CPG 19
CPG 19
PRACTICE GUIDELINE
Care of the
Contact Lens
Patient
OPTOMETRY:
THE PRIMARY EYE CARE PROFESSION
The mission of the profession of optometry is to fulfill the vision and eye
care needs of the public through clinical care, research, and education, all
of which enhance the quality of life.
OPTOMETRIC CLINICAL PRACTICE GUIDELINE
CARE OF THE CONTACT LENS PATIENT
Printed in U.S.A
NOTE: Clinicians should not rely on the Clinical Guideline
alone for patient care and management. Refer to the
listed references and other sources for a more
detailed analysis and discussion of research and
patient care information. The information in the
Guideline is current as of the date of publication. It
will be reviewed periodically and revised as needed.
Contact Lens iii
TABLE OF CONTENTS
INTRODUCTION
I. STATEMENT OF THE PROBLEM
A. History and Epidemiology of the Use of Contact Lenses
B. General Considerations
CONCLUSION
III. REFERENCES
IV. APPENDIX
Figure 1: Corneal Stain
Figure 2: Corneal Infiltrates
Figure 3: Conjunctival Injection (Conjunctivitis)
Figure 4: Contact Lens Induced Corneal Hypoxia
Figure 5: "3/9" or Juxtaposition Corneal Stain
Figure 6: Corneal Abrasion
Figure 7: Giant Papillary Conjunctivitis
Abbreviations of Commonly Used Terms
Glossary
Introduction 1
INTRODUCTION
This Optometric Clinical Practice Guideline for the Care of the Contact
Lens Patient describes appropriate examination and treatment procedures
for patients wearing contact lenses (CLs). It contains recommendations
for timely diagnosis and treatment, and, when needed, referral for
consultation with or treatment by another health care provider. This
Guideline will assist optometrists in achieving the following goals:
B. General Considerations
A. Pre-Fitting Considerations
1. Indications
Some factors that suggest whether a patient is a good candidate for CL
wear involve optical, physiologic, and cosmetic considerations. The
following indications should be considered in the evaluation of a
patient’s potential for successful CL use. (Table 1)
a. Optical Factors
Contact lenses improve visual function by neutralizing ametropia, or
minimizing distortion, especially when the patient suffers from more
than a modest spherical refractive error or astigmatism, regular or
irregular. Myopic patients benefit from the increased magnification
provided by CLs, compared with their spectacle corrections. The reverse
is true for both hyperopic and aphakic patients; however, such patients
benefit from enhanced fields of vision with CLs.10 For anisometropic
patients, aniseikonia and prismatic effects may be reduced or eliminated
with CL wear.
b. Presbyopia
Although many patients with presbyopia wear CLs, presbyopia is not
specifically an indication for CL correction. Presbyopic patients may
wear distance CLs and use additional reading spectacles of various types
to address their presbyopia. Alternatively, presbyopes (especially
emerging presbyopes) often successfully use what has been termed
“monovision” correction in which one eye wears a CL to correct for
distance vision and the other wears a CL to correct for near vision.
Various bifocal CLs are available in either RGP or hydrogel materials.
The Care Process 5
c. Therapeutic Potential
Contact lenses have been used to manage both aphakia and binocular
vision problems, especially accommodative esotropia and convergence
excess.11,12 Contact lenses, particularly rigid CLs, can optically smooth
an anterior corneal surface made irregular by disease (e.g., keratoconus
or corneal microbial infection), trauma, or surgery (e.g., penetrating
keratoplasty or ineffective refractive surgery). Hydrogel lenses are used
as ophthalmic bandages13 following corneal trauma or refractive corneal
surgery. Rigid CLs also have been used to manage14 or reduce15 myopia.
Both clear and tinted rigid and soft contact lenses have been used for
treatment by occlusion in cases of diplopia and amblyopia.16
d. Cosmetic Effect
Correcting ametropia by placing a lens directly on the corneal surface
improves cosmesis by eliminating the need for a spectacle frame and
often unattractive corrective ophthalmic lenses. Some patients elect to
wear colored CLs simply to change the appearance of their eye color.
