Catalogo de Delta Ohm de Agua
Catalogo de Delta Ohm de Agua
Catalogo de Delta Ohm de Agua
pag. 338 pag. 342 pag. 345 pag. 346 pag. 352 pag. 357 pag. 360 pag. 364 pag. 366
The qualitative level of our instruments is the result of a continuous evolving of the product itself. This way bring to slight differences between what reported on this manual and the instruments you bought. We can not completely exclude the presence of errors for which we apologise. Data, images and descriptions included in this catalogue can not be enforced legally. We reserve the right to perform any modification and correction at any time without notice.
Water Analysis
HD 2105.1 HD 2105.2
INSTRUMENT TECHNICAL CHARACTERISTICS Instrument Dimensions (Length xWidth xHeight) 185x90x40mm Weight 470g (complete with batteries) Materials ABS, rubber Display 2x4 digits plus symbols Visible area: 52x42mm Operating conditions Operating temperature Storage temperature Working relative humidity Protection degree Power Batteries Autonomy Power absorbed with instrument off Mains Security of memorized data Time Date and time Accuracy -550C -2565C 090%RH without condensation IP67 4 1.5V type AA batteries 200 hours with 1800mAh alkaline batteries 20A Output mains adapter 12Vdc / 1000mA Unlimited, independent of battery charge conditions Schedule in real time 1min/month max drift
Measured values storage - model HD2105.2 Type 2000 pages containing 17 samples each Quantity Total of 34000 samples Storage interval 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1h. Serial interface RS232C Type Baud rate Data bit Parity Stop bit Flow Control Serial cable length Selectable print interval RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1 Xon/Xoff Max 15m 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1h. 1.1 - 2.0 electrically isolated
USB interface - model HD2105.2 Type Connections Input module for the temperature probes pH/mV input Serial interface and USB Mains adapter Measurement of pH by Instrument Measurement range Resolution Accuracy Input impedance Calibration error @25C Temperature compensation automatic/manual
8-pole male DIN45326 connector Female BNC 8-pole MiniDin connector 2-pole connector (positive at centre) -2.000+19.999pH 0.01 or 0.001pH selectable from menu 0.001pH >1012 |Offset|>20mV Slope<50mV/pH or Slope>63mV/pH Sensitivity < 85% or Sensitivity > 106.5% -50+150C
HD2110CSNM
HD2101/USB
HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm. pH Electrodes KP 20: Gel pH combined electrode for general use , with S7 screw connector, EPOXY body. KP 30: Gel pH combined electrode for general use, 1m cable with BNC, EPOXY body . KP 50: Gel pH combined electrode, porous Teflon ring junction, suitable for emulsions, demineralised water, with S7 screw connector, glass body. KP 61: 3 diaphragm liquid filled pH combined electrode for wine, milk, cream, etc., S7 screw connector, liquid reference filling, glass body. KP 62: 1 diaphragm gel pH combined electrode for pure water, varnishes, gel filled, S7 screw connector, glass body. KP 63: 1 liquid filled pH combined electrode for general use, varnishes, 1m cable with BNC, glass body. KP 64: Liquid filled pH combined electrode ,Teflon ring diaphragm, for wine, varnishes, emulsions, S7 screw connector, glass body. KP 70: Pointed gel combined pH microelectrode diam. 6 x L=70 mm., with S7 screw connector, EPOXY body, glass tip, open junction. KP 80: Pointed gel pH combined electrode, with S7 screw connector, glass body, for cream, milk, viscous material, open junction. KP100: Flat membrane gel combined pH electrode with S7 screw connector, glass body, for skin, leather, paper. Characteristics and dimensions of the probes at page 401 CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 5: 5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 10: 10m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 15: 15m extension cable with BNC/S7 connector for electrode without cable, thread S7. CE : S7 screw connector for pH electrode. BNC: female BNC for extension cable ORP Electrodes KP 90: REDOX PLATINUM liquid filled electrode with S7 screw connector, glass body. KP 91: Gel REDOX PLATINUM electrode, 1m cable with BNC, EPOXY body Characteristics and dimensions of the probes at page 397 pH Buffer solutions HD8642: Buffer solution 4.01pH - 200cc. HD8672: Buffer solution 6.86pH - 200cc. HD8692: Buffer solution 9.18pH - 200cc. Redox Buffer solutions HDR220: Redox buffer solution 220mV 0.5 l. HDR468: Redox buffer solution 468mV 0.5 l. Electrolyte solutions KCL3M Ready to use solution for electrode refilling 100 cc Cleaning and maintenance HD62PT: Diaphragm cleaning (tiourea in HCl) - 500ml. HD62PP: Protein cleaning (pepsin in HCl) - 500ml. HD62RF: Regeneration (fl uorhydric acid) - 100ml. HD62SC: Solution for electrode preservation - 200ml. Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP47.100: Direct 4 wires Pt100 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 4 wires with connector, length 2 m. TP47.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 2 wires with connector, length 2 m. TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes.
Measurement of temperature by Instrument Pt100 measurement range -200+650C Pt1000 measurement range -200+650C Resolution 0.1C Accuracy 0.1C Drift after 1 year 0.1C/year TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT Temperature probes Pt100 sensor using SICRAM module
Model TP87 TP472I.0 TP473P.0 TP474C.0 TP475A.0 TP472I.5 TP472I.10 Type Immersion Immersion Penetration Contact Air Immersion Immersion Application range -50C+200C -50C+400C -50C+400C -50C+400C -50C+250C -50C+400C -50C+400C Accuracy 0.25C (-50C+200C) 0.25C (-50C+350C) 0.4C (+350C+400C) 0.25C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+250C) 0.3C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C)
0.003%/C
0.005%/C
ORDER CODES HD2105.1: The kit is composed of: instrument HD2105.1, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. Electrodes, temperature probe, calibration solutions, data transfer cable for PC or printer have to be ordered separately. HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C. HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin. C.206: Cable for instruments of the series HD211 and .2 to connect directly to USB input of PC. DeltaLog9: Software for download and management of the data on PC using Windows 98 to Vista operating systems. SWD10: Stabilized power supply at 230Vac/12Vdc-1A mains voltage. HD40.1: The kit includes: 24-columm portable thermal printer, serial interface RS232, 57mm paper width, four NiMh 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls BAT-40: Spare battery pack for HD40.1 printer with built-in temperature sensor. RCT: The kit includes 4 thermal paper rolls 57mm wide and 32mm diameter.
HD22.3
Water Analysis
HD 2305.0
INSTRUMENT TECHNICAL CHARACTERISTICS Measured quantities: pH, mV, C, F Instrument Dimensions (Length xWidth xHeight) 140x88x38mm Weight 160g (complete with batteries) Materials ABS Display 2x4 digits plus symbols Visible area: 52x42mm Working conditions Operating temperature Storage temperature Working relative humidity Protection degree Power Batteries Autonomy Power absorbed with instrument off Connections Input module for the temperature probes pH/mV input Measurement of pH by Instrument Measurement range Resolution Accuracy Input impedance Calibration error @25C Measurement of mV by Instrument Measurement range Resolution Accuracy Drift after 1 year Measurement of temperature by Instrument Pt100 measurement range Pt1000 measurement range Resolution Accuracy Drift after 1 year -550C -2565C 090%RH without condensation IP67 3 1.5V type AA batteries 200 hours with 1800mAh alkaline batteries 20A
8-pole male DIN45326 connector Female BNC -2.000+19.999pH 0.01 0.01pH1 digit >1012 |Offset|>20mV Slope<50mV/pH or Slope>63mV/pH Sensitivity < 85% or Sensitivity > 106.5% -1999.9+1,999.9mV 0.1mV 0.1mV 0.5mV/year -200+650C -200+650C 0.1C 0.1C 0.1C/year
TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT Temperature probes Pt100 sensor using SICRAM module
Model TP87 TP472I.0 TP473P.0 TP474C.0 TP475A.0 TP472I.5 TP472I.10 Type Immersion Immersion Penetration Contact Air Immersion Immersion Application range -50C+200C -50C+400C -50C+400C -50C+400C -50C+250C -50C+400C -50C+400C Accuracy 0.25C (-50C+200C) 0.25C (-50C+350C) 0.4C (+350C+400C) 0.25C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+250C) 0.3C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C)
0.003%/C
0.005%/C
HD8642
HD8672
HD8692
ORDER CODES HD2305.0: The kit is composed of: instrument HD2305.0, 3 1.5V alkaline batteries, operating manual, case. HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm. pH Electrodes KP 20: Gel pH filled combined electrode for general use , with S7 screw connector, EPOXY body. KP 30: Gel pH combined electrode for general use, 1m cable with BNC, EPOXY body . KP 50: Gel pH combined electrode, porous Teflon ring junction, suitable for emulsions, demineralised water, with S7 screw connector, glass body. KP 61: 3 diaphragm liquid filled pH combined electrode for wine, milk, cream, etc., S7 screw connector, liquid reference filling, glass body. KP 62: 1 diaphragm gel pH combined electrode for pure water, varnishes, gel filled, S7 screw connector, glass body. KP 63: 1 liquid filled pH combined electrode for general use, varnishes, 1m cable with BNC, glass body. KP 64: Liquid filled pH combined electrode ,Teflon ring diaphragm, for wine, varnishes, emulsions, S7 screw connector, glass body. KP 70: Pointed gel combined pH microelectrode diam. 6 x L=70 mm., with S7 screw connector, EPOXY body, glass tip, open junction. KP 80: Pointed gel pH combined electrode, with S7 screw connector, glass body, for cream, milk, viscous material, open junction. KP100: Flat membrane gel combined pH electrode with S7 screw connector, glass body, for skin, leather, paper. Characteristics and dimensions of the probes at page 401 CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 5: 5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 10: 10m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 15: 15m extension cable with BNC/S7 connector for electrode without cable, thread S7. CE : S7 screw connector for pH electrode. BNC: female BNC for extension cable
ORP Electrodes KP 90: REDOX PLATINUM liquid filled electrode with S7 screw connector, glass body. KP 91: Gel REDOX PLATINUM electrode, 1m cable with BNC, EPOXY body Characteristics and dimensions of the probes at page 402 pH Buffer solutions HD8642: Buffer solution 4.01pH - 200cc. HD8672: Buffer solution 6.86pH - 200cc. HD8692: Buffer solution 9.18pH - 200cc. Redox Buffer solutions HDR220: Redox buffer solution 220mV 0.5 l. HDR468: Redox buffer solution 468mV 0.5 l. Electrolyte solutions KCL3M Ready to use solution for electrode refilling 100 cc Cleaning and maintenance HD62PT: Diaphragm cleaning (tiourea in HCl) - 500ml. HD62PP: Protein cleaning (pepsin in HCl) - 500ml. HD62RF: Regeneration (fl uorhydric acid) - 100ml. HD62SC: Solution for electrode preservation - 200ml. Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP47.100: Direct 4 wires Pt100 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 4 wires with connector, length 2 m. TP47.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 2 wires with connector, length 2 m. TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes.
