem, the X-10 power line interface, and the control modules are options.)
CHAPTER 1: INTRODUCTION
1-7
Introduction
interface turning the devices on or off as needed. Each device that the Octopus controls, such as a heater or chiller, has its own control module. As many as 256 addresses and a variety of devices can be controlled by the Octopus, using X-10 control modules The Octopus 3000 incorporates several mechanisms to protect your aquatic environment from controller failure. For example, the controller is engineered to overcome problems associated with power fluctuations through the periodic re-sending of control module commands to make sure that a X-10 control is latched in the correct state. In addition to controlling peripheral equipment to maintain water quality, the H O/S software supports other optional features that can be enabled on the control menus:
2
Octopus 3000
Modem
X-10
Power Supply
Surge Suppressor
Probe Cables Octopus Power Transformer Plugged Into Surge Suppressor Octopus Control Module Device (E.g. Heater)
Introduction
Example Lighting Scheme
Timer 1 is for the fluorescent lamps. These lamps are the first lights to come on in the morning and the last to go off late at night. Timer 2 is for the metal halide lamps. To approximate the sun rising, these lamps come on in the late morning and go off in the late afternoon. Timer 3 is for an additional halide lamp. This lamp will come on for about one to two hours at noon to simulate the midday sun. Timer 4 is for a red bulb during the night to allow observation of nocturnal creatures that come out when the lights are out, and or moonlight simulation. If you have the 16-channel Lighting System and are not using all 16 channels, you can use the additional channels to control other devices. 24-hour period. The Octopus 3000 comes standard with 4-channel lighting software. A optional 16 channel system is available as an option. With separate lighting timers, there is a high degree of flexibility in setting up your lighting system. For example, in your application you may wish to use a combination of metal halide, fluorescent, and incandescent lighting. And each lighting system may require a dedicated ON and OFF time. The Octopus 3000 lighting system can manage these and other varied timing needs under software control. (Figure 1.5).
System Hold
The H2O/S system hold software is a feature designed to aid in the feeding of fish and invertebrates. When the Octopus controller is put into feed mode, all systems, except the Lighting and WaveMaker systems, are shut down for a period of time selected by the user. At the end of the specified time (up to 60 minutes) all systems are turned back on and will restart within 5 minutes. If desired, the system hold option may be turned off by setting the feed time to 0 minutes. The Octopus controller will then no longer prompt you to enter the system hold mode.
WaveMaker System
The H2O/S Dual Channel WaveMaker software feature allows the Octopus 3000 to control two independent wave or surge generation systems. Each system supports two pumps which are driven by Control Modules. The cycle time for each pump is selected by the user, from 1-999 minutes (over 16 hours).
CHAPTER 1: INTRODUCTION
1-9
Introduction
Alarm System
The H20/S alarm system software provides visual and audible notification in the event that a monitored parameter exceeds the user defined range. The optional pager alarm software upgrade used in conjunction with a modem provides remote alarm notification via a digital pager.
1-10
CHAPTER 1: INTRODUCTION
Chapter
2
Installing the Components
!
CAUTION
2-2
!
CAUTION
In addition to a surge suppressor, a Ground Fault Circuit Interrupter (GFCI) should be part of you installation. A GFCI is a sensitive device which immediately cuts off the electricity flow when an electric current leak is detected. All devices that are part of the Octopus 3000 system should be connected to a GFCI protected circuit. If you do not have one installed already, you should purchase one. They are inexpensive and are available at any electronics or hardware store. Dont wait, a GFCI can save your life!
2-3
!
CAUTION
Not all systems require the same installation steps. The steps required for each individual system will depend on: Whether the system is monitoring only or monitoring and control. Which optional components are included in the system.
Table 2.1 indicates which components installation procedures your system requires. For example, for Monitoring Only systems, only installation steps 1, 2, and 7 would be required.
Monitoring System
N/A N/A N/A Optional Optional
!
CAUTION
To prevent damage to the ORP probe sensor, FIRST REMOVE THE TAPE OVER THE VENT SEAL on the protective cap. Then remove the protective cap. Remove the protective caps from the remaining probes and install per Step 2 instructions.
OTHER CONSIDERATIONS
Note
The Octopus is a precision instrument that has been designed to interface with highquality laboratory grade probes manufactured in the USA to Aquadyne specifications. The use of other probes (especially inexpensive imported probes) not approved by Aquadyne is not recommended and will not be supported. A genuine Aquadyne probe can be identified easily by the presence of date code stamped on the top of the probe shaft.
pH and ORP probes can require a minimum of 48 hours to stabilize after installation. DO NOT ENABLE OCTOPUS CONTROL OF THESE PARAMETERS FOR 48 HOURS. Be sure probes are plugged into the correct Octopus probe ports. If they are plugged into the wrong ports, the system will not monitor properly. Place temperature, pH, & ORP probes within 6 inches of each other. The conductivity probe should be located at least 18 away from the temperature probe.
2-5
2. Remove the vent seal on the ORP probe protective cap, then remove all protective caps. 3. Plug the probe connector cables into the designated port at the bottom of the Octopus 3000 controller.
Temp.
pH
ORP
Conductivity Probe
To Pump
2-6
Octopus 3000
pH
OR
p em
Probe
Compression Fitting
Shutoff Valve
Shutoff Valve
Fig.2.4 In-Line Location Using Tee Fitting Figure 2.3 In-Filter Location Using Float
CHAPTER 2: INSTALLING THE COMPONENTS 2-7
In-Line Location
The in-line installation shown in Figure 2.6 is the preferred installation method because the tee-fitting is a natural conductivity cell and it provides the most accurate measurement environment. The general procedure is the same as that described above for temperature, pH and ORP probes. However, the conductivity cell, is utilized. Be sure that the installation provides that about 10% of the probe protrudes into the cell. Aquadyne provides a score-mark on all of its conductivity probes that guides installation and assures proper insertion depth.
Probe
Note
If the conductivity probe is not installed in a conductivity cell or if the probe is exposed to electrical interference from pumps, motors, or electrical ballasts, unreliable readings will result.
Conductivity Cell
2-8
The water flows in and out of the cup at a rate of about 1 cup per minute.
Installation "Vase d'Ecoulement"
Note
Verse l'Octopus
Pompe
Sonde de Conductivit
The Octopus has been calibrated at the factory to pre-set standards. Since each probe has its own unique measuring characteristics, it is always recommended that you perform a manual calibration allowing sufficient time for the probes to stabilize in the calibration solution.
Vase
Filtre/Bac
In a drip cup installation, there is no compression fitting to control insertion depth you must carefully position the probe so that the end of the probe is approximately 1 inch from the bottom of the cup.