Opaque contact lenses also may be used for their prosthetic effect (e.g.,
masking an unattractive corneal scar or damaged iris or providing an
artificial pupil in the treatment of aniridia).17
Insert Table 1
2. Cautions
Any patient whose clinical situation suggests increased risk of ocular
infection or inflammation, but who insists on cosmetic CL fitting, should
give formal informed consent before the clinician provides CLs.18
Several factors could limit a patient’s suitability for CL wear, as
discussed below. (Table 2)
a. Ocular Considerations
Cosmetic CL wear should be approached cautiously with patients who
present with any active anterior segment disease, especially ocular (or
adnexal) inflammation, infection, or severe dry eye conditions, because
of the possible increased risk of complications, especially corneal NV or
infection. Such diseases include acne rosacea, Sjögren syndrome, atopic
6 Contact Lens
Placing the lens directly in the precorneal tear film increases the risk of
tissue compromise. CL use should therefore be approached cautiously
for either the monocular patient (because of risk to the patient’s only
useful eye) or for the patient who is engaged in an avocation or vocation
with exposure to a particularly dirty or dry environment. Such
individuals may be advised to wear protective spectacles.
b. Systemic Considerations
Other indications for caution include the patient’s inability to manipulate
and care for CLs appropriately or to return for appropriate professional
supervision. Contact lens wear should be approached cautiously with the
patient who has immunosuppressive disease (e.g., AIDS), rheumatoid
arthritis, or diabetes, which may lead to insufficient lacrimation or
increased risk for corneal NV and infections.19,20
c. Noncompliant Patients
Clinicians should exercise caution, and occasionally exercise restraint,
when considering CL fitting for patients known or suspected to be so
noncompliant with appropriate CL care and general hygiene as to place
themselves at increased risk for severe complications (See Section IIE1).
Insert Table 2
The Care Process 7
a. Hydrogel Lenses
Spherical hydrogel CLs are indicated for the correction of myopia and
hyperopia when astigmatism is limited to less than 1.00 diopter (D)21,22
and tears are sufficient. Stock optical powers are commonly available
between +6.00 D and -20.00 D; higher custom powers are also available
(e.g., for cases of aphakia). Some hydrogel CLs, depending upon their
power and thickness profiles, may be difficult for some patients to insert
and remove.
The U.S. Food and Drug Administration (FDA) has classified all
hydrogel materials into four groups,23 which are believed to behave the
same chemically (Table 3). Oxygen permeability (Dk) of the hydrogel
materials in all groups increases with water content (WC).24 Oxygen
transmissibility (Dk/t) is lens specific, thus directly dependent on both
the WC (hence Dk) of the CL’s material and the reciprocal of its
individual thickness (t) profile.25-28 Another class of hydrogel CL
materials, in which silicone (for enhanced Dk) is blended with hydrogel
materials (for comfort), is also available.29
Insert Table 3
Toric hydrogel lenses30-32 are indicated for patients who are otherwise
good candidates for hydrogel CLs and who wish to use CLs for cosmetic
correction of refractive error, including visually significant astigmatism
(usually greater than 0.75 D). Standard designs frequently correct
astigmatism up to about 2.00 D; some custom hydrogel CL designs are
available to correct up to about 8.00 D. Toric hydrogel lenses are more
expensive than the spherical designs, and may not provide universally
stable visual results.33
Variable optical results and comfort levels may occur in patients who
have insufficient tears with all types of hydrogel CLs, especially toric
8 Contact Lens
b. Rigid Lenses
Rigid corneal CLs usually provide better visual results than do hydrogel
CLs in situations of either regular or irregular astigmatism of the corneal
surface. Insufficient tears usually will not affect the optics of rigid CLs,
but this condition does increase the prevalence of both intolerance and
some physiological complications. Rigid gas permeable CL materials
(Table 4) are available in a wide range of optical powers, oxygen
permeability,35 plastic "hardness," wettability, and specific gravity, all of
which affect lens design and positioning.36 Usually, the more oxygen
permeable the plastic, the more fragile the finished CL. PMMA CLs are
occasionally useful, although the clinician must recognize that this
material has virtually no oxygen permeability and that corneal
metabolism is totally dependent on tear exchange when CLs made of this
material are worn. Concern about hypoxia in patients with corneal grafts
or previous superficial pannus, possibly from the use of hydrogel CLs of
optical powers in excess of -10.00 D,37 is an indication for the use of
RGPs. Clinicians should note that the use of rigid CLs may be less
successful in dusty environments.
Scleral or haptic high-Dk RGP (or even PMMA) CLs can be used in the
management of keratoconus or other therapeutic cases such as ocular
cicatricial pemphigoid or Stevens-Johnson Syndrome.
Insert Table 4
use, such lenses are extremely helpful in rare cases of regular or irregular
corneal astigmatism (including keratoconus) or aphakia.