CP
HD22.3
Water Analysis
HD 2156.1 HD 2156.2
INSTRUMENT TECHNICAL CHARACTERISTICS Measured quantities: pH, mV, , , TDS, Nacl, C, F Instrument Dimensions (Length x Width x Height) 185x90x40mm Weight 470g (complete with batteries) Materials ABS, rubber Display 2x4 digits plus symbols Visible area: 52x42mm Operating conditions Working temperature Storage temperature Working relative humidity Protection degree -550C -2565C 090%RH without condensation IP67
Power Batteries 4 1.5V type AA batteries Autonomy 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20A Mains Output mains adapter 9Vdc / 250mA Security of memorized data Time Date and time Accuracy Unlimited, independent of battery charge conditions Schedule in real time 1min/month max error
Measured values storage - model HD2156.2 Type 2000 pages containing 10 samples each Quantity 20,000 sets of three measurements composed of pH or mV, or or TDS or salinity and temperature. Storage interval 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1h. Serial interface RS232C Type Baud rate Data bit Parity Stop bit Flow Control Serial cable length Immediate print interval USB interface - model HD2156.2 Type Connections pH/mV input Conductivity input Serial interface and USB Mains adapter Measurement of pH by Instrument Measurement range Resolution Accuracy Input impedance Calibration error @25C Measurement of mV by Instrument Measurement range Resolution Accuracy Drift after 1 year RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1 Xon/Xoff Max 15m 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1h. 1.1 - 2.0 electrically isolated Female BNC connector 8-pole male DIN45326 connector 8-pole MiniDin connector 2-pole connector (positive at centre) -2.000+19.999pH 0.01 or 0.001pH selectable from menu 0.001pH 1digit >1012 |Offset| > 20mV Slope > 63mV/pH or Slope < 50mV/pH Sensitivity > 106.5% or Sensitivity < 85% -1999.9+1999.9mV 0.1mV 0.1mV 1digit 0.5mV/year
Measurement of conductivity Measuring range 0.0019.99S/cm Kcell=0.1 Measuring range 0.0199.9S/cm Kcell=1 2001999S/cm 2.0019.99mS/cm 20.0199.9mS/cm Measuring range 2001999mS/cm Kcell=10 Accuracy (conductivity) 0.5%1digit Measurement of resistivity Measuring range Kcell=0.1 Measuring range Kcell=1 Measuring range Kcell=10 Accuracy (resistivity) till 100Mcm/(*) 5.0199.9cm 200999cm 1.00k19.99kcm 20.0k99.9kcm 100k999kcm 110Mcm 0.55.0cm 0.5%1digit
Measurement of salinity Measurement range 0.0001.999g/l 2.0019.99g/l 20.0199.9g/l Accuracy (salinity) 0.5%1digit Temperature compensation automatic/manual Reference temperature / TDS Conversion factor Cell constant K (cm-1)
0100C with T selectable from 0.00 to 4.00%/C 20C or 25C 0.40.8 0.1, 0.7, 1.0 and 10.0 147S/cm 1413S/cm 12880S/cm 111800S/cm -50+200C -50+200C 0.1C 0.25C 0.1C/anno K=0,01 - K=0,1 - K=1, K=10
Measurement of total dissolved solids (with coefficient /TDS=0.5) Measuring range 0.0019.99mg/l Kcell=0.1 Measuring range 0.0199.9mg/l Kcell=1 2001999mg/l 2.0019.99g/l 20.099.9g/l Measuring range 100999g/l Kcell=10 Accuracy (conductivity)
Standard solutions automatically detected @25C Measurement of temperature Pt100 measuring range Pt1000 measuring range Resolution Accuracy Drift after 1 year Preset cell constant values:
(*) The resistivity measurement is obtained from the reciprocal of conductivity measurement. Close to the bottom of the scale, the indication of resistivity appears like reported in the table below:
K cell = 0.01 cm-1 Conductivity (S/cm) Resistivity (Mcm) 0.001 S/cm 1000 Mcm 0.002 S/cm 500 Mcm 0.003 S/cm 333 Mcm 0.004 S/cm 250 Mcm K cell = 0.1 cm-1 Conductivity (S/cm) Resistivity(Mcm) 0.01 S/cm 100 Mcm 0.02 S/cm 50 Mcm 0.03 S/cm 33 Mcm 0.04 S/cm 25 Mcm
TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT 2 and 4 electrode conductivity probes
ORDER CODE MEASUREMENT RANGE
16 L=1.5m D=5 12 50
DIMENSIONS
156 20 17
SP06T
~72
40
1/2
16.2
30
14.5
27
17
56
35
120 12
SPT01G
L=1.5m D=5.5
35
130 12
SPT1G
L=1.5m D=5.5
35
130 12
SPT10G
L=1.5m D=5.5
Water Analysis
Temperature probes with connector 4 wire Pt100 and 2 wire Pt1000 sensor
Model TP47.100 TP47.1000 TP87.100 TP87.1000 Type Pt100 4 wires Pt1000 2 wires Pt100 4 wires Pt1000 2 wires Working range -50+200C -50+200C -50+200C -50+200C Accuracy Class A Class A Class A Class A
ORP Electrodes KP 90: REDOX PLATINUM liquid filled electrode with S7 screw connector, glass body. KP 91: Gel REDOX PLATINUM electrode, 1m cable with BNC, EPOXY body Characteristics and dimensions of the probes at page 402 pH Buffer solutions HD8642: Buffer solution 4.01pH - 200cc. HD8672: Buffer solution 6.86pH - 200cc. HD8692: Buffer solution 9.18pH - 200cc. Redox Buffer solutions HDR220: Redox buffer solution 220mV 0.5 l. HDR468: Redox buffer solution 468mV 0.5 l. Electrolyte solutions KCL3M Ready to use solution for electrode refilling 100 cc Cleaning and maintenance HD62PT: Diaphragm cleaning (tiourea in HCl) - 500ml. HD62PP: Protein cleaning (pepsin in HCl) - 500ml. HD62RF: Regeneration (fl uorhydric acid) - 100ml. HD62SC: Solution for electrode preservation - 200ml. Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP47.100: Direct 4 wires Pt100 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 4 wires with connector, length 2 m. TP47.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 2 wires with connector, length 2 m. TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes.
0.005%/C
ORDER CODES HD2156.1: The kit is composed of: instrument HD2156.1, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. Other pH electrodes, conductivity and temperature probes must be ordered separately. HD2156.2K: The kit is composed of: instrument HD2156.2 datalogger, 4 1.5V alkaline batteries, operating software DeltaLog9. Other pH electrodes, conductivity and temperature probes must be ordered separately. pH/mV probes, conductivity probes, temperature probes, standard calibration solutions for various types of measurements, connection cables for pH electrodes with S7 connector, cables for data transfer to PC or printer have to be ordered separately. HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C. C.206: Serial connection cable with USB connector for PC and 8-pole MiniDin male connector for the instrument. HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin (not suitable for HD2156.1K). DeltaLog9: Software for download and management of the data on PC using Windows 98 to Vista operating systems. SWD10: Stabilized power supply 100-240 Vac/12Vdc-1A mains voltage HD40.1: 24-column portable thermal printer, serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls. RCT: The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter. BAT-40: Spare battery pack for HD40.1 printer with built-in temperature sensor. HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by bench-top meters of the series HD22with cable HD22.2.1 (optional) or power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm. pH Electrodes KP 20: Gel pH filled combined electrode for general use , with S7 screw connector, EPOXY body. KP 30: Gel pH combined electrode for general use, 1m cable with BNC, EPOXY body . KP 50: Gel pH combined electrode, porous Teflon ring junction, suitable for emulsions, demineralised water, with S7 screw connector, glass body. KP 61: 3 diaphragm liquid filled pH combined electrode for wine, milk, cream, etc., S7 screw connector, liquid reference filling, glass body. KP 62: 1 diaphragm gel pH combined electrode for pure water, varnishes, gel filled, S7 screw connector, glass body. KP 63: 1 liquid filled pH combined electrode for general use, varnishes, 1m cable with BNC, glass body. KP 64: Liquid filled pH combined electrode ,Teflon ring diaphragm, for wine, varnishes, emulsions, S7 screw connector, glass body. KP 70: Pointed gel combined pH microelectrode diam. 6 x L=70 mm., with S7 screw connector, EPOXY body, glass tip, open junction. KP 80: Pointed gel pH combined electrode, with S7 screw connector, glass body, for cream, milk, viscous material, open junction. KP100: Flat membrane gel combined pH electrode with S7 screw connector, glass body, for skin, leather, paper. Characteristics and dimensions of the probes at page 401 CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 5: 5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 10: 10m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 15: 15m extension cable with BNC/S7 connector for electrode without cable, thread S7. CE : S7 screw connector for pH electrode. BNC: female BNC for extension cable
HD 9609
TECHNICAL CHARACTERISTICS pH simulation: pH resolution: pH accuracy 2025C: Thermal drift: mV simulation: mV resolution: mV accuracy: Thermal drift mV scale: mV thermal drift: Noise 010 Hz: Simulation of compensation temperature: Output impedance: Display: Symbols: Signals: Working temperature: Power supply: Consumption (instrument only): Autonomy: Dimensions: Weight:
014 pH 0.1 pH 0.002 pH 0.0005 pH/C from -5C to 20C and from 25C to 50 1999 mV 1 mV 100 V -199.9 ... +199.9: 0.01 mV/C from -5 to 20C and from 25 to 50C -1999 ... +1999: 0.05 mV/C from -5 to 20C and from 25 to 50C 1V peak/peak -20 to 150C (-4 to 302F) 100 K 1%, 1G 5% (no leading load capacity) LCD 2 lines, 3 digits. Figure height approx. 12.5 mm. pH, mV, C, F, HI imp., LO imp., 0.1 pH, 1 pH, 1 mV, 10 mV LOU, ER1, CAL -5 to 50C (23 to 122F) 9 Vdc alkaline battery. Low battery indication. 5 mA lit, 20 A off about 200 hours 187 x 72 x 38 mm. 300 gr
Water Analysis
ORDER CODES HD 9609: Kit composed of the instrument HD 9609, adapter cables CP 9509BNC, CP 9509 T, carrying case CP 9509BNC: Adapter cable L = 1 mt, male BNC connector on both ends CP 9509 T: Adapter cable L = 1 mt, BNC connector on only one end CP9509S7: Adapter cable L = 1 mt, BNC wall connector one end, S7 male connector on the other end.
DO 9403T-R1
pH 15 14 Pt100 4 wires
9 mA B
GND 8 4...20mA 7 4 3 2 1
13 12 11 10
temp
pH
15 14 Pt100 4 wires
9 mA B 8 7 4 3 2 1
13 12 11 10
temp
pH
13 12 11 10
temp
Min 9Vcc
9 8 7
4 3 2 1
+
A
pH
15 14 Pt100 4 wires
mA B
Regulator
13 12 11 10
4...20 mA
temp
Min 9Vcc
5 6
9 8 7
B 4 3 2 1
+
A
pH
15 14 Pt100 2 wires
13 12 11 10
4...20 mA
V+
Regulator
temp
Fig.4
OK CAL
Confirms the programming parameters, or the relay SET values, and stores them. - When pressed together with the C/F key it activates the manual temperature setting function. - When pressed together with the pH/mV key it activates the pH calibration function. - Key used to confirm pH calibration and manual temperature setting. - Key for increasing the value displayed in the parameter programming phase. - During the relay SET point programming phase. - During the calibration phase. - Key for decreasing the value displayed in the parameter programming phase. - During the relay SET point programming phase. - During the calibration phase.
- Using the and keys, adjust the pH value measured as a function of the liquid temperature. - Press CAL to confirm this value. The symbol switches off. NOTE: If you want to quit without storing the new calibration, press the C/F key. N.B.: The instrument can automatically recognize three standard calibration solutions: 4.01 pH, 6.86 pH and 9.18 pH. Programming the parameters P1 Relay control unit and analog output, pH or mV. P2 pH/mV value corresponding to 4 mA at output. May be set between -1.00 pH and 15.00 pH or between -1999 mV and +1999 mV. P3 pH/mV value corresponding to 20 mA at output. May be set between -1.00 pH and 15.00 pH or between -1999 mV and +1999 mV. P4 Delay time in the intervention of relay A. May be set between 0 and 250 seconds. P5 Delay time in the intervention of relay B. May be set between 0 and 250 seconds. P6 Calibration of Pt100 probe. P7 4 mA output current calibration. P8 20 mA output current calibration. P9 Input voltage calibration. P10 Display of the offset voltage value and of the slope value of the electrode. To change one of these parameters press key PRG until the message corresponding to the parameter to be changed appears on the screen. Press OK to show the value of the parameter. Using the and keys, bring the parameter displayed to the desired value. Press OK again to confirm. Parameter P10 cannot be altered, it can only be displayed. NOTE P6-P7-P8-P9: calibration procedure to be carried out at a laboratory by skilled workers. Pt100 probe calibration (100 at 0C) - Connect the Pt100 probe to the instrument. Press the PRG key until the message P6 appears on the display. - Press the OK key; the currently measured temperature appears on the display. - Immerse the Pt100 probe and a precision thermometer for reference in the zero calibration bath. Wait long enough for the reading to become stable.