2-9
Octopus 3000
Modem
Figure 2.7 Float Switch with 8-pin Connector Install the Float Switch in Tank or Filter
The float switch should be installed in the tank or filter at the desired level, according to the manufacturers instructions. (Many float switches will require drilling into the tank at the desired float location.) Be sure the float is installed in the position shown in Figure 2.7 (with the arrow on the side of the switch pointing up).
Probe Cables
X-10
Power Supply
Surge Suppressor
2-10
GFCI
15
15
AQUADYNE
TM
3 5
3 5
13 11 9 UNIT A 7
13
E
O M K
I ID
Heater
AQUADYNE
TM
11
15 13 11 9 UNIT A 7 1 3 5
SATELLITE
TM
Control Module
9 UNIT A
Adresse: A-3
C E
O M K
C E
O M K
I ID
TM
Control Module
I ID
Figure 2.10 Addressing a Control Module Figure 2.9 Control Module Installation
CHAPTER 2: INSTALLING THE COMPONENTS 2-11
Note
The X-10 Power Line Interface and all Control modules must be installed on circuits of the SAME PHASE. If the X-10 Power Line Interface and the Control modules are not on the same phase, they will not be able to communicate with each other, except by means of a X-10 Signal bridge. Do not plug Control Modules or the X-10 Power Line Interface into power strips equipped with surge suppression. Surge suppression will filter out the Octopus X-10 control command signals.
!
CAUTION
A sticker on each satellite module indicates the maximum load-rating of the device. DO NOT OVERLOAD. Equipment such as compressors and pump motors can exceed the load rating of a Control module. Verify with the manufacturer of your peripheral device that the surge current at startup does not exceed the load rating of the satellite module. Damage to the satellite unit could result! In addition, be sure you are using the correct plug configuration.
Timer 16
The alarm paging feature allows your pager to receive an alarm page when a water quality parameter goes outside the acceptable range or when a digital input switch signals an alarm condition. The pager message indicates which Octopus 3000 unit (via the last 3 digits of the unit serial number) is affected (in the event of multiple units), which parameter is involved, and whether the alarm represents a high or low condition. (Refer to Chapter 3, Step 23 for a discussion of pager message information.) As shown in Table 2.3, in order to have the alarm paging feature, you must install a modem to your system.
For a single PC to a single Octopus, a direct cable connection can be used (refer to Figure 2.11). Note that if a direct cable connection is used, alarm paging will not be available. (Both the direct cable connection and the modem connection require a serial port connection. And PCs typically have only one serial port available.) For remote access to the Octopus 3000 datalog or for networking two or more Octopus units, a modem connection is required.
Communication Feature
Alarm Paging Only AquaNet Only Both Alarm Paging & AquaNet
N/A
N/A
2-13
1.
Connect the Modem and the Octopus UnitFirst, plug your modem cable into the modem/serial port of the Octopus 3000 unit. Then connect the other end of the modem cable into the 25-pin connector on the back of the modem, as illustrated in Figure 2.12. The connectors are keyed and can only be installed one way. Make sure that the plugs on the cable are fully inserted into the jacks on the Octopus controller and the modem.
2-14
Telephone Line
Modem
2-15
There could be problems in modem operation and pager functionality if the telephone line is shared. For proper operation, the modem will required a dedicated telephone line.
!
CAUTION
Use of any cable other than that supplied by Aquadyne may result in an incomplete connection or, in extreme cases, damage to the Octopus or computer. The cable utilized for Direct Cable connection is a nonstandard cable. It can be ordered from your dealer or from Aquadyne.
To hook up directly to the PC, one end of the direct cable connects to the serial port on the Octopus 3000 unit and the other end connects to your PC com port. Aquadyne provides a optional 25 and 50-foot extension cables for longer runs.
pH Probe (BNC)
Diagnostics de dmarrage
OCTOPUS 3000 Aquadyne Inc. Ver. 2.0 Ser #97-1234567 Copyright (C) 1992-1997 Auto Test
Version du Logiciel et numro de serie
Auto Test ok Initialisation sorties num. Initialisation Termine Initialisation Modem: +++ Initialisation Modem: Envoie
Initialisation du Modem
08/15/97 11:45AM
pH = 7.8
Rdox= 350mV T= 78.2F pH= 8.20 Tous Systmes OK <--Affichage du changement de statu en rotation sur la ligne inferieur
Systme Redox ACTIVE
CD 45.4mS
ALARM pH mini
Probes (Temperature, ORP, pH and Conductivity). Digital Input Switch or Device (optional). X-10 Power Line Interface (Optional). Control Module(s) (Optional). Communications Equipment (Optional).
You then powered up the Octopus and installed the Backup Battery Pack.
Cables de Sondes Transformateur Branch dans Suppresseur de Tension
You are now ready for the second half of the installation process: programming the Octopus H2O/S software (Steps 9-25), presented in Chapter 3. Dont worry, you dont have to be a computer programmer to successfully program the Octopus. We not only give you the roadmap (the Menu Map), we guide you down the road as well.
9V Batterie Pack
!
CAUTION
Please note that the 9-Volt Backup Battery incorporates special hardware designed by Aquadyne specifically for the Octopus. The use of any other voltage source could cause damage to your Octopus.
2-18
Chapter
3
Programming the Software
pH = 7.8
CD 45.4mS
These menus are the first level down from the Top Level Display and comprise the Main Menu for the Octopus 3000. This menu is the doorway to all other sub-menus. To enter the Main Menu, simply push the Enter key from any position on the Top Level Display. Note that from the Top Level Display you will always enter the Main Menu at the same place: the Setpoints Menu. From there, you can scroll left or right in the Main Menu as desired, using the right or left arrows on the keypad.
Setpoint Menu
Operation Menu
Setup Menu
3-3
Setpoint Menu
Operation Menu
Setup Menu
Cond. Setpoint
Exit Setpoints
Temp Setpoint
ORP Setpoint
pH Setpoint
pH Setpoint Setpoint=x.xx
3-4
3-5
Note
Entering AQDYN at the new password prompt will display the system serial number and the version of H2O/S software installed in the Octopus. This information will be required should you need to contact technical support for any reason.
3-6
3-7
Conductivity Setpoints
To enter the conductivity optimal setpoint value, press Enter from any place on the Top Level Display. You will be at the Setpoints Menu. Press Enter and you will be at the Temperature Setpoints Menu. Using the left arrow key, scroll to the conductivity menu and press Enter. At the flashing
When these three values are entered, the system will operate as illustrated in Figure 3.5. As long as the water quality parameters remain within the acceptable operating range, no control or alarm will be activated. If, however, the parameter rises above or falls
3-8
Value Required
Conductivity Conductivity Setpoint Conductivity Range Low Conductivity Dose Time Low Conductivity Dwell Time High Conductivity Dose Time High Conductivity Dwell Time Temperature Temperature Setpoint Temperatrue Range pH pH Setpoint pH Range Dose Time Dwell Time CO2 Doser Time CO2 Dwell Time OR P ORP Setpoint
need to enter other values in the Setpoints Menu including dosing times for low and high conductivity, and dwell times. A dwell time is the length of time that the Octopus 3000 will wait until dosing can occur again. This feature prevents overdosing a system with a dosing solution and protects against short cycling of equipment.
