Manual Gse 250
Manual Gse 250
Published by:
GSE 1525 Fairlane Circle Allen Park, MI 48101 USA
Information in this Technical Reference Manual is subject to change without notice due to correction or enhancement. The information described in this manual is solely the property of GSE. This manual may not be distributed without written permission of GSE.
GSE Locations
GSE 1525 Fairlane Circle Allen Park, MI 48101 U.S.A. www.gse-inc.com GSE Amprobe Europe GmbH Phone: +49 (0) 2161-59906-11 Fax: +49 (0) 2161-59906-20
Table of Contents
Chapter 1: Introduction _________________________________________________ 9
Manual Set _______________________________________________________________________ 10 Document Conventions _____________________________________________________________ 10 Specifications _____________________________________________________________________ 11
Load Cell Signals ________________________________________________________________ 15 4 Wire Connection _____________________________________________________________ 15 6 Wire Connection _____________________________________________________________ 15 RS-232 Serial ___________________________________________________________________ 16 Printer Connection _______________________________________________________________ 17 Remote Key ____________________________________________________________________ 17 Remote Display _________________________________________________________________ 17 Model 250 _____________________________________________________________________ 18 Model 250X ____________________________________________________________________ 18
Display ________________________________________________________________________ 21 Status annunciators ______________________________________________________________ 22 Setup Mode Annunciators _________________________________________________________ 22 Selections and Options ____________________________________________________________ 23 Numeric Entry __________________________________________________________________ 23 Keypad ________________________________________________________________________ 23
Full Setup ______________________________________________________________________ 27 Safe Setup _____________________________________________________________________ 27 Setup Display Prompts ____________________________________________________________ 28 Setup Lock Out _________________________________________________________________ 28
Version 2.4
Page 5
GRP (Group) ___________________________________________________________________ 29 ITM (Item) _____________________________________________________________________ 29 Build _________________________________________________________________________ 29 Option ________________________________________________________________________ 30 Cal ___________________________________________________________________________ 31 Spec __________________________________________________________________________ 32 Serial _________________________________________________________________________ 34 Set.Pts or CheckW _______________________________________________________________ 35 Clock _________________________________________________________________________ 36 Test __________________________________________________________________________ 36 Factory ________________________________________________________________________ 37 End ___________________________________________________________________________ 37 None __________________________________________________________________________ 37 Test __________________________________________________________________________ 37 Count _________________________________________________________________________ 37 Units __________________________________________________________________________ 38 Hold __________________________________________________________________________ 38 Peak Hold______________________________________________________________________ 38 Live Weight Averaging ___________________________________________________________ 38 Show Total (Accumulation) ________________________________________________________ 39
Special Functions__________________________________________________________________ 37
Zero __________________________________________________________________________ 42 Span __________________________________________________________________________ 42 DIR.ZER (Direct Zero Calibration Entry) _____________________________________________ 42 DIR.SPN (Direct Span Calibration Entry) _____________________________________________ 43 ED.LIN (Edit Linearization Points) __________________________________________________ 43 CLR.LIN (Clear Linearization) _____________________________________________________ 44
Chapter 8: Communications_____________________________________________ 50
Automatic Weight Output ___________________________________________________________ 51
Printing __________________________________________________________________________ 52
Page 6
Using GSE View Software ________________________________________________________ 53 Protocol Summary _______________________________________________________________ 54 Example: Reading a Weight or Setpoint Value ________________________________________ 55 Example: Remote Keypress _______________________________________________________ 57 Example: Reading a Value in Hexadecimal ___________________________________________ 58 Example: Setting a Value _________________________________________________________ 61 Error Codes (Registers) ___________________________________________________________ 62 Error Messages _________________________________________________________________ 63 Weighing Errors _________________________________________________________________ 63 Setup and Calibration Errors _______________________________________________________ 63 Diagnostic Errors ________________________________________________________________ 64
Appendix _____________________________________________________________ 65
Dimensional Drawings ______________________________________________________________ 65 Setup Menu Quick Reference ________________________________________________________ 69
Index _________________________________________________________________________ 71
Version 2.4
Page 7
Page 8
Chapter 1: Introduction
The Model 250 series of indicators use the latest Sigma-Delta A/D technology to ensure fast and accurate weight readings. The setup and calibration of the Model 250 are stored digitally within non-volatile memory. All versions of the 250 provide a method of easily connecting an RS-232 cable for temporary communication for software upgrades or link to the GSE View software. Refer to page 53 for more information.
Version 2.4
Page 9
Model 250X
The Model 250X enclosure is rated as IP69K for severe washdown applications. The Model 250X may be operated from 9.6V, 12V or 24V batteries or a DC power source from 12V to 24V. This version of the Model 250 is only available in wall/desktop.
MANUAL SET
Technical Reference Manual - Contains detailed information on calibration and setup. This manual is intended for use by Scale Technicians who are installing the indicator. Operator Manual Used for day to day operation of the indicator. Quick Start Manual - Intended for Scale Technicians who are familiar with the indicator and simply need a quick reference for menu options and connection diagrams, etc.
DOCUMENT CONVENTIONS
The following document conventions (typographical) are used throughout this reference manual. Bold Text [KEY] <Key>
Bold text denotes words and phrases to note. A keypress while in the WEIGH MODE A keypress while in the SETUP MODE. These keypresses are designated by the display annunciators Note: In the Specifications section the < symbol means less than and the > symbol means greater than. This symbol denotes one space (used in Automatic Weight Output on page 51) Items marked with indicate that the setting is available only in Full Setup and is trade critical. When trade critical settings are changed the calibration counter will be incremented.
Page 10
SPECIFICATIONS
250
Resolution Zero Cancellation Span Adjustment Stability / Drift Excitation A/D Type A/D Conversion Rate Operation Environment Setup and Calibration Digital Filter Zero Range AC DC
250X
Up to 30,000 divisions, minimum of 0.25 V/division, 20 updates/second +/- 2.0 mV/V 0.1 mV/V to 3.0 mV/V full scale Zero: < 0.1 V/C (+ 8ppm of deadload max) 5 volts for up to 4 x 350 or 8 x 700 ohm load cells (4 or 6 wire plus shield). Maximum load cell resistance is 1000 ohms 24 bit Sigma Delta with +/- 8,388,608 internal counts 20Hz with FIR filtering > 80db Temperature: 14 to 122F Temperature: 14 to 122F Temperature: 14 to 122F ambient ambient ambient Humidity: <92% nonHumidity: <90% nonHumidity: 100% condensing condensing Storage: -4 to 122F ambient Storage: -4 to 122F ambient Storage: -4 to 122F ambient IP69K IP55 when panel mounted IP65 Full digital with visual prompting 0.1 to 4.0 seconds Adjustable from 2% to 20% of full capacity
Electrical
Wall transformer: 110/240 VAC to 12 VDC @ 0.5mA 12 to 24 VDC (2.5 VA max) 4 x AA batteries (Alkaline or rechargeable NiMH, NiCad) ABS, Silicon Rubber, Nylon, Acrylic (no halogen used) Desk, panel, wall or column Desk: 6.5 (164mm) x 3.8 (97 mm) x 3.2 (81mm) Panel: 6.5 (164mm) x 3.0 (77 mm) x 1.16 (29.5 mm) Wall transformer: 110/240 VAC to 12 VDC @ 0.5mA 12 to 24 VDC (2.5 VA max) 4 x AA batteries (Alkaline or rechargeable NiMH, NiCad) 100 240 VAC 50/60 MHz 12 VDC rechargeable NiMH (optional) battery pack.