*
Refer to the Optometric Clinical Practice Guideline for the Comprehensive Adult Eye
and Vision Examination.
10 Contact Lens
base curve and total diameter. The refractive astigmatic axis is stabilized
by prism, truncation, superior/inferior thin zones, or a combination of
methods.30-32 The astigmatic axis of the contact lens cylinder should be
prescribed as close as possible to the patient’s astigmatic axis, after
accounting for the estimated rotation of the lens on the eye;52 the optical
power of the patient’s astigmatism can often be undercorrected without
compromise to visual acuity, which may result in less visual disturbance
caused by any alignment variability or misrotation.32
When selecting the initial diagnostic RGP lens BCOR, the clinician
should begin with the previously measured corneal curvature values as
an initial guide. For many patients, in achieving a physically aligned fit,
the more spherical the Ks, the more likely that the optimum RGP CL
BCOR will be slightly flatter than the flat K. The more astigmatic the
Ks, the more likely it is that the appropriate base curve will be close to
the mean K. Some clinicians alternatively elect to achieve slight apical
vault by selection of BCOR/TD. Changes in the BCOR of RGP CLs will
directly affect the optical power of the CL/eye system and will require
direct optical power compensation.
In general, the flatter, more myopic, or more astigmatic the cornea, the
larger the TD that is required to achieve an optimum CL/cornea
relationship and vice-versa. A TD of approximately 9.0 mm is a good
starting point for most modern RGP CL designs, but clinicians
effectively prescribe RGPs with TDs ranging from less than 8.0 to
12 Contact Lens
greater than 11.0 mm. An optic zone that approximates the same value
as the BCOR (about 1.2 mm smaller than the TD) is common.
RGPs that ride low on the patient's cornea and move minimally should be
avoided. Adequate CL position and movement encourage the exchange
of tears, which pumps fresh oxygen from the air under the lens and
washes out debris and metabolic waste. Appropriate position and
adequate movement of the CL also minimize lens binding (in which
adherence to the underlying corneal surface leaves a physical impression
of the lens edge in the tissue). Lens binding may lead to 3/9 corneal
desiccation staining, which in turn can result in peripheral corneal
epithelial hypertrophy, vascularization, dellen, or even microbial
infection.34,60-64
The posterior peripheral curve system should be designed to lift the edge
of the CL gently off the corneal surface to provide a reservoir of tears for
exchange that maintains CL movement. This prevents chafing, due to
low edge lift, or drying of the peripheral cornea, due to high edge lift.65
The edge should also be well shaped and smooth.
c. Fenestrations
Small holes drilled through a CL are called fenestrations. They are
intended to improve oxygenation either directly or by encouraging tear
exchange.71
d. Blending
Smoothing or blending of the junctions between curvatures on the
posterior surface of RGP lenses may enhance comfort and reduces
corneal chafing or trauma.
4. Special Concerns
Some areas of CL application deserve additional discussion, especially
presbyopic correction, the use of CLs when eyes are “dry,” and extended
wear.
a. Presbyopia
Bifocal CLs for presbyopia are optically complex. Successful use is
subject to many patient-specific factors and the doctor's experience, skill,
and willingness to persist through fitting challenges. Thus, despite recent
improvements in both RGP and hydrogel designs, they still have limited
utility.
Two design philosophies guide distance and near correction with bifocal
contact lenses. “Simultaneous vision” bifocal (or multifocal) CLs
typically require consistent optimal positioning over the patient’s pupils.
In contrast, “alternating vision” lenses are intended to optimize distance
vision while the patient’s eyes are in the primary position, then reliably
move on the corneal surface so that a large portion of the near vision
optical zone covers the pupil in downgaze.
b. Dry Eye
Many patients with mild dry eyes may be helped to tolerate CLs.76 Both
systemic and ocular aspects of the dry eye condition should be managed
prior to and during contact lens wear. Instruction in lid hygiene and the
prescription of artificial tear drops, particularly unpreserved unit doses,
are two often helpful treatments. There is some evidence that contact
lens wear can cause or aggravate dry eye,77-79 increasing the importance
of such care.
c. Extended Wear
CLs have been provided for use on extended- or continuous-wear
schedules for many years. Both the prevalence and severity of all
complications, especially microbial infection,6-9 are increased when CL
wear is extended through one or more sleep cycles, as discussed in the
next section. Current FDA guidelines limit extended wear of approved
lenses to no more than six nights in succession.81 Because of increased
risk of complications, patients who elect extended CL wear should give
formal informed consent.