X X
X X X X X X
Conductivity Example
X X X X
X X
X X X X X X
Suppose that you desire an optimal setpoint value of 45 mS, an operating range of 5 mS and an alarm range of 3mS. Figure 3.5 illustrates the relationship between the optimal setpoint value, and the control range. For these systems, if the conductivity value remains within the target operating range, the parameter status light will not be illuminated. If the conductivity value enters the control range, the parameter status light will turn either red if the low conductivity dosing system is enabled, or green if the high conductivity dosing system is active. If the parameter status light is flashing red, the low conductivity dosing system is in a dwell state. A high conductivity dosing system dwell state is indicated by a flashing green parameter status light (See Figure 1.2).
Low Alarm ON
Low Control ON
High Control ON
High Alarm ON
Temperature Example
Suppose you have a setpoint value is 78.2 F, your target operating range is 0.1 F, and the alarm range is 0.1 F. (Figure 3.5 illustrates the relationship between the setpoint value, and the control and alarm ranges.) If the temperature stays within the target operating range (78.178.3 F), the parameter status light will remain off. If, however, the temperature enters the control range (between 78.1F on the low side and 78.3F on the high side), the parameter status light will either turn green if the chilling system is active, or red if the heating system is active. At the beginning of the control range, the Octopus will turn on the chiller or heater, as needed. These peripheral devices will remain activated to drive the temperature value back to the optimal setpoint, and once achieved, the peripheral device will be turned off.
Note
Note that the chiller system has a protective feature that prevents the chiller from turning on/off more often than every 10 minutes.
When you press Enter again the current temperature setpoint will begin flashing on the LCD display window. Pushing the right arrow will increase the setpoint value and pushing the left arrow will decrease the setpoint value. When the desired value is displayed, push the Enter key to select it. If the system is configured for monitoring and control, you will need to enter values for the temperature range in addition to the temperature setpoint. This determines the control range for heating and chilling devices. Since most heaters and chillers have their own thermostat, these thermostats can be used as a backup temperature control system. The heater thermostat should be set several degrees above your desired temperature and the chiller several degrees below. If your
Note
Configuration data such as setpoint values, are stored in permanent (nonvolatile) memory. Should primary or backup power to the Octopus be lost, the setpoint values will be restored to their previous values when power is restored to the Octopus.
3-10
Note
Note that if the dose time is set to 0, the dwell time entered will be ignored by the Octopus and the unit will continue to dose buffer until the pH returns to the Optimal Setpoint Value.
The last pH control selection is the CO2 dwell time. This is the time that the Octopus waits before allowing another CO2 gas injection to occur. This time is selectable from 1-255 minutes. If the Octopus 3000 senses that the pH is outside of the target operating range, it will
3-11
ORP Setpoints
The ORP setpoint defines the acceptable level of oxidation reduction potential in the aquatic environment. The ORP level is an indicator of ozone. Acceptable levels are between 200 and 450 mV. The ORP setpoint can be maintained through the use of an ozonizer in Monitoring and Control System. However, unlike the proceeding parameters, the ORP setpoint has a FIXED range of 5 mV.
Note
3-12 CHAPTER 3: PROGRAMMING THE SOFTWARE
The high or low alarm trigger points are computed by adding the control value and the alarm value to the setpoint value.
Table 3.5
An alarm will be triggered when the value measured falls outside of the control range. For example, suppose you have a optimal temperature setpoint value of 78.0 F and an acceptable operating range of 0.1. Figure 3.6 illustrates the relationship between the setpoint value and the acceptable operating range. When the parameter value stays within the acceptable operating range (77.978.1 F), the system monitoring function will not detect an alarm and the parameter status light will be off. If, however, the temperature drifts past the acceptable operating range (below 77.9 F or above 78.1 F), an audible alarm will be triggered, the monitored parameter will begin
Temperature Alarm
The temperature alarm range can be set from 0.10 degrees to 2.00 degrees. The temperature alarm range can only be set if the temperature alarm is enabled.
3-13
ORP Alarm
The ORP alarm range can be set from 5 mV to 100 mV. The ORP alarm range can only be set if the ORP alarm is enabled. If the Hold-off Timer value is greater than zero, the audible alarm will not sound until the alarm Hold-off Timer has expired. To reset the alarm you must push the Enter key once on the face of the controller. This action will silence the alarm and the alarm will not re-trigger until the measured value again falls outside the alarm range. The alarm range for ORP operates about the ORP Setpoint, and NOT the limits of the optimal operating range since the operating range is fixed at 5 mV. For example setting the ORP alarm range to 10 mV with a ORP setpoint of 200 mV, an alarm will occur at 190 mV and 210 mV. Set the Alarm Hold-off Timer The Hold-off Timer provides a delay before an audible alarm is sounded and before the remote Alarm Control Module (if the system includes it), and an alarm page is sent (if installed). This is useful to avoid bounce by allowing sufficient time for the trend to fully establish itself. The timer can be set from 0 minutes to 99 minutes. If you select 0 minutes then no delay will occur before a audible
pH Alarm
The pH alarm range is entered in the Setup Menu. Note that the pH alarm range can only be set if the pH alarm is enabled and it is expressed as a value around the pH setpoint. The pH alarm range value can be set from 0.05 to 1.0 around the setpoint control range and is adjustable in increments of 0.05 pH units. To reset the alarm you must push the Enter key on the front of the Octopus 3000 once. This action will silence the alarm. When the parameter returns to the acceptable operating range, the alarm will not re-trigger until the parameter again goes outside the acceptable operating range. There is a target control range in addition to the alarm range. For example, suppose your
3-14
No Alarm
Alarm is disabled.
3-15
Note
The reliability of the digital input circuit is dependent on the use of low impedance switches and wiring. For customers wiring multiple switches in series, the total resistance of the switches should not exceed 200 ohms.
3-16
CURRENT FLOW
Pump 2A Pump 2B
Hold prompt appears, press Enter to get to the Setpoints Menu. Use the arrows to scroll to the Setup Menu and press Enter. Use the arrow keys to scroll to the timer menu. When you press Enter again, you will be at the System Hold Time Menu. Press Enter and enter the desired hold time.
CURRENT FLOW
Pump 2A Pump 2B
3-17
Note
We recommend setting the alarm Hold-off period to at least 1 minute when using the pager option. This will avoid spurious pages due to small aberrations about the edge of the alarm range.
3-18
1 2 3 0 9
1 2 3 0 9
1 2 3 0 9
1 1 n/a 0 9
* For digital input devices, a 1 indicates an open condition and a 2 indicates a closed condition.