Enclosure
Material Mounting Dimensions (Refer to page 65) Stainless Steel face plate panel 7.8 (199mm) x 4.4 (112 mm) x 1.8 (44.5 mm) Food grade plastic with stainless steel mounting bracket Desk, wall, column 9.8 (249.mm) x 6.4 (163 mm) x 4.7 (119mm)
Features
Display Keypad GSE Link RS-232 Communication Setpoints Function Key Battery Backed Time/Date Linearization Approvals Shipping Weight LED backlit LCD with six 0.8 digits with 8 annunciators including units 6 key silicon rubber 6 key membrane 6-key capacitive Infra-red connector for optional GSE-Link PC cable (to RS-232 PC port) RS-232 automatic transmit, network or printer outputs. Transmission rate: 2400, 4800, or 9600 baud 2 isolated transistor drive 2 isolated transistor drive 3 isolated high side drive outputs outputs (300mA total at outputs (300mA total at (400mA each at 12 24 VDC) 50VDC) 50VDC) Unit switching, counting, manual hold, peak hold, live weight and accumulation Battery life 10 years minimum Five point linearity correction FCC, CE, NTEP FCC, CE, NTEP 1.0 lb 5.0 lb 10 point linearity correction CE, NTEP 4.4 lb
Version 2.4
Page 11
Page 12
Chapter 2: Installation
The following steps are required to set up the indicator. Inspect indicator to ensure it is in good condition. Use connection diagrams to wire up load cell, power and auxiliary cables as required. Use the template provided for hole locations. Connect power to indicator and press [ON/OFF] key to turn the Model 250 ON. Refer to the Setup Parameter section on page 29 for information on configuring and calibrating the Model 250. To turn the Model 250 OFF press and hold [ON/OFF] key for three seconds (until display blanks).
GENERAL WARNINGS
Do not subject the Model 250 indicator to shock, excessive vibration or extremes of temperature (before or after installation). Inputs are protected against electrical interference, but excessive levels of electro-magnetic radiation and RFI may affect the accuracy and stability. The indicator should be installed away from any sources of excessive electrical noise. The load cell cable is particularly sensitive to electrical noise and should be located well away from any power or switching circuits. For full EMC or for RFI immunity, termination of cable shields and correct grounding of the indicator is essential.
ELECTRICAL SAFETY
For your protection all electrical hardware must be rated for environmental conditions of use. Make sure all AC powered equipment is installed near an easily accessible power socket outlet. To avoid the possibility of electric shock or damage to the Model 250, always switch OFF or isolate the indicator from the power supply before maintenance is carried out.
CLEANING
To clean and maintain the Model 250, never use harsh abrasive cleaners or solvents. Wipe the indicator with a soft cloth slightly dampened with warm soapy water. For the stainless steel models, use a cleaner specifically for stainless steel.
Version 2.4
Page 13
CABLE CONNECTIONS
All cable connections are made to the rear of the Model 250 using screwless terminals. The Model 250SS and 250X indicators will need to have the rear cover removed to access the connection terminals. The Model 250SS uses screwless terminals where the Model 250X uses screw in terminal connectors. Wires must be stripped of insulation by at least 3/8 (10mm). It is not necessary to tin the ends of the wire with solder or to add crimp ferrules to the wires. However, these techniques are also compatible with the terminals and may ultimately make for a neater job. To install wires in the Model 250 and Model 250SS, depress the lever located beside the terminal required and push wire into the hole. Release the lever and pull gently on the wire to ensure it is secure in the terminal. Refer to the corresponding drawing below. To install wires in the Model 250X, remove the load cell connector and loosen the screws. Insert the wires in the designated position of the connector and tighten the screws. Gently pull on the wire to ensure it is secure in the connector. Refer to the corresponding drawing below. NOTE: For wash down and severe wash down environments, be sure to avoid splicing communication cables, load cell cables or home run cables. Splicing cables may cause water wicking of the cable and jeopardize the integrity of stainless steel and IP69K enclosures. Make sure all cables are sealed on both ends to prevent water damage.
Page 14
4 WIRE CONNECTION
The minimum connectivity requirements are the connection of four wires (i.e. Excitation + and along with Signal + and -). Internally the indicator has a precision analog switch that can be used to connect the Sense + and - lines directly to the Excitation + and - lines. Any addition to the load cell manufacturer's cable length using 4-wire connection, is only recommended for short cable runs. Where long additions to cable lengths are needed, a 6 wire extension is required. The BUILD:CABLE parameter must be set to 4 to allow for 4-wire connection. Refer to CABLE (4-Wire or 6-Wire) page 30.
6 WIRE CONNECTION
The excitation and signal lines are connected the same as for a 4-wire installation. The extra two wires should be connected to the Sense + and Sense terminals. The BUILD:CABLE parameter must be set to 6 (the default) to allow for 6-wire connection. Refer to CABLE (4-Wire or 6-Wire) page 30.
Version 2.4
Page 15
RS-232 SERIAL
Connection between the indicator and computer DB9 or DB25 serial port (RXD, TXD, GND)
Page 16
PRINTER CONNECTION
Connection between the indicator and printer DB25 serial port (RXD/TXD, GND and DTR))
REMOTE KEY
The indicator requires a voltage free contact between TXD and RXD to enable the remote key input and the SPEC:REM.FN parameter must be set. Refer to page 34 for configuration details. NOTE: The remote input will not function when in setup or when using the Opto-Link.
WARNING
Do not connect voltage to the remote key input switch or connection terminals. Connection of any active circuitry may damage the indicator.
REMOTE DISPLAY
The remote display documentation should be referred to for connection details. Connect TXD to RXD and GND to GND on the remote display.
Version 2.4
Page 17
MODEL 250X
The output drivers for the instrument are high side drive and are capable of driving up to 400mA each. This configuration allows for the direct connection of the instrument outputs to most types of PLC. The voltage applied to the instruments positive power supply terminal (+) appears on the output lines (Over, Pass and Under) when the outputs are active. The outputs can then be connected directly to PLC inputs. To drive external loads (eg. relays), connect the output line directly to the positive side of the relay coil. Connect the negative side of the relay coil to COMMON. It is recommended that flyback diodes or transient suppressors be fitted across inductive loads such as relay coils to limit switching noise.
Page 18
WARNING
The GSE-LINK head contains a strong magnet and care should be taken with it's proximity to electronic media (e.g. credit cards, floppy disks, etc.) and/or other electronic instrumentation.
Version 2.4
Page 19
CONNECTING SHIELDS
To obtain full EMC or for RFI immunity, cable shields MUST be connected to the ground lug on the rear of the indicator.
Figure 5 illustrates an example of possible connections. Also shown are the connecting cables restrained using cable ties fastened around the cable strain relief anchors.
Figure 5: Cable Shield Connection Care should be taken when connecting shields to maximize EMC or RFI immunity and minimize ground loops and cross-talk (interference) between indicators. For full EMC or for RFI immunity, termination of the cable shields at the ground (earth) lug is very important. The ground lug of the indicator must be separately connected to ground potential via a reliable link. The Model 250 should only be connected to ground via a single reliable link to avoid ground loops. Where each indicator is separately grounded, interconnecting cable shields should be connected at one end only.
Page 20
Chapter 3: Configuration
BASIC OPERATION
The Model 250 provides a bright backlit LCD display and simplified menu system for easy setup and use.