The patient should be taught to perform the following steps in the care
and handling of a CL:
• Wash hands.
• Clean each CL by gently rubbing and thoroughly rinsing with an
appropriate solution.
• Store and disinfect CLs in fresh appropriate solution for an
appropriate time interval in a clean case until reinsertion in the
eyes.
The Care Process 17
D. Progress Evaluations
Followup visits are important for proper management of the patient with
CLs. Planned evaluation should occur during the initial weeks and
months of CL wear to allow any necessary mechanical or optical
refinements in lens prescription(s), to monitor adaptation and minimize
ocular complications, and to reinforce appropriate CL care. Subsequent
evaluations are usually indicated at 6-to-12 month intervals for healthy
patients wearing cosmetic CLs.82,83 It is advisable to see patients who
may be at additional risk for ocular compromise during CL wear more
often than every 6 months, perhaps every 3 or 4 months or even more
frequently. Such patients include those using CLs for extended wear,
those wearing CLs for treatment of eye disease (e.g., keratoconus), or
following corneal trauma or surgery, and children wearing CLs for the
prevention or treatment of myopia14 or for correction of aphakia,84 for
example.
1. General Considerations
The most effective way to address the complications of CL wear is to
prevent them from occurring. One method of precluding many
complications is to maintain CL care and hygiene, consistent with both
common sense and FDA-approved manufacturers’ guidelines.
Achieving and maintaining total patient compliance with recommended
CL care, however, is often difficult.89-92
2. Noninfectious Complications
By far, the most prevalent complications of CL wear are associated with
lens care and solutions and CL spoilage, particularly in the case of
hydrogel lenses (Table 6).95-99
__________
Insert Table 6
a. Solution Reactions
The majority of these problems are cell-mediated (Gell-Coombs type IV)
reactions to preservatives,130 but the anterior segment signs are often
nonspecific. Solution reactions often present with both fine corneal
staining, with or without infiltrates, and conjunctival injection and/or
edema. When the clinician suspects such a reaction, CL wear should be
discontinued, and appropriate treatment and professional observation
should be initiated. After reversal of the reaction, the clinician may
initially try substituting one product or class of product for another.
22 Contact Lens
When this measure fails, hydrogel CL wearers may be fitted with daily
disposable CLs to eliminate all solution issues. RGP wearers can use
aerosol-packaged nonpreserved saline to rinse their lenses copiously
prior to insertion. For fear of an Acanthamoeba infection, the use of tap
water or fresh water rinses is discouraged. If water is utilized for rinsing
RGPs, an additional rinse with sterile saline or conditioning solution is
recommended (See Appendix Figure 3).
b. Hypoxia
In the mid-1970s, all rigid CLs were made of non-oxygen-permeable
PMMA, and early hydrogel lenses all had modest oxygen
transmissibility. Hypoxia was a common complication. Most of the
RGP and hydrogel CLs now available, however, generally do not cause
corneal hypoxia under daily wear conditions. When there is clear
evidence of hypoxic corneal changes (e.g., epithelial or stromal edema,
corneal pannus37 greater than approximately 2 mm unrelated to 3/9
stain), conjunctival146 changes, or suspected CES,114 the clinician should
adjust the CL wear schedule or change the CL material or design to
enhance the availability of oxygen to the anterior corneal surface (See
Appendix Figure 4).
managed. When all attempts to solve the problem with RGPs are
unsuccessful, and when there are no contraindications, the prescription of
hydrogel CLs may also be considered34 (See Appendix Figure 5).
d. Corneal Abrasion
Corneal epithelial abrasion is a common occurrence among CL wearers.
The clinician can expect to spend approximately one percent of CL-
related office visits treating abrasions; more in practices where
keratoconic patients are numerous.127 Treatment consists of first ruling
out infection and temporarily discontinuing CL wear. Some clinicians
believe in prophylactic antibiotic treatment, while others prefer to
withhold antibiotics unless infection is suspected or proven. To decrease
the risk of precipitating or enhancing a microbial corneal infection, the
clinician should neither patch nor use topical steroids to treat a CL-
associated abrasion (See Section IIE3a). Close professional supervision
is prudent until the epithelial defect has closed and the etiology of the
abrasion should be considered before CL wear is resumed. For example,
when the cause of the abrasion appears to be the patient’s failure to insert
or remove the CLs properly, reinstruction in these procedures should
precede CL redispensing. Management of the patient with repeated
apical corneal abrasions,148 in particular the patient with keratoconus,
may require refitting of the CLs with steeper BCORs or the use of
piggyback CL systems39 (See Appendix Figure 6).