The W dial modifier is particularly important to make sure that there is a dial tone before initiating a page. Pager Phone Number Example
!
CAUTION
After an alarm page has been acknowledged by pressing Enter on the face of the Octopus, if the alarm condition still exists (e.g., temperature or some other parameter is still outside the acceptable operating range.) the alarm page will be triggered again and the Retransmit Alarm function will be activated.
The following is an example of one scenario where the pager number requires a bit more than the nine-digit phone number: To reach the pager, it is necessary to: Dial 9 for an outside line. Wait for the dial tone. Dial the paging company. Wait a couple of seconds for the paging company to answer the phone and accept a pager input.
3-19
!
CAUTION
There could be problems in modem operation or pager functionality if the modem telephone line is shared. Be sure to use a dedicated line for the modem.
3-20
Baud Rate
The pager baud rate is the actual speed in which the Octopus 3000 can transmit data to the modem. The Octopus 3000 baud rate can vary at rates from 300 baud to 19,200 baud. The default value is 19,200 baud, which is ideal and recommended. This value will work with 90% of the modems in use today including the Aquadyne modem that comes with the pager option. If you also have the AquaWeb option you must set the baud rate of the pager to 19,200 to match the fixed baud rate of AquaWeb.
Pager Enable/Disable
The pager enable/disable command allows you to enable or disable the paging system. If the system is disabled, no pages will be sent by the controller should an alarm condition occur.
Data Format
The pager data format register controls how information is sent to the modem. The controller defaults 8 data bits, 1 stop bit, and no parity (81N). Additional information on serial data formats is outside the scope of this manual. If using both a pager and AquaWeb, you must use the same data format for both.
3-21
3-22
Chapter
4
Operation & Maintenance
Note
Probes are delicate instruments that can be damaged by improper handling, installation or poor maintenance. Damaged probes will cause inaccurate parameter measurements and operational problems for systems configured for control. Please review Chapter 5 Probe Problems for troubleshooting tips.
In the rest of this chapter, we will discusses topics that relate to the operation and maintenance of the Octopus 3000. The topics covered include:
2. 1. Daily Check List 2. Cleaning the Probes 3. Calibrating the Probes 4. Replacing a Probe 5. Conducting a System Test 6. Changing a Control Module Addr. 7. AquaNet Reporting and Graphing 8. Operating the Infrared Printer 4. 5. 3.
4-2
!
CAUTION
The use of any sandpaper with a coarser grit rating than 600 (e.g., 300,100, etc.) will damage the probe. When polishing the Tip of the electrode make sure to use a smooth flat surface to prevent scoring the graphic elements in the probe.
Lemon
H2 0
Q-tip
4-3
Complete
Complete
4-4
Note
Make sure to use a clean dry towel to dry the tip of the conductivity probe. Oil of any kind on the tip of the electrode will cause errors in calibration and operation.
From the Operations Menu, use the arrows to scroll to the Probe Calibrate Menu. After you respond to the Are You Sure? prompt, use the arrow keys to scroll to the Conductivity Calibration menu, and press Enter. During the calibration process you will see the following menu prompts shown in Figure 4.2. You are directed to dry off the probe then push Enter and begin the dry calibration. When the probe is calibrating the Display Window will read Cond Probe is Settling.
2.
Note
To insure accurate reading it is critical that the unit be calibrated with a conductivity buffer consistent with the range that the conductivity system will be operated in. Aquadyne recommends calibrating the low range with a 718 uS buffer and the high range with a 58,640 uS buffer. Failure to do this will yield inconsistent readings.
pH Probe
4. 5. 6.
7.
The menus that appear on the Octopus display during calibration are shown in Figure 4.3.
pH Probe is Settling
(Alternating Message)
Note
It is important that the buffer solution in cup #1 is the HIGHER of the two buffers being used. For example, if you are using buffers 7 & 10, cup #1 should contain pH 10 buffer.
pH Probe is Settling
Proceed with the calibration procedure, as follows: 1. Dip the stir stick provided into the Quinhydrone powder and remove a heaping portion on the bottom 1/4-inch of the stick. Mix this powder into one of the buffer solutions (pH-4 or pH-7). Some of the Quinhydrone should remain undissolved. If the Quinhydrone dissolves completely, add more. Use a different stir stick and repeat this procedure for the second buffer solution. At the Octopus controller, press the Enter key from any place on the Top
ORP Calibration Enter to Select
(Alternating Message)
2.
3. 4.
4-7
4. REPLACING A PROBE
Replace your pH and ORP probes every 18 months for Aquadyne laboratory grade probes, 12 months for standard grade probes, or as needed for accuracy and reliability. Replacement probes can be purchased from the store where you purchased your Octopus controller. For a list of dealers nearest you, check Aquadynes Web site at http:// www.aquadyne.com or call Aquadyne Customer Service.
Note
If you leave the controller in the System Test Mode unattended for 10 minutes or more, the controller will automatically revert back its state prior to entering System Test.
To conduct a System Test, first press Enter from any position on the Top Level Display, you will be at the Setpoints Menu. Use the arrow keys to scroll to the Setup Menu then press Enter. You should now be at the System Test submenu. When you press Enter again, you will be at the Heater Test screen. Use the arrow keys to scroll to the system you desire to test. When you are at the desired screen, press Enter. This will default to the System Control OFF screen. Use the arrow key to scroll to the System Control ON screen and press Enter. This will test the desired system.
5. CONDUCTING A SYSTEM TEST (ONLY FOR SYSTEMS WITH AN X-10 POWER LINE INTERFACE AND CONTROL MODULES)
When you conduct a System Test, all configured Control Modules automatically go to an "off" condition. Once in the System Test mode, you can select any Control Module and manually turn it on or off. The purpose of this test is to make sure each Control Module is working and that its address is correct.
4-8
Printer Menus
The printer commands are accessed through the Operations Menu. This menu allows you to select the information you want to print. The printer menu provides you with the choice of three data formats, current, high/ low, and graph.
