A-Scan Biometry
Rhonda G Waldron, MMSc, COMT, CRA, ROUB, RDMS, Senior Associate in Ophthalmology, Diagnostic Echographer, Senior Technologist, Department of Ophthalmology, Emory University School of Medicine Thomas M Aaberg, Jr, MD, Clinical Assistant Professor, Department of Surgery, Michigan State University College of Human Medicine; Consulting Staff, Department of Ophthalmology, Associated Retinal Consultants Updated: Dec 12, 2008
Ultrasound Principles
Sound is defined as a vibratory disturbance within a solid or liquid that travels in a wave pattern. When the sound frequency is between 20 hertz (Hz) and 20,000 Hz, the sound is audible to the human ear. To be considered ultrasound, sound waves must have a frequency of greater than 20,000 Hz (20 KHz), rendering them too high in frequency to be audible to the human ear.[1 ]In ophthalmology, most A-scan and B-scan ultrasound probes use a frequency of approximately 10 million Hz (10 MHz) that is predesigned by the manufacturer. This extremely high frequency allows for not only restricted depth of penetration of the sound into the body but also excellent resolution of small structures. This meets unique needs, because, at times, the probe is placed directly on the organ to be examined, and its structures are quite small, requiring excellent resolution. The velocity of sound is determined completely by the density of the medium through which it passes. Sound travels faster through solids than through liquids, an important principle to understand because the eye is composed of both. In A-scan biometry, the sound travels through the solid cornea, the liquid aqueous, the solid lens, the liquid vitreous, the solid retina, choroid, sclera, and then orbital tissue; therefore, it continually changes velocity. The known sound velocity through the cornea and the lens (average lens velocity for the cataract age group, ie, approximately 50-65 y) is 1641 meters/second (m/s), and the velocity through the aqueous and vitreous is 1532 m/s. The average sound velocity through the phakic eye is 1550 m/s. The sound velocity through the aphakic eye is 1532 m/s, and the velocity through the pseudophakic eye is 1532 m/s plus the correction factor for the intraocular lens (IOL) material.[2 ]The cornea is not routinely factored in because of its thinness. If one were to consider 1641 m/s at about 0.5 mm, only 0.04 mm would need to be added to the total eye length, which in no way alters the IOL calculation. In A-scan biometry, one thin, parallel sound beam is emitted from the probe tip at its given frequency of approximately 10 MHz, with an echo bouncing back into the probe tip as the sound beam strikes each interface. An interface is the junction between any two media of different densities and velocities, which, in the eye, include the anterior corneal surface, the aqueous/anterior lens surface, the posterior lens capsule/anterior vitreous, the posterior vitreous/retinal surface, and the choroid/anterior scleral surface. The echoes received back into the probe from each of these interfaces are converted by the biometer to spikes arising from baseline. The greater the difference in the two media at each interface, the stronger the echo and the higher the spike.[2 ]If the difference at an interface is not great, the echo is weak and the displayed spike is short (eg, vitreous floaters, posterior vitreous detachments). No echoes are produced if the sound travels through media of identical densities and velocities, eg, young, normal vitreous or the nucleus of a noncataractous lens, in which the A-scan display goes down to baseline.
1 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
High quality contact A-scan of the phakic eye. Note the 5 high-amplitude spikes and the steeply rising retinal spike, as well as the good resolution of the separate retinal and scleral spikes.
In the case of a cataractous lens, multiple spikes occur within the central lens area as the sound beam strikes the differing densities within the lens nucleus. This spike height, or amplitude, is therefore what gives the information on which to base the quality of the measurements. In fact, the "A" in A-scan is from the word "amplitude." Spike height is not only affected by the difference in density as it travels through the eye but also by the angle of incidence, which is determined by the probe orientation along the visual axis. If the probe is held in a perpendicular manner to the visual axis, it is in the proper position to receive the echoes back into the probe tip so they can be converted to spikes. Because sound waves can be reflected and refracted the same as light rays, if the probe is held in a nonperpendicular manner, part of the echo is diverted at an angle away from the probe tip, and, therefore, is not received by the machine.[2 ]The more nonperpendicular the angle of incidence, the weaker the signal and the shorter the spike amplitude.
2 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
When the sound beam incidence is perpendicular to the visual axis (upper image), most returning echoes are received back into the probe tip to be interpreted on the display as high-amplitude spikes. When the sound beam incidence is nonperpendicular to the visual axis (lower image), part of the returning echo is reflected away from the probe tip, with only a portion received by the probe. As a result, the spikes will be compromised.
The shape and smoothness of each interface also affects spike quality. An irregularity in the surface of an interface causes reflection and refraction of the returning sound waves away from the probe tip and, therefore, weaker echoes.[2 ]That is why it is important to know whether macular pathology is present that could adversely affect spike quality. A perfect high, steeply rising retinal spike may be impossible when macular pathology is present (eg, macular edema, macular degeneration, epiretinal membranes, posterior staphylomas).
3 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
If the macular surface is smooth (upper image) more of the echoes are received back into the probe to be displayed as high-amplitude spikes. If the macular surface is convex (center image), as with macular edema or pigment epithelial detachments, part of the echoes is reflected away from the probe tip. If the macular surface is irregular (lower image), as in macular degeneration or epiretinal membranes, reflection of the echoes away from the probe tip also will occur.