DISPLAY
The display is a six-digit LCD display. Figure 6 shows the main elements of the front panel and display. The Model 250 has various main display sections for the visual output of weight information. Each display section is described below.
Weight Display
The weight display indicates the weight readings, setup information, errors and warnings.
Units Display
The units display shows the units of the weight reading as either grams (g), kilograms (kg), pounds (lb), tons (t), none ( ) or ounces (o) (250X only). If the Model 250 is set up for counting, the units display will show pieces (p).
Version 2.4
Page 21
STATUS ANNUNCIATORS
Refer to the following table for details. Table 1: Status Annunciators
Symbol
Name ZERO
NET MOTION
Description
Visible when the gross reading is within of a division of true zero. Visible when the displayed reading represents NET weight. Visible when the displayed reading is not stable. Visible when the setpoint weight is over the setpoint target. Red annunciator Visible when the setpoint weight is under the setpoint target. Yellow annunciator Visible when the setpoint weight is in between the high and low limits. Green annunciator Visible when the displayed weight is within the zero 'dead' band setting. (The zero band symbol shows near the top right corner of the display.) Visible when the displayed reading is held. Visible when battery voltage is too low and batteries need replacing or recharging. (The low battery symbol shows in the top right corner of the display.)
or or
or
When in Setup Mode the editing annunciators are shown to identify the function of the front panel keys (i.e. GRP, ITM, SEL, EDT and OK). For more information refer to Setup Mode Annunciators discussed below.
Page 22
Annunciator
GRP ITM SEL EDT OK
Key
ZERO TARE GROSS/NET PRINT f (FUNCTION)
Description
Steps through the list of Groups. Steps through the list of Items. Press this key to accept changes and return to the menus. (Also refer to the OK description below.) Moves the editing cursor in some editing modes. Steps through the available options when editing a particular item. Press this key to accept changes and return to the menus. (Also refer to the ITM description above.)
NUMERIC ENTRY
A numeric entry box allows the input of a number. When entering a number, the display will show digits with the currently selected digit flashing. The <SEL> key is pressed to select a digit to change. When the digit is selected the <EDT> key is pressed to change the digit from 0 through 9. The left most digit can also be changed to a dash (-) to enter a negative number. The <OK> key is pressed to accept the number that has been entered and return to the menu item. Upper and lower limits are placed on some entries and an entry outside this range will cause the indicator to display dashes (i.e. - - - - - -). Example: Setting Scale Capacity Press [ZERO] (<GRP>) repeatedly to display the BUILD group. Press [TARE] (<ITM>) repeatedly to display the CAP item. Press [GROSS/NET] (<SEL>) to select CAP and display the current setting (e.g. 0000.00lb). The currently chosen digit will be flashing. Press [GROSS/NET] (<SEL>) to advance to the next digit. When the digit to edit is flashing, press [PRINT] (<EDT>) repeatedly to cycle from 0 through 9. When the new digit to be set is flashing either press <SEL> to move to the next digit to edit and repeat the previous step; or press [ f ]<OK> or [TARE] <ITM> to accept all of the displayed digits (including the flashing digit) and re-display the menu item name.
KEYPAD
All Models of the 250 utilize the same six keys in the same configuration. The material used for each model keypad is different however. Each key performs a different function in the weigh mode and setup mode. Refer to this section for a list of the keys and their function. The standard Model 250 and panel mount have six silicon rubber keys. The Model 250SS comes with a six key membrane keypad. The Model 250X offers a capacitive keypad with large easy to use keys. The Model 250X keypad has been calibrated to work with human fingers. NOTE: Do not use blunt objects to press the keys. This will cause damage to the keypad. Key
POWER ZERO TARE GROSS/NET PRINT f
Version 2.4
Page 23
Stability Considerations
Once the [ZERO], [TARE] or [PRINT] key is pressed the Model 250 waits for a stable valid reading before performing the associated operation. If the weight readings remain unstable or invalid due to a diagnostic error for longer than 10 seconds, the operation is cancelled and the STABLE ERROR message is displayed. To improve the stability of the weight reading, increase the filtering or decrease the motion detection value. Refer to FILTER (Reading Average) and MOTION (Motion Detection) on page 30 for more information.
POWER Key
The [POWER] key is used to turn the Model 250 on and off. To initially turn the indicator on, press and hold the [POWER] key. The display will show the following: Display segments will light and then clear. Software Version (i.e. V1.0). Calibration Counter (i.e. C.00010). The current weight will then display. To turn the indicator off, press and hold the [POWER] key for three seconds. The Model 250 will turn OFF followed by the three-second countdown. When using batteries the backlight will automatically turn off to conserve battery power after a short period of inactivity. A short press of the [POWER] key will turn the backlight on again. Refer to B.LIGHT (Backlight Operation) page 33 for more information. Note: The [POWER] key can be locked to prevent the indicator from being turned off through the front keypad. Refer to KEY.LOC (Front Panel Key Locking) page 33 for more information.
ZERO Key
When an empty scale has drifted away from a true zero reading, this key is used to perform a zero adjustment on the scale display. The zero adjustment is stored when power is removed and is re-used when next powered up. The amount of weight that may be cancelled by the [ZERO] key is limited to the setting at parameter Z.RANGE (Allowable Zero Operating Range). Refer to page 31 for more information. When the indicator is set to Industrial mode a long press of the [ZERO] key will remove any stored zero adjustment. Refer to Chapter 6: Approvals on page 45 for more information on modes.
TARE Key
This key is used to temporarily set the scale to zero (such as cancelling the weight of a carton before performing a filling operation). The display will show the Net weight and the NET annunciator will be lit.
Page 24
The weight tared is deducted from the allowable range of the scale, reducing the maximum weight that can be displayed. The tare adjustment is stored when power is removed and is re-used when next powered up. The [TARE] key can operate in all modes (i.e. Industrial, OIML and NTEP). Refer to Chapter 6: Approvals on page 45 for more information on modes.
GROSS/NET Key
This key toggles the weight display between the Gross weight and the Net weight if a Tare weight has previously been acquired using the [TARE] key.
GSE-LINK Activation
This feature is used to temporarily connect a PC to the Model 250 for calibration and setup purposes. A long press of the [GROSS/NET] key from the weigh mode will toggle the GSE-LINK infrared communications On/Off. When the GSE-LINK has been enabled the following will occur: The indicator briefly displays the prompt opto-L. The editing annunciators (i.e. GRP, ITM, etc.) will flash for up to five minutes while the indicator searches for activity. During this time period, the Model 250 also disables the RS232 communications. If the indicator is successful in locating activity, the editing annunciators will continue to flash during the entire period of communications. If the indicator fails to locate activity, the GSE-LINK will be disabled and the editing annunciators will stop flashing. The Model 250 will also revert back to the normal RS-232 communications (i.e. the SERIAL:TYPE setting will be re-activated).
PRINT Key
If a printer or computer has been attached to the Model 250 and the manual print function has been selected, the [PRINT] key will trigger an output of the current weight reading. The PRINT prompt is displayed while waiting for the printer to accept data. If the printer is offline the PRINT prompt will remain for a maximum of 10 seconds before the operation is cancelled. Each weight printed is automatically added to an internal total weight. A long press of the [PRINT] key will print the accumulated total. The total weight is then cleared automatically.
FUNCTION Key
When leaving the factory, the [ f ] key has no function programmed. The primary function of this key can be selected from a number of different choices including peakhold, counting, etc. Refer to page 37 for details of the available functions. Each primary function has an associated overlay sticker that should be applied underneath the [ f ] key to label the function of the key. A long press of the [ f ] key may be used for certain functions depending on the primary function of the key.