Combined electrode input pH ORP Input impedance Cable length Accuracy Pt100 2/4 wires Transducer energizing Cable length Accuracy -1.00 pH...15.00 pH (-500...+500 mV) -1999...+1999 mV >10 Tohm <50 metres screened (about 5 nF) 0.1% of reading 1 digit 0.01% of pH per C of temperature drift -50...199.9C 0.5 mA DC <10 metres unscreened <20 metres screened (about 2 nF) 0.2C 0.1% of reading 2 digits 0.01C/C
Setting the relay SET point - Press the SET button; the symbol appears on the display. - The REL symbol and the letter A also light up on the display to indicate that the value shown corresponds to the switching on threshold of relay A. - To change this value press the and keys. - Press SET; the REL symbol flashes and the letter A remains lit to indicate that the value shown corresponds to the switching off threshold of relay A. - To change this value press the and keys. - Press the SET button; the REL symbol and the letter B light up to indicate that the value shown corresponds to the switching off threshold of relay B. - To change this value press the and keys. - Press SET; the REL symbol flashes and the letter B remains lit to indicate that the value shown corresponds to the switching off threshold of relay B. - To change this value press the and keys. - Press SET, the instrument stores the set parameters and returns to normal function. The REL and symbols disappear. NOTE: During the SET point setting phase (REL symbol lit or flashing) the instrument returns to normal function if no key is pressed for 2 minutes. Manual temperature setting If the temperature probe is not connected or if the probe is broken the measuring unit C or F flashes. In this case it is possible to set the temperature compensation value manually. - Press the CAL key and the C/F key together; the symbol appears and the manual temperature is shown with the measuring unit flashing. - Using the and keys, set the temperature value corresponding to the temperature of the liquid in which you wish to measure the pH. - Press CAL to confirm this value. The symbol switches off and the instrument returns to the previous display. Calibration of the pH electrode Calibration of the offset of the pH electrode: - Immerse the electrode in the buffer solution used for calibrating the offset (6.86 pH). - Press the CAL key and the pH/mV key together; the symbol lights up on the display. - Using the and keys, adjust the pH value measured as a function of the liquid temperature. - Press CAL to confirm this value. The symbol switches off. Calibration of the slope of the pH electrode: - Immerse the electrode in the buffer solution used for calibrating the slope (4.01 or 9.18 pH). - Press the CAL key and the pH/mV key together; the symbol lights up on the display.
Temperature input
Automatic
According to Nerst
Current output
4.00...20.00 mA Accuracy Insulation A and B Active Passive Operation Storage External dimensions Protection class
Programmable and proportional to the pH or mV value 0.5% of reading 0.02 mA 2500 Vac 1 minute Bistable, contact 3 A/230 Vac free potential 24 or 230 Vac -15/+10% 1 VA, 48...62 Hz, see fig. 1 420 mA, 2 wire configuration, 1035 V, see fig. 2 0...50C -20...70C, no condensation 120x80x56 mm IP64
56.0 mm
120.0 mm
80.0 mm
21.0 mm
Dimensions
Case
Temp.
Power supply
Relay output
Water Analysis
- Using the and keys, adjust the value of the temperature measured by the Pt100 probe so that it corresponds with the value on the reference thermometer. - Immerse the Pt100 probe and a precision thermometer in the full scale calibration bath. Wait long enough for the reading to become stable. - Using the and keys, adjust the value of the temperature measured by the Pt100 probe so that it corresponds with the value on the reference thermometer. - Press OK to confirm. To quit programming, press PRG repeatedly. N.B.: If the temperature shown by the instrument is between 12C, the instrument calibrates the probe offset, otherwise it calibrates the gain. Calibrating the analog output - Connect a precision milliammeter to the analog output. - Press the PRG key until the message P7 appears on the display. - Press OK; the message 4.0 appears on the display, indicating calibration at 4 mA. - Using the and keys, adjust the value of the output current so as to have an indication of 4,00 mA on the precision milliammeter. - Press the PRG key until the message P8 appears on the display. - Press OK; the message 20.0 appears on the display, indicating calibration at 20 mA. - Using the and keys, adjust the value of the output current so as to have an indication of 20,00 mA on the precision milliammeter. - Press OK to confirm. To quit programming, press PRG repeatedly. Calibrating the voltage input - Press the PRG key until the message P9 appears on the display. - Press OK; the mV value of the input appears on the display. - Simulate a voltage of 0 mV at the input (if the value is between 25 mV the zero is calibrated, otherwise the full scale value is calibrated). - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Press the SET key, the REL symbol lights up on the display indicating that the instrument is measuring the voltage present at the input using the second measurement scale. - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Press the SET key, the REL symbol on the display switches off. - Simulate a voltage of 450 mV at the input, corresponding to the full scale value of the first scale. - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Simulate a voltage of 1800 mV at the input, corresponding to the full scale value of the second scale. - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Press OK to confirm. To quit programming, press PRG repeatedly. Display Symbol C F pH mV A B REL description indicates that the value shown is in C. indicates that the value shown is in F. indicates that the unit of the value shown is pH. Indicates that the unit of the value shown is milli Volts. indicates that the relay A is in closed status. indicates that the relay B is in closed status. - indicates that the value shown corresponds to the closing thresholds of the contacts of relay A or B; - indicates that the offset of the second voltage measurement scale is being calibrated. REL flashing indicates that the value shown corresponds to the opening thresholds of the contacts of relay A or B. - indicates that the instrument is in the parameter setting phase; - indicates that the closing and opening thresholds of relays A and B are being changed; - indicates that the manual compensation temperature is being changed; - indicates that the pH electrode is being calibrated.
Error signals OFL - Warning which appears during measurement when the value to be displayed is out of scale. E1 - Error warning which appears during pH calibration to indicate that the offset value of the electrode is too high in absolute value. E2 - Error warning which appears during pH calibration to indicate that the difference between the mV readings given by the two buffer solutions used for calibration is too great. E3 - Error warning which appears during pH calibration to indicate that the mV readings given by the two buffer solutions used for calibration are too close (about 50 mV at 25C). E4 - Reading error on the EEPROM. E5 - Error warning indicating that the slope calculation gives a value 20% lower than the nominal value or gives a negative value. E6 - Error warning indicating that the slope calculation gives a value 150% lower than the nominal value. Order code DO 9403T-R1: pH transmitter 420 mA passive or active, power supply 24 Vac, 120x80x56 mm for use in the field. HD 882 M100/300: Temperature probe with Pt100 sensor, miniature head, shaft 6x300 mm. HD 882 M100/600: Temperature probe with Pt100 sensor, DIN B head, shaft 6x600 mm. HD 8642: Buffer solution 4.01 pH. HD 8672: Buffer solution 6.86 pH. HD 8692: Buffer solution 9.18 pH. HD 220: Buffer solution redox 220 mV HD 468: Buffer solution redox 468 mV HD62PT: Diaphragm cleaning (tiourea in HCl) - 500ml. HD62PP: Protein cleaning (pepsin in HCl) - 500ml. HD62RF: Regeneration (fluorhydric acid) - 100ml. HD62SC: Solution for electrode preservation - 200ml CP5: Extension cable. Connector S/wire - TERMINAL BOARD. CP5/10: Extension cable L=10m.Connector BNC/S7. CP5S: Extension cable L=5m. Connector BNC/S7. CP5S/10: Extension cable L=10m. Connector BNC/S7. KPI 10: Combined industrial electrode, S7 brass 1 connector, glass body, Ag/AgCl sat KCI, 12x120, temperature 0130C, porous Teflon fitting. KPI 11: Combined industrial electrode, S7 brass 1 connector, Rytron body, Ag/AgCl sat KCI, temperature 0100C, porous Teflon fitting. KPI 12: Platinum electrode for Redox measurement,S PG13,5 connector, pressure 6 bar. KPI 13: Platinum electrode for Redox measurement, Rytron body, S PG13,5 connector, Ag/AgCl sat KCl. Electrode dimensions at page 351
DO 9785T DO 9765T
Technical characteristics
pH ORP Input impedance Cable length Accuracy Temperature input Pt100 2/4 wires Transducer energizing Cable length Accuracy Automatic pH electrode compensation temperature -1,00 pH...15,00 pH (-500...+500 mV) -1999...+1999 mV >10 Tohm <50 metres screened (about 5 nF) 0.1% of reading 1 digit 0.01% pH per C of drift in temperature -50...199,9C 0,5 mA DC <10 metres unscreened <50 metres screened (about 2 nF) 0.2C 0.1% of reading 2 digits 0.01C/C According to Nernst
Manual
-50200C
Current output
Programmable and proportional to the pH or mV value 0.5% of reading 0.02 mA 2500 Vac 1 minute RLmax = Vdc-10 0,022 RLmax = 636 @Vdc = 24 Vdc
Bistable, contact 3A/230 Vac free potential 420 mA, 2 wire configuration, 1035 V, see fig. 2 24 or 230 Vac - 15/+10% 1 VA, 48...62 Hz, see fig. 1 120x122x56 mm IP64 96x96x126 mm IP54
Power supply
Key functions PRG Programming of the parameters is activated by pressing the PRG key plus the and keys. The message P1 appears on the display, indicating that the parameter P1 is being programmed. When the PRG key is pressed continuously, the messages P2, P3, P4, P5, P6, P7 and the corresponding parameters are displayed in sequence. After P7 the instrument returns to normal function. SET Key for setting the relay intervention threshold. The ON or OFF symbol appears on the display, indicating the switching on or off threshold of relay A or of relay B. C/F - If this key is pressed it changes the temperature measuring unit to degrees Celsius or degrees Fahrenheit. - When pressed together with the CAL key it activates the manual temperature setting function. - If pressed during the conductivity calibration function it quits the calibration function without storing the calibration. pH/mV - If this key is pressed it changes the measuring unit to mV or pH. - When pressed together with the CAL key it activates the pH calibration function. OK Confirms the programming parameters, or the relay SET values, and stores them. CAL - When pressed together with the C/F key it activates the manual temperature setting function. - When pressed together with the pH/mV key it activates the pH calibration function. - Key used to confirm pH calibration and manual temperature calibration. - Key for increasing the value displayed in the parameter programming phase. - During the relay SET point programming phase. - During the calibration phase. - Key for decreasing the value displayed in the parameter programming phase. - During the relay SET point programming phase. - During the calibration phase.
Water Analysis
DO 9785T case
DO 9765T case
Setting the relay SET point - Press the SET button; the ON symbol appears on the display with the letter A to indicate that the value shown corresponds to the switching on threshold of relay A. - To change this value press the and keys. - Press SET; the OFF symbol appears with the letter A to indicate that the switching off threshold of relay A is being displayed. - To change this value press the and keys. - Press the SET button; the ON symbol appears on the display with the letter B to indicate that the value shown corresponds to the switching on threshold of relay B. - To change this value press the and keys. - Press SET; the OFF symbol appears with the letter B to indicate that the switching off threshold of relay B is being displayed. - To change this value press the and keys. - Press SET, the instrument stores the values and returns to normal function. NOTE: During the SET point setting phase (symbols ON or OFF lit) the instrument returns to normal function if no key is pressed for 2 minutes. Temperature setting for manual compensation If the temperature probe is not connected or if the probe is broken the measuring unit C or F flashes. In this case it is possible to set the temperature compensation value manually. - Press the CAL key and the C/F key together; the message CAL appears at the bottom of the display. - Using the and keys, set the temperature value corresponding to the temperature of the liquid in which you wish to measure the pH value. - Press CAL to confirm this value. The message CAL disappears. Calibration of the DO 9785T/DO 9765T with pH electrode Calibration of the offset of the pH electrode: - Immerse the electrode in the buffer solution used for calibrating the offset (6.86 pH). - Press the CAL key and the pH/mV key together; the message CAL appears at the top of the display. - Using the and keys, adjust the pH value measured as a function of the liquid temperature. - Press CAL to confirm this value. The message CAL disappears. Calibration of the slope of the pH electrode: - Immerse the electrode in the buffer solution used for calibrating the slope (4.01 or 9.18 pH). - Press the CAL key and the pH/mV key together; the message CAL appears at the top of the display. - Using the and keys, adjust the pH value measured as a function of the liquid temperature. - Press CAL to confirm this value. The message CAL disappears. NOTE: If you want to quit without storing the new calibration, press the C/F key. N.B.: The instrument can automatically recognize three standard calibration solutions: 4.01 pH, 6.86 pH and 9.18 pH. Programming the parameters P1 Relay control unit and analog output, pH or mV. P2 pH/mV value corresponding to 4 mA at output. May be set between -1.00 pH and 15.00 pH or between -1999 mV and +1999 mV. P3 pH/mV value corresponding to 20 mA at output.. May be set between -1.00 pH and 15.00 pH or between -1999 mV and +1999 mV. P4 Delay time in the intervention of relay A. May be set between 0 and 255 seconds. P5 Delay time in the intervention of relay B. May be set between 0 and 255 seconds. P6 Calibration of Pt100 probe, calibration of output in current, calibration of output in voltage. (Calibration procedure to be carried out at a laboratory by skilled personnel). P7 Display of the offset voltage value and of the slope value of the electrode. To change one of these parameters press key PRG until the message corresponding to the parameter to be changed appears on the screen. Using the and keys, bring the parameter displayed to the desired value. Press OK to confirm. Parameter P7 cannot be altered. Calibrating the voltage input (calibration procedure to be carried out at a laboratory by skilled workers) - Press the PRG key until the message P6 appears on the display. - Press the CAL key four times; the message CAL appears at the top of the display and the mV value of the input at the bottom. - Simulate a voltage of 0 mV at the input (if the value is between 25 mV the zero is calibrated, otherwise the full scale value is calibrated). - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Press the SET button, the ON symbol appears on the display to indicate that the instruments is measuring the voltage of the input using the second scale of measurement.