If at all possible, the patient diagnosed with GPC should first discontinue
CL wear until he or she is symptom (itching) free and the signs (mucus,
inflammatory tarsal conjunctival papillae) are subsiding. CL wear may
then be resumed cautiously, with improved CL cleaning (e.g., more
frequent, increased use of enzyme cleaner). The use of peroxide
24 Contact Lens
3. Infectious Complications
Corneal microbial infection, which has an incidence of about 21 of every
10,000 people using CLs for extended-wear and about 4 per 10,000
people using CLs for daily-wear per year,6,7 is probably the
CL-associated complication of most concern to both patients and
practitioners. Microbial corneal infections are identified by the
symptoms of ocular pain and photophobia and by the observation of
clinical signs such as corneal epithelial defects in association with
underlying inflammatory infiltration, often accompanied by anterior
chamber reaction (including hypopyon in some cases), conjunctival
discharge, lid swelling, and conjunctival injection.99,132
a. Bacterial Infections
Corneal infections associated with CL wear are usually bacterial,
primarily attributable to the Gram-negative Pseudomonas aeruginosa,
but also commonly due to Gram-positive Staphylococcus aureus and
Staphylococcus epidermidis.4,6,7,20,132,154-158 Other bacteria are also
occasionally cultured from such lesions. Bacterial corneal infection has
been primarily associated with wearing CLs through one or more sleep
cycles (extended or continuous wear).
The Care Process 25
b. Acanthamoeba Infections
The clinician should always consider the possibility of Acanthamoeba
species infections in any CL-related keratitis, especially in cases of
chronic disease with initially negative culture results and failure to
respond to antibiotic therapy. Clinical suspicion should be increased
when the patient reports extreme ocular pain and/or a history of exposing
the CLs to nonsterile water, or when an unusual epitheliopathy
(reminiscent of herpetic epithelial disease) or peripheral corneal radial
neuropathy is observed.159,172-174 Special culture techniques are available
for Acanthamoeba infections, but tissue biopsy is often necessary.
c. Fungal Infections
Fungal corneal infections are extremely rare among cosmetic CL
wearers. Most cases reported in the literature have involved the use of
bandage CLs or chronic treatment with topical steroids in patients
suffering from concurrent ocular disease (e.g., neurotrophic epithelial
defects, diabetes, trauma).177,178 Antifungal pharmaceutical agents (both
commercial and custom-made) are available, but medical treatment is
often quite difficult and prone to failure. It is important to note that
atypical mycobacterium and Acanthamoeba infections often mimic
fungal corneal ulcers and vice-versa.
The Care Process 27
d. Viral Infections
Concomitant viral corneal infections, of which adenovirus and herpes
simplex virus are of principal concern, can occur during CL wear. No
causative association has been uncovered for such viral infections, but
CL wear should be discontinued during viral infections unless the CL is
being used in a treatment protocol. Adenovirus infection is usually
successfully managed by supportive therapy such as tear supplements
and topical decongestants or by steroid therapy, as the clinical condition
indicates. Effective antiviral agents are available for the treatment of
herpetic eye disease. The clinician who observes an apparent herpetic
keratitis in association with use of CLs, however, should always consider
the possibility of Acanthamoeba as an alternative infectious agent.
It is prudent to consider discarding CLs that have been worn during the
period of viral infection and dispense new CLs once the infection has
resolved. Otherwise,, some effort should be made to disinfect CLs,
perhaps by soaking them in an appropriate disinfecting solution (e.g.,
CL-grade 3% hydrogen peroxide) for 10-15 minutes.
CONCLUSION
Complications that can threaten vision and persist after CL removal, such
as active microbial keratitis and deep stromal NV, are rare. Limiting the
use of prescribed CLs to daily wear, with adequate professional
supervision, and patient compliance with both the principles of good
personal hygiene and the published advice of the manufacturers of CLs
and solutions, results in CL wear that is safe for the vast majority of
patients.
References 29
III. REFERENCES
6. Schein OD, Glynn RJ, Poggio ED, et al. The relative risk of
ulcerative keratitis among users of daily wear and extended wear
soft contact lenses. New Engl J Med 1989; 321:773-8.
10. Westheimer G. The visual world of the new contact lens wearer. J
Am Optom Assoc 1962; 34:135-8.
30 Contact Lens
11. Robertson DM, Ogle KN, Dyer JA. Influence of contact lenses on
accommodation. Am J Ophthalmol 1967; 64:860-71.