In addition, sound is absorbed by everything through which it passes before it travels on to the next interface. The greater the density of the structure it is passing through, the greater the amount of absorption. This principle explains why retinal spike quality is reduced in the case of an extremely dense cataract; the lens absorbs much of the sound and less sound actually reaches the retinal surface.
4 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
When the gain setting is too high, the resolution of the separate retinal and scleral spikes is lost, resulting in 1 thick, flattened spike.
Gates are electronic calipers on the display screen that measure between two points.[2 ]Biometers are designed so that between each pair of gates a measurement is rendered. Biometers vary in the appearance of these gates, with some units not displaying them at all. Gates should be readily visible for accurate editing of the scans, because if any one of them is aligned along an incorrect spike the entire eye length measurement will be erroneous. The biometer automatically places a gate on what it believes to be the corneal spike, the anterior lens spike, the posterior lens spike, and the retinal spike, and it is programmed to measure the distance between each pair of gates at a given velocity.
5 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Gates are electronic calipers on the display (see arrows) that between each pair render a measurement. In this 4-gate system, each of the 3 sections of the eye is measured individually at its correct velocity, then added together for the total eye length. Equipment will vary in the appearance of these gates.
Ultrasound is measured based on how long it takes the sound to travel from one point to the next at a given velocity. The formula, distance = velocity X time, is programmed into biometers to calculate the distance between each pair of gates. Then, the formula is divided by 2 because the sound also must echo back into the probe tip. By selecting the eye type in the measurement mode (phakic, aphakic, or pseudophakic), the equipment is instructed to use this distance formula with the proper velocities between each gate pair for that particular eye type.[2 ] For example, in the phakic mode, the machine has been programmed to measure distance between the first pair of gates using a velocity of 1532 m/s, the velocity through the anterior chamber. The velocity of 1641 m/s is used between the second and third gates because this is the velocity through the lens. The velocity of 1532 m/s is once again used in the formula between the third and fourth gates because this is the velocity through the vitreous cavity. The most accurate machines measure each of these 3 sections of the eye individually at the proper sound velocity and then add them together for the total eye length. If any one gate is incorrectly placed, the machine will calculate the 2 sections involved using the incorrect velocity and time, which will make the total length erroneous. Proper gate placement is on the ascending edge of each appropriate spike. If a gate has been placed incorrectly, move it to the appropriate spike before it is stored and used in the calculation. Equipment varies greatly in the ability to see and move gates; refer to the manual for each specific unit to determine if it is possible to move the gates, and, if so, the steps required to do so. If the biometer does not allow for movement of gates, scans must be repeated until they automatically align properly. When the machine is set for phakic average, only 2 gates are present, measuring the total eye at its average velocity of 1550 m/s. The 2 gates should align along the corneal surface and the retinal surface, respectively. The disadvantages to this setting are that the anterior chamber depth and the lens thickness cannot be monitored and that the use of average sound velocity is simply not as accurate. The average sound velocity of 1550 m/s is only accurate through an average length eye. In eyes that are shorter or longer than average, this method of measuring produces an innate error. When setting the measurement mode to aphakic, 2 gates will be present (on the respective corneal and retinal surfaces), and the
6 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
biometer will calculate the distance at a velocity of 1532 m/s, the correct velocity for the aqueous and vitreous. When setting the measurement mode to pseudophakic, depending on how many pseudophakic options the equipment possesses, the eye length is calculated using 1532 m/s for the aqueous and vitreous, then the correction factor for the given implant material is added. If only one pseudophakic mode option is available, it only will be accurate for polymethyl methacrylate (PMMA) IOLs. Routinely using automatic mode on most equipment increases the risk of error because every biometer will capture poor quality scans. Biometers are programmed to capture any scans with spikes that are of high amplitude within their given area. However, they often cannot determine if the spike arose steeply from baseline or if a slope or step is present in the spike origin. Manual mode is sometimes preferable, in which the examiner presses a foot switch to capture the scan when it is seen to be of high quality. Equipment varies greatly with some manufacturers only using a 4-gate system on automatic mode, which means that anterior chamber depth can be monitored only in automatic mode. If this is the case, automatic mode is preferable, but the examiner must carefully edit the scans stored by the machine.
7 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
The Holladay II IOL Consultant formula is considered by most experts to be the best no matter the axial length. The Holladay II IOL Consultant formula is also the only formula that calculates for piggyback IOL procedures (ie, when 2 IOLs are implanted, which may be necessary when the eye is so small that 1 implant does not contain enough converging power, or when a piggyback lens is inserted to correct a postoperative surprise that results from the first implant), and is most recommended for patients that have had previous refractive surgery. The Haigis formula is also proving to be accurate no matter the axial length but must be optimized first, which requires 200-300 eyes, preferably. Optimization is done in the United States by Warren Hill, MD, and in Europe by Dr. Wolfgang Haigis.
8 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Immersion scan of the phakic eye. The probe and cornea are now separate spikes since they are not in contact with each other, and the corneal spike demonstrates 2 peaks (see arrow), representing the epithelium and endothelium. If these 2 peaks are not both high, the sound beam is not aligned through the corneal vertex. Gain must be reduced enough to appreciate and resolve these 2 peaks.