1 WEEK HIGH/LOW DATA High Low pH = High Low ORP = High Low Temp = pH = 8.32 7.95 ORP = 165 mv 280 mv Temp = 78.0F 77.5F
Current Data
When you select current data, you will receive a printout showing the current readings for each of the systems that you are monitoring or controlling with the Octopus . Your printout will look like this:
Graph Data
In the graph mode, the Octopus plots the data that it has been collected over the week of operation. The printer will automatically scale the Y-axis to make sure that the plot will always display all of the data points. The X-axis tick marks are labeled in hours if you are plotting in 24-hour or 48-hour mode. If you plot the last week's data, the tick marks represent one day of data collection. The most recent data sample is always shown on the far right hand side of the graph. On all the plots, each mark represents the hourly collection point. The graphs shown in Figures 4.9, 4.10, and 4.11 are rough approximations of the type of charting that the
4-10
4-12
Chapter
5
Setting up AquaWeb for Windows 95
Setting up AquaWeb
AQUAWEB FOR WINDOWS 9X - INTRODUCTION
A key element of Aquadynes product strategy is AquaWeb. AquaWeb is an integrated desktop application that offers networking and remote access to one or a network of Aquadyne computers. Once installed and properly configured, AquaWeb harvests water-quality data and node configuration from one or many Aquadyne computers. The data is stored Microsoft Access database --on a PC or a centralize server where the data can be shared and analyzed by one or many users. AquaWeb is a powerful analytical tool. After downloading the datalog, AquaWeb will display the data in a graphical format using simple line or statistical graphs that identify underlying trends, giving you the information you need to fine-tune parameter set points or alram ranges. AquaWebs security system allows only authorized users access to your systems and your data. All Aquadyne computers are shipped with software that enables communications with AquaWeb. There are three versions of AquaWeb available: Standard Edition, AquaWebSE Database Edition, AquaWebDB Network Edition, AquaWebNE The Standard Edition, AquaWebSE, is available free of charge to all our customers or to anyone who wishes to evaluate the product. The features installed in AquaWebSE allow you to view and graph data that resides in the datalog of any Aquadyne computer with H2O/S software. AquaWebSE does not include support for a Microsoft Access database --but it is a fully functional product in every other respect. Data is downloaded and graphed in real-time using the datalog stored in the remotely installed Aquadyne computer This typically 7 days for the Octopus and 30 days of hourly data for the AquaNode and AquaGuard. AquaWeb is designed around a powerful graphing system that supports simple line and bar graphs, or more sophisticated best curve line and points and spline graphs. The Database Edition, AquaWebDB includes all of the features of AquaWebSE plus support for a Microsoft Access database (run-time version included), that can be located on a your PC or on a centralized network server. The database receives data from AquaWeb and permanently store all data and configuation information that has be been download from a node. The Network Edition, AquaWebNE, adds the additional capabiity to store data for multiple Aquadyne nodes. This version is designed for customers who require remote access to a wide-area network (WAN) of a few or a few thousand Aquadyne nodes. Nodes are graphically displayed using an intuitively simple hierarchical treestructured view that is very similar to the Explorer view that is familar to PC and MAC users AquaWeb uses the Windows 95 Dial Up Networking system and and an integrated scheduler to poll each unit and download the node internal datalog automatically at user-define intervals. The CD-ROM demo version of AquaWebDB is shipped with every Aquadyne computer. You will have the full use of the software for a period of 30 days, after which time the trial period will expire. For Aquadyne customers, AquaWeb will revert to AquaWebSE allowing you to continue to use the software with your Aquadyne computer --but without the Microsoft Access Database. Customers wishing to upgrade to AquaWebDB or AquaWebSE must contact Aquadyne to register their software, tender the appropriate payment and receive a permanent software unlock code.
5-2
Setting up AquaWeb
PC REQUIREMENTS
AquaWeb is a 32-bit Windows 95 application, and is NOT compatible with Windows NT, Windows 3.1, MS-DOS, or MAC O/S. To install AquaWeb, you will need a PC with the following minimum configuration: At least 24 megabytes (MB) of RAM and 10 megabytes (MB) of available hard disk space. (32MB RAM or more is recommended.) At least 100MB disk space after all applica tions are opened. The capability to connect to the Internet. AquaWeb works on any properlyconfigured Windows 95 system where dial-up networking (such as CAUTION Microsoft Network) is working for other applications. Before you set up AquaWeb, verify that the dial-up networking is operational. If it is operational you would be able to dial up America Online (AOL), CompuServe, or other internet service provider.
Even after the system has been operational for some while, if you experience problems with AquaWeb, go back to the Windows 95 environment and confirm that the dial-up networking is operational before trying to resolve an apparent problem with AquaWeb. Contact the vendor of your networking software to troubleshoot your unsuccessful Internet connection. For example, if you use the Windows Networking software, first visit www.microsoft.com/support for late-breaking updates and technical information. Or, on the Windows 95 taskbar, click the Start button, click Help, click Troubleshooting, and then double-click a topic such as "If you have trouble using Dial-Up Networking." Another resource is the Readme file on your Windows CD-ROM and/or your printed Windows documentation. Lastly, US and Canadian customers can contact Microsoft Technical Support, and all other customers can contact Microsoft Product Support Services at the Microsoft
NOTE:
In order to properly install and operate AquaWeb, we recommend that you follow, step by step, the installation procedure describe in this chapter: STEP 1 CONFIGURING WINDOWS 95 COMMUNICATION PORTS SETTING TO SUPPORT AQUAWEB. STEP 2 INSTALLING AQUAWEB ON YOUR PC FROM A CD-ROM OR VIA INTERNET. STEP 3 STARTING AND CONFIGURING AQUAWEB FOR DIRECT WIRE CONNECTION OR FOR A MODEM CONNECTION.
5-3
Setting up AquaWeb
STEP 1CONFIGURING WINDOWS 95 COMMUNICATION PORTS SETTING TO SUPPORT AQUAWEB.
Communications between an Aquadyne comptuer and the AquaWeb PC can be made using either a Modem or Direct Wire Connection but not both simultaneously. If you are communicating via direct wire connection, an Aquadyne Direct Connect Cable must be used. If you plan to use AquaWeb with a modem connection, iIt is required that your PC modem be compatible with and properly installed under Windows 9x as a recognized Plug and Play device. We strongly recommend the use of a US Robotics Modem Sportster since this is the only modem that we are able to offer technical support. Your modem also MUST be installed on a separate COM port such as COM3, with its own independent IRQ assigned. To make sure your machine meets these requirements, consult with a your computer technician Attn. International customers: Please contact your distributor to determine which modem is being supported in your country. Before getting started, the first step is to properly configure the port settings. To do so: 1. 2. Go to the Start Menu and select Settings and then select the Control Panel. Select the Modem Icon if you are using a modem in a remote data application or if you are connecting up directly to a comm port on your PC go to the Systems Icon and under the Device Manager tab, select the comm port you are using for your Octopus (fig. 5.1). For a direct cable connection, click on Properties and select the Port Settings tab. Click on Advanced. The Advances port Settings dialog box will appear on the screen. (Fig 5.2)
5.
6. Click Ok to change the communications settings and exit out of the communications control and proceed with the starting of AquaWeb.
3.
Setting up AquaWeb
STEP 2INSTALLING AQUAWEB
AquaWeb installs quickly on your PC using Microsoft's latest software insallation utility for Windows 95compliant products.
6. 7.
5-5
Setting up AquaWeb
8. 9. The Aquadyne license agreement box will appear. After reading it and upon agreement click OK. You will be then asked to review the ReadMe file before the installation. It is VERY important that you review the RedMe file BEFORE installation since there are many installation related issues that you need to be aware of. 12. When the dialog box says AquaWeb Setup was completed successfully, click OK.