Version 2.4
Page 25
Page 26
Chapter 4: Setup
ACCESSING THE SETUP MODE
The Model 250 has two levels of access code protection to provide a security lock on accessing Setup via the keypad. For further details of menu items available in each setup mode, refer to the Setup Menu Quick Reference on page 69. Full Setup Access code Safe Setup Access code The Full Setup access code can also be used to access Safe Setup
FULL SETUP
The Full Setup method provides access to all functions in Setup, including legal for trade and calibration sensitive settings. Changes in Full Setup mode may result in the calibration counter being incremented. Changes to access codes and restoring the factory default settings can only be accessed in Full Setup mode. These items will however not increment the calibration counter. If an attempt is made to enter Full Setup using the incorrect access code, the indicator will respond with the message ENTRY DENIED. Setting an access code for Full Setup restricts any access to Full Setup. Refer to FULL.PC (Full Security Access code for Digital Setup) page 33. To access the Full Setup mode, first ensure the indicator is on. For the 250 and 250SS press and hold both the [POWER] and [f ] keys together for two seconds. For the 250X press and hold the [POWER] and [GROSS/NET] keys together for two seconds.
WARNING
All menus and items will be enabled in Full Setup. Care should be taken to avoid inadvertently altering the Build or Calibration settings.
SAFE SETUP
The Safe Setup method restricts access to the trade critical settings. Changes made in this mode will not increment the calibration or setup counter. If an attempt is made to enter Safe Setup mode using the incorrect access code, or if an attempt is made to alter a trade critical setting while in Safe Setup, the indicator will respond with the message ENTRY DENIED. Setting an access code for Safe Setup restricts access to Safe Setup functions. Refer to SAFE.PC (Safe Security Access code for Digital Setup) page 33. To access Safe Setup, first ensure the indicator is on. For the 250 and 250SS press and hold both the [POWER] and [ZERO] keys together for two seconds. For the 250X press and hold the [POWER] and [TARE] keys together for two seconds.
Version 2.4
Page 27
Method 1: Press and hold both the [POWER] and [f ] keys together for two seconds. Method 2: Press and hold both the [POWER] and [ZERO] keys together for two seconds. Method 3: Press the [ZERO] (GRP) key until End is displayed and then press [TARE] (ITM).
The indicator will beep and then display the following: SAVING Software Version (i.e. V1.0). Calibration Counter (i.e. C.00010) Setup Mode Counter (i.e. F.00003) if NTEP is enabled at OPTION:USE The current weight will then display. Warning: If the power is interrupted while in setup (i.e. by disconnecting the power cable or pressing the [POWER] key), unsaved settings will be lost.
Page 28
GRP (GROUP)
Setup is divided into a series of Groups. Each group has a distinctive group title. All options in any one group have related functions. The <GRP> (ZERO) key can be used to cycle through the available groups.
ITM (ITEM)
Each group is divided into individual Items. Each item represents a parameter that can be changed. Pressing the <ITM> (TARE) key will enter the displayed group, allowing access to the items within the group. The <ITM> (TARE) key can be used to cycle through the available items. The <SEL> (GROSS/NET) key is then used to edit the item.
SETUP PARAMETERS
The following sections describe the setup parameters of each of the Groups and Items in Setup.
BUILD
Settings within this Group are used to configure the indicator to suit the current application. It is important to fully set the options within this group before calibration is attempted. Later changes to items within this group may invalidate the current calibration data.
Version 2.4
Page 29
OPTION
Items within this Group are used to configure the operating parameters of the scale.
Page 30
CAL
Items within this group perform various calibration routines. For detailed scale calibration procedures refer to Chapter 5: Calibration on page 41. Certain items in the Scale Build can affect the calibration of the scale. Always check that these sections are correctly configured to suit the current application before attempting to calibrate the scale.
Version 2.4
Page 31
SPEC
Settings within this group control features including access codes, key locking, key functions and display settings.
Page 32
WARNING
Be sure to remember your password. Reflashing the unit will not clear out the password. The indicator will need to be sent back to the factory in order to regain access to the setup mode.
Version 2.4
Page 33
SERIAL
Settings within this Group determine the serial and printing outputs. Refer to Automatic Weight Output and Printing on page 51 for more information on serial configuration.
Page 34
SET.PTS OR CHECKW
Settings within this group configure the setpoint (250 & 250SS) or checkweighing (250X) system. Refer to Setpoints page 49 for a detailed explanation of setpoints.
Version 2.4
Page 35
CLOCK
Items within this group set date and time related functions.
TEST
Items within this Group allow access to the testing routines for the Model 250. With these routines the scale base output can be monitored and the inputs and outputs can be tested.
Page 36
FACTORY
This is the factory adjustment menu.
END
The parameter can be used for exiting the setup mode. Refer to page 28 for details.
SPECIAL FUNCTIONS
The Model 250 has a special function key on the front panel. The function of this key can be configured to any of the key functions detailed below. Refer to KEY.FN (Key Functions) on page 33 for configuration details. When leaving the factory, the [ f ] key has no primary function pre-programmed. Each primary function has an associated overlay sticker (supplied) that should be applied to the function key to label the function. Ensure the keypad is clean and dry before affixing the sticker. Refer to Cleaning page 13 for more information.
NONE
When set to NONE the special function key is not used during normal operation. This is the default setting.
TEST
The [TEST] key is used to test the functionality of the indicators display. Press the [TEST] key to clear the display and show all segments of the display. The display will be cleared again before returning to the weigh mode.
COUNT
The [COUNT] key is used for counting pieces. Pressing the [COUNT] key will toggle between the count and weight display. The annunciator for the count mode will be distinguished by a p for pieces.
Version 2.4
Page 37
To perform a new sample: 1. 2. 3. 4. Access the weigh mode. If a container is used, tare off the container weight by pressing the [TARE]. Place the sample pieces on the scale. Press and hold the [COUNT] key for two seconds. The default sample size will be displayed. If this is the first sample with the indicator SPEC will be displayed. Use the <SEL> and <EDT> keys to enter a new sample size. The <EDT> key will enter a number and the <SEL> key will select to the next number. Be sure to use leading zeroes if sample size is less than 1000. 5. Press <OK> and the selected sample size will be stored. 6. If printing is enabled the sample quantity and weight will be printed.
UNITS
Use the [UNITS] key to switch the displayed reading between lb and kg or other selected unit. This will affect the displayed and printed weight but will not affect entered data.
HOLD
Use the [HOLD] key to hold a particular weight on the display regardless of a variance of weight. The hold function will be designated on the display with the annunciator. Press the [HOLD] key again to unlock the hold function.
PEAK HOLD
Use the [PEAK] key to hold the largest absolute weight. Either positive or negative weight is stored in the peak value (e.g. -30 is larger than 25). The Hold annunciator is active when the display is showing the held weight. To clear the stored peak weight, hold the [PEAK] key for 2 seconds.
Page 38
Version 2.4
Page 39
Page 40
Chapter 5: Calibration
The calibration results are stored in permanent memory for use each time the Model 250 is powered up. Note: Some of the parameters in the setup mode can affect calibration. OPTION settings MUST be configured before calibration is attempted. The BUILD and
To perform a calibration, when in Full Setup select the CAL Group using the <GRP> key. The calibration routine will automatically prevent the indicator from being calibrated into an application outside of its specification. If an attempt is made to calibrate outside of the permitted range, an error message will be displayed and the calibration will be aborted. Refer to Error Messages on page 63. Note: It should not be assumed that just because the Model 250 has successfully calibrated a scale, that the scale is correct for Legal for Trade use. Refer to Chapter 6 for details on approval specifics. Always check the BUILD parameter against the approval specification.
