Shimadzu

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

C122-E058A

AA-7000 Series

Shimadzu Atomic Absorption Spectrophotometers

AA - 7000 Series
Shimadzu Atomic Absorption Spectrophotometers Reaching Even Greater Heights

Enhanced Flame Analysis World-Class High-Sensitivity Furnace Improved Dual Atomizer System Advanced Safety Technology
P 18 - Supports System Management and Accuracy Management P 20 - Standard Parts / PC / Software

Contents

P 06 - Enhanced Flame Analysis

P 08 - World-Class High-Sensitivity Furnace

P 10 - Improved Dual Atomizer System

P 21 - Optional Accessories / Related Products

P 12 - Advanced Safety Technology

P 24 - Specifications

P 14 -Dual-Background Correction Functions

P 26 - Installation Conditions

P 16 - Clear, Easy-to-Use Software

System Configuration can Evolve with Your


The AA-7000 Series can be upgraded by adding units to allow the system to handle the analysis targets.
ASC

*1

AA-7000F Manual Burner System

AAC

AA-7000F

AAC

ASC

GFA

*3

*2

AA-7000F Auto Burner System

ASC

*2

AA-7000G

GFA-7000

AA-7000G Auto Sampler System

Major Fields of Application


AA-7000 Series supports a wide range of analysis applications.

Analysis Sensitivities
Flame Furnace

Concentration limit of detection

Environment
Seawater, river water, effluent, sludge, air-borne dust

(ppm)

(ppb)

H Li Be B C N P O S

He F Ne Cl Ar

Metals, Semiconductors, Ceramics


Metals, minerals, glass, ceramics, IC chips

Na Mg K Ca Sc Ti Rb Sr Y

Al Si

V Cr Mn Fe Co Ni Cu Zn Ga a Ge As Se Br Kr I Xe

Petroleum, Chemicals, Polymers


Petroleum, oil, catalysts, chemical products, biodiesel

Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te

Cs Ba La Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn Fr Ra Ac Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
* May differ according to coexisting substances in the sample.

Medical, Biology, Pharmaceuticals


Blood, animals, plants, drugs, food products

Needs

AUTO

AA-7000F Dual Atomizer System


AAC Auto Atomizer Changer GFA Graphite Furnace Atomizer AUTO ASC Autosampler

Dual Atomizer System Automatic adjustment of atomizer Automatic serach of burner height

*1 Requires ASC stand kit. *2 Requires ASK-7000. *3 Automatic burner height search function can be used.

Flame Selection
Flame and Measurement Procedures

Air-C2H2 N2O-C2H2 HVG MVU

H Li Be Na Mg K Ca Sc Ti Rb Sr Y B C N P O S

He F Ne Cl Ar
AA-7000F Flame Model

Al Si

V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr I Xe

Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te

Cs Ba La Hf Ta W Re Os Ir Pt Au u Hg Tl Pb Bi Po At Rn Fr Ra Ac Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
AA-7000G Furnace Model (with options attached)

Enhanced Flame Analysis

Monochromator

Detector

Chopper mirror

Reference beam
D2 lamp

Sample beam
Beam splitter

Burner head (or graphite tube)

Hollow cathode lamp

Double-Beam Optics and Stable Hardware Achieve Superior Stability


The AA-7000 Series incorporates newly developed 3D double-beam optics. The optical system has been designed to produce its maximum performance for each measurement method through optimal adjustment of the light beam and light beam digital filter, and by using optical components that restrict light losses.

Advantages of the Double-Beam System


Long-term stability
The graph shows the results of measurements on 2ppm copper (Cu) conducted over at least one hour. (The plot shows mean values for 11 repeated measurements.) Over the course of more than 600 measurements, the instrument achieved a relative standard deviation within 1%.
Absorbance

High sensitivity
The graph shows the direct measurement results for 0.1 ppm lead (Pb).

A b s

Measurement time
Data Showing Long-Term Stability for Flame Cu Measurements

Conc(ppm)
0.1ppm to 0.4ppm Calibration Curve for Lead (Pb)

Support for Developing Analytical Conditions


Automatic gas flow rate optimization
Automatic searching for optimal fuel gas flow rate (Japanese Patent 2099886). It is important to determine the optimal gas flow rate for the flame when using an organic solvent or after changing the burner height. The AA-7000F Series automatically optimizes the gas flow rate by measuring the changes in absorbance between a blank and a standard sample. The difference between the two is displayed on the screen. The gas flow rate achieving the highest sensitivity is detected and this value is automatically set as the gas flow rate value.

Automatic burner height optimization (with options attached)


The absorption sensitivity for flame analysis is also affected by the burner height. This results from variations in the flame temperature due to the burner height. The flame temperature is also affected by the matrix components. The optional AAC-7000 adjusts the burner height in 0.5mm steps to automatically search for the optimal conditions.

Optional Autosampler Reduces Analysis Workload


Low carryover
Great care must be taken to avoid carryover during flame analysis. The new ASC-7000 autosampler rinses the nozzle at the rinsing port on the overflow mechanism after each sample measurement is complete. This ensures 10-4 max. carryover during the measurements of multiple samples. The graph shows the results of consecutive analyses of 10ppm, 20ppm, and 50ppm sodium (Na) standard solutions in the EMISSION mode. No carryover could be detected when measuring the 10ppm standard sample immediately after the measuring the 50ppm Na standard sample.
Discharge

Conventional method

Overflow method

Data showing low carryover

Trace Sample Analysis Using Micro Sampling


Micro sampling
At least 1mL (1000L) volume of a liquid sample is required for the continuous intake of sample during normal flame analysis. With micro sampling, however, approximately 50 to 90L of sample is injected in one shot into the flame and quantitation is based on the height and area of the peak signal obtained. This method offers the advantages below.