Setting up AquaWeb
2.
3.
Type both exactly as spelled and capitalized here because the login procedure is Case-sensitive. 4. Click OK on the login screen to open AquaWeb. AquaWeb is ready when you see a sample Network map displayed as a "tree" list (Fig.5.7)
5-7
Setting up AquaWeb
access to the port. 4. Select Dialing Properties and fill up the My Location dialog tab.
2. 3.
5-8
Chapter
6
ADLoader
Upgrading theH2O/S Software In your AquadyneComputer
ADLoader
UPGRADING THE H2O/S SOFTWARE IN YOUR AQUADYNE COMPUTER.
Your Aquadyne computer is designed using a Flash Memory architecture that allows the H2O/S system software in the computer to be upgraded at anytime. The Flash Memory can be reprogrammed using a special software utility available from Aquadyne called ADLoader. This software utility is available on CDROM that accompanies each Aquadyne product, or it can be downloaded from our WEB page at the following addresses: http://www.aquadyne.com 5. Once you selected the drive, locate the file called ADLoader.exe and double click on it. 6. 7. In the WinZip Self-Extractor dialog box, click Setup. The WinZip program will "unzip" a series of component files. Then a Setup message box will report further status. Follow the instructions provided within the installation application to complete the installation process. When the dialog box says ADLoader Setup was completed successfully, click OK.
8.
2.
3. On Aquadyne home page, select Support 4. On the Support page, click on Download the ADloader software 5. 6. In the download page, locate and click on the file ADLoader.EXE. When asked whether you want to open ADLoader.EXE or save it to disk, choose Save it to disk. We recommend that you save the file to your TEMP directory --usually \windows\temp. In the Windows 95 Save As dialog box, select the location you want to install ADLoader, leave the file name as ADLoader in the Name text box, and then click Save. The file is about 5MB. When the download is complete simply run the application ADLoader.exe. Follow the step 6 to 8 under Installing the ADLoader from a CDROM.
!
CAUTION
In rare situations, it may be desireable or necessary to force the H2O/S software into the download mode so that it can be re-programmed with new software. You can force the computer into the download mode by holding the Enter key and the Right button key at the same time. The LCD screen will confirm that it is in the download mode when the top half of the LCD screen shows 16 solid squares.
7.
8.
6-2
CHAPITRE 6: ADLOADER
ADLoader
Running the ADLoader Software
Once ADLoader is installed on your PC, upgrading your Aquadyne computer is a simple matter of running the ADLoader application. The entire process is automatic and requires no user intervention. ADLoader is designed to automatically connect with our Web server, and to check for a new release of the H2O/S system software. If there is a new release, the ADLoader will automatically download it from our FTP. ADLoader is capable of upgrading your Aquadyne Computer either by connecting the Aquadyne computer directly to you PC serial port, or by connecting to the Aquadyne computer to a PC using a modem. Refer to the Section 1 of this manual for information on both connection methods. The ADLoader software is very easy to use and will prompt you through the upgrade process. To begin the installation process, you will need to obtain an installation download key from Aquadyne. Both the serial number and the download key are required to complete the installation process. The download key can be obtained by sending a email to:
Fig. 6.2 Entering the Serial number and the download key
During the Download process all system functions such as monitoring, control and alarming are terminated. Make sure that the systems that the Octopus is managing are disabled during the download process.
!
CAUTION
email: [email protected]. Be sure to include the serial number of your Aquadyne Computer along with your name and full address. During the download sequence the Aquadyne computer is no longer running the H2O/S software. Once the
!
CAUTION
Telephone line conditions can frequently make it difficult (and sometimes impossible), to update your H2O/S software via a modem connection. If problems persist, we recommend that you use ADLoader with a direct wire connection only.
CHAPTER 6: ADLOADER
6-3
ADLoader
download process is started it MUST be allowed to finish successfully. If the process is terminated for any reason the Aqua-Node will automatically attempt to restart to installation process. Until the download process is completed successfully, the Aquadyne computer will NOT be able to perform any Monitoring or Control operations! After the download process is completed without error, the computer will perform a memory check an then it will re-boot itself. As the computer re-boots, you will see the version number of the new software just installed in the LCD screen. This confirms the proper installation of the H2O/S software.
6-4
CHAPITRE 6: ADLOADER
Chapter
7
Troubleshooting
Troubleshooting
The most common troubleshooting questions are in the following three areas: Probe readings, Peripheral equipment response, and Modem/pager feature.
PROBE TROUBLESHOOTING
Symptom
1. Probe reading fluctuates or only <<< or >>> are displayed a.
Possible Cause
Probe cap has not been removed
Solution
Verify that all protective caps have been removed from the probes.
b.
The pH and ORP probes must be located within 6 inches of the temperature probe.
Probes location must have sufficient water movement. Verify that probes are plugged into the proper Octopus port. c. Induced voltage interference Place pH, ORP, and temp probes in a cup of water. If readings stabilize, the problem may be due to induced voltage interference. (If probe readings do NOT stabilize, refer to Possible Cause #4) Motors, pumps, high energy halide lamp, electronic ballasts can leak voltage into the body of water. This can affect probe accuracy and longevity. The Octopus compensates for induced voltage up to 4 volts peak to peak. Voltage exceeding this level must be removed via the addition of a solution ground placed directly next to the equipment causing the leak.
7-2
CHAPTER 7: TROUBLESHOOTING
Troubleshooting
PROBE TROUBLESHOOTINGContd
Symptom
1. Probe reading fluctuates or only <<< or >>> are displayed Cont'd d.
Possible Cause
RF (Radio frequency) interference
Solution
Turn off all equipment other than the Octopus unit. If the probe readings stabilize, the problem may be a result of RF interference. (If probes DO NOT stabilize, call Aquadyne Technical Support at 619-569-2082.) High output equipment such as ozonizers, chillers, and electronic ballasts emit RF under normal operation. (RF can also be generated by household equipment such as TVs, refrigerators, dryers, etc.) The RF noise can be received by the probes or probe cables and will cause inaccurate readings to be displayed. If this is the problem attempt to relocate the probes as far away from the known source as possible. If this does not resolve the problem, the probes must be shielded suing metal conduit which has been grounded. If this does not resolve the problem, contact Aquadyne for assistance.
e.
Calibrate the probe using the procedure described in Chapter 4. Clean probe if Octopus message reads, "Clean or Replace Probe." If this does not resolve the problem, the probe may require replacement.
2.
a.