pH pH Pt100 4 wire 15 14 13 12
=
mA B
11
GND 10 4...20mA 9
8 7 6 5 4 3
temp
pH
13 12 11 10
temp
Min 9Vcc V 11 10
Regulator
9 4...20 mA
Regulator
Fig.3
56.0 mm 121.5 mm
120.0 mm
PG7
DO 9765T
126.0 mm
96.0 mm
96.0 mm
21.0 mm
DO 9785T
15.0 mm
Dimensions
- Using the and keys adjust the voltage value so as to have the correct voltage value on the display. - Press the set button, on the display the symbol ON disappears - Simulate a voltage of 450 mV at the input, corresponding to the full value of the first scale. - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Simulate a voltage of 1800 mV at the input, corresponding to the full value of the second scale. - Using the and keys, adjust the voltage value so as to have the correct voltage value on the display. - Press OK to confirm. Pt100 probe calibration (100 at 0C) (calibration procedure to be carried out at a laboratory by skilled workers) - Connect the Pt100 probe to the instrument. Press the PRG key until the message P6 appears on the display. - Press the CAL key; the message CAL appears at the bottom of the display and the temperature is shown at the top. - Immerse the Pt100 probe and a precision thermometer for reference in the zero calibration bath. Wait long enough for the reading to become stable. - Using the and keys, adjust the value of the temperature measured by the Pt100 probe so that it corresponds with the value on the reference thermometer. - Immerse the Pt100 probe and a precision thermometer in the full scale calibration bath. Wait long enough for the reading to become stable. - Using the and keys, adjust the value of the temperature measured by the Pt100 probe so that it corresponds with the value on the reference thermometer. - Press OK to confirm. N.B.: If the temperature shown by the instrument is between 12C, the instrument calibrates the probe offset, otherwise it calibrates the gain. Calibrating the analog output (calibration procedure to be carried out at a laboratory by skilled workers) - Press the PRG key until the message P6 appears on the display. - Connect a precision milliammeter to the analog output. - Press the CAL key twice; the message CAL appears at the top of the display and the message 4.0 at the bottom, indicating calibration at 4 mA. - Using the and keys, adjust the value of the output current so as to have an indication of 4.0 mA on the precision milliammeter. - Press the CAL key; the message CAL appears at the top of the display and the message 20.0 at the bottom, indicating calibration at 20 mA. - Using the and keys, adjust the value of the output current so as to have an indication of 20.0 mA on the precision milliammeter. - Press OK to confirm. Display Symbol C F pH mV A B ON OFF description the value shown is in C. the value shown is in F. the unit of the value shown is pH. the unit of the value shown is milli Volts. the relay A is in closed status. the relay B is in closed status. the value shown corresponds to the closing thresholds of the contacts of relay A or B. the value shown corresponds to the opening thresholds of the contacts of relay A or B.
HD 8642: Buffer solution 4.01 pH. HD 8672: Buffer solution 6.86 pH. HD 8692: Buffer solution 9.18 pH. HDR 220: Buffer solution Redox 220 mV 0,5l. HDR 468: Buffer solution Redox 468 mV 0,5l. HD62PT: Diaphragm cleaning (tiourea in HCl) - 500ml. HD62PP: Protein cleaning (pepsin in HCl) - 500ml. HD62RF: Regeneration (fluorhydric acid) - 100ml. HD62SC: Solution for electrode preservation - 200ml. CP5/10: Extension cable for connecting the electrode to the DO9403T or to the DO9765T (S7-wire-TERMINAL BOARD) L=10m. CP5: Extension for connecting the electrode to the DO 9403T-R1 or to the DO 9765T (S7-wire-TERMINAL BOARD) L=5m. CP5S: Extension for connecting the electrode to the DO 9785T (BNC-S7) L=5m. CP5S/10: Extension cable L=10m. Connector BNC/S7. KPI 10: Combined industrial electrode, S7 PG13.5 connector, refillable, glass body, Ag/AgCl sat KCl 12x120 mm, temperature 0130C, porous Teflon fitting. KPI 11: Combined industrial electrode, S7 brass 1 connector, refillable, Rytron body, Ag/AgCl sat KCl, temperature 0100C, porous Teflon fitting. KPI 12: Platinum electrode for Redox measurement,S PG13,5 connector, pressure 6 bar. KPI 13: Platinum electrode for Redox measurement, Rytron body, S PG13,5 connector, Ag/AgCl sat KCl.
Error signal OFL - Warning which appears during measurement when the value to be displayed is out of scale. E1 - Error warning which appears during pH calibration to indicate that the offset value of the electrode is too high in absolute value. E2 - Error warning which appears during pH calibration to indicate that the difference between the mV readings given by the two buffer solutions used for calibration is too great. E3 - Error warning which appears during pH calibration to indicate that the mV readings given by the two buffer solutions used for calibration are too close (about 50 mV at 25C). E4 - Reading error on the EEPROM. E5 - Error warning indicating that the slope calculation gives a value 20% lower than the nominal value or gives a negative value. E6 - Error warning indicating that the slope calculation gives a value 150% higher than the nominal value. Order code DO 9785T: pH transmitter 420 mA passive or active, power supply 24 Vac with double display 96x96 mm, for panel mounting. DO 9765T: pH transmitter 420 mA passive or active, power supply 24 Vac with double display 122x120, for use on the field. HD 882 M100/300: Temperature probe with Pt100 sensor, miniature head, shaft 6x300 mm. Water Analysis 351
Water Analysis
HD 2106.1 HD 2106.2
INSTRUMENT TECHNICAL CHARACTERISTICS Measured quantities: , , TDS, Nacl, C, F Instrument Dimensions (Length x Width x Height) 185x90x40mm Weight 470g (complete with batteries) Materials ABS, rubber Display 2x4 digits plus symbols Visible area: 52x42mm Operating conditions Working temperature Storage temperature Working relative humidity Protection degree Power Batteries Autonomy -550C -2565C 090%RH without condensation IP67
4 1.5V type AA batteries 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20A Mains Output mains adapter 12Vdc / 1A Security of memorized data Time Date and time Accuracy Unlimited, independent of battery charge conditions Schedule in real time 1min/month max error
Measured values storage - model HD2106.2 Type 2000 pages containing 18 samples each Quantity 36000 pairs of measurements [-C], [-C], [TDS-C] or [Sal-C] Selectable storage interval 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour) Serial interface RS232C Type RS232C electrically isolated Baud rate Can be set from 1200 to 38400 baud Data bit 8 Parity None Stop bit 1 Flow Control Xon/Xoff Serial cable length Max 15m Selectable print interval 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour) USB interface - model HD2106.2 Type Connections Conductivity input Input module for the temperature probes Serial interface and USB Mains adapter 1.1 - 2.0 electrically isolated 8-pole male DIN45326 connector 8-pole male DIN45326 connector 8-pole MiniDin connector 2-pole connector (positive at centre) Resolution 0.01S/cm 0.1S/cm 1S/cm 0.01mS/cm 0.1mS/cm 1mS/cm
Measurement of conductivity Measuring range 0.0019.99S/cm Kcell=0.1 Measuring range 0.0199.9S/cm Kcell=1 2001999S/cm 2.0019.99mS/cm 20.0199.9mS/cm Measuring range 2001999mS/cm Kcell=10 Accuracy (conductivity) 0.5%1digit
HD40.1
SWD10
Measurement of resistivity Measuring range till 100Mcm/(*) Kcell=0.1 Measuring range 5.0199.9cm Kcell=1 200999cm 1.00k19.99kcm 20.0k99.9kcm 100k999kcm 110Mcm Measuring range 0.55.0cm Kcell=10 Accuracy (resistivity) 0.5%1digit
Temperature compensation automatic/manual Reference temperature / TDS Conversion factor Cell constant K (cm-1)
0100C with T selectable from 0.00 to 4.00%/C 20C or 25C 0.40.8 0.1, 0.7, 1.0 and 10.0
Standard solutions automatically detected @25C 147S/cm 1413S/cm 12880S/cm 111800S/cm Preset cell constant values:
K cell = 0.01 cm-1 Conductivity (S/cm) 0.001 S/cm 0.002 S/cm 0.003 S/cm 0.004 S/cm Resistivity (Mcm) 1000 Mcm 500 Mcm 333 Mcm 250 Mcm
Measurement of total dissolved solids (with coefficient /TDS=0.5) Measuring range 0.0019.99mg/l 0.05mg/l Kcell=0.1 Measuring range 0.0199.9mg/l 0.5mg/l Kcell=1 2001999mg/l 1mg/l 2.0019.99g/l 0.01g/l 20.099.9g/l 0.1g/l Measuring range 100999g/l 1g/l Kcell=10 Accuracy (conductivity) 0.5%1digit Measurement of salinity Measurement range 0.0001.999g/ 2.0019.99g/l 20.0199.9g/l Accuracy (salinity) 0.5%1digit Measurement of temperature Pt100 measuring range Pt1000 measuring range Resolution Accuracy Drift after 1 year -50+200C -50+200C 0.1C 0.25C 0.1C/year Resolution l 1mg/l 10mg/l 0.1g/l
(*) The resistivity measurement is obtained from the reciprocal of conductivity measurement. Close to the bottom of the scale, the indication of resistivity appears like reported in the table below: 4 wire Pt100 and 2 wire Pt1000 Temperature probes
Model TP47.100 TP47.1000 TP87.100 TP87.1000 Type Pt100 4 wires Pt1000 2 wires Pt100 4 wires Pt1000 2 wires Working range -50+200C -50+200C -50+200C -50+200C Accuracy Class A Class A Class A Class A
0.005%/C
Water Analysis
ORDER CODES HD2106.1: The kit is composed of: instrument HD2106.1, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. HD2106.2: The kit is composed of: instrument HD2106.2 datalogger, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. Conductivity probes, temperature probes, standard calibration solutions, cables for data transfer to PC or printer have to be ordered separately. HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C. C.206: Serial connection cable with USB connector for PC and 8-pole MiniDin male connector for the instrument HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin (not suitable for HD2106.1K). DeltaLog9: Software for download and management of the data on PC using Windows 98 to Vista operating systems. SWD10: Stabilized power supply 100-240 Vac/12Vdc-1A mains voltage HD40.1: 24-column portable thermal printer, serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls. RCT: The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter. BAT-40: Spare battery pack for HD40.1 printer with built-in temperature sensor. HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by bench-top meters of the series HD22with cable HD22.2.1 (optional) or power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm. Conductivity probes Please see the order codes reported in the probes technical specifications.
Standard conductivity calibration solutions HD8747: Standard calibration solution 0.001mol/l equal to 147S/cm @25C, 200cc. HD8714: Standard calibration solution 0.01mol/l equal to 1413S/cm @25C, 200cc. HD8712: Standard calibration solution 0.1mol/l equal to 12880S/cm @25C, 200cc. HD87111: Standard calibration solution 1mol/l equal to 111800S/cm @25C, 200cc. Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes
TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT 2 and 4 electrode conductivity probes
ORDER CODE MEASUREMENT RANGE
16 L=1.5m D=5 12 50
DIMENSIONS
156 20 17
~72 40 1/2 16.2 30 14.5 27 17 56
SP06T
35
120 12
SPT01G
L=1.5m D=5.5
35
130 12
SPT1G
L=1.5m D=5.5
35
130 12
SPT10G
L=1.5m D=5.5
HD 2306.0
Connections Conductivity input/temperature probes 8-pole male DIN45326 connector Measurement of conductivity Measuring range 0.0019.99S/cm Kcell=0.1 Measuring range 0.0199.9S/cm Kcell=1 2001999S/cm 2.0019.99mS/cm 20.0199.9mS/cm Measuring range 2001999mS/cm Kcell=10 Accuracy (conductivity) 0.5%1digit Measurement of resistivity Measuring range Kcell=0.1 Measuring range Kcell=1 Measuring range Kcell=10 Accuracy (resistivity) till 100Mcm/(*) 5.0199.9cm 200999cm 1.00k19.99kcm 20.0k99.9kcm 100k999kcm 110Mcm 0.55.0cm 0.5%1digit Resolution l 1mg/l 10mg/l 0.1g/l 0.1cm 1cm 0.01kcm 0.1kcm 1kcm 1Mcm 0.1cm Resolution 0.01S/cm 0.1S/cm 1S/cm 0.01mS/cm 0.1mS/cm 1mS/cm
Measurement of salinity Measurement range 0.0001.999g/ 2.0019.99g/l 20.0199.9g/l Accuracy (salinity) 0.5%1digit Measurement of total dissolved solids (with coefficient /TDS=0.5) Measuring range 0.0019.99mg/l Kcell=0.1 Measuring range 0.0199.9mg/l Kcell=1 2001999mg/l 2.0019.99g/l 20.099.9g/l Measuring range 100999g/l Kcell=10 Accuracy (conductivity) 0.5%1digit Measurement of temperature Pt100 measuring range Pt1000 measuring range Resolution Accuracy Drift after 1 year Temperature compensation automatic/manual Reference temperature / TDS Conversion factor Cell constant K (cm-1) -50+200C -50+200C 0.1C 0.25C 0.1C/year
0100C with T selectable from 0.00 to 4.00%/C 20C or 25C 0.40.8 0.1, 0.7, 1.0 and 10.0
Standard solutions automatically detected @25C 147S/cm 1413S/cm 12880S/cm 111800S/cm Preset cell constant values: K=0,01 - K=0,1 - K=1, K=10
Water Analysis
(*) The resistivity measurement is obtained from the reciprocal of conductivity measurement. Close to the bottom of the scale, the indication of resistivity appears like reported in the table below:
K cell = 0.1 cm-1 Conductivity (S/cm) 0.01 S/cm 0.02 S/cm 0.03 S/cm 0.04 S/cm Resistivity (Mcm) 100 Mcm 50 Mcm 33 Mcm 25 Mcm
Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes.