15. Polse KA, Brand RJ, Vastine DW, Schwalbe JS. Corneal change
accompanying orthokeratology. Plastic or elastic? Results of a
randomized clinical trial. Arch Ophthalmol 1983; 101:1873-8.
16. Lee JR, Refojo MF, Leong FL. Tinted hydrogel lenses for
cosmetic and optical problems. Cont Lens Intraocular Med J 1977;
3:22-6.
23. Thompson, TT. FDA Four Lens Groups. Tyler’s Quarterly Soft
Contact Lens Parameter Guide 1999; 16:first index page.
28. Weissman BA, Phan C. The L in Dk/L. Optom Vis Sci 1992;
69:639-44.
35. Refojo MF. The siloxane bond in contact lens materials: effect of
methyl and phenyl content on oxygen permeability of silicone
lenses. Int Contact Lens Clin 1984; 11:83-6.
37. Chan WK, Weissman BA. Corneal pannus associated with contact
lens wear. Am J Ophthalmol 1996; 121:540-6.
38. Baldone JA, Clark WB. Contact lens in the aphakic child. Cont
Lens Med Bull 1970; 3:25-7.
40. Nelson LB, Cutler SI, Calhoun JH, et al. Silsoft extended wear
contact lenses in pediatric aphakia. Ophthalmology 1985;
92:1529-31.
41. Maguen E, Martinez M, Rosner IR, et al. The use of Saturn II lens
in keratoconus. CLAO J 1991; 17:41-3.
References 33
43. Varga JH, Rubinfeld RS, Wolf TC, et al. Topical anesthetic abuse
ring keratitis. Report of four cases. Cornea 1997; 16:424-9.
44. Olson M, Sarver MD. Apical bearing and success with the Bausch
& Lomb Soflens (polymacon) contact lens. Am J Optom Physiol
Opt 1976; 53:173-6.
47. Roseman MJ, Frost A, Lawley ME. Effects of base curve on the
fit of thin, mid-water contact lenses. Int Cont Lens Clin 1993;
20:95-101.
55. Holden BA, Mertz GW. Critical oxygen levels to avoid corneal
edema for daily and extended wear contact lenses. Invest
Ophthalmol Vis Sci 1984; 25:1161-7.
56. Klyce SD. Stromal lactate accumulation can account for corneal
edema osmotically following epithelial hypoxia in the rabbit. J
Physiol 1981; 321:49-64.
57. Bonnano JA, Polse KA. Corneal acidosis during contact lens
wear: effects of hypoxia and CO2. Invest Ophthalmol Vis Sci
1987; 28:1514-20.
59. Korb DR, Korb JE. A new concept in contact lens design - Part I
and Part II. J Am Optom Assoc 1970; 41:1023-32.
61. Sarver MD, Nelson JL, Polse KA. Peripheral corneal staining
accompanying contact lens wear. J Am Optom Assoc 1969;
40:310-3.
References 35
63. Grohe RM, Lebow KA. Vascularized limbal keratitis. Int Cont
Lens Clin 1989; 16:197-208.
64. Swarbrick HA, Holden BA. Rigid gas permeable lens adherence:
a patient dependant phenomenon. Optom Vis Sci 1989; 66:269-
75.
66. Sarver MD. A toric base curve corneal contact lens with spherical
power effect. J Am Optom Assoc 1963; 34:1136-7.
67. Weissman BA, Chun MW. The use of spherical power effect
bitoric rigid contact lenses in hospital practice. J Am Optom
Assoc 1987; 58:626-30.
69. Bailey NJ. Special contact lenses and their applications - prism
contact lenses. Opt J Rev Optom 1960; 97:54-6.
70. Knoll HA. On the stability of the shape of the human cornea. Am
J Optom Physiol Opt 1976; 53:359-61.
72. Back AP, Holden BA, Hine NA. Correction of presbyopia with
contact lenses: comparative success rates with three systems.
Optom Vis Sci 1989; 66:518-25.
36 Contact Lens
75. Harris MG. Informed consent for presbyopic contact lens patients.
J Am Optom Assoc 1990; 60:717-23.
76. Lowther GE. Dryness, tears, and contact lens wear: Clinical
practice in contact lenses. Boston: Butterworth-Heinemann, 1997.
77. Farris RL. The dry eye: its mechanism and treatment; evidence
that contact lenses are a cause. CLAO J 1986; 12:234-60.
