MFA160 :: ROAL Living Energy
MFA160 :: ROAL Living Energy
MFA160 :: ROAL Living Energy
DESCRIPTION
The MFA160-USr-2 is a series of high efficiency, small form factor AC-DC power supplies for use in Class II medical applications. Offering up to 160 W of regulated DC power from an open-frame 2 x 4 x 1 standard form factor, the MFA160-USr-2 series make easier its integration into space constrained systems. Thanks to its creepage and clearance greater than 8 mm, it can be operated over 4000 m of altitude without de-rating. By converting energy at 91% typical efficiency, the MFA160-USr-2 series generate less heat facilitating thermal management. The series is available in four different output voltages at 5, 12, 24 and 48 V, and is equipped with an auxiliary low power 12 V. The MFA160-USr-2 range comply with the 3rd edition of the IEC 60601-1and ANSI/AAMI ES60601-1 safety standards for Class II medical equipment. The series meets the EN 55022 EMC limits of Class B for conducted and Class A for radiated emissions as well as the IEC/EN 61000-3 and IEC/EN 61000-4 EMC standards.
KEY FEATURES
Universal input voltage range 160 W active PFC power supply Low leakage current (<100 A) Very small form factor (2 x 4 x 1) in High efficiency (91% typical) 5, 12, 24 and 48V standard output variants Class II isolation Over temperature protection Output over-voltage protection Over current, and short circuit protections Auxiliary fan +12 V output 4000 m altitude operation IEC/EN 60601-1 3rd ed., ANSI/AAMI ES60601-1 RoHS-6 compliant (EU directive 2011/65/EU)
V2 [V] 12 12 12 12
5 12 24 48
The combined output power of V1 and V2 must not exceed 70 W for the 5V model or 100 W for the 12, 24 and 48 V when natural convection cooled. and 100 W and 160 W respectively when forced air cooled at 500 LFM, up to 50 C ambient. Above 50 C output de-rating applies (see details on the output specifications). 2 Peak-to-Peak measured at 20 MHz Bandwidth.
INPUT SPECIFICATIONS
Specification AC Input Voltage Input Frequency DC Input Voltage Input Current Inrush Current Fusing Efficiency No load Power Consumption Power Factor Harmonic Current Fluctuations and Flicker Leakage Current Test Conditions / Notes PS starts and operates at 90 VAC at all load conditions Min. 90 47 170 Nominal 100/240 50/60 2.5 85 90 91 2.5 2.3 Max. 264 63 370 2.3 Units VAC Hz VDC A A A % W
RMS at 90 VAC, maximum load 230 VAC, cold start, no damage 2.5 A, Time Lag, 250 V on L and N 5V, 12, 24, 48V, 115 VAC 12, 24 , 48V, 230 VAC 115 VAC 230 VAC At full rated load, 115 VAC, 60 Hz 0.99 230 VAC, 50 Hz 0.89 Complies with EN-61000-3-2 Class D at 230 VAC 50 Hz. Complies with EN-61000-3-3 at nominal voltages and full load. 264 VAC, 60 Hz, normal condition -
100
OUTPUT SPECIFICATIONS
Specification V1 Set Point Accuracy V1 Output Power Rating Test Conditions / Notes 5V, natural convection 5V, at 500 LFM forced air 12, 24, 48V, natural convection 12, 24, 48V, at 500 LFM forced air All models (15% accuracy) All models VAC: nominal voltages V1 Load: 0 100% rated V2 Load: 0 0.5 A VAC: 90 264 VRMS V1: 0 100% load (V2 at 50% load) V2: 0 0.5 A load (V1 at 50% load) VAC: 90 264 VRMS 50% load changes at 0.1 A/s Recovery to regulation band within 1 ms All models, Peak-to-peak, 20 MHz BW. 470 pF ceramic and 22 F tantalum caps at the load (resistive). 90<VIN <264, any load conditions. V1 in regulation after AC is applied Min. 10.2 0.2 16 0 Nom. 1 12 10 20 1.2 Max. 70 100 100 160 13.8 0.5 5 1 5 1 15 0.1 5 1 20 1000 Units % W V A %V1 %V1 %V2 %V1 %V2 %V1 %V1 %V1 ms ms %V1 %V2 ms A mV/C
Load-Line Cross Regulation V1 Line Regulation V1 Transient Response (Voltage Deviation) V1 Ripple and Noise
Start-up Rise Time Start-up Delay Turn-on Overshoot Hold-up Time Minimum Load Temperature Drift
PROTECTION FEATURES
Specification Input Under Voltage Lockout Input Fuse Over Current Short Circuit Over Voltage Over Temperature I-O isolation Isolation V1/V2 Creepage and Clearance Test Conditions / Notes No damage, auto recovery Time Lag 2.5 A, 250 V on L and N Hiccup mode, auto-recovery Hiccup mode, auto-recovery Shut down, latch off mode Shut-down, auto-recovery Reinforced Min. 60 110 110 4000 100 8 Nominal 75 Max. 150 130 Units VAC %I1MAX %VNOM VAC VDC mm
ENVIRONMENTAL SPECIFICATIONS
Specification Operating Temperature Storage Temperature Range Humidity Operating Altitude Shock Vibration MTBF Cooling Test Conditions / Notes No de-rating up to 50 C Linearly de-rate above 50 C RH, Non-condensing Operating Non-operating Min -20 -20 Nominal Max 70 80 90 95 4000 Units C C % % m
Operating: 10 g, 11 ms, half sine, one shock input in each axes Operating, sinusoidal: 0.5 g peak-to-peak, 10-300 Hz, 3 axes 75% Full Load, Nominal VAC, 25 C ambient MIL-HDBK-217-E-1 Natural convection 10 Forced air cooling 500
LFM
Agency Files References E304543-A5-UL NO69561 (CB test certificate) No. P12215594 (Nemko)
Forced air cooling: Air flow direction, longitudinal or transverse, must be coplanar to the PCB no matter its orientation.
Connector AC Input Connector J1 J1 Mating Connector Output Connector J2 J2 Mating Connector Auxiliary Connector J3 J3 Mating Connector
Manufacturer and Part Number Molex 26-60-4030 or equivalent Molex 09-93-0300 (Crimp Terminal Housing) Molex 08-50-0105 (Crimp Terminal, 18-24 AWG) Molex 26-60-4080 or equivalent Molex 09-91-0800 (Crimp Terminal Housing) Molex 08-50-0105 (Crimp Terminal, 18-24 AWG) Molex 22-23-2041 or equivalent Molex 22-01-2047 (Crimp Terminal Housing) Molex 08-50-0113 (Crimp Terminal, 22-24 AWG)
1 4: V1 RTN; 5 8: +V1
1, 2: GND; 3, 4: +V2
Roal Electronics, S.p.A. may change product specifications and accordingly the information presented in this document. Customers are responsible for their products and applications using Roal Electronics, S.p.A. products. Roal Electronics, S.p.A. assumes no liability from the use of its products outside of specifications. No license is granted to any intellectual property rights by this document. ROAL ELECTRONICS, S.P.A. DISCLAIMS ALL REPRESENTATIONS AND WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, IMPLIED WARRANTIES OF NONINFRINGEMENT, MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Europe, Middle East and Africa ROAL Electronics S.p.A Via Jesina 56/A 60022 Castelfidardo (AN) Italy Tel:+39 071 721461 Fax:+ 39 071 72146 480
North America ROAL Electronics USA, Inc. 701, Main St. Suite 405 Stroudsburg, PA18360 Phone: + 1 570 421 5750 Fax: +1 570 421 5687 Rev 03 _ October 22, 2013
www.roallivingenergy.com