NOTE: Calibration points (Zero, Span and Linearization) must be spaced by at least 2% of full scale from each other.
Version 2.4
Page 41
ZERO
Press the <SEL> key to start. The display will show the current weight. Remove all weight from the scale. Press <SEL>, <EDT> or <OK> to execute a Zero Calibration. The display will show Z.in.P to indicate that zeroing is in progress. When the process is complete the weight will be displayed to allow the zero to be checked. Press the <ITM> key to exit the Zero routine or press <SEL>, <EDT> or <OK> to repeat the operation.
SPAN
Press <SEL> or <OK> to start. The display will show the current weight. Add the calibration test weight to the scale. The minimum acceptable span calibration weight is 2% of the scale range. A weight this small may limit calibration accuracy. The closer the test weight is to full scale the better the accuracy. Press <SEL> or <OK> to show the calibration weight value. calibration weight using the <SEL> and <EDT> keys. Change this to the correct
Press <ITM> or <OK> to trigger the Span Calibration routine. The display will show S.in P to indicate that spanning is in progress. When the process is complete the weight will be displayed to allow the new weight reading to be checked. When the Span Calibration is complete, press the <ITM> key to exit the Span routine or press <SEL>, <EDT> or <OK> to re-edit the calibration weight and repeat the operation.
Page 42
USING LINEARIZATION
This section provides instructions on the use of the linearization. approximate the weight output to a non-linear scale. Linearization is used to The chart below shows a non-linear characteristic for the load cell output. From the chart, it can be seen that the trace with no linearization applied may not be linear at approximately 750 kg. By applying one or more linearization points, more accurate weight readings can be achieved. To perform a linearization, a calibration of the zero and full scale span points must have been performed. Both the zero and full scale calibration points are used in the linearization of the scale base. These two points are assumed to be accurately set and thus have no linearization error.
3500 3000 2500 Weight (Kg) 2000 1500 1000 500 0 0 2 Zero P oint 4 Linearis ation P oint 6 8 10
Actu a l L o a d C e ll C h a ra cte ris tic N o L in e a ris a tio n L in e a ris a tio n Ap p lie d
S pan P oint
Loa d Ce ll O utput (m V )
A maximum of five linearization points can be set independently between zero and full scale. Unused or unwanted points may also be cleared. The maximum correction that can be applied using a linearization point is + / - 2%.
Version 2.4
Page 43
Press <OK> to change the linearization point selected or press <ITM> to exit without making any changes. After pressing <OK>, the current weight reading is displayed. Add the calibration test weight to the scale. The closer the test weight is to the point of maximum error in linearity the more effective the correction will be. Press <OK> to enter a corrected weight value for this point or <ITM> to exit without making changes. Use the <SEL> and <EDT> keys to enter the correct value of the calibration weight being used. Press the <OK> key to trigger the linearization routine. The display will show L.in P to show that linearization is in progress. When the process is complete the display will show the weight to allow the new weight reading to be checked before returning to the menus. Press <ITM> to leave the routine or <OK> to repeat the operation.
RESET CALIBRATION
To reset the calibration to factory condition, access the CAL parameter. Refer to FAC.CAL (Restore Default Factory Calibration) page 32. The menu choice will only be available when the Full Setup mode is accessed.
Page 44
Chapter 6: Approvals
REGULAR AND LEGAL FOR TRADE MODES
The Model 250 may be operated in Industrial, OIML or NTEP mode. (Note: For NSC requirements, use the OIML setting.) The industrial mode operates without an reference to trade regulations. The OIML and NTEP modes restrict certain aspects of indicator operation to ensure compliance with the respective trade certified standards. The following table lists the operation differences for each of these modes. Condition
Underload Overload Tare Test Modes
Industrial
105% of full scale 105% of full scale No restrictions on Tare Unlimited time allowed
NTEP
1% or 2% of full scale depending on zero range setting 105% of full scale Tare values must be > 0 and rounded to the nearest graduation Limited to five seconds
OIML
1% or 2% of full scale depending on zero range setting Full scale +9 divisions Tare values must be > 0 Limited to five seconds
The Model 250 factory default parameter setup does not ensure compliance with legal-for-trade installations as mandated by local weights and measures authorities. This chapter contains information on NTEP and OIML regulations, sealing and audit trails, and other requirements. Since legal-for-trade requirements may vary, you must ensure that the Model 250 is installed in accordance with all local regulations. For details on parameter setup refer to the USE (Scale Use) section page 30.
NTEP REQUIREMENTS
The National Type Evaluation Program (NTEP) is a widely accepted weights and measures standard in the United States, with most states abiding by some or all of the NTEP requirements. A complete list of these regulations is available in the "Handbook 44" publication distributed by the National Institute of Standards and Technology (NIST). For more information, call (301) 9753058, or visit http://www.nist.gov.
In order to configure the Model 250 to comply with NTEP requirements, the OPTION:USE parameter must be set to NTEP. This ensures the following: Serial data will not be received while in the Setup Mode. Received alpha characters will not be displayed. Numeric tare entries cannot be received through the serial port. Pressing [TARE] with a gross weight of zero (0) or keying in a tare value of zero (0) will not automatically switch to the net mode. Negative tare values are not accepted.
Version 2.4
Page 45
Tare rounding is enforced. When the tare value is zero, the net mode is not selectable.
Where applicable, enabling the NTEP parameter will override the current setting of other parameters.
OIML REQUIREMENTS
The International Organization of Legal Metrology is an intergovernmental body which harmonizes the national metrology regulations of its world wide members. A list of regulation publications may be obtained from the Bureau International de Mtrologie Lgale (BIML) in Paris, France. In order to configure the Model 250 to comply with OIML requirements, the OPTION:USE parameter must be set to OIML. This ensures the following: An overload condition will result when the gross weight exceeds nine graduations over the full scale capacity. Full scale capacity is always referenced from the last zero calibration reference, not the last zero acquired by pressing [ZERO].
AUDIT TRAIL
Audit trail parameters are designated on the Model 250 at power up. When in NTEP mode the two individual parameters will be displayed (calibration counter and setup configuration) to indicate changes to trade specific parameters. When the OIML mode is enabled, only the calibration counter will be displayed. Follow the instructions below to view the audit trail counter(s). 1. Press and hold the power key for 3 seconds to turn off the indicator. 2. Press the power key to turn the indicator back on. 3. The event counters will be displayed during the power up sequence for approximately two seconds. The calibration counter will be identified by C.xxxxx (example: C.00005). The setup configuration counter will be identified by F.xxxxx (example: F.00005).
PHYSICAL SEAL
The most common sealing method is a lead-wire seal. The Model 250 provides an easy means of applying this type of seal as shown in Figure 8. The calibration mode is enabled or disabled by pushing a button marked setup located on the rear of the indicator. To seal the Model 250, the R.ENTRY parameter must be set to ON. Refer to page 31 for more details. Follow the steps below to access and set the R.ENTRY parameter and place the lead seal.
Page 46
1. 2. 3. 4. 5. 6. 7. 8. 9.