Probe is bad
CHAPTER 7: TROUBLESHOOTING
7-3
Troubleshooting
PROBE TROUBLESHOOTINGContd
Symptom
3. Abnormally low conductivity readings
Possible Cause
a. In-line probe has been inserted too far into the flow stream portion of the teefitting
Solution
Reposition the probe so that the probe extends only about 10% into the flow stream. Be sure the compression fitting is secure and that the probe is supported independesntly from the compression fitting.
b.
c.
Control Module may be plugged into a power strip with surge protection
7-4
CHAPTER 7: TROUBLESHOOTING
Troubleshooting
PERIPHERAL EQUIPMENT RESPONSE TROUBLESHOOTINGContd
Symptom
1. Equipment does not turn on/off at the proper time Cont'd d.
Possible Cause
Control Module and Octopus are on a different phase of power source.
Solution
Plug the Octopus, the X-10 Power Line Interface and all Control modules into power outlets that are on the SAME PHASE of the electrical sub-panel. If wish to place equipment on multiple phases, a signal bridge is required. If possible, attempt to use a different satellite and/or address to see if this resolves the problem.
e.
Malfunction with a particular Control Module or Module Address. RF (radio frequency) interference
f.
Refer to the RF interference discussion in the "Probe Problems" troubleshooting section above.
MODEM/PAGER TROUBLESHOOTING
1. No response from the modem a. Modem and Octopus controller are not talking Check to insure that the cables are fully plugged into the Octopus controller and the data rates and formats are set for the modem being used. The standard data rate is 19,200 baud with 8 data bits, 1 stop bit, and no parity (81N). Verify that power is being supplied to the modem when the Octopus controller is sending. 2. No modem dial tone a. The modem attempted to dial the phone but when it checked for dial tone, none was present Verify that the phone line is live and that the phone cable is plugged into the jack labeled LINE.
CHAPTER 7: TROUBLESHOOTING
7-5
Troubleshooting
MODEM/PAGER TROUBLESHOOTINGContd
Symptom
3. Bad Response message a.
Possible Cause
The Octopus controller was unable to understand the message sent back from the modem
Solution
Verify that the cables are plugged securely into the Octopus and that the data rates and formats are set for the modem being used. The standard data rate is 19,200 baud with 8 data bits, 1 stop bit, and no parity (81N). The modem should not be connected to a multi-line telephone system (such as PBX) as this may interfere with proper communication of the modem. Connect the modem to a single telephone direct line. When the alarm pager is activated, it dials the number that was entered in the Octopus softare. Then when it begins to transmit the page message, it first sends the serial number of the Octopus that is alarming. Depending on any extra features (such as voice mail) provided by your paging service,the first digit in the seriial number may activate that extra service (such as voice mail), the alarm page is forwarded to the voice mail, and the page is never sent. Verify with your pager service that there are no extra service features that could interfere with the alarm page.
4.
a.
b.
An extra feature (such as voice mailbox, etc.) on the paging service may be interfering with pager function
c.
7-6
CHAPTER 7: TROUBLESHOOTING
Troubleshooting
AQUAWEB TROUBLESHOOTING
Symptom
1. Octopus not responding to AquaWeb
Possible Cause
a. Incorrect Serial Data interface cable.
Solution
Verify that you are using the correct serial interface cable to connect between the Octopus and the computer.
Make sure that the datalogger options in the Octopus are set as follows: Data Interval = 0 Min Baud Rate = 19,200 Data Format 8,1,N
Many applications that make use of a communications port in the computer may attach themselves to the port and not allow other applications such as AquaWeb to use the port. Make sure to shut down all applications such as fax managers, and communications applications while trying to use AquaWeb.
CHAPTER 7: TROUBLESHOOTING
7-7
Troubleshooting
NOTES:
7-8
CHAPTER 7: TROUBLESHOOTING
Appendix
A
Menu Maps
Changes In System status Rotate thru Bottom Line Heater System is Active of the Display ORP System is Active pH Low Alarm
Back to Top Level Display Low Conductivity Dose Sec. = X System Hold XX:XX Remaining System Hold Start Hold? NO Buffer Doser Dose Min. = X
YES
APPENDIX A - 2
APPENDIX A - 3
ORP Calibration Enter to Select Unlock Now Enter to Select Exit Lock-Unlock Enter to Select Lock Now Enter to Select
APPENDIX A - 4
Exit Cond. Menu Enter to Select Exit Temp Scale Enter to Select
APPENDIX A - 5
A
Low Cond. Ctrl. Enter to Select High Cond. Ctrl. Enter to Select Exit Cond. Ctrl. Enter to Select
A
Timer 1 Control Enter to Select Timer 2 Control Enter to Select
A
Timer 3 Control Enter to Select
A
Timer 4 Control Enter to Select
A
Exit Timers Enter to Select
APPENDIX A - 6
APPENDIX A - 7
A A
XX ID ID = Y Unit = ZZ
APPENDIX A - 8
APPENDIX A - 9
XX/XX/XX XX:XX:XX AM
See Timer 1
See Timer 1
See Timer 1
XX/XX/XX Correct? Yes/No No Yes XX/XX/XX Enter Date Yes XX/XX/XX Correct? Yes/No No Yes No No
APPENDIX A - 10
XX Alarm Enabled
XX Alarm Disabled
APPENDIX A - 11
PAGER MENU
Pager Menu Enter to Select Pager Tone/Pulse Enter to Select Baud Rate Enter to Select Data Format Enter to Select Page Now Enter to Select Exit Pager Menu Enter to Select Pager Enable Enter to Select Pager Ph. Number Enter to Select Pager Delay Enter to Select Pager Retry Enter to Select Pager Retransmit Enter to Select
APPENDIX A - 12
Appendix
B
Understanding The Water Quality Parameters
Water Quality
UNDERSTANDING THE WATER QUALITY PARAMETERS
Increasing your understanding of basic water chemistry will enhance your ability to interpret the data generated by your Octopus 3000 controller. With the controller's assistance you can cultivate the ultimate aquatic environment. The following chapter covers four components vital to creating the optimal aquatic habitat: pH, ORP, temperature and conductivity. Temperature should be altered gradually, with changes of no more than 2 F or 1 C in a 24-hour period. Your Octopus controller is ideal for maintaining water temperature since it makes constant incremental alterations.