. Permits analysis of small sample volumes . Multi-element analysis of small sample volumes . No blockage of burner slot with samples having a high salt concentration . Synchronized with the autosampler for auto-dilution measurements
* Synchronization with the autosampler requires the optional micro sampling kit.

Advantages

Sampling port

Sampling port

Solenoid valve (closed) 1. Sample injected into micro sampling port.

Solenoid valve (opened) 2. Solenoid valve opens to inject sample in one shot into the flame.

Example of micro sampling analysis


Examples of the measured waveforms (overlaid) and calibration curve for micro sampling analysis are shown to the right. A 2-ppm Cu standard solution was prepared by auto-dilution using the autosampler. The autosampler can also be used to conduct dilution measurements of the sample. (Injection volume: 90L)

World-Class High-Sensitivity Furnace

Achieves World-Class Lower Limits of Detection


Advances in optics and a new graphite furnace design achieve improved lower limits of detection for furnace analysis (comparison with previous Shimadzu models). This superb analysis performance is possible in all fields.
Enhanced Sensitivity due to Graphite Furnace
Previous Pb 0.08ppb Mn 0.02ppb New 0 .0 5 p p b 0 .0 1 p p b

High-Performance Autosampler
Mix and inject up to four sample types (diluted solution, standard solution, sample, matrix modifier, etc.). (Of course, unmixed samples can also be injected.) ected.) Select a fluororesin tube or pipette tip as the injection nozzle. Automatic dilution and re-analysis if the calibration curve concentration is exceeded.

Example of Automatic Dilution and Re-Analysis


Analysis of cadmium (Cd): The maximum concentration of the calibration curve has been set to 1ppb. At 1.8ppb, the sample of [003] has exceeded the maximum concentration. As a resultit has been automatically diluted by 4 times and re-measured.

High-Sensitivity Application Data (Analysis of Selenium (Se))


Independent control of the gas flow through the graphite tube during atomization achieves extremely high-sensitivity measurements. (Japanese Patent 2067563)

Measurement Example Showing High Sensitivity


Analysis of 0 to 5ppb selenium (Se): Sensitivity is adequate for 1 ppb measurements. (20L injected volume, Pb modifier, pyro-coated graphite tube)

Digital temperature control and electronic gas flow control enhance stability
Combination of a high-sensitivity optical sensor and unique digital temperature control technology achieves highly accurate temperature control in all temperature regions from drying to atomization. An electronic flow controller can precisely control the inner gas flow rate in 0.01L/min units. These control techniques significantly enhance both the sensitivity and the data stability.
Measurement Example Showing High Stability
Analysis of 0.5ppb lead (Pb): Approx. 5% relative standard deviation after 5 repeated measurements is adequate for quantitative analysis.

Selecting the Graphite Tube


High-density graphite tube
(P/N 206-50587) Can be used for all elements. Especially effective for low-boiling point elements (Cd, Pb, Na, K, Zn, Mg, etc.).

Pyro-coated graphite tube


(P/N 206-50588) Effective for elements that readily form carbides (Ni, Fe, Cu, Ca, Ti, Si, V, Mo, etc.).

Platform tube
(P/N 206-50887-02) Restricts chemical interference due to coexisting substances. Effective for the analysis of environmental samples and biosamples, such as sea water and industrial waste.

* Depending on the state of the sample, some other combinations may be appropriate.

Improved Dual Atomizer System

Single Unit Handles High Concentration to Trace Analysis Samples


When attached to the AA-7000F, the optional AAC-7000 Auto Atomizer Changer* creates an integral burner and furnace that remains permanently installed in the combustion chamber. It supports both flame analysis of high concentration samples and furnace analysis of trace samples, without changing the atomizer unit. The integral construction with the atomizer saves space.
* The AAC-7000 cannot be attached to an AA-7000G instrument.

10

New Atomizer Design Permits Easy Switching Between Flame and Furnace Measurements
The new drive mechanism halves the switching time between flame and furnace measurements (comparison with previous Shimadzu models). The atomizer unit is switched automatically by software operation. No troublesome replacement of the atomizer unit is required. No need to disconnect pipes or wires.

Flame measurement: Burner head intersects the light axis (red line).

Furnace measurement: Furnace intersects the light axis (red line).

A single autosampler can be used for both flame and furnace measurements. Simply slide it left or right. No need to purchase two autosamplers.

Flame measurement

Furnace measurement

Slim, Easy-to-Use Design


Smallest footprint for a full system
The system with a dual atomizer fits into a 960 mm width (including power supply for the graphite furnace atomizer), achieving the world's smallest installation footprint* for a full system.
* Shimadzu in-house investigation. Current as of November 2008.

Easily attach an autosampler


Direct connection of the autosampler to the AA instrument minimizes the installation footprint and simplifies maintenance.

Lamp house
Just slide the cover to the side for easy replacement of the hollow cathode lamps through the large opening.
AA-6300

AA-7000F Dual Atomizer System


* Dimensions do not include ASC-7000. Height (H) does not include protrusions.