Press [POWER] + [ f ] to access the setup mode. Press <GRP> until the OPTION menu is displayed. Press <ITM> to the R.ENTRY parameter. Press <SEL> to view the parameter. Press <EDT> to turn the setup button ON if set to OFF. Press <OK> to accept change Press [POWER] + [ f ] to save the change and exit the setup mode. Press the setup button to access the setup mode and make changes or calibrate. Press [POWER] + [ f ] to save the change and exit the setup mode. The setup mode cannot be accessed without pressing the Setup button again. 10. Place the physical seal plastic cover over the load cell and comm port connectors, and then install the drill head screw. 11. Thread the wire security seal through the indentations in the plastic cover and the drill head screw.
Figure 8: Model 250 and Model 250SS Panel Mount Physical Seal
Version 2.4
Page 47
Page 48
Chapter 7: Setpoints
The Model 250 is capable of working with two internal setpoints. The status of these setpoints is displayed on the LCD. Each setpoint is associated with a physical output driver which allows control of an external light or buzzer.
SETPOINT CONNECTION
Refer to Setpoint Output Connection page 18 for the method of connection of the external output drivers.
Version 2.4
Page 49
Page 50
CHAPTER 8: COMMUNICATIONS
The Model 250 offers one bi-directional RS-232 communication port or an optional fiberoptic link (GSE-Link) for seamless communication. This provides a number of serial output options allowing communications with external devices such as printers, computers, PLCs or remote displays. For wiring connections and pinouts, refer to Communication Port Connection page 16. The serial available functions include: Automatic Weight Output (two formats available) Network Communications Master / Slave Printing
Refer to SERIAL (Serial Communications Options) page 34 for communication parameter setup details. Be advised that the calibration counter is incremented when the calibration related steps are accessed via RS-232 port or the GSE-LINK.
Description
<STX> <SIGN> <WEIGHT(7)> <STATUS> <ETX> <STX> <SIGN> <WEIGHT(7)> <S1> <S2> <S3> <S4> <UNITS(3)> <ETX>
Description
Start of transmission character (ASCII 02) End of transmission character (ASCII 03) The sign of the weight reading (space for positive, dash (-) for negative) A seven character string containing the current weight including the decimal point. If there is no decimal point, then the first character is a space. Leading zero blanking applies Provides information on the weight reading. The characters G/N/U/O/M/E represent Gross / Net / Underload / Overload / Motion / Error, respectively A three character string, the first character being a space, followed by the actual units (e.g. ^lb, ^kg or ^^t). If the weight reading is not stable, the unit string is sent as ^^^ Displays G/N/U/O/E representing Gross / Net / Underload / Overload / Error, respectively Displays M/^ representing Motion / Stable, respectively Displays Z/^ representing center of Zero / Non-Zero, respectively Displays - representing single range
Version 2.4
Page 51
PRINTING
The Model 250 comes standard with a single fixed printing format that is as follows:
000048 12/05/2006 15:10 121.4 lb G 5.3 lb T 116.1 lb N
The first line contains a six digit sequential ID number that is automatically incremented with each printing, up to a maximum of 999999. The first line also contains the current date and time. The second line shows the displayed reading, units and G for Gross or N for Net. If counting is active, the printout has a third line for the counting information (p for pieces) as follows: Counting:
000168 06/03/2007 12:58 8.27 lb G 10 p G 0.00 lb T 8.27 lb N
Whenever a new sample is taken, the indicator prints the sample in the following format:
SAMPLE: 10 p = 6.17 lb
In addition to the printout, the indicator accumulates the weight, prints the accumulated total and number of accumulations. A long press of the [PRINT] key will print a ticket and clear these totals. The six digit sequential ID number is not cleared. Accumulation:
GROSS ITEMS: 3 TOTAL GROSS: 29.68 lb NET ITEMS: 2 TOTAL NET: 18.82 lb
If the count mode is enabled each of the individual printouts would have a third line as discussed above and the TOTAL would have an additional QTY line as follows:
Page 52
ADVANCED COMMUNICATION
USING GSE VIEW SOFTWARE
Software is available for the Model 250 which makes parameter setup simple. Simply connect the GSE Link cable or an RS-232 cable to the indicator and computer. In order to use GSE View with the Model 250X, the software revision must be 1.5 or higher. Click on the GSE View 200 icon to launch the GSE View Software. A screen similar to the one below should be displayed.
Description
Shows the display of the connected Model 250 View setup parameters and make changes Calibrate the Model 250 Plot out weight readings of 4 separate digital inputs Use as a communication terminal. Send commands to the indicator
Version 2.4
Page 53
It is possible to view the parameter settings, make changes to the parameters, and save the settings as a new file.
Figure 10: Settings Screen Under the Settings tab are all of the Groups of the parameter setup. Click on that particular Group to view individual items (parameters).
PROTOCOL SUMMARY
NOTE: The values used with the Read Final and Write Final commands are hexadecimal. The Model 250 indicator contains a number of registers. The indicator is configured by reading and writing the information stored in these registers. Information such as gross weight is obtained by reading the register. The Model 250 has registers of different types to hold weights, menu options, etc.
20
mm
rrrr
optional_value
Commands
05
Value to write (in hexadecimal)
Gross Weight Net Weight Target High Target Low Keypress System Status System Error
Page 54
Registers
Program
GSE View Communication Plus Other terminal program
20
05
0027
Commands
05
No value needed
Gross Weight Net Weight Target High Target Low Keypress System Status System Error
The indicator will respond with a message similar to the following: 9F050027: 4.23 lb N A Net weight of 4.23 lbs was returned
Registers
Code
0001 0002 0003 0004 0005 0009 0010 0011 0012 0013 0014
Example
V1.0 GSE 2006 250 V1.20 3123456 0100005B6D3F
Version 2.4
Page 55
Register Description
Counts each new reading Indicator status Indicator error status Absolute signal in mV/V Displayed weight reading Gross or Net weight Gross weight Net weight Tare weight Peak held weight Manual held weight Total weight reading Livestock weight Index for first register data Index for second register data Index for third register data Actual full access code Actual safe access code Weight used for calibration Calibrated zero mV/V Calibrated span weight Calibrated span mV/V Weight at linearization points >00100000 means no point Fullscale setting Resolution setting Decimal point position Units selection 4/6 wire operation High resolution (x10) mode Trade or industrial operation Filtering setting Motion setting Zero range setting Zero tracking setting Initial zero operation Zero band setting Type of serial output Baud rate Parity and stop bits Network address Reset printer sequence command Full date/time string Date format selection Date Month Year Hour Minute
Code
0020 0021 0022 0023 0024 0025 0026 0027 0028 0029 002A 002B 002C 0042 0043 0044 00D0 00D1 0100 0111 0112 0113 0114 0118 0121 0122 0128 0129 012A 012B 0130 0131 0132 0133 0134 0135 0136 0140 0141 0142 0143 0144 0150 0151 0152 0153 0154 0155 0156
Example
00169196 00000200 00000000 00006187 000005DC 000005DC 000005DC 000001F4 000003E8 000009C4 00000000 000009C4 0000001F4 00000001 00000002 00000005 000004D2 000009C4 000005DC FFFFFFFF 000009C4 000061A8 08000001 00000BB8 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 0000001F
Page 56
Register Description
Second Key locking Special function setting Auto power off setting Backlight options Remote key function setting Setpoint target high Setpoint target low Counting sample quantity
Code
0157 0160 0161 0162 0163 0164 0171 0172 0180
Example
00000015 00000000 00000000 00000000 00000000 00000000 000007D0 000003E8 0000000A
Program
GSE View Communication Plus Other terminal program
20
12
0008
8003
Commands
05
Keypress value to write (in hexadecimal)
Gross Weight Net Weight Target High Target Low Keypress System Status System Error
The indicator will respond with a message similar to the following: 9F120008:0000 Where the value after the : indicates no key was pressed. See the Codes for Keypad Related Registers chart for a reference of related codes.