pH
The pH of water refers to the degree to which an aquatic environment is acidic or alkaline. The pH scale ranges from 0 to 14.0, with a pH of 7.0 being neutral. Below 7.0 is acidic, and above 7.0 is basic. Natural sea water, for example, has a pH of about 8.3 which is slightly alkaline. The typical pH range for tropical marine fish is between 7.6 and 8.5 The pH probe must be calibrated with calibration solutions to assure the accuracy of the reading. Because the pH changes on a logarithmic scale, not a linear one, the calibration solutions used should be on both sides of the average pH for the environment. For saltwater environments with an ideal pH of 8.3, the calibration solutions used should be 7 and 9 or 7 and 10. The pH of an aquatic environment fluctuates naturally from day to night. The pH tends to fall at night and rise, due to photosynthesis by plants, during the day. Therefore, a range of readings over the course of the day is normal, and a single test may not mean very much. Wide fluctuations in pH can be fatal to fish and marine organisms. Therefore, small changes of no more than 0.1 unit of pH a day can be made by adding a buffer solution. With help from your OCTOPUS 3000 controller, maintaining proper pH levels becomes easy. The controller monitors your pH levels and instructs the doser control system to
TEMPERATURE
It is frequently desirable that the temperature of an aquatic environment remain stable over the course of a day. Varying a degree or two (Fahrenheit) is sometimes acceptable, but larger fluctuations can cause problems. A heater or chiller is very helpful maintaining correct temperatures. It may be sufficient, however, simply to heat or air condition the room the aquatic environment is in. The body temperature of your aquatic pets and all other cold blooded animals is determined by their environment. As the temperature cools, their metabolic processes slow down. The ideal temperature is one at which your fish are active, eating and growing. The ideal temperature range for salt water environments is 75 to 78 F. The ideal range for fresh water environments is 76 to 80 F. If you have questions about the requirements of a particular species, refer to the specialist at your local fish store.
APPENDIX B - 2
Water Quality
make necessary adjustments to replenish your pH depleted aqua system. salinity will gradually rise over time. A decreasing salinity is often reflected by changes in the growth of algae (Esp. blue-green or other slimy algae, and a decline in the health of invertebrates.) In Appendix C on Page 5 shows the relationship between conductivity and salinity is shown. For freshwater applications, conductivity is generally used to determine the general water quality. In hydroponic applications conductivity is useful in measuring the concentration of nutrient solutions. Proper conductivity levels for various types of plants, in varying stages of growth are listed in the table below.
ORP
Oxidation Reduction Potential (ORP) is a measurement of the potential for chemical reactions which occur in the water. This potential is determined by measuring the millivolts (mV) present in the water. ORP provides a useful indication of water purity. High ORP values indicate oxidation reactions greatly exceed reducing reactions. Low ORP values indicate reducing reactions greatly exceed oxidation reactions. ORP values in natural sea water (NSW) typically range from 350 to 400 mV. This corresponds to the ideal range for aquariums as well. Artificial manipulation of ORP is accomplished with ozonization of the water.
CONDUCTIVITY
As its name implies conductivity measures the ability of a sample to conduct an electrical current. For liquids, conductivity is dependent on the amount of ionic compounds dissolved in the sample. Conductivity is useful for measuring total dissolved solids (TDS) and salinity. As a general rule, as ion concentration increases, conductivity rises. Although the conductivity measurement can not differentiate between the contributions from specific ions, it does give a general indication of the total ion concentration. In saltwater, conductivity is generally used to determine the approximate salinity equivalent. Because all aquatic systems will experience a loss of water through evaporation,
APPENDIX B - 3
Water Quality
APPENDIX B - 4
Appendix
C
Conversion Tables
Conversion Tables
Conductivity (mS) 65 60 55 50 45
Conductivity at 30 C
40
Conductivity at 25 C
35 30 25 20 15
10
15
20
25
30
35
40
APPENDIX C
Conversion Tables
Conductivity at 25 C (mS/cm) 40.0 4 1. 0 42.0 43.0 44.0 45.0 46.0 47.0 48.0 49.0 50.0 5 1. 0 52.0 53.0 54.0 55.0
Salinity at at 25 C (0/00) 25.2 26.2 26.9 27.7 28.4 29.1 29.8 30.5 31.3 32.0 32.7 33.4 34.1 34.8 35.5 36.1
D e ns ity at 25 C (0/00) (g/cm3) 1.0187 1.0193 1.0198 1.0204 1. 0 2 0 9 1.0214 1. 0 2 2 0 1.0225 1.0232 1.0238 1.0244 1.0250 1.0256 1.0262 1.0268 1.0274
APPENDIX C
Conversion Tables
APPENDIX C
Appendix
D
Product Warranty
Product Warranty
Aquadyne Computer Corporation warrants articles of equipment manufactured by it to be free from defects in material and workmanship under normal use and service. Aquadynes obligation under this warranty is limited to making good at its factory any defective controller or satellite which is returned intact to Aquadyne or to one of Aquadynes authorized service centers within one year of delivery of the product to the original purchaser. The shipping must be prepaid until the product has been determined to be defective. Probes supplied by Aquadyne carry a 90-day warranty from date of purchase with the same conditions as the controller and satellite warranty. Aquadyne warrants any repair or replacement of its products for thirty (30) days or for the remainder of the original warranty, whichever period is longer. This warranty shall not apply to any article of equipment which has been repaired or altered outside the Aquadyne Computer Corporation factory or authorized service centers, nor which has been subject to misuse, negligence or accident, incorrect wiring by others, or installation or use not in accord with instructions furnished by the manufacturer. Aquadynes warranty obligation has limited transportation expenses. Products believed by buyer to be defective shall be returned to Aquadyne, transportation and insurance prepaid by buyer. If Aquadyne determines the product is defective in operation, workmanship or material, the product will be repaired or replaced and returned to the buyer with transportation and insurance prepaid. If Aquadyne determines that the product is operating normally and the buyer was misusing the product, Aquadyne may charge the buyer for labor and shipping charges. Aquadyne will give the buyer every possible consideration and will notify the buyer of products not subject to warranty allowance; and promptly upon receiving such notice, the buyer will notify Aquadyne as to whether the product should be repaired, returned to the buyer without repair or otherwise disposed of. Products left in Aquadynes possession longer than thirty (30) days without the buyers notice of disposition will be subject to a stocking and handling charge. For replaced items, the buyer will pay the established Aquadyne price. For repair labor, the buyer will pay Aquadynes established maintenance service rate. Under all circumstances, a product not subject to warranty coverage will free Aquadyne of all transportation charges.
APPENDIX D - 2
Product Warranty
EXCEPT FOR THE EXPRESS WARRANTY SET FORTH HEREIN, AQUADYNE COMPUTER CORPORATION GRANTS NO WARRANTIES, EITHER EXPRESS OR IMPLIED, OF MERCHANTABILITY AND FITNESS. THE STATED EXPRESS WARRANTY IS IN LIEU OF ALL LIABILITIES OR OBLIGATIONS OF AQUADYNE COMPUTER CORPORATION FOR DAMAGES INCLUDING, BUT NOT LIMITED TO, CONSEQUENTIAL DAMAGES OCCURRING OUT OF OR IN CONNECTION WITH THE DELIVERY, USE OR PERFORMANCE OF AQUADYNE COMPUTER CORPORATION PRODUCTS. Buyers remedies for breach of warranty shall be limited to repair, or replacement subject to adjustment as stated herein, and full or partial adjustment to purchase price.
APPENDIX D - 3