0.005%/C
ORDER CODES HD2306.0K: The kit is composed of: instrument HD2306.0, 3 1.5V alkaline batteries, operating manual, case. Other conductivity probes, temperature probes, calibration solutions must be ordered separately. HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by bench-top meters of the series HD22with cable HD22.2.1 (optional) or power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm. Conductivity probes Please see the order codes reported in the probes technical specifications. Standard conductivity calibration solutions HD8747: Standard calibration solution 0.001mol/l equal to 147S/cm @25C, 200cc. HD8714: Standard calibration solution 0.01mol/l equal to 1413S/cm @25C, 200cc. HD8712: Standard calibration solution 0.1mol/l equal to 12880S/cm @25C, 200cc. HD87111: Standard calibration solution 1mol/l equal to 111800S/cm @25C, 200cc.
HD8747
HD8714
HD8712
HD8711
TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT 2 and 4 electrode conductivity probes
ORDER CODE MEASUREMENT RANGE
16 L=1.5m D=5 12 50
DIMENSIONS
156 20 17
SP06T
35
120 12
SPT01G
L=1.5m D=5.5
35
130 12
SPT1G
L=1.5m D=5.5
35
130 12
SPT10G
L=1.5m D=5.5
HD 2156.1 HD 2156.2
INSTRUMENT TECHNICAL CHARACTERISTICS Measured quantities: pH, mV, , , TDS, Nacl, C, F Instrument Dimensions (Length x Width x Height) 185x90x40mm Weight 470g (comple te with batteries) Materials ABS, rubber Display 2x4 digits plus symbols Visible area: 52x42mm Operating conditions Working temperature Storage temperature Working relative humidity Protection degree -550C -2565C 090%RH without condensation IP67
Power Batteries 4 1.5V type AA batteries Autonomy 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20A Mains (SWD10) Output mains adapter 12Vdc / 1A Security of memorized data Unlimited, independent of battery charge conditions Time Date and time Accuracy Schedule in real time 1min/month max error
Serial interface RS232C Type Baud rate Data bit Parity Stop bit Flow Control Serial cable length Selectable print interval
RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1 Xon/Xoff Max 15m 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour) 1.1 - 2.0 electrically isolated Female BNC connector 8-pole male DIN45326 connector 8-pole MiniDin connector 2-pole connector (positive at centre) -2.000+19.999pH 0.01 or 0.001pH selectable from menu 0.001pH 1digit >1012 |Offset| > 20mV Slope > 63mV/pH or Slope < 50mV/pH Sensitivity > 106.5% or Sensitivity < 85% -1999.9+1999.9mV 0.1mV 0.1mV 1digit 0.5mV/year
USB interface - model HD2156.2 Type Connections pH/mV input Conductivity input Serial interface and USB Mains adapter Measurement of pH by Instrument Measurement range Resolution Accuracy Input impedance Calibration error @25C Measurement of mV by Instrument Measurement range Resolution Accuracy Drift after 1 year
Water Analysis
Measured values storage - model HD2156.2 Type 2000 pages containing 10 samples each Quantity 20,000 sets of three measurements composed of pH or mV, or or TDS or salinity and temperature. Selectable storage interval 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour)
Measurement of conductivity Measuring range 0.0019.99S/cm Kcell=0.1 Measuring range 0.0199.9S/cm Kcell=1 2001999S/cm 2.0019.99mS/cm 20.0199.9mS/cm Measuring range 2001999mS/cm Kcell=10 Accuracy (conductivity) 0.5%1digit Measurement of resitivity Measuring range till 100Mcm/(*) Kcell=0.1 Measuring range 5.0199.9cm Kcell=1 200999cm 1.00k19.99kcm 20.0k99.9kcm 100k999kcm 110Mcm Measuring range 0.55.0cm Kcell=10 Accuracy (resistivity) 0.5%1digit Measurement of total dissolved solids (with coefficient /TDS=0.5) Measuring range 0.0019.99mg/l Kcell=0.1 Measuring range 0.0199.9mg/l Kcell=1 2001999mg/l 2.0019.99g/l 20.099.9g/l Measuring range 100999g/l Kcell=10 Accuracy (conductivity) 0.5%1digit
Measurement of salinity Measurement range 0.0001.999g/l 2.0019.99g/l 20.0199.9g/l Accuracy (salinity) 0.5%1digit Temperature compensation automatic/manual Reference temperature / TDS Conversion factor Cell constant K (cm-1)
0100C with T selectable from 0.00 to 4.00%/C 20C or 25C 0.40.8 0.1, 0.7, 1.0 and 10.0
Standard solutions automatically detected @25C 147S/cm 1413S/cm 12880S/cm 111800S/cm Measurement of temperature Pt100 measuring range -50+200C Pt1000 measuring range -50+200C Resolution 0.1C Accuracy 0.25C Drift after 1 year 0.1C/anno Preset cell constant values: K=0,01 - K=0,1 - K=1, K=10 The resistivity measurement is obtained from the reciprocal of conductivity measurement. Close to the bottom of the scale, the indication of resistivity appears like reported in the table below:
K cell = 0.01 cm-1 Conductivity (S/cm) Resistivity (Mcm) 0.001 S/cm 1000 Mcm 0.002 S/cm 500 Mcm 0.003 S/cm 333 Mcm 0.004 S/cm 250 Mcm K cell = 0.1 cm-1 Conductivity (S/cm) Resistivity (Mcm) 0.01 S/cm 100 Mcm 0.02 S/cm 50 Mcm 0.03 S/cm 33 Mcm 0.04 S/cm 25 Mcm
TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT 2 and 4 electrode conductivity probes
ORDER CODE MEASUREMENT RANGE
16 L=1.5m D=5 12 50
DIMENSIONS
156 20 17
~72 40 1/2 16.2 30 14.5 27 17 56
SP06T
35
120 12
SPT01G
L=1.5m D=5.5
35
130 12
SPT1G
L=1.5m D=5.5
35
130 12
SPT10G
L=1.5m D=5.5
Temperature probes with 4 wire Pt100 and 2 wire Pt1000 connector sensor
Model TP47.100 TP47.1000 TP87.100 TP87.1000 Type Pt100 4 wires Pt1000 2 wires Pt100 4 wires Pt1000 2 wires Working range -50+200C -50+200C -50+200C -50+200C Accuracy Class A Class A Class A Class A
ORP Electrodes KP 90: REDOX PLATINUM liquid filled electrode with S7 screw connector, glass body. KP 91: Gel REDOX PLATINUM electrode, 1m cable with BNC, EPOXY body Characteristics and dimensions of the probes at page 402 pH Buffer solutions HD8642: Buffer solution 4.01pH - 200cc. HD8672: Buffer solution 6.86pH - 200cc. HD8692: Buffer solution 9.18pH - 200cc. Redox Buffer solutions HDR220: Redox buffer solution 220mV 0.5 l. HDR468: Redox buffer solution 468mV 0.5 l. Electrolyte solutions KCL3M Ready to use solution for electrode refilling 100 cc Cleaning and maintenance HD62PT: Diaphragm cleaning (tiourea in HCl) - 500ml. HD62PP: Protein cleaning (pepsin in HCl) - 500ml. HD62RF: Regeneration (fl uorhydric acid) - 100ml. HD62SC: Solution for electrode preservation - 200ml. Conductivity probes See oder codes reported in the table at page 358. Standard conductivity calibration solutions HD8747: Standard calibration solution 0.001mol/l equal to 147S/cm @25C, 200cc. HD8714: Standard calibration solution 0.01mol/l equal to 1413S/cm @25C, 200cc. HD8712: Standard calibration solution 0.1mol/l equal to 12880S/cm @25C, 200cc. HD87111: Standard calibration solution 1mol/l equal to 111800S/cm @25C, 200cc. Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes.
0.005%/C
pH Electrodes KP 20: Gel pH filled combined electrode for general use , with S7 screw connector, EPOXY body. KP 30: Gel pH combined electrode for general use, 1m cable with BNC, EPOXY body . KP 50: Gel pH combined electrode, porous Teflon ring junction, suitable for emulsions, demineralised water, with S7 screw connector, glass body. KP 61: 3 diaphragm liquid filled pH combined electrode for wine, milk, cream, etc., S7 screw connector, liquid reference filling, glass body. KP 62: 1 diaphragm gel pH combined electrode for pure water, varnishes, gel filled, S7 screw connector, glass body. KP 63: 1 liquid filled pH combined electrode for general use, varnishes, 1m cable with BNC, glass body. KP 64: Liquid filled pH combined electrode ,Teflon ring diaphragm, for wine, varnishes, emulsions, S7 screw connector, glass body. KP 70: Pointed gel combined pH microelectrode diam. 6 x L=70 mm., with S7 screw connector, EPOXY body, glass tip, open junction. KP 80: Pointed gel pH combined electrode, with S7 screw connector, glass body, for cream, milk, viscous material, open junction. KP100: Flat membrane gel combined pH electrode with S7 screw connector, glass body, for skin, leather, paper. Characteristics and dimensions of the probes at page 401 CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP: 1.5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 5: 5m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 10: 10m extension cable with BNC/S7 connector for electrode without cable, thread S7. CP 15: 15m extension cable with BNC/S7 connector for electrode without cable, thread S7. CE : S7 screw connector for pH electrode. BNC: female BNC for extension cable
Water Analysis
ORDER CODES HD2156.1: The kit is composed of: instrument HD2156.1, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. HD2156.2: The kit is composed of: instrument HD2156.2 datalogger, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. pH/mV probes, conductivity probes, temperature probes, standard calibration solutions for various types of measurements, connection cables for pH electrodes with S7 connector, cables for data transfer to PC or printer have to be ordered separately. HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C. C.206: Serial connection cable with USB connector for PC and 8-pole MiniDin male connector for the instrument HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin (not suitable for HD2156.1K). DeltaLog9: Software for download and management of the data on PC using Windows 98 to Vista operating systems. SWD10: Stabilized power supply 100-240 Vac/12Vdc-1A mains voltage HD40.1: 24-column portable thermal printer, serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls. RCT: The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter. BAT-40: Spare battery pack for HD40.1 printer with built-in temperature sensor. HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by bench-top meters of the series HD22with cable HD22.2.1 (optional) or power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm.
DO 9786T - R1 DO 9766T - R1
Technical characteristics
Measuring range 2/4 electrodes Input conductivity Transducer energizing Input impedance Cable length Accuracy Pt100 2/4 wires Input temperature Transducer energizing Cable length Accuracy Compensation temperature None manual automatic 4.00...20.00 mA Accuracy Insulation A and B Passive Active External dimensions Protection class External dimensions Protection class Linear 0.00...4.00%/C -50...+200C Linear 0.00...4.00%/C -50...+200C Programmable and proportional to conductivity 0.5% of reading 0.02 mA 2500 Vac 1 minute Bistable, contact 3A/230 Vac free potential 420 mA, 2 wire configuration, 1035 V, see fig. 2 24/230 Vac - 15/+10% 1 VA, 48...62 Hz, see fig. 1 120x122x56 mm IP64 96x96x126 mm IP44 0.0...199.9 mS Configurable cell constant 0.01...199.9 cm-1 Square wave 10...1000 mV, depending on conductivity, 200...1600 Hz, depending on conductivity >100 Mohm <10 metres unscreened <50 metres screened (about 2 nF) 0.5% of reading 2 digits 0.01% per C of drift in temperature -50...199.9C 0.5 mA dc <10 metres unscreened <50 metres screened (about 5 nF) 0.2C 0.1% of reading 0.01C/C of drift in temperature
Key functions PRG Programming of the parameters is activated by pressing the PRG key plus the and keys. The message P1 appears on the display, indicating that the parameter P1 is being programmed. When the PRG key is pressed continuously, the messages P2, P3, P4, P5, P6, P7, P8 and the corresponding parameters are displayed in sequence. After P8 the instrument returns to normal function. SET Key for setting the relay intervention threshold. The ON or OFF symbol appears on the display, indicating the switching on or off threshold of relay A or of relay B. C/F - If this key is pressed it changes the temperature measuring unit to degrees Celsius or degrees Fahrenheit. - When pressed together with the CAL key it activates the manual temperature setting function. - If pressed during the conductivity calibration function it quits the calibration function without storing the calibration. When pressed together with the CAL key it activates the conductivity calibration function. OK Confirms the programming parameters, or the relay SET values, and stores them. CAL - When pressed together with the C/F key it activates the manual temperature setting function. - When pressed together with the key it activates the conductivity calibration function. - Key used to confirm conductivity calibration and manual temperature calibration. - Key for increasing the value displayed in the parameter programming phase. - During the relay SET point programming phase. - During the calibration phase. - Key for decreasing the value displayed in the parameter programming phase. - During the relay SET point programming phase. - During the calibration phase. Setting the relay SET point - Press the SET button; the ON symbol appears on the display with the letter A to indicate that the value shown corresponds to the switching on threshold of relay A. - To change this value press the and keys. - Press SET; the OFF symbol appears with the letter A to indicate that the switching off threshold of relay A is being displayed. - To change this value press the and keys. - Press the SET button; the ON symbol appears on the display with the letter B to indicate that the value shown corresponds to the switching on threshold of relay B. - To change this value press the and keys. - Press SET; the OFF symbol appears with the letter B to indicate that the switching off threshold of relay B is being displayed. - To change this value press the and keys. - Press SET, the instrument stores the values and returns to normal function. NOTE: During the SET point setting phase (symbols ON or OFF lit) the instrument returns to normal function if no key is pressed for 2 minutes.
Power supply
Relay output
Current output
Manual temperature setting If the temperature probe is not connected or if the probe is broken the measuring unit C or F flashes. In this case it is possible to set the temperature compensation value manually. - Press the CAL key and the C/F key together; the message CAL appears at the bottom of the display. - Using the and keys, set the temperature value corresponding to the temperature of the liquid in which you wish to measure conductivity. - Press CAL to confirm this value. The message CAL disappears. Calibration of the DO 9786T-R1 / DO 9766T-R1 with conductivity probes Calibration of the DO 9786T-R1 / DO 9766T-R1 transmitters with conductivity probes: - Immerse the probe in the buffer solution used for calibration. - Press the CAL key and the key together; the message CAL appears at the top of the display. - The instrument can automatically recognize two standard calibration solutions: a 0.1 molar solution of KCl and a 0.01 molar solution of KCl. The instrument proposes the conductivity value as a function of the measured temperature if the temperature probe is connected, or the manually set temperature. - Using the and keys, adjust the conductivity value measured as a function of the liquid temperature. - Press CAL to confirm this value. The message CAL disappears. NOTE: If you want to quit without storing the new calibration, press the C/F key. N.B.: Before calibrating the probe set a cell constant close to the cell constant of the probe that you wish to calibrate with key PRG, function P2. If the message E1 appears during calibration, the instrument is indicating that the probe gain is too high; quit programming (C/F button) and increase the value of the cell constant. Likewise, if E2 appears, the instrument is indicating that the probe gain is too low; quit calibration and decrease the cell constant. Repeat the calibration operation. Programming the parameters P1 Temperature coefficient. May be set between 0 and 4.0%/C (0 and 2.2%/C). P2 Cell constant. May be set between 0.01 and 199.9. P3 Conductivity value corresponding to 4 mA at output. May be set between 0 and 199.9 mS. P4 Conductivity value corresponding to 20 mA at output. May be set between 0 and 199.9 mS. P5 Delay time in the intervention of relay A. May be set between 0 and 250 seconds. P6 Delay time in the intervention of relay B. May be set between 0 and 250 seconds. P7 Reference temperature of the conductivity measurement. May be set between the values 20.0 or 25.0C. P8 Calibration of Pt100 probe and calibration of analog output in current (see Pt100 probe calibration and analog output calibration). To change one of these parameters (except P8) press key PRG until the message corresponding to the parameter to be changed appears on the screen. Using the and keys, bring the parameter displayed to the desired value. Press OK to confirm. Pt100 probe calibration - Connect the Pt100 probe to the instrument. Press the PRG key until the message P8 appears on the display. - Press the CAL key; the message CAL appears at the bottom of the display and the temperature is shown at the top. - Immerse the Pt100 probe and a precision thermometer for reference in the zero calibration bath. Wait long enough for the reading to become stable. - Using the and keys, adjust the value of the temperature measured by the Pt100 probe so that it corresponds with the value on the reference thermometer. - Immerse the Pt100 probe and a precision thermometer in the full scale calibration bath. Wait long enough for the reading to become stable. - Using the and keys, adjust the value of the temperature measured by the Pt100 probe so that it corresponds with the value on the reference thermometer. - Press OK to confirm. N.B.: If the temperature shown by the instrument is between +12C, the instrument calibrates the probe offset, otherwise it calibrates the gain. Calibrating the analog output - Press the PRG key until the message P8 appears on the display. - Connect a precision milliammeter to the analog output. - Press the CAL key twice; the message CAL appears at the top of the display and the message 4.0 at the bottom, indicating calibration at 4 mA. - Using the and keys, adjust the value of the output current so as to have an indication of 4,000 mA on the precision milliammeter. - Press the CAL key; the message CAL appears at the top of the display and the message 20.0 at the bottom, indicating calibration at 20 mA. - Using the and keys, adjust the value of the output current so as to have an indication of 20,000 mA on the precision milliammeter. - Press OK to confirm. Display Symbol C F S mS A B Description indicates that the value shown is in C. indicates that the value shown is in F. indicates that the unit of the value shown is micro Siemens. indicates that the unit of the value shown is milli Siemens. indicates that the relay A is in closed status. indicates that the relay B is in closed status.
20 19 18 17 S mS / cm 16
=
mA B
11
GND 10 4...20mA 9
8 7 6 5 4 3
Pt100 4 wires
15 14 13 12
temp
conductivity 2 electrodes
1 20 19 18 17 S mS / cm 16
11 10 4...20mA V+ 9
8
Pt100 2 wires
15 14 13 12
B temp A
7 6 4 5 3
24 Vac
0 Vac
Min 9Vdc V
1 24 Vac 2
11 10
Regulator
9 4...20 mA
Regulator
120.0 mm
PG7
DO 9766T
96.0 mm 126.0 mm
96.0 mm
15.0 mm
DO 9786T
21.0 mm
Water Analysis
ON OFF
indicates that the value shown corresponds to the closing thresholds of the contacts of relay A or B. indicates that the value shown corresponds to the opening thresholds of the contacts of relay A or B.
Error signals OFL - Warning which appears during measurement when the value to be displayed is out of scale. E1 - Error warning which appears during conductivity calibration to indicate that the probe gain is too high. Press P2 to increase the cell constant value. E2 - Error warning which appears during conductivity calibration to indicate that the probe gain is too low. Press P2 to decrease the cell constant value. E3 - Error warning which appears to indicate that the instrument is unable to recognize the buffer solution used for automatic calibration. Press the or key to make this indication disappear. E4 - Reading error on the EEPROM. APPENDIX Table of compatibility between range and sensor
Conductivity Range 019.99 S 0199.9 S 0 1999 S 0199.9 S 019.99 mS 0199.9 mS 0 1999 mS 0.010.2 Nominal cell constant 0.22 220 20199.9
the instrument. This ensures greater accuracy of calibration. Disable the temperature compensation during the calibration of the instrument with the precision resistences. Calibration with standard solutions In this case too, for the calibration of the instrument, cable and measuring probes in a standard solution, the greatest attention must be paid to the temperature of the solutions and the cleanness of the measuring cell. It is advised not to carry out calibration below 500 S/cm. Solutions with low conductivity must be kept closed in their containers. Contact with the air increases their value due to the absorption of CO2. The regulations for the preparation of standard solutions with a base of KCl dissolved in water with a high degree of purity supply the method and percentages of KCl and water to be mixed. DELTA OHM supplies four solutions for calibration: HD8747: Standard calibration solution 0.001mol/l equal to 147S/cm @25C, 200cc. HD8714: Standard calibration solution 0.01mol/l equal to 1413S/cm @25C, 200cc. HD8712: Standard calibration solution 0.1mol/l equal to 12880S/cm @25C, 200cc. HD87111: Standard calibration solution 1mol/l equal to 111800S/cm @25C, 200cc. Care and maintenance of the conductivity cell In conductivity measurement systems in industrial plants, if the installation is correctly made, readings are generally reliable for a long time. The important thing is to carry out correct, programmed maintenance of the measuring cell. Abrasion of the cable due to continued swinging movements must be avoided, as must the formation of deposits and scale on the cell which can change its geometrical structure. The cell must always be immersed in the liquid that is to be measured. In the industrial field, measurements may range from highly pure water to sewage or water contaminated by corrosive substances. It is good practice to check the compatibility of the materials of which the cell and the connecting cable are made with the liquid in which the measurement is to be taken. Check that there are no floating bodies, suspended granules that may be more or less conductive, or which could get stuck inside the cell, thus leading to incorrect measurements. To clean the cell use detergents or substances suitable for the material of which the cell is made. Selecting the cell constant and installation The measurement range of the liquid to be examined determines the choice of the cell constant to be used. Installation of the cell will vary according to the application. On the whole, the following points must be considered: - Choose the correct cell and cell constant, suitable for the application. - Use suitable materials, cable, cell, supports, so as to resist corrosion and the influence of atmospheric agents. - The sensor/cell must be firmly fixed, in a place where they are easily accessible for maintenance. - The liquid in which the sensor is immersed must be a representative part of the whole that is to be measured. - There must be a moderate flow of liquid so that an updated sample of liquid arrives at the electrodes. Excessive movement or flow causes turbulence and air bubbles between the electrodes. As an air bubble is not conductive, it modifies the volume of the cell and changes the constant. - Install the sensor in such a way that sludge or particles of material cannot be deposited inside it. - If installed in containers where high currents are circulating, the conductivity cell may present measuring problems. - The maintenance and cleaning interval depends on the quality of the liquid in which the cell is installed. Order code DO 9786T: Conductivity transmitter 420 mA passive or active, power supply 24 Vac with double display 96x96 mm, for panel mounting. DO 9766T: Conductivity transmitter 420 mA passive or active, power supply 24 Vac with double display 122x120 mm, for use on the field. SPT 86: Combined industrial conductivity and temperature probe in POCAN with 4 platinum electrodes, cell constant K = 0.7, 1.5 meters cable, Pt100 with 2 wires. Temperature 090C. SPTKI 10: Combined industrial conductivity probe in Glass with 2 black oxidized platinum electrodes, cell constant K = 1, S7/PG13 screw-joint, 2 wires output: eurostandard S7. Temperature 0100C. SPTKI 11: Combined industrial conductivity and temperature probe in Rytron with 2 graphite electrodes, cell constant K = 1, 5 meters cable, Pt100 with four wires. Temperature 050C. SPTKI 12: Combined industrial conductivity and temperature probe in Rytron with 2 platinum electrodes, cell constant K = 01, 5 meters cable, Pt100 with four wires. Temperature 050C. SPTKI 13: Combined industrial conductivity and temperature probe in Rytron with 2 platinum electrodes, cell constant K = 10, 5 meters cable, Pt100 with four wires. Temperature 050C. HD 882 M100/300: Temperature probe with Pt100 sensor, miniature head, shaft 6x300 mm. HD 882 M100/600: Temperature probe with Pt100 sensor, miniature head, shaft 6x600 mm. HD 8747: Calibration solution 0.001 mol/l corresponding to 147 S/cm at 25C, 200cc. HD 8712: Calibration solution 0.1 mol/l corresponding to 12,880 S/cm at 25C, 200cc. HD 8714: Calibration solution 0.01 mol/l corresponding to 1413 S/cm at 25C, 200cc. HD 87111: Calibration solution 1 mol/l corresponding to 111800 S/cm at 25c, 200cc.
Temperature sensor
Temperature -50C -25C 0C 25C 50C 75C Pt100 80.25 90.15 100.00 109.73 119.40 128.98 Temperature 100C 125C 150C 175C 199C
Calculating the temperature coefficient of a solution If the temperature coefficient of the solution is not known, it may be determined using the DO 9786T/DO 9766T. - Set the temperature coefficient at 0.0%/C (parameter P1). The following measurements should be taken as close as possible to the work point, between 5C and 70C, for greater accuracy. - Immerse the probe in the testing liquid. Wait for the measurement to become stable. - Take note of the temperature and of the conductivity. - Increase the solution temperature by at least 10C. - Take note of the temperature and of the conductivity. - Calculate the temperature coefficient using the following formula: = Where:
N.B.: if the reference temperature is 25C, replace 20 with 25. - Set the temperature coefficient with the value calculated as above (parameter P1). Calibration of the instrument for measuring conductivity The conductivity measurement depends strongly on the temperature of the liquid that is to be measured; this relationship must be considered during calibration. Calibration of the instrument alone by means of a precision resistance This is a sure and accurate method for calibrating the instrument alone, but it does not allow for the variations of the cell constant that may occur, nor of the state of efficiency and cleanness of the cell. The precision resistance used for calibration will be chosen according to the scale that you want to calibrate. Typical values are the following:
Conductivity 100,0 S 500,0 S 1000 S 5000 S 10,00 mS 50,00 mS 100,0 mS 500,0 mS 1000 mS Resistance 10.000 2.000 1.000 200 100 20 10 2 1
The precision resistance will be connected to the end of the cable that connects the probe to
~72
40
1/2
16.2
30
14.5
27
17
56
K 0.7
17
L=1.5 m
39
120
4.6
S7
PG13
M 20
12
21
85
20
41
K1
Pt100
66
Brown Green Yellow Gray Pink White
L=5m
Screen
21
85
20
70
3/4 G
K 0.1
Pt100
L=5m
Shield
21
85
20
107
12
K 10
Pt100
29
3/4 G
L=5m
Screen
Temperature sensor Pt100 2 wire Pt100 4 wire Pt100 4 wire Pt100 4 wire
12
Water Analysis
Pt100
12 3/4 G
34
K1
HD 2109.1 HD 2109.2
INSTRUMENT TECHNICAL CHARACTERISTICS Measured quantities: mg/l O2, sat.% O2, mbar, C, F Instrument Dimensions (Length x Width x Height) 185x90x40mm Weight 470g (complete with batteries) Materials ABS, rubber Display 2x4 digits plus symbols Visible area: 52x42mm Operating conditions Working temperature Storage temperature Working relative humidity Protection degree Power Batteries Autonomy Power absorbed with instrument off With dissolved oxygen probe Mains (SWD10) Security of memorized data Time Date and time Accuracy -550C -2565C 090%RH without condensation IP67 4 1.5V type AA batteries 200 hours with 1800mAh alkaline batteries 20A 40A Output mains adapter 12Vdc / 1A Unlimited, independent of battery charge conditions Schedule in real time 1min/month max error
Measured values storage - model HD2109.2 Type 2000 pages containing 9 samples each Quantity 18,000 samples composed of 4 parameters: mg/l O2 - %O2 - mbar - (C or F) Selectable storage interval 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour) Serial interface RS232C Type Baud rate Data bit Parity Stop bit Flow Control Serial cable length selectable print interval immediate RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1 Xon/Xoff Max 15m 1s, 5s, 10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s (30min) and 3600s (1hour) 1.1 - 2.0 electrically isolated
USB interface - model HD2109.2 Type Connections Input for Oxygen and temperature probes Serial and USB interface Mains adapter
8-pole male DIN45326 connector 8-pole MiniDin connector 2-pole connector (positive at centre)
Measurement of the concentration of dissolved Oxygen Measurement range 0.0090.00mg/l Resolution 0.01mg/l Accuracy (60110%, 1013mbar, 2025C) 0.03mg/l1digit
HD40.1
SWD10
Measurement of the saturation index of dissolved Oxygen Measurement range 0.0600.0% Resolution 0.1% Accuracy 0.3% 1digit (in the range 0,0199,9%) 1% 1digit (in the range 200,0600,0%) Measurement of barometric pressure Measurement range Resolution Accuracy Setting the salinity Setting range Resolution 0.01100.0mbar 0.1mbar 2mbar1digit between 18 and 25C (2mbar+0.1mbar/C) in the remaining range 0.070.0mg/l 0.1mg/l
Temperature measurement with the sensor inside the dissolved Oxygen probe Measurement range 0+45C Resolution 0.1C Accuracy 0.1C Drift after 1 year 0.1C/year Temperature measurement by Instrument with Pt100 probe Pt100 measurement range -200+650C Resolution 0.1C Accuracy 0.21C Drift after 1 year 0.1C/year Automatic temperatur compensation 050C TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENT Temperature probes Pt100 sensor using SICRAM module
Model TP87 TP472I TP472I.0 TP473P.0 TP474C.0 TP472I.5 TP472I.10 Type Immersion Immersion Immersion Penetration Contact Immersion Immersion Working range -50C+200C -196C+500C -50C+400C -50C+400C -50C+400C -50C+400C -50C+400C Accuracy 0.25C (-50C+200C) 0.25C (-196C+350C) 0.4C (+350C+500C) 0.25C (-50C+350C) 0.4C (+350C+400C) 0.25C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C) 0.3C (-50C+350C) 0.4C (+350C+400C)
ORDER CODES HD2109.1: The kit is composed of: instrument HD2109.1, calibrator DO9709/20, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes and data transfer cable must be ordered separately. HD2109.2: The kit is composed of: instrument HD2109.2 datalogger, calibrator DO9709/20, connection cable for serial output HD2101/USB, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes and data transfer cable must be ordered separately. HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C. C.206: Serial connection cable with USB connector for PC and 8-pole MiniDin male connector for the instrument HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin (not suitable for HD2109.1K). DeltaLog9: Software for download and management of the data on PC using Windows 98 to Vista operating systems. SWD10: Stabilized power supply 100-240 Vac/12Vdc-1A mains voltage HD40.1: 24-column portable thermal printer, serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls. RCT: The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter. BAT-40: Spare battery pack for HD40.1 printer with built-in temperature sensor. HD22.2: Laboratory electrode holder composed of base plate with built-in magnetic stirrer, shaft and replaceable electrode holder. Suitable diameter 12mm. Powered by bench-top meters of the series HD22with cable HD22.2.1 (optional) or power supplier SWD10 (optional). HD22.3: Laboratory electrode holder composed of base plate. Flexible arm for free positioning. Suitable for electrodes with diameter 12mm. Solutions DO9700: zero solution. DO9701: electrolyte solution. Combined dissolved Oxygen/temperature probes DO9709 SS: The kit includes: combined probe for measurement of O2 and temperature with replaceable membrane, three membranes, 50ml of zero solution, 50ml of electrolyte solution. Cable length 2m. 12mm x 120mm. DO9709 SS.5: The kit includes: combined probe for measurement of O2 and temperature, replaceable membrane, 12mm x 12mm. Cable length 5m, three membranes, 50ml of zero solution, 50ml of electrolyte solution. Temperature probes complete with SICRAM module TP87: PT100 sensor immersion probe. Stem 3 mm, length 70 mm. Cable length 1 m. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 m. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 m. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 m. TP475A.0: Air probe, sensor Pt100. Stem 4mm, length 230mm. Cable length 2 m. TP472I.5: Immersion probe, sensor Pt100. Stem 6mm, length 500 mm. Cable length 2 m. TP472I.10: Immersion probe, sensor Pt100. Stem 6mm, length 1,000mm. Cable length 2 m. Temperature probes without SICRAM module TP47.100: Direct 4 wires Pt100 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 4 wires with connector, length 2 m. TP47.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 230mm. Connection cable 2 wires with connector, length 2 m. TP87.100: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. 4 wire connection cable with connector, length 1 m. TP87.1000: Pt1000 sensor immersion probe. Probes stem 3mm, length 70mm. 2-wire connection cable with connector, length 1 m. TP47: Module for the connection of Pt100 4-wire and Pt1000 2-wire probes
0.003%/C
Accuracy Class A
0.003%/C
DO9709SS DO9709SS.5 Polarographic probe, Silver anode, Platinum cathode 0.0060.00mg/l 045C 1% FS Replaceable 2m DO9709SS
120
12
5m
L=2m
DO9709SS.5
120
12
L=5m
DO9700
DO9701
Water Analysis
HD 98569
The serial connection RS232C can be used for direct printing of labels with a 24 column printer (HD40.1 or HD40.2). The software DeltaLog11 (vers. 2.0 and subsequent ones) allows instrument management and configuration, and data processing on PC. Technical characteristics of HD 98569 Measured values pH - mV - - TDS - NaCl mg/l O2 - %O2 C - F Instrument Dimensions (LengthxWidthxHeight) Weight Materials Display Operating conditions Working temperature Storage temperature Working relative humidity Protection degree Power Batteries Autonomy (with probes connected) Mains (cod. SWD10) Security of memorized data Time Date and hour Accuracy Continuous storage (LOG key) Quantity Type Storage interval Storage on command (MEM key) Quantity Type
250x100x50mm 640g (complete with batteries) ABS, rubber Graphic, back lighted LCD, 56x38mm. 128x64 points -5 50C -25 65C 0 90% RH without condensate IP66 4 batteries 1.5V type AA 25 hours with 1800mAh alkaline batteries 12Vdc/1A (positive at centre) Unlimited Schedule in real time 1min/month max. departure 9000 samples of the three inputs organised in 1800 pages containing 5 samples each 1s 999s 200 samples of the three inputs organised in 200 pages containing 1 sample each
Only conductivity probes with SICRAM module. Input for O2 and temperature probes or for only temperature probes with SICRAM module. Input for pH, mV, pH and temperature probes or for only temperature probes with SICRAM module. External Power supply. RS232 or USB interface.
Calibration storage pH and Dissolved Oxygen Conductivity RS232C serial interface Type Baud rate Data bit Parity Stop bit Flow control Length of serial cable USB interface Typ Bluetooth interface
Last 8 pH and dissolved oxygen calibrations. The last 2 are saved in the SICRAM memory of the probe as well. Last calibration is saved in the SICRAM memory of the probe. RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1 Xon/Xoff Max 15m 1.1 - 2.0 electrically isolated optional for PCs fitted with Bluetooth input or HDUSB. KL1 Bluetooth / RS232 adapter. The interface can be installed in Delta Ohm only.