11

Advanced Safety Technology

Comprehensive Safety Functions in Every System


Safety is an extremely important consideration for flame atomic absorption spectrophotometers using acetylene gas. The AA-7000 Series instruments are the first* ones in the world fitted with a vibration sensor as standard. They also incorporate a comprehensive range of safety mechanisms, including gas leak detectors.
* Shimadzu in-house investigation. Current as of November 2008.

12

Safety Functions for Flame-Specification Instruments


World first! flame extinguished automatically by vibration sensor
The flame is automatically extinguished when the new built-in sensor detects vibrations. Instruments are completely safe during severe shaking caused by earthquakes.

Multimode automatic gas leak check


The fuel gas pipes inside the instrument gas controller are automatically checked for gas leaks when the power is turned on and when the flame is extinguished. Should a gas leak be detected, a warning buzzer sounds and a warning is displayed on the screen.

Automatic flame ignition and extinguishing


Flame ignition and extinguishing operations are extremely simple. The Air-C2H2 flame priority ignition mechanism prevents flashback.

Automatic Air-N2O switching system with acetylene flow-rate monitor


After the ignition of an Air-C2H2 flame, the flame automatically switches to an N2O-C2H2 flame. If the C2H2 flow rate does not increase due to solenoid valve trouble, for example, switching to the support gas is disabled to prevent flashback.

The connected options and operation of safety mechanisms are checked during instrument initialization.

Pressure monitor prevents flashback Drain tank level monitor

Automatic gas shut-off by flame monitor Mechanism to prevent misuse of the burner

Sudden power interruption detection and re-ignition safety measures

Safety Functions for Furnace-Specification Instruments


Cooling water flow rate monitor Argon gas pressure monitor Furnace cooling check Overcurrent protection unit (double-checked by circuit protector and optical sensor)

Safety-Conscious Design
Flame-retardant materials
UL 94V-0-rated flame-retardant materials are used for the instrument external covers and atomizer unit. This superior design pays due consideration to safety.

Safe ignition switches


The switches perfectly fit finger contours and are positioned near the window. To prevent inadvertent operation, both switches must be pressed simultaneously to start ignition.

Durable pipes and reliable couplings


All pipes are selected for durability and used with highly reliable couplings.

Clear window
The large window permits observation of the flame, and the burner can be easily accessed through the large opening. The window closes when released to eliminate the chance of it being left open.

13

Dual-Background Correction Functions


The optimal background correction methods are installed as standard: high-speed self-reversal These functions can correct for spectral interference in flame measurement. Selecting the optimal background correction method for each sample ensures accurate and reliable analysis results.
Samples suitable for the SR method
Samples with a complex matrix (Containing a large quantity of a specific element as the main component)

Samples suitable for the D 2 method


Purified water, tap water, environmental water, etc. Samples with a relatively simple matrix

SR (high-speed self-reversal) method accurate background correction over a wide range


Features
1. High-speed self-reversal (SR) correction is generally more accurate than deuterium lamp (D2) correction. As both atomic absorption and background absorption can be measured using a single lamp, the correction errors due to light-axis misalignment are extremely small. This is ideal for the quantitation of trace components in a matrix exhibiting complex background absorption, such as bio-samples and metals. 2. Permits background correction over the entire wavelength range from 185 nm to 900 nm. 3. This method can correct for spectral interference due to neighboring lines that can occur when a resonance line for another element exists near the analytical line for the target element. (See table on next page.) 4. As no polarizer is used, measurements are possible with low light losses and a high S/N ratio. 5. The rapid lamp lighting permits accurate measurement unaffected by emission noise in the atomizer.
* Hollow cathode lamp L-2433 is required to use the SR method. Hollow cathode lamp L-2433 can also be used for the D2 method.
IH spectrum Background absorption Background Atomic absorption Lamp current IH Atomic absorption

Principle
A small current IL (approx. 10 mA) and a large current IH (approx. 500 mA) are alternately passed through the hollow cathode lamp. The lamp emission spectrum when the large current flows has a depression in the center (self-reverse), due to self-absorption of the large number of sputtered atoms in the atom cloud, as shown in the diagram to the left. No significant atomic absorption is apparent and background absorption mainly occurs. Conversely, the lamp emission spectrum when the small current flows comprises a single narrow peak resulting from both atomic absorption and background absorption. By determining the difference between the two types of absorption, it is possible to accurately correct for the background absorption and measure the true atomic absorption.

Wavelength IL IL spectrum

Wavelength

Absorption layer Background

Atomic absorption

Time

Atomic absorption

(Lamp energy)

(Sample measurement)

(Energy component ratio)

Examples suitable for SR method (where differences result between SR and D2 methods) Example: Measurement of trace levels of zinc in iron (analysis of Zn in Fe solution)
BGC-SR method
Atomic absorption signal Background signal

BGC-D 2 method
Zn 0.25ppm

Atomic absorption signal Background signal

Zn 0.50ppm Fe 0.1% Fe 0.5% Fe 0.5% Fe 0.5% Fe 0.75% Zn 0.25ppm Zn 0.5ppm Zn 0.3ppm

The identical 0.5 ppm Zn solution is accurately corrected to the same absorbance at (2) and (6).

Due to inadequate correction, the absorbance is higher at (6) than at (2) for the identical 0.5 ppm Zn solution.

14

method (SR method) and deuterium lamp method (D2 method).