Version 2.4
Registers
Page 57
Page 58
20
11
0026
Commands
05
Value to write (in hexadecimal)
Gross Weight Net Weight Target High Target Low Keypress System Status System Error
The indicator will respond with a message similar to the following: 9F110026:00000127 Where the value after the ":" (00000127) is the hexadecimal value of the gross weight without any decimal places or units. In decimal, this value is 295. Register Description
Version of the protocol Copyright message Model of indicator Software version Indicator serial number Contents of LCD display memory Save settings command Start of menu definitions OIML calibration counter NTEP calibration counter NTEP configuration counter Counts each new reading Indicator status Indicator error status Absolute signal in mV/V Displayed weight reading Gross or Net weight Gross weight Net weight Tare weight Peak held weight Manual held weight Total weight reading Livestock weight Index for first register data Index for second register data Index for third register data
Registers
Code
0001 0002 0003 0004 0005 0009 0010 0011 0012 0013 0014 0020 0021 0022 0023 0024 0025 0026 0027 0028 0029 002A 002B 002C 0042 0043 0044
Example
V1.0 GSE 2006 250 V1.20 3123456 0100005B6D3F 00000033 C.00011 F.00010 00169196 00000200 00000000 00006187 000005DC 000005DC 000005DC 000001F4 000003E8 000009C4 00000000 000009C4 0000001F4 00000001 00000002 00000005
Version 2.4
Page 59
Register Description
Actual full access code Actual safe access code Weight used for calibration Calibrated zero mV/V Calibrated span weight Calibrated span mV/V Weight at linearization points >00100000 means no point Fullscale setting Resolution setting Decimal point position Units selection 4/6 wire operation High resolution (x10) mode Trade or industrial operation Filtering setting Motion setting Zero range setting Zero tracking setting Initial zero operation Zero band setting Type of serial output Baud rate Parity and stop bits Network address Reset printer sequence command Full date/time string Date format selection Date Month Year Hour Minute Second Key locking Special function setting Auto power off setting Backlight options Remote key function setting Setpoint target high Setpoint target low Counting sample quantity
Code
00D0 00D1 0100 0111 0112 0113 0114 0118 0121 0122 0128 0129 012A 012B 0130 0131 0132 0133 0134 0135 0136 0140 0141 0142 0143 0144 0150 0151 0152 0153 0154 0155 0156 0157 0160 0161 0162 0163 0164 0171 0172 0180
Example
000004D2 000009C4 000005DC FFFFFFFF 000009C4 000061A8 08000001 00000BB8 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 0000001F 10/16/2005 10:32 00000000 00000011 0000000A 000007D3 0000000A 00000020 00000015 00000000 00000000 00000000 00000000 00000000 000007D0 000003E8 0000000A
Page 60
Program
GSE View Communication Plus Other terminal program
20
12
0171
1F4
Commands
05
Set high target value of 500 decimal (in hexadecimal)
Gross Weight Net Weight Target High Target Low Keypress System Status System Error
The indicator will respond with a message similar to the following: 81120171:0000 Where the "0000" value after the ":" indicates no errors. See the Error Codes (Registers) chart for a reference of related codes.
Version 2.4
Registers
Page 61
Code
0000H 8000H 4000H 2000H 1000H 0800H 0400H 0200H 0100H 0080H 0040H 0020H 0010H 0008H 0004H 0002H 0001H
Description
No errors occurred Always set to show an error Unknown error internal use only Feature not implemented on this device Action not permitted for the current access rights Value less than RangeMin Value greater than RangeMax Value not compatible with data type Operation not defined Write operation failed Parameter not valid Cannot modify register values while SETUP menus are active Reserved for future use Reserved for future use Reserved for future use Reserved for future use Internal data error
Page 62
Chapter 9: Troubleshooting
ERROR MESSAGES
A number of error messages may be displayed to warn of operation outside of the acceptable limits. These messages are described below. Short messages (XXXXX) will appear as a single message on the display. Longer messages (XXXXX) (YYYYY) will appear on the display in two parts, first the (XXXXX) part, then the (YYYYY) part.
WEIGHING ERRORS
These messages show status messages or errors that may occur during normal weighing operation (weigh mode). Error Message
(U - - - - -)
Description
The weight is below the minimum allowable weight reading. The weight is above the maximum allowable weight reading.
Solution
Increase the weight or decrease the minimum allowable weight reading. Warning - overloading may damage mechanical scale elements. Remove weight from load receiving element. Check the condition of load cell connections. Check for damaged load cell. The operation of the [ZERO] key is limited in the setup during installation. The indicator cannot be zeroed at this weight. Increase the Zero Range (Z.RANGE) or use the [TARE] key instead. Wait for motion to cease and try the operation again once the scale is stable.
(O - - - - -)
The weight reading is beyond the limit set for Zero operation. Scale motion has prevented a [ZERO] or [TARE] operation from occurring on command.
Solution
Access Full Setup to edit the parameter. Refer to Setup Lock-Out on page 28 for more information. Turn the Model 250 off. When the indicator is turned back on, enter the correct access code to access setup. Entered linearization point must be between zero and full scale. Refer to Using Linearization on page 43 for more information. Re-enter the calibration point. Points must be spaced by at least 2% of full scale from each other. Check the resolution (count-by) and capacity settings. Check the resolution (count-by) and capacity settings.
Version 2.4
Page 63
Error Message
(LO) (SPAN) (SPAN) (HI) (ZERO) (LO) (ZERO) (HIGH)
Description
Solution
Incorrect span weight entered (must be between zero and full scale). Scale wiring incorrect. Wrong load cell capacity (too large). Wrong or no calibration weight added to scale. Incorrect span weight entered (must be between zero and full scale). Scale wiring incorrect. Load cell capacity too small for application. Scale wiring incorrect Remove all weight from scale. Scale wiring incorrect.
The load cell signal range (span) is too small for the current settings. The load cell signal range (span) is too large for these settings. An attempt has been made to calibrate zero below -2mV/V. An attempt has been made to calibrate zero above +2mV/V.
DIAGNOSTIC ERRORS
The Model 250 will display specific codes for faults or out of tolerance conditions. These errors will be displayed as an E type error message. Error Message
(E0001) (E0002) (E0020) (E0100) (E0200) (E0300) (E0400) (E0800) (E2000) (E4000)
Description
The power supply voltage is too low. The power supply voltage is too high. The number of graduations has been set too low or too high. The digital setup information has been lost. The calibration information has been lost. All setup information has been lost The factory information has been lost. The EEPROM memory storage chip has failed ADC Out of Range Error. This may be caused from a broken load cell cable. The battery backed RAM data has lost data.