Other advantages of the immersion technique are that it is a faster method than the contact technique and that it reduces technician dependency. When using the contact technique, axial lengths will vary on subsequent scans by the same technician, as well as between technicians depending on the amount of corneal compression. When using the immersion technique, as long as the spikes are of high quality, the axial lengths will not vary from one scan to the next or from one biometrist to the next. Any practice that changes from the contact technique to the immersion technique must repersonalize their IOL constants in their IOL calculations, since they are achieving true rather than slightly shortened eye length measurements. A study should be completed of at least 20 eyes with the new technique, the same surgeon, and the same IOL to determine the correct personalization of the constant. Another current method for highly accurate axial length measurements uses optical coherent light rather than ultrasound. In this method, optical coherent light passes through the visual axis and reflects back from the retinal pigment epithelium. Interferometry is then used to measure the distance from the cornea to the retina. Because this is a noncontact method, as is immersion ultrasound, its accuracy is superior to contact ultrasound biometry. In fact, when it can be used it is as accurate as immersion ultrasound and was calibrated to immersion ultrasound in its development by Dr. Wolfgang Haigis. However, this method cannot be used in the event of significant media opacity (eg, dense cataracts or corneal or vitreal opacity) due to absorption of the light, or an inability of the patient to fixate on the target. Optical coherence cannot measure anterior chamber depth or lens thickness; however, a slit beam is in place for anterior chamber depth measurements. For highly accurate anterior chamber depth and lens thickness measurements, immersion ultrasound should be used. For those practices using optical coherence, immersion ultrasound will be necessary for patients who cannot be measured by optical coherence to ensure the same high level of accuracy.
9 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Other reasons to measure the pseudophakic eye include patients who are scheduled to undergo an IOL exchange or checking an unwanted postoperative refractive error. Measurement through the pseudophakic eye will result in multiple reverberation echoes in the vitreous cavity that tend to decrease in amplitude from left to right. The number and strength of these reverberations is dependent on the IOL material. Decreasing the gain in the pseudophakic eye is helpful, so that spike amplification of these artifacts is reduced, reinforcing correct gate placement on the retinal spike.
Reverberation artifacts in the vitreous cavity resulting from intraocular lens. The left image demonstrates the longer chain of artifact spikes from polymethyl methacrylate implants. The image on the right demonstrates the shorter chain of artifact spikes from foldable implants.
To obtain accurate measurements through the pseudophakic eye, knowledge of the implant material is essential. Most IOLs are currently made of PMMA, acrylic, or silicone. The velocity of sound through each of these materials is different because of their differing densities and if measured with the wrong modality can result in significant error. If an eye with an acrylic IOL is measured on pseudophakic PMMA mode, a 0.2 mm error will occur. If an eye with some silicone IOLs is measured on PMMA mode, a 1.2 mm error will occur. The velocity of sound through the pseudophakic eye is 1532 m/s plus the correction factor for the implant material. The velocity through PMMA is 2718 m/s, through acrylic is 2120 m/s, and through silicone is 980-1107 m/s, depending on the silicone used.[2 ] (Because acrylic and silicone lenses are foldable, they are not as dense as PMMA and, therefore, have a slower velocity.) If a biometer only has one pseudophakic setting, it will be accurate for PMMA only because it was manufactured when PMMA was the only implant material being used, and some biometers do not have upgrades for the various IOL materials. If this is the case, the way to achieve accurate measurements is to use the aphakic setting, which uses a sound velocity of 1532 m/s. Then, the examiner should manually add the correction factor for the IOL material to the results obtained on aphakic mode. The correction factor is +0.4 mm for PMMA, +0.2 mm for acrylic, and -0.4 mm to -0.8 mm for silicone, depending on the silicone velocity.[2 ] Therefore, if an eye measured 23.32 mm on aphakic mode and the IOL is made of PMMA, the correct axial length is 23.72 mm. If the IOL is acrylic, the correct axial length is 23.52 mm. If it is low-velocity silicone, the correct length is 22.52 mm. When any new implant material is produced, the correction factor can be calculated using the CT of the IOL and the sound velocity of the material at body temperature (35C), which must be supplied by the manufacturer. The formula for this calculation is the CT multiplied by 1 minus 1532 divided by the velocity of that material, or CT X (1-1532/vel).[2 ]For example, if the IOL has a CT of 0.8 mm, and the sound velocity of the material is found to be 1040 m/s, then 0.8 X (1-1532/1040) = 0.8 X -0.473 = -0.378. Therefore, the correction factor for this eye is -0.378 from the length obtained on aphakic setting. Another problem arises when the implant material is unknown. If the patient has a wallet card showing the implant used, the manufacturer may need to be called to determine implant material if the model is unfamiliar to the examiner. If the patient does not have a wallet card, contact the surgeon's office to determine the implant used. If the patient cannot recall the surgeon's name, it may be necessary to contact a family member in the case of an IOL exchange. However, the implant reverberation pattern may prove helpful because PMMA has a longer chain of reverberation echoes, followed by acrylic and then silicone.
10 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Corneal compression is demonstrated in the A-scan on the right. Note the more shallow anterior chamber depth of 2.63 mm as compared to the scan of the same eye on the left, with an anterior chamber depth of 3.20 mm, indicating 0.57 mm of corneal compression. Note also that the total eye length is shortened from 24.60 mm in the scan on the left to 24.18 mm in the scan on the right. This error would result in an unwanted postoperative refractive error of about -1.25 D.