Measurement of resistivity by instrument Measurement range (K cell=0.01) Up to 1Gcm Measurement range (K cell=0.1) Up to 100Mcm Measurement range (K cell=1) 5.0199.9cm 200999cm 1.00k19.99kcm 20.0k99.9kcm 100k999kcm 110Mcm Measurement range (K cell=10) 0.55.0cm Accuracy (resistivity) instrument 0.5% 1digit
(*) The resistivity measurement is obtained from the reciprocal of conductivity measurement. Close to the bottom of the scale, the indication of resistivity appears like reported in the table below:
K cell = 0.01 cm-1 Conductivity Resistivity (S/cm) (Mcm) 0.001 S/cm 1000 Mcm 0.002 S/cm 500 Mcm 0.003 S/cm 333 Mcm 0.004 S/cm 250 Mcm K cell = 0.1 cm-1 Conductivity Resistivity (S/cm) (Mcm) 0.01 S/cm 100 Mcm 0.02 S/cm 50 Mcm 0.03 S/cm 33 Mcm 0.04 S/cm 25 Mcm
Connections Enabled inputs for temperature probes with SICRAM module Input for pH/temperature with SICRAM module Input for conductivity/temperature with SICRAM module
pH/mV and O2 inputs. 8-pole male DIN45326 connector 8-pole male DIN45326 connector
Input for dissolved oxygen/temperature with SICRAM module 8-pole male DIN45326 connector RS232C / USB interface Bluetooth Mains adapter Measurement of pH by instrument Measuring range Resolution Accuracy Input impedance Calibration error @25C Calibration points Temperature compensation Automatically detected standard solutions @25C Measurement of mV by instrument Measuring range Resolution Accuracy Drift after 1 year 8-pole MiniDin female connector Optional 2-pole(5.5mm- 2.1mm). Positive at centre (e.g. SWD10). -9.999+19.999pH 0.01 o 0.001pH selectable from menu 0.001pH 1digit >1012 |Offset| > 20mV Slope > 63mV/pH or Slope < 50mV/pH Sensitivity > 106.5% or Sensitivity < 85% Up to 5 points from a list of 8 automatically detected buffers -50...150C 1.679pH - 4.000pH - 4.010pH 6.860pH - 7.000pH - 7.648pH 9.180pH - 10.010pH -1999.9+1999.9mV 0.1mV 0.1mV 1digit 0.5mV/year Resolution 0.001S/cm 0.01S/cm 0.1S/cm 1S/cm 0.01mS/cm 0.1mS/cm 1mS/cm
Measurement of salinity Measurement range 0.0001.999g/l 2.0019.99g/l 20.0199.9 g/l Accuracy (salinity) instrument 0.5% 1digit
Automatic/manual temperature compensation 0...100C with T = 0.004.00%/C Reference temperature Conversion factor /TDS Admitted cell constants K (cm-1) 050C (Default values 20C or 25C) 0.40.8 0.0120.00
Measurement of conductivity by instrument Measurement range (K cell=0.01) 0.0001.999S/cm Measurement range (K cell=0.1) 0.0019.99S/cm Measurement range (K cell=1) 0.0199.9S/cm 2001999S/cm 2.0019.99mS/cm 20.0199.9mS/cm Measurement range (K cell=10) 2001999mS/cm Accuracy (conductivity) instrument 0.5% 1digit
Water Analysis
Measurement of total dissolved solids Resolution (with coefficient /TDS=0.5) Measurement range (K cell=0.01) 0.001.999mg/l Measurement range (K cell=0.1) 0.0019.99mg/l Measurement range (K cell=1) 0.0199.9 mg/l 2001999 mg/l 2.0019.99 g/l 20.0199.9 g/l Measurement range (K cell=10) 100999 g/l Accuracy (total dissolved solids) instrument 0.5% 1digit
0.005mg/l 0.05mg/l 0.5 mg/l 1 mg/l 0.01 g/l 0.1 g/l 1 g/l
Measurement of concentration of dissolved oxygen Measurement range 0.0090.00mg/l Resolution 0.01mg/l Accuracy instrument 0.03mg/l 1digit(60...110%,1013mbar, 20...25C) Measurement of saturation index of dissolved oxygen Measurement range 0.0600.0% Resolution 0.1% Accuracy instrument 0.3% 1digit (in the range 0.0199.9%) 1% 1digit (in the range 200.0600.0%) Salinity setting Setting Setting range Resolution directly from menu or automatically by conductivity measurement 0.070.0g/l 0.1g/l
Temperature measurement with the sensor inside the O2 probe Measurement range 0.050.0C Resolution 0.1C Accuracy instrument 0.1C Drift after 1 year 0.1C/year Automatic temperature compensation 050C Measurement of temperature by instrument Pt100 Measurement range -50+150C Resolution 0.1C Accuracy instrument 0.1C 1digit Drift after 1 year 0.1C/year 24 column printing example
HD 98569 pH / chi / Oxy / temperature Ser num=12345678 2007 - 01 - 31 12:00:00 LAB POSITION #1 Operator = Amministratore SAMPLE ID = 00000001 pH EL sernum = 01234567 pH = 7.010 pH out of calibration ! O2 EL sernum = 76543210 mg/l O2 = 5.59 chi EL sernum = 98756410 mS = 2.177 Temp = 25.0C ATC
HD40.1: 24-column portable thermal printer, serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls. HD40.2: 24-column portable thermal printer, Bluetooth and serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls. Requires the module HD22BT (optional) or the cable HD 2110 CSNM (optional). RCT: The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter. BAT-40: Spare battery pack for HD40.1 printer with built-in temperature sensor. HD22.2: Laboratory electrode holder composed of basis plate with incorporated magnetic stirrer, staff and replaceable electrode holder. Height max. 380mm. For 12mm electrodes. Powered by bench top meters of series HD22 with cable HD22.2.1 (optional) or supplier SWD10 (optional) HD22.3: Laboratory electrode holder with metal basis plate. Flexible electrode holder for free positioning. For 12mm probes. HD22BT: Bluetooth module for wireless data transmission from instrument PC. The fitting of the module into the instrument is made exclusively by Delta Ohm, at the time of placing the order. HDUSB.KL1: USB/Bluetooth converter to be connected to the PC for wireless data transmission from the instrument with HD22BT module. SICRAM Modules with S7 input for pH electrodes KP471.1: SICRAM module for pH electrodes with S7 standard connection, cable L=1m. KP471.2: SICRAM module for pH electrodes with S7 standard connection, cable L=2m. KP471.5: SICRAM module for pH electrodes with S7 standard connection, cable L=5m.
pH Electrodes to be connected to KP471 SICRAM module KP20: Combined pH electrode for general use, GEL-filled, with screw connector S7, body in Epoxy,
120
S7
16
12
KP 50: Combined pH electrode pH for general use, varnishes, emulsions, gel-filled, with S7 screw connector, body in glass.
120
S7
16
12
KP 61: Combined pH electrode, 3 diaphragms for milk, cream, etc. gel-filled, with screw connector S7, body in glass.
120
Ordering codes HD 98569: The kit is composed of: instrument data logger HD 98569 for measurement of pH - redox - conductivity - resistivity - TDS - salinity - concentration of dissolved oxygen- saturation index - temperature, 4 1.5V batteries type AA, instructions manual, software DeltaLog11 (vers. 2.0 and subsequent ones), carrying case and SICRAM module pH471.1 (cable 1 meter). The pH/mV electrodes, conductivity probes, dissolved oxygen probes, temperature probes, standard reference solutions for different measurement types, connection cables for data download to PC or printer have to be ordered separately. HD2110CSNM: 8-pole connection cable Mini Din - Sub D 9-pole female for RS232C, for connection to PC with RS232C USB input. HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole Mini Din for connection to PC with USB input. DeltaLog11: Further unit of software (vers. 2.0 and subsequent ones) for data download and management on PC using Windows 98 to Vista operating systems. SWD10: Stabilized power supply at 100-240Vac/12Vdc-1A mains voltage.
S7
16
12
KP 62: Combined pH electrode, 1 diaphragm for pure water, paints, etc. GEL-filled, with screw connector S7, body in glass
130
S7
16
12
KP 64: Combined pH electrode for water, varnishes, emulsions, etc. reference filling solution KCl 3M, with S7 screw connector, body in glass.
120
Electrolyte solutions KCL 3M: 50cc ready for use solution for refilling of electrodes. Cleaning and maintenance HD62PT: Diaphragm cleaning (tiourea in HCl) - 200cc. HD62PP: Protein cleaning (pepsin in HCl) - 200cc. HD62RF: Regeneration (fluorhydric acid) - 100cc. HD62SC: Solution for electrode preservation - 200cc. Combined conductivity and temperature probes with SICRAM module SP06TS: Combined conductivity and temperature 4-electrode cell, body in Pocan. Cell constant K=0.7. Measurement range 5S/cm 200mS/cm, 090C.
156
S7
16
12
KP 70: Combined pH electrode, micro diam. 6 x L=70mm, GEL-filled, for paste, bread, cheese, etc, with S7 connector, body in glass.
90 50
16
50
20
S7
L=1.5m
16 15 6.5
D=5.5 12 17
KP 80: Combined pointed pH electrode, gel-filled, with screw connector S7, body in glass.
120
SPT401.001S: Combined conductivity and temperature 2-electrode cell in stainless steel AISI 316. Cell constant K=0.01. Cable 2m. Measurement range 0.04S/cm 20S/cm, 0120C. Measurement in closed-ell.
16 12
~72 40
S7
1/2
16.2
ORP Electrodes to be connected to KP471 SICRAM module KP90: REDOX PLATINUM electrode, with screw connector S7, reference filling solution KCl 3M, body in glass.
120
30
SPT01GS: Combined conductivity and temperature 2-electrode Platinum-wire cell, body in glass. Cell constant K=0.1. Measurement range 0.1S/cm 500S/cm, 080C.
S7 16 12
35 120
pH Electrodes with SICRAM module KP 50TS: Combined pH/temperature electrode, Pt100 sensor, gel-filled, with SICRAM module, body in glass, general use, varnishes, emulsions. Cable length 1m.
16
D=5.5
16
120
SPT1GS: Combined conductivity and temperature 2-electrode Platinum-wire cell, body in glass. Cell constant K=1. Measurement range 10S/cm 10mS/cm, 080C.
12
L=1.5m
35 130
KP63TS: Combined pH/temperature electrode, Pt100 sensor, gel-filled, with SICRAM module, body in glass, Ag/AgCl sat KCl.
16
D=5.5
16
120
12
SPT10GS: Combined conductivity and temperature 2-electrode Platinum-wire cell, body in glass. Cell constant K=10. Measurement range 500S/cm 200mS/cm, 080C.
35 130
pH buffer solutions HD8642: Buffer solution 4.01pH - 200cc. HD8672: Buffer solution 6.86pH - 200cc. HD8692: Buffer solution 9.18pH - 200cc. Redox buffer solutions HDR220: Redox buffer solution 220mV 500cc. HDR468: Redox buffer solution 468mV 500cc.
D=5.5
16
12
L=1.5m
12
12
L=1.5m
Water Analysis
14.5
27
17
56
Standard calibration solutions HD8747: Standard calibration solution 0.001mol/l equal to 147S/cm @25C - 200cc. HD8714: Standard calibration solution 0.01mol/l equal to 1413S/cm @25C - 200cc. HD8712: Standard calibration solution 0.1mol/l equal to 12880S/cm @25C - 200cc. HD87111: Standard calibration solution 1mol/l equal to 111800S/cm @25C - 200cc. Combined dissolved oxygen/temperature probes DO9709 SS: The kit includes: combined probe for the measurement of O2 and temperature with replaceable membrane, three membranes totally. 50ml of zero solution, 50ml of electrolyte solution. Cable length 2m. 12mm x 120mm.
16
Accessori es for combined dissolved oxygen/temperature probes DO9709 SSK: Accessory kit for the DO9709 SS probe consisting of three membranes, 50ml of zero solution, 50ml of electrolyte solution. DO9709.20: Calibrator for polarographic probes DO9709SS and DO9709SS.5. Temperature probes with SICRAM module TP87: Pt100 sensor immersion probe. Probes stem 3mm, length 70mm. Cable length 1 metre. TP472I.0: Pt100 sensor immersion probe. Stem 3 mm, length 230 mm. Cable length 2 metres. TP473P.0: Pt100 sensor penetration probe. Stem 4mm, length 150 mm. Cable length 2 metres. TP474C.0: Pt100 sensor contact probe. Stem 4mm, length 230mm, contact surface 5mm. Cable length 2 metres. TP475A.0: Pt100 sensor air probe. Stem 4mm, length 230mm. Cable length 2 metres. TP472I.5: Pt100 sensor immersion probe. Stem 6mm, length 500 mm. Cable length 2 metres. TP472I.10: Pt100 sensor immersion probe. Stem 6mm, length 1,000mm. Cable length 2 metres.
120
12
L=2m
DO9709 SS.5: The kit includes: combined probe for the measurement of O2 and temperature with replaceable membrane, three membranes totally. 50ml of zero solution, 50ml of electrolyte solution. Cable length 5m. 12mm x 120mm.
120
12
L=5m
HD 3405.2
Technical characteristics HD3405.2 pH - mV - C/F measurement Instrument Dimensions (Length x Width x Height) Weight Materials Display Operating conditions Working temperature Storage temperature Working relative humidity Protection degree Power Batteries Autonomy (only batteries) Mains (cod. SWD10) Security of memorized data Selectable storage interval Time Date and hour Accuracy Serial interface RS232C Type Baud rate Data bit Parity Stop bit Flow Control Serial cable length Selectable print interval USB Interface Type Connections Serial interface and USB Mains adapter (cod. SWD10) Storage of measured values Type Quantity
220x120x55mm 460g (complete with batteries) ABS, rubber 2x4 characters plus symbols visible area: 52x42mm -5 50C -25 65C 0 90% RH without condensation IP66 3 batteries 1.5V type AA 100 hours with 1800mAh alkaline batteries Output mains adapter 100-240Vac/ 12Vdc-1A Unlimited 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1hour Schedule in real time 1min/month max departure RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1 Xon/Xoff Max 15m immediate or 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1hour 1.1 - 2.0 electrically isolated 8-pole MiniDin connector 2-pole connector (positive at centre) 12Vdc/1A 2000 pages of 17 samples each 34,000 sets of measures made up of [pH or mV] or and [C or F].
Measurement connections Temperature probe input with SICRAM module or TP47 module pH/mV input
Vdc in
Female BNC
GRD
Water Analysis