81. Tolbert M, Lippman RE. Contact lenses: the better the care the
safer the wear. In: FDA Consumer Magazine (DHHS Pub. no.
[FDA] 91-4220): revised April 1991.
83. Gruber E. Post-fitting care of soft lenses. In: Dabezies OH, ed.
Contact lenses: CLAO guide to basic science and clinical practice.
Orlando: Grune & Stratton, 1984:41.1-41.2.
85. Mandell RB. Contact lens practice, 3rd ed. Springfield: Thomas,
1981:349.
91. Collins MJ, Carney LG. Patient compliance and its influence on
contact lens wearing problems. Am J Optom Physiol Opt 1986;
63:952-6.
92. Donzis PB, Mondino BJ, Weissman BA, Bruckner DA. Microbial
contamination of contact lens care systems. Am J Ophthalmol
1987; 104:325-33.
97. Tripathi RC, Tripathi BJ, Ruben M. The pathology of soft contact
lens soilage. Ophthalmology 1980; 187:365-80.
106. Madigan MC, Holden BA, Kwok LS. Extended wear of contact
lenses can compromise corneal epithelial adhesion. Curr Eye Res
1987; 6:1257-60.
110. Hirji NK, Larke JR. Corneal thickness in extended wear of soft
contact lenses. Br J Ophthalmol 1979; 63:274-6.
111. Zantos SG, Holden BA. Transient endothelial changes soon after
wearing soft contact lenses. Am J Optom Physiol Opt 1977;
54:856-8.
117. Rengstorff RH. The Fort Dix report - a longitudinal study of the
effects of contact lenses. Am J Optom Arch Am Acad Optom
1965; 42:153-63.
119. Wilson SE, Lin DTC, Klyce SD, et al. Topographic changes in
contact lens induced corneal warpage. Ophthalmology 1980;
97:734-44.
121. Schanzer MC, Mehta RS, Arnold TP, et al. Irregular astigmatism
induced by annular tinted contact lenses. CLAO J 1989; 15:207-
11.
122. Kline LN, DeLuca TI. Pitting stain with soft contact lenses. J Am
Optom Assoc 1977; 48:372-6.
125. Orsborn GN, Zantos SG. Corneal desiccation staining with thin
high water content contact lenses. CLAO J 1988; 14:81-5.
References 41
126. Lansche RK, Lee RC. Acute complications from present day
corneal contact lenses. Arch Ophthalmol 1960; 64:275-85.
132. Stein RM, Clinch TE, Cohen EJ, et al. Infected vs sterile corneal
infiltrates in contact lens wearers. Am J Ophthalmol 1988;
105:632-6.
134. Allansmith MR, Korb DR, Greiner JV, et al. Giant papillary
conjunctivitis in contact lens wearers. Am J Ophthalmol 1977;
83:697-708.
146. Papas EB, Vajdic CM, Aust R, Holden BA. High oxygen
transmissibility soft contact lenses do not induce limbal hyperemia.
Curr Eye Res 1997; 16:942-8.
References 43
147. Schnider CM, Terry RL, Holden BA. Effect of lens design on
peripheral corneal desiccation. J Am Optom Assoc 1997; 68:163-
7.
148. Korb DR, Finnemore VM, Herman JP. Apical changes and
scarring in keratoconus as related to contact lens fitting techniques.
J Am Optom Assoc 1982; 53:199-205.
149. Meisler DM, Berzins UJ, Krachmer JH, et al. Cromolyn treatment
of giant papillary conjunctivitis. Arch Ophthalmol 1982;
100:1608-10.
152. Bucci FA, Lopatynsky MO, Jenkins PL, et al. Comparison of the
clinical performance of the Acuvue disposable contact lens and
CSI lens in patients with giant papillary conjunctivitis. Am J
Ophthalmol 1993; 115:454-9.
155. Cooper RL, Constable IJ. Infective keratitis in soft contact lens
wearers. Br J Ophthalmol 1977; 61:250-4.
44 Contact Lens
157. Weissman BA, Mondino BJ, Pettit TH, et al. Corneal ulcers
associated with extended wear soft contact lenses. Am J
Ophthalmol 1984; 97:476-81.
158. Mondino BJ, Weissman BA, Farb MD, et al. Corneal ulcers
associated with daily wear and extended wear contact lenses. Am
J Ophthalmol 1986; 102:58-65.
160. Baum JL, Jones DB. Initial therapy of suspected microbial corneal
ulcers. Surv Ophthalmol 1979; 24:97-116.