The second most common error is misalignment, either by not obtaining perpendicularity to the macular surface or by not directing the sound beam through the visual axis. Perpendicularity to the macular surface is achieved when the retinal spike and scleral spike are of high amplitude, and the retinal spike arises steeply from baseline. No sloping of the retinal spike should be present and no jags, humps, or steps should be present on the ascending edge of that spike.
11 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Misalignment errors. The A-scan on the left demonstrates a contact scan with a sloping retinal spike. The A-scan on the right demonstrates an immersion A-scan with steps in the initial rise of the retinal spike.
If either the posterior or anterior lens spike are not of high amplitude, the sound beam could be misaligned at an angle through the lens and, therefore, not through the visual axis. The posterior lens spike may be slightly shorter than the anterior lens spike because the convex curvature of the posterior lens is steeper than the convex curvature of the anterior lens surface, allowing for reflection of the echoes away from the probe tip. Also, if a dense nuclear sclerotic cataract is present, more sound absorption could occur within the lens, causing the posterior lens spike to be shorter. In these instances, the gain can be increased to obtain better posterior lens and retinal/scleral spikes.
12 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Misalignment demonstrated by the decreased amplitude of the posterior lens spike (arrow). When either of the lens spikes is too short, the sound beam is aligned at an angle through the lens rather than through its center, and thus not aligned along the visual axis.
Misalignment along the optic nerve is an error that is easily recognized, since the scleral spike will be absent. The retinal spike will be present and of high amplitude and can even appear steeply rising, but, if the scleral spike is not as high in amplitude as the retina, the sound beam is misaligned along the nerve. No sclera is present at the optic nerve; the sound beam is passing through the nerve cord with only short amplitude echoes present, because the sound beam is striking blood vessels within the nerve cord. In the normal eye, there will generally not be a great difference in axial length when aligned along the optic nerve, but, in cases of a full optic disc, papilledema, or optic disc drusen, this will result in an erroneously short axial length measurement. In cases of optic nerve cupping, as seen in glaucomatous eyes, this error results in an erroneously long axial length measurement.
13 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Misalignment along the optic nerve. Note the missing scleral spike.
Another possible error in the contact method is a fluid meniscus between the probe tip and the cornea caused from ointment use, methylcellulose in the eye from previous examinations, or abnormally thick tear film. If either of these is suspected, rinse the eye with sterile saline prior to biometry. Extremely dense cataracts can be a challenge because of absorption of the sound beam as it passes through the lens. A higher gain setting may be necessary to achieve high-amplitude spikes from the retina and sclera. Improper gate placement also can occur easily, because a dense cataract produces multiple spikes within the lens. The posterior lens gate may erroneously align along one of the echoes within the lens nucleus, resulting in an erroneously thin lens thickness and erroneously long vitreous length, which results in an error in the total length of the eye. In this case, manually realign the gate to the correct posterior lens spike, and if the equipment does not allow for manual gate placement, repeat scans until the gates automatically align properly. Posterior staphylomas are among the greatest biometry challenges. These occur primarily in high myopes, where the globes are so elongated and thin that the posterior uvea bulges outward into the sclera, most commonly in the posterior pole. This causes the macula to be sloped in configuration, which in turn causes reflection of the sound beam away from the probe tip and a poor retinal spike. It is impossible to obtain perpendicularity to a macular surface that is sloped; thus, it is impossible to obtain a proper retinal spike. Also because of the sloped surface, the measurements will be not only long but extremely variable. Patients must be alerted that because their eye is misshapen, they have a higher risk of the postoperative result not being as accurate as a patient with a normally shaped, round globe. In these cases, a B-scan examination is necessary, with a horizontal macular scan performed and the axial length measured from the B-scan. The proper B-scan probe position for this measurement is to have the patient in primary gaze with the B-scan probe face (using a generous amount of gel-type tear solution) centered on the corneal vertex and the probe marker aimed nasally. (The probe marker is either a line or a dot on one side of the probe, near the probe face.) When this probe position is achieved, the B-scan display will demonstrate the epithelial and endothelial corneal echoes centered to the left, the posterior lens surface just to the right, and the optic nerve void just above the center to the far right. The macula will lie centered on the right, about 4.5 mm below the center of the optic disc. Simply place calipers on the vertex of the epithelial corneal echo and on the macula to measure the axial length at average sound velocity of 1550 m/s. Compare this axial length
14 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
measurement to the various biometry measurements, and use the measurement that has the most comparable vitreous length in the IOL calculation, preferably within 0.1 mm.
High myope with a long eye. Note the poor quality of the retinal spike on the A-scan. The axial length from the anterior corneal surface to the central macula was measured on the B-scan (left image), and compared to the variable A-scan measurements. The A-scan that was within 0.1 mm of the B-scan measurement was used in the intraocular lens calculation.
Optical coherence biometry has been shown to be beneficial in the case of the highly myopic globe because it measures to the fixation point of the patient and the lack of perpendicularity is not prohibitive. But again, optical methods are not always usable if the patient has a dense cataract or other media opacity or the inability to fixate, so ultrasound techniques still need to be used in these situations. Known macular retinal detachments are yet another challenging situation. In retina practices, the physician may decide to remove the cataract while the patient is undergoing retinal detachment repair and will need accurate IOL calculations. In these cases, the retinal spike will appear further to the left in the vitreous cavity, depending on the elevation of the macula. In these instances, the retinal gate should be moved from the detached retina to the next more posterior spike, since the retina should lie back into this position once repaired. The examiner should inquire if the surgeon plans to place a scleral buckle around the globe to repair the detachment, and if so manually add another 0.5-1.0 mm to the total eye length to account for lengthening of the globe by the buckle postoperatively.