Examples of elements and wavelengths causing spectral interference problems due to neighboring lines
Measured Element Analytical Line (nm) Coexisting Element Absorption Line (nm) Al As Ca Cd Cu Ga Mg Ni Pb Sb Se Si Zn 309. 28 193. 76 422. 67 228. 80 324. 75 294. 36 285. 21 232. 00 217. 00 217. 58 196. 03 251. 61 213. 856 Mg Fe Fe Ni Fe Ni Fe Fe Fe Fe Fe Fe Fe 3 0 9 .3 0 1 9 3 .7 3 4 2 2 .6 4 2 2 8 .8 4 3 2 4 .7 3 2 9 4 .3 9 2 8 5 .1 8 2 3 2 .0 4 2 1 6 .9 5 2 1 7 .5 5 1 9 6 .0 5 2 5 1 .6 9 2 1 3 .8 5 8 9

The SR method is suitable for samples containing elements that cause problems with spectral interference, as shown in the table to the left.

D2 (deuterium lamp) method highly sensitive background correction


Features
1. Detection sensitivity is superior to the SR method. Therefore, this method is suitable for the analysis of samples with a simple matrix requiring high sensitivity, such as the measurement of trace levels of impurities in ultrapure water or environmental analyses. 2. As the lighting frequency is higher than with the SR method, it can eliminate noise due to emission components of the flame or graphite tube to permit accurate atomic absorption measurements. 3. The original hollow cathode lamp can be used.
D2 lamp spectrum Background absorption

Principle
Atomic absorption

The deuterium lamp method involves lighting the hollow cathode lamp and the deuterium lamp alternately at high speed. After separation by the monochromator, the light from the deuterium lamp has a bandwidth from 0.1 to 5 nm. Therefore, an atomic absorption with a line width of only about 1/1000 nm is almost unobservable compared to the background absorption due to wide-bandwidth molecular absorption. However, as the light from the hollow cathode lamp has approximately the same bandwidth as the atomic absorption band, the total of the atomic absorption and the background absorption can be observed. With the deuterium lamp (D2) method, light from both sources passes through the atomizer. The difference in absorbance is determined to conduct background correction.

Hollow cathode lamp spectrum

Atomic absorption + background absorption

Examples suitable for D2 method (where no difference results between SR and D2 methods) Example: Measurement of trace levels of lead in 2% NaCl solution by molecular absorption (analysis of Pb in 2% NaCl solution)
BGC-SR method BGC-D 2 method

Background signal

Atomic absorption signal

Spike

0ppb

2ppb

4ppb

Spike

0ppb

2ppb

4ppb

It can be seen that the sensitivity is higher with the BGC-D2 method.

15

Clear, Easy-to-Use Software


Features of the WizAArd Software The WizAArd software used with the AA-7000 runs under Windows Vista/XP. Wizard Functions Make Setting the Conditions So Simple p That Even a Novice Can Do It.
Just set the measurement conditions using WizAArd to complete the general settings. The initial settings can also be completed simply by following the Wizard procedures.

Start Wizard.

Select element.

Set calibration curve and sample parameters.

Set monochromator parameters.

Connect to instrument.

Measurement Screen Layout Shows Measurement Status at a Glance


Measured element Real-time signal monitor Signal profile display Calibration curve display

MRT (Measured Results Table) The worksheet shows sample names, absorbance, concentrations, and correction calibration results.

Lamp History Function Assists with Lamp Management


The accumulated operation time for each lamp displayed on the lamp registration screen assists with lamp service life management. Multiple lamps for the same element are differentiated using lamp IDs.

16

Software Provides Comprehensive Information Displays and Flexible Settings Identical software operation for flame and furnace measurements
Even if the calibration curve is created after sample measurements, concentration calculations can be conducted on past data. This allows trial measurements to be directly used as actual measurements. Extrapolation of calibration curve range. If the sample concentration exceeds the calibration curve range, additional measurements can be performed on standard solutions to re-calculate the concentration. Use existing calibration curves for measurements. Useful for sample concentration order checks and for checking the instrument sensitivity. One-touch re-measurement operation. No need to repeat the measurement from the beginning. Create and save multiple calibration curves on a single worksheet. Select the appropriate calibration curve for the sample concentration and composition.

g Combining a high-performance autosampler simplifies advanced settings


Real-time drying and ashing progress display for furnace measurement A higher temperature cleaning operation than the cleaning stage can be added for each measurement. This reduces the graphite tube memory effect. Tube deterioration is reduced by minimizing the cleaning operations. Add optional autosampler rinsing operations to ensure adequate rinsing of the autosampler after completing the measurement of high-concentration samples. Set up sample dilution and spike-and-recovery tests on the same screen.

Overlay function allows size and shape comparison of peak profiles. Large-volume injection permits high-sensitivity measurements. Precision control of drying and ashing temperatures reduces analysis time.

17

Supports System Management and Accuracy


FDA 21 CFR Part 11 Compliance
A combination of enhanced WizAArd software with Shimadzu's network-compatible CLASS-Agent data management software supports FDA 21 CFR Part 11 Compliance. WizAArd used alone provides comprehensive system policy setting, user management, log browser, audit trail, and electronic signature system management functions. Hardware validation software installed as standard.

Comprehensive Data Management Functions


System Policy
The system policy settings allow advanced password and lockout settings. The levels can be set in stages from "unrestricted" to "Part 11 compatible".

CLASS-Agent (Option)
Connection to CLASS-Agent permits efficient, long-term database management of large amounts of analysis data.

User Management
User management restricts the users of the system. It registers each user to determine how each one uses the software.

Log Browser
The log browser easily displays a variety of logs to check the system modification history and other log information.

