EAL-G12: Traceability of Measuring and Test Equipment To National Standards
EAL-G12: Traceability of Measuring and Test Equipment To National Standards
EAL-G12: Traceability of Measuring and Test Equipment To National Standards
Publication Reference
EAL-G12
PURPOSE
To give guidance on the calibration and maintenance of measuring equipment in meeting the requirements of the ISO 9000 series of standards for quality systems, and the EN 45001 standard for the operation of testing laboratories.
EDITION 1 Z NOVEMBER 1995 PAGE 1 OF 16
Authorship This publication has been written by EAL Committee 2. Official language The text may be translated into other languages as required. The English language version remains the definitive version. Copyright The copyright of this text is held by EAL. The text may not be copied for resale. Further information For further information about this publication, contact your National member of EAL, whose telephone and facsimile numbers are given below:
Calib National member Testing National member Austria BMwA Tel: +431 711 02 233 Fax: +431 714 35 82 BKO/OBE Tel: +32 2 206 47 35 Fax: +32 2 206 57 45 DANAK Tel: +45 35 86 86 86 Fax: +45 35 86 86 87 FINAS Tel: +358 0 616 71 Fax: +358 0 616 7341 BMwA Tel: +431 711 02 248 Fax: +431 714 35 82 BELTEST Tel: +32 2 206 46 80 Fax: +32 2 206 57 42 DANAK Tel: +45 35 86 82 80 Fax: +45 35 86 86 87 FINAS Tel: +358 0 616 71 Fax: +358 0 616 7467 member Italy Calib National member Fax: +353 1 607 3109 SIT Tel: +39 11 348 8933 Fax: +39 11 348 6384 Testing National Fax: +353 1 607 3109 SINAL Tel: +39 6 487 1176 Fax: +39 6 481 4563 RvA Tel: +31 30 239 4500 Fax: +31 30 239 4539 NA Tel: +47 22 20 0226 Fax: +47 22 20 7772 IPQ Tel: +351 1 294 8201 Fax: +351 1 294 8202 ENAC Tel: +34 1 457 32 89 Fax: +34 1 458 62 80 SWEDAC Tel: +46 84 02 00 71 Fax: +46 87 91 89 29 SAS Tel: +41 31 963 3412 Fax: +41 31 963 3210 UKAS Tel: +44 181 943 6266 Fax: +44 181 943 7134
Belgium
Denmark
Netherlands RvA Tel: +31 30 239 4500 Fax: +31 30 239 4539 Norway NA Tel: +47 22 20 0226 Fax: +47 22 20 7772 IPQ Tel: +351 1 294 8201 Fax: +351 1 294 8202 ENAC Tel: +34 1 457 32 89 Fax: +34 1 458 62 80 SWEDAC Tel: +46 33 17 77 30 Fax: +46 33 10 13 92
Finland
Portugal
France
COFRAC COFRAC Tel: +33 1 44 68 82 33 Tel: +33 1 44 68 82 40 Fax: +33 1 44 68 82 22 Fax: +33 1 44 68 82 23 DKD DAR Tel: +49 531 592 8320 Tel: +49 30 8104 1710 Fax: +49 531 592 9292 Fax: +49 30 8104 1717 Ministry of Commerce Tel: +30 1 38 14 168 Fax: +30 1 38 42 642 ISAC Tel: +354 1 681 122 Fax: +354 1 685 998 NAB Tel: +353 1 607 3003 ELOT Tel: +30 1 201 5025 Fax: +30 1 202 0776 ISAC Tel: +354 1 681 122 Fax: +354 1 685 998 NAB Tel: +353 1 607 3003
Spain
Germany
Sweden
Greece
Iceland
Switzerland SAS Tel: +41 31 963 3381 Fax: +41 31 963 3210 United Kingdom UKAS Tel: +44 181 943 7068 Fax: +44 181 943 7134
Ireland
PAGE 2 OF 16
Contents
Section 0 1 2 3 4 5 6 7 Introduction Scope and field of application Measuring and test equipment in production and in testing laboratories Calibration, traceability Why are calibrations and traceability necessary? Elements of traceability Calibration hierarchy Terminology in the hierarchy of standards Checklist for the assessment of the calibration and traceability of measuring and test equipment Page 4 4 4 5 6 6 7 11 13
Appendix A
PAGE 3 OF 16
0
0.1
Introduction
The quality of products and services is increasingly dependent on reliable measurements. The importance attached to measurements is reflected in relevant standards by the requirement that measurements must be traceable to national or international standards. Different definitions and explanations of the term traceability exist in standards and literature, giving rise to differing interpretations (and misinterpretations). There is more to traceability than can be demonstrated in a written declaration, and the objective of this guidance publication is to give a clearer picture of the principles of traceability and how it may be achieved.
0.2
1
1.1
1.2
1.3
2
2.1
2.2
2.3
PAGE 4 OF 16
organisations that are building up a quality system. It is therefore addressed primarily to staff members responsible for quality in industry. 2.4 However, the calibration of measuring and test equipment, and the traceability of the measurements to national standards, is also an important requirement for the operation of calibration and testing laboratories and is a pre-requisite for their accreditation in accordance with European standards of the EN 45000 series and equivalent international publications. Therefore, EAL-G12 is also addressed to the operators of calibration and testing laboratories and those who assess them as part of the accreditation procedure.
3
3.1
Calibration, traceability
Calibration means determining and documenting the deviation of the indication of a measuring instrument (or the stated value of a material measure) from the conventional true value of the measurand. The term traceability means a process whereby the indication of a measuring instrument (or a material measure) can be compared, in one or more stages, with a national standard for the measurand in question. In each of these stages, a calibration has been performed using a standard with a metrological quality already determined by calibration with a higher level standard. There is therefore a calibration hierarchy, as shown in Fig. 1. The figure illustrates in particular how an in-house calibration system (right hand side of the diagram) may interact with the existing metrological infrastructure (left hand side of the diagram). The exact definitions of calibration and traceability are given in the International vocabulary of basic and general terms in metrology (VIM), BIPM et al, 1993.
3.2
3.3
3.4
Fig. 1
EDITION 1 Z NOVEMBER 1995 PAGE 5 OF 16
4
4.1
4.2
4.3
If it becomes necessary to prove absence of liability, the producer must be able to demonstrate, by reference to a systematic and fully documented system, that adequate measuring and test equipment was chosen, was in proper working order and was used correctly for controlling a product. There are similar technical and legal reasons why calibration and testing laboratory operators should have consistent control of measuring and test equipment in the manner described.
4.4
5
5.1
Elements of traceability
Traceability is characterised by a number of essential elements: (a) an unbroken chain of comparisons going back to a standard acceptable to the parties, usually a national or international standard; (b) measurement uncertainty; the measurement uncertainty for each step in the traceability chain must be calculated according to agreed methods and must be stated so that an overall uncertainty for the whole chain may be calculated; (c) documentation; each step in the chain must be performed according to documented and generally acknowledged procedures; the results must equally be documented; (d) competence; the laboratories or bodies performing one or more steps in the chain must supply evidence for their technical competence, eg by demonstrating that they are accredited; (e) reference to SI units; the chain of comparisons must end at primary standards for the realization of the SI units; (f) re-calibrations; calibrations must be repeated at appropriate intervals; the length of these intervals will depend on a number of variables, eg uncertainty required, frequency of use, way of use, stability of the equipment.
PAGE 6 OF 16
5.2
In many fields, reference materials take the position of physical reference standards. It is equally important that such reference materials are traceable to relevant SI units. Certification of reference materials is a method that is often used to demonstrate traceability to SI units.
6
6.1
6.1.1
Calibration hierarchy
International level
At the international level, decisions concerning the International System of Units (SI) and the realization of the primary standards are taken by the Confrence Gnrale des Poids et Mesures (CGPM). The Bureau International des Poids et Mesures (BIPM) is in charge with coordinating the development and maintenance of primary standards and organises intercomparisons on the highest level.
6.2
6.2.1
6.3
6.3.1
6.3.2
6.3.3
Many accredited laboratories carry out calibrations for third parties on request, eg for firms that do not have calibration and measurement facilities with suitable equipment, and for private test laboratories working in the field of product certification. Note: If an accredited laboratory is to be engaged to carry out a particular calibration, the client must ensure that the measurement uncertainty achieved is appropriate for the intended use of the calibrated instrument.
6.3.4 6.3.5
The calibration results are documented in a calibration certificate. The European Calibration Laboratory Accreditation Bodies cooperate in the European cooperation for Accreditation of Laboratories (EAL). One main goal of EAL is to ensure that calibration certificates, issued by any accredited laboratory, are accepted in the other countries. This implies that the laboratories and the Accreditation Bodies have to operate in a compatible way. An evaluation of the Accreditation Bodies and the results of interlaboratory comparisons have resulted in an EAL Multilateral Agreement on the mutual acceptance of calibration certificates. This means that official certificates of calibration (displaying the logo of the accreditation scheme) issued by a calibration laboratory accredited by one of the signatories of the Multilateral Agreement, are equivalent with the certificates issued by laboratories accredited by any of the other signatories.
6.4
6.4.1
6.4.2
6.4.3
PAGE 8 OF 16
Fig. 2
Fig. 3
EDITION 1 Z NOVEMBER 1995 PAGE 9 OF 16
Fig. 4
Fig. 5
PAGE 10 OF 16 EDITION 1 Z NOVEMBER 1995
7
7.1
PAGE 11 OF 16
Some CRMs and RMs (see below) have properties which, because they cannot be correlated with an established chemical structure or for other reasons, cannot be determined by exactly defined physical and chemical measurement methods. Such materials include certain biological materials such as vaccines to which an International unit has been assigned by the World Health Organisation.
Reference material (RM): Material or substance one or more of whose property values are sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials. Note: A RM may be in the form of a pure or mixed gas, liquid or solid. Examples are water for the calibration of viscometers, sapphire as a heat-capacity calibrant in calorimetry, and solutions used for calibration in chemical analysis.
General note: The value of any standard has an uncertainty. In the calibration hierarchy, the higher ranking standard has a smaller uncertainty. Each additional subordinate level therefore leads to an increase in the uncertainty of measurements.
PAGE 12 OF 16
Appendix A
Checklist for the assessment of the calibration and traceability of measuring and test equipment.
(CAUTION: negative answers may require further questions). A0 A0.1 A0.2 General remarks The assessor for measuring equipment must have sufficient knowledge in the fields of metrology and calibration. The assessment refers only to the testing activities for which the laboratory is to be accredited; it does not refer to the operation of a purely calibration laboratory already accredited by another body. Appropriate calibration of measuring equipment Is a calibration prescribed for all measuring instruments: (a) appropriate with respect to the measurement uncertainty of the measuring equipment; (b) appropriate with respect to the influence of the measured quantity on the test result? A1.2 Is an appropriate functional test determined for those measuring instruments that are based on natural constants (eg defined wave lengths)? Bodies performing calibration of measuring equipment Is the calibration carried out by an external body that is generally responsible for calibrations or accredited or otherwise accepted for that purpose? (a) By a National Metrology Institute? (b) By an accredited calibration laboratory? A2.2 Is the calibration carried out internally or externally by a laboratory not falling into the categories mentioned in A2.1(a) and A2.1(b)? (a) By a competent internal body of the institute operating the test laboratory? (b) By a competent staff group or single person in the test laboratory? (c) By the user of the measuring equipment himself? (d) By an external body the competence of which is demonstrated by an assessment?
A1 A1.1
A2 A2.1
PAGE 13 OF 16
A3
Calibration facilities (A3 is applicable only if the answer is yes to one of the questions in A2.2)
A3.1
Are internal reference standards and, if appropriate, working standards available for all measuring and test instruments, and measured quantities which are relevant for the measurement and test results? Are the reference standards directly or indirectly linked, in any case by an unbroken chain documented by certificates, to national standards and labelled accordingly by a calibration label? Are all calibration equipment instruments properly identified? Is each calibration described in a procedure, eg by switching diagrams or flow charts? Is the calibration procedure described step-by-step? Are defined environmental conditions ensured during calibrations? Are relevant environmental conditions recorded during calibrations? Are procedures for the calculation of the measurement uncertainty of the calibration equipment specified and are they followed? Are re-calibration intervals fixed in accordance with the intended use and the properties of the equipment, and are there programmes for regular recalibrations? Evaluation and documentation of results Are the calibration results and the associated uncertainties documented? Is the observance of fixed re-calibration intervals supervised? Where calibrations have to be performed before each measurement, are these cases clearly identified? Are the measuring instruments labelled accordingly? Are the results of calibrations (including environmental conditions, if applicable) documented and filed? Are they available to the user of the measuring instrument? Is a calibration label used as a visible indication of an established confirmation system for the measuring equipment? Is the calibration label supported by a calibration certificate and is the number of the certificate referred to on the label? Are controls for calibration and adjustment, which should not be adjusted by the user, sealed?
A3.2
PAGE 14 OF 16
A5 A5.1
Specified calibration procedures Is the measuring equipment of a self-calibration type? (a) Is the internal reference calibrated? (b) Is the process of self-calibration checked?
A5.2
Does the measuring equipment include an internal calibration of a less stable component by means of an internal reference? (a) Is the internal reference calibrated? (b) Is the procedure of internal calibration checked? (c) Is the internal calibration performed regularly, eg before each use of the measuring equipment?
A5.3
Is the complete measuring system calibrated as a whole? (a) Are the single components of the measuring system adjusted, especially with respect to zero setting? (b) How is the labelling performed for a complete measuring system?
A5.4
Are single components of a measuring system calibrated? (a) Are the calibration parameters for the complete measuring system determined from the values of the single components?
A5.5
What is done in the case of disposable measuring devices which cannot be calibrated individually (eg strain gauge transducers)? (a) Are samples calibrated? Is continuous sample testing practiced? (b) Which body is performing sample testing? (c) Does the body according to A5.5(b) fulfil the requirements of EN 45001 or ISO/IEC Guide 25? (d) Is the body accredited according to EN 45001 or ISO/IEC Guide 25?
A5.6
Are reference materials used for the calibration? (a) Are the reference materials certified?
A5.7
Are the calibrations computer-aided? (a) Is the software validated? (b) By which method?
PAGE 15 OF 16
Responsibilities; administrative aspects of calibration of measuring equipment Is each user of measuring equipment aware that he is responsible for the calibration status of his measuring equipment? Is each new measuring equipment calibrated before use? Are measuring instruments brought to re-calibration by a confirmation system when the calibration validity period is expired? Is the responsibility for the working standards, and the internal reference standards and their traceable calibrations clearly defined? Is the responsibility for the reliability of calibration software clearly defined?
PAGE 16 OF 16