15 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
If the patient's previous medical records and measurements are unobtainable, the contact lens method of determining corneal curvature can be used. This method can also be used for comparison even when other methods are utilized. The formula for this method is as follows: K = Base Curve + (Difference in Refractive Error From Without Contact Lens to With Contact Lens) First, refract the patient without a contact lens and note the spherical equivalent. Then, place a hard contact lens (PMMA) of known base curve and plano power on the eye, and refract the patient again. The difference in refractive error is algebraically added to the known base curve of the contact lens. For example, if a patient refracts to a -1.5 D without a contact lens, and a plano contact lens with a base curve of 40.0 D is placed on the eye and the patient now refracts to a -4.25 D, that is a difference in refraction of 3.75 D in the minus direction. The formula K = 40.0 D + (-3.75) = 36.25 D is used as the K reading in the IOL calculation. Care must be taken to note the direction of the change in refraction. For example, if another patient refracts to a -2.0 D without a contact lens and then +0.25 D with the same 40.0 D contact lens, the difference in refractive error is 2.25 D in the plus direction, so K = 40.0 D + (+2.25) = 42.25 D to be used in the calculation. Regardless of the base curve of the contact lens used for the refraction, this formula results in the same K power because the difference in base curve of the contact lens causes the refraction with it in place to vary. Many experts suggest using a consensus approach for determining corneal curvature after refractive surgery, rather than one method alone.[4 ] Other methods currently being used include the corneal power correction equation by Shammas.[3 ]This equation involves measuring the K's now and then averaging them into one number. So, the corrected K = 1.14 (average K) - 6.8. Another method, referred to as the topography method, as published by Wang et al, involves measuring the eye with topography, multiplying that number by 1.114, and then subtracting the correction factor of 6.1. Gaining confidence is the method from Wake Forest University, sometimes known as the corneal bypass method. In this method, the original K's prior to the refractive surgery are entered into the calculation formula along with today's axial length measurement. In order to achieve emmetropia, however, the refractive target entered into the calculation is the amount of correction from the refractive surgery. In their study, all patients had better outcomes than using the standard clinical history method and the implants were calculated to be closer to the perfect IOL power.[5 ] Another method faring well in recent studies is the Masket Formula.[6 ]This formula involves an adjustment in the final IOL power rather than an adjustment of the keratometry readings in the refractive surgery patient. In Dr. Masket's study, keratometry and biometry were performed with the IOL Master as with standard cataract patients, and then the following formula was used to adjust the calculated power from those measurements: IOL adjustment = LSE X (-.326) + 0.101, where LSE = the spherical equivalent of change after laser vision correction. If the patient had myopic correction, the LSE should be a negative number, and if hyperopic correction, a positive number. In other words, if a myopic refractive patient has standard calculations of +16.0 D, and LSE was 6.0 D of myopia, -6.0 X (-0.326) + 0.101 = +2.057, which should be added to the calculated +16.0 D for a final IOL power of +18.0 D. If a patient had hyperopic refractive surgery and the calculation yielded a power of +22.0 D, and LSE was 3.0 D of hyperopia, +3.0 X (-0.326) + 0.101 = -0.877, which is added to the calculated implant power of +22.0 D, for a final IOL power of +21.0 D. On a final note, using a formula where the patient's actual preoperative anterior chamber depth is entered into the calculation is best (ie, the optimized Haigis formula, the Holladay II IOL Consultant formula). Formulas that only take into account axial length and keratometry falsely assume that these patients with such flat K's after myopic correction must also have shallow anterior chamber depths, resulting in incorrect IOL powers. Also, the history method has been shown to be not as accurate in the postradial keratotomy patient, since the postoperative refraction never completely stabilizes.
Velocity Conversion
The velocity conversion equation is helpful in many biometric circumstances, including cases of silicone oil in the vitreous, using an incorrect velocity setting on the biometer or measuring an eye filled with silicone oil.[2 ]The equation is as follows: Velocity (correct)/Velocity (measured) X Apparent Length = True Length In the event of an incorrect eye type setting, this equation is quite simple to use and will preclude the need for the patient to return for repeat measurements. For example, an aphakic eye was measured incorrectly with the phakic average setting. The correct velocity for this eye is 1532 m/s. The velocity used was 1550 m/s. If the eye length obtained was 24.1 mm on the phakic average setting, then 1532/1550 X 24.1 = 23.82 mm = true eye length.
16 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
For eyes that have silicone oil in the vitreous cavity, this formula is used to determine the true vitreous length. Silicone oil is used surgically to replace the vitreous in some cases of recurrent retinal detachment and macular hole. The oil is removed several months later, but while in the eye, it causes a cataract, often requiring the removal of the cataract at the time of oil removal. The velocity conversion equation is necessary because the velocity through silicone oil is only 980 m/s, much slower than the 1532 m/s the biometer uses in determining vitreous length. Therefore, the biometer measures the vitreous erroneously long, and, consequently, the total length also is erroneously long. In a 4-gate system with silicone oil, the ACD and lens thickness are accurate, so they should be subtracted from the total length to isolate the erroneous apparent vitreous length. Then, the formula used is as follows: 980/1532 X Apparent Vitreous Length = True Vitreous Length The corrected vitreous length is now added back to the anterior chamber depth and lens thickness for an accurate total eye length. Biometry is best performed with these patients sitting upright so the bubble will not separate from the retinal surface, causing a spike to arise at the back of the bubble, which can be confused easily with the retinal spike. If an aphakic patient has silicone oil in the eye, it must be determined whether or not the oil is in the anterior chamber or only the posterior chamber. If the oil is only in the posterior chamber, the ACD should be subtracted from the total length, the vitreous length corrected using the velocity conversion equation, then added back to the ACD.
An eye with silicone oil in the vitreous cavity. The erroneous vitreous length was isolated from the measurements and the velocity conversion equation used to correct for the sound velocity through silicone oil. The corrected vitreous length was added back to the anterior chamber depth and lens thickness for a true eye length measurement.
If the oil reaches anteriorly to the back of the corneal endothelium, the entire eye length should be corrected using the following equation: 980/1532 X Apparent Eye Length = True Eye Length If oil is in the vitreous cavity of the pseudophakic patient, the ACD should be subtracted to isolate the posterior chamber, using the
17 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
following equation:
980/1532 X Apparent Vitreous Length = True Vitreous Length Then, add back to the ACD along with the correction factor for the IOL in place. Some newer ultrasound units have adjustable velocities or silicone filled eye types. In these cases, the vitreous velocity is simply changed to the appropriate velocity through the silicone oil, and the measurement is taken while the patient is sitting upright. Also, optical coherence biometry has been shown to be effective in measuring these eyes, although the cataract must be mild enough for it to be used and the patient must have the ability to fixate. If the silicone oil is to be left in the eye permanently, power must be added to account for the index of refraction of the silicone oil. Consulting the Holladay II IOL Consultant formula for this calculation is best. Also, a silicone IOL should not be implanted in these eyes, because it will adhere to the silicone oil.
Multimedia
Media file 1: High quality contact A-scan of the phakic eye. Note the 5 high-amplitude spikes and the steeply rising retinal spike, as well as the good resolution of the separate retinal and scleral spikes.
18 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 2: When the sound beam incidence is perpendicular to the visual axis (upper image), most returning echoes are received back into the probe tip to be interpreted on the display as high-amplitude spikes. When the sound beam incidence is nonperpendicular to the visual axis (lower image), part of the returning echo is reflected away from the probe tip, with only a portion received by the probe. As a result, the spikes will be compromised.
19 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 3: If the macular surface is smooth (upper image) more of the echoes are received back into the probe to be displayed as high-amplitude spikes. If the macular surface is convex (center image), as with macular edema or pigment epithelial detachments, part of the echoes is reflected away from the probe tip. If the macular surface is irregular (lower image), as in macular degeneration or epiretinal membranes, reflection of the echoes away from the probe tip also will occur.
20 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 4: When the gain setting is too high, the resolution of the separate retinal and scleral spikes is lost, resulting in 1 thick, flattened spike.
21 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 5: Gates are electronic calipers on the display (see arrows) that between each pair render a measurement. In this 4-gate system, each of the 3 sections of the eye is measured individually at its correct velocity, then added together for the total eye length. Equipment will vary in the appearance of these gates.
Media file 6: Immersion scan of the phakic eye. The probe and cornea are now separate spikes since they are not in contact with each other, and the corneal spike demonstrates 2 peaks (see arrow), representing the epithelium and endothelium. If these 2 peaks are not both high, the sound beam is not aligned through the corneal vertex. Gain must be reduced enough to appreciate and resolve these 2 peaks.
Media file 7: Reverberation artifacts in the vitreous cavity resulting from intraocular lens. The left image demonstrates the longer chain of artifact spikes from polymethyl methacrylate implants. The image on the right demonstrates the shorter chain of artifact spikes from foldable implants.
22 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 8: Corneal compression is demonstrated in the A-scan on the right. Note the more shallow anterior chamber depth of 2.63 mm as compared to the scan of the same eye on the left, with an anterior chamber depth of 3.20 mm, indicating 0.57 mm of corneal compression. Note also that the total eye length is shortened from 24.60 mm in the scan on the left to 24.18 mm in the scan on the right. This error would result in an unwanted postoperative refractive error of about -1.25 D.
Media file 9: Misalignment errors. The A-scan on the left demonstrates a contact scan with a sloping retinal spike. The A-scan on the right demonstrates an immersion A-scan with steps in the initial rise of the retinal spike.
23 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 10: Misalignment demonstrated by the decreased amplitude of the posterior lens spike (arrow). When either of the lens spikes is too short, the sound beam is aligned at an angle through the lens rather than through its center, and thus not aligned along the visual axis.
24 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Media file 11: Misalignment along the optic nerve. Note the missing scleral spike.
Media file 12: High myope with a long eye. Note the poor quality of the retinal spike on the A-scan. The axial length from the anterior corneal surface to the central macula was measured on the B-scan (left image), and compared to the variable A-scan measurements. The A-scan that was within 0.1 mm of the B-scan measurement was used in the intraocular lens calculation.
Media file 13: An eye with silicone oil in the vitreous cavity. The erroneous vitreous length was isolated from the measurements and the velocity conversion equation used to correct for the sound velocity through silicone oil. The corrected vitreous length was added back to the anterior chamber depth and lens thickness for a true eye length measurement.
References
1. Byrne SF, Green RL. Ultrasound of the Eye and Orbit. 2nd ed. St Louis, MO: Mosby, Inc;. 2002. 2. Byrne SF. A-Scan Axial Eye Length Measurements: A Handbook for IOL Calculations. Mars Hill, NC:. Grove Park Publishers;1995. 3. Shammas, H. John. Intraocular Lens Power Calculations. 6900 Grove Road, Thorofare, NJ 08086 USA: SLACK Incorporated; 2004. 4. Randleman JB, Foster JB, Loupe DN, Song CD, Stulting RD. Intraocular lens power calculations after refractive surgery:
25 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Consensus-K technique. J Cataract Refract Surg. November 2007;33:1892-1898. 5. Walter KA, Gagnon MR, Hoopes PC, Dickinson PJ. Accurate Intraocular Lens Power Calculation after Myopic Laser in Situ Keratomileusis (LASIK), Bypassing Corneal Power. J Cataract Refract Surg. March 2006;32:425-429. 6. Masket S, Masket SE. Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation. J Cataract Refract Surg. March 2006;32:430-434. 7. Artaria LG. Messung der bulbuslange mit verschiednen ultraschall-geraten. Klin Monatsbl Augenheilkd. Vol 188. 1986: 492-4. 8. Binkhorst RD. A Guide to the Author's TI-58-59 IOL Power Module. Intraocular Lens Power Calculation Manual. 2nd ed. New York;. 1981. 9. Binkhorst RD. The accuracy of ultrasonic measurement of the axial length of the eye. Ophthalmic Surg. May 1981;12(5):363-5. [Medline]. 10. Burton TC, Herron BE, Ossoinig KC. Axial length changes after retinal detachment surgery. Am J Ophthalmol . Jan 1977;83(1):59-62. [Medline]. 11. Byrne SF. Improving A-Scans. Ocul Surg News. 1994;12(6):86-7. 12. Clemens S, Kroll P, Rochels R. Ultrasonic findings after treatment of retinal detachment by intravitreal silicone instillation. Am J Ophthalmol . Sep 15 1984;98(3):369-73. [Medline]. 13. Hoffer KJ. Axial dimension of the human cataractous lens [published erratum appears in Arch Ophthalmol 1993 Dec;111(12):1626]. Arch Ophthalmol . Jul 1993;111(7):914-8. [Medline]. 14. Hoffer KJ. Biometry of 7,500 cataractous eyes. Am J Ophthalmol . Sep 1980;90(3):360-8. [Medline]. 15. Hoffer KJ. Take velocity of ultrasound into consideration in IOL calculations. Ophthalmology Times. 1993;18(17):40. 16. Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. Nov 1993;19(6):700-12. [Medline]. 17. Hoffer KJ. Ultrasound velocities for axial eye length measurement. J Cataract Refract Surg. Sep 1994;20(5):554-62. [Medline]. 18. Holladay JT. Cataract surgery in patients with previous keratorefractive surgery (RK and LASIK). Ophthalmic Practice. 1997;15:238-44. 19. Holladay JT. How do you calculate the intraocular lens power after radial keratotomy?. Ophthalmol . 1996;28:66-9. 20. Holladay JT. Intraocular lens power calculations for the refractive surgeon. Operative Techniques in Cataract and Refractive Surg. 1998;1:105-17. 21. Holladay JT. IOL calculations following radial keratotomy surgery. Refractive & Corneal Surgery. 1989;5(3). 22. Holladay JT, Prager TC. Accurate ultrasonic biometry in pseudophakia. Am J Ophthalmol . Apr 15 1993;115(4):536-7. [Medline]. 23. Holladay JT, Prager TC. Accurate ultrasonic biometry in pseudophakia. Am J Ophthalmol . Feb 15 1989;107(2):189-90. [Medline]. 24. Holladay JT, Prager TC, Chandler TY, et al. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. Jan 1988;14(1):17-24. [Medline]. 25. Jansson F, Kock E. Determination of the velocity of ultrasound in the human lens and vitreous. Acta Ophthalmol (Copenh). 1962;40:420-33. [Medline]. 26. Jansson F, Sundmark E. Determination of the velocity of ultrasound in ocular tissues at different temperatures. Acta Ophthalmol (Copenh). 1961;39:899-910. [Medline]. 27. Kendall, CJ. Ophthalmic Echography. Thorofare, NJ:. Slack Incorporated;1990.
26 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
28. Meldrum ML, Davis JL, Patel A. Clinical application of the theoretically derived IOL power selection method in silicone oil-filled eyes. Symposium on Cataract, IOL, and Refractive Surg. 1995;Abstract 164:41. 29. Olsen T, Nielsen PJ. Immersion versus contact technique in the measurement of axial length by ultrasound. Acta Ophthalmol (Copenh). Feb 1989;67(1):101-2. [Medline]. 30. Olsen T, Thim K, Corydon L. Accuracy of the newer generation intraocular lens power calculation formulas in long and short eyes. J Cataract Refract Surg. Mar 1991;17(2):187-93. [Medline]. 31. Olsen T, Thim K, Corydon L. Theoretical versus SRK I and SRK II calculation of intraocular lens power. J Cataract Refract Surg. Mar 1990;16(2):217-25. [Medline]. 32. Ossoinig KC. Standardized echography: basic principles, clinical applications, and results. Int Ophthalmol Clin. Winter 1979;19(4):127-210. [Medline]. 33. Ossoinig KC. Standardized Ophthalmic Echography of the Eye, Orbit, and Periorbital Region. A Comprehensive Slide Set and Study Guide. 3rd ed. Goodfellow, Iowa City:1985:12. 34. Packer M, Fine IH, Hoffman RS, et al. Immersion A-scan compared with partial coherence interferometry: outcomes analysis. J Cataract Refract Surg. Feb 2002;28(2):239-42. [Medline]. 35. Patel AS. IOL power selection for eye with silicone oil used as vitreous replacement. Symposium on Cataract and Refractive Surg. 1995;Abstract 163:41. 36. Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula [published erratum appears in J Cataract Refract Surg 1990 Jul;16(4):528]. J Cataract Refract Surg. May 1990;16(3):333-40. [Medline]. 37. Retzlaff JA, Sanders DR, Kraff MC. Lens Power Calculation - A Manual for Ophthalmologists and Biometrists. 3rd ed. Thorofare, NJ:. Slack Incorporated;1990. 38. Sanders DR, Retzlaff JA, Kraff MC, et al. Comparison of the SRK/T formula and other theoretical and regression formulas. J Cataract Refract Surg. May 1990;16(3):341-6. [Medline]. 39. Schelenz J, Kammann J. Comparison of contact and immersion techniques for axial length measurement and implant power calculation. J Cataract Refract Surg. Jul 1989;15(4):425-8. [Medline]. 40. Shammas HJ. A comparison of immersion and contact techniques for axial length measurement. J Am Intraocul Implant Soc. Fall 1984;10(4):444-7. [Medline]. 41. Shugar JK, de Juan E Jr, McCuen BW 2nd, et al. Ultrasonic examination of the silicone-filled eye: theoretical and practical considerations. Graefes Arch Clin Exp Ophthalmol . 1986;224(4):361-7. [Medline]. 42. Yang S, Lang A, Makker H, Zaleski E. Effect of silicone sound speed and intraocular lens thickness on pseudophakic axial length corrections. J Cataract Refract Surg. Jul 1995;21(4):442-6. [Medline].
Keywords
A-scan ultrasound, intraocular lens calculations, IOL calculations, axial length measurements, immersion, biometry
Rhonda G Waldron, MMSc, COMT, CRA, ROUB, RDMS, Senior Associate in Ophthalmology, Diagnostic Echographer, Senior Technologist, Department of Ophthalmology, Emory University School of Medicine Rhonda G Waldron, MMSc, COMT, CRA, ROUB, RDMS is a member of the following medical societies: Association for Technical Personnel in Ophthalmology, International Society for Ophthalmic Ultrasound, and Ophthalmic Photographer's Society Disclosure: Quantel Medical Honoraria Speaking and teaching; Quantel Medical Consulting fee Testing new equipment; Accutome Honoraria Speaking and teaching; Alcon Laboratories Consulting fee Ultrasound expert for animal research project
Coauthor(s)
27 de 28
21-10-2010 11:46
http://emedicine.medscape.com/article/1228447-print
Thomas M Aaberg, Jr, MD, Clinical Assistant Professor, Department of Surgery, Michigan State University College of Human Medicine; Consulting Staff, Department of Ophthalmology, Associated Retinal Consultants Thomas M Aaberg, Jr, MD is a member of the following medical societies: Alpha Omega Alpha, American Academy of Ophthalmology, American Medical Association, American Society of Retina Specialists, Michigan Society of Eye Physicians & Surgeons, and Retina Society Disclosure: Novartis Consulting fee Consulting
Medical Editor
Richard W Allinson, MD, Associate Professor, Department of Surgery, Texas A&M University Health Science Center; Senior Staff Ophthalmologist, Scott and White Clinic Richard W Allinson, MD is a member of the following medical societies: American Academy of Ophthalmology, American Medical Association, and Texas Medical Association Disclosure: Nothing to disclose.
Pharmacy Editor
Mary L Windle, PharmD, Adjunct Assistant Professor, University of Nebraska Medical Center College of Pharmacy, Pharmacy Editor, eMedicine Disclosure: Pfizer Inc Stock Investment from broker recommendation; Avanir Pharma Stock Investment from broker recommendation
Managing Editor
J James Rowsey, MD, Former Director of Corneal Services, St Luke's Cataract and Laser Institute, Florida J James Rowsey, MD is a member of the following medical societies: American Academy of Ophthalmology, American Association for the Advancement of Science, American Medical Association, Association for Research in Vision and Ophthalmology, Florida Medical Association, Pan-American Association of Ophthalmology, Sigma Xi, and Southern Medical Association Disclosure: Nothing to disclose.
CME Editor
Lance L Brown, OD, MD, Ophthalmologist, Affiliated With Freeman Hospital and St John's Hospital, Regional Eye Center, Joplin, Missouri Disclosure: Nothing to disclose.
Chief Editor
Rick Kulkarni, MD, Medical Director, Assistant Professor of Surgery, Section of Emergency Medicine, Yale-New Haven Hospital Rick Kulkarni, MD is a member of the following medical societies: Alpha Omega Alpha, American Academy of Emergency Medicine, American College of Emergency Physicians, American Medical Association, American Medical Informatics Association, Phi Beta Kappa, and Society for Academic Emergency Medicine Disclosure: WebMD Salary Employment
Further Reading 1994-2010 by Medscape. All Rights Reserved (http://www.medscape.com/public/copyright)
28 de 28
21-10-2010 11:46