18

Management
Accuracy Management (QA/QC) Functions Permit User Level Management, Audit Trail, and Electronic Signatures
QA/QC compatibility
The QA/QC functions permit accuracy management.

Hardware validation software installed as standard ard


The hardware validation software makes it simple to evaluate the instrument performance. When combined with the autosampler, it automatically determines the wavelength accuracy, noise level, baseline drift, absorbance and repeatability, and prints out the results of comparison with the pass criteria.

* Data from all Shimadzu analytical instruments can be centrally managed by CLASS-Agent.

19

Standard Parts / PC / Software

Common to All AA-7000 Series Models


Part Name Cable set Cards (set of 10) Instruction manual Inspection report Safety inspection sheet (safety instruction) Declaration of conformity, explanatory notes (ISO-9001) Warranty label Registration form Serial number label Quantity 1 1 1 1 1 1 1 1 3 Cat. No. 071-60816-12 206-52046-91 206-97176 206-77551-02 206-97225 228-30164 037-70238-01 221-40500 206-57770

* The PC, monitor, printer, hollow cathode lamps, high-temperature burner head, pressure regulator and compressor are not included in the standard configuration.

AA-7000F Standard Parts List


Part Name Hose ASSY (for air supply) Hose ASSY (for C2H2 gas supply) Hose clamp (16 mm) Cleaning wire Sampling tube (PTFE) Sampling tube (for organic solvent samples) Polyethylene tubing, No. 3 Polyethylene tubing, 8 1 (drain tubing) Drain ASSY Grease (in cup) Funnel Declaration of conformity, English version (ISO-9001) Quantity 1 1 2 1 2 1 1 ( 0.3m ) 1 ( 2.4m ) 1 1 1 1 Cat. No. 206-50389-91 206-50389-92 037-61019 201-79229-01 204-05899-01 206-50772-91 200-31328-01 016-43201-02 206-77413-91 206-50442-91 206-77243-91 206-84934-66

AA-7000G Standard Parts List


Part Name Declaration of conformity, English version (ISO-9001) Furnace mounting plate * GFA-7000 is not included in the standard configuration. Quantity 1 1 Cat. No. 206-84934-67 206-77704

Personal Computer / Monitor


Operating system CPU RAM Monitor Storage device I/O port Microsoft Windows Vista Business or XP Professional Intel Celeron 420 (1.60GHz) or higher 1 GB or higher (Vista) or 512MB or higher (XP) XGA (1024768 dots) or higher One CD-ROM drive (for installing software) Requires 60MB min. hard disk space for installation One serial port (for AA control)

* Windows is a registered trademark of Microsoft Corp., U.S.A.

20

Optional Accessories / Related Products

For Flame Analysis


Part Name High-temperature burner head Cat. No. 206-77530-91 Remarks Made of pure titanium. Air-cooled. 5 cm slot for N2O-C2H2 flame

.Extremely corrosion-resistant .Can also be used for Air-C H flame.


2 2

Flow meter kit Sample platform Air compressor Low-noise air compressor Mist separator kit YR-71 compressed gas regulator MAF-85S compressed gas regulator Micro sampling kit O-ring set

206-77617-91 206-77655-91 208-91753-91 208-91750-31 206-52458-91 040-72020-01 040-72019-11 206-77540-91 206-77620-92

Float-type flow meter for support gas Dimensions: 250 W 130 D 170 H mm Vial mounting platform: 220 W 95 D mm Vial positions: 5-step switching 100 VAC, 50/60 Hz, with mist separator 100 VAC, 50/60 Hz, with mist separator Required if using an air compressor other than above. For C2H2 For dinitrogen oxide gas Required to use the flame micro sampling method. ASC-7000 and ASK-7000 (or ASC stand kit) are also required. O-ring set for organic solvents

For Furnace Analysis


Part Name GFA-7000 graphite furnace atomizer Cat. No. 206-77700-XX Remarks ISpecifications Heating control system: Drying: Digital current control (with automatic temperature calibration function) Ashing, atomization: Digital temperature control via optical sensor Heating temperature range: Ambient to 3000 C Inner gas type: Dual automatic switching type Inner gas flow rate: 0 to 1.50 L/min, 0.01 L/min increments Ar gas: 3.5 L/min max. Cooling water: Cooling water circulation unit or tap for process water Water temperature: 10 to 30 C, flow rate: 0.6 to 1.5 L/min
* Separately order the parts below.

High-density graphite tube Pyro-coated graphite tube Platform tube MAF-106S compressed gas regulator CA-1112 cooling water circulation unit Cooler connection kit Cooling water tube ASSY Regulator ASSY

206-50587 206-50588 206-50887-02 040-72019-21 044-01809-03 206-84373-91 206-51028-91 206-86147-91 Argon gas cylinder For cooling GFA, 100 VAC, 50/60 Hz Operational temperature range: -10 to +40 C For connecting GFA and CA-1112 Connecting tubes when using tap water to cool GFA Decompression valve when using tap water to cool GFA Select one of the three types of graphite tubes according to the aim of the analysis.

Parts for cooling water connections

Autosampler
Part Name ASC-7000 autosampler Cat. No. 206-77600-XX ISpecifications System: Flame / furnace Function: Zero-point detection, auto rinse, auto diagnosis, random access Maximum reagent / sample positions: Reagents: 8 positions, samples: 60 positions
* Separately order the parts below.

Remarks

For flame analysis

For furnace analysis

ASC stand kit ASK-7000 extension unit for furnace analysis Nozzle ASSY, HVG

206-77650-91 206-77550-91 206-67563

Required to mount ASC-7000 to conduct flame analysis only. Permits both flame and furnace analysis. Required to use ASC-7000 and HVG-1.