161. Levey SB, Katz HR, Abrams DA, et al. The role of cultures in the
management of ulcerative keratitis. Cornea 1997; 16:383-6.
162. Clemons CS, Cohen EJ, Arentsen JJ, et al. Pseudomonas ulcers
following patching of corneal abrasions associated with contact
lens wear. CLAO J 1987; 13:161-4.
165. O’Brien TP, Maguire MG, Fink NE, Alfonso E, et al. Efficacy of
ofloxacin vs cefazolin and tobramycin in the therapy of bacterial
keratitis. Arch Ophthalmol 1995; 113:1257-65.
References 45
175. Berger ST, Mondino BJ, Hoft RH, et al. Successful medical
management of Acanthamoeba keratitis. Am J Ophthalmol 1990;
110:395-403.
177. Wilson LA, Ahearn DG. Association of fungi with extended wear
soft contact lenses. Am J Ophthalmol 1986; 101:434-6.
178. Kent HD, Cohen EJ, Laibson PR, et al. Microbial keratitis and
corneal ulceration associated with therapeutic soft contact lenses.
CLAO J 1990; 16:49-52.
Appendix 47
Figure 8
ICD-9-CM CODES
Hypermetropia 367.0
Far-sightedness Hyperopia
Myopia 367.1
Near-sightedness
Astigmatism 367.2
Anisometropia 367.31
Aniseikonia 367.32
Presbyopia 367.4
Keratoconus 371.6
Descemetocele 371.72
Aphakia 379.31
50 Contact Lens
Cornea 918.1
Corneal abrasion
Superficial laceration
Dk Oxygen permeability
HEMA Hydroxyethylmethacrylate
IgE Immunoglobulin E
NV Neovascularization
TD Total diameter
WC Water content
Appendix 53
GLOSSARY
Back optical diameter (or zone) The central optical posterior surface
of the contact lens.
Back toric lens A contact lens which has a back surface cylinder and
spherical front surface for toric cornea fitting.
54 Contact Lens
Daily wear lens A contact lens requiring daily or more frequent removal
for cleaning and other purposes.
Edge lift The distance between an extension of the BCOR and the
absolute edge of the lens; when measured parallel to the optical axis,
56 Contact Lens
axial edge lift (AEL); when measured along the radius, radial edge lift
(REL).
Front optic diameter (or zone) The anterior optical surface of a contact
lens.
Front surface toric lens (FST) Contact lens with toric optics on only its
front surface and a spherical base curve , intended to correct residual
astigmatism.
Ocular rosacea Acne rosacea involving the eye or its adnexa, that may
include any or all of these chronic eye signs: blepharitis, meibomitis,
telangiectasia of the lids; insufficient tears; bulbar and corneal
epitheliopathies, corneal scarring and melting.
Scleral (or haptic) contact lens A large contact lens, covering most of
the front of the eye, including the bulbar conjunctiva as well as the
cornea.
Stromal striae Fine parallel lines seen in the deep stroma during corneal
swelling from contact lens-associated hypoxia, early Fuchs dystrophy, or
keratoconus. (Deeper frank folds in Descemet’s membrane usually are
not related to contact lens wear and are called “striate keratopathy”).
Toric lens A lens which has one surface with two meridians of
curvature, least and greatest curvatures, located at right angles to each
other; astigmatism.
__________________________________________________________
Sources:
__________________________________________________________
Cosmetic
Refractive error: anisometropia, myopia, hyperopia, regular
astigmatism
Prosthetic use
Therapeutic
Myopia management
Reduction (i.e., orthokeratology)
Maintenance
Aphakia
Keratoconus
Corneal irregularity secondary to trauma, disease, surgery
Bandage
Occlusion
Treatment of accommodative esotropia or convergence excess
__________________________________________________________
Table 2
__________________________________________________________
Ocular (local)
Active anterior segment disease, especially infection (e.g., severe
blepharitis or dacryocystitis)
Dry eye* possibly associated with Sjögren syndrome secondary to
rheumatoid arthritis, lupus, thryroid disease
Acne rosacea
Atopic dermatitis
Active filtering blebs
Decreased corneal sensitivity (e.g., neurotrophic)
Systemic
The presence of only one visually useful eye
Diabetes
Immunosuppression
Inability to care for CLs or to present periodically for
professional care
__________________________________________________________
Source:
1999 Food and Drug Administration (FDA) Four Lens Groups. In:
Thompson, TT. Tyler’s Quarterly Soft Contact Lens Parameter Guide
1999; 16:first index page.
Table 4
Based on: