Manual On Quality Standads For HIV Testing Laboratories Naco
Manual On Quality Standads For HIV Testing Laboratories Naco
Manual On Quality Standads For HIV Testing Laboratories Naco
March 2007
TABLE OF CONTENTS
S.NO. CHAPTERS Table of contents Foreword Acknowledgements List of abbreviations 1. Basics of HIV Structure Transmission Immunopathogenesis Susceptibility Laboratory diagnosis of HIV / AIDS and National HIV testing strategies Introduction Laboratory investigation for detection of HIV infection Causes of false positive and false negative results Strategies/algorithms of HIV testing Detection of p24 antigen Polymerase chain reaction (PCR) Virus culture Viral load assay HIV Testing at counseling and testing sites (ICTCs/VCTCs and PPTCTC's), etc. Introduction Types and Technologies of rapid tests - Immunoconcentration - Agglutination - Immunocomb assay - Immunochromatography - Possible outcomes of these rapid tests Accuracy of rapid tests Strategies/algorithms for rapid tests Uses of rapid tests Quality assurance practices Total quality management Management requirements Technical requirements Laboratory management Laboratory configuration Stock/inventory management of reagents and consumables Data management 1---5
2.
6--15
3.
16--20
4.
21--25
5.
26--27
6.
Quality control Introduction Scientific principles involved Definition of terms Essential components Categories of controls Dispensation and storage of quality control samples Determination of acceptable range of quality control to validate each test run Variations in the quality control validation guidelines Importance of 'E' ratios over OD values of external controls to check validity of runs Interpretation of aberrant result of QC samples: shift and trend Collection, transport & storage of specimens for HIV testing Performing venipuncture Transport and storage of samples\ Standard operative procedures (SOP) The structure of SOP An example of SOP Documentation and document control Selection of HIV test kits ELISA Rapid tests Details of some of the commercially available kits Evaluation of HIV test kits Introduction Preparation of serum panel Composition of serum panel Reference test Testing conditions Parameters for performance characteristics of a kit Selection of HIV test kits Equipment maintenance and calibration Introduction Checklist before deciding on purchase of instrument Maintenance and calibration of different equipment Calendar of activities for maintenance & calibration of equipment Function check Documentation Preventive maintenance Operation of ELISA Reader (Multiskan plus)
28--46
7.
47--50
8.
51--52
9. 10.
53--54 55--59
11.
60--66
12.
67--75
ii
13.
Troubleshooting Introduction Specific assay problems: some examples and their possible solutions Variations in test results Common errors in HIV testing and resolution of the same Vigilance / supervision Record keeping Implementation of external quality assessment programme at different levels of the HIV/AIDS testing network and job definitions of each level Equipment, supplies and reagents for practicing quality control in HIV testing Biosafety practices in HIV testing laboratory Introduction Modes of exposure to blood borne pathogens in the laboratory Essential steps of biosafety Biosafety guidelines for laboratory workers Chemical disinfection commonly used in HIV laboratory Transport of specimens Disposal of laboratory waste Processing of syringes, needles and gloves for reuse Packing, storage and transport of treated (disinfected) waste Final disposal of laboratory waste Container and color coding for disposal of bio medical waste Spills and accidents Post exposure prophylaxis against HBV and HIV Selected list of references consulted Annexures: Annexure I: Consent form for HIV testing Annexure II: Proforma for requisitioning HIV test Annexure III: Worksheet for ELISA Annexure IV: HIV test report form Annexure V A & B: Formats for reporting to SACS and NACO Annexure VI: Proforma for External quality assessment programme for HIV testing Annexure VII: Sample EQA Questionnaire for lab supervisors Annexure VIII: Sample EQA Questionnaire for lab staff Annexure IX: States allotted to the national reference centers Annexure X: Pipetting techniques List of Experts For Revised Manual List of Experts involved in Field Testing of the Original Manual
76--80
14.
81--84
15.
85--86
16.
87--98
17. 18.
99--100
101 102 103 104 105--106 107--111 112--115 116-118 119 120-122 123 124-125
iii
K. Sujatha Rao
Additional Secretary & Director General National AIDS Control Organisation, Ministry of Health and Family Welfare, Government of India
FOREWORD
In order to ensure that the result produced by a HIV Testing Laboratory (ICTC/VCTC, PPTCT, and blood bank laboratory) is reliable, accurate and reproducible, it is essential that laboratory personnel have the knowledge about all aspects of HIV testing and undertake stringent quality control measures every day. There is a vast network of HIV testing laboratories in the country; however there is a need to institute uniform testing and quality control protocols in these laboratories across the Nation especially in view of the wide variety of tests kits used by them. The generation of false positive and false negative results is associated with immense social, ethical and legal implications. It is indeed critical to assure quality in HIV testing so that accurate results are produced every time the test is performed. Manual on "Quality Standards for HIV Testing Laboratories" will provide guidelines for practicing quality assurance required for HIV testing on a day to day basis; training, implementation and establishment of quality assurance programs in all the HIV testing facilities across the country (blood banks, ICTCs/VCTCs, PPTCTs and Blood Bank labs); evaluation and selection of HIV test kits and standard formats for requisitioning, performing and reporting the test results. The guidelines in this manual conform to the recommended quality standards for HIV testing laboratories as well as the needs of the country based on available infrastructure. We appreciate the efforts of all the expert who prepared the document; those who field tested the manual and those who updated the manual.
9th Floor, Chandralok Building, 36 Janpath, New Delhi - 110001 Phone : 011-23325331 Fax : 011-23731746 Email : [email protected]
viuh ,pvkbZoh voLFkk tkusa; fudVre ljdkjh vLirky esa eqr lykg o tkp ik,A
Know your HIV status; go to the nearest Government Hospital for free Voluntary Counselling and Testing.
ACKNOWLEDGEMENTS
The manual entitled Quality Standards in HIV Testing Laboratories is intended for use by laboratory specialists, laboratory supervisors/in charges and technical staff engaged in HIV testing across the country .The manual provides background knowledge of HIV; details of laboratory diagnosis of HIV; step by step practices of quality assurance; HIV testing procedure; evaluation and selection of HIV kits; troubleshooting and biosafety practices. The document also includes samples of standard formats used for HIV testing and reporting. The manual can be referred to for preparing Standard Operative Procedures for any HIV test, practicing quality assurance and producing accurate results. We thank Ms. Sujatha Rao, Additional Secretary and Director General, NACO for giving us the opportunity to revise and update the existing manual. We are extremely grateful to Dr. Jotna Sokhey, Addl. Project Director NACO, Dr M. Shaukat, Joint Director, NACO, Dr. D. Bachani, Joint Director, NACO and to all the contributors and experts and the team at Clinton Foundation who contributed, field tested, and updated this manual to bring the manual to its final form.
The Contributors
Dr. Jotna Sokhey, Addl. Project Director, NACO Dr. M. Shaukat, Joint Director, NACO Dr. D. Bachani, Joint Director, NACO Dr. Sandhya Kabra, Deputy Director, NACO Dr. D. Kasana, former Deputy Director ,R & D , NACO Dr. Preeti Mehta, KEM College, Mumbai Dr. Madhuri Thakar, NARI, Pune Dr. R. K. Lodha, AIIMS, Delhi Dr. Mario Tamburini, WHO, SEARO Dr. Usha K. Baveja, former Lab Specialist, Clinton Foundation Dr. Namita Singh , Lab Specialist, Clinton Foundation Dr. Sanjay Sarin, former Lab Specialist, Clinton Foundation Dr. S. N. Misra, Clinton Foundation
vi vii
vi vii
vi vii
123
124
8.
Dr. V.Ravi
9.
Dr D.K. Neogi
10.
11.
12.
Senior Scientist Department of Blood products & Microbiology, National Institute of Biologicals NOIDA 201301
13.
Dr.Ramesh Paranjape
Office-in-charge National AIDS Research Institute, Plot No. 73, G Block, MIDC Bhosari, Pune 411026
14.
Microbiologist National AIDS Research Institute, Plot No. 73, G Block, MIDC Bhosari, Pune 411026
15.
125
LIST OF ABBREVIATIONS
BTC CCR-2, 3, 5, CXCR4 cDNA CMV CD4 CSF CV DNA EIA ELISA EQA E ratio E/R gp HBIg HBV HCW HCV HIV HIV 1 HIV 2 HIV -I M HIV -I O : : : : : : : : : : : : : : : : : : : : : : : Blood testing centre Co-receptors on T cells/ macrophages for HIV complementary DNA Cytomegalovirus Cluster determinant 4 Cerebrospinal fluid Coefficient of variation Deoxyribonucleic acid Enzyme Immuno Assay Enzyme Linked Immunosorbent Assay External Quality Assessment ELISA ratio ELISA / Rapid Glycoprotein Hepatitis B Immunoglobulin Hepatitis B virus Health care worker Hepatitis C virus Human Immunodeficiency Virus Human Immunodeficiency Virus Type -1 Human Immunodeficiency Virus Type -2 HIV -1 Major group HIV -1 Outlier group
vi
HIV I N HTLV IB NPV PPV NRL SRL OD PCR PEP PHA PI PPTCTC QA QAP RIA RNA SD SOP TQM UV VCTC ICTC WB
: : : : : : : : : : : : : : : : : : : : : : : :
HIV -1 New group Human T lymphotropic virus Immunoblot Negative Predictive Value Positive Predictive Value National Reference Laboratory State Reference Laboratory Optical density Polymerase Chain Reaction Post Exposure Prophylaxis Passive haemagglutination Protease inhibitors Prevention of parent to child transmission centre Quality Assurance Quality Assurance Programme Radio Immunoassay Ribonucleic acid Standard deviation Standard Operative Procedure Total Quality Management Ultra violet Voluntary Counseling & Testing Centre Integrated Counselling and Testing Centers Western blot
vii
Structure
HIV is a 90-120 nm icosahedral, enveloped, RNA virus. HIV comprises of an outer envelop consisting of a lipid bilayer with uniformly arranged 72 spikes or knobs of glycoprotein (gp) 120 and gp 41 in HIV-1 and gp140 and gp 36 in HIV- 2, respectively. Glycoprotein 120/140 protrudes out on the surface of the virus and gp 41/36 is embedded in the lipid matrix. Inside is the p24 protein core surrounding two copies of RNA. This core also contains viral enzymes: reverse transcriptase, integrase and protease, all essential for viral replication and maturation. Proteins p7 and p9 are bound to the RNA and are believed to be involved in regulation of gene expression.
Genetic structure
The genetic structure of virus contains both highly conserved and highly variable regions. The high variability of the virus accounts for drug resistance and evasion from immune response. This also poses problems for development of a successful vaccine. In an infected individual, quasi-species of a particular viral subtype may be found on account of constant variability. As a result many recombinant strains like AE, AG and AB may be present in infected individuals. HIV has structural and regulatory genes coding for structural and regulatory products, respectively. Structural genes direct the synthesis of physical components of the virus and are also responsible for viral size, shape, structural integrity and its compartmentalization in host cell. The regulatory genes direct synthesis of proteins that affect the synthesis of viral components and viral replication. Structural genes are gag, pol and env. HIV also has some accessory genes vpu, vpr and vif, etc. which increase infectivity and budding.
Replication
Glycoprotein 120/140 of HIV binds to a receptor / receptors on HIV permissive host cell. Predominant receptor is the CD4 molecule present on T lymphocytes and macrophages, though others such as galactosyl-ceramide (gal C) have 1
also been proposed. Receptors are molecules (proteins and / or glycoproteins) present on the surface of host cells which facilitate the attachment and entry of viruses into the cell. Entry of virus into the host cell requires certain cellular co-receptors / factors e.g. CCR-5, CXCR-4, CCR-2 and CCR-3, etc. designated collectively as cell infectivity factors (CIF). CIF may be a co-receptor or enzyme helping in virus interaction with host cell. Most convincing candidate is the chemokine receptor related protein, fusin (CXCR-4). Once the gp41/36 of the virus fuses with the host cell membrane, the capsid is uncoated and a ribonucleoprotein complex capable of reverse transcription is formed. During the process of reverse transcription cDNA is formed under the effect of viral enzyme, the reverse transcriptase. Reverse transcription is inefficient in quiescent cells suggesting the involvement of host components in the process. The nucleoprotein complex formed after transcription comprises of linear double stranded DNA, the gag matrix (MA) protein, the accessory vpr protein and the viral integrase (IN). This is called pre-integration complex and is transported into the host cell nucleus. It mediates a complex series of enzymatic steps and integration occurs at cellular loci with open chromatin structure. Integration probably is an essential step for viral replication. The integrated virus is called provirus. The virus may not be expressed in many cells and is considered latent. Virus expression can be stimulated by many viral, cellular and exogenous factors. Other co-existent viral infections e.g. cytomegalovirus, herpes virus etc. can make the non permissive cells permissive. Maturation of virus also takes place after virus assembly and budding.
Immunopathogenesis
HIV enters the circulation and targets onto cells expressing CD4 on the surface in various organs including brain and lymphoid tissues. Infected CD4 lymphocytes migrate to lymph nodes, become activated, proliferate and start releasing HIV. The antigen driven migration and accumulation of CD4 cells within the lymphoid tissues may account for dramatic decrease in CD4 cells and generalized lymphadenopathy characteristic of acute retroviral syndrome. Cellular and humoral responses are established within one week to three months (window period) and host enters the phase of clinical latency characterized by lack of symptoms and almost initial normal levels of CD4cells. There is gradual deterioration of the immune system due to depletion of CD4 cells. CD4 cells are destroyed by HIV mediated cell lysis; formation of syncytia and multinucleated giant cells (infected as well as uninfected cells are destroyed); virus specific immune responses; superantigen mediated activation of T cells making them permissible to HIV and programmed cell death (apoptosis). The ability of HIV to replicate in T cells depends on the state of activation of cells. So, a vicious cycle is established leading to depletion of CD4 cells, destruction of lymphnode architecture and progressive increase in the viral load. The progression of HIV disease also depends upon the biological properties of the infecting HIV strain (high vs. low replicating, syncytium vs. nonsyncytium forming virus). The HIV disease progresses fast in infants infected in utero because of the immature immune system and rapid spread of HIV in the body. Most cases vertically infected have a shorter incubation period and a more rapid, fulminant disease process.
Transmission
Risk factors for transmission of HIV include: unsafe sexual practices with multiple sexual partners, recipient of multiple 2
blood transfusions, use of infected/unsterilised needles or syringes (whether therapeutically or for recreational drug use), transmission from an infected mother to the foetus/infant before, during, or shortly after delivery and during breastfeeding by an untreated mother who is infected (perinatal factors that increase transmission include prematurity, maternal STI, episiotomy, and prolonged rupture of membranes and others). The efficiency of transmission of HIV is determined by the amount of virus in a body fluid and the extent of contact. High concentration of free infectious virus and virus infected cells have been reported in blood, genital fluids and cerebrospinal fluid. Breast milk and saliva yield varying numbers, whereas, other body fluids have a low viral content. Saliva in adults contains some non-specific inhibitory substances like fibronectins and glycoproteins which could prevent cell to cell transfer of virus. Thus, saliva is not a likely vehicle of transmission. Urine, sweat, broncho-alveolar lavage fluid, amniotic fluid, synovial fluid, faeces and tears have been reported to yield zero or few HIV particles. Hence, these vehicles also do not appear to be important in virus transmission, though standard work precautions should be undertaken while handling all the body fluids Breast milk at the time of primary infection in a feeding mother has a high content of virus and may transmit the infection to the baby. In contrast, cerebrospinal fluid (CSF) also has a high content of virus particularly in individuals with neurological disease; CSF is not a natural source of virus transmission. Table1 Efficiency of different routes of HIV transmission and their contribution to total number of cases reported in India. Exposure route Percent efficiency * Percentage of total Worldwide Blood transfusion Perinatal Sexual intercourse Vaginal Anal Oral Injecting drugs use Needle stick exposure Exposure unspecified
* see references Percentage of total in India from NACO website
India 2.5
7.3
4.2
The most efficient vehicle of HIV transmission is blood (Table 1). However, the risk of infection via blood transfusion is now extremely low due to strict HIV screening of donated blood. The most common route of transmission is unprotected, penetrative sexual contact. Different forms of sexual practices carry a variable risk gradient of acquiring HIV. Cell associated rather than free virus is responsible for disease transmission. Anal intercourse carries a high risk of transmission because of bowel mucosa which acts as a portal of entry for virus, and also because of a greater chance of injury to the mucosa. Risk to insertive partner is through infection of lymphocytes and macrophages in the foreskin or along the urethral canal. In females, HIV transmission occurs when infected cells in the semen gain entry 3
into the female genital tract, and infect the resident lymphocytes, macrophages and probably the uterine epithelial cells. The transmission from infected mother to child appears to occur in 20-40% children born to HIV positive mothers without PPTCT intervention. The source of virus in the new-born is controversial. HIV infection can occur via amniotic fluid, genital secretions, maternal blood and through the breast milk. Transmission to the baby can occur inutero, and during or after delivery. Transmission of HIV infection to Health Care Professionals (HCP) is extremely uncommon. Pooled data from 20 prospective studies suggests that risk associated with needle stick injury from HIV infected blood is approximately 0.3 percent. Further, the risk associated with mucocutaneous contact is too low to be reliably estimated. The risk from mucosal or non-intact skin is also minimal. Transmission of HIV through casual contact, sharing utensils, lavatories and through insect bites has not been documented. Prospective studies offer a conclusive evidence that family members and close household contacts of HIV infected individuals are not at risk of acquiring HIV infection through casual human contact (shaking hands, kissing and by sharing of utensils, toilet, bed linen etc.) or by providing routine nursing care. Mosquito bite does not transmit HIV.
Susceptibility of HIV
Sterilization Autoclaving at 121C at 15 lb pressure for 20 minutes Dry heat 170 C for 1 hr. (holding time) Boiling for 20 minutes (this kills HBV and HCV also ).
Sensitivity
It is the accuracy with which a test can confirm the presence of an infection in an infected individual. Tests with high sensitivity show few false negative results and are meant to be used to screen blood prior to transfusion and ensure blood safety. It is recommended that 3rd or 4th generation HIV test kits which are 100% sensitive should be preferred for use at blood banks for screening donated blood. 6
Specificity
It is the accuracy with which a test can confirm the absence of an infection in an uninfected individual. Tests with high specificity show few false positives and are to be preferred for the diagnosis of HIV infection in an individual.
(iii)
Screening tests
ELISA/EIA: (Enzyme Linked Immunosorbent Assay/Enzyme Immunoassay) It is the most commonly performed test at blood banks and tertiary labs to detect HIV antibodies. There are various kinds of ELISA based on the principle of test: Indirect ELISA Competitive ELISA Sandwich ELISA Immuno Capture ELISA ELISA is also classified on the basis of the antigens utilized, into: 1st generation: Infected cell lysate is used as the antigen. 2nd generation: Glycopeptides (recombinant antigens) are used as the antigen. 3rd generation: Synthetic peptides are used as the antigen. 4th generation: Antigen and antibodies are detected simultaneously. The assays may use a combination of recombinant and synthetic peptides as antigens. When a serum sample tests reactive once by a system of ELISA / Rapid (E/R) test, the test is to be repeated immediately by a different system in order to confirm the diagnosis at VCTCs and PPTCTCs. If it tests reactive a second time, the sample is to be taken up further for supplemental tests to confirm the diagnosis. Supplemental tests are E/R based on different principles of test or different antigen systems. Sometimes, WB, etc. may have to be used to resolve the sero-discordant results. ELISA takes up to three hours to yield results. It has a major advantage of being economical. Although rapid tests give results within minutes, these are relatively expensive per test. Commercial kits are available for both ELISA and rapid tests. Rapid tests include: Dot blot assays (immunoconcentration based) Particle agglutination (gelatin, RBC, latex, microbeads) Dip stick and comb tests, etc (EL ISA technology based). Immunochromatography based tests (lateral flow of reagents) Tests which detect antibody to HIV 1 and 2 and all the subtypes of HIV1 and HIV2, are to be employed
Supplement tests
Second and third ELISA/Rapid Western blot (WB)
WB is expensive, time consuming and requires expertise to perform. This is mostly done to confirm the diagnosis on samples which given discordant results in E/R and in legal cases.
multiple myeloma, primary biliary cirrhosis, alcoholic hepatitis, chronic renal failure, positive RPR (rapid plasma regain) test and false negative results have been reported in cases of window period prior to seroconversion, immunosuppressive therapy, malignant disorders, B-cell dysfunction and bone marrow transplantation, etc.
Strategy I (algorithm I)
Blood/plasma/serum is subjected once to E/R for HIV. If negative, the sample is considered free of HIV and if positive, the sample is taken as HIV infected for all practical purposes. This strategy is used for ensuring donation safety (blood, organ, tissues, sperms, etc.). Units of blood testing positive by the screening E/R are destroyed as per guidelines. A donor who gives consent to know the result of HIV test is informed about the result. In case the test is positive, the donor is informed about the possible reactive nature of the result and is advised that the same requires confirmation. The individual is accordingly referred to VCTC for counseling, testing and confirmation of the test result.
Strategy II A (algorithm II A)
A serum sample is considered negative for HIV if the first ELISA or rapid (screening) test is negative, but if reactive, the same sample is subjected to a second ELISA or rapid test which utilizes a system different from the first one. i.e. either the principle of test and/or the antigen used is different. It is reported reactive only if the second ELISA/ rapid test confirms the positive report of the first test. In case the second E/R is non reactive, then the result is taken as negative for surveillance purposes (II A). Strategy II A is used for sentinel surveillance (anonymous, unlinked testing).
Strategy II B (algorithm II B)
This strategy is used for detection of HIV infection in symtomatic individuals with symptoms of AIDS clinically. The sample is processed as in Strategy IIA, but a sample reactive with first assay and non-reactive with second assays subjected to the third tie breaker E/R. If the third E/R is reactive, the samples reported as indeterminate and patient is called back for repeat testing after 2-4 weeks. If the third test is negative, it is reported as negative. Two to three different test kits with different antigen system or different principle of the test are required to follow this Strategy IIB of testing. In this strategy if the first two consecutive tests are positive for presence of HIV antibodies, a positive report can be given to the patient.
A1 + Consider Positive2 (Destroy the unit of blood as per guidelines refer to VCTC for confirmation of status after consent) 10
A1 Consider Negative
Strategy/Algorithm II A
A1 +
A1 Report Negative
A21
A1 + A2 + Report positive
(Diagnosis of an individual with AIDS indicator disease symptoms.) 3 Test kits required. A11
A1 +
A1 Report Negative
A21
A1 + A2 -
A31
A1 + A2 - A3 + Indeterminate
3
A1 + A2 - A3 Report Negative 11
Strategy/Algorithm III [To detect HIV infection in asymptomatic individuals (VCTCs, PPTCTs) ] 3 Test kits required
A11
A1 +
A1 Report Negative
A2
A1 + A2 +
A1 + A2 -
A31
A1 + A2
A3 +
A1 + A2 + A3 Indeterminate3 A31
A1 + A2 - A3 + Indeterminate3
A1 + A2 - A3 Report Negative
The screening test selected preferably should be fourth generation and the sensitivity of the test should be 100%. The supplemental/confirmatory tests selected should be >99.8% sensitive and >98.5% specific. Highly sensitive tests reduce the occurrence of false negative result and fourth generation HIV tests reduce the window period to 2-3 weeks, thus playing an important part in ensuring blood safety. Highly specific tests reduce the occurrence of false positive result and thus ensure that a person testing positive as per the strategy II/ III is actually HIV infected.
1 2
Assays A1, A2, A3 represent 3 different assays Such a result is not adequate for diagnostic purposes: use strategies IIB or III. Whatever the final diagnosis,
donations which were initially reactive should not be used for transfusions or transplants. Refer to VCTC after
3
informed consent for confirmation of HIV status Testing should be repeated on a second sample taken after 14-28 days. In case the serological results continue to be
indeterminate, then the sample is to be subjected to a Western blot /PCR if facilities are available or refer to the National Reference Laboratory for further testing. 12
EIA for HIV-1 antigen detects primarily uncomplexed/dissociated p24 antigen, in serum, plasma, CSF or cell culture. It indicates active infection and allows diagnosis before seroconversion. Quantitative test can predict prognosis and may be useful for monitoring response to therapy. Disadvantages of antigen detection assays include: poor sensitivity (only 69% in patients with AIDS and low in neonates < 1 month old); detection is not possible in patients with high titers of p24 antibody (which complexes with the antigen); and failure to detect HIV-2 antigen.
Virus culture
Virus culture is another method for identifying HIV infection. Positive culture rates of up to 98% are reported in confirmed seropositive individuals. The culture method is, however, expensive, labour-intensive, can take weeks for complete results and potentially exposes laboratory workers to high concentrations of HIV. Virus culture is used for research (drug sensitivity, vaccine studies, etc.) only.
therapy and predicting progression and clinical outcome. Because baseline HIV viral load is predictive of survival at 10 years in patients with nearly identical CD4 counts, assessment of baseline viraemia prior to initiation of therapy may be useful in patient management.
Details about enumeration of CD4 cells and the uses of this marker in management of HIV infected individuals are given in Guidelines on enumeration of CD4 cells using single platform technology.
Mandatory testing
When testing is done without the consent of the patient and when the data could be linked to identify the person, it is called mandatory testing. Mandatory testing is recommended only for screening donors of semen, organs or tissues (to ensure transplantation safety) in order to prevent transmission of HIV to the recipient of the biological products.
15
CHAPTER 3 HIV TESTING AT COUNSELLING AND TESTING SITES ( VCTCs AND PPTCTCs) USING RAPID TESTS
Introduction:
Rapid tests are in vitro qualitative tests for the detection of antibodies to Human Immunodeficiency Viruses (HIV) types 1 and 2 in human serum, plasma, whole blood saliva and urine. Currently HIV testing in India is performed on serum/whole blood (fingerprick), and plasma. This is because the HIV testing on urine and saliva samples has not been evaluated and validated in India. In recent times, a large number and wide range of rapid tests of high quality have become available and are being currently used in laboratories under the following conditions: Facilities to perform ELISA test are absent, In emergency cases Remote blood banks where the collection volume is low and the facilities for ELISA are not available (ELISA is the test of choice for HIV testing at blood banks) . Point of care settings like VCTCs, PPTCTCs, and also intervention healthcare sites, etc.
Immunoconcentration:
This technology is available in the form of dot blot assays. These are multi step tests, very rapid to perform and results are obtained within a few minutes. The antigens (recombinant / synthetic peptides) are passively blotted on the nitro cellulose membrane matrix which is bound to the solid support and contains the adsorbent pads to collect the serum and reagents after their addition. Inbuilt control (internal control) in the form of a spot or line to indicate the validity of the test is available with each test.
Agglutination:
Agglutination assays are easy to perform and require no wash procedures. The antigens are adsorbed passively on the carriers (red blood cells, latex particles, gelatin particles and or micro beads). The antibody present in the serum sample reacts with the antigen adsorbed carriers, resulting in clumping of agglutinated particles. However, if in the reaction mixture the concentration of antigen and antibody is not optimum, phenomenon of prozone reaction (antibody excess) occurs which may lead to a false negative reaction. To overcome this initial dilution of the test sample is to be done as recommended by the manufacturer. Reactions are read visually. The control (uncoated or un sensitized particles) is run to detect non specific agglutination. If agglutination is detected in both the control and the test samples the assay needs to be repeated. Examples of these tests are Capillus and Serodia. 16
Reactive
Latex Aggregation White Clumping
Non-reactive
No Latex Aggregation no white clumping
Immunocomb assays:
The immunocomb is a rapid test intended for the qualitative and differential detection of antibodies to HIV-1 and HIV-2 in human serum or plasma. The immunocomb is an indirect solid phase enzyme immunoassay (EIA). The solid phase is a comb with 12 projections. Each tooth has 3 spots: Upper spot: Goat antibodies to human immunoglobulin (control to validate the test) Middle spot: HIV-1 synthetic peptides Lower spot: HIV-2 synthetic peptides The developing plate has 6 rows (12 wells each) with each row containing a reagent solution ready for use at different steps in the assay. Immunoglobulin present in the testing samples is captured by the anti human immunoglobulin antibody on the upper spot (internal control). Unbound components are washed away. IgG from the sample is captured on the teeth and reacts with antihuman IgG antibody labeled with alkaline phosphatase which react with chromogenic components and the results are seen as gray blue spots on the surface of the teeth of the comb Interpretation of controls: Appearance of upper spot- Negative control Appearance of all 3 spots- Positive control Upper spot does not appear - Invalid result Interpretation of results: Appearance of upper spot - Non reactive sample Appearance of upper and middle spot- Reactive for HIV-1 Appearance of upper and lower spot- Reactive for HIV-2 Appearance of all 3 spots - Reactive for HIV-1 and HIV-2 Post-assay activities: After the assay, discard the test devices in the box containing 1% sodium hypochlorite Discard the used filter paper into the bio hazard bags Swab the workbench and all equipments after use with 1% sodium hypochlorite followed by 70% alcohol 17
Immunochromatography:
These are lateral flow one step tests. These tests are usually temperature stable and can be stored at the room temperature. For HIV testing the rapid tests are used to detect the HIV antibodies. Examples of kits based on this technology are Determine, Unigold and Hemastrip.
Reactive
2 lines of any intensity appear in both the control and patient areas.
Non-reactive
1 line appears in the control area and no line in the patient area.
Invalid
No line appears in the control area. Do not report invalid results. Repeat test with a new test device even if a line appears in the patient area.
Possible outcomes of these rapid tests( Immunoconcentration, Immunofiltration, Dipsticks and Immunocombs) : Reactive or positive : Appearance of test band and control band Non reactive or negative : Appearance of control band only Invalid : Absence of control band. Test has failed. Repeat the Test with a new device
Advantages:
No equipment is required to perform the test. Micropipettes may be needed to dilute the sample for some assays eg. agglutination assays Limited infrastructure is needed. Some of the rapid tests can be stored at room temperature. These tests have wide temperature range stability. Rapid assays can be used in remote peripheral labs and for circumstances where same day results are required e g VCTCs and PPTCTs 18
The rapid assays are versatile as far as sample stability is concerned. These can be used on whole blood, serum, plasma, saliva or urine. Different kits are available for use on these different samples Most of the rapid tests are one to five step procedures Reading is subjective and visual with naked eye examination. However now rapid test readers are available for some tests, allowing us to perform semiquantitative analysis. Rapid immunoassay results can now be read objectively using CCD chips (as in digital cameras). Rapid tests are now more robust Management of waste generated by performing rapid tests is easier to manage as compared to that generated by performing ELISA Rapid tests work out more cost effective overall particularly in set ups like that in VCT and PPTCTCs as they are more robust, there is lesser cost of the requisite lab infrastructure and lesser cost for sample collection
Disadvantages
More expensive because of single test packaging and performance. Test performance may vary by the product. Refrigeration required for some of the products. Each test card cannot be quality controlled with an external quality control sample. There may be issues of sensitivity, specificity, negative predictive value and positive predictive value in relation to the reference gold standard. These should be carefully considered while procuring the product.
19
20
Technical requirements
Personnel Civil Infrastructure Environment Equipment Sampling Handling of test and calibration items Assuring quality of test and the results Reporting of results Biosafety and safe waste disposal
The organization provides the managerial and technical personnel who have the resources and authority needed to carry out the assigned functions i.e. HIV testing in this case. The organization ensures that policies and procedures are in place right from the collection of the sample to the reporting of results. The quality system practiced is able to identify the occurrence of departure from standard procedures and errors and takes corrective actions as and when required. The organization is responsible for assigning responsibilities, ensuring smooth inter relationship of personnel who manage, perform and verify the laboratory activities. Organization also provides supervision and has overall responsibility for technical operations and provision of resources (equipment, space and supplies) and manpower to carry out quality HIV testing.
Quality system
Quality System also dictates that there will be a quality manual which gives details of all processes and procedures to be followed for HIV testing; a biosafety manual which will give details of standard work precautions to carry out the procedures and tests in the lab and guidelines on safe disposal of waste.
Document control
The documents form a part of the quality system. There are a number of documents generated in the laboratory ranging from the policy documents, procedure documents, worksheets, instruction sheets, equipment documents, reports and records, etc. A system has to be in place to control all these documents. The documents should be numbered and available with the organization. All the documents are available and accessible to the laboratory personnel. Documents with controlled access should be clearly listed. Authorized, current documents should be available at all locations where operations are performed. The old documents should be archived as per the laboratory norms for historical reference. Documents on procedures, equipment, etc. should be revised from time to time as required. Details on documentation are given in the relevant chapters .
Maintenance contracts
All equipment in the laboratory must be under the annual maintenance contract (AMC) so that there are no interruptions in the functionality due to sudden breakdowns. AMC should be inbuilt in the procurement process.
Internal audit
Internal audits should be conducted periodically and in accordance with a predetermined schedule and procedure to verify and ensure that the laboratory procedures fulfill the requirements of the quality system. Internal audit serves to identify any lapses/errors in the system which can be corrected by corrective actions. The internal audit report should be tabled and discussed with the lab administration/ management so as to ensure adequate addressal of the nonconformances raised during the audit. 22
Complaints
There should be a procedure and policy in place to resolve the complaints received from a client. These may include delay in reporting the result, rude behavior of personnel, break in confidentiality and wrong result, etc. There should be an investigation following the complaints, the responsibility should be fixed and corrective action taken to resolve the complaint. A note on corrective action taken should be made in the records.
Preventive action
Preventive Action is a pro-active process to identify opportunities for improvement to prevent breakdown in equipment, work and quality system.
Corrective action
A problem with the quality system or with technical operations of the laboratory may be identified during internal/ external audit, management reviews, feedback from clients or staff observations. The corrective action has to be taken to resolve the problem. For this, an investigation to identify the cause of problem should be carried out by analysis of all potential sources of the problem. Many a times this may be the most difficult task. However once the root cause/ causes of the problem are identified corrective action should be taken (such as repair of an equipment, training of staff), documented and effect of corrective action must be monitored
Civil infrastructure
The laboratory should have adequate space to carry out the procedures, house the equipment and space for personnel. There should be enough lighting and all power requirements should be met. Surfaces should be smooth and disinfectable. There should be a pest control system in place. Working surfaces preferably should be acid resistant.
Environment
A clean environment should be provided for the laboratory. The temperature and humidity should be maintained as per the requirements of the testing procedure and the equipment maintenance etc .The environmental conditions should not adversely affect the required quality of test or invalidate the test. Smoking, eating and drinking should be prohibited in the laboratory and signs to this effect must be displayed. 23
Equipment
It must be ensured that the equipment needed for the required testing is available and functional. The list of equipment required for the HIV testing is given in the relevant chapter . The equipment should be maintained as per the instructions of the manufacturer and calibrated as recommended. The equipment should be operated by authorized staff and the documents on equipment should be a part of document control. The details to be documented about equipment include name of equipment, name of manufacturer, serial number or any other unique identification, record of checks as per specifications, manufacturer's instruction manual, and the contract for maintenance, malfunction modification/ updating, etc. The procedure for calibrating the measuring equipment should be available. Other details on equipment procurement, maintenance and use are given in the chapter on equipment maintenance and calibration.
Sampling
Sampling is a defined procedure whereby part of a substance, material or product is taken to provide for testing or calibration of a representative sample of the whole. The sampling procedure should describe the selection, sampling plan, withdrawal and preparation of a sample from a substance, material or product to yield the required information. For example before the HIV test kits are imported in the country ,the Drug Controller General of India lifts a sample of kits for evaluation by the statutary institute. .
The report should include patients ID, name and address of the laboratory, mention of the sample, name of the test carried out, name of the test performed, name of the person who performed the test and the name of the person who reviewed the test result. All results should be reported with appropriate units and with applicable reference range.
Biosafety
It is essential to have the guidelines on standard work precautions in the laboratory. The bio safety practices are extremely important to prevent transmission of blood borne viruses to the technologist. These precautions should be practiced diligently at all times while working in the laboratory and/or providing other services in the health care facility. The bio safety practices are detailed in a separate chapter.
25
Laboratory Configuration
The key to a reliable HIV testing lab is for it to be optimally configured. This should involve the following: Equipment layout: The lay out of the equipment should be in line with the testing requirements of the lab. This may vary for a lab with high or a low testing throughput. This would include layouts for refrigerator, water bath, incubator, centrifuge, ELISA reader, etc. Workflow staging and direction: Work areas should be arranged to allow uni-directional sample flow and defined space for each test step. Defined areas are needed for: Specimen collection Specimen receiving and storage Specimen preparation Specimen testing-instrument Result production, validation and release Reagent and consumable storage Electrical requirements: These should be in line with the testing requirements of the laboratory. Staffing requirements: Decisions on staffing should give due consideration to the required level of expertise in terms of the type of laboratory and the tests to be performed. It should also take into account the number of trained staff required on the basis of the operational volumes.
26
Data management
Proper record keeping of patient results is vital for providing optimal patient care and gaining knowledge from patient data collected. A well defined data management system should: Ensure reliable and rapid delivery of results to clients / clinic sites Ensure clinics have systems for receiving and processing result data Ensure the laboratory maintains records of result data for defined periods Use standard reporting formats Ensure that dedicated human and other resources for data management are assigned
In addition, the data management system should work to ensure that: Laboratories examine all specimens accepted to ascertain that they meet the proper criteria for data entry and processing Laboratories verify that the results are for the intended patient Laboratories enter and process data correctly, to ensure that the results are given out in a correct and efficient manner. This can be achieved in the following manner : Results should be entered on both the worksheet and patient result form Worksheets should be filed by date in the laboratory for easy result retrieval Patient report form results should be entered and processed. All patient reports should be signed by the concerned technologist / technician and the same should be verified by the lab head. The release of results in the correct form is as important as conducting the test. The results should only be handed over to the concerned individual after confirming all the antecedents / particulars of the concerned individual & posttest counseling. Confidentiality should be maintained in all respects. Another important aspect of data management system is the archiving of results. Archives can be be electronic or paper-based. It should use a consistent system for data storage based on one or more variables (collection date, clinic site, patient ID) that allows easy reference and retrieval of records.
27
28
Quality assurance
Quality assurance is defined as the means by which the laboratory ensures that the final results reported are correct and are achieved in a standard, reproducible and traceable manner. Quality assurance helps ensure avoidance of mistakes; consistency of performance; data integrity; efficiency and cost effectiveness; customer satisfaction and laboratory esteem and credibility. Quality assurance practices are important to acquire standards to national and international accreditation.
Quality assessment
Quality assessment also known as proficiency testing and external quality assessment (EQA) is a means to determine the quality of results generated by the laboratory, which involves incorporation of proficiency panels prepared by an external agency into routine testing and analysis of the results produced after the testing of this panel. The process is known as the External Quality Assessment Programme. These results are then assessed and compared with the standard results, a performance score is developed and feed back is given to the participating laboratories. The aim is to assess the performance of the laboratory for specific tests, identify the factors responsible for incorrect results (material, processes, performance) and try to correct these through corrective actions and training. Specimens of EQAP (external quality assessment programme) must be handled in the same way as a specimen is handled for diagnostic purposes in the laboratory. The EQA process involves the following elements:Gathering information from the participant laboratories regarding the kits/assays being used. Preparation of serum panel. Distribution of panel Collection of results from the participating laboratories Collation and analysis of results Sending the analysis report to the participating laboratories Taking corrective actions for the participating laboratory performing poorly on EQA panel through locating the problem and resolving the same. Issuing participation certificate to all the participating laboratories.
Essential components
The following essential components of quality control must be performed during every assay: Each test run must include one full set of controls that should yield results within the limits of the standard for acceptability and validity of the run. The values of controls obtained for the first run of the day cannot be used for subsequent runs. The controls have to be included in each run. Any run not having at least the minimum recommended number of controls falling within the acceptable range is invalid and should be repeated. All test kits should be used before the expiry date. Physical parameters of the test such as incubation time and temperature must be followed as per manufacturer's instructions.
External controls
These are a set of controls included from outside i.e. not belonging to the kit. They are applied for each run/assay to continually monitor assay performance (validity). A model Quality Control Programme that incorporates the external controls has the following main components: Making an external control with desirable anti HIV titer available Determination of acceptable ranges Inclusion of external controls in each run Data collection Analysis Reporting
The steps for preparing a borderline reactive external quality control sample are:
Selection of a high titer HIV positive plasma/serum. Serial dilution of the high titer HIV positive sample using normal HIV negative human serum. Selection of a suitable sample dilution to achieve the desired titer for use as external control with borderline reactivity. Batch production: preparation of bulk external (low positive) control sample. Batch validation. Dispensation and storage. Validation after storage.
Centrifugation is carried out at 1500 rpm for 20 min. The serum is separated aseptically. A small volume of serum is tested for HIV antibodies by screening and confirmatory tests (as well as for HBV and HCV). The bulk (i.e. remaining quantity of serum) can be stored in smaller and
convenient volumes at -20C, after labeling. If plasma is more than 7 days old, thrombin should be added to the plasma before re-calcification.
Selection of diluents
The recommended diluent is normal human serum rather than normal saline or other sample diluents. This is to keep the antibodies in natural serum protein environment. Adequate quantity of serum for use as diluent can be obtained from a pool of normal healthy individuals (HIV, HBV and HCV negative) or from the unit of blood negative for HIV, HBV and HCV. In the later instance, plasma has to be recalcified to obtain serum as per the procedure mentioned earlier.
Discard 1 Tube No 11 Positive serum Diluent Final dilution (reciprocal) Log2 dilution 2 3 4 5 6 7 8 9 10
200ml N 0
100ml 2 1
100ml 4 2
100ml 8 3
100ml 16 4
100ml 32 5
100ml 64 6
100ml 128 7
100ml 256 8
100ml 512 9
100ml 1024 10
32
Fig. 2 Conducting EIA with the dilutions of (in triplicates) HIV positive serum.
1 A B C D E F G H
CP = positive control (kit controls); CN = negative control (kit controls) 1, 2, 3...12 = Indicate sample (pre-diluted) numbers Note: Controls in the plate map are displayed as per the protocol of Innotest ELISA.
2 1
5 2
10
11 1
12
CP CP CP CN CN CN
3 1 2 3 3 2
coated antigens, etc. However, such factors influencing the OD values of the controls would also expectedly influence the OD values of the test samples in similar direction. Thus the relative reactivity of a given test sample in relation to cut off value would not vary much. This relative reactivity of a test sample in relation to cut off value in a particular run is expressed and termed as E ratio. This is the ratio between the sample OD and cut off OD. Hence it is more appropriate to express the degree of ELISA reactivity in terms of E ratio rather than individual OD values as explained. Following are the examples of two unknown samples, designated X and Y, run on different dates using kits from same manufacturer and same batch (Table 1 and 2).
1.
Indirect EIA
Innogenetics NV, Belgium Biochem Immuno System Inc, Canada United Biomedical Inc., USA Cambridge Biotech Ltd., Ireland
2.
Detect HIV
Indirect EIA
0.160 - 0.240
3.
Indirect EIA
0.125 - 0.205
4.
HIV-1
Indirect EIA
0.393 - 0.518
Table - 2 Variations in the EIA values of two samples X and Y run in different EIA kits
Sample number X 1 Y X 2 Y X 3 Y 0.037 0.032 1.778 0.518 0.071 0.029 1.512 0.417 0.076 3.432 0.363 0.066 3.625 Run day Sample OD 1.262 Cut off OD E ratio 3.476
As evident from above example, the variation of OD values expressed as 'E' ratio is much less compared to OD value above.
Fig. 3 EIA reactivity (E ratio) of HIV-1 positive serum by indirect EIA Innotest HIV-1/HIV-2, (Innogenetics kit) in relation to serial dilutions in normal serum.
10 9 8 7 6 o 5 4 3 2 1 0 1 2 3 4 5 6 7 8 9 10 11
Solid line indicates the cut-off limit for interpretation of result Arrow indicates suitable dilutions for preparation of external control
Procedure
HIV positive sample and diluent are filtered to remove bacteria and fungi. However, filtration of recalcified serum is often difficult as filters tend to clog by fibrin clots. Hence the positive sample should be centrifuged, prefiltered with a 0.8m biological filter to remove aggregated proteins or debris and then filtered through a 0.22 m biological filter to remove contaminating bacteria, although there will be inevitable reduction in volume of the valuable specimens selected as controls. The positive sample is centrifuged and filtered (0.22 m) followed by heat inactivation at 56C for 20 minutes. Positive sample and volume of diluent i.e. normal human serum containing preservative (e.g. Bronidox) is mixed in the volumes required to yield the required titer (as described above) and kept for 16-20 hours at 4C on a magnetic stirrer. The required titer is ensured by running EIA for 10 runs for 10 days. The batch thus produced is aliquoted into large polypropylene storage containers (e.g. 250 ml). Polyethylene containers are not recommended as these absorb antibodies and thus may affect the titer of the QC sample. 35
An I.D number is assigned to the lot with date of production mentioned. All containers to be used to produce the batch should be autoclaved.
Table - 3 An example of values obtained while testing batch variation of 25 batches of external control
36
Mean
The mean (expressed as X) is calculated by adding individual observations (X1+X2+X 3------Xn or X) and then dividing by the number of observations (expressed as n). Thus mean is calculated as follows:
In the data shown in table, the mean 'E' ratio calculated as the sum of individual 'E' ratios divided by number of aliquots is found to be as follows:
Standard deviation
Standard deviation (SD) is a measure of the variation or dispersion of observations about the mean and defines the expected range of a control in relation to the mean value.
37
Coefficient of variation
The coefficient of variation (CV) is expressed as percentage and the following formula is used.
Fig. 4 Batch validation (inter-aliquot variation) of 25 aliquots of external control sample; coefficient of variation (CV) is calculated to be 6.82%.
A Levy-Jennings chart in which E ratios are plotted against the Y axis and the individual samples (aliquots) plotted against the X axis are shown in Figure - 5. 38
Fig. 5 Levy-Jennings chart to plot a QC graph on the basis of values ('E' ratios) obtained in 20 consecutive days run of external control.
Coefficient of variation less than 10% is considered to be an indication of little inter aliquot (batch) variation and thus consistency can be expected in the performance of various aliquots as external quality control in daily run.
Procedure
The external QC sample is tested in at least 20 runs (e.g. in 20 consecutive days) to make statistically significant observations. 39
The mean and standard deviation of 'E' ratio of the external controls are calculated (as per method mentioned above) in each of the 20 runs separately. The coefficient of variation (CV) of the external controls is calculated on the basis of mean and standard deviation of the 'E' ratios.
It is important that coefficient of variation (CV) of the 'E' ratios observed on different dates is minimal (i.e. <20%). An example of values obtained in 20 consecutive runs of external controls is shown in Table - 4.
Data analysis
As per the standard methods of calculations the following observations are made about the above mentioned runs of EIA: Mean ( X) Standard deviation (SD) Limit of Mean 1SD Limit of Mean 2SD Coefficient of variation (%) = = = = = 1.90 0.19 1.71-2.09 1.52-2.28 11.6% 40
Having assured the minimal batch variation (inter aliquot variation) i.e. CV < 20%, as tested in at least 20 consecutive runs, the external control is considered to be suitable for inclusion as external control in subsequent daily runs of EIA.
'E' ratio =
The observed 'E' ratio is matched with the Levy-Jennings chart prepared (as described above).
Interpretation
If the 'E' ratio of the external control falls beyond 2 SD limit of mean in Levy-Jennings chart, the run of EIA is treated as quality control failed or invalid even though the internal kit control values are within acceptable limits. On the other hand, if it falls within the limit mentioned above, the run is considered as valid.
41
According to some authors, setting up the acceptability limit as Mean 2 SD is likely to lead to perceptible degree of rejection due to random error. Hence, they recommend the extension of the limit of SD to + 2.5 instead of 2 with the assumption that 5% or less of any set of control values will fall outside the range due to chance.
A relatively more stringent measure of quality control recommends incorporation of external controls (or even internal controls) in triplicate at 3 random wells in the ELISA plate in each run to check intrarun variation.
Importance of 'E' ratio over OD value of external controls to check validity of runs
There are two methods by which the reference QC graph may be represented. The first one is to plot the OD values of the external control on the Y axis and the consecutive run members (i.e. run dates in case one plate is run per day) on the X axis. The second option is to plot the 'E' ratios i.e. ratio between the OD value of external control and the cut off value on the Y axis instead of the OD value of the external control alone, the determinants on the X and Y axis remaining as in the first method. The second method is considered to be more accurate alternative since any general change in the conditions during the run of EIA on a specific day/run will affect the OD value of the external control as well as kit controls towards a similar direction (explained in details previously in section 2). This is exemplified in the following example of six EIA runs on consecutive dates using permissible limit of (i.e. mean 2SD) in the values of 20 runs mentioned in example for setting the limit of acceptability of external controls for validity of runs. The acceptability limits of external controls for validation of test results calculated as below (based on OD values of 20 runs above). Mean (X) of OD SD of OD Limit of Mean 1SD Limit of Mean 2SD = = = = 0.428 0.055 0.373 - 0.483 0.318 - 0.538
On the other hand, as shown in Table -4, the following values are obtained if the 'E' ratios in the same runs of EIA are taken into consideration to validate the run. Mean (X) of 'E' ratio SD of 'E' ratio Limit of Mean 1SD Limit of Mean 2SD = = = = 1.90 0.19 1.71 - 2.09 1.52 - 2.28
Now, let it be presumed that the following six consecutive runs of EIA are evaluated for their validity based on the performance of external control Table 5.
42
It can be observed from Figure - 6 and 7 that when the OD values of the external controls on different days of test run are matched against the acceptability limit (mean 2SD) drawn on the basis of OD values of the external control alone, the run on day 2 appears to be invalid since the observed OD value on that particular day (i.e. day 2) lies beyond acceptability limit i.e. more than mean 2 SD limit or 0.538). On the other hand, if the corresponding 'E' ratios are matched against the acceptability limit (i.e. mean 2SD ) drawn on the basis of 'E' ratios of the external control, the run on day 2 appears to be within the acceptability limit, while the run on day 4 appears to be invalid (Figures - 6 and 7.).
Fig. 6 Validity of runs evaluated on the basis of OD values of the external control alone. Cut off OD values are shown at the bottom as broken line
Fig. 7 Validity of runs evaluated on the basis of ELISA ratios (E ratio i.e. external control OD/ cut off OD) of the external control.
If the OD values are further examined for the runs in question, it is clear that, on day 2 although the OD value of the external control went up, there was some degree of rise in cut off OD value as well. Thus the net change in the ratio between the external control sample OD and cut off OD (i.e. 'E ratio) did not fall outside the acceptability range when 43
matched in terms of 'E' ratio. On the other hand, on day 4 of run, there was some degree of elevation of external control OD value while cut off value did not show any such change rather it showed change on reverse side which is not as per expectation (as explained in earlier section). Thus there is a rational ground to consider the run on day 4 as invalid Table - 3 & 4, Figure - 6 and 7
Fig. 8 Interpretation of aberrant results of QC: shift and trend Illustrations of shift and trends
Trend
Six successive points distributed in one general direction (towards either higher result or lower result) within the acceptable range. Can be due to the following: Deteriorations of reagents Slowly faltering equipments e.g. pipette
The date of expiry of kit/reagents and/or any indications of deterioration/or contamination. The performance of the operator of the test to ensure the assay is being performed correctly. Whether or not the aberrant results were produced with a particular batch of kit.
Flow CHART FOR PERFORMING HIV TESTING USING QUALITY CONTROL ALIQUOT Performance of EIA with external control (HIV)
45
FLOW CHART FOR PERFORMING RAPID HIV TESTING USING QUALITY CONTROL ALIQUOT
* *
Licensed Quality checked As per manufacturers guidelines Internal controls are checked Include one positive external control daily Examine for either coloured dot/dots, or coloured line/lines and/or agglutination Ig spot should always give positive result. Positive control should give positive result (colored dot, line and/or agglutination
* *
Inclusion of external control in daily run Examine cartridge, comb, etc. for result
* *
-Ig spot not seen -Positive control result does not show as positive
46
Performing venipuncture:
Gloves should be worn and sterilised /disposable syringes and needles should be used. For avoiding soiling, a piece of linen with a layer of dressing pad ( a sheet of absorbent cotton between two layers of gauze piece) or simply a big piece of absorbent cotton may be placed below the forearm before commencing veni-puncture. After collecting 3- 5 ml of blood aseptically, it should be carefully transferred from the syringe without squirting into a sterile plastic leak proof specimen container preferably screw capped. The containers should be labelled before commencement of venipuncture. The cap may be tightly screwed after the blood has been transferred to the vial. After blood is collected, the tourniquet is removed and the needle is withdrawn. The patient is given a dry sterile cotton swab to press over the site of venipuncture. Elbow may be flexed to keep the cotton swab in place till the blood stops. Any blood spill is carefully wiped with 70% ethanol/10% bleach solution. All the swabs and cotton pieces are placed in plastic bags for disposal. If the outside of the vial is visibly contaminated with blood, it should be cleaned with 10% freshly prepared sodium hypochlorite solution. The blood is allowed to clot for 30 minutes (not more than 2 hours) at room temperature. The clot may be gently broken if necessary using sterile pasteur pipettes. Alternatively vacuum based blood collection systems (vacutainers) can also be used. These vacuum based collections are relatively safe for usage and harbour minimum risk of unwanted exposure to infected blood.
47
The specimen vial is un-stoppered, the serum is drawn off by sterile Pasteur pipette and transferred to a sterile plastic screw capped leak proof tube.
Addition of preservative:
The usual preservative should not be added since it inactivates conjugates and gives rise to false serological results. If necessary, 5 bromo, 5 nitro, 1-3 dioxane in propylene glycol at a final concentration of 0.05% is recommended as preservative. Thiomersal at a final concentration of 0.01% is effective only for a few weeks as it loses activity when exposed to light.
Sample transportation
These instructions are recommended for specimen transportation: The shipment of infectious agents is regulated by the Transportation of Dangerous Goods Act and the International Air Transport Association (IATA) dangerous goods regulations. HIV infected specimens are classified as infectious class 6.2 substances under the United Nations (UN) no. 2814. The packaging must adhere to UN class 6.2 specifications. Packaging requires a 3 layer system as described below (see Fig. 1 for a diagrammatic representation): The specimen tube, in which serum is to be transported, should not have cracks / leakage. It should preferably be made of plastic and be screw capped. The outside of the container should be checked for any visible contamination with blood which should be disinfected. Place the tube containing the specimen in a leak-proof container (e.g. a sealed plastic bag with a zip lock or alternatively the bag may be stapled and taped) and pack this container inside a cardboard canister / box containing sufficient material (cotton gauze) to absorb all the blood should the tube break or leak. Cap the canister/box tightly. Fasten the request slip securely to the outside of this canister. This request slip should have all details i.e. name, age , sex , risk factors, history of previous testing, etc. and should accompany the specimen. The request slip should be placed in a plastic ziplock bag to prevent smudging on accountsof spillage. 48
For mailing, this canister/box should be placed inside another box containing the mailing label and biohazard sign.
The diagram ( fig 1) depicts the method of sample transport for a single/ few ( 2-3) samples that could fit into the secondary container shown in the diagram. The size of the primary sample container will vary with the number of samples being transported. For a larger number of samples ,a tube rack (or some such container ) may be used wherein the samples can be transported in the upright position and at appropriate temperature. The packaging instructions for the transport of a larger number of samples are given below: The specimen should be carefully packaged to protect it from breakage and insulated from extreme temperature Label appropriately and mention the test/s being requested for that sample. The collection site should make use of a unique identification number as sample identity. Names of the patients should be avoided to prevent confusion on account of duplication of names as well as to maintain confidentiality. Secure the vacutainer cap carefully and seal it further with sticking tape ( placed so that it covers the lower part of the cap and some part of the tube stem. During packaging, the tubes containing specimens should be placed in a tube rack and packed inside a cool box (plastic or thermocol) with cool/ refrigerated / frozen gel packs ( as appropriate to keep the sample at the recommended temperature for the test ) placed below and on the sides of the tube rack. Place some cotton or other packaging material between the tubes to ensure that they do not move or rattle while in transit.. Cool box required for transportation could be a plastic bread box or a vaccine carrier. Seal/secure the lid of the cool box . This cool box should then be placed in a secure transport bag for purposes of shipping to the testing facility. The request slips should be placed in a plastic zip lock bag and fastened securely to the outside of the cool box with a rubber band and sticking tape. A biohazard label should be pasted on the visible outer surface of the package containing the samples. The package must be marked with arrows indicating the 'up' and 'down'side of the package Samples should be transported to the receiving laboratory by commercial courier or be hand delivered by a trained delivery person. The collection site must have prior knowledge of the designated testing days of the laboratory to which the samples are being sent. No transport should be done during weekends and holidays or non-testing days of the testing laboratory unless prior arrangement has been made with the receiving laboratory.
Note: Use overnight carriers with an established record of consistent overnight delivery to ensure arrival of specimen within the specified time.
49
50
SOP should be available at testing site and should be followed every time the test is performed. The SOP must be simple to follow and should be in the form of flow diagram / chart, one step leading into the next and easy to interpret. SOP must also include all the pre-analytic, analytic and post-analytic requirements for performing the test keeping the quality issue in view.
A sample SOP
Name of the test procedure. Intended use: detection of HIV-1 and HIV-2 antibodies. Principle of the test procedure: details as per the manufacturer's kit insert have to be given. Details of the kit contents like conjugates, negative and positive controls, plate / cartridge / comb etc. List the materials which are required for performance of the test but are not provided in the kit. Storage instructions.
51
The precautions to be taken while performing the test, standard work precautions and any other precautions specific to the test.
Specimen collection and storage of whole blood, serum or plasma; How the serum / plasma should be separated; How the specimen is to be used and how it should be stored.
Quality controls required to be put to validate the tests. Details of the test procedure as per the instructions in the kit insert. Details about how to validate and interpret the results as per instructions of the kit insert. Limitations of the test in detecting infection. Standard work precautions. Details about disposal of left over samples and other wastes generated. Details about recording the result and reporting.
52
Types of documents
Customer complaint document. Corrective and preventive action document. Equipment maintenance, calibration, and monitoring document (details in chapter on equipment maintenance and calibration). Equipment inventory document. Reagent inventory document. Safety manual. List of suppliers of equipment, reagents and other materials used in the laboratory. Documents detailing the contracts to perform special activities (e.g. research projects, etc). Samples of contracts to be stored. Documents on standards being used by the organization.
For example, corrective and prevention action document will detail what preventive action needs to be taken to avoid breakdown of an equipment. Preventive actions are defined as the proactive processes to identify the potential sources of non-conformance. This is discussed in details in the chapter on equipment maintenance and calibration. The corrective actions taken to repair equipment, improve a procedure, improve performance, etc. must also be documented.
Document control
Certain documents need to be controlled i.e. such documents may not be accessible to everyone in the lab. All such documents which need to be controlled should be clearly specified and the accessibility must also be defined. These may include financial records of the lab and all those documents which the management feels are essential to be controlled. At the same time, control of documents should be exercised in a manner that it does not interfere with the quality management system of the lab and should not affect the quality of the result.
54
The performance characteristics to be considered for the selection of kits are as under:
Sensitivity Specificity Efficiency Positive predictive value Negative predictive value Stability Ease of performance of test.
The parameter's values should be within those recommended by the technical expert committee, G.O.I. on this subject. The final approval of the test kit is given by DCGI on the basis of the evaluation reports submitted by the identified evaluation centres Sensitivity: Specificity: Efficiency: Predictive values: Sensitivity of a test is defined as its ability to detect truly infected individuals as also its ability to detect very small amounts of analyte. This is the ability of a test to correctly identify all the uninfected individuals i.e. there should be no false positives. It is the overall ability of a test to correctly identify all positives as positive and all negatives as negative. This is the measure of value of a test in relation to the prevalence of the disease in the population. The value of a test in addition to the parameters described above depends on the population being tested. Positive predictive value (PPV) is the ability of the test to identify actually infected individuals among all persons giving a positive result with the kit being evaluated. Negative predictive value (NPV) is the ability of the test to correctly identify the really non infected individuals from among all the persons giving negative result. 55
If the prevalence is high, the PPV of a test will be high i.e. an individual testing positive will mean actual infection. Whereas, in low prevalence area the chance that an individual testing positive is actually infected is lower.
The selection criteria 2-9 are identical to those applicable as above for HIV (ELISA) test its general specifications
The selection criteria 3-9 are identical to those applicable as above for General Specifications
6. 7. 8. 9. 10. 11.
Adequate literature dealing the components, methodologies, validity criteria and performance characteristics of the product should be provided with each kit. Should have sensitivity level at 99.8% and above and specificity level at 98% and above. The supplier/local agent should have facility for storage of kits at 2C - 8C The total procedure time should not be more than 30 minutes. Provision shall be made for conducting single test at a time. The tests should be easy to perform and the kits should preferably be stable at room temperature (22 25C.
The selection criteria 3-11 are identical to those applicable as above for HIV Rapid test kits (general specifications)
The selection criteria 3-11 are identical to those applicable as above for HIV Rapid test kits (general considerations)
HIV Rapid test kits - any other principle excluding Enzyme Immuno Assay and Agglutination
1. Should be solid phase/particle coated with synthetic/recombinant HIV I & HIV-II antigens. The assay should detect antibodies to HIV I & HIV II by any other principle excluding Enzyme Immuno Assay and Agglutination. The selection criteria 3-10 are identical to those applicable as above for HIV Rapid test kits-general specifications
2.
Immunofiltration*
Immunoassay
57
Agglutination
Different manufacturers use different raw materials as matrix for adsorbing the HIV peptides. The immunofiltration involves the vertical flow of the reagents. As the reagents pass through the matrix the reactants get adsorbed and concentrated (immunoconcentration) on the viral peptides on the matrix in case of HIV positive sample to produce a colored spot /line
ELISA TESTS
58
Examples
I. II. III. IV. Tridot/Pareekshak/ HIV Spot + Immunocomb HIV 1& 2Bispot/HIV EIA Comb/CombAIDS RS + Capillus/NEVA Tridot/Pareekshak/ HIV Spot + Immunocomb HIV 1& 2Bispot/HIV EIA Comb/CombAIDS RS + SD Bioline HIV 3.0/ Retrocheck HIV/ Precise Tridot/Pareekshak/ HIV Spot + Capillus/NEVA + SD Bioline HIV 3.0/ Retrocheck HIV/ Precise Immunocomb HIV 1& 2Bispot/HIV EIA Comb/CombAIDS RS + Capillus/NEVA + SD Bioline HIV 3.0/ Retrocheck HIV/ Precise Care has to be taken that atleast two kits selected should be able to differentiate between HIV 1 & 2 For example: Do not select Pareekshak + Comb AIDS RS + Capillus as none of these kits can differentiate between HIV 1 & 2, instead Tridot + HIV EIA Comb + Capillus can be selected
59
As far as the first category is concerned DCGI asks the importer to get the representative sample of kits evaluated, using a locally made and well characterized panel by the Institutes identified for this purpose. Currently NIB is the statutory institute which perform evaluations. If found suitable, only then the kits should be allowed to be imported for distribution, both in Government as well as private sector. In second case, when the kits are manufactured indigenously, the performance characteristics of the kits should be ascertained by evaluating different batches of kits (the final product) with the standard serum panel for evaluation of kits at least at two to three different geographical sites and consistent results should be obtained. Only the kits performing to the standards (sensitivity and specificity, etc.) as laid down should be licensed for marketing. In the third case, the licensed, and/or approved kit's performance may be questioned particularly when different centers report inconsistent results. This can happen due to break in cold chain during transport and storage of HIV kits. To resolve this problem the kits have to be evaluated by the identified centers to ensure optimal performance. This chapter gives the basic guidelines to be followed to ensure proper evaluation of HIV test kits.
Information about the performance characteristics (sensitivity, specificity, predictive values, etc.) of the kit is ascertained from the manufacturer. The information helps to select the right type of evaluation panel. The evaluations can be done by a single site or multiple sites. In case of multi-site evaluations, a QC sample should be included to assess inter-laboratory variations.
Evaluation panel
A serum panel is defined as a collection of well characterized serum samples for a specific reference/purpose. Serum panel can be prepared and used for different purposes. These are: Panel for evaluation of kits i.e. evaluation panel including seroconversion panel Panel for proficiency testing (external quality assurance) Panel for quality control in the laboratory (internal quality control)
Aliquoting of serum
Repeat freezing and thawing of specimens affects the quality of the panel. To avoid this, samples should be aliquoted in volumes that may meet the particular panel requirement at one time. Suitable anti-bacterial agents e.g. Bronidox (0.05%) should be added to panel samples to prevent contamination.
Storage
Aliquots must be stored at or below -70C. Glass vials should not be used as they may eventually crack with long term storage in frozen condition. 61
Transport of panel
To provide safety for laboratory staff, couriers and community, specimens should be transported under International Air Transport Association (IATA) regulation, for this purpose. Specimens should be packed, labeled and documented as classed under 6.2 UN 2814 to minimize the risk of infection to anyone during transit.
Reference test
A well established, reference gold standard test must be available to the laboratory for comparison, the results of which should correlate with the actual status of the patient. Currently, such a test in serologic testing for HIV infection is considered to be Western blot (WB) test. However, sometimes WB may give indeterminate results in which case more sophisticated tests like synthetic peptide based line immunoassay, polymerase chain reaction (PCR) and virus culture can be undertaken. The tests are expensive, labor intensive and require higher level of technical expertise. Thus for the moment WB usually is regarded as the Reference/Gold standard test. When a number of kits are being evaluated the tie breaker for false positive/false negative is usually a WB test, sometimes supported by more sophisticated tests as mentioned above.
Testing condition
All evaluations must be performed under identical conditions. The tests must be conducted in the same laboratory using same equipment (pipettes, instruments, etc.), by the same technician, using kits which have not expired and have been stored properly and using the exact protocols provided in the package insert.
62
The technicians must not know the results obtained on the panel with other kits i.e. the evaluation testing is done blindly.
Sensitivity
Sensitivity of a test is defined as its ability to detect truly infected individuals as also its ability to detect very small amounts of analyte. Sensitivity can be calculated by the following formula: True positives Sensitivity = ----------------------------------------------------- X 100 True positives + false negatives Or Number of samples detected as positives by the kit under evaluation ----------------------------------------------------------------------------------------------X 100 Total number of positives by the reference test employed
Sensitivity =
Example Total sera tested= 100 HIV infected = 10 HIV noninfected = 90 Results with the reference test
Results with kit under evaluation: 9 out 10 HIV infected detected as positive 9 Sensitivity of the kit under evaluation will equal to --------- X 100 = 90% 9+1
Specificity
This is the ability of an assay to correctly identify all the uninfected individuals i.e. there should be no false positives (reactive result in the absence of disease). The specificity can be calculated by the following formula: True negatives ------------------------------------------------ X 100 True negatives + False positives
Specificity =
63
Specificity =
Or Number detected as negatives by the kit under evaluation -------------------------------------------------------------------------------------- X 100 Total number of negatives detected by reference test employed
Example Total sera tested= 100 HIV infected = 10 HIV noninfected = 90 Results with the reference test
The results obtained with kit under evaluation, 9 positive out of 10 HIV infected and 1 positive out of 90 HIV noninfected. 89 Specificity of the kit under evaluation will equal to -------------X 100 = 98.9 % (approximately) 89+1
Efficiency
It is the overall ability of a test to correctly identify all positives as positive and all negatives as negative. Efficiency is determined as below: True positives + True negatives Efficiency = ----------------------------------------------------------------------X 100 True positives + False negatives + True negatives + False positive Or Total number detected as positives + negatives by test kit ---------------------------------------------------------------------------------------------------- X 100 Total number of positives + negatives detected by reference test employed
Efficiency =
Example Total sera tested= 100 HIV infected = 10 HIV noninfected = 90 Results with the reference test
The kit being evaluated detected 9 out of 10 HIV infected as positive and one out 90 non-infected also as positive. 9+ 89 So efficiency of the kit will equal to ------------------- X 100 = 98% 9+1+89+1
Delta value
This is a statistical measure of sensitivity and specificity and is calculated from the actual OD values generated by the test. This also helps to determine how close the O.D. value of false positive and or false negative is to the real positive and or real negative value, respectively. If these values are too close the test is flawed. Delta values are defined as the distance of the mean O.D. ratio of the sample population from the cut off (C.O.) measured in standard deviation units.
64
Delta values =
Mean O.D. / Cut off ratio (log 10) -------------------------------------------------------- X 100 Standard deviation
Positive delta value is calculated on mean of ODs of positive samples and negative delta value is calculated on mean ODs of negative serum samples. When two tests are being compared, the test that has a higher positive delta value would characterize positive samples more accurately; similarly the test with higher negative delta value will better classify the negative samples.
Example
Positive predictive values This is the measure of value of a test in relation to the prevalence of the disease in the population. The value of a test in addition to the parameters described above depends on the population being tested. Positive predictive value (PPV) is the ability of the test to identify actually infected individuals among all persons giving a positive result with the kit being evaluated. True positives -------------------------------------------------- X 100 True positives + False positives
PPV =
Negative predictive value (NPV) is the ability of the test to correctly identify the actually non infected individuals from among all the persons giving negative result. True negatives NPV = -------------------------------------------------- X 100 True negatives + False negatives If the prevalence is high the PPV of a test will be high i.e. an individual testing positive will mean actual infection. Whereas, in low prevalence area the chance that an individual testing positive is actually infected is lower. Example Low prevalence area 1% Total sera tested = HIV infected = HIV noninfected =
1000 10 990 65
Results obtained with kit under evaluation: Positive results obtained in 8 individuals, two from HIV infected group and 6 false positives from HIV non infected group. Negative results obtained in 992 individuals, 984 from the non infected group and 6 false negatives from HIV infected group. 2 ------- X 100 = 25% 8 984 ------------- X 100 = 99.3% 992
The PPV =
High Prevalence area 10% Total sera tested = 1000 HIV infected = 100 HIV noninfected = 900
Results obtained with HIV kit under evaluation; 105 tested positive, 95 from HIV infected group and 10 from HIV non infected group i.e. gave false positive result. Negative result was obtained in a total of 900 cases. 892 of 900 were from HIV non infected group and 8 were from HIV infected group. 95 --------------- X 100 = 90.5% (approx.) 105 892 --------------- X 100 = 99.12% 900
PPV =
NPV =
All the above parameters are ascertained by the laboratory carrying out the evaluation of kits. The parameter values should be within those recommended by the Technical Expert Committee, G.O.I. on this subject. The final approval of the test kit is given on the basis of the evaluation reports submitted by the identified evaluation centers. For further information on test kit evaluation, see also: Guidelines for Assuring the Accuracy and Reliability of HIV Rapid Testing: Applying a Quality System Approach. 2005. Centers for Disease Control/WHO.
66
Turn the speed control slowly up and down. Stop the centrifuge immediately if it makes an abnormal noise. After use, the buckets should be inverted to drain dry. After any sample spillage, wipe and disinfect immediately. Clean the centrifuge at short intervals (preferably daily) because it is one of the most frequently used instruments. Check mounting and replace if necessary. Check brushes and bearings every 3 months. Replace if necessary. Check for corrosion and clean and repaint if necessary. Calibration: use a pre-calibrated tachometer to check centrifuge speed.
Refrigerators
The following general advice may be helpful for maintenance: Refrigerators must be so placed that sufficient air can pass the condenser (at the back of the Daily Checks Check temperature daily. It should not exceed 12C. Application of battery driven mobile or stationary thermometers is recommended, preferably those including continuous printing or plotting of temperature measured when heatsensitive reagents are stored for long periods. Monthly Checks Clean cool chamber and defrost the evaporator monthly. Clean refrigerator from outside. Clean condenser of dust. Clean door gasket. refrigerator) for exchange of heat and also to facilitate cleaning of the condenser. The refrigerator door must seal perfectly to prevent warm outside air from entering the cool chamber. Calibration: Use a pre-calibrated thermometer to ensure accuracy of temperature check.
Autoclave
Bacteria cannot survive in an autoclave environment; however, sterilization does not necessarily kill viruses. It should be borne in mind that autoclaves need careful handling and must be regularly inspected. They can be hazardous and can seriously injure a person with hot steam accidentally escaping from the instrument. The main factors influencing perfect steam sterilization are saturated steam temperature time Use of Autoclaves Prepare the material for autoclaving with indicator paper. Fill the bottom of the autoclave with demineralized water up to the support. Before placing the material in the autoclave, ensure that there is an adequate amount of water. Close the lid and make sure that the rubber washer is in its groove. Screw down the clamps firmly. Open the air outlet valve. Turn on the heating (electric element). Do not leave the autoclave unattended. Close the outlet valve when a constant jet of steam starts to escape. Reduce the heat as not to heat too quickly. Once the expected temperature is reached, reduce the heat to maintain the temperature. Do not touch drainage tap, outlet or safety valve while heating under pressure. When the time required for sterilization has passed, turn off the heat completely. When the temperature falls below 1000C, open the outlet valve slowly. Do not leave the outlet valve unopened for too long. Never unscrew the lid clamps or open the lid except after the whistling sound stops. Leave the sterilized material to cool before its removal from the autoclave. Check the autoclave tape (used in the preparation of the material to be sterilized) which must have turned black and the covering paper brown (not yellow or black). Calibration: use a pre-calibrated pressure gauge to check accuracy of pressure in the autoclave.
Adequate sterilization by autoclaves and dry ovens should be monitored by the weekly use of a biological (spore suspension) or chemical indicator.
69
Incubators
Incubators should be subjected to continuous recording of temperature. However, if it is not possible, the temperature must be recorded every day and before opening of the incubator. The incubator should have a fan to maintain uniform temperature. Incubators are used for bacterial culture by laboratories working in microbiology. The incubator must maintain a constant temperature of 37 1 C. Temperature should be daily recorded in incubators. Like all laboratory instruments, incubators must be cleaned and disinfected routinely at short intervals (at least every fortnight) and after spilling of infectious material. The actual temperature must correspond to the thermometer control when the instrument is used. Calibration: use a pre-calibrated thermometer to ensure accuracy of temperature checks.
Water bath
It is important that the water bath maintains a constant temperature within a narrow range ( 0.5C) when used for kinetic measurements such as determination of enzyme activities or clotting assays. Inadequate adjustment and insufficient stabilization of the temperature will strongly affect the results of kinetic measurements. Use of water baths The level of water in the water bath must be above the level of the solution to be incubated. Open containers, vials or tubes must not be incubated in a water bath with an open lid to avoid contamination and dilution of the incubated material by condensed water. The water bath must be refilled regularly to prevent growth of algae and bacteria.
Pipettes
Each pipette, whether manual, semi-automated or automated, must be tested periodically to determine if it is delivering the correct volume. Pipettes are used for volumetric measurements of liquids. They may be made of glass or plastic and may have different volumes, ranging from 2 mL to 100 mL. They may be calibrated or non-calibrated. Glass pipettes are either calibrated to contain (In) or to deliver (Ex). The variety of designs for these devices is so numerous that it is difficult to discuss the subject systematically. It must be noted, however, that the replacement of broken conventional glass pipettes can be very costly, and that there may be other economic advantages in using mechanical pipettes. Mechanical pipettes are preferably used to pipette small volumes (4L to 1000L); they have a higher precision and are less subject to errors in pipetting than glass pipettes. Volumes less than 500L should be pipetted with mechanical pipettes. However, mechanical micropipettes can only be recommended where a reliable supply of standardized disposable tips can be guaranteed. They are usually of air displacement (indirect), or direct displacement design. To avoid contamination during pipetting samples, most designs use a disposable tip which is discarded after each delivery. The practice of washing and reusing disposable tips is to be discouraged, as any cleaning procedure will change the wettability of the plastic. In addition, drying even to only slightly elevated temperatures may distort the tip. This will prevent a good pneumatic seal with the pipette body and change the volume of liquid to be pipetted. 70
After each use, the mechanical pipettes must be kept in an upright position and thoroughly cleaned at periodic intervals. Pipettes are precise and important basic instruments of the laboratory, and as such, need to be calibrated at least every 3-6 months. Several methods of pipette calibration are available (see annexure X)
pH meters
A pH meter needs to be standardized before each run with a standard buffer of pH 7.0. However, in instances when the work is related to a pH range of less then 6.0, it is advisable to use a standard buffer of pH 4.0. The buffer solution should be checked monthly with another pH meter and discarded if the pH deviates by more than + 0.4 or if the buffer is contaminated with microorganisms. Glass electrodes must be stored in buffer solutions at pH 4 to pH 8. The buffer solution must be regularly changed at short intervals. Glass electrodes which have been stored for longer periods must be soaked in 0.1 molar HCI for at least four hours. Thereafter they must be carefully washed with distilled water. The same procedure must be applied for dried-up electrodes. The life-span of properly maintained glass electrodes is about two years. Thereafter they should be replaced. Aging of an electrode is indicated when the constant electrode potential does not develop 20 seconds after insertion of the electrodes into the ion solution. Glass electrodes are sensitive to mechanical damage. Calibration of pH Meters To obtain a precise measurement of pH, the pH meter must be calibrated with two different buffers at pH 4 and pH 7 every day (two-point calibration). For the calibration of a pH meter special buffer solutions must be used, the pH of which should be near the pH of the solution to be measured. Phosphate buffers and acetate buffers are preferable. Calibration measurements should be done using plastic containers. Use of the pH Meter Switch off the measuring circuit Wash the electrode with deionized water Standardize the electrode at the start of the day with two standard buffers (pH 7.0, pH 4.0) Transfer the electrode to a small beaker containing the standard buffer, pH 7.0 Measure the pH of the standard buffer and adjust the buffer control knob to reading Wash the electrode and gently wipe with fibre-free material. Transfer the electrode to another beaker containing the standard buffer, pH 4.0 Read the pH of the standard buffer. It should be very close to expected pH value (pH 4.0) Use fresh standard buffer every day. Discard standard buffer if it is contaminated or cloudy
Measuring the pH of the test solution The following steps must be taken: Wash the electrode with deionized water and transfer it to the test solution. With proper electrodes the potential establishes 5 to 20 seconds after insertion of the electrodes into the sample solution. Air bubbles at the electrodes must be avoided since they cause a drift of the electrode potential. Compensate for the temperature at each run. Measure the temperature and adjust the temperature dial to the reading. 71
Do not read the pH before the reading is stable. Record the pH reading of the pH solution. Wash the electrode and store it in a container with storage solutions.
Balances
Balances are used to measure weight and mass. The principle of weighing is based on attracting forces between two separate masses. In daily life this is the attracting force between mass and the earth. In laboratories, weighing is an essential step in preparing defined quantities of reagents and reaction mixtures. There are two main categories of balance Mechanical balances Electromagnetic balances (battery driven or connected to the mains)
The following factors influence weighing and can cause errors in measurement Temperature Moisture (atmospheric humidity) Electrostatic effects Magnetism Gravitational forces Air
Maintenance of balances The following guidelines are worth remembering: The balance should be placed on a solid vibration-free surface, free from dust and at even temperature, away from sunlight. The instrument must be placed in an exactly horizontal position. The balance should be zeroed prior to each use. Use the smallest possible vessel for weighing. Avoid weighing in vessels made of plastic, because they can become electrostatically charged. Use instead glass vessels or weighing paper, as applicable. The weighing vessel and the sample to be weighed should be at the ambient temperature. Never put your hand into the weighing chamber so as not to warm it up. Place the sample to be weighed in a weighing vessel in the middle of the weighing pan, to avoid corner-load error. Liquids or powders should never be directly weighed on the pan. The weight of the weighing vessel needs to be determined prior to placing the substance to be weighed. A tweezer is useful as a substitute for hands in the weighing chamber. Use clean weighing vessels. Keep the working place, weighing chamber and weighing pan clean. To avoid any possible corroding effect of chemicals, any spillage must be cleaned immediately. Biological materials may be a source of infection. Disinfection can be done with 70% alcohol. After completing weighing return the balance to zero weight. Keep the working place at the balance as clean as possible. Weight of the material to be weighed should be within the range of the balance.
72
Function checks
It is essential that laboratory personnel know and document that all equipment is in good condition each day of use. This can be accomplished by undertaking function checks, often referred to as calibration and validation. 73
Calibration That process which is applied to quantitative measuring or metering of equipment to assure its accurate operation throughout its measuring limits. Validation The steps taken to confirm and record the proper operation of equipment at a given point of time in the range in which tests are performed. Documentation The assurance that a piece of equipment is operating properly can best be judged by examining its performance over time. Records of performance parameters, therefore, are a vital element in the proper operation of laboratory equipment. Some suggested information is provided below: Name and serial number of instrument Elements to be checked and kind of data to be collected Frequency of checking Record of data Changes made to restore accuracy and precision, if any Signature with date of the person performing these tasks.
Preventive maintenance
Maintenance of equipment is an extremely important function in the microbiology laboratory. Unfortunately, this is often grossly neglected because of indifference on the part of laboratory workers and on the erroneous belief that it is too costly. The expense of such maintenance policies as inspection, lubrication and adjustment of instruments is insignificant when compared with the cost of emergency repairs, rebuilding or overhauling equipment, and the additional personnel time and materials involved in producing test results when equipment is down. Preventive maintenance is defined as a programme of scheduled inspections of equipment and instruments resulting in minor adjustments or repairs for the purpose of delaying or avoiding major repairs and emergency or premature replacements. It provides the following advantages over breakdown maintenance. Better quality results Identification of components showing excessive wear Greater safety Fewer interruptions in services Lower repair costs Less stand by equipment requirements
ELISA Reader
Regular maintenance of the ELISA machine should be carried out according to the manufacturers' instructions and may vary between brands. The machine itself should not be opened.
74
Note: The filters must be protected from moisture and fungal growth. Keep Silica gel packet in the filter box.
Plate washer
Plate washers are critical in ELISA assay performance. The washer works on a simple principle and comprises of wash fluid, waste fluid reservoirs, pressure and vaccum pumps, a dispense manifold and a plate carrier. The wash fluid is pressurized and a valve opens and allows the fluid through a manifold and into assay microwells. The waste fluid under vaccum is aspirated back through the manifold to the waste container. A number of cycles of dispense and aspiration comprises the washing of a plate. At the end of the wash procedure the wells are empty of the fluid. After use care Fill the rinse bottle with about 500 ml of distilled water. Dispose off the unused wash buffer. Rinse with distilled water, a couple of times and leave about 500 ml in the wash bottle. Fix the cap tightly. After using the washer switch off power.
75
CHAPTER 13 TROUBLESHOOTING
Introduction
In spite of being vigilant and following the SOP, many a times a test run fails. The reason may be apparent in some cases but not in others. Troubleshooting refers to measures undertaken to determine why a run has failed.
General approach to trouble shooting Factors responsible for some common problems / errors
Specimen: lysed blood, lipaemia, deterioration, volume not sufficient. Clerical/transcription: wrong name/ID no/wrong result transcribed, labeling, etc. Kit related: conjugate gone bad (deterioration of reagents), storage not proper and inter lot variations, etc. Laboratory variations: due to use of different kit, different testing procedure (SOP), different technique. This kind of variation may be inter-laboratory as well as intra-laboratory. So, results are non-reproducible. Environmental conditions: higher temperature, humidity. Equipment problems: ELISA reader, washer, calibration of pipettes, water bath and refrigerator, etc. Technical errors: carelessness, not following SOP, pipetting errors, worksheets not maintained. Borderline reactor/grey zone reactors. Laboratory does not practice quality control procedures.
2)
If the reagent indicator cells or carrier particles in an agglutination assay appear clumped or otherwise heterogeneous - this reagent may have been contaminated or may have deteriorated. Most package inserts describe the appearance of the reagents contained within the kit, and thus ensure comparison of the actual appearance with what is expected.
Specimen problems
1. 2. Error Insufficient volume for repeating the test in case of reactive result. Haemolysis can cause false positive or false negative result in ELISA and background noise on spot test. Lipaemia causes pipetting error, otherwise it causes no error in the test result. Bacterial contamination degrades antibody and will affect the borderline reactive result. Resolution Collect appropriate volume (3-5 ml) of blood depending upon the objective of testing and strategy of testing. Repeat sample. In case it is not possible record "sample haemolysed" in the result. Buy widebore pipettes and pipette tips. Record "sample lipaemic" in the result. Refrigerate the samples for short storage (one week) and store at -20C for long storage(>one week). 77
3. 4.
5. 6.
Freeze/thaw cycle may affect the borderline positive samples. Aliquoting errors
Avoid freeze-thawing as far as possible till the test are performed. All aliquots should be labeled appropriately. A uniform method of storing samples/aliquots should be followed in the laboratory. Organise the specimen in racks and leave a tube space empty between aliquots made from different specimens.
Clinical/clerical errors
Error 1. 2. Logging in specimen : there may be mixing of names, wrong history, wrong number, etc. Result print out : the well number on worksheet and the print out not matched properly (transcription error). Errors during translating of results from the worksheet to the report form. Reporting to the wrong person or result communicated without post test counselling and confidentiality. Resolution Set up the log books/records properly ; collect second sample from the same patient and perform testing. Transcribe the result from the print out sheet to the worksheet carefully. Repeat the test for resolution of the result. Supervisory review/vigilance of results by a second person can correct the error. Report result to the correct person after post test counselling and maintain the confidentiality. Reporting should be as per the "National HIV Testing Policy".
3. 4.
2. 3. 4.
Mixing of reagents from different kits/lot numbers. Performance characteristics (sensitivity, specificity and delta values) not satisfactory. Deterioration/contamination of one or more component or reagents of the kit. (faulty transport/storage). Intra-lot and inter-lot variation in kit performance due to faulty manufacturing practices.
5.
78
2. 3. 4. 5.
Scratching off the coated antigen from the well. Inconsistent technique for test and controls particularly quality control samples. Mixing reagents from different lots of kits. Mixing of samples.
6. 7.
SOP not followed to perform the test. Wrong pipetting due to carelessness.
3. 4. 5. 6.
ELISA washer not working satisfactory. Use of incorrect U.V. filters. Improper reader Refrigerator/deep freeze not working to optimum level.
Vigilance / Supervision
Periodic monitoring of the performance of laboratory workers is essential for ensuring quality of testing. This can be done by officer incharge of laboratory by resubmitting an already tested specimen as a test sample without the knowledge of the laboratory worker. If the results of the resubmitted sample are consistent with the previous results that means quality is being maintained. This should be done in a healthy way and under no circumstances the laboratory worker should be intimidated/reprimanded/ticked. Vigilance means scrutiny of laboratory and every step of the test to ensure accurate results every time. Scrutiny has to be random and of each step of the test right from collection, transport of specimen to reporting of result to the right person in the right manner. Every step of quality control has to be practised every day to ensure that the procedures and results are accurate and no false positive or false negative results are generated by the laboratory. This may be done by regular internal audit of the laboratory.
Record keeping
Monitor the records from collection and logging in of the specimen in the laboratory to reporting of the results. This will include the following details: Name of individual or specimen identification number. Date of collection of specimen and date received in laboratory. Name of requesting physician. The test requested. Worksheet. Result of the test.
The worksheet records the date of test, type of kit (Lot no. etc.) used, the test samples (nos.), quality control samples (nos.), name of laboratory worker, O.D. values obtained and cut-off values, etc. The worksheets are an important document for quality performance of tests. A model of worksheet is given in annexure III.
80
CHAPTER 14 IMPLEMENTATION OF EXTERNAL QUALITY ASSEMENT PROGRAMME AT DIFFERENT LEVELS OF THE HIV TESTING NETWORK AND JOB DEFINITIONS OF EACH LEVEL
A wide network of HIV testing laboratories operating at different levels exist in the country. The network includes National and State Reference Laboratories (NRLs / SRLs,), Voluntary Counseling and Testing Centers (VCTCs), PPTCTCs and blood bank HIV testing laboratories. These centers and laboratories are uniformly distributed across the country. Each and every facility / laboratory engaged in HIV testing has to practice quality assurance to ensure that every time the testing is performed the results produced are accurate, within the limits of the procedure and the reagents / kits used. The network of laboratories all over India is huge and scattered. So, the Quality Assurance Programme (QAP) is being implemented in three phases.
Phase I
All the National Reference Laboratories participate in this phase of QAP. Apex centre i.e. NIB distributes the characterized serum panel (10) to the National Reference Lab along with the procedural proforma. The participating labs process the External Quality Assessment serum panel exactly the same way as the routine samples and send the results to NIB within 15 days for analysis and feedback. This exercise is repeated once every 6 months. NIB also conducts EQA training program for the NRLs. Those National Reference Laboratories which are already participating in any of the External Quality Assessment Programmes being conducted by WHO Reference / collaborative laboratories may continue to do so.
Phase II
All the NRLs conduct training programme on EQAP and Internal Quality control as well as aspects of HIV Testing for the identified State Reference Laboratories (SRL) and distribute the EQA panel. Composition of serum panel is the same as that from apex reference laboratory. The SRLs test the EQA panel in the same way, under same conditions and using same kits which are used for testing client / patient sample within two to four weeks. The reports are sent back to the respective NRLs for analysis and feedback.
Phase III
Each National Reference Laboratory has been allotted states for implementation of EQA for all the HIV testing facilities existing in the respective states. For this purpose the National Reference Labs and or SRLs prepare the External Quality Assessment (EQA) serum panel of ten well characterized sera as per details already given. This panel comprising of 10 sera is sent to each and every HIV testing facility namely VCTC, PPTCTC & blood banks by the NRL/SRL in the respective allotted states. National AIDS Control Organization, Government of India facilitates the conduction of EQAP at various levels by 81
providing funds to NRLs/SRLs and issuing directives to SACS, NRL and participating VCTC, PPTCT and blood bank labs, etc for compliance, emphasizing the importance of EQAP. The participating laboratories in the states test the EQA panel sent by NRL/SRL as routine test samples and send the results to the assigned NRL/SRL within 2 weeks. NRL/SRL analyse the data and provide feed back to the participating labs. In case of discordance of result the experts from NRL visit the defaulting centre / laboratory for trouble shooting. NRL/SRL also conduct trainings for VCTCs and other HIV testing labs in the state. The National Reference Labs/SRL liase with the Project Director, State AIDS Control Society to train the staff from VCTC, PPTCTC and Blood Banks to participate in the External Quality Assessment Programme. The training programme is funded by respective Project Directors. It is a five days hands on training programme. The modus operandi for implementation of EQAP in states at grass route level can be decided by the Officer In charges, NRLs/SRLs corresponding to the size of states, etc. The panel is sent to the participating laboratories maintaining the cold chain (2-80C). The serum panel comprises of a total of 10 samples (4 positive, 4 negative and 2 borderline positive). The NRL sends all the relevant papers, proformas (Annexure VI) for performing the test and reporting back of the results obtained by the participating laboratories to the SRL/ NRL. The networking of HIV testing facilities for EQAP is given in annexure IX. Each National Reference Laboratory has been allotted SRLs/HIV testing cnentres in various states as mentioned in the annexure. The NRLs/SRLs cater to the ever expanding HIV testing facilities in their respective regions for EQAP.
Training of SRLs and HIV testing laboratories. Visiting SRLs and HIV testing labs for trouble shooting
VCTCs, PPTCTCs and blood bank laboratories, Job definition Testing of external quality control panel sent by NRL/SRL. Generation of data from blood bank labs, VCTCs and PPTCTCs. Referring lab related problems and problem sera (sero-status not resolved) to respective SRL Sending representative samples to SRL for quality assurance (20% of all positives and borderline positive sera and 5% of all negative sera). Should possess an independent HIV testing laboratory with adequate floor space, personnel for carrying out quality assurance work.
Transportation of panel
All the panels are to be transported from one laboratory to another in the temperature range of 2-8C. The panel should consist of 10 serum samples and each individual sample should have a volume of at least 100 microlitres.
Panel evaluation
Apex laboratory as well as the National Reference labs will carry out the evaluation of panel and analysis of the results. NACO will inform the apex laboratory and NRLs about the details of HIV test kits that are likely to be purchased and distributed to the blood banks and blood testing centers. 83
Kit evaluation
The apex laboratory and the National Reference Laboratories will have the responsibility of evaluation of kits as per requirement of NACO and other national bodies. It is also suggested that each fresh batch of kits purchased by NACO should be evaluated for sensitivity, specificity and delta values before it is distributed to the blood banks and other laboratories so that the quality of the kits supplied is assured. The apex laboratory and the NRL should put together efforts to prepare panel of sera to be used for evaluation of kits. This panel should include (i) sera from different geographical locations in the country (ii) sera against prevalent types and subtypes (iii) sera from patients with chronic diseases such as tuberculosis, leprosy, kala azar, auto immune diseases, malaria, etc. and (iv) sera from HIV sero-negative individual with risk behaviour for acquiring HIV infection. The panel established by the Apex Laboratory and National Reference laboratories will be tested for thermostability and reproducibility in the Apex Laboratory and NRLs. It is also recommended that regular surveillance should be maintained at each level. Five percent of negative samples, 20% of positive samples and all border line reactive samples may be monitored at the respective higher levels; grey zone samples should be repeated.
84
CHAPTER 15 EQUIPMENT, SUPPLIES AND REAGENTS FOR PRACTICING QUALITY CONTROL IN HIV TESTING*
Biosafety " " " " " " " " " " " Collection of serum samples from individual/plasma from blood bags (from blood banks). " " " " " " " " " Preparation of external control " " " " " " " " " " " " " " " " Laboratory coats/gowns (front closed). Chappals, slippers (top covered). Rubber gloves. Sodium hypochlorite / ethanol / glutaraldehyde / any other appropriate disinfectant. Discarding jar. Disposable paper sheets. Disposable bags. Autoclave, preferably incinerator. Lysol/labolene (for disinfecting floor) . Formalin (for disinfecting rooms). Needle destroyer Disposable syringes and needles. Screwcapped plastic tubes. Centrifuge for separation of serum / plasma. Vacutainers (vacuum based blood collection systems) CaCl 2. 6H2O; MgCl2. 6H2O. Centrifuge bottles (250 ml, presterilised, screwcapped). Water bath. Vacuum pump, seitz filter with filter pads, biological filters ( 0.8m, 0.6m). Magnetic stirrer. Water bath (35-44C). Incubator BOD ( 35-44C). Balance (physical / electrical). Rocker (40 - 60 rockings / minute for running Western Blot). ELISA reader filters (405nm, 492nm, 540nm and paper role). ELISA washing system. ELISA / rapid kits for HIV testing. Glass pipettes - 2ml, 5 ml, 10 ml. Microtips ( capacity 25l, 50l, 100l). Multichannel pipettes ( 50-200l). Micro pipettes ( 5-50l, 50-200l, 200-1000l). Microtip container. Pasteur pipette, rubber teats. Plastic forceps, plastic troughs. Tube racks. Graph paper. 85
" Sterilisation " " " " " " " " " "
Calculator (scientific). Hot air oven. Autoclave. Refrigerator (domestic with freezer compartment). Deep freeze (-30C, -70C). Polypropylene storage bottles ( 250 ml.). Plastic storage vials . Syringes for filtration through biological filters (50 ml.). Racks for holding storage vials. Biological filters ( 0.8m, 0.2 m). Bronidox (5 - bromo - 5 - nitro - 1, 3dioxane) and Propylene glycol (solvent for Bronidox)/ Thiomersal. Sample carrier ( e.g. Thermocol box). Water distillation plant. Glass ware (e.g. Beakers, flasks, measuring cylinders, etc). Markers, parafilm (for sealing storage vials). Air conditioner. UPS (uninterrupted power supply). Biological safety cabinet. ELISA washer (automated / semi-automated).
Storage / preservations
86
CHAPTER 16 BIOSAFETY FOR HEALTH CARE WORKERS WITH SPECIAL REFERENCE TO HIV
Introduction
Ever since AIDS has been recognized in 1981, the concern about bio safety in medical science has attained a new dimension. However, any measure regarding safety, needs assessment of the extent of risk, failing which, there is possibility of undue panic or carelessness due to over-estimation or under-estimation of the risk respectively. Fortunately, the threat of HIV infection to health care workers (HCW) has never assumed so much of problem. This has been mainly due to the fact that HIV is known for its fragile nature, having very low power of surviving in natural conditions like drying, temperature, etc. as well as due to its relatively lower concentration in blood and body fluids. Blood is the single most important source of HIV, HBV, HCV and other blood borne infections for HCWs. The risk of infection varies with the following factors: Type of exposure e.g. exposure of intact skin, nonintact skin, mucous membrane or needle stick injury. The amount of blood involved in the exposure; hollow bore needles, cannula, etc. carry more blood than IM (intramuscular) needles. The amount of virus in patient's blood at the time of exposure (maximum during acute viraemia). Prevalence of infection in the population. Number of exposures in case of needlestick injuries from individuals with unknown serostatus. Timely availability of post exposure prophylaxis (PEP).
Table1.
87
Table2.
The essential biosafety measures in relation to various steps of activities in a health care setting are as follows.
Bio safety guidelines for laboratory workers General precautions for laboratory workers
Wear gloves for all manipulations of infectious materials or where there is possibility of exposure to blood or body fluids containing blood. Ensure generous supply of good-quality gloves. Discard gloves whenever they are thought to have become contaminated; wash your hands, and put on new gloves. Do not touch your eyes, nose, or other exposed membranes or skin with gloved hands. Do not leave the workplace or walk around the laboratory wearing gloves. Wash your hands with soap and water immediately after any contamination and after work is completed. If gloves are torn, wash your hands with soap and water after removing the gloves. Wear a laboratory gown, smock, or uniform when in the laboratory, wrap-around gowns are preferable. Remove this protective clothing and leave it in the laboratory when leaving the laboratory. When work with material potentially infected with HIV is in progress, close the laboratory door and restrict access to the laboratory. The door should have a biohazard symbol and No Admittance sign posted. Keep the laboratory clean, neat, and free of extraneous materials and equipment. Disinfect work surfaces when procedures are completed at the end of each working day. An effective all-purpose disinfectant is a 1% hypochlorite solution. Place used needles, syringes, and other sharp instruments and objects in a puncture-resistant container. Do not recap used needles and do not remove needles from syringes. Never use mouth pipetting.
88
Perform all technical procedures in a way that minimizes the creation of aerosols, droplets, splashes, or spills. Do not eat, drink, smoke, store food or personal items, or apply cosmetics in the laboratory. Have an effective insect and rodent control programme as per standard recommendation.
Guidelines
Inspect your hands for cuts, scratches, or other breaks in the skin. If the skin is broken, cover with occlusive bandage and wear gloves. Take care to avoid contaminating your hands during the collection of blood. If blood gets on gloves, these should be discarded. Place used needles and syringes in a puncture-resistant container. Do not recap used needles. Do not remove needles from syringes. Do not use needle clippers. Seal specimen containers securely. Wipe the exterior of the container free of any blood contamination with a disinfectant. Wash your hands with soap and water immediately after any blood contamination and after work is completed. Wear the laboratory gown. In the event of needlestick or other skin puncture or wound, wash the wound thoroughly with soap and water. Encourage bleeding. Report any contamination of the hands or body with blood, any puncture wound, or cut to the supervisor and the health service.
89
Chlorine - sodium hypochlorite For wiping work benches and specimen containers, cleaning gloves, etc. 1% freshly prepared sodium hypochlorite solution is used. For heavily soiled equipment, blood spills, etc. 10% sodium hypochlorite solution is needed. Sodium hypochlorite solutions gradually lose their strength, requiring daily preparation of fresh solutions for use. The effectiveness of hypochlorite solutions is neutralized by blood, serum, and other proteinaceous material. Daily replacement is required. Chlorine - calcium hypochlorite Available in powder, granule, or tablet form. Decomposes at a slower rate than sodium hypochlorite. Availability of chlorine - (70%) 0.1% solution is obtained by dissolving 1.4 g dry chemical in 1 litre of water. 1.0% solution is obtained by dissolving 14 g dry. chemical in 1 litre of water. Chlorine - sodium dichloroisocyanurate (NaDCC ) Available as tablet. 60% available chlorine. 0.1% available chlorine solution is obtained by a dilution of 1.7 g stock in 1 litre of water. 1.0% solution is obtained by a dilution of 17 g in 1 litre of water. More stable than sodium and calcium hypochlorite. Chloramine
Available in powder or tablet form 25% available chlorine Releases chlorine at a slower rate than other chlorine compounds. Higher concentrations are required for effective disinfection. General purpose - 20 g stock in 1 litre of water. 40 g stock in 1 litre of water for disinfecting heavily soiled equipment, blood spills, etc. Chloramine is more stable than sodium and calcium hypochlorite.
Alcohol-ethanol (ethyl alcohol) and 2-propanol (isopropyl alcohol) Ethanol and 2-propanol are effective disinfectants, particularly for cleaning surfaces such as the exterior of specimen containers and bench tops. For maximum effectiveness they should be used in a 70% concentration (70% alcohol, 30% water). Iodophore - povidone iodine (PVI) Disinfectant activity is similar to that of hypochlorite solutions. More stable and less corrosive (however, do not use on aluminium or copper). Formaldehyde -formalin Excellent disinfectant Formalin containing 35%-40% formaldehyde, 10% methanol and water. Should be diluted 1:10 (a solution containing 3.5%-4% formaldehyde) for disinfection. Rinse the equipment before reuse.
90
Glutaraldehyde High-level disinfectant. Extremely useful for the disinfection of non-disposable heat-sensitive equipment and instruments. Usually available as a 2% aqueous solution. Immersion in the solution destroys vegetative bacteria, fungi, and viruses in less than 30 minutes. Ten hours' immersion is required for the destruction of spores. Treated equipment must be thoroughly rinsed. Prepared solutions should not be kept more than 2 weeks, or according to the manufacturer's instructions.
The selection of an appropriate disinfectant will also be weighed in terms of contact time, physical characteristics like irritant and toxic properties etc. In a developing country like India cost is another important consideration.
Transport of specimens
Transport of specimens (e.g. serum, blood etc.) is a frequently performed activity for a laboratory engaged in HIV diagnosis. The broad guidelines are as follows: General guidelines Specimen containers should be screwcapped, leak-proof resistant plastic or glass. After the container is closed and sealed it should be wiped with a disinfectant - e.g. 1% sodium hypochlorite solution. On receipt of a specimen, before the container is opened it should be wiped with a disinfectant as above. Within the health care facility and laboratory, the specimen containers should be placed in racks to maintain them in an upright position. The racks are to be transported in leak-proof plastic, metal or rigid thermocol containers. From field collection sites or between laboratories, leak-proof plastic or metal boxes with secure, tight-fitting covers should be sought.
Disposal of laboratory waste Waste should be segregated into different categories at the site of generation i.e. in the laboratory. Non infectious waste is collected in black bags and disposed as general waste. Infectious waste should be decontaminated prior to its storage, transport and disposal. Solid infectious waste is disposed as follows Sharps needles and syringe nozzles are shredded in needle-destroyer (if available); if not available decontaminated as in below. Scalpel blades/ lancets / broken glass should be put in an impermeable container with bleach that should be filled to three fourth only, transferred to plastic/ cardboard boxes, sealed to prevent spillage and transported to incinerator. Glass ware should be disinfected and sterilised by autoclaving followed by cleaning; never attempt cleaning before autoclaving. Swabs are chemically disinfected followed by incineration.
Disposable items include single use products (syringes, gloves, sharps, etc.). As these items are often recycled and have the risk of being reused illegally, these should be disinfected by dipping in freshly prepared 1% sodium 91
hypochlorite for 30 minutes to 1 hour. Bins which can be used for this purpose are a set of twin bins, one inside the other with the inner one being perforated and easily extractable. This minimizes contact when the contents are being removed. Disposable items like gloves, syringes, etc. should be shredded, cut or mutilated before disposal followed by deep burial or properly accounted before disposal. Extreme care should be taken while handling the needles. Liquid wastes generated by the laboratory are either pathological or chemical in nature and are disposed of as follows: Non infectious chemical waste should first be neutralised with reagents and then flushed into conventional sewer system. The liquid infectious waste should be treated with a chemical disinfectant for decontamination then neutralized and flushed into the sewer. Solid wastes are collected in leak-resistant single heavy duty bags or double bags may be used. Bags having different colour with red labels mentioning date and details of waste are recommended. The bags are tied tightly after they are three-fourths full. Processing of glass syringes, needles and gloves for reuse These are the items most frequently handled in any laboratory handling blood or other infectious fluids/materials. Hence procedures for their reuse are mentioned below. Reuse of needles and glass syringes Gloves must be worn. Extreme care must be exercised to prevent needle stick injuries and/or cuts. Leave the needle attached to the syringe. Aspirate hypochlorite or another suitable disinfectant into the syringe. Immerse the syringe and the attached needle in disinfectant solution, horizontally in a flat tray. Leave them immersed in disinfectant solution for 20 minutes. Discharge the disinfectant solution from the syringe and needle. Rinse the syringe and needle, filling and emptying several times. Examine needles and syringes for needle barbs, the rubber ring, the needle hub fit to the syringe, readable syringe markings, etc. Disinfect or sterilize the syringe and needle by autoclaving (in a steam steriliser) or boiling in water prior to reuse. Reuse of gloves Generally recommended for laboratory workers who handle blood specimens or other suspect fluids potentially infected with HIV. Gloves reduce the risk of contamination of the hands with blood, but will not prevent the occurrence of penetrating injuries or cuts caused by needles, other sharp instruments, or broken glass or plastic. It is important to remember that gloves are meant to supplement, not replace good infection control practices, including the hygienic practice of proper hand washing. Test the gloves for peel, cracks, punctures, etc. Thoroughly rinse your gloved hands in a hypochlorite disinfectant solution. Rinse your gloved hands in clear water. Wash your gloved hands with soap and water and rinse. Remove the gloves and hang them up to dry. Wash your hands. 92
Packing, storage and transport of treated (disinfected) wastes All segregated and disinfected waste should be packed in proper containers and colour-coded bags with red label biohazard signs (Table - 4). All containers used for storage of such waste shall be provided with a properly covered lid to avoid spillage during handling and transit of such waste. Such containers should be inaccessible to scavengers and protected against insects, birds, animals and rain. The waste should be transported in vehicles authorised for this purpose only. No such waste should be stored in the place where it is generated for a period of more than two days.
Table 4.
ks
93
ks
The exposure must be reported to the appropriate authority and condition must be treated as an emergency. Prompt reporting is absolutely essential.
* HBIG dose 0.06 ml/kg IM @ Adequate anti-HBs Ag is > 10 SRU by RIA or positive by EIA.
Hepatitis C (HCV)
Hepatitis C is most commonly transmitted by parenteral route, i.e., through infected blood transfusion or sharing of needles and syringes in IV drug users. It is a common agent of co-infection with HIV. Like the other bloodborne infections, the mainstay of preventing occupational infection is through prevention of exposure. Risk of transmission after needlestick injury is 2-3%. There is no vaccination at this time, and no recommended chemoprophylaxis. For HCW exposed to Hepatitis C, follow up care is recommended.
95
HIV
Management of Exposure : Steps to be taken on accidental exposure to blood (or body fluid containing blood) are: Wash wound immediately with running water and soap Inform the lab /hospital management and document occupational accident Consult with nearest ART centre/ resource for Post-exposure prophylaxis, evaluation, and follow-up (as per the National guidelines on PEP) Counselling and collection of blood for testing from the exposed HCW with written informed consent must be done. Whenever possible confidential counselling and testing of source for Hepatitis, HIV etc must be done. A history should be taken as well to ascertain likely risk of the source. (PEP should be provided to the exposed HCW until report of source is available and confirmed negative). Risk of infection and transmission must be evaluated Never delay start of therapy due to debate over regimen. Begin with basic 2-drug regimen, and change if warranted, once expert advice is obtained Reevaluation of the exposed person should be considered within 72 hours post exposure, especially as additional information about the exposure or source person becomes available. The exposed person is advised to seek medical evaluation for any febrile illness that occurs within 12 weeks of exposure. Administer PEP for 4 weeks .PEP should be provided until result of the source's test is available and confirmed negative or until course completed ,if source positive or unknown A repeat HIV test of the exposed individual should be performed at 6 weeks, 12 weeks and 6 months postexposure, regardless of whether or not PEP was taken
Ideally, prophylaxis should be begun within 2 hours of exposure. Late PEP (within 72 hours) may still be useful as early treatment of HIV infection, in case infection has occurred. Donts: Do not panic! Dos: Do not reflexively place pricked finger into mouth Do not squeeze blood from wound, this causes trauma and inflammation, increasing risk of transmission Do not use bleach, alcohol, betadine, or iodine, which may be caustic, also causing trauma
Remove gloves, if appropriate Wash site thoroughly with running water. Irrigate thoroughly with water or saline if splashes have gone into the eye or mouth.
Seroconversion after occupational exposure If transmission of a bloodborne infection occurs after occupational exposure which has been documented, the HCP has a right to receive treatment and care for this illness. It is also the right of all persons infected thusly, to freedom from discrimination regarding their working conditions. Such persons are entitled to have all of their human rights respected, beginning firstly with the right to confidentiality regarding their health care.
96
of
Nelfinavir 750 mg three times daily or any other boosted protease inhibitor. (for higher risk categories - consult expert).
In case the source patient is on ART and considered to be having drug resistance (clinically) then the ART 97 124 125 124 125
physician should be consulted and appropriate ARVs should be given for PEP. Schedule of necessary measures regarding therapy and follow up in case of PEP The person should be provided with pre test counseling and PEP be started as discussed above. Before starting PEP, 3-5 ml of person's reference blood sample is to be taken and sent to the laboratory for testing and storage. It is important that a serum sample is collected from the HCW as soon as possible (zero hour) after exposure for HIV testing, failing which it may be difficult to attribute the acquired infection due to exposure in occupational setting. This may have bearing on the claims of compensation from health authorities. The first sample for HIV testing is collected immediately after exposure and 2nd at 6 weeks, 3rd at 12 weeks and last at 6 months after the exposure. During the follow-up period, especially the first 6-12 weeks when most infected persons are expected to show signs of infection, the recommendations for preventing transmission of HIV are to be followed by the HCW. These include refraining from blood, semen, organ donation and abstaining from sexual intercourse. In case sexual intercourse is undertaken a latex condom must be used correctly and consistently. This reduces the risk of HIV transmission. In addition, women should not breast-feed their infants during the follow-up period after exposure to prevent exposing their infants to HIV in breast milk. The antiretroviral drugs for PEP are to be given for four weeks. Government of India has already made the resources available with various State AIDS Societies to meet with the expense of PEP for HCWs. The drugs for PEP must be available round the clock. The report of exposure and PEP has to be sent to Addl. Director (Technical) National AIDS Control Organisation, GOI. Pregnancy and PEP If the HCW is pregnant at the time of occupational exposure to HIV, the designated authority/physician must be consulted about the use of anti-viral drugs for post exposure treatment. Facts known about the safety and side effects of these drugs Most of the information known about the safety and side effects of these drugs is based on studies of their use in HIVinfected individuals. For these individuals, ZDV and 3TC have usually been tolerated well except for nausea, vomiting, diarrhoea, tiredness, or headache for people taking ZDV. Steps to be undertaken by the infection control officer on receiving information about occupational exposure All the needle stick injuries should be reported to the State AIDS Society giving the details of the type of exposure. and the measures taken to manage the same.The State AIDS Societies should in turn inform NACO about the cases periodically. A registry is available with NACO for follow-up of all such cases. Infection control officers in all hospital have been directed to ensure that anti retroviral drugs for PEP are available in casualty at all the times.
98
3.
17.
AIDS Etiology Diagnosis Treatment and Prevention 4th Ed. Curran J., Esses M., and Fauci A.S, Pippincott Raven, 1997. HIV Testing Manual: Laboratory Diagnosis, Biosafety and Quality Control. NACO document. NABL documents on accreditation. ISO 17025 and other ISO documents.
18. 19 20
100
Annexure I
Signature: Date:
Note:
1. It may be noted that general consent obtained for carrying out procedures in hospital does not include HIV consent. 2. 3. In case of minor, the consent should be obtained from the parents. In case of unconscious patients, where there is a need for diagnosis of HIV for management of the patient, consent should be obtained from the parents, spouse/closest relative available at that time. 4. In case no attendant is available, the test, if necessary for management of the patient, may be carried out on recommendations of two attending doctors.
101
Annexure II
102
Annexure III
103
Annexure IV
104
105
Additional Project Director (Technical) N.A.C.O., Ministry of Health & Family Welfare
106
Annexure VI
107
Annexure VI contd.
HIV ANTIBODY TEST RESULTS ELISA RESULTS (SRL/BLOOD BANKS/OTHER LABS PERFORMING ELISA)
Please record the reading for the sample (column A) and the appropriate cutoff value/rate (column B) and calculate the Sample Ratio/Cut off Ratio (A-B) for each panel sample. Please record the test interpretation given to the sample after the testing.
108
Annexure VI contd.
109
Annexure VI contd.
110
Annexure VI contd.
111
Annexure VII Name of the NRL / SRL Conducting the Training Programme (Sample Questionnaire) Workshop on External Quality Assessment Programme for HIV Testing Participant Group: Laboratory supportive personnel Pre-Test / Post-Test Evaluation Time :30 min. Name of the Participant: Name of the Institute:
TICK (
1.
The choice of protocol for serological testing does not depend on the following: (i) Objective of HIV testing (ii) Sensitivity and specificity of the test (iii) Prevalence of HIV infection in (iv) Time limit for performance of test the population (v) Serum or plasma as specimen In multiple testing strategy for HIV infection, the strategy II B is employed for (i) Blood bank (ii) Diagnosis of individual with AIDS defining symptoms (iii) Detection of HIV status in VCTC The most important limitation of p24 antigen for diagnosis of HIV infection is: (i) Poor specificity (ii) Poor sensitivity (iii) Poor correlation with therapy (iv) Inability to diagnose HIV-2 infection External quality assessment scheme is usually organized by (i) Local officer-in-charge of the (ii) Local supportive laboratory personnel laboratory (iii) External agency / National / Regional reference laboratory Recommended method for conversion of plasma to serum is (i) Calcification (ii) Thrombinisation (iii) Heating the plasma at 56C in (iv) Incubating at 37C water bath Recommended diluent for making serial dilution of strong reactive serum for preparing level control is following expert (i) Normal saline (iii) Serum negative for infectious markers (like HIV, HBV, HCV etc.) (ii) Phosphate buffer saline at pH 7.2 (iv) Distilled water 112
2.
3.
4.
5.
6.
Annexure VII.....contd. 7. In order to find out the extent of variation of reactivity of a sample over 7 consecutive days in ELISA kits of same batch number, it is preferable to draw inference from (i) OD values of the samples (ii) Positive control values (iii) Negative control values (iv) 'E' ratio of the sample 8. Any given HIV positive serum when subjected to 4 kits of different brands following things are likely to be observed (i) 'E' ratio differs (ii) 'E' ratio does not differ Recommended range of 'E' ratios of an external control in indirect ELISA is (i) 10-20 (ii) 1.5-2 (iii) 0.01-0.1 (iv) 0.1-1 Levy Jenning's chart for validation of test result usually employees the range of (i) Mean 1SD (ii) Mean 2SD (iii) Mean 3 SD (iv) Mean 4 SD Six consecutive points of external control values gradually moving in one direction up or below the mean line in the Levy Jenning's chart possibly indicate: (i) Sudden change in the batch of kit (ii) Change in the technician Performing the test (iii) Change of incubator with (iv) Gradually deteriorating reagents different temperature set point 12. For storage of the aliquot of external control sample for daily use over 7 days Temperature required is (i) 0C (ii) 2-8C (iii) <-30C (iv) <-70C The recommended preservative for long term preservation of external control serum is (i) Sodium azide (ii) Thiomersal (iii) Bronidox Standard quality control practice requires (i) Inclusion of external control in (ii) (iii) Changing the working hand (performing the test frequently)
9.
10.
11.
13.
14.
15.
Two populations X and Y were subjected to a screening test (p) for HIV infection with sensitivity and specificity of 99.5% and 99% respectively. Prevalence of HIV infection in population X is 20 per 1000 which the same in population Y is 5 / 1000. The positive predictive value of the screening test (p) in population X will be: (i) Lower than that in population Y (iii) Identical compare to that in population Y (ii) Higher than that in population Y
113
Annexure VII.....contd. 16. While choosing a kit for diagnosis of HIV infection, three kits were evaluated with a panel of 200 sera comprising of 100 true positive and 100 true negatives the following characteristics were obtained. Mark the one with maximum positive Delta value.
17.
The most important limitation of a rapid test for diagnosis of HIV infection is (i) Poor sensitivity (ii) Poor specificity You want to buy one micropipette for your laboratory for HIV testing. Placed below is a checklist of major (basic) specifications that you need to consider before the purchase. Specify (mark ) if any of these is considered unnecessary by you. Name sticker to avoid mixing among personnel (iii) Colour of the micropipette black or grey (v) Slim design to negotiate through narrow neck containers upto a volume of 50 ml. (vii) Simple for cleaning and servicing matching stickers for (i) Devise for comfortable resting on hand (iv) Easy tip purging (vi) Autoclavable (ii)
18.
19.
Some common methods for calibration of micropipette are the following with the exception of (Mark ). (i) Pipetting of coloured solutions (ii) Weighing distilled water or followed by spectrophotometric mercury analysis (iii) Pipetting volumes of radioisotopes (iv) Commercially available spectrophotometric calibration kit (i) Weighing the micropipette at weekly interval for 4 week The 99% Sensitivity of an HIV test means that: The test will correctly identify 99 to 100 infected individuals (iii) The technician must be careful in running the test (v) The test will rarely give a false positive results (i) The test will be negative in 99% of uninfected persons (iv) The accuracy of the test divided by its cost (ii)
20. (A)
114
Annexure VII.....contd. (B) The 99% Specificity of an HIV test means that: (i) The test will correctly identify (ii) 99 to 100 infected individuals (iii) The technician must be careful in (iv) running the test (v) The test will rarely give a false positive results The test will be negative in 99% of uninfected persons The accuracy of the test divided by its cost
115
Annexure VIII Name of the NRL / SRL Conducting the Training Programme (Sample Questionnaire) Workshop on External Quality Assessment Programme for HIV Testing Participant Group: Laboratory supportive personnel Pre-Test / Post-Test Evaluation Time :30 min. Name of the Participant: Name of the Institute:
TICK (
1.
3rd generation ELISA uses the following as antigen of HIV (i) Infected cell lysate (ii) Recombinant glycopeptides (iii) Synthetic peptides Internal kit controls (serum) refers to following: (i) Controls supplied with the kit (ii) Controls prepared out of own Laboratory samples (iii) Controls procured from commercial (iv) Controls procured from the agency reference laboratory External control (serum) refers to (i) Controls supplied with the kit (iii) Control procured from commercial agents
2.
3.
Controls prepared out of own laboratory samples (iv) Controls procured from the reference laboratory
(ii)
4.
ELISA ratio or E ratio refers to (i) Ratio of sample OD/ cut off OD (iii) Ratio of sample OD upon positive control OD
(ii)
5.
Calculation of standard deviation of external control values over 20 observations requires. (i) Mean value of the 20 observations (ii) Square of the individual values
Annexure VIII.....contd. 6. The recommended storage temperature of most of the HIV diagnostic kits is (i) 0C (iii) <-30C 7. (ii) 2-8C (iv) <-70C
For general purpose use the recommended concentration of sodium hypochlorite is (i) 0.01% (ii) 0.1% (iii) 1% (iv) 10% To mop up accidental spillage of blood, the recommended concentration of Sodium hypochlorite is (i) 0.01% (ii) 0.1% (iii) 1% (iv) 10% In the multiple test strategy for diagnosis of HIV infection in an individual (i) (ii) The specimen is tested thrice by a particular rapid test with high sensitivity. The specimen is tested by indirect EIA from three different companies that employ identical antigens for
8.
9.
coating the EIA plate. (iii) The specimen is tested by three tests (EIA/rapid) based on different principles / different antigen preparations 10. Ideal choice of a kit for use in a blood bank is the one with (i) Maximum sensitivity(100%) (ii) Maximum specificity Third generation/fourth generation Discrimination between HIV-1 and HIV-2 infections is possible by (i) All EIA kits (ii) All rapid / visual kits (iii) Some rapid / visual kits On a particular day while running routine EIA for detection of HIV reactive units among donated blood samples it was observed that all the wells in the plate including the control well did not show any colour at the end of the run. The possibilities considered are: (Mark ) the one you think valid. (i) The plate is not coated with antigen Washing has not been proper or omitted at the stage following addition of conjugate (iv) Plate has been incubated more than the recommended time (vi) The plate has been exposed to light for long time after addition of substrate (ii)
11.
12.
(iii) Substrate has been broken down before addition (v) Stop solution has been added late in the final stage 13.
On a day of EIA run, the whole plate, including the positive control wells show colour. The possibilities considered are as follows: (Mark ) the one you consider valid: (i) Washing has not been done (ii) Substrate has lost its potency properly (iii) Conjugate has been deteriorated (iv) Incubation has been carried out for less than recommended period 117
Annexure VIII.....contd. 14. Most commonly employed laboratory tests for diagnosis of HIV infection. (i) Serological diagnosis by (ii) HIV isolation antibody detection (iii) Serological diagnosis by antigen (iv) Haematological studies detection (iv) Skin test Collection of blood for diagnosis of HIV infection by ELISA (i) Clotted blood (ii) Heparinised blood (iii) Blood in tissue culture media Mention the appropriate storage temperature for 3 days for serum for serological test of HIV infection: (i) 2-8C (ii) 0C (iii) Less than 20C Mention the storage temperature of ELISA kits for HIV (i) 2-8C (ii) 0C (iii) Less than 20C The most commonly used disinfectant in a blood bank / HIV testing laboratory is (i) Absolute alcohol (ii) Formaldehyde (iii) Dettol (iv) Sodium hypochlorite The recommended working concentration of sodium hypochlorite is (put tick against the appropriate concentration) (i) ELISA plate / tips used in routine serology (ii) Blood spillage .001% 100% .001% 100% .01% 0.1% .01% 0.1% 1% 1% 5-10% 5-10% 1
15.
16.
17.
18.
19.
20.
ELISA positivity in an HIV infected means that person has AIDS: (i) True (ii) False
118
Annexure IX
6.
7.
8. 9.
119
ANNEXURE X
PIPETTING TECHNIQUE
The following information is consolidated from operational instruction manuals from several pipette manufacturers. Complete information and more detailed instructions may be seen in the specific pipette instruction manuals. Read the manufacturer's manual carefully before beginning the pipetting procedure. Choose the appropriate disposable pipette tips for your application. Hold the pipette in one hand and use a slight twisting movement to seat the tip firmly on the shaft of the pipette to ensure an air-tight seal Select the desired volume (with manual pipettes, higher volumes should be set first; if adjusting from a lower to a higher volume, first surpass the desired volume and then slowly decrease the volume until the required setting is reached). If applicable, select the desired mode (e.g., reverse pipette). This is recommended for optimal precision and reproducibility. Reverse pipetting can be done with a manual pipette by pressing the control button slightly past the first stop when aspirating, taking up more liquid than will be dispensed, then pressing the control button only to the first stop when dispensing. A small volume will remain in the tip after dispensing. Select an appropriate tip.
Pre-rinsing
The following procedures will help ensure optimal precision and accuracy. Volumes >10 L: Prerinse pipette 23 times for each new tip (this involves aspirating and dispensing liquid several times). Reasons for prerinsing include the following: to compensate for system pressure, for slight differences in temperature between pipette and liquid, and for properties of the liquid; to clear the thin film formed by the liquid on the inside of the pipette. Without prerinsing, retention of a thin film on the inside wall of the tip would cause the first volume to be too small. The thickness and nature of this film, and therefore the potential source of error, will vary depending on the nature of the liquid being pipetted. Volumes <10 L: Do not prerinse pipette, but rinse tip after dispensing to ensure that the whole volume was dispensed. For smaller volumes, prerinsing is not recommended because the dispensed volume may be higher than the requisite volume.
Filling
Make sure tip is securely attached. Hold pipette upright. When aspirating, try to keep the tip at a constant depth below the surface of the liquid. Glide control button slowly and smoothly (electronic pipettes perform this step automatically). When pipetting viscous liquids (e.g., whole blood), leave the tip in the liquid for 12 seconds after aspirating before withdrawing it. After liquid is in the tip, never lay the pipette on its side.
120
Dispensing
Hold the tip at an angle, against the inside wall of the vessel/ tube if possible. Glide control button slowly and smoothly (electronic pipettes perform this step automatically).
121
Other Recommendations :
To ensure optimal performance, the temperature of the pipetted solution and the pipette and tips should be the same (volume errors may occur because of changes in air displacement and viscosity of the liquid). Do not pipette liquids with temperatures >70C. Volume errors may also occur with liquids that have a high vapor pressure or a density/viscosity that differs greatly from water. Water is most commonly used to calibrate pipettes and to check inaccuracy and imprecision. A pipette could possibly be recalibrated for liquids with densities that vary greatly from that of water. Pipettes should be checked regularly for precision and accuracy. Regular maintenance (e.g., cleaning) should be performed either by the user or a service technician according to manufacturer's instructions. All calibration and maintenance records must be maintained.
122