21

Optional Accessories / Related Products

Dual Atomizer System


Part Name AAC-7000 auto atomizer changer Cat. No. 206-77701-91 Remarks Required to add GFA-7000 to AA-7000F to conduct furnace analysis. Permits automatic burner position setting for flame analysis with AA-7000F.

Hollow Cathode Lamps


Part Name L-233 series L-2433 series Cat. No. 200-38422-XX 200-38456-XX For SR method Remarks

Other Accessories
Part Name Analog output cable Cat. No. 206-77707-91 Remarks Used for analog output to a pen recorder, etc. One cable required per channel (atomic absorption/energy signal, background signal).

For High-Sensitivity As, Se, Sb Analysis


Part Name HVG-1 hydride vapor generator Cat. No. 206-17143-XX Remarks Environmental standards prescribe the hydride generation method as one method of As, Se, and Sb analysis. 1. Used with the AA-7000 Series, HVG-1 permits the rapid and accurate quantitation of elements such as As, Se, Sb, Sn, Te, and Bi at several-ppb levels. 2. Used with an ASC-7000 autosampler, it permits the automated serial analysis of up to 60 samples.
* Nozzle ASSY, HVG (P/N: 206-67563) is required to use HVG-1 with an ASC-7000 Series instrument. Order separately.

ISpecifications Measurement method: continuous flow Sample consumption: 0 to 7 mL/min, variable Reagent consumption: 0 to 2.5 mL/min, variable Atomizer: Heated absorption cell (heated by Air-C2H2 flame in standard system) Carrier gas: Ar pressure: 3.2 kg/cm2, consumption: 70 mL/min Power requirements: 100, 115, 220, 240 VAC, 30 VA, 50/60 Hz Dimensions: 340 W 220 D 200 H mm IStandard Major Items Hydride vapor generator, absorption cell (P/N: 200-66584-03) Reagent bottles (P/N: 204-09363-01), gas hose, drain tube, etc.

For Higher Sensitivity with the Hydride Generation Method


Part Name SARF-16C atomic muffle furnace (Electronic Cell Heater) Cat. No. 208-97249 Remarks 1. This dedicated furnace permits higher sensitivity measurements using the hydride vapor generator than the flame heating method. 2. The temperature controller provides optimal control of the quartz cell temperature. 3. Good temperature reproducibility. Extends cell life by preventing damage due to overheating.
* The atomic muffler cannot be used when an AAC-7000 is installed on an AA-7000F. Te Temperature sensor Cell he heater Controller

ISpecifications Furnace Dimensions: 170 W 110 D 110 H mm I.D.: 25 170 mm Heater: PYROMAX DS Operational temperature range: Ambient to 1100 C Temperature regulation range: 2.5 C Load capacity: 1000 W max. Power requirements: 100 VAC, 400 VA, 50/60 Hz

Controller

HVG-1 hydride vapor generator

Atomic absorption spectrophotometer

Computer

* Use a stepdown transformer in regions that do not have a 100 V power supply.

Mounting adaptor

206-52135-91 206-83755-91

For AA-7000F For AA-7000G

22

For High-Sensitivity Hg Analysis


Part Name MVU-1A mercury vaporizer unit Cat. No. 206-21932-XX Remarks This mercury vaporizer unit permits analysis using the reduction vaporization atomic absorption method. It permits easy, high-sensitivity water quality analysis. ISpecifications Vaporization method: reduction vaporization with a reducing agent Measurement method: recirculation Flow cell: optical path length 100 mm (with quartz window) Sample volume: 250 mL max. Exhaust contamination prevention: adsorption trapping in mercury trap bottle Dimensions: 200 W 288 D 287 H mm Weight: Approx. 10 kg IStandard Major Items MVU-1A unit: 1 Reaction vial (P/N: 200-93018): 5 Reaction vial lid (P/N: 204-21989): 2 Stirrer tip (P/N: 046-00617-06): 10 Mercury trap bottle (P/N: 038-00218): 1
*Separately order the parts below. . Gas flow cell (P/N: 201-98687) . Holder for gas flow cell (P/N: 206-77703-91) . Hg hollow cathode lamp (P/N: 200-38422-28)

ICP Emission Spectrometers


ICPE-9000 Remarks Offers both high throughput and easy analysis. Does not require processing of large data volumes, such as spectral line selection, which was problematic with the conventional Multi-ICP. Various assistant functions allow anyone to rapidly perform accurate analysis. Light source: Axial observation, compatible with mini torch Monochromator / detector: Echelle CCD Wavelength range: 167 to 800 nm (using vacuum monochromator)

23

Specifications
Main Unit
Basics
Wavelength range Monochromator Bandwidth Detector Optics Optics Background correction method Number of HC lamps Lamp mode Software requirements Parameter setting Measurement mode Concentration computation mode Repeat analysis Baseline correction Signal processing segment setting Sensitivity correction Data processing Analog output Tabular data processing Recall of parameters Procedure/result display Report generation QA/QC Re-analysis Digital recording Power requirements Dimensions and weight Ambient temperature / humidity 185.0 to 900.0 nm Aberration-corrected Czerny-Turner mounting, Number of grating grooves: 1800 lines / mm, Focal length: 298 mm 0.2, 0.7, 1.3, 2.0L nm (4-step automatic switching)

. .

Photomultiplier tube Optical double-beam BGC-SR (high-speed self-reversal method) (185.0 to 900.0 nm) BGC-D2 (D2 lamp method) (185.0 to 430.0 nm)

6-lamp turret, 2 lamps simultaneously lit (1 for measurement, 1 warming up for next measurement) EMISSION, NON-BGC, BGC-D2, BGC-SR Microsoft Windows Vista Business / XP Professional Wizard method Flame continuous method, flame micro sampling method, furnace method, flame emission method

. Calibration curve method (select primary, secondary, tertiary) . Standard addition method, simple standard addition method (primary expression)
Up to 20 repetitions. Mean value, standard deviation (SD) and coefficient of variation (RSD) display Automatic exclusion of deviant values by setting SD and %RSD Automatic correction of baseline drift by offset correction in peak height / peak area modes. Signal processing segments can be changed in peak height / peak area modes. Automatic calibration curve correction function using sensitivity monitoring 2 channels (atomic absorption/energy signal, background signal) Output range: 5.0, 2.5, 1.25, 0.625 Abs./V (each settable in 4 stages) Fixed at 1 V F.S. in EMISSION mode. Final concentration calculations based on sampled volume, dilution rate, fixed volume, and factor inputs Template functions available MRT (Measurement Results Table) worksheet Summary report Select whether to continue or discontinue measurements based on results of evaluation on coefficient of correlation, %RSD, ICV . ICB, CCV . CCB, PB, LCS, SPK, PDS, and DUP.

. Select whether on not to conduct re-analysis. . Automatic dilution and re-analysis of unknown samples via autosampler (flame micro sampling method, furnace method) . Management by login ID and password . Control user access authority by user level . Log record . Audit trail . Electronic signatures
Choose from 100, 120, 220, or 230 VAC, 50/60 Hz (Power is required separately for the personal computer.) AA-7000F: 700 W 588 D 714 H mm, 72 kg AA-7000G: 700 W 580 D 538 H mm, 65 kg (Protruding parts and optional equipment are not included.) 10 to 35 C, 20 to 80% (less than 70% when temperature is higher than 30 C)

. . . . .

Flame
Type Burner head Nebulizer Chamber Burner unit Positioning Air-cooled pre-mix type Pt-lr capillary PTFE orifice Titanium 10 cm slot (5 cm titanium slot for N2O-C2H2 flame available as an option) ceramic impact bead (capable of handling hydrofluoric acid) Engineering plastics AA-7000F/AA-7000G

. Lateral/vertical manual adjustment . Automatic flame/furnace switching by motor . Automatic search of optimum burner height
AA-7000F (with an optional auto atomizer changer mounted) 0 to 90 (Angle adjustment is not possible if the optional auto atomizer changer and GFA-7000 are both installed.)

Angle adjustment Flow rate control Gas control Safety measures

. Automatic fuel gas flow rate setting (0.1 L/min step) . Automatic search of optimum gas flow rate . Automatic gas leak check . Automatic Air-N O switching as C H flow rate increases . Flame monitor . Prevention of wrong burner head use . Gas pressure monitor . Drain tank level monitor . Automatic flame extinction upon power outage or sudden power interruption . Automatic flame extinction via flame vibration sensor . Internal fan stop sensor
2 2 2

Furnace (GFA-7000)
Heating temperature range Heating control system

Setting heating conditions

Safety measures

. Ambient to 3,000 C . Drying: Digital current control with automatic temperature calibration function . Ashing, Atomization: Digital temperature control via optical sensor . Maximum 20 stages . Heating mode: RAMP/STEP . Inner gas type: Dual automatic switching type . High-sensitivity mode setting . Enrichment in furnace: Maximum 20 times . Optimum temperature program search support function . Inner gas flow rate: 0 to 1.50 L/min . Cooling water flow rate monitor . Gas pressure monitor . Overcurrent protection unit (double check by circuit protector and optical sensor) . Furnace block cooling check

Positioning

. .

. Lateral/vertical manual adjustment AA-7000F (with an optional auto atomizer changer mounted) . Automatic flame/furnace switching by motor

AA-7000F/AA-7000G

Power requirements Dimensions and weight 24

200, 220, 230, or 240 VAC 5%, 6000 VA, 50/60 Hz 260 W 560 D 510 H mm, 46 kg

Autosampler
Common Specifications
Control Maximum reagent / sample positions Rinse water bottle Power requirements RS-232C communication control from AA unit Reagents: 8 positions 2L Choose from 100, 120, 220, or 230 VAC, 50 VA, 50/60 Hz Samples: 60 positions (random access available with reagents or samples)

Flame Continuous Method


Functions Sample volume Nozzle rinse Dimensions and weight Zero-point detection, auto diagnosis, auto rinse, random access Sample vials: 15 mL, reagent vials: 53 mL Solvent rinse aspiration method Sampler: 340 W 280 D 270 H mm, 8 kg Controller (including stand): 260 W 200 D 320 H mm, 7 kg