Solution
Check supply Check scale / cables Enter the correct graduation size at BUILD:DP and BUILD:RES enter setup recalibrate Enter setup and calibrate Return for Service Return for Service Check BUILD:CABLE setting. Check load cell cable, wiring, etc. Re-enter setup
Page 64
Appendix
DIMENSIONAL DRAWINGS
Model 250 (Plastic Enclosure) Panel Mount Version Dimensions shown in mm 3D View Front View
Side View
Back View
Version 2.4
Page 65
Model 250 (Plastic Enclosure) Desk Mount with Battery Compartment Version Dimensions shown in mm 3D View Front View
Side View
Back View
Page 66
Model 250 (Stainless Steel Enclosure) Panel Mount Version Dimensions shown in mm
Version 2.4
Page 67
Model 250X IP69K version Dimensions shown in mm Wall Mount Indicator with Swivel Bracket
3D View Front View
Side View
Bottom View
Page 68
Page
29 29 30 30 30 30 30 30 30 31 31 31 31 31 31 32 32 32 32 32 32 32 33 33 33 33 33 34 34 34 35 35 35 35 35 35 36 36 36 36 36 36 36 36 37 37 37
BUILD
OPTION
CAL
SPEC
SERIAL
SET.PTS
CLOCK
Version 2.4
Page 69
Page 70
INDEX
4
4-Wire Connection, 7 NTEP, 47 OIML, 47 CAP, 21, 62 Capacity, 21 Cleaning, 5 CLOCK, 28, 62 CLR.LIN, 24, 36, 62 Connecting shields, 12 Consecutive number reset, 27 Count, 29 Count by resolution, 22, 62
6
6-Wire Connection, 7
A
Access codes, 19 full setup, 25 safe setup, 24 Accessing the setup mode, 19 full setup, 19 safe setup, 19 Accumulation, 31 Address, 27 Advanced communication, 45 Annunciators setup mode, 14 status, 14 Audit trail, 38 AUT.OFF, 25, 62 Automatic weight output, 43
D
Date. See Time & Date Date format, 28, 62 DAY, 28, 62 DC power supply, 6 Decimal point, 21, 62 DEFLT, 29, 62 Diagnostic errors, 56 Dimensional drawings, 57 DIR.SPN, 24, 34, 35, 62 DIR.ZER, 24, 34, 62 Document conventions, 2 DP, 21, 62
B
B.LIGHT, 25, 62 Basic operation, 13 BAT.VLT, 26 BAUD, 26, 48, 52, 62 BITS, 27, 48, 52, 62 BUILD, 15, 21, 62
E
ED.LIN, 24, 35, 62 Edit linearization points, 24, 35, 62 Electrical safety, 5 END, 29, 62 Error messages, 55 Exit the setup mode, 20, 29, 62
C
CABLE, 7, 22, 62, See Cable connections, 6, 7 CAL, 33, 62 Calibration, 23, 33, 62 clear linearization, 36 clear linearization points, 24 direct span, 24, 35 direct zero, 24, 34 edit linearization points, 24, 35 errors, 55 linearization, 35 mV/V entry, 34 restore, 24, 36 span, 24, 34 with test weights, 33 zero, 23, 34 Calibration counter, 16, 20, 38
F
FAC.CAL, 24, 62 Factory default, 24, 28, 29 FACTRY, 28, 62 FILTER, 22, 62 Firmware update, 18 Force output, 62 Force outputs, 28 FORMAT, 28, 62 FRC.OUT, 28, 62 Front panel key locking, 25, 62 FULL.PC, 25, 62 Function key, 17
Version 2.4
Page 71
G
General warnings, 5 Gross/net GSE-link activation, 17 Gross/net key, 15, 17 gse view communication, 45 Group, 21 Groups and items, 21 GRP, 21 GSE view software, 45
M
Manual set, 2 Master serial output, 45 MINUTE, 28, 62 MONTH, 28, 62 MOTION, 14, 22, 62 Motion detection, 14, 22
N
Net annunciator, 14 NONE, 29 Numeric entry, 15
H
HI.RES, 22, 62 High resolution, 22, 62 HOLD, 30 annunciator, 14 HOUR, 28, 62
O
Operator manual, 2 Option GSE-link, 11 OPTION, 22, 62
I
Industrial mode, 37 INIT.Z, 23, 62 Initial zero, 62 Initial zero, 23 Item, 21 ITM, 21
P
Panel mount, 5 PEAK HOLD, 30 Physical seal, 38 Power key, 16 Print key, 17 Printer connection, 9 Printing, 44 count, 44 master serial output, 45 sample, 44 sequential ID number, 44 type, 26 Protocol summary, 46
K
Key functions, 25, 62 Key locking front panel, 25, 62 KEY.FN, 25, 62 KEY.LOC, 25, 62 Keypad, 15 function key, 17 Gross/net key, 17 power key, 16 print key, 17 related register codes, 50 tare key, 16 weigh mode, 16 zero key, 16
Q
Quick reference setup menu, 62 Quick start manual, 2
R
Reading average, 22, 62 Reflash, 18 REM.FN, 26, 62 Remote display, 9 Remote function, 26, 62 Remote input, 9 Remote key, 9 RES, 22, 62 Reset printed consecutive number, 27 Resolution, 22, 62 high, 22, 62 Restore factory settings, 24, 28, 62 Restory factory calibration, 62 RS-232 serial, 8 RST.CON, 27, 62
L
L.in P, 24, 36 Legal for trade, 37 audit trail, 38 NTEP, 37 OIML, 38 physical seal, 38 Live weight averaging, 30 LIVE.WT, 25, 30 Load cell connection, 7 4-wire, 7 6-wire, 7 Load cell signals, 7 Low battery annunciator, 14
Page 72
S
S.in P, 24 Safe setup access code, 19, 24 SAFE.PC, 24, 62 Scale setting capacity, 15 SCALE, 28, 62 Scale base test display, 28, 62 Scale Build, 23 Sealing audit trail, 38 physical seal, 38 Selections and options, 15 Serial baud rate, 26 master output, 45 output type, 26 parity, 27 RS-232, 8 SERIAL, 26, 62 Set hour, 28, 62 Set minute, 62 Set minutes, 28 Set month, 28, 62 Set year, 28 SET.PTS, 27, 62 Setpoint connection, 41 high target, 27 low target, 27 operation, 41 output connection, 10 Setpoint settings, 27, 62 Setpoints weight source, 27 Setup access button, 23 Setup display prompts, 20 Setup errors, 55 Setup lock out, 20 Setup menu quick reference, 62 Setup mode annunciators, 14 Setup parameters, 21 SHOW.T, 31 Sigma-Delta A/D converter, 1 SPAN, 24, 34, 62 SPEC, 24, 62 Special functions, 29 accumulation, 31 count, 29 hold, 30 live weight averaging, 30 peak hold, 30 test, 29 units, 30 Specifications, 2 Stability considerations, 16 Status annunciators, 14
T
Tare legal for trade, 37 Tare key, 16 setup mode, 15 weigh mode, 15 TARG.HI, 62 TARG.LO, 62 Target high, 27, 62 low, 27, 62 Target annunciators, 14 Targets, 41 Technical reference manual, 2 Test scale base, 28 TEST, 28, 29, 62 Time & Date, 3 TYPE, 26, 62
U
Units, 30 UNITS, 22, 30, 62 Units display, 13 USE, 22, 62
W
Weighing errors, 55 Weight display, 13
Y
YEAR, 28, 62
Z
Z.BAND, 23, 62 Z.in P, 34 Z.RANGE, 23, 62 Z.TRAC, 23, 62 Zero initial, 23 ZERO, 34, 62 Zero annunciator, 14 Zero band, 23, 62 Zero band annunciator, 14 Zero calibration, 23 Zero key, 16 setup mode, 15 weigh mode, 15 Zero operating range, 23 Zero tracking, 23
Version 2.4
Page 73