Furnace Method / Flame Micro Sampling Method


Functions Sample volume Nozzle rinse Sampling functions Syringe Injection volume Repeatability Carryover Mixing port rinse Mixing function Maximum number of mixing reagents added Zero-point detection, auto diagnosis, auto rinse, auto mixing, random access Sample vials: 2 mL, reagent vials: 20 mL Solvent rinse discharge method Dilution function, reagent addition function 250 L 2 to 90 L 1%RSD (20 L) Rinse port: Less than 0.0001 Mixing port: Less than 0.00001 Solvent rinse discharge method, solvent discharge and rinse with next sample Performed in mixing port. Maximum mixture volume: 0.6 mL Up to 4 solutions Number of solutions possible for mixing . Calibration curve method: 5 solutions max. (sample + 4 types of reagents) . Standard addition method: 6 solutions max. (sample + standard solution + 4 types of reagents) For measurement result on unknown samples: . If extrapolation of calibration curve is possible: automatic calculation of dilution rate and dilution to bring concentration within calibration curve range . If extrapolation of calibration curve is not possible: dilution rate fixed at 10x Sampler (including sliding parts): 340 W 280 D 400 H mm, 11 kg Controller: 235 W 110 D 240 H mm (built into GFA-7000), 6 kg

Auto dilution / re-analysis

Dimensions and weight

Required Software Operation Environment


Shimadzu recommends the PC and monitor in the PC Set for AA-7000 Series with WizAArd Software Pre-Installed. Purchase a printer separately. The personal computer and monitor specifications are as follows. Operating system CPU RAM Monitor Storage device I/O port Peripheral device Microsoft Windows Vista Business / XP Professional Intel Celeron 420 (1.60 GHz or higher) 1 GB or higher (Vista) or 512 MB or higher (XP) XGA (1024 768 dots) or higher One CD-ROM drive (for installing a software) Requires 60 MB min. hard disk space for AA installation. One serial port (for AA control) Monitor, keyboard, mouse, printer

* Notes . The PC, monitor and printer are not included in the standard configuration. Prepare them separately. . The ASC-7000, ASK-7000 and GFA-7000 are optional accessories. . The separately available micro sampling kit is required for the flame micro sampling method. . Microsoft Windows Vista Business or XP Professional is required to run the AA-7000 Series software. . Microsoft Windows Vista Business and XP Professional are registered trademarks of Microsoft Corp., U.S.A. . The AA-7000 Series does not conform to IEC60601 Medical Electrical Equipment.

25

Installation Conditions
Dimensions

Wall
200 465 870 200 max. 280 150 to 200 Damper 150 to 200 Exhaust fan 588 150

Table

Damper Approx. 500 500

415 1100 to 1200

415

714

390

340 700 960 960

Both for flame analysis and furnace analysis

900 to 1000

For furnace analysis

Above dimensions do not include PC or printer.

Unit: mm S=1: 2 0

26

Make sure all required equipment is available before starting instrument installation. See the installation guidelines for details.

Example of recommended piping for the atomic absorption system


Stainless steel piping (inner diameter greater than 7 mm) Stop cock Stop cock Setting pressure 0.11 MPa

Operating environment Testing bench Gas tube

Temperature range 10 to 35 C Humidity range 20 to 80% (less than 70% when temperature is higher than 30 C) 1500 mm min. (W) 700 mm min. (D) Withstand load: 200 kg *1, 2 Material Dimensions Acetylene Stainless *3 7 mm min. I.D. 80 mm min. (L) Cylinder: 0.11 MPa Lab: 0.09 MPa Compressor: 0.4 MPa Lab: 0.35 MPa Cylinder: 0.4 MPa Lab: 0.35 MPa Cylinder: 0.4 MPa Lab: 0.35 MPa Stainless For flame: Approx. 500 mm W 500 mm D For furnace: 150 mm to 200 mm For flame: 600 to 1200 m3/hr For furnace: 10 to 180 m3/hr Cooling water circulation unit or tap-water equipment *4

Pressure regulator Pressure range 0.3 MPa Setting pressure 0.09 MPa Acetylene gas Setting pressure 0.4 MPa

Gas supply Air pressure (set secondary Dinitrogen oxide pressure) Argon

Setting pressure 0.4 MPa Pressure range 1 MPa Setting pressure 0.35 MPa Dinitrogen oxide gas

Material Exhaust duct Dimensions Intake capacity

Pressure range 1 MPa Setting pressure 0.35 MPa Mist separator Pressure range 1 MPa Setting pressure 0.35 MPa Lab 80 mm max.

Argon gas

Cooling water (GFA-7000)

Setting pressure 0.4 MPa

*1 Atomic absorption spectrophotometer stand recommended *2 Maintain a free maintenance space of 150 to 200 mm to the sides and rear of the instrument. *3 Do not use pipes containing copper, silver, gold, mercury (or alloys containing these metals) as pipes for acetylene. *4 If tap-water equipment is used, ensure that it meets the specifications below. Compatible faucet Faucet 13 or 12 to 15 mm O.D. rimmed faucet 10 to 30 C 0.6 to 1.5 L/min 0.08 to 0.15 MPa *5 7 mm min.

Outdoor Air compressor

Water temperature Water flow rate Supply pressure Supply port

*5 If the supply pressure exceeds 0.17 MPa, use the optional Regulator ASSY.

27

JQA-0376

Founded in 1875, Shimadzu Corporation, a leader in the development of advanced technologies, has a distinguished history of innovation built on the foundation of contributing to society through science and technology. We maintain a global network of sales, service, technical support and applications centers on six continents, and have established long-term relationships with a host of highly trained distributors located in over 100 countries. For information about Shimadzu, and to contact your local office, please visit our Web site at

www.shimadzu.com

SHIMADZU CORPORATION. International Marketing Division


3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641 Fax. 81(3)3219-5710

URL http://www.shimadzu.com
Printed in Japan 3295-04902-50ANS

The contents of this brochure are subject to change without notice.

You might also like